NOTE:
PANADENT LTD
149, Sevenoaks Way, Orpington, Kent, Br5 3AQ, ENGLAND
Tel. +44 (0) 1689 881788 - Fax. +44 (0) 1689 881789
E-mail: info@panadentltd.co.uk web: www.panadent.net
01 - Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 01
02 - Standard and Professional Face-Bow registration procedure . . . . . . . . 02
2.1 - Dentate and partially dentate patients . . . . . . . . . . . . . . . . . . . . . . 02
2.2 - Totally edentureless patient . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 04
03 - Mounting the casts on the Articulator . . . . . . . . . . . . . . . . . . . . . . . . . . . 05
3.1 - Preparing the Articulator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 05
3.1.1 - Articulator Model 4000. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 05
3.1.2 - Articulator Model 2000. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 06
3.1.3 - Articulator Model A7 Plus . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 06
3.1.4 - Articulator Model A7 Fix. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 06
3.1.5 - Articulator Model EVA Plus . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 07
3.1.6 - Articulator Model EVA Fix . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 07
3.2 - Mounting the upper cast . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 08
3.2.1 - Standard Face-Bow. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 08
3.2.2 - Professional Face-Bow . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 09
3.3 - Interocclusal Registration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
3.4 - Mounting the Mandibular Cast . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
04 - Special features of the Bio-Art Articulators . . . . . . . . . . . . . . . . . . . . . . . 13
05 - Special features of the Standard and Professional Face-Bows . . . . . . . 14
06 - Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
07 - Safety Rules . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
08 - Technical Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
09 - Garantee. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
10 - Part List Articulator 2000. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
11 - Part List Articulator 4000. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
12 - Part List Articulator A7 Plus . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
13 - Part List Articulator A7 Fix . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
14 - Part List Articulator EVA Plus . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
15 - Part List Articulator EVA Fix . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
16 - Part List Standard Face-Bow . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
17 - Part List Professional Face-Bow. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
1 - INTRODUCTION
IMPORTANT NOTE: Before using your BIO-ART articulator, please read carefully all the
Instructions Manual.
The face-bow is an instrument used to register the position of the patient's dental arcade in
relation to the skull and transfer this record to the articulator.
The BIO-ART semi-adjustable articulators are Arcon type conventional equipment (models
2000, 4000, A7 Plus and A7 Fix) and Non-Arcon type (models EVA Fix and EVA Plus).
Therefore, the mounting technical procedures are the same for these articulators.
This Manual contains only basic and simplified information regarding the use of the
equipment, which do not replace the need of a specific course of Occlusion and/or
Prosthesis, offered in many Teaching Institutions and Dental Faculties, as a compulsory
discipline.
IMPORTANT NOTE: The use of the equipment by professionals not qualified can cause
damage or harm to the product and/or patient.
The use of the semi-adjustable articulator and face-bow is therefore recommended for most
prosthetic, occlusal and rehabilitation work. The technique is simple, fast and easy, offering
highly satisfactory results for both the patient and professional.
01
2 - STANDARD AND PROFESSIONAL FACE-BOW
REGISTRATION PROCEDURE
2.1 - DENTATE AND PARTIALLY DENTATE PATIENTS
b) Position the bite fork so that the midline of the fork handle
is aligned with the midline of the maxilla and place it on
FIG. 1 the upper teeth, holding it firmly in place until the
registration material hardens (fig. 2). Only a small
amount of registration material should be used since the
purpose is to record only the cusp tips of the teeth while
keeping the fork as immobile as possible (fig. 3).
Afterwards it is advisable to try the cast on the
registration to check its stability (absence of clearance)
(fig. 4).
Note: Alternatively, a pre-impression of the teeth can be
FIG. 2 made on the upper model and then refined in the mouth.
In the case of patients who have partially jagged teeth, it
is important to locate points that hold the fork stably in
place, despite the missing teeth and the points used for
the transfer.
