Académique Documents
Professionnel Documents
Culture Documents
Statement of problem. Little information is available on the optimum technique for soldering high-
palladium alloys, which have gained considerable popularity for prosthodontic applications.
Purpose. The objective of this study was to compare the flexural stress at the proportional limit of four
noble dental alloy specimens soldered with torch, oven, and infrared techniques.
Materials and methods. The high-palladium alloys studied were Legacy XT (Jelenko), Freedom Plus
(Jelenko), and IS 85 (Williams/Ivoclar). A gold-palladium alloy, Olympia (Jelenko), served as the control.
Thirty round bars, 18 × 3 mm, were cast from each alloy, cut in half, aligned, and joined using Olympia
Pre solder (Jelenko) for the gas-oxygen torch and the infrared technique and Alboro LF solder (Jelenko)
for the oven technique. Each soldered bar was subjected to three-point bending, and the maximum elastic
stress or strength of the solder joint was calculated at the proportional limit. Data were analyzed by two-
way ANOVA and the Ryan-Einot-Gabriel-Welsch (REGW) multiple range test at the 0.05 level of signifi-
cance.
Results. There was no significant difference between torch and oven-soldering, but both were signifi-
cantly different from the infrared technique. ANOVA showed a significant difference between alloys, but
this difference could not be detected with the REGW test. SEM examination of the fracture surfaces
revealed grooves associated with the path of crack propagation. X-ray energy-dispersive spectroscopic
analysis failed to detect copper in the solders, and there were no significant changes in the solder composi-
tions after the melting procedures.
Conclusions. All three techniques can yield satisfactory solder joints in high-palladium alloys. These
joints should be well-polished to achieve optimal strength. (J Prosthet Dent 1998;79:677-84.)
CLINICAL IMPLICATIONS
The results of this investigation indicate that the strengths of high-palladium alloy
joints properly prepared by the torch, oven, and infrared soldering techniques are ac-
ceptable for clinical use and have generally comparable values. The current results
also demonstrate the importance of having solder joints that are properly polished. The
apparent notch-sensitivity of solder joints in high-palladium alloy castings that con-
tain finishing grooves suggests that there should be some concern when cyclic loads are
applied over a clinically appropriate period of time.
tures (FPDs), the best adaptation was produced after mens according to ANSI/ADA Specification No. 5.15
soldering. Several studies have shown that, as the length The ends of the plastic patterns were cut with a dia-
of the prosthesis increases or involves the curvature of mond disk, leaving round bars 3 mm in diameter and
the arch, the potential for a poorly fitting prosthesis in- 20 mm in length. The specimens were then sprued with
creases.4-6 With the recent development of implant-re- wax (10 bars per casting ring) and invested with a car-
tained prostheses, which may often be long and may bon-free phosphate-bonded investment (High Span II,
need the use of increased amount of casting alloy, fabri- J.F. Jelenko and Co., Armonk, N.Y.). The manufacturer’s
cating a one-piece casting incorporating machined parts recommendations were followed for the special liquid-
and having these machined surfaces accurately and pas- to-powder ratio and burnout schedule for the invest-
sively fit the implant superstructure becomes even more ment.
difficult. As a result, the practitioner must have an ef- The high-palladium alloys used to cast the specimens
fective method of compensating for poorly fitting cast- were Freedom Plus, Legacy XT (Jelenko) and IS 85 (Wil-
ings. liams/Ivoclar, Amherst, N.Y.). A gold-palladium alloy
Soldering provides a means of improving the fit of a (Olympia, Jelenko) served as the control. The alloy com-
prosthesis and, at the same time, joining the components positions (wt%) are provided in Table I. Melting was
rigidly.7 The strength of solder joints is of utmost impor- performed with a multiorifice gas-oxygen torch, and the
tance for clinical success of the prostheses. Many studies alloys were cast with a broken arm centrifugal casting
have reported measurements of tensile strength for sol- machine. The castings were permitted to bench cool
der joints, with mixed results. Some investigators have before devestment. After air abrasion with 50 µm alu-
found that preceramic solder joints are stronger than minum oxide, castings were numbered at each end and
postceramic solder joints.8 Monday and Asgar9 and randomly assigned to treatment groups. Specimens were
Lorenzana et al.10 reported that there was no significant then sectioned at their midpoint with a low-speed dia-
difference in the ultimate tensile strength of postceramic mond saw (VR/50, Leco Corp., St. Joseph, Mich.) and
and preceramic solder joints. However, other authors11-13 water coolant. After sectioning, the bars were ultrasoni-
indicated that the ultimate tensile strength of postceramic cally cleaned with acetone and dried.
