Académique Documents
Professionnel Documents
Culture Documents
1/2
1/2
0 1/2 1/2
Nuclear Spins in B0
In the absence of applied magnetic field. All spin
states of a given nucleus are of equivalent
energy(degenerate), and in a collection of
atoms all of the spin states should be almost
equally populated, with the same number of
atoms having each of the allowed spins
E.g. within a collection of 1H and 13C atoms,
nuclear spins are completely random in
orientation.
When placed in a strong external magnetic field
of strength B0, however, interaction between
nuclear spins and the applied magnetic field is
quantized. The result is that only certain
orientations of nuclear magnetic moments are 4
Nuclear Spins in B0
for 1H and 13C, only two orientations
are allowed.
Nuclear Spins in B0
The applied magnetic resonance phenomenon
occurs when nuclei with an applied field are
induced to absorb and change their spin
orientation with respect to the applied field
In an applied field strength of 7.05T the
difference in energy between nuclear spin
states for
1H is approximately 0.120 J (0.0286 cal)/mol,
which corresponds to a frequency of 300
MHz (300,000,000 Hz).
13C is approximately 0.030 J (0.00715
cal)/mol, which corresponds to a frequency
of 75MHz (75,000,000 Hz).
Nuclear Spin in B0
The energy difference between allowed
spin states increases linearly with applied
field strength.
Values shown here are for 1H nuclei.
Nuclear Magnetic
Resonance
If the precessing nucleus is
irradiated with electromagnetic
radiation of the same frequency as
the rate of precession,
the two frequencies couple
energy is absorbed
the nuclear spin is flipped from spin
state +1/2 (with the applied field) to
-1/2 (against the applied field).
9
Energy Absorption
The nuclear magnetic resonance phenomenon occurs when
nuclei aligned with an applied field are induced to absorb
energy and change their spin orientation with respect to the
applied field.
The energy absorption is quantized process, and the energy
absorbed must equal the energy difference between the two
states involved.
Eabsorbed = (E-2 state E+2 state ) = h
In practice, this energy difference is a function of the strength
of the applied magnetic field, Bo . The stronger the applied
magnetic field, the greater the energy difference between the
two states.
10
Energy Absorption
The magnitude of the energy- level separation also
depends on the particular nucleus involved. Each
nucleus (hydrogen, chlorine, and so on) has a
different ratio of magnetic to angular momentum,
since each has different charge and mass. This ratio,
called the magnetogyric ratio, , is a constant for
each nucleus and determines the energy
dependence on the magnetic field:
E = f(Bo ) = h
Since the angular momentum of the nucleus is
quantized in units of h/2, the final equation takes
the form
E = (h/2)Bo = h
= ( /2)Bo
11
12
Nuclear Magnetic
Resonance
If we were dealing
with H nuclei isolated from
1
300 x 10 Hz
10
= 1.2 ppm
15
CH3 Si CH3
CH3
Tetramethylsilane (TMS)
NMR Spectrometer
Essentials of an NMR spectrometer are a
powerful magnet, a radio-frequency
generator, and a radio-frequency
detector.
The sample is dissolved in a solvent,
most commonly CDCl3 or D2O, and
placed in a sample tube which is then
suspended in the magnetic field and set
spinning.
Using a Fourier transform NMR (FT-NMR)
spectrometer, a spectrum can be
recorded in about 2 seconds.
17
NMR Spectrometer
Schematic diagram of a nuclear
magnetic resonance spectrometer.
18
19
NMR Spectrum
1H-NMR spectrum of methyl acetate.
Equivalent Hydrogens
Equivalent hydrogens: Hydrogens that
have the same chemical environment.
A molecule with 1 set of equivalent
hydrogens gives 1 NMR signal.
