Académique Documents
Professionnel Documents
Culture Documents
Instruments
Service Manual Adjustment Procedure Circuit Diagrams Alignment Procedure for HM5010/HM 5011 ......... A Control and Adjustment of Supply Voltages ............................................... B Final Alignment - Tuner ..................................... C Alignment - IF-Unit ........................................... D Linearity of Frequency Display ......................... HM5010 XY-board ........................................... Tuner ............................................................... Tracking-Generator .......................................... E Alignment of HM5011 Tracking Generator ...... Tracking-Generator ................................................. Tuner RA-Board ...................................................... Tuner RB-Board ...................................................... IF-Amplifier ............................................................. Main board ............................................................. X-Y Board ............................................................... FC-Board ................................................................ PA-Board ................................................................ CRT-Board .............................................................. Power supply Board ............................................... RA-Board ................................................................ Tracking-Generator ................................................. RB-Board ................................................................ Main board ............................................................. Main board ............................................................. XY-Board ................................................................. XY-Board ................................................................. FC-Board ................................................................ FC-Board ................................................................ PA-Board ................................................................ CRT-Board .............................................................. PS-Board ................................................................ PS-Board ................................................................ PS-Board ................................................................ Block Diagram HM5010/HM5011 ..........................
Subject to change without notice
5 5 6 6 7 10 11 12 13 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39
HM5010/5011
SPECTRUM ANALYZER
Specifications
Frequency
Frequency range: 0.15MHz to 1050MHz (-3dB) Center frequency display accuracy: 100kHz Marker accuracy: (0.1% span + 100kHz) Frequency display res.: 100kHz (4 digit LED) Frequency scanwidth: 100kHz/div. to 100MHz/div. in 1-2-5 steps and 0Hz/div. (Zero Scan) Frequency scanwidth accuracy: 10% Frequency stability: better than 150kHz / hour IF Bandwidth (-3dB): Resolution: 400kHz and 20kHz; Video-Filter on: 4kHz Sweep rate: 43Hz
Amplitude
Amplitude range: -100dBm to +13dBm Screen display range: 80dB (10dB / div.) Reference level: -27dBm to +13dBm (in 10dB steps) Reference level accuracy: 2dB Average noise level: -99dBm (20kHz BW) Distortion: <-75dBc; 2nd and 3rd harmonic 3rd order intermod.: -70dBc (two signals >3MHz apart) Sensitivity: <5dB above average noise level Log scale fidelity: 2dB (without attn.) Ref.: 250MHz IF gain: 10dB adjustment range
Input
Input impedance: 50 Input connector: BNC Input attenuator: 0 to 40 dB (4 x 10dB steps) Input attenuator accuracy: 1dB/10dB step Max. input level: +10dBm, 25VDC (0dB attenuation) +20dBm (40dB attenuation)
Tracking Generator
Output level range: -50dBm to +1dBm (in 10dB steps and var.) Output attenuator: 0 to 40dB (4 x 10dB steps) Output attenuator accuracy: 1dB Output impedance: 50 (BNC) Frequency range: 0.15MHz to 1050MHz Frequency response: 1.5dB Radio Frequency Interference (RFI): <20dBc
Divers
AM-Demodulator output for head-sets. Permissible load impedance >8
General
Display: CRT. 6 inch, 8 x 10 div. intern. graticule Trace rotation: Adjustable on front panel Line voltage: 115 / 230V 10%, 50-60Hz Power consumption: approx. 20W Operating ambient temperature: 0C..+40C Protective system: Safety Class I (IEC 1010-1) Weight: approx. 7kg Cabinet: W 285, H 125, D 380 mm
Subject to change without notice 5/96
Accessories supplied: Line Cord, Operators Manual. Optional accessories, 50-feedthrough termination HZ22 Viewing Hood HZ47, Near Field Probe Set HZ530, Carrying Case HZ96-2, Transient Limiter HZ560
EMC-MEASUREMENT EQUIPMENT
Specifications
Frequency Frequency range: 0.1MHz to 1000MHz (lower frequency limit depends on probe type) Output impedance: 50 Output connector: BNC-jack Input capacitance: 2pF (high imped. probe) Max. Input Level: +10dBm (without destruction) 1dB-compression point: -2dBm (frequency range dependent) DC-input voltage: 20V max. Supply Voltage: 6V DC 4 AA size batteries Supply-power of HM5010/5011 Supply Current: 8mA (H-Field Probe) 15mA (E-FieldProbe) 24mA(High imp.Probe) Probe Dimensions: 40x19x195mm (WxDxL) Housing: Plastic; (electrically shielded internally) Package contents: Carrying case 1 H-Field Probe 1 E-Field Probe 1 High Impedance Probe 1 BNC cable (1.5m) 1 Power Supply Cable (Batteries or Ni-Cads are not included)
SCALE = 10dB/DIV.
