Vous êtes sur la page 1sur 14

Validation of analytical procedures

1.0 Introduction Regulated laboratories must perform method validation in order to be in compliance with 1) US Food and Drug Administration (FDA) regulations 2) US Pharmacopeia 3) International Conference on Harmonization Unfortunately, there is no single source or final guideline on analytical method validation Validation is customized by choosing necessary tests and acceptance criteria for a given method. The comprehensiveness of validation is based upon the type of method and its requirements. Due to legal status of USP and NF standards, it is essential, therefore, that proposals for adoption of new or revised compendial analytical procedures be supported by sufficient laboratory data to document their validity. 2.0 Pre-validation requirements The first step in the development of a method validation protocol is to determine the objective of the method. What is the method intended to demonstrate? 2.1 Analytical equipment qualification Vendors site Structural and software validation ISO / 21 CFR 11 Vendors Responsibility Users site Functional validation Users site Calibration Maintenance System suitability PQ IQ OQ PQ

IQ

OQ

Before use Vendor assistance

Users Responsibility

Installation Qualification (IQ), Operational Qualification (OQ), Performance Qualification (PQ)

IQ - Instrument is received as designed and specified and that it is properly installed

OQ - specific modules of the system are operating according to the defined specifications
PQ - PQ testing is conducted under actual running conditions across the anticipated working. For HPLC, the PQ test should use a method with a well-characterized analyte mixture. After the instrument is placed on-line in the laboratory, and after a set period of use, regulations require maintenance followed by calibration and standardization. A system suitability test provides assurance that a systems performance still is appropriate for use.

2.2 Stability of analytical solutions


To generate reproducible and reliable results, the stability of sample solutions, standards, reagents and mobile phases must be determined prior to initiate the method validation studies. For assay methods samples and standards should be tested over at least a 24-h period (depends on need), and quantitation of components should be determined by comparison to freshly prepared standards. Acceptable stability is 2% change in standard or sample response, relative to freshly prepared standards. The mobile phase is considered to have acceptable stability if aged mobile phase produces equivalent chromatography (capacity factors, resolution, or tailing factor) and assay results are within 2% of the value obtained with fresh mobile phase. For impurity methods, the sample, standard solutions and mobile phase will be stable for 24 h under defined storage conditions. Acceptable stability is 20% change in standard or sample response at the limit of quantitation, relative to freshly prepared standards. The mobile phase is considered to have acceptable stability if aged mobile phase produces equivalent chromatography and if impurity results at the limit of quantitation are within 20% of the values obtained with fresh mobile phase. If a solution is not stable at room temperature, then decreasing the temperature to 28 8C can improve stability of samples and standards. 2.3 System suitability test System suitability tests are based on the concept that the equipment, electronics, analytical operations, and samples constitute an integral system that can be evaluated as a whole. System suitability is the checking of a system to ensure system performance before or during the analysis of unknowns.

The parameters to be measured and their recommended limits obtained from the analysis of the system suitability are shown as follows

System suitability parameters and recommendations


Parameter Capacity factor (k) Repeatability Relative retention Recommendation The peak should be well-resolved from other peaks and the void volume, generally k>2.0 RSD 1% for N 5 is desirable Not essential as long as the resolution is stated

Resolution (Rs )

Rs of >2 between the peak of interest and the closest Eluting potential interference ( impurity, excipient, degradation product, internal standard, etc.
T of 2 In general should be > 2000

Tailing factor (T ) Theoretical plates (N) 3.0 Validation

Validation of an analytical procedure is the process by which it is established, by laboratory studies, that the performance characteristics of the procedure meet the requirements for the intended analytical applications.
3.1 Analytical Performance Characteristics 3.1.1 Accuracy Definition The accuracy of an analytical procedure is the closeness of test results obtained by that procedure to the true value.

The accuracy of an analytical procedure should be established across specified range of the analytical procedure
Accuracy is usually determined in one of four ways. 1) In the case of assay of drug substance accuracy can be assessed by analyzing a sample of known purity (reference materials) and comparing the measured value to the true value. 2) Comparison of the results of the proposed analytical procedure with those of a second well-characterized procedure, the accuracy of which is stated and/or defined

