Académique Documents
Professionnel Documents
Culture Documents
Heat Sterilizers”
By
Anjana Sinha
MPH/10036/18
(M.Pharm QA)
3
VALIDATION
• Purpose of Validation.?
4
When should be validated ?
• Any Change in established method New method developed for
a particular problem Established method used in different
laboratories, different equipment or different staff Out-of-control
situations within internal quality assurance Demonstration of
equivalence between two methods (e.g. a rapid new test
against a standard method)
5
Who should validate?
6
• The vendor or the user has the ultimate responsibility for the
accuracy of the analysis results and also for equipment
qualification.
• DQ should always be done by the user.
• While IQ for a small and low cost instrument is usually done by
the user, IQ for large, complex and high cost instruments should
be done by the vendor.
• OQ can be done by either the user or the vendor.
• PQ should always be done by the user because it is very
application specific and the vendor may not be familiar with
these.
7
CALIBRATION
8
QUALIFICATION
VALIDATION
QUALIFICATION
9
QUALIFICATION
• The action of proving that any equipment Works correctly and
leads to the expected results.
10
Installation Operational Performance Design
Qualification(IQ) Qualification(OQ) Qualification(PQ) Qualification(DQ)
11
12
STAGES OF QUALIFICATION
Design Qualification
Installation Qualification
Operational Qualification
Performance Qualification
Change Control
13
Sterilization
• “The act or process, physical or chemical, that destroys or
eliminates all viable microbes including resistant bacterial
spores from a fluid or a solid.”
14
Gas (Ethylene Other (UV, Steam
Dry Heat Moist Heat Radiation (Gamma oxide, Hydrogen
or electron) & Formaldehyde,
peroxide) Filteration
15
Dry Heat Sterilization
• Dry heat is the most commonly used methods to sterilize
and/or depyrogenate pharmaceutical components
(glasswares) and products. Most often, it is used for
depyrogenation of parenteral containers is performed utilizing
a dry heat oven.
• Dry heat, as the name indicates, utilizes hot air that is either
free from water vapor, and where this moisture plays minimal
or no role in the process of sterilization.
• The process of sterilization within a chamber or hot air tunnel
is a critical process and there is regulatory requirement for
validation of the process.
16
Types of dry heat sterilizers
Commonly employed dry heat sterilizers in pharmaceutical
industry are-
• Forced Convention batch sterilizers.
• Laser/Plasma sterilizers.
17
Types of Dry Heat Sterilizers
Commonly utilized dry heat sterilizers in pharmaceutical industry are-
• Used for:
• glassware and product containers used in aseptic manufacture, non-
aqueous thermo stable powders and liquids (oils)
19
Image reference :www.slideshare.net
Validation Test Equipment
Equipment used for validation testing of oven are listed below:
Resistance temperature detectors
Thermocouples
Data logger
Stopwatch
Voltmeter
Optical tachometer 20
Qualification Approach
21
Basic Validation Approach
Design qualification (DQ):
• The DQ outline the key features of the system designed to address
the user requirement, regulatory compliance and selection rationale
of a particular supplier.
22
Design Qualification
(DQ)
23
24
Image reference : www.slideshare.net
Installation Qualification
(IQ)
• It is carried out after or concurrently with the installation of the equipment
at the user’s premises.
Contd.
25
• Electricals- Proper identification, safety cutoff
• Air or natural gas- Check that the source and type of supply are
consistent with the manufacturer’s recommendations.
• Blowers- Check for use of correct fan belt & that is in good condition.
26
Operational Qualification
(OQ)
• It is documented verification that the system or subsystem
performs as intended throughout all specified operating range
27
Each of the following process components must be identified & the
operating performance & range determined.
• Cycle timer - The accuracy of timer must be determined, so that assurance is
provided for cycle time.
• Door interlocks - If a unit is equipped with double doors, the interlocks must
operate such that the door leading to the aseptic area cannot be opened if the door
to the non-aseptic area is opened.
• Blowers - The air velocity consistent and motor speed of blowers should be noted
in the OQ records.
28
• Cooling Coils- If coils are present, the type and size of the coils and temperature of
the cooling medium at the inlet and outlet of the coils should be recorded.
• Chamber leaks- The perimeter of the doors for batch sterilizers should be checked
for air leakage while operating.
29
Performance Qualification
(PQ)
• Verifies that the equipment performs according to design
specifications and user defined requirements in a reliable &
reproducible manner under normal production conditions
• Physical -
• Heat penetration studies on empty chamber
• Heat distribution study on loaded chamber
• Heat penetration study on loaded chamber
• • Microbiological-
• Bio-challenge/ Pyro -challenge studies
30
Physical
• Heat Penetration Studies On • Heat Distribution Study on
Empty Chamber Loaded Chamber
To identify heat distribution patterns . Multiple thermocouples throughout
including slowest heating points chamber to determine effect of load
configuration on temperature
Multiple temperature sensing distribution.
devices should be
used(Thermocouples) Temperature distribution for all
loads using all container sizes used
Temperature profile locate hot/cold in production should be tested.
areas in the sterilizer by mapping
temperature at various location. Repeat runs should be performed
to check variability.
Temperature profile for each
chamber load configuration should
be documented
31
Cont..
