Vous êtes sur la page 1sur 8

Milos Novotny Stephen R.

Springston
Chemistry Department Indiana University Bloomington, Ind. 47405

Paul A. Peaden John C. Fjeldsted Milton L. Lee


Chemistry Department Brigham Young University Provo, Utah 84602

Capillary Supercritical Fluid Chromatography


Rapid separations of respectable efficiency for nonvolatile solutes can be achieved by use of capillary columns in conj u n c t ion w i t h super cr i ti ca 1 f 1u i d s

Capillary columns were first described for gas chromatography (GC) in 1957 by M.J.E. Golay, who utilized long, open tubular columns uniformly coated with a thin layer of stationary phase ( I ) . Although many technological improvements were necessary in the years following Golays invention, the method has recently revolutionized many areas of scientific research. This is primarily due to its high resolving power, allowing the separation, detection, and quantitation of hundreds of components from a sample in a single chromatogram. The practical separating capability of capillary gas chromatography is presently unparalleled. However, the method is somewhat restricted by the limited volatility and thermal stability of many organic compounds. Mixtures of less volatile compounds can be analyzed by high performance liquid chromatography (HPLC),but large numbers of theoretical plates can only be obtained a t the expense of a long analysis time. Achievement of the maximum number of theoretical plates in capillary HPLC is presently complicated by the extremely small column diameters required for low plate heights. This requirement is a consequence of solute diffusivities and mobile phase viscosities. These are the two most important properties governing chromatographic efficiency (2, 3 ) and are primarily a function of the mobile phase and its physical properties. While encouraging results were recently obtained ( 4 ) with open microtubular columns as small as 30 pm id., theoretical analysis by Knox and
0003-270018 11035 1-407A$01.OO/O @ 1981 American Chemical Society

Gilbert ( 5 ) indicates the need to work with columns of about 10 pm and detectors with volumes of a few nL to maximize efficiency, a formidable challenge at present. Clearly, other alternatives are needed for highly efficient separations of large, less volatile molecules. A system is needed that provides somewhat faster solute diffusion than in liquids (permitting larger, more reasonable column diameters) and mobile phase viscosities sufficiently low so as to permit long columns to be exploited. Simultaneously, such a system should be suitable for the separation of large and thermally labile molecules. Qualitatively, these conditions are met by use of open capillary columns in conjunction with a supercritical fluid as the mobile phase. It is reasoned in this article and supported by preliminary data that, under certain experimental conditions, this combination can provide a nearly ideal chromatographic system for the above-mentioned molecular species. The pioneering work on supercritical fluid chromatography in packed columns by Klesper et al. (6) clearly demonstrated the potential of this approach in separating thermally labile substances. Later work of Giddings et al. (7-9) and Sie and Rijnders (10-13) with both dense gases and supercritical fluids extended this interesting direction in separation science both theoretically and experimentally. However, these developments of the late 1960s were largely overshadowed by the advent of HPLC. Furthermore, technical difficulties with supercritical

fluids discouraged additional research. The most recent advances have been reviewed by Klesper (14). A very serious limitation of supercritical fluid chromatography with packed columns is the pressure gradient generated by the column packing, as discussed by Novotny et al. ( 1 5 ) . Ironically, while a decrease in particle size leads to much greater efficiency in liquid chromatography (where the mobile phase is practically noncompressible), the reverse is observed with supercritical fluids ( 1 5 ) .This point was reemphasized by Gouw and Jentoft (I6), compared the effects of who an increasing pressure gradient along the column in supercritical fluid chromatography to a decreasing temperature program in gas chromatography. Both have disastrous consequences as far as the column efficiency is concerned. The advantages of capillary column openness are well known in GC and will become even more evident with supercritical fluid chromatography, if the whole system (from the point of sampling to the point of detection) can be maintained under proper conditions of high pressure and low pressure drop. Supercritical fluid chromatography should provide superior migration of labile and less volatile substances through a capillary column when compared to GC. Fluids compressed to their critical points and above exhibit an extraction effect on chromatographed samples similar to solvation in HPLC. The effect of extraction may be predicted based on the second virial coefficient of the system (10) or
407A

