Académique Documents
Professionnel Documents
Culture Documents
The Nutech MST Manual Stack Sampling System is designed to meet all EPA specifications as
outlined in the Office of Air Programs Publication No. APTD-0576. Manufactured with the finest
components, the MST is designed to provide maximum performance under demanding field
conditions. The MST Control Console, and many of the individual components, is extremely
versatile. The MST Control Console is engineered to conduct a wide variety of sampling
applications. These applications include but are not limited to U.S. EPA Methods 4, 5, 8, 10, 13,
23, 26A, 29, 201A, and 301.
Table of Contents
Chapter 1 - Introduction Chapter 4 -Preparation for Sampling
1.1 Principal of Operation 4.0 Pre-Sampling Operations
4.1 Apparatus Check and
Chapter 2 - System Description Calibration
2.1 General Description 4.1.1 Probe Liner and Nozzle
2.2 Probe Assembly 4.1.2 Glassware Preparation
2.3 Filter Box 4.1.3 Dry Gas Meter
2.4 Impinger Box 4.1.4 Filter Preparation
2.5 Umbilical Cable
Chapter 5 - Sampling Operations
2.6 Pump Housing
5.0 On-Site Operations
2.7 Control Console
5.1 Handling of Equipment
Chapter 3 - Maintenance 5.2 Sampling
3.1 Control Console 5.2.1 Site Selection
3.1.1 Casing 5.2.2 Sample Point Location
3.1.2 Manometer 5.2.3 Traversing
3.1.2.1 Pitot Tube Lines 5.2.4 Sample Case Support
3.1.2.2 Orifice Meter Lines 5.2.4.1 Monorail
3.1.3 Timer 5.2.4.2 Slide Box
3.1.4 Temperature Controllers 5.2.5 Vertical Traversing
3.1.5 Temperature Display
3.1.6 Quick Connects 5.3 Pre-Test Calculations
3.1.7 Dry Gas Meter 5.3.1 Molecular Weight
3.1.8 Inlet Filter 5.3.2 Static Pressure
3.1.9 Control Box Leak Check 5.3.3 Absolute Stack Pressure
3.2 Pump Housing 5.3.4 Average Delta P
3.2.1 Muffler Jar 5.3.5 Average Gas Velocity
3.2.2 Oiler Jar 5.3.6 Absolute Meter Pressure
3.2.3 Pump Cleaning 5.3.7 Moisture Content
3.2.4 Fiber Vane Replacement 5.3.8 Stack Gas Molecular
3.2.5 Setting Rotor Tolerance Weight
3.2.6 Thermal Switch 5.3.9 Nozzle Size
3.3 Umbilical Cable 5.3.10 K-Factor
3.3.1 Amphenol Connectors 5.4 Sample Train Preparation
3.3.2 Quick Connects 5.5 Modular Sample Case
3.3.3 Thermocouple Plugs 5.6 Sample Probe Assembly
3.3.4 Sample Line Flushing 5.6.1 Nozzle Attachment
3.4 Sample Probes 5.6.2 Probe Mounting
3.4.1 Heating Mantle 5.6.3 Visual Inspection
3.4.2 Thermocouples 5.7 Sample Run Calculations
3.4.3 Pitot Tip
3.4.4 Nozzle Calibration 5.8 Post Test Calculations
5.8.1 Total volume of Dry Gas
5.8.2 Volume of Water Vapor at
Standard Conditions
Chapter 1
Introduction
1.1 Principal of Operation
The MST is designed to sample gas stream effluents isokinetically in
accordance with the U.S. Environmental Protection Agency (EPA)
standards as outlined in the Code of Federal Regulations, Title 40, Part 60
(40CFR60), Appendix A. Stack gases typically are extracted through a
nozzle and heated probe, into a heated filter chamber where particulates are
removed. The hot gases are then passed through a series of cold impingers
where condensable are removed and the gasses are cooled before going to
the pump, dry gas meter and orifice.
Maintenance
3.1 Control Console
Periodic inspection of the control console components is a must to insure
accurate sampling. This is especially true after the control console has
undergone rough treatment during field use or shipment. Although seldom
necessary, the casing can be removed by unscrewing the base securing
screws and the eight front panel securing screws on the sides and top. The
bottom screws on the front panels should remain in place.
