Vous êtes sur la page 1sur 8

CHAVES ET AL THE JOURNAL OF PROSTHETIC DENTISTRY

Effects of three soldering techniques on the strength of high-palladium alloy


solder joints
Marisol Chaves, DDS, MS,a Stanley G. Vermilyea, DMD, MS,b Efstratios
Papazoglou, DDS, MS,c and William A. Brantley, PhDd
College of Dentistry, The Ohio State University, Columbus, Ohio

Statement of problem. Little information is available on the optimum technique for soldering high-
palladium alloys, which have gained considerable popularity for prosthodontic applications.
Purpose. The objective of this study was to compare the flexural stress at the proportional limit of four
noble dental alloy specimens soldered with torch, oven, and infrared techniques.
Materials and methods. The high-palladium alloys studied were Legacy XT (Jelenko), Freedom Plus
(Jelenko), and IS 85 (Williams/Ivoclar). A gold-palladium alloy, Olympia (Jelenko), served as the control.
Thirty round bars, 18 × 3 mm, were cast from each alloy, cut in half, aligned, and joined using Olympia
Pre solder (Jelenko) for the gas-oxygen torch and the infrared technique and Alboro LF solder (Jelenko)
for the oven technique. Each soldered bar was subjected to three-point bending, and the maximum elastic
stress or strength of the solder joint was calculated at the proportional limit. Data were analyzed by two-
way ANOVA and the Ryan-Einot-Gabriel-Welsch (REGW) multiple range test at the 0.05 level of signifi-
cance.
Results. There was no significant difference between torch and oven-soldering, but both were signifi-
cantly different from the infrared technique. ANOVA showed a significant difference between alloys, but
this difference could not be detected with the REGW test. SEM examination of the fracture surfaces
revealed grooves associated with the path of crack propagation. X-ray energy-dispersive spectroscopic
analysis failed to detect copper in the solders, and there were no significant changes in the solder composi-
tions after the melting procedures.
Conclusions. All three techniques can yield satisfactory solder joints in high-palladium alloys. These
joints should be well-polished to achieve optimal strength. (J Prosthet Dent 1998;79:677-84.)

CLINICAL IMPLICATIONS
The results of this investigation indicate that the strengths of high-palladium alloy
joints properly prepared by the torch, oven, and infrared soldering techniques are ac-
ceptable for clinical use and have generally comparable values. The current results
also demonstrate the importance of having solder joints that are properly polished. The
apparent notch-sensitivity of solder joints in high-palladium alloy castings that con-
tain finishing grooves suggests that there should be some concern when cyclic loads are
applied over a clinically appropriate period of time.

This article is based on a thesis submitted in partial fulfillment of the


requirements for the degree of Master of Science in the Gradu-
ate School of The Ohio State University, 1997.
H igh-palladium alloys have become popular for
the fabrication of metal ceramic restorations and im-
Presented at the AADR meeting in Orlando, Florida as a finalist for plant superstructures. Among the reasons for the great
the Arthur R. Frechette award competition, March 1997. Presented interest in these alloys are their cost, excellent mechani-
at the Carl O. Boucher Conference, April 1997.
Supported by the Greater New York Academy of Prosthodontics and
cal properties, and good bonding with dental porcelain.1
by NIDR Research Grant DE10147. The alloys and the investments used to fabricate cast
a
Senior Resident, Advanced Education Program in Prosthodontics. restorations are selected carefully to compensate for di-
b
Associate Professor and Chair, Section of Primary Care. mensional changes that occur during the fabrication pro-
c
Clinical Assistant Professor, Section of Restorative Dentistry, Prosth- cess. For three-unit prostheses, the marginal adaptation,
odontics and Endodontics; and Doctoral Graduate Student in Oral
Biology.
described by Ziebert et al.,2 was the same for soldered
d
Professor, Section of Restorative Dentistry, Prosthodontics and En- restorations or one-piece castings. However, Gegauff and
dodontics; and Director, Graduate Program in Dental Materials. Rosenstiel3 found that for three-unit fixed partial den-

JUNE 1998 THE JOURNAL OF PROSTHETIC DENTISTRY 677


THE JOURNAL OF PROSTHETIC DENTISTRY CHAVES ET AL

Table I. Nominal compositions (wt%) of alloys*


Alloy Pd Cu Ga In Au Ag Ru Sn

Freedom Plus** 78 8.0 5.0 6.0 2.0 — <1.0 —


Legacy XT** 75.5 — 6.0 6.0 2.0 10.0 <1.0 —
IS 85*** 82 — 6.0 3.5 2.5 2.5 — 3.5
Olympia** 38.5 — 1.5 8.5 51.5 — <1.0 —

*Values were obtained from the product information literature.


