Vous êtes sur la page 1sur 5

Journal of Medical and Biological Engineering, 26(4): 149-153 149

Heterodyne Interferometric Surface Plasmon


Resonance Biosensor
Shen-Fen Joe Li-Zen Hsieh* Liann-Be Chang1 Chih-Chung Hsieh2

Chien-Ming Wu2
Department of Electrical Engineering, Chung-Cheng Institute of Technology, National Defense University, Tao-Yuan, Taiwan, 335 ROC
1
Department of Electronic Engineering, Chang Gung University, Tao-Yuan, Taiwan, 333, ROC
2
Department of Biomedical Engineering & Environmental Sciences, National Tsing-Hua University, Hsin-Chu, Taiwan, 300 ROC

Received 30 Jun 2006; Accepted 3 Nov 2006

Abstract
A high-sensitivity surface plasmon resonance (SPR) biosensor is constructed by using a common-path heterodyne
interferometric system. The beam combiner is employed to combine the modulated He-Ne TE-wave and TM-wave into
a heterodyne light source with a frequency difference of 60 kHz. Using the common-path heterodyne interferometric
system to obtain the phase shift between the two waves has the advantage of high-sensitivity real-time phase detection
for monitoring molecular interactions. By detecting the phase shift, the proposed SPR biosensor is employed to monitor
the interactions between the sheep IgG covalently immobilized on the sensor chip surface and the anti-sheep IgG
contained in the running buffer.

Keywords: Surface Plasmon Resonance, Biosensor, Heterodyne interferometry

occur, thus resulting in changes in the refractive index and


Introduction thickness of the dielectric biosensing layer. Therefore,
Surface plasmon resonance (SPR) has the advantages of examining the changes in the intensity of the reflected light
being highly sensitive, requires no fluorescent-labeled analytes can reveal the molecule interactions [5-7]. Methods for
or receptors, and is capable of monitoring biomolecular detecting the intensity of reflected light have been applied to
interactions in real time. This technique has been employed in SPR biosensor [8-9]. The angle of incidence and wavelength
commercially available optical biosensors [1-2]. SPR is the of the reflected light at its lowest intensity are taken as the
collective oscillation of electrons that occurs at the interface resonance angle and resonance wavelength, respectively.
between a dielectric and a good conductor. To make it resonant, During resonance, there is also a significant phase change of
the parallel polarization wave vector of the incident light the reflected TM-wave [10]. The phase detection is more
matches the wave vector of the surface plasmon [3-4]. Factors sensitive than those of angle or wavelength measurement [11].
affecting the resonance include the incident angle and There have been a few studies on the phase measurement
wavelength of the light source, as well as the permittivity of [12-16], and among these, the common-path interferometers
the dielectric material near the metal surface. During method has been found to be less affected by external
resonance, electrons from the metal absorb most of the energy, disturbance [17]. The TE-wave and TM-wave of different
thus rapidly decreasing the intensity of the reflected light. frequencies are employed to obtain the reference and measured
According to this principle, with other parameters of the beating frequency signals. Most of the noise is eliminated by
reflected light held constant, changes in the intensity of the means of a high-quality filter, thus enhancing the purity of the
reflected light will indicate the changes in the refractive index signal [18].
of the metal-adjoining dielectric material. In this study, the SPR biosensor is developed using a
In SPR biosensing, the ligand is immobilized on a metal common-path heterodyne interferometer that is employed to
surface to form a biosensing film. The sample containing the detect the interactions between the sheep IgG and its antibody
target analyte is passed over the surface of the biosensing film, at concentrations ranging from 0.055 to 110µg/ml. Two
where interactions between the target analyte and ligand may acousto-optical modulators are utilized to modulate the
TE-wave and TM-wave, thus forming a heterodyne light
source with a frequency difference of 60 kHz. Observing the
* Corresponding author: Li-Zen Hsieh
Tel: +866-3-3800301 ext.265; Fax: +886-3-3801407 phase difference between the reference and measured signals
E-mail: lzhsieh@ccit.edu.tw obtained by the dual-phase lock-in amplifier can reveal the
150 J. Med. Biol. Eng., Vol. 26. No. 4 2006

phase shift between the TE-wave and TM-wave. To examine


the performance of the proposed SPR biosensor, comparison is
also made with the commercial optical sensor Biacore X.