3.1.1 Casing
Inspect all screws and rivets used in the assembly of the control console.
Any loose rivets should be replaced and loose screws tightened.
Replacement screws are available from Graseby. The thread size for the
screws attaching the latches is 6-32. All other screws are in the outer case
are 8-32.
3.1.2 Manometer
Visually inspect the orifice and pitot lines.
They should be free of fluid. Any fluid found
should be drained. The manometer should be
cleaned with soap and water or kerosene
Filler
only. Check for leaks around the oil level Screw
adjust knobs and filler screws. Any leaks
around the oil level adjust knobs can be fixed
by replacing the entire adjust assembly. Oil Level
Adjust Knobs
Replacement filler screws are also available.
Replacement manometer oil must have a Drain
Screw
specific gravity of 0.826. Adding additional
oil to the manometer is accomplished by Figure 3-1 Manometer
removing the oil filler screw at the left end
of the horizontal scale. A long blade type screwdriver inserted through the
hole in the casing eliminates the need to remove the manometer from the
front panel for this operation.
• Note: During shipment the manometer lines should be
disconnected. If for any reason the control unit has been inverted, be sure
the floating check valves of the manometer have returned to their normal
position. These valves are located under the manometer ports and must be
in the normal position to facilitate the use of the manometer.
• Note: The null switch on the electrical panel allows the user to
check the zero of the orifice lines during testing.
3.1.2.1 Pitot Tube Lines
To be certain the pitot lines are free of obstructions, gently blow through
the pitot tube quick connects. The pitot tube gauge should respond. Be
careful not to put more than ten inches (254mm) of water pressure on the
system.
3.1.2.2 Orifice Meter Lines
The following procedure should be used to check that the orifice lines are
clean of obstructions. Turn off all switches except the POWER and PUMP
switches. Carefully open the coarse and fine valves. The manometer
should respond with movement. If no movement is observed check to be
sure that all lines are connected and valves open. Also, move the Null
switch to check to see that the solenoid valves are operating properly. You
should hear a click when the Null switch is energized and the manometer
will move to zero.
3.1.3 Timer
The digital timer on the MST control
displays elapsed time in minutes and
seconds. The red button on the timer is
the reset button. There are no user
serviceable parts in this temperature Figure 3-2 Digital
controller. If power is lost during a Timer
sample run, the internal battery will
display the time at which power was
disconnected.
3.1.4 Temperature Controller
The dial type temperature controllers are °F
ON
300
set by turning the dial to the desired 150 450
temperature setting. When the unit is
functioning properly, the light in the upper 0 600
left corner will remain on only while
TOS Autonics
power is being output to the attached
heater. There are no user serviceable parts
in this temperature controller . Figure 3-3 Temperature
Controller
3.1.5 Temperature Display
The digital temperature display should Always on Top 2 pins = °F
two left pins Bottom 2 pins = °C
be periodically calibrated as described
in the steps below. To change the
digital temperature display from °F to
°C remove the red plastic cover.
Switching the two lower and the far
right jumpers will convert the unit to
display Celsius. Two left pins =°F High Low
Two right pins =°C Calibration Calibration
NOTE: Figure 3-4 illustrates the unit
in its configuration for °F. Figure 3-4 Temperature
Display
The following steps should be followed
to calibrate your digital temperature display:
1. Warm up the display for a period of 1 hour
2. Remove the plastic red cover from the display
3. Attach a Standard Thermocouple Calibrator to any of the input jacks
on the front panel of the MST control console and adjust the selector
switch to that channel
4. Set the calibrator to each temperature setting as requested on data
sheet DS-1 and record the set point and display temperatures.
5. Complete the calibration before any adjustments to the display are
made
6. Using the calibration screws as shown in Figure 3-4, adjust the hi and
low settings as needed
7. Repeat steps three and four until the display and calibrator agree within
± 5.4°F (3°C)
3.1.6 Quick Connects
A drop of penetrating oil helps keep the quick connects in good working
order.
3.1.7 Dry Gas Meter
The meter and orfice are calibrated before leaving the factory. A record of
the initial calibration is kept at Andersen Instruments Inc.for a period of 1
year. When required recalibrate the dry gas meter and orifice according to
the procedure on the following page.