solder joints was greater than that of preceramic solder Corresponding halves of each casting were placed in
joints. Although tensile tests are useful in the evaluation a lathe for proper alignment. A thickness gauge was used
of soldering effectiveness, dental prostheses are subjected to provide a 0.5 mm gap. Autopolymerizing resin (GC
mainly to flexural loading during clinical use.14 A popular Corporation, Tokyo, Japan) was used to unite the halves
method to obtain the flexural strength of a material is of each bar. Each bar was then invested in a silicone
three-point bending, where a concentrated load is applied mold (Coltene, Whaledent Inc., Mahwah, N.J.) with Hi-
to the center of a uniform beam that is supported near Heat soldering investment (Whip Mix, Louisville, Ky.).
each end. The flexural strength is often termed the modu- The area to be soldered was left exposed during the in-
lus of rupture or transverse strength. vesting procedure, and an airway was created under the
In this study, torch, oven, and infrared soldering tech- soldering area to provide uniform heating of the joint.
niques were compared for three high-palladium dental The invested blocks were preheated according to the
alloys to determine the best method to join these met- alloy manufacturer’s instructions or those of the solder-
als. A well-known gold-palladium alloy served as a con- ing equipment manufacturer (infrared technique). Olym-
trol for the experiments. A literature review indicated pia Pre Solder (Jelenko) was chosen for the torch and
that there are no published studies comparing these sol- infrared-soldering groups, and Alboro LF solder
dering techniques for the high-palladium alloys. (Jelenko) for the oven technique groups.
For the torch-soldering, the investment blocks were
MATERIAL AND METHODS
preheated to 1100° F, and a gas-oxygen torch with a
The specimens were fabricated from polystyrene plas- No. 2 orifice point was kept moving at an oblique angle
tic patterns used for the fabrication of tensile test speci- to the preheated investment surface. A strip of approxi-
mately 5 mm of solder was placed over the joint, and ing moment (Fl/4 for three-point bending), I is the
the solder was allowed to flow through the gap. moment of inertia (πd4/64 for a circular cross-section),
The oven-soldering experimental groups were sub- and c is the radius of the bar.16
jected to the Vita Omega cycle (Vident, Baldwin Park, Ten test specimens were prepared for each combina-
Calif.) under vacuum for porcelain application (two tion of alloy and soldering technique for a total of 120
opaque bakes, two body bakes and a low glaze bake), specimens. A power analysis17 performed on the experi-
before the bar was cut with the diamond saw. After the mental data, assuming α = 0.05 and a power of 80%
blocks were preheated in the oven, a 5 mm strip of fluxed (β = 0.20), indicated that a sample size of 10 would
solder was placed in contact with the parts to be sol- show significant differences of 185 MPa in the mean
dered. Then the temperature in the oven was raised to proportional limit or fracture stress between sample
the melting temperature of the solder, allowing it to flow groups. Bartlett’s test17 was performed to determine the
through the joint space. The soldering procedure was homogeneity of the variances for all of the specimen
performed using the vacuum available with the porce- groups, and the mean values of flexural stress (propor-
lain furnace. tional limit or fracture stress) values for all the groups
For the infrared-soldering groups, the preheated were ranked and subjected to two-way analysis of vari-
blocks were placed on the platform of the infrared sol- ance (ANOVA) to examine the overall effects of alloy
dering machine (Ney Infrared Unit, J.M. Ney Co., and soldering technique, as well as their interactions.
Bloomfield, Conn.). The invested specimen and the sol- The mean values for the specimen groups were ranked,
der strip were fluxed, and this solder strip was placed in and two-way ANOVA on the ranks18,19 showed that there
contact with the parts to be soldered. The joint area was were highly significant differences between techniques.
placed 1 mm below the alignment rod, and the infrared The Ryan-Einot-Gabriel-Welsch (REGW) multiple
lamp was activated and held at maximum level until the range test20 with a significance level of α = 0.05 was
solder flowed. The soldered specimens were permitted used on the ranks to determine which specific specimen
to bench cool before devesting. The specimens were groups were significantly different from each other. The
cleaned of residual investment, and the solder joints were REGW test is considered to have less likelihood of Type
finished flush with the remainder of the bar. Mounted II statistical errors than the well-known Tukey multiple
stones and rubber wheels were used with a low-speed range test. In addition, representative fracture surfaces
dental handpiece, and the bars and the handpiece were were randomly selected from each group and examined
attached to the lathe. The diameter of each soldered joint with a SEM (JSM-820, Jeol Ltd., Tokyo, Japan).
