Académique Documents
Professionnel Documents
Culture Documents
A COMPARATIVE AND
OF
STUDY AUGMENTED
TOTAL
OF STANDARD,
HIP
REPLACEMENTS
LONG POSTERIOR WALL
ACETABULAR
COMPONENTS
RICHARD
M.
NICHOLAS,
JOHN
F.
ORR,
RAYMOND
A.
B.
MOLLAN,
JAMES
W.
CALDERWOOD,
JAMES
R. NIXON,
PETER
WATSON
From
Queens
University
and
Musgrave
Park
Hospital,
Belfast
Augmentation
of the acetabular
of total
hip replacements
is a method
of increasing
stability
and preventing recurrent dislocation. turning moments and angles required prostheses.
a series of mechanical experiments designed to evaluate the standard, long posterior wall and two different augmented
Dislocation
following
total
hip
replacement
is a serious
complication for both patient and surgeon. The incidence of dislocation has ranged from less than 1% to 8% (Bergstrom et al 1973; Charnley and Cupic 1973; Coventry et al 1974 ; Fraser and Wroblewski 1981 ; Khan, Brakenbury and Reynolds 1981 ; Dorr et al 1983). Many factors are involved ; they may act independently or may summate and have been discussed by many authors. Previous operations or injuries at the hip may have disrupted the soft tissues around the arthroplasty (Carlsson and Gentz 1977; Lewinnek Ct al 1978; Fackler and Poss l90; Woo and Morrey 1982; Dorr et al 1983). Woo and Morrey (1982) showed that a posterior surgical approach was associated with an increased dislocation rate. have Acetabular both been of and femoral shown to be component important malposition factors. Retromay be at operation. femoral due to Too
Internal rotation the femoral component acetabular component causing has Fluid been
the
posterior dislocation (Coventry et al 1974). collection within the pseudocapsule of the hip implicated (Ritter 1980), while mental and disturbances can contribute. a higher 1980) primary 1982). hip flexion complication (Khan et Revision rate al 1981) surgery such as is associAndersson
neurological Parkinsonism ated with and Herberts (20%) than and Mallory increasing
(Ahnfelt,
and a much higher dislocation rate replacement (Williams, G#{246}ttesman Coventry et al (1974) suggested that due to a gradual stretching of the by hip 1983; but was
pseudocapsule could account for late dislocations. Management of recurrent dislocation. Treatment brace has been described (Clayton and Thirupathi Dorr et al 1983) for recurrent instability
the acetabular component positioning of the patient little the anteversion likelihood of the of instability
or too increase
component on movement
;
satisfactory only in selected cases. Williams et al (1982) described a six week period of immobilisation in an above knee hip spica after open or closed reduction ; they had success in 1 5 of 16 selected cases. The design of a standard Charnley prosthesis prevents dislocation by any uniaxial load, unless the neck of the margin femoral (Fig. component 1). In theory, impinges on the acetabular the larger the head of the
of the hip (Lewinnek et al 1978; Khan and Morrey 1982; Dorr et al 1983).
et al 1981
Woo
MB BCh, BAO, Senior House Officer MIMechE, Senior Bioengineer FRCS, FRCSI, Professor of Orthopaedics
J. W. Calderwood, FRCS, Consultant Orthopaedic Surgeon J. R. Nixon, MCh(Orth), FRCS, Consultant Orthopaedic Surgeon
P. Watson, Musgrave Correspondence PhD, AMIMechE, Affiliate ASME, Research Fellow Park Hospital, Belfast BT9 7JB, Northern Ireland. should be sent to Dr R. M. Nicholas. Joint Surgery
component the more stable the hip replacement be, since dislocation would require the head to be through a greater distance. Woo and Morrey that this of femoral can factor head was of less importance diameter to neck width. either by augmenting the the distance through before dislocating, or thus removing
be prevented
1990
British Editorial Society of Bone and 0301-620X/90/3075 $2.00 JBonefointSurgfBr] 1990; 72-B: 418-22.
movement,
418
THE
JOURNAL
OF BONE
AND
JOINT
SURGERY
DISLOCATION
OF TOTAL
HIP
REPLACEMENTS
419
of impingement. by narrowing
the
for a
et al (personal augmentation
tion of the acetabular another identical with two or three further dislocations though opposite one oftheir direction.