O
CH3 CCH3
ClCH2 CH2 Cl
C
H3 C
Propanone
(Acetone)
CH3
H3 C
1,2-Dichloroethane
Cyclopentane
C
CH3
2,3-Dimethyl2-butene
21
Equivalent Hydrogens
Cl
O
Cyclopentanone
(2 signals)
CH3
C C
H
H
(Z)-1-Chloropropene
(3 signals)
Cyclohexene
(3 signals)
22
23
Chemical
Shifts
1
H-NMR
Chemical
Type of
Hydrogen
Shift ()
(CH3 ) 4 Si
0 (by definition)
RCH3
0.8-1.0
RCH2 R
1.2-1.4
R3 CH
1.4-1.7
R2 C=CRCHR2 1.6-2.6
RC CH
2.0-3.0
ArCH3
2.2-2.5
ArCH2 R
2.3-2.8
ROH
0.5-6.0
RCH2 OH
3.4-4.0
RCH2 OR
3.3-4.0
R2 NH
0.5-5.0
O
RCCH3
2.1-2.3
O
RCCH2 R
2.2-2.6
Type of
Hydrogen
O
RCOCH3
O
RCOCH2 R
RCH2 I
RCH2 Br
RCH2 Cl
RCH2 F
ArOH
R2 C=CH2
R2 C=CHR
ArH
O
RCH
O
RCOH
Chemical
Shift ()
3.7-3.9
4.1-4.7
3.1-3.3
3.4-3.6
3.6-3.8
4.4-4.5
4.5-4.7
4.6-5.0
5.0-5.7
6.5-8.5
9.5-10.1
10-13
24
Chemical Shift
Chemical shift depends on the (1)
electronegativity of nearby atoms, (2)
hybridization of adjacent atoms, and (3)
diamagnetic effects from adjacent pi bonds.
Electronegativity
CH3-X
Electronegativity of X
Chemical
Shift ()
CH3F
CH3OH
CH3Cl
CH3Br
CH3I
4.0
3.5
3.1
2.8
2.5
4.26
3.47
3.05
2.68
2.16
(CH3) 4C
(CH3) 4Si
2.1
1.8
0.86
0.00
25
Chemical Shift
Hybridization of adjacent atoms.
Type of Hydrogen
(R = alkyl)
Name of
Hydrogen
Chemical
Shift ()
RCH3 , R2 CH2 , R3 CH
Alkyl
0.8 - 1.7
R2 C=C(R)CHR2
Allylic
1.6 - 2.6
RC CH
Acetylenic
2.0 - 3.0
R2 C=CHR, R2 C=CH2
Vinylic
4.6 - 5.7
RCHO
Aldehydic
9.5-10.1
26
Chemical Shift
Diamagnetic effects of pi bonds
A carbon-carbon triple bond shields an
acetylenic hydrogen and shifts its signal to
lower frequency (to the right) to a smaller
value.
A carbon-carbon double bond deshields
vinylic hydrogens and shifts their signal to
higher frequency (to the left) to a larger
Chemical
value.Type of H
Name
Shift ()
RCH3
RC CH
R2 C=CH2
Alkyl
Acetylenic
Vinylic
0.8- 1.0
2.0 - 3.0
4.6 - 5.7
27
28
29
30
31
Signal Splitting (n + 1)
1H-NMR spectrum of 1,1dichloroethane.
32
34
36
37
Coupling Constants
Coupling constant (J): The distance between
peaks in a split signal, expressed in hertz.
The value is a quantitative measure of the
magnetic interaction of nuclei with coupled
spins.
Ha
Ha Hb
C C
6-8 Hz
Ha
Hb
11-18 Hz
Ha
H
Hb a
Hb
Hb
8-14 Hz
Ha
0-5 Hz
Ha
Ha
Hb
0-5 Hz
Hb
8-11 Hz
Hb
C
5-10 Hz
0-5 Hz
38
39
40
41
Coupling Constants
An important factor in vicinal coupling is
the angle between the C-H sigma bonds
and whether or not it is fixed.
Coupling is a maximum when is 0 and
180; it is a minimum when is 90.
