Installation

Copyrighte 1999 by General Electric Company

Operating Documentation
GE MEDICAL SYSTEMS TGP ROM KIT INSTALLATION
REV 1 2246983

D THIS SERVICE MANUAL IS AVAILABLE IN ENGLISH ONLY.


WARNING D IF A CUSTOMER’S SERVICE PROVIDER REQUIRES A LANGUAGE OTHER
THAN ENGLISH, IT IS THE CUSTOMER’S RESPONSIBILITY TO PROVIDE
TRANSLATION SERVICES.
D DO NOT ATTEMPT TO SERVICE THE EQUIPMENT UNLESS THIS SERVICE
MANUAL HAS BEEN CONSULTED AND IS UNDERSTOOD.
D FAILURE TO HEED THIS WARNING MAY RESULT IN INJURY TO THE SERVICE
PROVIDER, OPERATOR OR PATIENT FROM ELECTRIC SHOCK,
MECHANICAL OR OTHER HAZARDS.

D CE MANUEL DE MAINTENANCE N’EST DISPONIBLE QU’EN ANGLAIS.


AVERTISSEMENT D SI LE TECHNICIEN DU CLIENT A BESOIN DE CE MANUEL DANS UNE AUTRE
LANGUE QUE L’ANGLAIS, C’EST AU CLIENT QU’IL INCOMBE DE LE FAIRE
TRADUIRE.
D NE PAS TENTER D’INTERVENTION SUR LES ÉQUIPEMENTS TANT QUE LE
MANUEL SERVICE N’A PAS ÉTÉ CONSULTÉ ET COMPRIS.
D LE NON-RESPECT DE CET AVERTISSEMENT PEUT ENTRAÎNER CHEZ LE
TECHNICIEN, L’OPÉRATEUR OU LE PATIENT DES BLESSURES DUES À DES
DANGERS ÉLECTRIQUES, MÉCANIQUES OU AUTRES.

D DIESES KUNDENDIENST–HANDBUCH EXISTIERT NUR IN


ENGLISCHER SPRACHE.
WARNUNG
D FALLS EIN FREMDER KUNDENDIENST EINE ANDERE SPRACHE BENÖTIGT,
IST ES AUFGABE DES KUNDEN FÜR EINE ENTSPRECHENDE ÜBERSETZUNG
ZU SORGEN.
D VERSUCHEN SIE NICHT, DAS GERÄT ZU REPARIEREN, BEVOR DIESES
KUNDENDIENST–HANDBUCH NICHT ZU RATE GEZOGEN UND VERSTANDEN
WURDE.
D WIRD DIESE WARNUNG NICHT BEACHTET, SO KANN ES ZU VERLETZUNGEN
DES KUNDENDIENSTTECHNIKERS, DES BEDIENERS ODER DES PATIENTEN
DURCH ELEKTRISCHE SCHLÄGE, MECHANISCHE ODER SONSTIGE
GEFAHREN KOMMEN.

D ESTE MANUAL DE SERVICIO SÓLO EXISTE EN INGLÉS.


AVISO D SI ALGÚN PROVEEDOR DE SERVICIOS AJENO A GEMS SOLICITA UN IDIOMA
QUE NO SEA EL INGLÉS, ES RESPONSABILIDAD DEL CLIENTE OFRECER UN
SERVICIO DE TRADUCCIÓN.
D NO SE DEBERÁ DAR SERVICIO TÉCNICO AL EQUIPO, SIN HABER
CONSULTADO Y COMPRENDIDO ESTE MANUAL DE SERVICIO.
D LA NO OBSERVANCIA DEL PRESENTE AVISO PUEDE DAR LUGAR A QUE EL
PROVEEDOR DE SERVICIOS, EL OPERADOR O EL PACIENTE SUFRAN
LESIONES PROVOCADAS POR CAUSAS ELÉCTRICAS, MECÁNICAS O DE
OTRA NATURALEZA.

a
GE MEDICAL SYSTEMS TGP ROM KIT INSTALLATION
REV 1 2246983

D ESTE MANUAL DE ASSISTÊNCIA TÉCNICA SÓ SE ENCONTRA


DISPONÍVEL EM INGLÊS.
ATENÇÃO
D SE QUALQUER OUTRO SERVIÇO DE ASSISTÊNCIA TÉCNICA, QUE NÃO A
GEMS, SOLICITAR ESTES MANUAIS NOUTRO IDIOMA, É DA
RESPONSABILIDADE DO CLIENTE FORNECER OS SERVIÇOS DE TRADUÇÃO.
D NÃO TENTE REPARAR O EQUIPAMENTO SEM TER CONSULTADO E
COMPREENDIDO ESTE MANUAL DE ASSISTÊNCIA TÉCNICA.
D O NÃO CUMPRIMENTO DESTE AVISO PODE POR EM PERIGO A SEGURANÇA
DO TÉCNICO, OPERADOR OU PACIENTE DEVIDO A‘ CHOQUES ELÉTRICOS,
MECÂNICOS OU OUTROS.

