Académique Documents
Professionnel Documents
Culture Documents
Abstract
The purpose of this experiment is to perform analysis on a cantilever beam. Using a flexor
frame and loading screw, the applied load and shear force will be calculated from strain
measurements. The shear force and moment relationships are found using two stress
determinations, and using this information the linearity of the strain along the longitudinal
axis of the beam is found. Using these techniques, strain at points spaced at 1, 4, and 7
inches along the beam was measured to be 1736, 1111, and 581 microstrain respectively.
Calculated strain was found to be 2004.64, 1336.43, and 668.31 with an average error of
16%.
2
Richard Dyar 2/7/2011
Background
Cantilever beams are commonly designated as a structure supported or fixed at only one
principles. Forces such as shear, bending moments, and reaction, can all be found using
simple problems to represent more complicated structures. From these shear and moment
diagrams, as well as shear stress calculations can be found in order to find the weakest
σx =6P(L-x)bt2 (1)
Where σ, is stress along the beams axis, P is the load applied to the beam at length L. The
respectively. Equation 1 indicates that normal stress on the surface of the beam is tensile;
because of this the surface does not experience shear stress. This allows the use of Hooke’s
Law, equation 2, because of uniaxial stress. Again σ is stress, E being modulus of elasticity,
and ε is strain.
σx =Eεx (2)
Equation 3 is the centroid of the beam, usually given or found through tables. Where I
I=bt312 (3)
3
Richard Dyar 2/7/2011
Using these equations and some algebra, equation 4 is found. Values hold the same with the
addition of δend which is found using the initial and final deflections found during the
experiment.
δend =δfinal-δinitial
P=3EIδend L3 (4)
Procedure
Beam Dimensions
The beam has a gage factor of 2.085 and a
Width(b) 1 inch
transverse sensitivity of -+1, these values are
Thickness(t) .25
inches used to set the strain indicator.
Fixed point to strain gage 1 inch Load Estimate 1, using equation 1 solved
1(x1) for P:
Fixed point to strain gage 4 inches
2(x2) P1,2=-Ebt26*ε1-ε2x1-x2=-10.4*106
P2,3=-Ebt26*ε2-ε3x2-x3=-10.4*106
*1*.2526*600*10-6-3=21.66
P=P1,2+P2,32=24.34 lb
Individual Strains:
ε=εinitial-εfinal
4
Richard Dyar 2/7/2011
ε1=0--1736=1736
ε2=-1614--2574=960
ε3=-1789--2370=581
Strain Gradient:
The graph shows the gradient of the strain across the surface of the member used
in the experiment. By plotting the points and taking the slope of the trend line the
Load Estimate 2:
P=-Ebt26*-192.5*10-6=-10.4*106 *1*.2526*-192.5*10-6=20.854 lb
Load Estimate 3:
δend =.7511-.157=.549
I=1*.25312=.00130 (3)
P=3*10.4*106*.00130*.549103=24.13 lb
σx =6P(L-x)bt2=6*24.34*91*.252=21029.76 psi
5
Richard Dyar 2/7/2011
σ1 =Eε=6P(L-x)bt2
ε1=6P(L-x)Ebt2=6*24.13*(10-1)10.4*106 *1*.252=2004.64
%error =εcalc-εmeasuredεmeasured*100=2004.64-17361736*100=15%
Results
Stress Estimate 1
Estimate 1 Estimate 3
Stress Estimate
Gag
6
Richard Dyar 2/7/2011
1 18054.4
psi
2 11554.4
psi
3 6042.4 psi
Stress Comparisons
Estimate
1 psi
2 psi
Strain Calculations
Conclusion
The values calculated were found to be consistent in each estimate, and close to the
measured values. The margins of error are fairly large due to the strain indicator’s loose
connections, and it is most likely not getting an accurate reading of strain. There is also
slight operator error in this experiment within micrometer measurement as well as other
recorded values and calculations. Other than those errors the calculations show that
7
Richard Dyar 2/7/2011
multiple ways of finding stresses and strains can be found given a few different values, and
References
*Gilbert, J. A and C. L. Carmen. "Chapter 7 – Poisson's Ratio Flexure Test." MAE/CE 370