**J.F. Jelenko and Co., Armonk, N.Y.
***Williams Division/Ivoclar North America, Amherst, N.Y.

tures (FPDs), the best adaptation was produced after mens according to ANSI/ADA Specification No. 5.15
soldering. Several studies have shown that, as the length The ends of the plastic patterns were cut with a dia-
of the prosthesis increases or involves the curvature of mond disk, leaving round bars 3 mm in diameter and
the arch, the potential for a poorly fitting prosthesis in- 20 mm in length. The specimens were then sprued with
creases.4-6 With the recent development of implant-re- wax (10 bars per casting ring) and invested with a car-
tained prostheses, which may often be long and may bon-free phosphate-bonded investment (High Span II,
need the use of increased amount of casting alloy, fabri- J.F. Jelenko and Co., Armonk, N.Y.). The manufacturer’s
cating a one-piece casting incorporating machined parts recommendations were followed for the special liquid-
and having these machined surfaces accurately and pas- to-powder ratio and burnout schedule for the invest-
sively fit the implant superstructure becomes even more ment.
difficult. As a result, the practitioner must have an ef- The high-palladium alloys used to cast the specimens
fective method of compensating for poorly fitting cast- were Freedom Plus, Legacy XT (Jelenko) and IS 85 (Wil-
ings. liams/Ivoclar, Amherst, N.Y.). A gold-palladium alloy
Soldering provides a means of improving the fit of a (Olympia, Jelenko) served as the control. The alloy com-
prosthesis and, at the same time, joining the components positions (wt%) are provided in Table I. Melting was
rigidly.7 The strength of solder joints is of utmost impor- performed with a multiorifice gas-oxygen torch, and the
tance for clinical success of the prostheses. Many studies alloys were cast with a broken arm centrifugal casting
have reported measurements of tensile strength for sol- machine. The castings were permitted to bench cool
der joints, with mixed results. Some investigators have before devestment. After air abrasion with 50 µm alu-
found that preceramic solder joints are stronger than minum oxide, castings were numbered at each end and
postceramic solder joints.8 Monday and Asgar9 and randomly assigned to treatment groups. Specimens were
Lorenzana et al.10 reported that there was no significant then sectioned at their midpoint with a low-speed dia-
difference in the ultimate tensile strength of postceramic mond saw (VR/50, Leco Corp., St. Joseph, Mich.) and
and preceramic solder joints. However, other authors11-13 water coolant. After sectioning, the bars were ultrasoni-
indicated that the ultimate tensile strength of postceramic cally cleaned with acetone and dried.
solder joints was greater than that of preceramic solder Corresponding halves of each casting were placed in
joints. Although tensile tests are useful in the evaluation a lathe for proper alignment. A thickness gauge was used
of soldering effectiveness, dental prostheses are subjected to provide a 0.5 mm gap. Autopolymerizing resin (GC
mainly to flexural loading during clinical use.14 A popular Corporation, Tokyo, Japan) was used to unite the halves
method to obtain the flexural strength of a material is of each bar. Each bar was then invested in a silicone
three-point bending, where a concentrated load is applied mold (Coltene, Whaledent Inc., Mahwah, N.J.) with Hi-
to the center of a uniform beam that is supported near Heat soldering investment (Whip Mix, Louisville, Ky.).
each end. The flexural strength is often termed the modu- The area to be soldered was left exposed during the in-
lus of rupture or transverse strength. vesting procedure, and an airway was created under the
In this study, torch, oven, and infrared soldering tech- soldering area to provide uniform heating of the joint.
niques were compared for three high-palladium dental The invested blocks were preheated according to the
alloys to determine the best method to join these met- alloy manufacturer’s instructions or those of the solder-
als. A well-known gold-palladium alloy served as a con- ing equipment manufacturer (infrared technique). Olym-
trol for the experiments. A literature review indicated pia Pre Solder (Jelenko) was chosen for the torch and
that there are no published studies comparing these sol- infrared-soldering groups, and Alboro LF solder
dering techniques for the high-palladium alloys. (Jelenko) for the oven technique groups.
For the torch-soldering, the investment blocks were
MATERIAL AND METHODS
preheated to 1100° F, and a gas-oxygen torch with a
The specimens were fabricated from polystyrene plas- No. 2 orifice point was kept moving at an oblique angle
tic patterns used for the fabrication of tensile test speci- to the preheated investment surface. A strip of approxi-

