Vous êtes sur la page 1sur 2

Vascular Intervention

PRO-Kinetic Energy

Technical Specifications

PRO-Kinetic Energy Cobalt Chromium Coronary Stent System


Technical Specification

Device Description
The PRO-Kinetic Energy coronary stent system consists of a coronary stent premounted on a fast-
exchange delivery system. The tubular, balloon-expandable coronary stent is sculpted by laser from a
single tube of L-605 cobalt chromium alloy (CoCr) and is completely coated with PROBIO® amorphous
silicon carbide (a-SiC:H).

Table 1: PRO-Kinetic Energy Stent characteristics


Stent 2.0 2.5 3.0 3.5 4.0 4.5 5.0
Stent material Cobalt Chromium, L-605
Stent design Double helix
Coating PROBIO® Amorphous Silicon Carbide Coating
Strut thickness 60 μm (0.0024”) 80 μm (0.0031”) 120 μm (0.0047”)
Foreshortening 0% -0.7% 0.0%
Recoil 4.7% 4.6% 4.6%
Metal surface area 20% 16% 13% 12% 11% 12% 11%

Table 2: PRO-Kinetic Energy delivery system characteristics


Delivery system
Catheter type Rapid exchange
Balloon material Semi crystalline co-polymer material
Coating (distal shaft) Hydrophilic coating
Marker bands Two swaged platinum-iridium markers
Guide wire diameter 0.014”
Lesion entry profile 0.017”
Crossing Profile Ø 2.00 mm: 0.87 mm (0.034”) Ø 3.50 mm: 1.08 mm (0.043")
Ø 2.25 mm: 0.88 mm (0.035") Ø 4.00 mm: 1.10 mm (0.043")
Ø 2.50 mm: 0.89 mm (0.035") Ø 4.50 mm: 1.21 mm (0.048")
Ø 2.75 mm: 0.93 mm (0.037") Ø 5.00 mm: 1.22 mm (0.048")
Ø 3.00 mm: 0.95 mm (0.037")
Usable catheter length 140 cm
Proximal shaft diameter 2.0 F
Distal shaft diameter 2.5F: Ø 2.0 – 3.5 mm / 2.8F: Ø 4.0 – 5.0 mm
Nominal pressure 9 atm
Rated burst pressure 16 atm: Ø 2.0 – 4.0 mm / 14 atm: Ø 4.5 – 5.0 mm
Table 3: PRO-Kinetic Energy Unique Selling Points
Feature Benefits
For the thinnest strut design, optimal radiopacity and excellent vessel
Cobalt Chromium
support
Helical meanders gives flexibility to the stent for excellent delivery and
Helical stent design
allows for a smooth crimped profile
Three stent designs Optimal scaffolding with thinnest possible struts for excellent results
Acts as a diffusion barrier, eliminating contact between blood/tissue and
PROBIO coating
allergenic metal components for improved haemo- and biocompatibility
Pantera® balloon More flexible materials and new balloon shaping technology gives
technology greater crossability and trackability
Stent and delivery device designed with lower profiles to track and cross
Low profile system
with ease
Double welded
Enhanced Force Unique joint between the hypotube and distal tubing for unsurpassed
Transmission (EFT) device pushability and kink resistance
shaft
Soft tip and minimal
Additional safety during direct stenting procedures
balloon overhang
Multi-step thermal crimping process yields high retention forces for
Thermal crimping
safety

Table 4: PRO-Kinetic Energy Cobalt Chromium Stent System


Ø (mm) Stent lengths (mm)
9 13 15 18 20 22 26 30 35 40
2.00 360490 360497 360506 360515 360524
2.25 360491 360498 360507 360516 360525
2.50 360492 360499 360508 360517 360526 360533
2.75 360493 360500 360509 360518 360527 360534 360538 360544
3.00 360494 360501 360510 360519 360528 360535 360539 360545
3.50 360495 360502 360511 360520 360529 360536 360540 360546 360550 360552
4.00 360496 360503 360512 360521 360530 360537 360541 360547 360551 360553
4.50 360504 360513 360522 360531 360542 360548 360554
5.00 360505 360514 360523 360532 360543 360549 360555