42
43
In ethyl propenoate, an
unsymmetrical terminal alkene, the
three vinylic hydrogens are
nonequivalent.
44
45
46
47
49
13
C-NMR Spectroscopy
50
52
C-NMR Spectroscopy
13
13
C-NMR Spectroscopy
54
55
Chemical
Shift ()
RCH3
RCH2R
10-40
15-55
20-60
R3CH
RCH2I
RCH2Br
0-40
25-65
RCH2Cl
R3COH
35-80
R3COR
40-80
65-85
RC CR
R2C=CR2
40-80
100-150
C-NMR
13
Type of
Carbon
C R
Chemical
Shift ()
110-160
O
RCOR
160 - 180
O
RCNR2
165 - 180
O
RCCOH
165 - 185
O
O
RCH, RCR
180 - 215
56
57
59
61
63
64
65
Introduction
Two-dimensional
2D
NMR is a set ofto
NMR
methods which give data
nuclear
plotted
in a magnetic resonance
spectroscopy
(2Dfrequency
NMR) axes rather than
space
defined by two
one.
Examples of types of 2D NMR
correlation spectroscopy (COSY),
exchange spectroscopy (EXSY), and
Nuclear Overhauser effect spectroscopy (NOESY).
Two-dimensional NMR spectra provide more
information about
a molecule than one-dimensional NMR spectra
and are
especially useful in determining the structure of a
molecule,
particularly for molecules that are too complicated
to work with
using one-dimensional NMR.
66
Introduction to 2D NMR
Each
experiment
consists ofexperiments
a sequence have
of radio
Almost
all two-dimensional
four
frequency
stages:
pulses with
delay
in between
them.
(RF)
preparation
period
- aperiods
magnetization
coherence
It isisthe
created
timing,
frequencies,
intensities of these
through
a set of RFand
pulses;
pulses
that period, a determined length of time
evolution
distinguish different NMR experiments from one
during
another
which no pulses are delivered and the nuclear
spins are
allowed to freely precess (rotate);
mixing period, where the coherence is
manipulated by
another series of pulses into a state which will
give an
observable signal; and
67
detection period, in which the free induction
Introduction to 2D NMR
Introduction to 2D NMR
Types of 2D NMR
1. Homonuclear through-bond
correlation
methods
In
these methods,
magnetization transfer occurs
between nuclei of the same type, through j-coupling of
nuclei connected by up to a few bonds.
69
70
Mass Spectroscopy
Mass Spectrometry is an analytical spectroscopic tool
primarily concerned with the separation of molecular
(and atomic) species according to their mass.
What information can be determined?
Molecular weight
Molecular formula (HRMS)
Structure (from fragmentation fingerprint)
Isotopic incorporation / distribution
Protein sequence (MS-MS)
72
73
Mass Spectroscopy
Mass spectrometry is the study of systems causing
the
formation of gaseous ions, with or without
fragmentation,
which are then characterized by their mass to
charge ratios
(m/z) and relative abundances.
In MS, compounds are ionized, ionized molecule
decomposes
into smaller ions/radicals/radical ions/ neutrals.
One way to ionize molecules is to extract electrons
from a
molecule.
The positively charged fragments produced are
74
separated,
75
Diagram of mass
spectrometer
instrument
Sample
Inlet
System
Ion
Source
Data
System
Mass
Analyzer
Ion
Detector
Vacuum
Pumps
Data
76
77
Electron Ionization
Electron Ionization (EI) is the most common
ionization technique
used for mass spectrometry.
EI works well for many gas phase molecules, but it
does have
some limitations.
Although the mass spectra are very reproducible
and are widely
used for spectral libraries, EI causes extensive
fragmentation so
that the molecular ion is not observed for many
compounds.
Electron
Ionization Source
79
Electron Ionization
The electrons used for ionization are produced by
passing a
current through a wire filament (see above Figure).
The amount of current controls the number of
electrons emitted
by the filament.
