Académique Documents
Professionnel Documents
Culture Documents
Contents
Contents..................................................................................................................... 1
1.0 AN INTRODUCTION TO LEVEL/ INTERFACE MEASUREMENT...................................2 2.0 INSTRUMENT RANGING VS CALIBRATION...............................................................3 3.0 ARCHIMEDES PRINCIPLE AND DISPLACERS ..........................................................5 4.0 RANGING DISPLACER LEVEL TRANSMITTERS: MASONEILAN DRESSER 12300........8 5.0 RANGING INTERFACE LEVEL TRANSMITTERS: MASONEILAN DRESSER 12300......10 6.0 SPECIAL CASES...................................................................................................11 7.0 A LAST WORD.....................................................................................................12
1. Water level in a bucket could be determined. 2. If the same bucket contained water and oil, the level of water could still be determined. Case 1 would require level measurement. Case 2 would require the interface level between water and oil. The oil would float on the water with an interface between water and oil. These two types of measurements define level (1) and interface (2) measurement and are readily available in process industries. An example of industrial interface measurement is found in separators for oil and gas production where interface measurement indicates the water level in the separator vessel. Figure 1. Shows the contents of a typical 3 phase separator
LRV = 0m URV = 10m Since different applications require different LRV and URV, ranging of instruments is left to the user. Measuring the level of kerosene in a tank 2m height is different application from measuring level of palm oil in tank of 5ft height. Ranging is to define the limits to which your instrument will measure. Usually, instruments have their own range as specified by their manufacturer. An instrument may be capable of measuring up to a maximum of 2 ft level but it can be ranged for to measure 1ft level maximum. Calibration, on the other hand, is to make sure that the input/output from your instrument transmitter is as correct as possible. Instrument calibration should be performed periodically by instrument technicians to ensure the accuracy and precision of instruments. A sensor may become fatigued after years of use, thus, degrading in accuracy. A pressure sensor may read 10.7 bars from an actual pressure of 11 bars. In essence, this sensor requires calibration if this deviation of 0.3 bars is still within the tolerance stated in the sensor datasheet. If 0.3 bar pressure is beyond the tolerance, then its high time to replace the sensor.
The same philosophy applies to transmitter outputs. Consider measuring maximum 2ft liquid level, the transmitter output is 4mA when the tank is empty and 20 mA when tank is 2ft full of liquid. In reality, however, there is always a deviation between the ideal and the actual output. It could be 4.1mA instead of 4mA at your low range value (LRV). The deviation should not exceed the stated value on the instrument manufacturer data sheet after calibration. For more reading on instrument calibration, visit www.isa.org to order your copy of Calibration: A technicians guide from International society for Automation. The point to note here is the difference between instrument ranging and calibration. Ranging defines the application while calibration checks the transmitter output accuracy. In analog instrument transmitters, calibration and ranging are done at the same time. This is not so for digital instrument transmitters. I have been confused over these concepts in my beginning years as an instrument technician. As such, I believe it is worth mentioning here. Whenever, you are required to install an instrument, please simply range it for your application. Most of the time, these instruments are already factory calibrated so you need not bother yourself. From my experience, ranging could be quite tricky for level/ interface applications. Level depends on what you have defined. A vessel may have a 4
height of 1.5 m but its measured level of interest may be between 0.8 and 1.5 m. Just 0.7m!!!! Sometimes, the instrument is mounted through an extended nozzle from the tank or vessel as shown below
FIG. 2: Nozzle and Cage Level Assembly Two nozzles extend from the tank, Upper and lower. The range is, as shown, between the top of the lower nozzle to the bottom of the upper nozzle. The nozzle and cage assembly is a form of the U tube. The top nozzle ensures that the pressure on both the cage and the tank are equal thereby leading to equal levels in tank and cage. I have seen a lot of instrument technicians get confused over what the true level range is a couple of times. The truth is if you dont know your range, who will? It will certainly be the beginning of your troubles and frustration. Before reading, please ensure that you understand what your expected level range is before attempting work on level/ interface level instruments.
