Académique Documents
Professionnel Documents
Culture Documents
Funcional Check Update
Funcional Check Update
Publications
2205009
Revision 13
CT ProSpeed Series
Functional Check /Adjustment
Operating Documentation
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
a
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
b
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
In performing all electrical work on these products, GE will use its own specially trained field engineers. All of GE’s electrical
work on these products will comply with the requirements of the applicable electrical codes.
The purchaser of GE equipment shall only utilize qualified personnel (i.e., GE’s field engineers, personnel of third-party
service companies with equivalent training, or licensed electricians) to perform electrical servicing on the equipment.
c
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
DAMAGE IN TRANSPORTATION
All packages should be closely examined at time of delivery. If damage is apparent, have notation “damage in
shipment” written on all copies of the freight or express bill before delivery is accepted or “signed for” by a General
Electric representative or a hospital receiving agent. Whether noted or concealed, damage MUST be reported to the
carrier immediately upon discovery, or in any event, within 14 days after receipt, and the contents and containers held
for inspection by the carrier. A transportation company will not pay a claim for damage if an inspection is not requested
within this 14 day period.
Call Traffic and Transportation, Milwaukee, WI (414) 827–3449 / 8*285–3449 immediately after damage is found. At
this time be ready to supply name of carrier, delivery date, consignee name, freight or express bill number, item
damaged and extent of damage.
Complete instructions regarding claim procedure are found in Section “S” of the Policy & Procedure Bulletins.
GE personnel, please use the GEMS CQA Process to report all omissions, errors, and defects in this documentation.
Customers, please contact your GE Sales or Service represenatives.
CAUTION
Do not use the following devices near this equipment. Use of these devices near this equipment could cause
this equipment to malfunction.
Devices which intrinsically transmit radio waves such as; cellular phone, radio transceiver, mobile radio transmitter,
radio–controlled toy, etc.
Keep power to these devices turned off when near this equipment.
Medical staff in charge of this equipment is required to instruct technicians, patients an
d
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 13 2205009
REVISION HISTORY
PAGE REV PAGE REV PAGE REV PAGE REV PAGE REV
Title page . . . . . . . . 13 5–1 to 5–10 . . . . . . . . 0 3–1 . . . . . . . . . . . . . . . 0 3–1 to 3–2 . . . . . . . . . 0 1–4 . . . . . . . . . . . blank
Title page rear . blank 5–11 . . . . . . . . . . . . . . 7 3–2 . . . . . . . . . . . . . . . 7
a to d . . . . . . . . . . . . . 0 5–12 to 5–15 . . . . . . 0 3–3 to 3–4 . . . . . . . . . 0 Tab 6 (DAS/Detector) Tab 9 (Switch/Jump-
A . . . . . . . . . . . . . . . . 13 5–16 . . . . . . . . . . . . . . 7 3–5 to 3–6 . . . . . . . . . 2 i.................. 0 er Setting of Boards/
B . . . . . . . . . . . . . blank 5–17 to 5–44 . . . . . . 0 3–7 to 3–9 . . . . . . . . . 0 ii . . . . . . . . . . . . . blank Devices)
i to ii . . . . . . . . . . . . . . 5 6–1 to 6–47 . . . . . . . . 0 3–10 . . . . . . . . . . blank 1–1 to 1–5 . . . . . . . . . 0 i to iii . . . . . . . . . . . . 13
iii to v . . . . . . . . . . . . 10 6–48 . . . . . . . . . . blank 4–1 . . . . . . . . . . . . . . . 0 1–6 . . . . . . . . . . . blank iv . . . . . . . . . . . . . blank
vi . . . . . . . . . . . . . blank 7–1 to 7–20 . . . . . . . . 0 4–2 . . . . . . . . . . . blank 2–1 to 2–9 . . . . . . . . . 0 1–1 to 1–65 . . . . . . . 13
7–21 to 7–23 . . . . . . 1 5–1 . . . . . . . . . . . . . . . 0 2–10 . . . . . . . . . . blank 1–66 . . . . . . . . . . blank
7–24 to 7–27 . . . . . . 0 5–2 . . . . . . . . . . . . . . . 7
Tab 1 (Introduction) 2–1 to 2–71 . . . . . . . 12
7–28 . . . . . . . . . . blank 5–3 . . . . . . . . . . . . . . . 0 Tab 7 (X–ray Genera-
i.................. 0 2–72 . . . . . . . . . . blank
5–4 . . . . . . . . . . . . . . . 7 tor)
ii . . . . . . . . . . . . . blank 3–1 to 3–10 . . . . . . . . 0
5–5 to 5–11 . . . . . . . . 0 i.................. 0
1–1 to 1–6 . . . . . . . . . 0 Tab 3 (Operator Con- 4–1 to 4–2 . . . . . . . . . 0
5–12 . . . . . . . . . . blank ii . . . . . . . . . . . . . blank
2–1 to 2–5 . . . . . . . . . 0 sole) 4–3 . . . . . . . . . . . . . . . 8
6–1 to 6–11 . . . . . . . . 0 1–1 to 1–9 . . . . . . . . . 0
2–6 . . . . . . . . . . . blank i.................. 0 4–4 to 4–6 . . . . . . . . . 0
6–12 . . . . . . . . . . blank 1–10 . . . . . . . . . . blank
3–1 to 3–2 . . . . . . . . . 0 ii . . . . . . . . . . . . . blank 5–1 to 5–7 . . . . . . . . . 0
7–1 to 7–17 . . . . . . . . 0 2–1 to 2–8 . . . . . . . . . 0
1–1 to 1–14 . . . . . . . . 0 5–8 . . . . . . . . . . . blank
7–18 . . . . . . . . . . blank 3–1 to 3–3 . . . . . . . . . 0
2–1 to 2–11 . . . . . . . . 0 6–1 to 6–3 . . . . . . . . . 0
Tab 2 (System) 3–4 . . . . . . . . . . . blank 6–4 . . . . . . . . . . . blank
2–12 . . . . . . . . . . blank
i to ii . . . . . . . . . . . . . 11 4–1 . . . . . . . . . . . . . . . 4
1–1 to 1–22 . . . . . . . . 0 Tab 5 (Table) 4–2 . . . . . . . . . . . blank
Tab APPENDIX
2–1 . . . . . . . . . . . . . . . 5 Tab 4 (Gantry) i.................. 0
A–1 to A–4 . . . . . . . . 0
2–2 . . . . . . . . . . . blank i to ii . . . . . . . . . . . . . . 0 ii . . . . . . . . . . . . . blank Tab 8 (DC III)
3–1 to 3–49 . . . . . . . 11 1–1 to 1–9 . . . . . . . . . 0 1–1 to 1–11 . . . . . . . . 0 i.................. 0
3–50 . . . . . . . . . . blank 1–10 . . . . . . . . . . blank 1–12 . . . . . . . . . . blank ii . . . . . . . . . . . . . blank
4–1 to 4–50 . . . . . . . 11 2–1 to 2–14 . . . . . . . . 0 2–1 to 2–14 . . . . . . . . 0 1–1 to 1–3 . . . . . . . . . 0 Blank/Rear cover . . . –
A
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
blank 2205009
B
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 5 2205009
CONTENTS
INTRODUCTION (TAB 1)
General
Torque Value for Screw Tightening
Common Procedures
SYSTEM (TAB 2)
Checks during System Installation
Software Loading (Qj, Qj–Base, Qj–Hlt, Zj–Adv)
Software Loading (Zj)
X–ray Alignment
System Calibration
Image Performance Verification
OPERATOR CONSOLE (TAB 3)
OC Components
OC Operation
GANTRY (TAB 4)
DC Power Supplies
Rotational Mechanism
Rotational Operation
Slip Rings and Brushes
Tilt Operation
Collimator
Table and Gantry
i
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 5 2205009
CONTENTS (continued)
TABLE (TAB 5)
Elevation Function
Longitudinal Function
Table Accessories
DAS/DETECTOR (TAB 6)
Detector Related Devices
DAS Operation
X–RAY GENERATOR (TAB 7)
X–ray Tube
KV and MA
PDU
Power Supply Adjustment
DC III (TAB 8)
DC III
SWITCH/JUMPER SETTING OF BOARDS/DEVICES (TAB 9)
Operator Console (Qj, Qj–Base, Qj–Hlt, Zj–Adv)
Operator Console (Zj)
Gantry
DAS
X–ray Generator
PDU (Power Distribution Unit)
APPENDIX
Symbols and Classification
ii
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 10 2205009
NOTE (continued)
iii
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 10 2205009
NOTE (continued)
Conversion Formula:
xx (kgfScm) = xxxx (NSm) x 10.2
Conversion Table:
iv
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 10 2205009
NOTE (continued)
1 NSm=0.737 lbfSft
Coversion: NSm –> lbfSft
NSm 0 1 2 3 4 5 6 7 8 9
lbfSft
0 0 0.737 1.474 2.211 2.948 3.685 4.422 5.159 5.896 6.633
10 7.370 8.107 8.844 9.581 10.318 11.055 11.792 12.529 13.266 14.003
20 14.740 15.477 16.214 16.951 17.688 18.425 19.162 19.899 20.636 21.373
30 22.110 22.847 23.584 24.321 25.058 25.795 26.532 27.269 28.006 28.743
40 29.480 30.217 30.954 31.691 32.428 33.165 33.902 34.639 35.376 36.113
50 36.850 37.587 38.324 39.061 39.798 40.535 41.272 42.009 42.746 43.483
60 44.220 44.957 45.694 46.431 47.168 47.905 48.642 49.379 50.116 50.853
70 51.590 52.327 53.064 53.801 54.538 55.275 56.012 56.749 57.486 58.223
80 58.960 59.697 60.434 61.171 61.908 62.645 63.382 64.119 64.856 65.593
90 66.330 67.067 67.804 68.541 69.278 70.015 70.752 71.489 72.226 72.963
100 73.700
v
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
blank 2205009
vi
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
INTRODUCTION
TABLE OF CONTENTS
SECTION PAGE
i INTRODUCTION
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
blank 2205009
ii INTRODUCTION
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
SECTION 1 – GENERAL
WARNING!
DEATH OR SERIOUS INJURY!!
READ AND THOROUGHLY UNDERSTAND THE ‘PROSPEED SERIES SAFETY GUIDELINES’
MANUAL (P9011AW) BEFORE PERFORMING ANY PROCEDURE IN THIS ‘FUNCTIONAL
CHECK/ADJUSTMENT.’
1–1 INTRODUCTION
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
Each of the code names represents a product family; for example, ProSpeed S, ProSpeed SX, and Lemage constitute
a product family, called ‘Qj.’
In this manual, product names (ProSpeed, Lemage, ...) or product family names (i.e., code names; Qj, Zj–Adv, ...)
are used to indicate some sections, illustrations, or other types of information which are only applied to those products
indicated.
1–2 INTRODUCTION
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
Xe Hilight
D A during– and post–installation check to ensure that the entire system operates properly.
D As a reference source for Periodic Maintenance which mainly contains abbreviated lists.
D A post–diagnostic check to verify system operation after repairs, before returning it to the customer to use.
1–3 INTRODUCTION
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
1-2-1 Gantry
WARNING!
ROTATION AND ELECTROCUTION HAZARDS!
SWITCH OFF THE ‘ROTATE’ AND ‘XG (SAFETY LOOP)’ SWITCHES LOCATED AT THE REAR
OF THE GANTRY BASE BEFORE INSERTING ANY OBJECT INTO OR WORKING ON THE
GANTRY.
WARNING!
ROTATION HAZARD!
ALWAYS INSERT THE GANTRY AZIMUTH LOCK PIN BEFORE REPLACING THE TUBE, DAS,
DETECTOR, HV TANK, OR OTHER WEIGHTY COMPONENTS.
WARNING!
ELECTROCUTION!
DANGEROUS VOLTAGES ARE PRESENT ON UNSHIELDED TERMINALS. USE EXTREME
CAUTION WHEN WORKING ON LIVE EQUIPMENT.
WARNING!
ROTATION HAZARD!
ANY OF THE FOLLOWINGS WILL INITIATE THE GANTRY ROTATION:
1. POWERING ON THE SYSTEM OR GANTRY.
2. SWITCHING ON THE ‘CONTROL’ SWITCH.
3. PUSHING THE RESET SWITCH ON THE TGP BOARD.
4. PUSHING THE POSITIONING LIGHT BUTTON.
STAND CLEAR OF THE ROTATING GANTRY.
1–4 INTRODUCTION
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
WARNING!
ELECTRICAL SHOCK HAZARD!
DANGEROUS VOLTAGES EXIST ON EXPOSED SURFACES. USE EXTREME CAUTION
WHEN WORKING ON LIVE EQUIPMENT.
CAUTION
Rotating fan assemblies!
Do not place any item into rotating blades. Use extra care when working near fan assemblies.
NOTICE
1–5 INTRODUCTION
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
CE MARK-
ING LABEL
OC REAR
VIEW
1–6 INTRODUCTION
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
CAUTION
Use locktite for screws which are installed to the Gantry rotation block. Use stud bolt with
locktite or nylon nut to prevent loosening if a screw is removed in service maintenance.
CAUTION
When you secure stud bolts with locktite, you must keep the screw torque shown in Table
‘Stud Bolt (Steel) with Locktite.’
Do not rotate the Gantry until the locktite hardens. (75 % of maximum strength after 2 hours;
100 % after 16 hours.)
If it is not possible to wait for the time period mentioned above, use nylon nuts.
2–1 INTRODUCTION
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
Style
2–2 INTRODUCTION
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
M 1.7 – – –
M2 1.6 1.7 2.1
M 2.2 2.0 2.2 2.7
M 2.3 2.5 2.7 3.4
M 2.5 3.2 3.4 4.2
M 2.6 3.6 3.9 4.8
M3 5.6 6.0 7.5
M 3.5 8.7 9.3 11.5
M4 13 13.5 17
M 4.5 18.5 20 25
M5 26 27.5 34.5
M6 44 47 58.5
M8 107 114 143
Style
2–3 INTRODUCTION
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
2–4 INTRODUCTION
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
Style
or
L L
2–5 INTRODUCTION
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
blank 2205009
2–6 INTRODUCTION
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
4. Enter its positions (x POSN: 0.00 cm, y POSN: 0.00 cm) and diameter dimensions (Dia: 2.00 cm).
6. Press the [END] key several times to exit the Image Measurement menu.
3–1 INTRODUCTION
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
1. Enter the Management menu by pressing the key.
>
3–2 INTRODUCTION
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 11 2205009
SYSTEM
TABLE OF CONTENTS
SECTION PAGE
i SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 11 2205009
TABLE OF CONTENTS
SECTION PAGE
ii SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
1-1 GENERAL
This ‘Checks during System Installation’ is performed during the following:
System Installation
Periodic Maintenance.
For System Installation, it is required that all wiring and cabling are complete and equipment is installed.
Proceed to Section 1-2 ‘Checks prior to Applying Power’, and then, to Section 1-3 ‘Ground Continuity’, Section 1-4
‘Temperature,’ ... , in the section order.
1–1 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
CAUTION
Leave all power to the system OFF and tagged until power is called for in specific steps.
D 5 Switches (‘Rotate,’ ‘XG,’ ‘Table/Tilt,’ ‘Control,’ and ’Slip Ring 115V’) and 1 breaker (NFB) on the Gantry
D
(Qj, Qj–Base, Qj–Hlt)
D
(Zj, Zj–Adv)
1–2 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
P9180AB/AF
2121798
ÅÅÅÅÅ
ÅÅÅÅÅ
ÅÅÅÅÅ
1–3 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
VIEW B NFB
SUB
BOARD TGP
BOARD
A
DETAIL OF `A'
SERVICE
Rotate XG Table/Tilt Slip Ring SW
Control 115V SYS-OFF-MNL
(Safety Loop) SW
(DAS/DET)
1–4 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
1–5 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
(TOP VIEW)
1–6 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
2. Use an earth bonding test meter or equivalent device to measure the DC resistance between the following com-
ponents in the system, and verify that each is less than 0.1 ohms. Record the measured resistance values.
RESISTANCE
FROM TO (OHMS)
GND (OC)
* GND (MT)
GND (OC)
* GND (MFC)
GND (OC)
* GND (OD)
GND (OC)
*GND
(ADVANTAGE WINĆ
DOWS or DC III)
1–7 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
1-4 TEMPERATURE
During the installation and start–up of the system you must take continuous temperature checks to make sure the
environment meets specifications listed in Section 3 of the Preinstallation manual (2124010). Excessive temperature,
coupled with reduced air flow from clogged filters, may overstress components and cause failures.
Filters
Check and clean all electronic equipment dust filters. GE recommends initially doing this on a weekly basis, because
new installations always have dust and dirt present at this stage. Over a period of time, you may increase the mainte-
nance interval as site environment improves.
Specifications
During installation, use the temperature/humidity recorder and a thermometer to record the temperature and relative
humidity in the scan room. Make sure the data you collect meets the following temperature and humidity
specifications.
D Scan Room
Action
Contact the air conditioning supplier or your local GE service department when your data fails to meet any
specifications.
1–8 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
Verify that an input selection on the Power Distribution Unit (See Illustration 1–4) meets the input power voltage. If not,
correct them.
The ‘input power voltage’ here means the rating voltage specified on the power distribution box (hospital supply), not
the measured voltage on the power distribution box.
F4 F5 F6
R S T
FOR 480 V INPUT
F4 F5 F6
R S T
FOR 415 V INPUT
F4 F5 F6
TB37
R S T
FOR 380 V INPUT
SW20
FRONT VIEW
1–9 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
Instrument
Use the same instrument (Dranetz, etc.) to monitor line voltage during this check that you used during the preinstalla-
tion check. The recommended line analyzer is the Dranetz Model 606–3 (3 channel) with the 101 frequency option, or
Dranetz Model line monitor plug in modules. You can use a similar analyzer, as long as it can monitor and report mean
voltage, surges, sags, short term impulses, and frequency drift.
D Verify that the current line voltage equals the voltage measured during the preinstallation check.
D Verify that mean line voltage does not fall outside of equipment specifications during the day to day opera-
tion of other site equipment during the installation period.
D Look for any line voltage disturbances and check for a correlation between disturbances and system mal-
functions such as “Aborts” or component failures.
From the printouts, ensure that the power supply meets the following criteria :
D Line Voltage : 200 VAC, 208 VAC, 380 VAC, 400 VAC, 415 VAC, 440 VAC, 460 VAC or
480 VAC
(50 Hz/60 Hz) (Normal Steady State)
D Daily Voltage Variation : +10% to –5% from nominal steady state (50 Hz)
+6% to –9% from nominal steady state (60 Hz)
Action
Investigate any condition that falls out of spec. Identify the problem source(s) and resolve the problem.
1–10 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
WARNING!
DANGEROUS VOLTAGES EXIST WITHIN THE UNIT, USE EXTREME CARE TO AVOID INJURY
TO PERSONNEL AND EQUIPMENT.
Follow this procedure exactly in the order it was written. This procedure was developed with your safety in mind.
Follow this sequence to protect you and the equipment from irreparable damage.
d. Verify again that all breakers and switches are OFF according to ‘Checks prior to Applying Power.’
f. Switch ON two breakers (NFB1, NFB2) on the OC. (Qj, Qj–Base, Qj–Hlt)
1–11 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
Verify the ‘Power’ lamp (green) in the front upper left corner of the PDU turns ON.
5. Measure 115 VAC between terminals 15 and 16 on TB37 on the PDU. It should be 103.5 ~ 126.5 VAC. See
Illustration 1–4.
6. Switch ON the ‘Slip Ring 115V (DAS/DET)’ switch at the Gantry rear base.
Verify that the x–ray generator on the rotating assembly is powered ON; this can be verified by checking that
LEDs on the XG controller boards turn ON. Also verify that the x–ray tube fan and pump start to operate.
Verify that the voltage between terminals 11 and 12 on TM 1 of the OC is 90 ~ 110 VAC. See Illustration 1–3.
Verify that the voltage between terminals 1 and 2, and between terminals 2 and 3 on TM1 of the OC is 90 ~ 110
VAC. See Illustration 1–3.
9. Measure 103.5 ~ 126.5 VAC between terminals 10 and 11 on TB37 on the PDU.
10. Verify that the Servo Amplifier breaker is in the OFF position.
12. Measure 180 ~ 220 VAC on the Servo Amplifier input (between R & S, S & T, and T & R terminals). See Illustration
1–5.
1–12 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
a. Remove the top and front covers from the PDU. (A ground cable is connected on each cover)
d. Verify again that all breakers and switches are OFF according to ‘Checks prior to Applying Power.’
f. Switch ON two breakers (NFB1, NFB2) on the OC. (Qj, Qj–Base, Qj–Hlt)
Verify the ‘Power’ lamp in the front upper left corner of the PDU turns ON.
5. Measure 103.5 ~ 126.5 VAC between terminals 8 and 9 on the front side of the PDU. See Illustration 1–6.
TERMINALS FOR
MEASUREMENT
1–13 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
Verify that the x–ray generator on the rotating assembly is powered ON; this can be verified by checking that
LEDs on the XG controller boards turn ON. Also verify that the x–ray tube fan and pump start to operate.
Verify that the voltage between terminals 11 and 12 on TM 1 of the OC is 90 ~ 110 VAC. See Illustration 1–3.
Verify that the voltage between terminals 1 and 2, and between terminals 2 and 3 on TM1 of the OC is 90 ~ 110
VAC. See Illustration 1–3.
9. Measure 103.5 ~ 126.5 VAC between terminals 11 and 12 on the front side of the PDU. See Illustration 1–6.
10. Verify that the Servo Amplifier breaker is in the OFF position.
12. Measure 180 ~ 220 VAC on the Servo Amplifier input (between R & S, S & T, and T & R terminals). See Illustration
1–5.
1–14 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
WARNING!
2. Remove the front and both side covers from the Gantry.
3. Set the safety bracket located on the right side of the Gantry.
6. Rotate the Gantry manually in the CW direction to verify that nothing interferes the Gantry during rotation (cables,
etc.).
Verify that the Gantry starts to rotate, and it stops at the home position (x–ray tube at 12 o’clock position) before
making two revolutions.
1–15 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
Verify that power is removed from the Gantry and Table; check the following:
Verify that power is applied to the Gantry and Table; check the following:
(Zj, Zj–Adv)
4. Press the Emergency Reset switch next to the emergency switch. (Use a fine tip of some object)
Verify that power is applied to the Gantry and Table; check the following:
5. Repeat the above check for emergency switches located on the followings:
PDU (one)
Gantry front cover (two; right and left)
Gantry frames (four; all the Gantry covers should be removed)
1–16 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
1-9 FILTER
(Qj, Qj–Base, Qj–Hlt)
There are one filter on the OC and two on the PDU.
(Zj, Zj–Adv)
There are two filters on the OC and two on the PDU.
1. Slide the filter at the bottom of the OC forward until it is removed from the OC.
(The OC front cover does not need to be removed)
2. Remove the front and rear covers (or right and left side covers) of PDU and gently pull the foam out of each enclo-
sure.
1–17 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
1. Switch OFF the ‘Rotate’ switch at the Gantry rear base. (The other switches should be set to ON)
2. Remove the front and both side covers from the Gantry.
b. Rotate the Gantry manually in the CW direction to the home position (x–ray tube at the 12 o’clock position).
Verify that a scan can not be performed and that the “XG ERROR 148” message is reported on the CRT status
screen.
CONNECTOR
(BROWN)
1–18 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
Warm Up Scan
1. Make sure that the ‘New Patient’ and the ‘Warm Up’ buttons are blinking on the touch panel screen.
Note
‘Warm Up’ should blink following the completion of the system power–up sequence.
2. Make sure that nothing is in the scan area in the Gantry room.
3. Touch the ‘Warm Up’ button to start the x–ray tube warm up program.
Scoutview Scan
4. Perform a series of scout scans to check scout scan operations:
FOV : 50 cm
THICKNESS : 1 mm
KV : 120 kV
MA : 60, 80, 100, 130, or 160 mA
Make sure that the software prohibits a scout scan with the Gantry tilted.
Make sure that you can hear and understand each voice.
Verify that the phantom diameter (In–Out direction) on the image is 266 mm ±3 mm.
1–19 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
Make sure that you can hear and understand each voice.
Release the ‘Scan’ button before the whole image is displayed and verify that the scan continues.
Push the ‘Abort’ key and verify that the cradle and scan stop.
Localize
5. Perform a Localize function:
1–20 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
Axial Scan
6. Perform a series of axial scans to check axial scan operations:
Acquisition No. 1 2 3 4 5 6 7 8 9 10
Scan Time [sec] 0.7 0.7 1 1 1.5 1.5 2 2 3 3
(For ProSpeed SX, ProSpeed SX Power, Lemage, Lemage SX, Lemage SX/E, ProSpeed SX Advan-
tage, Lemage Supreme, ProSpeed Advantage, ProSeed SA, ProSeed SA Libra, ProSpeed, ProSpeed
Plus, ProSeed Accell, ProSeed Accell EI, ProSeed, ProSeed EF)
For systems where 0.8 sec scan is available, use the following, instead of the above.
Acquisition No. 1 2 3 4 5 6 7 8 9 10 11 12
Scan Time [sec] 0.5 0.5 0.8 0.8 1 1 1.5 1.5 2 2 3 3
Acquisition No. 1 2 3 4 5 6 7 8
Scan Time [sec] 1 1 1.5 1.5 2 2 3 3
Acquisition No. 1 2 3 4
Scan Time [sec] 2 2 3 3
THICKNESS : 10 mm
FOCUS : LARGE
KV : 120 kV
MA : 160 mA
INTERVAL : 10 mm
Select each of the auto voice messages (Start 1, Start 2, Start 3, and End 1).
1–21 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
(For ProSpeed SX, ProSpeed SX Power, Lemage, Lemage SX, Lemage SX/E, ProSpeed SX Advan-
tage, Lemage Supreme, ProSpeed Advantage, ProSeed SA, ProSeed SA Libra, ProSpeed, ProSpeed
Plus, ProSeed Accell, ProSeed Accell EI, ProSeed, ProSeed EF)
Recon Time
Verify that the time between the scan end and the image appearance on the CRT monitor is within 9 sec.
Abort
Perform the above scan and press the ‘Abort’ button before the scan completes.
Verify that the scan is terminated.
Emergency OFF
Perform the above scan and press the Emergency button on the Console before the scan completes.
Verify that the scan is terminated and that the Gantry comes to a stop within two inertial revolutions.
1–22 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 5 2205009
SECTION 2 – (BLANK)
2–1 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
blank 2205009
2–2 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 11 2205009
3-1 GENERAL
This ‘Software Loading’ describes the Load–From–Cold (LFC) procedure for a full system software loading from an
MOD (Magnetic Optical Disk) and some FDs (Floppy Disks) to the system hard disk. Perform this LFC procedure
when:
D A failure or crash corrupts the hard disk so the existing software can not be used. A Load–From–Cold
rebuilds the operating software on the disk and restores normal system operation.
D CT Engineering has revised the current software to incorporate improvements or correct existing prob-
lems. Replace the old software by performing a Load–From–Cold with the new software.
CAUTION
Save existing patient data any time you perform the LFC procedure or you will lose this data.
The saving procedure is described in the ‘Pre–ATINST Procedure’ menu.
3–1 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 11 2205009
START
ATINST program
DATA FILE SAVE
DISK INITIALIZE
SOFTWARE LOADING (STANDARD,
OPTION)
FILE ALLOCATION
DATA FILE LOADING
END
3–2 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 11 2205009
D System Software Disks listed in Table 3–1 ∼ 3–15 (see the table below)
The table lists the Volume Names of an MOD which contains the system software, of some floppy disks
which contain application software, and of other floppy disks required for the system.
D Option Software Disks (which the customer already own) listed in Table 3–16.
This table shows all the currently available option software.
Normally, option software disks and also some application software disks included in the system software
(“PG00” disk, “PZ00” disk, ...) are not required in a version up or a re–loading procedure. However, these
disks are required if the ATINST program (described later) failed, or when the system hard disk drive is
replaced with a new one.
D Option Software Disks (which the customer bought and have not yet installed to the system) listed in Table
3–16.
3–3 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 11 2205009
BOOT 5 (2196037)
MQ00 5 System Software V/R 6.00 (2196039) (MOD)
V602 0 System Software (2213371) (for Version 6.02)
V603 0 System Software (2224498) (for Version 6.03)
PG00 0 Paging (P9200QH)
PZ00 0 Pan-zoom (P9200QJ)
DIND 0 Display in Display (P9200QE)
RTRF 0 Real-time Reformation (P9200QK)
CVRF 0 Curved Reformation (P9200QL)
BOOT 5 (2196037)
MQ00 5 System Software V/R 6.00 (2196039) (MOD)
V602 0 System Software (2213371) (for Version 6.02)
V603 0 System Software (2224498) (for Version 6.03)
CLV3 0 Cooling V3 (2123692)
STM1 0 One Sec Scan (2123687)
PG00 0 Paging (P9200QH)
PZ00 0 Pan-zoom (P9200QJ)
DIND 0 Display in Display (P9200QE)
RTRF 0 Real-time Reformation (P9200QK)
CVRF 0 Curved Reformation (P9200QL)
3–4 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 11 2205009
BOOT 5 (2196037) O O O O
MQ00 5 System Software V/R 6.00 (2196039) (MOD) O O O O
V602 0 System Software (2213371) (for Version 6.02) O O O O
V603 0 System Software (2224498) (for Version 6.03) O O O O
FRN1 0 Fast Recon 5-7 (2123690) O O O O
CLV3 0 Cooling V3 (2123692) O O - -
CLV4 0 Cooling V4 (2123693) - - O O
STM1 0 One Sec Scan (2123687) - O O O
PG00 0 Paging (P9200QH) O O O O
PZ00 0 Pan-zoom (P9200QJ) O O O O
DIND 0 Display in Display (P9200QE) O O O O
RTRF 0 Real-time Reformation (P9200QK) O O O O
CVRF 0 Curved Reformation (P9200QL) O O O O
HLCL 0 Helical Scan (P9200QS) - O O O
CINE 0 Cine Retro Recon (P9200QM) - O O O
Model B: Systems with a metal-free cradle installed as an option, or systems not including a metal-free cradle.
Model E: Systems with a metal-free cradle as standard and a Gantry not marked with an `R' at its rear base.
Model R: Systems with a Gantry marked with an `R' at its rear base.
Model R2: Systems with a Gantry marked with an `R2' at its rear base.
O: The system includes the software.
-: The system does not include the software.
3–5 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 11 2205009
BOOT 5 (2196037) O O O O
MQ00 5 System Software V/R 6.00 (2196039) (MOD) O O O O
V602 0 System Software (2213371) (for Version 6.02) O O O O
V603 0 System Software (2224498) (for Version 6.03) O O O O
FRN1 0 Fast Recon 5-7 (2123690) O O O O
CLV3 0 Cooling V3 (2123692) O - - -
CLV4 0 Cooling V4 (2123693) - O O O
STM1 0 One Sec Scan (2123687) - O O O
PG00 0 Paging (P9200QH) O O O O
PZ00 0 Pan-zoom (P9200QJ) O O O O
DIND 0 Display in Display (P9200QE) O O O O
RTRF 0 Real-time Reformation (P9200QK) O O O O
CVRF 0 Curved Reformation (P9200QL) O O O O
HLCL 0 Helical Scan (P9200QS) - O O O
CINE 0 Cine Retro Recon (P9200QM) - O O O
Model B: Systems with a metal-free cradle installed as an option, or systems not including a metal-free cradle.
Model E: Systems with a metal-free cradle as standard and a Gantry not marked with an `R' at its rear base.
Model R: Systems with a Gantry marked with an `R' at its rear base.
Model R2: Systems with a Gantry marked with an `R2' at its rear base.
O: The system includes the software.
-: The system does not include the software.
3–6 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 11 2205009
BOOT 5 (2196037) O O O O
MQ00 5 System Software V/R 6.00 (2196039) (MOD) O O O O
V602 0 System Software (2213371) (for Version 6.02) O O O O
V603 0 System Software (2224498) (for Version 6.03) O O O O
FRN1 0 Fast Recon 5-7 (2123690) O O O O
CLV3 0 Cooling V3 (2123692) O O - -
CLV4 0 Cooling V4 (2123693) - - O O
STM1 0 One Sec Scan (2123687) - O O O
PG00 0 Paging (P9200QH) O O O O
PZ00 0 Pan-zoom (P9200QJ) O O O O
DIND 0 Display in Display (P9200QE) O O O O
RTRF 0 Real-time Reformation (P9200QK) O O O O
CVRF 0 Curved Reformation (P9200QL) O O O O
HLCL 0 Helical Scan (P9200QS) - O O O
CINE 0 Cine Retro Recon (P9200QM) - O O O
HILT 0 HILIGHT (2123453) O O O O
Model B: Systems with a metal-free cradle installed as an option, or systems not including a metal-free cradle.
Model E: Systems with a metal-free cradle as standard and a Gantry not marked with an `R' at its rear base.
Model R: Systems with a Gantry marked with an `R' at its rear base.
Model R2: Systems with a Gantry marked with an `R2' at its rear base.