D IL PRESENTE MANUALE DI MANUTENZIONE È DISPONIBILE


SOLTANTO IN INGLESE.
AVVERTENZA
D SE UN ADDETTO ALLA MANUTENZIONE ESTERNO ALLA GEMS RICHIEDE IL
MANUALE IN UNA LINGUA DIVERSA, IL CLIENTE È TENUTO A PROVVEDERE
DIRETTAMENTE ALLA TRADUZIONE.
D SI PROCEDA ALLA MANUTENZIONE DELL’APPARECCHIATURA SOLO DOPO
AVER CONSULTATO IL PRESENTE MANUALE ED AVERNE COMPRESO IL
CONTENUTO.
D NON TENERE CONTO DELLA PRESENTE AVVERTENZA POTREBBE FAR
COMPIERE OPERAZIONI DA CUI DERIVINO LESIONI ALL’ADDETTO ALLA
MANUTENZIONE, ALL’UTILIZZATORE ED AL PAZIENTE PER
FOLGORAZIONE ELETTRICA, PER URTI MECCANICI OD ALTRI RISCHI.

b
GE MEDICAL SYSTEMS TGP ROM KIT INSTALLATION
REV 1 2246983

IMPORTANT! . . . X-RAY PROTECTION


X-ray equipment if not properly used may cause injury. Accordingly, the instructions herein contained should
be thoroughly read and understood by everyone who will use the equipment before you attempt to place this
equipment in operation. The General Electric Company, Medical Systems Group, will be glad to assist and
cooperate in placing this equipment in use.
Although this apparatus incorporates a high degree of protection against x-radiation other than the useful beam, no
practical design of equipment can provide complete protection. Nor can any practical design compel the operator to
take adequate precautions to prevent the possibility of any persons carelessly exposing themselves or others to
radiation.
It is important that everyone having anything to do with x-radiation be properly trained and fully acquainted with the
recommendations of the National Council on Radiation Protection and Measurements as published in NCRP Reports
available from NCRP Publications, 7910 Woodmont Avenue, Room 1016, Bethesda, Maryland 20814, and of the
International Commission on Radiation Protection, and take adequate steps to protect against injury.
The equipment is sold with the understanding that the General Electric Company, Medical Systems Group, its agents,
and representatives have no responsibility for injury or damage which may result from improper use of the equipment.
Various protective material and devices are available. It is urged that such materials or devices be used.

CERTIFIED ELECTRICAL CONTRACTOR STATEMENT


All electrical installations that are preliminary to positioning of the equipment at the site prepared for the equipment shall be
performed by licensed electrical contractors. In addition, electrical feeds into the Power Distribution Unit shall be performed
by licensed electrical contractors. Other connections between pieces of electrical equipment, calibrations, and testing shall
be performed by qualified GE Medical personnel. The products involved (and the accompanying electrical installations) are
highly sophisticated, and special engineering competence is required.

In performing all electrical work on these products, GE will use its own specially trained field engineers. All of GE’s electrical
work on these products will comply with the requirements of the applicable electrical codes.

The purchaser of GE equipment shall only utilize qualified personnel (i.e., GE’s field engineers, personnel of third-party
service companies with equivalent training, or licensed electricians) to perform electrical servicing on the equipment.

c
GE MEDICAL SYSTEMS TGP ROM KIT INSTALLATION
REV 1 2246983

DAMAGE IN TRANSPORTATION

All packages should be closely examined at time of delivery. If damage is apparent, have notation “damage in
shipment” written on all copies of the freight or express bill before delivery is accepted or “signed for” by a General
Electric representative or a hospital receiving agent. Whether noted or concealed, damage MUST be reported to the
carrier immediately upon discovery, or in any event, within 14 days after receipt, and the contents and containers held
for inspection by the carrier. A transportation company will not pay a claim for damage if an inspection is not requested
within this 14 day period.

Call Traffic and Transportation, Milwaukee, WI (414) 827–3449 / 8*285–3449 immediately after damage is found. At
this time be ready to supply name of carrier, delivery date, consignee name, freight or express bill number, item
damaged and extent of damage.

Complete instructions regarding claim procedure are found in Section “S” of the Policy & Procedure Bulletins.

OMISSIONS & ERRORS

GE personnel, please use the GEMS CQA Process to report all omissions, errors, and defects in this documentation.
Customers, please contact your GE Sales or Service represenatives.