SCALE = 10dB/DIV.
adjust: adjust: adjust: adjust: check: check: check: check: check: check:
The corresponding voltage test points to check direct voltage can be measured at the measuring connector strip (see PIC 2). Basic adjustment: When Y-pos. knob is snapped in on front board, adjust beam approx. 2 mm below bottom graticule line via R801.
4 3
1 2 3 4 5 6
PIC 1. PS-Board 5
CRT-Board
The tuner is already aligned by the factory. When changing the 1st mixer, it might be necessary to realign cavity filter: Set signal to 500MHz -27dBm to input, center frequency to 500MHz, Scanwidth to 0.5MHz, turn all three M3 screws to max. amplitude. In case curve is not uniform, coils L1, L2, and L3 can be adjusted to maximum uniformity (fig. 22) by slightly bending them. In case signal in frequency range is shacky, 2nd local oscillator might not be snapped in correctly. The correct snapin of the PLL is visible via the lock detect LED (D2), which has to be lit constantly without flickering if PLL is snappedin correctly. The oscillator signal has to be fixed (crystal dependent) at 1.32GHz, it may not drift. The PLL is snapped-in when the tuning voltage VT (measure at PAD1) is between 0.5 and 4.5V. The alignment will bring the tuning voltage to the center of the tuning range. VT of the 2nd LO has to be between 2V and 2.5V. Constantly check tuning voltage while aligning tuner. Alignment is necessary in two cases: VT<2.0V and VT>2.5V. Case 1 VT<2.0V Correction: a ) Solder all adjustment areas to connect trace from C1 to C2. b ) Remove excess solder at center conductor of resonator pad. Case 2 VT>2.5V Correction: a ) Adjustment areas may not be soldered to connect trace from C1 to C2. b ) Add solder to resonator pad.
Alignment of IF-Unit
Measurement setup: Apply two different signals to the input of the HM5011 via BNC T-connector and 10dB attenuators. Adjust HM5010/5011: Center frequency to 500MHz; all attenuators on (40 dB) Bandwidth 400kHz; video filter off; scanwidth 0.5MHz/div.; marker off. Adjust Tracking Generator: Attenuator 20dB; level max. (HM5011 only) RF-generator: frequency 500MHz; level 5dBm
The output voltage of the tracking generator is visible on the screen by an overlay of the fixed frequency. A "zero" point fig. 23) is visible within the signal to the right or to the left of the 500MHz spectral line (fig. 23 (5011 only). IF Filter Curve at 400kHz Bandwidth
Step 1
Align (with plastic screw driver) Coil L1 (PIC 5) to maximum output amplitude and symmetry to Y-axis. Step 2 Align (with plastic screw driver) Coils 3 + 4 + 13 + 14 (PIC 5) to symmetrie to Y-axis. The "zero" point must be at the maximum (at 500MHz) (fig. 23). fig. Stept 3 If necessary, repeat step 1 and 2. VR 1
L 10 L 13 L 11 L 14 L 12
L7
L3 L1
L8 L4 L9
VR 4
VR 2
align:
Step 1
Align (with plastic screw driver) Coils L7+8+9+10+11+12 until "zero" point is exactly centered at 500MHz (as with 400kHz Bandwidth). Watch for symmetry of Filter Curve.