ANALYTICAL CHEMISTRY, VOL. 53, NO. 3, MARCH 1981

the Hildehrand solubility theory (9), although such predictions are first approximations a t best. The point of practical interest is that while ordinary gases can he compressed to produce these effects (9), similar conditions can he met with some other fluids at pressures and temperatures well within the limits of ordinary HPLC technology. Some examples are shown in Table I. Attention will he focused on fluids with low critical temperatures, since they will he useful for separating thermally labile compounds. Optimum conditions for capillary supercritical fluid chromatography will now he briefly considered. If we introduce the initial restriction of a maximum allowable pressure drop, P, between the inlet and outlet of a , . , capillary column of radius r , the maximum column length, L. , is given by the integrated Poiseuille equation:

stituting equations (l), and (4) (3), into equation (2) yields the maximum number of plates that can he generated from a column of fixed length:

where 7 is the mobile phase viscosity and F is the average linear velocity of the fluid. The number of theoretical plates, N, for such a column is:

Since both L and H (plate height) are functiuns o f i , the maximum efficiency, Nma,,can easily he found (3). However, the maximum number of plates applies here to conditions of infinite length and zero flow. instead, If, the column is operated under conditions of optimum efficiency per unit length, more realistic efficiencies may he calculated. The minimum plate height, H,,,, can he derived from the Colay equation, assuming that resistance to mass transfer in the stationary phase is negligible:

Hm,"= 0.577 r
This cundition occurs at:

(3)

where D , is the diffusion coefficient of a solute in the mobile phase. Sub408 A

This equation is similar to that derived earlier hy Giddings (3) for the limiting number of plates used to compare the relative merits of gas and liquid chromatography. The difference is that equation (5) assumes ii = popt, whereas Giddings's expression holds only for L = m and i = 0. For a i given pressure drop, operation at ii = poptrather than L = m and = 0 decreases the maximum efficiency by a factor of two. While the above considerations are only valid for a nonretained chromatographic peak (capacity ratio, k = O), some interesting observations can he made. Estimated efficiencies per unit time are quite impressive since D , and q are more favorable to a rapid chromatographic separation in a supercritical fluid than in a liquid mobile phase. Whereas literature data on both these physical quantities for supercritical fluids are rare, only approximate calculations can he made a t this time. The diffusion coefficient of n-pentane in carbon dioxide a t 48 atm and 40 "C was estimated to he 1.5 X cmz s-l (17), while a viscosity value of 3.59 X g cm-1 s--1 at 90 atm and 40 "C is available from another literature source (18).Actual diffusion coefficients may he as much as one or two orders of magnitude smaller for large solutes a t supercritical pressures, hut no one seems to have measured such quantities as yet. If we arbitrarily choose a pressure gradient restriction of 1atm, a maximum of 8.1 X lo6 theoretical plates could be generated in 18 h for a nonretaiced component with a 0.24 mm i.d. capillary column (length, 570 m). Through reduction of the column diameter to 100 pm, 1.4 X 106 theoretical plates could be obtained in just 0.54 h (length, 41 m). Practical realization of optimized capillary supercritical fluid chromatography is dependent on at least three crucial criteria:

The system must he designed so the entire column and the detector both operate at the same high pressure: instrumental contributions to hand broadening must he minimal, and * stationary phase films of optimum thickness must he permanently bonded to the capillary wall so the resistance to mass transfer in the stationary phase is small, and so the phase cannot he stripped off by the passing fluid. Development of capillary supercritical fluid chromatography has recently become feasible. Designs of microcolnmn HPLC equipment (19-22) have made certain instrumental components more readily adaptable to capillary supercritical fluid chromatography. Among them, a low-volume loop injector and a miniaturized spectroscopic cell are particularly useful. A number of fluids are available, such as n-pentane, isopropanol, or carbon dioxide, that allow the migration of relatively heavy molecules under reasonable pressures and temperatures. These mobile phases, combined with the small-volume requirements of capillary techniques, make safety hazards almost negligible. Glass capillary columns with wall thicknesses similar to those used in conventional capillary gas chromatography can be easily used. Advantage can also he taken of recently developed bonded phases
(22).

