Académique Documents
Professionnel Documents
Culture Documents
Ananya
Contents
Classification
History
Introduction
Indications
Key terms- compliance , ventilatory work
Components
Control mechanism
Variables
Triggering
Factors to consider in mechanical ventilation
Wave-forms
Classification
According to Robert chatburn
LOAD-
The pressure required to deliver the tidal volume is referred
to as the load that the muscles or ventilator must work
against.
load elastic ( α volume & inv. Prop t0 compliance)
resistance (α Raw & inspiratory flow)
Equation of motion for respiratory
system
Muscle pressure + ventilator pressure =
(volume / compliance)+ (resistance x flow)
It includes pistons
bellows
reducing valves
pneumatic circuits
Piston mechanism
Bellows mechanism
Pneumatic mechanism
Pneumatic circuits- uses pressurized gas as power
source.
these are microprocessor controlled with solenoid
valves.
use programmed algorithms in microprocessor to
open and close solenoid valves to mimic any flow or
pressure wave pattern.
Control circuit
Its the system that governs the ventilator drive
mechanism or output control valve.
Classified as-
Open circuits- desired output is selected and venti.
achieves it without any further input from clinician.
Closed circuits- desired output is selected and venti.
Measures a specific parameter (flow/vol/press)
continuously and input is constantly adjusted to
match desired output.
a.k.a SERVO controlled.
Control parameters
Pressure
Volume
Flow
Time
Ventilators deliver gas to the lungs using positive
pressure at a certain rate. The amount of gas
delivered can be limited by time, pressure or
volume. The duration can be cycled by time,
pressure or flow.
If volume is set, pressure varies…..if pressure is set,
volume varies…..
….according to the compliance…...
Mechanical- employs levers or pulleys to control drive
mechanism.
Pneumatic
Fluidic- applies gas flows and pressure to control
direction of other gas flows and to perform logic
functions based on the COANDA effect.
Electronic- uses resistors and diodes and integrated
circuits to provide control over the drive mechanism.
Pressure controller
Ventilator controls the trans-respiratory system
pressure .
This trans-respiratory system gradient determines the
depth or volume of respiration.
Based on this a ventilator can be positive or negative
pressure ventilator.
Volume controller
Volume cycled ventilation delivers a:
set volume;
with a variable Pressure - determined by resistance,
compliance and inspiratory effort
Flow controller
Allows pressure to vary with changes in patient s
compliance and resistance while controlling flow.
This flow is measured by vortex sensors or venturi
pnemotachometers.
Time controller
measures and controls inspiratory and expiratory time.
These ventilators are used in newborns and infants
Inspiratory time is a combination of the inspiratory flow
period and time taken for inspiratory pause. The following
diagram depicts how the addition of an inspiratory pause
extends total inspiratory time.
Normal inspiratory time of a spontaneously breathing healthy adult is approximately 0.
8- 1.2 seconds, with an inspiratory expiratory (I: E) ratio of 1:1.5 to 1:2 2.
Its advantageous to extend the inspiratory time in order to:
• improve oxygenation - through the addition of an inspiratory pause; or to
•increase tidal volume - in pressure controlled ventilation
Adverse effects of excessively long inspiratory times are haemodynamic compromise,
patient ventilator dysynchrony, and the development of autoPEEP.
Phase variables
A. Trigger …….
What causes the breath to begin?
B C
B. Limit ……
What regulates gas flow during the breath?
A
C. Cycle …….
What causes the breath to end?
Phases of ventilator supported breath
inspiration
change from inspiration to expiration
expiration
change from expiration to inspiration
Types of ventilator breaths-
Mandatory breath
Assisted breath
Spontaneous breath
Trigger variable
It’s the variable that determines start of inspiration
Triggering refers to the mechanism through which the
ventilator senses inspiratory effort and delivers gas flow or
a machine breath in concert with the patient’s inspiratory
effort.
Can use pressure or volume or time or flow as a trigger.
In modern ventilators the demand valve is triggered by
either a fall in pressure (pressure triggered) or a change in
flow (flow triggered).
