Académique Documents
Professionnel Documents
Culture Documents
Mass / CD/ORD
NMR
IR
35 Marks
40 Marks
8 Marks
X Ray
10 Marks
UV
7 Marks
Total
100 Marks
Mass Spectrometry
Sensitivity:
Selectivity:
High Speed:
-e
M+.
-N.
C+
-Nc
M+.
A+
A+. + N
-Na
B+
-Nb
etc.
Multiple pathways
Simple fragmentation
Complex Fragmentation
Simple spectrum
Complex spectrum
1.
Abundances of ions
x M+.
y M+.
sufficient internal energies
insufficient
internal
energies
fragment further to
A+, B+, C+ etc
Table 1.1. Normalized ion abundances in the mass spectrum of 1.3-dimethylbenzene ( % RA)
m/z
Abundance
m/z
Abundance
m/z
Abundance
38
1.2
65
5.5
102
1.1
39
7.1
74
1.1
103
6.1
51
2.8
77
11.3
104
2.6
52
9.1
78
5.1
105
28.7
53
5.0
79
6.5
106
61.7
62
1.6
91
100.0
107
5.4
63
4.2
92
7.8
a
ion abundances less than one per cent of the base peak have been omitted.
Table 1.2 ion abundances in the mass spectrum of 1,3-dimethylbenzene as %
TIC
m/z
Abundance
m/z
Abundance
m/z
Abundance
39
2.5
65
1.9
92
2.7
50
1.0
77
3.9
103
2.1
51
3.2
78
1.8
105
10.0
52
1.7
79
2.2
106
21.4
63
1.5
91
34.8
107
1.9
a
ion abundances of less than one per cent are omitted
A+ + B+
M2+
C
2+
+N
Isotopes:
Hydrogen
Boron
Carbon
Nitrogen
Oxygen
Fluorine
Silicon
Phosphorus
Sulphur
Chlorine
Bromine
Iodine
31
H (99.99)
10
B (19.8)
12
C (98.9)
14
N (99.6)
16
O (99.8)
19
F (100.0)
28
Si (92.2)
P (100.0)
32
S (95.0)
35
CI (75.5)
79
Br (50.5)
127
I (100.0)
B (80.2)
13
C (1.1)
15
N (0.4)
18
O (0.2)
11
Si (4.7)
30
Si (3.1)
S (0.7)
37
CI (24.5)
81
Br (49.5)
34
S (4.2)
29
33
Metastable ions
The ions, which at the detector, have less than the full
kinetic energy orginally imparted at the ion source.
Ions spend 10-6 s in the ion source and take 10-5 s after
leaving the source to reach the detector.
Ions,
i)
with large excess of int. E have high rates of
decompn. (> 106/sec) and fragment in the ion
source.
ii)
small excess of int E have low rates of
decompn(< 105/sec) they are stable within the
life time of the analysis.
iii)
intermediate int E have rate consts b/w 105106/sec give rise to metastable ions.
MASS SPECTROMETRY
1. Sample introduction
2. Methods of sample ionization
3. Analysis and separation of sample
ions
4. Detection and recording of sample
ions.
MASS SPECTROMETER
1. Sample introduction
The method of sample introduction to the ionisation
source often depends on the ionisation method being
used, as well as the type and complexity of the sample.
The sample can be inserted directly into the ionisation
source, or can undergo some type of chromatography en
route to the ionisation source. This latter method of
sample introduction usually involves the mass
spectrometer being coupled directly to a high pressure
liquid chromatography (HPLC), gas chromatography
(GC) or capillary electrophoresis (CE) separation
column, and hence the sample is separated into a series
of components which then enter the mass spectrometer
sequentially for individual analysis.
(EI Source)
At pressure not more than 10-5 torr to 10-7 torr to avoid:
1) the collision of the ions with each other and with the molecules,
as this would produce inter-ionic intermolecular or ion molecular
reactions,
ii) the erosion of the electron producing filament.