The capillary supercritical fluid chromatograph operating in our lahoratory possesses the basic features illustrated in Figure l. A high-pressure syringe pump operating a t rmm temperature delivers, a t constant pressure, a liquid (n-pentane) or compressed gas (carbon dioxide) that is converted into a supercritical fluid in a preheating coil. The fluid suhsequently sweeps through a loop valve where the sample is introduced directly into the glass capillary column. The column is heated to temperatures consistent with the desired fluid conditions in a thermostat. The thermostat also contains a flow-through cell (either a UV or spectrofluorimetric detector) used to monitor chromatographic effluents under high pressure. A length of 50-pm glass capillary after the detection cell is used to maintain system pressure. Solutes can also he detected directly in the last section of the column. As this comprises the optical compartment, no hand spreading occurs due to interconnection deadvolume. Increasing the pressure is the most convenient means to influence solute retention (15-27): therefore, pressure programming will ultimately he used for resolution of complex mixtures as shown in Figure 1.

ANALYTICAL CHEMISTRY, VOL. 53. NO. 3. MARCH 1981

AZcXeC
HIGH VACUUM TECHNOLOGY

Single source convenience and responsibility for all your high vacuum needs. Heres how!
Choose from Alcatels complete line of high vacuum technolog equipment with proven r e l i a b i i t y and backed by w o r l d w i d e sales and service. H e l i u m leak detectors Ultra-sensitive performance, combined with exclusive single point tuning for extremely simple operation. Three models to choose from: ASM-10 portable leak test station; ASM-51 with manual throttling;

Direct drive W mechanical pumps The international standard Tor reliability and quality with over 20 models to choose from. Standard Series (3.2 to 150 cfml for general needs; Hermetic Seried(7.0to 40 cfm) for precious gases; Chemical Series (3.2 to 90 cfm) for corrosive gases. Send for Bulletin 10080

r
I

pushbutton selection of H z , He3, He4 for nuclear and crvos!enic a~~lications. Send f i r Bulleiin 102-80

13 -

Turbomolecular Pumps Exclusive desig featuring air bearines and

120 and 450 lisec. models provide silent, vibration free operation. Send for Bulletin 10480 Complete line of standard vacuum accessories Full range of standard traps, fittings, couplings, etc. simplify installation and assure optimum system performance. Send for Bulletin 10580

Crystal d i f f u s i o n pumps and pumping systems The new generation offering outstanding performance with ultra-low

--

Alcatel offers application assistance and maintenance from service locations throughout the United States and the world. Write or call for our free descriptive brochures, price, and delivery information to Alcatel Vacuum Products, 60 Sharp Street, Hingham, MA 02043. Telephone: 617-337-7910.

able as a

Available as a component or as a packaged pumping system. Simple design and rugged construction provide long,

See these and other Alcatel high vacuum products at Booth 1026 & 1028, Pittsburgh Conference; and Booth 407, SSE 81.
CIRCLE 5 ON READER SERVICE CARD

-I

Canada:

755 de Miniac, St; Laurent, Montr


2 Bcrrvmm S. Toronto. Ontwia I. t Europe: IT. Alcatel High Vacuum Division 3, Rue Emeriau - 75125 Paris Cede Iwldwide sales and Y

I I

6-

I
Flgure 1. Block diagram showing the components of a capillary supercritical fluid

clean-up technique is simpler


Rheodyne'sTechnical Note 2 tells how to prepare samples with greater speedand lower cost than purification steps using liquid/liquid partition, column chromatography or commercial separation cartridges. The rapid clean-up technique uses Rheodyne's Model 7125 Syringe Loading Sample Injector with a short column in the sample loop. It allows complexsamplesto beinjected directly into the chromatographic system. This technique can be used to: * Eliminate interferences from complex matrices Concentrate samples to improve detection limits * Shorten analysis time by trapping late-eluting peaks
0

chromatograph We wish to report some preliminary results on capillary supercritical fluid chromatography, using n-pentane as the mobile phase a t 210 "C and 32 atm. The glass capillary columns used were fabricated with phenylmethyl polysiloxane film bonded to the capillary wall similar to the procedure of Blomberg and W h n m a n (22).As a compromise with currently available sampling conditions, 0.2 and 0.3 mm i.d. glass capillary columns were used. The plate height vs. linear velocity curves were measured for the fluorimetrically detected pyrene peak (A, = 335 nm; ,A , = 425 nm, 10-nm bandwidth). Figure 2 shows the curves obtained for the two column diameters. While pyrene is only slightly retained in the phase system used, the plate height values are larger than theory predicts. Reasons for this discrepancy are not currently obvious; while it would he useful to have more accurate values for the diffusion coefficient, D,, under the conditions used, we believe that the current design of the injection device is the largest contributor to the plate height increase. Efforts are being made a t present to eliminate such problems. The direct insertion of the capillary tube into the spectral measurement cell reduces similar difficulties at the outlet. A model mixture of polycyclic aromatic hydrocarbons is shown in Figure 3. A 58 m X 0.2 mm i.d. glass capillary column was used to separate this mixture under isobaric conditions with n-pentane as the carrier fluid. While this result is only a preliminary one [capillary gas chromatography can achieve better separation of such compounds (23)],capillary supercritical fluid chromatography clearly merits further exploration.