CAUTION

Do not use the following devices near this equipment. Use of these devices near this equipment could cause
this equipment to malfunction.

Devices not to be used near this equipment:

Devices which intrinsically transmit radio waves such as; cellular phone, radio transceiver, mobile radio transmitter,
radio–controlled toy, etc.

Keep power to these devices turned off when near this equipment.

Medical staff in charge of this equipment is required to instruct technicians, patients and other people who may be
around this equipment to fully comply with the above regulation.

d
GE MEDICAL SYSTEMS TGP ROM KIT INSTALLATION
REV 2 2246983

REVISION HISTORY

REV DATE PRIMARY REASON FOR CHANGE


0 . . . . June 17, 1999 . . . Initial release.
1 . . . . . July 23, 1999 . . . Wj, Vj, Lj, SLj, XMj information added
2 . October 13, 1999 . . . Patch procedures added

LIST OF EFFECTIVE PAGES

PAGE REV PAGE REV PAGE REV PAGE REV PAGE REV
Title page . . . . . . . . 2 i................. 2 Tab 2 Tab 3
Title page rear . blank ii . . . . . . . . . . . . . blank 2–1 . . . . . . . . . . . . . . 1 3–1 . . . . . . . . . . . . . . 1 Blank/Rear cover . . –
a to d . . . . . . . . . . . . 1 2–2 . . . . . . . . . . . blank 3–2 to 3–3 . . . . . . . . 2
A ................ 2 Tab 1 3–4 . . . . . . . . . . . blank
B . . . . . . . . . . . . . blank 1–1 to 1–2 . . . . . . . . 1

A
GE MEDICAL SYSTEMS TGP ROM KIT INSTALLATION
blank 2246983

B
GE MEDICAL SYSTEMS TGP ROM KIT INSTALLATION
REV 2 2246983

TABLE OF CONTENTS

SECTION PAGE

SECTION 1 – INTRODUCTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1–1

SECTION 2 – MATERIAL LIST . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2–1

SECTION 3 – INSTALLATION PROCEDURE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3–1


3-1 REPLACING ROMS ON THE TGP BOARD . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3–1
3-2 PATCH PROCEDURES . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3–3
3-3 FUNCTIONAL CHECK . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3–3

i
GE MEDICAL SYSTEMS TGP ROM KIT INSTALLATION
blank 2246983

ii
GE MEDICAL SYSTEMS TGP ROM KIT INSTALLATION
REV 1 2246983

SECTION 1 – INTRODUCTION

1-1 TGP ROM KIT FOR ZJ/QJ


This is the kit for the Zj/Qj systems with the following TGP board (other than P9183XA):

D For HLT system:


2123546. 2123546–2, 2123546–3. 2123546–4. . 2123546–5.
2166454, 2166454–2

D For Xe system:
2106126, 2106126–2, 2106126–3, 2106126–4, 2106126–5,
2176944, 2176944–2

This kit can resolve the problems below.

D Surge current on the slip ring 115V line occurs.

D Tilt angle is changed without tilt operation.

D Scout scan angle at Biopsy mode moves.

1-2 TGP ROM KIT FOR XMJ


This is the kit for the XMj systems with the following TGP board:

D 2107711, 2107711–2, 2107711–3, 2107711–4, 2107711–5)

This kit can resolve the problems below.

D Surge current on the slip ring 115V line occurs.

D Tilt angle is changed without tilt operation.

D Scout scan angle at Biopsy mode moves.

1–1
GE MEDICAL SYSTEMS TGP ROM KIT INSTALLATION
REV 1 2246983

1-3 TGP ROM KIT FOR WJ–C, VJ, LJ, SLJ


This is the kit for the Wj–C, Vj, Lj, and SLj systems with the following TGP board:

D Wj–C (System Version: 10.xx or later)


P9163XG–05

D Vj (System Version: 6.xx or later)


P9153XG–06, P9153XG–07, P9153XA–05

D Lj (System Version: 4.xx or later)


P9153XG–06, P9153XG–07, P9153XA–05

D SLj (System Version: 3.xx or later)


2108988

This kit can resolve the problems below.

D Tilt angle is changed without tilt operation.

1–2
GE MEDICAL SYSTEMS TGP ROM KIT INSTALLATION
REV 1 2246983

SECTION 2 – MATERIAL LIST

TGP ROM Kit


2246378

ITEM PART NO. DESCRIPTION QTY.


1 2123543–6 GP ROM Assy (for HLT Gantry) 1
2 2106127–6 GP ROM Assy (for Xe Gantry) 1
3 2106128–5 TP ROM Assy 1
4 2106129–5 MP ROM Assy 1
5 2246379 TGP Board Label (2123546–6) 1
6 2246380 TGP Board Label (2166454–3) 1
7 2246381 TGP Board Label (2106126–6) 1
8 2246382 TGP Board Label (2176944–3) 1
9 2246983 Installation instruction (This manual) 1

TGP ROM Kit


2249350

ITEM PART NO. DESCRIPTION QTY.