678 VOLUME 79 NUMBER 6


CHAVES ET AL THE JOURNAL OF PROSTHETIC DENTISTRY

mately 5 mm of solder was placed over the joint, and ing moment (Fl/4 for three-point bending), I is the
the solder was allowed to flow through the gap. moment of inertia (πd4/64 for a circular cross-section),
The oven-soldering experimental groups were sub- and c is the radius of the bar.16
jected to the Vita Omega cycle (Vident, Baldwin Park, Ten test specimens were prepared for each combina-
Calif.) under vacuum for porcelain application (two tion of alloy and soldering technique for a total of 120
opaque bakes, two body bakes and a low glaze bake), specimens. A power analysis17 performed on the experi-
before the bar was cut with the diamond saw. After the mental data, assuming α = 0.05 and a power of 80%
blocks were preheated in the oven, a 5 mm strip of fluxed (β = 0.20), indicated that a sample size of 10 would
solder was placed in contact with the parts to be sol- show significant differences of 185 MPa in the mean
dered. Then the temperature in the oven was raised to proportional limit or fracture stress between sample
the melting temperature of the solder, allowing it to flow groups. Bartlett’s test17 was performed to determine the
through the joint space. The soldering procedure was homogeneity of the variances for all of the specimen
performed using the vacuum available with the porce- groups, and the mean values of flexural stress (propor-
lain furnace. tional limit or fracture stress) values for all the groups
For the infrared-soldering groups, the preheated were ranked and subjected to two-way analysis of vari-
blocks were placed on the platform of the infrared sol- ance (ANOVA) to examine the overall effects of alloy
dering machine (Ney Infrared Unit, J.M. Ney Co., and soldering technique, as well as their interactions.
Bloomfield, Conn.). The invested specimen and the sol- The mean values for the specimen groups were ranked,
der strip were fluxed, and this solder strip was placed in and two-way ANOVA on the ranks18,19 showed that there
contact with the parts to be soldered. The joint area was were highly significant differences between techniques.
placed 1 mm below the alignment rod, and the infrared The Ryan-Einot-Gabriel-Welsch (REGW) multiple
lamp was activated and held at maximum level until the range test20 with a significance level of α = 0.05 was
solder flowed. The soldered specimens were permitted used on the ranks to determine which specific specimen
to bench cool before devesting. The specimens were groups were significantly different from each other. The
cleaned of residual investment, and the solder joints were REGW test is considered to have less likelihood of Type
finished flush with the remainder of the bar. Mounted II statistical errors than the well-known Tukey multiple
stones and rubber wheels were used with a low-speed range test. In addition, representative fracture surfaces
dental handpiece, and the bars and the handpiece were were randomly selected from each group and examined
attached to the lathe. The diameter of each soldered joint with a SEM (JSM-820, Jeol Ltd., Tokyo, Japan).
was measured, and external defects for each specimen The two solders were also melted individually using the
were noted. three techniques, embedded in epoxy resin, ground with
The specimens were subjected to a three-point flex- 400 and 600 grit metallographic paper, and polished with
ural load in a screw-driven mechanical testing machine a series of alumina slurries (from 15 to 0.05 µm particle
(model 4204, Instron Corp., Canton, Mass.) at a cross- size). Quantitative elemental composition information was
head speed of 0.25 mm per minute. The load and posi- provided by x-ray energy-dispersive spectroscopic analyses
tion of the crosshead were recorded by means of a (EDS), using a Link eXL microanalysis system with a
Pentium personal computer, with the software program PentaFET detector and an ultrathin window (Oxford In-
Labview for Windows (National Instruments, Austin, struments Group, High Wycombe, England) coupled to
Texas) and a data acquisition card with a frequency of the SEM. A similar procedure was performed for the as-
1 Hz. The raw data from each test bar were stored in received unmelted solder, and the changes in the composi-
the computer as a text file. The software program Excel tions of the solders after the melting procedures were de-
(Microsoft Corporation, Redmond, Wash.) was used to termined.
plot and manipulate the data. The midspan flexural stress
was calculated at the proportional limit (just before the RESULTS
point of initial nonlinear deformation as determined from The mean values with standard deviations of the pro-
a graphic representation of load versus crosshead move- portional limit for ductile solder joints or the fracture
ment) or at the point of failure, whichever was lower. strength of brittle solder joints for the various specimen
The equation used for the calculation of the maximum groups are summarized in Figure 1. Bartlett’s test showed
elastic flexural stress (σmax) developed at the midpoint of that the variances were significantly different from each
a centrally loaded round beam was as follows: other. For this reason, the technique of ranking the ob-
σmax = 8Fl/πd3 servations was used. Two-way ANOVA on the ranks
showed that there were highly significant differences
where F = load, l = distance between support points (test between techniques and a significant difference between
span length of 12.5 mm) and d = diameter. This equa- alloys (Table II). There was no significant interaction
tion was developed from the well-known elastic flexure between the alloys and soldering techniques. The sensi-
formula σmax = Mc/I, where M is the maximum bend- tive REGW multiple range test showed that, even though

