Académique Documents
Professionnel Documents
Culture Documents
ATEX
ATEX system
The 3662Ex is utilized when it is necessary to mount both sensor and instrument inside a hazardous area. The complete system is designed to prevent any form of spark. It has current, voltage, and RS485 outputs that should be connected to an isolated repeater to send the signal to the safe area. The 3660Ex instrument is suitable for installation outside of the hazardous area. This instrument receives a signal from a sensor which is installed in the hazardous area up to 1 km away. The instrument has an intrinsically safe current output to stay in the hazardous area and current, alarms, and serial outputs (RS232 or RS 485) for connection in the safe area.
Sa f e ar e a
32956 Safety barrier
230/240 VAC 110/127 VAC 24/48 VDC 0/4 - 20mA
32 95 4
329 56
Ha za r d ou s a re a
3662 x instrument sample stream
6.5-13.5VDC @50 mA
32959
3295 9
RS485/RS232 converter
RS 485 RS 485
max. cable length: 1000 m
RS 485
O2 . H2 . O3
EEx Intrinsincally safe apparatus built to CENELEC European standards. ia Type of protection: the highest category, based on a safety factor 1.5 on two faults. No combination of two faults in the analyzer can produce a spark, or heating, causing ignition of an explosive atmosphere. IIC Gas group: corresponds to the most ammable gases, including acethylene and hydrogen. T6 Temperature class: maximum surface temperature of 85C for an ambient temperature of 45C.
Accessories
Model n
29501 311xxE.xx 312xxE-xx 313xxE.xx 32001.xxx 32003 32301
Operating Temperature
Item Description
Sensor socket pipe weld adapter Oxygen sensor Hydrogen sensor Ozone sensor Flow chamber for 0.25 or 6 mm ProAcc insertion/extraction valve for 1-4 pipe. Electrochemical sensor cleaning and regeneration center External pressure sensor cable, (mm = length, standard length 3 m) Sensor cable, (mm = length) RS-232 cable for 32959 converter External pressure sensor, 0-3.5 bar (50 psia) ATEX 3660 Windows program software Isolator repeater for 4/20mA output, installed in safe area if 4/20mA is used (for 3662EX) Power supply safety barrier for 3662 Ex (X = A : 115 VAC, B : 230 VAC, C : 24/48 VDC) RS 485 isolator for 3662 Ex RS485/RS232 converter or 3662 Ex
0 to +45 C IP 65/NEMA 4X, all stainless steel EMC Standards: EN 61326-1 (1997) +A1 (1998) 200 x 200 x 100 mm 2.5 kg 1 km
Enclosure rating CE certification Dimensions ( W x H x D ) Weight Sensor cable max length 3662 Ex Power requirements Serial link (hazardous area)* Current output (hazardous area)*
6.5 to 13.5 VDC @ 50 mA, 0.8 W through a Zener barrier RS485 (requires 32958 isolator to send the signal to the safe area) 0/4 20 mA, software adjustable, non-isolated, Rmax = 100 (requires 32954 isolator/repeater to send the signal to the safe area). Available in 0-4 mA current output range 0 2 V, software adjustable, non-isolated, Rmin = 100k .
32514E.mm
3660 Ex Power requirements Serial link (safe area) Alarm relays (safe area) 95 to 130 VAC/207 to 253 VAC 10%, 50/60 Hz/20 VA 18 to 36 VDC, 10 Wmax RS 232 or RS485 3 relays: high and low limits, system alarm, NO and NC contacts, current 1 A, voltage 24 VDC, power 24 W, hysteresis: 1% of high or low limit 0/4 20 mA, software adjustable, non-isolated, Rmax = 500 0/4 20 mA, software adjustable, non-isolated, Rmax = 350
32956.X
32958 32959
Instrument congurations
366x Ex / 1xx
02*
H2
Dissolved Gaseous
ppb/ppm ; ppm ; mg/l Pa/kPa ; mbar/bar ; % ; % sat ; ppm / % ppb/ppm ; cc/kg Pa/kPa ; mbar/bar ; % ppb/ppm ; ppm
366x Ex / 2xx
Dissolved Gaseous
366x Ex / 3xx
03
Dissolved
In the interest of continued product development, Orbisphere reserves the right to make improvements to this literature and/or the products it describes, without notice or obligation.
orbisphere
salesinfo@hachultra.com Tel. ++41 22 855 91 00 Fax ++41 22 855 91 99 6, route de Compois, C.P.212 CH-1222 Vsenaz, Geneva Switzerland
Represented By
Model
Gas
Phase
Concentration units