IF Filter Curve at 20kHz Bandwidth (HM5011) Pre-Alignment: Remove T-connector. Connect tracking generator module directly to the input of the HM5011. Turn off attenuators on tracking generator module and on HM5011 (0 dB). The output voltage of the tracking generator module is now visible as horizontal line (with slight ripple). adjust: Step 1 Step 2 Adjust (with plastic screw driver) Coils 7 + 8 + 9 + 10 + 11 + 12 to maximum screen height of displayed output voltage. Add attenuators of HM5010/5011 as soon as "line" has reached the middle of the screen (30 dB). This alignment has to be performed repeatedly in order to optimize settings.
adjust:
Fine Alignment: Connect BNC-T-connector again as in the beginning of the IF alignment. Set attenuator of tracking generator module to 20dB, scanwidth to 0.5 MHz/div. If necessary, re-adjust center frequency (adjust 500MHz spectral line to screen center). "Zero" point is now visible to the right or to the left of screen center (amplitude maximum) (fig. 24). fig.
adjust: adjust:
Step 1 Step 2
Align (with plastic screw driver) Coils L7+8+9+10+11+12 until "zero" point reaches maximum. Watch for symmetry of Filter Curve. This alignment has to be performed repeatedly in order to optimize settings. fig. 24
IF Gain - Adjustment of different Bandwidths Set scanwidth to 0.2MHz/div. Switch bandwidth repeatedly between 400kHz and 20kHz. The distance between the output voltage (line) of the tracking generator and tge 400MHz amplitude may not vary. If the amplitude values do not match exactly, adjust by means of R-trimmer VR2 (PIC5) (PIC5). Linearity of IF-Amplifier Gain The linearity of the IF-amplifier gain has to be checked through the entire display range. Apply a 400MHz (-27dBm) signal directly to the input of the HM5010/5011. Adjust scanwidth to 5MHz/div, release attenuator switched (0 dB), select filter bandwidth of 400kHz. The spectral line should reach the upper screen edge. Use attenuators to reduce signal in 10db steps, whereby each individual attenuation step has to reduce the level by 10dB 1dB. In case deviation is as follows: - A the drop of the individual attenuation steps is larger than 10dB 1dB, or - B the drop of the individual attenuation steps is smaller than 10dB 1dB, the linearity of the attenuators has to be adjusted as follows: Set attenuator to -40dB. Adjust spectral line exactly to -40dB (center line) via VR1, VR3, VR4. Then set attenuators back to 0dB und adjust spectral line by means of trimpots VR802A and VR801 (XY board) to zero point and base line to bottom graticule line. This procedure has to be repeated until the settings at -40dB and 0dB are correct.
Settings for HM5010/5011: Center frequency at 500MHz; all attenuator switches released; filter bandwidth 400kHz; video filter off; scanwidth to 100MHz/div.; marker off. Measurement Setup: Apply signal of 500MHz 27dBm to input of HM 5010/5011.
Check: Check:
Basic setting: The HM5010/5011 has to be adjusted that the noise level touches the bottom graticule line. The spectral line of 500MHz -27dBm reaches the top graticule line and is situated in the center of the screen. Adjust 500MHz spectral line exactly to screen center via X-pos knob (on front of the unit). Upper frequency limit: Check if frequency of minimum 1050 can be set as center frequency.
Adjust HM5010/5011 to basic setting (see above). Apply input signal of 100MHz, level +7dBm Overriding the input allows for the harmonics of the input signal to become visible (fig. 7) This fig. 7). simplifies the adjustment of individual spectral lines in horizontal direction.
adjust:
Spectral line at 400MHz Turn X-ampl. knob on front of instrument to have the 4 spectral lines matched with the corresponding graticule lines on the screen(fig. 8 fig. 8).
fig. 7
adjust:
Spectral line at 100MHz With Trimpot RV171 align to match this spectral line with the corresponding graticule line on the screen (fig. 9 fig. 9). Zero Peak With Trimpot RV173 align to match this spectral line with the corresponding graticule line on the screen (fig. 9 fig. 9).
fig. 8
adjust:
adjust: adjust:
Spectral line at 600-1000MHz With Trimpot RV186 adjust to match 700MHz to the correct graticule line. Spectral line at 800 MHz; With Trimpots RV181 + RV183 for 900MHz and RV197 for 1000MHz (PIC 3) adjust to match this spectral line with the corresponding graticule line on the screen.
fig. 9
adjust:
Beam length With Trimpot RV101 adjust beam to end approx. 1mm beyond the right CRT boundary (graticule) (Fig. 11 - base line too short) Fig. (Fig. 10 - 300 to 1000MHz spectral lines not correct ) Fig.
