Académique Documents
Professionnel Documents
Culture Documents
X-ray
Spectrometry
Charles Lyman
in the AEM
Based on presentations developed for Lehigh University
semester courses and for the Lehigh Microscopy School
● Spatial resolution
» 2-20 nm (103 times better than SEM/EPMA)
● Elemental detectability
» 0.1 wt% - 1 wt%, depending on the specimen (~same as SEM/EDS)
● Straightforward microanalysis
» Qualitative analysis => Which element is present?
» Quantitative analysis => How much of the element is present?
» Easy x-ray mapping
2 Types of X-rays
» Characteristic x-rays
– elemental identification
– quantitative analysis
» Continuum x-rays
– background radiation
– must be subtracted for quantitative
analysis
● The good
» The shape of the continuum is a valuable
check on correct operation
● The not-so-good
» I bkg increases as ib increases
» I bkg is proportional to Zmean of specimen
» I maxbkg rises as beam energy rises
● Peak-to-background ratio
Continuum
» Ratio of Ichar / Ibkg sets limit on elemental detectability x-rays
Absorption of
continuum
● Example
» Vacancy in K-shell
» Vacancy filled from L-shell
» Emission of a Kα x-ray (or
a KLL Auger electron)
● Important uses
» Qualitative use x-ray energy to identify
elements
» Quantitative use integrated peak intensity
to determine amounts of elements
Beam
electron
If beam E > EK, then a
K-electron may be ionized: β
X-ray energy is the difference between α
two energy levels:
E Ka 1 = E K - E L III K L M
Energy levels EK and EL3 are in Bearden's "Tables
of X-ray Wavelengths and X-ray Atomic Energy
Levels" in older editions of the CRC Handbook of
Chemistry and Physics
For Na only see one peak since the Kβ is only 26 eV from the Kα line
PASI - Electron Microscopy - Lyman - EDS Qual 6
Families of Lines
If K-series excited, will
also have L-series
Note: this is a simplified version of Goldstein Figure 6.9 showing only lines seen in EDS
PASI - Electron Microscopy - Lyman - EDS Qual 7
Fluorescence Yield ω
ω = fraction of ionization events producing characteristic x-rays
• the rest produce Auger e–
·
● ω increases with Z
» ω K typical values are:
– 0.03 for carbon (12) K-series @ 0.3 keV
– 0.54 for germanium (32) K-series @ 9.9 keV
– 0.96 for gold (79) K-series @ 67 keV
» X-ray production is inefficient for low Z lines (e.g., O, N, C) since mostly Augers
produced
● LN dewar is most
recognizable part
» To cool FET and crystal
y gr en E
Within the intrinsic region
» Li compensates for impurity holes Acceptor
» Ideally # electrons = # holes
» # electron-hole pairs is proportional Hole
to energy of detected x-ray Valence band
For Cu Kα :
8048 eV/3.8 eV = 2118 e-h pairs
(after drawing by J. H. Scott)
Anti-reflective Ice?
Al coating 30 nm Gold electrode
Gold
Silicon inactive layer (p-type) ~100 nm electrode
20 nm
X-ray
Holes Electrons
(–) (+)
Window
Be, BN, Active silicon
diamond,
polymer (intrinsic)
0.1 µ m — 7 µ m 3 mm
(after drawing by J. H. Scott)
–1000 V bias
PASI - Electron Microscopy - Lyman - EDS Qual 12
X-ray Pulses to Spectrum
spectrum
charge
staircase slow amplifier
fast amplifier
from Goldstein et al., Scanning Electron Microscopy and X-ray Microanalysis, Springer, 2003.
