Luca Mastrogirolamo
Luca Mastrogirolamo © Hitachi Medical Systems Europe 2017 All rights reserved.
Contents
Luca Mastrogirolamo © Hitachi Medical Systems Europe 2017 All rights reserved.
Concept of the Preset
Embedded
Dedicated on
3D Preset Anatomy 3D Setting in the
With 4 QSS Preset normal anatomy QSS
- 2D OR
- 3D (Anatomy) Dedicated 3D QSS.
- Skeleton Attention! 2D settings
- Corpus Callosum is effecting 3D
performances
Luca Mastrogirolamo © Hitachi Medical Systems Europe 2017 All rights reserved. 3
Recommended Preset for Surface rendering
Luca Mastrogirolamo © Hitachi Medical Systems Europe 2017 All rights reserved. 4
2D Setting
Pre-Formed Curves or
defined by user
Luca Mastrogirolamo © Hitachi Medical Systems Europe 2017 All rights reserved. 6
Hi Def 3D
Scan Angle
Golden Preset
Line Density 1 Line Density 4 Line Density 2
Frame Density Low Frame Density High Frame Density Medium
Luca Mastrogirolamo © Hitachi Medical Systems Europe 2017 All rights reserved. 8
Hi Def 3D – Scan Angle
Luca Mastrogirolamo
Scan Angle Affects the quality of 3D
© Hitachi Medical Systems Europe 2017 All rights reserved. 9
4D Mode
Line Density 1 -Frame Density Low Line Density 4 -Frame Density High
Scan Angle 70% -> 3.9VPS Scan Angle 100% -> 0.8VPS
Up to 9.5 with Scan angle 10%
Luca Mastrogirolamo © Hitachi Medical Systems Europe 2017 All rights reserved. 10
4D Mode
Preferred for anatomy Designed for fetal bone For surface rendering
Default in Golden sctructures. that sharpens the
Presets Requires more structures.
smoothing on the set Lower 3D Brightness
up is needed
Good for “Dark” 2D
Image
Luca Mastrogirolamo © Hitachi Medical Systems Europe 2017 All rights reserved. 12
Renderings
MIP renders the image at the Renders the image at the Reflective surface rendering.
highest echo intensity that is average echo intensity that is Enhance the three-dimensions
generated over the ray (the line generated over the ray (the line with a sense of depth using
of sight from the camera of sight from the camera lighting.
position). position). Useful for solid structures,
Semi-transparent image, useful Renders an image that is more internal organs and gynecology
for observing fetal skeletal transparent compared to “MIP”
structures and internal organs. Useful for Internal organs and
CNS
Luca Mastrogirolamo © Hitachi Medical Systems Europe 2017 All rights reserved. 13
Renderings
Luca Mastrogirolamo © Hitachi Medical Systems Europe 2017 All rights reserved. 14
Main Parameters to adjust on 3D4D
Opacity
Control
Smoothing Lower
(3D) Threshold
Gamma 3D
(3D) Brightness
Luca Mastrogirolamo © Hitachi Medical Systems Europe 2017 All rights reserved. 15
Opacity = Main Parameter
The gauge that lets you adjust the surface opacity level for a 3D image is called Opacity Control.
A high opacity level will display the surface of the image, while a low opacity level will render the
surface more transparent / translucent.
GP
Lower Threshold is filtering Low Brightness Echo Noise. It represents one of the most important
parameters to adjust image quality and amount of information to show
GP
GP
Configures the level for changing the brightness in the depth direction to apply effective contrast
to the 3D image. Configures the contrast level for displaying regions close to the viewpoint as
bright and those farther away as dark
Color is used from upper edge of Clipping Box to back to enhance the sense of depth in 3D
images.This function is available only for RMM Surf. Sharp + Lighted & Surf. Smooth + Lighted
Color Depth OFF Color Depth On Color Depth OFF Color Depth On
Luca Mastrogirolamo © Hitachi Medical Systems Europe 2017 All rights reserved. 20
Color Depth Shift
Color Depth shift: The lower the value, the further the color at the back will extend to the front and
the larger the value, the smaller the back color area will be.
