Académique Documents
Professionnel Documents
Culture Documents
Publications
2308529–100
Revision 0
do not duplicate
WARNING
X–RAY EQUIPMENT IS DANGEROUS TO BOTH PATIENT AND OPERATOR
UNLESS MEASURES OF PROTECTION ARE STRICTLY OBSERVED
Though this equipment is built to the highest standards of electrical and mechanical safety, the useful x–ray beam becomes a source of
danger in the hands of the unauthorized or unqualified operator. Excessive exposure to x–radiation causes damage to human tissue.
Therefore, adequate precautions must be taken to prevent unauthorized or unqualified persons from operating this equipment or exposing
themselves or others to its radiation.
Before operation, persons qualified and authorized to operate this equipment should be familiar with the Recommendations of the Interna-
tional Commission on Radiological Protection, contained in Annals Number 26 of the ICRP, and with applicable national standards.
ATENCION
LOS APARATOS DE RAYOS X SON PELIGROSOS PARA EL PACIENTE Y EL MANIPULADOR
CUANDO LAS NORMAS DE PROTECCION NO ESTAN OBSERVADAS
Aunque este aparato está construido según las normas de seguridad más estrictas, la radiación X constituye un peligro al ser manipulado
por personas no autorizadas o incompetentes. Una exposición excesiva a la radiación X puede causar daños al organismo.
Por consiguiente, se deberán tomar todas las precauciones necesarias para evitar que las personas incompetentes o no autorizadas
utilicen este aparato, lo que sería un peligro para los demás y para sí mismas.
Antes de efectuar las manipulaciones, las personas habilitadas y competentes en el uso de este aparato, deberán informarse sobre las
normas de protección fijadas por la Comisión Internacional de la Protección Radiológica, Anales No 26: Recomendaciónes de la Comi-
sión Internacional sobre la Protección Radiológica y normas nacionales.
ACHTUNG
RÖNTGENAPPARATE SIND EINE GEFAHR FÜR PATIENTEN SOWIE BEDIENUNGSPERSONAL,
WENN DIE GELTENDEN SICHERHEITSVORKEHRUNGEN NICHT GENAU BEACHTET WERDEN
Dieser Apparat entspricht in seiner Bauweise strengsten elektrischen und mechanischen Sichereitsnormen, doch in den Händen unbe-
fugter oder unqualifizierter Personen wird er zu einer Gefahrenquelle. Übermäßige Röntgenbestrahlung ist für den menschlichen Orga-
nismus schädlich.
Deswegen sind hinreichende Vorsichtsmaßnahmen erforderlich, um zu verhindern, daßunbefugte oder unqualifizierte Personen solche
Geräte bedienen oder sich selbst und andere Personen deren Bestrahlung aussetzen können.
Vor Inbetriebnahme dieses Apparats sollte sich das qualifizierte und befugte Bedienungspersonal mit den geltenden Kriterien für den ge-
fahrlosen Strahleneinsatz durch sorgfältiges Studium des Hefts Nr. 26 der Internationalen Kommission für Strahlenschutz (ICRP) vertraut
machen: Empfehlungen der Internationalen Kommission für Strahlenschutz und anderer nationaler Normenbehörden.
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
TABLE OF CONTENTS
REVISION HISTORY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . v
LIST OF EFFECTIVE PAGES . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . v
1 INSTALLATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1–1
SECTION 1 – INTRODUCTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1–1
1–1 Chapter Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1–1
1–2 Inspection for Damage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1–1
SECTION 2 – DESCRIPTION OF DELIVERED EQUIPMENT . . . . . . . . . . . . . . . . . . . . . . . . 1–1
SECTION 3 – HOW TO USE THE CONSOLE IN INSTALLATION MODE . . . . . . . . . . . . . . 1–1
3–1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1–1
3–2 Application and Installation modes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1–1
3–3 Description of the Console . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1–2
3–4 Accessing the Generator or Gantry Installation Mode from the Console . . . . . . . . . . . . 1–4
3–5 Installation Menu Tree Structure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1–9
3–6 Accessing the Different Installation Menus in the Tree Structure . . . . . . . . . . . . . . . . . 1–23
3–7 Saved Parameters and Checksum . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1–23
3–8 Re–loading of Default (Initial) Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1–24
3–9 Returning to the Application Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1–25
3–10 Generator Configuration Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1–25
3–10–1 Gantry Present/Absent (PRS_A) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1–25
3–10–2 Display of Parameters Following an Exposure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1–26
3–10–3 Calendar and Clock . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1–27
3–10–4 Visual Representation of Numerical Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1–27
3–10–5 Visual Display of Parameter Values . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1–28
3–10–6 Modification of a Parameter Value or Entry of a Measurement . . . . . . . . . . . . . . . . . . . 1–28
3–10–7 Entry of an Alpha–Numeric Value . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1–29
3–11 Principles of Generator Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1–30
3–11–1 Vocabulary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1–30
3–11–2 Use of the First Trigger/Second Trigger Buttons During Generator Installation . . . . . . 1–30
3–11–3 Inhibit Conditions Due to Gantry Absence . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1–30
3–11–4 Inhibit Conditions Due to Grid Absence/Presence . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1–31
3–11–5 Displayed Messages During Exposure–Producing Calibration Procedures . . . . . . . . . . 1–31
3–11–6 Other Displayed Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1–31
SECTION 4 – SITE PREPARATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1–31
SECTION 5 – INSTALLATION STEERING . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1–32
i
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
2 CALIBRATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2–1
SECTION 1 – INTRODUCTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2–1
SECTION 2 – GENERATOR CALIBRATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2–1
SECTION 3 – GANTRY CALIBRATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2–1
SECTION 4 – AEC CALIBRATION FOR A GIVEN SCREEN PAIR . . . . . . . . . . . . . . . . . . . . 2–2
SECTION 5 – AEC CALIBRATION ON STEREOTIX 2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2–2
SECTION 6 – RE–CALIBRATION OF REFERENCE OPTICAL DENSITY IN
NON–STEREOTIX 2 MODE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2–2
SECTION 7 – CUSTOMER REQUESTED RE–CALIBRATION OF STEREOTIX 2 MODE
REFERENCE OPTICAL DENSITY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2–3
ii
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
iii
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
iv
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
REVISION HISTORY
)%' '() ) %$ % '( %$ , ) %""%, $ *&'(
)' ! )
'&"#$) % ) )* - )* $
*&' % ) (%),' )% +
%' ! )
*&' % ) (%),' )% + %$"-
%),' + &'%+ ( ) *(' , ) ) %&&%')*$ )- )% $'( ) -
" ,
)* #- $ )% %# %#&" $) '' $ )*
%*)&*) $ (-()# '(%"*) %$
Title page 0
i through vi 0
v
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
Blank page
vi
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
CHAPTER 1 – INSTALLATION
INSTALLATION
SECTION 1
INTRODUCTION
This chapter presents the Senographe DMR installation procedures. The wiring and room
layout requirements for the product are covered in the Pre–Installation Manual.
The Senographe DMR was completely inspected for proper operation and appearance before
shipment. However, it is necessary to inspect the product after the shipment is received.
Visually inspect the packages for any apparent damage. If there are signs of damage, refer the
matter to the shipping company immediately.
Open the packages and refer to the Product Delivery Instructions (PDI). Verify that items on
the list are present in the package. Carefully examine the contents for small parts.
SECTION 2
DESCRIPTION OF DELIVERED EQUIPMENT
3–1 Introduction
This section describes how to use the Senographe DMR console during installation.
Throughout the Senographe DMR documentation, reference is made to the application mode
and to the installation mode to differentiate between normal operation of the product by a
doctor and operation of the product during installation.
Each time the Senographe DMR is turned on the software automatically starts up in
application mode.
For information on using the Senographe DMR in application mode, refer to the Operator
Manual.
Two major Senographe DMR components must be installed – the generator and the gantry.
Therefore, there are two separate installation phases, one for the generator and one for the
gantry (see Section 3–4).
1–1
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
Illustration 1–1 shows the Senographe DMR console. When the Senographe DMR is in
application mode, the six keys below the display window correspond to the functions that
their respective symbol markings indicate (focal spot, track, filter, etc. – refer to the Operator
Manual for explicit details).
On the other hand, when the Senographe DMR is in installation mode, the function of each of
these six keys depends on the procedure being performed and is displayed on the bottom line
of the display window adjacent to the corresponding key.
To enter installation mode, press the SETUP key once. To return immediately to
Illustration 1–3 shows the console before and after entry to the installation mode.
Section 1 of this illustration shows a typical console display during normal application mode
use.
Section 2 shows the appearance of the display window when, having started from application
Section 4 shows the appearance of the display window when the password has already been
entered once, and starting from application mode, the SETUP key is pressed.
1–2
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
ILLUSTRATION 1–1
SENOGRAPHE DMR CONSOLE
INSTALLATION
1–3
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
3–4 Accessing the Generator or Gantry Installation Mode from the Console
INSTALLATION
To access the generator or gantry installation mode starting from application mode, follow
this sequence:
2. For access to the generator and gantry installation menus, change the position of the
installation menu enable switch (switch B1–8 on generator CPUboard 400–PL3) from
its position prior to power-up to the opposite position (see Illustration 1–2).
The installation menus can also be accessed by password. Select the PASSWORD key,
enter your password (given at the training course) (see Illustration 1–3, Section 3).
6. You now have access to the installation menus of the sub–assembly selected (generator
or gantry).
Note: The position of the installation menu enable switch is memorized each time the
Senographe DMR is turned on. Until the machine is again turned off (or the default
parameters are re–loaded via the generator’s CLEAR menu, see Section 3–8), this
memorized position is the installation-menu-access-disabled position, and the opposite
position is the installation-menu-access-enabled position. The significance of the
switch’s position thus depends upon its position at power–up.
When you have finished using the Senographe DMR installation mode, you must
CAUTION turn off the Senographe DMR before handing over the machine to the user. In this
way, when the machine is turned on again, access to the installation mode will be
automatically locked out (see Note above).
Illustration 1–3 shows the sequence for accessing generator installation mode.
1–4
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
ILLUSTRATION 1–2
LOCATION OF INSTALLATION MENU ENABLE SWITCH
INSTALLATION
GENE CPU BOARD 400–PL3
B1
?
P1 P2
1–5
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
ILLUSTRATION 1–3
SEQUENCE FOR ENTERING GENERATOR INSTALLATION MODE
INSTALLATION
1
APPLICATION
SETUP
Nb of exposures
ÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉ
VIEW MEDICAL LANGUAGE PASSWD
ÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉ
ÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉ
ÉÉÉÉÉÉ
ÉÉÉÉÉÉ
ÉÉÉÉÉÉ
1–6
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
INSTALLATION
APPLICATION
SETUP
Nb of exposures
ENTER YOUR PASSWORD:
1 2 3 4 5 6
PASSWORD
4
APPLICATION
SETUP
Nb of exposures
ÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉ
GENE MEDICAL LANGUAGE ARM CLINIC
GENE ARM
ÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉ
ÉÉÉ
ÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉ
ÉÉÉ
ÉÉÉ
1–7
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
5
INSTALLATION
APPLICATION
SETUP
GEMS-E VX.XX GEN DD.MM.YY
INSTAL MAINT
ÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉ
GENE ARM
ÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉ
ÉÉÉÉÉÉÉÉÉÉÉÉ
ÉÉÉÉÉÉÉÉÉÉÉÉ
ÉÉÉÉÉÉÉÉÉÉÉÉ INSTAL
6
APPLICATION
SETUP
LAST INSTAL (D/M/Y H:M) 28/03/96 10:24
CKSUM GENE
ÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉÉ
ÉÉÉÉÉÉÉÉÉÉÉ
ÉÉÉÉÉÉÉÉÉÉÉ INSTAL
ÉÉÉÉÉÉÉÉÉÉÉ
CKSUM GENE
1–8
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
INSTALLATION
The various installation menus are accessible via a tree structure which is shown in
Illustrations 1-5 through 1–13.
1–9
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
Blank page
INSTALLATION
1–10
L These selections appear only when
the password has been entered.
Select arm/gene install
GENE L MEDICAL LANGUAGE ARM L PASSWORD
GENERATOR CALIBRATION
18 24 PARAM CHANGE offs= (values)
CONFIG GENE AOP DOSE PROD
NEXT VALID
WARNING
WILL ENTAIL RE–CALIBRATING
THE TUBE HEATER CURRENT.
USE THE FUNCTION ONLY WHEN
YOU HAVE REPLACED THE X–
B C
RAY TUBE.
section 3–10–3
RE–INITIALISATION DATE and TIME UPDATE WITH THE BOARD 300PL3
INIT_TUBE INIT_ENERGY DAY HOUR YES NO FILM (AOP)
CELL ALGO COPY TYPE
E
B–SYSTEMS PRESENCE TUBE RE–INITIALISATION
PRS–DS PRS–IDC VALID
NCE
O
DIGITAL STATION PRESENCE
YES NO
IMAGE DETECTOR CTRL PRES.
YES NO
VALID
D
1–12
C
Job Card IST 001
HTR_DELTA Parameters HTR_SCL HARDWARE TRACK 1 PARAM HTR_SCL track 1 x=(2 values)
LF/MO SF/MO LF/RH SF/RH 1st pt 2nd pt calcul CHANGE
nu
as
MO LF/Mo HTR_DELTA display DEL_x=(3 values) HTR_SCL CALIB TRACK1 PT1 ref=(value)
CMDE CALIB Measi THE MENU ICH_M, WHICH IS USED TO ENTER
THE MEASURED FILAMENT CURRENT VALUE
FOUND IN THE CALIB MENU, IS ONLY ACCESSI-
BLE AFTER USE OF THE CALIB MENU.
es kV/A)
HTR_SCL TRACK1 PT1 modif ref=(value) HTR_SCL TRACK1 1st point meas=(value)
NEXT VALID NEXT VALID
lue) (value)
mA MEASUREMENT
B CALIB PARAM
THIS MENU IS ONLY ACCESSIBLE WHEN THIS MENU IS ONLY ACCESSIBLE WHEN
BOTH FOCAL BIAS VOLTAGE VALUES HAVE BOTH HV VALUES HAVE BEEN MEASURED
BEEN MEASURED AND ENTERED VIA THE AND ENTERED VIA THE ”1ST PT” AND ”2ND PRESSING THE ”MEASUR” KEY UP-
”1ST PT” AND ”2ND PT” MENUS. PT” MENUS. DATES THE DISPLAYED VALUE OF
Same menu THE THEORETICAL MEASURED X–
BIAS VOLT calcul structure as RAY TUBE CURRENT VALUE
for 1st pt
VALID
mA MEASUREMENT PARAMETER f/mA=(value)
CHANGE
FOCAL BIAS VOLT Display Focus/Track kV CALIB 1st point ref=(value) kV
WHEN
HAS CHANGE
ENU. REF CALIB kV_M mA MEAS PARAM MODIF f/mA=(value)
AL BIAS VOLTAGE modif x/y (value) kV REFERENCE 1st point ref=(value) kV kV CALIB 1st point MEASURED kV 1st meas=(value)
XT VALID NEXT VALID NEXT VALID
1–14
Job Card IST 007
CELL CALIBRATION
HV/DAC PM/HV STER
E
Job Card IST 004 SOURCE CHOICE(fsc)/(scr/nscr) TARGET CHOICE (fsc)/(scr/nscr) ENABLE COPY
ALIBRATION Job Card IST 005 SCREEN FSC SCREEN FSC VALID
_HV/DAC PARAM display x=(2 values) STEREOTIX CELL calib grid present/absent
HANGE CALIB PARAM
_HV/DAC PARAM change x=(value) STEREOTIX CELL calib STEREOTIX PARAM display x=(4 values)
EXT VALID CELL CALIB CALC CHANGE
CHANGE CHANGE
Same menu
M change x=(value) structure as
THE VALUE OF GAMMA IS for OD_R
WARNING CALCULATED DURING THE REFERENCE ENERGY change OD_r=(value)
CALIBRATION PROCEDURE
AND MUST NEVER BE MODI- NEXT VALID
FIED MANUALLY FOR ANY
REASON WHATSOEVER.
1–16
E E
Job Card IST 015
MAGNIFICATION CORRECTION
Job Card IST 013 CALIB PARAM
TIME RECIPROCITY FAILURE fsc name
HEATER REFEN LNRT PARAM
CONTACT CONFIGURATION
TIME RECIPROCITY FAILURE THICK.=x cm YES NO
CALIB OD_mea CALC
x=(7 values) MAGNIFICATION CALIB. x cm
THICK CALIB
RECIPROCITY FAIL.:calc
TIME RECIPROCITY FAILURE
VALID
(value) Select plexi thickness MAGNIFICATION CALIB.
2 CM 4 CM 6 CM
MEASURED OPTICAL DENSITY=(value)
NEXT VALID
5220 5230
5250
TALLATION TESTS ARM CONFIGURATION
ARM CALIBRATION
TRENGTH CUTOFF TUBE COLLIM CASSET BUCKY CAMERA ANGLE
THICK COMP FORMAT GRID HOUSING ANGLE
09999
Not used CAMERA FORMAT
5290
6x6 8x8
GTH
5227
CALIB PARAM
G 5446
POWER CUTOFF TEST 18x24 / 24x30 HEIGHT DIFF.: NO
VALID Job Card IST 029 5310 YES NO
TUBE HOUSING ANGULATION
5445
MENU CONSISTS OF PRODUCTION
PROCEDURES THAT ARE OF NO USE
HE SERVICE ENGINEER.
G H J K CALIB PARAM
CASSETTE DETECT PRESENCE
YES NO
5311 5232
HOUSING ANGULATION CALIB COLLIMATOR CONFIGURATION
EXPOSE MEASURE S.I.D Calcul FRONT_B TYPE MARG_L MARG_R MARG_F
5314
HOUSING ANGULATION PARAM
A–REF Same menu
structure as
MARG_L
5444
FRONT BLADE PRESENCE
YES NO
1–18
5251 Job Card IST 025 Job Card IST 023 5270 Job Card IST 030 5290
BREAST THICKNESS CALIB COMPRESSION CALIBRATION X–RAY FORMAT ’COMPRESSION ARM ROTATION
CONTACT MAGNIF FORCE LIMIT CALIB PARAM CALIB PARAM
J H K A
MIN ANGLE CALIB MAX ANGLE CALIB
* Job Card IST 024 (Value) (Value)
NEXT VALID NEXT VALID
52 5257 5271
ICKNESS ’MAGNIF MODE’ THICKNESS FORCE DETECTOR CALIB
CALIB PARAM CALIB PARAM
5276
5259 5273/5283 COMP MOTOR CURRENT LIMIT
’MAGNIF MODE’ THICK PARAM FORCE DETECTOR PARAMETERS CALIB PARAM
OFFSET ELASTIC 0daN 3daN. . . . . . . . . . . .27daN
5282
5279 5280
FORCE DETECTOR MEASURES
IN_COMPRESSION LIMIT % : (value) OUT OF COMP LIMIT % : (value)
15daN 18daN 21daN 24daN 27daN 2>1
+ – VALID + – VALID
5335
Menu structure Menu structure LIGHT COVERAGE: (value in Motor Steps)
identical to identical to
REF_R REF_R + – VALID
5293
5336
LATERAL X–RAY FORMAT REF
MEASURE S.I.D Calcul LIGHT ALIGNMENT : (value in Motor Steps)
+ – VALID
5295
REF MEASURE (mm) : (value)
VALID 5343
CORR REAR FORMAT
MEASURE S.I.D Calcul
5301
MAGE DISTANCE mm : (value) 5345
VALID SOURCE IMAGE DISTANCE mm=(value) 5338
5339
5344 ARM INSTALLATION exposure
REAR DIFF MEASURE mm=(value) VALID
NEXT VALID
0006
ARM: ADJUST EXPOSURE
5340
5241 5342
SOURCE IMAGE DISTANCE mm : (value) 0005
LATERAL DIFF MEASURE mm=(value)
ARM INSTALLATION exposure end
VALID NEXT VALID
VALID
1–20
Job Card IST 030 5290
X–RAY FORMAT REFERENCE set SID to 560 !
CALIB PARAM
H 5236
X-RAY FORMAT PARAMETERS
DELTA_X COLLIM LIGHT
5135
FRONT COLL OPENING (mm): (value)
NEXT VALID
Job Card IST 030 5290
X–RAY FORMAT
CALIB PARAM
G J H A G
Job Card IST 026 5285
GRID TRAVEL
CALIB PARAM
5287
GRID PARAMETERS DISPLAY
OFFSET GAIN MIN MAX TESTS
+4.000E+0
PARAMS VMAX PARAMS SYNC
T VALUES, GIVEN FOR INFORMATION ONLY. DO NOT ATTEMPT TO CHANGE THESE PARAMETERS!
1–22
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
INSTALLATION
Once you have followed the installation mode access procedure given in Section 3–4, you can
go down into the various installation menus by selecting them progressively using the six
keys below the console display window (see Section 3–5).
To move upward in the tree structure, press the SETUP key once for each change
of level desired. This is useful for going from one branch of the tree structure to another
during installation procedures, or if you accidentally go into an unwanted branch.
1–23
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
To re–load the generator or gantry default parameters (prior to any installation or calibration
procedures), use the appropriate CLEAR menu. The use of this menu executes a software
power–up that is comparable to the very first hardware power–up of the generator or gantry.
When the generator CLEAR menu is used, the DMR leaves the installation menus and returns
to the application mode. In order to re–access the installation mode after performing a
generator CLEAR, you must change the position of the installation menu enable switch (see
Section 3–4).
When the gantry CLEAR menu is used, the DMR returns to the gantry installation entry level
(SETUP/ARM/VALID).
Following use of the CLEAR menu, the software displays a CHECKSUM ERROR message.
All generator or gantry calibration procedures must be performed, followed by a CKSUM to
enable use of the new parameters.
1–24
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
INSTALLATION
To return to the application mode press the SETUP key n times, where n is the
number of levels that you have gone down the installation tree structure.
Note: If you have performed any calibration procedures, or otherwise changed any saved
parameter values, you must perform a saved parameter memory checksum before
returning to the application mode (see Section 3–7) – if you do not perform this
checksum, then you will receive a CHECKSUM ERROR message.
Note: If you switch off the DMR for any reason and then switch it on again, you must once
again change the position of the installation menu enable switch before being able to get
access to the installation mode (see Section 3–4).
The service engineer may set PRS_A to NO to isolate the generator from the gantry during
troubleshooting (from the point of view of the generator software, the gantry becomes a
simple tube stand).
Following a generator CLEAR (see section 3–8) PRS_A takes on the value NO.
PRS_A is kept in the saved parameter memory, which means that its value is retained when
power is turned off. This means that if, during a service call, the service engineer has to set
PRS_A to NO it must be set back to YES before handing back the DMR to the customer.
The value of PRS_A can be set by following the menu access path shown above, selecting
YES or NO and performing a CKSUM.
1–25
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
To simplify the service engineer’s work, the software can display certain parameters
following an exposure. These parameters are displayed on the console display window in two
groups which appear alternately:
and
Where:
This DISP parameter is turned on and off by following the menu access path shown above.
Select YES or NO. A CKSUM is NOT necessary because this parameter is not a saved
parameter. When power is turned off, the value of DISP is lost and reverts to NO at the next
power–up.
1–26
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
INSTALLATION
Menu access path: SETUP/GENE/INSTAL/GENE/CONFIG/DATE
The generator CPU is equipped with a calendar and clock that are used for the following
functions:
D Thermal protection cooling algorithms, whether or not the generator is turned on.
D Memorization of the date and time of an equipment failure.
D Last date of access to the installation mode.
The last function makes it possible for a service engineer to verify the date and time of the last
service visit and also determine whether access to the installation modes was made by
unauthorized persons since that visit.
The date is modified by following the menu access path shown above, selecting DATE, then
NEXT and, finally, rotating the kv dial to obtain the desired day. Select NEXT
again and rotate the kv dial to obtain the desired month. Repeat once more to obtain the
desired year, then select VALID to confirm the date.
Note: The date is entered in the form DD/MM/YY (for example, if the date to be entered is
February 14th, 2001, you must enter 14/02/01).
The time is modified by following the menu access path shown above and then selecting
HOUR. Enter the time of day as explained above for the date, using the NEXT, kv dial and
VALID functions.
Note: The time of day is entered in 24-hour format, therefore, if the time of day is 3:13:27 pm,
you must enter 15:13:27.
Note: Following a generator CLEAR of saved parameters, the calendar and clock, as well as
the last installation mode access date, are initialized to the release date of the generator
software version at 0:00:00.
When displayed on the console, most numerical values appear in the form of real numbers in
scientific notation floating point format with four significant figures.
–1.250E+3 = –1250.00
+4.345E–1 = 0.4345
+1.000E+0 = 1.000
1–27
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
A parameter value can be displayed on the DMR console by accessing the calibration menus
for the procedures that compute the parameter in question. Typically this is done when you
are at a menu choice between CALIB and PARAM. Select PARAM to display the parameter
value.
Note: If the calibration procedure computes more than one parameter, turning the kv
As an example, looking at the values of scale factor K and offset voltage VOF of tube track 1
which are computed by the heater current scale factor calibration routine: follow the menu
access path SETUP/GENE/INSTAL/GENE/GENE/HTR_SCL/TRACK1/PARAM.
On the right of the first console display line appears the value of scale factor K. Rotate the kv
dial clockwise by one position. Now offset voltage VOF appears instead.
When at the parameter value modification or measurement entry level in the menu structure,
there are two possible selections: NEXT and VALID. Begin by selecting NEXT. This enables
modification of the first character to be entered. Successive selections of NEXT will enable
modification of each character to be entered.
Use the kv dial to modify the character being entered. Select VALID when all
NEVER modify a parameter value on a DMR that has already been calibrated
CAUTION unless you have a specific reason. Doing so, in general, will necessitate entering the
default parameter value(s) and completely re–calibrating the sub–assembly
(generator or gantry) for the parameter value(s) in question.
1–28
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
As an example, set the maximum exposure time to 5.7 (+5.700E+0) seconds, follow the menu
access path SETUP/GENE/INSTAL/GENE/CONFIG/CONFIG/T_MAX, select NEXT
INSTALLATION
and turn the kv dial to set the sign of the value to be entered (in this example, the
value is positive +). Select ”NEXT again. Turn the kv dial to set the value of the first digit to 5.
Continue selecting NEXT and turning the kv dial until you have input +5.700E+0 (mantissa
sign and four digits plus exponent sign and one digit). If you make a mistake, select NEXT
repeatedly until you arrive at the faulty value, then correct it by rotating the kV dial. If you
truly wish to enter the parameter value, select VALID, keeping the above CAUTION in mind
before doing so! To exit the modification routine, press SETUP once.
Note: Pressing the VALID key is effective ONLY when the entire value, including exponent, is
displayed. This means that if you correct a digit after making a mistake, as described
above, you must press the NEXT key repeatedly until the entire parameter value is
visible before being able to enter the new value using the VALID key.
This is the case when entering screen pair names (seven characters maximum allowed).
The principle is identical to that used for parameter value entry as described in Section
1–29
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
3–11–1 Vocabulary
The following key words are found throughout the generator installation menus:
CALIB designates a menu that steers you to the calibration procedure itself.
PARAM designates a menu that steers you to other menus that allow you to display and
modify parameters.
REF designates a menu within a calibration procedure that allows you to modify the
command input (mAs for reference energy, photo-cell HV for the frequency/HV calibration,
heater current for the heater current scale factor calibration, etc...).
CALCUL designates a menu that steers you to the execution of the computations necessary to
obtain the parameter values.
3–11–2 Use of the First Trigger/Second Trigger Buttons During Generator Installation
In general, calibration computations are started by pressing the SECOND TRIGGER button
(the FIRST TRIGGER button is not used during installation).
After going down through the different menus, the final CALIB selection steers you to a
menu that has no second line displayed on the console display window (i.e. there are no more
menus to select). It is at this level that the calibration computation is started by pressing the
SECOND TRIGGER button.
This general rule does not apply, for practical reasons, when measurements on the gantry
are necessary (because the gantry and generator/console may be physically separated, the
service engineer cannot necessarily be near both at the same time). For example:
– HV/DAC calibration. The computation is started upon selection of the START key and
stops automatically. A message appears on the console display to indicate the end of
the computation.
The AOP calibration procedures (CELL and ALGO) need to exchange information with the
gantry CPU.
Therefore, before performing these calibration procedures, it is necessary to set the value of
PRS_A to YES (see Section 3–10–1) otherwise, error messages indicating software absence
of the gantry will appear on the console display and access to the CALIB menus is inhibited
(only display and modification of parameters will be possible).
1–30
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
INSTALLATION
Many of the AEC calibration procedures are performed with grid and without grid. For such
procedures, the gantry CPU indicates to the generator CPU whether or not the grid is present.
The calibration software will verify agreement between physical presence/absence of the
grid and the service engineer’s choice of calibration with grid or without grid. If there is
disagreement, error messages will appear and access to the CALIB menu will be inhibited.
The following messages are displayed in the upper right–hand corner of the console display
during calibration procedures that make X–ray exposures:
Certain calibration procedures that make X–ray exposures insert, in alternation with the
above messages, specific messages such as:
In general the software displays, as much as possible, messages indicating the phases of
calibration such as:
Calibration end
Calcul done
SECTION 4
SITE PREPARATION
1–31
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
SECTION 5
INSTALLATION STEERING
INSTALLATION
Note: Use the installation flow chart (illustration 1–16) as a memo to assist you during the
installation.
Use calibration form (IST 043) to record all values obtained.
1. Unpack generator, gantry and accessories. See Job Card IST 008 UNPACKING THE
EQUIPMENT.
2. Physically place/anchor gantry, generator, console, protective glass screen, pedals, etc.
(see Job Card IST 009 PHYSICAL INSTALLATION).
3. Attach appropriate safety labelling to the equipment according to local regulations (see
Job Card IST 038 SAFETY LABELLING).
4. Remove all side covers and trim panels from generator and gantry (see Job Card IST
032 REMOVAL AND REPLACEMENT OF SIDE COVERS AND TRIM PANELS).
5. Inspect inside of generator and gantry for loose or disconnected cables, wires,
connectors, etc. Refer to MIS maps if necessary.