FIG. 4
02
d) Place the nasion relator on the face-bow cross bar, when
using the face-bow Standard or on the “W” of the face-
bow, when using the Professional Face-Bow. For both
models, centre the nosepiece on the patient's nasion. The
earpieces should now be carefully positioned as deeply
as possible in the patient's auditory meatus and the
nosepiece should be gently pressed against the patient's
nasion while tightening the wing nuts of the nazion relator
assembly (fig. 8). Afterwards, tighten the three screws of
the Standard face-bow or the central screw when using
the Professional model (fig. 9 and 14).
Central Screw
Intercondylar
distance
Intercondylar Distance
Distance Indicator STANDARD FACE-BOW PROFESSIONAL FACE-BOW
FIG. 12 FIG. 13 FIG. 14
03
g) Loosen the wing nut of the nosepiece and remove the nazion relator assembly from the
face-bow. Then, loosen the central wing nut of the face-bow and hold the cross bar of the
bow at the same time the patient opens the mouth slowly, removing the whole set carefully
(fig. 15, 16 and 17)
a) When registering totally edentureless patients with the face-bow, use the special
edentureless fork, or Conti fork (optional), to maintain the prove plate and roller wax.
b) After the patient's registration on the wax roller, heat both handles of the fork slightly in a
flame and press them on the upper wax roller, which was previously joined to the lower
wax roller, or not, with the midline of the fork handle aligned with the midline of the patient
(fig. 18, 19).
c) Place the set (fork + wax rollers) in the patient's mouth inserting it on the alveolar edges
(fig. 20)
Note: When carrying out the registration with wax rollers which were previously joined,
ask the patient to keep it in position and close his mouth.
04
3 - MOUNTING THE CASTS ON THE ARTICULATOR
Adjust the condylar guides at a 50º angle and the immediate side shift at 0 (zero) millimeters.
Close the central lock assuring that there are no lateral movements and fasten the condylar
guide elastic in the lower frame (fig. 26, 27 and 28).
PS.: In order to make sure that the Articulator A7 Plus is totally locked at the centric position,
check if the 2 locks (left and right) are at the locked position.
06
3.1.5 - Articulador Modelo EVA Plus:
Adjust the condylar guide at a 30° angle and the “Benett” angle at a 0° (zero) (fig. 35 and 36).
Tighten the right and left screws, assuring that there are no lateral movements (fig. 37)
Tighten the right and left screws, assuring that there are no lateral movements (fig. 38)
FIG. 38
IMPORTANT NOTE:
The horizontal slide bar lock screw of the Articulators Model EVA PLus and EVA Fix (see this
item on the Part List), are used to calibrate the equipment at the plant and should not be
unfasten or removed. This procedure can disadjust the equipment.
07
3.2 - MOUNTING THE UPPER CAST
08
3.2.2 - Professional face-bow
The main feature of this model is the Jig Transfer Assembly for mounting the casts on the
articulator (fig. 45). This assembly is placed in the lower frame of the articulator through the
Jig assembly lower base thus eliminating the need for the face-bow arm while transferring the
patient's registration onto the articulator and providing quick mounting of the cast, easy
handling, greater stability, precision and reliability.
FIG. 45
a) Remove the incisal guiding table from the lower frame of the articulator and insert the jig
transfer lower base, assuring that the guide pin is properly touching the end of the slot
(reference point) (fig. 46,47 and 48).
b) Remove the jig transfer assembly from the face-bow by releasing its respective wing nut,
connect it to the jig transfer lower base and fasten the wing nut (fig. 49 and 50).
c) Insert the upper frame support on the jig transfer connection rod, supporting the upper
frame and assuring the parallelism among the articulator frames (fig. 51).