With pressure triggered a preset pressure sensitivity has to
be achieved before the ventilator delivers fresh gas into the
inspiratory circuit. With flow triggered a preset flow
sensitivity is employed as the trigger mechanism.
Time triggering
Pressure Triggering
Breath is delivered when ventilator senses patients spontaneous
inspiratory effort.
sensitivity refers to the amount of negative pressure the patient
must generate to receive a breath/gas flow.
If the sensitivity is set at 1 cm then the patient must generate 1
cm H2O of negative pressure for the machine to sense the
patient's effort and deliver a breath.
Acceptable range - -1 to -5 cm H2O below patient s baseline
pressure
If the sensitivity is too high the patient's work of breathing will
be unnecessarily increased. It is not a reasonable course of
action to increase the sensitivity to reduce the patient's
respiratory rate as it only increases their work of breathing.
Flow Triggering
The flow triggered system has two preset variables for
triggering, the base flow and flow sensitivity.
The base flow consists of fresh gas that flows
continuously through the circuit. The patient’s earliest
demand for flow is satisfied by the base flow.
The flow sensitivity is computed as the difference
between the base flow and the exhaled flow
Here delivered flow= base flow- returned flow
Hence the flow sensitivity is the magnitude of the
flow diverted from the exhalation circuit into the
patient’s lungs. As the subject inhales and the set flow
sensitivity is reached the flow pressure control
algorithm is activated, the proportional valve opens,
and fresh gas is delivered.
•Flow trigger
Advantages -
-The time taken for the onset of inspiratory effort to the onset of
inspiratory flow is considerably less.
-decreases the work involved in initiating a breath.
Limit variable
Cycle variable
Defined as the length of one complete breathing cycle.
Inspiration ends when a specific cycle variable is
reached.
This variable is used as a feedback signal to end
inspiratory flow delivery which then allows exhalation
to start.
Most new ventilators measure flow and use it as a
feedback signal.
So volume becomes a function of flow and time
Volume= flow x inspiratory time
Baseline variable
Plateau Pressure-
The plateau pressure is defined as the end inspiratory
pressure during a period of no gas flow. The plateau
pressure reflects lung and chest wall compliance.
As the plateau pressure is the pressure when there is
no flow within the circuit and patient airways it most
closely represents the alveolar pressure and thus is of
considerable significance as it desirable to limit the
pressure that the alveoli are subjected to.
Excessive pressure may result in extrapulmonary air
(eg pneumothorax) and acute lung injury.
An increase in airways resistance (including ETT
resistance) will result in an increase in PIP.
An increase in resistance will result in a widening of
the difference between PIP and plateau pressure.
A fall in compliance will elevate both PIP and plateau
pressure.
It is generally believed that end inspiratory occlusion
pressure (ie plateau pressure) is the best clinically
applicable estimate of average peak alveolar pressure.
Although controversial it has been generally
recommended that the plateau pressure should be
limited to 35 cms H2O.
Mean Airway Pressure-
The mean airway pressure is an average of the system
pressure over the entire ventilatory period.
Improves ventilation
Inspiratory Time
In most volume cycled ventilators used in the intensive
care environment it is not possible to set the inspiratory
time.
The inspiratory time is determined by the peak inspiratory
flow rate, flow waveform and inspiratory pause. Where
inspiratory time is able to be set, flow becomes dependent
on inspiratory time and tidal volume.
The following example illustrate how these parameters effect
inspiratory time.
Ventilator settings
· Tidal volume 1000mls
· Peak Flow 60 lpm
· Flow Waveform square / constant
· Insp. Pause 0 secs
The inspiratory time for this patient would be 1 second because
gas is constantly being delivered at a flow rate of 60 lpm,
which equals 1 litre per second. If an inspiratory pause of 0.5
seconds were applied then the inspiratory time would be
increased to 1.5 seconds.
Changing the patients flow waveform from a square to a
decelerating flow waveform, without changing the flow rate,
will result in an increase in inspiratory time, because the flow
of gas is only initially set at 60 lpm and decreases throughout
inspiration
Output waveforms
Graphical representation of the control or phase
variables in relation to time.
presented as pressure
flow waveforms
volume