fig. 10
RV 112
RV113
fig. 11
RV 149 RV 101 RV 183 RV 181 RV 186 RV 197 RV 171 RV 173 PIC 3 - MB-Board Set Trimpot RV113 that frequency will not display below 990MHz when center frequency is set to lowest frequency
Subject to change without notice
Adjust Marker. Settings on HM5010/5011: Center frequency to 500MHz, all attenuators in off position, filter bandwidth to 400kHz, video filter off. Scanwidth to 100MHz, Marker on. Signal 100MHz +7dBm. Adjust Marker to 500MHz display and set to 500MHz mark on screen via RV149 (PIC 4). Turn Marker left to max. and adjust to 990MHz via RV112. VR 3 VR 2
L 10 L 13 L 11 L 14 VR 4 L 12
L7
L3 L1
VR 1
L8 L4 L9
RV 112
RV 113
RV 149
RV 101
PIC 3 - Main-Board 9
XY-Board
RV 802A
1
1 2 3 4 5 6
RV 801
10
1st Mixer
Tracking-Generator
RV4
RV3
C68 C69
MPAD2
12
E 1. 1.1 1.2 1.3 1.4 1.5 2. 2.1 2.2 2.3 2.4 2.5
Alignment of HM5011 Tracking Generator Required Instruments: Spectrum Analyzer - minimum 1000MHz Completely assembled HM5011 Multimeter to measure DC voltage Oscilloscope, i.e. HM303 Coax cabel Preparation: Assemble Spectrum Analyzer completely (without case). All sub-assemblies have to be pre-checked. HM5011 has to be warmed up. Connect VCO output of RA board to VCO input (MCX connector). Connect 12MHz reference clock of TG board to RCA connector ST1 on RB board.
3.
Check Signalling Lines Cable connection W1 Pin Nr. 1 2 3 Description -12V +12V +5V Signal Supply OP Supply Supply PLL
4.
Check Supply and Bias Voltages Pin Nr. 1 1 3 5 3 3 Description 1. VCO Amp. 2. VCO Amp. fix LO Amp. Var. Amp. 1. Power Amp. 2. Power Amp. Voltage +5.8 V +5.8 V +5.9 V +5.6 V +3.6 V +4.8 V
5.
Alignment of fixed LO
5.1 The 2nd LO oscillates at a frequency of 1.35 GHz. The level at the mixer input IC3 Pin2 is -17dBm. Oscillation of the 2nd LO can also be measured with the H or E-Field Probe (HZ530). 5.2 Lock-Detect LED (D4) will confirm the correct PLL lock when lighting up without flickering or going out. The oscillator signal has to remain fixed crystal dependent) at 1.35GHz and may not drift. PLL is locked if tuning voltage VT (measure at PAD2) is between 0.5 and 4.5V. Align tuning voltage to medium tuning range. VT of fixed LO should be between 2 and 2.5V. The tuning volage has to be check constantly during alignment. Place coax resonator CR2 bottom side to the edge of soldermask next to center conductor, and solder on the left and the right side. 5.3 Alignment is necessary in two cases: VT<2.0V and VT>2.5V. Case 1 VT<2.0V Correction: a ) Solder all adjustment areas to connect trace from C1 to C2. b ) Remove excess solder at center conductor of resonator pad. Case 2 VT>2.5V Correction: a ) Adjustment areas may not be soldered to connect trace from C1 to C2. b ) Add solder to resonator pad.