O: The system includes the software.
-: The system does not include the software.
3–7 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 11 2205009
BOOT 5 (2196037) O O O O
MQ00 5 System Software V/R 6.00 (2196039) (MOD) O O O O
V602 0 System Software (2213371) (for Version 6.02) O O O O
V603 0 System Software (2224498) (for Version 6.03) O O O O
FRN1 0 Fast Recon 5-7 (2123690) O O O O
CLV3 0 Cooling V3 (2123692) O - - -
CLV5 0 Cooling V5 (2141556) - O O O
STM1 0 One Sec Scan (2123687) O O O O
MA35 0 350 mA (2123689) O O O O
PG00 0 Paging (P9200QH) O O O O
PZ00 0 Pan-zoom (P9200QJ) O O O O
DIND 0 Display in Display (P9200QE) O O O O
RTRF 0 Real-time Reformation (P9200QK) O O O O
CVRF 0 Curved Reformation (P9200QL) O O O O
HLCL 0 Helical Scan (P9200QS) - O O O
CINE 0 Cine Retro Recon (P9200QM) - O O O
Model B: Systems with a metal-free cradle installed as an option, or systems not including a metal-free cradle.
Model E: Systems with a metal-free cradle as standard and a Gantry not marked with an `R' at its rear base.
Model R: Systems with a Gantry marked with an `R' at its rear base.
Model R2: Systems with a Gantry marked with an `R2' at its rear base.
O: The system includes the software.
-: The system does not include the software.
3–8 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 11 2205009
Table 3–7 System Software Disks (ProSpeed SX Power with Hilight Detector)
BOOT 5 (2196037) O O O O
MQ00 5 System Software V/R 6.00 (2196039) (MOD) O O O O
V602 0 System Software (2213371) (for Version 6.02) O O O O
V603 0 System Software (2224498) (for Version 6.03) O O O O
FRN1 0 Fast Recon 5-7 (2123690) O O O O
CLV3 0 Cooling V3 (2123692) O - - -
CLV5 0 Cooling V5 (2141556) - O O O
STM1 0 One Sec Scan (2123687) O O O O
MA35 0 350 mA (2123689) O O O O
PG00 0 Paging (P9200QH) O O O O
PZ00 0 Pan-zoom (P9200QJ) O O O O
DIND 0 Display in Display (P9200QE) O O O O
RTRF 0 Real-time Reformation (P9200QK) O O O O
CVRF 0 Curved Reformation (P9200QL) O O O O
HLCL 0 Helical Scan (P9200QS) - O O O
CINE 0 Cine Retro Recon (P9200QM) - O O O
HILT 0 HILIGHT (2123453) O O O O
Model B: Systems with a metal-free cradle installed as an option, or systems not including a metal-free cradle.
Model E: Systems with a metal-free cradle as standard and a Gantry not marked with an `R' at its rear base.
Model R: Systems with a Gantry marked with an `R' at its rear base.
Model R2: Systems with a Gantry marked with an `R2' at its rear base.
O: The system includes the software.
-: The system does not include the software.
3–9 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 11 2205009
BOOT 5 (2196037) O O O
MQ00 5 System Software V/R 6.00 (2196039) (MOD) O O O
V602 0 System Software (2213371) (for Version 6.02) O O O
V603 0 System Software (2224498) (for Version 6.03) O O O
FRN2 0 Fast Recon 3-5 (2123691) O O O
MA35 0 350 mA (2123689) O O O
CLV5 0 Cooling V5 (2141556) O O O
STM1 0 One Sec Scan (2123687) O O O
FELK 0 Fast E-Link (2149250) O O O
PG00 0 Paging (P9200QH) O O O
PZ00 0 Pan-zoom (P9200QJ) O O O
DIND 0 Display in Display (P9200QE) O O O
RTRF 0 Real-time Reformation (P9200QK) O O O
CVRF 0 Curved Reformation (P9200QL) O O O
HLCL 0 Helical Scan (P9200QS) O O O
CINE 0 Cine Retro Recon (P9200QM) O O O
HILT 0 HILIGHT (2123453) O O O
Model E: Systems with a metal-free cradle as standard and a Gantry not marked with an `R' at its rear base.
Model R: Systems with a Gantry marked with an `R' at its rear base.
Model R2: Systems with a Gantry marked with an `R2' at its rear base.
O: The system includes the software.
-: The system does not include the software.
3–10 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 11 2205009
BOOT 5 (2196037)
MQ00 5 System Software V/R 6.00 (2196038) (MOD)
V602 0 System Software (2213371) (for Version 6.02)
V603 0 System Software (2224498) (for Version 6.03)
STM1 0 One Sec Scan (2123687)
CLV4 0 Cooling V4 (2123693)
PG00 0 Paging (P9200QH)
PZ00 0 Pan-zoom (P9200QJ)
DIND 0 Display in Display (P9200QE)
HLCL 0 Helical Scan (P9200QS)
CINE 0 Cine Retro Recon (P9200QM)
3–11 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 11 2205009
BOOT 5 (2196037) O O O O
MQ00 5 System Software V/R 6.00 (2196038) (MOD) O O O O
V602 0 System Software (2213371) (for Version 6.02) O O O O
V603 0 System Software (2224498) (for Version 6.03) O O O O
STM1 0 One Sec Scan (2123687) O O O O
MA30 0 300 mA (2123688) O O O O
FRN1 0 Fast Recon 5-7 (2123690) O O O O
CLV4 0 Cooling V4 (2123693) O - - -
CLV5 0 Cooling V5 (2141556) - O O O
PG00 0 Paging (P9200QH) O O O O
PZ00 0 Pan-zoom (P9200QJ) O O O O
DIND 0 Display in Display (P9200QE) O O O O
HLCL 0 Helical Scan (P9200QS) O O O O
CINE 0 Cine Retro Recon (P9200QM) O O O O
MOD4 0 Quad Density MOD (2169037) - - - O
Model B: Systems with a metal-free cradle installed as an option, or systems not including a metal-free cradle.
Model E: Systems with a metal-free cradle as standard and a Gantry not marked with an `R' at its rear base.
Model R: Systems with a Gantry marked with an `R' at its rear base.
Model R2: Systems with a Gantry marked with an `R2' at its rear base.
O: The system includes the software.
-: The system does not include the software.
3–12 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 11 2205009
BOOT 5 (2196037) O O O O
MQ00 5 System Software V/R 6.00 (2196038) (MOD) O O O O
V602 0 System Software (2213371) (for Version 6.02) O O O O
V603 0 System Software (2224498) (for Version 6.03) O O O O
STM1 0 One Sec Scan (2123687) O O O O
FRN1 0 Fast Recon 5-7 (2123690) - O O O
CLV4 0 Cooling V4 (2123693) O O O O
PG00 0 Paging (P9200QH) O O O O
PZ00 0 Pan-zoom (P9200QJ) O O O O
DIND 0 Display in Display (P9200QE) O O O O
HLCL 0 Helical Scan (P9200QS) O O O O
CINE 0 Cine Retro Recon (P9200QM) O O O O
MOD4 0 Quad Density MOD (2169037) - - - O
Model B: Systems with a metal-free cradle installed as an option, or systems not including a metal-free cradle.
Model E: Systems with a metal-free cradle as standard and a Gantry not marked with an `R' at its rear base.
Model R: Systems with a Gantry marked with an `R' at its rear base.
Model R2: Systems with a Gantry marked with an `R2' at its rear base.
O: The system includes the software.
-: The system does not include the software.
3–13 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 11 2205009
BOOT 5 (2196037) O O O O O
MQ00 5 System Software V/R 6.00 (2196038) (MOD) O O O O O
V602 0 System Software (2213371) (for Version 6.02) O O O O O
V603 0 System Software (2224498) (for Version 6.03) O O O O O
STM1 0 One Sec Scan (2123687) O O O O O
MA30 0 300 mA (2123688) O O O O O
FRN1 0 Fast Recon 5-7 (2123690) O O O - O
FRN2 0 Fast Recon 3-5 (2123691) - - - O -
CLV4 0 Cooling V4 (2123693) O - - - -
CLV5 0 Cooling V5 (2141556) - O O O O
PG00 0 Paging (P9200QH) O O O O O
PZ00 0 Pan-zoom (P9200QJ) O O O O O
DIND 0 Display in Display (P9200QE) O O O O O
HLCL 0 Helical Scan (P9200QS) O O O O O
CINE 0 Cine Retro Recon (P9200QM) O O O O O
HILT 0 HILIGHT (2123453) O O O O O
SRCN 0 Smart Recon (2167710) - - - O -
MOD4 0 Quad Density MOD (2169037) - - - - O
Model B: Systems with a metal-free cradle installed as an option, or systems not including a metal-free cradle.
Model E: Systems with a metal-free cradle as standard and a Gantry not marked with an `R' at its rear base.
Model R: Systems with a Gantry marked with an `R' at its rear base.
Model L: Lemage Supreme (2168716); this model includes Smart Recon software.
Model R2: Systems with a Gantry marked with an `R2' at its rear base.
O: The system includes the software.
-: The system does not include the software.
3–14 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 11 2205009
Model B: Systems with a metal-free cradle installed as an option, or systems not including a metal-free cradle.
Model E: Systems with a metal-free cradle as standard and a Gantry not marked with an `R' at its rear base.
Model R: Systems with a Gantry marked with an `R' at its rear base.
Model R2: Systems with a Gantry marked with an `R2' at its rear base.
O: The system includes the software.
-: The system does not include the software.
3–15 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 11 2205009
Model R: Systems with a Gantry marked with an `R' at its rear base.
Model R2: Systems with a Gantry marked with an `R2' at its rear base.
O: The system includes the software.
-: The system does not include the software.
3–16 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 11 2205009
CD00 0 Co-ordinate
CCI0 0 CCI
CB00 0 CBF
BMD0 0 BMD
ELNK 0 Ethernet
MRS0 0 MARS
PG00 0 Paging
PZ00 0 Pan-zoom
HILT 0 HILIGHT
MA30 0 300 mA
MA35 0 350 mA
3–17 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 11 2205009
About MOD Cartridge Type. For data save purpose, use MOD cartridges compatible with the Hitachi MOD drive in
Japanese sites; or use MOD cartridges compatible with the Pioneer MOD drive in except Japanese sites.
About Floppy Disks. For data save purpose, use blank floppy disks, or initialize them with a volume name ‘FPAR’,
even though they are already initialized with this volume name, otherwise, the previously saved data on the floppy
disks will persist (not be overwritten)!
Refer to Section 3-1-2, ‘Floppy Disk Initialize’, of Tab ‘Introduction.’
Archive existing patient image data to MOD, floppy disks, tape, or optical disk before proceeding, or you will lose
this data. Use the Image File Management function to save data. (Refer to Operator Manual)
For any system on which ELINK is installed, save Communication Data on a floppy disk as follows, otherwise, the
data will be deleted during LFC.
...
3–18 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 11 2205009
CAUTION
If you intend to save raw data onto MOD, be sure to start with a blank cartridge , or one that
does not need to be preserved. This procedure will erase all previous data contained on the
MOD cartridge.
Archive raw data files to an MOD or floppy disks if the customer wants to save them. Otherwise, these files are
destroyed during LFC.
Before proceeding to the following save procedure, write down all the raw data file ID numbers to be saved.
If you use floppy disks, format them before proceeding.
a. Insert the BOOT floppy into the disk drive. (This step is not required for systems installed in Japan)
i. Enter all the raw data file ID Numbers. Then press the [END] key.
...
3–19 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 11 2205009
Note
Usually the Custom Parameter data does not need to be recorded, since the data will be saved
automatically in the ATINST program (Data File Save Function). However, in this LFC (Re–loading
of the current system software), at least the following parameters should be recorded:
Network (since you will be requested to select the current setting parameter in the CHANGE PA-
RAMETERS mode later)
Also, it is recommended to check which option software are currently installed on the system, so that
you can confirm later that those option software have been correctly retrieved.
In addition, if you have all the records, they can serve as backup in the event there is trouble with
the ATINST program.
** CUSTOM PARAMETERS **
1. DISPLAY PARAMETERS
2. CHANGE PARAMETERS
Select No. = 1 ‘CR’
f. Custom Parameters will be displayed one by one, as listed in Table 3–17. Use this table to record the parame-
ters.
Custom Parameters include FUNCTION MENU which does not show in the Display Parameters mode.
Therefore, enter the Change Parameters mode (BOOT floppy is required) and record the parameters in Table
3–18.
3–20 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 11 2205009
Frequency 50 . 60
MACHINE No.
HOSPITAL NAME
Disk Size
3–21 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 11 2205009
Apex Lateral
OPTION
3–22 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 11 2205009
OPTION
FUNCTION MENU
AUTO STORE DEVICE OPTION OD. MT. MOD
OD SPECIAL STORE OPTION NON. AUTO
AUTO DELETE OPTION ON. OFF
MIN VACANT IMAGE SPACE FOR SCAN
FD PACKING OPTION DPCM. ORIGINAL
MT PACKING OPTION DPCM. ORIGINAL
OD PACKING OPTION DPCM. ORIGINAL
FD SYSTEM NO.
MT SYSTEM NO.
OD SYSTEM NO.
MT IMAGE MAX. NO. 128. 256. 384. 512. 640. 768
DEVICE TYPE OF OPTICAL DISK OD. MOD (See Note1.)
Note 1 This parameter appears only if both OD and MOD are installed.
3–23 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 11 2205009
Note
Usually the sets of information below do not need to be saved since they will be automatically saved
in the ATINST program. However, if you save them, they can serve as backup in the event there is
trouble with the ATINST program.
Save the following data files on floppy disks, following the steps below.
1. Scan Parameter
2. Patients Information
3. Series Description
4. Calibration Files
5. Station Number Information
6. Voice Information
7. System Parameters (including Custom Parameters)
...
c. Insert the BOOT floppy into the disk drive. (This step is not required for systems installed in Japan)
f. Select SYSTEM PARAMETERS DUMP & LOAD from the Maintenance menu to save System Parameters
(including Custom Parameters).
3–24 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 11 2205009
For systems with a CBF option, record the user setting parameters, otherwise, the data will be deleted during
the ATINST procedure.
For systems with a Functional Image option, record the user setting parameters, otherwise, the data will be de-
leted during the ATINST procedure.
For systems where the user sets `Quick Filming ON', the Quick Filming Delay Time must be recorded, other-
wise, the data will be deleted during the ATINST procedure.
Mgt
a. Press the key.
e. Press the [End] key and repeat this until exiting the Management menu.
Quick Filming ON
Delay Time (sec)
3–25 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 11 2205009
D Disk Initialize
D File Allocation
File Allocation creates Variable files for the system hard disk, such as image directory, calibration files,
and image data files. It consists of two phases as follows:
3–26 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 11 2205009
D FAIL LED: If lit, this LED indicates that a bus error has occurred. (Abnormal state)
D HALT LED: If lit, this LED indicates that the System Operating System has terminated the current
command due to some error. (Abnormal state)
The following two kinds of errors could occur during the execution of the ATINST program:
D An error of the ATINST program itself (Program abort due to a program bug or hardware failure)
If either of the above errors occur, the following menu will be displayed:
Retry of ATINST
After terminating the ATINST program by selecting ‘1’ in the menu, retry ATINST.
The ATINST program will start as ‘Reloading Mode’, or ‘First Loading Mode’, depending on when the program failed.
After the ATINST program is successfully finished, restore the following data files if they were saved as backup, AF-
TER performing step 1 (Restore the InSite software) and step 2 (Restore the Custom Parameters) in ‘Post–ATINST
Procedure’; since the System Parameters HAVE to be loaded AFTER InSite and Elink (if system has) software are
restored.
Scan Parameter
Patients Information
Series Description
Calibration Files
Station Number Information
Voice Information
System Parameters (including Custom Parameters)
3–27 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 11 2205009
3-4-3 Preparation
CAUTION
Save existing patient data any time you perform an ATINST procedure or you will lose this
data.
The saving procedure is described in ‘Pre–ATINST Procedure.’
1. Set the SW2 and SW4 of the “BOOT” switches on the CPW Board (Slot #1) to the ON position. See Illustration
3–2.
ON
1 2 3 4
DNP4
OFF
3. Pull the reset switch located under the right side of the Operator Console.
The system is reset and “>” will appear on the monitor.
4. Set both the A side and B side of the MOD (contains system software) “not write protected.”
3–28 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 11 2205009
FHS ERROR C2
> ATINST ‘CR’
HD and BOOT PARITY No. is same : 04.00
Install as RELOADING MODE ? <0:OK, 1:Exit>=0 ‘CR’
(This message appears if the hard disk already contains the system
software whose Version number is equal to the one of the BOOT
floppy. The ATINST program will reload the system software from
MOD to hard disk by entering 0.)
Delete IMAGE & RAW Data ? <0:OK, 1:Exit> =0 ‘CR’
** INSTALLATION START **
Please Insert <<MQ00_[x] >> MOD and Press <ENTR>
The following operations are automatically carried out in the order described:
D Disk Initialize
D File Allocation
It will take approximately 40 minutes for the ATINST program to complete all its operations.
For systems with two system hard disk drives, it will take approximately 80 minutes. (Qj, Qj–Base, Qj–Hlt)
It will take approximately 80 minutes for the ATINST program to complete all its operations. (Zj–Adv)
During operations, the BUSY indicator on the MOD drive repeats On & OFF, and you can know what operations
are being performed by looking at the CRT monitor screen; however, ignore all the inquires that appear, except
the following:
1. PROTECTED MOD
If a Sony MOD drive is used, and if it is not the ‘First Loading Mode’, the following message will appear after
inserting the MOD.
** PROTECTED MOD
After insert Unprotected MOD and Press <ENTR>
This is because that the system software can not write on a MOD media of single density (644 MB) with the
Sony MOD drive.
In this case, proceed to Section 3-4-6, ‘For Sony MOD.’
3–29 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 11 2205009
To resume ATINST, you have only to press the [End] key ONCE.
Model Name
Select the appropriate Model Name. (for ProSpeed S, ProSpeed SX, Lemage, ProSpeed Advantage,
ProSeed SA, ProSeed SA Libra)
3–30 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 11 2205009
1 2 3 4
OFF
DNP4
OFF
8. Pull the system reset switch located under the right side of the Operator Console.
9. Check that the system starts up normally and that the correct version indication and logo appear on the CRT
monitor.
3–31 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 11 2205009
1 2 3 4
OFF
DNP4
OFF
3. Pull the system reset switch located under the right side of the Operator Console.
Note
Prepare approximately 10 blank floppy disks for data storage.
4. Perform steps 4 (Custom Parameters) and 5 (data files) in Section 3-3, ‘Pre–ATINST Procedure’, since we will
execute the ATINST in the ‘First Loading Mode.’
6. Enter $$.
10. Restore the data files (Scan Parameter, Patients Information, ...).
3–32 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 11 2205009
The InSite program must be restored since Version–up procedures using ATINST delete it.
b. Insert the BOOT floppy disk. (This step is not required for systems installed in Japan)
>
3–33 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 11 2205009
Use the CHANGE PARAMETERS function of the CUSTOM PARAMETERS program for the following purposes:
to change the custom parameters;
to load additional option software;
to change the number of raw data files.
Note
The custom parameter data will be restored automatically in the ATINST program (Data File Loading
Function). However, regarding the following, it is necessary to manually restore due to a software
bug:
Note
As for the Recon Kernel Type parameter, select Normal for systems with Xe Detector (it seems that
you can select Enhanced, but regardless of the selection, the system will still use the Normal type
recon kernel); for systems with Hilight Detector, Normal or Enhanced can be selected.
(For systems with Version 6.xx or 4.xx (ProSeed SA, Libra) and with a Xe Detector, this parameter
does not appear)
Note
As for the Collimator Type parameter, select ‘7mm Thickness AVAILABLE Type’ for systems where
either of the following conditions is met.
1. An ‘R2’ mark is attached on the Gantry rear base and the software version is at 6.xx or at 4.xx
(ProSeed SA, Libra).
2. ‘7 mm Thickness Collimator’ Option is installed.
Note
For the Network Option parameter, you are requested to select one of ‘Unavailable’, ‘YMS E–Link’,
or ‘YMS Fast E–Link’; therefore, you have to know the correct setting for the system, prior to entering
the CHANGE PARAMETERS menu. To know which of ‘YMS E–Link’ or ‘YMS Fast E–Link’ the sys-
tem is equipped with, check it in the DISPLAY PARAMETERS mode, or check that which of ETC
Board (for YMS E–Link) and ETC2 Board (for YMS Fast E–Link) is installed on the OC.
3–34 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 11 2205009
** CUSTOM PARAMETERS **
1. DISPLAY PARAMETERS
2. CHANGE PARAMETERS
Select No. = 2 ‘CR’
f. Verify or change the Custom Parameters which are displayed one by one as listed in Table 3–20, with the
previously recorded Tables 3–17 and 3–18.
Note
If a parameter is not required to be changed, press the [END] key to proceed to the next item.
All the option software to be installed should be loaded during this step.
At the end of this ‘Change Parameters’ menu, you are instructed to pull the reset switch.
h. Pull the reset switch located under the right side of the Operator Console.
Verify that the system starts up normally and that the correct version indication and logo appear on the CRT
monitor.
i. Using ‘DISPLAY PARAMETERS’ of the CUSTOM PARAMETERS menu, check if all the optional software
you installed are displayed as “Available.”
3–35 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 11 2205009
Current Model Name This parameter does not appear if the hard disk
contains images.
Optical Disk
Magnetic Tape
Frequency
MACHINE No. This parameter does not appear if the hard disk
contains images.
HOSPITAL NAME This parameter does not appear if the hard disk
contains images.
Collimator Type
THICKNESS 2/3 mm SET The default setting is `3'. If the customer's system
is adjusted to 2mm, `2' must be selected.
See Note 7.
This parameter does not appear if `Collimator
Type' is set to `7mm Thickness AVAILABLE Type.'
SET NO. RAW DATA This parameter does not appear if the hard disk
contains images.
See Caution and Note 6.
3–36 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 11 2205009
Note 2 This parameter will not appear if Gantry" is selected as the Kind of Base in DIRECTION
OF SCOUT IMAGE (Lateral).
Note 3 FUNCTION The following menu with the current settings is displayed:
** FUNCTION MENU **
CURRENT
1. AUTO STORE DEVICE OPTION = (OD or MT or MOD)
2. OD SPECIAL STORE OPTION = (NON or AUTO)
3. AUTO DELETE OPTION = (ON or OFF)
4. MIN VACANT IMAGE SPACE FOR SCAN = (XXXX)
5. FD PACKING OPTION = (DPCM or ORIGINAL)
6. MT PACKING OPTION = (DPCM or ORIGINAL)
7. OD PACKING OPTION = (DPCM or ORIGINAL)
8. FD SYSTEM No. = (XXXXXX)
9. MT SYSTEM No. = (XXXXXX)
10. OD SYSTEM No. = (XXXXXX)
11. MT IMAGE MAX No. = (XXX)
Select No. =
2. OD SPECIAL STORE OPTION AUTO: Images are automatically stored to the OD (AUTO STORE
DEVICE), when the number of vacant images on the system HD is decreased to below the number
(4. MIN VACANT IMAGE SPACE FOR SCAN).
3. AUTO DELETE OPTION ON: When the AUTO STORE function is enabled, studies which are
already stored (and not the retrieved ones) are automatically deleted from the HD to reserve the
specified image space (4. MIN VACANT IMAGE SPACE FOR SCAN). Delete operation is performed
on files starting from the study having the smallest study No. on the HD through to the study which
will allow the specified space (4. MIN VACANT IMAGE SPACE FOR SCAN) to be reserved.
4. MIN VACANT IMAGE SPACE FOR SCAN During an image retrieve operation, if the number
of vacant images on the HD becomes equal to this specified number, the retrieve operation is termiĆ
nated and a DK FULL" message is displayed on the Status screen.
NOTICE
The Tube parameter allows selection of the proper reconstruction and cooling algorithms for
use with the appropriate Tube. DO NOT select the wrong Tube, as system and/or Tube dam-
age will result.
Note 5 Enter the appropriate No. of the option software you want to install. Repeat this proceĆ
dure until all the software options you want to install are loaded. The software options loaded will
be indicated by Installed."
3–37 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 11 2205009
CAUTION
For systems with Hilight Detector, and if the ATINST was in the “First Loading Mode” (this is
performed when replacing the hard disk drive, etc.), RE–ENTER the number, even if the de-
fault number seems appropriate; and this procedure must be performed AFTER the HILT op-
tion software has been installed. Failure to this instruction will cause image data corruption.
Note 6 If the number of raw data files needs to be changed, select `0' and press the [ENTER]
key. Then enter the appropriate value. Up to 300 can be entered.
Enter 50 for the inquiry on Program Area."
It will take several minutes to complete data file reservation. If there are many bad sectors on the
disk, a `Too Many Bad Sector' message and the new appropriate (maximum possible) number of
Raw Data to be entered appear on the monitor. Enter the new number again if it is considered as
appropriate.
Note 7 If Thickness is changed, the air and phantom calibrations and CT number adjustment
must be performed; if the system has the Hilight Detector, the Hilight Calibration also must be perĆ
formed. However, for the GEMS-E region, read the following Notice:
NOTICE
3–38 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 11 2205009
...
If you archived patient image data prior to performing the ATINST procedure, restore it to the disk. Use the Image
File Management function to restore it. (Refer to Operator Manual)
If you saved raw data files prior to performing the ATINST procedure, restore them to the disk. Use the
SUPPORT program to retrieve them.
a. Insert the BOOT floppy into the disk drive. (This step is not required for systems installed in Japan)
3–39 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 11 2205009
(The MOD raw data files will have the following ID No. on the disk.
0:0.XX.1;1 (STA:STD.SER.ACQ;IMG - digits other than the series No. are fixed)
Designate series Nos. for the MOD raw data files.
Press the [END] key, and the following message appears:)
3–40 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 11 2205009
Option Misc
a. Enter the Miscellaneous Application menu by pressing Soft or Appli key.
c. Press the [Enter] key and repeat this until the CBF Processing menu appears.
This step is required for systems with a Functional Image option only.
Option Image
a. Press Soft or Compst key.
3–41 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 11 2205009
The version 6.00 system software should be installed prior to this procedure.
b. Insert the BOOT floppy disk. (This step is not required for systems installed in Japan)
>
>RU ‘CR’
SELECT No.= 2S
WHAT FILE NAME ?
SOURCE: V602 ‘CR’
OBJECT: SYS ‘CR’
DO YOU WANT SELECT ? N ‘CR’
IF FILE EXISTS, OVERWRITE OK? Y ‘CR’
COPY START !
.
TCOP: TERMINATED
>
3–42 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 11 2205009
> RT ‘CR’
1.CALIBRATION
** CUSTOM PARAMETERS **
1.DISPLAY PARAMETERS
2.CHANGE PARAMETERS
p. Press the [End] button to skip a menu and repeat this until the following message appears.
***********************************
* PULL THE SYSTEM RESET BUTTON !! *
***********************************
Check that the system starts up normally and that the “VER. 06.02” and logo appear on the CRT monitor.
3–43 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 11 2205009
The version 6.02 system software should be installed prior to this procedure.
b. Insert the BOOT floppy disk. (This step is not required for systems installed in Japan)
>
>RU ‘CR’
SELECT No.= 2S
WHAT FILE NAME ?
SOURCE: V603 ‘CR’
OBJECT: SYS ‘CR’
DO YOU WANT SELECT ? N ‘CR’
IF FILE EXISTS, OVERWRITE OK? Y ‘CR’
COPY START !
.
TCOP: TERMINATED
>
3–44 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 11 2205009
> RT ‘CR’
Check that the system starts up normally and that the “VER. 06.03” and logo appear on the CRT monitor.
3–45 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 11 2205009
The version 4.00 system software should be installed prior to this procedure.
b. Insert the BOOT floppy disk. (This step is not required for systems installed in Japan)
>
d. Remove the BOOT disk.
SELECT No.= 2S
WHAT FILE NAME ?
SOURCE: V401 ‘CR’
OBJECT: SYS ‘CR’
DO YOU WANT SELECT ? N ‘CR’
IF FILE EXISTS, OVERWRITE OK? Y ‘CR’
COPY START!
.
.
TCOPY TERMINATED
> RT ‘CR’
h. Pull the system reset switch located under the right side of the Operator Console.
Check that the system starts up normally and that the “VER. 04.01” and logo appear on the CRT monitor.
3–46 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 11 2205009
Option Misc
1. Enter the Miscellaneous Application menu by pressing Soft or Appli key.
3. Press the [Enter] key and repeat this until the CBF Processing menu appears.
3–47 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 11 2205009
Option Image
1. Press Soft or Compst key.
4. Select FILE.
8. Repeat step 6.
9. Continue the above until user parameters for all of Menu 1 through Menu 8 are recorded.
10. Press the [End] key and repeat this until exiting this program.
Menu 1 [ ]
Appearance Cal. Mask
Iterations Smooth Matrix
Recirc. Cut CT# Lower
Reciro. Mask CT# Upper
Menu 2 [ ]
Appearance Cal. Mask
Iterations Smooth Matrix
Recirc. Cut CT# Lower
Reciro. Mask CT# Upper
Menu 3 [ ]
Appearance Cal. Mask
Iterations Smooth Matrix
Recirc. Cut CT# Lower
Reciro. Mask CT# Upper
3–48 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 11 2205009
Menu 4 [ ]
Appearance Cal. Mask
Iterations Smooth Matrix
Recirc. Cut CT# Lower
Reciro. Mask CT# Upper
Menu 5 [ ]
Appearance Cal. Mask
Iterations Smooth Matrix
Recirc. Cut CT# Lower
Reciro. Mask CT# Upper
Menu 6 [ ]
Appearance Cal. Mask
Iterations Smooth Matrix
Recirc. Cut CT# Lower
Reciro. Mask CT# Upper
Menu 7 [ ]
Appearance Cal. Mask
Iterations Smooth Matrix
Recirc. Cut CT# Lower
Reciro. Mask CT# Upper
Menu 8 [ ]
Appearance Cal. Mask
Iterations Smooth Matrix
Recirc. Cut CT# Lower
Reciro. Mask CT# Upper
3–49 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
blank 2205009
3–50 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 11 2205009
4-1 GENERAL
This ‘Software Loading’ describes the Load–From–Cold (LFC) procedure for a full system software loading from an
MOD (Magnetic Optical Disk) and some FDs (Floppy Disks) to the system hard disk.
Perform this LFC procedure when:
D A failure or crash corrupts the hard disk so the existing software can not be used. A Load–From–Cold
rebuilds the operating software on the disk and restores normal system operation.
D CT Engineering has revised the current software to incorporate improvements or correct existing prob-
lems. Replace the old software by performing a Load–From–Cold with the new software.
4–1 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 11 2205009
START
ATINST program
DATA FILE SAVE
DISK INITIALIZE
SOFTWARE LOADING
(STANDARD, OPTION)
FILE ALLOCATION
DATA FILE LOADING
RESET FMP1 DIP SW
END
4–2 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 11 2205009
D System Software Disks listed in Table 4–1 ∼ 4–7 (see the table below)
The table lists the Volume Names of an MOD which contains the system software, of some floppy disks
which contain application software, and of other floppy disks required for the system.
D Option Software Disks (which the customer already owns) listed in Table 4–8
This table shows all the currently available option software.
Normally, option software disks and also some application software disks included in the system software
(“PG00” disk, “PZ00” disk, ...) are not required in a version up or a re–loading procedure. However, these
disks are required if the ATINST program (described later) failed, or when the system hard disk drive is
replaced with a new one.
However, in V/R 4.01 Upgrade, V/R 5.00 Upgrade, and V/R 5.02 Upgrade, the following option disks are
required (if the customer owns).
