Vous êtes sur la page 1sur 66

BASICS OF WAVEFORM

INTERPRETATION
RET 2284
Principles of Mechanical Ventilation

Objectives
Identify graphic display options
provided by mechanical ventilators.
Describe how to use graphics to
more appropriately adjust the patient
ventilator interface.

Introduction
Monitoring and analysis of graphic display
of curves and loops during mechanical
ventilation has become a useful way to
determine not only how patient are being
ventilated but also a way to assess
problems occurring during ventilation.

Uses of Flow, Volume, and


Pressure Graphic Display
Confirm mode functions

Detect auto-PEEP
Determine patient-ventilator synchrony
Assess and adjust trigger levels
Measure the work of breathing
Adjust tidal volume and minimize overdistension
Assess the effect of bronchodilator administration
Detect equipment malfunctions
Determine appropriate PEEP level

Uses of Flow, Volume, and


Pressure Graphic Display
Evaluate adequacy of inspiratory time in pressure
control ventilation
Detect the presence and rate of continuous leaks
Assess inspiratory termination criteria during
Pressure Support Ventilation
Determine appropriate Rise Time

Measured Parameters
Flow
Pressure
Volume
Time

Most Commonly Used


Waveforms (Scalars)
Pressure vs. Time
Flow vs. Time
Volume vs. Time

Pressure vs. Time Curve


30
A

C
PIP
Baseline

P
aw
cmH O

Mean Airway Pressure

Sec

1
-10

Pressure-Time Curve
20

Volume Ventilation

Pressure Ventilation

Paw
cmH2O

Sec

Patient Triggering
30

P
aw
cmH O

Sec

1
-10

Adequate Flow During


Volume-Control Ventilation
30

Adequate flow

Paw
cmH2O
1
-10

Time (s)

Inadequate Flow During


Volume-Control Ventilation

30

Adequate flow

Paw
cmH2O

Flow set too low

1
-10

Time (s)

Patient/Ventilator Synchrony
Volume Ventilator Delivering a Preset Flow and Volume

Adequate Flow

Paw

Sec

cmH2O
1

-20

Patient/Ventilator Synchrony
The Patient Outbreathing the Set Flow

Air Starvation

Paw

Sec

cmH2O
1

-20

Plateau Time
30

Inadequate plateau time

Paw
cmH2O

-20

SEC

Plateau Time
30

Adequate Plateau Time

Paw
cmH2O

-20

SEC

Flow vs.Time Curve


120
INSP

Inspiration

.
V

LPM

120

SEC

EXH

Flow vs.Time Curve


120
INSP

Inspiration

.
V

LPM

SEC

Expiration
120

EXH

Flow vs.Time Curve


120

Constant Flow

Descending Ramp
INSP

Inspiration

.
V

LPM

120

SEC

EXH

Flow-Time Curve
120
INSP

Insp. Pause

LPM

SEC

Expiration

120

EXH

Inspiratory Time
Short Normal Long

Expiratory Flow Rate and Changes


in Expiratory Resistance
120

SEC

LPM

-120

A Higher Expiratory Flow Rate and a


Decreased Expiratory Time Denote a Lower
Expiratory Resistance
120

SEC

LPM

120

Obstructed Lung

Delayed flow return

Pressure-Time and Flow-Time Curves

20

Volume Ventilation

Paw

Expiration

cmH2O

Sec

V.

Pressure-Time and Flow-Time Curves


Different Inspiratory Flow Patterns
20

Volume Ventilation

Paw

Expiration

Inspiration

cmH2O

Sec

V.

Pressure-Time and Flow-Time Curves

20

Pressure Ventilation

Volume Ventilation

Inspiratory Time

Paw
cmH2O

Sec

V.

