3D4D ARIETTA V70 and V60

Set Up of the system and new features

29th – 30th August 2017

Luca Mastrogirolamo

Luca Mastrogirolamo © Hitachi Medical Systems Europe 2017 All rights reserved.
Contents

1.  Concept of Presets


2.  Define the quality of Hi Def 3D and 4D
3.  Renderings and how to use them
4.  Main Parameters to adjust in 3D4D
5.  How to set up the system in different clinical applications

Luca Mastrogirolamo © Hitachi Medical Systems Europe 2017 All rights reserved.
Concept of the Preset

To set up 3D4D on ARIETTA V70 & V60 we have two options:

Embedded
Dedicated on
3D Preset Anatomy 3D Setting in the
With 4 QSS Preset normal anatomy QSS
- 2D OR
- 3D (Anatomy) Dedicated 3D QSS.
- Skeleton Attention! 2D settings
- Corpus Callosum is effecting 3D
performances

Luca Mastrogirolamo © Hitachi Medical Systems Europe 2017 All rights reserved. 3
Recommended Preset for Surface rendering

Set Up of Basic 2D Image (2nd Trimester)


•  Slightly higher Contrast
•  THI Mode -> FmT (HdTHI if easy patient)
•  Freq / Receiving Filter -> Low / Reso (to be changed if needed)
•  HiRez ->5
•  Persist -> 0
•  Line Density ->3
•  Echo Enhancement ->2
•  Color Map -> E
•  Post Process -> S Curve Gamma 1.3 Reject 5
•  Compound ->On Type 1

Luca Mastrogirolamo © Hitachi Medical Systems Europe 2017 All rights reserved. 4
2D Setting

Recommended presets Vs Modified preset

HiRez 1 – HdTHI – Graymap 10


Luca Mastrogirolamo © Hitachi Medical Systems Europe 2017 All rights reserved. 5
3D ROI and Curve Shapes

How to set the ROI for 3D4D + Curve Shapes

Pre-Formed Curves or
defined by user

Luca Mastrogirolamo © Hitachi Medical Systems Europe 2017 All rights reserved. 6
Hi Def 3D

Image Parameters to define the quality

Line Density Frame Density


(3D) (3D)

Scan Angle

Frame Density changes the frame density of the


The combination of the parameters is
mechanical scan direction.The lower the frame
density, the higher the volume rate.
affecting the Quality of the 3D
Luca Mastrogirolamo
Reconstruction
© Hitachi Medical Systems Europe 2017 All rights reserved. 7
Hi Def 3D

Golden Preset
Line Density 1 Line Density 4 Line Density 2
Frame Density Low Frame Density High Frame Density Medium

Fast Acquisition Slow Acquisition Good Balance of


And Low Quality And High Quality Acquisition time and Quality

Luca Mastrogirolamo © Hitachi Medical Systems Europe 2017 All rights reserved. 8
Hi Def 3D – Scan Angle

Scan Angle 30% Scan Angle 100%


Small FOV. Good for early pregnancy Wide FOV for 2nd/3rd Trimester

Luca Mastrogirolamo
Scan Angle Affects the quality of 3D
© Hitachi Medical Systems Europe 2017 All rights reserved. 9
4D Mode

4D allow to see 3D reconstruction in real time.


Volume Rate (VPS) is affecting the quality of the reconstruction

Line Density 1 -Frame Density Low Line Density 4 -Frame Density High
Scan Angle 70% -> 3.9VPS Scan Angle 100% -> 0.8VPS
Up to 9.5 with Scan angle 10%
Luca Mastrogirolamo © Hitachi Medical Systems Europe 2017 All rights reserved. 10
4D Mode

Golden Presets to balance the image quality and the VPS

Line Density 2 – Frame Density High – Scan Angle 90% ->1.8VPS


Luca Mastrogirolamo © Hitachi Medical Systems Europe 2017 All rights reserved. 11
Renderings