JUNE 1998 679


THE JOURNAL OF PROSTHETIC DENTISTRY CHAVES ET AL

Fig. 3. SEM photomicrograph of fracture surface of IS 85 alloy


for oven-soldered specimen shows ductile fracture character-
istics of this alloy. (Bar = 10 µm.)
Fig. 1. Values of proportional limit or fracture strength for sol-
dered joints of high-palladium alloys obtained with three-point
bending test.

Fig. 4. SEM photomicrograph of torch-soldered Olympia speci-


men. (Bar = 1 mm.)

Fig. 2. SEM photomicrograph of fracture surface for oven-sol-


The specimens soldered with the oven technique sepa-
dered IS 85 specimen. (Bar = 1 µm.)
rated completely after being subjected to the three-point
bending test. No sign of bending or lines of fracture were
there were differences in the mean values for the four visually evident. The fracture path appeared to be com-
alloys that ANOVA (Table II) indicated were significant pletely flat, giving the impression that an adhesive frac-
(p = 0.048), these differences were not statistically sig- ture had occurred between the parent metal and solder.
nificant when the results for the three soldering tech- Flux was used on the oven-soldered specimens (as
niques were pooled (Table III). The REGW test was recommended by the manufacturer) and on the infra-
also used to analyze differences between the soldering red-soldered specimens because of the impossibility for
techniques used in this study where the results for the the solder to flow without the flux. The results showed
four alloys were pooled. It showed no statistically sig- that the flux did not perceptibly affect the solderability
nificant differences between the torch-soldering and of the alloys, but further quantitative analysis of the ef-
oven-soldering (Table IV). However, these two solder- fect of the flux would be useful.
ing techniques were significantly different from the in- Figure 2 is a photomicrograph of the fracture surface
frared-soldering technique. No specimen was rejected for an oven-soldered IS 85 specimen. SEM examination
from the statistical analysis after being fractured, regard- revealed that the specimen fractured cohesively through
less of the amount of porosity. the solder, even though visual examination suggested

680 VOLUME 79 NUMBER 6


CHAVES ET AL THE JOURNAL OF PROSTHETIC DENTISTRY

Table II. Two-way ANOVA results based on ranking the observations


Source DF ANOVA SS Mean square F value Prob. > F

Alloy 3 7784.65 2594.88333 2.72 0.04833


Technique 2 22231.95 11115.97500 11.63 0.00003
Alloy*technique 6 10776.10 1796.01667 1.88 0.09078
Error 108 103192.80 955.48889 — —