13
Control of fixed LO-Level To check 2nd LO-level, cap C20 (22pF) has to be removed and 50Ohm Coax-cable has to be connected to 13dB attenuator of output. Connect other end of coax cable with adequate test analyzer. Adjust test analyzer to 1.35 GHz center frequency; the LO-level to be measured has to be -17dBm (1 dB) (consider attenuation of cable). If cap is not removed and coax cable is connected parallel to mixer to the VCO-branch, a level of approx. -24 dBm (1dB) should be measured due to decreased load resistance. Check of VCO Adjust HM5011 to Zero Scan, center frequency to 500MHz. Remove Cap C14 (22pF), connect one side to attenuator output, other side stand up in the air. Connect Cap side sticking up in the air to center conductor of 50 Ohm cable. Solder shielding of cable directly next to Cap to ground. Connect other end of Cap to Analyzer input (i.e. Advantest R3361A). Start 1.35GHz Stop 2.35GHz reference level: 0dBm. VCO signal which can be observed on the test analyzer has to shift according to center frequency (tuning voltage). Tuning range has to be at least from 1350 to 2350 Mhz. continuously and without interruption. It should not go under absolute level of +7dBm (coax cable of 1 mtr. has 1-2 dB of attenuation at 2GHz). Level has to be between +7 dBm and +10 dBm. Connect coax cable parallel to 1st mixer to VCO-branch. A level of +5 dBm to +8 dBm due to decreased load resistance over frequency band. Check of Attenuators: Set -30dB attenuation to Tuner and to 0dB attenuation to TG. Set Center Frequency to 500MHz, 100MHz/Div. span. Connect tracking generator to tuner input. TG-line has to be visible. Switch on additional attenuators. Attenuation for each attenuator switch has to be 10dB (1 dB) TG Level Adjustment: Connect TG to reference analyzer. Set HM5011 to Zero Scan and 500MHz Center Frequency. Release all attenuator buttons of TG (0-dBm) and turn TG-level to maximum. Adjust amplitude to +1dBm with Pot RV4 on TG. turn TG-level to minimum. Adjust amplitude to -10dBm with Pot RV3 on TG. Amplitude has to be adjustable now between +1dB and -10dB. Final Check
7.4
7.5
8. 8.1 8.2 8.3 8.4 8.5 9. 9.1 9.2 9.3 9.4 9.5 9.6 9.7 10.
10.1 Position of TG-level has to align in both bandwidths (400kHz and 20kHz). If necessary, align with IF-Amp. 10.2 TG-level knob has to be adjustable from +1 to -10dBm.
14
Tracking-Generator
15
Tuner RA-Board
16
Tuner RB-Board
17
IF AMP IF PCB
+12V R8 221R R5 220R L1101 +12V L1102 R9 3K32 R6 10K0 4.7V= R7 10K0 KA1102 T1 BC547 C8 10n TO XYF PCB P803 S.3 VID BW F3 RF2 C11 82p +5V 100u 6 5 4 3 2 1 6 5 4 3 2 1 +12V GND +5V GND VIDEO SIGNAL BW 100u KA1101
C9 0.1u
C10 10n
IF IN
30.0 MHz
IF
C4 8p2
C1 8p2
C3
22p
R1 100k
250mV
C2 47p
F1 RF3
R10
2.2k
500mVpp
T3 BC557
D1 1N4148
C15 22p
C19 1p8
D2
R11 4K75