6. Connect cables between generator, console, and gantry and check the vent screw on the
HV tank(see Job Card IST 010).
7. Connect mains supply to generator and configure generator for mains voltage,
frequency and number of phases (see Job Card IST 011 ”CONNECTION OF MAINS
SUPPLY).
Do not attempt to turn on the DMR at this point. Continue following the
CAUTION Installation Steering in the chronological order provided.
8. Measure mains line resistance and determine maximum allowable generator power (see
Job Card IST 012 LINE RESISTANCE MEASUREMENT).
Note: If a high-power installation (50 kW or more) is known to exist on the same mains circuit
as the DMR being installed, it is not necessary to measure line resistance. In this case, set
maximum generator power to 5.0 kW (maximum possible value) when you reach step
13.–a.–ii below. DO NOT SET MAXIMUM GENERATOR POWER NOW. FOLLOW
THE CHRONOLOGICAL ORDER OF THE INSTALLATION STEERING.
1–32
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
9. Check basic supply voltages (see Job Card IST 022 CHECKING BASIC SUPPLY
VOLTAGES).
INSTALLATION
DO NOT ACCESS THE INSTALLATION SETUP MENUS DURING THIS
WARNING CHECK. FOLLOW THE PROCEDURE IN THE ORDER GIVEN. DO NOT
PERFORM ANY CHECKSUM OPERATIONS. EXTREMELY IMPORTANT
INFORMATION ABOUT THE INTEGRITY OF THE DMR RAM CONTENTS
CAN BE PERMANENTLY LOST IF THESE INSTRUCTIONS ARE NOT
FOLLOWED CORRECTLY.
10. Check jumpers and switches (see Job Card IST 037 JUMPERS AND SWITCHES).
11. Measure free space between floor and ceiling of the room.
If the available height is less than 2.5 meters (about 8.2 feet), perform job card IST 40
Setting of elevator upper travel limit.
12. Perform the initial power–up check.
c. Change the position of the installation menu enable switch (switch B1–8 on
generator CPU board 400–PL3, see section 3–4).
1–33
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
Starting from application mode, change the position of the installation menu en-
able switch (switch B1 on generator CPU board 400–PL3, see section 3–4).
Note: For compatibility with the information contained in this manual concerning menu paths,
select ENGL for installation of the DMR. When installation is complete, the language
will be changed to that of the country of installation.
Note: The default (and maximum possible) value is +6.000E+2 (600) mAs.
Note: The default (and maximum possible) value is +2.000E+1 (20) seconds.
v. Check X–ray tube (model, focus, number of anode metals and anode angles)
type configuration:
1–34
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
INSTALLATION
the corresponding value:
GS412: for DMR V1 and V2
70TH–F: for DMR+ and V1 or V2 upgraded with Software V6.21 and DMR+
tube
70TH–D: only SENO 2000D
If the configuration is correct, skip directly to step vii below. Otherwise, con-
tinue with step vi.
Select SETUP on the console, then verify that the console display now shows
the correct tube type as specified in step v above.
vii. Check AOP type configuration. Starting from step v above, select: SETUP/
SETUP/SETUP/AOP/TYPE
Check that Status shown is: A.
The following information on the choice of AOP type should be transmitted
to the physician. Three types are available:
Type A is AOP 4.01, as installed by default on DMR+ and DMR V2. It imple-
ments new AOP tables, with variable composition.
Type B is AOP 4.01, without variable composition. It is not widely used.
Type C is AOP 3.01, as installed on DMR V1. (no variable composition).
Recommended choice:
If there is already a DMR on site, select the same configuration.
If you have free choice, select AOP 4.01 with variable composition (type A).
If the physician requires it, select AOP 3.01 (type C)
– Enter NO when buckies 18x24 have the same thickness than the bucky
24x30 (Standard choice for DMR V1 and DMR+),
– Enter YES when buckies 18x24 are more higher than the bucky 24x30
(Standard choice for DMR V2).
1–35
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
CALIB/GRID/PARAMS/TESTS/ACCEL
– Set ACCEL parameter to 1 (1.000E+0) for DMR+.
– Set ACCEL parameter to 1 (4.000E+0) for DMR V1_P and V2_P.
iv. Set X–ray tube housing configuration.
Starting from application mode, select SETUP/ARM/VALID/INSTAL/
CONFIG/TUBE/YES on the console (the X–ray tube housing is inclinable
on all DMRs).
Perform a gantry CKSUM and return to application mode. Check that NONE
of the following error messages appears on the console display:
vii. Set lateral margin of X–ray beam. Access the arm configuration menu. Start-
ing from application mode, select:
SETUP/ARM/VALID/INSTAL/CONFIG/COLLIM/MARG_L
Then use +/– keys to set margin at desired value and select VALID.
See the table below for standards and recommendations to get full film black-
ening.
1–36
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
INSTALLATION
Then use +/– keys to set margin at desired value and select VALID.
See the table below for standards and recommendations to get full film black-
ening.
Ask the doctor for the desired maximum compression force in daN. Select
SETUP/FORCE on the console. Select + or – on the console repeatedly until
the desired force value is reached (possible values range from 4 to 20 daN in
steps of 1 daN). Once the desired value is reached, select VAL on the console
to confirm the desired value.
Note: The ACR recommendation is 18daN (40lbs).
1–37
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
Ask the doctor if a beep tone is desired. Select SETUP/BEEP on the console.
Select YES on the console if a beep tone is desired, or NO if it is not desired.
Ask the doctor for the desired maximum height during automatic decompres-
sion (between 10 and 40 cm). Select SETUP/HEIGHT on the console.
Select + or – on the console repeatedly until the desired height value is
reached. Then select VALID to confirm the value.
Ask the doctor for the desired cassette detection mode (default is “YES”).
Select SETUP/FILM on the console. Select “NO” ONLY if no cassette
detection is desired.
Note: The parameter film is reset to yes at each power up of the DMR.
The service engineer is not needed for selecting the above–mentioned gantry medical
parameters (SPEED, FORCE, DECOMP, BEEP, HEIGHT and FILM); the doctor can
easily change them at will at any time following installation of the DMR.
Note 2 : In case of error messages during the following steps, see IST 41 for explanation of these
messages.
19. Calibrate AEC for each screen pair (repeat steps a. to i. below for each screen pair).
1–38
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
INSTALLATION
CONFIGURATION.
PROCESSOR
SCREEN BRAND AND CYCLE TIME
TYPE FILM BRAND AND TYPE (SEC) A0 A1 A2
KODAK MIN–R KODAK MIN–R MA 90 +1.000E+0 +1.329E–1 +1.333E–2
” ” ” KODAK MIN–R E 90 +1.000E+0 +9.797E–2 +1.124E–2
” ” ” KODAK MIN–R H 90 +1.000E+0 +1.432E–1 +1.880E–2
” ” ” KODAK MIN–R M 90 +1.000E+0 +1.698E–1 +2.469E–2
” ” ” KODAK MIN–R 2000 90 +1.000E+0 +1.219E–1 +1.491E–2
” ” ” 3M TRIMAX HM 90 +1.000E+0 +8.479E–2 +1.210E–2
AGFA MR detail S AGFA MR5 150 +1.000E+0 +1.491E–1 +1.796E–2
” ” ” MR3–II 120 +1.000E+0 +1.170E–1 +9.586E–3
1–39
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
Enter manually the three reciprocity law failure compensation parameters (A0, A1
and A2) either from table 1 or from your own notes from a previous installation of
INSTALLATION
the same screen pair on another DMR for the screen and film of the screen pair be-
ing calibrated:
i. If the AEC calibration for the screen pair being calibrated is successful and
complete at this point (which is normally the case) be sure that this screen pair is
enabled for use in application mode (see Job Card IST 006 SCREEN PAIR
SELECTION AND CONFIGURATION.
1–40
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
INSTALLATION
21. Change console dialog display language to that of the country of installation (if other
than English):
Starting from application mode, select SETUP/GENE/MEDICAL/LANGUAGE/
language on the console, where language is the desired console dialog display lan-
guage (FRANC, ENGL, DEUTS, ESPAN, ITALI or PORTU).
22. At this step, a calibration backup can be done. Then, in case of a system crash, the
calibration backup can be used or the system can be completely recalibrated following
the process in chapter 2.
23. Perform safety checks (collimator size for example) if applicable from local
regulations.
24. Switch off the DMR to disable access to installation menus.
25. Connect room lights (optionnal), see Job card IST042 for Senographe DMR+ and
DMR V2+. For DMR V1 consult Job Card IST010.
26. Refit all covers.
1–41
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
ILLUSTRATION 1–16
DMR STEERING GUIDE FLOWCHART
INSTALLATION
Do software configuration
* of medical menus:
Unpacking the equipment ...... Language
Compression speed
IST 008 Compression max. force ......
Decomp. auto
Beep
Physical installation Film detection
......
IST 009
Do/check software configuration
Safety labelling of install menus:
...... Generator: Pmax
IST 038 MaSMax
T. Max
Removal of side covers AOP type
IST 032 ...... Tube type ......
& general inspection Gantry:Cassette detection
Bucky settings
Tube housing
Connecting the cables + ...... Collimator settings
HV tank check IST 010 Paddle deformation settings (IST 044)
1–42
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
ILLUSTRATION 1–17
FLOW CHART OF AEC/AOP CALIBRATION
INSTALLATION
DMR : FLOW CHART OF AEC/AOP CALIBRATION
CAUTION: ONE NEED
FILM/SCREEN ONLY
DO NOT PERFORM IST 017
? COUPLE
ALREADY ? YES
? TO SHIFT
AVERAGE
? NO
AT FIRST INSTALLATION CALIBRATED DENSITY
NO YES
CHECK OF VIEWBOX
& SENSITOMETRIC
STABILTY IST 047 & IST 048
A
DIFFERENT
? SCREEN
?
kV SCALE FACTOR IST 017
COMMON YES
TO HV_PM INTERNAL COUPLE CHOICE NO
VOLTMETER IST 004 (NAME, VALIDATION)
ALL CALIBRATION parag.5
COUPLES IST 006
DAC IST 004 COUPLE CHOICE
Time : 20’ CALIBRATION parag.6
Time : 5’ (NAME, VALIDATION
IST 006
PM/HV_PM
PARAMETERS DUPLICATION Time : 5’
CALIBRATION Copy only parameters without
IST 005
screen
PARAMETERS
COUPLE CHOICE IST 007 section 8 step 1 to 5
DUPLICATION IST 007
(NAME, VALIDATION Time : 5’ section 8
IST 006 step 1
Time : 5’ PM YIELD See Note 1 below to 10
PM YIELD IST 007 Perform only calibration Time : 5’
See Note 2 with screen
IST 007 parag. 6.5, 6.6, 6.7, 6.10
Time : 40’ and 6.11
kV RISE TIME IST 046 Time : 20’
A
LNRT
PARAMS
? ALREADY
? NO
YES SEE RECOMMENDATION # 1
KNOWN COUPLE CHOICE
HEATER
ENTER KNOWN CALIBRATION AT 30 mA
PARAMETERS
(A0, A1, A2) IST 013/5.5 Time : 5’
from steering Time : 10’
guide table SEE RECOMMENDATION # 2
REFERENCE ENERGY
path = AOP/ALGO/FSC = x..
/CALIB/LNRT.. REFERENCE ENERGY IST 014
/PARAM/CHANGE IST 013/5.6 Time : 20’
Time : 5’ Time : 25’
LNRT OD OPTIMIZATION
REFERENCE ENERGY SEE RECOMMENDATION # 3 Contact mode
IST 050
IST 014 IST 013/5.7 Time : 30’ to 90’
Time : 20’ Time : 45’ to 80’ OD OPTIMIZATION
Mag. mode IST 051
AOP STRATEGY
Time :0’ Automatic (See installation
section 5, parag.19.F)
REFERENCE
ENERGY
OD OPTIMIZATION FOR STEREOTIX
Time : 30’ to 90’ Contact mode IST 050 (IF APPLICABLE)
Time : 15’
STEREOTIX NOTE 1
? TO BE ? DO NOT FORGET : DO BOTH COPY I
CALIBRATED YES
STEREOTIX CELL E.G. : A no screen TO B no screen
CALIBRATION IST 018 AND
NO Time : 15’ A with screen TO B with screen
REFERENCE ENERGY
FOR STEREOTIX IST 019
NOTE 2
Time : 20’
IF LNRT PARAMS ARE UNKNOWN
PERFORM IST 13 IMMEDIATELY
CALIBRATE AFTER IST 07 (SO, WITH HOT TUBE)
END ? ANOTHER ?
NO YES
COUPLE
Time (in minutes) is an average value for indication purpose only
1–43
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
INSTALLATION
Whenever Reference O.D. is determined by the customer, take in account the following:
• Site dedicated to cancer screening campaign. See applicable standards,
To perform IST 013, it is possible to use an O.D. value other than the one expected by the doctor.
• In this case, once IST 013 has been performed, O.D. value must be adjusted with IST 014,
• This O.D. (used to complete IST 013) needs to be chosen on the linear part of the sensitometric curve with the
recommended gamma (See Job Card IST 048). Film Reciprocity Law Failure Compensation algorithm of DMR
uses a calculation range around O.D. ± 0.2.
Note: Take care to have OD ± 0.2 range inside the same linear part (i.e. with the same slope).
Note: The target O.D. is between two density steps. Use these two values to calculate the gamma value:
Gamma = (D(n+1) – D(n))/0.3. The DMR will do the calculation this way.
The result shall be identical or very close to the value found with Job card IST 048.
If not, modify the choice of O.D. steps, by shifting the step values.
– Calibrate the Reference Energy (step 5.6 of Job Card IST 013).
• Choose O.D. according to the recommendation # 1.
• For a good result, it is recommended to find manually the exact mAs value to reach the expected O.D.
with 28 kV/4 cm acrylic.
– Perform Film Reciprocity Law failure Compensation procedure: step 5.7 of IST 013.
– CAUTION: If an error 068/xxx is displayed by the Senographe console, it means that the parameters may be
erroneous. Do not attempt to force the calibration. Reset the generator and perform again the calibration.
The optical density must not vary by more than 0.2 between any of the three films; if it does, the entire
calibration must be repeated.
1–44
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
• If the O.D. used in IST 013 is different from O.D. desired by customer, perform IST 014 : Calibration of
the REFERENCE ENERGY FOR A GIVEN SCREEN/FILM PAIR.
INSTALLATION
Note: Check IST 039. Verify if the strategy computed by DMR is compatible with kV and
mAs values:
– If you decrease the strategy, e.g., Very Fast to Fast, for a same AOP mode configu-
ration, the choice of kV will be higher.
1–45
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
Blank page
INSTALLATION
1–46
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
CHAPTER 2 – CALIBRATION
SECTION 1
INTRODUCTION
This chapter is a resume of the procedures to follow when a complete re–calibration of the
generator and/or gantry is deemed necessary, when a new screen pair is being introduced,
when the Stereotix 2 option is being installed after the initial installation, and when the
customer wishes to change the reference optical density in AEC and AOP modes.
CALIBRATION
This chapter is divided into 7 sections:
SECTION 2
GENERATOR CALIBRATION
SECTION 3
GANTRY CALIBRATION
2–1
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
8. Calibrate collimator blades reference positions – see Job Card IST 030.
9. Calibrate light centering device– see Job Card IST 036 and 028.
10. Test for absence of grid lines on exposed film – see Job Card IST 031.
SECTION 4
AEC CALIBRATION FOR A GIVEN SCREEN PAIR
When a new screen pair is being introduced, follow the sequence Calibrate AEC for each
screen pair given in the Installation Steering in chapter 1, and the AEC flow chart (see
illustration NO TAG in Chapter 1).
CALIBRATION
SECTION 5
AEC CALIBRATION ON STEREOTIX 2
When the Stereotix 2 option is installed after initial installation of a Senographe DMR, follow
the sequence given below:
1. set up and calibrate the Stereotix 2 according to the Stereotix 2 Service Manual
2. calibrate photocell in Stereotix 2 positions (see Job Card IST 018)
3. Calibrate reference energy in Stereotix 2 mode (see Job Card IST 019).
SECTION 6
RE–CALIBRATION OF REFERENCE OPTICAL DENSITY IN NON–STEREOTIX 2 MODE
2–2
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
4. Perform the section entitled Check Results in Job Card IST 014.
SECTION 7
CUSTOMER REQUESTED RE–CALIBRATION OF STEREOTIX 2 MODE REFERENCE OPTICAL DENSITY
When the customer requests a re–calibration of Stereotix 2 mode reference optical density,
perform Job Card IST 019 in its entirety.
CALIBRATION
2–3
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
Blank page
CALIBRATION
2–4
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
JOB CARD IST 001 – CALIBRATION OF HEATER CURRENT SCALE FACTOR . . . . . . . . . . . . . . . . . . . . . . . . 3–3
JOB CARD IST 002 – CALIBRATION OF X–RAY TUBE FOCAL BIAS VOLTAGE SCALE FACTOR . . . . . . . 3–9
JOB CARD IST 003 – CALIBRATION OF X–RAY TUBE HEATER CURRENT . . . . . . . . . . . . . . . . . . . . . . . . . . 3–15
JOB CARD IST 004 – CALIBRATION OF PHOTO-CELL HV MEASUREMENT AND SCALE FACTOR . . . . . 3–19
JOB CARD IST 005 – CALIBRATION OF PHOTO CELL SENSITIVITY AS A FUNCTION OF ITS VOLTAGE 3–23
JOB CARD IST 006 – SCREEN PAIR SELECTION AND CONFIGURATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3–27
JOB CARD IST 007 – CALIBRATION OF PHOTO-CELL GAIN FOR A GIVEN SCREEN PAIR . . . . . . . . . . . . . 3–31
JOB CARD IST 008 – UNPACKING THE EQUIPMENT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3–51
JOB CARD IST 009 – PHYSICAL INSTALLATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3–55
JOB CARD IST 010 – WIRING . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3–57
JOB CARD IST 011 – CONNECTION OF MAINS SUPPLY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3–71
JOB CARD IST 012 – LINE RESISTANCE MEASUREMENT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3–79
JOB CARD IST 013 – CALIBRATION OF FILM RECIPROCITY LAW FAILURE COMPENSATION
JOB CARDS
FOR A GIVEN SCREEN PAIR . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3–83
JOB CARD IST 014 – CALIBRATION OF REFERENCE ENERGY FOR A GIVEN SCREEN PAIR . . . . . . . . . . 3–93
JOB CARD IST 015 – MAGNIFICATION CORRECTION FOR A GIVEN PAIR . . . . . . . . . . . . . . . . . . . . . . . . . . 3–101
JOB CARD IST 016 – CALIBRATION OF X–RAY TUBE MA MEASUREMENT . . . . . . . . . . . . . . . . . . . . . . . . . 3–107
JOB CARD IST 017 – CALIBRATION OF KV SCALE FACTOR . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3–111
JOB CARD IST 018 – CALIBRATION OF PHOTOCELL IN STEREOTIX 2 POSITIONS . . . . . . . . . . . . . . . . . . . 3–117
JOB CARD IST 019 – CALIBRATION OF REFERENCE ENERGY IN STEREOTIX 2 MODE
FOR A GIVEN SCREEN PAIR . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3–123
JOB CARD IST 020 – AUTOMATIC X–RAY TUBE WARM–UP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3–131
JOB CARD IST 021 – CHECKING AND ADJUSMENT OF LATERAL CENTERING OF X–RAY BEAM . . . . . 3–133
JOB CARD IST 022 – CHECKING BASIC SUPPLY VOLTAGES . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3–139
JOB CARD IST 023 – CALIBRATION OF COMPRESSION FORCE DETECTOR . . . . . . . . . . . . . . . . . . . . . . . . . 3–143
JOB CARD IST 024 – MINIMUM COMPRESSION/DECOMPRESSION FORCE TESTS . . . . . . . . . . . . . . . . . . . 3–145
JOB CARD IST 025 – CALIBRATION OF BREAST THICKNESS MEASUREMENT . . . . . . . . . . . . . . . . . . . . . . 3–149
JOB CARD IST 026 – CALIBRATION OF GRID POSITION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3–153
JOB CARD IST 027 – CHECKING X–RAY FILM FORMAT GEOMETRY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3–155
JOB CARD IST 027 B – CHECKING X–RAY FILM FORMAT GEOMETRY FOR NON MOTORIZED
FROM BLADE SYSTEM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3–169
JOB CARD IST 028 – ADJUSTMENT OF FRONT COLLIMATOR BLADE POSITION . . . . . . . . . . . . . . . . . . . . 3–183
JOB CARD IST 029 – CALIBRATION OF X–RAY TUBE REFERENCE ANGLE . . . . . . . . . . . . . . . . . . . . . . . . . 3–191
JOB CARD IST 030 – CALIBRATION OF LATERAL, REAR, AND FRONT, COLLIMATOR BLADE
REFERENCE POSITIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3–195
JOB CARD IST 031 – TEST FOR ABSENCE OF GRID LINES ON EXPOSED FILM . . . . . . . . . . . . . . . . . . . . . . 3–199
JOB CARD IST 032 – REMOVAL AND REPLACEMENT OF SIDE COVERS AND PANELS . . . . . . . . . . . . . . . 3–203
JOB CARD IST 033 – PRELIMINARY GENERATOR AND GANTRY TESTING . . . . . . . . . . . . . . . . . . . . . . . . . 3–207
JOB CARD IST 034 – CHECKING SAFE–LIGHTING AND FILM PROCESSING . . . . . . . . . . . . . . . . . . . . . . . . 3–213
JOB CARD IST 035 – REPLACEMENT OF CENTERING DEVICE LAMP (NEW VERSION) . . . . . . . . . . . . . . 3–217
3–1
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
JOB CARD IST 035 B – REPLACEMENT OF CENTERING DEVICE LAMP (OLD VERSION) . . . . . . . . . . . . . 3–221
JOB CARD IST 036 – ADJUSTMENT OF CENTERING DEVICE LAMP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3–225
JOB CARD IST 037 – JUMPERS AND SWITCHES . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3–235
JOB CARD IST 038 – SAFETY LABELLING . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3–237
JOB CARD IST 039 – MANUAL DETERMINATION OF AOP STRATEGY FOR A GIVEN SCREEN PAIR . . . 3–241
JOB CARD IST 040 – SETTING OF ELEVATOR UPPER TRAVEL LIMIT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3–245
JOB CARD IST 041 – GENERAL ERRORS DURING THE CALIBRATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3–249
JOB CARD IST 043 – DMR INSTALLATION FORM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3–257
JOB CARD IST 044 – SETTING PADDLE DEFORMATION PARAMETERS . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3–271
JOB CARD IST 045 – PREVENTIVE MAINTENANCE FOLLOW–UP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3–273
JOB CARD IST 046 – CHECKS ON PRE-EXPOSURE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3–277
JOB CARD IST 047 – CHECK OF VIEWBOX(ES) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3–283
JOB CARD IST 048 – CHECK OF SENSITOMETRIC STABILITY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3–287
JOB CARD IST 050 – A0 OPTIMIZATION FOR CONTACT MODE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3–291
JOB CARD IST 051 – A0 OPTIMIZATION FOR MAGNIFICATION MODE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3–293
JOB CARDS
3–2
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
SECTION 1
SUPPLIES
None
SECTION 2
TOOLS
D True R.M.S. AC ammeter
e.g: Fluke 8060, Fluke 87, Fluke 45. HP 3457A, Tektronix DMM 254.
JOB CARDS
D HV unit / cable adaptor, Part number 36003399.
IMPORTANT NOTE:
To ensure an accurate measurement:
D For ALL Digital Multimeters, the Autorange function MUST NOT be used, as it can lead to severe mea-
surement errors (>40% !!).
D For ALL Digital Multimeters, the Relative Measurement function MUST NOT be used to reset to zero
the natural offset displayed when there is no current measured. The offset is automatically reset to zero
by the principle of the rms measurement.
D Measurement configurations to be used are as follows:
a. Fluke 8060: used with 20–A shunt in the 200–mV AC range (shunt reference: 80J10).
b. Fluke 87: no shunt, in the 10–A AC range.
SECTION 3
SAFETY PRECAUTIONS
SECTION 4
PREREQUISITES
None
3–3
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
SECTION 5
CALIBRATION OF HEATER CURRENT SCALE FACTOR
The objective is to determine the scale factor K and fixed offset VOF between the X–ray tube filament current
command from the software and the real filament current value obtained. Each tube track has its own filament
supply circuit, so this calibration procedure is to be performed for each of the two tracks.
DEFAULT PARAMETER VALUES ARE K = +9.450E+0 (9.45) AND VOF = +0.000E+0 (0.0).
5.1.2 Open the HV converter control signal inputs by disconnecting flat cable WG8 from generator com-
mand board 400–PL1 connector XJ7. This eliminates any possibility of energy build–up in the HV
JOB CARDS
5.1.6 Connect the true r.m.s. AC ammeter in series with the filament being calibrated
on the HV unit / cable adaptor; and place a jumper on the other track (see illustrations 1 and 2).
Note: Keep the other filament connected in order to avoid error messages when the Senographe DMR
is switched on.
3–4
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
ILLUSTRATION 1
HV UNIT HV CABLE ADAPTOR 36003399
Track 2 (Rh): jumper or ammeter
From HV tank To X Ray tube
HV cable
shield
–HT –HT
TRACK 2 (Rh) C C TRACK 2 (Rh)
TRACK 1 (Mo) TRACK 1 (Mo)
L S S L
BIAS BIAS
G G
Ground (+HV) Ground (+HV)
JOB CARDS
C G
Bias measurement
Track 1 (Mo): jumper or ammeter
3–5
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
ILLUSTRATION 2
HV CABLE CONNECTOR PIN CONNECTION DIAGRAM
L
NOTE: THIS IS A HEAD–ON VIEW
OF THE HV CABLE CONNECTOR
AFTER BEING DISCONNECTED
FROM THE HV UNIT.
G
POLARIZING TAB C
S
JOB CARDS
The HV cable connector is marked with the standard X–ray industry markings (L, S, C and G).
However, the functions of the corresponding conductors in the Senographe DMR are as follows:
5.1.8 Change the position of the installation menu enable switch (switch B1 on generator CPU board
400–PL3, see ”Accessing the Different Installation Menus in the Tree Structure” in chapter 1).
5.2 Perform the calibration procedure for the desired X–ray tube track.
Note: The default filament current command value for this first point is +4.000E+00 (4.0) A.
3–6
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
5.2.2 Press the 2nd trigger button (no X–rays are produced) and hold it until the ammeter reading stabi-
lizes (this may take a few seconds).
5.2.3 Note the filament current value measured from the ammeter.
5.2.6 Enter and validate the measured filament current value from the ammeter for this first calibration
point.
5.2.7 Select SETUP/SETUP/2nd pt/CALIB on the console for the first calibration point.
Note: The default filament current command value for this second point is +5.000E+00 (5.0) A.
JOB CARDS
5.2.8 Press the 2nd trigger button (no X–rays are produced) and hold it until the ammeter reading stabi-
lizes (this may take a few seconds).
5.2.9 Note the filament current value measured from the ammeter.
5.2.12 Enter and validate the measured filament current value from the ammeter for this second calibration
point.
5.2.13 Select SETUP/SETUP/calcul/VALID on the console. This executes the calculation of the scale
factor and fixed offset of the filament current command from the software.
Note: Again, the default filament current command value for this first point is +4.000E+00 (4.0) A.
5.3.2 Press the 2nd trigger button (no X–rays are produced) and hold it until the ammeter reading stabi-
lizes (this may take a few seconds).
5.3.3 Note the filament current value measured from the ammeter.
3–7
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
Note: Again, the default filament current command value for this second point is +5.000E+00 (5.0) A.
5.3.6 Press the 2nd trigger button (no X–rays are produced) and hold it until the ammeter reading stabi-
lizes (this may take a few seconds).
5.3.7 Note the filament current value measured from the ammeter.
5.3.9 The expected results are 4.0 ± 0.02 A for the first point and 5.0 ± 0.02 A for the second point. If the
results are within these tolerances, the calibration for this X–ray tube track is finished. If the results
are out of tolerance, repeat the calibration procedure in section 5.2 and the check procedure in sec-
tion 5.3 for the same X–ray tube track.
JOB CARDS
Note: Performing the calibration procedure once is normally sufficient for correct calibration.
Note: If it is necessary to perform the calibration procedure more than once, the measured values of cur-
rent for the two points (4.0 and 5.0 A) should become increasingly accurate. If not, the procedure
is probably not being followed correctly.
5.3.10 Select SETUP/SETUP/SETUP/PARAM on the console. Note down the calculated values of scale
factor K and offset VOF. Each of these two values can be displayed alternately by rotating the kV
dial on the console.
If the value of K is less than +8.000E+0 (8.0), either the calibration procedure
CAUTION was done incorrectly or there is a hardware defect.
5.3.11 Repeat the calibration and check procedures on the other tube track, if necessary.
5.3.13 Re–connect flat cable WG8 to generator command board 400–PL1 connector XJ7, remove high
voltage cable adapter. Do not forget to secure the HV ring with the Allen screw, to insure proper
ground continuity.
3–8
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
SECTION 1
SUPPLIES
None
SECTION 2
TOOLS
D Voltmeter
D HV unit / cable adaptor, part number 36003399
JOB CARDS
SECTION 3
SAFETY PRECAUTIONS
SECTION 4
PREREQUISITES
D Job Card IST 001 CALIBRATION OF HEATER CURRENT SCALE FACTOR
3–9
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
SECTION 5
CALIBRATION OF X–RAY TUBE FOCAL BIAS VOLTAGE SCALE FACTOR
The objective is to determine the scale factor alpha and offset beta between the X–ray tube focal bias voltage
command from the software and the real focal bias voltage value obtained. The focal bias voltage applied to
the X–ray tube electron beam concentration device establishes the focal spot width.
DEFAULT PARAMETER VALUES ARE ALPHA = +9.389E+1 (93.89) AND BETA = +3.801E–2
(0.03801).