1. Connect the control unit to a
calibration device (Standard dry
gas meter, wet test meter,
calibrated orifice, etc.)
2. Run the control unit for a period
of 15 minutes to allow all
components to warm up. If using
a wet test meter, it is very
important that the water be at
Figure 3-5 Calibration Setup
room temperature to avoid
errors. It is also necessary that
room temperature not fluctuate.
3. Adjust the flow rate to the desired flow. Before leaving the factory,
Andersen Instruments Inc.calibrates each control unit at the following
manometer settings: 0.75, 2.0 and 6.0 at a vacuum of approximately 5”
Hg. For a post test calibration, the flow rate should be set to the
average ∆H achieved during the sample runs. If using calibrated
orifices, open the main and coarse valves completely to maximum
vaccum
4. Do not adjust the flow rate, turn the pump power off and record the
dry gas meter volume and calibration device volume.
5. Reset the timer.
6. Turn the pump and timer switches on simultaneously.
7. Record all data as requested on Data Sheet DS-2
8. A minimum of 5 cubic feet (0.15 m3) are required at each flow rate.
9. Calculate the γ and ∆H@ for each flow setting.
10. A calibration is acceptable if all of the individual values are ± 2% of the
average. Contact the Andersen Instruments Inc.Customer Service
department if the meter is not functioning properly. If the dry gas
meter coefficient values obtained before and after a test series differ by
more than 5%, recalibrate the meter over the full range of settings.
• NOTE: Any tests performed using a meter failing the 5% criteria
should be voided or calculations performed using the γ (before or after)
that gives the lower value of total sample volume.
3.1.8 Control Console Inlet Filter
Located just behind the sample line quick connect, sample gas passes
through the filter jar immediately as it enters the Control Console to
remove any foreign matter before it enters the pump housing. The felts in
this jar should be replaced regularly. To replace these felts, remove the
glass catch jar and remove the plastic retaining washer from below the
bottom felt. Remove and replace all three felts. Before reassembling the jar
inspect the gasket located inside the metal cap for and drying or cracks.
Replace if necessary. A few wraps of Teflon tape around the glass threads
will help protect against leaks.
3.1.9 Control Console Leak Check
Before any calibration or sampling procedure, a leak check should be
conducted. The meter system should be leak free. Both positive
(pressure) and negative (vacuum) leak checks should be performed. The
following is a pressure leak check procedure that will check the metering
system from the sample inlet quick connect to the orifice outlet including
the inclined manometer.
1) Disconnect the orifice meter
line from the downstream
orifice pressure tap (the one Vacuum Gauge
closest to the orifice exhaust), Orifice Tube Fine Control Coarse Control
and plug this tap . Valve Valve
Pre-Sampling Operations
4.0 Pre-Sampling Operations
Strict adherence to the specifications listed in this section are necessary to
maintain sample integrity. Each of these items are generally completed
before traveling to the sampling site.
4.1 Apparatus Check and Calibration
A pretest check will have to be made on most of the sampling apparatus. A
copy of all calibration data should be kept with the equipment as well as
filed in a safe location.
4.1.1 Probe Liner and Nozzle
Clean the probe and nozzle interior by brushing first with tap water, then
with Deionized distilled water, and finally with acetone; allow both to dry
in a clean environment. Some sampling methods will require different
reagents be used. Regardless of the reagent used, the objective is to leave
the nozzle and liner free from contaminants. After drying, the probe liner
should be sealed against further contamination. The probe heating system
should be checked to see that it is operating properly.
4.1.2 Impingers, Filter Holder and Glass Connectors
All glassware should be cleaned first with detergent and tap water and then
with Deionized distilled water. After drying, all glassware should be sealed
against further contamination. A visual inspection of the glassware should
be made for cracks or breakage and then repaired or discarded if defective.
Note: Some sampling methods require the use of different reagents for the
cleaning process.
4.1.3 Dry Gas Meter
The dry gas meter should be calibrated as described in Section 3.1.7 before
each sampling procedure.
