Vous êtes sur la page 1sur 4

Scaling table analogue output

S 400 standard range

S 400 outputs the volumetric flow to the analogue output. The scaling of
the analogue output (4...20 mA) depends on the selected tube diameter.
The corresponding volumetric flow representing 20 mA can be found in
the table below.

Volumetric flow (reference: 1000 hPa, 20 °C)


Tube Air Nitrogen
Inch mm m³/h m³/min cfm m³/h m³/min cfm
1/2" 16.1 45.6 0.76 26.8 44.1 0.73 25.9
3/4" 21.7 89.1 1.49 52.4 86.2 1.44 50.7
1" 25 122.2 2.04 71.9 118.2 1.97 69.5
26 132.9 2.21 78.2 128.5 2.14 75.6
27.3 147.7 2.46 86.9 142.8 2.38 84.0
28.5 162 2.70 95.3 156.7 2.61 92.2
30 180.9 3.02 106.5 175.0 2.92 103.0
1 1/4" 32.8 218.8 3.65 128.8 211.6 3.53 124.5
36 266.3 4.44 156.7 257.5 4.29 151.6
36.3 270.9 4.52 159.5 262.0 4.37 154.2
1 1/2" 39.3 321.0 5.34 188.7 310.0 5.17 182.5
40 332.6 5.54 195.7 321.6 5.36 189.3
41.9 366.7 6.11 215.8 354.6 5.91 208.7
43.1 389.5 6.50 229.2 376.6 6.28 221.7
45.8 442 7.37 260.1 427.4 7.12 251.6
2" 50 530.8 8.85 312.4 513.2 8.55 302.0
51.2 557.2 9.29 328.0 538.8 8.98 317.1
53.1 600.1 10.01 353.2 580.3 9.67 341.5
54.5 632.9 10.55 372.5 612.0 10.20 360.2
57.5 708.9 11.81 417.2 685.5 11.42 403.4
60 773.7 12.89 455.3 748.2 12.47 440.3
64.2 889.1 14.82 523.2 859.7 14.33 505.9
2 1/2" 65 911.8 15.20 536.6 881.6 14.69 518.9
68.9 1026.5 17.11 604.1 992.6 16.54 584.2
70.3 1071.2 17.85 630.4 1035.9 17.26 609.6
71.1 1095.8 18.26 644.9 1059.6 17.66 623.6
76.1 1258.3 20.97 740.5 1216.8 20.28 716.1

1 Scaling analogue S 400 English, V1.07, 09.04.15


Scaling table analogue output

S 400 standard range

Volumetric flow (reference:1000 hPa, 20 °C)


Tube Air Nitrogen
Inch mm m³/h m³/min cfm m³/h m³/min cfm
3" 80.9 1424.4 23.74 838.3 1377.4 22.96 810.6
82.5 1482.4 24.71 872.4 1433.5 23.89 843.6
84.9 1569.9 26.16 923.9 1518.1 25.30 893.4
90 1766.4 29.44 1039.5 1708.0 28.47 1005.2
4" 100 2183.3 36.4 1284.9 2111.2 35.19 1242.4
107.1 2507.3 41.79 1475.6 2424.6 40.41 1426.9
110 2645 44.08 1556.6 2557.6 42.63 1505.2
5" 125 3419.6 57 2012.4 3306.7 55.11 1945.9
133.7 3912.2 65.2 2302.3 3783.0 63.05 2226.3
6" 150 4930.1 82.17 2901.4 4767.3 79.45 2805.6
159.3 5560.4 92.67 3272.3 5376.8 89.61 3164.2
182.5 7306.7 121.77 4300.6 7065.4 117.75 4158.0
190 7919.6 132 4660.7 7658.1 127.63 4506.8
8" 200 8785.6 146.4 5170.3 8495.5 141.59 4999.6
206.5 9366 156.09 5511.9 9056.7 150.94 5329.9
10" 250 13743.9 229.1 8088.3 13290.1 221.50 7821.2
260.4 14914.9 248.57 8777.4 14422.4 240.36 8487.6
12" 300 19814.8 330.23 11661.0 19160.5 319.33 11275.9
309.7 21116.9 351.9 12427.3 20419.6 340.31 12016.94
339.6 25391.1 423.2 14942.7 24552.8 409.2 14449.3
400 35226.3 587.1 20730.7 34063.2 567.7 20046.2
500 55041.1 917.3 32391.7 53223.7 887.03 31322.1
600 79259.2 1320.9 46644.0 76642.1 1277.32 45103.9
700 107880.5 1797.95 63487.7 104318.4 1738.57 61391.4
800 140905.2 2348.33 82922.7 136252.6 2270.79 80184.7
900 178333.1 2972.1 104949.0 172444.7 2874 101484
1000 220164.3 3699.3 129566.7 212894.7 3548.1 125288.5

2 Scaling analogue S 400 English, V1.07, 09.04.15


Scaling table analogue output
S 400 max range

S 400 outputs the volumetric flow to the analogue output. The scaling of
the analogue output (4...20 mA) depends on the selected tube diameter.
The corresponding volumetric flow representing 20 mA can be found in
the table below.