component, using a sector cut from acetabular component, and secured cortical screws. They reported no in the augmented region of the cup patients sustained a dislocation is a relatively arthroplasty recurrent 1985). Gie, in the simple with dislocation Mogensen, its We Augmentation
MATERIALS
AND
METHODS
turning
technique in comparison to revision high incidence (33% to 44%) of (Woo and Morrey 1982; Coventry Arnason (1989) and J#{243}nsson (1986), and have described augmentation
standard,
replacements (inverted cup and metal backed). A compression testing machine simulated joint loading of 300 N to the articulated prostheses. This load was chosen for convenience and represented the weight of the compression testing machine anvil. A head, with the standard Charnley neck) placed testing femoral prosthesis (22 mm standard the neck compression was mounted in a fixed position vertically in the axis of loading of machine. A standard Charnley diameter) was a tight-fitting
acetabular mounted
Secondary
Impjngement
nylon retaining collar (Fig. 2). The block was then mounted using 0.5 inch diameter steel stub-axles between a pair of radial ball bearings whose axis of rotation coincided with the centre of the acetabular prosthesis. This assembly was therefore free to rotate around the centre of rotation ofthe static femoral head under turning moments wheel. minimise inaccuracy dislocation. A spring The applied entire lateral and by means apparatus loading to cope was of an aluminium was placed on caused with attached by any displacements loading rollers to machining during wheel
balance
to the loading
so that the turning moment applied to the cup could be calculated. Attached to the loading wheel was a 360#{176} protractor to measure angles and ranges of movement. Primary impingement occurred when the neck contacted Secondary the femoral of the nents the inner impingement component margin (P1) was of the femoral (SI) occurred first contacted deemed to component component. neck of margin when the the outer have first
of the acetabular
1). The joint compountil secondary unstable moments recorded. dislocations could have : sudden at primary
impingement
dislocation followed. The and secondary impingement Experimental dislocations. 1) Repeated may damage dislocation. the inner
of a cup given an
erroneous turning moment on subsequent dislocations. A standard acetabular component was therefore dislocated in the apparatus 18 times in succession, noting the
Fig. The compression standard acetabular test machine components. 2 showing femoral and
turning moment at which dislocation occurred. The angles of primary and secondary impingement and the range of movement before and after the series of dislocations were also noted.
VOL.
72-B,
No.
3, MAY
1990
420
R. M. NICHOLAS,
J. F. ORR.
R. A. B. MOLLAN,
J. W. CALDERWOOD,
J. R. NIXON,
P. WATSON
2) Standard acetabular component. Five new standard cups were dislocated once each, noting the angles of primary and secondary impingement, the range of movement and the turning moment to dislocation.
3) Long posterior wall cups were primary movement femoral containing and stall component. dislocated once secondary turning always Five new long posterior each, noting the angles of the range of to dislocation. out of the The sector
regression, the line of best fit had the equation Y = 0.0003 X + 0.229. The gradient of 0.0003 is therefore virtually horizontal up to 1 8 dislocations. However the range ofdata points were large (0.22 to 0.26 Nm) and the correlation Before the this seem became coefficient experiment,
-
did
to be significant,
on the acetabular
multiple
dislocations.
o5:
E z
C a,
-,
Consecutive
I
dislocations
of one
component
//
//
E
0
\ /\
\\
#{149} .
/\
I
E a)
C C I-
/
\., /
\
\\\\
2nd Impingement
/
\
/
\__.
\\
,/
II
Fig.
Dislocation
The methods of augmentation used with the inverted cup method (left) and the metal-backed UHMW polyethylene device (right). These are both shown mounted on long posterior wall components.
Table
and
secondary
of Olerud and Karlstr#{246}m (1985). Three AO cortical screws were used (Fig. 3). Each augmented cup was dislocated once, noting the angles of primary and secondary impingement, the range of movement and the turning were only moment used be used because with this to dislocation. the type Long of component. posterior procedure wall cups could augmentation
5) Metal-backed UHM W polyethylene augmented acetabular component. A standard cup was augmented using this device. Five AO screws were used to secure the device to the cup (Fig. 3). Even when the apparatus was loaded to the maximum turning moment dislocation did not occur. Therefore only the angles to primary impingement and the range of movement were noted. Secondary impingement could not be achieved. are RESULTS The effect ofrepeateddislocation on acetabular component to dislocate number of to linear
allowed by all the prostheses The largest mean range for the standard acetabular had a mean cup allowed
of
movement
(106.2#{176}) as w
4: the turning moment not decrease with the the data were subjected
68.8#{176} the metal-backed and The angles between primary ment are shown in Table I ; they
THE JOURNAL
OF BONE
DISLOCATION
OF TOTAL
HIP
REPLACEMENTS
421
of
40%. This restriction authors to provide Karlstr#{246}m 1985; The turning posterior those for wall the
of range increased
primary and secondary 6. That for secondary wall prostheses (0.34 augmented cup devices UHMW polyethylene be dislocated.
impingement are shown in Figure impingement in the long posterior Nm) was greater than that in the (0.29 Nm). The metal-backed could not augmentation devices
could
(Olerud
Karlstr#{246}m 1985
Gie
at al 1988).