Methods:
i)-a) Direct insertion
FIELD IONISATION
An organic compound in the gas phase can be ionized when
the molecules pass near a sharp metal anode carrying a high
electric field. Electrons are sucked from the sample molecules into
incomplete orbitals in the metal, and the resulting molecular ions are
then propelled towards a slit cathode. Primary focusing takes place
at the cathode slit before the ions pass through the entrance slit of
the mass spectrometer to be focused magnetically and
electrostatically as in electron-impact studies.
The principal advantages of field ionization from an organic
chemists point of view are the increased abundance of molecular
ions and the minimization of complex fragmentations and
rearrangements. Disadvantages are the lower sensitivity and
resolution obtained.
2. FIELD IONIZATION:
A very high local electric field
( 2x108 V/cm or 2V/Ao)
M+.
APPLICATION OF FI/FD
FI has the advantage over CI that only M+ + (M+H)+ ions are formed;
adduct ions, (M+X)+ in which the nature of X may not be certain do
not usually occur.
FD is the method of choice for determining the Mol.wt. of thermally
labile molecules e.g:
CARBOHYDRATES
Monosaccharides
Methyl derivs
Disaccharides
Nucleosides
Porphyrins
Triglycerides
Hydrocarbons
Chloroalkanes
Pesticide molecules
Abundant (M+1)+
more abundant (M+1)+
(With FI sometime decompn) but
Methyl derivatives
M+ abundant frags. In FI due to
fission of the glycoside Bond.
Xenon atoms
ionized Xe+. Ions
Accelerated Fast Ions
Xe+. + Xe (thermal)
(Fast) (gas chamber)
Xe
(Fast)
Xe+.
(Thermal)
access
is
deflected.
Thermal Vaporization:
due to the localized heating caused by the impact
of a single H.E. particle.
Sputtering:
Molecules interact with the impacting particles as
hard spheres and desorption occurs during a very
short time interval following impact and prior to
equilibration of energy.
Common Matrices
Glycerol MW 92
1-S Glycerol MW 108
m-Nitrobenzyl Alcohol MW 153
Sample must be soluble
In the matrix
Matrix must have low volatility
Additives can
spectra
be used to enhance
Pos Ion
Neg Ion
NaCl
MNa+
MCl-
HCl
MH+
MCl-
NH4OH
MNH4+
M-H-
Chemical Ionization
For organic chemists, Chemical Ionization (CI) is especially
useful technique when no molecular ion is observed in EI mass
spectrum, and also in the case of confirming the mass to charge
ratio of the molecular ion. Chemical ionization technique uses
virtually the same ion source device as in electron impact,
except, CI uses tight ion source, and reagent gas. Reagent gas
(e.g. ammonia) is first subjected to electron impact. Sample ions
are formed by the interaction of reagent gas ions and sample
molecules. This phenomenon is called ion-molecule reactions.
Reagent gas molecules are present in the ratio of about 100:1
with respect to sample molecules. Positive ions and negative
ions are formed in the CI process. Depending on the setup of the
instrument (source voltages, detector, etc...) only positive ions or
only negative ions are recorded.
Unlike molecular ions obtained in EI method, MH + and [M-H]detection occurs in high yield and less fragment ions are
observed.
1.CHEMICAL IONIZATION:
CH4 + e
CH4+. + 2e
CH4 + CH4+.
CH5+ + CH3.
Reagent gas
1Torr; eV 100-500
ion chamber
Sample gas
R+e
10-3 Torr
[R+.]*
excited
primary reactant
gas ions
.
collisional
H-transfer
(R+H)
equilibrated sec
reactant gas ion
(R+H)+
reactant
gas ion
(M+H)+ + R
molecule
under
investigation
(M-H)+
quasi-mol
(tertiary)
ions
CH4
affords
principally
Isobutane
NH3
H2
+RH2
neutral
species
CH5+
C4H9+ Sec.
ions
NH4+
H 3+
Chemical Ionization
Similar to EI (Electron Ionization)
Used to analyze same types of compounds (must be volatile)
Use same sample inlets (solids probe or GC)
Requires only minor changes to the ion source
Different from EI
Gives less fragmentation, charge-induced fragmentation only
Can determine the number of labile hydrogens
Can selectively ionize amines
Sample introduction
Methods of sample ionization
Comparison b/w Cl and El; source modifications required to
increase pressure
Why is high reagent gas pressure required?