Surface Skeleton S. Sharp

Preferred for anatomy Designed for fetal bone For surface rendering
Default in Golden sctructures. that sharpens the
Presets Requires more structures.
smoothing on the set Lower 3D Brightness
up is needed
Good for “Dark” 2D
Image
Luca Mastrogirolamo © Hitachi Medical Systems Europe 2017 All rights reserved. 12
Renderings

MIP X Ray Lighted

MIP renders the image at the Renders the image at the Reflective surface rendering.
highest echo intensity that is average echo intensity that is Enhance the three-dimensions
generated over the ray (the line generated over the ray (the line with a sense of depth using
of sight from the camera of sight from the camera lighting.
position). position). Useful for solid structures,
Semi-transparent image, useful Renders an image that is more internal organs and gynecology
for observing fetal skeletal transparent compared to “MIP”
structures and internal organs. Useful for Internal organs and
CNS
Luca Mastrogirolamo © Hitachi Medical Systems Europe 2017 All rights reserved. 13
Renderings

S Smooth Other (RMM) How to Set Up RMM


•  Select [Rendering Mode Mix] on
the touch panel.
•  Select a combination from [Surf.
Sharp + Lighted], [Surf. Smooth
+ Lighted], [Surf. Sharp + MIP] or
[Surf. Smooth + MIP] on the
touch panel.
•  Set a mix ratio in [Mix Ratio] on
the touch panel.
Example: When the combination
type is [Surf. Sharp + Lighted] and
the [Mix Ratio] is “30/70” “S. Sharp”
For surface rendering that Rendering Mode Mix S.Smooth
will be displayed at 30% while
smoothen the structures. + MIP – Mix Ratio 50/50
Surface with less power Rendering for internal “Lighted” will be displayed at 70%.
Good for contrasted 2D image parenchymal organs.
0/100 30/70 50/50 70/30 100/0

Luca Mastrogirolamo © Hitachi Medical Systems Europe 2017 All rights reserved. 14
Main Parameters to adjust on 3D4D

Opacity
Control

Smoothing Lower
(3D) Threshold

Gamma 3D
(3D) Brightness

Luca Mastrogirolamo © Hitachi Medical Systems Europe 2017 All rights reserved. 15
Opacity = Main Parameter

The gauge that lets you adjust the surface opacity level for a 3D image is called Opacity Control.
A high opacity level will display the surface of the image, while a low opacity level will render the
surface more transparent / translucent.
GP

Opacity 1 GP Opacity 12 Opacity 16


Less information of surface Good balance of Less filtered. Risk of lines
More translucent. information of surface and of the surface
Not for surface clear image.
Luca Mastrogirolamo © Hitachi Medical Systems Europe 2017 All rights reserved. 16
Lower Threshold

Lower Threshold is filtering Low Brightness Echo Noise. It represents one of the most important
parameters to adjust image quality and amount of information to show

GP

Lower Threshold 9 GP Lower Threshold 40 Lower Threshold 96


Low filtering -> noisy image Good balance for 2nd trim Too much filtering. Risk of
Surface. Removes noise removing too much
and clear the structures information
Luca Mastrogirolamo © Hitachi Medical Systems Europe 2017 All rights reserved. 17
3D Brightness

Adjust the brightness of a 3D image on a rendered 3D image

GP

3D Brightness 1 GP 3D Brightness 8 3D Brightness 16


Low Brightness on 3D Good balance of brightness High bright image. Too
image on surface rendering intense for surface.
Suitable for S.Sharp
Luca Mastrogirolamo © Hitachi Medical Systems Europe 2017 All rights reserved. 18
Depth Attenuation

Configures the level for changing the brightness in the depth direction to apply effective contrast
to the 3D image. Configures the contrast level for displaying regions close to the viewpoint as
bright and those farther away as dark

Depth Attenuation 1 Depth Attenuation 5 Depth Attenuation 10


Low effect of depth. Mid Level of depth Big effect of shadowing in
shadowing the deep points. Fading
background effect
Luca Mastrogirolamo © Hitachi Medical Systems Europe 2017 All rights reserved. 19
Color Depth

Color is used from upper edge of Clipping Box to back to enhance the sense of depth in 3D
images.This function is available only for RMM Surf. Sharp + Lighted & Surf. Smooth + Lighted

Color Depth OFF Color Depth On Color Depth OFF Color Depth On

Luca Mastrogirolamo © Hitachi Medical Systems Europe 2017 All rights reserved. 20
Color Depth Shift

Color Depth shift: The lower the value, the further the color at the back will extend to the front and
the larger the value, the smaller the back color area will be.