Luca Mastrogirolamo © Hitachi Medical Systems Europe 2017 All rights reserved. 21
Color Depth Contrast
Color Depth contrast: The higher the value, the greater the difference between the basic color
and color depth in the image.
A B C D E
F G H I J
Luca Mastrogirolamo © Hitachi Medical Systems Europe 2017 All rights reserved. 23
Smoothing (3D)
Smoothing (3D) applies smoothing to volume data which has been created. This setting only
applies to 3D images. It does not apply to cross-sectional images.
New algorithm from previous version
GP
Gamma (3D) changes the brightness of the 3D. It adjust the echo brightness of the reconstructed
image.
GP
Realistic rendering that displays the surface like a fetoscopy. More natural effect and lifelike skin
tones are used.
A Light Source is available to highlight and use the shadowing for a more realistic effect.
Luca Mastrogirolamo © Hitachi Medical Systems Europe 2017 All rights reserved. 26
Shadowing
Shadowing in 4DShading changes the directionality of light. The higher the value, the higher the
directionality. Diffuse to direct light effect
• Light Enhancement Front adjusts the front brightness of a 3D image. Changes the power of the lighting in
the front part
• Light Enhancement Back adjusts the back brightness of a 3D image and change the lighting in the back of
the 3D image
GP
GP
Set Higher Quality (Line Density 4 – Frame Density High – Scan Angle approx 50%).
Lower Threshold 28/30 ( Lower value to increase information) – Smoothing MID2/High
Luca Mastrogirolamo © Hitachi Medical Systems Europe 2017 All rights reserved. 29
1st Trimester
• To get a nice transparent image of the baby use 4DShading with the backlight
• The high quality set up gives good image quality of MPR plans
Luca Mastrogirolamo © Hitachi Medical Systems Europe 2017 All rights reserved. 30
1st Trimester
Luca Mastrogirolamo © Hitachi Medical Systems Europe 2017 All rights reserved. 31
3rd Trimester
The Set up for 3rd Trimester is similar to the one for the 2nd trimester.
An enlarged FOV is needed and slightly lower frequency.
2D Setup
• Slightly higher Contrast
• THI Mode -> FmT (HdTHI if easy patient)
• Freq / Receiving Filter -> Low / Low (to be changed if needed)
• HiRez ->5
• Persist -> 0
• Line Density ->3
3D Setup
• Scan Angle 90/100%
• Line Density 2
• Frame Density Medium
Luca Mastrogirolamo © Hitachi Medical Systems Europe 2017 All rights reserved. 32
Spine
Gamma 1 Gamma 7
Luca Mastrogirolamo © Hitachi Medical Systems Europe 2017 All rights reserved. 33
Fetal Brain – Corpus Callosum
For Fetal Brain examination use a setup similar to 3rd Trimester scanning:
2D Setup
• Slightly higher Contrast
• THI Mode -> FmT (HdTHI if easy patient)
• Freq / Receiving Filter -> Low / Low (to see well the fetal brain structures)
• HiRez ->5
• Persist -> 0-2
• Line Density ->3/4
Luca Mastrogirolamo © Hitachi Medical Systems Europe 2017 All rights reserved. 34
Fetal Brain – Corpus Callosum
For X Ray Rendering set higher Gamma (3D) and no change of the gain
Luca Mastrogirolamo © Hitachi Medical Systems Europe 2017 All rights reserved. 35
Fetal Brain – Corpus Callosum
21W
MSI (Multi Slice Imaging) with VSI view is useful to explore the whole brain
Setup of MSI with VSI:
• RefC (or the plane needed)
• MSI Slice Thickness 1.5-2.0mm
• VSI Thickness 8-10mm
• Gamma(3D) 7-1
Luca Mastrogirolamo © Hitachi Medical Systems Europe 2017 All rights reserved. 37
Fetal Brain – Corpus Callosum
FMPR (Free Axis of MPR) is useful to quickly search for cross-sectional images of a specific
structure.
Luca Mastrogirolamo © Hitachi Medical Systems Europe 2017 All rights reserved. 38
Fetal Face and Palate
Luca Mastrogirolamo © Hitachi Medical Systems Europe 2017 All rights reserved. 39
Luca Mastrogirolamo © Hitachi Medical Systems Europe 2017 All rights reserved.