that separation was between the solder and alloy. An in- Table III. Ryan-Einot-Gabriel-Welsch test results for the alloys*
tergranular fracture surface can be seen as well, with Alloy Number of specimens Mean (MPa) REGW grouping
evident separation between some grains. This fracture
surface was characteristic of the solder. There was no Freedom Plus 30 902.40 A
evidence of microvoids or the dimpled rupture struc- Legacy XT 30 834.70 A
ture indicative of ductile fracture.21 IS 85 30 803.44 A
Olympia 30 800.64 A
Oven-soldered Olympia alloy specimens were also
randomly selected after fracture and observed at the same *Alloys with the same letter are not statistically different at the α =0.05 level.
range of magnifications used for the IS 85 specimens.
The fracture surfaces for these specimens were similar
to Figure 2, which would be expected because the same Table IV. Ryan-Einot-Gabriel-Welsch test results for the solder-
solder was used (Alboro LF), and the fracture mode was ing techniques*
again cohesive through the solder.
Technique Number of specimens Mean (MPa) REGW grouping
An area near the edge of an oven-soldered IS 85 speci-
men is depicted in Figure 3. This photomicrograph in- Torch 40 881.62 A
dicates that cohesive ductile fracture also occurred Oven 40 868.79 A
through the IS 85 alloy, because of the characteristic Infrared 40 755.48 B
dimpled rupture appearance resulting from the *Techniques with the same letter are not statistically different at the α = 0.05
microvoids. Previously, Stewart et al.22 found that this level.
alloy exhibited excellent ductility, as reported by the
manufacturer.
SEM photomicrographs of fractured solder joints for A low-magnification photomicrograph of an infrared-
Legacy XT and Freedom Plus specimens that had been soldered specimen of Olympia after fracture is shown in
oven-soldered showed characteristics similar to those of Figure 6. This photomicrograph shows that the princi-
the fractured solder joints for IS 85, which follows from pal path of fracture was in the alloy below the solder
the cohesive nature of the fracture process through the (white band in the center of the photomicrograph), i.e.,
same low-fusing solder. cohesive fracture through the alloy. Numerous voids in
The fracture of the torch-soldered and infrared-sol- the solder are also visible in Figure 6.
dered specimens visually appeared to have initiated at In the low-magnification photomicrograph (Fig. 4)
the tension side and propagated through the solder. Low- of a fractured Olympia specimen that had been torch
power SEM examination (Fig. 4) confirmed this obser- soldered, an indentation in the solder where the bend-
vation. For some of these specimens, fracture also oc- ing load was applied can be seen (center left edge of
curred through the parent metal, but no separation of photomicrograph). The width of the solder joint appears
the specimen halves was observed. In contrast, the oven- to be extended in a vertical direction near the right edge
soldered specimen halves of the alloys separated com- of the photomicrograph (toward the tension side in
pletely during testing. bending), due in part to crack propagation. However,
The fracture surface for a torch-soldered IS 85 speci- the indentation and this tapering suggest some ductility
men is depicted in Figure 5. Substantial porosity and a of the solder. These observations were also made for
rounded structure for the solder below a void are evi- another fractured specimen that had been soldered by
dent. Parallel grooves on the surface resulted from the the infrared technique.
experimental procedure of holding each soldered speci- X-ray EDS analyses were performed on the two sol-
men in the lathe and polishing with stone followed by ders to determine their compositions. It was found that
rubber wheel (with a handpiece). An association of the there was no copper in the composition of either solder.
fracture boundary with the polishing grooves is appar- The compositions of the solders before melting and af-
ent. This same pattern was observed on a representative ter the melting procedures with the three techniques
fracture surface of an infrared-soldered Legacy XT speci- are summarized in Table V. Only slight changes were
men. Again, there was a close association between the found in the compositions of the solders after the melt-
polishing grooves and the path of fracture in the solder. ing procedures.

JUNE 1998 681


THE JOURNAL OF PROSTHETIC DENTISTRY CHAVES ET AL

Fig. 5. SEM photomicrograph of torch-soldered IS 85 speci- Fig. 6. SEM photomicrograph of fracture path through parent
men demonstrates solder porosity. (Bar = 10 µm.) metal of infrared-soldered Olympia specimen. (Bar = 100 µm.)

Table V. Energy-dispersive spectroscopic analysis (wt.%) for composition of solders


Solder Au Ag Zn Pd Sn

Alboro LF* 62.9 (0.1) 23.2 (0.5) 13.5 (0.7) — 0.4 (0.1)
Olympia Pre* 70.5 (0.3) 16.5 (0.5) 0.9 (0.2) 10.9 (0.4) 1.1 (0.1)
Alboro LF (oven)** 62.3 (0.6) 24.1 (0.6) 12.1 (0.5) 0.9 (0.4) 0.6 (0.2)
Olympia Pre (infrared)** 72.1 (0.1) 16.7 (0.4) — 10.3 (0.6) 0.9 (0.1)
Olympia Pre (torch)** 70.9 (0.8) 16.7 (0.5) — 10.3 (0.2) 1.3 (0.4)

*Solder as received from the manufacturer.