Reagent gas; common gases; reagent ions; common reactions
Proton transfer; hydride abstraction; adduct formation
Proton Affinity; how does it relate to ion excitation energy and
fragmentation
Use proton affinity to predict PT or A in ammonia Cl
Use ammonia Cl to determine the number of exchangeable H
Explain apparent M+ in ammonia Cl spectra
Fast Atom Bombardment Ionization (FAB) and
Liquid Secondary lon Mass Spectrometry (LSIMS)
6.Electrospray ionisation
This gas, usually nitrogen, helps to direct the spray emerging from
the capillary tip towards the mass spectrometer. The charged
droplets diminish in size by solvent evaporation, assisted by a
warm flow of nitrogen known as the drying gas which passes across
the front of the ionisation source. Eventually charged sample ions,
free from solvent, are released from the droplets, some of which pass
through a sampling cone or orifice into an intermediate vacuum
region, and from there through a small aperture into the analyser of
the mass spectrometer, which is held under high vacuum.
Electrospray ionisation
In ESI, samples (M) with molecular
masses up to ca. 1200 Da give rise to
singly charged molecular-related ions,
usually protonated molecular ions of the
formula (M+H)+ in positive ionisation
mode, and deprotonated molecular ions
of the formula (M-H)- in negative
ionisation mode.
4
ESI
n1 charge of m/z1
n2 charge of m/z2
n 2 = n1 + 1
n1 = m/z2 - mH
m/z1 m/z2
Suppose 1st peak is at m/z 1001
2nd peak is at m/z 501
n1 = (501-1)/1001-501
= 500/500 = 1
Mr = 1x (1001-1) = 1000
And doubly protonated ion is at m/z 501 b/c
15
1000+2/2 = 501
Calculation of MW in ESI
Using formula:
n1 = m/z2 - mH
m/z1 m/z2
712.5-1
949.66-712.5
We know,
MW = (m/z x n) n
MW = (949.66 x 3) 3 = 2846.06
17
= 3.00 i.e. n1 = 3
11
13
In
negative
ionisation
mode
the
deprotonated molecular ions (M-H)- are
usually the most abundant species,
accompanied by some salt adducts and
possibly traces of dimeric or doubly charged
materials. Negative ionisation can be used
for the analysis of oligonucleotides and
oligosaccharides.
16
17
13
14
18
19
11
12
37
10
19
20
Magnetic-Sector Mass
Spectrometry
THEORY:
The ion source accelerates ions to a kinetic energy given by:
KE = mv2 = eV
Where m is the mass of the ion, v is its velocity, e is the charge on the
ion, and V is the applied voltage of the ion optics.
The ions enter the flight tube and are deflected by the magnetic
field, B.
Only ions of mass-to-charge ratio that have equal centripetal and
centrifugal forces pass through the flight tube:
mv2 /r = BeV, where r is the radius of curvature
mv2 /r = BeV
By rearranging the equation and eliminating the velocity term using
the previous equations, r = mv/eB = 1/B(2Vm/e)1/2
Therefore, m/e = B2r2/(2V)
This equation shows that the m/e ratio of the ions that reach the
detector can be varied by changing either the magnetic field (B) or
the applied voltage of the ion optics (V).