4.1.4 Filter Preparation
Check the filters visually against light for irregularities, flaws, and pinhole
leaks. Either label the filters on the backside near the edge using a felt tip
marker, or label the petri dishes and keep the filters in their respective
dishes except during actual sampling and weighing. Before weighing, the
filters must be desiccated at 68° ±10°F (20° ±5.6°C) and at ambient
pressure for at least twenty four hours. Weigh each filter at six hour
intervals until weight changes of less than 0.5mg from the previous
weightings are achieved. During each weighing, the filter must not be
exposed to the laboratory atmosphere for greater than two minutes or to a
relative humidity of greater than 50%. An alternative procedure to 24 hour
desiccation is oven drying the filters at 220°F(105°C) for two to three
hours followed by desiccation for two hours and by weighing to a constant
weight, as described above. The balance should be calibrated according to
the manufacture’s specifications before each set of weights are taken. If
the balance does not agree to within ±0.5g maintenance procedures should
be performed to correct any errors. A record of calibration should be kept
with the balance at all times.
Chapter 5
On-Site Operations
5.0 On-Site Operation
The on-site activities include transporting equipment to the test site,
unpacking and assembling the equipment, making duct measurements,
performing the velocity traverse, determining molecular weights and stack
gas moisture contents, sampling, recording data and calculating results.
Blank data forms are located in Appendix A for the convenience of the
Nutech MST user.
5.1 Handling of Equipment
The most efficient means of transporting or moving the equipment from
ground level to the sampling site should be decided during the preliminary
site visit (or prior correspondence). Care should be exercised to prevent
damage to the test equipment or injury to personnel during the moving
phase. A "laboratory" area should be designated for assembling the
sampling train, placing the filter in the filter holder, charging the impingers,
recovering the sample, and documenting the results. This area should be
clean and free of excessive drafts.
5.2 Sampling
The on-site sampling includes site selection, preliminary measurements and
setup, placing the filter in the filter holder, setting up the sampling train,
preparing the probe, checking for leaks along the entire train, inserting the
probe into the stack, sealing the port, checking the temperature of the
probe, sampling at designated points, and recording the data. A final leak
check must always be performed upon completion of the sampling.
5.2.1 Site Selection
Many times you will find your sample site already chosen, often based on
inadequate information. If your location does not fit the following criteria,
it would be best to relocate the sampling site in order to make the data
more acceptable. Other factors to be considered in selecting the sampling
site are that the distance from the probe to the bottom of the sample case is
about 13 inches, and the minimum recommended sample port size is three
inches.
A sampling site should be chosen where there are at least two diameters
downstream and 0.5 diameters upstream from any expansion, contraction,
bend, obstruction, visible flame, or exit. This location, however, requires
sampling at 24 different points within the duct as shown in Figure 5-1. The
optimum location is at least eight diameters downstream and two upstream,
requiring only 12 sampling points. For rectangular cross sections,
determine an equivalent diameter from the following equation:
(length) (width)
equivalent diameter = 2
length + width
To use Figure 5-1, first measure the distance from the chosen sampling
location to the nearest upstream and downstream disturbances. Determine
the corresponding number of sampling points for each distance from Figure
5-1, and select the higher of the two numbers. Then round off the number
for a round stack to the nearest multiple of four. For a square or
rectangular stack, divide the cross section into as many equal rectangular
areas as sample points, such that the ratio of the length to width of each
elemental area is between one and two. Data Sheet DS-4, located in
Appendix A is provided to aid in determining sample points.
B
30 24 or 25a
Disturbance
20
20 16
0
2 3 4 5 6 7 8 9 10
Duct Diameters Downstream from Flow Disturbance* (Distance B)
Traverse % of Location of 6
Point Diameter Sample 5
Point
1 4.4 2.64 4
2 14.6 8.76
3 29.6 17.76 3
4 70.4 42.24
2
5 85.4 51.24
6 95.6 57.36 1
240”
Note: When marking the sample points on your probe, do not forget to
add the port nipple length to the distance for each point
5.2.3 Traversing
Sampling should always be done traversing perpendicular to the flow of
stack gasses. A vertical stack or duct requires a horizontal traverse. A
horizontal stack may be traversed horizontally or vertically. For stacks or
ducts at offset angles, horizontal traversing is recommended.
The sampling site must not only meet the requirements of the least
disturbance to flow, but also allow accessibility of the sampling equipment.
There should be at least 13” below the centerline of the port, 14” above the
centerline and 25” should be added to the probe to allow for full traversing.
