OIML R 49-1
Water meters intended for the metering of cold potable water and hot water meters
International Standards
Major International Standards:
International Organization of Legal Metrology (OIML)
OIML R49-1 OIML R49-2 OIML R49-3
International Standards
Major International Standards (cont.):
o Comite Europen de Normalisation (CEN)
EN 14154-1 EN 14154-2 EN 14154-3
National Standards
Examples of National Standards:
NMI 49-1 (Australia) NIST Handbook 44, Section 3.36 (USA) BS 5728 (United Kingdom) Draft Weights and Measures Regulation, Division XIV (Canada) National Measurement Law (Japan)
The 2006 edition is a revision of the 2003 version with the inclusion of R 72 (hot water meters) and combination meters The United Kingdom had been the secretariat of OIML TC8/SC5 since its inception.
Smallest difference between indications of an indicating device that can be meaningfully distinguished For a digital device, this is the change in the indication when the least significant digit changes by one step.
Note: The list may be extended to higher or lower values in the series.
3.1.5 The ratio Q2 / Q1 shall be 1.6. until 30 April, 2009, Q2 / Q1 may also be 2.5, or 4, or 6.3, provided that Q3 / Q2 > 5. 3.1.6 The ratio Q4 / Q3 shall be 1.25.
Classification Comparison
Old classification system - 15 mm meter Class A Class B Class C Class D
New classification system q3 = 1.6 m3/h q3/q1 = 20 q3/q1 = 80 q3/q1 = 100 q3/q1 = 200
Q4 Q3 Q2 Q1
The maximum permissible error for the lower flowrate zone (Q1 Q < Q2) is 3 %. Accuracy class 1 designation shall be applied only to water meters with Q3 100 m3/h. Note: Class 1 meters not common for customer billing
Re-verification In-service
mpeu 2 % mpel 5 % mpeu 4.0 % mpel n/a mpeu 2.5 % mpel 6.0 % mpeu 2 % mpel 5 % mpeu 4 % mpel 10 % mpeu 4 % mpel 10 % n/a mpeu 4 % mpel 10 %
Sampling
mpeu 3 % mpel 8 % n/a n/a tbd
(h) (i)
(j) (k)
Shall include visual means of testing and calibration May include additional elements for testing or calibration Indicated volume shall be expressed in cubic metres
Q3m3/h
Example # 1 2 3
Flowrate: Water temperature: Ambient temperature: Ambient relative humidity Ambient atmospheric pressure
0.7 x (Q2 + Q3) 0.03 x(Q2 + Q3); Within 5 C of reference value(s) of Table 1 (20 5) C (60 15) %; 86 kPa to 106 kPa [0.86 to 1.06 bar].
OIML R 49-1
Water meters intended for the metering of cold potable water and hot water meters
Questions or Comments