3) The third approach, based on the recovery of known amounts of analyte, is performed by spiking analyte in blank matrices. The analyte levels in the spiked samples should be determined using the same quantitation procedure as will be used in the final method procedure. The percent recovery is then calculated. For assay methods Spiked samples are prepared in triplicate at three levels over a range of 50150% of the target concentration. For impurity methodsSpiked samples are prepared in triplicate at three levels over a range that covers the expected impurity content of the sample, such as 0.12.5% (v/w). 4) The fourth approach is the technique of standard additions, which can also be used to determine recovery of spiked analyte. This approach is used if it is not possible to prepare a blank sample matrix without the presence of the analyte. For assay methods The mean recovery should be 100 2% at each concentration over the range of 80120% of the target concentration. For impurity methodsThe mean recovery should be within 0.1% absolute of the theoretical concentration or 10% relative, whichever is greater, for impurities in the range of 0.12.5% (v/w) To document accuracy the ICH guideline recommends collecting data from a minimum of nine determinations over a minimum of three concentration levels covering the specified range (for example, three concentrations, three replicates each). 3.1.2 Precision The precision of an analytical procedure expresses the closeness of agreement (degree of scatter) between a series of measurements obtained from multiple sampling of the same homogeneous sample under the prescribed conditions. It is normally expressed as the percent relative standard deviation (RSD) for a statistically significant number of samples. According to the ICH precision should be performed at three different levels: repeatability, intermediate precision, and reproducibility.

3.1.2.1 Repeatability Repeatability expresses the precision under the same operating conditions over a short interval of time. Repeatability is also termed intra-assay precision. It should be determined from a) A minimum of 9 determinations covering the specified range for the procedure (e.g., 3 concentrations/3 replicates each) b) A minimum of 6 determinations at 100% of the test concentration. 3.1.2.2 Intermediate precision Intermediate precision expresses within-laboratories variations : different days, different analysts, different equipment, etc. In determining intermediate precision, experimental design should be employed so that the effects (if any) of the individual variables can be monitored. For assay method The instrument precision (RSD) will be 1% and the intra-assay precision will be 2%.

For impurity assay


At the limit of quantitation, the instrument precision will be 5% and the intra -assay precision will be 10%. 3.1.2.2 Reproducibility Reproducibility expresses the precision between laboratories . Statistical equivalence is often used as a measure of acceptable inter-laboratory results. An alternative, more practical approach, is the use of analytical equivalence, in which a range of acceptable results is chosen prior to the study and used to judge the acceptability of the results obtained from the different laboratories. For assay method The assay results obtained in multiple laboratories will be statistically equivalent or the mean results will be within 2% of the value or the mean results will be within 2% of the value obtained by the primary testing lab. For impurity assay The results obtained in multiple laboratories will be statistically equivalent or the mean results

will be within 10% (relative) of the value obtained by the primary testing lab, % (w/w) within 25% for impurities from 0.1 to 1.0 % (w/w).

Reproducibility is not normally expected if intermediate precision is per formed.


3.1.3 Specificity Specificity is the ability to assess unequivocally the analyte in the presence of components which may be expected to be present. Typically these might include impurities, degradants, matrix, etc. 3.1.3.1 Identification Suitable identification tests should be able to discriminate between compounds of closely related structures which are likely to be present. The discrimination of a procedure may be confirmed by obtaining positive results (perhaps by comparison with a known reference material) from samples containing the analyte, coupled with negative results from samples which do not contain the analyte. The choice of such potentially interfering materials should be based on sound scientific judgement with a consideration of the interferences that could occur. 3.1.3.2 Assay and Impurity Test(s) For chromatographic procedures, representative chromatograms should be used to demonstrate specificity and individual components should be appropriately labeled. Similar considerations should be given to other separation techniques. Critical separations in chromatography should be investigated at an appropriate level. For critical separations, specificity can be demonstrated by the resolution of the two components which elute closest to each other. In cases where a non-specific assay is used, other supporting analytical procedures should be used to demonstrate overall specificity. For example, where a titration is adopted to assay the drug substance for release, the combination of the assay and a suitable test for impurities can be used. 3.1.3.3 Impurities are available For the assay , this should involve demonstration of the discrimination of the analyte in the presence of impurities and/or excipients; practically, this can be done by spiking pure substances (drug substance or drug product) with appropriate levels of impurities and/or excipients and demonstrating that the assay result is unaffected by the presence of these materials (by comparison with the assay result obtained on unspiked samples). For the impurity test, the discrimination may be established by spiking drug substance or drug

product with appropriate levels of impurities and demonstrating the separation of these impurities individually and/or from other components in the sample matrix.

