Calibration and Measurement Capabilities

Acoustics, Ultrasound and Vibration, Mexico, CENAM (Centro Nacional de Metrologia)


Measurement Conditions/Independent Variable

Calibration or Measurement Service

Measurand Level or Range

Expanded Uncertainty

Quantity

Instrument or Artifact

Instrument Minimum Type or Method value Reciprocity ASA S2.2-1959 ISO16063-1 Laser interferometry ISO 16063-11 Laser interferometry ISO 16063-11 Laser interferometry ISO 16063-11 Laser interferometry ISO 16063-11 Comparison, ISO 5347-3 Laser interferometry ISO 16063-11 Laser interferometry ISO 16063-11 Comparison, ISO 5347-3

Maximum value

Units

Parameter

Specifications

Value

Coverage Units factor

Level of Confidence

Is the expanded uncertainty a relative one? Yes

Comments

NMI Internal Service Identifier 510-F010001

Charge sensitivity (magnitude)

Accelerometer

2 C/(m/s )

Frequency

100 Hz to 4 kHz

95%

Sinusoidal excitation

Charge sensitivity (magnitude)

Accelerometer

C/(m/s2)

Frequency

50 Hz to 1 kHz

0.5

95%

Yes

Sinusoidal excitation

510-F011001

Charge sensitivity (magnitude)

Accelerometer

C/(m/s )

Frequency

1 kHz to 5 kHz

0.8

95%

Yes

Sinusoidal excitation

510-F011001

Charge sensitivity (magnitude)

Accelerometer

C/(m/s )

Frequency

5 kHz to 8 kHz

95%

Yes

Sinusoidal excitation

510-F011001

Charge sensitivity (phase shift)

Accelerometer

360

Frequency

50 Hz to 5 kHz

95%

No

Sinusoidal excitation

510-F011001

Charge sensitivity (magnitude)

Accelerometer

C/(m/s2)

Frequency

50 Hz to 5 kHz

1.2

95%

Yes

Sinusoidal excitation

510-F012001

Voltage sensitivity Acceleration (magnitude) measuring chain

V/(m/s 2)

Frequency

50 Hz to 1 kHz

0.5

95%

Yes

Sinusoidal excitation

510-F011001

Voltage sensitivity Acceleration (phase shift) measuring chain Voltage sensitivity Acceleration (magnitude) measuring chain

360

Frequency

50 Hz to 5 kHz

95%

No

Sinusoidal excitation Sinusoidal excitation

510-F011001 510-F012001

2 V/(m/s )

Frequency

50 Hz to 5 kHz

1.2

95%

Yes

The BIPM key comparison database, April 2005

1/5

Calibration and Measurement Capabilities

Acoustics, Ultrasound and Vibration, Mexico, CENAM (Centro Nacional de Metrologia)


Measurement Conditions/Independent Variable

Calibration or Measurement Service

Measurand Level or Range

Expanded Uncertainty

Quantity

Instrument or Artifact

Instrument Minimum Type or Method value

Maximum value

Units

Parameter

Specifications

Value

Coverage Units factor

Level of Confidence

Is the expanded uncertainty a relative one? Yes

Comments

NMI Internal Service Identifier 510-F012015 510-F011001 510-F011001 510-F019001

Voltage sensitivity Acceleration (magnitude) measuring chain Acceleration measuring instrument Acceleration measuring instrument Acceleration measuring instrument Acceleration measuring instrument Calibrators

Comparison, ISO 5347-3 Laser interferometry ISO 16063-11 Laser interferometry ISO 16063-11 Comparison, ISO 5347-3 Comparison, ISO 5347-3 Comparison, ISO 5347-3 Comparison, ISO 5347-3 Pressure reciprocity IEC 1094-2:1992 Pressure reciprocity IEC 1094-2:1992 Pressure reciprocity IEC 1094-2:1992