was measured, and external defects for each specimen The two solders were also melted individually using the
were noted. three techniques, embedded in epoxy resin, ground with
The specimens were subjected to a three-point flex- 400 and 600 grit metallographic paper, and polished with
ural load in a screw-driven mechanical testing machine a series of alumina slurries (from 15 to 0.05 µm particle
(model 4204, Instron Corp., Canton, Mass.) at a cross- size). Quantitative elemental composition information was
head speed of 0.25 mm per minute. The load and posi- provided by x-ray energy-dispersive spectroscopic analyses
tion of the crosshead were recorded by means of a (EDS), using a Link eXL microanalysis system with a
Pentium personal computer, with the software program PentaFET detector and an ultrathin window (Oxford In-
Labview for Windows (National Instruments, Austin, struments Group, High Wycombe, England) coupled to
Texas) and a data acquisition card with a frequency of the SEM. A similar procedure was performed for the as-
1 Hz. The raw data from each test bar were stored in received unmelted solder, and the changes in the composi-
the computer as a text file. The software program Excel tions of the solders after the melting procedures were de-
(Microsoft Corporation, Redmond, Wash.) was used to termined.
plot and manipulate the data. The midspan flexural stress
was calculated at the proportional limit (just before the RESULTS
point of initial nonlinear deformation as determined from The mean values with standard deviations of the pro-
a graphic representation of load versus crosshead move- portional limit for ductile solder joints or the fracture
ment) or at the point of failure, whichever was lower. strength of brittle solder joints for the various specimen
The equation used for the calculation of the maximum groups are summarized in Figure 1. Bartlett’s test showed
elastic flexural stress (σmax) developed at the midpoint of that the variances were significantly different from each
a centrally loaded round beam was as follows: other. For this reason, the technique of ranking the ob-
σmax = 8Fl/πd3 servations was used. Two-way ANOVA on the ranks
showed that there were highly significant differences
where F = load, l = distance between support points (test between techniques and a significant difference between
span length of 12.5 mm) and d = diameter. This equa- alloys (Table II). There was no significant interaction
tion was developed from the well-known elastic flexure between the alloys and soldering techniques. The sensi-
formula σmax = Mc/I, where M is the maximum bend- tive REGW multiple range test showed that, even though
that separation was between the solder and alloy. An in- Table III. Ryan-Einot-Gabriel-Welsch test results for the alloys*
tergranular fracture surface can be seen as well, with Alloy Number of specimens Mean (MPa) REGW grouping
evident separation between some grains. This fracture
surface was characteristic of the solder. There was no Freedom Plus 30 902.40 A
evidence of microvoids or the dimpled rupture struc- Legacy XT 30 834.70 A
ture indicative of ductile fracture.21 IS 85 30 803.44 A
Olympia 30 800.64 A
Oven-soldered Olympia alloy specimens were also
randomly selected after fracture and observed at the same *Alloys with the same letter are not statistically different at the α =0.05 level.
range of magnifications used for the IS 85 specimens.
The fracture surfaces for these specimens were similar
to Figure 2, which would be expected because the same Table IV. Ryan-Einot-Gabriel-Welsch test results for the solder-
solder was used (Alboro LF), and the fracture mode was ing techniques*
again cohesive through the solder.
Technique Number of specimens Mean (MPa) REGW grouping
An area near the edge of an oven-soldered IS 85 speci-
men is depicted in Figure 3. This photomicrograph in- Torch 40 881.62 A
dicates that cohesive ductile fracture also occurred Oven 40 868.79 A
through the IS 85 alloy, because of the characteristic Infrared 40 755.48 B
dimpled rupture appearance resulting from the *Techniques with the same letter are not statistically different at the α = 0.05
microvoids. Previously, Stewart et al.22 found that this level.
alloy exhibited excellent ductility, as reported by the
manufacturer.