5.1.2 Open the HV converter control signal inputs by disconnecting flat cable WG8 from generator com-
mand board 400–PL1 connector XJ7. This eliminates any possibility of energy build–up in the HV
JOB CARDS
5.1.6 Connect a DC voltmeter between common (C) and the focal bias voltage wire (G) (see illustrations
1 and 2).
3–10
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
ILLUSTRATION 1
HV UNIT HV CABLE ADAPTOR 36003399
–HT –HT
TRACK 2 (Rh) C C TRACK 2 (Rh)
TRACK 1 (Mo) TRACK 1 (Mo)
L S S L
BIAS BIAS
JOB CARDS
G G
Ground (+HV) Ground (+HV)
C G
Bias measurement
Track 1 (Mo): jumper or ammeter
3–11
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
ILLUSTRATION 2
HV CABLE CONNECTOR PIN CONNECTION DIAGRAM
L
NOTE: THIS IS A HEAD–ON VIEW
OF THE HV CABLE CONNECTOR
AFTER BEING DISCONNECTED
FROM THE HV UNIT.
G
POLARIZING TAB C
S
JOB CARDS
The HV cable connector is marked with the standard X–ray industry markings (L, S, C and G).
However, the functions of the corresponding conductors in the Senographe DMR+ are as fol-
lows:
5.1.8 Change the position of the installation menu enable switch (switch B1 on generator CPU board
400–PL3, see Accessing the Different Installation Menus in the Tree Structure in chapter 1).
3–12
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
5.1.10 Rotate the kV dial in order to check that all four focal bias voltage command values are the same
as the values specified on the label on the X-ray tube housing:
MO/LF = . . . . . . . . . . .
RH/LF = . . . . . . . . . . . .
MO/SF = . . . . . . . . . . .
RH/SF = . . . . . . . . . . . .
Modify any values that appear different from these on the tube label by using the
CAUTION CHANGE key, rotating the kV dial and using the NEXT, VALID and SETUP
keys. See ”Visual Display of Parameter Values” and ”Modification of a Parameter
Value or Entry of a Measurement” in chapter 1.
JOB CARDS
5.2 Perform the calibration procedure.
5.2.1 Select SETUP/CALIB/1st pt/CALIB on the console for the first calibration point.
5.2.2 Press the 2nd trigger button (no X–rays are produced) and hold it until the voltmeter reading stabi-
lizes.
5.2.3 Note the focal bias voltage value measured from the voltmeter.
5.2.6 Enter and validate the absolute value of measured focal bias voltage value from the voltmeter for
this first calibration point. See ”Entry of an Alpha–Numeric Value” in chapter 1 for information on
entering alpha–numeric values.
5.2.7 Select SETUP/SETUP/2nd pt/CALIB on the console for the second calibration point.
5.2.8 Press the 2nd trigger button (no X–rays are produced) and hold it until the voltmeter reading stabi-
lizes.
5.2.9 Note the focal bias voltage value measured from the voltmeter.
5.2.12 Enter and validate the measured focal bias voltage value from the voltmeter for this second calibra-
tion point.
3–13
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
5.2.13 Select SETUP/SETUP/calcul/VALID on the console. This executes the calculation of the parame-
ters alpha and beta. The message ”CALCUL DONE” appears on the console display.
5.3.2 Press the 2nd trigger button (no X–rays are produced) and hold it until the voltmeter reading stabi-
lizes.
5.3.3 Note the focal bias voltage value measured from the voltmeter for the first point.
5.3.4 Release the 2nd trigger button. DO NOT ENTER THIS MEASURED VALUE.
5.3.6 Press the 2nd trigger button (no X–rays are produced) and hold it until the voltmeter reading stabi-
lizes.
5.3.7 Note the focal bias voltage value measured from the voltmeter for the second point.
5.3.8 Release the 2nd trigger button. DO NOT ENTER THIS MEASURED VALUE.
5.3.9 The expected results for the first point are the value of MO/LF ±5% for the first point and the value
of MO/SF ±5% for the second point (see section 5.1.10 for these values). If the results are within
these tolerances, the focal bias voltage calibration is finished. If the results are out of tolerance,
repeat the calibration procedure in section 5.2 and the check procedure in section 5.3.
Note: Performing the calibration procedure once is normally sufficient for correct calibration.
Note: If it is necessary to perform the calibration procedure more than once, the measured values of focal
bias voltage for the two points should become increasingly accurate. If not, the procedure is prob-
ably not being followed correctly.
5.3.10 Select SETUP/SETUP/SETUP/PARAM on the console. Note down the calculated values of
parameters alpha and beta. Each of these two values can be displayed alternately by rotating the kV
dial on the console.
5.3.12 Re–connect flat cable WG8 to generator command board 400–PL1 connector XJ7, remove high
voltage cable adapter. Do not forget to secure the HV ring with the Allen screw, to insure proper
ground continuity.
3–14
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
SECTION 1
SUPPLIES
None
SECTION 2
TOOLS
None
JOB CARDS
SECTION 3
SAFETY PRECAUTIONS
SECTION 4
PREREQUISITES
The following procedures must have been performed in the following order:
D Job Card IST 001 CALIBRATION OF HEATER CURRENT SCALE FACTOR
D Job Card IST 002 CALIBRATION OF X–RAY TUBE FOCAL BIAS VOLTAGE SCALE FACTOR
D Job Card IST 016 CALIBRATION OF X–RAY TUBE MA MEASUREMENT
D Job Card IST 017 CALIBRATION OF kV SCALE FACTOR
3–15
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
SECTION 5
CALIBRATION OF X–RAY TUBE HEATER CURRENT
The objective is to determine the X–ray tube heater current necessary for producing the desired X–ray tube
anode current at a given kV value.
The calibration procedure must be performed on each focal size/track combination. Four repetitions of the cal-
ibration procedure are therefore necessary.
Three exposures are taken automatically at different kV values. Using the results, the software calculates the
necessary parameters to determine the X–ray tube heater current needed to produce the desired X–ray tube
anode current for all possible kV values.
5.1.1 Starting from application mode, change the position of the installation menu enable switch (switch
B1 on generator CPU board 400–PL3) or enter password on console.
JOB CARDS
Note: This must only be done if the Senographe DMR has been switched off since the last access to the
service menu.
Note: /focal size/track/ must be SF/RH for the first combination performed.
To avoid excessive X-ray doses on the photo-cell tube, place an empty cassette
CAUTION on the top of the image receptor.
5.1.3 Press the 2nd trigger button and hold it down. The messages EXPOSURE NO 1, EXPOSURE NO
2, EXPOSURE NO 3 and CALIBRATION END appear sequentially on the console display. The
software automatically performs the three exposures and the calculations.
Note: Calibration time varies according to tube temperature. The message TOO HOT appears on the
console display when tube temperature is excessive for taking the next exposure. It is not neces-
sary to hold the 2nd trigger button down while waiting for the tube to cool. However, do not exit
the calibration menu while waiting. As soon as the TOO HOT message disappears, continue the
calibration by pressing the 2nd trigger button and holding it down.
Note: In case of arcing, the software re–attempts the same exposure automatically.
Note: The calibration procedure includes an exposure at 40 kV. If for some reason this exposure is
impossible to perform (e.g. excessive arcing), tube warm–up may be necessary (see Job Card
IST 020 AUTOMATIC X–RAY TUBE WARM–UP).
3–16
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
5.1.4 Repeat sections 5.1.2 to 5.1.3 for each of the remaining focal size/track combinations in the follow-
ing order: SF/Mo, LF/Rh, LF/Mo.
1. Access the DMR installation menu to read the heater parameters (the path is: SETUP/GENE/
INSTAL/GENE/GENE/TUBE/HEATER/LF/MO/Param/I–Value). If necessary, see DMR
menu structure in service manual (Ill. 1–7).
JOB CARDS
2. Explore the 28 kV/A combinations and note the kV values showing a current value equal to
5.600 A.
3. Repeat Step 2. for the LF/RH combination (there is a very low probability of current limitation
occurring in small focus).
5. Set the DMR for 2–point exposure technique at 60 mAs. Make an exposure for each kV point
detected at 5.600 A (LF/MO and LF/RH). Allow at least 60 seconds between each exposure.
6. Repeat Steps 1. and 2.: If no more kV point are set to 5.600 A, the problem has been resolved.
If there is still one or more points showing a current value of 5.600 A make Job Card IST 001,
then restart this procedure from Step 1 above. If the problem recurs, replace the X–ray tube.
5.2.1 Be sure that the gantry presence parameter PRES_A is set to YES before proceeding (see Gantry
Present/Absent in Chapter 1).
5.2.2 Activate the display of parameters following an exposure (see Display of Parameters Following an
Exposure in Chapter 1).
5.2.3 Perform a CKSUM and return to application mode (Path is: SETUP/GENE/INSTALL/CKSUM/
CKSUM.
5.2.4 Set up the Senographe DMR for a manual (2–point) exposure at 30 kV and 63 mAs. For each of
the four focal size/track combination, take an exposure following a 3-second preparation time. Note
the resulting X–ray tube anode currents displayed on the console display.
Note: A lower mAs value cannot would decrease the measurement accuracy.
3–17
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
5.2.5 The table below shows the expected X–ray tube anode current for each focal size/track combina-
tion:
Value available for Maxiray 70TH–F tube:
LF/RH – 62 mA ± 5% ± DVM accuracy
SF/RH – 35 mA ± 5% ± DVM accuracy
LF/MO – 100 mA ± 5% ± DVM accuracy
SF/MO – 40 mA ± 5% ± DVM accuracy
5.2.6 If you are handing the machine back to the user at this point, switch the Senographe DMR off, then
on again.
3–18
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
SECTION 1
SUPPLIES
None
SECTION 2
TOOLS
D Voltmeter
JOB CARDS
SECTION 3
SAFETY PRECAUTIONS
These procedures must be done with cables W14 and W15 connected to
CAUTION gantry photo-cell board 800–PL4 connectors XJ3 and XJ2, respectively.
Note: These two calibration procedures must always be done together and in the order presented. Follow
the procedure in section 5 and then the procedure in section 6.
SECTION 4
PREREQUISITES
None
3–19
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
SECTION 5
CALIBRATION OF PHOTO-CELL HV MEASUREMENT
The objective is to determine the scale factor F/HV between the real photo-cell HV value and the photo-cell
HV measurement input to the software. This is accomplished by sending a theoretical photo-cell HV command
value of +7.000E+2 (700) V from the software and measuring the real voltage obtained. The measured value
is entered and the software calculates the value of F/HV.
DEFAULT PARAMETER VALUE IS F/HV = +1.797E+2 (179.7) Hz/V
5.1.1 Starting from application mode, change the position of the installation menu enable switch (switch
B1 on generator CPU board 400–PL3) or enter password on console.
Note: This must only be done if the Senographe DMR has been switched off since the last access to SET-
JOB CARDS
UP/GENE/INSTAL or SETUP/ARM/VALID/INSTAL
5.1.2 Carefully connect a voltmeter set to 1000 VDC to the photo-cell HV (+ve lead to the terminal lug
on the frame of the gantry photo-cell board 800–PL4 connector XJ2, and –ve lead to the gantry pho-
to-cell board 800–PL4 test point TP1, marked –HT).
3–20
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
5.2.2 Wait for the voltmeter reading to stabilize, and note its absolute (positive) value.
5.2.3 Select STOP – the message Done appears on the console display.
5.2.5 Enter and validate the absolute (positive) value of the voltmeter reading by using the NEXT and
VALID keys and the kV dial (see Visual Display of Parameter Values and Modification of a Parame-
ter Value or Entry of a Measurement in Chapter 1).
5.2.6 Select SETUP/CALC/VALID – the message Calcul done appears on the console display.
JOB CARDS
5.3 Check the calibration accuracy.
5.3.1 Activate the display of parameters following an exposure by selecting SETUP six times, then
CONFIG/DISP/YES (see Chapter 1, Display of Parameters Following an Exposure).
5.3.3 Press the 1st trigger button and hold it down until the voltmeter reading stabilizes. Note the absolute
(positive) value of the voltmeter reading. Release the 1st trigger button.
5.3.4 Check that the photo-cell HV value displayed on the console display is within ±1%, ±0.1V, plus
voltmeter error, of the voltmeter reading taken in step 5.3.3.
SECTION 6
CALIBRATION OF PHOTO-CELL HV SCALE FACTOR
The objective is to determine the scale factor alpha and offset beta between the photo-cell HV command value
from the software and the real photo-cell HV value obtained. The calibration is fully automatic.
DEFAULT PARAMETER VALUES ARE ALPHA = +2.567E–1 (0.2567) AND BETA = +2.184E+0 (2.184).
6.1.1 Select SETUP/SETUP/ HV/DAC /CALIB/START – the messages Working, then Done appear
on the console display.
6.1.2 Select SETUP/PARAM. Write down the calculated values of scale factor alpha and offset beta.
Each of these two values is displayed alternately by rotating the kV dial on the console.
3–21
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
6.2.2 Enter and validate a photo-cell HV reference value of +4.000E+2 (400) V by using the NEXT and
VALID keys and the kV dial (see Visual Display of Parameter Values and Modification of a Parame-
ter Value or Entry of a Measurement in Chapter 1).
6.2.3 Select SETUP/CALIB/START – the message Working appears on the console display.
6.2.4 Wait for the voltmeter reading to stabilize. The voltmeter reading must be 400 ±5V. Do not enter
this measured value.
6.2.5 Select STOP – the message Done appears on the console display.
6.2.7 Enter and validate a photo-cell HV reference value of +6.000E+2 (600) V by using the NEXT and
VALID keys and the kV dial.
6.2.8 Select SETUP/CALIB/START – the message Working appears on the console display.
6.2.9 Wait for the voltmeter reading to stabilize. The voltmeter reading must be 600 ±5V. Do not enter
this measured value.
6.2.10 Select STOP – the message Done appears on the console display.
6.2.11 Repeat steps 6.2.6 to 6.2.10 using a photo-cell HV reference value of +8.000E+2 (800) V. In this
case, the resulting voltmeter reading must be 800 ±5V.
6.2.12 Perform a CKSUM and return to application mode. If you are handing the machine back to the user
at this point, switch the Senographe DMR off, then on again.
3–22
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
SECTION 1
SUPPLIES
None
SECTION 2
TOOLS
D Plexiglass in thickness increments of 0.5 cm (minimum plexiglass dimensions 20 x 20 cm to insure com-
plete covering of the photo cell)
JOB CARDS
SECTION 3
SAFETY PRECAUTIONS
SECTION 4
PREREQUISITES
D Job Card IST 004 ”CALIBRATION OF PHOTO CELL HV MEASUREMENT AND SCALE FAC-
TOR”
D Remove the protective film from bucky top plate.
Note: In case of error messages during this calibration, see IST 041 for explanations.
3–23
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
SECTION 5
CALIBRATION OF PHOTO CELL SENSITIVITY AS A FUNCTION OF ITS VOLTAGE
The photo cell’s sensitivity to light varies as a function of its voltage. This phenomenon is exploited by the
Senographe DMR to compensate for all the different possible configurations encountered in application mode
(kV, SID, etc.). The Senographe DMR determines the needed photo cell sensitivity for a given configuration
and applies the appropriate photo cell HV.
The objective here is to determine the coefficients beta and gamma that define the relationship between photo
cell HV and photo cell sensitivity. A third coefficient, alpha, is always equal to zero.
DEFAULT PARAMETER VALUES ARE ALPHA = +0.000E+0 (0), BETA = +7.377E+0 (7.377) AND
GAMMA = –4.931E+1 (–49.31).
To avoid excessive X–Ray doses on the photo–cell tube, place an empty cas-
CAUTION sette on the top of the image receptor.
5.1.2 Starting from application mode, change the position of the installation menu enable switch (switch
B1 on generator CPU board 400–PL3).
Note: This must only be done if the Senographe DMR has been switched off since the last access to SET-
UP/GENE/INSTAL or SETUP/ARM/VALID/INSTAL
Note: To insure that the photo cell is fully covered by the plexiglass, be sure that it overlaps the front
edge of the cassette holder by about 1 cm.
3–24
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
5.2.2 Press the 2nd trigger button and hold it down. A series of X–ray exposures is taken. If more, or less,
plexiglass is needed to successfully perform the calibration, the message ”not enough plex” or ”too
much plexi” appears on the console display, and the series of exposures is halted.
If this happens:
– Release the 2nd trigger button and, according to the message, either add or remove
1 cm of plexiglass,
– Select SETUP/CALIB on the console,
– Press the 2nd trigger button and hold it down again,
JOB CARDS
– Continue the procedure.
5.2.3 When the calibration is successful and complete, the series of X–ray exposures halts automatically
and the message ”calibration end” appears on the console display. At this point, release the 2nd trig-
ger button.
5.2.4 Select SETUP/PARAM on the console. Note down the calculated values of parameters beta and
gamma. Each of these values is displayed by rotating the kV dial on the console. Since the third
parameter, alpha, always has a value of zero, it is not necessary to note it down.
Note: The value of alpha MUST be zero. Do not modify it under any circumstances.
5.2.5 Perform a ”CKSUM” and return to application mode. If you’re turning the machine back over to
the user at this point, switch the Senographe DMR off, then on again.
3–25
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
Blank page
JOB CARDS
3–26
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
SECTION 1
SUPPLIES
None
SECTION 2
TOOLS
None
JOB CARDS
SECTION 3
SAFETY PRECAUTIONS
SECTION 4
PREREQUISITES
None
3–27
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
SECTION 5
SCREEN PAIR SELECTION AND CONFIGURATION
The objective is to select and configure a screen pair prior to performing AEC calibrations on it.
Note: This must only be done if the Senographe DMR has been switched off since the last access to SET-
UP/GENE/INSTAL or SETUP/ARM/VALID/INSTAL
to be calibrated.
Note: Do not confuse the screen pair indicator with the screen pair name, which is chosen via the proce-
dure in section 5.6. While the 5 screen pair indicators cannot be changed and do not appear in
application mode, each of them can be assigned a screen pair name corresponding to the film/cas-
sette combination used, which appears on the console display in application AOP (0–point) and
AEC (1–point) modes.
Note: This default AOP strategy value will be verified and changed, if necessary, at the end of screen
pair calibration. This value has no effect on screen pair calibration.
3–28
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
Note: The default screen pair name is the screen pair indicator. See section 5.2.
Typically, this name is an abbreviation of the film and cassette type used to constitute the screen pair being
used. It is a good idea to ask the doctor for suggestions for the screen pair name.
Note: This name can be changed at any time in the future without producing the need to re-perform any
calibrations on the screen pair in question.
JOB CARDS
Once you have completed the procedures in this Job Card, perform a CKSUM, return to application mode and
proceed directly to Job Card IST 007, Calibration of Photo Cell Gain for a Given Screen Pair.
3–29
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
Blank page
JOB CARDS
3–30
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
SECTION 1
SUPPLIES
D Normal cassette of type used in the screen pair being calibrated (same cassette for all the procedure)
D Film of type used in the screen pair being calibrated.
SECTION 2
TOOLS
D Screen–less cassette simulation tool OR cassette of type used in the screen pair being calibrated, but with-
JOB CARDS
out its screen (see section 5.2)
D Plexiglass in thickness increments of 10 mm (minimum plexiglass dimensions 200 x 200 mm to ensure
complete coverage of the photo-cell)
Tool reference : 45 203 014 or 46–286893P1 (US pole).
Thickness accuracy of plexiglass plates is mandatory (10 0.2 mm)
SECTION 3
SAFETY PRECAUTIONS
SECTION 4
PREREQUISITES
D Job Card IST 006 SCREEN PAIR SELECTION AND CONFIGURATION
D About checksum, see chapter 1, para. 3–7.
D kV accuracy check (if applicable) followed by kVp recalibration must be performed before this job
card (see IST 033, para. 5.1.11).
D Job Cards IST 004 and IST005 must have been successfully performed.
D In case of error messages during this calibration, see IST 041 for explanations.
3–31
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
SECTION 5
ADVANCE PREPARATION AND INFORMATION
START
YES YES
DMR+ set–up for the contact mode:
Sections 6.1/6.2
END
END
3–32
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
5.2.1 Determine whether or not the screen–less cassette simulation tool can be used.
The screen–less simulation tool can be used only if the cassette of the screen pair being calibrated
is among the following:
– KODAK Min–R 2
– FUJI MA
– AGFA–GEVAERT MAMORAY
– 3M
– DUPONT
– KONICA
If the cassette of the screen pair being installed is not among these models, the screen–less cassette
JOB CARDS
simulation tool cannot be used. In this case, a normal cassette of the type used in the screen pair
being calibrated must be used, but without its screen.
Note: If a screen–less cassette simulation tool is not available, any of the cassette models listed above,
without its screen*, can be used during calibration of any screen pair that uses any of the models
in the list. For example, if you’re calibrating a screen pair that uses the 3M cassette, you can use
a KODAK Min–R 2 cassette without its screen. However, if you are calibrating a screen pair that
uses a cassette model that is not on this list, you must use a cassette of that exact model, without
its screen.
* When removing the screen from a cassette, be sure to leave the layer of plastic foam intact in
the cassette.
3–33
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
Note: It is normal for the film not to lie flat at the rear edge of the film holding area. The rear edge of
the film holding area was in fact designed to insure that the FRONT edge of the film butts up per-
fectly against the front edge of the film holding area. This is critical, because the photo-cell is
located towards the front edge of the bucky and must be maximally covered by the film.
The loaded screen–less cassette simulation tool must be installed in the Senographe DMR cassette
holder or in the bucky in the same way as an ordinary cassette.
Be sure to orient the simulation tool in the cassette holder with its front edge
CAUTION coinciding with the front edge of the bucky, as shown in illustration 1.
JOB CARDS
Note: Whenever the instructions in this Job Card instruct you to install the loaded cassette (without
screen) install the loaded simulation tool instead, if permitted (see section 5.2.1 above).
5.2.3 Each of the two cassettes (with and without screen) used in these procedures must be loaded with
undeveloped film of the type used in the screen pair being calibrated.
– The same two cassettes must be used for all procedures in this Job Card.
– The same undeveloped films can be left in the cassettes for all procedures in this Job
Card.
– Mark the cassette without screen clearly so as not to confuse it with the cassette with
screen (not applicable if screen–less cassette simulation tool is used).
3–34
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
ILLUSTRATION 1
SCREEN–LESS CASSETTE SIMULATION TOOL
Rear edge of
film does not
lie flat against
simulation tool
JOB CARDS
Front edge (towards patient)
Undeveloped film
3–35
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
5.4 General instructions to be observed during all procedures in this Job Card
5.4.1 The photo-cell must be in its most forward position (towards the patient) throughout the procedures
in this Job Card.
Note: Number the plexiglass plates from 1 to 6 and use the same plates for the same thickness.
5.4.2 When selecting plexiglass thicknesses (2, 4 or 6 cm) in the menus between exposure sequences,
never go up in the menu structure higher than indicated in the procedure being performed. Doing
so will reset stored intermediate values and ruin the calibration. Also, once all three thicknesses
have been calibrated, follow the instructions immediately for performing the calculation. Select
SETUP no more than the precise number of times indicated, again at the risk of losing intermediate
values prior to the calculation.
5.4.3 On the other hand, if an error is made during the calibration of one of the three plexiglass thicknesses
JOB CARDS
(for example, the plexiglass was incorrectly placed or is not the correct thickness), then you must
go upward in the menus (by selecting SETUP a certain number of times) at least to the level that
gives the choices GRID, SCREEN and CALIB (just below PM_YLD). Here you reset the inter-
mediate values intentionally, and you must repeat all three thickness calibrations.
5.4.4 Normally, this Job Card is used when a new screen pair is being introduced. In this case, follow the
procedures in this Job Card in the exact order shown.
5.4.5 If, however, only part of the photo-cell gain calibration is being done (e.g. because of a mistake dur-
ing a certain part of the calibration), the procedures can be performed independently of the others,
but be sure to read and carry out all preliminary preparations and procedures associated with the
procedure in question.
5.4.6 When placing plexiglass on the cassette holder or 1.5 magnification device, the plexiglass must
always overlap the front edge (towards the patient) by about 1 cm to insure that the photo-cell is
fully covered.
5.4.7 When taking a series of calibration exposures, if arcing occurs, the arcing message appears on the
console display. This does not mean that the series must be repeated (i.e. the software keeps track
of which exposures in the series were successful even if it must repeat one due to arcing). Only if
there were a serious arcing problem would the software abort the series, in which case it would have
to be repeated.
5.4.8 If you are in the middle of IST 007, and must stop, continue at another time, there are certain steps
that you need not repeat. For example:
D if you have completed step 6.4.13, then you can start with step 6.5
D if you have completed step 6.6.13, then you can start with step 6.7
D if you have completed step 6.9.12, then you can start with step 6.10.
3–36
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
SECTION 6
CALIBRATION OF PHOTO-CELL GAIN
The objective is to determine the X–ray absorption of the screen. The calibration software achieves this by sub-
tracting the photo-cell gain obtained with a screen from the photo-cell gain obtained without a screen.
Therefore, each configuration must be calibrated twice – once with and once without a screen.
6 coefficients are calculated for each of 5 track/filter combinations. Thus, 30 coefficients are calculated for
each of 4 different cases (2 configurations, each calibrated once with and once without a screen), for a total
of 120 coefficients.
6.1.1 Starting from application mode, change the position of the installation menu enable switch (switch
B1 on generator CPU board 400–PL3).
JOB CARDS
Note: This must only be done if the Senographe DMR has been switched off since the last access to SET-
UP/GENE/INSTAL or SETUP/ARM/VALID/INSTAL
6.1.2 Set the SID to that used by the user, 610 mm or the most common 660 mm.
6.2 Prepare the Senographe DMR for the procedures using contact mode with grid (bucky 18 x 24 ).
6.2.3 Select CALIB/PM_yld/GRID/YES on the console. This sets up the calibration for the with grid
procedure.
3–37
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
6.3 Perform the procedure using contact mode with grid (bucky 18 x 24 ) and cassette without screen.
6.3.1 Be sure to be familiar with the advance preparation and information given in section 5 before pro-
ceeding, especially the information concerning the use, if permitted, of the screen–less cassette sim-
ulation tool. Also, be sure to prepare the Senographe DMR by following sections 6.1 and 6.2 before
proceeding.
6.3.2 Select SETUP/SCREEN/NO on the console. This sets up the calibration for the cassette without
screen procedure.
6.3.6 Select SETUP/CALIB/CALIB/THICK/2 cm on the console. This tells the software the amount
of plexiglass present.
6.3.8 Press the 2nd trigger button and hold it down. A series of X–ray exposures is taken (at least 25).
6.3.9 When the series of exposures is complete, the message thick plexi end appears on the console dis-
play. At this point, release the 2nd trigger button.
6.3.10 Raise the compression paddle and increase the amount of plexiglass to 4 cm. Select SETUP/
THICK/4 cm on the console.
6.3.11 Lower the compression paddle onto the plexiglass. Repeat sections 6.3.7 to 6.3.9.
6.3.12 Raise the compression paddle and increase the amount of plexiglass to 6 cm. Select SETUP/
THICK/6 cm on the console.
6.3.13 Lower the compression paddle onto the plexiglass. Repeat sections 6.3.7 to 6.3.9.
6.3.14 Select SETUP/CALC/VALID on the console. This executes the calculation of the 6 coefficients
(A0, A1, A2, B0, B1, and B2) for each of 5 track/filter combinations. The message PM_yld cal end
appears on the console display.
3–38
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
6.4 Check the contact mode with grid (bucky 18 x 24 ) and cassette without screen calibration accuracy.
Note: This check must always be made immediately after performing the calibration procedure in sec-
tion 6.3.
6.4.1 Copy the without screen parameters for the screen pair being calibrated to the with screen parameter
location of an unused screen pair (generally E) by following this procedure:
– Starting from application mode, select SETUP/GENE/INSTAL/GENE/AOP/
FILM/COPY/SOURCE/SCREEN/NO on the console.
– Select SETUP/FSC/FSC=x, where x is the screen pair indicator corresponding to the
screen pair being calibrated.
– Select SETUP/SETUP/TARG/SCREEN/YES.
– Select SETUP/FSC/FSC=y, where y is the indicator of the unused screen pair.
JOB CARDS
Be very sure of the choice of unused screen pair indicator y (A, B, C, D or E).
CAUTION If a mistake is made, there is a risk of erasing good parameter values on
another screen pair that is in use.
– Once you are SURE that the screen pair indicator y is correct (see CAUTION above),
copy the parameters by selecting SETUP/SETUP/COPY/VALID. The message
COPY EXECUTED appears on the console display.
6.4.2 If necessary, enable use of the unused screen pair. (From application mode, select SETUP/GENE/
INSTAL/GENE/AOP/FILM/ALGO/FSC=y/VALID/YES on the console.)
6.4.3 Enable visual display of parameter values when in application mode. See Display of Parameters
Following an Exposure in chapter 1, section 3–10–2.
6.4.5 Set up the Senographe DMR in the configuration given in section 6.2.1.
3–39
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
6.4.10 Take a short exposure. It is not necessary for the exposure to be complete, so the 2nd trigger button
can be released as soon as X–rays are being produced.
Note: Even if exposure is aborted due to insufficient available mAs, the thickness displayed on the con-
sole is useable. It is NOT USEFUL to wait for the tube to cool and repeat the exposure.
JOB CARDS
6.4.11 Note in the Job Card IST043 (IST007 data section) the thickness displayed on the console after the
exposure. It must be equal to the plexiglass thickness ±2 mm.
6.4.13 If the displayed thickness is outside the tolerance given in section 6.4.11, first check that the correct
screen pair was selected in section 6.4.9. Otherwise, the corresponding calibration procedure (con-
tact mode with grid and cassette without screen, section 6.3) was probably done incorrectly. That
procedure, as well as this check procedure immediately following it, would have to be repeated (this
can be done independently of the other procedures in this Job Card).
6.5 Perform the procedure using contact mode with grid (bucky 18 x 24 ) and cassette with screen.
6.5.1 Be sure to be familiar with the advance preparation and information given in section 5 before pro-
ceeding. Also, be sure to prepare the Senographe DMR by following sections 6.1 and 6.2 before
proceeding.