Co–ordinate (CD00)
Paging (PG00)
Real–time Reformation (RTRF)
Curved Reformation (CVRF)
3–D Imaging (DIM3)
Calculator (CALC)
MIP (MIP0)
YMS Link (YLNK)
YMS E–Link (ELNK)
D Option Software Disks (which the customer bought and have not yet installed to the system) listed in Table
4–8
4–3 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 11 2205009
4–4 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 11 2205009
4–5 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 11 2205009
BOOT 0 (P9180PH)
ST00 0 (P9180PL)
MZ00 0 System Software V/R 4.00 (P9180PF) (MOD)
V411 0 System Software (2118916) ( for Version 4.01)
V412 0
V500 0 System Software (2133783) (for Version 5.00)
V501 0 System Software (2144870) (for Version 5.01)
V502 0 System Software (2167519) (for Version 5.02)
V503 0 System Software (2236433) (for Version 5.03)
PG00 0 Paging (P9180QA)
PZ00 0 Pan-zoom
RTRF 2 Real-time Reformation
CVRF 1 Curved Reformation
DIND 0 Display in Display
HSRC 0 High Speed Recon (P9180QW)
Latest System Software: V/R 5.03
4–6 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 11 2205009
4–7 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 11 2205009
BOOT 0 (P9180PH)
BSE0 0 (P9180PK)
MZ00 0 System Software V/R 4.00 (P9180PE) (MOD)
V411 0 System Software (2118916) ( for Version 4.01)
V412 0
V500 0 System Software (2133783) (for Version 5.00)
V501 0 System Software (2144870) (for Version 5.01)
V502 0 System Software (2167519) (for Version 5.02)
V503 0 System Software (2236433) (for Version 5.03)
PG00 0 Paging (P9180QA)
PZ00 0 Pan-zoom
DIND 0 Display in Display
HLCL 0 Helical Scan (P9180QP)
CINE 0 Cine Retro Recon (P9180QU)
Latest System Software: V/R 5.03
4–8 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 11 2205009
4–9 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 11 2205009
BOOT 0 (P9180PH)
ST00 0 (P9180PL)
MZ00 0 System Software V/R 4.00 (P9180PE) (MOD)
V411 0 System Software (2118916) ( for Version 4.01)
V412 0
V500 0 System Software (2133783) (for Version 5.00)
V501 0 System Software (2144870) (for Version 5.01)
V502 0 System Software (2167519) (for Version 5.02)
V503 0 System Software (2236433) (for Version 5.03)
PG00 0 Paging (P9180QA)
PZ00 0 Pan-zoom
RTRF 2 Real-time Reformation
CVRF 1 Curved Reformation
DIND 0 Display in Display
DIM3 1 3-D Imaging
MIP0 0 MIP (P9180QX)
HLCL 0 Helical Scan (P9180QP)
CINE 0 Cine Retro Recon (P9180QU)
HSRC 0 High Speed Recon (P9180QW)
Latest System Software: V/R 5.03
4–10 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 11 2205009
CD00 1 Co-ordinate
CCIO 1 CCI
CB00 1 CBF
BMD0 0 BMD
ELNK 0 Ethernet
MRS0 0 MARS
MIP0 0 MIP
4–11 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 11 2205009
About MOD Cartridge Type. For data save purpose, use MOD cartridges compatible with the Hitachi MOD drive in
Japanese sites; or use MOD cartridges compatible with the Pioneer MOD drive in except Japanese sites.
About Floppy Disks. For data save purpose, use blank floppy disks, or initialize them with a volume name ‘FPAR’,
even though they are already initialized with this volume name, otherwise, the previously saved data on the floppy
disks will persist (not be overwritten)!
Refer to Section 3-1-2, ‘Floppy Disk Initialize’, of Tab ‘Introduction.’
Archive existing patient image data to MOD, floppy disks, tape, or optical disk before proceeding, or you will lose
this data. Use the Image File Management function to save data. (Refer to Operator Manual)
For any system on which ELINK or YLINK is installed, save Communication Data on a floppy disk, otherwise, the
data will be deleted during LFC.
h. Insert a formatted floppy disk into the disk drive, then save Communication Data onto it.
Note
For systems with V/R 5.xx or later system software, the ‘Change Interval time’ item disappears from
the menu. Therefore, this item MUST be set to the (appropriate) default value (20 seconds) prior
to starting version–up procedure.
‘Interval time’ means a period of time between events that the system checks the list of transmission.
If this value is set to a few seconds (e.g. 1 sec., 2 sec....), system error might occur.
4–12 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 11 2205009
CAUTION
If you intend to save raw data onto MOD, be sure to start with a blank cartridge , or one that
does not need to be preserved. This procedure will erase all previous data contained on the
MOD cartridge.
Archive raw data files to MOD, floppy disks, tape, or optical disk if the customer wants to save them. Otherwise,
these files are destroyed during LFC.
Before proceeding to the following save procedure, write down all the raw data file ID numbers to be saved.
If you use floppy disks, format them before proceeding.
a. Insert the BOOT floppy into the disk drive. (This step is not required for systems installed in Japan)
i. Enter all the raw data file ID Numbers. Then press the [END] key.
...
4–13 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 11 2205009
a. Insert the BOOT floppy into the disk drive. (This step is not required for systems installed in Japan)
** CUSTOM PARAMETERS **
1. DISPLAY PARAMETERS
2. CHANGE PARAMETERS
Select No. = 1 ‘CR’
f. Custom Parameters will be displayed one by one, as listed in Table 4–9. Use this table to record the parame-
ters.
Note
Usually the Custom Parameters data does not need to be recorded, since the data will be saved auto-
matically in the ATINST program (Data File Save Function). However, in this LFC (Re–loading of
the current system software), you have to record at least the items specified at steps i and ii.
In addition, if you have all the records, they can serve as backup in the event there is trouble with
the ATINST program.
ii. Record all the option software that are currently installed on the system.
(This step is required to confirm later that all the option software are correctly retrieved, and to re–install
some option software due to some currently existing software problem)
4–14 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 11 2205009
Frequency 50 . 60
MACHINE No.
HOSPITAL NAME
Disk Size
Apex Lateral
Thickness 2/3 mm 2 mm . 3 mm
4–15 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 11 2205009
OPTION
4–16 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 11 2205009
Note
Usually the sets of information above do not need to be saved since they will be automatically saved
in the ATINST program. However, if you save them, they can serve as backup in the event there is
trouble with the ATINST program.
Save the following data files on floppy disks, following the steps below.
1. Scan Parameter
2. Patients Information
3. Series Description
4. Calibration Files
5. Station Number Information
6. Voice Information
7. System Parameters (including Custom Parameters)
...
c. Insert the BOOT floppy into the disk drive. (This step is not required for systems installed in Japan)
f. Select SYSTEM PARAMETERS DUMP & LOAD from the Maintenance menu to save System Parameters
(including Custom Parameters).
4–17 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 11 2205009
For systems with a CBF option, record the user setting parameters, otherwise, the data will be deleted during
the ATINST procedure.
For systems with a Functional Image option, record the user setting parameters, otherwise, the data will be de-
leted during the ATINST procedure.
For systems where the user sets `Quick Filming ON', the Quick Filming Delay Time must be recorded, other-
wise, the data will be deleted during the ATINST procedure.
Mgt
a. Press the key.
e. Press the [End] key and repeat this until exiting the Management menu.
Quick Filming ON
Delay Time (sec)
4–18 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 11 2205009
D Disk Initialize
D File Allocation
File Allocation creates Variable files for the system hard disk, such as image directory, calibration files,
and image data files. It consists of two phases as follows:
4–19 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 11 2205009
D DMA LED: When this is blinking it indicates that the program is accessing the hard disk or floppy disk.
It shows that the system is operating normally.
D BER LED: If lit, this LED indicates that a bus error has occurred. (Abnormal state)
D HLT LED: If lit, this LED indicates that the System Operating System has terminated the current
command due to some error. (Abnormal state)
The following two kinds of errors could occur during the execution of the ATINST program:
D An error of the ATINST program itself (Program abort due to a program bug or hardware failure)
If either of the above errors occurs, the following menu will be displayed:
Use Menu 3 and Menu 4 to compare the Log File and Text File.
4–20 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 11 2205009
Retry of ATINST
After terminating the ATINST program by selecting ‘1’ in the menu, retry ATINST.
The ATINST program will start as ‘Reloading Mode’, or ‘First Loading Mode’, depending on when the program failed.
After the ATINST program is successfully finished, perform the following in the order described; since the System
Parameters HAVE to be loaded AFTER InSite and Elink or Ylink (if system has) software are restored.
4–21 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 11 2205009
4-4-3 Preparation
CAUTION
Save existing patient data any time you perform the ATINST procedure or you will lose this
data.
(The saving procedure is described in the ‘Pre–ATINST Procedure.’)
1. Set the SW2 and SW4 of the ‘BOOT’ switches on the FMP1 Board (Slot #30) to the ON position. See Illustration
4–2.
ON
ËË
ËË
ËË
1
2
ËË
ËË
3
ËË
4
ËË
2. Insert the BOOT floppy into the disk drive. (The BOOT floppy disk should not be removed until the ATINST
program finishes)
3. Pull the reset switch located under the right side of the Operator Console.
The system is reset and ‘F>’ will appear on the monitor.
4. Set both the A side and B side of the MOD (contains system software) “not write protected.”
4–22 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 11 2205009
FHS ERROR C2
> ATINST ‘CR’
HD and BOOT PARITY No. is same : 04.00
Install as RELOADING MODE ? <0:OK, 1:Exit>=0 ‘CR’
(This message appears if the hard disk already contains the system
software whose Version number is equal to the one of the BOOT
floppy. The ATINST program will reload the system software from
MOD to hard disk by entering 0.)
Delete IMAGE & RAW Data ? <0:OK, 1:Exit> =0 ‘CR’
** INSTALLATION START **
Please Insert <<MZ00_[x] >> MOD and Press <ENTR>
The following operations are automatically carried out in the order described:
D Disk Initialize
D File Allocation
It will take approximately 40 minutes for the ATINST program to complete all its operations.
During operations, the BUSY indicator on the MOD drive repeats On & OFF, and you can know what operations
are being performed by looking at the CRT monitor screen; however, ignore all the inquires that appear, except
the following:
1. PROTECTED MOD
If a Sony MOD drive is used, and if it is not the ‘First Loading Mode’, the following message will appear after
inserting the MOD.
** PROTECTED MOD
After insert Unprotected MOD and Press <ENTR>
This is because that the system software can not write on a MOD media of single density (644 MB) with the
Sony MOD drive.
In this case, proceed to Section 4-4-6, ‘For Sony MOD.’
4–23 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 11 2205009
Model Name
Select the appropriate Model Name.
You will be instructed to insert the ‘ST00’ or ‘BSE0’ floppy.
VINIT
Respond as follows, to inquiries.
What Device Name ?
Main Disk HD00 ‘CR’
Auxiliary Disk HD01 ‘CR’
Do You Want to Use LOT Info. of ‘FD’ (Y/N/Q) ?
Main Disk N ‘CR’
Auxiliary Disk N ‘CR’
Do You Want to Format Disk ? N ‘CR’
Do You Want to Validate Sector ? N ‘CR’
Do You Want to Append Bad Sectors to LOT ? N ‘CR’
Do You Want to Lock Out Disk ? N ‘CR’
Do You Want to Initialize Disk ? Y ‘CR’
Volume Name ?
Main Disk SYS ‘CR’
Auxiliary Disk SYS1 ‘CR’
User No. ? 0 ‘CR’
Description ? PROSPEED 0400 940701 ‘CR’
(Example)
Do You Want System Dump Area ?
Main Disk Y ‘CR’
What File Name ? DUMP ‘CR’
How Many Sectors ? 4 ‘CR’
Auxiliary Disk N ‘CR’
Do You Want IPLT Loading ?
Main Disk Y ‘CR’
What Volume Involving IPLT ? BOOT ‘CR’
Auxiliary Disk N ‘CR’
Do You Want to Dump LOT to ‘FD’ ? N ‘CR’
4–24 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 11 2205009
OFF
ËË
1
ËË
ËË
ËË
2
ËË
3
4
ËË
7. Remove the MOD from the drive.
8. Pull the reset switch located under the right side of the Operator Console.
Check that the system starts up normally and that the correct version indication and logo appear on the CRT
monitor.
4–25 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 11 2205009
OFF
ËË
1
ËË
ËË
ËË
2
ËË
3
4
ËË
2. Remove the MOD from the drive.
3. Pull the system reset switch located under the right side of the Operator Console.
Note
Prepare approximately 10 blank floppy disks for data storage.
4. Perform steps 4 (Custom Parameters) and 5 (data files) in Section 4-3, ‘Pre–ATINST Procedure’, since we will
execute the ATINST in the ‘First Loading Mode.’
6. Enter $$.
10. Restore the data files (Scan Parameter, Patients Information, ...).
4–26 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 11 2205009
b. Insert the BOOT floppy disk. (This step is not required for systems installed in Japan)
4–27 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 11 2205009
F> RU ‘CR’
.
.
SELECT NO. =2S ‘CR’
WHAT FILE NAME ?
SOURCE: V412 ‘CR’
OBJECT: SYS ‘CR’
DO YOU WANT A SELECT ? Y/N/Q N ‘CR’
IF FILE EXISTS, OVERWRITE OK ? Y/N/Q Y ‘CR’
COPY START !
.
.
F> RT
Note
Do not reset the system yet.
If the system is unintentionally reset and if ELINK Option is installed on the system, a system hang–up
may occur. If this happens, the system must be started up using a Special Procedure, written in Sec-
tion 4-6.
The InSite program must be restored since Version–up procedures using ATINST delete it.
b. Insert the BOOT floppy disk. (This step is not required for systems installed in Japan)
F>
4–28 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 11 2205009
g. Pull the system reset switch located under the right side of the Operator Console.
Check that the system starts up normally and that the correct version indication and logo appear on the CRT
monitor.
Use the CHANGE PARAMETERS function of the CUSTOM PARAMETERS program for the following purposes:
to change the custom parameters;
to load additional option software;
to change the number of raw data files.
c. Enter $$ as follows;
** CUSTOM PARAMETERS **
1. DISPLAY PARAMETERS
2. CHANGE PARAMETERS
Select No. = 2 ‘CR’
4–29 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 11 2205009
Note
If a parameter is not required to be changed, press the [END] key to proceed to the next item.
The custom parameter data will be restored automatically in the ATINST program (Data File Loading Func-
tion). However, at least the following steps (i, ii, iii) must be performed due to software problems.
i. Set the following parameters to as recorded in step 4 in Section 4-3, ‘Pre–ATINST Procedure.’
NOTICE
The Tube parameter allows selection of the proper reconstruction and cooling algorithms for
use with the appropriate Tube. DO NOT select the wrong Tube, as system and/or Tube dam-
age will result.
NOTICE
Perform the following step; do not skip, since the V/R 4.01 system software requires the re–
entering of the following parameter.
ii. Re–set the following parameter, by entering ‘1’ or ‘2’, according to the table below.
4–30 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 11 2205009
NOTICE
Perform the following step; do not skip, since the V/R 4.01 system software requires the re–
installation of the following option software.
iii. Re–install any of the following option software, that the customer owns.
(Although the options will be indicated as ‘Installed’, still, however, the re–installation must be performed.
To know whether your system includes any of these, see ‘OPTION’ in Table 4–9.
Co–ordinate (CD00)
Paging (PG00)
Real–time Reformation (RTRF)
Curved Reformation (CVRF)
3–D Imaging (DIM3)
Calculator (CALC)
MIP (MIP0)
YMS Link (YLNK)
YMS E–Link (ELNK)
All the option software to be installed should be loaded during this step.
At the end of this ‘Change Parameters’ menu, you are instructed to pull the reset switch.
h. Pull the reset switch located under the right side of the Operator Console.
i. Using ‘DISPLAY PARAMETERS’ of the CUSTOM PARAMETERS menu, check if all the optional software
you installed are displayed as “Available.”
4–31 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 11 2205009
Current Model Name This parameter does not appear if the hard disk
contains images.
Optical Disk
Magnetic Tape
Frequency
MACHINE No. This parameter does not appear if the hard disk
contains images.
HOSPITAL NAME This parameter does not appear if the hard disk
contains images.
INSITE
No. of Raw Data Files This parameter does not appear if the hard disk
contains images.
See Note 6.
4–32 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 11 2205009
Note 2 When `DIRECTION OF SCOUT IMAGE' is saved and loaded automatically using ATINST,
the default data will be rewritten against the original data. The indication visible in the display might
be the same as the original one, but data has been rewritten. So, enter the original data recorded
at step 4 in Section 4-3, ‘Pre–ATINST Procedure’ to New Data.
Note 3 When `Gantry Direction from Operator' is saved and loaded automatically using ATINST,
the default data will be rewritten against the original data. The indication visible in the display might
be the same as the original one, but data has been rewritten. So, enter the original data recorded
at step 4 in Section 4-3, ‘Pre–ATINST Procedure’ to New Data.
This parameter will not appear if Gantry" is selected as the Kind of Base in DIRECTION OF SCOUT
IMAGE (Lateral).
NOTICE
The Tube parameter allows selection of the proper reconstruction and cooling algorithms for
use with the appropriate Tube. DO NOT select the wrong Tube, as system and/or Tube dam-
age will result.
Note 5 Enter the appropriate No. of the option software you want to install. Repeat this proceĆ
dure until all the software options you want to install are loaded. The software options loaded will
be indicated by Installed."
Note 6 If the number of raw data files needs to be changed, select `0' and press the [ENTER]
key. Then enter the appropriate value. It will take several minutes to complete the data file reservaĆ
tion. If there are many bad sectors on the disk, a `Too Many Bad Sector' message and the new
appropriate (maximum possible) number of raw data to be entered appear on the monitor. Enter
the new number again if it is considered as appropriate.
4–33 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 11 2205009
If you archived Patient Image Data before this LFC procedure, restore it to the disk. Use the Image File
Management function to retrieve it (Refer to Operator Manual).
5. Set ‘ACQ. REVERSE’ data (ON or OFF) in Set Display to the data as recorded at step 6 in Section 4-3, ‘Pre–
ATINST Procedure.’
6. Set ‘Preset Voice Timing’ in the Maintenance menu to the data recorded at step 7 in Section 4-3, ‘Pre–ATINST
Procedure.’
If you saved raw data files before this LFC procedure, restore them to the disk. Use the Support program to
retrieve them.
a. Insert the BOOT floppy into the disk drive. (This step is not required for systems installed in Japan)
4–34 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 11 2205009
(The MOD raw data files will have the following ID No. on the disk.
0:0.XX.1;1 (STA:STD.SER.ACQ;IMG - digits other than the series No. are fixed)
Designate series Nos. for the MOD raw data files.
Press the [END] key, and the following message appears:)
8. Pull the reset switch located under the right side of the Operator Console.
Check that the system starts up normally and that ‘VER 04.01’ amd the system logo appear on the CRT monitor.
4–35 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 11 2205009
b. Insert the BOOT floppy disk. (This step is not required for systems installed in Japan)
D>
d. Remove the BOOT disk.
Insert firstly the Version 5.00 system software floppy disk (1/5).
D>RU ‘CR’
.
.
SELECT No.= 2S
WHAT FILE NAME ?
SOURCE: V500 ‘CR’
OBJECT: SYS ‘CR’
DO YOU WANT SELECT ? N ‘CR’
IF FILE EXISTS, OVERWRITE OK? Y ‘CR’
COPY START !
.
.
g. Repeat step f until all of the Version 5.00 system software (5 floppies) are loaded.
D> RT ‘CR’
10. Press the [End] key to exit the Management menu.
4–36 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 11 2205009
Co–ordinate (CD00)
Paging (PG00)
Real–time Reformation (RTRF)
Curved Reformation (CVRF)
3–D Imaging (DIM3)
Calculator (CALC)
MIP (MIP0)
YMS Link (YLNK)
YMS E–Link (ELNK)
** CUSTOM PARAMETERS **
1. DISPLAY PARAMETERS
2. CHANGE PARAMETERS
Select No. = 2 ‘CR’
4–37 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 11 2205009
NOTICE
Perform the following step; do not skip, since the V/R 5.00 system software requires the re–
installation of the stated option software.
(The same procedure was performed during the V/R 4.01 upgrade, and again, this must be
performed.)
g. Insert the option floppy disk into the drive and enter the number of the option software.
(Although the options will be indicated as ‘Installed’, still, however, the re–installation must be performed.)
h. Repeat the above step until all of the options are installed.
i. Press the [End] key to skip a menu and repeat this until the following message appears.
Check that the system starts up normally and that the “VER. 05.00” and logo appear on the CRT monitor.
This step is required for systems with an ELINK or YLINK option only.
Use the floppy disk saved at step 2 in Section 4-3, ‘Pre–ATINST Procedure.’
...
13. Restore System Parameters using the Maintenance menu (System Parameter Dump & Load).
Check that the system starts up normally and that the logo appears on the CRT monitor with no error.
4–38 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 11 2205009
b. Insert the BOOT floppy disk. (This step is not required for systems installed in Japan)
F>
F>RU ‘CR’
SELECT No.= 2S
WHAT FILE NAME ?
SOURCE: V501 ‘CR’
OBJECT: SYS ‘CR’
DO YOU WANT SELECT ? N ‘CR’
IF FILE EXISTS, OVERWRITE OK? Y ‘CR’
COPY START !
.
.
F> RT ‘CR’
4–39 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 11 2205009
** CUSTOM PARAMETERS **
1. DISPLAY PARAMETERS
2. CHANGE PARAMETERS
Select No. = 2 ‘CR’
f. Press the [End] key to skip a menu and repeat this until the Tube Installation menu is displayed.
NOTICE
The Tube parameter allows selection of the proper reconstruction and cooling algorithms for
use with the appropriate Tube. DO NOT select the wrong Tube, as system and/or Tube dam-
age will result.
NOTICE
Perform the following step; do not skip, since the V/R 5.01 system software requires the re–
entering of the following parameter.
(The same procedure was performed during the V/R 4.01 upgrade, and again, this must be
performed)
4–40 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 11 2205009
Check that the system starts up normally and that the ‘VER. 05.01’ and logo appear on the CRT monitor.
4–41 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 11 2205009
b. Insert the BOOT floppy disk. (This step is not required for systems installed in Japan)
F>
F>RU ‘CR’
SELECT No.= 2S
WHAT FILE NAME ?
SOURCE: V502 ‘CR’
OBJECT: SYS ‘CR’
DO YOU WANT SELECT ? N ‘CR’
IF FILE EXISTS, OVERWRITE OK? Y ‘CR’
COPY START !
.
.
F> RT ‘CR’
4–42 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 11 2205009
** CUSTOM PARAMETERS **
1. DISPLAY PARAMETERS
2. CHANGE PARAMETERS
Select No. = 2 ‘CR’
f. Press the [End] key to skip a menu and repeat this until ‘FD Contents Option’ is displayed.
NOTICE
Perform the following step; do not skip, since the V/R 5.02 system software requires the re–
installation of the stated option software.
(The same procedure was performed during the V/R 5.00 upgrade, and again, this must be
performed)
h. Insert the option floppy disk (YLNK or ELNK) into the drive and enter the number of the option software.
(Although the options will be indicated as ‘Installed’, still, however, the re–installation must be performed)
i. Repeat the above step until all of the options are installed.
j. Press the [End] key to skip a menu and repeat this until the following message appears.
Check that the system starts up normally and that the “VER. 05.02” and logo appear on the CRT monitor.
4–43 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 11 2205009
b. Insert the BOOT floppy disk. (This step is not required for systems installed in Japan)
F>
F>RU ‘CR’
SELECT No.= 2S
WHAT FILE NAME ?
SOURCE: V503 ‘CR’
OBJECT: SYS ‘CR’
DO YOU WANT SELECT ? N ‘CR’
IF FILE EXISTS, OVERWRITE OK? Y ‘CR’
COPY START !
.
.
F> RT ‘CR’
Check that the system starts up normally and that the “VER. 05.03” and logo appear on the CRT monitor.
4–44 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 11 2205009
Option Misc
a. Enter the Miscellaneous Application menu by pressing Soft or Appli key.
c. Press the [Enter] key and repeat this until the CBF Processing menu appears.
This step is required for systems with a Functional Image option only.
Option Image
a. Press Soft or Compst key.
24. Restore the Quick Filming Delay Time user setting parameter.
4–45 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 11 2205009
1. Set the SW2 and SW4 of the ‘BOOT’ switches on the FMP1 Board (Slot #30) to the ON position.
ON
1 ËË
ËËËË
ËËËË
2
3
4 ËË
2. Insert the BOOT floppy into the disk drive.
3. Pull the reset switch located under the right side of the Operator Console.
SELECT No.= 2S
WHAT FILE NAME ?
SOURCE: SYS:0.ELNKFMP. ‘CR’
OBJECT: SYS:0.FMP. ‘CR’
DO YOU WANT SELECT ? N ‘CR’
IF FILE EXISTS, OVERWRITE OK? Y ‘CR’
COPY START !
.
.
4–46 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 11 2205009
OFF
1 ËË
2 ËË
ËË
3
4 ËË
ËË
6. Pull the system reset switch located under the right side of the Operator Console.
4–47 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 11 2205009
Option Misc
1. Enter the Miscellaneous Application menu by pressing Soft or Appli key.
3. Press the [Enter] key and repeat this until the CBF Processing menu appears.
4–48 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 11 2205009
Option Image
1. Press Soft or Compst key.
4. Select FILE.
8. Repeat step 6.
9. Continue the above until user parameters for all of Menu 1 through Menu 8 are recorded.
10. Press the [End] key and repeat this until exiting this program.
Menu 1 [ ]
Appearance Cal. Mask
Iterations Smooth Matrix
Recirc. Cut CT# Lower
Reciro. Mask CT# Upper
Menu 2 [ ]
Appearance Cal. Mask
Iterations Smooth Matrix
Recirc. Cut CT# Lower
Reciro. Mask CT# Upper
Menu 3 [ ]
Appearance Cal. Mask
Iterations Smooth Matrix
Recirc. Cut CT# Lower
Reciro. Mask CT# Upper
4–49 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 11 2205009
Menu 4 [ ]
Appearance Cal. Mask
Iterations Smooth Matrix
Recirc. Cut CT# Lower
Reciro. Mask CT# Upper
Menu 5 [ ]
Appearance Cal. Mask
Iterations Smooth Matrix
Recirc. Cut CT# Lower
Reciro. Mask CT# Upper
Menu 6 [ ]
Appearance Cal. Mask
Iterations Smooth Matrix
Recirc. Cut CT# Lower
Reciro. Mask CT# Upper
Menu 7 [ ]
Appearance Cal. Mask
Iterations Smooth Matrix
Recirc. Cut CT# Lower
Reciro. Mask CT# Upper
Menu 8 [ ]
Appearance Cal. Mask
Iterations Smooth Matrix
Recirc. Cut CT# Lower
Reciro. Mask CT# Upper
4–50 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
5-1 GENERAL
#$- ,2 '$")( ). *)-$-.- *! .# !*''*1$)" # &%/-.( ). +,* /, -
') *! *..$*)
. .*, (*)$)*1
,0$.2
-* ). , '$")( ).
$'. , ). , '$")( ).
#
,0$.2 +,* /, $- # & +,* /, !*, .# ).,2 ,*..$)" ( #)$-( ) 0$,.$*)- .
,!*,( .#$- # & $! *)-$ , ) --,2
! 2*/ , -/, .#. -*( +,* /, - * )*. #0 .* + ,!*,( $) .#$- - .# +*--$' ''*1 +,* /,
*($).$*)*, , -#*/' *) *! .# !*''*1$)"
→ → → *,
→ → *,
→ *,
WARNING!
ROTATION HAZARD! IN THIS ‘X–RAY ALIGNMENT’, GANTRY COMPONENTS ARE
ADJUSTED AND SCANS ARE PERFORMED. ALWAYS SWITCH OFF THE ‘ROTATE’ SWITCH
BEFORE SERVICING OR ADJUSTING THE GANTRY. ALSO VERIFY THAT NO PEOPLE ARE
IN THE SCAN ROOM WHENEVER A SCAN IS PERFORMED.
5–1 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
Requirements
The critical components for X–ray Alignment are all mounted directly to the rotating Gantry; they are:
D X–ray tube
D Collimator (includes a bow–tie filter and an adjustable beam defining exit slit)
D Detector
D !" $!%
#
$% &&$% "$#(#$( !#"#& (Qj, Qj–Base, Zj)
(Qj–Hlt, Zj–Adv)
D !" $!% # $% &&$% "$#(#$( !#"#&
D !#"#& &$$!
5–2 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
Gantry Geometry
(Qj–Hlt, Zj–Adv)
5–3 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
2. Break the chemical pod over a table edge, or use a squeeze roller as follows:
5–4 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
1. Preparation:
c. Remove the front and both side Gantry covers. (The rear cover should remain installed to attach a Polaroid
film onto it later)
d. Remove the bow–tie filter from the collimator. (Refer to the Component Replacement manual)
a. Mark a ‘T’ on the film to identify the Table side. See Illustration 5–2.
b. Attach the film on the bottom ring of the Gantry rear cover using adhesive tape. And put a lead plate on the
film. See Illustration 5–1. (If a lead plate is not available, use a piece of solder)
5–5 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
Perform two off–line stationary scans; one with 0 deg, the other with 180 deg.
a. Insert the BOOT floppy into the disk drive. (This step is not required for systems installed in Japan)
Scan 1 Scan 2
AZIMUTH : 0 deg AZIMUTH : 180 deg
SCAN TIME : 2.0 sec SCAN TIME : 2.0 sec
THICKNESS : 1 mm THICKNESS : 1 mm
FOCUS : SMALL FOCUS : SMALL
KV : 120 kV KV : 120 kV
MA : 60 mA MA : 60 mA
(Others) : (defaults) (Others) : (defaults)
h. Develop the film. Wait the required film development time described on the film pack.
5–6 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
≤ 0.5 mm –––> proceed to step 6.
> 0.5 mm –––> proceed to the next step 5.
!
!
5–7 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
d. Locate the Adjustment Nuts for Z–Axis; note that there are two Z–Axis nuts for adjustment.
e. Loosen the nut that faces the direction in which you want to shift the tube.
f. Turn the other nut in the direction in which you want to shift the tube.
One turn ≈ 1 mm tube shift. The required tube shift (mm) is measured by the following:
Tube Shift =
See Illustration 5–2; in this example, the tube should be shifted axially toward the Table to center the wide
beam over the narrow beam.
j. Return to step 2.
5–8 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
7. Check if the narrow exposure lies within the wider exposure, and the center lines of the two beams coincide within
0.75 mm:
≤ 0.75 mm –––> proceed to step 8.
> 0.75 mm –––> adjust the tube position per step 5, and then go to step 6.
5–9 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
1. Preparation:
You do not have to perform the following, if you have just completed the ‘Plane of Rotation’ procedure.
c. Remove the bow–tie filter from the collimator. (Refer to the Component Replacement manual)
5–10 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 7 2205009
(Anode heat is indicated as ‘A’ or ‘Anode’ at the upper right of the touch screen)
After the above scans are completed, the Gantry will not halt at the home position. Set the TGP service switches
‘ABT/SVE/FIX’, ‘CONT/HOME/90°’, and ‘SYS/OFF/MNL’ to FIX, HOME, and MNL, respectively, in this order,
to position the Gantry to the home position (x–ray tube at the 12 0’clock position). Then set ‘SYS/OFF/MNL’ to
SYS.
b. Insert a Polaroid film to the two film holders with the side marked “THIS SIDE TOWARD LENS” facing down
and the “INSERT THIS END” toward the Gantry when placed on the Detector. Make sure that the films are
fully inserted into the holders, otherwise you may lose part of the image. One film holder is for the right side of
the Detector, and the other is for the left side. See Illustration 5–4.
Film Holder
Each film holder has a side tab to position it to the window slot of the Detector, and two end tabs to align it to the
end of the detector housing.
c. Mount the two film holders with film onto both ends of the Detector.
5–11 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
4. Ensure that the x–ray tube anode is at 40% or higher heat level; and then, perform a stationary scan on the films:
a. Insert the BOOT floppy into the disk drive. (This step is not required for systems installed in Japan)
#
""
!
"
5–12 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
j. Develop the film. Wait the required film development time described on the film pack.
5. Examine the films for axial centering of the x–ray beam to the center line of the Detector window:
See Illustration 5–5. You should see both the top and bottom edges of x–ray. The edges should be well defined.
< 1 mm, < 1 mm
< 1 mm
5–13 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
e. Turn the Z–Axis adjustment nut in the direction you want to shift the Detector (i. e., toward Table or toward
Gantry) and over the distance required to meet specification using vernier calipers to measure Z–Axis shift
measured in step 5.
g. Make this adjustment at each of the four mounting locations. The measurement for all the four locations
should be the same.
i. Return to step 2.
5–14 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
5–15 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 7 2205009
1. Preparation:
You do not have to perform the following, if you have just completed the ‘Plane of Rotation’ procedure.
c. Remove the bow–tie filter from the collimator. (Refer to the Component Replacement manual)
(Case heat is indicated as ‘C’ or ‘Case’ at the upper right of the touch screen)
After the above scans are completed, the Gantry will not halt at the home position. Set the TGP service switches
‘ABT/SVE/FIX’, ‘CONT/HOME/90°’, and ‘SYS/OFF/MNL’ to FIX, HOME, and MNL, respectively, in this order,
to position the Gantry to the home position (x–ray tube at the 12 0’clock position). Then set ‘SYS/OFF/MNL’ to
SYS.
5–16 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
b. Insert a Polaroid film to the three film holders with the side marked “THIS SIDE TOWARD LENS” facing down
and the “INSERT THIS END” toward the Gantry when placed on the Detector. Make sure that the films are
fully inserted into the holders, otherwise you may lose part of the image. See Illustration 5–7.
Film Holders
Since the Z–Axis position of the Detector is adjusted on three locations (right, left, and center of the Detector),
two kinds of film holders are provided: two for the right and left ends of the Detector, and one for center.
c. Mount the three film holders with film onto the right and left ends and the center, of the Detector. When instal-
ling the center film holder, push it toward the Gantry until its center tab touches the center mounting support of
the Detector. See Illustrations 5–7 and 5–8.
