Technical
Publication
Direction 5125372-100
Revision 1
GE Medical Systems
2
Copyright 2004 by General Electric Company Inc. All Rights Reserved.
GE Medical Systems
Important Information
LANGUAGE
Important Information
3
Copyright 2004 by General Electric Company Inc. All Rights Reserved.
GE Medical Systems
Important Information
Copyright 2004 by General Electric Company Inc. All Rights Reserved.
GE Medical Systems
Important Information
Copyright 2004 by General Electric Company Inc. All Rights Reserved.
GE Medical Systems
Important Information
Copyright 2004 by General Electric Company Inc. All Rights Reserved.
GE Medical Systems
Revision History
Rev
Date
Oct. 4, 2004
Revision History
Copyright 2004 by General Electric Company Inc. All Rights Reserved.
GE Medical Systems
Revision History
Copyright 2004 by General Electric Company Inc. All Rights Reserved.
GE Medical Systems
Table of Contents
1 LCD Video Monitor Overview - NEC 1980SXi
1.1 General Monitor Information
1.1.1 Cleaning
. . . . . . . . . . . . . . . . . . . . . . 11
. . . . . . . . . . . . . . . . . . . . . . . . . . . 11
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
. . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
. . . . . . . . . . . . . . . . . . . . 12
Table of Contents
Copyright 2004 by General Electric Company Inc. All Rights Reserved.
GE Medical Systems
10
Table of Contents
Copyright 2004 by General Electric Company Inc. All Rights Reserved.
GE Medical Systems
Avoid using any cleaning solution or glass cleaner! When cleaning, press lightly
against the LCD monitor surface to avoid damage to the screen.
Clean the LCD monitor surface with a lint-free, non-abrasive cloth and ONLY water.
1.1.3.2 Storage
1.1.4 Weight
9.7 kg (21.4 lbs)
11
GE Medical Systems
- +).
To rotate OSM between Landscape and Portrait modes, press the RESET button.
NOTE:
Menu
EXIT
- + (Adjust )
Moves the highlighted area of main menu right to select one of the controls.
RESET
NOTE:
When RESET is pressed in the main and sub-menu, a warning window will appear allowing
you to cancel the RESET function by pressing the EXIT button.
12
GE Medical Systems
OUT OF RANGE: This function gives a recommendation of the optimized resolution and
refresh rate. After the power is turned on or there is a change of input signal or the video
signal doesnt have proper timing, the Out Of Range menu will appear.
CHECK CABLE: This function will advise you to check all Video Inputs on the monitor and
computer to make sure they are properly connected.
13
GE Medical Systems
Preliminary Reqs
Procedure
Finalization
0 mins
20 mins
0 mins
2.2 Overview
This procedure outlines the process required to reset and set-up the MultiSync LCD 1980SXi
NEC 19 inch Monitor. If a system has more than one of these monitors, each must be setup.
The procedure time listed above is for each monitor. An active video input signal must be
connected to each monitor to perform this process.
2.3.2 Consumables
No consumables required.
2.3.4 Safety
No general safety information.
2.4 Procedure
2.4.1 Monitor Connections and Powering Up
1. Connect the 15-pin Sub-D connector cable(s) between the Operator Consoles HOST
and each monitors blue connector INPUT 2 D-SUB. See Illustration 1.
NOTE:
14
GE Medical Systems
2. Connect the AC in line cord from the Operator Console to each monitor's rear side using
the GE supplied console power cord. See Illustration 2.
NOTE:
3. Apply power to each monitor by setting the Vacation Switch on the left side of the monitor
to the ON position. See Illustration 3.
NOTE:
The Vacation Switch is a true on/off switch. If the switch is in the OFF position, the monitor
cannot be turned on using the power button.
15
GE Medical Systems
4. Press the front panel far right button to apply power to each monitor. Verify the green
LED is illuminated. See Illustration 4.
Illustration 4: Monitor Power Button and LED
The monitor may perform an Auto Adjust referred to as the No Touch Auto Adjust
function. This function is activated only when a new video signal (Analog input only) is
received for the first time.