3–40
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
6.5.6 Select CALIB/CALIB/THICK/2 cm on the console. This tells the software the amount of plexi-
glass present.
6.5.8 Press the 2nd trigger button and hold it down. A series of X–ray exposures is taken (at least 25).
6.5.9 When the series of exposures is complete, the message ”thick calib end” appears on the console dis-
play. At this point, release the 2nd trigger button.
6.5.10 Raise the compression paddle and increase the amount of plexiglass to 4 cm. Select SETUP/
THICK/4 cm on the console.
JOB CARDS
6.5.11 Lower the compression paddle onto the plexiglass. Repeat sections 6.5.7 to 6.5.9.
6.5.12 Raise the compression paddle and increase the amount of plexiglass to 6 cm. Select SETUP/
THICK/6 cm on the console.
6.5.13 Lower the compression paddle onto the plexiglass. Repeat sections 6.5.7 to 6.5.9.
6.5.14 Select SETUP/CALC/VALID on the console. This executes the calculation of the 6 coefficients
(A0, A1, A2, B0, B1, and B2) for each of 5 track/filter combinations. The message PM_yld cal end
appears on the console display.
6.6 Check the contact mode with grid (bucky 18 x 24 ) and cassette with screen calibration accuracy.
Note: This check must always be made immediately after performing the calibration procedure in sec-
tion 6.5.
6.6.1 Enable visual display of parameter values when in application mode. Display of Parameters Fol-
lowing an Exposure in chapter 1, section 3–10–2.
6.6.2 If necessary, enable use of the screen pair being calibrated. (From application mode, select SETUP/
GENE/INSTAL/GENE/AOP/FILM/ALGO/FSC=x/VALID/YES on the console.)
6.6.4 Select the screen pair being calibrated and verify that its name is as entered in Job Card IST 006
SCREEN PAIR SELECTION AND CONFIGURATION
6.6.5 Set up the Senographe DMR in the configuration given in section 6.2.1.
3–41
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
6.6.10 Take a short exposure. It is not necessary for the exposure to be complete, so the 2nd trigger button
can be released as soon as X–rays are being produced.
Note: Even if exposure is aborted due to insufficient available mAs, the thickness displayed on the con-
sole is useable. It is NOT USEFUL to wait for the tube to cool and repeat the exposure.
6.6.11 Note in the Job Card IST043 (IST007 data section) the thickness displayed on the console after the
exposure. It must be equal to the plexiglass thickness ±2 mm.
6.6.13 If the displayed thickness is outside the tolerance given in section 6.6.11, first check that the correct
screen pair was selected in section 6.6.9. Otherwise, the corresponding calibration procedure (con-
tact mode with grid and cassette with screen, section 6.5) was probably done incorrectly. That pro-
cedure, as well as this check procedure immediately following it, would have to be repeated (this
can be done independently of the other procedures in this Job Card).
6.7 Prepare the Senographe DMR for the procedure using magnification mode without grid.
Set up the Senographe DMR in the following configuration:
D 18 x 24 cm compression paddle installed .
D Magnification device installed (always in the bottom holes)
Note: Magnification of 1.5 if the SID = 660 and Magnification of 1.4 if the SID = 610
D Cassette holder 18 x 24 in place ( no grid ).
D SID as indicated in section 6.1.2
D Perform all the magnification calibration (with and without screen) in maximal collimation size.
At each time you move the magnification plate, the collimation is set by default in minimal size 9x9.
3–42
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
6.8 Perform the procedure using magnification mode without grid (cassette holder 18 x 24 ) and cassette
without screen.
6.8.1 Be sure to be familiar with the advance preparation and information given in section 5 before pro-
ceeding, especially the information concerning the use, if permitted, of the screen–less cassette sim-
ulation tool. Also, be sure to prepare the Senographe DMR by following sections 6.1 and 6.7 before
proceeding.
6.8.3 Select CALIB/PM_YLD/GRID/NO on the console. This sets up the calibration for the without
grid procedure.
JOB CARDS
6.8.4 Select SETUP/SCREEN/NO on the console. This sets up the calibration for the cassette without
screen procedure.
6.8.9 Select SETUP/CALIB/CALIB/THICK/2 cm on the console. This tells the software the amount
of plexiglass present.
6.8.11 Press the 2nd trigger button and hold it down. A series of X–ray exposures is taken (at least 25).
6.8.12 When the series of exposures is complete, the message ”thick calib end” appears on the console dis-
play. At this point, release the 2nd trigger button.
6.8.13 Raise the compression paddle and increase the amount of plexiglass to 40 mm. Select SETUP/
THICK/4 cm on the console.
6.8.14 Lower the compression paddle onto the plexiglass. Repeat sections 6.8.10 to 6.8.12.
6.8.15 Raise the compression paddle and increase the amount of plexiglass to 6 cm. Select SETUP/
THICK/6 cm on the console.
3–43
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
6.8.16 Lower the compression paddle onto the plexiglass. Repeat sections 6.8.10 to 6.8.12.
6.8.17 Select SETUP/CALC/VALID on the console. This executes the calculation of the 6 coefficients
(A0, A1, A2, B0, B1, and B2) for each of 5 track/filter combinations. The message PM_yld cal end
appears on the console display.
6.9 Check the magnification mode without grid (cassette holder 18 x 24 ) and cassette without screen calibra-
tion accuracy.
Note: This check must always be made immediately after performing the calibration procedure in sec-
tion 6.8.
6.9.1 Copy the without screen parameters for the screen pair being calibrated to the with screen parameter
location of an unused screen pair (generally E) by following this procedure:
JOB CARDS
Be very sure of the choice of unused screen pair indicator y (A, B, C, D or E).
CAUTION If a mistake is made, there is a risk of erasing good parameter values on
another screen pair that is in use.
– Once you are SURE that the screen pair indicator y is correct (see CAUTION above),
copy the parameters by selecting SETUP/SETUP/COPY/VALID. The message
COPY EXECUTED appears on the console display.
6.9.2 Enable visual display of parameter values when in application mode. See Display of Parameters
Following an Exposure in chapter 1, section 3–10–2
6.9.4 Set up the Senographe DMR in the configuration given in section 6.7.
3–44
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
6.9.9 Take a short exposure. It is not necessary for the exposure to be complete, so the 2nd trigger button
can be released as soon as X–rays are being produced.
Note: Even if exposure is aborted due to insufficient available mAs, the thickness displayed on the con-
JOB CARDS
sole is useable. It is NOT USEFUL to wait for the tube to cool and repeat the exposure.
6.9.10 Note in the Job Card IST043 (IST007 data section) the thickness displayed on the console after the
exposure. It must be equal to the plexiglass thickness ±2 mm.
6.9.12 If the displayed thickness is outside the tolerance given in section 6.9.10, first check that the correct
screen pair was selected in section 6.9.8. Otherwise, the corresponding calibration procedure (1.5
magnification mode without grid and cassette without screen, section 6.8) was probably done incor-
rectly. That procedure, as well as this check procedure immediately following it, would have to be
repeated (this can be done independently of the contact mode procedures in this Job Card).
6.10 Perform the procedure using magnification mode without grid (cassette holder 18 x 24 ) and cassette with
screen.
6.10.1 Be sure to be familiar with the advance preparation and information given in section 5 before pro-
ceeding. Also, be sure to prepare the Senographe DMR by following sections 6.1 and 6.7 before
proceeding.
6.10.3 Select CALIB/PM_YLD/GRID/NO on the console. This sets up the calibration for the without
grid procedure.
3–45
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
6.10.8 Select SETUP/CALIB/CALIB/THICK/2 cm on the console. This tells the software the amount
of plexiglass present.
JOB CARDS
6.10.10 Press the 2nd trigger button and hold it down. A series of X–ray exposures is taken (at least 25).
6.10.11 When the series of exposures is complete, the message ”thick calib end” appears on the console dis-
play. At this point, release the 2nd trigger button.
6.10.12 Raise the compression paddle and increase the amount of plexiglass to 4 cm. Select SETUP/
THICK/4 cm on the console.
6.10.13 Lower the compression paddle onto the plexiglass. Repeat sections 6.10.9 to 6.10.11.
6.10.14 Raise the compression paddle and increase the amount of plexiglass to 6 cm. Select SETUP/
THICK/6 cm on the console.
6.10.15 Lower the compression paddle onto the plexiglass. Repeat sections 6.10.9 to 6.10.11.
6.10.16 Select SETUP/CALC/VALID on the console. This executes the calculation of the 6 coefficients
(A0, A1, A2, B0, B1, and B2) for each of 5 track/filter combinations. The message PM_yld cal end
appears on the console display.
6.11 Check the magnification mode without grid (cassette holder 18 x 24 ) and cassette with screen calibration
accuracy.
Note: This check must always be made immediately after performing the calibration procedure in sec-
tion 6.10.
6.11.1 Enable visual display of parameter values when in application mode. See Display of Parameters
Following an Exposure in chapter 1, section 3–10–2.
3–46
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
6.11.2 If necessary, enable use of the screen pair being calibrated. (From application mode, select SETUP/
GENE/INSTAL/GENE/AOP/FILM/ALGO/FSC=x/VALID/YES on the console.)
6.11.4 Select the screen pair being calibrated and verify that its name is as entered in Job Card IST 006
SCREEN PAIR SELECTION AND CONFIGURATION
6.11.5 Set up the Senographe DMR in the configuration given in section 6.7.
JOB CARDS
6.11.9 Set up the following exposure:
Mode: AEC (1–point)
Screen pair: Select the name of the screen pair being calibrated
Focal spot: SMALL (IMPORTANT for thickness accuracy)
Track: MO
Filter: MO
HV: 30 kV
Collimation: maximal size.
6.11.10 Take a short exposure. It is not necessary for the exposure to be complete, so the 2nd trigger button
can be released as soon as X–rays are being produced.
Note: Even if exposure is aborted due to insufficient available mAs, the thickness displayed on the con-
sole is useable. It is NOT USEFUL to wait for the tube to cool and repeat the exposure.
6.11.11 Note in Job Card IST043 the thickness displayed on the console after the exposure. It must be equal
to the plexiglass thickness ±2 mm.
3–47
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
6.11.13 If the displayed thickness is outside the tolerance given in section 6.11.11, first check that the correct
screen pair was selected in section 6.11.9. – otherwise, the corresponding calibration procedure (1.5
magnification mode without grid and cassette with-screen, section 6.10) was probably done incor-
rectly. That procedure, as well as this check procedure immediately following it, will have to be
repeated (it can be done independently of the other procedures in this Job Card).
6.11.14 It is ESSENTIAL to disable the UNUSED screen pair y that was used to test the without-screen
parameters (from application mode, select SETUP/GENE/INSTAL/GENE/AOP/FILM/
ALGO/FSC=y/VALID/NO).
6.11.15 If necessary, disable use of the screen pair being calibrated (from application mode, select SETUP/
JOB CARDS
SECTION 7
CALIBRATING THE 5TH COUPLE (ONLY)
Perform this calibration only when you need to calibrate the 5th couple.
Presuming that the last free couple is E the procedure is as shown below.
7.1 Perform the no-screen/contact-mode calibration (similar to the procedure in Section 6.3)
7.2 Copy E no-screen to E with-screen and test (similar to the procedure in Section 6.4)
7.3 Repeat 7.1 and 7.2 in magnification-mode (similar to the procedure in Sections 6.8 & 6.9)
7.4 Perform the with-screen/contact-mode calibration and test (similar to the procedure in Section 6.6 & 6.5)
7.5 Repeat 7.4 in magnification-mode and test (similar to the procedure in Section 6.10 & 6.11).
Note: The above method could be applied to all couples, but is more time-consuming.
SECTION 8
COPY OF PARAMETERS FOR ANOTHER FILM/SCREEN COMBINATION
3–48
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
2. Select SETUP/FSC/FSC=x, where x is the screen pair indicator corresponding to an existing calibrated
screen pair using the same cassette as the screen pair currently being calibrated.
3. Select SETUP/SETUP/TARG/SCREEN/NO.
4. Select SETUP/FSC/FSC=y, where y is the indicator of the screen pair being calibrated.
5. Once you are SURE that the screen pair indicator y is correct, copy the without screen parameters by select-
ing SETUP/SETUP/COPY/VALID. The message COPY EXECUTED appears on the console display.
JOB CARDS
7. Select SETUP/FSC/FSC=x, where x is the screen pair indicator corresponding to an existing calibrated
screen pair using the same cassette as the screen pair currently being calibrated.
8. Select SETUP/SETUP/TARGET/SCREEN/YES.
9. Select SETUP/FSC/FSC=y, where y is the indicator of the screen pair being calibrated.
10. Once you are SURE that the screen pair indicator y is correct, copy the with screen parameters by selecting
SETUP/SETUP/COPY/VALID. The message COPY EXECUTED appears on the console display.
3–49
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
Blank page
JOB CARDS
3–50
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
SECTION 1
SUPPLIES
None
SECTION 2
TOOLS
D Ratchet wrench with 22 mm socket
D 6 mm Allen wrench
JOB CARDS
D Claw hammer or other suitable tool for removing packing nails
SECTION 3
SAFETY PRECAUTIONS
D Follow standard safety practices for handling and moving large machines.
SECTION 4
PREREQUISITES
None
3–51
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
SECTION 5
UNPACKING THE EQUIPMENT
The objective is to unpack the gantry, generator and accessories and move them to the mammography room.
5.2.3 Remove the protective screen and the other accessories from the crate and move them into the mam-
mography room.
5.3.1 Use the 6 mm Allen wrench to remove the four gantry baseplate shipping hold–down screws.
5.3.2 Thread the four special jacking bolts (provided in the accessory pouch) into their respective holes
in the gantry baseplate. Place the four flat round jack feet (provided in the accessory pouch) on the
floor of the crate with their nipples facing upwards. Continue turning the jacking bolts downwards
until their hollowed ends engage with the nipples of the jack feet.
5.3.3 Use the ratchet wrench with a 22 mm socket to progressively jack the gantry up via the four jacking
bolts.
5.3.4 When gantry baseplate is high enough to be clear of the wooden support blocks, remove these
blocks.
5.3.5 Lower the gantry progressively via the four jacking bolts until it is fully supported by the castors.
5.3.6 Leave the four special jacking bolts threaded into the gantry baseplate and keep the jack feet aside
(the jack feet will be used again when installing the gantry).
3–52
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
5.3.7 Carefully roll the gantry down the ramp and into the mammography room.
5.4.2 Carefully roll the generator down the ramp and into the mammography room.
JOB CARDS
3–53
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
Blank page
JOB CARDS
3–54
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
SECTION 1
SUPPLIES
None
SECTION 2
TOOLS
D Ratchet wrench with 22 mm socket
D 8 mm Allen wrench
JOB CARDS
D Drill
D Screwdrivers
SECTION 3
SAFETY PRECAUTIONS
D Follow standard safety practices for handling and moving large machines.
SECTION 4
PREREQUISITES
D Job Card IST 008 ”UNPACKING THE EQUIPMENT”
SECTION 5
PHYSICAL INSTALLATION
The objective is to place and anchor physically the gantry in its permanent position, and to assemble the protec-
tive screen.
Note: The generator is not permanently installed in this Job Card because it must remain mobile until
the wiring between it and the gantry is complete (Job Card IST 010 ”WIRING”). The generator
must therefore remain on its castors for the moment.
The ”CE” marking label, located on one of the large side panels of the generator, must remain vis-
ible after installation of the generator.
3–55
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
PHYSICAL INSTALLATION
Job Card IST 009 2 of 2
5.1 Install and anchor the gantry into its permanent position.
D Before you proceed, it is recommended to perform a final verification of the information given in the Pre–
Installation Manual, particularly concerning clearances around the gantry when installed in its permanent
position.
D If no Stereotix 2 is to be installed, the anchoring is not mandatory, only recommended (except in seismic
area).
5.1.1 Place the drilling template (provided in the installation kit) flat on the floor, align it with the desired
permanent position and mark the three gantry mounting hole centers on the floor.
5.1.2 Drill the three gantry mounting holes (depth and diameter depend on the type of insert to be used
– see below).
5.1.3 Place the appropriate type of insert (not provided) into each of the three mounting holes (depending
JOB CARDS
on the type and thickness of floor). The inserts should be able to receive the 10 x 80 mm screws (pro-
vided), or, for U.S. installations, the closest U.S. equivalent (not provided).
5.1.4 Wheel the gantry so that its three baseplate mounting holes are aligned with the three holes drilled
in the floor.
5.1.5 Place the four flat round jack feet on the floor with their nipples facing upwards.
5.1.6 Turn the four jacking bolts downwards until their hollowed ends engage with the nipples of the four
jack feet.
5.1.7 Use the ratchet wrench with a 22 mm socket to progressively jack the gantry up via the four jacking
bolts until the castors are clear of the floor.
5.1.9 Lower the gantry progressively via the four jacking bolts until it is fully supported by its baseplate.
Remove the four jacking bolts completely.
Note: The four jack feet remain below the gantry baseplate. If the floor is not perfectly even, level the
DMR base using the 4 screws PN 45562657 (provided in installation kit). Insert them in place
of the jacking blots to push on the jack feet.
5.1.10 To provide access to all three gantry baseplate mounting holes (one of the holes is located inside
the gantry column), remove the two column side panels (see IST 032 if necessary).
5.1.11 Bolt the gantry to the floor with the three 10 x 80 mm diameter screws (provided) or, for U.S. instal-
lations, the closest U.S. equivalent (not provided).
5.1.12 Cover all unused gantry baseplate holes with the plastic plugs provided.
3–56
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
SECTION 1
SUPPLIES
D HV grease
SECTION 2
TOOLS
D Standard set of socket wrenches and/or nut drivers
JOB CARDS
D Standard set of open end wrenches
SECTION 3
SAFETY PRECAUTIONS
None
SECTION 4
PREREQUISITES
The chronological order given in Installation Steering in chapter 1 must be followed.
3–57
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
WIRING
Job Card IST 010 2 of 14
SECTION 5
CONNECTION OF CABLES BETWEEN GANTRY, GENERATOR AND CONSOLE
The objective is to connect all cables between the generator, gantry and console, and to connect X–ray lamp
and room lamp wires to the generator, if used.
Note: The mains cable is NOT connected in this Job Card. This operation is performed in
Job Card IST 011 ”CONNECTION OF MAINS SUPPLY”.
Note: THE ROUTING PROCEDURE FOR THE CABLE BUNDLE IS VERY CRITICAL. TO AVOID
JOB CARDS
HAVING TO UN–DO YOUR WORK AND CORRECT IT, BE SURE TO FOLLOW THE
INSTRUCTIONS IN THE EXACT ORDER GIVEN. READ EACH STEP COMPLETELY
BEFORE STARTING TO PERFORM IT.
5.1.1 Be sure that the gantry is placed exactly in its permanent position.
5.1.2 Be sure that the generator is placed approximately in its permanent position (it may be necessary
to move the generator slightly in order to route and attach certain wires and cables inside the genera-
tor cabinet).
Note: The generator cabinet should still be easy to move as it should still be on its castors at this point.
Note: Be sure that the generator is placed with HV unit 701 closest to the gantry.
5.1.3 Free the cable bundle from the gantry and lay it out flat and free of kinks and twists.
3–58
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
WIRING
Job Card IST 010 3 of 14
5.1.4 Identify inner bundles 1 and 2 at the free end of the cable bundle.
– Inner bundle 1 consists of:
– HV cable
– X–ray tube anode current return wire (”BLOC +HT”)
– Emergency ”OFF” switch wires (”300 PL3 XJ 5”)
– Anode starter cable (”G300 PL2 XJ2”) with its green/yellow ground wire
– Inner bundle 2 consists of:
– Green/yellow ground wire (”45553495”)
– Gantry +27 VDC supply (”200 PL2 XJ6”)
– Gantry +27 VDC current return (”200 PL2 XJ7”)
JOB CARDS
– 4 optical fibers (”400 PL2 XJ6”, ”400 PL2 XJ7”, ”400 PL2 XJ8”, and ”400 PL2 XJ9”)
5.1.5 Feed the free end of the cable bundle UNDER the generator cabinet (see illustration 1):
7. Pass the free end of the cable bundle via the narrow side of the generator cabinet closest to HV
unit 701.
8. Carefully pull the free end of the cable bundle via one of the two wide sides of the generator cabi-
net until all slack between the gantry and generator is taken up.
Note: The choice of which wide side of the generator cabinet must be made by the service engineer who
must judge which side will facilitate cable bundle installation. Illustration 1 shows both possibili-
ties.
Note: When this step is finished, the entire cable bundle should be on the floor of the mammography
room. All excess cable bundle should be laid out flat and free of kinks and twists next to the wide
side of the generator cabinet.
3–59
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
WIRING
Job Card IST 010 4 of 14
5.1.6 Roll up excess cable bundle as shown in illustration 1, taking the following guidelines into account:
– Be sure to choose the direction of spiral rotation according to your choice of generator
side, as shown in illustration 1.
– The resulting roll must consist of only one layer (i.e. all the rolled up excess cable bundle
must be flat on the floor).
– The diameter of the roll must not exceed 35 cm.
– The orientation of inner bundles 1 and 2 must be EXACTLY as shown in illustration 1
(i.e. inner bundle 1 towards the HV unit side of the generator cabinet and inner bundle
2 towards the opposite side).
– Inner bundle 2 must pass BELOW inner bundle 1.
If any of the above guidelines is not met, re–arrange (and re–roll, if necessary), the excess cable
bundle until all the guidelines are met. THIS IS THE CRITICAL POINT WHICH DETER-
JOB CARDS
5.1.7 Slide the excess cable bundle roll under the generator so that it protrudes from the opposite side of
the cabinet by about 10 cm. Be sure that inner bundles 1 and 2 are still oriented as given in the guide-
lines in section 5.1.6 above.
5.1.8 Slide the entire excess cable bundle roll completely under the generator cabinet.
3–60
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
WIRING
Job Card IST 010 5 of 14
ILLUSTRATION 1
INITIAL PREPARATION OF GANTRY–GENERATOR CABLE BUNDLE (TOP VIEW)
GANTRY 800
GENERATOR CABINET
HV UNIT
701
JOB CARDS
INNER BUNDLE 2
INNER BUNDLE 1
35 CM MAX.
OR
INNER BUNDLE 2
INNER BUNDLE 1
35 CM MAX.
HV UNIT
701
GENERATOR CABINET
GANTRY 800
3–61
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
WIRING
Job Card IST 010 6 of 14
5.2 Route and connect the wires and cables from inner bundle 1 (and X–ray lamp and room lamp wires
if used) inside the generator cabinet (illustration 2).
Note: It is recommended to wait until all cables and wires in inner bundle 1 (and X–ray lamp and room
lamp wires if used) are routed and connected inside the generator cabinet before attaching any
wires or cables with tie–wraps.
1. Route inner bundle 1 out from beneath the generator cabinet on the HV unit side. Separate HV
cable from other cables in inner bundle 1.
3. Route the HV cable above the horizontal cross–member of the generator cabinet, then down
towards HV unit 701. See illustration 2.
JOB CARDS
5. Tighten carefully the HV connector ring. Tighten the small 1.5 mm screw located on the ring,
to insure proper grounding of the HV cable.
Note: In case of no proper grounding of HV cable, intermittent console lockup and random error mes-
sages could occur. Always check grounding at both ends of HV cable in case of problem.
5.2.2 Route and connect X–ray tube anode current return wire.
1. Separate X–ray tube anode current return wire (marked ”BLOC +HT”) from inner bundle 1.
3. Locate the threaded stud marked with a ground symbol ” ” on the top cover of HV unit 701.
4. Connect X–ray tube anode current return wire to this stud by carefully removing the retaining
nut with a nut driver or socket wrench, adding this wire to the others, and replacing the nut.
Do not tighten this nut excessively, as HV unit 701 is sealed and can only be
CAUTION serviced in the factory.
3–62
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
WIRING
Job Card IST 010 7 of 14
1. Separate emergency off switch wire (marked ”300 PL3 XJ5”) from inner bundle 1.
3. Locate connector XJ5 on generator anode security board 300 PL3 and connect this wire to it.
5.2.4 Route and connect anode starter cable and its ground wire.
1. Separate anode starter cable (marked ”G300 PL2 XJ2”) from inner bundle 1.
3. Locate connector XJ2 on generator anode starter board 300 PL2 and connect this cable to it.
4. Use a socket wrench or nut driver to remove CAREFULLY the ground screw located on the
JOB CARDS
sheet metal frame of generator power conversion assembly 300, just below generator anode
starter board 300 PL2. Add the yellow/green ground wire from the anode starter cable to the
other existing ground wires, and then replace the ground screw.
Do not tighten this screw excessively, as there is no nut on the opposite side
CAUTION of the sheet metal ; there is only a threaded hole in the sheet metal.
5.2.5 Route and connect X–ray lamp and/or room lamp wires (field supplied – if used).
2. Locate connectors XJ12 and XJ11 on generator mains distribution board 200 PL1.
3. Connect X–ray lamp wires to connector XJ12 and room lamp wires to connector XJ11.
3–63
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
WIRING
Job Card IST 010 8 of 14
5.2.6 Attach inner bundle 1, X–ray lamp and room lamp wires and cables to generator cabinet.
1. Arrange HV cable neatly as shown in illustration 2, then attach it to the lower and middle genera-
tor cabinet cross–members with two tie–wraps as shown (one tie–wrap on each cross–member).
2. Arrange the remaining cables and wires from inner bundle 1 as shown in illustration 2, then
attach them to the lower and middle generator cabinet cross–members with two tie–wraps as
shown (one tie–wrap on each cross–member).
Note: X–ray tube anode current return wire ”BLOC +HT” is attached only to the lower generator cabinet
cross member.
3. Arrange X–ray lamp and room lamp wires (if used) as shown in illustration 2, then attach them
to the lower and middle generator cabinet cross–members with two tie–wraps as shown (one
tie–wrap on each cross–member).
JOB CARDS
3–64
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
WIRING
Job Card IST 010 9 of 14
ILLUSTRATION 2
ROUTING OF CABLES WITHIN INNER BUNDLE 1 AND X–RAY AND ROOM LAMP WIRES
JOB CARDS
EMERGENCY OFF
SWITCH WIRE
”300 PL3 XJ5” ANODE STARTER
300 PL3 CABLE ”G300 PL2 XJ2”
ÈÈ
ÈÈ ÈÈ
X–RAY LAMP AND
ROOM LAMP WIRES
(FIELD SUPPLIED)
ÈÈ
ÈÈ ÈÈ HV CABLE
ÈÈ
WIRE ”BLOC +HT”
ÈÈ = TIE–WRAP
ÈÈ
ÈÈ
ÈÈ
ÈÈ
ÈÈ
ÈÈ INNER BUNDLE 1
3–65
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
WIRING
Job Card IST 010 10 of 14
5.3 Route and connect the wires and cables from inner bundle 2 and console cable inside the generator
cabinet (illustration 3).
Note: It is recommended to wait until all cables and wires in inner bundle 2 and console cable are routed
and connected inside the generator cabinet before attaching any wires or cables with tie–wraps.
1. Route inner bundle 2 out from beneath the generator cabinet on the mains input (i.e. non–HV
unit) side. Separate yellow/green generator–gantry ground wire (marked ”45553495”) from
other wires and cables in inner bundle 2.
3. Locate the grounding strip just behind generator Mains Supply Switch 200–S1.
JOB CARDS
4. Use a nut driver or socket wrench to remove one of the AVAILABLE grounding screws from
this grounding strip.
5. Attach the generator–gantry ground wire to the grounding strip with this screw, being sure to
place the lockwasher onto the screw shaft before re–installing it.
Do not tighten this screw excessively, as there is no nut on the opposite side
CAUTION of the sheet metal; there is only a threaded hole in the sheet metal.
5.3.2 Route and connect gantry +27 VDC supply and current return wires.
1. Separate gantry +27 VDC supply and current return wires (marked ”200 PL2 XJ6” and
”200 PL2 XJ7”, respectively) from other wires and cables in inner bundle 2.
3. Use a nut driver or socket wrench to remove carefully the retaining nuts from connectors XJ6
and XJ7 of generator Mains Control board 200 PL2. Add each of these two wires to its respective
connector and replace the retaining nuts, tightening them gently with the nut driver or socket
wrench.
3–66
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
WIRING
Job Card IST 010 11 of 14
Optical fiber cables and their connectors are more delicate than ordinary
CAUTION electrical cables. Handle them with care.
1. Separate the black flexible tubing containing the four optical fiber cables (marked
”400 PL2 XJ6”, ”400 PL2 XJ7”, ”400 PL2 XJ8”, and ”400 PL2 XJ9”) from other wires and
cables in inner bundle 2.
3. Locate optical fiber connectors XJ6, XJ7, XJ8 and XJ9 on generator interface board 400 PL2.
CAREFULLY connect each of these four optical fiber cables to its respective connector.
JOB CARDS
5.3.4 Route and connect console cable.
2. Locate connector XJ8 on generator command board 400 PL1 and connect this cable to it.
1. Leaving sufficient slack to avoid stress on the optical fiber connectors, attach optical fiber cable
tubing to the middle generator cabinet cross–member with a tie–wrap as shown in illustration 3.
2. Arrange optical fiber cable tubing and console cable neatly as shown in illustration 3, then
attach them to the lower generator cabinet cross–member with a tie–wrap as shown.
3. Arrange generator–gantry ground wire and gantry +27VDC supply and current return wires
neatly as shown in illustration 3, then attach them to the lower generator cabinet cross–member
with a tie–wrap as shown.
3–67
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
WIRING
Job Card IST 010 12 of 14
ILLUSTRATION 3
ROUTING OF CABLES WITHIN INNER BUNDLE 2 AND CONSOLE CABLE
400 PL3
300 PL1
JOB CARDS
400 PL2
CONSOLE CABLE
200 PL2
ÈÈ
GANTRY +27VDC
SUPPLY AND CUR-
RENT RETURN
ÈÈ TIE–WRAP MUST
WIRES ”200 PL2 XJ6” BE PLACED
AND ”200 PL2 XJ7” AROUND THE
OPTICAL FIBER
GENERATOR– CABLE COMMON
GANTRY TUBING
GROUND WIRE
”45553495” 200 PL1
ÈÈ
CONSOLE CABLE
ÈÈ = TIE–WRAP
3–68
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
WIRING
Job Card IST 010 13 of 14
2. Using an open–end wrench, turn the four jack–screws at the bottom of the generator cabinet so
that all four castors are clear of the mammography room floor and the generator cabinet is stand-
ing stable and level on the jack–screws.