1N4148
F19
R19 4K75
R21 100k
RF2
C18 120p
R40 2K21
C17 0.1u
CD1 1nF
CD2 1nF
0V(250 kHz)
IF-Amplifier
+12V
BW
+8V(12.5 kHz)
31
32
+VREF +28V L-D IC106B LM324 R135 P109 5 47k5 -6V2 1 C140 * R168 2 10 8 9 R137 1M5 R136 8k25 1 C125 14 (VT) +5V 12 IC106D LM324 1 3 +VREF 100R R141 9k09 +28V R143 100R R133 56k2 4 C115 P110 Z Y X1 X0 X 14 15 4 1 2 3 1 2 3 GND BLANK GND 2 0.1u 1 R124 1k50 SAWTH IC101A TL074 C105 22nF R114 5k11 -6V2 R142 15k0 13 12 1 1 IC106A LM324 0.1u R170 * Y1 Y0 1 2 R127 150k 3 Z1 Z0 C116 0.1u 1 C109 3 5 R145 * C143 470p TO CRT W701 S.7 R147 * R146 * PCB 10k0 9 10 11 6 C B A INH +5V +28V R144 * +VREF IC105 HEF4053 C110 10nF C131 0.1u D144 * D110 BAS16 C126 1000u /16V * 3 IC107 * I N ADJ R100 0R O U 2 R166 * R165 * 1 C134 * D128 BAV99 3 C132 * D180 * R198 56k2 * TO PS P1006 S.8 C139 0.1u 7 6 R134 C128 1000u /16V R138 1M5 2 G OU N IN D -12V (+12V) GND (-5V) +5V +32V 3 R162 100R IC128 LM79L05 RV197 2k5 R179 * R196 10k0 RV180 * 49k9 * R181 0R 30k1 R187 3k32 R185 * L-A1 301k 4k99 RV171 5K0 R192 20k0 R172 2k74 R173 47k5 R176 * D171 BAV70 2 R194 * 13 D197 BAS16 RV173 500K L-A2 R175 * R182 4k75 3k01 R189 1k00 RT102 * R186 221k D181 BAW56 2 RT101 2k2 3 1 R193 61k9 301R R140 562R 499R 0R R183 10k0 RV183 100K L-C2 R197 47k5 IC106C LM324 R180 * R171 1k50 RV176 * R174 * R191 30k0 R195 0R R184 0R R188 L-B RV186 10K R178 * 10k0 L-C1 RV181 5K0 6 5 4 3 2 1 6 5 4 3 2 1
ANALOG SECTION
R105
IC103 HEF4051
R106
R107
G D C B A T N V D C
Main board
P107
X0 X1 X2 X3 X4 X5 X6 X7
13 14 15 12 1 5 2 4
R104
MOLEX6
1 2 3 4 5 6
INH
1 2 3 4 5 6
A B C
6 11 10 9
R103
R102
R101
IC102 HEF4051
R110
R108
+VREF
X0 X1 X2 X3 X4 X5 X6 X7
13 14 15 12 1 5 2 4
R109
CF
INH
A B C
6 11 10 9
R130 39R
C127 0.1u
R112 10k0
+28V
+5V
R177 2k74
RV113
1k0
R111 33k2
CF-MIN +5V R156 221k R157 8k25 IC104B LM393/M 6 7 LM393/M 8 -5V +5V R158 1M00 R167 22k1 R160 100R -6V2 R159 1k50 C129 0.1u C130 4n7 5 D103 BAS16 R161 51R1 -6V2
-6V2
1 1
2 C120 +VREF SW-AMPL +28V R150 47k5 4 2 1 3 IC104A 0.1u RV101 100K R120 22k1 R117 22k1 D109 BAS16 R123 274k R126 10k0 T101 BC860C R153 * R113 6M8 C107 R125 8k25 9 8 10 IC101C TL074 R190 6M8 3 1 R148 10k0 RV149 10K MKR-Offs. 2 R149 10k0 +VREF -6V2 C136 0.1uF x100V -12V R10b 8k25 T102 BC850C -6V2 C138 3n3 R119 22k1 C121 220p R154 22k1 * R118 4k75 R155 47k5 -6V2 R10a 10k0 14 IC101D TL074 R115 10k0 R10c 6M8 R10d 6M8
T108 BC850C
+20V 60mA
C137 0.1u
C141 0.1u
R116 100k
12
IC108 LM317
I N ADJ O U 2
13
+VREF
D102 BAS16
R163 2k05
R128
14k3
C108 1nF
RV110
P108
MKR
10K 2
/10turns
R129 6k19
1 2 3 4 5 6
VREF
C135 0.1u
R152
1k00
RV112 1K0
-5V +5V R121 221R R122 68k1 IC101B TL074 6 7 5 0.1u 0.1u 0.1u 0.1u C102 C104 C111 C133
MK-MIN
C119 0.1u
R151 18k2
0.1u
0.1u
-6V2