**Solder after melting by the indicated technique.
Entries are mean values with standard deviations in parentheses. Five analyses were performed on each type of solder and condition.

DISCUSSION
It has been previously noted that a power analysis
The development of the strength for a solder joint showed that the sample size of 10 used in this study
involves melting, flowing, and wetting of the solder by would be able to detect a significant difference (α = 0.05)
capillary forces between the parent metal and the sol- between specimen groups of approximately 185 MPa
der.23 All three techniques used in this study were con- with a power of 80% (β = 0.20). This difference was
sidered to be acceptable for soldering the three repre- approximately 28% greater than the pooled standard
sentative high-palladium alloys. It is important to con- deviation of 144 MPa for all 12 specimen groups. For
sider the expertise of the operator for achieving solder example, sample sizes of 15 and 41 would have been
joints of similar characteristics. The problems that have required to show significant differences of 150 MPa and
been reported with the commonly used gas-oxygen torch 90 MPa, respectively, at the same α and β levels, and
for soldering metal ceramic alloys are gas inclusions, would have necessitated much larger numbers of 180
voids, improper melting of the solder, and excessive oxi- and 492 soldered specimens for the entire study. Al-
dation of the pieces to be soldered.24 The main advan- though a larger sample size might have revealed statisti-
tages with this technique are the availability of the gas- cally significant differences between some additional
oxygen torch in the dental laboratories and the flexibil- sample groups compared with the results obtained with
ity of this method for use with the preceramic and the current sample size, the general observations and
postceramic solders. clinical recommendations from this study would not be
It was initially expected that the torch soldering would substantially altered.
exhibit the greatest variation in results because of the rela- An alternative approach to the conventional torch or
tively poor control in the temperature during the solder- oven soldering is the use of an infrared heat source. In
ing procedure. However, this result was not always found. the commercially available unit (Ney) used in this study,
For example, the flexural strength of the Olympia speci- infrared energy for soldering is supplied by a quartz-
mens that were torch soldered had one of the smaller iodine-tungsten-filament lamp. Similar to the gas-oxy-
standard deviations of all the groups (Fig. 1). gen torch, flux is required with the solder in this tech-