.--
90
80 70
C) .
showed that this system However, it also suggested loading in vivo (possibly movement) such a rigid the cement fixation and/or device. the /
,/
provided the greatest stability. that in a situation of extreme caused by the restriction in system could lead to failure of fracture ofthe screws locating
a, a, 0) a, 0 a, 0) C
60;
50 40
augmentation
30
20
o45T
0 .4 0.35
. --...
-----
-.--.
io;
0
.
...T:_
E z
0.3.
I I
. . . . . . ...,
r
.
0.25 0.2
.
E 0)0.15
C C
. .
. . .
5 of all prostheses.
0.1.
I.
.
.
0.05k 0..--...
:
.
:
..
I.--
I
Long posterior
.-
wall
Metal-backed
UHMW
polyethylene augmentation device
Clinical
results.
We
have
used
acetabular total
for three patients ments with well posterior After further tated dislocation reduction dislocations by minor
who had Charnley aligned components. in falls under general (four, two trauma. from
a standing
and
secondary
Augmentation
sectioning a long posterior wall prosthesis then secured to the acetabular component cortical further screws. dislocations. None of these patients have
Augmentation has a role in the management of recurrent dislocation of total hip arthroplasties, since it involves a shorter operation time, a smaller incision, less soft-tissue damage These advantages and would no disruption clearly be of most hard tissues. beneficial in
DISCUSSION Multiple that the dislocation. dislocations number did This of the same not significantly but component affect clinically showed the ease of there is an initial rather
metal-backed augmenin that it may be used wall cost cups. and It is also flexibility low
posterior
is reassuring,
in accommodating
of cup.
wear of the prosthesis. The ranges of movement allowed by the augmentation devices were significantly lower than those of standard and long posterior wall prostheses, by almost
VOL. 72-B, No. 3, MAY 1990
We would like to thank Mr V. Rimmer of Chas. F. Thackray Ltd. for providing materials which were used in this study. No benefits in any form have been received or will be received from a commercial party related directly or indirectly to the subject of this article.
422
R. M. NICHOLAS,
J. F. ORR,
R. A. B. MOLLAN,
J. W. CALDERWOOD,
J. R. NIXON,
P. WATSON
REFERENCES
Ahnfelt L, Andersson G, Herberts P. Reoperation tiker. Lkartidningen 1980; 77:30-1. Bergstrom B, after total seriesofcases. Carlsson AS, Brunswick harnley Clayton
av totala
Fraser
Lindberg
hip
L, Persson BM, Onnerf#{228}lt R. arthroplasty according to Charnley C/in Orthop 1973; 95 :91-5.
dislocation C/in Orthop
GA, Wroblewski BM. Revision of arthroplasty for recurrent or irreducible Surg[Br] 1981 ; 63-B :552-5. A, Scott replacement T, [lag RSM. dislocation. PH, J Bone Cup J Bone
Gie
Khan
for recurrent hip 1989; 71-B :338. following :214-8. J total JR. Bone for
J, Cupic Z. The nine and ten arthroplasty of the hip. Clin Orthop ML, Thirupathi RG. plasty : management by Orthop 1983; 177:154-9. Late dislocations J BoneJoint
Lewinnek
arthroC/in hip
GE, Lewis JL, Tan R, Compere Dislocations after total hip-replacement Joint Surg [Am] 1978 ; 60-A :217-20. Arnason H, J#{243}nsson GT. total hip. Acta Orthop Scand
Socket 1986;
in patients with Charnley Surg[Am] 1985; 67-A :832-41. course and 56-A :273-84.
total
arthroplasties : a study of postoperative cations. J Bone Joint Surg [Am] 1974, Dorr
S, Karlstr#{246}m G. Recurrent dislocation ment : treatment by fixing an additional component. J Bone Joint Surg [Br] 1985 MA. A treatment plan for the dislocated C/inOrthop 1980; 153:153-5. RYG, Joint Money Surg [Am] BF. Dislocations
;
after total hip replacesector to the acetabular 67-B :402-5. total hip arthroplasty. J Bone hip C/in
Ritter Woo
LD, Wolf AW, Chandler R, Conaty JP. Classification treatment of dislocations of total hip arthroplasty. C/in 1983; 173:151-8. CD, Poss R. Dislocation 1980; 151:169-78. in total hip arthroplasties. C/in
after :1295-306.
total
hip arthroplasty.
1982
64-A
Williams
Fackler
JF, Gottesman MJ, Mallory arthroplasty : treatment with an Orthop 1982. 171 :53-8.
THE
JOURNAL
OF BONE
AND
JOINT
SURGERY