09
d) Place the upper cast, with retentions and previously
hydrated, on the fork registration. Lift the Upper Frame of
the articulator, depositing a small amount of plaster on the
Upper Mounting Plate and some on the top of the Upper
Cast (fig. 52). Then, using one of your hands, keep the fork
and the cast in position, avoiding any movement of the
Bite Fork and carefully hinge the Upper Frame until it
Fork support (optional) touches the support. Wait for the plaster to harden.
FIG. 52
10
3.4 - MOUNTING THE MANDIBULAR CAST
FIG. 60
11
3.5 - FEATURES OF THE RAIL MOUNTING PLATE
The possibility of reusing the Mounting Plate without breaking the model requires care with the
side finishing of the plaster on the plate, i.e., the plate should not have plaster at the lateral
border (picture 61).
In order to take off the model, the plaster should be removed of from the retention hole (picture
62) and then the model should be hold while tapping the plate towards the indicated arrows
located on the lower face of it (picture 63).
PS.: Once the model has been taken off the Mounting Plate, we can not guarantee the
returning of it in the initial position with precision, since it depends on the technical quality of the
work and the materials used (especially the plaster).
12
4 - SPECIAL FEATURES OF THE BIO-ART ARTICULATORS
Model
2000 4000 A7 Fix A7 Eva Fix Eva Plus
Features
Classification Arcon Arcon Arcon Arcon No-Arcon No Arcon
Curve Flat Adjustable / Flat
Condylar Guide Adjustable / Curve Adjustable / Flat Adjustable / Curve
Fixed at 30° Fixed at 30°
Bennet Angle Fixed at 15° Adjustable Fixed at 15° Adjustable Fixed at 15° Adjustable
Intercondylar Distance Fixed at 110 mm Adjustable * Fixed at 110 mm Fixed at 110 mm Fixed at 110 mm Fixed at 110 mm
Central Lock Yes No Yes Yes Yes Yes
Rubber Ring (O'ring) Spring in the Spring in the
Stabilization System ** Central Spring Magnetic *** Silicone Connection
condyle guide condyle guide
Immediate Side Shift Adjustable No No No No No
Protrusion Adjustment No No No No No Yes
* Adjustment made through the micrometrical expanding spindle. (Bio-Art's exclusive system)
** Joining system between the upper and lower frames provides stability during articulating movements.
Mod. 2000 Mod. 4000 Mod. A7 Fix Mod. A7 Mod. EVA Fix Mod. EVA Plus
Interchangeable Articulators (Optional)
Bio-Art's interchangeable articulators are standardized (calibrated) by the manufacturer to allow for interchangeability of the casts among these
articulator models. Only models 2000, 5000 and EVA can be supplied in the standardized option.
13
5 - SPECIAL FEATURES OF THE STANDARD AND
PROFESSIONAL FACE-BOWS
The Standard Face Bow can be used with the all Articulators
models, its use is forbidden with the Models EVA Fix and EVA
Plus.
The main feature of this model is the Jig Transfer Assembly for
mounting the casts on the articulator, eliminating the need for
the face-bow arm while transferring the patient's registration,
providing greater quality, stability, precision and reliability (ref.
chapter 3, item 3.2.2). The design of the face-bow ensures
more stability concerning the opening and closing movement
in relation to the fork, or in other words, the fork is fixed when
the face-bow moves. The auricles are anatomic, removable
and chemically sterilized ensuring comfort for the patient.
FIG. 64
FIG. 65
14
6 - ACCESSORIES
7 - SAFETY RULES
• This manual only contains basic and simplified information regarding the use of the
instrument and under no circumstances it is a substitute for a proper training course. The
articulator and face-bow are products aimed at the exclusive use of qualified dental
professionals.
• The Bio-Art face-bow is designed for use with Bio-Art's own articulator and vice-versa.
Although, the Standard face-bow is compatible with other articulator models, its use with
similar models should be consulted. The Professional face-bow is exclusively designed for
use with the Bio-Art articulator. Therefore, Bio- Art does not guarantee precision when its
face-bows are used with other products.