ÁÁÁÁ
ÁÁÁÁ
ÁÁÁÁ ÁÁÁÁ
ÁÁÁÁ ÁÁÁÁ
ÁÁÁÁÁ
ÁÁÁÁÁ
5–17 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
4. Ensure that the x–ray tube case is at 20% or higher heat level; and then, perform a stationary scan on the films:
a. Insert the BOOT floppy into the disk drive. (This step is not required for systems installed in Japan)
!
5–18 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
j. Develop the film. Wait the required film development time described on the film pack.
k. Mark ‘Left’, ‘Right’, or ‘Center’ and ‘Table’ or ‘Gantry’ on the film as appropriate.
5. Examine the films for axial centering of the x–ray beam to the center line of the Detector window:
See Illustration 5–9. You should see both the top and bottom edges of x–ray. The edges should be well defined.
Measure if the x–ray beam centering falls within the following values:
< 1 mm < 1 mm < 1 mm
< 1 mm
5–19 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
c. Remove the outer Detector mounting nut at each of the three locations. See Illustration 5–10.
5–20 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
e. Loosen the bowl adjustment lock nut at each of the three locations.
f. Turn the bowl adjustment to shift the Detector along the Z–Axis according to below:
CW: away from the Table
CCW: toward the Table
One Turn: approximately 2mm shift.
Distance to be shifted: Distance measured in step 5, i.e.;
g. Tighten first the bowl adjustment lock nut, and then, the inner Detector mounting nut at each of the three loca-
tions. The center mounting nut must be tightened first, and then the right and left nuts.
h. Install and tighten the outer Detector mounting nut at at each of the three locations.
j. Return to step 2.
7. Proceed to Section 5-5, ‘Gravity SAG’, or Section 5-6 or 5-7, ‘Iso Center Alignment.’
5–21 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
You do not have to perform the following, if you have just completed the ‘Detector Beam–on–Window’ procedure.
c. Remove the bow–tie filter from the collimator. (Refer to the Component Replacement manual)
2. Perform three air scans with different (Scan Time, Focus, MA) parameters:
a. Insert the BOOT floppy into the disk drive. (This step is not required for systems installed in Japan)
Scan 1
SCAN TIME : 1.0 sec or 2.0 sec (if 1.0 sec is available, select 1.0 sec)
THICKNESS : 10 mm
FOCUS : SMALL
KV : 120 kV
MA : 160 mA
SERIES NO. : 1
Scan 2 Scan 3
SCAN TIME : 3.0 sec SCAN TIME : 3.0 sec
THICKNESS : 10 mm THICKNESS : 10 mm
FOCUS : SMALL FOCUS : LARGE
KV : 120 kV KV : 120 kV
MA : 60 mA MA : 60 mA
SERIES NO. : 2 SERIES NO. : 3
5–22 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
c. Rotate the Gantry manually in the CW direction until the x–ray tube is positioned at the bottom.
e. Move the G–SAG pin into the position shown in Illustration 5–11.
4. Perform three pin scans with different (Scan Time, Focus, MA) parameters:
Scan 4
SCAN TIME : 1.0 sec or 2.0 sec (the selection should be the same as Scan 1)
THICKNESS : 10 mm
FOCUS : SMALL
KV : 120 kV
MA : 160 mA
SERIES NO. : 4
Scan 5 Scan 6
SCAN TIME : 3.0 sec SCAN TIME : 3.0 sec
THICKNESS : 10 mm THICKNESS : 10 mm
FOCUS : SMALL FOCUS : LARGE
KV : 120 kV KV : 120 kV
MA : 60 mA MA : 60 mA
SERIES NO. : 5 SERIES NO. : 6
5–23 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
** START
CENTER CH. = aaa.aaa
AVERAGE CH. = XXX.XXX
g. Repeat the above steps for the other raw data files to obtain mean files No. 3 and NO. 5. according to the
following table:
5–24 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
Perform the steps below (g, h, and i) to show a graph for Channels beyond 500.
j. Overdraw the mean file No. 3 data (located on the third line), to check the following:
Perform the above, by similarly repeating steps c through i. (The mean file used this time is No. 3)
k. Overdraw the mean file No. 5 data (located on the fifth line), to check the following:
Perform the above, by similarly repeating steps c through i. (The mean file used this time is No. 5)
5–25 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
m. Replace the G–SAG pin to the original position. See illustration 5–11.
5–26 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
7. If the Gravity SAG data is out of specifications, verify that the Detector mounting bolts and Tube mounting bolts
are correctly torqued, and then perform this ‘Gravity SAG’ procedure again.
8. Proceed to Section 5-6 or 5-7, ‘Iso Center Alignment’ if you performed the ‘Detector Beam–on–Window’ proce-
dure prior to this ‘Gravity SAG’ procedure; or you do not perform ‘Iso Center Alignment’ or ‘Filter Center Align-
ment’ (Section 5-8), perform the following:
Switch OFF the ‘Rotate’ switch.
Installed the bow–tie filter.
5–27 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
1. Preparation:
You do not have to perform the following, if you have just completed the ‘Detector Beam–on–Window’ or ‘Gravity
SAG’ procedure.
c. Remove the bow–tie filter from the collimator. (Refer to the Component Replacement manual)
5–28 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
a. Insert the BOOT floppy into the disk drive. (This step is not required for systems installed in Japan)
Scan 1 Scan 2
SCAN TIME : 3.0 sec SCAN TIME : 3.0 sec
THICKNESS : 10 mm THICKNESS : 10 mm
FOCUS : SMALL FOCUS : LARGE
KV : 120 kV KV : 120 kV
MA : 60 mA MA : 60 mA
SERIES NO. : 7 SERIES NO. : 8
a. Attach a steel pin, approximately 6 mm in diameter, (ex. drill bit or phillips screw driver) to the phantom holder
at a position approximately 5 cm away from the scan axis center, as shown in Illustration 5–16.
Scan 3 Scan 4
SCAN TIME : 3.0 sec SCAN TIME : 3.0 sec
THICKNESS : 10 mm THICKNESS : 10 mm
FOCUS : SMALL FOCUS : LARGE
KV : 120 kV KV : 120 kV
MA : 60 mA MA : 60 mA
SERIES NO. : 9 SERIES NO. : 10
5–29 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
5–30 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
** START
ISO AVERAGE CH. SMALL = XXX.XXX
LARGE = XXX.XXX
MEAN = XXX.XXX (406.75 ± 0.02)
MOVE TUBE = C.CCC MM (D.DDD Inch) RIGHT
5–31 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
d. Locate the Adjustment Nuts for Theta; note that there are two Theta nuts for adjustment.
e. Loosen the nut that faces the direction in which you want to shift the tube.
f. Turn the other nut in the direction in which you want to shift the tube.
j. Return to step 2.
5–32 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
" !
%&
!"
"
! #
%
!!
" !
Skew Alignment
!/, 020./9 ,9,*957 08 ;,7> 8,48090;, 95 9/, +07,*9054 5- 9/, =7(> ),(3 6(790*:2(72> 0- 9/, ,9,*957 *,22 08 897(0./9
570,49,+ 95 9/, 9:), -5*:8 4 9/08 *(8, ( 820./9 8/0-9 04 9/, -5*:8 65809054 3(> *(:8, (4 (790-(*9 54 03(.,8 57 9/08
7,(854 9/, ,9,*957 08 6:7658,2> 04*204,+ 820./92> 95 9/, $ +07,*9054 !/, 85 *,49,7 */(44,2 86,*0-0*(9054 -57 9/08
(20.43,49 08
± /
5–33 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
You do not have to perform the following, if you have just completed the ‘Detector Beam–on–Window’ or ‘Gravity
SAG’ procedure.
c. Remove the bow–tie filter from the collimator. (Refer to the Component Replacement manual)
a. Insert the BOOT floppy into the disk drive. (This step is not required for systems installed in Japan)
Scan 1 Scan 2
SCAN TIME : 3.0 sec SCAN TIME : 3.0 sec
THICKNESS : 10 mm THICKNESS : 10 mm
FOCUS : SMALL FOCUS : LARGE
KV : 120 kV KV : 120 kV
MA : 60 mA MA : 60 mA
SERIES NO. : 7 SERIES NO. : 8
5–34 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
a. Attach a steel pin, approximately 6 mm in diameter, (ex. drill bit or phillips screw driver) to the phantom holder
at a position approximately 5 cm away from the scan axis center, as shown in Illustration 5–16.
Scan 3 Scan 4
SCAN TIME : 3.0 sec SCAN TIME : 3.0 sec
THICKNESS : 10 mm THICKNESS : 10 mm
FOCUS : SMALL FOCUS : LARGE
KV : 120 kV KV : 120 kV
MA : 60 mA MA : 60 mA
SERIES NO. : 9 SERIES NO. : 10
4. Check that the calculated Iso Center Channel data is within a specification:
(The displayed specification is 458.75 ±0.02; ignore this at this ‘Iso Center Rough Alignment’)
5–35 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
d. Locate the Adjustment Nuts for Theta; note that there are two Theta nuts for adjustment.
e. Loosen the nut that faces the direction in which you want to shift the tube.
f. Turn the other nut in the direction in which you want to shift the tube.
j. Return to step 2.
Skew Alignment
6. Wait at least 30 minutes since the last scan.
5–36 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
b. Install the Skew Tool Assy (part number: 2127793) by fixing the pin and tightening the screw. See Illustration
5–18.
5–37 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
*** START
AVERAGE CENTROID OF THE PIN SCAN= XXX.XXX (462.37 ± 0.20)
MOVE DETECTOR X.XXX mm (X.XXXX inch) CCW
d. If the AVERAGE value is out of 462.37 0.2, perform the following step (10).
If the AVERAGE value is 462.37 ±0.2, perform the following:
iii. if you performed the following step (10), remove the Dial Gauge.
5–38 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
c. Remove the outer Detector mounting nut at each of the three locations. See illustration 5–10.
d. Loosen the inner Detector mounting nut at each of the three locations.
e. Install the Dial Gauge (part number: U0051BF) onto the metal board located below the left end of the Detector,
using two butterfly fixing nuts. See Illustration 5–19.
g. Adjust the Detector position by turning the adjustment screw by the amount instructed on the CRT screen.
i. Tighten the inner Detector mounting nut at each of the three locations. The center mounting nut must be tight-
ened first, and then the right and left nuts.
j. Install and tighten the outer Detector mounting nut at at each of the three locations.
m. Return to step 6.
5–39 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
→
→
5–40 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
1. Preparation:
You do not have to perform the following, if you have just completed the ‘Iso Center Alignment’ procedure.
c. Remove the bow–tie filter from the collimator. (Refer to the Component Replacement manual)
5–41 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
a. Insert the BOOT floppy into the disk drive. (This step is not required for systems installed in Japan)
5–42 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
(Qj–Hlt, Zj–Adv)
** START
d.
!
± " (Qj, Qj–Base, Zj)
!
±
e.
!
± " (Qj–Hlt, Zj–Adv)
!
±
5–43 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
d. Rotate the adjustment screw to shift the filter for the amount instructed on the CRT screen.
f. Return to step 3.
6. Wrap–up:
5–44 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
6-1 GENERAL
This ‘System Calibration’ describes the following check/calibration/adjustment:
D CT Number Adjustment
The above calibration and adjustment should be performed if any of the following replacements and/or system adjust-
ments are performed since the last calibration/adjustment.
‘CAM Amplifier Linearity Check’ may be performed before air/phantom calibrations.
D Detector replacement
6–1 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
(Qj–Hlt, Zj–Adv)
The following table shows that $ Hilight calibration (Q-cal, XT-cal, AV-cal, or DG-cal) should be performed
for above occasions and also the calibration/adjustment to be followed.
Required Phantoms
Use the following phantoms (furnished with each system) during system calibration.
Make sure they are clean and contrast–free.
6–2 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
6–3 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
(Qj–Hlt, Zj–Adv)
" ! #
"
1. Perform two off–line scans; one with 100 mA, the other with 200 mA:
a. Insert the BOOT floppy into the disk drive. (This step is not required for systems installed in Japan)
Scan 1 Scan 2
AZIMUTH : 0 deg AZIMUTH : 0 deg
SCAN TIME : 3.0 sec SCAN TIME : 3.0 sec
THICKNESS : 5 mm THICKNESS : 5 mm
KV : 120 kV KV : 120 kV
MA : 100 mA MA : 200 mA
FOCUS : LARGE FOCUS : LARGE
SERIES NO. : 1 SERIES NO. : 2
(Others) : (defaults) (Others) : (defaults)
6–4 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
Obtain the following mean files 1 and 2, by repeating the steps below.
(Qj–Hlt, Zj–Adv)
e.
OFFSET FILE CORRECTION <0:YES, 1:NO> (1)=0
REFERENCE CH. CORRECTION <0:6ch, 1:16block, 2:NO> (2)=1
NATURAL LOG. CONVERSION <0:YES, 1:NO> (1)=1
6–5 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
Check that the non-linearity (shown as E" in the illustration) is less than (0.1 %). In the illustration,
A" and B" indicate the average values of approximately 10 channels each. The difference between B" and
A" (= E") shows the magnitude of the non-linearity of the amplifiers.
(Qj–Hlt, Zj–Adv)
Check that the non-linearity (shown as E" in the illustration) is less than (0.1 %) for 363 ~ 563 Ch.
and less than (0.2 %) for the other channels. In the illustration, A" and B" indicate the average values
of approximately 10 channels each. The difference between B" and A" (= E") shows the magnitude of the
non-linearity of the amplifiers.
Perform the steps below (f, g, and h) to show a graph for Channels beyond 500.
h. Repeat step d.
6–6 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
6–7 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
6-3-1 General
This Hilight calibration should be performed prior to the air and phantom calibrations; and the air and phantom calibra-
tions should be performed if once the Hilight calibration is performed.
The x–ray tube must be sufficiently cool (the tube case temperature must be below 10 % heat level).
# !
"! ! ! !$
! !
# !
"! ! ! !$
! !
6–8 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
6-3-2 Q Calibration
The focus position of the x–ray tube shifts with the temperature change. It will consequently shift the x–ray beam
position along the Z–axis. In this calibration, the following scans are automatically performed:
D four scans (Thickness: 1, 3, 5, and 10 mm) (when the x–ray tube is cold)
D warm–up scans
D four scans (Thickness: 1, 3, 5, and 10 mm) (when the x–ray tube is hot)
This calibration takes approximately 10 minutes.
1. Preparation:
6–9 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
6-3-3 XT Calibration
The Lumex crystals of the Detector cells generate light when they are hit by x–rays. Since the crystals are placed close
to the crystals of the adjacent channels, the light from one channel can interfere with adjacent channels, causing
crosstalk noise. This procedure creates the calibration file to compensate this interference.
This calibration takes approximately 15 minutes.
1. Preparation:
4. Perform air scans, following the instructions displayed on the CRT screen.
6–10 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
ROTATION HAZARD! WHEN YOU PRESS THE ‘POSN LIGHT’ BUTTONS (LARGE BUTTON
AND SMALL BUTTON), THE GANTRY QUICKLY ROTATES TO THE 0_ POSITION. IF IT IS
NECESSARY TO PLACE ANY PART OF YOUR BODY INSIDE OR NEAR THE GANTRY, SWITCH
OFF THE ‘SCAN’ SWITCH FIRST.
c. Position the phantom at the scan center; align the positioning lights and a cross marked on the phantom sur-
face.
d. Hold down the ‘Height’ button to display the current height (ex. ‘–198’), and press the ‘Up’ button to raise the
Table by 180 mm (ex. Until the Display reads ‘–018’).
If the Table can not be raised to the specified height due to the interlock function, switch ON (up) the T3 Table
dip switch on the TGP Board to disable the interlock function, and continue to raise the Table. In this case, be
careful not to have the Table collie against anything.
6–11 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
Do not forget to switch OFF the T3 Table dip switch, otherwise there will be collision hazard, since
the interlock function does not work, if T3 is set to ON.
6–12 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
6-3-4 AV Calibration
The light emission of the Lumex Crystal persists after the x–ray radiation stops. This phenomenon is called ‘After
Glow.’ This procedure creates a file called After Glow Vector, used to compensate the After Glow.
This calibration takes approximately 2 minutes.
1. Preparation:
1. CALIBRATION
2. HILIGHT CALIBRATION
Select No. ? 2 ‘CR’
6–13 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
6-3-5 DG Calibration
If some scans are performed with the same conditions but scan times, the CT numbers may not be the same for these
scans. This is due to CAM Board gain changes caused by different scan times. The DG calibration reduces the CT
number changes.
This calibration takes approximately 3 minutes.
1. Preparation:
6–14 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
Precondition
D
!
(Qj, Qj–Base)
1. CALIBRATION
.
.
Select No. ? 1 ‘CR’
b. Follow the instructions displayed on the CRT screen. The scans performed in this automatic calibration are
described in Table 6–1.
6–15 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
6-4 AIR AND PHANTOM CALIBRATIONS (continued) (Qj, Qj–Base, Qj–Hlt, Zj–Adv)
Warm Up Scan -
Air Scan 17
Phantom Scan 6
Phantom: 25 cm Water
FOV: 25 cm
Phantom Scan 6
Phantom: 35 cm Polypropylene
FOV: 35 cm
Phantom Scan 6
Phantom: 42 cm Polypropylene
FOV: 42 cm
Phantom Scan 6
Phantom: 42 cm Polypropylene
FOV: 50 cm
At the beginning of each of the above phantom scans, you are
instructed to center the phantom.
Warm Up Scan -
Air Scan 17
Phantom Scan 6
Phantom: 25 cm Water
FOV: 25 cm
Phantom Scan 6
Phantom: 35 cm Polypropylene
FOV: 35 cm
Phantom Scan 6
Phantom: 42 cm Polypropylene
FOV: 50 cm
At the beginning of each of the above phantom scans, you are
instructed to center the phantom.
6–16 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
Precondition
D The x–ray tube anode should be at 10% or below heat level.
(Anode heat is indicated as ‘Anode’ at the upper right of the touch screen)
After the warm–up scans are finished, promptly proceed to the next step.
6–17 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
c. Perform an AP (0_) ScoutView scan and an LT (90_) ScoutView scan to check that the phantom is positioned
straight.
h. Center the phantom, following the instructions displayed on the CRT screen.
j. Repeat the above steps (a through i; in step g, select the correct FOV listed below) for the following phantoms
and FOVs in the order written:
6–18 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
Note
The CT# of water will change according to its temperature by a factor of –0.4/_C.
a. Touch the [Preset Prmtr] key while the [New Patient] key flashes.
b. Select SCAN PARAMETER from the Preset Parameters menu on the CRT screen.
For a MA parameter, set this parameter after a Thickness parameter is set, otherwise, the MA parameter may
be automatically modified by the system according to the thickness parameter selected.
! # # " !#!" "!
! " %
! " %
! " %
#! # # " !#!" "!
#! " %
#! " %
#! " %
$#
6–19 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
3. Perform the 25 cm water phantom scans, using the preset scan parameters (step 1):
e. Center the water phantom, following the instructions displayed on the CRT screen.
4. Record the CT number of each of all the air scan images acquired:
b. Display a cursor specified in the table below, at the reconstruction center (xPOSN = yPOSN = 0 cm).
Refer to Section 3-1-1, ‘Set Cursor’, of Tab ‘Introduction.’
D Table 6–3
D Table 6–4
D Table 6–5
D Table 6–6
5. Record the CT number of each of all the water scan images acquired, similarly as step 4.
6–20 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
a. Insert the BOOT floppy into the disk drive. (This step is not required for systems installed in Japan)
** ADJUST CT# **
1. CHANGE CT# PARAMETER
2. INITIALIZE CT# PARAMETER
Select No.=1 ’CR’
6–21 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
80 kV 120 kV 140 kV
FOV TH FOV TH FOV TH
1. 25 cm 10 mm 17. 25 cm 10 mm 33. 25 cm 10 mm
2. 25 cm 5 mm 18. 25 cm 5 mm 34. 25 cm 5 mm
3. 25 cm 3 mm 19. 25 cm 3 mm 35. 25 cm 3 mm
4. 25 cm 1 mm 20. 25 cm 1 mm 36. 25 cm 1 mm
5. 35 cm 10 mm 21. 35 cm 10 mm 37. 35 cm 10 mm
6. 35 cm 5 mm 22. 35 cm 5 mm 38. 35 cm 5 mm
7. 35 cm 3 mm 23. 35 cm 3 mm 39. 35 cm 3 mm
8. 35 cm 1 mm 24. 35 cm 1 mm 40. 35 cm 1 mm
9. 42 cm 10 mm 25. 42 cm 10 mm 41. 42 cm 10 mm
10. 42 cm 5 mm 26. 42 cm 5 mm 42. 42 cm 5 mm
11. 42 cm 3 mm 27. 42 cm 3 mm 43. 42 cm 3 mm
12. 42 cm 1 mm 28. 42 cm 1 mm 44. 42 cm 1 mm
<xx>
<xx–yy>
Select No. <xx,yy,zz>=1 ’CR’
6–22 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
D Table 6–3
D Table 6–4
D Table 6–5
D Table 6–6
KV 80 kV
FOV 25 cm
TH 10 mm
Enter AIR CT# (–1000.00)=–998.50 ’CR’
Enter WATER CT# ( 0.00)=2.43 ’CR’
b <OLD> 7.84 E–04 <NEW> 7.83 E–04
c <OLD> 0.00 <NEW> –2.43
OK?<0:Yes, 1:No>0 ’CR’
.
.
.
6–23 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
The reconstructed image will be stored with an Image ID: (Study#).1.1;2 (example).
e. Use the same ROIs for CT number measurement as were used for scanned images.
Refer to the following tables:
D Table 6–3
D Table 6–4
D Table 6–5
D Table 6–6
If the Retro Recon FOV is the same as for the last scan FOV, the CT number will not be updated. Perform
Retro Recons with a different FOV as was used in the last scan’
6–24 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
*: For systems where 7 mm thickness can be selected. SCAN PARAMETERS SCAN PARAMETERS
TIME: 3 SEC TIME: 2 SEC
MA: 60 mA MA: See above.
ROI SIZE (BOX): ROI SIZE (BOX):
1.00 cm width x 24.0 cm height 6 cm width x 6 cm height
6–25 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
*: For systems where 7 mm thickness can be selected. SCAN PARAMETERS SCAN PARAMETERS
TIME: 3 SEC TIME: 2 SEC
MA: 60 mA MA: See above.
ROI SIZE (BOX): ROI SIZE (BOX):
1.41 cm width x 33.8 cm height 6 cm width x 6 cm height
6–26 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
*: For systems where 7 mm thickness can be selected. SCAN PARAMETERS SCAN PARAMETERS
TIME: 3 SEC TIME: 2 SEC
MA: 60 mA MA: See above.
ROI SIZE (BOX): ROI SIZE (BOX):
1.67 cm width x 40.16 cm 6 cm width x 6 cm height
height
6–27 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
*: For systems where 7 mm thickness can be selected. SCAN PARAMETERS SCAN PARAMETERS
TIME: 3 SEC TIME: 2 SEC
MA: 60 mA MA: See above.
ROI SIZE (BOX): ROI SIZE (BOX):
2.00 cm width x 48.0 cm height 6 cm width x 6 cm height
6–28 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
Note
The CT# of water will change according to its temperature by a factor of –0.4/_C.
a. Touch the [Preset Prmtr] key while the [New Patient] key flashes.
b. Select SCAN PARAMETER from the Preset Parameters menu on the CRT screen.
For a MA parameter, set this parameter after a Thickness parameter is set, otherwise, the MA parameter may
be automatically modified by the system according to the thickness parameter selected.
! # # " !#!" "!
6–29 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
3. Perform 25 cm water phantom scans, using the preset scan parameters (step 1):
e. Center the water phantom, following the instructions displayed on the CRT screen.
4. Record the CT number of each of all the air scan images acquired:
b. Display a cursor specified in the table below, at the reconstruction center (xPOSN = yPOSN = 0 cm).
Refer to Section 3-1-1, ‘Set Cursor’, of Tab ‘Introduction.’
D Table 6–8
D Table 6–9
D Table 6–10
5. Record the CT number of each of all the water scan images acquired, similarly as step 4.
6–30 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
a. Insert the BOOT floppy into the disk drive. (This step is not required for systems installed in Japan)
** ADJUST CT# **
1. CHANGE CT# PARAMETER
2. INITIALIZE CT# PARAMETER
Select No.=1 ’CR’
6–31 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
80 kV 120 kV 140 kV
FOV TH FOV TH FOV TH
1. 25 cm 10 mm 13. 25 cm 10 mm 25. 25 cm 10 mm
2. 25 cm 5 mm 14. 25 cm 5 mm 26. 25 cm 5 mm
3. 25 cm 3 mm 15. 25 cm 3 mm 27. 25 cm 3 mm
4. 25 cm 1 mm 16. 25 cm 1 mm 28. 25 cm 1 mm
5. 35 cm 10 mm 17. 35 cm 10 mm 29. 35 cm 10 mm
6. 35 cm 5 mm 18. 35 cm 5 mm 30. 35 cm 5 mm
7. 35 cm 3 mm 19. 35 cm 3 mm 31. 35 cm 3 mm
8. 35 cm 1 mm 20. 35 cm 1 mm 32. 35 cm 1 mm
9. 50 cm 10 mm 21. 50 cm 10 mm 33. 50 cm 10 mm
10. 50 cm 5 mm 22. 50 cm 5 mm 34. 50 cm 5 mm
11. 50 cm 3 mm 23. 50 cm 3 mm 35. 50 cm 3 mm
12. 50 cm 1 mm 24. 50 cm 1 mm 36. 50 cm 1 mm
<xx>
<xx–yy>
Select No. <xx,yy,zz>=1 ’CR’
f. Specify each CAL file one by one (<xx>) and enter air CT# and water CT#, from the following tables:
D Table 6–8
D Table 6–9
D Table 6–10
KV 80 kV
FOV 25 cm
TH 10 mm
Enter AIR CT# (–1000.00)=–998.50 ’CR’
Enter WATER CT# ( 0.00)=2.43 ’CR’
b <OLD> 7.84 E–04 <NEW> 7.83 E–04
c <OLD> 0.00 <NEW> –2.43
OK?<0:Yes, 1:No>0 ’CR’
.
.
.
6–32 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
The reconstructed image will be stored with an Image ID: (Study#).1.1;2 (example).
e. Use the same ROIs for CT number measurement as were used for scanned images.
Refer to the following tables:
D Table 6–8
D Table 6–9
D Table 6–10
If the Retro Recon FOV is the same as for the last scan FOV, the CT number will not be updated. Perform
Retro Recons with a different FOV as was used in the last scan’
6–33 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
*: For systems where 7 mm thickness can be selected. SCAN PARAMETERS SCAN PARAMETERS
TIME: 3 SEC TIME: 2 SEC
MA: 60 mA MA: See above.
ROI SIZE (BOX): ROI SIZE (BOX):
1.00 cm width x 24.0 cm height 6 cm width x 6 cm height
6–34 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
*: For systems where 7 mm thickness can be selected. SCAN PARAMETERS SCAN PARAMETERS
TIME: 3 SEC TIME: 2 SEC
MA: 60 mA MA: See above.
ROI SIZE (BOX): ROI SIZE (BOX):
1.41 cm width x 33.8 cm height 6 cm width x 6 cm height
6–35 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
*: For systems where 7 mm thickness can be selected. SCAN PARAMETERS SCAN PARAMETERS
TIME: 3 SEC TIME: 2 SEC
MA: 60 mA MA: See above.
ROI SIZE (BOX): ROI SIZE (BOX):
2.00 cm width x 48.0 cm height 6 cm width x 6 cm height
6–36 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
Note
The CT# of water will change according to its temperature by a factor of –0.4/_C.
a. Touch the [Maint.] key while the [New Patient] key flashes.
b. Select SCAN PRESET PARAMETER from the Maintenance menu on the CRT screen.
For a MA parameter, set this parameter after a Thickness parameter is set, otherwise, the MA parameter may
be automatically modified by the system according to the thickness parameter selected.
25 (Air Scan) Set the scan parameters described in Table 6-12
35 (Air Scan) Same as above
50 (Air Scan) Same as above
25 (Water Scan) Set the scan parameters described in Table 6-12
35 (Water Scan) Same as above
50 (Water Scan) Same as above
Slice Count:12
INTVAL: 0
6–37 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
3. Perform 25 cm water phantom scans, using the preset scan parameters (step 1):
e. Center the water phantom, following the instructions displayed on the CRT screen.
4. Record the CT number of each of all the air scan images acquired:
b. Display a cursor specified in the table below, at the reconstruction center (xPOSN = yPOSN = 0 cm).
Refer to Section 3-1-1, ‘Set Cursor’, of Tab ‘Introduction.’
D Table 6–12
D Table 6–13
D Table 6–14
5. Record the CT number of each of all the water scan images acquired, similarly as step 4.
6–38 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
a. Insert the BOOT floppy into the disk drive. (This step is not required for systems installed in Japan)
6–39 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
80 kV 120 kV 140 kV
FOV TH FOV TH FOV TH
1. 25 cm 10 mm 13. 25 cm 10 mm 25. 25 cm 10 mm
2. 25 cm 5 mm 14. 25 cm 5 mm 26. 25 cm 5 mm
3. 25 cm 3 mm 15. 25 cm 3 mm 27. 25 cm 3 mm
4. 25 cm 1 mm 16. 25 cm 1 mm 28. 25 cm 1 mm
5. 35 cm 10 mm 17. 35 cm 10 mm 29. 35 cm 10 mm
6. 35 cm 5 mm 18. 35 cm 5 mm 30. 35 cm 5 mm
7. 35 cm 3 mm 19. 35 cm 3 mm 31. 35 cm 3 mm
8. 35 cm 1 mm 20. 35 cm 1 mm 32. 35 cm 1 mm
9. 50 cm 10 mm 21. 50 cm 10 mm 33. 50 cm 10 mm
10. 50 cm 5 mm 22. 50 cm 5 mm 34. 50 cm 5 mm
11. 50 cm 3 mm 23. 50 cm 3 mm 35. 50 cm 3 mm
12. 50 cm 1 mm 24. 50 cm 1 mm 36. 50 cm 1 mm
<xx>
<xx–yy>
Select No. <xx,yy,zz>=1 ’CR’
g. Specify each CAL file one by one (<xx>) and enter air CT# and water CT#, from the following tables:
D Table 6–12
D Table 6–13
D Table 6–14
KV 80 kV
FOV 25 cm
TH 10 mm
Enter AIR CT# (–1000.00)=–998.50 ’CR’
Enter WATER CT# ( 0.00)=2.43 ’CR’
b <OLD> 7.84 E–04 <NEW> 7.83 E–04
c <OLD> 0.00 <NEW> –2.43
OK?<0:Yes, 1:No>0 ’CR’
.
.
.
6–40 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
The reconstructed image will be stored with an Image ID: (Study#).1.1;2 (example).
e. Use the same ROIs for CT number measurement as were used for scanned images.
Refer to the following tables:
D Table 6–12
D Table 6–13
D Table 6–14
If the Retro Recon FOV is the same as for the last scan FOV, the CT number will not be updated. Perform
Retro Recons with a different FOV as was used in the last scan’
6–41 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
6–42 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
6–43 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
6–44 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
6–45 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
6–46 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
6–47 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
blank 2205009
6–48 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
7-1 GENERAL
This ‘Image Performance Verification’ contains tables of scan techniques and specifications for evaluating image
performance.
General Requirements
Perform the following before continuing:
1. Verify that system calibration (air and phantom calibrations, ...) has been completed.
All image performance scans must use calibration data that are less than 8 hours old.
2. Warm up the system at least 2.0 hours (minimum) prior to data collection.
Required Phantoms
Use the following phantoms (furnished with each system) during image performance verification.
Make sure they are clean and contrast–free.
D
7–1 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
7–2 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
D Visually check each image for rings, bands, streaks, shadings or artifacts near the image center.
D Position a circular ROI at the locations listed below (see also Illustration 7–1) and write down the data on
the table.
D The ROI size should be 2 cm in diameter. (ProSpeed S, ProSpeed SX, Lemage, Qj–Base, Zj)
(Qj–Hlt, Zj–Adv)
D The ROI size (diameter) should be changed according to the FOV size as described below:
25 cm 2 cm
35 cm 3 cm
50 cm 4 cm
7–3 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
+Y ROI CIRCLE (2 cm
IN DIAMETER)
IMAGE
-X +X
Mc (0.0 cm) Mo (10.0 cm)
SD (0.0 cm)
-Y
-X +X
Mc (0.0 cm) Mo (10.0 cm)
SD (0.0 cm)
-Y
7–4 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
7–5 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
The operator’s manual instructs customers to perform QA checks on a daily basis. Service persons, however, shall
perform QA checks after any of the following events has occurred:
D System installation
D Detector replacement
D KV/MA calibration
D Periodic maintenance.
Note
The service person shall complete all of the procedures in the Operator’s Manual.
7–6 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
7–7 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
7–8 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
7–9 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
Table 7–2 Image Performance Check (continued) (ProSpeed SX, Lemage, Qj–Base)
7–10 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
42 cm PP 50 120 1.5 250 10 M10 ≤ 16.0 M10 ± 3 –120 ≤ M10 ≤ –100 (Note 1)
42 cm PP 42 120 1.5 250 10 M20 ≤ 16.0 M20 ± 3 –120 ≤ M20 ≤ –100 (Note 1)
7–11 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
Table 7–3 Image Performance Check (continued) (ProSpeed SX, Lemage, Qj–Base)
7–12 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
Table 7–3 Image Performance Check (continued) (ProSpeed SX, Lemage, Qj–Base)
Note 1: For systems where the maximum MA available is 300 mA or below only.