Volumetric flow (reference: 1000 hPa, 20 °C)


Tube Air Nitrogen
Inch mm m³/h m³/min cfm m³/h m³/min cfm
1/2" 16.1 91.0 1.52 53.5 88.0 1.47 51.8
3/4" 21.7 177.8 2.96 104.7 172.0 2.87 101.2
1" 25 243.9 4.06 143.5 235.8 3.93 138.8
26 265.2 4.42 156.1 256.4 4.27 150.9
27.3 294.7 4.91 173.4 285.0 4.75 167.7
28.5 323.3 5.39 190.3 312.7 5.21 184.0
30 361.1 6.02 212.5 349.2 5.82 205.5
1 1/4" 32.8 436.7 7.28 257.0 422.3 7.04 248.5
36 531.5 8.86 312.8 513.9 8.57 302.4
36.3 540.7 9.01 318.2 522.9 8.71 307.7
1 1/2" 39.3 639.8 10.66 376.5 618.7 10.31 364.1
40 663.7 11.06 390.6 641.8 10.70 377.7
41.9 731.9 12.20 430.7 707.7 11.79 416.5
43.1 777.3 12.95 457.5 751.7 12.53 442.4
45.8 882.2 14.70 519.2 853.0 14.22 502.0
2" 50 1059.2 17.65 623.4 1024.2 17.07 602.8
51.2 1112.0 18.53 654.4 1075.3 17.92 632.8
53.1 1197.6 19.96 704.8 1158.0 19.30 681.5
54.5 1263.1 21.05 743.3 1221.4 20.36 718.8
57.5 1414.7 23.58 832.5 1367.9 22.80 805.0
60 1544.1 25.73 908.7 1493.1 24.88 878.7
64.2 1774.3 29.57 1044.2 1715.7 28.60 1009.7
2 1/2" 65 1819.6 30.32 1070.8 1759.5 29.32 1035.5
68.9 2048.6 34.14 1205.6 1980.9 33.01 1165.8
70.3 2137.8 35.63 1258.1 2067.3 34.45 1216.6
71.1 2186.8 36.44 1286.9 2114.6 35.24 1244.4
76.1 2511.2 41.85 1477.9 2428.3 40.47 1429.1

3 Scaling analogue S 400 English, V1.07, 09.04.15


Scaling table analogue output
S 400 max range

Volumetric flow (reference: 1000 hPa, 20 °C)


Tube Air Nitrogen
Inch mm m³/h m³/min cfm m³/h m³/min cfm
3" 80.9 2842.7 47.38 1672.9 2748.8 45.81 1617.7
82.5 2958.5 49.31 1741.1 2860.8 47.68 1683.6
84.9 3133.1 52.22 1843.8 3029.7 50.49 1783.0
90 3525.1 58.75 2074.5 3408.7 56.81 2006.0
4" 100 4357.2 72.62 2564.2 4213.3 70.22 2479.5
107.1 5003.9 83.40 2944.8 4838.6 80.64 2847.5
110 5278.5 87.98 3106.4 5104.2 85.07 3003.8
5" 125 6824.4 113.74 4016.2 6599.1 109.98 3883.6
133.7 7807.4 130.12 4594.7 7549.7 125.82 4442.9
6" 150 9838.9 163.98 5790.2 9514.1 158.56 5599.0
159.3 11096.8 184.94 6530.5 10730.4 178.83 6314.8
182.5 14581.8 243.02 8581.4 14100.3 235 8298.0
190 15804.9 263.4 9301.2 15283.1 254.71 8994.1
8" 200 17533.3 292.21 10318.4 16954.4 282.56 9977.7
206.5 18691.5 311.51 10999.9 18074.3 301.23 10636.7
10" 250 27428.5 457.12 16141.7 26522.8 442.03 15608.7
260.4 29765.4 496.07 17516.9 28782.5 479.69 16938.5
12" 300 39544.1 659.04 23271.7 38238.4 637.28 22503.3
309.7 42142.6 702.35 24800.9 40751.1 679.16 23982.0
339.6 50672.7 844.51 29820.9 48999.6 816.63 28836.3
400 70300.6 1171.63 41371.9 67979.3 1132.94 40005.8
500 109844.7 1830.67 64643.6 106217.7 1770.22 62509.1
600 158176.3 2636.17 93086.8 152953.5 2549.12 90013.1
700 215295.5 3588.11 126701.4 208186.7 3469.64 122517.9
800 281202.3 4686.52 165487.6 271917.3 4531.77 160023.3
900 355896.7 5931.37 209445.2 344145.4 5735.53 202529.5
1000 439378.6 7322.68 258574.3 424870.8 7080.9 250036.5

4 Scaling analogue S 400 English, V1.07, 09.04.15