• Bio-Art articulators are not interchangeable, i.e. the casts mounted on one articulator
model should not be transferred (mounted) to any other articulator. Therefore, Bio-Art does
not guarantee precision when casts mounted on one articulator are transferred to another.
• The only Bio-Art articulators, which are interchangeable, are those which are calibrated in
the factory. Thus, for safe and precise interchangeability, casts should be transferred only
between calibrated articulators.
• Before using the articulator and face-bow, the professional should check the instrument for
possible damage, distortion of the incisal pin, face-bow's parallelism, proper centricity of
the incisal pin in relation to the lower frame, etc. Should any abnormality be found, Bio-Art's
Technical assistance or Authorized Representative should be contacted immediately.
• The face-bow should be cleaned with warm water prior to its use. The Bite Fork must be
autoclaved and the Earpieces disinfected with a bactericide or with 70° alcohol.
• Bio-Art recommends only warm water for the general cleaning of the articulator and face-
bow.
• When using “godiva” or any other registration material, special care should be taken to
avoid excessive heating of the material, which might burn the patient's mouth. Bio-Art
urges the user to carefully follow the instructions of the registration material manufacturer.
• Because the articulator is a precision instrument, it should be handled, transported and
stored carefully. Bio-Art provides a plastic case (optional accessory) exclusively designed
to accommodate the articulator and to facilitate its transportation.
15
8 - TECHNICAL SERVICE
For your safety, the technical service of this product should be made by people/companies
authorized by BIO-ART. Please contact us or the distributors, where the product was
acquired.
Our web site is: www.bioart.com.br
9 - GUATANTEE
BIOART Dentistry Equipment Ltd guarantees the ARTICULATOR for one year. This
guarantee covers any manufacturing fault in the case of repairing the product.
a) The product had to have been used correctly according to the instructions. We
remind you that the ARTICULATOR should be handled, transported and stored
carefully. Dropping or knocking the device means bad use and this will result in
losing the guarantee.
b) The receipt of the purchase of the product must be attached to the complaint
To make use of this guarantee, the consumer should send the product (at his/her own
cost) to the nearest authorized BIOART Technical Assistance or to the following
address:
Natural wear and tear of the product, as well as damage caused by transportation
(both ways) are excluded from this guarantee.
16
10 - PART LIST - ARTICULATOR 2000
Item Qt Code Description
01 2 SCRE0008 Condylar Guide Wing Screw
02 2 FFIX0057 Condylar Guide Pin
03 1 FCTZ0068 Centric Pin
04 1 FCDL0072 Condylar Guide (Right)
05 2 FPIG0062 Face Bow's Guide Pin
06 2 CPAR0231 Elastic's Screw of Condylar Guide
07 4 CORI0029 Condylar Guide Elastic