GE Medical Systems
NOTE:
If the Advanced Setup menu (Tabs 1-9 at the top) is displayed instead of the OSM menu,
the monitor must be powered OFF and then ON again to exit the Advanced Setup menu
mode and enable the OSM menu mode.
2. Press the > button five (5) times to highlight the Tools 2 Menu. See Illustration 6.
Illustration 6: Tools 2 Menu Highlighted
3. Press the SELECT/12 button to enter the Tools 2 Menu. See Illustration 7.
17
GE Medical Systems
4. Press the < button one (1) time to highlight the FACTORY PRESET symbol. See
Illustration 8.
Illustration 8: Factory Preset Symbol
18
GE Medical Systems
7. Press the EXIT button. The initial OSM Menu appears. See Illustration 5.
8. Press the front panel far right button (see Illustration 4) to remove power. Verify the
green LED is not illuminated.
2.4.2.2 Color Temperature Setup
1. Press the front panel far right button to apply power to each monitor. Verify the green
LED is illuminated. See Illustration 4.
2. Press the > button to display the initial OSM Menu. See Illustration 5.
3. Press the > button three (3) times to highlight the RGB Menu.
4. Press the SELECT/12 button to enter the RGB Menu. The [N] may be highlighted
initially. See Illustration 10.
NOTE:
If the monitor has been already set-up the [1] may be highlighted.
19
GE Medical Systems
5. Press the > button two (2) times to highlight setting 1. Verify it is 9300K. See Illustration
11.
Illustration 11: RGB Menu - Selection 1 Highlighted
6. (For setting 1 at 9300K) Press EXIT twice and continue with Step 8.
7. (For setting 1 NOT at 9300K) If necessary, press the SELECT/12 button to enter
inside the setting 1 color temperature menu. Change the setting to 9300K as follows:
a. Press the or + buttons as necessary to set the TEMPERATURE to 9300K. See
Illustration 12.
20
GE Medical Systems
The monitor must be powered OFF and then ON again to access the Advanced Setup
menu in the next section.
8. Press the front panel far right button to remove power. Verify the green LED is not
illuminated. See Illustration 4.
21
GE Medical Systems
4. Press the > button four (4) times to highlight selection 5, GAMMA SELECTION.
NOTE:
First time out of the box monitors will not follow the SELECT/12 step. Instead, press the +
button three (3) times to scroll down the Menu and then continue with the > button once to
highlight CUSTOM VALUE.
5. Press the SELECT/12 button to enter inside the GAMMA SELECTION Menu.
6. Press the > button once to highlight CUSTOM VALUE. See Illustration 14.
22
GE Medical Systems
7. If necessary, press the or + buttons to set the CUSTOM VALUE to 2.6 (K). See
Illustration 15.
Illustration 15: Gamma Selection Menu - Final Setting
23
GE Medical Systems
The monitor must be powered OFF and then ON again to exit the Advanced Setup menu
mode and enable the OSM menu mode.
13. Press the front panel far right button to remove power. Verify the green LED is not
illuminated.
14. Press the front panel far right button to apply power to each monitor. Verify the green
LED is illuminated.
24
GE Medical Systems
4. Remove the stand cover by sliding the top/bottom pieces off the stand. Remove the 4
screws connecting the monitor to the stand and lift off the stand assembly. The monitor is
now ready for mounting in an alternate manner. See Illustration 20.
Illustration 20: Remove Stand From Monitor
25
GE Medical Systems
Please use the attached screws (4 pieces) when mounting. To fulfill the safety
requirements the monitor must be mounted to an arm which guaranties the
necessary stability under consideration of the weight of the monitor. The LCD
monitor shall only be used with an approved arm (e.g. GS mark).
NOTE:
2.5 Finalization
No finalization steps.
26
GE MEDICAL SYSTEMS
GE MEDICAL SYSTEMS - AMERICAS: FAX 262.312.7434
3000 N. GRANDVIEW BLVD., WAUKESHA, WI 53188 U.S.A.
GE MEDICAL SYSTEMS - EUROPE: FAX 33.1.40.93.33.33
PARIS, FRANCE
GE MEDICAL SYSTEMS - ASIA: FAX 65.291.7006
SINGAPORE
http://www.gemedicalsystems.com