An electric field accelerates these electrons across
the source
region to produce a beam of high energy electrons.
When an analyte molecule passes through this
electron
a
Knockbeam,
an electron out of the molecule generates
valence shell electron can be removed
from the
the Molecular ion (M+.)
M+. fragments
molecule to
All charged species are accelerated through a
produce an ion.
magnetic field
based
on further
their mass-to-charge
Depending
on the
energy
fragmentation
ratio
(m/z)
1
eV
(per
molecule
may
take
place or atom) = 96.415 220 6
80
kJ/mole
Chemical Ionization
In Chemical Ionization the source is enclosed in a small cell with
openings for
the electron beam, the reagent gas and the sample.
The reagent gas is added to this cell at approximately 10 Pa
(0.1 torr) pressure.
This is higher than the 10-3 Pa (10-5 torr) pressure typical for a
mass
spectrometer source.
At 10-3 Pa the mean free path between collisions is
approximately 2 meters and
ion-molecule reactions are unlikely. In the CI source, however,
the mean free
path between collisions is only 10-4 meters and analyte
molecules undergo
many collisions with the reagent gas.
The reagent gas in the CI source is ionized with an electron
beam to produce a
+
CH5in
ion. cloud react and
cloud of ions. The reagent gas ions
this
81
produce adduct ions
Chemical Ionization
When analyte molecules (M) are introduced to a source
region with
this cloud of ions, the reagent gas ions donate a proton to
the analyte
molecule and produce MH+ ions.
The energetics of the proton transfer is controlled by using
different
reagent gases.
The most common reagent gases are methane, isobutane
and ammonia.
Methane is the strongest proton donor commonly used with
a proton
affinity (PA) of 5.7 eV. For softer ionization, isobutane (PA
8.5 eV) and
ammonia (PA 9.0 eV) are frequently used.
Fragmentation is minimized in CI by reducing the amount of
excess
energy produced by the reaction.
82
Because the adduct ions have little excess energy and are
A) EI of reagent gas to
form ions:
B) Reaction
CH4 + e of reagent
CH4++gas ions to form
adducts:
+2e
OR
83
Electrospray ionization:
Pressure = 1 atm
Inner tube diam. = 100 um
Partial
vacuu
m
MH+
N2
Sample in solution
N2 gas
+
+ ++
++
++++
++
+
+ ++
++
+
++
+
++
+
++
+
++
+
++
+
++
+ +
+
+
+
+
+
MH2+
MH3+
High voltage applied
to metal sheath (~4 kV)
Charged droplets
87
88
MALDI Ionization
Summary
Ionisation methods
Electron Impact ionisation (EI)
Strong ionisation method
Useful for identification
Structural information due to excessive
fragmentation
Commercial libraries
Chemical Ionisation (CI)
Soft ionisation method
Useful for molecular weight
determination
Reactant gas required
Less structural information
92
Ionisation methods
Atmospheric Pressure Chemical Ionisation
(APCI)
Soft ionisation
For aqueous samples and smaller
molecules
No commercial libraries
Electrospray Ionisation (ESI)
Soft ionisation
For aqueous samples and large
molecules
Multiply charged ions large
molecules can be measured (MW
100,000)
No commercial libraries
93
Ionisation methods
Atmospheric Pressure Photoionisation
(APPI)
Soft ionisation
For non-polar molecules
No commercial libraries available
New technique (introduced in 2000)
Matrix-Assisted Laser Desorption
Ionisation (MALDI)
Sample mixed with suitable matrix
Sample is hit with energetic laser beam
ions are formed and desorbed from
the sample
Soft ionisation
Large molecules can be ionised
94
MASS ANALYZERS:
After ions are formed in the source region they are accelerated
into the mass
analyzer by an electric field. The mass analyzer separates
these ions according
to their m/z value.
The selection of a mass analyzer depends upon the resolution,
mass range,
scan rate and
detectiondescribed
limits required
for an
application.
Analyzers
are typically
as either
continuous
or pulsed.