THEORY:
KE=eV when electrons are accelerated through an
electric field
KE of ion is mv2, so eV= mv2 and velocity is
inversely proportional to mass
Transit time (t) is L/v, where L is drift tube length
and v is velocity
So t=L/(2V/m/e) can be solved for charge-mass ratio
Background Info
Quadrupole Mass Analyzers
are in use since the 1950s
most commonly used mass spec today
sometimes referred to as mass filters because
ions of only a single mass to charge (m/z) ratio pass
through the apparatus
separate ions based on oscillations in an electric
field (the quadrupole field) using AC and DC
currents
Benefits
Quadrupole Mass Analyzers
easy to use
simple construction
fast
low cost
can achieve unit to 0.1 m/q resolution
Obtaining A Spectrum
Quadrupole Mass Analyzers
a mass spectrum is obtained by varying the voltages on the
rods and monitoring which ions pass through the filter
two methods for varying rod voltages:
vary while holding U and V constant
vary U and V but keep the ratio U/V fixed
remember: U= DC voltage, V=AC voltage, = angular
velocity of alternating voltage
The resolution is determined by the magnitude of U/V ratio
beam currents for individual m/e ratios may be as low as
10-13 A and so electron multipliers are usually used
Applications
Quadrupole Mass Analyzers
Summary
32
SUMMARY
33
A.
a)
Often (M-1)+ and sometime (M-2)+, (M-3)+ etc. are
observed due to losses of H atoms from M+.
c)
d)
B)
If metastable ions which implicate the candidate peak as being due to product ions of
some higher mass species, then it can not be the true M+ peak.
Metastable ions in the routine spectrum; recall that the position of the metastable ion
(m) is given by,
m22
m = ______
m1
where,
C)
i)
ii)
iii)
12
100x10.9 = 10
1.1x100
D)
E)
F)
i)
ii)
14
15
STRUCTURE ELUCIDATION
Without rationalization
Mass spectra of unknown structure may be compared with the
published spectra compilations either:
i) Manually
tedious ; or
ii) Through a computer and automatic library searching
Possible only if the mass spectrum has already been recorded and
available in DATA.
By rationalization
(Most in use):
i) Functional-group approach
M+-18
M+-60
M+-31
26
(loss of water)
(loss of C2H4O2)
(loss of OCH3 or-CH2OH)
(C7H7+)
m/z 105
17
R1
C1
R2
C2
R1
C
(R3)
C1
(R3)
X
C2
R1
18
X+
or
R2
or
C
(R3)
+
X
19
CH3
CH3
20
CH
CH2
O +
CH
CH3
CH2
+
HC
+
CH3 + CH2
HC
CH
CH2
+
O
CH2
CH
+
CH2
21
R1
H
CH
(R )
CH2
C
CH2
+
HX
R1
H
CH
+ R1CHCH2CH2CR2
(R3)
RE1
+
X
(R )
CH2
C
CH2
XH
RE2
R1CH
CH2 + CH2
CR2
23
SUMMARY
24
(a)
one-electron
mechanism
(b)
e
+
two-electron
mechanism
O+
O+
O+
H2N
HO
RO
ester
II
+
X
H
+
amide
IV
carboxylic acid
III
X
R
even
B+
odd
odd
even
even
even
Radical ions, being odd-electron species, can extrude a neutral molecule, leaving a radical
ion as coproduct.
.
A+
B+
odd
odd
even
odd
B+
even
C (or B + C+)
odd odd
even
A)
R2
+
R1
C
(R3)
C+
R1
X+
or
C1
R2
C2
C1
(R3)
X
C2
R1
R2
or
C
(R3)
+
X
10
11
12
13
Alpha-Cleavage
-Cleavage of Acetone Molecular Ion
The EI mass spectrum of acetone is comparatively
simple. It basically shows three important peaks at m/z
58, 43, and 15. According to the formula C3H6O, the
peak at m/z 58 corresponds to the molecular ion. The
base peak at m/z 43 is related to this signal by a
difference of 15 u, a neutral loss which can almost
always be assigned to loss of a methyl radical, CH3. The
m/z 15 peak may then be expected to correspond to the
ionic counterpart of the methyl radical, i.e., to the CH3+
carbonium ion. The question remains, as to whether this
mass spectrum can be rationalized in terms of ion
chemistry. Let us therefore consider the steps of electron
ionization and subsequent fragmentation in greater detail
14
15
16
17
When a ketone grows larger it does not necessarily imply that it has
two identical alkyl groups at the carbonyl. In case of different alkyls
at the carbonyl, Stevenson's rule may also be applied to decide
which of them will dominantly be detected as part of the acylium ion
and which should preferably give rise to a carbenium ion. Overall, a
nonsymmetrical ketone will yield four primary fragment ions in its EI
mass spectrum.
