SPECIFICATION FOR TRANSMISSION ELECTRON MICROSCOPE

1. System control PC based software control system; all the components like electron gun, the optical elements, vacuum system and the stage should be completely controlled by computer operation. It should be able to perform Dual-axis Tomography, Electron Tomography, Single Particle Analysis, Viral Load Monitoring, Cellular Tomography, Macromolecular Localization, Host-pathogen Interactions, Pathology and Medical Diagnostics, Cell-to-Cell Interfacing, Particle Analysis. 0.24 nm or better 0.14 nm or better 0.14 nm or better 100 kV and above, HT stability 1.0 ppm Tungsten/Lanthanum hexaboride 25 x -750 Kx or better High Magnification, low magnification, diffraction, nano probe, dark field, low dose mode, multistage condenser lens system, variable aperture, fully electromagnetic user alignment system Two stage objective lenses along with standard anti contamination device; Focal length: 2.7 mm Minimum focus step (nm): 3.0, Cs objective: 2.0 mm, Cc objective: 2.0 mm, SA diffraction camera length: 50 7000 mm, Maximum diffraction angle: 10, Maximum eucentric tilt: 70or better. Should be computer controlled Eucentric, side entry Goniometer, at least 2 axis motorized (X and Y) fine control for Z movement, motor driven and software controlled tilt capability, maximum Holder Tilt Angle should be at least 70o or better, Goniometer stage should accept variety of holders with heating, cooling and low background double tilt holders, cold trap in specimen area. X,Y = 2 mm and Z = 0.4 mm or better (Fully computer controlled with specimen recall facility), specimen recall reproducibility should be 0.3 um(XY) and 0.1 degree tilt. Fast exchange single and double tilt specimen holder with 7.4 mm or better pole piece gap. Slow scan retractable high resolution embedded CCD camera (2048x2048 pixal or better), software with external computer, interface accessed with PC; plate camera for direct photography. Software for capturing, processing and editing the image in attached PC and output in JPG or tiff format. Separate ion getter pumps (IGP) for specimen area and liner tubes, liquid-nitrogen cooled cold trap in specimen area, the accelerator and gun section of the column should separately be evacuated by IGP, fully automatic sequential control, automatic controlled Air Lock, fully automatic microprocessor controlled with manual over

2.

Applications

3. 4. 5. 6. 7. 8. 9.

Point resolution Lattice resolution Line resolution HT control Filament type Magnification Illumination system

10.

Objective lens

11.

Specimen chamber

12.

Specimen movement

13. 14.

Specimen Holders Imaging system

15. 16.

Software compatibility Vacuum system

17. 18.

Cooling System Computer environment

19. 20. 21. 22. 23. 24.

Electrical requirement Power backup Training Warranty Annual Maintenance Contract (AMC) Buy-back offer

riding options to attain 10-6 - 10-7 torr or better at gun area. Efficient automatic temperature controlled water cooling system. Compatible computer with 3.2 GHz Intel Pentium IV processor or higher, at least 3 GB DDR RAM, Microsoft Windows 7 operating system, Minimum 320 GB HD, DVD+/-R/RW, high brightness 19 flat LCD screen monitor, laser color printer. 220 V, 50 Hz. Suitable UPS having one hour backup to the complete system Should provide onsite training as well as training at factory for two persons At least 03 years from the date of installation on site Rate for AMC after expiry of warranty period Offer for buy-back of the existing JEOL 100S-TEM

FEI
1. System control PC based software control system; all the components like electron gun, the optical elements, vacuum system and the

2.

Applications

3. 4. 5. 6. 7. 8. 9.

Point resolution Lattice resolution Line resolution HT control Filament type Magnification Illumination system

10.

Objective lens

11.

Specimen chamber

12.

Specimen movement

13. 14.

Specimen Holders Imaging system

15. 16.

Software compatibility Vacuum system

17. 18.

Cooling System Computer environment

stage should be completely controlled by computer operation. It should be able to perform Dual-axis Tomography, Electron Tomography, Single Particle Analysis, Viral Load Monitoring, Cellular Tomography, Macromolecular Localization, Host-pathogen Interactions, Pathology and Medical Diagnostics, Cell-to-Cell Interfacing, Particle Analysis. 0.24 nm or better (.20nm) 0.14 nm or better 0.14 nm or better (.2nm) 100 kV and above, HT stability 1(2).0 ppm/? Tungsten/Lanthanum hexaboride 25 x -750(650) Kx or better (10 laks) 650Kx (FEI) High Magnification, low magnification, diffraction, nano probe, dark field, low dose mode, multistage condenser lens system, variable aperture, fully electromagnetic user alignment system Two stage objective lenses along with standard anti contamination device; Focal length: 2.7 mm (2.8mm FEI) Minimum focus step (nm): 3.0, Cs objective: 2.0 (2.2 Fei) mm, Cc objective: 2.0 (2.2 Fei) mm, SA diffraction camera length: 50 7000 mm, Maximum diffraction angle: 10, Maximum eucentric tilt: 70or better. Should be computer controlled Eucentric, side entry Goniometer, at least 2 axis motorized (X and Y) fine control for Z movement, motor driven and software controlled tilt capability, maximum Holder Tilt Angle should be at least 70o or better, Goniometer stage should accept variety of holders with heating, cooling and low background double tilt holders, cold trap in specimen area. X,Y = 2 mm and Z = 0.4 mm or better (Fully computer controlled with specimen recall facility), specimen recall reproducibility should be 0.3 um(XY) and 0.1 degree tilt. Fast exchange single and double tilt specimen holder with 7.4 mm or better pole piece gap. Slow scan retractable high resolution embedded CCD camera (2048x2048 pixal or better), software with external computer, interface accessed with PC; plate camera for direct photography. Software for capturing, processing and editing the image in attached PC and output in JPG or tiff format. Separate ion getter pumps (IGP) for specimen area and liner tubes, liquid-nitrogen cooled cold trap in specimen area, the accelerator and gun section of the column should separately be evacuated by IGP, fully automatic sequential control, automatic controlled Air Lock, fully automatic microprocessor controlled with manual over riding options to attain 10-6 - 10-7 torr (Pascal?) or better at gun area. Efficient automatic temperature controlled water cooling system. Compatible computer with 3.2 GHz Intel Pentium IV processor or higher, at least 3 GB DDR RAM, Microsoft

19. 20. 21. 22. 23. 24.

Electrical requirement Power backup Training Warranty Annual Maintenance Contract (AMC) Buy-back offer

Windows 7 operating system, Minimum 320 GB HD, DVD+/-R/RW, high brightness 19 flat LCD screen monitor, laser color printer. 220 V, 50 Hz. Suitable UPS having one hour backup to the complete system Should provide onsite training as well as training at factory for two persons At least 03 years from the date of installation on site Rate for AMC after expiry of warranty period Offer for buy-back of the existing JEOL 100S-TEM