Note 2: For systems where the maximum MA available is 350 mA only.
Note 3: For systems where the maximum MA available is 250 mA only.
Note 4: For systems where the maximum MA available is 300 mA or 350 mA only.
Note 5: For systems where 0.8 sec scan is available only.
Note 6: For systems which have a 60 continuous scan capability only.
Note 7: For systems where 7 mm thickness can not be selected, ignore checks for 7 mm and 2 mm; for systems where 7 mm thickness can be selected also, however,
checks for 7 mm and 2 mm can be skipped.
7–13 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
7–14 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
7–15 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
25 cm WTR 50 120 1.5 160 10 0±3 ≤ 4.0 XC ± 3 Xo should be measured 9 cm off cen-
ter ROI.
7–16 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
7–17 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
7–18 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
7–19 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
7–20 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 1 2205009
42 cm PP 50 120 0.8 300 LARGE 10 M10 ± 2 ≤ 15.6 – – for systems where 0.8 sec scan is
available only
35 cm PP 35 120 0.8 350 LARGE 10 M20 ± 2 ≤ 11.2 – – for systems where 0.8 sec scan is
available only
7–21 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 1 2205009
25 cm WTR 50 120 1 250 LARGE 10 0±3 ≤ 4.0 XC ± 3 Xo should be measured 9 cm off cen-
ter ROI.
25 cm WTR 25 120 0.8 350 LARGE 10 M30 ± 2 ≤ 4.7 – – for systems where 0.8 sec scan is
available only
7–22 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 1 2205009
Note 1: For systems where 7 mm thickness can not be selected, ignore checks for 7 mm and 2 mm; for systems where 7 mm thickness can be selected also, however,
checks for 7 mm and 2 mm can be skipped.
7–23 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
Table 7–8 Image Performance Check (ProSpeed, ProSpeed Plus, ProSeed, ProSeed EF)
7–24 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
7–25 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
Table 7–9 Image Performance Check (ProSpeed VX, ProSeed Accell, ProSeed Accell EI)
35 cm PP 50 120 1.5 250 LARGE 10 M10 ≤ 8.7 M10 ± 3 –120 ≤ M10 ≤ –100
7–26 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
7–27 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
blank 2205009
7–28 SYSTEM
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
OPERATOR CONSOLE
TABLE OF CONTENTS
SECTION PAGE
i OPERATOR CONSOLE
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
blank 2205009
ii OPERATOR CONSOLE
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
SECTION 1 – OC COMPONENTS
Ground Connection
If this is the first time to check the OC power supplies after system installation, perform this ‘Ground Connection’
check.
a. Connect a digital multimeter between the OC ground stud and ground (FG terminal) on one power supply.
(Multimeter: resistance measure mode) See Illustration 1–2.
Voltage Check
4. Check the output of the +5 VDC power supply:
a. Connect a digital multimeter across the capacitor (C1) at the front edge of the IPU2 Board (Slot #2). See
Illustration 1–1.
(Multimeter: voltage measure mode)
c. Measure the voltage and check that it is within +5.0 VDC ∼ +5.1 VDC.
If it falls outside these limits, adjust it by turning the potentiometer. See Illustration 1–2.
e. Observe the voltage and check that the ripple is less than 200 mVp–p.
If it falls outside this range, replace the power supply.
VOLTAGE ADJUSTING
POTENTIOMETERS
PWR
a. Connect a digital multimeter at the terminal strip on the front of the power supply.
b. Measure the voltage and check that it is within +11.8 VDC ∼ +12.4 VDC.
If it falls outside these limits, adjust it by turning the potentiometer. See Illustration 1–2.
d. Observe the voltage and check that the ripple is less than 100 mVp–p.
If it falls outside this range, replace the power supply.
Ground Connection
If this is the first time to check the OC power supplies after system installation, perform this ‘Ground Connection’
check.
a. Connect a digital multimeter between the OC ground stud and ground (FG terminal) on one power supply.
(Multimeter: resistance measure mode)
Voltage Check
4. Check the output of the +5 VDC power supply:
a. Connect a digital multimeter across the capacitor (C1) at the front edge of the IPU2 Board (Slot #32). See
Illustration 1–1.
(Multimeter: voltage measure mode)
c. Measure the voltage and check that it is within +4.9 VDC ∼ +5.1 VDC.
If it falls outside these limits, adjust it by turning the potentiometer. See Illustration 1–3; the potentiometer can
be accessed through a hole on the protection cover (the potentiometer is a white one, and located next to an
LED).
e. Observe the voltage and check that the ripple is less than 200 mVp–p.
If it falls outside this range, replace the power supply.
6. Check the output of the +12 VDC power supply the same way as step 4; specifications are given below:
Ground Connection
If this is the first time to check the OC power supplies after system installation, perform this ‘Ground Connection’
check.
a. Connect a digital multimeter between the OC ground stud and ground (FG terminal) on one power supply.
(Multimeter: resistance measure mode)
Voltage Check
4. Check the output of the +5 VDC power supply:
a. Connect a digital multimeter across the capacitor at the front edge of the FCP2A Board (Slot #36). See Il-
lustration 1–4.
(Multimeter: voltage measure mode)
CAPACITOR
FCP2A BOARD
(FRONT VIEW)
c. Measure the voltage and check that it is within +4.9 VDC ∼ +5.1 VDC.
If it falls outside these limits, adjust it by turning the potentiometer. See Illustration 1–5; the potentiometer can
be accessed through a hole on the protection cover (the potentiometer is a white one, and located next to an
LED).
e. Observe the voltage and check that the ripple is less than 200 mVp–p.
If it falls outside this range, replace the power supply.
a. Connect a digital multimeter at the terminal strip on the front of the power supply. See Illustration 1–5.
(Multimeter: voltage measure mode)
b. Measure the voltage and check that it is within –4.9 VDC ∼ –5.1 VDC.
If it falls outside these limits, adjust it by turning the potentiometer.
d. Observe the voltage and check that the ripple is less than 200 mVp–p.
If it falls outside this range, replace the power supply.
6. Check the output of the +24 VDC power supply the same way as step 5; specifications are given below:
7. Check the output of the +12 VDC power supply the same way as step 5; specifications are given below:
1-4 FANS
1. Locate and verify the operation of the following fans. If any fan is defective, replace the unit.
(Qj, Qj–Base, Qj–Hlt)
! ! " !
! ! !! ! !
! # "$
(Zj–Adv, Zj)
! ! ! $
! ! " !
! ! !! ! !
! # "$
! # "$
! # "$
ËËËËËËË
ËËËËËËË
ËËËËËËË
ËËËËËËË
ËËËËËËË
VIDEO VIDEO
VIDEO OUT IN
TERMINATION
1. Remove the rear lower cover of the touch panel, one screw.
2. Insert a screwdriver into a hole of one adjuster knob. See Illustration 1–8.
(Support the touch panel while adjusting.)
Rotate the driver counterclockwise to loosen the knob.
Loosen the other knob.
Adjust the tilt angle.
Tighten the knobs using a screwdriver.
(tighten the knobs such that the pin remains inserted in the two holes of part A and B; see the illustration)
PIN
A OC REAR VIEW
TIGHTEN B
INSERT SCREWDRIVER
LOOSEN
3. Set bit 8 of the dip switch 6A1 on the PDC Board (slot #37 (Zj–Adv) or slot #43 (Zj)) to ON (Calibration Call), as
illustrated below.
5. Follow the instructions displayed on the plasma display to perform the calibration procedure.
1. Disconnect the video signal cable from the Operator Console to the MFC at the MFC side.
EI AEF DD
To change the video signal source (CN3 or CN5), press the MFC control [RESET] button on the console that
is currently sending the image, then press the MFC control [RESET] button on the other console.
If not, adjust the potentiometer VR11 or VR21 on the MV AMP II Board, accordingly (VR11: for OC, VR21: for
DC III).
SECTION 2 – OC OPERATION
Power–up Sequence
The processor of the Operator Console starts with a power–on self test right after powered on, as follows:
[CPW Test – Phase 1] → [DISP (or DISP2) Test – Phase 1] → [CPW Test – Phase 2] → [IPU2 Test] →
[DISP (or DISP2) Test – Phase 2] → [MISC2 Test]
The test progress information is displayed on the CRT monitor for [CPW Test – Phase 2] and the subsequent tests.
Either ‘Power–on Test 1’ or ‘Power–on Test 2’ is performed as the power–on test, according to the dip switch setting on
the CPW Board;
Power–on Test 1: Takes 4 minutes. This test is selected for normal operations.
Power–on Test 2: Takes more than 45 minutes. This test is a more extensive one than Power–on Test 1.
Note
The above power–on test 2 will fail, if both of the following two statements are true:
The CPW Board part number is 2110738.
The IPU2 Board is equipped with less than 144 MB memory.
In this case, a test ‘IPU (33) MEMORY MODULE TEST’ will fail approximately 50 minutes after the
power–on test start, generating a “CPU TIMEOUT ERROR” message.
If the CPW Board is of 2110738–02 or later revision, the problem does not occur.
After the power–on test completes with no error occurred, the processor loads the operation and application software
from the hard disk into memory and starts the application software.
When no error is found in these processes, the system logo message will be shown on the CRT monitor.
3. Verify that the Boot Mode Select switch (location: 10A1) on the CPW Board is set as illustrated below.
(This is the normal setting)
4. Verify that the Test Mode Select switch (location: 8A1) on the CPW Board is set as illustrated below.
This is the normal setting which selects Power–on Test 1. If Power–on Test 2 is desired, set bit 3 to ON.
1 2 3 4 5 6 7 8 OFF: ‘0’
OFF
BP8L
This is the normal setting except for bit 1 which is normally set to ‘ON’, i.e., the PDC power–on test is skipped.
In this procedure, however, perform the PDC power–on test (set bit 1 to ‘OFF’).
ON
`1'
1 2 3 4 5 6 7 8 `0'
6. Switch ON the Operator Console or pull the reset switch located on the underside of the Console.
Verify that “POWER ON TEST 1” or “POWER ON TEST 2” appears on the CRT monitor screen.
Approximately 4 minutes (Test 1) or more than 45 minutes (Test 2) later, “SYSTEM PREPARATION” appears on
the CRT monitor.
System logo message, with software revision, appears on the CRT monitor.
Press the STATUS key and check for error messages.
If the other subsystems (Gantry, Generator, etc.) have been turned ON, the initial screen appears on the touch
panel screen.
7. Set the Test Mode switch (location: 6A1) on the PDC Board as illustrated below. (Zj–Adv)
(Set bit 1 to `ON')
ON
`1'
1 2 3 4 5 6 7 8 `0'
Power–up Sequence
The processors (FMP1 and DMP4) of the Operator Console start with a power–on self test right after powered on.
The File processor (FMP1) and Scan/Display processor (DMP4) each perform a self test independently, and also
perform a communication test with each other.
After the power–on test completes with no error occurred, the File processor loads the operation and application soft-
ware from the hard disk into memory and starts the application software.
When no error is found in these processes, the system logo message will be shown on the CRT monitor.
A firmware program of FMP1 and DMP4 oversees these operations and displays their status on eight bit LEDs on each
board. These status codes become error codes if the system fails during any operation.
3. Verify that Boot Mode Select switch (Location: 16A) on the FMP1 Board (slot #30) is set as illustrated below.
(This is the normal setting)
DMP2 (slot #24; this board is used for Optional DC III only) – SW4A1 (Location: 4A)
5. Switch On the Operator Console or pull the reset switch located on the underside of the Console.
Verify that the “SYSTEM TEST IN PROGRESS” appears on the CRT monitor screen and on the plasma screen.
System logo message, with software revision, appears on the CRT monitor.
Press the [Status] key and check for error messages.
If the other subsystems (Gantry, Generator, etc.) have been turned On, the initial screen appears on the touch
panel screen.
2. Move the AUDIO2 Board to a position such that jumpers and variable resistors can be accessed while adjusting
audio functions:
c. Remove the screws of the base plate for the AUDIO2 Board, and move this board from the nest to an appropri-
ate position.
For the AUDIO2 Board location, see Illustration 1–2 in Section 1-1, ‘General’, of Tab ‘Switch Jumper Setting.’
AUDIO2 Board
VR1 – Record Level VR2 – OC Mic Level VR3 – Patient Mic Level
6. Adjust VR2 until someone standing (or laying) anywhere in the Gantry room can hear your voice when you step
on the OC footswitch and speak.
Make sure that this person can not hear your voice when you speak without stepping on the footswitch.
7. Adjust VR3 until you can hear Gantry room conversation (spoken approximately one meter apart from the Gantry
microphone) while standing in front of the OC.
Make sure you cannot hear Gantry room conversation when you step on the footswitch.
2. Install an extension board to adjust the jumpers and variable resistors on the OC INTF3 Board:
OC INTF3 Board
5. Adjust VR3 until someone standing (or laying) anywhere in the Gantry room can hear your voice when you step
on the OC footswitch and speak.
Make sure that this person can not hear your voice when you speak without stepping on the footswitch.
6. Adjust VR2 until you can hear Gantry room conversation (spoken approximately one meter apart from the Gantry
microphone) while standing in front of the OC.
Make sure you cannot hear Gantry room conversation when you step on the footswitch.
11. Select an appropriate volume for the Auto Voice for the OC speaker, by selecting one of the three positions (MIN,
MEDIUM, MAX) on J8.
12. Adjust VR1 so that the volume of the auto voice messages defined by the customer is equal to the volume of
the standard auto voice messages.
GANTRY
TABLE OF CONTENTS
SECTION PAGE
i GANTRY
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
TABLE OF CONTENTS
SECTION PAGE
ii GANTRY
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
3. Check DC power supply voltages on the TGP Board, according to Table below.
!%'
⋅
⋅
,
GANTRY
CONTROL
SWITCH
1–1 GANTRY
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
b. Switch OFF the ‘Rotate’ and ‘XG’ switches at the Gantry rear base.
c. Remove all the Gantry covers (front, rear, and both side covers).
e. Rotate the Gantry in the clockwise direction by hand until the x–ray tube is positioned at approximately 6
o’clock.
2. Remove the DAS rear cover. See Illustration 1–2. For more information to remove the rear cover, refer to the
Component Replacement manual.
ÉÉ
ÉÉ
ÉÉ
1–2 GANTRY
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
Prior to removing the DAS CN1 or CIF CN4 connector, remove the BNC connector from the
Detector Bias Power Supply. Also, before reconnecting the BNC connector, switch OFF the
‘Slip Ring 115V’ switch and wait at least two minutes.
3. Check DC power supply voltages on the connector pins shown in Table 1–2 or Table 1–3. (Qj–Base)
&&.,-+-$)(, 1–2 or 1–6.
Verify that the power supply voltages are within specifications. If not, adjust the voltage with the appropriate
potentiometer. See Illustrations 1–2 and 1–7.
4. # % *)0 + ,.**&1 /)&-" , )( -# )(( -)+ *$(, ,#)0( $( Table 1–2. (Qj, Zj)
&&.,-+-$)(, 1–2 or 1–6.
Verify that the power supply voltages are within specifications. If not, adjust the voltage with the appropriate
potentiometer. See Illustrations 1–2 and 1–7.
1–3 GANTRY
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
Table 1–3 Power Supply Specifications (for Gantry 2123xxx. 213796x, 2175xxx only)
(High) (Low)
+24 V (for DTRE) DTRE J1-4 (*1) DTRE J1-3 24 V ± 0.5 V PS51
12 V ± 0.5 V(*2)
+5 V (for DTRE) DTRE J1-2 (*1) DTRE J1-1 5.1 V ± 0.1 V PS52
+5V (for DAS) CIF CN4-7 CIF CN4-8 5.2 V ± 0.1 V PS53
+15 V (for DAS) CIF CN4-1 CIF CN4-2 +15.1 V ± 0.1 V PS54
-15 V (for DAS) CIF CN4-4 CIF CN4-5 -15.1 V ± 0.1 V PS55
Note (*1): For accessing the DTRE Board, refer to the Component Replacement manual.
Note (*2): For Gantry 2175xxx only.
DTR2 CN2
9 8 7
6 5 4
3 2 1
OGP CN1 4 3 2 1
B R B Y
1–4 GANTRY
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
ÄÄÄÄÄÄÄÄ
ÄÄÄÄÄÄÄÄ
ÄÄÄÄÄÄÄÄ
ÄÄÄÄÄÄÄÄ
ÄÄÄÄÄÄÄÄ
ÄÄÄÄÄÄÄÄ
ÄÄÄÄÄÄÄÄ
1–5 GANTRY
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
b. Switch OFF the ‘Rotate’ and ‘XG’ and switches at the Gantry rear base.
c. Remove all the Gantry covers (front, rear, and both side covers).
e. Rotate the Gantry in the clockwise direction by hand until the x–ray tube is positioned at approximately 6
o’clock.
2. Remove the DAS left bottom cover, 2 nuts and 6 Allen screws. See Illustration 1–8.
ÉÉÉ
1–6 GANTRY
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
Verify that the power supply voltages are within specifications. If not, adjust the voltage using the appropriate
potentiometer. See Illustrations: 1–8, 1–9, and 1–10.
Table 1–4 Power Supply Specifications (for Gantry 2123xxx. 2137xxx, 2117xxx, 2175xxx only)
+24 V (for DTRE) DTRE J1-4 (*) DTRE J1-3 24 V ± 0.5 V PS51
12 V ± 0.5 V**
+5 V (for DTRE) DTRE J1-2 (*) DTRE J1-1 5.1 V ± 0.1 V PS52
+5V (for DAS) CIF CN4-3 CIF CN4-6 5.2 V ± 0.1 V PS53
+15 V (for DAS) CIF CN4-1 CIF CN4-4 +15.1 V ± 0.1 V PS54
-15 V (for DAS) CIF CN4-5 CIF CN4-2 -15.1 V ± 0.1 V PS55
+24 V (for DAS FAN) DAS FAN-1 DAS FAN-2 24 V ± 0.5 V PS58
Note (*): For accessing the DTRE Board, refer to the Component Replacement manual.
Note (**): For Gantry 2175xxx only.
1–7 GANTRY
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
+24 V (for DTR2) DTR2 CN2-3 (*) DTR2 CN2-6 24 V ± 0.5 V PS51
+5 V (for DTR2) DTR2 CN2-8 (*) DTR2 CN2-9 5.1 V ± 0.1 V PS52
+5V (for DAS) CIF CN4-3 CIF CN4-6 5.2 V ± 0.1 V PS53
+15 V (for DAS) CIF CN4-1 CIF CN4-4 +15.1 V ± 0.1 V PS54
-15 V (for DAS) CIF CN4-5 CIF CN4-2 -15.1 V ± 0.1 V PS55
+24 V (for DAS FAN) DAS FAN-1 DAS FAN-2 24 V ± 0.5 V PS58
Note (*): For accessing the DTR2 Board, remove the DAS right bottom cover.
1–8 GANTRY
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
Illustration 1–9 Power Supply Potentiometers (except for PS53 & 58)
ÄÄÄÄÄÄÄÄ
ÄÄÄÄÄÄÄÄ
ÄÄÄÄÄÄÄÄ
ÄÄÄÄÄÄÄÄ
ÄÄÄÄÄÄÄÄ
ÄÄÄÄÄÄÄÄ
ÄÄÄÄÄÄÄÄ
ÄÄÄÄÄÄÄÄ
ÄÄÄÄÄÄÄÄ
ÄÄÄÄÄÄÄÄ
ÄÄÄÄÄÄÄÄ
ÄÄÄÄÄÄÄÄ
ÄÄÄÄÄÄÄÄ
ÄÄÄÄÄÄÄÄ
Illustration 1–10 Power Supply Potentiometers (for PS53 & 58)
ÉÉÉÉÉÉ
ÉÉÉÉÉÉ
ÉÉÉÉ
ÉÉÉÉÉÉ
ÉÉÉÉÉÉÉ
ÉÉÉÉÉÉ
ÉÉÉÉÉÉÉ
ÉÉÉÉÉÉ
ÉÉÉÉÉÉÉ
ÉÉÉÉÉÉÉ
1–9 GANTRY
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
blank 2205009
1–10 GANTRY
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
a. Switch OFF the ‘Rotate’ and ‘XG’ switches at the Gantry rear base.
b. Remove the front, rear, and both side covers of the Gantry.
2. Verify that the clearance between each rotating component part and all fixed parts is more than 10mm, while
rotating the Gantry in the CW direction by hand.
2–1 GANTRY
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
Ensure proper clearance (10 mm) between the x–ray tube cables and the stationary frame/component.
Ensure clearance of 20 mm or more between the pump hose and the stationary frame/component.
2–2 GANTRY
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
2–3 GANTRY
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
ÂÂÂÂ
ÂÂ ÂÂÂÂ
ÂÂ ÂÂÂÂ
ÂÂÂÂ
ÂÂ
ÂÂ ÂÂÂ
ÂÂÂÂÂ
ÂÂ ÂÂÂÂÂ
ÂÂ
ÂÂÂÂÂ
ÂÂÂ
ÂÂÂÂÂ
g. Anode and cathode HV tanks and Top cover
2–4 GANTRY
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
Special Equipment
D Spring Balance (Fish Scale): See Illustration 2–8.
1. Preparation:
a. Switch OFF the ‘Rotate’ and ‘XG (Safety Loop)’ switches at the Gantry rear base.
a. Switch ON the Service SW of the SUB Board, and rotate the Gantry CW until the x–ray tube is in the home
position.
c. Check or set the main weight as follows. See also Illustration 2–7.
(Qj, Qj–Base, Zj)
MAIN REAR
2 kg x 6, 0.5 kg x 2, total 13 kg.
MAIN FRONT
20 kg x1, 5 kg x 2, 1 kg x 2, total 32 kg.
TABLE SIDE (Fixed weight; do not adjust)
MAIN REAR
2 kg x 3, 0.5 kg x 5, total 8.5 kg.
MAIN FRONT
20 kg x1, 5 kg x 2, 1 kg x 5, 0.5 kg x 2,
total 36 kg. (Fixed weight; do not adjust)
TABLE SIDE
(For Gantry 21231xx, 213796x only)
2–5 GANTRY
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
MAIN REAR
3.6 kg x 1 (P/N: 2119818), total 3.6 kg.
MAIN FRONT
20 kg x1, 1 kg x 4, 0.5 kg x 4, total 26 kg.
(Fixed weight; do not adjust)
X–ray
Tube
SIDE WEIGHT
MAIN
WEIGHT
65_
2–6 GANTRY
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
b. Rotate the Gantry CW until the side weight is in the bottom position.
d. Check or set the side weight as follows. See also Illustration 2–7.
(Qj, Qj–Base, Zj)
SIDE REAR
2 kg x 1, 0.5 kg x 8, total 6 kg.
(Fixed weight; do not adjust)
SIDE FRONT
5 kg x 1, 2 kg x 2, total 9 kg.
SIDE REAR
2 kg x 2, 1 kg x 1, 0.5 kg x 1, total 5.5 kg.
(Fixed weight; do not adjust)
SIDE FRONT
5 kg x 1, 2 kg x 1, 1 kg x 3, 0.5 kg x 2,
total 11 kg.
(For Gantry 21231xx, 213796x only)
(Qj–Hlt, Zj–Adv)
SIDE REAR
2 kg x 1, 0.5 kg x 8, total 6 kg.
(Fixed weight; do not adjust)
SIDE FRONT
5 kg x 1, 2 kg x 1, 0.5 kg x 5, total 9.5 kg.
2–7 GANTRY
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
Rotate the Gantry CW until the x–ray tube is in the home position.
b. Verify that the Gantry does not rotate due to any unbalance.
c. Rotate the Gantry CW until the side weight is in the bottom position.
d. Verify that the Gantry does not rotate due to any unbalance.
a. Measure the force required to initiate Gantry rotation by pulling the Gantry with a spring weight balance as
shown in Illustration 2–8. See Illustration 2–9 for the measurement/pulling positions.
Measure force values F1, F2, F3, and F4 (all in kg); and enter these values.
2–8 GANTRY
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
Adjustment required for the main rear weight = (0.552 x (F4 – F3)) [kg]
Adjustment required for the side front weight = (0.515 x (F1 – F2)) [kg]
c. Increase or decrease the main rear weight and/or side front weight, according to the calculated values at
step b. Note the following:
D If the sign of the value is positive, increase the weight, if the sign is negative, decrease the weight.
D The minimum provided weight increment is 0.5 kg. Round to the nearest multiple of 0.5 kg as shown be-
low.
ACTUAL
ADJUSTMENT (kg) 2.0
1.0
-2.0
2–9 GANTRY
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
F1
F2
SIDE WEIGHT
MAIN WEIGHT
65_
F3
F4
MAIN WEIGHT
65_
SIDE WEIGHT
2–10 GANTRY
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
a. Switch OFF the ‘Rotate’ and ‘XG’ switches at the Gantry rear base.
2. Rotate the Gantry in the CW direction by hand until both LEDs on the G. pulse sensors turn OFF (one on the
stationary side and the other on the rotational side). See Illustration 2–10.
3. Measure the angle of the x–ray tube mounting plate. Verify that the angle is 0 ±0.5 deg.
c. Adjust the positions of the photo sensors so that both LEDs on the G. pulse sensors turn OFF (one on the
stationary side and the other on the rotational side); in this adjustment, the conditions described in Illustrations
2–11 and 2–12 should be met.
APPROXIMATE LOCATION
OF G. PULSE SENSOR
SLIP RINGS PHOTO SENSOR (FOR
ROTATION COUNT)
PHOTO PLATE
PHOTO SENSOR
(FOR G. PULSE)
LED
(VIEWED FROM
OF GANTRY)
2–11 GANTRY
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
PHOTO SENSOR
PHOTO PLATE
6 mm ±0.5 mm
(Min: 3 mm)
STATIONARY
SIDE
A = H/2 ±1.5 mm
A PHOTO PLATE
H H
A
(SIDE VIEW)
A = H/2 ±1.5 mm
ROTATIONAL
SIDE
2–12 GANTRY
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
a. Switch OFF the ‘Rotate’ and ‘XG’ switches at the Gantry rear base.
b. Loosen the set nut, and then rotate the adjusting bolt in the CCW direction until the distance D (between part A
and part B) is approximately 2 ∼ 3mm.
Over tension. After part A and part B are just touching, do not rotate the adjusting bolt; other-
wise, this will cause over tension and will shorten the life of the axial drive belt.
c. Turn the adjusting bolt in the CW direction until part A and part B are just touching.
2–13 GANTRY
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
ADJUSTING
BOLT
D JUST ZERO
(JUST TOUCH)
SET NUT B
A
D
2–14 GANTRY
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
b. Remove the front and both side covers from the Gantry.
c. Set the safety bracket located on the right side of the Gantry.
3–1 GANTRY
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 7 2205009
CAUTION
Gantry may experience sudden high speed rotation due to order from the OC. Be very care-
ful; stand clear of the rotating Gantry.
a. Be ready to hit the Emergency OFF switch in case of any rotational problems, and then, switch ON the Gantry
switches in the following sequence:
Note
The Gantry rotation alarm sounds indicate that Gantry rotation is enabled.
The Gantry starts rotating at 15 sec/Rot and stops at a position whose angle from Home is symmetrical to the
angle of the start position, after one or two rotations (during which Gantry Initialize Sequence is performed).
The Gantry Display shows the Test Sequence pattern. See Illustration 3–4.
During the Gantry Initialize Sequence, if the Gantry can not rotate (ex. TGP service switch (SYS/OFF/MNL) is
set to OFF), LEDs ‘ERRG’ and ‘LOOPG’ are lit on alternately on the TGP Board. (Waiting for Gantry rotation)
To release this status, push the reset button, after setting the service switches properly.
3–2 GANTRY
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
-
- - - - - - - - -
3–3 GANTRY
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
b. Remove the front, and both side covers from the Gantry.
ABT/SVE/FIX : SVE ↓
SPD2/SPD1/SPD0 : SPD0 ↓
FAST/MID/SLOW : SLOW ↓
CONT/HOME/90_ : CONT ↓
SYS/OFF/MNL : MNL
3. Check the Gantry rotation movement by setting all the possible speeds:
a. Set the ABT/SVE/FIX switch to ABT, and keep depressing it until the Gantry rotates for more than one rota-
tion.
Verify that the Gantry rotates smoothly, and that there is no abnormal noise.
3–4 GANTRY
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 2 2205009
2. Enter the TGP off–line function #400 (Gantry azimuth position display mode):
a. Set the G6 Gantry dip switch (location 7S) to ON on the TGP Board.
c. Verify that the Gantry Display panel displays the test number 400 and the azimuth position of the Gantry.
(ex. 180 indicates the home position, 270 indicates the 90_ CW position from home position)
ABT/SVE/FIX : SVE ↓
SPD2/SPD1/SPD0 : SPD1 ↓
FAST/MID/SLOW : MID ↓
CONT/HOME/90_ : CONT ↓
SYS/OFF/MNL : MNL
After one minute, verify that the Gantry Display panel displays exactly 180 (the number is in degrees); other-
wise, perform the following:
i. If the Gantry drifts in the CCW direction, adjust VZERO in the CW direction (approx. 32 degrees/turn of
VZERO).
If the Gantry drifts in the CW direction, adjust VZERO in the CCW direction.
(VZERO: location 10M on the TGP Board)
4. Set the G6 Gantry dip switch (location 7S) to OFF on the TGP Board.
3–5 GANTRY
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 2 2205009
1. Preparation:
b. Switch ON the ‘Rotate’, ‘Control’ and ‘Slip Ring 115V’ switches at the Gantry rear base.
3. Enter the TGP off–line function #401 (Gantry rotation speed display mode):
a. Set the G6 Gantry dip switch (location 7S) to ON on the TGP Board.
The Gantry rotates to the home position, and the Gantry Display panel displays a test number 400.
c. Press the test switch once on the TGP Board. The Gantry Display will change from 400 to 401.
(401: Gantry rotation speed display mode)
4. Rotate the Gantry at a specified speed and check/adjust the Gantry rotation speed displayed on the the Gantry
Display:
a. Set the TGP service switches as follows to rotate the Gantry (2.0 sec/Rot).
ABT/SVE/FIX : FIX ↓
SPD2/SPD1/SPD0 : SPD1 ↓
FAST/MID/SLOW : FAST ↓
CONT/HOME/90_ : CONT ↓
SYS/OFF/MNL : MNL
b. After Gantry rotation is stabilized (a few rotations are required), verify that the speed value displayed is within
the following range:
200 ±2
Note
If the specification above can not be achieved, adjust the potentiometer VR2 of the Servo Amp. How-
ever, care must be taken not to rotate too much.
3–6 GANTRY
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
7. Set the G6 Gantry dip switch (location 7S) to OFF on the TGP Board.
3–7 GANTRY
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
1. Preparation:
b. Switch ON the ‘Rotate’, ‘Control’ and ‘Slip Ring 115V’ switches at the Gantry rear base.
3–8 GANTRY
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
Deceleration Check/Adjustment
3. Rotate the Gantry at a specified speed and check/adjust the Gantry rotation deceleration:
a. Set the TGP service switches as follows to rotate the Gantry (2.0 sec/Rot).
ABT/SVE/FIX : FIX ↓
SPD2/SPD1/SPD0 : SPD1 ↓
FAST/MID/SLOW : FAST ↓
CONT/HOME/90_ : CONT ↓
SYS/OFF/MNL : MNL
b. After Gantry rotation is stabilized (a few rotations are required), set the CONT/HOME/90_ switch to HOME to
stop the Gantry.
Repeat steps a and b until achieving the shortest stop time without an overshoot.
3–9 GANTRY
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
blank 2205009
3–10 GANTRY
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
Note
For the slip rings and their associated components (brushes, etc.) information, see the ‘CT Slip Ring
Maintenance Manual (2154943).’
4–1 GANTRY
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
blank 2205009
4–2 GANTRY
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
3. Return the cradle to the furthest (Out) position from the Gantry.
4. Set the Table height to approximately ‘–100 mm’ according to the Gantry Display.
Verify that the Gantry tilts to the FWD 25.5 deg. position safely.
6. Rotate the Gantry while in the FWD 25.5 deg. tilt position by setting the TGP Service switches as follows:
ABT/SVE/FIX : FIX ↓
SPD2/SPD1/SPD0 : SPD2 ↓
FAST/MID/SLOW : MID ↓ (1.0 sec/Rot)
CONT/HOME/90_ : CONT ↓
SYS/OFF/MNL : MNL
Verify that the Gantry tilts to the BWD 25.5 deg. position safely.