For example the 70 e V EI mass spectrum of butanone shows an
ethyl loss, m/z 43, which is largely preferred over methyl loss, m/z
57. Furthermore, the C2H5+ ion, m/z 29, is more abundant than the
less stable CH3+ ion, mlz 15. If the alkyls become larger than ethyl,
another pathway of ion dissociation will occur in addition.
18
19
NH2
+
CH2=NH2
m/z=30
20
21
22
SUMMARY
Fragmentation Processes
23
McLafferty Rearrangement
Figure: Typical appearance of mass spectra (a) with three mirror galvanometers
scanning simultaneously and (b) with a logarithmic galvanometer. Note the
metastable ions at m/z 43.4 and m/z 60.2 in (a)
OC2H5
+
O
77
O
C +
105
8
C2H5O
51
C2H5OC
McLafferty Rearrangement
+
OH
Cl
+
O
OH
Cl
+
m/z 56
18
Detection
limit (pmol)
Common
application
field (Da)
Precision
(%)
Analyzer
FABMS
1-50
6000
0.05
Magnetic or
quadrupole
ESIMS
0.01-5
>130000
0.01
Magnetic or
quadrupole
MALDI-MS
0.001-1
>300000
0.05
Time-offlight
Mass Spectrometry
Points to remember:
1. Ions are detected at a certain m/z value, and therefore only ionic species may
be correlated with peaks. Neutral losses are exclusively indicated by the mass
difference between peaks.
2. Even-electron rule: Odd-electron ions (such as molecular ions and fragment
ions formed by rearrangements) may eliminate either a radical or an evenelectron neutral species, but even electron ions (such as protonated molecules
or fragments formed by a single bond cleavage) will not usually lose a
radical to form an odd-electron cation. In other words, the successive loss of
radicals is forbidden.
3. It is important to assign the correct charge and radical state to all species
encountered and to carefully track them through a fragmentation scheme.
Otherwise, impossible fragmentation pathway may be formulated, thereby
misleading the assignment of elemental composition and molecular
constitution.
4. The detection of the positive halogen ion and a less intensive peak due to the
hydrogen halogenide is a generally observed characteristic of halogenated
compounds. In accordance with the relative electronegativities of the
halogens their intensities follow the order I+ > Br+ > CI+ > F+. In case of Br
21and CI the isotopic patterns give additional evidence for the presence of the
respective halogen.
22
23
24
25
11. A distonic ion is a positive radical ion, which would formally arise by
ionization of a zwitterion or a diradical by isomerization or fragmentation of
classical molecular ion, or by ion-molecule reactions. Consequently, distonic
ions have charge and radical at separate atoms in a conventional valence bond
description.
12. Note: The mere occurrence of the [C7H7]+ ion especially when it is of low
intensity is not sufficient to prove that the analyte belongs to the
phenylalkanes. This ion and its characteristic fragments may also be observed
if there is some way at all to generate a [C7H7]+ fragment ion.
26
27
18. Phthalates are commonly used plasticizers in synthetic polymers. Especially di2-ethyl hexylphthalate (also known as dioctyl phthalate, DOP) accounts for 45%
of all plasticizer usage. Unfortunately, they are extracted from the polymer by
elongated exposure to solvents such as dichloromethane, chloroform or toluene,
e.g. from syringes, tubing, vials etc. Therefore, they are often detected as
impurities. They are easily recognized from their typical peaks at m/z 149 (often
base peak), 167 and [M-(R-2H)]+ (m/z 279 for DOP). The molecular ions are
often absent in their EI spectra.
28
11
FATTY ACIDS
(Derivatives for MS Structure Determinations)
Benzyl Esters
Fragments characteristic of the acylium cations
are present at m/z (M-107) formed by loss of
benzyloxy radicals and at m/z (M-91) (loss of the
benzyl radical) and m/z (M-91-18) (loss of the
benzyl radical and water molecule).