(PS) 110...210ns 15 12 %10 R222 51R1 9 IC206C 74HC02 1 0 IC203 HEF4020 12 HEF4011 8 10 9 R267 1k00 13 R262 51R1 IC204C R249 51R1 1 1 1 0 IC206D 74HC02 R S T C L K 13 11 1 1 1 1 3 2 1 5 4 2 3 6 4 5 7 9 R233 51R1 11 13 9 3 1 IC204A HEF4011 %10k 1 R257 51R1 11 10 15 12 14 CKA CKB CLR IC202B 74HC390 J201 COAX R224 51R1 TO TG J402 S.6 8 MHz OUT %4 2 D225B %2 BAW56 3 5 R255 51R1 D225A 3 IC205C 74HCU04 6 R266 51R1 R225 2k21 %8 R223 51R1 UNC D265 BAS16 5 IC204B HEF4011 6 IC206B 74HC02 R258 51R1 C265 0.1u PE QA QB QC QD 13 11 10 9 C262 4n7 LM 317 R254 2k21 IC205D 74HCU04 8 IC204D HEF4011 2 %2k %1k 6 4 5 4 R264 51R1 LE D263 BAS16 R263 100k Q Q Q Q Q 1 1 1 1 1 Q Q Q Q Q Q Q 4 3 2 1 0 9 8 7 6 5 4 1 12 8 1 R221 51R1 CKA CKB CLR QA QB QC QD 14 13 11 10 9 R250 22k1 2
+5V
TOP VIEW
3 R259 51R1 BAV70 2 1 3 IC206A 74HC02 R261 51R1 IC205A 74HCU04 1 4 2 BAW56 CKA CKB CLR IC202A 74HC390 R256 51R1 %4 1 2 QA QB QC QD 3 5 6 7
C201
R211 51R1
0.1u
%2
C202 0.1u
1 4
CKA CKB
IC201B 74HC390
CLR
QA QB QC QD
3 5 6 7
C203 0.1u
IC201A 74HC390
C L k
IC205F 74HCU04
C205 0.1u
C216 10nF
R216 51R1
C206 0.1u
IC205B 74HCU04
R169 3
LM 7805
1k00
R199 1k00
1 2
2 1
XT254 8MHz
C199 0.1u
P201 MOLEX6
1 2 3 4 5 6
R253 10k0
MMS0204 / 0.25W / 5%
C253 27p
C254 56p
+ V T G G P 2 T V N N S 0 C D D V
R265 4k75 1 2 3 4 5 6
P202 MOLEX6
HAMEG INSTRUMENTS GERMANY Title MB PCB HM5010/11 C L P U G G L E E N N N K C D D TO FC PCB W303 S.5 Size C Date: November 6, 1995 Document Number MB50101B.SCH Sheet 4 of REV
DIGITAL SECTION
W802
W801 6-Pol
YF1 NC YF2 CHECK-PTS. R817 +140V 10R +140V +73V R815 C824 0.1u/250V 47u/250V R868 1.5mA R816 1K82 +140V R828 10R0 R850 10R0 C827 C804 0.1u 0.1u R843 100R T809 BFS20 T814 BF422 X AMPL T813 BF422 2 R855 1K82 3 R857 100R 1 R819 1K82 RV803 4K7 A R856 1K82 R860 4K75 R846 22K1 R822 2k74 R805A 100R C803 0.1u C823 10u/25V 3 10u/25V BAS16 R862 1K50 R 0R R829 10R R847 10R -12VA -12V C812 0.1u R861 3K01 C828 C809 0.1u T807 BC847B D802 1 R818 1K82 R870 4K75 R844 0R R842 4K75 C811 0.1u R820 100R R848 22K1 R859 2 T808 BFS20 100R 3 R858 100R R849 100R 1 X POS RV801 10K RV804 4K7 A 1 +12VY YP-sym T802 BFS20 R805 100k C802 0.1uF 3 2 R823 100R R845 22K1 10u/25V C810 C822 10u/25V 47K5 1W * C826 C825 0.1u/250V R821 100R R854 100R +20V...+125V R867 47K5 1W P810 2.5MB6 +73V +12V TO CRT PCB P702 S.7 +140V +32V GND +5V +12V -12V 1 2 3 4 5 6 1 2 3 4 5 6
1 2 3
Y AMP
1 2 3 4 5 6
X AMP
3-Pol
+73V
+73V
R814 100R
R866 30K1 1W
R865 30K1 1W
R810 100R
2.0mA
+133V
R825
R806 100R
R824 100R
100R
T803 BFS20
Y AMPL
T811 BF422
T812 BF422
R826 *
R807 100R
R812 1K50
R813 100R
R809
1K50
RV802A 4K7
R811 1K50
R804 2K21
R808
1K50
R836 10K0
R801 4K75
T801 BC847B
R802 1K0
D801 BAS16
C801 0.1u
R830 10R
+12V
R841 100R
R833 100K
R838 2K74
C806 0.1u
T806 BC860C
R853 10R
3 2 1
C807
22p
T805 BC850C
R839 1K50
R834 12K1
T804 BC850C
R852
10K0
R835 12K1
R832 22K1
R840 *
R837 10K0
C808 22p
R851 6K81