SEM photomicrographs of fractured solder joints for A low-magnification photomicrograph of an infrared-
Legacy XT and Freedom Plus specimens that had been soldered specimen of Olympia after fracture is shown in
oven-soldered showed characteristics similar to those of Figure 6. This photomicrograph shows that the princi-
the fractured solder joints for IS 85, which follows from pal path of fracture was in the alloy below the solder
the cohesive nature of the fracture process through the (white band in the center of the photomicrograph), i.e.,
same low-fusing solder. cohesive fracture through the alloy. Numerous voids in
The fracture of the torch-soldered and infrared-sol- the solder are also visible in Figure 6.
dered specimens visually appeared to have initiated at In the low-magnification photomicrograph (Fig. 4)
the tension side and propagated through the solder. Low- of a fractured Olympia specimen that had been torch
power SEM examination (Fig. 4) confirmed this obser- soldered, an indentation in the solder where the bend-
vation. For some of these specimens, fracture also oc- ing load was applied can be seen (center left edge of
curred through the parent metal, but no separation of photomicrograph). The width of the solder joint appears
the specimen halves was observed. In contrast, the oven- to be extended in a vertical direction near the right edge
soldered specimen halves of the alloys separated com- of the photomicrograph (toward the tension side in
pletely during testing. bending), due in part to crack propagation. However,
The fracture surface for a torch-soldered IS 85 speci- the indentation and this tapering suggest some ductility
men is depicted in Figure 5. Substantial porosity and a of the solder. These observations were also made for
rounded structure for the solder below a void are evi- another fractured specimen that had been soldered by
dent. Parallel grooves on the surface resulted from the the infrared technique.
experimental procedure of holding each soldered speci- X-ray EDS analyses were performed on the two sol-
men in the lathe and polishing with stone followed by ders to determine their compositions. It was found that
rubber wheel (with a handpiece). An association of the there was no copper in the composition of either solder.
fracture boundary with the polishing grooves is appar- The compositions of the solders before melting and af-
ent. This same pattern was observed on a representative ter the melting procedures with the three techniques
fracture surface of an infrared-soldered Legacy XT speci- are summarized in Table V. Only slight changes were
men. Again, there was a close association between the found in the compositions of the solders after the melt-
polishing grooves and the path of fracture in the solder. ing procedures.
Fig. 5. SEM photomicrograph of torch-soldered IS 85 speci- Fig. 6. SEM photomicrograph of fracture path through parent
men demonstrates solder porosity. (Bar = 10 µm.) metal of infrared-soldered Olympia specimen. (Bar = 100 µm.)
Alboro LF* 62.9 (0.1) 23.2 (0.5) 13.5 (0.7) — 0.4 (0.1)
Olympia Pre* 70.5 (0.3) 16.5 (0.5) 0.9 (0.2) 10.9 (0.4) 1.1 (0.1)
Alboro LF (oven)** 62.3 (0.6) 24.1 (0.6) 12.1 (0.5) 0.9 (0.4) 0.6 (0.2)
Olympia Pre (infrared)** 72.1 (0.1) 16.7 (0.4) — 10.3 (0.6) 0.9 (0.1)
Olympia Pre (torch)** 70.9 (0.8) 16.7 (0.5) — 10.3 (0.2) 1.3 (0.4)
DISCUSSION
It has been previously noted that a power analysis
The development of the strength for a solder joint showed that the sample size of 10 used in this study
involves melting, flowing, and wetting of the solder by would be able to detect a significant difference (α = 0.05)
capillary forces between the parent metal and the sol- between specimen groups of approximately 185 MPa
der.23 All three techniques used in this study were con- with a power of 80% (β = 0.20). This difference was
sidered to be acceptable for soldering the three repre- approximately 28% greater than the pooled standard
sentative high-palladium alloys. It is important to con- deviation of 144 MPa for all 12 specimen groups. For
sider the expertise of the operator for achieving solder example, sample sizes of 15 and 41 would have been
joints of similar characteristics. The problems that have required to show significant differences of 150 MPa and
been reported with the commonly used gas-oxygen torch 90 MPa, respectively, at the same α and β levels, and
for soldering metal ceramic alloys are gas inclusions, would have necessitated much larger numbers of 180
voids, improper melting of the solder, and excessive oxi- and 492 soldered specimens for the entire study. Al-
dation of the pieces to be soldered.24 The main advan- though a larger sample size might have revealed statisti-
tages with this technique are the availability of the gas- cally significant differences between some additional
oxygen torch in the dental laboratories and the flexibil- sample groups compared with the results obtained with
ity of this method for use with the preceramic and the current sample size, the general observations and
postceramic solders. clinical recommendations from this study would not be
It was initially expected that the torch soldering would substantially altered.