Principle
θ
Figure 1 displays the five-layer Kretschmann-Raether Prism
configuration of a SPR sensor chip. The five layers from top to Gold film
bottom are prism, gold film, linker, hydrogel and liquid, all of Linker
which serve different functions. The prism provides total Hydrogel
internal reflection for inducing SPR. The linker layer, which is
attached to the gold film, provides a hydrophilic environment Liquid
for the immobilization of molecules passing through the
Figure 1. The five-layer Kretschmann-Raether configuration
hydrogel layer. The antigen and antibody from the solution
interact and are bound covalently when passing through the
M1 AOM1 PBS1
hydrogel layer. When the incident light reaches the interface
between the prism and the metal, the component of the wave Laser
vector in the x direction is
kx = ko np Sinθ (1)
AOM2
where k0=2 π λ , np is the refractive index of the prism,
and θ is the angle of incidence. The wave vector of surface PBS2
plasmon resonance is M2
ω ε mε d Flow cell
K SP = ε m +ε d (2) Rotary stage
c
BS
where ε m is the permittivity of gold and ε d is the Pol2
permittivity of the dielectric sample.
Pol1 PhD2
When both the TE-wave and TM-wave hit the interface of
the dielectric material, the reflection coefficient of the
PhD1 Lock-in amplifier
TE-wave, rs, and the reflection coefficient of the TM-wave, rp,
can be expressed respectively as Figure 2. Schematic diagram of the common-path heterodyne
iδ s interferometer
rs = rs e (3)

donkey anti-sheep IgG (whole molecule) were obtained from


rp = rp eδ p (4)
i
SIGMA (St. Louis, MO, USA).
3.2 Experimental Setup
where δs and δp denote the phase shift of the TE-wave and
Figure 2 is the schematic diagram of the common-path
TM-wave, respectively. Since the TE-wave cannot induce SPR,
heterodyne interferometer used for detecting molecular
there is little variation in the phase shift near the resonance
interactions. The light source is a linear polarized stabilized
angle; hence, δs can be regarded as constant. In fact, phase shift
He-Ne laser of 632 nm wavelength. The incident light is split
of the TM-wave δp shows great variation near the resonance
equally into two beams, TE-wave and TM-wave, by the
angle. The phase difference, Φ, between the TM-wave and
polarization beam splitter (PBS1). The TE-wave and TM-wave
TE-wave can be expressed as
are then modulated by two acoustooptic modulators AOM1 and
Φ=δp-δs. (5) AOM2 (both are InterAction Corp, 400 series), with
frequencies of 40 MHz and 40.06 MHz, respectively. After
Since there exists a linear relationship between the phase modulation, the two beams are combined into the heterodyne
difference, Φ, and phase shift of TM-wave, δp, molecular light source with a frequency difference of 60 KHz by the beam
interactions on the sensor surface can be detected by measuring combiner (PBS2). Passing through a beam splitter (BS), the
the phase difference. heterodyne light source is again divided into two beams. One
passes through a polarizer (Pol1) and enters the photo-detector
Experiment (PhD1) as a reference signal, Ir, for the lock-in amplifier
(Standford Research System, model CR850). The other is
3.1 Buffers and reagents transmitted through two prisms, both of which are fixed on the
Sensor chip CM5, amine-coupling kits, HBS Buffer rotary stage (Newport, URM100PP). One prism has its side in
(0.01M HEPES pH=7.4, 0.15M NaCl, 3mM EDTA, 0.005% close contact with the sensor chip and flow cell, while the other
Surfactant P20), sodium acetate (pH=4.0,10mM) and provides total internal reflection of the incident light beam. In
glycine-HCl (10mM pH= 2.0) were all purchased from this way, the incident light beam transmitted onto the rotary
Pharmacia Biosensor AB (Uppsala, Sweden). Sheep IgG and stage will be parallel to that reflected from the rotary stage.
Heterodyne Interferometric SPR Biosensor 151

Finally, this beam passes through another polarizer (Pol2) and


enters the photo-detector (PhD2) to become a measurement
signal, Im, for the lock-in amplifier in detecting phase shift. The
reference signal, Ir, and measurement signal, Im, can be
expressed respectively as

I r ∝ Eso Epo Cos( ∆ ωt) (6)

I m ∝ r Ep0Ep0 Cos( ∆ ωt+ Φ )



(7)

Eso and Epo denote the amplitudes of the TE-wave and


TM-wave, respectively. ∆ ω is the difference in beating
frequency between the TE-wave and TM-wave. Φ , the phase
difference between TM-wave and TE-wave, is detected by the Figure 3. Phase shift of sheep IgG immobilized on the sensor chip
senor chip and that after being reflected by the prism. surface using the amine-coupling method
Therefore, the difference between the measurement signal and
the reference signal obtained by the dual-phase lock-in
amplifier can provide real-time detection of the interaction
between biological molecules on the sensor chip surface.