Rise Time
How quickly set pressure is reached

Flow Acceleration Percent


Rise Time
Minimal Pressure Overshoot

P
Slow rise

Moderate rise

Fast rise

.
V
Pressure Relief
Time

Patient / Ventilator Synchrony

Volume Ventilation Delivering a Preset Flow and Volume


30

Adequate Flow

Paw

Sec

cmH2O
1

-20

Patient -Ventilator Synchrony


The Patient Is Outbreathing the Set Flow

30

Air Starvation

Paw

Sec

cmH2O
1

-20

If Peak Flow Remains the Same, I-Time


Increases: Could Cause Asynchrony
120

.
V

SEC

LPM

-120

Changing Flow Waveform in Volume


Ventilation: Effect on Inspiratory Time
120

.
V

SEC

LPM

-120

Increased Peak Flow: Decreased


Inspiratory Time
120

.
V

SEC

LPM

-120

Detecting Auto-PEEP
120

.
V

SEC

LPM

Zero flow at end exhalation indicates


equilibration of lung and circuit pressure
-120

Note: There can still be pressure in the lung behind


airways that are completely obstructed

Detecting Auto-PEEP
120

.
V

SEC

LPM

120

The transition from expiratory to inspiratory


occurs without the expiratory flow returning
to zero

Volume vs.Time Curve


800 ml

Inspiration

VT

SEC

Volume vs.Time Curve


800 ml

Expiration
VT
SEC

Typical Volume Curve


I-Time
E-Time

1.2

A
VT
Liters

SEC

-0.4
A = inspiratory volume
B = expiratory volume

Leaks
1.2
A

VT
Liters

SEC

-0.4
A = exhalation that does not return to zero

Setting Appropriate I-Time


600 cc

450 cc

VT

SEC

120

.
V
LPM

SEC

1
120

Setting Appropriate I-Time


600 cc

450 cc

VT

500 cc
SEC

120

.
V
LPM

Lost VT
SEC

1
120

Loops

Pressure-Volume Loops
Flow-Volume Loops

Pressure-Volume Loop
VT
LITERS

0.6

0.4

0.2

Paw
cmH2O -60

40

20

20

40

60

Mandatory Breath
VT
LITERS

0.6

0.4

Inspiration
0.2

Paw
cmH2O -60

40

20

20

40

60

Mandatory Breath
VT

Counterclockwise

LITERS

0.6

Expiration
0.4

Inspiration
0.2

Paw
cmH2O -60

40

20

20

40

60

Spontaneous Breath
VT

Clockwise

LITERS

0.6

0.4

Inspiration

0.2

Paw
cmH2O -60

40

20

20

40

60

Spontaneous Breath
VT

Clockwise

LITERS

0.6

0.4

Inspiration

Expiration
0.2

Paw
cmH2O -60

40

20

20

40

60

Work of Breathing
VT
LITERS

0.6

0.4

0.2

Paw
cmH2O

-60

-40

-20

20

40

60

Assisted Breath
VT
LITERS

0.6

0.4

Assisted
Breath

0.2

Paw
cmH2O -60

40

20

20

40

60

Assisted Breath
VT
LITERS

0.6

0.4

Assisted
Breath

0.2

Inspiration

Paw
cmH2O -60

40

20

20

40

60

Assisted Breath
VT

Clockwise to Counterclockwise

LITERS

0.6

Expiration
0.4

Assisted
Breath

0.2

Inspiration

Paw
cmH2O -60

40

20

20

40

60

Pressure-Volume Loop Changes


VT
LITERS

0.6

0.4

0.2

Paw
cmH2O

-60

-40

-20

20

40

60

Changes in Compliances
VT
LITERS

Indicates a drop in compliance


(higher pressure for the same
volume)

0.6

0.4

0.2

Paw
cmH2O

-60

40

20

20

40

60

Overdistension
VT

A = inspiratory pressure

LITERS

B = upper inflection point

0.6

C = lower inflection point


0.4

0.2

Paw
cmH2O

C
-60

-40

-20

20

40

60

Lung Overdistension

Normal Flow-Volume Loops

Flow -Volume Loops


Volume Control

Flow

Tidal Volume

Inspiration
Volume
Expiration

Flow -Volume Loops


Volume Control
Tidal Volume

Peak Inspiratory Flow

Flow

Peak Expiratory Flow

Inspiration
Volume
Expiration

ETT or Circuit Leaks

Obstructive Pattern

Bronchodilator Response
BEFORE
3
2

V
LPS
1
2
3

V
LPS

Bronchodilator Response
BEFORE

AFTER
Worse

V
LPS

V
LPS

Bronchodilator Response
BEFORE

AFTER
Better

Worse
3

3
INSP

V
LPS

V
LPS

V
LPS

VT

EXH

Remember!
Waveforms and loops are graphical representation of the
data generated by the ventilator.
Typical Tracings
Pressure-time,
Flow-time,
Volume -time
Loops
Pressure-Volume
Flow-Volume

Assessment of pressure, flow and volume


waveforms is a critical tool in the management
of the mechanically ventilated patient.