Académique Documents
Professionnel Documents
Culture Documents
John Pauly
1
ky a) b) ky
ky,max RF
Gz
Asymmetrically
Sampled Gy
kx
Gx
ky,s s (t)
Symmetrically
Sampled kx
-ky,s
Figure 3: Pulse sequence with a reduced echo time,
resulting in a partial echo readout.
-ky,max
Symmetrically Asymmetrically
ky
Sampled Sampled
Figure 2: Partial k-space acquisition for reducing scan
time by reducing the number of phase encodes required.
2
1.2 Applications to Other k-Space Ac- 2 Direct Partial k-Space Recon-
quisition Methods struction
Partial k-space acquisitions are also of interest for 2.1 Trivial Reconstruction by Zero
most other k-space acquisition methods. For EPI Padding
a partial k-space acquisition reduces the echo time,
which can be quite long for a fully symmetric ac- The simplest way to reconstruct a partial k-space
quisition. Reconstruction of this type of partial data set is to simply fill the uncollected data
k-space EPI data is very similar to the reduced (phase-encodes or readout samples) with zeroes.
phase-encode 2DFT case. Then, perform the 2DFT and display the magni-
tude. This works acceptably if the collected k-
Other trajectories are more interesting. Projec- space fraction is close to 1, and works poorly as
tion acquisitions can exploit partial k-space sym- this fraction approaches 0.5. This is illustrated in
metries, as can spiral acquisitions. The proper way Fig. 5 for a k-space fraction of 9/16th s. The recon-
to reconstruct these data sets is still an open re- struction of the full k-space data is shown in (a),
search question. We will briefly describe the issues and the reconstruction of the zero-padded partial
and possible solutions below. k-space data in (b). The result is significant blur-
ring in the phase-encode direction. Clearly this
is unacceptable, and this motivates the search for
other solutions.
3
a) W(ky)
ky
abs{w(y)}
real{w(y)}
y
imag{w(y)}
y
b)
4
image domain is a convolution in the frequency
domain, and the size of this convolution function
can be significant. The fact that the convolution
is operating on zero data for the uncollected phase
Initial partial k-space
encodes produces errors near the boundary. Be- data set Mpk(kx,ky)
low we will describe a conjugate synthesis method
based on k-space weighting that reduces this prob-
lem, and present some examples of reconstructed Ms(kx,ky)
images. Symmetric
Data DFT-1
5
Im
W(ky)
n
ni
ky
S+n
Symmetric (W(ky)+W*(-ky))/2
S
ky
Re
Antisymmetric (W(ky)-W*(-ky))/2
magnitude of |n| is given in [1]. Figure 10: The weighting function that truncates a
full k-space data set to a partial k-space data set can
be decomposed into symmetric and antisymmetric com-
ponents, corresponding to the real and imaginary com-
2.3 Homodyne Reconstruction ponents of the impulse response, shown in Fig. 6.
6
abs(mf(x,y)) Re{mf(x,y)p*(x,y)} difference*16
Figure 9: A full k-space reconstruction (left), and the real part of the phase corrected image. The phase
correction function was computed using ±1/16th of the k-space data (corresponding to a 9/16th s k-space re-
construction). The difference image shows vessels, as well as areas of rapid local change in phase, such as the
air-tissue interfaces in the sinuses.
7
Full k-Space Step Weighting Ramp Weighting
Figure 13: Comparison of the performance of different homodyne k-space weighting functions for a 9/16th s
data set. This is aggressive for phase gradients in this data set. On the left is the full k-space reconstruction. In
the middle a step k-space weighting has been used. Note the distinct ghosts from the subcutaneous fat in the
scalp (arrow). Using a ramp weighting, right, effectively eliminates this artifact. However, an additional artifact
appears above the auditory canal, where the image phase is rapidly changing (arrow).
8
On Resonance p/2 Phase shift in 5 voxels
Pre-Weighting
Initial partial k-space Function W(ky)
data set Mpk(kx,ky) Voxel a Voxel a
Voxel b Voxel b
Symmetric
Data
X
a+b a+b
DFT-1 DFT-1
ms(x,y) mpk(x,y)*w(x,y)
Phase
Corrected
9
Full k-Space Step Weighting Ramp Weighting
Figure 16: Comparison of the performance of different k-space weighting functions for a 5/8th s data set,
for a phase compensated conjugate synthesis reconstruction where the conjugate synthesis is done using the
same k-space weighting technique as is used in the homodyne reconstruction. On the left is the full k-space
reconstruction. In the middle a step k-space weighting has been used On the right a ramp has been used. Each
of these results in reasonable reconstructions. At 9/16th s ghosting is produced with either weighting function
(not shown), similar to the step weighted homodyne example of Fig. 13.
10
reconstruction has fewer artifacts than the step-
Zero Matrix
windowed (shown) or ramp windowed homodyne
reconstructions.
DFT-1
abs(mi(x,y))
ms(x,y) 5 References
X
p(x,y) = 1. D.C. Noll, D.G. Nishimura, and A. Macovski,
exp(i arg{ms(x,y)}) IEEE Transactions on Medical Imag., MI-
10(2), 154, (1991).
11
HD 9/16
HD 5/8
POCS 9/16
POCS 5/8
Image error*4
Figure 18: Comparison of partial k-space reconstructions of the axial head data of Fig. 1. The top two
are homodyne reconstructions at 9/16th s and 5/8th s k-space, and the lower two are POCS reconstructions for
the same k-space fractions. Difference images are computed relative to the full-k-space reconstructions. The
homodyne reconstructions at 9/16th s has clear ghosting. The POCS reconstruction at the same k-space fraction
produces fewer artifacts. Either works well at 5/8th s. 12
Phase Correct Full abs(PCF - Full)
Homodyne abs(HD-PCF)
POCS abs(POCS-PCF)