JOB CARDS
3–69
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
WIRING
Job Card IST 010 14 of 14
Blank page
JOB CARDS
3–70
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
SECTION 1
SUPPLIES
None
SECTION 2
TOOLS
D a.c. voltmeter
D torque measurement wrench (optional)
JOB CARDS
SECTION 3
SAFETY PRECAUTIONS
Follow standard safety techniques and procedures for handling equipment and circuits that can come into con-
tact with mains supply voltages.
SECTION 4
PREREQUISITES
The chronological order given in Installation Steering in chapter 1 must be followed.
SECTION 5
CONNECTION OF MAINS SUPPLY
The objective is to determine the characteristics (i.e. nominal voltage, frequency and number of phases) of the
available power source, configure the Senographe DMR generator accordingly, and connect it to the power
source.
Note: Unless otherwise specified, all boards and components referred to in this Job Card are contained
in Mains Interface Assembly 200 of the Senographe DMR generator.
Note: Recommended torque to tighten all lug nuts is 4 N.m (Newton. meter)
3–71
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
Position and connect mains voltage selection wire ”SEL” to the lug on generator Mains Distribution board
200–PL1 corresponding to the nominal (NOT measured) line voltage determined in section 5.2. See illustra-
tion 1 and table 1 below.
Note: The ”0V” lug is reserved exclusively for performing the procedure given in Job Card IST 012
”LINE RESISTANCE MEASUREMENT”. The Senographe DMR cannot function normally
when the ”SEL” wire is in this position.
3–72
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
TABLE 1
POSITION OF VOLTAGE SELECTION BOARDS IN1 AND IN2 AND ”SEL” WIRE ACCORDING TO NOMINAL MAINS
VOLTAGE AND NUMBER OF PHASES
JOB CARDS
440V MONO–PHASE
use transformer reference : 45563370 and 380V monophase
p
configuration
480V MONO–PHASE
Note: 415V TRI–PHASE connection cannot be used. Use 415V MONOPHASE instead.
3–73
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
ILLUSTRATION 1
PLACEMENT OF VOLTAGE SELECTION BOARDS IN1 AND IN2 AND ”SEL” WIRE ON GENERATOR MAINS DISTRIBU-
TION BOARD 200–PL1 ACCORDING TO NOMINAL LINE VOLTAGE AND NUMBER OF PHASES
IN1 ÉÉÉÉÉ
ÉÉÉÉÉ
BT ÇÇÇÇÇ
ÇÇÇÇÇ
HT
Any Voltage
ÉÉÉÉÉ
ÉÉÉÉÉ IN2
ÇÇÇÇÇ
ÇÇÇÇÇ IN2
ÇÇÇÇ
except R1 R4
3ø 415 V
ÇÇÇÇ
ÇÇÇÇ
JOB CARDS
ÇÇÇÇ
MONO
ÇÇÇÇ
IN1 415V
380V
240V
ÇÇ 220V
208V
200V
0V
”SEL” WIRE
SEE SECTION 5.5
X1
3–74
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
1. Verify that generator Mains Supply Switch 200–S1 is open (i.e. in the ”0” position) and that
mains source is turned OFF.
2. Route mains cable into the generator cabinet, via the mains input (i.e. non–HV unit) side, as
shown in illustration 2.
3. Connect mains cable conductors (two or three phases plus ground) to terminals on generator
Mains Supply Switch 200–S1 and grounding terminal as shown in illustration 3.
Note: To connect the ground wire to the grounding terminal, insert a flat–blade screwdriver into the slot
in front of the terminal and push firmly. Twist the handle of the screwdriver while still pushing
firmly to open the terminal. Insert the ground wire into the open terminal and release the screw-
driver to close the terminal on it.
JOB CARDS
4. Arrange mains cable conductors neatly as shown in illustration 2, and attach mains cable to the
lower cross–member of the generator cabinet using a tie–wrap AROUND THE JACKETED
PART OF THE CABLE, as shown.
3–75
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
ILLUSTRATION 2
ROUTING OF MAINS CABLE
400 PL3
300 PL1
JOB CARDS
400 PL2
200 PL2
200 PL1
GENERATOR
MAINS SUPPLY
SWITCH 200–S1
ÈÈ
ÈÈ
MAINS CABLE
ÈÈ
TIE–WRAP MUST BE
PLACED AROUND
THE JACKETED PART
OF THE MAINS CABLE ÈÈ = TIE–WRAP
3–76
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
ILLUSTRATION 3
CONNECTION OF MAINS SUPPLY AND GROUND WIRES TO GENERATOR MAINS SUPPLY SWITCH 200–S1 AND CON-
ÈÈÈÈ
FIGURATION OF FUSES F1, F2 AND F3
ÈÈÈÈ
ÈÈÈÈ
ÈÈÈÈ ÉÉÉ
ÈÈÈÈ ÉÉÉ
ÈÈÈÈ ÉÉÉ
ÈÈÈÈ
ÈÈÈÈ ÉÉÉ PH3 WIRE
F1 F2 F3
ÇÇÄÄÄÄÄ
ÈÈÈÈ
ÇÇÇ ÇÇ
CON-
ÄÄÄÄÄ
ÈÈÈÈ
NECTED
ÈÈÈ
ÇÇ
FOR 3Φ
ÈÈÈÈ
JOB CARDS
ONLY
ÈÈÈ
ÈÈÈ
ÈÈÈ
ÈÈÈ
S1 MUST BE
OFF
ÈÈÈ
BEFORE
WIRES ARE
K1 ÈÈÈ S1 CON-
NECTED
REPLACE F1
WITH METAL CYL-
INSERT A FLAT–BLADE SCREWDRIVER
INDER IF (AND
FIRMLY INTO THIS SLOT AND TWIST
ONLY IF) POWER
TO OPEN GROUND WIRE TERMINAL
SOURCE CON-
SISTS OF ONE
PHASE AND NEU-
TRAL
3–77
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
Blank page
JOB CARDS
3–78
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
SECTION 1
SUPPLIES
None
SECTION 2
TOOLS
D a.c. voltmeter
JOB CARDS
SECTION 3
SAFETY PRECAUTIONS
During the procedures in this Job Card, you will be in close proximity to cir-
CAUTION cuits operating at mains voltage. Be sure to take appropriate precautions.
SECTION 4
PREREQUISITES
The chronological order given in Installation Steering in chapter 1 must be followed.
Note: Be sure that the console is connected to the Senographe DMR generator.
SECTION 5
LINE RESISTANCE MEASUREMENT
The objective is to determine the mains line resistance at the mains input to the Senographe DMR generator
cabinet. The result of this measurement may have an impact on the maximum allowable generator power.
(Maximum allowable generator power, once determined by the procedure in this Job Card, will be set later
in the sequence given in Installation Steering in chapter 1).
5.1.1 Switch off the Senographe DMR and open generator Mains Supply Switch 200–S1 (”0” position).
3–79
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
5.1.2 Disconnect mains voltage selection wire ”SEL” from its normal lug on generator Mains Distribu-
tion Board 200–PL1, and reconnect it to the ”0V” lug.
Note: The ”0V” position of the ”SEL” wire is reserved EXCLUSIVELY for this line resistance measure-
ment. The Senographe DMR cannot function normally when the ”SEL” wire is in this position.
DO NOT ATTEMPT TO TURN ON THE SENOGRAPHE DMR AT THIS POINT. FOLLOW
THE INSTRUCTIONS BELOW.
5.1.3 Connect the ac voltmeter to the input (i.e. line) end of generator mains fuses F1 and F2.
5.2.1 Turn mains source on, close generator Mains Supply Switch 200–S1 (”1” position) and note down
the open circuit line voltage ”V1” from the ac voltmeter.
JOB CARDS
Note: IN THE NEXT STEP, YOU WILL HAVE ONLY 3 SECONDS TO TAKE A READING ON
THE AC VOLTMETER. FAILING TO DO SO WILL MEAN A LOSS OF SEVERAL MIN-
UTES OF INSTALLATION TIME. READ THE NEXT STEP COMPLETELY BEFORE PER-
FORMING IT.
5.2.2 While looking carefully at the ac voltmeter reading, press the Senographe DMR ”on” switch (either
on the console or, more conveniently, on the generator Mains Control Board 200–PL2). During a
period of three seconds, a resistive load (power resistors R1 and R4 on generator Mains Distribution
Board 200–PL1) will be effectively connected across the Senographe DMR mains input, causing
a drop in the voltage read on the ac voltmeter. Read and note down this closed circuit voltage ”V2”.
Note: Immediately following this three–second period, red LED DE5 on generator Mains Control Board
200–PL2 will light, indicating that an attempt to repeat this closed circuit voltage reading is tem-
porarily impossible due to heating of the resistive load. After about 15 minutes, the LED will go
out, indicating that the resistive load has sufficiently cooled down, and the reading can be repeated
if necessary.
3–80
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
JOB CARDS
5.2.4 Once line resistance RL is determined, find the maximum allowable generator power from table 2.
Example 1: Imagine that for a 240V mono–phase installation, the line resistance is determined to
be 0.41 ohms. Looking across the ”240V MONO–PHASE” row of table 2, we find the HIGHEST
maximum allowable generator power whose maximum line resistance value exceeds 0.41 ohms.
The result is thus a maximum allowable generator power of 3.5 kW.
Example 2: Imagine that for a 380V tri–phase installation, the line resistance is determined to be
2.46 ohms. Looking at the ”380V TRI–PHASE” row of table 2, we find that installation of the Seno-
graphe DMR is IMPOSSIBLE with such a high line resistance; even at a maximum allowable gen-
erator power of 3.0 kW, the maximum line resistance permitted is 2.40 ohms. In such a case, it would
be necessary to install a new line before continuing with the installation.
Note: The maximum allowable generator power as determined here will be set later in the sequence
given in Installation Steering in chapter 1.
3–81
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
TABLE 2
MAXIMUM LINE RESISTANCE VALUES (IN OHMS) FOR A GIVEN NOMINAL MAINS VOLTAGE AND NUMBER OF PHASES
AND MAXIMUM ALLOWABLE GENERATOR POWER
5.2.6 Open generator Mains Supply Switch 200–S1 (”0” position) and turn mains source OFF.
5.2.7 Disconnect mains voltage selection wire ”SEL” from the ”0V” lug on generator Mains Distribution
Board 200–PL1 and reconnect it to its normal lug, i.e. the lug which corresponds to the NOMINAL
(NOT measured) mains voltage.
3–82
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
SECTION 1
SUPPLIES
None
SECTION 2
TOOLS
D Normal cassette and a supply of unexposed films of type used in the screen pair being calibrated
D Sensitometer and densitometer.
JOB CARDS
SECTION 3
SAFETY PRECAUTIONS
SECTION 4
PREREQUISITES
You must strictly follow the order of AEC calibration for each screen pair given in Installation Steering in
Chapter 1. In particular, the three film reciprocity law failure compensation parameters A0, A1 and A2 for the
screen pair being calibrated must be unknown. If they are known, it is useless to follow the procedures in this
Job Card (see AEC calibration sequence and “Recommendations for AEC/AOP calibration procedure”
given in Installation Steering of Chapter 1).
Note: In the case of error messages during this calibration, see IST 041 for the explanation.
3–83
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
SECTION 5
PROCEDURE.
To avoid excessive X–Ray doses on the photo–cell tube, place an empty cas-
CAUTION sette on the top of the image receptor.
5.3.1 Perform the calibration procedure on the LF/RH focal size/track combination, then on the others
in any order (SF/RH, LF/MO, SF/MO).
5.3.2 Starting from application mode, change the position of the installation menu enable switch (switch
B1 on generator CPU board 400–PL3).
Note: This must only be done if the Senographe DMR has been switched off since the last access to SET-
UP/GENE/INSTAL or SETUP/ARM/VALID/INSTAL
3–84
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
Note: /focal size/track/ must be LF/RH for the first combination performed.
To avoid excessive X-ray doses on the photo-cell tube, place an empty cassette
CAUTION on the top of the image receptor.
5.3.4 Press the 2nd trigger button and hold it down. The messages EXPOSURE NO 1, EXPOSURE NO
2, EXPOSURE NO 3 and CALIBRATION END appear sequentially on the console display. The
software automatically performs the three exposures and the calculations.
Note: Calibration time varies according to tube temperature. The message TOO WARM appears on the
console display when tube temperature is excessive for taking the next exposure. It is not neces-
sary to hold the 2nd trigger button down while waiting for the tube to cool. However, do not exit
the calibration menu while waiting. As soon as the TOO WARM message disappears, continue
the calibration by pressing the 2nd trigger button and holding it down.
JOB CARDS
Note: In case of arcing, the software re–attempts the same exposure automatically.
Note: The calibration procedure includes an exposure at 40 kV. If for some reason this exposure is
impossible to perform (e.g. excessive arcing), tube warm–up may be necessary (see Job Card IST
020 AUTOMATIC X–RAY TUBE WARM–UP).
5.3.5 Repeat sections 5.3.3 to 5.3.4 for each of the remaining focal size/track combinations.
Note: This must be done only if the Senographe DMR has been switched off since the last access to SET-
UP/GENE/INSTAL or SETUP/ARM/VALID/INSTAL
2. Determine the film reciprocity law failure compensation parameters A0, A1 and A2 for the screen pair
being calibrated. These parameters are determined jointly by the screen pair and the characteristics of the
Senographe DMR product. This determination is done in section 5.7.
3–85
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
Note: This calibration does not depend on the screen pair being calibrated, and thus needs to be per-
formed only once for the first screen pair to be calibrated. For subsequent screen pairs, skip this
section and go directly to section 5.6.
Note: This calibration does not depend on any particular configuration of the Senographe DMR, so it
can be performed without any particular attention paid to presence or absence of accessories such
as compression paddle, cassette, grid, etc.
JOB CARDS
5.6.1 Select SETUP/SETUP/PARAM on the console. Verify the initial film reciprocity law failure com-
pensation parameters A0, A1 and A2 by rotating the kV dial through the three values. The initial
parameter values are given below:
A0init = +1.008E+0 (1.008)
A1init = +1.244E–1 (0.1244)
A2init = +2.277E–2 (0.02277)
If any of the parameter values are different from the initial values given above, correct them
manually by pressing the CHANGE key, then enter the correct value using the NEXT and
VALID keys and the kV dial (see Visual Display of Parameter Values and Modification of a
Parameter Value or Entry of a Measurement in Chapter 1). Once the value is entered and vali-
dated, press the SETUP key once to continue checking the remaining parameter values.
Note: With experience installing several Senographe DMRs, you may find sets of initial parameter val-
ues for certain screen pair types (i.e. high or low speed films, certain screen types) that result in
fewer films being used to complete this calibration.
5.6.2 Ask the doctor for the desired reference optical density (Target net density); generally
OD_R(Optical Density–Reference) is between 1.0 and 1.6.
Starting from application mode, select SETUP/GENE/INSTAL/GENE/AOP/ALGO/
FSC=x/CALIB/LNRT/REFEN/OD_ref/CHANGE on the console, where x is the screen pair
indicator (A,B,C,D,E) corresponding to the screen pair being calibrated (see Job Card IST 006).
Enter the desired reference optical density by rotating the kV dial and using the NEXT and
VALID keys (see Visual Display of Parameter Values and Modification of a Parameter Value or
Entry of a Measurement in Chapter 1).
3–86
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
5.6.3 Perform a sensitometry test on the fresh film (unexposed film) being used with the screen pair being
calibrated.
5.6.4 Select SETUP/SETUP/SCALE on the console. Rotate the kV dial until the value OD_3 is dis-
played on the console screen. Select CHANGE on the console.
Enter the optical density from scale number 3 of the sensitometry test by using the NEXT and
VALID keys and rotating the kV dial.
5.6.5 Press the SETUP key on the console. Rotate the kV dial until the value OD_5 is displayed on the
console screen. Select CHANGE on the console.
Enter the optical density from scale number 5 of the sensitometry test by using the NEXT and
VALID keys and rotating the kV dial.
JOB CARDS
5.6.6 Repeat section 5.6.5 for OD_7, OD_9, OD_11, OD_13 and OD_15 from the sensitometry test.
Note: If the message “Scale error” is displayed, the entered values of optical density did not increase as
expected; repeat the test and check that the correct values have been entered.
5.6.8 Install the cassette loaded with fresh film corresponding to the screen pair being calibrated.
5.6.9 IF this is the FIRST exposure, select SETUP/SETUP/CALIB/CALIB on the console. For SUB-
SEQUENT exposures (if any) select only SETUP/CALIB on the console. Press the 2nd trigger
button and hold it down. A single X-ray exposure is taken. Release the 2nd trigger button.
Note: By default, the first exposure is taken at +7.000E+1 (70) mAs. For certain slower speed films, this
calibration can be done using fewer films if the initial mAs value is increased. To do this, prior
to pressing the 2nd trigger button, select SETUP/SETUP/MAS/CHANGE on the console and
enter the desired initial mAs value using the NEXT and VALID keys and rotating the kV dial. Then
select SETUP/SETUP/CALIB/CALIB on the console and proceed with the exposure.
Note: The grid is not activated during this calibration mode exposure, so it is normal to see grid lines
on the developed film.
3–87
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
5.6.11 Select SETUP/OD_mea on the console. Use the NEXT and VALID keys and the kV dial to enter
the measured optical density value (see Visual Display of Parameter Values and Modification of a
Parameter Value or Entry of a Measurement in Chapter 1.
5.6.12 Select SETUP/CALC/VALID on the console to execute the calculation. At this point, either the
result is satisfactory, and the software displays CALIB END on the console display or, if another
exposure and re-calculation are necessary, the software displays ANOTHER FILM.
In the latter case, re-load the cassette with fresh film and repeat sections 5.6.8 to 5.6.12 without
manually changing the mAs value – the software will have calculated a new mAs value for this
exposure. Normally, no more than 2 or 3 films need to be used before CALIB END appears on the
console display to indicate a satisfactory result.
Note: It is useful to note the updated mAs value determined by the software before taking any subsequent
exposures. This can give you an idea of the initial mAs values needed for certain film types to
reduce the number of films needed during future Senographe DMR installations (see Note in sec-
JOB CARDS
tion 5.6.9).
Note: There is a delay of about 2 seconds between the moment when the LNRT key is pressed and the
appearance of the succeeding sub-menu titles. This is due to a calculation that the software per-
forms to determine, among other things, the plexiglass thickness necessary for the calibration.
5.7.3 Place the amount of plexiglass indicated on the console display on the bucky and install the cassette
loaded with fresh film corresponding to the screen pair being calibrated.
5.7.4 Press the CALIB button, then press the 2nd trigger button and hold it down. A single X-ray exposure
is taken. Release the 2nd trigger button.
Note: The exposure time can be very short (20 ms) or very long (9 s), so be sure to hold the 2nd trigger
button down constantly until the exposure is complete.
3–88
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
Note: The grid is not activated during this calibration mode exposure, so it is normal to see grid lines
on the developed film.
5.7.6 Select SETUP/OD_mea on the console. Use the NEXT and VALID keys and the kV dial to enter
the measured optical density value (see Visual Display of Parameter Values and Modification of a
Parameter Value or Entry of a Measurement in Chapter 1.
5.7.7 Select SETUP/CALC/VALID on the console to execute the calculation. At this point, either the
result is satisfactory, and the software displays CALIB END on the console display, or, if another
exposure and re-calculation is necessary, the software displays ANOTHER FILM.
In the latter case, press the SETUP key once and repeat sections 5.7.3 to 5.7.7. As many as 15 films
may have to be used before CALIB END appears on the console display following the calculation,
indicating a satisfactory result.
JOB CARDS
Note: Be sure to update the amount of plexiglass on the cassette holder (if necessary) according to the
indication on the console display.
5.7.8 Select SETUP/SETUP/PARAM on the console and note down the three film reciprocity law fail-
ure compensation parameters A0, A1 and A2 by rotating the kV dial.
Note: Keep these three parameter values with the table in Installation Steering in Chapter 1, once you
have verified them in section 5.9 below. They can be used in the future to avoid repeating this Job
Card during installation of other Senographe DMRs with the same screen pair. The Installation
Steering section explains how to perform the calibration when these parameters are known.
5.8.1 Ask the doctor for the desired manual density correction step size.
For example, if the step size is chosen to be +1.000E–1 (0.1), then, assuming that the reference opti-
cal density of 1.0, the available range of corrected optical density is from 0.5 to 1.5 in steps of 0.1
in both AEC and AOP application modes.
3–89
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
5.9.1 If necessary, enable use of the screen pair being calibrated by selecting SETUP/SETUP
/SETUP/SETUP/SETUP/VALID/YES on the console.
5.9.5 Take three exposures, using the screen pair being calibrated, with the following kV and plexiglass
thicknesses, measuring and noting down the optical density of the film, the mAs and the time ob-
tained in Job Card IST043, (IST013 data section) for each one:
– 28 kV with 40 mm of plexiglass
– 32 kV with 15 mm of plexiglass
– 28 kV with 55 mm of plexiglass
Note: Since these exposures are taken in application mode, the grid is activated during the exposures and
thus no grid lines should be visible on the developed films.
The optical density must not vary by more than 0.2 between any two of the three films. If it does,
the entire calibration must be repeated.
3–90
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
Once the variation in optical density between the three films is satisfactory, check that the optical
density of the 28kV/40 mm film is inside the tolerance of the reference optical density ±0.1 (the ref-
erence optical density was chosen in section 5.6.2).
If the optical density is outside this tolerance, the reference energy must be calibrated as indicated
in Job Card IST 014. To do this, perform the following sequence:
1. Perform the section entitled Perform calibration of reference energy in Job Card IST 014.
Note: The reference optical density must be re–entered following the menu sequence given in Job Card
IST 014.
Note: For the initial calibration attempt, do not use the default mAs value of +7.000E+1 (70). Instead,
use the mAs value obtained above from the 28 kV/40 mm check exposure.
2. Perform the section entitled Check results in Job Card IST 014.
JOB CARDS
5.9.6 If necessary, disable use of the screen pair being calibrated (from application mode, select SETUP/
GENE/INSTAL/GENE/AOP/ALGO/FSC=x/VALID/NO on the console – but see CAUTION
in section 3).
3–91
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
Blank page
JOB CARDS
3–92
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
SECTION 1
SUPPLIES
D Normal cassette and a supply of unexposed films of type used in the screen pair being calibrated
SECTION 2
TOOLS
D Densitometer
JOB CARDS
D Sensitometer
D plexiglasss in thickness increments of 5 mm (minimum dimensions 200 x 200 mm to ensure complete
covering of the photo cell)
SECTION 3
SAFETY PRECAUTIONS
Note: In the event of error messages during this calibration, see IST 041 for an explanation.
SECTION 4
PREREQUISITES
The chronological order of AEC calibration given in Installation Steering in chapter 1 must be strictly followed
for each screen pair.
This Job Card is used ONLY when the three reciprocity law failure
CAUTION compensation parameters given (A0, A1 and A2) have been entered manu-
ally as explained in Installation Steering in Chapter 1.
3–93
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
SECTION 5
PREPARATION
D To avoid confusing results (for you and the software), take care to avoid putting the single-sided film
upside-down in the cassette.
D Whenever the optical density of a film is measured, the measurement is always relative to the film
base fog level (net density). Thus, you must null out the densitometer on the film base fog,(unex-
posed area of the film); then measure the net density in the middle of the film.
To avoid excessive X–Ray doses on the photo–cell tube, place an empty cas-
CAUTION sette on the top of the image receptor.
D Starting from application mode, change the position of the installation menu enable switch B1 on genera-
tor CPU board 400–PL3.
Note: This is necessary only when the Senographe DMR has been switched off since the last access to
SETUP/GENE/INSTAL or SETUP/ARM/VALID/INSTAL
3–94
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
5.3.1 Ask the doctor for the desired reference optical density OD_R (Optical Density–Reference, the
Target Net Density). This is generally between 1.0 and 1.6.
5.3.3 Enter the desired reference optical density by using the NEXT and VALID keys and rotating the
kV dial, see ”visual display of parameters values” and ”modification of a parameter value or
entry of a measurement” in chapter 1.
JOB CARDS
5.3.4 Perform a sensitometry test on a fresh film (unexposed film) being used with the screen pair being
calibrated.
5.3.6 Rotate the kV dial until the value OD_3 is displayed on the console screen.
5.3.8 Enter the optical density from scale number 3 of the sensitometry test by using the NEXT and
VALID keys and rotating the kV dial.
5.3.10 Rotate the kV dial until the value OD_5 is displayed on the console screen.
5.3.12 Enter the optical density from scale number 5 of the sensitometry test by using the NEXT and
VALID keys and rotating the kV dial.
5.3.13 Repeat section 5.3.9 through 5.3.12 for OD_7, OD_9, OD_11, OD_13 and OD_15 from the sensi-
tometry test.
5.3.14 Select SETUP/SETUP/SETUP/CONFIG/NORMAL on the console (this tells the software that
the calibration is being done for the normal, and not Stereotix 2, mode).
3–95
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
5.3.16 Install the cassette loaded with fresh film corresponding to the screen pair being calibrated.
5.3.17 For an initial calibration, use the default mAs value of +7.000E+1 (70) by selecting SETUP/
CALIB/CALIB on the console. For any subsequent re-calibration, follow the instructions given
for entering a mAs value other than the default value (see IST013 Section 5.6.9).
5.3.18 Press the 2nd trigger button and hold it down. A single X–ray exposure is taken. Release the 2nd
trigger button.
5.3.19 Develop the film and measure its optical density (OD) (See section 5.1).
The measured OD should be greater than or equal to (OD_R – 0.2) but lower than (OD_R + 0.8),
where OD_R is the reference optical density chosen in section 5.3.1.
For example: If OD_R = 1.1 then the optical density must be between 0.9 and 1.9
Note: The grid is not activated during this calibration mode exposure, so it is normal to see grid lines
JOB CARDS
5.3.20 If the measured value is not in the required range, first change the mAs value and repeat the OD
measurement.
If the film is too dark (measured OD greater than OD_R + 0.8), the mAs must be reduced; if the
film is too light (measured OD less than OD_R – 0.2), the mAs must be increased.
Calculate the required change in mAs value as 8% for each 0.1 of OD variation required.
Note: The value of 8% is based on a film gamma of 3.
For example: with an OD_R of 1.1 and 70 mAs the OD = 1.6, which you want to reduce by 0.5;
then 8% 5 = 40%, and 40% of 70 mAs = 28 mAs.
N to decrease the OD by 0.5, you should enter a mAs value of (70 – 28) = 42 mAs.
Repeat the OD measurement with the calculated new mAs value, as follows:
2. Enter the new mAs value, using the NEXT and VALID keys and the kV dial
Note: Be careful not to press the SETUP key more than once after entering the mAs value – a second
press will cause the value to revert to the default of +7.000E+1 (70).
5.3.21 When the measured OD value is in the correct range, select SETUP/OD_MS; use the NEXT and
VALID keys and the kV dial to enter the measured value into the DMR.
5.3.22 Select SETUP/CALC/VALID on the console to execute the calculation; Calcul Done appears
on the console display.
3–96
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
5.4.1 Ask the doctor the desired manual density correction step size.
For example, if the step size is chosen to be +1.000E–1 (0.1), then assuming that the reference opti-
JOB CARDS
cal density is 1.0 the available range of corrected optical density is from 0.5 to 1.5 in steps of 0.1
in AEC and AOP Application Modes.
5.5.1 If necessary, enable use of the screen pair being calibrated by selecting
SETUP/SETUP/SETUP/SETUP/SETUP/VALID/YES on the console.
5.5.4 Keep the Senographe DMR in the same configuration as given in section 5.3.15.
5.5.5 Take an exposure, using the screen pair being calibrated, note down the resulting mAs value and
measure the optical density of the film.
Note: Since this exposure is taken in Application Mode, the grid is activated during the exposure and
thus no grid lines should be visible on the developed film.
3–97
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
5.5.6 The optical density of the film must not be outside the tolerance of the reference optical density ±0.1
(the reference optical density was chosen in section 5.3.1). If the optical density is within this toler-
ance, skip directly to section 5.5.8 below.
Note: The reference optical density chosen and entered in section 5.3.1 can no longer be displayed at
this point; it was erased and replaced by the default value when the DMR was returned to Applica-
tion Mode for these checks.
5.5.7 If the optical density of the film is outside this tolerance, repeat the calibration procedure, but use
the mAs value obtained from the check exposure taken in section 5.5.5. To do this, perform the fol-
lowing sequence:
1. Select SETUP/GENE/INSTAL/GENE/AOP/ALGO/FSC=x/CALIB/Energ./PARAM/
OD_R/CHANGE on the console. Re-enter the desired reference optical density by using
the NEXT and VALID keys and rotating the kV dial.
JOB CARDS
Note: If the measured optical density from the check exposure is greater (less) than the desired reference
density, it is preferable to use a mAs value slightly lower (higher) than the value obtained from
the check exposure.
Note: Be careful not to press the SETUP key more than one time after entering the mAs value. Doing
so would cause this value to revert to the default mAs value.
3. Repeat Sections 5.3.15 to 5.3.23, EXCEPT THAT in Section 5.3.17, IT WILL NOW ONLY BE
NECESSARY TO SELECT SETUP/CALIB AND NOT SETUP/CALIB/CALIB ON THE
CONSOLE.
Note: The greater the difference between the optical density measured during the CALIBRATION and
the desired reference density, the greater the final error in optical density when in automatic expo-
sure Application Mode.
5.5.8 If you have entered the parameters A0, A1, and A2 manually, make the following check (otherwise
go to Section 5.5.9).
Take two exposures using the screen pair being calibrated with the following kV and plexiglass
thickness, measuring and noting the optical density of the film, the mAs and the time obtained in
Job Card IST043 (IST014 data section) for each one:
– 32 kV with 1.5 cm of pexiglass.