Each
analyzer analyzers
has
Continuous
include quadrupole, filters and magnetic
very different operating characteristics and the selection of an
sectors.
instrument
These analyzers are similar to a filter or monochromator used for
involves
optical important tradeoffs.
spectroscopy. They transmit a single selected m/z to the
detector and the
mass spectrum is obtained by scanning the analyzer so that
different mass
to charge ratio ions are detected.
While a certain m/z is selected, any ions at other m/z ratios are
lost,
95
reducing the S/N for continuous analyzers. Single Ion
96
m1
m2
m4
m3
m4
m3 m2
m1
m1
m2
m4
m3
m2
m2
m2
m2
detector
Source
98
Magnetic Sectors
Magnetic Sectors are the traditional means of
separating ions of different
masses using the principle that when a beam of ions
passes through a
magnetic field they are deflected through an arc whose
radius is related to
the ratio of mass to charge.
By varying the field in a known manner, i.e. sweeping from
high to low field
with time, ions of different mass arrive at the detector
sequentially and
produce the mass spectrum.
99
Magnetic Sector
Magnetic Sector instruments have higher
resolution and
greater mass range than quadrupole instruments,
but they
require larger vacuum pumps and often scan more
slowly.
The typical mass range is to m/z 5000, but this may
be extended
to m/z 30,000.
Magnetic sector instruments are often used in
series with an
electric sector, described above, for high resolution
and tandem
mass spectrometry experiments.
Magnetic
Sector Mass
Spectrometer.
The low(Figure below)
Magnetic
sector
instruments
m/z ion (B+) is separated from the high m/z
100
separate
ions
in
a
ion (A+).
Magnetic Sector
Ions are accelerated from the source region into
the magnetic
sector by a 1 to 10 kV electric field. This
acceleration is
significantly greater than the 100 V acceleration
typical for a
quadrupole instrument.
Since the ions are charged, as they move through
the magnetic
sector, the magnetic field bends the ion beam in
an arc. This is
the same principal that causes electric motors to
turn.
The radius of this arc (r) depends upon the
101
momentum of the
Magnetic Sector
Ions with greater momentum will follow an arc
with a larger
radius. This separates ions according to their
momentum, so
magnetic sectors are often called momentum
analyzers.
The momentum of the ion is the product of the
mass (m)
and the velocity (v). The charge of the ion is
the product of
The
velocity
an ion is
by the
the
chargeofnumber
of determined
the ion (z) and
the
acceleration
charge of an
voltage
in (e).
the source
regionthese
(V) and
the mass
electron
Substituting
variables
intoto
charge
ratio
eq. 1 yields
(m/z) of the ion. Equation 2 rearranges to give
102
the m/z ion
Top View
R in g E le c t r o d e
E n tra n c e
E ndcap
E le c t r o d e
E x it E n d c a p
E le c tr o d e
R in g E le c tr o d e
Benefits
High sensitivity
Multi-stage mass spectrometry
Compact and cheap mass analyzer
Limitations
Poor quantitation
Poor dynamic range
Subject to space charge effects and ion molecule
reactions ([M+H]+ formation in EI); at least older
GC instruments
105
- Linear Quadrupole
Ion Trap (LIT)
- -Also called 3D ion
trap
107
Others
108
Summary
..\..\Files from old computer\
English+books
\Spectrometric Identification of Orga
nic
Compunds 7th ed - R[.pdf
109
Detectors
These are usually electron multipliers which
produce up to 106 electrons for every ion that hits
the detectors. This produces a current which can be
amplified and measured.
Electron multipliers
Electron multipliers...
111
sources: http://www.chm.bris.ac.uk/ms/theory/detection.html &
Spectrum of air
Contains: H2O (MW 18), N2 (MW 28), O2
(MW
32), CO2 (MW 44)
112
100%
X-axis:mass/charge (m/z);
Note: not just mass of the ion
VERIFIN 113
Molecular ion
114
Molecular Ion
The molecular ion provides the molecular mass of the analyte
and is the first clue used to interpret a mass spectrum.
