Académique Documents
Professionnel Documents
Culture Documents
Basic Principles
CUP
LAAQ-B-LC001B 1
Polarity of Substances
H –
H O
O
C H C C –
H H O
+ H
H H
H
CUP
LAAQ-B-LC001B Methane Water Acetic acid 2
Nonpolar (Hydrophobic) Functional Groups and Polar
(Hydrophilic) Functional Groups
CUP
LAAQ-B-LC001B 3
Partition Chromatography
CUP
LAAQ-B-LC001B 4
Normal Phase / Reversed Phase
Stationary
Mobile phase
phase
Normal High polarity Low polarity
phase (hydrophilic) (hydrophobic)
CUP
LAAQ-B-LC001B 5
Reversed Phase Chromatography
• Stationary phase: Low polarity
– Octadecyl group-bonded silical gel (ODS)
• Mobile phase: High polarity
– Water, methanol, acetonitrile
– Salt is sometimes added.
CUP
LAAQ-B-LC001B 6
Separation Column for Reversed Phase
Chromatography
• C18 (ODS) type • Phenyl type
• C8 (octyl) type • TMS type
• C4 (butyl) type • Cyano type
C18 (ODS)
CUP
LAAQ-B-LC001B 7
Effect of Chain Length of Stationary
Phase
C8
Medium
C18 (ODS)
Strong C4
Weak
CUP
LAAQ-B-LC001B 8
Hydrophobic Interaction
H2O H2O
H2O
H2O H2O …the nonpolar substance
H2O H2O is pushed to a nonpolar
location.
Nonpolar solute
CUP
LAAQ-B-LC001B Nonpolar stationary phase 9
Relationship Between Retention Time
and Polarity
C18 (ODS) OH
Weak
Strong
CH3
CUP
LAAQ-B-LC001B 10
Basic Settings for Eluent Used in
Reversed Phase Mode
• Water (buffer solution) + water-soluble organic
solvent
– Water-soluble organic solvent: Methanol
Acetonitrile
Tetrahydrofuran etc.
– The mixing ratio of the water (buffer solution) and organic
solvent has the greatest influence on separation.
– If a buffer solution is used, its pH value is an important
separation parameter.
CUP
LAAQ-B-LC001B 11
Difference in Solute Retention Strengths for Water
and Water-Soluble Organic Solvents
CH3OH
Nonpolar solute CH3OH
Nonpolar solute
60/40
70/30
80/20
CUP
LAAQ-B-LC001B 13
Consideration of Analytical
Conditions
CUP
LAAQ-B-LC001B 14
Guidelines for Setting Mobile Phase Conditions (1)
Neutral (Nonionic) Substances
• Eluent Composition
– Water / acetonitrile
– Water / methanol
• Separation Adjustment
– Changing the mixing ratio of the water and organic
solvent
– Changing the type of organic solvent
CUP
LAAQ-B-LC001B 15
pH of Eluent and Retention of Ionic
Solutes
COOH
Acidic
Increased
hydrophobicity
pH of eluent
COO
Alkaline
Increased
hydrophilicity
+
H
CUP
LAAQ-B-LC001B 16
Guidelines for Setting Mobile Phase Conditions
Acidic (Anionic) Substances
• Eluent Composition
– Acidic buffer solution / acetonitrile
– Acidic buffer solution / methanol
CUP
LAAQ-B-LC001B 17
Analysis of Basic Substances (1)
Problems Encountered with Alkaline Eluents
OH Si
O
OH Si OH
…silica gel dissolves in alkalis,
so the packing material
OH deteriorates rapidly.
OH
OH
CUP
LAAQ-B-LC001B 18
Analysis of Basic Substances (2)
Influence of Residual Silanol Groups
Si
O
Si -O-Si-O
Residual silanol group N+
O H
Si
CUP
LAAQ-B-LC001B 19
Analysis of Basic Substances (3)
Addition of Sodium Perchlorate
Si
O
CUP
LAAQ-B-LC001B 20
Guidelines for Setting Mobile Phase Conditions (3)
Basic Substances (Cationic Substances)
• Eluent Composition
– Acidic buffer solution containing anions with a low
charge density (e.g., perchlorate ions) / acetonitrile
– As above / methanol
Making eluent acidic
Suppresses dissociation of residual silanol groups
Prevents tailing!