Send for Tech Note #2


The simplified procedures are fully described in this well-illustrated 4-page technical note. Contact Rheodyne,Inc., PO. Box 996, Cotati, Calif. 94928, U.S.A. Phone (707) 664-9050.

ST2 RHEODYNE
~ ~

~.
~

l i l t LC CONNNtCTION COMPANY

CIRCLE 180 ON READER SERVICE CARD

410 A

ANALYTICAL CHEMISTRY, VOL. 53. NO. 3. MARCH 1981

The excitation source, ICP or DCP, is simply one component of a total analytical system. The source is complemented by a monochromator and a detection array. SpectraMetrics offers you a choice,high-resolution DCP or ICP emission systems. Sample types, chemical matrices, and elements analyzed govern the source chosen. More critical in the final analysis is the systems monochromator, more specifically spectral resolution. And that is where the Spectraspan family outstrips the competition. The reason our competitors never talk about resolution is simple; their conventional, medium resolution plasma emission systems cant resolve wavelengths as the SpectraSpans Echelle monochromator can. Their answer to the question What do I do when your medium resolution system cant resolve the lines that I want to use? is Choose other lines. Lines that are neither as sensitive nor as linear as your first choice. You just have to compromise. Compromise??. . .Never! With the Echelle you can resolve those lines that medium resolution instruments can t .High-resolution combined with the high-dispersion of the Echelle system virtually eliminates spectral overlap.Ask a competitor to do the scan at the right.

Furthermore, we give you the choice; High-Resolution DCP, High Resolution ICP, a High-Resolution ICP-MAES that complies with the proposed amendment to the Clean Water Act of 1979, and an ICP system for general analyses. So compromise.. .NEVER, not with a Spectraspan System, because at SpectraMetrics we will accomodate you in never compromising when it comes to resolultion. For our demonstrative report on resolution please circle the number or write for,Resolution-A Comparison of Spectral Bandpasses.

WHERE YOU HAVE A CHOICE


SpectraMetrics, Inc A Subsidiary of Beckman instruments, Inc 204 Andover Street Andover, MA 01810 (617)475-7015 Telex 94-7134
CIRCLE 206 ON READER SERVICE CARD

ANALYTICAL CHEMISTRY, VOL. 53, NO. 3, MARCH 1981

411A

I
0

n
5
6

lure 2. Plot of HETP vs. linear velocity for pyrene on a 0.30 mm 1.d. (blue) and 1.20 mm i.d. (green) column. The mbile phase was Rpentane at 210 "C and 32

R EAGENTGRADE SOLVENTS
Users of Corco chromatographic. electronic and spectrophometric reagentgrade solvents know they can count on Corco for highest quality. Reliability. And delivery. In pints, gallons, five-gallons, or drums. At Corco, we've been specialists in reagent-grade solvents since 1953and have built our reputation on meeting the requirements and specifications of our customers, ACS and ASTM. Corco can also provide reagentgrade acids, bases, and specialty chemicals. Write or call regarding your specific solvent requirements and our complete solvent listing in Bulletin 10. Or circle the number.

n The high efficiency demonstrated ove with capillary supercritical fluid romatography is a good reason for vsuing this research. Other facts of iportance also appear. For separaIns where both selectivity and high lumn efficiency are needed (e.g., resition of certain isomeric submces), it may be of advantage that e interaction of solute and mobile w e molecules ean he sensitively adited by pressure. In addition, while adient elution techniques are needed "program" solute retention in liq-

1 -

uid chromatography, pressure programming in supercritical fluid chromatography can he used to meet similar goals. As the former programming technique is incompatible with certain detection systems, fewer problems are anticipated with supercritical fluids under increasing pressure. For example, one can envision a complex mixture of nonvolatile substances in supercritical carbon disulfide being separated hy pressure programming, while the resolved components are introduced into a high pressure cell of an infrared detector. Some attention has already been given to the coupling of supercritical fluid chromatography with maas spectrometry through a jet molecular separator (24). While the technological problems that exist here are formidable, they appear no more serious than with the now extensively studied