The mass of the molecular ion is based upon the mass of the
most abundant
isotope for each element in the molecule. This is not the atomic
weight from
the periodic table. Since many mass spectrometers have unit
mass
resolution, the atomic mass is rounded to the nearest whole
number, this is
called the nominal mass. For example the molecular ion for
CHBr3 is observed
at m/z 250, not at the formula weight of 253 (CHBr3 ; (12 + 1 +
379) = 250).
In many mass spectra, the molecular ion is easily identified as
the ion with the
highest mass to charge ratio. However, this assignment should
be made with
115
caution because the highest mass to charge ion be an impurity
Fragmentation
Although the molecular ion is useful for identification, it does
not provide any
structural information about an unknown.
The structural information is obtained from the fragmentation
patterns of the
mass spectrum. Identifying an unknown without analyzing
the fragmentation
patterns is like putting together a jigsaw puzzle without the
picture.
Fragmentation patterns are often complex, but they fit
together like pieces of
the puzzle to identify the structure of the molecule.
Fragmentation mechanisms
116
Fragmentation
After a molecule is ionized, the molecular ion retains the
excess ionization
energy. If this excess energy is greater than the energy
required to break a
chemical bond, the molecule can fragment.
The fragmentation processes are typically categorized as
direct cleavage or
Rearrangement Cleavage reactions are simply the breaking
of a bond to
produce two fragments. These reactions usually produce an
even electron
ion (AB+).
The even electron ion is detected at an odd m/z value
(assuming no
nitrogen) and a neutral odd electron radical.* Since the
radical is a neutral
fragment it is not observed in the mass spectrum.
Rearrangements are more complex reactions that involve
both making and
117
breaking bonds. These reactions are thermodynamically
Fragmentation
Odd electron ions are radical species with an unpaired
electron. They are
produced by removing a lone pair electron or a bonding
electron from a
molecule. For example, water is ionized by removing a nonbonding electron
from oxygen to produces H2O+, . This is an example of an odd
electron ion.
Odd electron ions have an even m/z value. The exception to
this is if the ion has
an odd number of nitrogen atoms. Calculate the m/z value for
some ions to
verify this statement.
Even electron ions have all paired electrons. Even electron ions
are usually
produced by cleavage reactions that result in loss of a neutral
radical. An
.
example of this is the cleavage of CH3 CH2 CH3+. to form CH
1183
+
Isotope Peaks
The peaks from the isotopes
The intensity ratios (relative intensities) in the
isotope patterns
are arising from the natural abundance of the
isotopes, thus are
valuable to ascertain the atomic composition of ions.
M+1 peaks are primarily due the presence of 13C in
the sample
M+2 peaks useful for N, O calculation (< seven C
cpds)
M+2, M+4, .. indicative of presence of Br, Cl, S;
(79Br : 81Br 1:1, 35Cl : 37Cl 3:1)
119
120
121
122
123
124
125
Index of Hydrogen
Deficiency
IHD = number of pairs of
hydrogen that must be
removed from the
corresponding saturated
hydrocarbon.
IHD = The SUM of the number
of rings, double bonds, or 2*the
number of triple bonds.
Formula
Used to determine an
approximate structure
once all of the atoms
have been identified.
IHD = (2C + 2 H
X + N)/2 or
R + DB = C (H +
X)/2 + N/2 + 1 (round
down)
126
Structural Interpretation
Rings Plus Double Bonds Rule
Examples:
Ion Formula
CH3COCH2+
=1
Calculation
R+DB = 3 (5+0)/2 + 0/2 + 1 = 1.5
C5H4NCl+
R+DB = 5 (4-1)/2 + + 1 = 4
127
Structural Interpretation
Formation of Ions
Three factors determine which bond in a molecular ion will
break and which ions will form:
-stability of the ions formed
128
Structural Interpretation
Libraries
NIST/EPA/NIH library contains about 130,000
spectra with 10 million peaks.