– 28 kV with 5.5 cm of plexiglass.
3–98
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
Check the optical density value obtained previously with 4 cm, with 1.5 cm, and with 5.5 cm. It must
not vary by more than 0.2 between any two of the three films. If it does:
1. Check the reciprocity low failure compensation parameters A0, A1, and A2.
– If the parameters are not correct, re–enter them manually as explained in Chapter 1 of
the Installation Steering Guide, and perform this Job Card, IST 014, from the beginning.
– If the parameters are correct, check Job Card IST 007.
If Job Card IST 007 does not check correct, perform the entire Job Card IST 007 before
performing Job Card IST 014 again.
2. If all the previous checks are correct, check the automatic film processor by making several sen-
sitometry tests and comparing the results (see Job Card IST 034, Section 5.2).
3. If all the previous checks are correct, perform Job Card IST 013.
JOB CARDS
5.5.9 If necessary, disable use of the screen pair being calibrated (from Application Mode, select SETUP/
GENE/INSTAL/GENE/AOP/ALGO/FSC=x/VALID/NO on the console) (see CAUTION in
Section 3).
3–99
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
Blank page
JOB CARDS
3–100
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
SECTION 1
GENERAL INFORMATION:
D This Job Card is applicable for senographe DMR V1 and DMR V2 with Carbon cassette holder
18x24 CDRH with top (45561633).
D This Job Card is NOT applicable for senographe DMR V2 with the new Plastic cassette holder
18x24 CDRH (2147919) without top.
SECTION 2
JOB CARDS
SUPPLIES
D Normal cassette loaded with one unexposed film of type used in the screen pair being calibrated
SECTION 3
TOOLS
D Plexiglass in thickness increments of 0.5 cm (minimum plexiglass dimensions 20 x 20 cm to ensure com-
plete coverage of the photo cell)
SECTION 4
SPECIAL SAFETY PRECAUTIONS
SECTION 5
PREREQUISITES
You must follow the chronological order of AEC calibration given in Installation Steering in Chapter 1.
3–101
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
SECTION 6
PREPARATION
Starting from application mode, change the position of the installation menu enable switch (switch B1 on gen-
erator CPU board 400–PL3, see Accessing the Generator or Gantry Installation Mode from the Console in
Chapter 1).
Note: This must only be done if the Senographe DMR has been switched off since the last access to SET-
UP/GENE/INSTAL or SETUP/ARM/VALID/INSTAL
The objective is to calculate two co-efficients that correct the photocell gain when the Senographe DMR is
being used in magnification mode.
3–102
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
6.2.6 Press the 2nd trigger button and hold it down. One or more exposures are taken. Continue holding
the 2nd trigger button down until CALIBRATION END appears on the console display, then release
the 2nd trigger button.
JOB CARDS
Do NOT attempt to execute the calculation here. The calculation will be per-
CAUTION formed in section 6.2.25.
6.2.14 Select SETUP/SETUP/CONFIG/NO on the console (this tells the software that the calibration in
1.5 magnification mode is being done).
6.2.18 Press the 2nd trigger button and hold it down. One or more exposures are taken. Continue holding
the 2nd trigger button down until CALIBRATION END appears on the console display, then release
the 2nd trigger button.
3–103
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
6.2.25 Select SETUP/SETUP/CALC/VALID on the console to execute the calculation. CORR CAL
END appears on the console display.
6.3.1 If necessary, enable use of the screen pair being calibrated by selecting SETUP/SETUP/SETUP/
SETUP/VALID/YES on the console.
6.3.2 Enable visual display of parameter values when in application mode. See ”Display of parameters
following an exposure” in chapter 1, section 3–10–2.
JOB CARDS
6.3.8 Take an exposure and note down the estimated thickness from the console display.
3–104
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
6.3.10 Take an exposure and note down the estimated thickness from the console display.
6.3.12 Keep the exposure configuration given in section 6.3.5. Leave the cassette loaded with undeveloped
film installed.
6.3.14 Take an exposure and note down the estimated thickness from the console display.
6.3.16 Take an exposure and note down the estimated thickness from the console display.
6.3.17 Compare the estimated thicknesses for 20 mm of plexiglass in magnification mode and in contact
JOB CARDS
mode. The difference must be less than 1.5 mm.
6.3.18 Compare the estimated thicknesses for 40 mm of plexiglass in magnification mode and in contact
mode. The difference must be less than 1.5 mm.
6.3.19 Normally, at this point, the calibration of the screen pair is finished and the Senographe DMR can
be left to the user with the screen pair enabled. If for some reason the calibration of the screen pair
is not finished, and if the machine must be left unattended at this point, (see CAUTION in section
4) disable use of the screen pair being calibrated. (From application mode, select SETUP/GENE/
INSTAL/GENE/AOP/ALGO/FSC=x/VALID/NO on the console.)
6.3.20 If you are handing the Senographe DMR back to the user at this point, switch the Senographe DMR
off, then on again.
3–105
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
Blank page
JOB CARDS
3–106
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
SECTION 1
SUPPLIES
None
SECTION 2
TOOLS
D Milliammeter
JOB CARDS
SECTION 3
SPECIAL SAFETY PRECAUTIONS
None
SECTION 4
PREREQUISITES
None
3–107
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
SECTION 5
Calibration of XRT mA measurement
The objective is to determine the scale factor F/mA between the X–ray tube mA measurement input to the soft-
ware and the real X–ray tube mA value.
The default parameter value is F/mA = +1.000E+3 (1000) Hz/mA
5.1 Preparation
5.1.2 Remove strap between terminal lugs MES_M_NEG and MES_M_POS of generator kV/mA board
701–PL1.
5.1.3 Connect milliammeter between test points MES_M_NEG and MES_M_POS of generator kV/mA
board 701–PL1. Set the milliammeter to its 100 mA DC scale.
5.1.5 Change the position of the installation menu enable switch (switch B1 on generator CPU board
400–PL3, see Accessing the Generator or Gantry Installation Mode from the Console in Chapter 1).
5.2.2 Select SETUP/I_MEAS on the console. Enter the value of current (in milliamperes) read from the
milliammeter by using the NEXT and VALID keys and rotating the kV dial. See Visual Display of
Parameter Values and Modification of a Parameter Value or Entry of a Measurement in chapter 1.
Note: Only values between +4.000E+1 (40) and +6.000E+1 (60) mA are accepted by the calibration
software. An error message appears on the console display if a value outside of this range is
entered.
5.2.3 Select SETUP/CALCUL/VALID on the console to execute the calculation. If the calculated F/mA
value is between 850 and 1150 mA calib END appears on the console display – if it is not, then
an error message appears.
3–108
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
5.3.1 Select SETUP/CALIB on the console. Press pushbutton SW2 on generator kV/mA board 701–PL1
and HOLD IT DOWN. Read and note down the value measured on the milliammeter. While still
holding SW2 down, select MEASUR on the console. Verify that the measurement on the console
display now corresponds to the value read on the milliammeter ±400 µA ± milliammeter error.
Release pushbutton SW2. DO NOT ENTER THIS MEASURED VALUE.
5.3.2 Change the milliammeter scale to 200 mA DC. Press pushbutton SW1 on generator kV/mA board
701–PL1 and HOLD IT DOWN. Read and note down the value measured on the milliammeter.
(The current should be approximately 120 mA.) While still holding SW1 down, select MEASUR
on the console. Verify that the difference between the measurement on the console display and the
value read on the milliammeter is less than 2.4 mA. Release pushbutton SW1. DO NOT ENTER
THIS MEASURED VALUE.
JOB CARDS
5.3.3 Select SETUP/SETUP/PARAM on the console and note down the displayed value of F/mA.
5.3.6 Disconnect the milliammeter and replace the strap between terminal lugs MES_M_NEG and
MES_M_POS of generator kV/mA board 701–PL1.
3–109
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
Blank page
JOB CARDS
3–110
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
SECTION 1
SUPPLIES
None
SECTION 2
TOOLS
D Non invasive: RMI 232 kVp–meter, calibrated with known calibration factors.
D or Invasive: Machlett Dynalyser III divider and Fluke multimeter.
JOB CARDS
Note: MQSA (US) regulation requires kVp accuracy better than 5%. It means that the tool used for the
following measurements should have an accuracy much better than 5%, and the kVp limit values
displayed by the tool should depend from its own accuracy. The table 1 is given below to detail
these limit values vs. tool accuracy.
Note: RMI 232 must be used with its calibration factors. If not its accuracy (when verified only) is not
sufficient (from 6.0% at 25 kVp to 4.5% at 40 kVp).
SECTION 3
SPECIAL SAFETY PRECAUTIONS
SECTION 4
PREREQUISITES
Note: The Senographe DMR is calibrated at factory in accordance with MQSA requirements (5%).
Therefore, this Job Card should not be performed at the first installation of the system.
The following procedures must have been performed in the following order:
D Job Card IST 001
D Job Card IST 002
D Job Card IST 016
3–111
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
SECTION 5
CALIBRATION OF KV SCALE FACTOR
The objective is to determine the scale factor G and offset VOF between the X–ray tube kV command from
the software and the real kV value obtained.
DEFAULT PARAMETER VALUES ARE G = +6.554E+1 (65.54) and VOF = +0.000E+0 (0.0)
Since MQSA requirement is 5% accuracy, the operator should take into account the accuracy of the measure-
ment method himself to define the read values not to exceed. See Table 1 which gives the limit values for each
used tool.
JOB CARDS
– Place kV peak meter on the magnification device or cassette holder. Turn it on and set it to the
MO/MO position. Set the waveform selector to CP (constant potential).
– Use the Senographe DMR light centering device to center the kV peak meter target with the X–ray
zone.
– Starting from application mode, change the position of the installation menu enable switch
(switch B1 on generator CPU board 400–PL3).
Note: This must only be done if the Senographe DMR has been switched off since the last access to SET-
UP/GENE/INSTAL or SETUP/ARM/VALID/INSTAL
3–112
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
– For each displayed value on the multimeter, multiply it by 10 to get the kVp value (the divider
ratio is 10000 and the high voltage is expressed in kilo–volts).
Note: By default, the kV command value sent by the software for this 1st point is 25 kV.
5.2.2 Press the 2nd trigger button and hold it down. A single 1–second exposure is taken. ”CALIBRA-
TION END” appears on the console display. Release the 2nd trigger button.
. If using a RMI 232, read the displayed value, correct it by the correction factor given in the
calibration factor table and note down this result. The calibration factor table gives values by 1 kV
steps; for intermediate values, use a linear interpolation.
. If using the Divider and Multimeter, read the displayed value, multiply it by 10 and note down
JOB CARDS
this result.
Note: If for some reason, this exposure has to be repeated select SETUP/CALIB on the console before
pressing the 2nd trigger button again
5.2.3 Select SETUP/kV_M on the console. Enter the kV value read from the kV peak meter by using
the NEXT and VALID keys and rotating the kV dial. See Visual Display of Parameter Values” and
Modification of a Parameter Value or Entry of a Measurement in Chapter 1.
Note: By default, the kV command value sent by the software for this 2nd point is 40 kV. Due to DMR
software, the Mo filter remains selected.
5.2.5 Press the 2nd trigger button and hold it down. A single 1–second exposure is taken. ”CALIBRA-
TION END” appears on the console display. Release the 2nd trigger button.
. If using a RMI 232, read the displayed value, correct it by the correction factor given in the
calibration factor table and note down this result. The calibration factor table gives values by 1 kV
steps; for intermediate values, use a linear interpolation.
. If using the Divider and Multimeter, read the displayed value, multiply it by 10 and note down
this result.
Note: If for some reason, this exposure has to be repeated select SETUP/CALIB on the console before
pressing the 2nd trigger button again.
3–113
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
5.2.6 Select SETUP/kV_M on the console. Enter the kV value read from the kV peak meter by using
the NEXT and VALID keys and rotating the kV dial. See ”Visual Display of Parameter Values” and
”Modification of a Parameter Value or Entry of a Measurement” in chapter 1.
5.2.7 Select SETUP/SETUP/calcul/VALID on the console to execute the calculation. Normally, ”KV
CALIB END” appears on the console display. If ”OUT OF RANGE” appears on the console dis-
play, this indicates either faulty kV measurement(s) in the steps above, or a hardware failure.
5.3.1 Leave the Senographe DMRsenographe in the same conditions as described in section 5.1 (with the
kVp meter or divider–multimeter in place), and select SETUP/1st pt/CALIB on the console.
5.3.2 Press the 2nd trigger button and hold it down. A single 1-second exposure is taken. ”CALIBRA-
TION END” appears on the console display. Release the 2nd trigger button.
JOB CARDS
. If using a RMI 232, read the displayed value, correct it by the correction factor given in the
calibration factor table and note down this result. The calibration factor table gives values by 1 kV
steps; for intermediate values, use a linear interpolation.
. If using the Divider and Multimeter, read the displayed value, multiply it by 10 and note down
this result.
5.3.3 The kV value read on the kV peak meter should be 25 kV ± 0.3 kV.
Note: The generator kVp accuracy is ± 1% (one per cent) when measured just after a calibration.
5.3.5 Press the 2nd trigger button and hold it down. A single 1–second exposure is taken. ”CALIBRA-
TION END” appears on the console display. Release the 2nd trigger button.
. If using a RMI 232, read the displayed value, correct it by the correction factor given in the
calibration factor table and note down this result. The calibration factor table gives values by 1 kV
steps; for intermediate values, use a linear interpolation.
. If using the Divider and Multimeter, read the displayed value, multiply it by 10 and note down
this result.
3–114
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
5.3.6 The measured value should be within limits which depend on the measurement tool, as given in
Table 1 below:
Measurement tool 25 kV 40 kV
Min. Limit Max. Limit Min. Limit Max. Limit
RMI 232 24.12 25.86 38.58 41.37
Dynalyser III + Fluke 87 24.02 25.96 38.43 41.54
Table 1
Note: The generator kVp accuracy is ± 1% (two per cent) when measured just after a calibration.
Note: If the need should arise to check this point in application mode, it is not possible to do so because
exposures at 40 kV are prohibited with the Mo filter in application mode.
JOB CARDS
5.3.7 Select SETUP/SETUP/SETUP/PARAM on the console. Note down the displayed values of G and
VOF by rotating the kV dial.
5.3.8 Job Card IST 003 ”CALIBRATION OF X–RAY TUBE HEATER CURRENT” must be done
immediately following this Job Card, but if you are turning the machine back over to the user at this
point, switch off, then on again.
5.3.9 Perform calibration check of IST007 (section 6.6 only). If thickness accuracy is outside the toler-
ance, IST007 must be performed completely for all already adjusted film/screen couples.
3–115
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
Blank page
JOB CARDS
3–116
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
SECTION 1
SUPPLIES
D The cassette that will be used by the doctor for Stereotix 2 mode
Note: Presence or absence of film in the cassette is of no importance in this Job Card.
SECTION 2
TOOLS
JOB CARDS
D Plexiglass in thickness increments of 5 mm (minimum plexiglass dimensions 200 x 200 mm to ensure
complete covering of the photo cell)
SECTION 3
SAFETY PRECAUTIONS
3–117
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
SECTION 4
PREREQUISITES
The Senographe DMR being installed must be equipped with the Stereotix 2 option that has been set up and
calibrated according to the Stereotix 2 Service Manual.
Also, Job Card IST 004 must have been performed before this Job Card.
Note: In case of error messages during this calibration, see IST 041 for explanations.
SECTION 5
CALIBRATION OF PHOTO CELL IN STEREOTIX 2 POSITIONS
The objective is to calibrate the photocell in its Stereotix 2 positions. Because of the different angles between
the tube housing and the photocell when in Stereotix 2 mode, the photocell does not receive the same dose of
radiation as it does when in non–Stereotix 2 mode.
JOB CARDS
Note: Since this calibration does not depend on a certain screen pair, it is not necessary to perform it each
time a new screen pair is introduced.
This Job Card consists of four interdependent calibration procedures which are performed in the following
order:
D Normal, which is done without the Stereotix 2 installed as a reference calibration for the following three
calibrations.
D Stereotix Left, which is done with the Stereotix 2 installed and the tube–carrying arm rotated to the right
side as viewed from the front of the gantry.
D Stereotix Center, which is done with the Stereotix 2 installed and the tube–carrying arm rotated to the
center.
D Stereotix Right, which is done with the Stereotix 2 installed and the tube–carrying arm rotated to the left
side as viewed from the front of the gantry.
3–118
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
5.1 General instructions to be observed during all procedures in this Job Card
5.1.1 The photo cell must be in its most forward position (towards the patient: position1).
5.1.2 When placing plexiglass on the photocell or Stereotix 2 cassette holder, the plexiglass must always
overlap the front edge (towards the patient) by about 10 mm to ensure that the photocell is fully cov-
ered.
5.1.3 Error messages will appear on the console display when the tube–carrying arm position is not in
agreement with the position being calibrated.
JOB CARDS
Note: This must only be done if the Senographe DMR has been switched off since the last access to SET-
UP/GENE/INSTAL or SETUP/ARM/VALID/INSTAL
Note: Initially, do not put any plexiglass on the photocell. However, it may be necessary to add plexiglass
during these calibration procedures.
3–119
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
5.3.4 Press the 2nd trigger button and hold it down. One or more X–ray exposures are taken. Either the
message ”CALIBRATION END” or a message requesting a change in plexiglass thickness appears
on the console display. In either case, release the second trigger button.
If a message requesting a change in plexiglass thickness appeared on the console display, select
SETUP/CALIB on the console and repeat the exposure after changing the amount of plexiglass
by an increment of 10 mm.
If the message ”CALIBRATION END” appeared on the console display, proceed directly to the
Stereotix Left calibration in section 5.4.
5.4.2 Set the Stereotix 2 compression plate to its highest position. Set the needle holder to the center of
its lateral (x–axis) range and to the maximum depth (y–axis) range, i.e. as far as possible from the
patient.
5.4.3 Keep the same plexiglass thickness (if any) as used in Normal calibration section 5.3.
5.4.5 Rotate tube–carrying arm to its left Stereotix position (i.e. to the right side as viewed from the front
of the gantry). NORMAL CELL ERR appears on the console display.
5.4.6 Select CELL/LEFT/SETUP/CALIB on the console. STEREO. LEFT appears on the console dis-
play.
5.4.7 Place cassette (without film) on right side of Stereotix 2 cassette holder as viewed from the front
of the gantry.
5.4.8 Press the 2nd trigger button and hold it down. One or more X–ray exposures are taken. Either the
message CALIBRATION END or a message requesting a change in plexiglass thickness appears
on the console display. In either case, release the second trigger button.
If a message requesting a change in plexiglass thickness appeared on the console display:
– Repeat the normal calibration in section 5.3 using the new amount of plexiglass (gener-
ally the thickness is changed in increments of 10 mm).
– Repeat this Stereotix left calibration using the same new amount of plexiglass.
3–120
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
– Continue repeating the above two steps (if necessary) until a satisfactory plexiglass
thickness is found for both procedures.
– Proceed to the Stereotix Center calibration in section 5.5 using the same new amount
of plexiglass.
If the message CALIBRATION END appeared on the console display, proceed directly to the Ster-
eotix Center calibration in section 5.5.
5.5.1 Keep the Senographe DMR set up as outlined in sections 5.4.1 and 5.4.2.
5.5.2 Keep the same plexiglass thickness (if any) as used in Normal calibration section 5.3 and Stereotix
left calibration section 5.4.
JOB CARDS
5.5.4 Rotate tube–carrying arm to its center (i.e. non–Stereotix 2) position. STER. LEFT ERR appears
on the console display.
5.5.5 Select CELL/CENT/SETUP/CALIB on the console. STEREO. CENTER appears on the console
display.
5.5.7 Press the 2nd trigger button and hold it down. One or more X–ray exposures are taken. Either the
message CALIBRATION END or a message requesting a change in plexiglass thickness appears
on the console display. In either case, release the second trigger button.
If a message requesting a change in plexiglass thickness appeared on the console display:
– Repeat the normal calibration in section 5.3 using the new amount of plexiglass (gener-
ally the thickness is changed in increments of 1 cm).
– Repeat the Stereotix left calibration in section 5.4 using the same new amount of plexi-
glass.
– Repeat this Stereotix center calibration using the same new amount of plexiglass.
– Continue repeating the above three steps (if necessary) until a satisfactory plexiglass
thickness is found for all three procedures.
– Proceed to the Stereotix Right calibration in section 5.6 using the same new amount of
plexiglass.
If the message CALIBRATION END appeared on the console display, proceed directly to the Ster-
eotix Right calibration in section 5.6.
3–121
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
5.6.1 Keep the Senographe DMR set up as outlined in sections 5.4.1 and 5.4.2.
5.6.2 Keep the same plexiglass thickness (if any) as used in Normal calibration section 5.3, Stereotix left
calibration section 5.4 and Stereotix center calibration section 5.5.
5.6.4 Rotate tube–carrying arm to its right Stereotix position (i.e. to the left side as viewed from the front
of the gantry). STER. CENT ERR appears on the console display.
5.6.5 Select CELL/RIGHT/SETUP/CALIB on the console. STEREO. RIGHT appears on the console
display.
5.6.6 Move cassette to left side of Stereotix 2 cassette holder as viewed from the front of the gantry.
JOB CARDS
5.6.7 Press the 2nd trigger button and hold it down. One or more X–ray exposures are taken. Either the
message CALIBRATION END or a message requesting a change in plexiglass thickness appears
on the console display. In either case, release the second trigger button.
If a message requesting a change in plexiglass thickness appeared on the console display:
– Repeat the Normal calibration in section 5.3 using the new amount of plexiglass (gener-
ally the thickness is changed in increments of 1 cm).
– Repeat the Stereotix left calibration in section 5.4 using the same new amount of plexi-
glass.
– Repeat the Stereotix center calibration in section 5.5 using the same new amount of
plexiglass.
– Repeat this Stereotix right calibration using the same new amount of plexiglass.
– Continue repeating the above four steps (if necessary) until a satisfactory plexiglass
thickness is found for all four procedures.
– Be sure to execute the calculation as described in the next paragraph as soon as all four
procedures are satisfactorily completed.
If the message CALIBRATION END appeared on the console display, execute the calculation by
selecting SETUP/CALC/VALID on the console.
5.6.8 Perform a CKSUM and return to application mode. If you are handing the machine back to the user
at this point, switch the Senographe DMR off, then on again.
3–122
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
SECTION 1
SUPPLIES
D The cassette that will be used by the doctor for Stereotix 2 mode
Note: This MUST be the same cassette that was used in Job Card IST 018
D A supply of unexposed films of type used in the screen pair being calibrated
SECTION 2
JOB CARDS
TOOLS
D Densitometer
D Plexiglass in thickness increments of 5 mm (minimum plexiglass dimensions 200 x 200 cm to ensure
complete coverage of the photocell)
SECTION 3
SAFETY PRECAUTIONS
3–123
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
SECTION 4
PREREQUISITES
The Senographe DMR being installed must be equipped with the Stereotix 2 option that has been set up and
calibrated according to the Stereotix 2 Service Manual. The chronological order of AEC calibration given in
Installation Steering in chapter 1 must be strictly followed for each screen pair.
In case of Stereotix 2 add–on to an already–installed Senographe DMR, follow the instructions given in the
section entitled AEC Calibration Upon Stereotix 2 Add–on in Chapter 2.
SECTION 5
PREPARATION
GENERAL INSTRUCTIONS:
D The photo cell must be in its most forward position (towards the patient) throughout the procedures in
this Job Card.
JOB CARDS
D When placing plexiglass on the Stereotix 2 cassette holder, the plexiglass must always overlap the front
edge (towards the patient) by about 1 cm to insure that the photo cell is fully covered.
D Whenever the optical density of a film is measured, the measurement is always relative to the film
base fog level (Net density). Thus, you must null out the densitometer on the film base fog (unex-
posed area of the film), then measure the net density in the middle of the film.
D When measuring optical density of a Stereotix 2 film, the point of measure must be at about 5 mm below
the horizontal reference line, and in line with the vertical reference line.
5.1 Starting from application mode, change the position of the installation menu enable switch (switch B1 on
generator CPU board 400–PL3, see ”Accessing the Generator or Gantry Installation Mode from the
Console” in chapter 1).
Note: This must only be done if the Senographe DMR has been switched off since the last access to SET-
UP/GENE/INSTAL or SETUP/ARM/VALID/INSTAL
3–124
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
5.2.3 Load cassette with undeveloped film and install it in the cassette holder.
JOB CARDS
5.3 Perform calibration of reference energy in Stereotix 2 mode.
The objective is to calibrate the Senographe DMR for correct film blackening when in AEC (1–point) mode
with Stereotix 2.
Do not go any higher in the menu tree structure than given in the various
CAUTION steps in this section. If, for any reason, you do go higher in the tree structure,
it is essential to go into the CONFIG/STEREO menu first to avoid acciden-
tally using the default NORMAL mode, before proceeding with the calibra-
tion.
5.3.2 Select SETUP/PARAM/OD_R/CHANGE on the console. Enter the desired reference optical
density by using the NEXT and VALID keys and rotating the kV dial. See ”Visual Display of Param-
eter Values” and ”Modification of a Parameter Value or Entry of a Measurement” in chapter 1.
Note: Normally, this value should be the same as the value used for the non–Stereotix 2 calibration in
Job Card IST 014.
3–125
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
5.3.3 Perform a sensitometry test on the film being used with the screen pair being calibrated.
5.3.4 Select SETUP/SETUP/SCALE on the console. Rotate the kV dial until the value OD_3 is dis-
played on the console screen. Select CHANGE on the console. Enter the optical density from scale
number 3 of the sensitometry test by using the NEXT and VALID keys and rotating the kV dial.
See ”Visual Display of Parameter Values” and ”Modification of a Parameter Value or Entry of a
Measurement” in chapter 1.
5.3.5 Press the SETUP key on the console. Rotate the kV dial until the value OD_5 is displayed on the
console screen. Select CHANGE on the console. Enter the optical density from scale number 5 of
the sensitometry test by using the NEXT and VALID keys and rotating the kV dial.
5.3.6 Repeat section 5.3.5 for OD_7, OD_9, OD_11, OD_13 and OD_15 from the sensitometry test.
– SID = 510 mm
– No cassette holder / No bucky installed
– Stereotix 2 installed
– Tube–carrying arm in right Stereotix position (i.e. rotated to the left side as viewed
from the front of the gantry)
– Stereotix 2 compression paddle installed
– Stereotix 2 needle holder set to the center of its lateral (x–axis) range and to the maxi-
mum depth (y–axis) range, i.e. as far as possible from the patient
– 4 cm plexiglass on the Stereotix 2 cassette holder
Note: If the mAs value obtained at section 5.2.4 was under 40 mAs, use 5 cm plexiglass.
– Stereotix 2 compression paddle in contact with plexiglass
5.3.8 Install the cassette loaded with fresh film corresponding to the screen pair being calibrated. Place
it on the left side of the Stereotix 2 cassette holder as viewed from the front of the gantry.
Note: Be careful not to press the SETUP key more than one time after entering the mAs value. Doing
so would cause this value to revert to the default mAs value of +7.000E+1 (70).
3–126
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
5.3.11 Press the 2nd trigger button and hold it down. A single X–ray exposure is taken. Release the 2nd
trigger button.
5.3.12 Develop the film and measure its optical density according to the technique given at the beginning
of section 5.
If the optical density of the film is greater than or equal to (OD_R – 0.2), enter this value by
selecting SETUP/OD_MS and using the NEXT and VALID keys and the kV dial to enter the
measured optical density value, then skip directly to section below 5.3.13.
If the optical density of the film is less than (OD_R – 0.2), increase the mAs value by selecting
SETUP/mAs on the console, then entering the increased mAs value by using the NEXT and
VALID keys and the kV dial. Then re–load the cassette and place it on the left side of the Stereo-
tix 2 cassette holder as viewed from the front of the gantry. Select SETUP/CALIB on the con-
JOB CARDS
sole and repeat sections 5.3.11 and 5.3.12.
Note: Be careful not to press the SETUP key more than one time after entering the mAs value. Doing
so would cause this value to revert to the default mAs value of +7.000E+1 (70).
5.3.13 Select SETUP/CALC/VALID on the console to execute the calculation. CALCUL DONE appears
on the console display.
Note: This is NOT the film gamma, but another parameter that is calculated by the software.
– The reference energy (select SETUP/ENR_R on the console)
3–127
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
5.4.1 If necessary, enable use of the screen pair being calibrated by selecting SETUP/SETUP/SETUP/
SETUP/VALID/YES on the console.
HV: 28 kV
5.4.4 Keep the Senographe DMR in the same configuration as given in section 5.3.7.
5.4.5 Install the cassette loaded with fresh film corresponding to the screen pair being calibrated. Place
it on the left side of the Stereotix 2 cassette holder as viewed from the front of the gantry.
5.4.6 Take an exposure and note down the resulting mAs value.
5.4.7 Slide the cassette to the right side of the Stereotix 2 cassette holder.
5.4.8 Rotate tube–carrying arm to left Stereotix position (i.e. rotated to the right side as viewed from the
front of the gantry).
5.4.10 Develop the film and measure the optical density of the right Stereotix 2 image according to the
technique given at the beginning of section 5.
5.4.11 The optical density of the right Stereotix 2 image must not be outside the tolerance of the reference
optical density ±0.1 (the reference optical density was chosen in section 5.3.2). If the optical density
is within this tolerance, skip directly to section 5.4.13 below.
Note: The reference optical density chosen and entered in section 5.3.2 can no longer be displayed at
this point; it was erased and replaced by the default value when the machine was returned to
application mode for these checks.
3–128
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
5.4.12 If the optical density of the right Stereotix 2 image on the film is outside this tolerance, repeat the
calibration procedure, but, instead of entering the mAs value given in section 5.3.9, use the mAs
value obtained in section 5.4.6 above. To do this, perform the following sequence:
3. Repeat sections 5.3.7 to 5.3.14 EXCEPT THAT IN SECTION 5.3.9 YOU MUST ENTER THE
RESULTING MAS VALUE OBTAINED FROM THE CHECK EXPOSURE TAKEN IN SEC-
TION 5.4.6. DO NOT ENTER THE VALUE GIVEN IN SECTION 5.3.9.