I
5 3 0 4 5

LCIMS. As research in capillary supercritical fluid chromatography proceeds, different fluids and pressure regions can he explored to achieve chromatographic migration of various nonvolatile and unstable molecules. As pointed out hy Giddings and co-workers (9),the Hildebrand solubility parameters can be employed to predict approximate suitability of mobile phasea for various molecular separations. As our capability of performing precise measurements in supercritical fluid chromatography gradually improves, meaningful measurements of solut* solvent interactions in the vicinity of the critical point and beyond will a b become available. References
(1) Golay, M.J. E A w l . Chem. 1957,29, . 92-32,

Time (min)

COW0 CHEMICAL CORPORATION


Man~ifacUmS of R w n f & Ekcwonk C M s T@um Road 6 Csdsr LMe FIltb6.5 Hiiis, R. 19030 (215) 2955M16
CIRCLE 39 ON READER SERVICE CARD

Dum 3. isobaric run of several PAH andards in a 0.20 mm 1.d. X 58 m lpillary column. Conditions were as In jure 2. The standards were: 1) anracene. 2) pyrene, 3) benzo[k]fluonthene, 4) benzo[e]pyrene, 5)dlbenztbclanthracene* 6, benzo[ghilpe~ le, and 7 ) caonene

1'

412A

ANALYTiCAL CHEMISTRY, VOL. 53, NO. 3 MARCH 1981 .

Titration bonus: extra backup in everv box.


Truly a prize package: unsurpassed excellence in titration equipment. Plus total, solid support from the Fisher team of experts. The kind of backup only Fisher can provide - becausewere the only manufacturer to design and build our titration equipment in the USA., then distribute it through our own worldwide network. You get immediate technical aid, advice and literature. Directly from your local Fisher branch. Your Fisher rep has available dozens of Applications Bulletins,for specific titrations that arent covered in any handbook. All developed and proven in use by Fisher applications chemists. With innovative procedures in petrochemical, polymer, environmental, water/wastewater and pharmaceutical analyses. Including many applications where titration just hasnt been a viable method till now.

Get long-term support from Fisher titration headquarters. Our versatility means your investment in Fisher equipment gets more valuable daily. Because were constantly expanding the uses to which you can put it. Our engineers handle over 1200 requests every year, with specific advice and brand new procedures for literally hundreds of applications.To help you speed up routine lab testing and crack the toughest analytical enigmas. Everything you need and everything youre going to need. First, our uniquely flexible line of titration equipment. It lets you handle non-aqueous samples as easily as aqueous. Pinpoints true equivalence with a volumebased derivative, for top accuracy in minimal time.

Next, modular add-ons that let your system grow and change without obsoleting your present equipment. Supported by the most complete line of reagents and supplies anywhere. With nationwideservice and parts availability. The final bonus: our brains are yours. If youre into unchartedareas of titration, dont sa it cant be done until youve checkedYwith us. Feel free to consult our experts. Call this number: (412) 5628426. Get a Fisher applications chemist on the line. Check our files; pick our brains. Challenge us - and well take the dare. Because, if it can be done by titration, Fisher can do it.

Flhersciimc

CIRCLE 67 ON READER SERVICE CARD

(2) Giddings, J. C. Dynamics of Chroma-

tography; Marcel Dekker: New York,


1QfiS

(3) Giddings, J. C. Anal. Chem. 1964.36. 1890-92. (4) Tsuda, T.; Nakagawa, G. J. Chromatogr. 1980,199,249-58. (5) Knox, J. H.; Gilbert. M. T. J. Chromatog?. 1979.186. 405-18. (6) Klesper, E.; Corwin, A. H.;Turner, D. A. J . Ore. Chem. 1962,27,7WI. (7) Myers. M. N.; Giddings, J. C. Separation Sci. 1966. I, 761-76, (8) McLaren. L.; Myers. M. N.;Giddings, J . C . Srimee 1968,159,197-9. (9)Giddings. J. C.; Myers, M. N.; MeLaren, I..: Keller, R. A. Science 1968. 162,
(10) Sie, S. T.: van Beersum, W.; Rijnders,

67-73.