CUP
LAAQ-B-LC001B 23
Points to Note Concerning the Use of
Ion Pairs
• Selection of Ion Pair Reagent
– In general, the retention strength increases with the length of the
alkyl chain.
• pH of Eluent
– The retention strength changes according to whether or not
ionization takes place.
• Concentration of Ion Pair Reagent
– In general, the retention strength increases with the ion pair
concentration, but there is an upper limit.
• Proportion of Organic Solvent in Eluent
– Optimize the separation conditions by considering the type and
concentration of the ion pair reagent.
CUP
LAAQ-B-LC001B 24
HPLC Separation Modes
CUP
LAAQ-B-LC001B 25
HPLC Separation Modes
• Adsorption (liquid-solid) chromatography
• Partition (liquid-liquid) chromatography
– Normal phase partition chromatography
– Reversed phase partition chromatography
• Ion exchange chromatography
• Size exclusion chromatography
CUP
LAAQ-B-LC001B 26
Adsorption Chromatography
• A solid such as silica gel is used as the
stationary phase, and differences, mainly in
the degree of adsorption to its surface, are
used to separate the solutes.
• Liquid-solid chromatography
• The retention strength increases with the
hydrophilicity of the solute.
CUP
LAAQ-B-LC001B 27
Partition Chromatography
CUP
LAAQ-B-LC001B 28
Normal Phase and Reversed Phase
CUP
LAAQ-B-LC001B 29
Normal Phase (Partition)
Chromatography
• Partition chromatography in which the stationary
phase has a high polarity (hydrophilic) and the
mobile phase has a low polarity (hydrophobic)
• Essentially based on the same separation
mechanism as adsorption chromatography in
which the stationary phase has a hydrophilic base,
such as silica gel
CUP
LAAQ-B-LC001B 30
Stationary Phase and Mobile Phase Used in Normal
Phase Mode
• Stationary Phase
– Silica gel: -Si-OH
– Cyano type: -Si-CH2CH2CH2CN
– Amino type: -Si-CH2CH2CH2NH2
– Diol type: -Si-CH2CH2CH2OCH(OH)-CH2OH
• Mobile Phase
– Basic solvents: Aliphatic hydrocarbons,
aromatic hydrocarbons, etc.
– Additional solvents: Alcohols, ethers, etc.
CUP
LAAQ-B-LC001B 31
Relationship between Hydrogen Bonding and
Retention Time in Normal Phase Mode
SiOH HO
Strong
SiOH
Weak
Very weak
OH
Steric hindrance
CUP
LAAQ-B-LC001B 32
Relationship Between Eluent Polarity and Retention
Time in Normal Phase Mode
Eluent: Hexane/methanol
100/0
98/2
95/5
CUP
LAAQ-B-LC001B 33
Comparison of Normal Phase and
Reversed Phase
CUP
LAAQ-B-LC001B 34
Ion Exchange Chromatography
R
Anion exchange N+ R
R
++++
Cation exchange SO3- + +
++++
Electrostatic interaction
(Coulomb force)
CUP
LAAQ-B-LC001B 35
Stationary Phase Used in Ion Exchange
Mode
• Base Material
– Resin is often used.
– Silica gel is also used.
CUP
LAAQ-B-LC001B 36
Dependence of Exchange Capacity of Ion Exchanger
on pH of Eluent
Exchange capacity
Weakly acidic
cation exchanger Weakly basic
anion exchanger
0 7 14 0 7 14
pH pH
Cation exchange mode Anion exchange mode
CUP
LAAQ-B-LC001B 37
Relationship between Retention Time and Salt
Concentration of Eluent in Ion Exchange Mode
Solute ions and eluent ions compete for ion exchange groups.
If the salt concentration of the eluent increases, the solutes are eluted sooner.
CUP
LAAQ-B-LC001B 38
Ion Exclusion Chromatography
H+
H+
H+
CUP
LAAQ-B-LC001B 39
HPLC Hardware
CUP
LAAQ-B-LC001B 40
Detection Condition Requirements
• Sensitivity
– The detector must have the appropriate level of
sensitivity.
• Selectivity
– The detector must be able to detect the target
substance without, if possible, detecting other
substances.
• Adaptability to separation conditions
• Operability, etc.