The McLafferty-rearranged ion at m/z 150 is
observed only in low abundance, whereas the
corresponding ions at m/z 74 usually form the
base peak in spectra of straight-chain methyl
esters.
Generally, spectra of benzyl esters appear to give
less information about the hydrocarbon skeleton
of the acid than those of the corresponding methyl
esters. As a typical example, the mass spectrum of
benzyl
dodecanoate
is
shown
(fig.8-23).
13
14
16
CI Spectrum
20
21
22
23
24
25
26
189
70
-18
-H
207
HO
COOH
248
-45
203
-70
133
27
28
COOH
CH3O
+H
167
29
The presence of an oxo or hydroxyl substituent at
C-6 as in methyl 6-oxo-3-hydroxy-12-oleanen-28- oate
and methyl sumaresinolate), respectively, leads to
scission of the 7,8-and 9,10-bonds with
formation of
an ion of mass 302 consisting of rings
C, D, and E)
30
32
33
Flavonoids
34
3'
2'
8
6'
and/or
5'
2
4'
HC
+O
+
+ O
A
O
Flavonols
38
39
40
RESOLUTION
Magnetic instruments function at constant resolution
R = m/m
or
m m
As a result, d m varies proportionality to m.
In the low-mass range m is small.
In the high-mass range m is large.
Suppose the resolution is adjusted to 1000, then for an ion with mass exactly
100.00,
m = m/R
= 100/1000 = 0.1
For an ion with mass 1000,
m = 1000/1000 = 1
i.e. for ion with mass100, the instrument is scanning from mass 99.95 to 100.05
for ion with mass 1000, this ranges from 999.5 to 1000.5.
DETECTORS
Photographic plates
The ions impinge on a metal plate connected to
earth
through
a
resistor.
The
resulting
neutralization of the charge on the ions leads to a
flow of current through the resistor, which can be
detected and affords a measure of ion abundance
(disadvantage lack of sensitivity).
Electron multiplier
Consists of a series of 10 electrodes; provides gain
of the order of 106.
A+
even
B+
odd
odd
even
even
even
Name:
Q.1.
Q.2.
Q.3.
Q.4.
EI.
Q.5.
Q.6.
Q.7.
Figure 1.6 contributions of carbon and chlorine isotopes to the pattern of peaks in the
molecular ion region of a dichlorobenzene. The molecular ion (M+*) is usually
considered to be the peak having contributions from the isotopes of lowest mass but
strictly all six ions are molecular ions.
REFERENCES
3'
2'
8
7
5'
2
4'
6'
and/or
A
C
HC
35
3'
2'
8
6'
and/or
5'
2
4'
HC
+O
+
+ O
A
O
1)
4+
2)
5+
3)
6+
4)
7+
5)
8+
6)
9+
= Mr + 4/4
= 5068/4
= 1267
[1267 x 4 = 5068 - 4 = 5064]
= 5064 + 5/5
= 1013.8
[1013.8 x 5 = 5069 - 5 = 5064]
= 5064 + 6/6
= 845
[845 x 6 = 5070 - 6 = 5064]
= 5064 + 7/7
= 724.428
[724.43 x 7 = 5071 - 7 = 5064]
= 5064 + 8/8
= 634
[634 x 8 = 5072 - 8 = 5064]
= 5064 + 9/9
= 563.666
[563.666 x 9 = 5073 - 9 = 5064]
P.1 of 8
Q.1. (b) and (c) are the spectra of two isomeric hydrocarbons with molecular mass 150
(not observed in the spectra). Explain the ions at m/z 43, 57 and 71 with respect to their
relative abundances for the two compounds (2 Marks).
O
CH3
CH2 C
Q.3. Explain the ions at m/z 49, 51, 56, 57, 77 and 79 in the following
spectrum. Draw the formation of ion at m/z 56 showing electron
transfer (2.5 Marks).