2 V/(m/s )

Frequency

50 Hz to 5 kHz

1.7

95%

Sinusoidal excitation Sinusoidal excitation Sinusoidal excitation Sinusoidal excitation Sinusoidal excitation Sinusoidal excitation Sinusoidal excitation

Acceleration

10

100

m/s2

Frequency

50 Hz to 1 kHz

0.5

95%

Yes

Acceleration

10

100

m/s

Frequency

1 kHz to 5 kHz

0.8

95%

Yes

Acceleration

10

50

m/s

Frequency

50 Hz to 5 kHz

1.2

95%

Yes

Acceleration

10

50

m/s

Frequency

50 Hz to 1 kHz

95%

Yes

510-G'03-002

Acceleration

10

100

m/s

Frequency

80 Hz to 160 Hz

95%

Yes

510-F012009 510-F012009

Acceleration

Calibrators

10

100

m/s

Frequency

50 Hz to 5 kHz

95%

Yes

Pressure sensitivity level

Measurement microphone type LS1 Measurement microphone type LS1 Measurement microphone type LS1

dB (re 1 V/Pa)

Frequency

31.5 Hz to 250 Hz

0.04

dB

95%

No

510-F013001

Pressure sensitivity level

dB (re 1 V/Pa)

Frequency

500 Hz to 3.1 kHz

0.03

dB

95%

No

510-F013001

Pressure sensitivity level

dB (re 1 V/Pa)

Frequency

4 kHz to 5 kHz

0.05

dB

95%

No

510-F013001

The BIPM key comparison database, April 2005

2/5

Calibration and Measurement Capabilities

Acoustics, Ultrasound and Vibration, Mexico, CENAM (Centro Nacional de Metrologia)


Measurement Conditions/Independent Variable

Calibration or Measurement Service

Measurand Level or Range

Expanded Uncertainty

Quantity

Instrument or Artifact

Instrument Minimum Type or Method value

Maximum value

Units

Parameter

Specifications

Value

Coverage Units factor

Level of Confidence

Is the expanded uncertainty a relative one?

Comments

NMI Internal Service Identifier

Pressure sensitivity level

Measurement microphone type LS1 Measurement microphone type LS1 Measurement microphone type LS1 Measurement microphone type LS2 Measurement microphone type LS2 Measurement microphone type LS2 Measurement microphone type LS2 Measurement microphone type LS2 Measurement microphone type LS2

Pressure reciprocity IEC 1094-2:1992 Pressure reciprocity IEC 1094-2:1992 Pressure reciprocity IEC 1094-2:1992 Pressure reciprocity IEC 1094-2:1992 Pressure reciprocity IEC 1094-2:1992 Pressure reciprocity IEC 1094-2:1992 Pressure reciprocity IEC 1094-2:1992 Pressure reciprocity IEC 1094-2:1992 Pressure reciprocity IEC 1094-2:1992

dB (re 1 V/Pa)

Frequency

6.3 kHz

0.06

dB

95%

No

510-F013001

Pressure sensitivity level

dB (re 1 V/Pa)

Frequency

8 kHz to 9 kHz

0.1

dB

95%

No

510-F013001

Pressure sensitivity level

dB (re 1 V/Pa)

Frequency

10 kHz

0.15

dB

95%

No

510-F013001

Pressure sensitivity level

dB (re 1 V/Pa)

Frequency

63 Hz to 3.15 kHz

0.05

dB

95%

No

510-F013001

Pressure sensitivity level

dB (re 1 V/Pa)

Frequency

4 kHz to 8 kHz

0.06

dB

95%

No

510-F013001

Pressure sensitivity level

dB (re 1 V/Pa)

Frequency

9 kHz to 10 kHz

0.1

dB

95%

No

510-F013001

Pressure sensitivity level

dB (re 1 V/Pa)

Frequency

12.5 kHz

0.12

dB

95%

No

510-F013001

Pressure sensitivity level

dB (re 1 V/Pa)