PASI - Electron Microscopy - Lyman - EDS Qual 13
Slow EDS Pulse Processing
● Detector Performance
» Energy resolution (stamped on detector)
» Incomplete charge collection (low energy tails)
» Detector window (thin window allows low-energy x-ray detection)
» Detector contamination (ice and hydrocarbon)
» Count rate linearity (counts vs. beam current)
» Energy calibration (usually auto routine)
» Maximum throughput (set pulse processor time constant to collect the
most x-rays in a given clock time with some decrease in energy resolution)
2 2 1/ 2
FWHM » (C E + N )
C = 2.35(Fe)1/ 2
E = x-ray energy
N = electronic noise
F = Fano factor (~0.1 for Si)
● Measured with 1000 cps at 5.9 keV
»
E K= 3.8
Mn α
eV/electron-hole pair
line
Mn Kα line
●
Sum peaks
Two photons of same energy enter detector simultaneously
» Count of twice the energy
» Only for high count rates
»
●
Si internal fluorescence peak
Photon generated in dead layer
» Detected in active region
»
Escape peaks
Si
interna
l fluor.
System Sum
peaks peaks
from Goldstein et al., Scanning Electron Microscopy and X-ray Microanalysis, Springer, 2003.
● We want x-rays to come from just under the electron probe, BUT…
t or
from Williams and Carter, Transmission
etec
Electron Microscopy, Springer, 1996 D
Top view of disc
PASI - Electron Microscopy - Lyman - EDS Qual 24
Spurious X-rays in the
Microscope
● Pre-Specimen Effects
» spurious x-rays => hole count due to column x-rays and stray
electrons
» spurious x-rays => poor beam shape from large C2 aperture
● Post-Specimen Scatter
» system x-rays => elements in specimen stage, cold finger,
apertures, etc.
» spurious x-rays => excited by electrons and x-rays generated in
specimen
● Coherent Bremsstrahlung
» extra peaks from specimen effects on beam-generated
continuous radiation
Ι ον
ΜοΚ
ΝιΟ φιλµ
% hole count = α
ον ΝιΟ φιλµ ξ 100%
Ι ΝιΚ
α
Mo grid bar
(Νοτε
- “on film”:results
∀ον ΝιΟ φιλµresults
~ “in hole” χουντ∀ανδ
∀ιν ηολεχουντ∀αρε σιµιλαρ
)
- Usually take inverse “hole count” ratio
Hard x-ray from
illumination system
Better
Less good
● Scan over large area of single phase => avoid spot mode
Specimen
» Use thin foils, flakes, or films rather than self-supporting disks to reduce spurious x-rays (not
always possible)
» Orient specimen so that EDS detector is on the side of the specimen hole opposite where you
take your analysis
Microscope Column
» Use highest kV of microscope
» Use clean, top-hat Pt aperture in C2 to minimize “hole count”
effect
» Minimize beam tails
– (C2 aperture or VOA should properly limit beam angle)
» Use ~ 1 nA probe current to maximize count rate
– This may enlarge the electron beam (analyze smaller regions
later)
» Remove the objective aperture
X-ray Spectrometer
» Ensure that detector is cranked into position
» Keep detector shutter closed until you are ready to analyze
» Use widest energy range available (0-20 keV is normal)
– 0–40 keV for Si(Li) detector
– 0–80 keV for intrinsic Ge detector
» Choose short detector time constant (for maximum countrate)
» Count for a long time – 100-500 live sec
● Important:
» Use more than one peak for identification
» If peak too small to "see", collect more counts or forget about
identifying that peak; peak should be greater than 3B1/2
M-series
L-series
K-series
M-series
L-series
K-series
At 2.0 keV:
Z = 15 (P) Kα @ 2.013 keV K-series
Z = 40 (Zr) Lα @ 2.042
Z = 77 (Ir) Mα @ 1.977
Energy (keV)
PASI - Electron Microscopy - Lyman - EDS Qual 39
Data Analysis for Unknown #1
26
4.5 6.4
Fe(26)
Kα
4.9 7.0
Fe(26)
Kβ
from xray.optics.rochester.edu/.../spr04/pavel/
W Mα
Elements present
major: W, Ru
trace: Fe, Cu, C
W Lα
Ru Lα
W Mζ W Ll
Fe Kα ,β W Lβ 1,2
C Cu W Lγ 1,3