682 VOLUME 79 NUMBER 6


CHAVES ET AL THE JOURNAL OF PROSTHETIC DENTISTRY

nique.25 Some of the reported advantages of infrared CONCLUSIONS


soldering are lack of gas inclusions, limitation of heat-
Under the conditions of this study, the following con-
ing area and less working time.25,26 Cheng et al.25 found
clusions were drawn.
that solder joints of a cobalt-chromium base metal alloy
1. All three techniques used in this study were judged
produced by infrared soldering had higher ultimate ten-
to be adequate for soldering high-palladium alloys.
sile strength than those produced by the gas-oxygen
2. Infrared soldering yielded the lowest values of flex-
torch. Other investigators reported little difference in
ural strength among the three techniques used.
joint integrity and strength, whether the thermal energy
3. X-ray energy-dispersive spectroscopic analysis in-
was delivered by a torch flame or infrared apparatus.27,28
dicated that there is no copper in either of the two sol-
Most of the infrared solder joints in this study frac-
ders used.
tured in a cohesive mode through the solder, but some
4. Elemental energy-dispersive spectroscopic analysis
showed failure through the parent metal (Fig. 6). This
also showed that there are only minor changes in the
soldering technique requires meticulous control of the
composition of each solder after the melting procedures.
focal point because, if this is not achieved, the solder
5. Microscopic grooves on finished specimens were
will melt unevenly and result in weaker joints, or not at
revealed by the scanning electron microscope to have a
all. Careful attention to voltage fluctuations is necessary
close association with the paths of crack propagation.
when using the infrared unit, because such changes could
This observation indicates that solder joints should be
alter the efficiency of the soldering procedure. The speci-
well-polished to achieve optimal strength.
mens soldered with the infrared technique were the weak-
est for each alloy (Fig. 1 and Table IV), but they were We thank John C. Mitchell, Senior Electron Microscopist, Depart-
judged to be sufficiently strong for dental prostheses. ment of Geological Sciences, for expert technical assistance in per-
forming the SEM/EDS analyses.
Oven soldering (postceramic soldering) is the best
choice when porcelain application has been performed. REFERENCES
The maximum temperature can be controlled to allow 1. Carr AB, Brantley WA. New high-palladium casting alloys: part 1. Overview
solder flow without distorting the porcelain. Staffanou and initial studies. Int J Prosthodont 1991;4:265-75.
et al.11 observed that more consistent connectors, in 2. Ziebert GJ, Hurtado A, Glapa C, Schiffleger BE. Accuracy of one-piece cast-
ings, preceramic and postceramic soldering. J Prosthet Dent 1986;55:312-
strength and size, were obtained with oven soldering 7.
for a wide range of dental alloys, when compared with 3. Gegauff A, Rosenstiel S. The seating of one-piece and soldered fixed partial
torch soldering. Stade et al.29 demonstrated that joints dentures. J Prosthet Dent 1989:62:292-7.
4. Bruce RW. Evaluation of multiple unit castings for fixed partial dentures. J
of similar or superior strength to the parent metal were
Prosthet Dent 1964;14:939-43.
obtained when oven soldering was used. However, oven 5. Huling JS, Clark RE. Comparative distortion in three-unit fixed prostheses
soldering requires a well-calibrated oven to achieve the joined by laser welding, conventional soldering, or casting in one piece. J
Dent Res 1977;56:128-34.
proper temperature for melting the solder, and the sol-
6. Garlapo DA, Lee S-H, Choung CK, Sorensen SE. Spatial changes occurring
der should contact both sides of the parent metal to be in fixed partial dentures made as one-piece castings. J Prosthet Dent
soldered. This technique can only be used for 1983;49:781-5.
7. Padilla MT, Bailey JH. Margin configuration, die spacers, fitting of retainers/
postceramic soldering, because most of the ovens do
crowns, and soldering. Dent Clin N Am 1992;36:743-64.
not achieve temperatures necessar y to melt the 8. Squire BE. The relative strength of high and low fusing solders. [MS thesis.]
preceramic solder. In this study, all of the oven-sol- Indianapolis: Indiana University, College of Dentistry; 1971.
9. Monday JJL, Asgar K. Tensile strength comparison of presoldered and
dered joints principally fractured cohesively through
postsoldered joints. J Prosthet Dent 1986;55:23-7.
the solder and generally had few voids when observed 10. Lorenzana RE, Staffanou RS, Marker VA, Okabe T. Strength properties of
with the SEM. soldered joints for a gold-palladium alloy and a palladium alloy. J Prosthet
Dent 1987;57:450-4.
Although all the alloys soldered well, producing joints
11. Staffanou RS, Radke RA, Jendresen MD. Strength properties of soldered joints
that were judged strong enough to resist intraoral forces, from various ceramic-metal combinations. J Prosthet Dent 1980;43:31-9.
SEM observations revealed that the fracture path of the 12. Rasmussen EJ, Goodkind RJ, Gerberich WW. An investigation of tensile
strength of dental solder joints. J Prosthet Dent 1979;41:418-23.
solder joints followed microscopic grooves produced by
13. Rosen H. Ceramic/metal solder connectors. J Prosthet Dent 1986;56:671-
the finishing procedures for the specimens. Another study 7.
should be performed to examine the relationships be- 14. Anusavice KJ, Okabe T, Galloway SE, Hoyt DJ, Morse PK. Flexure test evalu-
ation of presoldered base metal alloys. J Prosthet Dent 1985;54:507-17.
tween finishing procedures and strength of the solder
15. Council on Dental Materials, Instruments and Equipment. Revised Ameri-
joints. can National Standard/American Dental Association Specification No. 5
The general composition of dental solders is reported23 for Dental Casting Alloys. Chicago: American Dental Association; 1988.
16. Popov EP. Introduction to mechanics of solids. Englewood Cliffs (NJ):
to consist primarily of the elements gold, silver, and cop-
Prentice-Hall; 1968. p. 25-7, 181-5, 186-8.
per. However, as previously noted, no copper was de- 17. Sokal RR, Rohlf FJ. Biometry. 3rd ed. New York: WH Freeman; 1995. p.
tected in the two solders used in this study (Table V). 260-5, 396-401.
18. Conover WJ. Practical nonparametric statistics. 2nd ed. New York: John
Other components, such as zinc (found in both solders),
Wiley; 1980. p. 294-338.
tin, and phosphorus, are included to reduce the fusion 19. Conover WJ, Iman RL. Rank transformation as a bridge between parametric
temperature and improve flow. and nonparametric statistics. Am Stat 1981;35:124-9.