JOB CARDS
5.4.13 Measure the optical density of the left Stereotix 2 image according to the technique given at the
beginning of section 5.
The difference between this density and the density of the right Stereotix 2 image must be less
than 0.1. If this is not the case, there is a problem with the calibration done in Job Card IST 018
”CALIBRATION OF PHOTO CELL IN STEREOTIX POSITIONS”. In this case, repeat Job
Card IST 018, then this Job Card in its entirety.
5.4.15 Install the cassette loaded with fresh film in the center of the Stereotix 2 cassette holder.
5.4.16 Keeping the Senographe DMR set up as given in section 5.4.3, take an exposure.
5.4.17 Develop the film and measure its optical density according to the technique given at the beginning
of section 5.
The difference between this density and the density of either the right Stereotix 2 image OR the
left Stereotix 2 image measured above must be less than 0.1. If this is not the case, there is a prob-
lem with the calibration done in Job Card IST 018 ”CALIBRATION OF PHOTO CELL IN
STEREOTIX POSITIONS”. In this case, repeat Job Card IST 018, then this Job Card in its
entirety.
5.4.18 If necessary, disable use of the screen pair being calibrated. (From application mode, select
SETUP/GENE/INSTAL/GENE/AOP/ALGO/FSC=x/VALID/NO on the console.) (See CAU-
TION in section 3.)
3–129
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
Blank page
JOB CARDS
3–130
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
SECTION 1
SUPPLIES
None
SECTION 2
TOOLS
None
JOB CARDS
SECTION 3
SAFETY PRECAUTIONS
SECTION 4
PREREQUISITES
This Job Card should only be used when the performance of Job Card IST 003 ”CALIBRATION OF X–RAY
TUBE HEATER CURRENT” has been impossible due to excessive arcing.
SECTION 5
AUTOMATIC X–RAY TUBE WARM–UP
This Job Card provides the procedure for re–starting an X–ray tube that was known to be good upon leaving
the factory, but which now exhibits arcing after a long period of inactivity.
The principle is to take a series of 12 pre–determined exposures, starting at low kV values and progressively
increasing towards high kV values. If arcing occurs during an exposure, some or all exposures are automati-
cally repeated. If persistent arcing occurs, the software automatically terminates the warm–up procedure.
3–131
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
5.1.1 Starting from application mode, change the position of the installation menu enable switch (switch
B1 on generator CPU board 400–PL3).
Note: This must only be done if the Senographe DMR has been switched off since the last access to SET-
UP/GENE/INSTAL or SETUP/ARM/VALID/INSTAL
5.2.1 Press the 2nd trigger button and HOLD IT DOWN while the exposures are taken.
JOB CARDS
Note: There is a time delay of 10 seconds between exposures. You can release the 2nd trigger button dur-
ing this 10 second period, then press it again to continue, but the minimum time between exposures
will still be 10 seconds. If the tube overheats at some point, exposures will be inhibited until it
cools down again. During the cooling period, the 2nd trigger button can be released; as soon as
the display shows that the X–ray tube has sufficiently cooled, press the 2nd trigger button to con-
tinue.
5.2.2 When the warm–up procedure is finished, the message ”Tube OK” normally appears on the console
display. If persistent arcing occurred during the procedures, the message ”Too many spits” appears
on the console display.
3–132
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
SECTION 1
SUPPLIES
D ”Ready Pack” or ”cardboard cassette” films (24 x 30 format)
SECTION 2
TOOLS
D Ruler calibrated in mm (minimum length 30 cm)
JOB CARDS
D Pen or pencil capable of marking the surface of an X–ray film
D Thumb tacks or push pins
D 2.5 mm Allen wrench
D 8 mm hollow offset tubular socket wrench
SECTION 3
SAFETY PRECAUTIONS
SECTION 4
PREREQUISITES
None
SECTION 5
CHECKING AND ADJUSTMENT OF LATERAL CENTERING OF X–RAY BEAM
The objective is to check (and adjust if necessary) the lateral alignment of the following three points:
D X–ray tube focal spot
D Center point between lateral collimator blades
D Center of image receptor support
3–133
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
5.1.2 Take a new cardboard cassette or Ready Pack and insert three thumb tacks through it, FROM THE
SIDE THAT WILL FACE THE X–RAY TUBE, in the following way:
– Two thumb tacks inserted along the (long) edge of the film that will be at the front edge
JOB CARDS
of the image receptor support, at approximately 2 cm in from this edge and approxi-
mately 4 cm inwards from the two shorter edges
– The third thumb tack along the (short) edge of the film that will be on the left side of
the image receptor support, approximately 2 cm in from this edge and approximately
12 cm from the rear edge of the film.
3–134
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
ILLUSTRATION 1
PLACEMENT OF THREE THUMB TACKS TO BE INSERTED IN THE READY PACK
APPROX 12 CM
READY PACK
TUBE SIDE
JOB CARDS
APPROX 2 CM
APPROX 4 CM
APPROX 4 CM
APPROX 2 CM
APPROX 2 CM
Note: It is NOT important to insert the three thumb tacks in a perfectly ”squared” alignment with the
edges of the Ready Pack.
5.1.3 Place the Ready Pack on the image receptor support so that the points of the two ”front” thumb tacks
butt perfectly up against the front edge of the image receptor support and the third thumb tack butts
perfectly up against the left edge of the image receptor support. THE POSITIONING OF THE
THREE THUMB TACKS AGAINST THE IMAGE RECEPTOR SUPPORT IS VERY IMPOR-
TANT TO THE SUCCESS OF THESE CHECKS.
3–135
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
5.2.1 Using normal application mode, set up a manual (2–point) exposure as follows:
– 25 kV
– 20 mAs
5.2.4 Using the ruler and a pen or pencil, carefully draw a straight line between the rearward edges (i.e.
the edges of the holes closer to the front edge of the image receptor support during the exposure)
of the two holes left by the thumb tacks on the front edge of the film. (See illustrations 2 and 3
below.) THIS LINE IS THE REFERENCE FOR ALL THE FOLLOWING GEOMETRICAL
JOB CARDS
MEASUREMENTS.
5.2.5 Attach the developed film to a sheet of graph paper (or other support marked with a series of perpen-
dicular lines), carefully aligning the front edge reference line with a convenient line on the graph
paper. See illustration 3 below.
ILLUSTRATION 2
DETAIL OF WHERE TO DRAW REFERENCE LINE ON FILM WITH RESPECT TO HOLES LEFT BY THUMB TACKS
TOWARDS
FRONT EDGE OF
HOLE LEFT IN FILM FILM
BY THUMB TACK
3–136
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
ILLUSTRATION 3
CHECKING EXPOSED FILM FOR LATERAL CENTERING OF X–RAY TUBE FOCAL SPOT WITH RESPECT TO THE
IMAGE RECEPTOR SUPPORT
DEVELOPED READY PACK FILM GRAPH PAPER EXPOSED AREA OF FILM
JOB CARDS
t2 t1
approx. 100 mm
3–137
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
5.3.1 Use the ruler to measure carefully distances t1 and t2 in mm in parallel with the reference line and
at approximately 100 mm rearward from the front edge of the film (see illustration 3).
5.3.3 IF E ≥ 3 mm, proceed to section 5.4 below. If not, this is the end of the Job Card.
5.4 Correct lateral centering (if necessary) of X–ray beam on image receptor support.
5.4.1 Determine, according to whether E is positive or negative, from which side of the X–ray tube hous-
ing the adjustment will be made:
JOB CARDS
5.4.2 Adjust lateral position of X–ray tube focal spot by following the procedure below:
1. Place the 8 mm hollow offset tubular socket wrench around the adjustment screw locknut
located on the side of the housing determined in section 5.4.1 above.
2. Hold the adjustment screw in the center of the locknut in place with the 2.5 mm Allen wrench.
4. Use the 2.5 mm Allen wrench to UNSCREW (turn COUNTER–clockwise) the adjustment
screw according to the formula NUMBER OF TURNS = 0.5 * E.
6. Loosen the corresponding locknut on the OPPOSITE side of the housing and TIGHTEN (turn
CLOCKWISE) the adjustment screw until snug, then re–tighten the locknut.
3–138
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
SECTION 1
SUPPLIES
None
SECTION 2
TOOLS
D Accurate digital voltmeter (Fluke 8060 or equivalent)
JOB CARDS
SECTION 3
SAFETY PRECAUTIONS
None
SECTION 4
PREREQUISITES
None
SECTION 5
FIRST POWER–ON CHECK
The objective in this section is to detect (before it’s too late) an incorrect mains supply configuration.
5.1.1 Check that generator Mains Supply Switch 200–S1 is open (i.e. in the ”0” position) and that mains
source is turned OFF.
5.1.2 Connect the voltmeter (set to 20 VDC scale) to generator mains power board 200–PL1 between test
points PT6 (+VE) and PT7 (0VE). See illustration 1 below.
3–139
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
ILLUSTRATION 1
LOCATION OF TEST POINTS PT6 AND PT7 FOR CHECKING KEEP–ALIVE SUPPLY VOLTAGE ON GENERATOR MAINS
DISTRIBUTION BOARD 200–PL1
JOB CARDS
PT6 PT7
5.1.3 Close generator Mains Supply Switch 200–S1 (i.e. turn it to the ”1” position) and read the voltmeter
immediately. It should read 8 ± 2 VDC.
If the voltmeter reading is outside this range, turn mains source OFF
CAUTION IMMEDIATELY and check configuration of voltage selection board IN2 on
generator mains distribution board 200–PL1 (see Job Card IST 011 ”CON-
NECTION OF MAINS SUPPLY”).
5.1.4 Connect the voltmeter (set to 500 VAC scale) to generator mains power board 200–PL1 between
terminals 1 (0V) and 4 (220V).
5.1.5 Turn console on and read the voltmeter immediately. It should read 220 ± 20 VAC.
If the voltmeter reading is outside this range, open generator Mains Supply
CAUTION Switch 200–S1 IMMEDIATELY (i.e. turn it to the ”0” position) and check
configuration of voltage selection board IN1 and mains voltage selection wire
”SEL” on generator mains distribution board 200–PL1
(see Job Card IST 011 ”CONNECTION OF MAINS SUPPLY”).
3–140
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
SECTION 6
CHECKING GENERATOR AND GANTRY DC SUPPLY VOLTAGES
6.1.1 On generator CPU board 400–PL3, measure DC voltage (+5 V nominal) between TP110 and TP111
(GND). The voltage should be in the range +5.1 V +/–0.05.
Note: This voltage can be adjusted via the potentiometer located on the switching power supply, just
beside the connector.
6.1.2 On generator interface board 400–PL2, measure DC voltage (+12 V nominal) between PT3 and
(PT4SERIG). The voltage should be in the range +11.5 to +12.5 VDC.
6.1.3 On generator interface board 400–PL2, measure DC voltage (–12 V nominal) between PT5 and
JOB CARDS
(PT4SERIG). The voltage should be in the range –11.5 to –12.5 VDC.
Note: The voltages measured in sections 6.1.2 and 6.1.3 above depend on the +5 V adjustment men-
tioned in section 6.1.1.
6.2.1 On gantry power board 800–PL1, measure DC voltage (+27V nominal) between terminals 3 and
1 (GND) of connector TB1. The voltage should be in the range +22 to +35 VDC.
6.2.2 On gantry CPU board 800–PL3, measure DC voltage (+5V nominal) between TP110 and TP111
(GND). The voltage should be in the range +4.75 to +5.25 VDC.
Note: Unlike the +5V supply on the generator CPU board, the +5V supply on the gantry CPU board can-
not be adjusted.
3–141
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
Blank page
JOB CARDS
3–142
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
SECTION 1
SUPPLIES
None
SECTION 2
TOOLS
D Calibrated force detector capable of measuring compression force between the compression paddle and
the cassette holder, required accuracy ±0.5 daN – the Mammo compression scale Part No.
JOB CARDS
46-194427P407 is recommended.
SECTION 3
SPECIAL SAFETY PRECAUTIONS
None
SECTION 4
PREREQUISITES
None
SECTION 5
CALIBRATION OF COMPRESSION FORCE DETECTOR
The objective is to calibrate the compression force detector.
WARNING
If a power line failure occurs during the calibration (before a checksum is done), it is mandatory to repeat this
calibration from the beginning.
5.1 Preparation
Starting from application mode, change the position of the installation menu enable switch (switch B1 on gen-
erator CPU board 400–PL3).
Note: This must only be done if the Senographe DMR has been switched off since the last access to SET-
UP/GENE/INSTAL or SETUP/ARM/VALID/INSTAL
3–143
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
5.2.2 Raise the compression paddle and place the force detector on the cassette holder.
5.2.4 With the compression paddle completely raised and out of contact (i.e. in its home position), press
the ”0daN” key on the console.
JOB CARDS
5.2.5 Lower the compression paddle into contact with the force detector. Using the compression fine–ad-
justment knob, adjust the compression to exactly 3 daN on the force detector.
5.2.7 Still using the compression fine–adjustment knob, increase the compression to exactly 6 daN on
the force detector.
5.2.9 Continue this process for 9 and 12 daN, then press the ”ETC” key on the console and proceed with
15, 18, 21, 24 and 27 daN.
5.2.10 Select SETUP/CALCUL on the console. This executes the calculation of the new force detector
calibration parameters.
5.2.12 Request a total decompression of the compression paddle (the paddle must return to its home posi-
tion).
5.3.1 With the arm at 0° (i.e. X–ray tube above the cassette holder), bring the compression paddle into
contact once again with the force detector. Increase the compression until maximum force is
reached (i.e. compression cannot be further increased). Read the force detector measurement and
check that it is within ±1 daN of the maximum compression force chosen in the SETUP/ARM/
VALID/MEDICAL/FORCE menu.
5.3.2 Repeat the above check with the arm rotated to 180° (i.e. X–ray tube below the cassette holder).
Note: The maximum compression force must never exceed 30 daN. This limit must always be checked
after any maintenance (preventive or corrective) of the compression assembly to insure patient
safety.
3–144
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
SECTION 1
SUPPLIES
None
SECTION 2
TOOLS
D A 3 kg weight whose dimensions allow it to be placed on top of the compression paddle
JOB CARDS
SECTION 3
SAFETY PRECAUTIONS
None
SECTION 4
PREREQUISITES
D Job Card IST 023 ”CALIBRATION OF COMPRESSION FORCE DETECTOR”
SECTION 5
MINIMUM COMPRESSION/DECOMPRESSION FORCE TESTS
The objectives are:
D Check that a compression force (limit IC: In Compression) of at least 22 daN can be reached when the
arm is rotated by 180° with the paddle drive motor current limit value set between 50% ard 70% (motor’s
maximum current rating).
D Check that a decompression force (limit OC: Out of Compression) of at least 3 daN can be reached when
the arm is at 0° with the paddle drive motor current limit value set to 65% of the motor’s maximum current
rating. (This ensures that the compression paddle will rise and stay up when decompression is requested.)
3–145
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
MINIMUM COMPRESSION/DECOMPRESSION
FORCE TESTS Job Card IST 024 2 of 4
5.1.1 Starting from application mode, change the position of the installation menu enable switch (switch
B1 on generator CPU board 400–PL3).
Note: This must be done only if the Senographe DMR has been switched off since the last access to
SETUP/GENE/INSTAL or SETUP/ARM/VALID/INSTAL
5.1.2 Set up the Senographe DMR in contact mode with a compression paddle installed.
5.1.3 Inhibit the cassette detection in medical menu. Starting from application mode:
SETUP/MEDICAL/FILM/NO
5.2.2 Rotate the arm to 180° (i.e, X–ray tube below the cassette holder).
5.2.3 Bring the compression paddle into contact with the cassette holder.
Note: Be ready to read the compression force displayed on the gantry readout unit, the paddle should
release and fall towards the tube housing for a compression close to 22 daN.
5.2.4 Increase compression to 22 daN as displayed on the gantry display panel. You must be able to reach
22 daN without the compression paddle releasing. If the compression paddle releases for a force
lower or higher than 22 daN, perform the adjustment in step 5.2.5; otherwise go to step 5.3.
5.2.5 Set up the limit IC to its minimum: Starting from application mode, select
SETUP/ARM/VALID/INSTAL/CALIB/COMP/LIMIT/CALIB/LIMIT_IC on the con-
sole.Use the + and - key on the console to set the motor current value to 50%. Enter this values with
the VAL key on the console.
5.2.6 Increase compression to 22 daN as displayed on the gantry display panel. You must be able to reach
22 daN without the compression paddle releasing.
5.2.7 If the compression paddle releases before a force of 22 daN is reached, increase the compression
motor current value by 5%, then try again to reach 22 daN. Continue increasing the maximum in
compression motor current value in steps of 5%, if necessary, until 22 daN can be reached.
Note: Between each compression, request a total decompression of the compression paddle, using the
appropriate footswitch.
3–146
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
MINIMUM COMPRESSION/DECOMPRESSION
FORCE TESTS Job Card IST 024 3 of 4
5.3.1 Rotate the arm to 0° (i.e, X–ray tube above the cassette holder).
5.3.2 Bring the compression paddle into contact with the cassette holder.
5.3.3 Select SETUP/LIMIT_OC on the console. The maximum motor current value (in decompression
mode) should be set to 65% otherwuise use the + and - key on the console to set this value. Enter
this value with the VAL key on the console and return to the application menu.
5.3.5 Request a total decompression of the compression paddle using the appropriate pedal. The com-
pression paddle must rise smoothly to its home position.
5.3.6 If the compression paddle slips or releases, check that the compression paddle drive belt is not slip-
JOB CARDS
ping. If it is slipping, adjust or replace the belt.
Note: The maximum out of compression motor current adjustment range is limited to 65%.
5.3.8 If you are handing the machine back to the user at this point, turn the Senographe DMR off, then
back on again.
3–147
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
MINIMUM COMPRESSION/DECOMPRESSION
FORCE TESTS Job Card IST 024 4 of 4
Blank page
JOB CARDS
3–148
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
SECTION 1
SUPPLIES
None
SECTION 2
TOOLS
D Plexiglass in thickness increments of 0.5 cm (minimum plexiglass dimensions 20 x 20 cm)
JOB CARDS
SECTION 3
SAFETY PRECAUTIONS
None
SECTION 4
PREREQUISITES
D Job Card IST 023 ”CALIBRATION OF COMPRESSION FORCE DETECTOR”
D Check the configuration of the senographe with a Plastic bucky or a Carbon bucky (see section 5 installa-
tion steering, step 14.b.ii).
SECTION 5
CALIBRATION OF BREAST THICKNESS MEASUREMENT
The objective is to calibrate the breast thickness measurement. Since the compression paddle and its drive
train, as well as the cassette holder or magnification device undergo deflection when under compression, a
compensation factor must be calculated to correct for changes in compression force at a fixed breast thickness.
Two calibrations must in fact be done, one for contact mode and one for magnification mode.
Note: This must only be done if the Senographe DMR has been switched off since the last access to SET-
UP/GENE/INSTAL or SETUP/ARM/VALID/INSTAL
3–149
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
5.2.1 Set up the Senographe DMR in contact mode with the 18 x 24 compression paddle installed.
5.2.2 Request a total decompression of the compression paddle to its home position.
Note: This MUST be done before entering the gantry installation menus.
5.2.4 Lower the compression paddle into contact with the cassette holder. Using the compression fine–
adjustment knob, adjust the compression to between 3 and 5 daN as read on the gantry display panel.
Press the ”<5daN” key on the console.
5.2.5 Increase the compression to between 15 and 20 daN as read on the gantry display panel. Press the
JOB CARDS
”>15daN” key on the console. The ”Calcul” key appears on the console display.
5.2.6 Press the ”Calcul” key on the console. This executes the calibration calculation for contact mode.
5.3.2 Keeping the Senographe DMR in contact mode, place 4 cm of plexiglass in the center of the bucky.
5.3.3 Lower the compression paddle into contact with the plexiglass.
5.3.4 Verify for any compression force between zero and 20 daN of compression (as displayed on the
gantry display panel), that the breast thickness displayed on the gantry display panel does not vary
outside the range of 35 to 45 mm (i.e. plexiglass thickness of 40 mm, ± 5 mm).
5.4.3 Request a total decompression of the compression paddle to its home position, using the appropriate
pedal.
5.4.4 Lower the compression paddle into contact with the magnification device. Using the compression
fine–adjustment knob, adjust the compression to between 3 and 5 daN as read on the gantry display
panel. Press the ”<5daN” key on the console.
5.4.5 Increase the compression to between 15 and 20 daN as read on the gantry display panel. Press the
”>15daN” key on the console. The ”Calcul” key appears on the console display.
3–150
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
5.4.6 Press the ”Calcul” key on the console. This executes the calibration calculation for magnification
mode.
5.5.3 Lower the compression paddle into contact with the plexiglass.
5.5.4 Verify for any compression force between zero and 20 daN of compression (as displayed on the
gantry display panel), that the breast thickness displayed on the gantry display panel does not vary
outside the range of 37 to 43 mm (i.e. plexiglass thickness of 40 mm, ± 3 mm).
JOB CARDS
5.6.1 Perform a CKSUM and return to application mode.
5.6.2 If you are turning the Senographe DMR back over to the user at this point, turn the Senographe
DMR off, then back on again.
3–151
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
Blank page
JOB CARDS
3–152
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
SECTION 1
SUPPLIES
None
SECTION 2
TOOLS
None
JOB CARDS
SECTION 3
SAFETY PRECAUTIONS
None
SECTION 4
PREREQUISITES
D Job Card IST 015 “CALIBRATION OF COMPRESSION ARM ANGULATION”. Arm angulation must
be correctly calibrated to ensure that grid movement is made in the correct direction when the angulation
is positive or negative.
SECTION 5
CALIBRATION OF GRID POSITION
The objective is to calculate the gain and offset necessary for determining the grid position. The calibration
is fully automatic. A check is included at the end of this automatic calibration to see if the calculated gain and
offset are within a certain range. If they are not, an error message appears on the console display.
5.1 Prepare the Senographe DMR for the calibration of grid position.
5.1.1 Starting from application mode, change the position of the installation menu enable switch (switch
B1 on generator CPU board 400–PL3).
Note: This must only be done if the Senographe DMR has been switched off since the last access to SET-
UP/GENE/INSTAL or SETUP/ARM/VALID/INSTAL
5.1.2 Set up the Senographe DMR in contact mode with the 18 x 24 bucky.
3–153
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
5.2.2 Select CALIB on the console. If the calibration is successful, ”CALIBRATION OK” appears on
the console display within a few seconds.
– If “Invalid gain/offset” or other error message is displayed, the problem must be cor-
rected.
5.2.3 On the console, select each of the following calibration parameters in turn; note the values and
check them against the acceptable values given in the table below:
OFFSET; GAIN; MIN; MAX
– If any of these values is out of range, the problem with the bucky must be corrected.
JOB CARDS
5.2.4 Repeat steps 5.2.2 and 5.2.3 with the arm rotated to +90°.
5.2.5 Repeat steps 5.2.2 and 5.2.3 with the arm rotated to –90°.
5.2.7 Return to application mode (a CKSUM is not necessary when grid position parameters have been
changed).
5.2.8 If you are handing the Senographe DMR back to the user at this point, turn the Senographe DMR
off, then back on again.
3–154
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
Note: This procedure is only applicable for Senographes DMR V1_P which have
been upgraded with motorized front blade kit.
For DMR with non–motorized front blade, perform Job Card IST 027B in
place of IST 027.
SECTION 1
SUPPLIES
D Ready Pack or cardboard cassette films (24 x 30 format)
JOB CARDS
SECTION 2
TOOLS
D Ruler calibrated in mm (minimum length 30 cm).
D Writing implement capable of marking the surface of an X–ray film.
D Thumb tacks or push pins (at least 20 mm long)
D Plexiglas plates in thickness increments of 5mm. Plates used for AEC calibration are adequate.
SECTION 3
SAFETY PRECAUTIONS
SECTION 4
PREREQUISITES
Job Card IST 029 must be done before commencing this procedure.
3–155
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
SECTION 5
CHECKING X–RAY FILM FORMAT GEOMETRY
The objective is to verify that the exposed area on the X–ray film is within the required tolerances for the fol-
lowing properties:
Parallelism between front collimator blade and front edge of image receptor support
Perpendicularity between left collimator blade and front edge of image receptor support
Perpendicularity between right collimator blade and front edge of image receptor support
Parallelism between rear collimator blade and front edge of image receptor support
Lateral centering of X–ray beam on image receptor support
Maximum overlap of X–ray beam beyond front edge of image receptor support (MQSA requirement)
Dimensions of exposed area (width and depth).
These checks are performed by making various geometrical measurements on an exposed cardboard cassette
JOB CARDS
1. Starting from application mode, change the position of the installation menu enable switch
(switch B1 on generator CPU board 400–PL3).
Note: This must only be done if the Senographe DMR has been switched off since the last access to
SETUP/GENE/INSTAL or SETUP/ARM/VALID/INSTAL
3. Press the + or – key on the console, repeatedly if necessary, until the desired lateral collimator
margin in millimetres is obtained (see the table below for standards and recommendations to
get full film blackening).
3–156
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
6. Press the + or – key on the console, repeatedly if necessary, until the desired rear collimator mar-
gin in millimetres is obtained (see the table below for standards and recommendations to get full
film blackening).
JOB CARDS
** Standard authorizes the full film blackening and in these tables, we indicate the margin configu-
ration to get it. But the user can also use margins inside the film.
9. Press the + or – key on the console, repeatedly if necessary, until the desired front collimator
margin in millimetres is obtained (see the table below for standards and recommendations to
get full film blackening).
3–157
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
5.1.2 Take a new cardboard cassette or Ready Pack and insert three thumb tacks through it, from the side
that will face the x–ray tube, in the following way:
– two thumb tacks inserted along the (long) edge of the film that will be at the front edge
of the image receptor support, at approximately 20mm in from this edge and approxi-
mately 40mm in from the two shorter edges
– the third thumb tack along the (short) edge of the film that will be on the left side of the
image receptor support, approximately 20mm in from this edge and approximately
120mm from the rear edge of the film.
Refer to Illustration 1 below.
ILLUSTRATION 1
PLACEMENT OF THREE THUMB TACKS TO BE INSERTED IN THE READY PACK
JOB CARDS
APPROX 120mm
READY PACK
TUBE SIDE
APPROX 20mm
APPROX 40mm
APPROX 40mm
APPROX 20mm
APPROX
20mm
Note: It is not important to insert the three thumb tacks in a perfectly squared alignment with the edges
of the Ready Pack.
3–158
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
5.1.3 Place 15 mm of plexiglas on the image receptor top as shown in illustration 2, to raise the Ready
pack film.
Note: This arrangement is done to simulate the exact film placement when a Bucky and a cassette are
used. This configuration is mandatory to measure accurately the X-Ray area in step 5.8, 5.9
&5.10 (particularly critical for front edge position).
The reason for such criticity is the tolerance decreasing (now 1% of SID for front edge overlap)
induced by various country regulations.
ILLUSTRATION 2
TOP VIEW
IMAGE RECEPTOR
TOP
JOB CARDS
THUMB TACK
PLEXIGLASS PLATE
(15 MM)
(Placed between film
and image receptor)
5.1.4 Place the Ready Pack on the plexiglas plate support so that the points of the two front thumb tacks
butt perfectly up against the front edge of the support and the third thumb tack butts perfectly up
against the left edge of the support. Correct positioning of all three thumb tacks is very important
for the success of the ensuing checks.
3–159
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
5.2.1 Using normal application mode, set up a manual (2–point) exposure as follows:
– Maximum format
– Mo track
– Large focal spot
– 25 kV
– 20 mAs
CAUTION It is critical that the TRACK is MO and that the FOCAL SPOT is LARGE.
JOB CARDS
5.2.4 Using the ruler and a pen or pencil, carefully draw a straight line between the rearward edges (i.e.
the edges of the holes closer to the front edge of the image receptor support during the exposure)
of the two holes left by the thumb tacks on the front edge of the film (see Illustration 3 below).
Draw another line parallel to the first one and shifted by 3 mm inside the film. This line represents
the edge of the film inside a cassette and is the reference for all the following measurements.
5.2.5 Attach the developed film to a sheet of graph paper (or other support marked with a series of perpen-
dicular lines), carefully aligning the front edge reference line with a convenient line on the graph
paper. See illustration 4 below.
ILLUSTRATION 3
DETAIL OF WHERE TO DRAW REFERENCE LINE ON FILM WITH RESPECT TO HOLES LEFT BY THUMB TACKS
REFERENCE LINE DRAWN LINE DETERMINED
IN INK OR PENCIL BY HOLES
3 mm
TOWARDS
FRONT EDGE OF
FILM
3–160
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
5.3 Check parallelism between the front collimator blade and the front edge of the image receptor support.
Use the ruler to measure carefully distances y1 and y2 (see illustration 4).
CRITERION: The difference between y1 and y2 must be LESS THAN 2 mm.
If the criterion is not met, loosen the collimator assembly hold–down screws and rotate the assembly to correct
the parallelism. Re–tighten the screws. Check by repeating the exposure. Once the parallelism is satisfactory,
proceed to the next check below (use the NEW film obtained from the most recent check exposure in all the
following checks).
5.4 Check perpendicularity between left collimator blade and front edge of image receptor support.
Use the ruler to measure carefully distances z1 and z2. (See illustration 5.)
CRITERION: The difference between z1 and z2 must be LESS THAN 1.5 mm.
JOB CARDS
Note: It is important to notice that z1 and z2 are measured between the left edge of the exposed area of
the film and ANY convenient perpendicular line on the graph paper.
If the criterion is not met, exchange the collimator. No field adjustment is possible.
5.5 Check perpendicularity between right collimator blade and front edge of image receptor support.
Use the ruler to measure carefully distances z1’ and z2’. (See illustration 5.)
CRITERION: The difference between z1’ and z2’ must be LESS THAN 1.5 mm.
Note: It is important to notice that z1’ and z2’ are measured between the right edge of the exposed area
of the film and ANY convenient perpendicular line on the graph paper.