Results and Discussion


4.1 Sheep IgG immobilization
To immobilize IgG involves binding it covalently to the
CM5 sensor chip using the amine-coupling method. All
reagents are injected by a peristaltic pump. First, the running
buffer, HEPES, is injected into the sensor chip, providing a
standard for future comparison. The sensor chip is then
activated by a mixture of EDC and NHS at a 1:1 ratio. The
sheep IgG is mixed with sodium acetate (pH = 4.0) at a 1:100
Figure 4 The sensorgram obtained using the common-path
ratio, and then injected into the flow cell for covalent heterodyne interferometer
immobilization on the CM5 sensor chip. The running buffer,
HEPES, is again injected into the flow cell to remove free 4.3 Measurement with different concentration
sheep IgG not bounded on the sensor chip. To prevent Figure 4 displays the binding of anti-sheep IgG at various
non-specific binding, ethanolamine is employed to deactivate concentrations to immobilized sheep IgG obtained using the
the remaining active esters of the unbound sheep IgG. phase-measurement-based SPR biosensor. As can be seen,
Non-specific or unwanted binding can reduce overall assay there are increases in phase shift of 4, 14, 33, 62 and 107
performance and sensitivity at extremely low detection degrees at anti-sheep IgG concentrations of 0.055, 0.11, 1.1, 11
concentrations. and 110 µg/ml. At concentrations below 1.1 µg/ml, the
Figure 3 shows the phase shift and time taken for increase is slow and gradual, while at concentrations of 110
immobilization of sheep IgG on the sensor chip surface. As µg/ml, the increase is sharp and marked.
can be seen, there is an increase of 85 degree in phase shift as Figure 5 shows the relationship between phase shift and
a result of sheep IgG immobilization, indicating that the sheep concentration obtained using the developed SPR biosensor. As
IgG has been successfully immobilized on the sensor chip can be seen, there exists a linear relationship between the
surface. phase difference and various concentrations of anti-sheep IgG.
4-2 Biosensing Protocol Their correlation can be expressed as Y = 5.595 logX + 38.225
To analyze the antigen-antibody interaction, the running with R2 =0.9686, where X is the concentration of anti-sheep
buffer HEPES is first injected into the flow cell at a rate of 150 IgG, and Y is the phase difference.
ml/min, followed by the injection of the anti-sheep IgG of T h e sa m e e xp e r i me n t wa s c o n d uc t e d usi n g a
different concentrations stored in the sample loop after 4 commercially available optical biosensor, Biacore X (Biacore
minutes. Finally, glycine-HCl (pH = 2.0) is injected to International SA, Uppsala, Sweden), and the results are shown
disassociate the anti-sheep IgG immobilized on the sensor chip in Figure 6. As can be seen, no change in the resonance unit
to measure the phase shift at various concentrations of can be detected at anti-sheep IgG concentrations of 55ng/ml
anti-sheep IgG. Regeneration of the sensor surface with and 110ng/ml. However, at anti-sheep IgG concentrations of
non-specifically bound anti-body removed is important for 1.1, 11 and 110 µg/ml, there are significant changes in the
maintaining binding reproducibility. resonance unit that can be attributed to molecular interactions.
152 J. Med. Biol. Eng., Vol. 26. No. 4 2006

Comparing the results shown in Figure 5 and Figure 7


reveals that the phase-measurement-based SPR biosensor is
superior in its capacity to detect phase difference at lower
concentrations of anti-sheep IgG, while the results obtained
from Biacore X better fit the linear relationship between
changes in resonance unit and concentrations of IgG.

Conclusion
This study develops a sensitive SPR biosensor with the
common-path heterodyne interferometric system to detect the
molecular interactions between sheep IgG and its antibody.
The proposed biosensor utilizes a pair of acoustooptic
modulators to modulate the TE-wave and TM-wave, thus
forming a heterodyne light source with a frequency difference
Figure 5 Plot depicting the relationship between phase difference
and concentration of the sensorgram data from Fig.4 of 60 kHz. The phase difference between the TE-wave and
TM-wave can be obtained by using the dual-phase lock-in
amplifier to obtain the difference in beating frequency between
the reference signal and the measured signal. The obtained
phase difference of 4, 14, 33, 62 and 107 degrees between the
sheep IgG immobilized on the sensor chip and the anti-sheep
IgG with a corresponding concentration ranges from 0.055 to
110 µg/ml. The proposed SPR biosensor constructed with the
common-path heterodyne interferometric system does not
require fluorescent-labeled analytes or receptors. Not only can
it reduce disturbance from the surrounding environment, it can
also provide high-sensitivity real-time phase detection for
monitoring molecular interactions. Compared with the
commercial optical biosensor, Biacore X, this proposed SPR
biosensor has better performance in detecting the anti-sheep
IgG at concentrations as low as 55ng/ml.
Figure 6 The sensorgram obtained using Biacore X

Acknowledgement
This work was supported by the National Science Council
of ROC under Grant No. NSC-94-2215-E-007-015.