Q.4. The following compound has two hydroxyl substituents. Place these
in rings A/B in the light of fragment ions at m/z 136 and 118 (2 Marks).
3'
2'
8
5'
2
4'
6'
and/or
A
C
B
HC
C15H10O4
M ; m/z 254
Q.5. The following compound has the molecular ion peak at m/z 488
and besides the OH at C-3 and a COOH at C-18, you have to place two
more OH groups. The mass spectrum has significant ions at m/z 248,
203, 133, 239, 221 and 203. Suggest in which of the rings A/B, C, D or E,
these OH groups should be placed (2 Marks).
E
18
D
COOH
+ 2 x OH
HO
C30H48O5 ; M+ 488
11
home | disclaimerCopyright 2000-2009 IonSource All rights reserved. Last updated: Saturday, February 26, 2005
01:43:20 PM
21
r = mv
(1)
eH
that is,
22
r2 = m2v2
e2H2
(2)
v2 = r2e2H2
m2
(3)
Now given that the mass of the ions is m and their charge e, and that
their initial kinetic energy is negligible then their kinetic energy after
they have been accelerated to velocity v must be equal to the
electrostatic energy acquired in passing through the applied voltage V.
That is,
m2 = eV
or
2 = 2eV
m
(4)
(5)
(6)
24
Benefits
25
Limitations
Not well-suited for pulsed ionization
methods (e.g. MALDI)
Usually larger and higher cost than other
mass analyzers
Linked scan MS/MS gives either limited
precursor selectivity with unit product-ion
resolution, or unit precursor selection with
poor product-ion resolution
26
Applications
All organic MS analysis methods
Accurate mass measurements
27
28
29
19
Peak Matching
Comparing the masses of two compounds simultaneously ionized in
the source.
One unknown
One known with accurate mass.
b/c
mk
2 Vk
munk
20
r2 B2
r2 B2
2 Vunk
16
Mass Analysis
Single focusing
(Low resolution)
Double focusing
(High resolution)
MS/MS EXPERIMENTS
These experiments can do more than
that and a diverse type of data can be
obtained instead of only obtaining a
mass-to-charge ratios. This requires
consideration of analyzer parameters
and principles involved in the analysis
of charged particles to understand the
actual physical property on which the
instrument operation is based.
2
The kinetic energy of an ion after it has been fully accelerated from the ion
source is:
(1)
m2 = eV
2
or m2 = 2eV
m = mass of a single ion
= velocity of the ion
e = charge of the ion
V = potential drop though which ion is accelerated
Electric Sectors
The equation of motion of an ion in an electric field is:
m2
r
eE
(2)
r = radius of deflection
E = electric field strength
Rearranging wil give:
m2
= Er
e
In coventional operation at a:
(3)
Therefore the ions formed in the ion source travel in a path that
follows this radius of curvature.
If an ion fragments after it has been accelerated, but in a non-accelerating field, its
mass will change
will be same
Magnetic Sectors
An ion in a magnetic field B (or H) travels in a circular path in a plane normal to the direction of
the magnetic field. The motion is described by:
m2/r = eB
(4)
m/e = Br
(5)
By substituting the expression for the ion velocity from equation (1)
m2/2 = eV
(1)
eV
m2
r
(1)
eB
(4)
From eq (1),
if = const
Vd is increased to Vp, then md must increase to mp
If,
mp fragments to md prior to the first sector,
The mass
ion is determined from the equation:
m V of
/ Vthe=parent
m
8
Linked Scans
Two of the fields are scanned simultaneously.
When an ion fragments in a reaction region and experiences no acceleration or
deceleration, then,
d = p
The trajectory of an ion through a magnetic field and an electric field is a function of both
m and
By maintaining the ratio of the fields such that the ion velocity necessary to pass through
both sectors is constant, a daughter ion scan is performed
Division of eq. (5) by (3) gives,
m/e = Br
(5)
m2/e = Er
(3)
= E/B
or
9
B/E = 1/
to determine precursor
ions of a chosen product
B/E
B2/E
10