1 2108989 GP ROM Assy 1
2 2241702 TP ROM Assy 1
3 2249351 TGP Board Label (2241699) 1
4 2249352 TGP Board Label (2108988–2) 1
5 2249353 TGP Board Label (2241698) 1
6 2246983 Installation instruction 1

TGP ROM Kit


2248907

ITEM PART NO. DESCRIPTION QTY.


1 2107712–6 GP ROM Assy 1
2 2107713–4 TP ROM Assy 1
3 2106129–5 MP ROM Assy 1
4 2248908 TGP Board Label (2107711–6) 1
5 2246983 Installation instruction 1

2–1
GE MEDICAL SYSTEMS TGP ROM KIT INSTALLATION
blank 2246983

2–2
GE MEDICAL SYSTEMS TGP ROM KIT INSTALLATION
REV 1 2246983

SECTION 3 – INSTALLATION PROCEDURE

3-1 REPLACING ROMS ON THE TGP BOARD

CAUTION

D Wear the wrist strap and place yourself at ground potential before accessing any board.

D ROM/PAL Handling Caution! Take care not to bend the pins of ROM or PAL during removing or
installing. When installing a ROM or PAL, align its cut–out or groove with the cut–out of the
base.

1. Turn OFF the power switch of the Operator Console.

2. Remove the left side cover from the Gantry.

3. Switch OFF all switches at the Gantry rear base.

4. Replace the following ROMs on the TGP Board listed below with new ones.

When installing the TGP ROM KIT 2246378:


ROM ROM Location New ROM Part No. Remarks
GP ROM U7P 2123543–6 For HLT system
2106127–6 For Xe system
TP ROM U3P 2106128–5 –
MP ROM U3C 2106129–5 –

When installing the TGP ROM KIT 2249350:


ROM ROM Location New ROM Part No. Remarks
GP ROM U7J 2108989 TGP Board P9153XA–05 ONLY
TP ROM U4J 2241702 –

When installing the TGP ROM KIT 2248907:


ROM ROM Location New ROM Part No. Remarks
GP ROM U7P 2107712–6 –
TP ROM U3P 2107713–4 –
MP ROM U3C 2106129–5 –

3–1
GE MEDICAL SYSTEMS TGP ROM KIT INSTALLATION
REV 2 2246983

3-1 REPLACING ROMS ON THE TGP BOARD (Continued)


5. Attach the following Part No. label listed below. This depends on TGP board.

When installing the TGP ROM KIT 2246378:


Current TGP Part No. New TGP Part No.
2123546–xx 2123546–6
2166454–xx 2166454–3
2106126–xx 2106126–6
2176944–xx 2176944–3

When installing the TGP ROM KIT 2249350:


Current TGP Part No. New TGP Part No.
P9153XG–06, P9153XG–07 2241699
P9153XA–05, 2108988 2108988–2
P9163XG–05 2241698

When installing the TGP ROM KIT 2248907:


New TGP Part No.
2107711–6

6. Install the Gantry cover in the reverse order of removal.

7. Switch ON all switches at the Gantry rear base.

3–2
GE MEDICAL SYSTEMS TGP ROM KIT INSTALLATION
REV 2 2246983

3-2 PATCH PROCEDURES


For Qj, Zj–adv., and XMj (except Synergy) with the following version of the system software ONLY, the patch proce-
dure is required to move the tube to the daily potision after warm–up scans.

Qj series : V/R 3.02 – V/R 6.03


Zj–adv. series : V/R 1.03 – V/R 4.01
XMj series (except Synergy) : V/R 2.02 – V/R 3.03

For the systems other than the system version above, the patch procedure does not need to be performed.

1 Insert the BOOT floppy into the floppy disk drive.

2 Press ”Mgt” Key, then Enter ”$$”.


”>” appears on the CRT monitor.

3 Enter the following.

> FN SYS:0.DMP.OPTB.VC ’CR’


> FD 2FE 1 ’CR’
0002FE 00 05 00 02 00 ....
> FM ’CR’
00 ’CR’
0002FE 00 00 00 02 00 ....
> RT ’CR’

4 Press ”End” Key.

5 Pull the system reset switch.

3-3 FUNCTIONAL CHECK


1. Switch ON the main switch of the OC.

The test progress information is displayed on the CRT monitor.

Verify that the system starts up normally.

2. Perform Scan Test.

Verify that each scan completes with no error.

3–3
GE MEDICAL SYSTEMS TGP ROM KIT INSTALLATION
blank 2246983

3–4