exhibit the greatest variation in results because of the rela- An alternative approach to the conventional torch or
tively poor control in the temperature during the solder- oven soldering is the use of an infrared heat source. In
ing procedure. However, this result was not always found. the commercially available unit (Ney) used in this study,
For example, the flexural strength of the Olympia speci- infrared energy for soldering is supplied by a quartz-
mens that were torch soldered had one of the smaller iodine-tungsten-filament lamp. Similar to the gas-oxy-
standard deviations of all the groups (Fig. 1). gen torch, flux is required with the solder in this tech-
20. Welsch RE. Stepwise multiple comparison procedures. J Am Stat Assoc 29. Stade EH, Reisbick MH, Preston JD. Preceramic and postceramic solder
1977;72:566-75. joints. J Prosthet Dent 1975;34:527-32.
21. Reisbick MH, Brantley WA. Mechanical property and microstructural varia-
tions for recast low-gold alloy. Int J Prosthodont 1995;8:346-50. Reprint requests to:
22. Stewart RB, Gretz K, Brantley WA. A new high-palladium alloy for implant- DR. WILLIAM A. BRANTLEY
supported prostheses. [Abstract no. 423.] J Dent Res 1992;71:158. COLLEGE OF DENTISTRY
23. Craig RG, editor. Restorative dental materials. 9th ed. St Louis: Mosby; 1993. THE OHIO STATE UNIVERSITY
p. 402-7. 305 WEST 12TH AVE.
24. Carlberg T, Wictorin L. Soldering of dental alloys under vacuum by IR-heat- ROOM 3005-L POSTLE HALL
ing. Dent Mater 1986;2:279-83. COLUMBUS, OH 43210-1241
25. Cheng AC, Chai JY, Gilbert J, Jameson LM. Investigation of stiffness and
microstructure of joints soldered with gas-oxygen torch and infrared meth- Copyright © 1998 by The Editorial Council of The Journal of Prosthetic Den-
ods. J Prosthet Dent 1994;72:8-15. tistry.
26. Wictorin L, Fredriksson H. Microstructure of the solder-casting zone in 0022-3913/98/$5.00 + 0. 10/1/88849
bridges of dental gold alloys. Odont Rev 1976;27:187-96.
27. Tehini GE, Stein RS. Comparative analysis of two techniques for soldered CONTRIBUTING AUTHOR
connectors. J Prosthet Dent 1993;69:16-9.
28. Cattaneo G, Wagnild G, Marshall G, Watanabe L. Comparison of tensile William M. Johnston, PhD, Professor, Section of
strength of solder joints by infrared and conventional torch technique. J
Restorative Dentistry, Prosthodontics and Endodontics,
Prosthet Dent 1992;68:33-7.
College of Dentistry, The Ohio State University.
Purpose. Microleakage at the margins of dentin-bonded crowns may be a cause of failure of these
types of restorations. This in vitro study assessed the microleakage of dentin-bonded porcelain
crowns placed with three luting materials.
Material and Methods. Forty-five teeth were prepared by reducing the occlusal surface by 2 mm;
removing the convexity from the mesial, buccal, distal, and lingual surfaces; and forming a knife-
edge finish line. One half of the cervical margin was placed on dentin/cementum and the remain-
der of the margin was placed on enamel. Individual crowns were fabricated in feldspathic porcelain
and then cemented to the teeth with three luting materials, according to manufacturers recommen-
dations. The teeth were stored in water for 2 weeks at 37° C before evaluating the microleakage
with a dye penetration technique. Teeth with crowns were immersed in a 2% solution of methylene
blue for 48 hours at 37° C. Specimens were then thermocycled while in the dye for an additional 6
hours. The teeth were then sectioned in a bucco lingual plane and examined at ×10 magnification
with a ranking criteria ranging from 0 = no leakage to 4 = dye penetration grater than two thirds of
the length of the restoration-tooth interface.
Results. All groups exhibited leakage at the dentin and enamel margins and at the restoration-
cement interface. There were significant differences in leakage at the enamel margins for the three
groups but not at the margins of the finished dentin.
Conclusions. The resin cement combinations were unable to prevent microleakage completely.
Although there was no significant difference in leakage at the margin in dentin for all three materi-
als, Mirage ABC/FLC did not perform as well at the enamel margin. 30 References. — ME
RAZZOOG