Dialysis machine

What is dialysis machine


Dialysis machine is a electro-mechanicalhydraulics device which helps us to remove the unwanted particles (Eg.- K+, Na+, CL-, Calcium, Magnesium & etc.) from the patients blood through hollow fiber artificial dialyzer.

Principle of dialysis
Dialysis works on the principles of the diffusion and osmosis of solutes and fluid across a semi-permeable membrane. Blood flows by one side of a semi-permeable membrane, and a dialysate or fluid flows by the opposite side. Smaller solutes and fluid pass through the membrane. The blood flows in one direction and the dialysate flows in the opposite. The concentrations of undesired solutes (for example potassium, calcium, and urea) are high in the blood, but low or absent in the dialysis solution and constant replacement of the dialysate ensures that the concentration of undesired solutes is kept low on this side of the membrane. The dialysis solution has levels of minerals like potassium and calcium that are similar to their natural concentration in healthy blood. For another solute, bicarbonate, dialysis solution level is set at a slightly higher level than in normal blood, to encourage diffusion of bicarbonate into the blood, to neutralise the metabolic acidosis that is often present in these patients.

Types of dialysis
Hem filtration Peritoneal dialysis Hemo-dialysis

Hem filtration
Hemofiltration is a similar treatment to hemodialysis, but it makes use of a different principle. The blood is pumped through a dialyzer or "hemofilter" as in dialysis, but no dialysate is used. A pressure gradient is applied; as a result, water moves across the very permeable membrane rapidly, facilitating the transport of dissolved substances, importantly ones with large molecular weights, which are cleared less well by hemodialysis. Salts and water lost from the blood during this process are replaced with a "substitution fluid" that is infused into the extracorporeal circuit during the treatment. Hemodiafiltration is a term used to describe several methods of combining hemodialysis and hemofiltration in one process.

Peritoneal dialysis

Working principle of peritoneal dialysis


In peritoneal dialysis, a sterile solution containing minerals and glucose is run through a tube into the peritoneal cavity, the abdominal body cavity around the intestine, where the peritoneal membrane acts as a semipermeable membrane. The dialysate is left there for a period of time to absorb waste products, and then it is drained out through the tube and discarded. This cycle or "exchange" is normally repeated 4-5 times during the day, (sometimes more often overnight with an automated system). Ultrafiltration occurs via osmosis; the dialysis solution used contains a high concentration of glucose, and the resulting osmotic pressure causes fluid to move from the blood into the dialysate. As a result, more fluid is drained than was instilled. Peritoneal dialysis is less efficient than hemodialysis, but because it is carried out for a longer period of time the net effect in terms of removal of waste products and of salt and water are similar to hemodialysis. Peritoneal dialysis is carried out at home by the patient and it requires motivation. Although support is helpful, it is not essential. It does free patients from the routine of having to go to a dialysis clinic on a fixed schedule multiple times per week, and it can be done while travelling with a minimum of specialized equipment.

Hemodialysis

Principle of hemodialysis
In hemodialysis, the patient's blood is pumped through the blood compartment of a dialyzer, exposing it to a semipermeable membrane. The cleansed blood is then returned via the circuit back to the body. Ultrafiltration occurs by increasing the hydrostatic pressure across the dialyzer membrane. This usually is done by applying a negative pressure to the dialysate compartment of the dialyzer. This pressure gradient causes water and dissolved solutes to move from blood to dialysate, and allows removal of several litres of excess fluid during a typical 3 to 5 hour treatment. In the US, hemodialysis treatments are typically given in a dialysis center three times per week

How to know dialysis is required


The decision to initiate dialysis or hemofiltration in patients with renal failure can depend on several factors, which can be divided into acute or chronic indications. Acute indications for dialysis/hemofiltration: Hyperkalemia Metabolic acidosis Fluid overload (which usually manifests as pulmonary edema) Uremic Serositis complications, such as uremic pericarditis and uremic encephalopathy And in patients without renal failure, acute poisoning with a dialysable drug, such as lithium, or aspirin Chronic indications for dialysis: Symptomatic renal failure Low glomerular filtration rate (GFR) (RRT often recommended to commence at a GFR of less than 10-15 mls/min/1.73m2)

What is AV fistula

Phenomenon of AV fistula
AV (arteriovenous) fistulas are recognized as the preferred access method. To create a fistula, a vascular surgeon joins an artery and a vein together through anastomosis. Since this bypasses the capillaries, blood flows rapidly through the fistula. One can feel this by placing one's finger over a mature fistula. This is called feeling for "thrill" and produces a distinct 'buzzing' feeling over the fistula. Fistulas are usually created in the nondominant arm and may be situated on the hand (the 'snuffbox' fistula'), the forearm (usually a radiocephalic fistula, or so-called Brescia-Cimino fistula, in which the radial artery is anastomosed to the cephalic vein), or the elbow (usually a brachiocephalic fistula, where the brachial artery is anastomosed to the cephalic vein). A fistula will take a number of weeks to mature, on average perhaps 4-6 weeks. During treatment, two needles are inserted into the fistula, one to draw blood and one to return it.

AV GRAFT

What is dialyzer

Principle of dialyzer
The dialyzer is the piece of equipment that actually filters the blood. Almost all dialyzers in use today are of the hollow-fiber variety. A cylindrical bundle of hollow fibers, whose walls are composed of semi-permeable membrane, is anchored at each end into potting compound (a sort of glue). This assembly is then put into a clear plastic cylindrical shell with four openings. One opening or blood port at each end of the cylinder communicates with each end of the bundle of hollow fibers. This forms the "blood compartment" of the dialyzer. Two other ports are cut into the side of the cylinder. These communicate with the space around the hollow fibers, the "dialysate compartment." Blood is pumped via the blood ports through this bundle of very thin capillary-like tubes, and the dialysate is pumped through the space surrounding the fibers. Pressure gradients are applied when necessary to move fluid from the blood to the dialysate

Block diagram of dialysis

Basic function of dialysis


The hemodialysis machine pumps the patient's blood and the dialysate through the dialyzer. The newest dialysis machines on the market are highly computerized and continuously monitor an array of safety-critical parameters, including blood and dialysate flow rates; dialysis solution conductivity, temperature, and pH; and analysis of the dialysate for evidence of blood leakage or presence of air. Any reading that is out of normal range triggers an audible alarm to alert the patient-care technician who is monitoring the patient. Manufacturers of dialysis machines include companies such as Fresenius, Gambro, Baxter, B. Braun, and Bellco.