To summarize, this is a chronocoulometric method (potentiostatitc) with smart adjustment of the applied voltage. The coulometry is measured and when a value (set by the user) is reached the experiment is stopped.
E1 is the applied potential of the step 1; E2 is the applied potential of the step 2; ECI is the potential measured during the OCV period after 20 ms and n is the number of step needed to reach the charge limit.
This experiment is built with CA (chronocoulometry) and with OCVperiod. Loop technique is added to create step of potential.
The charge limit is controlled in the Experiment limit in the Advanced settings tab.
Advantage: This experiment is already available Drawback: The protocol is changed (ocv period reduce to 20 ms) The sampling rate is 200 s instead of 97 s.
C-2 with EC-Lab Express or DLL (based on Modular pulse technique) For this proposal, the potentio and galvano periods are performed with the Modular Pulse technique (available in V5.50 of EC-Lab Express and V5.27 of DLL).
R&D work: Add the charge limit Advantage: Sampling rate of 97 s can be reached. Drawback: The protocol is changed (ocv period reduce to 20 ms)
D- True protocol
If user cannot accept the previous proposal, only EC-Lab express can do the job (sampling rate is around 100 s).
R&D work: Add the charge limit (note the charge of interest is not the charge of the sequence but the total charge, charge of the whole technique) The applied potential in the step n is calculated from the OCV period. In the step n+1, it has to be possible to set smartly the applied potential of n+1 step. E2 has to be E1 + ECI. User should select the time at which the ECI is measured.