Frequency

14 kHz to 18 kHz

0.14

dB

95%

No

510-F013001

Pressure sensitivity level

dB (re 1 V/Pa)

Frequency

20 kHz

0.2

dB

95%

No

510-F013001

The BIPM key comparison database, April 2005

3/5

Calibration and Measurement Capabilities

Acoustics, Ultrasound and Vibration, Mexico, CENAM (Centro Nacional de Metrologia)


Measurement Conditions/Independent Variable

Calibration or Measurement Service

Measurand Level or Range

Expanded Uncertainty

Quantity

Instrument or Artifact

Instrument Minimum Type or Method value

Maximum value

Units

Parameter

Specifications

Value

Coverage Units factor

Level of Confidence

Is the expanded uncertainty a relative one? No

Comments

NMI Internal Service Identifier 510-F012002

Pressure sensitivity level

Measurement Substitution IEC microphone type 1094-5 LS2 Pistonphone, single frequency 250 Hz Pistonphone or sound calibrator, single frequency 125 Hz to 1 kHz Calibrated measurement microphone

dB (re 1 V/Pa)

Frequency

250 Hz

0.11

dB

95%

Sound pressure level

70

130

dB (re 20 Pa)

Microphone type

LS1P

0.09

dB

95%

No

510-F012003

Sound pressure level

Calibrated measurement microphone

70

130

dB (re 20 Pa)

Microphone type

LS2P

0.11

dB

95%

No

510-F012003

Sound pressure level

Sound calibrator, single frequency 125 to 1 kHz

Calibrated measurement microphone Calibrated measurement microphone

70

130

dB (re 20 Pa)

Microphone type

WS2P

0.16

dB

95%

No

510-F015003

Sound pressure level

Sound calibrator, multi-frequency

70

115

dB (re 20 Pa)

Frequency Microphone type

125 Hz to 1 kHz

0.16

dB

95%

No

510-F012008

LS2P 510-F012008

Sound pressure level

Sound calibrator, multi-frequency

Calibrated measurement microphone

70

115

dB (re 20 Pa)

Frequency Microphone type

31.5 Hz to 8 kHz

0.21

dB

95%

No

LS2P 250 Hz to 1 kHz 0.3 dB 2 95% No 510-F015001

Sound pressure response level

Sound level meters

Comparison IEC651 and IEC804

dB (re 20 Pa)

Frequency

The BIPM key comparison database, April 2005

4/5

Calibration and Measurement Capabilities

Acoustics, Ultrasound and Vibration, Mexico, CENAM (Centro Nacional de Metrologia)


Measurement Conditions/Independent Variable

Calibration or Measurement Service

Measurand Level or Range

Expanded Uncertainty

Quantity

Instrument or Artifact

Instrument Minimum Type or Method value

Maximum value

Units

Parameter

Specifications

Value

Coverage Units factor

Level of Confidence

Is the expanded uncertainty a relative one? No

Comments

NMI Internal Service Identifier 510-F016001

Air-conduction response level

Audiometers

IEC60645-1

dB HL (re threshold of hearing) dB HL (re threshold of hearing) dB (re reference equivalent threshold force) dB (re 1 N/(m/s) ) 0.01 1 W

Frequency

125 Hz to 8 kHz

0.38

dB

95%

Bone-conduction response level

Audiometers

IEC60645-1

Frequency

125 Hz to 4 kHz

0.4

dB

95%

No

510-F016001

Force response level

Mechanical coupler system

Comparison IEC 60373

Frequency

125 Hz to 4 kHz

0.3

dB

95%

No

510-F012011

Mechanical impedance Ultrasonic power

Mechanical coupler system Ultrasonic source transducer

Comparison IEC 60373 Radiation force balance IEC 1161

Frequency

125 Hz to 4 kHz

0.2

dB

95%

No

510-F012011 510-F018003

Frequency

1 MHz to 10 MHz

11

95%

Yes

The BIPM key comparison database, April 2005

5/5