• Heat Penetration Study on • Bio-Challenge/ Pyro-Challenge
Loaded Chamber Studies
Designed to determine the The challenge should
location of the slowest heating demonstrate the lethality delivered
point within a oven at various by the cycle with either
locations. microorganisms or endotoxins
Each size & type of material Resistant bacterial spores are
should be tested by penetration available primarily in the form of
studies. spore strips andSpore suspension.
Hot areas in the load are more
important for heat labile items.
32
Tablet Compression Machine
Image reference:
http://www.pharmainfo.net/equipment-validationarticles/laboratory-equipment-qualification 33
Introduction Of Compression Machine
• Compression is a critical step in the production of a tablet
dosage form. The materials being compressed will need to
have adequate flow and compression properties.
34
Qualification Approach
35
Installation Qualification
(IQ)
• Verity approved purchase order.
• Verity invoice Check manufacturer and supplier.
• Verity model number and serial number.
• Check for any physical damage.
• Confirm location and installation requirements per recommendation of
manufacturer.
• Verify that the required utilities are available.
• Installation shall be conducted per instructions provide in the manual.
36
Objective of IQ:
• To Check all the critical contact parts which directly affect quality
of the product either visually or by kit.
• To review proper installation as per checklist.
• To measure the dimension wherever possible
37
Installation Qualification check list :
• Compare the all specification and write • Feeder - by visual check Main electrical
the observation Machine height - Measure motor
with measuring tape • HP – Visually check by reading plate
• Overall Dimension - Measure with affixed on motor
measuring tape RPM of turret (max. &
min.) - • RPM - Visually check by reading plate
affixed on motor or Direction of motor (anti
• Check by tachometer Hopper clockwise) -by visual check
Conventional hopper - by visually check • Thickness controller
MOC (Material of construction) –
• Kit Height - Measure with measuring tape • Weight controller
• No. of station - Visually count no. of holes • Emergency switch
on turret • Dust Extraction and exhaust System Main
• Type of tooling - Using Venire caliper drive Available ON Indicator
check the die hole diameter • By visual Potentiometer check Selector
• Turret MOC – by visual check switch Emergency OFF button
• Circumference – Measure with measuring • Lubrication Oil pressure gauge Selector
tape switch for Auto/ Manual
• Distance between two dies - Using Venire
caliper
38
Operation Qualification
(OQ)
• Verify alarm control.
• Perform calibration requirements, identified in the manual or
established by the validation team.
• Operate the equipment at low ,medium and high speed per operation
manual to verify the operation control.
• Verify that all switches and push buttons are functioning properly.
• Establish procedures for operation, maintenance and calibration
• Establish training program for relevant staff.
• Run one pilot batch for each product
39
Objective of OQ :
• To operate machine as per proposed procedure given in manual and
record.
40
• Operational qualification check list :
1)Compare the all specification and write the observation
Main Switch – Check visually by operating the main switch
of the machine .
2)Start push button – (illuminated Green switch)By pressing
start button the Green switch glows and the main drive
motor should start.
3)Stop push button - (illuminated red switch) By pressing
stop button the red switch glows and the main drive motor
should stop.
41
4) Turret RPM challenge test – set in digital table counter by rotating
knob Check by tachometer Rotation direction ( clock wise) by visual
check
5)Emergency switch (Machine should be stop)
6) Selector/ isolator switch (Machine start/stop) By
7) Tablet thickness Visually & Hardness controls check By turning
the swing lever to right (anticlockwise) thickness increased/
hardness decrease vise versa
8 ) Tablet weight controls Dozer and wedge is moved clockwise
Decrease / increase / Anti clockwise weight
42
Performance Qualification
(PQ)
43
Objective of PQ:
44
• Performance qualification check list :
Sanitization:
• It should detailed the cleaning methods used for the equipment,
the SOPs covering each method, and the cleaning materials
utilized.
Change control-
• Changes to the equipment that might compromise the validation must be
brought to the attention of the group or individuals in charge of the change
control program.
Revalidation-
• It may be required after changes or repairs are made on the unit or an a
predetermined periodic interval.
47
Documentation
Qualifications
Raw Data:
• All original data, results, calculation and conclusions must be retained for
empty- and loaded chamber and biochallenge studies
Routine Monitoring:
• All change-control information and postvalidation mechanical changes are
recorded along with any revalidation work
49
Conclusion
• Equipment is one of the basic component of pharma processing and
therefore equipment validation is an important aspect.
50
References
1. Laurie BC, Gayle DH, Dry Heat Sterilization and Depyrogenation Validation and
Monitoring, In:, James AF, Carleton J, Editors. Validation Of Pharmaceutical
Process, informa healthcare; P. 223-239.
2. Potdar AM. Pharmaceutical Quality Assurance,1st ed. Published by Nirali
Prakashan; 2006.P. 8.13-8.20.
3. Syed Imtiaz Haider. Pharmaceutical Master Validation Plan. The Ultimate Guide to
FDA,GMP and GLP Compliance. informa healthcare;2010. P.138-139.
4. https://biologicalindicators.mesalabs.com/2015/12/14/validation-of-dry-heat-
sterilizationprocesses
5. https://www.pharmatutor.org/articles/validation-dry-heat-sterilization-methods.
6. http://www.mackpharmatech.com/doc.htm
7. http://www.ehow.com/list_6809815_equipment
51
52