Color Depth Shift 55 Color Depth Shift 63


Good balance of shift Almost no shift

Luca Mastrogirolamo © Hitachi Medical Systems Europe 2017 All rights reserved. 21
Color Depth Contrast

Color Depth contrast: The higher the value, the greater the difference between the basic color
and color depth in the image.

Color Depth Contrast 10 Color Depth Contrast 55 Color Depth Contrast 63


Faded color Depth on all Balanced contrast for this High color depth contrast to
3D image depth highlight the depth
Luca Mastrogirolamo differences
© Hitachi Medical Systems Europe 2017 All rights reserved. 22
Color Depth Maps

Multiple Color Depth Maps are available

A B C D E

F G H I J

Luca Mastrogirolamo © Hitachi Medical Systems Europe 2017 All rights reserved. 23
Smoothing (3D)

Smoothing (3D) applies smoothing to volume data which has been created. This setting only
applies to 3D images. It does not apply to cross-sectional images.
New algorithm from previous version
GP

Smoothing (3D) OFF GP Smoothing (3D) MID2 Smoothing (3D) HIGH


Surface rendering is raw Best option for Surface It applies high level of
rendering smoothing. Good for crispy
Luca Mastrogirolamo surface
© Hitachi Medical Systems Europe 2017 All rights reserved. 24
Gamma (3D)

Gamma (3D) changes the brightness of the 3D. It adjust the echo brightness of the reconstructed
image.
GP

Gamma (3D) 1 Gamma (3D) 4 Gamma (3D) 16


Default GP level. Good Higher level of brightness Very high brightness and
balance of brightness echo echo lines artifacts
Luca Mastrogirolamo © Hitachi Medical Systems Europe 2017 All rights reserved. 25
4D Shading

Realistic rendering that displays the surface like a fetoscopy. More natural effect and lifelike skin
tones are used.
A Light Source is available to highlight and use the shadowing for a more realistic effect.

Do not place the source of light in the front.


You lose the Shadowing effect and the color tones

Luca Mastrogirolamo © Hitachi Medical Systems Europe 2017 All rights reserved. 26
Shadowing

Shadowing in 4DShading changes the directionality of light. The higher the value, the higher the
directionality. Diffuse to direct light effect

Shadowing 0 Shadowing 4 Shadowing 10


Diffuse light effects. Faded Medium Lighting with more Direct light with high
shadow is applied directional effect directionality. Spotlight
effect
Luca Mastrogirolamo © Hitachi Medical Systems Europe 2017 All rights reserved. 27
Light Enhancement Front/Back

•  Light Enhancement Front adjusts the front brightness of a 3D image. Changes the power of the lighting in
the front part
•  Light Enhancement Back adjusts the back brightness of a 3D image and change the lighting in the back of
the 3D image
GP

Light Enh Front 3 Light Enh Front 7 Light Enh Front 10

GP

Light Enh Back 0 Light Enh Back 5 Light Enh Back 10


Luca Mastrogirolamo © Hitachi Medical Systems Europe 2017 All rights reserved. 28
1St Trimester

On 1st trimester scanning the Set up of the ROI is important.


