Académique Documents
Professionnel Documents
Culture Documents
Candela Corporation 530 Boston Post Road Wayland, MA 01778-1886 Telephone (508) 358-7637 Toll Free (800) 733-8550 (Technical Assistance) Toll Free (800) 73-LASER (Customer Service) 8502-00 -0869Revision A June 2004
TABLE OF CONTENTS
This "User Verification Guide" applies to the following Smoothbeam Delivery Systems:
* Introduction * Test #1 - Beam Alignment for 4 mm & 6 mm Delivery Systems * Test #2 - Cryogen Alignment for 4 mm Delivery Systems * Test #3 - Cryogen Alignment for 6 mm Delivery Systems * Test #4 - Cryogen Air Bubble Detection for 4 mm & 6 mm Delivery Systems * Warnings and Precautions
8502-00-0869 Revision A
Sheet 2 of
13
Introduction: The purpose of this "User Verification Guide" is to provide you with a series of simple tests that will assist you in determining that your Smoothbeam delivery systems are working optimally. In addition, a series of "warnings" have been included. Through these "warnings" you may obtain a better understanding of variables that may affect the durability and performance of the Smoothbeam Candela laser delivery systems. It is important that you carefully review and follow the procedures and instructions contained in this "User Verification Guide" and in the laser's operator manual, treatment guidelines, and related information. FAILURE TO USE THE LASER IN ACCORDANCE WITH SUCH PROCEDURES AND INSTRUCTIONS COULD RESULT IN SERIOUS INJURY TO THE OPERATOR, THE PATIENT AND/OR OTHERS, AS WELL AS DAMAGE TO THE LASER DELIVERY SYSTEM. The fact that your laser delivery system passes the applicable tests contained in this "User Verification Guide" does not guarantee that your laser delivery system is problem- free. Discontinue use of your laser delivery system if you still suspect a problem with your laser delivery system.
Overview of Tests: This guideline contains information regarding four (4) types of tests. There are three specific tests to be performed per handpiece. Each test should be performed for the indicated delivery systems at the beginning of each treatment day using the tests outlined in this manual. In addition, check the delivery system if there is any concern about the delivery system's performance or the delivery system has been dropped. Discontinue use of the delivery system if problems are noted in any of these tests or you suspect / observe other factors that may affect performance.
Note: Perform tests 1-3 for 6 mm handpiece (pages 5, 8, 9, and 10) and then for the 4 mm handpiece (pages 5, 6, 7, and 10). This will eliminate the need to switch back and forth between the 4 mm and the 6 mm handpieces while performing the procedure.
8502-00-0869 Revision A
Sheet 3 of
13
Introduction (continued)
Test # 1 2
Test Beam alignment Cryogen alignment Cryogen alignment Cryogen air bubble detection
Purpose Identify a misaligned distance gauge. Verify that the cryogen spray nozzle is properly aligned with the distance gauge ring. Verify the cryogen spray nozzle is properly aligned with the distance gauge ring. Check that air bubbles in the cryogen line are detected and the associated "fault" messages are displayed on the system.
6 mm
4 mm & 6 mm
WARNING PLEASE WEAR APPROPRIATE SAFETY GLASSES DURING THE TESTS You will need the following supplies to perform these tests:
Zap-It Laser Alignment paper part Number Z-48 Available from Zap-It Corporation, Andover, NH 03216, USA. Template (P/N 1301-00-8192) available from Candela Corporation.
8502-00-0869 Revision A
Sheet 4 of
13
6. 7.
Good Full
###################333333
Template
Bad Incomplete
Clipping
Results: Acceptable results: The laser spot should form a round circle within the 4 or 6 mm distance gauge. Use template (1301-00-8192) to ensure that the laser spot on the ZAP-IT is round and located within the 4 or 6 mm circle on the template. This ensures that the laser beam is aligned within the distance gauge. The laser spot does not have to fill the template. The template provides a visual reference for being round. Be sure to clean the window before proceeding with treatments. Unacceptable results: If the laser spot IS NOT a full circle and does not match the proper template, the distance gauge is clipping the laser beam. Replace the distance gauge and repeat the test. If the spot IS a full circle, discard the original distance gauge. If after repeating the test, the spot still IS NOT a full circle, contact Candela to return the entire delivery system. Smoothbeam replacement distance gauges and windows are listed below:
8502-00-0869 Revision A
Sheet 5 of
13
4 mm Smoothbeam Procedure:
1. 2. 3. Warning: Put on appropriate laser safety glasses . Install the 4 mm delivery system on the laser. Holding the handpiece, select a fluence of 10 J/cm2 and the DCD cryogen spray time of 60ms. Warning: Stay in Standby. While directing the handpiece in a safe direction, purge the cryogen approximately 5 times by nd depressing the Next Screen button, and pressing purge button on 2 screen. This will fill the cryogen line within the delivery system.