Wiley library contains about 233,000 with 8
million peaks.
There are other smaller libraries that specialize in
certain kinds of compounds, i,e. drugs, poisons,
pesticides, flavors, and fragrances.
Of course manufacturers supply smaller
databases with their instruments.
129
Structural Interpretation
Libraries
Search routines often use a presearch to compare
8 most intense peaks in the unknown with 16
intense peaks in each library spectrum.
This reduces the number of candidates, usually to
about 100.
Ions are weighted for significance based on their
intensity and mass, higher mass ions are
considered more important.
130
131
Cycloalkanes
Cycloalkanes generally form strong molecular ion
peaks. Fragmentation via the loss of a molecule of
ethene (M-28) is common
- Strong M+
- Fragment ions: M-28, a series of peaks, M-15, M-29, M43, M-47,
etc .
e.g. pp 409 410
Alkenes
-The mass spectra of most alkenes show distinct
molecular ion peaks.
- Apparently, electron bombardment removes one f the
electrons in the pi bond, leaving the carbon skeleton
relatively undisturbed.
- Fragmentation to form an allyl cation (m/e = 41) is
favored
133
- Strong M+
ynes
e mass spectra of alkynes are very similar to those of alkene
e molecular ion peaks tend to be rather intense, and
agmentation patterns parallel those of the alkenes
rong M+,
agment ions: M/e = 39, strong M-1 peak. E.g. p 413
134
Generally
135
136
Hyphenation Techniques
Techniques in which a separation device is
combined with a detector, such as gas
chromatography-Fourier transform infrared
spectroscopy (GC-FTIR and GC-mass
spectrometry (GC-MS), have already become well
accepted analytical tools
GCMS should be considered a relatively mature
technique. One does not see major changes in
the technique and practices of this method
appearing in the current literature.
LCMS is an almost mature technique. One
occasionally sees a new innovation concerning
how this experiment is done. By distinct contrast
LCNMR is a technique still transitioning between
infancy and adolescence
137
LC-MS
Thermospray Interface
Mass Spectrometric Detectors
Direct Introduction of Total
Fundamental Problem
Effluent
Large Solvent Volume of
Flow: up to 2 mL/min
HPLC
Liquid Vaporized in Heated
High Vacuum Requirement of
Stainless Steel Capillary
MS
Aerosol of Solvent &
Solutions
Analyte Produced
Column Split
Analyte Ionized by
Uses a Small Faction of
Ammonium Salt in Eluent
Eluent
Simple Spectra (Soft
Best with Microbore
Ionization)
Columns
Disadvantages:
Flow: 10 to 50 L/min
Only polar solvents
Moving Belt/Moving Wire
(for ammonium salt)
Only polar analytes
Effluent Dripped onto
Column Wire
Advantage - Excellent
Evaporator
Solvent Evaporated MSDesorptionChamber
LOD: 1 to 10 pg
Desorption-Ioniztion in MS
MovingWire
cartoonschematic
138
System Configuration
Liquid
Chromatography
Ionization
Very
important!
Many columns
Many solvent
ESI
APCI
APPI
Mass Analyzer
Detector/
Data
Collection
Triple Quadrapoles
Ion-Traps
Hybrids
systems
139
140
GCMS
Methods in GCMS
Sample
introduction
Split
Splitless
On-column
Thermodesorption
Air Samples
Dynamic
Ionisation
Electron impact
Chemical
Ionisation
Ammonia
Isobutane
Mass Analysis
Quadrupole
Triple
Quadrupole
Ion trap
Time-of-flight
Magnetic Sector
Methane
Other
head-space
SPME
Etc.
142
Ion m/z
3-Dimensional Data
Ion Count
Retention time
143
Ion m/z
3-Dimensional Data
Ion Count
Retention time
144
EIC
Ion m/z
Ion Count
Retention time
145