08 2 FECD0049 Condyle Element
09 1 FHST0058 Condyle Element Shaft
10 4 CPAR0230 Elastic's Screw of Lower Frame
11 1 FRAM0066 Lower Frame
12 4 FPLM0064 Mounting Plate
13 4 FPIG0076 Mounting Plate Guide Pin
14 2 SCRE0010 Mounting Plate Knob
15 1 SCRE0002 Incisal Table Wing Screw
16 1 FMES0063 Incisal Guiding Table (Polycarbonate)
17 1 STRA0016 Central Lock
18 2 CPAR0348 Central Lock Allen Screw
19 1 FPIN0053 Incisal Guide Pin
20 1 SCRE0011 Incisal Guide Pin Wing Screw
21 1 FALH0073 Side Shift Guide (Left)
22 1 FRAM0065 Upper Frame
23 1 FCDL0071 Condylar Guide (Left)
24 1 FALH0074 Side Shift Guide (Right)
25 2 CPAR0034 Centric Pin Screw
26 2 CANE0008 Washer for Mounting Plate Knob
27 2 SCRE0020 Condylar Guide Wing Screw
17
11 - PART LIST - ARTICULATOR 4000
18
12 - PART LIST - ARTICULATOR A7 PLUS
Item Qty Code Description
1 1 SCRE0078 Incisal table screw
2 1 FRAM0731 Lower frame
3 1 FMES0063 Incisal table (acrylic)
4 1 FTAP0761 Finishing cover of right column
5 1 FAGU0753 Occlusal plan indicator
6 1 FPIN0750 Incisal pin with hole
7 1 FRAM0740 Upper frame
8 1 SALH0075 Bennett angle adjusting device (Right)
9 1 SGUI0073 Condylar guide (right)
10 2 FPIN0744 Stabilizing elastic band pin
11 1 FPOS0759 Position indicator of Bennett angle adjusting device (right)
12 2 FPAR0742 Screw to fix the Bennett angle adjusting device
13 1 FPIN0743 Upper frame support pin
14 2 FTRA0765 Lock of upper frame
15 2 CANE0319 Condylar guide O-ring
16 1 CETQ1113 Model identification label
17 2 FFIX0766 Condylar guide fixer
18 2 SCRE0008 Condylar guide screw
19 1 FTAP0764 Pulley finishing cover
20 1 FPOS0758 Position indicator of Bennett angle adjusting device (left)
21 1 SGUI0072 Condylar guide (left)
22 1 SALH0074 Bennett angle adjusting device (Left)
23 4 FPIG0076 Mounting plate guide pin
24 2 FCAN0755 Magnetic fitting
25 2 CIMA1118 Mounting plate magnetic
26 2 SPLA0079 Mounting plate (track)
27 1 SCRE0011 Incisal pin screw
28 2 CPAR1119 Condylar element screw
29 2 CMOL1100 Stabilizing elastic band
30 2 FECD0734 Condylar element
31 2 FPIG0788 Face bow's guide pin
32 1 FTAP0760 Finishing cover of left column
33 1 SCRE0082 Mounting plate screw (lower frame)
34 1 SCRE0010 Mounting plate screw (upper frame)
19
13 - PART LIST - ARTICULATOR A7 FIX
20
14 - PART LIST - ARTICULATOR EVA PLUS
Item Qt Code Description
01 02 CPAR0548 Column Screw
02 02 FBAS0483 Column Base
03 02 SCRE0010 Mounting Plate Knob
04 01 FRAM0474 Lower Frame
05 02 CANE0008 Washer for Mounting Plate Knob
06 04 FPIG0076 Mounting Plate Guide Pin
07 04 FPLM0064 Metallic Mounting Plate
08 01 FCOL0500 Column (Right)
09 01 FCAB0499 Bennet Angle Adjusting Head (Right)
10 02 FARR0504 Adjusting Head Lock Washer
11 01 FTRA0503 Adjusting Head Lock Pin
12 01 FTRA0501 Condylar Guide Lock (Right)
13 01 FHST0514 Protrusion Adjustment Screw-Driver
14 02 FTRA0502 Condylar Guide Wedge Lock
15 02 FPAR0513 Protrusion Adjustment Screw
16 02 CANE0805 Rubber O’Ring
17 02 FESF0482 Condylar Sphere
18 02 FEIX0484 Horizontal Slide Bar
19 01 FGUI0498 Condylar Guide (Right)