JUNE 1998 683


THE JOURNAL OF PROSTHETIC DENTISTRY CHAVES ET AL

20. Welsch RE. Stepwise multiple comparison procedures. J Am Stat Assoc 29. Stade EH, Reisbick MH, Preston JD. Preceramic and postceramic solder
1977;72:566-75. joints. J Prosthet Dent 1975;34:527-32.
21. Reisbick MH, Brantley WA. Mechanical property and microstructural varia-
tions for recast low-gold alloy. Int J Prosthodont 1995;8:346-50. Reprint requests to:
22. Stewart RB, Gretz K, Brantley WA. A new high-palladium alloy for implant- DR. WILLIAM A. BRANTLEY
supported prostheses. [Abstract no. 423.] J Dent Res 1992;71:158. COLLEGE OF DENTISTRY
23. Craig RG, editor. Restorative dental materials. 9th ed. St Louis: Mosby; 1993. THE OHIO STATE UNIVERSITY
p. 402-7. 305 WEST 12TH AVE.
24. Carlberg T, Wictorin L. Soldering of dental alloys under vacuum by IR-heat- ROOM 3005-L POSTLE HALL
ing. Dent Mater 1986;2:279-83. COLUMBUS, OH 43210-1241
25. Cheng AC, Chai JY, Gilbert J, Jameson LM. Investigation of stiffness and
microstructure of joints soldered with gas-oxygen torch and infrared meth- Copyright © 1998 by The Editorial Council of The Journal of Prosthetic Den-
ods. J Prosthet Dent 1994;72:8-15. tistry.
26. Wictorin L, Fredriksson H. Microstructure of the solder-casting zone in 0022-3913/98/$5.00 + 0. 10/1/88849
bridges of dental gold alloys. Odont Rev 1976;27:187-96.
27. Tehini GE, Stein RS. Comparative analysis of two techniques for soldered CONTRIBUTING AUTHOR
connectors. J Prosthet Dent 1993;69:16-9.
28. Cattaneo G, Wagnild G, Marshall G, Watanabe L. Comparison of tensile William M. Johnston, PhD, Professor, Section of
strength of solder joints by infrared and conventional torch technique. J
Restorative Dentistry, Prosthodontics and Endodontics,
Prosthet Dent 1992;68:33-7.
College of Dentistry, The Ohio State University.

Noteworthy Abstracts Microleakage of dentin-bonded crowns placed with differ-


of the ent luting materials
Current Literature Patel S, Saunders WP, Burke FJT. Am J Dent 1997;10:179-83.

Purpose. Microleakage at the margins of dentin-bonded crowns may be a cause of failure of these
types of restorations. This in vitro study assessed the microleakage of dentin-bonded porcelain
crowns placed with three luting materials.
Material and Methods. Forty-five teeth were prepared by reducing the occlusal surface by 2 mm;
removing the convexity from the mesial, buccal, distal, and lingual surfaces; and forming a knife-
edge finish line. One half of the cervical margin was placed on dentin/cementum and the remain-
der of the margin was placed on enamel. Individual crowns were fabricated in feldspathic porcelain
and then cemented to the teeth with three luting materials, according to manufacturers recommen-
dations. The teeth were stored in water for 2 weeks at 37° C before evaluating the microleakage
with a dye penetration technique. Teeth with crowns were immersed in a 2% solution of methylene
blue for 48 hours at 37° C. Specimens were then thermocycled while in the dye for an additional 6
hours. The teeth were then sectioned in a bucco lingual plane and examined at ×10 magnification
with a ranking criteria ranging from 0 = no leakage to 4 = dye penetration grater than two thirds of
the length of the restoration-tooth interface.
Results. All groups exhibited leakage at the dentin and enamel margins and at the restoration-
cement interface. There were significant differences in leakage at the enamel margins for the three
groups but not at the margins of the finished dentin.
Conclusions. The resin cement combinations were unable to prevent microleakage completely.
Although there was no significant difference in leakage at the margin in dentin for all three materi-
als, Mirage ABC/FLC did not perform as well at the enamel margin. 30 References. — ME
RAZZOOG

684 VOLUME 79 NUMBER 6

Vous aimerez peut-être aussi