Big ROI includes lot of background structures. Small ROI defines better the anatomy of the baby

Big ROI Thin ROI

Set Higher Quality (Line Density 4 – Frame Density High – Scan Angle approx 50%).
Lower Threshold 28/30 ( Lower value to increase information) – Smoothing MID2/High
Luca Mastrogirolamo © Hitachi Medical Systems Europe 2017 All rights reserved. 29
1st Trimester

•  To get a nice transparent image of the baby use 4DShading with the backlight
•  The high quality set up gives good image quality of MPR plans

Transparent image of the Good MPR image quality


baby with backlight with High Quality setting

Luca Mastrogirolamo © Hitachi Medical Systems Europe 2017 All rights reserved. 30
1st Trimester

Color Depth with surface rendering is effective in 1st trimester.


Set Low Level of Lower Threshold and modify the smoothing according to the structure you have
to show. The smaller the structure, the lower the smoothing.

Embryo. Very early pregnancy 12 Weeks 13 Weeks

Luca Mastrogirolamo © Hitachi Medical Systems Europe 2017 All rights reserved. 31
3rd Trimester

The Set up for 3rd Trimester is similar to the one for the 2nd trimester.
An enlarged FOV is needed and slightly lower frequency.
2D Setup
•  Slightly higher Contrast
•  THI Mode -> FmT (HdTHI if easy patient)
•  Freq / Receiving Filter -> Low / Low (to be changed if needed)
•  HiRez ->5
•  Persist -> 0
•  Line Density ->3

3D Setup
•  Scan Angle 90/100%
•  Line Density 2
•  Frame Density Medium

Luca Mastrogirolamo © Hitachi Medical Systems Europe 2017 All rights reserved. 32
Spine

3D of the spine needs a specific rendering and some changes to preset:


Rendering Skeleton – Opacity 10-13 – Lower Threshold 35 or lower – Gamma 1-7

Gamma 1 Gamma 7
Luca Mastrogirolamo © Hitachi Medical Systems Europe 2017 All rights reserved. 33
Fetal Brain – Corpus Callosum

For Fetal Brain examination use a setup similar to 3rd Trimester scanning:
2D Setup
•  Slightly higher Contrast
•  THI Mode -> FmT (HdTHI if easy patient)
•  Freq / Receiving Filter -> Low / Low (to see well the fetal brain structures)
•  HiRez ->5
•  Persist -> 0-2
•  Line Density ->3/4

3D Setup (High Quality)


•  Scan Angle 90/100%
•  Line Density 3/4
•  Frame Density High

Luca Mastrogirolamo © Hitachi Medical Systems Europe 2017 All rights reserved. 34
Fetal Brain – Corpus Callosum

How to see Corpus Callosum from a surface rendering:


Fast solution
•  Start from normal 4 Views in Surface
•  Change to XRay rendering for a very fast visualization

For X Ray Rendering set higher Gamma (3D) and no change of the gain

Luca Mastrogirolamo © Hitachi Medical Systems Europe 2017 All rights reserved. 35
Fetal Brain – Corpus Callosum

How to see Corpus Callosum from a surface rendering:


VSI solution
•  VSI can be obtained with Xray, Surface and MIP renderings

21W

X Ray Rendering MIP Rendering


VSI Slice Thickness 8mm VSI Slice Thickness 8mm
Gamma 3D 7
Luca Mastrogirolamo © Hitachi Medical Systems Europe 2017 All rights reserved. 36
Fetal Brain – Corpus Callosum

MSI (Multi Slice Imaging) with VSI view is useful to explore the whole brain
Setup of MSI with VSI:
•  RefC (or the plane needed)
•  MSI Slice Thickness 1.5-2.0mm
•  VSI Thickness 8-10mm
•  Gamma(3D) 7-1

Luca Mastrogirolamo © Hitachi Medical Systems Europe 2017 All rights reserved. 37
Fetal Brain – Corpus Callosum

FMPR (Free Axis of MPR) is useful to quickly search for cross-sectional images of a specific
structure.

Luca Mastrogirolamo © Hitachi Medical Systems Europe 2017 All rights reserved. 38
Fetal Face and Palate

MSI Imaging is useful for fetal face and palate


Same setting of Fetal Brain

Luca Mastrogirolamo © Hitachi Medical Systems Europe 2017 All rights reserved. 39