9. Rotate the Gantry while in the BWD 25.5 deg. tilt position by setting the TGP Service switches as follows:
ABT/SVE/FIX : FIX ↓
SPD2/SPD1/SPD0 : SPD2 ↓
FAST/MID/SLOW : MID ↓ (1.0 sec/Rot)
CONT/HOME/90_ : CONT ↓
SYS/OFF/MNL : MNL
5–1 GANTRY
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 7 2205009
1. Preparation:
b. Switch OFF the ‘Rotate’ and ‘XG’ switches at the Gantry rear base.
c. Remove the front, rear, and both side covers from the Gantry.
e. Set the TGP service switches ‘ABT/SVE/FIX’, ‘CONT/HOME/90°’, and ‘SYS/OFF/MNL’ to FIX, HOME, and
MNL, respectively, in this order, to position the Gantry to the home position (x–ray tube at the 12 0’clock posi-
tion). Then set ‘SYS/OFF/MNL’ to SYS.
g. Set the Table height to approximately ‘–100 mm’ according to the Gantry Display.
2. Verify that the four tilt guide rollers can be freely rotated by hand with a gap of approximately 0.5 mm at every
tilt angle from FWD 25_ to BWD 25_.
If the rollers are not free at every angle, perform the following adjustment.
b. Loosen the main nuts of each guide roller, except for the front left guide roller, which is not adjustable.
(Three of the guide rollers are located on adjustment slots. The front left guide roller is fixed and not adjust-
able.)
c. Manually move the complete Gantry tilting frame assy, towards, or, away from the front left guide roller, such
that the arched frame just touches the front left guide roller, but also such that the front left guide roller can still
be rotated by hand.
d. Adjust the three guide rollers (front right, rear right and rear left) such that the gap between the rollers and
arched frames is approximately 0.5 mm.
g. Verify that the rollers can be rotated by hand at every tilt angle.
5–2 GANTRY
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
5–3 GANTRY
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 7 2205009
1. Preparation:
b. Switch OFF the ‘Rotate’ and ‘XG’ switches at the Gantry rear base.
c. Remove the front, rear, and both side covers from the Gantry.
e. Set the TGP service switches ‘ABT/SVE/FIX’, ‘CONT/HOME/90°’, and ‘SYS/OFF/MNL’ to FIX, HOME, and
MNL, respectively, in this order, to position the Gantry to the home position (x–ray tube at the 12 0’clock posi-
tion). Then set ‘SYS/OFF/MNL’ to SYS.
g. Set the Table height to approximately ‘–100 mm’ according to the Gantry Display.
2. Verify that the four tilt rollers under the arched frames rotate along with the frames during tilt operations.
a. Loosen the upper roller Lock Nuts (A) on both sides of the Gantry.
The Lock Nut is accessed from the outside the Gantry.
b. Loosen the set nuts for the Adjusting Bolts (E) on both sides of the Gantry, and then loosen the Adjusting Bolts.
c. Loosen the Lock Nut (B) on the left side only; do not loosen it too much.
The Lock Nut is accessed from the outside the Gantry.
d. Loosen the set nut and the Adjusting Bolt (D) on the left side.
5–4 GANTRY
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
g. Tighten the upper roller Lock Nuts (A) slightly on both sides of the Gantry.
h. Tighten the Adjusting Bolts (E) on both sides of the Gantry until you can barely rotate the upper rollers by hand.
i. Tighten the set nuts for (D) and (E) on both sides of the Gantry.
j. Tighten the upper roller Lock Nuts (A) on both sides of the Gantry.
5–5 GANTRY
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
2. Set the Table height to approximately ‘–100 mm’ according to the Gantry Display.
3. Set the Gantry to the BWD 25 deg. position using the ‘BWD’ button on the Gantry Panel.
4. Measure the time from BWD 25 deg. to FWD 25 deg., and from FWD 25 deg. to BWD 25 deg. Verify that their
values are in the following specified ranges:
±
Note
Actually, the tilt operation stops at 0 deg. when you tilt the Gantry from BWD 25 deg. to FWD 25 deg.,
or from FWD 25 deg. to BWD 25 deg. For this reason, measure the time in the following way:
5–6 GANTRY
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
a. Switch OFF the ‘Rotate’ and ‘XG’ switches at the Gantry rear base.
d. Rotate the Gantry in the CW direction by hand until the x–ray tube is positioned at approximately three o’clock
position.
f. Remove the Gantry inner bottom meshed plate with its stays, by removing the six hex screws. See Illustration
5–4.
ÉÉÉÉ
ÉÉÉÉ
ÉÉÉÉ
5–7 GANTRY
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
i. Rotate the four adjusters CCW to the adjuster intervals shown in Table below.
± ± ± ±
' #)"&' %)! ( #)"& $ #(&*!' &$" (
&( $# ! " (
5–8 GANTRY
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
Overtension. When changing the adjuster positions, make sure that you shift the right and
left adjusters the same distance and direction. If not, the Gantry right and left frame applied
torque could become unbalanced and the frame could be bent.
If the time is less than 27 sec, rotate the LEFT–FWD and RIGHT–FWD adjusters for 1.0 scale interval in the
CW (S) direction.
If the time is more than 33 sec, rotate the LEFT–FWD and RIGHT–FWD adjusters for 1.0 scale interval in the
CCW (F) direction.
If the time is less than 27 sec, rotate the LEFT–BWD and RIGHT–BWD adjusters for 1/5 of a scale interval in
the CW (S) direction.
If the time is more than 33 sec, rotate the LEFT–BWD and RIGHT–BWD adjusters for 1/5 of a scale interval in
the CCW (F) direction.
10. Install the Gantry inner bottom meshed plate, 6 hex screws.
5–9 GANTRY
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
Set the Table height to approximately ‘–100 mm’ according to the Gantry Display.
2. Press the ‘FWD’ button continually until the Gantry tilt to the +25 deg. position.
Verify that the tilt angle is 25 ±1 deg. and that the Tilt Display shows ‘FWD 25.0’.
4. Continually press ‘BWD’ until the Gantry stops. Verify that the Gantry stops at the 0_ tilt position and the Tilt
Display shows ‘0’.
5. Free the ‘BWD’ button, and then press it again until the Gantry tilts to –25 deg. position.
Verify that the tilt angle is –25 ±1 deg. and the Tilt Display shows ‘BWD 25.0’.
8. Enter the TGP off–line function #002 (Gantry tilt angle display mode):
Tilt 002
Position 255 (tilt angle; example only)
(Angle Display: 0.1_/digit
5: FWD 25 deg.
255: 0 deg.
505: BWD 25 deg.)
5–10 GANTRY
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
If the FWD tilt limit is reached (the Gantry tilt reads ‘0’ and stops) before the Gantry is tilted FWD 25 deg., turn
variable register T0 on the TGP Board in the CW direction to increase the tilt reading, and then continue to tilt
the Gantry to FWD 25 ±0.2 deg.
b. When the Gantry is tilted FWD 25 ±0.2 deg., adjust T0 to read ‘5’ on the Gantry Display.
If the BWD tilt limit is reached (the Gantry tilt reads ‘505’ and stops) before the Gantry is tilted BWD 25 deg.,
turn variable register TG on the TGP Board in the CCW direction to decrease the tilt reading, and then contin-
ue to tilt the Gantry to BWD 25 ±0.2 deg.
b. When the Gantry is tilted BWD 25 ±0.2 deg., adjust TG to read ‘505’ on the Gantry Display.
11. Set the T6 Table dip switch to OFF on the TGP Board.
13. Verify that the angle is ± 0.5 deg. measured using a clinometer.
5–11 GANTRY
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
blank 2205009
5–12 GANTRY
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
SECTION 6 – COLLIMATOR
!( **!&%(- ) " $ ( &+)
!( **!&%(- ) " $ ( &+)
!( **!&%(- ) " $ ( &+)
!( **!&%(- ) " $ ( &+)
!( **!&%(- ) " $ ( &+)
!( **!&%(- ) " $ ( &+)
&,*! !#*( %)*##
&* &( )-)*$) , (
$$ * !"%)) % )#* &%#- !) %&* ''#!) *& )*') .
#&,
a. Insert the BOOT floppy into the disk drive. (This step is not required for systems installed in Japan)
1. CALIBRATION
Select No. = ‘CR’
6–1 GANTRY
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
d. Select as follows:
OFFSET FILE CORRECTION : (Yes)
REFERENCE CH. CORRECTION : (No)
NATURAL LOG. CONVERSION : (No)
6–2 GANTRY
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
6–3 GANTRY
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
Enter 7 for DATA. (The mean file No. 7 is located on the seventh line)
Enter 0.16 for DATA MAX. (The specification for the mean file No. 7 is 0.10 ∼ 0.16)
f. Perform the steps below to show data for Channels beyond 500.
6–4 GANTRY
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
0.16
0.15
0.14
0.13
0.12
0.11
100.000 400.000
0.16
0.15
0.14
0.13
0.12
0.11
600.000
6–5 GANTRY
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
6–6 GANTRY
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
General Information
! ! ! $
D The DAS/Detector has been powered ON for more than two hours.
D The x–ray alignment (‘Plane of Rotation’, ‘Detector Bean–on–Window’, ‘Iso Center Alignment’, and ‘Filter
Center Alignment’) is proper.
!
! !
! " !! "
! ! ! ! ! $ ! ! %!& #& ! ! !'
$ "& !! & %& ! $ " ! !
1. Preparation:
2. Perform two air stationary scans with the filter installed and removed:
a. Insert the BOOT floppy into the disk drive. (This step is not required for systems installed in Japan)
1. CALIBRATION
.
.
Select No. = $$ ‘CR’
6–7 GANTRY
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
Scan 1 Scan 2
SCAN TIME : 3.0 sec SCAN TIME : 3.0 sec
THICKNESS : 10 mm THICKNESS : 10 mm
FOCUS : LARGE FOCUS : LARGE
KV : 120 kV KV : 120 kV
MA : 60 mA MA : 60 mA
SERIES NO. : 1 SERIES NO. : 2
(Others) : (defaults) (Others) : (defaults)
3. Obtain the following mean files 1 and 2 from the acquired raw data, by repeating the steps below.
d. Select as follows;
OFFSET FILE CORRECTION : (Yes)
REFERENCE CH. CORRECTION : (6ch)
NATURAL LOG. CONVERSION : (No)
6–8 GANTRY
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
Enter 3 for DATA. (The filter curve mean file is located on the third line)
6–9 GANTRY
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
Enter 3 for DATA. (The filter curve mean file is located on the third line)
6–10 GANTRY
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
6–11 GANTRY
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
blank 2205009
6–12 GANTRY
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
GANTRY
TABLE
1. Preparation:
c. Remove the front and both side covers from the Gantry.
7–1 GANTRY
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
WARNING!
ROTATION HAZARD! WHEN YOU PRESS THE ‘POSN LIGHT’ BUTTONS (LARGE BUTTON
AND SMALL BUTTON), THE GANTRY QUICKLY ROTATES TO THE 0 DEG. POSITION. IF IT IS
NECESSARY TO PLACE ANY PART OF YOUR BODY INSIDE OR NEAR THE GANTRY, SWITCH
OFF THE ‘ROTATE’ SWITCH FIRST.
c. Advance the cradle into the Gantry until the internal light cuts the approximately middle section of the head
holder.
d. Perform an axial scan of the head holder using the following technique:
7–2 GANTRY
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
7–3 GANTRY
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
0cm
L R
80 kV
60 mA
3.0 sec
35-CAL
0 cm
L R
80 kV
60 mA
3.0 sec
35-CAL
cm
L R
80 kV
60 mA <1> D= +11.07
3.0 sec H= 0.0d
35-CAL V= 89.0d
7–4 GANTRY
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
0c
m
L R
<1> D= +11.07
80 kV H= 0.0d
60 mA V= 89.0d
3.0 sec
35-CAL
L R
<2> D= +11.86
80 kV H= 0.0d
60 mA V= -89.0d
3.0 sec <1> D= +10.07
35-CAL H= 0.0d
V= 89.0d
| (1) – (2) | ≤ 2 mm
7–5 GANTRY
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
a. Switch OFF the ‘Rotate’ and ‘Table/Tilt’ switches at the Gantry rear base.
d. Move the Table to center the head holder in the scan plane, according to step 3.
g. Return to step 2.
7–6 GANTRY
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
GANTRY
GANTRY
TABLE TABLE
The positions of the Inner right and left lights, and Outer right and left lights are adjusted in the horizontal and vertical
directions.
The position of the Outer top light is adjusted in the right–left and horizontal directions.
7–7 GANTRY
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
WARNING!
ROTATION HAZARD! WHEN YOU PRESS THE ‘POSN LIGHT’ BUTTONS (LARGE BUTTON
AND SMALL BUTTON), THE GANTRY QUICKLY ROTATES TO THE 0 DEG. POSITION. IF IT IS
NECESSARY TO PLACE ANY PART OF YOUR BODY INSIDE OR NEAR THE GANTRY, SWITCH
OFF THE ‘ROTATE’ SWITCH FIRST.
2. Place a fine pin (ex. M3 size hexagonal wrench) on the center of the phantom holder. See Illustration 7–4.
7–8 GANTRY
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
7–9 GANTRY
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
b. Display a grid.
c. Measure the distance between the pin and the scan center in the following way:
ii. Move the cursor using the trackball to the center of the pin, and press the [Dist Angle] key to copy the
cursor over the pin location.
iii. Move the cursor vertically to the X axis; press [Dist Angle] again, to deposit the end location and display
the distance between the pin and the X axis (A). See Illustration 7–6.
iv. Perform the above to measure the distance between the pin and the Y axis (B).
d. Position the pin so that the pin is placed at the scan center, according to the distances measured at the above
step.
Use the up–down adjuster and right–left adjuster of the phantom holder (one turn: 1 mm movement). (The pin
direction is adjusted parallel with the Gantry axial direction, per step 3)
e. Repeat steps a through d, until the pin is positioned at the scan center within ±0.5 mm.
7–10 GANTRY
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
7. Adjust the inner right and left lights in the vertical direction using the ‘A’ screw so that the horizontal light and the
pin are aligned within ±0.5 mm. See Illustration 7–9.
Adjust the light angle using the ‘D’ lever, if necessary.
9. Adjust the outer right and left lights in the vertical direction using the ‘A’ screw so that the horizontal light and the
pin are aligned within ±0.5 mm.
Adjust the light angle using the ‘D’ lever, if necessary.
b. Verify that the horizontal line and the phantom surface are parallel. See Illustration 7–7.
VERIFY THAT
THESE TWO
LINES ARE
PARALLEL.
7–11 GANTRY
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
c. Adjust the direction of the phantom holder, using the front surface angle adjuster (see Illustration 7–4, accord-
ing to the image, so that the phantom surface is parallel with the vertical line.
d. Repeat steps b and c, until the line and the phantom surface are parallel.
e. Move the vertical line until it just touches the surface of the phantom scout image. See Illustration 7–8.
h. Select ‘Confirm’ on the Touch Panel screen and press the ‘Table set’ button.
The phantom surface is moved into the scan plane position.
VERTICAL LINE
7–12 GANTRY
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
14. Switch OFF the ‘Rotate’ switch at the Gantry rear base.
15. Adjust the inner right and left lights in the horizontal direction using the ‘B’ screw so that the vertical light and the
phantom surface are aligned within ±0.5 mm.
Adjust the light angle using the ‘D’ lever, if necessary.
16. Move the cradle in the Out direction over 175 mm.
17. Adjust the outer right and left lights in the horizontal direction using the ‘B’ screw so that the vertical light and the
phantom surface are aligned within ±0.5 mm.
Adjust the light angle using the ‘D’ lever, if necessary.
7–13 GANTRY
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
STEEL
A PLATE
ADJUSTING B
a
TUBE
b
2
1
2’
3'
1' 1
3'
D
3 d (A hex screw is installed inside this hole)
LIGHT
ADJUST ADJUST
2 SCREEN 1' DIRECTION SCREW/LEVER
1 - 1' A
2' 2 - 2' B
3 - 3' D
7–14 GANTRY
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
Move the cradle to the furthest (Out) position from the Gantry.
2. Push the ‘Down’ button until the Table reaches its lowest position.
Verify that the ‘cradle longitudinal position display’ changes to ‘Table height display’, and that the Table stops on
the way up with the ‘Table height display’ blinking.
Free the ‘UP’ button and hold it down again until the highest position is reached.
4. Return the Table to its lowest position using the ‘Down’ button.
6. Push the ‘Height’ button and verify that the ‘longitudinal display’ changes to ‘height display.’
2. Verify that the cradle moves In and Out smoothly when pressing the ‘In’ or ‘Out’ button on the Gantry Panel
Switch.
3. Press the ‘In’ or “Out’ button together with the ‘Fast’ button and verify that the cradle moves smoothly at approxi-
mately twice the speed as in step 2; check both In and Out.
2. Push the ‘Zero’ button. Verify that the ‘Table position’ display is ‘0.0’.
3. Push ‘Move to Scan’ and verify that the ‘In’ LED starts to blink.
4. Continually press the ‘In’ button until the cradle stops. Verify that the cradle moves 175 mm, and that the ‘Table
position’ display increments to “–175”, and then turns to ‘0.0’.
7–15 GANTRY
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
Verify that the Gantry rotates to the Home position and the 5 positioning lights are turned ON.
(The Gantry does not rotate, if it is already positioned at the Home position)
Verify that the positioning lights are turned OFF automatically after approximately 1 minute.
(If the positioning lights are not turned ON, push the reset switch on the TGP Board once and perform the above)
7-3-6 Home
1. Raise the Table to approximately –50 mm height.
Move the cradle to the 200 mm position in the In direction from the outer limit position.
7–16 GANTRY
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
Move the cradle to any position except the out limit position.
2. Move the Table down until it stops automatically; this is the interlock position.
Verify that the Table up/down function and Gantry tilt function are interlocked along the dashed line indicated in
Illustration 7–10, but that the cradle in/out function can still operate.
25
20
TILT ANGLE [deg.]
BWD
10
10
FWD
20 INTERLOCK AREA
25
7–17 GANTRY
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
blank 2205009
7–18 GANTRY
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
TABLE
TABLE OF CONTENTS
SECTION PAGE
i TABLE
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
blank 2205009
ii TABLE
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
1. Preparation:
a. Switch OFF the ‘Rotate’, ‘Control’, and ‘Table/Tilt’ switches at the Gantry rear base.
2. Set the T3 Table dip switch (location: 3S) on the TGP Board to ON.
(Cancels the Table–Gantry interlock function)
3. Press and hold the ‘Home Position’ button on the Gantry switch panel, and the following actions occur:
The Gantry tilts to 0 degrees.
The Table lowers to its lowest position.
The cradle moves to its furthest (Out) position from the Gantry.
Verify that the Table and Gantry motions are smooth and quiet.
Lowest Height
4. Verify that the Table height is more than 350 mm (more than 650 mm, for Compact Siting Mode 2; refer to Section
1-2, ‘Table Height’). See Illustration 1–1 for proper measuring points (measure ‘H’).
Release the ‘UP’ button and depress it again until the Table stops at its highest position.
1–1 TABLE
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
Highest Height
6. Verify that the Table height is more than 1000 mm.
NOTICE
Do not forget to switch OFF the T3 Table dip switch, otherwise there will be collision hazard, since
the interlock function does not work, if T3 is set to ON.
Note
After the above check procedure is finished and the front and rear bottom covers are re–installed,
verify that the Table up and down operations can be correctly performed by all the foot pedals each.
1–2 TABLE
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
a. Switch OFF the ‘Rotate’, ‘Control’, and ‘Table/Tilt’ switches at the Gantry rear base.
2. Enter the TGP off–line function #001 (Table height display mode):
a. Set the T6 Table dip switch (location: 3S) to ON on the TGP Board.
c. Push the test switch once after the Gantry Display panel displays a test number ‘000.’ The Gantry Display will
change from ‘000’ to ‘001.’
(001: Table height display mode)
1–3 TABLE
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
Tilt 001
Position 0520 (Table height; example only)
CASE A CASE B
((
((
If the Table can not be lowered down to 347 ~ 360 mm due to the interlock function, switch ON (up) the T3
Table dip switch to disable the interlock function, and continue to lower the Table. In this case, be careful not
to have the Table collide against anything (including the Table mechanical limit).
c. Adjust variable register H0 on the TGP Board so the Gantry Display XXXX equals (H – 345) X 2.
1–4 TABLE
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
If the Table can not be raised up to 940 ~ 953 mm due to the interlock function, switch ON (up) the T3 Table dip
switch to disable the interlock function, and continue to raise the Table. In this case, be careful not to have
the Table collide against anything (including the Table mechanical limit).
If the values of the Gantry Display are out of the specified range, return to step 3.
1–5 TABLE
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
WARNING!
ROTATION HAZARD! WHEN YOU PRESS THE ‘POSN LIGHT’ BUTTONS (LARGE BUTTON
AND SMALL BUTTON), THE GANTRY QUICKLY ROTATES TO THE 0_ POSITION. IF IT IS
NECESSARY TO PLACE ANY PART OF YOUR BODY INSIDE OR NEAR THE GANTRY, SWITCH
OFF THE ‘SCAN’ SWITCH FIRST.
When the Table is raised or lowered, the cradle is automatically moved in the Out or In direction to remain at the
same distance from the Gantry. The following procedure checks/adjusts this function.
a. Raise the Table until the Gantry Display displays approximately 1000 mm.
c. Mark the cradle longitudinal position; for example, attach tape on the cradle and mark a line on it, aligning with
positioning lights. See Illustration 1–2.
d. Raise the Table until the Gantry Display displays approximately ‘1200’ (raise the Table by 10 cm); or
Lower the Table until the Gantry Display displays approximately ‘800’ (lower the Table by 10 cm).
e. Verify that the longitudinal correction error falls within ±1 mm, that is, the mark line and the positioning light
align within ±1 mm.
Or;
When the Table is raised, if the cradle is moved in the In direction for 1 mm or more excess over the
correct amount (in this case, when the Table is lowered, the cradle will be moved in the Out direction for 1
mm or more excess over the correct amount);
turn VR HG a little in the CW direction.
ii. Raise or lower the Table for some height (10 mm, 20 mm, or any).
(This operation is required prior to repeating the above check)
1–6 TABLE
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
7. Set the T6 (and T3) Table dip switch(es) on the TGP Board to OFF (down position).
NOTICE
Do not forget to switch OFF the T3 Table dip switch, otherwise there will be collision hazard, since
the interlock function does not work, if T3 is set to ON.
Note
After the above check/adjustment procedure is finished and the front and rear bottom covers are re–
installed, verify that the Table up and down operations can be correctly performed by all the foot ped-
als each.
1–7 TABLE
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
a. Switch OFF the ‘Control’ and ‘Table/Tilt’ switches at the Gantry rear base.
2. Enter the TGP off–line function #001 (Table height display mode):
a. Set the T6 Table dip switch (location: 3S) to ON on the TGP Board.
c. Push the test switch once after the Gantry Display panel displays a test number ‘000.’ The Gantry Display will
change from ‘000’ to ‘001.’
(001: Table height display mode)
Tilt 001
Position 0520 (Table height; example only)
If the Table can not be lowered down to 645 ~ 655 mm due to the interlock function, switch ON (up) the T3
Table dip switch to disable the interlock function, and continue to lower the Table. In this case, be careful not
to have the Table collide against anything (including the Table mechanical limit).
c. Adjust variable register H0 on the TGP Board so the Gantry Display XXXX equals (H – 345) X 2. (This adjust-
ment is temporary; H0 will be re–adjusted later)
1–8 TABLE
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
If the Table can not be raised up to 895 ~ 905 mm due to the interlock function, switch ON (up) the T3 Table dip
switch to disable the interlock function, and continue to raise the Table. In this case, be careful not to have
the Table collide against anything (including the Table mechanical limit).
If the values of the Gantry Display are out of the specified range, return to step 3.
1–9 TABLE
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
WARNING!
ROTATION HAZARD! WHEN YOU PRESS THE ‘POSN LIGHT’ BUTTONS (LARGE BUTTON
AND SMALL BUTTON), THE GANTRY QUICKLY ROTATES TO THE 0_ POSITION. IF IT IS
NECESSARY TO PLACE ANY PART OF YOUR BODY INSIDE OR NEAR THE GANTRY, SWITCH
OFF THE ‘SCAN’ SWITCH FIRST.
When the Table is raised or lowered, the cradle is automatically moved in the Out or In direction to remain at the
same distance from the Gantry. The following procedure checks/adjusts this function.
a. Raise the Table until the Gantry Display displays approximately 1000 mm.
c. Mark the cradle longitudinal position; for example, attach tape on the cradle and mark a line on it aligning with
positioning lights. See Illustration 1–2.
d. Raise the Table until the Gantry Display displays approximately ‘1200’ (raise the Table by 10 cm); or
Lower the Table until the Gantry Display displays approximately ‘800’ (lower the Table by 10 cm).
e. Verify that the longitudinal correction error falls within ±1 mm, that is, the mark line and the positioning light
align within ±1 mm.
Or;
When the Table is raised, if the cradle is moved in the In direction for 1 mm or more excess over the
correct amount (in this case, when the Table is lowered, the cradle will be moved in the Out direction for 1
mm or more excess over the correct amount);
turn VR HG a little in the CW direction.
ii. Raise or lower the Table for some height (10 mm, 20 mm, or any).
(This operation is required prior to repeating the above check)
1–10 TABLE
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
NOTICE
Do not forget to switch OFF the T3 Table dip switch, otherwise there will be collision hazard, since
the interlock function does not work, if T3 is set to ON.
b. During lowering the Table, verify that the Table does not hit any objects and stops at an appropriate position.
c. Raise the Table to a height between the Up limit position and –150 mm position.
d. During moving the cradle in the In direction, verify that the cradle does not hit any objects and stops at an
appropriate position.
Note
After the above check/adjustment procedure is finished and the front and rear bottom covers are re–
installed, verify that the Table up and down operations can be correctly performed by all the foot ped-
als each.
1–11 TABLE
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
blank 2205009
1–12 TABLE
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
1. Move the cradle to the Out limit position by the ‘OUT’ button.
2. While moving the cradle in the In direction by continuously pushing the ‘IN’ button, perform the following:
Verify that the cradle stops moving, and that the Gantry Display goes blank; this indicating that the cradle is
unlatched.
Verify that the cradle starts moving and the Gantry Display changes to normal (displaying a position in millime-
ter).
3. Repeat the above same check for the Unlatch switch on the opposite side.
2–1 TABLE
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
4. Adjust the position of the bearings to meet the following conditions. See Illustration 2–2.
D The Table frame and the cradle are parallel; adjust so that the difference between gap DF (measured on
the In side) and gap DR (measured on the Out side) becomes 1.5 mm or less.
D A gap of 0.1 mm or more (‘A’) exists between the cradle and the bearing (at least one of the right and left
bearings); however, the total of the gaps on the right and left sides should be 2.5 mm or less.
2–2 TABLE
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
3. Observe the gap between the cradle and the cradle support bearing (see Illustration 2–3) while moving the cradle
with no load over the entire travel (In $ Out) by hand, and find out a travel point where the gap is the narrowest
and stop the cradle there.
4. Adjust the gap between the cradle and the cradle support bearing by the nuts, so that the gap becomes 0.2 mm or
less (however, it should be > 0).
5. Verify that the cradle does not touch the cradle support bearing while moving the cradle with no load over the
entire travel (In $ Out).
2–3 TABLE
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
3. Adjust the gap between the cradle and the table covers by loosening the cover attaching bolts, so that the gap
becomes 7 mm or less (however, it should be > 0). See Illustration 2–4.
2–4 TABLE
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
2. The cradle has the side play. Move the cradle to the left all the way by hand, and then check the gap between the
cradle and the sponge. It must be within the specifications below:
Approx. 5.5 mm
1 mm or less
3. Move the cradle to the right all the way by hand, and then check the gap between the cradle and the sponge. It
must be within the specifications in step 2 above. See Illustration 2–5.
4. Move the cradle IN or OUT, and then verify that both sponges should not be pulled by the cradle.
2–5 TABLE
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
3. Press the ‘Zero’ button, and verify that the Gantry Display indicates Zero.
4. Mark the Table cover with a marking tape or other at one end of the cradle. Define this position as a reference.
5. Move the cradle approximately 1000 mm (Display) by pushing the ‘In’ + ‘Fast’ buttons.
7. Perform the same check by returning the cradle in the Out direction approximately 1000 mm.
2–6 TABLE
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
2. Press the ‘Zero’ button, and verify that the Gantry Display indicates Zero.
3. Mark the Table cover with a marking tape or other at one end of the cradle. Define this position as a reference.
4. Move the cradle approximately 100 mm (Display) by pressing the ‘In’ button.
Repeat the above step (a) a total ten times, i.e., 1000 mm movement.
5. Return the cradle continuously without stopping until the Gantry Display shows ±1.0 (mm).
6. Verify that the cradle has returned to the reference point (per step 3). See below.
7. Move the cradle to the In limit position. Do the same check as above in the reverse direction.
2–7 TABLE
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
1. Preparation:
c. Set the Table height to more than –150 mm (on Gantry Display).
Adjustment at 5 mm
2. Set the cradle to the Out mechanical limit position.
Move the cradle in the In direction until the Gantry Display reads 5.0 (mm).
(If you can’t move the cradle with the switches, release the cradle and move it into position by hand)
3. Enter the TGP off–line function #000 (Cradle position display mode):
a. Set the T6 Table dip switch (location: 3S) to ON on the TGP Board.
Tilt 000
Position 11 (Cradle position; example only)
4. Adjust variable register P0 on the TGP Board so that the Gantry Display value equals 10.
5. Exit the TGP off–line function #000 by Setting the T6 Table dip switch to OFF on the TGP Board.
The Gantry Display will return to its normal status.
2–8 TABLE
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
Adjustment at 1525 mm
6. Move the cradle in the In direction until the Gantry Display reads 1520.0 (mm).
(If you can’t move the cradle with the switches, release the cradle and move it into position by hand)
8. Adjust variable register PG on the TGP Board so that the Gantry Display value equals 3050.
(= 1525.0 mm = 1520.0 + 5.0)
9. Exit the TGP off–line function #000 by Setting the T6 Table dip switch to OFF on the TGP Board.
The Gantry Display will return to its normal status.
2–9 TABLE
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
Special Equipment
D Spring Balance (Fish Scale): 20 kg maximum
2. Set the spring balance on the cradle end and cradle rail using wire. See Illustration 2–7.
3. Place a weight of approximately 20 kg at the front end of the cradle, or if such a thing is not available, ride on the
cradle, at location ‘A’, shown in Illustration 2–7.
This is to prevent the cradle roller from slipping which is equipped underneath the cradle at location ‘A’.
TABLE TOP VIEW
ÍÍÍ ÁÁ
ÍÍÍ ÁÁ
ÍÍÍ ÁÁ
ÁÁ
2–10 TABLE
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
c. Enter the TGP off–line function #019 (Cradle drive mode at specified speeds):
i. Set the T6 Table dip switch (location : 3S) to ON on the TGP Board.
iii. Push the test switch for a number of times until the Gantry Display shows ’ ’.
(This is ‘Cradle drive mode at specified speeds’)
iv. Push simultaneously both In and Fast, or both Out and Fast buttons so that the Gantry Display shows
’ ’.
(This sets the cradle speed to 0.333 mm/sec)
While the cradle is moving in the In direction, the motor will automatically stop at some position of the cradle;
read the spring balance at this position.
Verify that this value (maximum motor torque) is 17 ~ 18 kg at around 1000 mm from the Out limit position.
Note
Wait for the motor driver to cool before performing another motor torque measurement.
2–11 TABLE
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
NOTICE
Fragile component. Be careful not to break or bend the cradle position potentiometer cable
attached to the cradle end handle.
b. Remove the four 6 mm allen head screws from the cradle end handle, and remove the cradle.
(For details, refer to the Component Replacement manual)
c. Move the cradle end handle carefully to the Table end of the Gantry side.
d. Remove the screws which hold the upper Table covers together at the outer end of the Table, farthest from the
Gantry, and spread the covers slightly apart.
2–12 TABLE
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
2–13 TABLE
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
a. Verify that both of the latch switches for the cradle are set to ‘Latch.’
b. Connect a multimeter between the ‘CP+’ and ‘CP–’ terminals of the driver. See Illustration 2–8 or 2–9.
c. While moving the cradle at slow speed, turn the ‘RUN’ potentiometer appropriately to increase or decrease
the voltage between ‘CP+’ and ‘CP–’.
This voltage determines the maximum motor torque; to increase torque, increase the voltage.
(The standard set voltage is 7.43 V when the cradle is not moving)
Note
Since the ‘STOP’ potentiometer is factory set, do not adjust it.
d. Perform step 4.
7. Exit the TGP off–line function #019 by setting the T6 Table dip switch (location : 3S) to OFF on the TGP Board.
8. Shuttle the cradle over the entire travel both at the slow and fast speed, without the spring balance, to verify that
movement is normal.
2–14 TABLE
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
Purpose
This procedure checks whether the holders (head holder, coronal holder, etc.) properly fit on the cradle.
1. Verify that all the bracket bolts of the axial head holder are firmly installed; otherwise, tighten them. See Illustra-
tion 3–1.
3–1 TABLE
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
3. Try to move the holder by hand, back and forth, and upward, to check that the holder firmly fits on the cradle.
If the holder is moved (not secured) or taken out from the cradle, proceed to the next step; otherwise, go to step 7.