R831 10R
C805 0,1uF
-12V
SOT23
Top View
BC850
R-Chip 1% 1206
BAS16
BC860
R-Chip 5% 1206
nc
XY Board
BFS20
33
34
+5V IC321A 74HC74 4 D P R 6 CLK C L Q D340 BAW56 +5V 3 1 BAV70 4 1 ALPS R342 Q 2K21 1 C L Q 3 R344 221K +5V 8 1 3 +5V R359 100R 1 3 T340 BC850 2 3 RV310 10K A 9 CF FINE 2 MARKER OFF 2 1 3 2 1 0 D P R CLK 5 2 D341 TO MB PCB P108 S.4 6 3 1 SW305 R343 10K0 1 Q 5 /60mm W305 6pol 2 3
FC-Board
IC320 HEF4017
CLK ENA
15
RST
Q0 Q1 Q2 Q3 Q4 Q5 Q6 Q7 Q8 Q9
1 2 3 4 5 6
CO
D301
+5V
BAV70
D302 BAV70
L302 100uH
IC300 ICM7217N
7 6 5 4 D303 BAS16 6 6 1 6 6 1 1 6 1 1
18 17 16 15
T301 BC850
10
R301
14
4K75
C302 0.1u
24
19
10 9 8 5 4 2 3 7 10 9 8 5 4 2 3 7 10 9 8 5 4 2 3 7 10 9 8 5 4 2 3 7 10 9 8 5 4 2 3 7
9 IC301 HP3603 1000MHz IC302 HP3603 100MHz IC303 HP3603 10MHz IC304 HP3603 1MHz IC305 HP3603 0.1MHz
20 23 27 25 28 22 26 21
12
R302 10K0
R304 10K0
R341 475R
R336 39R2
+5V
R311 182R LED305 2 LED306 IC308 LED307 IC307 MC14029 +5V 74HC42 3 R313 22K1
T303 BC860
D304 BAS16
100
IC306A MC14011 1
R310 221R
50
D305 BAV99
20
R309
+5V
10
10K0
C310 0.1u
C311 0.1u
C309 0.1u
T304 BC850
0.5
R325 100R
R315 221R LED313 +5V LED314 8 10 LED315 T306 BC860 IC306C MC14011 9 R335 4k75
0.2
R323 82K5
R312 1M00 R326 100R IC306D 13 11 12 MC14011 VF BW +5V +5V R339 * 3 1 1 2 3 4 5 6 R322 1K00 R331 33K2 2 BW C339 * R338 * 1 2 3 4 5 6 W304 6pol /70mm 5 2 B A B C D T W V Z C TO MB PCB P107 S.4 4 1 5 4 2 1 SW304 6 3 6 SW303 3 T331 BC850 TO XYF PCB P802 S.3 R321 33K2 C308 0.1u W301 6pol R327 100R
5K0 TR
0.1
BAV70
D306
R316 221R
ZERO
R336 *
BAW56
D307
/300mm
R318 100K
R330 10K0
1 2 3 4 5 6
W303 6pol
/65mm
G G N N D D
SW301
12.5kHz
250kHz
ALPS
UP
SCANWIDTH
DOWN Title
BANDWIDTH 12.5kHz
VIDEOFILTER OFF
FC-PCB HM 5010/11 Size C Date: November 22, 1995 Document Number FC50101A.sch Sheet 5 of REV
PA-Board
22
23
D701 1 2 3 1 2 3 D14-364 TO MB PCB P110 S.4 13 9 R706 10K0 4 12 R710 3 2 1 5 IC701 6 CNY17 11 7 6 GND BLANC GND R701 681R KA701
CRT-Board
LL4151
D702 LL4151
2 R705 T701 BC850 10K0 C701 0.1u R704 1k0 D703 TLSG5101 1 1 4
51R1
KA702
1 2 3 4 5 6
1 2 3 4 5 6
TO PS P1002 S.8
R703 47k5
0.1u/100
P702
6 5 4 3 2 1
6 5 4 3 2 1
HAMEG INSTRUMENTS GERMANY Title CRT PCB HM5005/6 Size A Date: March 14, 1994 Document Number CRT5005.SCH Sheet 7 of 8 REV
24
RA-Board
25
Tracking-Generator
26
RB-Board
27
TO
Main Board
28
Main board
29
XY-Board
30
XY-Board
31
FC-Board
32
FC-Board
33
PA-Board
34
CRT-Board
35
PS-Board
36
PS-Board
37
PS-Board
38
39
Oscilloscopes Multimeters Counters Frequency Synthesizers Generators R- and LC-Meters Spectrum Analyzers Power Supplies Curve Tracers Time Standards
HAMEG GmbH Industriestrae 6 D-63533 Mainhausen Telefon: +49 (0) 6182 / 800-0 Telefax: +49 (0) 6182 / 800-100 E-mail: sales@hameg.de service@hameg.de
Internet:
www.hameg.de
4S-5010-0010
Printed in Germany