CUP
LAAQ-B-LC001B 41
Representative HPLC Detectors
CUP
LAAQ-B-LC001B 42
UV-VIS Absorbance Detector
C: Concentration
Detection cell
Ein Eout
A
l
CUP
LAAQ-B-LC001B 43
Optical System of UV-VIS
Absorbance Detector
Grating
Sample cell
l Ein Eout Photodiode
Reference cell
D2 / W lamp
CUP
LAAQ-B-LC001B 44
Spectrum and Selection of
Detection Wavelength
A single photodiode
D2 / W lamp measures the absorbance for
the corresponding wavelength
at a resolution of approx. 1 nm.
Photodiode array
CUP
LAAQ-B-LC001B 46
Data Obtained with a Photodiode
Array Detector
Spectrum
Chromatogram
Absorbance
CUP
LAAQ-B-LC001B
Retention time 47
Advantages of Photodiode Array
Detectors
• Peak Identification Using Spectra
– Complementation of identification based on
retention time
– Library searches
• Evaluation of Peak Purity
– Purity evaluation performed by comparison of the
shape of spectra from the peak detection start
point to the peak detection end point
CUP
LAAQ-B-LC001B 48
Fluorescence Detector
Excitation wavelength
+ hv1 *
* hv2 +
Fluorescence wavelength
Excited state
Quasi-excited state
hv1
hv2
Fluorescence
CUP
LAAQ-B-LC001B
Ground state 49
Optical System of Fluorescence Detector
Xenon lamp
Fluorescence
grating
Photomultiplier tube
Fluorescence
Excitation
Excitation grating Sample cell
light
CUP
LAAQ-B-LC001B 50
Fluorescence Derivatization Reagents
CHN2 CH2OCOR
9-anthryldiazomethane
(ADAM)
CUP
LAAQ-B-LC001B 51
Differential Refractive Index
Detector (Deflection-Type)
Light-receiving unit
Reference cell
Light
Sample cell
CUP
LAAQ-B-LC001B 52
Optical System of Differential Refractive Index
Detector (Deflection-Type)
Slit W lamp
Reference cell
Sample cell
The slit image moves if the
refractive index inside the
flow cell changes.
Photodiode
CUP
LAAQ-B-LC001B 53
Evaporative Light Scattering Detector
Light-receiving unit
Drift tube
Nebulizer
Column eluate
Nebulizer gas
Drain Assist gas
Light source
CUP
LAAQ-B-LC001B 54
Electrical Conductivity Detector
Cl- Na+
The bulb does not light with water. The bulb lights if there are ions.
CUP
LAAQ-B-LC001B 55
Principle of Electrical Conductivity Detector
I A
V K k
I
E L
L
k K
A
A A K: Electrical conductivity [S]
I: Electric current [A]
E: Voltage [V]
A: Electrode surface area [cm2]
L Electrode L: Distance between electrodes [cm]
k: Specific electrical conductivity [S•cm-1]
CUP
LAAQ-B-LC001B 56
Limiting Equivalent Ion Conductance, l
[S•cm2/mol], in Aqueous Solution (25ºC)
Cation l Anion l
H+ 349.8 OH– 198.3
Li+ 38.6 F– 55.4
Na+ 50.1 Cl– 76.3
K+ 73.5 Br– 78.1
NH4+ 73.5 NO3– 71.4
(CH3)3NH+ 47.2 CH3COO– 40.9
Mg2+ 53.0 C6H5COO– 32.3
Ca2+ 59.5 SO42– 80.0
CUP
LAAQ-B-LC001B 57
Electrochemical Detector
Electrode
HO R
HO
2e-
O R
+ 2H+
O
CUP
LAAQ-B-LC001B 58
Cell Structure of Electrochemical
Detector (Amperometric Type)
Eluent
Electrode couple
CUP
LAAQ-B-LC001B 59
Mass Spectrometer (LCMS)
Atmospheric
pressure High vacuum
Quadrupole MS analyzer
API probe
Electron
multiplier tube
RP TMP1 TMP2
(high vacuum pumps)
CUP
LAAQ-B-LC001B 60
Atmospheric Pressure Ionization
3) n E
Liquid
Liquid Samples
Sample
1)
+
2) of S
Io
C va
C
High
High Voltage
ou p
Neburaizing
Nebulizing Gas
Ev o l
ha
Voltage
lo ora
ap ve
rg
n
Gas
ed
ol nt
E x tio
at
D
cl n
io
ro
us
n
pl
io
et
n
Atmospheric Pressure Chemical Ionization (APCI)
m/z=150
C D
CUP
LAAQ-B-LC001B 62
Advantages of LCMS (2)
M/Z
M/Z
CUP
LAAQ-B-LC001B M/Z 63
Comparison of Detectors
Possibility of
Selectivity Sensitivity
Gradient System
Light-absorbing
Absorbance ng Possible
substances
Differential
None µg Impossible
refractive index
Evaporative light
Nonvolatile substances µg Possible
scattering
Electrical
Ionic substances ng Partially possible
conductivity
Oxidizing / reducing
Electrochemical pg Partially possible
substances
CUP
LAAQ-B-LC001B Note: The above table indicates general characteristics. There are exceptions.