5. Turn the support screws (two) of the holder CW until the gap becomes almost zero (a slight gap should exist).
See Illustrations 3–1 and 3–2.
6. Return to step 2.
Phantom Holder
Coronal Holder
Cradle Extender
Others (if the customer has other type holders, such are Low Profile Head Holder, etc.)
ÂÂÂÂ
ÂÂÂÂ ÂÂÂÂ
ÂÂÂÂ ÂÂÂÂ
ÂÂÂÂ ÂÂÂÂ
ÂÂÂÂ ÂÂÂÂ
ÂÂÂÂ
ÂÂÂÂ
ÂÂÂÂ
ÂÂÂÂ
SUPPORT SCREW SUPPORT SCREW
3–2 TABLE
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
DAS/DETECTOR
TABLE OF CONTENTS
SECTION PAGE
i DAS/DETECTOR
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
blank 2205009
ii DAS/DETECTOR
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
CAUTION
The Detector Bias voltage is approximately 1000 VDC ! Do not touch conducting parts of the
voltage, so you do not receive a shock.
1. Wait approximately one minute after switching OFF the “Slip Ring 115V” switch for the bias
voltage to bleed off to a safe level.
2. Use a proper range on the digital volt meter.
1. Preparation:
a. Switch OFF the ‘Rotate’ and ‘Slip Ring 115V’ switches at the Gantry rear base.
d. Rotate the Gantry by hand until the DAS/Detector reaches the 1 o’clock position. See Illustration 1–1.
2. Disconnect the cable from the BNC connector of the bias voltage power supply. See Illustration 1–2.
1–1 DAS/DETECTOR
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
- MEASURE POINTS
(GROUND)
BIAS VOLTAGE
POWER SUPPLY
Wait approximately one minute for voltage to bleed off before reconnecting the BNC connector cable.
7. Connect the BNC connector cable to the bias voltage power supply.
1–2 DAS/DETECTOR
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
1. Preparation:
d. Rotate the Gantry by hand until the X–ray tube reaches the 11 o’clock position.
3. Set a clamp–on meter at near the TEMP CONT box to measure the AC current supplied to the transformer of
the TEMP CONT box.
4. Monitor the ‘D3’ LED on the TEMP CONT Board. See Illustration 1–4.
Verify that the current increases by approximately 1 A when the ‘D3’ LED lights ON.
(This indicates that the Detector heater is turned ON)
1–3 DAS/DETECTOR
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
1–4 DAS/DETECTOR
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
1. Preparation:
a. Switch OFF the ‘Rotate’ and ‘Slip Ring 115V’ switches at the Gantry rear base.
d. Rotate the Gantry by hand until the X–ray tube reaches the 11 o’clock position.
Disconnect CN3 on the TEMP CONT Board. See Illustrations 1–3 and 1–4.
4. Read the thermistor resistance value written on a tag. See Illustration 1–4. (Qj, Qj–Base, Zj)
5. Read the thermistor resistance value which is written on a paper sheet attached on the Detector surface near
the TEMP CONT box. (Qj–Hlt, Zj–Adv)
6. Connect a digital meter between TP1 and TP2 (resistance measurement mode).
7. Adjust RV1 until the resistance between TP1 and TP2 matches the resistance read at step 4.
Connect CN3.
1–5 DAS/DETECTOR
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
blank 2205009
1–6 DAS/DETECTOR
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
1. CALIBRATION
.
.
Select No. = ‘CR’
2–1 DAS/DETECTOR
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
Precondition
(Qj, Qj–Base, Zj)
(Qj–Hlt, Zj–Adv)
1. Perform an axial scan with no x–rays (i.e., acquire a raw data file consisting of offset data):
a. Insert the BOOT floppy into the disk drive. (This step is not required for systems installed in Japan)
AXIAL SCAN
SCAN TIME : 1.0 or 2.0 sec (if 1.0 sec is available, select 1.0 sec)
XG, ROTOR : ROTOR ON
SERIES NO. : 1
(Others) : (defaults)
2. Verify that MEAN MAX & MIN data are within specifications without applying Offset File Correction on the ac-
quired raw data (offset data):
d. Select as follows;
OFFSET FILE CORRECTION : (No)
REFERENCE CH. CORRECTION : (No)
NATURAL LOG. CONVERSION : (No)
2–2 DAS/DETECTOR
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
(Qj–Hlt, Zj–Adv)
** START
MAX MAX xxxxxxx
MIN xxxxxxx
AVE xxxxxxx
MIN MAX xxxxxxx
MIN xxxxxxx
AVE xxxxxxx
MEAN MAX xxxxxxx ≤ 6000
MIN xxxxxxx ≥ 3000
AVE xxxxxxx
S.D. MAX xxxxxxx
MIN xxxxxxx
AVE xxxxxxx
2–3 DAS/DETECTOR
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
d. Select as follows;
OFFSET FILE CORRECTION : (Yes)
REFERENCE CH. CORRECTION : (No)
NATURAL LOG. CONVERSION : (No)
e. Verify that MEAN MAX & MIN and S.D. MAX & MIN data are within specifications as described below:
(Qj, Qj–Base, Zj)
** START
MAX MAX xxxxxxx
MIN xxxxxxx
AVE xxxxxxx
MIN MAX xxxxxxx
MIN xxxxxxx
AVE xxxxxxx
MEAN MAX xxxxxxx ≤ 0.7
MIN xxxxxxx ≥ –0.7
AVE xxxxxxx
S.D. MAX xxxxxxx ≤ 3.5
MIN xxxxxx x ≥ –3.5
AVE xxxxxxx
(Qj–Hlt, Zj–Adv)
** START
MAX MAX xxxxxxx
MIN xxxxxxx
AVE xxxxxxx
2–4 DAS/DETECTOR
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
Precondition
Ensure that the DAS/Detector has been powered ON for at least two hours prior to performing the following proce-
dure.
a. Insert the BOOT floppy into the disk drive. (This step is not required for systems installed in Japan)
STATIONARY SCAN
SCAN TIME : 2.0 sec
XG, ROTOR : BOTH OFF
NO OF SCAN : 2
DAS TEST : LOW X 64
SERIES NO. : 1
(Others) : (defaults)
(Two scans are performed, the data of the first scan will not be used, since the second data is considered to be
more stabilized)
g. Repeat the above scans with the following parameters (the other parameters are the same).
2–5 DAS/DETECTOR
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
Obtain the following mean files 1 through 6, by repeating the steps below.
b. Enter 1.2 (2.2 ∼ 6.2) for the raw data file ID No.
d. Select as follows;
OFFSET FILE CORRECTION : (Yes)
REFERENCE CH. CORRECTION : (No)
NATURAL LOG. CONVERSION : (No)
2–6 DAS/DETECTOR
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
Obtain the following mean files NO. 7, 8, and 9, by repeating the steps below.
2–7 DAS/DETECTOR
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
g. Perform the same check as above for the mean files No. 8 and No. 9. (At steps c and f–ii, enter 8 or 9)
2–8 DAS/DETECTOR
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
2–9 DAS/DETECTOR
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
blank 2205009
2–10 DAS/DETECTOR
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 4 2205009
X–RAY GENERATOR
TABLE OF CONTENTS
SECTION PAGE
i X–RAY GENERATOR
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
blank 2205009
ii X–RAY GENERATOR
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
a. Insert the BOOT floppy into the disk drive. (This step is not required for systems installed in Japan)
AXIAL SCAN
SCAN TIME : 3.0 sec
THICKNESS : 10 mm
FOCUS : LARGE
KV : 120 kV
MA : 60 mA
SERIES NO. : 1
(Others) : (defaults)
d. Select as follows:
OFFSET FILE CORRECTION : (Yes)
REFERENCE CH. CORRECTION : (No)
NATURAL LOG. CONVERSION : (No)
e. Verify that MEAN MAX data are within specifications as described below:
(Qj, Qj–Base, Zj)
** START
MAX MAX xxxxxxx
MIN xxxxxxx
AVE xxxxxxx
MIN MAX xxxxxxx
MIN xxxxxxx
AVE xxxxxxx
(Qj–Hlt, Zj–Adv)
** START
MAX MAX xxxxxxx
MIN xxxxxxx
AVE xxxxxxx
MIN MAX xxxxxxx
MIN xxxxxxx
AVE xxxxxxx
a. Insert the BOOT floppy into the disk drive. (This step is not required for systems installed in Japan)
** TUBE CALIBRATION **
1. CHANGE PARAMETER
2. RESET PARAMETER
Select No.=_
** CHANGE PARAMETER **
<SELECT SCAN KV, MA, FOCUS>
80 kV 120 kV 140 kV
1: 60 mA 7: 60 mA 12: 60 mA
LARGE 2: 160 mA 8: 160 mA
3: 350 mA
4: 60 mA – – – – – –
SMALL 5: 130 mA
6: 250 mA
< x >
< x–y >
< x, y, z >
< kV >
< L > or < S >
Select No. <Enter to All>=_
(Depending on systems, 20 mA also is shown or is shown instead of 60 mA, for Large Focus)
4. Select KV–MA stations that you want to calibrate either way of the following:
D Enter a number range. For example, entering 2–4 selects KV–MA stations of (2: 80 kV–160 mA), (3: 80
kV–350 mA) and (4: 80 kV–60 mA) ( < x–y > )
D Enter a KV (80, 120, or 140) to select all the MA stations for the selected KV. ( < kV > )
D Select Large Focus or Small Focus to select all the KV–MA station for the selected focus.
( < L > or < S > )
Verify that the following message appears when all the calibrations you selected are completed.
** CHANGE PARAMETER **
PARAMETER STORE IN PROGRESS
** TUBE CALIBRATION **
1. CHANGE PARAMETER
2. RESET PARAMETER
Select No.=_
If you want to revert the calibration data to the original data which is stored in ROM, select RESET PARAMETER.
The protocol shown in Table 1–1 is used for the heat soak procedure.
Enter all the parameters either to a scan screen, or (register) to Preset Scan Parameters. If you register the
parameters to Preset Scan Parameters, you can use it later without having to enter all the parameters one by
one.
To display the registration screen, push the following buttons in the order written:
(Qj, Qj–Base, Qj–Hlt)
[Preset Prmtr] → SCAN PARAMETER → I–EXTR → MULTI/DYNAMIC Menu No. 20
(If No. 20 is already used by the customer, use an other No.)
(Zj–Adv, Zj)
[Maint] → SCAN PRESET PARAMETER → AuxB → Infant → Para. Set → Multi Scan → SGL/MLT Menu. No.
10
(If No. 10 is already used by the customer, use an other No.)
3. Verify that ‘Anode’ and ‘Case’ heat indicators on the screen are at 0%.
To terminate the scan sequence before completion while scanning, press the <Abort> key.
If the system detects XG Fail or other error condition, this procedure will be terminated.
To retry the heat soak procedure, wait until the Anode/Case Heat falls to 0%.
Note
Do not perform a warm–up or scans for at least ten minutes after completing this heat soak proce-
dure.
(continued)
1. Preparation:
a. Switch OFF the ‘Rotate’, ‘XG (Safety Loop)’, and ‘Slip Ring 115V’ switches at the Gantry rear base.
2. Tighten the screws which fasten the x–ray tube housing together. See Illustration 1–1.
The total number of screws is 16 (4 of which also fix the heat exchanger brackets to the housing).
Use the dedicated hexagonal wrench (P9180GW) provided with the system.
X-RAY TUBE
8 SCREWS 8 SCREWS
(AROUND THE (AROUND THE
CIRCUMFERENCE) CIRCUMFERENCE)
PUMP
HEAT EXCHANGER
SECTION 2 – KV AND MA
e. Wait five minutes or more following the last scan prior to proceeding to the next step.
Verify that the KV and MA values displayed on the Gantry Display is within the following specifications:
KV: ±5 kV
MA: ±20 mA
(KV and MA values are displayed in real–time during an exposure, and ‘0’ is displayed after the exposure is fin-
ished)
TIME FOCUS KV MA
2 SEC LARGE 80 60 100 160 250
120 60 130 200
140 60 100 160
SMALL 80 60 130 160 250
120 60 130 200
140 60 100 160
(ProSpeed SX, ProSpeed S Fast, ProSpeed, Lemage, Lemage SX/E, ProSpeed VX)
TIME FOCUS KV MA
2 SEC LARGE 80 60 100 160 250 350
120 60 100 160 250
140 60 130 200
SMALL 80 60 130 160 250
120 60 130 200
140 60 100 160
TIME FOCUS KV MA
2 SEC LARGE 80 60 100 160 250 350
120 60 130 200 300
140 60 100 160 250
SMALL 80 60 130 160 250 250
120 60 130 200
140 60 100 160
TIME FOCUS KV MA
1.5 SEC LARGE 80 60 100 160 250 350
120 60 100 160 250 350
140 60 130 200 300
SMALL 80 60 130 160 250 250
120 60 130 200
140 60 100 160
c. Remove the front , rear, and both side covers of the Gantry.
e. Rotate the Gantry to the home position (x–ray tube at the 12 o’clock position).
f. Switch OFF the ‘Rotate’, ‘XG (Safety Loop)’, and ‘Slip Ring 115V’ switches at the Gantry rear base.
Electric shock! Use a ground bar to discharge the high voltage connector any time you dis-
connect it.
b. Connect the HV divider cables, which are provided with the system, to the cathode and anode tanks.
(Before connecting the cables, wipe clean the HV connectors, and lubricate them with silicon oil)
Rotate the Gantry by hand in the clockwise direction until the x–ray is positioned at the 3 o’clock
position.
d. Remove the HV cable connectors from the x–ray tube. (Both anode and cathode connectors)
e. Connect the HV divider cables whose other ends are connected to the HV divider to the x–ray tube.
(Before connecting the cables, wipe clean the HV connectors, and lubricate them with silicon oil)
g. Rotate the Gantry by hand in the clockwise direction approximately to the 5 o’clock position.
CATHODE HV
TANK WITH
INVERTOR
ANODE HV
TANK WITH
INVERTOR
TO ANODE TO TUBE
HV TANK ANODE
TO CATHODE TO TUBE
HV TANK CATHODE
KV Read
3. Connect a digital multimeter to the output terminals of the HV divider.
(The multimeter should be in the DC voltage measurement mode)
Place the multimeter so that it can be read from the Operator’s room.
NOTICE
a. Switch ON the ‘Slip Ring 115V’ and ‘XG (Safety Loop)’ switches at the Gantry rear base.
KV MA KV SPECIFICATION
80 250 74.4 ≤ ≤ 85.6
120 200 111.6 ≤ ≤ 128.4
140 160 130.2 ≤ ≤ 149.8
Stationary Scan (Present Degree)
Scan Time: 3.0 sec
Thickness: 1 mm
Focus: Large
2-3 MA CHECK
1. Preparation:
b. Wait five minutes or more following the last scan prior to proceeding to the next step.
c. Switch OFF the ‘Rotate’, ‘XG (Safety Loop)’, and ‘Slip Ring 115V’ switches at the Gantry rear base.
f. Rotate the Gantry by hand in the clockwise direction until the x–ray is positioned at the 2 o’clock position.
BURDEN
RESISTOR
BOARD
MA Check
4. Set the multimeter to the DC voltage measurement mode (mV).
Place the multimeter so that it can be read from the Operator’s room.
NOTICE
a. Switch ON the ‘Slip Ring 115V’ and ‘XG (Safety Loop)’ switches at the Gantry rear base.
Verify that the MA measurements ([measured voltage (mV)]/[10 (ohms)]) are within the specifications.
KV MA MA SPECIFICATION
80 250 237.5 ≤ ≤ 262.5
120 200 190.0 ≤ ≤ 210.0
140 160 152.0 ≤ ≤ 168.0
Stationary Scan (Present Degree)
Scan Time: 3.0 sec
Thickness: 1 mm
Focus: Large
SECTION 3 – PDU
1. Switch OFF the power for the PDU at the Power Distribution Board.
2. Remove the top and front covers of the PDU. See Illustration 3–1.
GROUND
REMOVE FOUR(4) CABLE
HOIST RINGS.
REMOVE THE
ÅÅÅÅÅ TOP COVER. ÅÅÅÅÅÅ
ÅÅÅÅÅ ÅÅÅÅÅÅ
ÅÅÅÅÅ ÅÅÅÅÅÅ
: FRAME GROUND POINT
REMOVE THE
FRONT COVER.
4. Switch ON the power for the PDU at the Power Distribution Board.
5. Verify that the relays K1, K31, K32, K48 are all OFF.
6. Connect a digital volt meter to the test terminals (Range: 1000 VDC).
(For PDUs which are equipped with test terminals instead of a volt meter only)
WARNING!
ELECTROCUTION ! 550 VDC ∼ 750 VDC IS GENERATED FROM THE PDU AND APPLIED TO
THE GANTRY SLIP RINGS. KEEP AWAY FROM THE CONDUCTING PARTS.
Verify that approximately five sec later, relays K48 and K1 are turned ON.
Read the voltage on the volt meter of the PDU or the digital volt meter, and verify that it is within 550 ∼ 750 V.
ÅÅÅÅÅÅÅÅÅ
BP TEST
ÅÅÅÅÅÅÅÅÅ
BUTTON
ÅÅÅÅÅÅÅÅÅ
ÅÅÅÅÅÅÅÅÅ
RELAY K48
RELAY K1
ÅÅÅÅ ÅÅÅÅÅÅÅÅÅ
ÅÅÅÅÅÅÅÅÅ
ÅÅÅÅ
ÅÅÅÅ ÅÅÅÅÅÅÅÅÅ
ÅÅÅÅÅÅÅÅÅ
ÅÅÅÅÅÅÅÅÅ
ÅÅÅÅÅÅÅÅÅ
SW2
OR
TEST
VOLT METER
TERMINALS
5 V Power Supply
GND +5 V
ÁÁÁÁ
ÁÁÁÁ
Filament Board
Electric Rack
DC III
TABLE OF CONTENTS
SECTION PAGE
i DC III
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
blank 2205009
ii DC III
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
a. Connect a digital multimeter to the output terminals of the power supply. See Illustration 1–1.
c. Measure the voltage and check that it is within +5.05 VDC ∼ +5.10 VDC.
e. Observe the voltage and check that the ripple is less than 200 mVp–p.
1–1 DC III
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
3. Verify that the ‘OC READY’ on the DC III panel lights ON.
4. Verify that the following lamps light OFF on the DC III panel.
DC III Check
7. Verity that the same kinds of diagnostic operation as on the OC can be performed on the DC III.
Verity that the [SLAVE] button lights ON and the same screen as on the OC is displayed on the DC III.
This is called ‘Slave Mode.’
1–2 DC III
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
ON B-4A
1 2 3 4
RD SD
SD RD
CTS RTS
DSR
DCD
RTS CTS
DTR DSR
N.C. DCD
1 : Not Connected
3 : MFC is connected to CN1
3 1
DC MFC
1 : Not Connected
3 : MFC is connected to CN3
3 1
DC MFC
1–3 DC III
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
blank 2205009
1–4 DC III
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 13 2205009
SWITCH/JUMPER SETTING
TABLE OF CONTENTS
SECTION PAGE
i SWITCH/JUMPER SETTING
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 13 2205009
TABLE OF CONTENTS
SECTION PAGE
ii SWITCH/JUMPER SETTING
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 13 2205009
TABLE OF CONTENTS
SECTION PAGE
iv SWITCH/JUMPER SETTING
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 13 2205009
1-1 GENERAL
NOTICE
NOTICE
Graphic Conventions
JUMPER
JUMPER PIN
SOCKET
D `01B2' indicates jumper or
switch location on a board.
01B2 D Left illustration indicates that a
jumper socket links pins 1 and 2.
1 2 3
Component Locations
The following illustration shows the locations of some components within the Operator Console, and slot # for the nest
chassis circuit boards.
INTER-
FACE NEST
PANEL
IF
SV AMP SLOT #01 – 09
AUDIO2
DAS
IF
INTER-
FACE NEST
PANEL
IF2
SV AMP
SLOT #01 – 10
AUDIO2
DAS
IF
+5 VDC
POWER
PS ADP SUPPLY
ASSY FD ADP
MOD ASSY FDD
+12 VDC
POWER
SUPPLY
SV AMP
SLOT #20 - 37
DAS
IF
AUDIO2
INTER-
FACE NEST
DIP SW 5A1
5A1
JUMPERS
CN3
8A1 DIP SW 8A1 8J1
8D1
DIP SW 10A1
LE DIP SW 12A1
D
JUMPERS
CN4
FUSE 72111
19J1
PUSH
SW
JUMPERS
CN5
30C1
30D1
BAT-
TERY
Switch Settings
D 5A1 – Not used.
ON B-8A
`1'
`0'
1 2 3 4 5 6 7 8
Set to all `0'.
1 2 3 4 5 6 7 8 OFF: ‘0’
OFF
BP8L
1 2 3 4 OFF: ‘0’
OFF
DNP4
ON B-4A
`1'
`0'
1 2 3 4
0XXX : One
HDD
1XXX : Two
HDDs
Jumper Settings
D 30D1 – Battery Back–up for SRAM on CPW Board – Normally set as illustrated below.
PON RST
T N
N : Enable
T : Disable
BTO
16 ms
4 ms
D 19J1 – INTVL1 (Bus Release every 16 word SCSI transfer) – Normally set as illustrated below.
INTVL1
D 8J1 – INTVL2 (Bus Release Time – see 19J1 jumper) – Normally set as illustrated below.
INTVL2
1 1 : 1 ms
2 2 : 2 ms
4 4 : 4 ms
8 8 : 8 ms
D 8A1 (for 2178399 CPW) – FDD Mode – Normally set as illustrated below.
3 2 1
IPU2 SW Jumpers
(S5A1)
9S1 9V1 9X1
MD EXT SW 0
(S7A1)
J10 J15 10
J5 J6 J7
INT VEC SW
(S9A1)
J20 J25
Switch Settings
D IPU2 SW S5A1 – Selects the recon processor.
(Qj, Lemage SX/E)
ON B-8A
`1'
OFF: ‘0’
`0'
1 2 3 4 5 6 7 8
(Zj–Adv)
ON B-8A
`1'
`0'
1 2 3 4 5 6 7 8
ON B-8A
`1' OFF: ‘0’
`0'
1 2 3 4 5 6 7 8 The left is the standard setting
ON B-8A
`1'
The left setting is for IPU2 Board with
`0'
‘144 MB Memory Option’
1 2 3 4 5 6 7 8
(Qj–Base, Qj–Hlt)
ON B-8A
`1'
`0' The left is the standard setting
1 2 3 4 5 6 7 8
OFF: ‘0’
ON B-8A
`1' The left setting is for IPU2 Board with ‘Extended Heli-
`0' cal Option’ (this option replaces the currently installed
1 2 3 4 5 6 7 8 memory with 144 MB memory)
(Zj–Adv)
ON B-8A
`1'
OFF: ‘0’
`0'
1 2 3 4 5 6 7 8
ON B-8A
`1'
OFF: ‘0’
`0'
1 2 3 4 5 6 7 8
Jumper Settings
D JP9S1 ∼ 11S2, JP9V1 ∼ 11V2, JP9X1 ∼ 11X1 – Adjusts Internal Clock Frequency.
Note
Do not change these settings. Each board has its own unique settings.
WHIT
E
ACRTC
VSYN
C
DAS
LD M40B M40A
Jumper Settings
D WHITE – OVL White Width Set – Sets the OVL dot timing.
It is normally set to ‘15’ as illustrated below.
10
15
20
WHITE
4
3
ACRTC
D VSYNC – VSYNC Interrupt Level Set on System Bus – Normally set as illustrated below.
4
3
VSYNC
4
3
DAS
Note
Do not change these settings. Each board has its own unique settings.
JPR0
JPR1
JPR2
JPR3
JPR4
Jumper Settings
Note
Do not change the setting of jumpers JPR0, JPR1, JPR2 and JPR3. Their settings are determined
by the manufacturer.
INT0
INT1
INT2
INT3
INT4
INT5
INT7
INT8
JPR4
ON ON
S1 S2
ON ON
S3 S4
ON
S5
JP7
JP4
JP5
JP6
JP8 JP9
Jumper Settings
Note
Do not change the setting of jumpers JP1, JP2 and JP3. Their settings are determined by the
manufacturer.
1-2-5 IPA Board (continued) (Qj–Base except Lemage SX/E, Qj–Hlt, Zj–Adv)
D JP6, JP7 – IOINT Interruption – Normally set as illustrated below.
JP8 JP9
Switch Settings
D S1, S2, S3, S4, S5 – Base Address
Note
These dip switch settings differ according to the IPA Board slot location.
LED
CN3
CN1
RESET SW
ABORT SW
HD 63484
CN4 HD 68000
CN5
Jumper Settings
D J21B2 – Periodic Interrupt Interval – Normally set as illustrated below.
4 1
Switch Settings
D SW04A1 – Mode Select– Normally set as illustrated below.
ON
`1'
`0'
1 2 3 4 5 6 7 8
JP5
CN9 CN3
CN4 CN5
SW1
JP6
RST
TST
SW2
CN6
JP7
PLCC
JP4, 3, 2, 1
CN7
CN8
JP8
Jumper Settings
D JP1 – TG SI01 – Normally set as illustrated below.
7C
MNL TST
D JP2 – TG SI02 – Normally set as illustrated below.
7C
MNL TST
D JP3 – TG SI01 (for future use) – Normally set as illustrated below.
7C
NML TST
7C
NML TST
1 2 3 4 5 6 7 8
C
DCE
1A B
DTE A
D JP6 – Communication Protocol Set for MFC SIO (RS232) – Normally set as illustrated below.
1 2 3 4 5 6 7 8
C
DCE
3,4A B
DTE A
D JP7 – Clock Frequency Det for TGP serial communication – Normally set as illustrated below.
1D2
9.6 KHz
4.8 KHz
2.4 KHz
1.2 KHz
1 2 3
8W2
Loudness Low
Loudness Mid
Loudness High
Switch Settings
D SW1 – MFC, KB – Not used. Set all bits to ‘OFF’.
ON
`1`
OFF: `0'
`0'
1 2 3 4 5 6 7 8
ON
`1`
OFF: `0'
`0'
1 2 3 4 5 6 7 8
D SW3 (RST) – Reset – Resets the MISC2 Board or other Boards controlled by MISC2.
D SW4 (TST) – Test – Starts the test selected on SW2 bit 5 & 6.
J3 DS2 J1
DS1 SW10B1
J5
J2
DS3
Switch Settings
D SW10B1 – Normally set as illustrated below.
ON ND 8
(Switches 1 to 4 are for J5 connector,
and switches 5 to 8 are for J3 connector)
1 2 3 4 5 6 7 8
CN3
2H2 2J1 2J3
CN1
RESET
ABORT
CN4
Jumper Settings
D 2H1, 2H2, 2H3, 2J1– IRQ Level – Normally all set to ‘2’ as illustrated below.
IRQ
BR
2 1
23G2
Switch Settings
D 09G1 – Base Address – Normally set to ‘50H’ as illustrated below.
ON B-8A
`1'
`0'
1 2 3 4 5 6 7 8
7 6 5 4 3 2 1 0
ON B-8A
`1'
`0'
1 2 3 4
LSB
ID
MSB
(Reserve)
D 27A1 – Test Mode – Normally set to ‘00(H)’ as illustrated below (No test).
ON B-8A
`1'
`0'
1 2 3 4 5 6 7 8
7 6 5 4 3 2 1 0
LSB
Test No. [00H - 0DH]
MSB
Engineering use only; set to `0'
LSB
Test Mode [00H - 03H]
MSB
00 : No test
01 : Signal test
02 : Loop test (error stop)
03 : Loop test (Error no-stop)
JP1B2
SW6A1
Jumper Settings
D JP1B2 – Power–on Reset Enable/Disable – Normally set as illustrated below.
N
Normal Setting: Enabled
For Low voltage (below 4.75 V) test: Disabled
Switch Settings
D SW6A1 – Test Mode – Normally set as illustrated below.
ON
`1'
1 2 3 4 5 6 7 8 `0'
Jumper Settings
No jumpers to set.
Switch Settings
No switches to set.
Jumper Settings
No jumpers to set.
Switch Settings
No switches to set.
RESET
0K2, 1K1, 1K2
STATUS
TIMER
4K1, 5K1, 5K2
10A1 11E1
11E2
14A1 17H1
15A1
25K1
GPIB
Jumper Settings
D J04K1, 05K1, 05K2 – Interrupt Request Level – Normally all set to ‘2’ as illustrated below.
6 5 4 3 2 1
4K1
IRQ
5K1
5K2
D J00K2, 01K1, 01K2 – Bus Request Level – Normally all set to ‘1’ as illustrated below.
0 1 2 3 4
0K2
1K1 BR
1K2
D J11E2 – DMA Block Time Interval – Normally set to ‘16 ms’ as illustrated below.
16 8 4 2 1 ms
11E2
16 8 4 2 1 word/block
11E1
CRC
Switch Settings
D SW17H1 – Base Address – Normally set as illustrated below.
ON
‘1’
‘0’
1 2 3 4 5 6 7 8
ON
‘1’
1 2 3 4 5 6 7 8 ‘0’
Switch Settings
D SW14A1 – GP–IB Address – Normally set to ‘10H’ as illustrated below.
ON
‘1’
1 2 3 4 5 6 7 8 ‘0’
ON
ON
OFF
1 2 3 4 5 6 7 8
1J1
1J2
2J1
CN3 CN1
3J1
DIP SW
CN2
LED
PUSH
SW
24A1 26E2
DIP SW
Jumper Settings
D J26E2 – ROM Select – Normally set to ‘64K’ as illustrated below.
- 64K ROM
- 128K ROM
1 2 3 4 5 6 7
0 1 2 3 4
1J1
BG
1J2
BR 2J1
0 1 2 3 4
Switch Settings
D SW24A1– Test Mode – Not used. It is normally set as illustrated below.
ON
`1'
`0'
1 2 3 4
ON
`0'
Note: OFF: `1'
1 2 3 4 5 6 7 8 `1'
SW10J
SW14A1
SW24K1
J24J1
J26K1 J26K2
J27K2
J27K1
J27K3
Jumper Settings
D J24J1 – COMX Interrupt Request Level – Normally set to ‘2’ as illustrated below.
1 2 3 4 5 6 7
D J26K, 27K2, 27K3 – COMX Bus Request Level – Normally set to ‘1’ as illustrated below.
0 1 2 3 4
26K2
27K2
27K3
0 1 2 3 4
16 8 4 2
26K1
D J27K1 – DMA Block Size – Normally set to ‘16 words/block’ as illustrated below.
16 8 4 2 1
27K1
Switch Settings
D SW24K1 – Interrupt Vector Number – Normally set to ‘C6H’ as illustrated below.
ON
`1'
`0'
1 2 3 4 5 6 7 8
ON
`1'
`0'
1 2 3 4 5 6 7 8
ON
`1'
`0'
1 2 3 4 5 6 7 8
0H1
0B1 0B2 0H2 0J1
1H1
1E1 1K1
CN1
12J1
LED
CN2
26C1
26H1 26J1
25A1
Jumper Settings
D 0J1 – NEMA Interrupt Request Level – Normally set to ‘2’ as illustrated below.
1 2 3 4 5 6 7
D 0H1, 0H2, 1H1 – NEMA Bus Request Level – Normally set to ‘1’ as illustrated below.
0 1 2 3 4
0H1
0H2
1H1
0 1 2 3 4
16 8 4 2
26J1
D 26H1 – DMA Block Size – Normally set to ‘16 words/block’ as illustrated below.
16 8 4 2 1
26H1
64 32 16 4
0B1 Wait before resend
4 3 2 1
D 26C1 – Strobe Input (STBI) Data Set–up – Normally set to ‘0 msec’ as illustrated below.
8 25 41 58
0 16 33 50
1
2
3
4
5
6
7
8
Switch Settings
D 1K1 – Interrupt Vector Set – Normally set to ‘C6H’ as illustrated below.
ON B-8A
`1'
`0'
1 2 3 4 5 6 7 8
D 12J1 – Base Address for Command Packet Area – Normally set to ‘27H’ as illustrated below.
ON B-8A
`1'
`0'
1 2 3 4 5 6 7 8
ON B-8A
`1'
`0'
1 2 3 4 5 6 7 8
OFFLINE TEST MODE
Effective when bit 3 = ON
bit 4= ON
B3 B4
OFF OFF On-line Mode
ON OFF
OFF ON Off-line Test Mode (non-active)
ON ON Off-line Test Mode (active)
SW4A
SW3L
LED SW6A SW4J
SW8A CN1
CN3
SW15L
PUSH SW
CN2
CN4
J27D
Switch Settings
D SW4J – Base Address of Command Packet – Normally set to ‘30H’ as illustrated below.
ON
`1'
`0'
1 2 3 4 5 6 7 8
ON
`1'
`0'
1 2 3 4 5 6 7 8
ON
`1'
`0'
1 2 3 4
D SW4A – Serial interface (RS232C) Baud Mode – Normally set to ‘00H’ as illustrated below.