Fig.4: Displacer measuring level in a vessel (Nozzle and Cage) Measuring level/ interface using displacers is based on 2 equations; one for liquid level and the other for liquid interface. From Archimedes principle, 6
Weight loss by displacer (Wd) = Weight of liquid displaced (Wl) .. 1 Wd = mg .2 Where m is equal to the mass of the liquid displaced and g is the acceleration due to gravity. But Mass = density X Volume = V 3 Putting (3) into (2) Wd = gV .. 4 Where is the density of the liquid displaced, g is acceleration due to free fall, and V is the volume of the liquid displaced. The multiplication between density and acceleration due to gravity gives the weight density. So (4) can be written as Wd = V . 5 The displacer in question is usually cylindrical as shown in figure 4. Hence, the volume in question is that of a cylinder. Such that Wd = (r2L) = r2 L . 6 Equation 6 is the guiding relationship for LIQUID LEVEL MEASUREMENT using displacers. You can see from equation 6 that there is a linear relationship between the weight loss experienced by the displacer and the level of liquid measured. More liquid level submerging the displacer will result in more weight loss experienced by the displacer. Equation 6 can be expanded for liquid interface measurement. Here, there are 2 liquids involved and consequently 2 weight densities involved too. So,
Wd = r2 (1L1 + 2L2) . .. 7 Equations 6 and 7 will guide our ranging methods for displacers. They are critical equations which must be understood by instrument technicians working with displacers. Note that in equation 7, the interface level is either L1 or L2 and not both. When L1 is increasing, L2 is decreasing and otherwise. The interface level is actually the level of the liquid with the higher weight density.
Fig. 5: Makeshift sight glass 4. Open transmitter terminal and connect hart communicator. Verify proper mounting configuration. Use flow chart below MAN CONFIG BASIC SETUP RIGHT/ LEFT
Right or left means the relative position of the transmitter head to the displacer. If the displacer transmitter is left of the displacer, then choose left.
5. MAN
Make sure the transmitter is configured to measure level not interface. Use the flow chart below CONFIG BASIC SETUP LEVEL.
6. Get to the submenu that will allow you change the specific gravity value to that of the liquid you wish to measure. Use the flow chart below MAN 7. CAL VAR SET CHG SER.
If you know the specific gravity of the liquid whose level you want to measure, then input it here under CHG SER. This means change service specific gravity. Changing the specific gravity in equation 6 is the same thing as matching the displacer loss in weight to the corresponding liquid level. 9
8. Fill the cage with the liquid of interest to the 100% percent range value(URV) while observing from the sight glass. Does your transmitter register 100% level? If yes, you have successfully ranged your transmitter. If no, then go back to the CHG SER submenu and try a different specific gravity value. From equation 6, you should know your direction of change. If your transmitter reads a higher value, say 120%, then you need to increase your SG to bring this value towards 100%. 9. A zero check is advisable. Drain your cage and see if your transmitter records 0%. If zero is not achieved, then you need to perform a specific gravity calibration according to the instrument manual. After SG calibration you need to start from step 1 again to be sure your transmitter reads the right range of your application.
3. Get to the submenu that will allow you change the specific gravity value to that of the liquid you wish to measure. Use the flow chart below MAN CAL VAR SET CHG SER.
4. Put in the specific gravity values for both the lower and upper liquids. Changing the specific gravities in equation 7 is the same thing as matching the loss in weight of the displacer to the interface level. Note that the interface level is the level of the liquid with the higher specific gravity.
5. Fill the cage with the lower SG liquid to full range. Does your transmitter read 0%? If yes, then the LRV ranging is successful. If not, then change the lower SG to achieve a zero percent from the transmitter. If the transmitter reads below 0%, then you should decrease your SG in small increments . The direction of the change is determined from equation 7. 6. Fill the cage with the higher SG liquid to full range. Does your transmitter read 1000%? If yes, then the URV ranging is successful. If not, then change the 10
lower SG to achieve a 100 percent from the transmitter. The direction of the change is determined from equation 7. 7. Repeat steps 5 and 6 at least once to be sure your range is correctly executed.
2. EMULSIONS: Emulsions occur when two liquids mix together. In this case, the resultant liquid combines the specific gravities of the 2 liquids. Interface applications using displacers suffer from emulsions. Practically, displacer transmitters specify the minimum difference between the lower and higher specific gravities. For masonelian dresser transmitters, the least difference is about 0.1. The transmitter goes to high alarm or unpredictable response if this difference goes smaller than that. A vessel with a layer of emulsion contains 2 interfaces which confuses your transmitter to start spiking between the LRV and URV. Make sure there are no emulsions in your vessel before attempting to range your displacer transmitters!!! Whenever emulsions appear in your vessel, displacer transmitters move out of control. So watch out for these symptoms on Oil and water separators.
11
3. EMPTY UPPER NOZZLE: Displacers required to measure interface level must be completely immersed in both liquids. The upper nozzle must be completely filled with the lower SG liquid else the displacer cannot differentiate between a low interface level and an empty upper nozzle. I have once ranged an interface level transmitter expected to measure the interface level between fuel oil condensate and glycol. Guess what? After the job, the transmitter measured falsely because there was no condensate in the vessel. Condensate was expected to flash off from the glycol in the running process. But this never happened!!!! Its important that you pay close attention also to the process in which your displacer is a part of.
12
Archibong A., is an instrument technician with a 5 year experience in both manufacturing and oil and gas industries. He is currently pursuing a CCST/ CAP with international society for Automation.
13