4.
5. 6. 7.
Insert the handpiece into the calport and change the DCD setting to 20ms. Enter Ready mode and calibrate the laser by depressing the footswitch. After calibration has been completed, remove the handpiece from the calport. Holding the distance gauge contact ring flat on the ZAP-IT paper, pulse once onto the black side of the ZAP-IT paper.
8502-00-0869 Revision A
Sheet 6 of
13
Test #2 Cryogen Alignment for 4 mm Delivery Systems continued 4 mm Smoothbeam Procedure (continued): The following are examples of acceptable and unacceptable burn patterns. Observe where the dark area is in relation to the light area. The dark area should be near the middle of the circle rather than near the edge of the circle.
Note: If the dark area is not seen or is faint; reduce the fluence until the dark area is visible on the zap-it paper.
Acceptable results:
Unacceptable results: Repeat test If the result is still unacceptable, discontinue use of the delivery system and replace the delivery system by contacting Candela Technical Support.
8502-00-0869 Revision A
Sheet 7 of
13
6 mm Smoothbeam Procedure:
1. 2. 3. Warning: Put on appropriate laser safety glasses . Install the 6 mm delivery system and insert the handpiece into the calport. Holding the handpiece, select a fluence of 10 J/cm2 and the DCD cryogen spray time of 60 ms. Warning: Stay in Standby. While directing the handpiece in a safe direction, purge the cryogen approximately 5 times by depressing the Next Screen button, and pressing purge button on 2nd screen. This will fill the cryogen line within the delivery system.
4.
5. 6. 7.
Insert the handpiece into the calport and change the DCD setting to 20ms. Enter Ready mode and calibrate the laser by depressing the footswitch. After calibration has been completed, remove the handpiece from the calport. Holding the distance gauge contact ring flat on the ZAP-IT paper, pulse once onto the black side of the ZAP-IT paper.
8502-00-0869 Revision A
Sheet 8 of
13
Test #3 Cryogen Alignment Test for the 6 mm Delivery System (continued) 6 mm Smoothbeam Procedure continued:
The following are examples of acceptable and unacceptable burn patterns. Observe where the dark area is in relation to the light area. The dark area should be near the middle of the circle rather than the near the edge of the circle.
Note: If the dark area is not seen or is faint; reduce the fluence until the dark area is visible on the zap-it paper.
Acceptable results:
Unacceptable results: Repeat test If the result is still unacceptable, discontinue use of the delivery system and replace the delivery system by contacting Candela Technical Support.
8502-00-0869 Revision A
Sheet 9 of
13
Cryogen Connector
5. 6. 7. 8. 9.
Set the DCD spray on the front panel to the highest setting (85 ms is typical). Set the FLUENCE to the lowest setting (this varies by spot size). Enter the Ready mode, and calibrate the laser by pressing on the footswitch. Remove the handpiece from the CALPORT and aim in direction away from personnel (such as the floor). Depress the footswitch to pulse laser continuously until a PURGE REQUIRED window box appears on the user screen and lasing ceases. This should happen, typically 20 pulses but could be as much as 50 pulses. Results Comments This particular handpiece passed the test This particular handpiece failed the test Action You may continue use of this handpiece Discontinue use of this handpiece and call Candela Tech Support at (508) 358-7637 ext. 336 or (800) 733-8550 ext. 336.
10. Reconnect the Cryogen Connector (see Figure 1) to use the system, or repeat the test on other Smoothbeam handpiece delivery systems. 11. Depress purge until the line is refilled.
8502-00-0869 Revision A
Sheet 10 of
13
8502-00-0869 Revision A
Sheet 11 of
13
8502-00-0869 Revision A
Sheet 12 of
13
8502-00-0869 Revision A
Sheet 13 of
13