ON
`1'
`0'
1 2 3 4 5 6 7 8
D SW6A – Ethernet Address and ETC Manufactured Data Set Mode – Normally set to ‘0H’ as illustrated
below.
ON
`1'
`0'
1 2 3 4
ON
`1'
`0'
1 2 3 4
Jumper Settings
D J27D – Debug Port. A port for connecting a debug machine. No setting to be made.
1 2
J27D
7 8
CN6
CN5
CN1
CN7 9C1
J1 J2
CN10
CN8
11C1
CN12 CN11
Jumper Settings
D J1, J2 – RS232C Control Mode – Normally set as illustrated below.
1 2 1 2 3
J1 J2
(open)
Switch Settings
D SW9C1– Parity Bit/Stop Bit – Normally set as illustrated below.
ON
`1'
`0'
1 2 3 4
D SW11C1 – Baud Rate/Touch Panel Scan Rate – Normally set as illustrated below.
ON
`0'
B3 B2 B1 B0 Interval
`1' 1 1 1 1 20 ms
1 2 3 4 5 6 7 8 1 1 1 0 30 ms
1 1 0 1 40 ms
B0 : 1 1 0 0 50 ms
1 0 1 1 60 ms
B1 : Touch Panel Scan Rate 1 0 1 0 70 ms
B2 : 1 0 0 1 80 ms
B3 : 1 0 0 0 90 ms
0 X X X 100 ms
B4 :
B5 : Baud Rate B7 B6 B5 B4 Baud Rate
B6 : 1 1 1 1 19200
1 1 1 0 9600
B7 : 1 1 0 1 4800
1 1 0 0 2400
1 0 1 1 1200
1 0 1 0 600
J1
Jumper Setting
D J1 – Normally set as illustrated below.
JP4
JP5
SW2
VR3
VR5
VR4
SW1
JP1 JP2
SW4
JP7
JP6 VR1
JP8 JP3
VR2
Jumper Settings
D JP1 – Rec In – Normally set as illustrated below.
1 2 3
4 5 6
MIC LINE
1 2 3 1 2 3
4 5 6 4 5 6
INT EXT INT EXT
JP3
OLD NEW
(Zj–Adv)
JP3
OLD NEW
JP4
1 2 HI
3 4
5 6 LO
JP5
1 2 MAX
3 4 MID
5 6 MIN
JP6
1 2 HI
3 4 MID
5 6 LO
JP7
1 2 MAX
3 4 MID
5 6 MIN
JP8
1 2 MAX
3 4 MID
5 6 MIN
Switch Settings
D SW1 – Normally set as illustrated below.
ON B-8A
`1'
`0'
1 2 3 4 5 6 7 8
SAM1
SAM2
CSEL1
CSEL2
MFMSCT
CPU/MNL
FOOT SV
REC/PLY
ON B-8A
`1'
`0'
1 2 3 4 5 6 7 8
J4
J3
J5 J2
Jumper Settings
D J2, J3, J4, J5 – Normally set as illustrated below.
1 2
J2
(OPEN)
1 2 3
J3
1 2
J4
(OPEN)
1 2
J5
(OPEN)
J031 J071
SBB ASSY
ILLUSTRATION 1
Note
There are no jumpers connected to AUX2 slot (option). Therefore, no setting is required when an
option Board is installed.
(Setting Example)
- BG4
Jumper Plug No. Slot No. Circuit Board Set
- BG3
J021 Slot# 02 IPU2 open
J031 Slot# 03 DISP open - BG2
J041 Slot# 04 SBPU open
J061 Slot# 06 MISC2 open - BG1
J071 Slot# 07 ISC open
J081 Slot# 08 SCSI open - BG0
No Plug Slot# 09 (AUX2) N/A
- IACK
(Open)
QJBB ASSY
ILLUSTRATION 2
Note
There are no jumpers connected to AUX2 slot (option). Therefore, no setting is required when an
option Board is installed.
Jumper Plug No. Slot No. Circuit Board Set (Setting Example)
J021 Slot# 02 IPU2 open - BG4
J031 Slot# 03 DISP2 open
J041 Slot# 04 IPA open (*)
J051 Slot# 05 IPA open - BG3
J071 Slot# 07 MISC2 open
J081 Slot# 08 ISC open - BG2
J091 Slot# 09 SCSI open
No Plug Slot# 10 (AUX2) N/A - BG1
- BG0
(*) These jumpers must be opened only if this second
IPA Board is installed. - IACK
(Open)
Jumper Plug No. Slot No. Circuit Board Set (Setting Example)
J041 Slot# 32 IPU2 open - BG4
J051 Slot# 31 DISP2 open
J061 Slot# 30 IPA open
J071 Slot# 29 IPA open - BG3
J081 Slot# 27 MISC2 open
J101 Slot# 25 ISC open - BG2
J111 Slot# 24 SCSI open
J121 Slot# 23 ODC2 open (*) - BG1
J131 Slot# 22 MTC open (*)
J141 Slot# 21 COMX/NEMA open (*) - BG0
(*) Any of these jumpers for ODC2, MTC, COMX/NEMA
must be opened only if the connected Board is installed. - IACK
(Open)
TERMINATOR
MODULE
POWER 8
SUPPLY 2 10
CONNECTOR CNH11
1 1 9
CNH10
SW1
SCSI 6 5
CONNECTOR
2 1
D CNH10 D CNH11
2 4 6 2 4 6 8 10
1 3 5
1 3 5 7 9
D SW1
1 2 3 4 5 6 7 8
ON
(`2' only to OFF)
OFF
The numbers for the jumpers and switches above correspond to the numbers
written in Illustration 1-25.
For CNH11, different numbers are printed on the Board from ones written in
Illustration 1-25; ignore the numbers printed on the Board.
Terminator Module
D Leave the terminator module installed on the Board.
IMPORTANT NOTE:
There are two types of HDD 2143341 in the field.
POWER
SUPPLY
CONNECTOR
PRINTED CIRCUIT BOARD
HARD DISK
DRIVE
CNH1
SCSI
CONNECTOR TERMINATOR MODULE
HD#0 HD#1
CNH1 1 3 5 7 1 3 5 7
2 4 6 8 2 4 6 8
ID = #0 ID = #1
Terminator Module INSTALLED REMOVED
CNH3
CNH1
HD#0 HD#1
CNH1 2 16 2 16
1 15 1 15
ID = #0 ID = #1
CNH3 2 2
4 4
1 3 1 3
SLIDING SWITCH
REAR VIEW
G
F
E
D
C
B
A
Compatibility Information
This FDD 2191376 or 2349231 must be used with 2178399 or later CPW Board. (Qj, Qj–Base, Qj–Hlt)
This FDD 2191376 or 2349231 must be used with 2201348 or later CPW Board. (Zj–Adv)
Floppy disks which were used on U0012ZA FDDs can be read/written on this FDD 2191376 or 2349231; however,
2DD type floppy disks can not be used on this FDD 2191376 or 2349231.
REAR VIEW
SCSI CONNECTOR
JA3
1 19
2 20
SCSI ADDRESS 20
SCSI ADDRESS 21
SCSI ADDRESS 22
2 16
JP2
1 15
2 16
JP6
1 15
REAR VIEW
SCSI CONNECTOR
SCSI JUMPER
1 2 3 4 5 6 7 8
SCSI ADDRESSES
PARITY CHECK
NOT USED
POWER ON RESPONSE
MODE
SCSI Terminator
D Leave the terminator installed.
BOTTOM VIEW
2-1 GENERAL
NOTICE
Graphic Conventions
JUMPER
JUMPER PIN
SOCKET
D `01B2' indicates jumper or
switch location on a board.
01B2 D Left illustration indicates that a
jumper socket links pins 1 and 2.
1 2 3
Component Locations
The following illustration shows the locations of some components within the Operator Console, and slot # for the nest
chassis circuit boards.
+5 VDC
POWER
SUPPLY FD ADP
PS ADP ASSY
MOD -5 VDC
ASSY POWER FDD
SUPPLY
2-2-1 MV AMP II
CN7
JP2
CN6 JP3
JP1
CN5
CN8 CN4
CN3
CN9
CN16 CN1 CN15 CN14 CN13 CN12 CN11 CN10
CN2
Jumper Settings
D JP1 – Normally set as illustrated below (open.
CN1 RO
M
IL1 IN1
SW SW
mPD7831
0
LED
CN2
Jumper Settings
D CN8 – Normally set as illustrated below.
DTE DCE
RD SD
SD RD
CTS RTS
DSR
DCD
RTS CTS
DTR DSR
N.C. DCD
Switch Settings
D SWIL1 – Normally set as illustrated below.
ON
`1'
`0'
1 2 3 4 5 6 7 8
ON
`1'
`0'
1 2 3 4 5 6 7 8
ROM
LED0A1
CN1
SW4A1
ROM
SW6A1
ROM
CN2
LED
LED13A1
LED14A1
CN11
Switch Settings
D SW13B1 – Normally set as illustrated below.
ON
`1'
`0'
1 2 3 4 5 6 7 8
ON
`1'
`0'
1 2 3 4
B0 : #4
B1 : #3 Set one of switches to '1'
B2 : #2
B3 : #1
ON
`1'
`0'
1 2 3 4 5 6 7 8
TMS34061
LED CN1
CN3
JP3
DIP
SW
GATE
CN4 ARRAY
CN2
CN5
27E1
HD63484
Resistor Module
17A1, 17A2, 19A1, 19A2 – Terminating Resistors for I/O Bus – Leave these resistor modules on the sockets.
Jumper Settings
D JP1, JP2 – LOCAL BUS Interrupt Request – Normally set as illustrated below.
JP1 JP2
JP1 : IRQ3
JP2 : IRQ4
JP3
JP3 : INTR1
Switch Settings
D SW15A1 – I/O BUS Address Mapping – Normally set as illustrated below.
ON
`1'
`0'
1 2 3 4
B3 B2 Address(Hex)
0 0 F70600 - F7063F
1 0 F70640 - F7067F
0 1 F70680 - F706BF
1 1 F706C0 - F706FF
D/A CONV
CN4
CN1
CN3
TMS34061
CN5
LED
CN2
Jumper Settings
No jumper to set.
Switch Settings
No switch to set.
TEST PIN
LED
CN1
CN3
CN11
CN13
DIP SW
CN4 CN1
CN12
LED
FCP22
Switch Settings
D SW14A1 – FPU Number – Normally set as illustrated below.
ON
`1'
`0'
1 2 3 4
CN1
CN3
CN4 CN2
Jumper Settings
No jumper to set.
Switch Settings
No switch to set.
LED
CN1
CN3
DIP SW
CN2
CN4
Switch Settings
D SW15A1 – Normally set as illustrated below.
ON
`1'
`0'
1 2 3 4
LED
CN1
CN3
DIP SW
CN2
CN4
Switch Settings
D SW08A1 – FMD Address Mapping – Normally set as illustrated below.
ON
`1'
`0'
1 2 3 4
Note
When FMD is installed in slot #18, check the switch settings on FCP3B in slot #17 and on DMP2 in slot
#24.
SW4A
SW3L
LED SW6A SW4J
SW8A CN1
CN3
SW15L
PUSH SW
CN2
CN4
J27D
Switch Settings
D SW4J – Base Address of Command Packet – Normally set to ‘30H’ as illustrated below.
ON
`1'
`0'
1 2 3 4 5 6 7 8
ON
`1'
`0'
1 2 3 4 5 6 7 8
ON
`1'
`0'
1 2 3 4
D SW4A – Serial interface (RS232C) Baud Mode – Normally set to ‘00H’ as illustrated below.
ON
`1'
`0'
1 2 3 4 5 6 7 8
D SW6A – Ethernet Address and ETC Manufactured Data Set Mode – Normally set to ‘0H’ as illustrated
below.
ON
`1'
`0'
1 2 3 4
ON
`1'
`0'
1 2 3 4
Jumper Settings
D J27D – Debug Port. A port for connecting a debug machine. No setting to be made.
1 2
J27D
7 8
SW10J
SW14A1
SW24K1
J24J1
J26K1 J26K2
J27K2
J27K1
J27K3
Jumper Settings
D J24J1 – COMX Interrupt Request Level – Normally set to ‘2’ as illustrated below.
1 2 3 4 5 6 7
D J26K, 27K2, 27K3 – COMX Bus Request Level – Normally set to ‘1’ as illustrated below.
0 1 2 3 4
26K2
27K2
27K3
0 1 2 3 4
16 8 4 2
26K1
D J27K1 – DMA Block Size – Normally set to ‘16 words/block’ as illustrated below.
16 8 4 2 1
27K1
Switch Settings
D SW24K1 – Interrupt Vector Number – Normally set to ‘C6H’ as illustrated below.
ON
`1'
`0'
1 2 3 4 5 6 7 8
ON
`1'
`0'
1 2 3 4 5 6 7 8
ON
`1'
`0'
1 2 3 4 5 6 7 8
03A4
04C2 05J1 05J2
LED
CN1
PUSH SW 07C2
06J2
LED 06J1
12D2
CN3
15C2
-
CN4
CN2
21C2
23E2
CN5
Note
This DMP2 Board is used for an optional DC III.
Jumper Settings
D J03A4 – PTM#1 Clock Input – Normally set as illustrated below (open).
8 - 4.0 sec
4 - 2.0 sec
2 - 1.0 sec
- 0.5 sec
1
T - TEST : CBF option enabled. When CBF option is installed on system, set to this side.
- 2Mbit ROM
- 4Mbit ROM
4MB
D J05J1, 05J2 – MPU Bus Request Level – Normally set to ‘3’ as illustrated below.
5J1
5J2
4 3 2 1 0
4 3 2 1 0 BR
BG
D J06J1, 06J2 – DMAC Bus Request Level – Normally set to ‘2’ as illustrated below.
6J1
6J2
4 3 2 1 0
4 3 2 1 0 BR
BG
Switch Settings
D SW6A2 – Processor Number – Normally set as illustrated below.
ON
`1'
`0'
1 2 3 4
B3 B2 B1 B0 Description
0 0 0 0 Not used
B0 : 0 0 0 1 DMP4 for OC
B1 : 0 0 1 0 DMP2 for DC III #1
B2 : 0 0 1 1 DMP2 for DC III #2
0 1 0 0 DMP2 for DC III #3
B3 : 0 1 0 1 DMP2 for DC III #4
ON
`1'
`0'
1 2 3 4 5 6 7 8
1J1
1J2
2J1
CN3 CN1
3J1
DIP SW
CN2
LED
PUSH
SW
24A1 26E2
DIP SW
Jumper Settings
D J26E2 – ROM Select – Normally set to ‘64K’ as illustrated below.
- 64K ROM
- 128K ROM
1 2 3 4 5 6 7
0 1 2 3 4
1J1
BG
1J2
BR 2J1
0 1 2 3 4
Switch Settings
D SW24A1– Test Mode – Not used. It is normally set as illustrated below.
ON
`1'
`0'
1 2 3 4
ON
`0'
Note: OFF: `1'
1 2 3 4 5 6 7 8 `1'
RESET
0K2, 1K1, 1K2
STATUS
TIMER
4K1, 5K1, 5K2
10A1 11E1
11E2
14A1 17H1
15A1
25K1
GPIB
Jumper Settings
D J04K1, 05K1, 05K2 – Interrupt Request Level – Normally all set to ‘2’ as illustrated below.
6 5 4 3 2 1
4K1
IRQ
5K1
5K2
D J00K2, 01K1, 01K2 – Bus Request Level – Normally all set to ‘1’ as illustrated below.
0 1 2 3 4
0K2
1K1 BR
1K2
D J11E2 – DMA Block Time Interval – Normally set to ‘16 ms’ as illustrated below.
16 8 4 2 1 ms
11E2
16 8 4 2 1 word/block
11E1
CRC
Switch Settings
D SW17H1 – Base Address – Normally set as illustrated below.
ON
‘1’
‘0’
1 2 3 4 5 6 7 8
ON
‘1’
1 2 3 4 5 6 7 8 ‘0’
Switch Settings
D SW14A1 – GP–IB Address – Normally set to ‘10H’ as illustrated below.
ON
‘1’
1 2 3 4 5 6 7 8 ‘0’
ON
ON
OFF
1 2 3 4 5 6 7 8
2-4-6 SCSI
CN3
2H2 2J1 2J3
CN1
RESET
ABORT
CN4
Jumper Settings
D 2H1, 2H2, 2H3, 2J1– IRQ Level – Normally all set to ‘2’ as illustrated below.
IRQ
BR
2 1
23G2
Switch Settings
D 09G1 – Base Address – Normally set to ‘50H’ as illustrated below.
ON B-8A
`1'
`0'
1 2 3 4 5 6 7 8
7 6 5 4 3 2 1 0
ON B-8A
`1'
`0'
1 2 3 4
LSB
ID
MSB
(Reserve)
D 27A1 – Test Mode – Normally set to ‘00(H)’ as illustrated below (No test).
ON B-8A
`1'
`0'
1 2 3 4 5 6 7 8
7 6 5 4 3 2 1 0
LSB
Test No. [00H - 0DH]
MSB
Engineering use only; set to `0'
LSB
Test Mode [00H - 03H]
MSB
00 : No test
01 : Signal test
02 : Loop test (error stop)
03 : Loop test (Error no-stop)
2-4-7 MRX
CN1
14A1
DIP SW 15A1
16A1
CN2
Switch Settings
D SW14A1/16A1 – System Bus/ Local Bus Local Memory Base Address –
Normally set to ‘080000H’ as illustrated below.
ON
`1'
`0'
1 2 3 4 5 6 7 8
D SW15A1 – System Bus Common Memory Base Address – Normally set to ‘D00000H’ as illustrated below.
ON
`1'
`0'
1 2 3 4 5 6 7 8
DC-DC CONV
02A2 03B1
LED
CN1
CN3
DIP SW
10D1
15D2
CN4 HD63463
HD68450
WRRD
CN5
CN7 17D3
CN2
CN6 17D4 CN8
24K3
JUMPER
Jumper Settings
-F - 28 -
- 24 -
-H
- 20 -
(Example Only)
- 16 -
- 12 -
D J24K1, 24K2, 24K3 - DMAC Bus Request Level - -8-
Normally set to `4' as illustrated below. -4-
-0-
24K3 17D3 17D4
24K1
Note
This adjustment is HDC dependent and
24K2 0 1 2 3 4 has been adjusted at Manufacturing. Do
not change this setting in the field.
0 1 2 3 4
Switch Settings
D SW04A1 – Drive Code – Normally set as illustrated below.
ON
`1'
`0'
1 2 3 4 5 6 7 8
B0
B2/B5 B1/B4 B0/B3 Disk Type
B1 HD#00 Drive Code 0 0 1 130MB
B2 0 1 0 260MB
B3 0 1 1 330MB
1 0 1 290MB
B4 HD#01 Drive Code 1 1 1 Not Connected
B5
B6 Interleave Factor IF0 B7 B0 Interleave Sector
0 0 0 (Contiguous)
B7 Interleave Factor IF1
0 1 1
1 0 2
1 1 3
1K1
2K1
10A1 16L2
16L1
16L3
Jumper Settings
D 16L1, 16L2, 16L3 – DMAC Bus Request Level – Normally set to ‘4’ as illustrated below.
4
3
2
1
0
16L1 16L2 16L3
Switch Settings
D 10A1 – Drive Code – Not used. Set all switches to ‘0.’
ON
`1'
`0'
1 2 3 4 5 6 7 8
ON
`0'
`1'
1 2 3 4 5 6 7 8
B0 : 1 ms
B1 : 2 ms
B2 : 4 ms
B2 : 8 ms
B4 : 16 ms
B5 : 32 ms
B2 : 64 ms
B7 : 128 ms
D 2K1 – Burst Transfer Words – Normally set to ‘16 words’ as illustrated below.
ON
`0'
`1'
1 2 3 4 5 6 7 8
B0 : 16 words
B1 : 32 words
B2 : 64 words
B2 : 128 words
B4 : 256 words
B5 : 512 words
B2 : Not used. Set to `0'
B7 : Not used. Set to `0'
2-4-10 FMP1
JUMPER EPROM
01B2
CN3
HD68000
LED
DIP SW
12A1 14H1
JUMPER
PUSH SW 15H1
DIP SW
16A1 15H2
HD68450
CN4
Jumper Settings
D J01B2 - Power-on Reset Enable/Disable - D J14H1,15H1,15H2 - Bus Request Level -
Normally set to `A' as illustrated below. Normally set to `4' as illustrated below.
A B 0 1 2 3 4
14H1
BG
15H1
B : Disable
BR 15H2
A : Enable
0 1 2 3 4
Switch Settings
D SW16A1 – Boot Mode Select – Normally set as illustrated below.
ON Note
`1'
Special Boot Device Select is only used for Special Boot.
`0'
1 2 3 4
Boot Mode : '1' - Special Boot, '0' - Normal Boot from Hard Disk
Not used. Set to '0'
Special Boot Device Select : '1' - Floppy Disk, '0' - 'YMON' firmware monitor
Not used. Set to '0'
ON
`1'
`0'
1 2 3 4 5 6 7 8
2-4-11 DMP4
03A4
04C2 05J1 05J2
LED
CN1
PUSH SW 07C2
06J2
LED 06J1
12D2
CN3
15C2
-
CN4
CN2
21C2
23E2
CN5
Jumper Settings
D J03A4 – PTM#1 Clock Input – Normally set as illustrated below (open).
8 - 4.0 sec
4 - 2.0 sec
2 - 1.0 sec
- 0.5 sec
1
T - TEST : CBF option enabled. When CBF option is installed on system, set to this side.
- 2Mbit ROM
- 4Mbit ROM
4MB
D J05J1, 05J2 – MPU Bus Request Level – Normally set to ‘3’ as illustrated below.
5J1
5J2
4 3 2 1 0
4 3 2 1 0 BR
BG
D J06J1, 06J2 – DMAC Bus Request Level – Normally set to ‘2’ as illustrated below.
6J1
6J2
4 3 2 1 0
4 3 2 1 0 BR
BG
Switch Settings
D SW6A2 – Processor Number – Normally set as illustrated below.
ON
`1'
`0'
1 2 3 4
B3 B2 B1 B0 Description
0 0 0 0 Not used
B0 : 0 0 0 1 DMP4 for OC
B1 : 0 0 1 0 DMP2 for DC III #1
B2 : 0 0 1 1 DMP2 for DC III #2
0 1 0 0 DMP2 for DC III #3
B3 : 0 1 0 1 DMP2 for DC III #4
ON
`1'
`0'
1 2 3 4 5 6 7 8
2-4-12 ISC
J3 DS2 J1
DS1 SW10B1
J5
J2
DS3
Switch Settings
D SW10B1 – Normally set as illustrated below.
ON ND 8
(Switches 1 to 4 are for J5 connector,
and switches 5 to 8 are for J3 connector)
1 2 3 4 5 6 7 8
2-4-13 OVL2
See Section 2-3-2.
2-4-14 IFB2
See Section 2-3-3.
LED
CN1
DIP SW
CN2
Switch Settings
D SW15A1 – Normally set as illustrated below.
ON
`1'
`0'
1 2 3 4
2-4-17 FCP1
See Section 2-3-5.
LED
CN1
CN3
DIP SW
CN2
CN4
Switch Settings
D SW15A1 – Normally set as illustrated below.
ON
`1'
`0'
1 2 3 4
SW1
JP1
Jumper Settings
D JP1 – Operation Mode. Normally set as illustrated below
JP1
(With this setting, this board operates as FCP3B + FMD;
FCP3B FCP4 MD capacity is 32 MB.
If the jumper is set to the FCP3B side, this board operates
as FCP3B; MD capacity is 8 MB.)
Switch Settings
D SW1 – Normally set as illustrated below.
ON
`1'
`0'
1 2 3 4 5 6 7 8
(B0 ~ B3 settings are effective when the JP1 is set to the FCP4 side; and
B4 & B5 settings are effective when the JP1 is set to the FCP3B side.)
LED
CN1
CN3
DIP SW
CN2
CN4
Switch Settings
D SW08A1 – FMD Address Mapping – Normally set as illustrated below.
ON
`1'
`0'
1 2 3 4
Note
When FMD is installed in slot #39, check the switch settings on FCP3B in slot #38 and on DMP4 in slot
#31.
Jumper Settings
No jumper to set.
Switch Settings
No switch to set.
SW1
Switch Settings
D SW1 – Normally set according to the table below
ON
`1'
`0'
1 2 3 4
Jumper Settings
No jumper to set.
Switch Settings
No switch to set.
M64W ASSY
U157
4
3
2
N 1
O
2-4-24 BPU
LED
CN1
DIP SW
CN2
Switch Settings
D SW15A1 – Normally set as illustrated below.
ON
`1'
`0'
1 2 3 4
2-4-25 PDC
JP1B2
SW6A1
Jumper Settings
D JP1B2 – Power–on Reset Enable/Disable – Normally set as illustrated below.
N
Normal Setting: Enabled
For Low voltage (below 4.75 V) test: Disabled
Switch Settings
D SW6A1 – Test Mode – Normally set as illustrated below.
ON
`1'
1 2 3 4 5 6 7 8 `0'
2-4-26 OC INTF3
SW3
J1 SW4
J3-2 J4
J10
J3-1 J6 J7
J9
J5
J2
J8
Switch Settings
D SW3, SW4 – Normally set as illustrated below.
ON ON
1 2 3 4 5 6 7 8 1 2 3 4 5 6 7 8
SW3 SW4
Jumper Settings
D J1 – Buzzer volume for Cooling Wait End – Normally set as illustrated below.
MAX 1 2
3 4
MIN 5 6
1 3 5
2 4 6
MAX MIN
NML NML
TEST TEST
J3-1 J3-2
1 3 5
2 4 6
HI LO
1 3 5
2 4 6
MAX MIN
2AMP
1 2 3
2AMP
1 2 3
D J8 – Volume for Auto Voice for the OC speaker. Normally set as illustrated below
1 3 5
2 4 6
MAX MIN
1 2 3 1 2 3
2-5 BACKBOARD
J31 ∼ J231
J152 J172
J162
(Setting Example)
- BG4 -
- BG3 -
- BG2 -
- BG1 -
- BG0 -
- IACK -
(Close) (Open)
Note
Open all six jumper plugs if that slot contains a circuit board.
MANUFACTURING
ADJUSTMENT ONLY
JP2
21 1
22 2 JP3
JP4
JP1
JP6
JP5 1
2
3 1
4 2
TERMINATOR
LED1
LED2
LED3
2-6 HARD DISK DRIVE (continued) (ProSpeed VX, ProSpeed, Accell, ProSeed)
22 21 16 15 4 3
2 1
14 13 D JP5
4 3
2 1 2 1
2 1
D JP4 D JP6
12 11 12 11 2 1
2 1 2 1
(For Standard Disk) (For Auxiliary Disk)
Terminator
D (For Standard Disk)
Leave all four terminators on the disk.
2-7 HARD DISK DRIVE (ProSpeed Plus, Accell EI, ProSeed EF)
TERMINATOR
MODULE
POWER 8
SUPPLY 2 10
CONNECTOR CNH11
1 1 9
CNH10
SW1
SCSI 6 5
CONNECTOR
2 1
1 3 5 1 3 5
1 3 5 7 9 1 3 5 7 9
D SW1 D SW1
1 2 3 4 5 6 7 8 1 2 3 4 5 6 7 8
ON ON
(`2' only to OFF) (`2' only to OFF)
OFF OFF
The numbers for the jumpers and switches above correspond to the numbers
written in Illustration 2-35.
For CNH11, different numbers are printed on the Board from ones written in
Illustration 2-35; ignore the numbers printed on the Board.
POWER
SUPPLY
CONNECTOR
PRINTED CIRCUIT BOARD
HARD DISK
DRIVE
CNH1
SCSI
CONNECTOR TERMINATOR MODULE
HD#0 HD#1
CNH1 1 3 5 7 1 3 5 7
2 4 6 8 2 4 6 8
ID = #0 ID = #1
Terminator Module INSTALLED REMOVED
CNH3
CNH1
HD#0 HD#1
CNH1 2 16 2 16
1 15 1 15
ID = #0 ID = #1
CNH3 2 2
4 4
1 3 1 3
SECTION 3 – GANTRY
WARNING!
UNCONTROLLED GANTRY RESPONSE! CONFIRM THAT THE DIP SWITCHES ON THE TGP
BOARD ARE SET AS DESCRIBED IN TABLE 3–1, UNLESS OTHERWISE SPECIFIED. IF ANY
OF THE SWITCHES ARE SET TO ON, THE GANTRY MIGHT BEHAVE DANGEROUSLY.
If any settings of the TGP dip switches are changed and then reset to their original settings, verify the following accord-
ing to Table 3–3.
2. Lower the Table, and verify that the Table stops at the specified height.
4. Move the cradle in the IN direction, and verify that the cradle stops at the specified position.
Normal Setting
∼
∼
OFF
ON G4 ONLY
°
NOTICE
Do not set the SYS/OFF/MNL switch to OFF except in emergency while the Gantry is rotating,
or the static brake will be damaged.
Note
1. Do not repeatedly accelerate and decelerate the Gantry over a limited period of time; this may
cause ‘OVER LOAD ERROR’ in the servo amplifier.
2. The Gantry might act abnormally if it has been rotated by hand. If this occurs, then press the reset
button on the TGP Board.
MID
The rotation speed is determined by a combination of the two switch settings as above.
SLOW
According to the ROM part number and the PDU model ( ) the system has, the dip switch set-
According to the PDU model ( ) the system has, the dip switch settings on the OGP Board
(Qj–Base)
(Qj–Hlt, Zj–Adv)
JP1
JP8 JP8
JP7
JP6
JP9
SECTION 4 – DAS
J3
J1 J2 J4
SW2 J5
SW1
J6 J7 J8 J9
CIF ASSY
$# ( $#' $& ( $* '(( #
D
$& ")'( (
$& !(&
D %&( #)"& $
$& $# ( ")'(
JUMPER NORMAL SETTING
#
$# ( $#' $& ( $* '(( #
D
$& ")'( (
$& $&
D %&( #)"& $
$& $# ( ")'(
Jumpers
Switches
D Push Switch S1: aborts the system software loading
Jumpers
! !
"
Switches
D Push Switch RAZ: resets the CPU Board
5-4 KV BOARD
Jumpers
Only one jumper ST1 exists on the kV Board. Refer to Illustration 5–6. ST1 has four possible positions:
XJ2 XJ1
ETIQUETTE
ST1
NOR
TAN
TANTCA
TCA
Jumpers
Normal settings of JP1 through JP8: ‘ZJ’ side
OPTICAL FIBER
CABLES
PDU
PDU
Direct Current
Emergency Stop
A–1 APPENDIX
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
417–5009 Standby
Start
Table Set
Abort
Intercom
A–2 APPENDIX
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
Symbol Description
Microphone (Mic)
Contrast
Brightness
_
System storage prior to installation:
Maintain storage temperature between –10° C and +60° C
_
A–3 APPENDIX
CT PROSPEED SERIES
GE MEDICAL SYSTEMS FUNCTIONAL CHECK / ADJUSTMENT
REV 0 2205009
CLASS 1 EQUIPMENT
Any permanently installed equipment containing operator or patient accessible surfaces must provide backup protec-
tion against electric shock,in case the BASIC INSULATION fails. In addition to BASIC INSULATION,Class1 equip-
ment contains a direct connection to a PROTECTIVE(EARTH) CONDUCTOR which prevents shocks when a person
touches a broken piece of equipment or touches two different equipment surfaces simultaneously.
TYPE B EQUIPMENT
CLASS I, II, or III EQUIPMENT or EQUIPMENT with INTERNAL ELECTRICAL POWER SOURCES provide an ade-
quate degree of protection against electric shock arising from (allowable) LEAKAGE CURRENTS or a breakdown
in the reliability of the protective earth connection.
ORDINARY EQUIPMENT
Enclosed EQUIPMENT without protection against the ingress of water.
OPERATION 0f EQUIPMENT
CONTINUOUS OPERATION WITH INTERMITTENT LOADING.
Operation in which EQUIPMENT is connected continuously to the SUPPLY MAINS. The stated permissible loading
time is so short that the long term on–load operating temperature is not attained. The ensuing interval in loading is,
however, not sufficiently long for cooling down to the long term no–load operating temperature.
EQUIPMENT not suitable for use in the presence of a FLAMMABLE ANESTHETIC MIXTURE WITH AIR or WITH
OXYGEN or NITROUS OXIDE
CLEANING
The ProSpeed S series system is NOT WATERPROOF. It is NOT designed to protect internal components against
the ingress of liquids.Clean external system surfaces(Gantry,table,consoles and accessories)with a soft cloth dipped
in hot water and wrung DAMP/DRY. (NOT dripping!) IF NECESSARY, use only mild (dish washing liquid) soap to re-
move dirt.
NOTICE
Avoid damage to equipment! Some ”spray and wipe”cleaners etch and permanently cloud
clear plastic surfaces!! Use only warm water and mild liquid soap to clean surfaces.
A–4 APPENDIX
GE Medical Systems: Telex 3797371
P.O. Box 414, Milwaukee, Wisconsin 53201 U.S.A.
(Asia, Pacific, Latin America, North America)