20 02 CLIM0554 Condylar Sphere Limiter
21 02 CMOL0031 Condylar Sphere Limiter Spring
22 03 SCRE0011 Incisal Guide Pin Wing Screw
23 01 CETQ0634 PVC Label
24 02 FPAR0485 Horizontal Slide Bar Lock Screw
25 01 FPIN0692 Stabilization Pin (Extension)
26 01 FGUI0506 Condylar Guide (Left)
27 01 FPIN0481 Support Pin
28 01 FRAM0475 Upper Frame
29 01 FTRA0520 Condylar Guide Lock (Left)
30 01 FTRA0519 Adjusting Head Lock Pin (Left)
31 01 FCAB0507 Bennet Angle Adjusting Head (Left)
32 01 FCOL0509 Column (Left)
33 01 FPIN0053 Incisal Guide Pin
34 01 FMES0063 Incisal Guiding Table (Polycarbonate)
35 01 SCRE0002 Incisal Table Wing Screw
36 01 CETQ0016 Identification Label
37 02 FALA0505 Condylar Guide Lever
38 02 FALA0521 Bennet Angle Lever
39 01 FHST0058 Horizontal Slide Bar Shaft
21
15 - PART LIST - ARTICULATOR EVA FIX
22
16 - PART LIST - STANDARD FACE-BOW
Item Qt Code Description
1 2 SBBL0001 Wing screw
2 1 FFIX0362 Piece to fix the cross bar (without hole)
3 2 FMOL0025 Elastic washer
4 1 SEXP0006 Horizontal slide bar
5 1 STAR0005 Cross bar
6 1 FPNZ0020 Nose piece support
7 5 SCRE0020 Face-bow wing screw
8 1 FHST0018 Nose piece shaft
9 1 FBLO0016 Nose piece
10 1 FARC0784 Face-bow arm (Right)
11 1 FAUR0033 Ear piece (Right)
12 1 CETQ0018 Model identification label
13 1 FAUR0032 Ear piece (Left)
14 1 CELA0009 Elastic band
15 1 FARC0785 Face-bow arm (Left)
16 1 FPOR0039 Center nut
17 1 SGDE0007 Bite fork
18 1 FBLO0008 Connection Block (internal)
19 2 FTMP0027 Stopping screw
20 1 FBLO0006 Connection block (external)
21 1 CARR0004 Washer for the block
22 1 FBLO0026 Fork´s connection block
23 1 CPIE0005 Affixing block pin
24 1 FFIX0022 Affixing block (with hole)
23
17 - PART LIST - PROFESSIONAL FACE-BOW
Item Qt Code Description
1 4 SBBL0001 Wing screw
2 1 FPIG0009 Guide pin of transference set (lower base)
3 1 SCRE0002 Incisal table screw
4 1 FBAS0002 Transference set lower base
5 2 CORI0002 Transference set column O-ring
6 1 FCOL0007 Transference set column
7 1 FTMP0027 Stopping screw
8 1 FBLO0560 Connection Block (internal)
9 1 SEXP0004 Horizontal slide bar
10 1 FBLO0559 Connection block (external)
11 1 CARR0004 Washer for the block
12 1 FBLO0561 Fork´s connection block
13 1 CPIE0005 Affixing block pin
14 1 CAPO1074 Upper frame support
15 2 FMOL0025 Elastic washer
16 1 FFIX0022 Affixing block (with hole)
17 1 SGDE0003 Bite fork
18 1 FARC0582 Face-bow (Left)
19 2 FAUR0019 Ear piece
20 1 CETQ0025 Model identification label
21 2 FPOR0021 Face-bow nut
22 1 FHST0018 Nose piece shaft
23 1 FBLO0016 Nose piece
24 1 FARC0581 Face-bow (Right)
25 3 SCRE0020 Face-bow wing screw
26 1 CPOR1023 Center nut
27 1 FPNZ0020 Nose piece support
28 2 CPAR0230 Upper base screw
29 1 FWWW0583 Face-bow "W"
30 2 FPAR0017 Joint screw
31 1 FBAS0003 Transference set upper base
32 1 FFIX0362 Piece to fix the cross bar (without hole)
24
Rua Teotônio Vilela, 120 - Jd. Tangará - CEP 13568-000 - São Carlos - SP - Brasil
Tel. +55 (16) 3371-6502 - Fax +55 (16) 3372-5953
Home Page: www.bioart.com.br E-mail: bioart@bioart.com.br