64
Post-Column Derivatization
Reaction
chamber
Pump
Reaction
solution
CUP
LAAQ-B-LC001B 65
Application Examples of Post-
Column Methods
• Amino Acids • Bromate Ions
– Orthophthalic acid, OPA – Tribromide ionization
(fluorescence) (ultraviolet absorption)
– o-Dianisidine
– Ninhydrin (visible absorption)
(visible absorption)
• Reducing Sugars • Cyanide Ions
– Arginine (fluorescence)
– Chlorination - pyrazolone
• Carbamate Pesticides (visible absorption)
– Alkaline hydrolysis - OPA • Transition Metal Ions
(fluorescence) – 4-(2-Pyridylazo) resorcinol,
PAR (visible absorption)
CUP
LAAQ-B-LC001B 66
Quantitative Analysis
CUP
LAAQ-B-LC001B 67
Qualitative Analysis
• Identification based on retention time
• Acquisition of spectra with detector
– UV spectra
– MS spectra
• Transfer to other analytical instruments after
preparative separation
CUP
LAAQ-B-LC001B 68
Quantitative Analysis
• Quantitation performed with peak area or
height.
• Calibration curve created beforehand using a
standard.
– Absolute calibration curve method
– Internal standard method
– Standard addition method
CUP
LAAQ-B-LC001B 69
Calibration Curve for Absolute
Calibration Curve Method
Area
Concentration
A1 Calibration curve
C1
A4
A2
Peak area
A3
C2
A2
A3
C3
A1
A4
C4 C1 C2 C3 C4
CUP
LAAQ-B-LC001B
Concentration 70
Calibration Curve for Internal
Standard Method
Concentration Area
A2 AIS A3 /AIS
C2 CIS
A2 /AIS
A3 AIS
C3 CIS
A1/AIS
A4 AIS
C4 CIS C1/CIS C2 /CIS C3 /CIS C4 /CIS
Concentration of target substance /
CUP
LAAQ-B-LC001B
Concentration of internal standard
71
Advantages of Internal
Standard Method (1)
• Not affected by inconsistencies in injection volume.
IS
X AX / AIS
10 µL
injected
Same area
ratio
IS
X
9 µL
injected
CX / CIS
CUP
LAAQ-B-LC001B 72
Advantages of Internal
Standard Method (2)
• Not affected by the pretreatment recovery rate.
IS
X
AX / AIS
100%
recovery
rate
Same area
ratio
IS
90% X
recovery
rate
CX / CIS
CUP
LAAQ-B-LC001B 73
Selection Criteria for Internal Standard
CUP
LAAQ-B-LC001B 74
Sample Pretreatment
CUP
LAAQ-B-LC001B 75
Objectives of Pretreatment
• To improve the accuracy of quantitative values
• To improve sensitivity and selectivity
• To protect and prevent the deterioration of
columns and analytical instruments
• To simplify measurement operations and
procedures
• To stabilize target substances
CUP
LAAQ-B-LC001B 76
Substances That Must Not Be
Injected into the Column
• Insoluble substances (e.g., microscopic particles
and precipitation)
• Substances that are precipitated in the eluent
• Substances that irreversibly adsorb to the packing
material
• Substances that dissolve, or chemically react,
with the packing material
CUP
LAAQ-B-LC001B 77
Filtration and Centrifugal Separation
CUP
LAAQ-B-LC001B 78
CUP
Deproteinization
• Precipitation
– Addition of organic solvent (e.g., acetonitrile)
– Addition of acid (e.g., trichloroacetic acid,
perchloric acid)
– Addition of heavy metal or neutral salt
• Ultrafiltration
CUP
LAAQ-B-LC001B 79
Solid Phase Extraction
Solvent with
high elution
strength
Target
component
Unwanted
components
CUP
LAAQ-B-LC001B 80
Pre-Column Derivatization