G. W. Separation Sei. 1966,I . 459A. wl (IyYSie, S. T.; Rijnders, G. W.A. Separa-

tion Sei. 1967.2, 729-53. (12) Sie, S.T.; Rijnders. G. W. A. Separation Sei. 1967,2, 755-77. (13) Sie. S.T.; Bleumen. J. P.A,; Rijnders.

G. W. A. In Gas Chroma raphy 1968; Harbourn, C. L.A.%d.; Institute of Petroleum: London, 1969: p 235. (14) Klesper. E Angew. Ckem. I n t . Ed. . Engl. 1978.17.738.16. (15) Novotny, M.; Rertseh, W.; Zlatkis, A. J . Ckromatogr. 1971,61,17-28. (16) Gouw. T. H.; Jentoft, R. E J . Chro. matogr. 1972.68, 303-23. (17) Sie. S T : . . Riinders. G. W. A. A n d . Chim.Aeto 1967.3R,31-44. (18) Kestin. J.; Whitelaw,J. H.;Zien,T. F. Phy.vieo 1964.30, 161-81. 119) Ishii. D.: Asai. K.: Hibi. K :Jonoku. . . chi, T.; Nagaya. M.J. Ckromatogr. 1977, 144, 157-68. (20) Hirata Y ;Novotny, M. J . Chromo. togr. 1979,186.521-8. (21) Scott. R. P.W.: Kucera. P.J . Chromotogr. 1979.169.51-72. (22) Blomberg. L ;Wannman, T J. Chro. . matogr. 1979.168.81-8. (23) Lee, M. L.; Novotny. M.; Bartle, K. D. Anal. Ckem. 1976.44 156672. (24) Randall, L. G.; Wshrhsftig, A. L. Anal. Chem. 1978.50, 1703-5.

COMBINES EASE WITH PRECISION.


8 8

Milos Novotny is currently professor of chemistry at Indiana University. A native of Czechoslovakia, he received his undergraduate education in chemistry and a doctoral degree in biochemistry at the University of Brno. Nouotnys research interests include capillary gas chromatography, HPLC, and GC-MS, as well as biomedical and environmental applications of these methods. John C. Fjeldsted received a BA in chemistry at Brigham Young University, and is also working toward his PhD degree in analytical chemistry. His research is concerned with the development of instrumentation and selective detection systems applicable to high resolution GC and supercritical fluid chromatography. Milton L. Lee received his BA at the University of Utah. and earned a PhD in analytical chemistry at Indiana University in 1975. He spent one year at MIT performing post-doctoral research before accepting his present position on the faculty at Brigham Young University. His research interests include the surface chemistry of glass and silica as it applies to GC, GC-MS. LC, and supercritical fluid chromatography,and the application of these techniques to the analysis of complex mixtures in environmental, coal, and coal-derived samples. Stephen R.Springstan receiued his BS degree from Virginia Polytechnic Institute and State University, Blacksburg, Va., in the field of chemistry. He is currently a doctoral degree student at Indiana University. His thesis research is concerned with the fundamental principles, instrumentation, and applications of supercritical fluid chromatography.

Molecular weights 100 to 25,000 Temperature range 25% to 13OOC Digital readout with analog recorder output Aqueous or organic solvents with no change in thermistors

The WESCAN Model 233 Molecular Weight Apparatus is a convenient


and easy means for routine determination of molecular weights up to 10,000, and with slightly greater care, up lo at least 25,000. It works on the well established principle of differential vapor pressure measurement using two thermistors in a closed chamber. Incorporation of the latest circuitry, and the same measurement chamber used in the CoronalWescan Model 232A, provides an instrument which is more precise and stable than earlier designs. Yet the Model 233 includes a number of operating conveniences which make it easy to use.

For more information,contacl


WESCAN INSTRUMENTS, Inc., 3018 Scott Boulevard, Santa Clara, CA 95050

P.A. Peaden

S. H. Springsfon

INSTRUMENTS,INC.
CIRCLE 231 ON READER SERVICE CARD

WESCAN
414A

Paul A. Peaden received his BS in chemistry at Brigham Young University, and is presently working on his PhD i n analytical chemistry. His research topic is the development of liquid Chromatographic and supercritical fluid chromatographic techniques for the analysis of high molecular weight polycyclic aromatic compounds.

ANALYTICAL CHEMISTRY, VOL. 53. NO. 3. MARCH 1981

Vous aimerez peut-être aussi