3.1.3.4 Impurities are not available


If impurity or degradation product standards are unavailable, specificity may be demonstrated by comparing the test results of samples containing impurities or degradation products to a second well-characterized procedure e.g.: pharmacopoeial method or other validated analytical procedure (independent procedure). As appropriate, this should include samples stored under relevant stress conditions: light, heat, humidity, acid/base hydrolysis and oxidation. For assay

The two results should be compared


For impurity tests The impurity profiles should be compared Analyte peak is evaluated for peak purity (diode array detector) and resolution from the nearest eluting peak. 3.1.4 Limit of detection The limit of detection (LOD) is defined as the lowest concentration of an analyte in a sample that can be detected, not quantitated. Several approaches for determining the detection limit are possible, depending on whether the procedure is a non-instrumental or instrumental. Approaches other than those listed below may be acceptable. 3.1.4.1 Based on Visual Evaluation Visual evaluation may be used for non-instrumental methods but may also be used with instrumental methods. The detection limit is determined by the analysis of samples with known concentrations of analyte and by establishing the minimum level at which the analyte can be reliably detected. 3.1.4.2 Based on Signal-to-Noise This approach can only be applied to analytical procedures which exhibit baseline noise. Determination of the signal-to-noise ratio is performed by comparing measured signals from samples with known low concentrations of analyte with those of blank samples and establishing the minimum concentration at which the analyte can be reliably detected. A signal-to-noise ratio between 3 or 2:1 is generally considered acceptable for estimating the LOD

3.1.4.3 Based on the Standard Deviation of the Response and the Slope

The detection limit (LOD) may be expressed as: 3.3 / S


where = the standard deviation of the response S = the slope of the calibration curve

The slope S may be estimated from the calibration curve of the analyte. The estimate of may be carried out in a variety of ways, for example: a) Based on the Standard Deviation of the Blank Measurement of the magnitude of analytical background response is performed by analyzing an appropriate number of blank samples and calculating the standard deviation of these responses. b) Based on the Calibration Curve A specific calibration curve should be studied using samples containing an analyte in the range of LOD. The residual standard deviation of a regression line or the standard deviation of y-intercepts of regression lines may be used as the standard deviation. Recommended Data The detection limit and the method used for determining the detection limit should be presented. If DL is determined based on visual evaluation or based on signal to noise ratio, the presentation of the relevant chromatograms is considered acceptable for justification. In cases where an estimated value for the detection limit is obtained by calculation or extrapolation, this estimate may subsequently be validated by the independent analysis of a suitable number of samples known to be near or prepared at the detection limit. 3.1.5 Limit of quantification The quantitation limit of an individual analytical procedure is the lowest amount of analyte in a sample which can be quantitatively determined with suitable precision and accuracy. The quantitation limit is a parameter of quantitative assays for low levels of compounds in sample matrices, and is used particularly for the determination of impurities and / or degradation products. 3.1.5.1 Based on Visual Evaluation Visual evaluation may be used for non-instrumental methods but may also be used with instrumental methods.

The quantitation limit is generally determined by the analysis of samples with known concentrations of analyte and by establishing the minimum level at which the analyte can be quantified with acceptable accuracy and precision. 3.1.5.2 Based on Signal-to-Noise Approach

This approach can only be applied to analytical procedures that exhibit baseline noise. Determination of the signal-to-noise ratio is performed by comparing measured signals from samples with known low concentrations of analyte with those of blank samples and by establishing the minimum concentration at which the analyte can be reliably quantified. A typical signal-to-noise ratio is 10:1.
3.1.5.3 Based on the Standard Deviation of the Response and the Slope The quantitation limit (LOQ may be expressed as: 10.0 / S where = the standard deviation of the response S = the slope of the calibration curve

The slope S may be estimated from the calibration curve of the analyte. The estimate of may be carried out in a variety of ways, for example:

a) Based on the Standard Deviation of the Blank Measurement of the magnitude of analytical background response is performed by analyzing an appropriate number of blank samples and calculating the standard deviation of these responses. b) Based on the Calibration Curve A specific calibration curve should be studied using samples containing an analyte in the range of LOD. The residual standard deviation of a regression line or the standard deviation of y-intercepts of regression lines may be used as the standard deviation. Recommended Data The detection limit and the method used for determining the detection limit should be presented. If DL is determined based on visual evaluation or based on signal to noise ratio, the presentation of the relevant chromatograms is considered acceptable for justification. In cases where an estimated value for the detection limit is obtained by calculation or extrapolation, this estimate may subsequently be validated by the independent analysis of a suitable number of samples known to be near or prepared at the detection limit.

3.1.6 Linearity and range

Vous aimerez peut-être aussi