5.2.4 Sample Case Support
Support of the sample case is a primary consideration in any stack sampling
job. The need for a traversing platform equal in length to the stack
diameter has brought about the development of several sample case
support systems. The two most common systems are the monorail support
and the slide box.
5.2.4.1 Monorail
The Modular Sample Case is designed in such a way that a monorail
support system can be used as in Figure 5-5. The advantages are light
weight and flexibility. The handle slips easily onto the monorail roller
swivel bracket. By rotating the sample case, the probe can be attached by
the operator while standing on a platform very close to the stack. The
sample case is then pushed out to its farthest position by holding the probe
which is then inserted in the sample port. The distance between the probe
centerline and the bottom of the filter box is approximately thirteen inches.
At times, an obstruction may not allow you to place the monorail at the
desired thirteen inches. If this occurs, greater distances can be achieved by
adding chain between the roller and Modular Sample Case.
Stack
2 x 4 for support
Support
Plywood
Blocks
Rooftop
Flow
∆p * ( 460 + Ts )
= 85.49 * Cp *
Ps * Ms
0.0358 * Qm * Pm ( Ts + 460 ) * Ms
Dn = *
( Tm + 460 ) * Cp * (1 - Bws) Ps * ∆p
Comments:
Appendix A
Meter Box Calibration Sheet
DS-2
Control Box Number Barometric Pressure Pb Calibration Device #
Meter Serial Number Ambient Temperature Calibration Device γ
Temperature Measurements
Calibration Dry Gas Meter
Orifice Gas Volume Gas Volume Average
Manometer of Calibration of Dry Gas Meter Meter Inlet Meter Outlet Meter Time Calculated
Setting (∆H) Meter, Vw Meter, Vd Tw Tdi Tdo Td (Minutes), θ Calculated γ ∆H
Average
γ ∆H@
2
Vw Pb (Td + 460)
0.0317 ∆H (Tw + 460) θ
Vd (
Pb +
∆H
13.6 ) ( Tw + 460) Pb (Td + 460) [ Vw ]
Appendix A
Determination of Sampling Points Worksheet
DS-4
Plant Name Job Number
City/State Location
Operator
Schematic of Location
Distance from Far Wall
to Outside of Port
Port Length
Stack Diameter (Ds)or Length
Stack Width if Rectangular
Distance Upstream (A)
Distance Downstream (B)
2LW
Equivalent Diameter = = =
(L + W)
Diameters Upstream =
Minimum # of Sampling Points for All Stacks with Diameters Greater than 24 Inches is 12
Minimum # of Sampling Points for Circular Stacks with Diameters Between 12 and 24 Inches is 8
Minimum # of Sampling Points for Rectangular Stacks with Diameters Between 12 and 24 Inches is 9
Appendix A
Pre Sampling Calculations
DS-5
0.0358 * Qm * Pm ( Ts + 460 ) * Ms
Ideal Nozzle Diameter = *
( Tm + 460 ) * Cp * (1 - Bws) Ps * ∆p
∆p * ( 460 + Ts )
Stack Gas Velocity ( vs ) = 85.49 * Cp *
Ps * Ms
Average α
Run Number
Date Prepared
Date Analyzed
Analyst
Impinger 1 ( )
Final Weight, g
Initial Weight, g
Water Catch, g
Impinger 2 ( )
Final Weight, g
Initial Weight, g
Water Catch, g
Impinger 3 ( )
Final Weight, g
Initial Weight, g
Water Catch, g
Impinger 4 ( )
Final Weight, g
Initial Weight, g
Water Catch, g
Comments:
Appendix A
Moisture Determination Worksheet
DS-8
Volume of Dry Gas Sampled at Standard Conditions, Dry Standard Cubic
Feet
Pbar + ( ∆H / 13.6 )
Vm(std) = 17.64 * Y * Vm * 460 + Tm
+( / 13.6 )
Vm(std) = 17.64 * * *
460 +
+
Vm(std) = *
Vm(std) = *
Vm(std) =
%H20 = 100 * / ( + )
%H20 = /
%H20 =
Appendix A
Field Data Sheet
DS-9
Comments:
Appendix A
Orsat Field Data
DS-10
Plant Name Date
Job Number Sampling Location
Operator Fuel Type
Average Average
Average Average
Average Average