References
[1] R. L. Rich and D. G. Myszka, “Survey of the 1999 surface
plasmon resonance biosensor literature,” J. Mol. Recogn.,
13:388-407, 2000.
[2] J. Homola, S. S. Yee and G. Gauglitz, “Surface plasmon
resonance sensors: review,” Sens. Actuators B, 54:3-15, 1999.
[3] A. Otto, ”Excitation of non-radiative surface plasma wave in
silver by the method of frustrated total reflection”, Z. Phys.,
216:398-410, 1968.
[4] E. Kretschmann, “The determination of optical constants of
metals by excitation of surface plasmons,” Z. Phys., 241:313-24,
Figure 7 Plot depicting the relationship between change in resonance
1971.
unit and concentration of the sensorgram data from Fig.6
[5] B. Liedberg, C. Nylander and I. Lundström, “Surface plasmon
resonance for gas detection and biosensing,” Sens. Actuators B,
4:299-304, 1983.
Figure 7 is the plot of concentration of anti-sheep IgG against [6] B. Liedberg, C. Nylander and I. Lundström, “Biosensing with
changes in resonance units, showing a linear relationship surface plasmon resonance-how it all started”, Biosens.
Bioelectron., 10:i-ix, 1995.
between the two. Their correlation can be expressed as Y =
[7] S. Löfås and B. Johnsson, “A novel hydrogel matrix on gold
1.5699 logX + 1.475 with R2 =0.9949, where X is the surfaces in surface plasmon resonance sensors for fast and
concentration of anti-sheep IgG and Y is the change in the efficient covalent immobilization of ligands,” J. Chem. Soc.
resonance unit. Chem. Commun., 21:1526-8, 1990.
Heterodyne Interferometric SPR Biosensor 153

[8] J. L. Elkind, D. I. Stimpson, Anita A. Strong, D.U. [13] C. Xiao and S. F. Sui, ”Characterization of surface plasmon
Bartholomew and J. L. Melendez, “Integrated analytical sensor: resonance biosensor,” Sens. Actuators B, 66:144-177, 2000.
the use of the TISPR-1 as a biosensor,” Sens. Actuators B, [14] R. Naraka and K. Kajikawa, ”Phase detection of surface
54:182-190, 1999. plasmon resonance using rotating analyzer method”, Sens.
[9] I. Stemmier, A. Brecht, and G. Gauglitz, “Compact surface Actuators B, 107:952-956, 2005.
plasmon resonance-transducers with spectral readout for [15] P. Westphal and A. Bornmann, ”biomolecular detection by
biosensing applications,” Sens. Actuators B, 54:98-105, 1999. surface plasmon enhanced ellipsometry,” Sens. Actuators B,
[10] [10] I. R. Hooper and J. R. Sambles, “Differential ellipsometric 84:278-282, 2002.
surface plasmon resonance sensors with liquid crystal [16] X. Yu, L. Zhao, H. Jiang, H. Wang, C. Yin, and S.
polarization modulators,” Appl. Phys. Lett., 85:3017-3019, Zhu, ”immunosensor based on optical heterodyne phase
2004. detection,” Sens. Actuators B, 76:199-202, 2001.
[11] S. G. Nelson, K. S. Johnston and S. S. Yee, “High sensitivity [17] D. Malacara, Optical shop testing. New York: John Wiley &
surface plasmon resonance sensor based on phase detection,” Son, ch.3: 95-122, 1992.
Sens. Actuators B, 35:187-91, 1996. [18] C. M. Wu, Z. C. Jian, S. F. Joe, and L. B.
[12] A. V. Kabashin and P. I. Nikitin, “Interferometer based on a Chang, ”High-sensitivity sensor based on surface plasmon
surface plasmon resonance for sensor applications,” Quantum resonance and heterodyne interferometry,” Sens. Actuators B,
Electron, 27:653-4, 1997. 92:133-136, 2003.

Vous aimerez peut-être aussi