If the criterion is not met, exchange the collimator. No field adjustment is possible.
5.6 Check parallelism between rear collimator blade and front edge of image receptor support
Use the ruler to measure carefully distances d1 and d2. (See illustration 6.)
CRITERION: The difference between d1 and d2 must be LESS THAN 2 mm.
Note: It is important to notice that d1 and d2 are measured between the rear edge of the exposed area and
the REFERENCE LINE.
If the criterion is not met, exchange the collimator. No field adjustment is possible.
3–161
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
EXAMPLE 1: If margin is set to +2 mm (CDRH Standard), then the criterion is: 239 mm < t1 < 245 mm.
EXAMPLE 2: If margin is set to +4 mm, then the criterion is: 243 mm < t1 < 249 mm.
If this criterion is not met, perform Job Cards IST 029 and IST 030, then repeat the check exposure (be sure
to re–configure the Senographe DMR as described in Section 5.1 above before taking the exposure) and
proceed to the next check below (use the NEW film obtained from this most recent check exposure in the
following check).
CRITERION: d3 must be in the range of (178 + MARG_R) ±3mm where MARG_R was determined and
entered in Section 5.1.
EXAMPLE: If Margin is set to +2 mm (CDRH standard) then the criterion is: 177 mm < d3 < 183 mm.
If this criterion is not met, perform Job Cards IST 029 and IST 030, then repeat the check exposure (be sure
to re–configure the Senographe DMR as given in THIS Job Card before taking the exposure) and proceed to
the next check bellow, using the film obtained with Job Card IST 030.
3–162
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
JOB CARDS
If these criteria are not met, perform Job Cards IST 029 and IST 030, then repeat the check exposure (be sure
to re–configure the Senographe DMR as described in Section 5.1 above before taking the exposure).
3–163
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
ILLUSTRATION 4
CHECKING EXPOSED FILM FOR PARALLELISM BETWEEN FRONT EDGE OF EXPOSED AREA AND FRONT EDGE OF
IMAGE RECEPTOR SUPPORT
y1 y2
3–164
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
ILLUSTRATION 5
CHECKING EXPOSED FILM FOR PERPENDICULARITY BETWEEN LEFT AND RIGHT COLLIMATOR BLADES AND
FRONT EDGE OF IMAGE RECEPTOR SUPPORT
DEVELOPED READY PACK FILM GRAPH PAPER EXPOSED AREA OF FILM
JOB CARDS
z2 Z2’
z1 z 1’
3–165
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
ILLUSTRATION 6
CHECKING EXPOSED FILM FOR PARALLELISM BETWEEN REAR COLLIMATOR BLADE AND FRONT EDGE OF
IMAGE RECEPTOR SUPPORT
DEVELOPED READY PACK FILM GRAPH PAPER EXPOSED AREA OF FILM
JOB CARDS
d1 d3 d2
3–166
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
ILLUSTRATION 7
CHECKING EXPOSED FILM FOR CENTERING OF LATERAL COLLIMATOR BLADES WITH RESPECT TO THE IMAGE
RECEPTOR SUPPORT
DEVELOPED READY PACK FILM GRAPH PAPER EXPOSED AREA OF FILM
JOB CARDS
t2 t1
approx. 100 mm
3–167
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
ILLUSTRATION 8
CHECKING EXPOSED FILM FOR MAXIMUM OVERLAP OF X–RAY BEAM BEYOND FRONT EDGE OF IMAGE RECEP-
TOR SUPPORT (MQSA REQUIREMENT)
DEVELOPED READY PACK FILM GRAPH PAPER EXPOSED AREA OF FILM
JOB CARDS
w1 w2
3–168
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
SECTION 1
SUPPLIES
D ”Ready Pack” or ”cardboard cassette” films (24 x 30 format)
SECTION 2
TOOLS
D Ruler calibrated in mm (minimum length 30 cm)
JOB CARDS
D Pen or pencil capable of marking the surface of an X–ray film
D Thumb tacks or push pins
SECTION 3
SAFETY PRECAUTIONS
SECTION 4
PREREQUISITES
None
3–169
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
SECTION 5
CHECKING X–RAY FILM FORMAT GEOMETRY
The objective is to verify that the exposed area on the X–ray film is within the required tolerances for the fol-
lowing properties:
Parallelism between front collimator blade and front edge of image receptor support
Perpendicularity between left collimator blade and front edge of image receptor support
Perpendicularity between right collimator blade and front edge of image receptor support
Parallelism between rear collimator blade and front edge of image receptor support
Lateral centering of X–ray beam on image receptor support
Maximum overlap of X–ray beam beyond front edge of image receptor support (CDRH requirement)
Dimensions of exposed area (width and depth)
These checks are performed by making various geometrical measurements on an exposed ”cardboard cas-
JOB CARDS
5.1.1 Set the lateral collimator margin according to local standards or the doctor’s preference:
1. Starting from application mode, change the position of the installation menu enable switch
(switch B1 on generator CPU board 400–PL3).
Note: This must only be done if the Senographe DMR has been switched off since the last access to SET-
UP/GENE/INSTAL or SETUP/ARM/VALID/INSTAL
3. Press the ”+” or ”–” key on the console, repeatedly if necessary, until the desired lateral collima-
tor margin (in mm) is obtained. (See table below for standards.)
Note: The default value for lateral collimator margin is that of the CDRH standard.
3–170
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
5.1.3 Take a new cardboard cassette or Ready Pack and insert three thumb tacks through it, FROM THE
SIDE THAT WILL FACE THE X–RAY TUBE, in the following way:
– Two thumb tacks inserted along the (long) edge of the film that will be at the front edge
of the image receptor support, at approximately 2 cm in from this edge and approxi-
mately 4 cm inwards from the two shorter edges
– The third thumb tack along the (short) edge of the film that will be on the left side of
the image receptor support, approximately 2 cm in from this edge and approximately
JOB CARDS
12 cm from the rear edge of the film.
Refer to illustration 1 below.
3–171
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
ILLUSTRATION 1
PLACEMENT OF THREE THUMB TACKS TO BE INSERTED IN THE READY PACK
APPROX 12 CM
READY PACK
TUBE SIDE
JOB CARDS
APPROX 2 CM
APPROX 4 CM
APPROX 4 CM
APPROX 2 CM
APPROX 2 CM
Note: It is NOT important to insert the three thumb tacks in a perfectly ”squared” alignment with the
edges of the Ready Pack.
5.1.4 Place the Ready Pack on the image receptor support so that the points of the two ”front” thumb tacks
butt perfectly up against the front edge of the image receptor support and the third thumb tack butts
perfectly up against the left edge of the image receptor support. THE POSITIONING OF THE
THREE THUMB TACKS AGAINST THE IMAGE RECEPTOR SUPPORT IS VERY IMPOR-
TANT TO THE SUCCESS OF THESE CHECKS.
3–172
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
5.2.1 Using normal application mode, set up a manual (2–point) exposure as follows:
– Maximum format
– Mo track
– Large focal spot
– 25 kV
– 20 mAs
CAUTION It is critical that the TRACK be MO and that the FOCAL SPOT be LARGE.
JOB CARDS
5.2.2 Take the exposure.
5.2.4 Using the ruler and a pen or pencil, carefully draw a straight line between the rearward edges (i.e.
the edges of the holes closer to the front edge of the image receptor support during the exposure)
of the two holes left by the thumb tacks on the front edge of the film. (See illustrations 3 and 4
below.) THIS LINE IS THE REFERENCE FOR ALL THE FOLLOWING GEOMETRICAL
MEASUREMENTS.
5.2.5 Attach the developed film to a sheet of graph paper (or other support marked with a series of perpen-
dicular lines), carefully aligning the front edge reference line with a convenient line on the graph
paper. See illustration 4 below.
ILLUSTRATION 2
DETAIL OF WHERE TO DRAW REFERENCE LINE ON FILM WITH RESPECT TO HOLES LEFT BY THUMB TACKS
REFERENCE LINE DRAWN
IN INK OR PENCIL
TOWARDS
FRONT EDGE OF
FILM
3–173
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
5.3 Check parallelism between front collimator blade and front edge of image receptor support.
Use the ruler to measure carefully distances y1 and y2. (See illustration 4.)
CRITERION: The difference between y1 and y2 must be LESS THAN 2 mm.
If the criterion is not met, loosen the collimator assembly hold–down screws and rotate the assembly to correct
the parallelism. Re–tighten the screws. Check by repeating the exposure. Once the parallelism is satisfactory,
proceed to the next check below (use the NEW film obtained from the most recent check exposure in all the
following checks).
5.4 Check perpendicularity between left collimator blade and front edge of image receptor support.
Use the ruler to measure carefully distances z1 and z2. (See illustration 5.)
CRITERION: The difference between z1 and z2 must be LESS THAN 1.5 mm.
Note: It is important to notice that z1 and z2 are measured between the left edge of the exposed area of
JOB CARDS
the film and ANY convenient perpendicular line on the graph paper.
If the criterion is not met, bend slightly the left collimator blade to correct the perpendicularity. Check by
repeating the exposure before proceeding. Once the perpendicularity is satisfactory, perform Job Card IST 030
”CALIBRATION OF LATERAL AND REAR COLLIMATOR BLADE REFERENCE POSITIONS”, then
repeat the check exposure one last time (be sure to re–configure the Senographe DMR as given in THIS Job
Card before taking the exposure) and proceed to the next check below (use the NEW film obtained from this
most recent check exposure in all the following checks).
5.5 Check perpendicularity between right collimator blade and front edge of image receptor support.
Use the ruler to measure carefully distances z1’ and z2’. (See illustration 5.)
CRITERION: The difference between z1’ and z2’ must be LESS THAN 1.5 mm.
Note: It is important to notice that z1’ and z2’ are measured between the right edge of the exposed area
of the film and ANY convenient perpendicular line on the graph paper.
If the criterion is not met, bend slightly the right collimator blade to correct the perpendicularity. Check by
repeating the exposure before proceeding. Once the perpendicularity is satisfactory, perform Job Card IST 030
”CALIBRATION OF LATERAL AND REAR COLLIMATOR BLADE REFERENCE POSITIONS”, then
repeat the check exposure one last time (be sure to re–configure the Senographe DMR as given in THIS Job
Card before taking the exposure) and proceed to the next check below (use the NEW film obtained from this
most recent check exposure in all the following checks).
3–174
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
5.6 Check parallelism between rear collimator blade and front edge of image receptor support
Use the ruler to measure carefully distances d1 and d2. (See illustration 6.)
CRITERION: The difference between d1 and d2 must be LESS THAN 2 mm.
Note: It is important to notice that d1 and d2 are measured between the rear edge of the exposed area and
the REFERENCE LINE.
If the criterion is not met, bend slightly the rear collimator blade to correct the parallelism. Check by repeating
the exposure before proceeding. Once the parallelism is satisfactory, perform Job Card IST 030 ”CALIBRA-
TION OF LATERAL AND REAR COLLIMATOR BLADE REFERENCE POSITIONS”, then repeat the
check exposure one last time (be sure to re–configure the Senographe DMR as given in THIS Job Card before
taking the exposure) and proceed to the next check below (use the NEW film obtained from this most recent
check exposure in all the following checks).
JOB CARDS
Use the ruler to measure carefully distances t1 and t2 at approximately 100 mm rearward from the front edge
of the film. (See illustration 7.)
CRITERION: 275 – (t1 + 2t2) must be LESS THAN 4 mm.
Where 275 mm is the width of the image receptor support and distances t1 and t2 are in mm.
If the criterion is not met, adjust the lateral position of the X–ray tube housing as described at the end of Job
Card IST 021 ”CHECKING AND ADJUSTMENT OF LATERAL CENTERING OF X–RAY BEAM”.
Check by repeating the exposure before proceeding (be sure to re–configure the Senographe DMR as given
in THIS Job Card before taking the exposure). Once the lateral centering is satisfactory, proceed to the next
check below (use the NEW film obtained from this most recent check exposure in all the following checks).
5.8 Check for maximum overlap of x–ray beam beyond front edge of image receptor support (CDRH
requirement).
Use the ruler to measure carefully distances w1 and w2. (See illustration 8.)
CRITERIA:
1. 8.5 mm < w1 < 11.5 mm
Note: If these two criteria are met, conformity with CDRH standards concerning beam overlap beyond
front edge of image receptor support is guaranteed for any track and SID (i.e. overlap < 2% of
SID).
3–175
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
If either or both of these criteria is not met, perform Job Card IST 028 ”ADJUSTMENT OF FRONT COLLI-
MATOR BLADE POSITION”. Check by repeating the exposure before proceeding (be sure to re–configure
the Senographe DMR as given in THIS Job Card before taking the exposure). Once the overlap is satisfactory,
proceed to the next check below (use the NEW film obtained from this most recent check exposure in all the
following checks).
Senographe DMR as given in THIS Job Card before taking the exposure) and proceed to the next check below
(use the NEW film obtained from this most recent check exposure in the following check).
3–176
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
ILLUSTRATION 3
CHECKING EXPOSED FILM FOR PARALLELISM BETWEEN FRONT EDGE OF EXPOSED AREA AND FRONT EDGE OF
IMAGE RECEPTOR SUPPORT
JOB CARDS
y1 y2
3–177
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
ILLUSTRATION 4
CHECKING EXPOSED FILM FOR PERPENDICULARITY BETWEEN LEFT AND RIGHT COLLIMATOR BLADES AND
FRONT EDGE OF IMAGE RECEPTOR SUPPORT
DEVELOPED READY PACK FILM GRAPH PAPER EXPOSED AREA OF FILM
JOB CARDS
z2 Z2’
z1 z 1’
3–178
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
ILLUSTRATION 5
CHECKING EXPOSED FILM FOR PARALLELISM BETWEEN REAR COLLIMATOR BLADE AND FRONT EDGE OF
IMAGE RECEPTOR SUPPORT
DEVELOPED READY PACK FILM GRAPH PAPER EXPOSED AREA OF FILM
JOB CARDS
d1 d3 d2
3–179
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
ILLUSTRATION 6
CHECKING EXPOSED FILM FOR CENTERING OF LATERAL COLLIMATOR BLADES WITH RESPECT TO THE IMAGE
RECEPTOR SUPPORT
DEVELOPED READY PACK FILM GRAPH PAPER EXPOSED AREA OF FILM
JOB CARDS
t2 t1
approx. 100 mm
3–180
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
ILLUSTRATION 7
CHECKING EXPOSED FILM FOR MAXIMUM OVERLAP OF X–RAY BEAM BEYOND FRONT EDGE OF IMAGE RECEP-
TOR SUPPORT (CDRH REQUIREMENT)
DEVELOPED READY PACK FILM GRAPH PAPER EXPOSED AREA OF FILM
JOB CARDS
w1 w2
3–181
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
Blank page
JOB CARDS
3–182
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
SECTION 1
SUPPLIES
D ”Ready Pack” or ”cardboard cassette” films (24 x 30 format)
SECTION 2
TOOLS
D Ruler calibrated in mm (minimum length 30 cm)
JOB CARDS
D Pen or pencil capable of marking the surface of an X–ray film
D Thumb tacks or push pins
D 5.5 and 7 mm Allen wrenches.
SECTION 3
SAFETY PRECAUTIONS
SECTION 4
PREREQUISITES
D Good parallelism between front collimator blade and front edge of image receptor support. See the
appropriate section of Job Card IST 027 ”CHECKING X–RAY FILM FORMAT GEOMETRY” or IST
027B “CHECKING X–RAY FILM FORMAT GEOMETRY FOR NON–MOTORIZED FRONT
BLADE SYSTEM”.
3–183
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
SECTION 5
CHECKING AND ADJUSTING (IF NECESSARY) FRONT COLLIMATOR BLADE POSITION
The objective is to check that the front collimator blade allows complete X–ray exposure up to the front edge
of the film without exceeding allowable overlap requirements.
If the front collimator blade is adjusted correctly for rhodium track, large focal spot and 660 mm SID, com-
pliance with CDRH requirements (overlap < 2% of SID) is guaranteed for all possible configurations.
5.1.2 Take a new cardboard cassette or Ready Pack and insert three thumb tacks through it, FROM THE
SIDE THAT WILL FACE THE X–RAY TUBE, in the following way:
– Two thumb tacks inserted along the (long) edge of the film that will be at the front edge
of the image receptor support, at approximately 2 cm in from this edge and approxi-
mately 4 cm inwards from the two shorter edges
– The third thumb tack along the (short) edge of the film that will be on the left side of
the image receptor support, approximately 2 cm in from this edge and approximately
12 cm from the rear edge of the film.
Refer to illustration 1 below.
3–184
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
ILLUSTRATION 1
PLACEMENT OF THREE THUMB TACKS TO BE INSERTED IN THE READY PACK
APPROX 12 CM
READY PACK
TUBE SIDE
JOB CARDS
APPROX 2 CM
APPROX 4 CM
APPROX 4 CM
APPROX 2 CM
APPROX 2 CM
Note: It is NOT important to insert the three thumb tacks in a perfectly ”squared” alignment with the
edges of the Ready Pack.
5.1.3 Place the Ready Pack on the image receptor support so that the points of the two ”front” thumb tacks
butt perfectly up against the front edge of the image receptor support and the third thumb tack butts
perfectly up against the left edge of the image receptor support. THE POSITIONING OF THE
THREE THUMB TACKS AGAINST THE IMAGE RECEPTOR SUPPORT IS VERY IMPOR-
TANT TO THE SUCCESS OF THESE CHECKS.
3–185
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
5.2.1 Using normal application mode, set up a manual (2–point) exposure as follows:
D Mo track
D Large focal spot
D 25 kV
D 20 mAs
CAUTION It is critical that the TRACK be Mo and that the FOCAL SPOT be LARGE.
JOB CARDS
5.2.4 Using the ruler and a pen or pencil, carefully draw a straight line between the rearward edges (i.e.
the edges of the holes closer to the front edge of the image receptor support during the exposure)
of the two holes left by the thumb tacks on the front edge of the film. (See illustrations 2 and 3
below.) THIS LINE IS THE REFERENCE FOR ALL THE FOLLOWING GEOMETRICAL
MEASUREMENTS.
5.3.1 Use the ruler to measure carefully distances w1 and w2. (See illustration 3 below.)
Note: w1 and w2 are acceptable as they are if they are both within the following limits:
If w1 and/or w2 are outside these limits, proceed to the next section. If w1 and w2 are both within
these limits, the next section is optional.
3–186
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
5.4.1 Use the 7 mm Allen wrench to loosen the two front collimator blade locking screws.
5.4.3 If the value of ”d” measured above is POSITIVE, use the 5.5 mm Allen wrench to turn the front
collimator blade adjustment screw COUNTER–CLOCKWISE.
5.4.4 If the value of ”d” measured above is NEGATIVE, use the 5.5 mm Allen wrench to turn the front
collimator blade adjustment screw CLOCKWISE.
5.4.5 Use the 7 mm Allen wrench to re–tighten the two front collimator blade locking screws.
JOB CARDS
5.4.6 Re–tighten the adjustment screw locknut.
5.4.7 Repeat the exposure and check results as given in sections 5.1 to 5.3.
ILLUSTRATION 2
DETAIL OF WHERE TO DRAW REFERENCE LINE ON FILM WITH RESPECT TO HOLES LEFT BY THUMB TACKS
TOWARDS
FRONT EDGE OF
HOLE LEFT IN FILM FILM
BY THUMB TACK
3–187
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
ILLUSTRATION 3
CHECKING EXPOSED FILM FOR MAXIMUM OVERLAP OF X–RAY BEAM BEYOND FRONT EDGE OF IMAGE RECEP-
TOR SUPPORT (CDRH REQUIREMENT)
DEVELOPED READY PACK FILM EXPOSED AREA OF FILM
JOB CARDS
w1 w2
3–188
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
ILLUSTRATION 4
ADJUSTMENT OF FRONT COLLIMATOR BLADE
JOB CARDS
FRONT COLLIMATOR
BLADE
Í
Í
ÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇ
ADJUSTMENT SCREW LOCKNUT
Í
ÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇ
Í
5.5 MM HEX SOCKET HEAD
ADJUSTMENT SCREW
FRONT SIDE OF
SENOGRAPHE DMR
3–189
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
Blank page
JOB CARDS
3–190
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
SECTION 1
SUPPLIES
D 18 x 24 mm film and cassette
SECTION 2
TOOLS
D Ruler calibrated in mm (minimum length 20 cm)
JOB CARDS
SECTION 3
SAFETY PRECAUTIONS
SECTION 4
PREREQUISITES
D None
SECTION 5
CALIBRATION OF X–RAY TUBE REFERENCE ANGLE
The objective is to determine the X–ray tube angle when in its reference position.
This calibration is done by taking an exposure (at a given SID) whose field coverage depth is limited by the
anode geometry and not by the rear collimator blade, measuring this field coverage depth, and entering this
depth into the Senographe DMR which subsequently calculates the X–ray tube angle when in its reference
position.
3–191
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
CAUTION The SID must not be changed from 560 mm until successful completion of
this Job Card.
JOB CARDS
5.1.5 Develop the film and use the ruler to measure the distance between the front edge (edge closest to
the patient) of the film and the rear edge of the blackened area.
Note: Be sure to measure from the front edge of the FILM, even if the blackened area does not extend
to it.
Note: Be sure that the rear edge of the blackened area truly represents the end of the anode, and not a
stuck rear collimator blade (normally the software commands ”full open” during the exposure)
or other obstacle. This can be determined by the abruptness of the change in contrast between the
blackened area and the clear area. If the change is gradual, the exposure is good. If it is very abrupt,
it is necessary to correct the problem and repeat the calibration.
Note: If the blackened area of the film extends all the way to its rear edge, the X–ray tube reference angle
is severely incorrect and this calibration cannot be performed.
5.1.6 Select SETUP/SETUP/SETUP/MEASURE on the console and enter the measured value in mm
using the NEXT and VALID keys.
3–192
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
5.1.7 Select SETUP/Calculation on the console to execute the X–ray tube angle calculation.
Normally, the message ”calib OK” appears on the console display.
If the message ”calib error” appears instead, this means that the calculated X–ray tube reference
angle value is out of range. To check this, select SETUP/SETUP/PARAM on the console. The dis-
played value of the calculated X–ray tube reference angle A_REF must be between +2.800E–1 and
+3.500E–1 (0.28 and 0.35) radians.
If the angle is out of range, correct the problem and repeat the calibration (check first that the SID
was effectively set to 560 mm).
5.1.9 If you’re turning the machine back over to the user at this point, turn the Senographe DMR off, then
back on again.
JOB CARDS
3–193
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
Blank page
JOB CARDS
3–194
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
SECTION 1
SUPPLIES
D 18 x 24 cm film and cassette
SECTION 2
TOOLS
D Ruler calibrated in mm (minimum length 20 cm)
JOB CARDS
SECTION 3
SAFETY PRECAUTIONS
SECTION 4
PREREQUISITES
D Job Cards IST 029 and IST 027 must have been done before commencing this procedure (these
prerequisites are given as a reminder only).
D Job Cards IST 028 if the collimator has a non–motorized front blade.
SECTION 5
CALIBRATION OF ALL COLLIMATOR BLADE REFERENCE POSITIONS
The objective is to determine the lateral, rear, and front, collimator reference positions. The result is that the
software will then be able to command and obtain any desired X–Ray format.
This calibration is done by taking three exposures (at a given SID):
– A first exposure with blades at optical fork reference.
– A second exposure as large as possible.
– A third exposure as small as possible.
The width and depth positions of the blackened areas of the exposed film are then measured and entered in
the Senographe DMR which in turn calculates and saves the reference positions.
3–195
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
Note: A warning message “Set Sid to 560” will appear in the sub–menu if the SID is not correctly set.
In this case, simply reset the SID to 560. The message dissapears when the menu level is changed
(up or down).
JOB CARDS
1. The SID must not be changed from 560 mm until successful completion of
CAUTION this Job Card – the value is now the default for SID in the following calibra-
tion menus.
2. This calibration uses the full surface area of the film. Check that no mask
or other object is present in the beam before proceeding.
5.1.3 Select:
SETUP/ARM/VALID/INSTAL/CALIB/FORMAT/CALIB/COLLIM/EXPOSE/VALID/
VALID
Note: The default value of 10 mAs can be modified if necessary; for normal cassettes use 5 mAs for each
exposure. The kV value is automatically set to 22 kV.
Too dark an exposure will produce a shadow around the exposed area, decreasing distance mea-
surement accuracy.
5.1.4 Take the three exposures in sequence; press the 2nd trigger button and hold it until the third
exposure is acquired (the messages: “1st exposure”, 2nd exposure”, “3rd exposure”, then “end of
acquisition”, are displayed).
5.1.5 Develop the film and use the ruler to measure the differents distances as shown in Illustration 1.
3–196
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
Note: During this procedure, warning messages Value too high or Value too low will be displayed if an
out-of-range value is entered. If this happens, check the value.
ILLUSTRATION 1
BLACKENED FILM AREA DIMENSIONING
L2
JOB CARDS
R2
L1
L3
R1
R3
5.2.3 Select SETUP/REF_R/MEASURE and enter the measured (R1) value in mm using the NEXT and
VALID keys. The valid range is 128.6 < R1 < 142.3.
5.2.5 Select SETUP/REF_F/MEASURE and enter the measured (A1) value in mm using the NEXT and
VALID keys. The valid range is 7.15 < A1 < 16.86.
5.2.7 L2 – L3 value must be entered as follows: Select SETUP/DIFF–L/MEASURE and enter the (L2
– L3) value in mm using the NEXT and VALID keys.
3–197
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
5.2.9 R2 – R3 value must be entered as follows: Select SETUP/DIFF–R/MEASURE and enter the (R2
– R3) value in mm using the NEXT and VALID keys.
5.2.11 The calibration procedure is now complete. Perform a checksum as follows: SETUP/GENE/
INSTAL/CKSUM.
Note: This calibration procedure adjusts the values of STP_L and STP_R to improve shutter positioning
accuracy. The new values after calibration should be close to the default values (below).
To check for correct operation of a collimator, or if you find a problem with the calibration, start
with the default parameter values given here:
3–198
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
SECTION 1
SUPPLIES
D Film and cassette of the type and size(s) used by the customer on the Senographe DMR being installed
(18 x 24 mm, plus 24 x 30 mm if this option is chosen by the customer)
SECTION 2
TOOLS
D Plexiglass in thickness increments of 0.5 cm (minimum plexiglass dimensions 20 x 20 cm to insure com-
JOB CARDS
plete covering of the photo cell)
SECTION 3
SAFETY PRECAUTIONS
SECTION 4
PREREQUISITES
D Job Card IST 026 ”CALIBRATION OF GRID POSITION”
D Complete AEC calibration for the screen pair to be used in this test (see Installation Steering in chapter 1).
SECTION 5
TEST FOR ABSENCE OF GRID LINES ON EXPOSED FILM
The objective is to test for correct grid movement by taking exposures and examining the developed films for
grid lines .
3–199
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
• Different people, whether customers or Service Engineers, have very different visual percep-
tions of grid lines.
• When the fault is not very pronounced, it is usually not present over the entire film, but only in
certain areas (at the edge of the film, as follows):
• The following display conditions are essential for grid line observation:
• Viewer light intensity must be high and uniform,
JOB CARDS
5.1 General instructions to be observed during all procedures in this Job Card
5.1.1 The photo cell must be in its most forward position (towards the patient) throughout the procedures
in this Job Card.
5.1.2 When placing Plexiglass on the cassette holder, the Plexiglass must always overlap the front edge
(towards the patient) by about 1 cm to insure that the photo cell is fully covered.
5.2 Perform the test for absence of grid lines on exposed film.
5.2.1 Set up the Senographe DMR in contact mode with the 18 x 24 mm bucky, grid and compression
paddle installed.
3–200
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
5.2.4 Enable visual display of parameter values when in application mode. See ”Visual Display of Param-
eter Values” in chapter 1.
5.2.5 Load the cassette with a fresh undeveloped film that corresponds to the screen pair being used for
the test and install it in the bucky.
JOB CARDS
thickness sure
(mm) time
(ms)
1 AEC MO MO 15 25
2 AOP–CNT 15
3 AOP–CNT 25
4 AOP–CNT 35
5.2.7 Note the exposure time as displayed on the console display. If it is within the range 0.200 to 0.300
S, develop the film and proceed directly to step 5.2.10. If it is above (below) this range, increase
(decrease) the HV by 1 kV and repeat the exposure using the same undeveloped film in the cassette.
Continue increasing or decreasing the HV by 1 kV until a satisfactory exposure time is reached, then
proceed to the next step.
Note: For an unusually slow film (i.e. if the exposure time is still greater than 300 mS at 30 kV), remove
the Plexiglass from the bucky and continue the search for a satisfactory exposure time.
Note: For an unusually fast film (i.e. if the exposure time is still less than 200 mS at 22 kV), increase
the amount of Plexiglass on the bucky to 2 cm and continue the search for a satisfactory exposure
time.
5.2.8 Re–load the cassette with a fresh undeveloped film that corresponds to the screen pair being used
for the test and install it in the cassette holder.
5.2.9 Keeping the same HV value as determined above, take an exposure and develop the film.
3–201
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
5.2.13 If the Senographe DMR being installed is equipped with the 24 x 30 mm bucky/grid option, per-
form an exposure in the conditions of Exposure N°4 (step 5.2.12) using 24X30 option.
5.2.14 Study the developed films on a light screen with the jeweler’s loop. Verify that there are no signs
of grid lines anywhere on the exposed area of the film. Visible grid lines indicate a problem in the
grid drive train or defective offsets and gain settings. If there are visible grid lines, proceed to steps
5.2.15 to 5.2.16.
5.2.15 Re–calibrate gain and offset following the bucky trouble shooting Job Card.
5.2.16 Check also there are any sticky or rough points by moving the grid by hand back and forth across
its full range of movement.
5.2.17 If you are turning the machine back over to the user at this point, switch the Senographe DMR
off, then back on again.
JOB CARDS
3–202
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
SECTION 1
SUPPLIES
None
SECTION 2
TOOLS
D Metric Allen wrenches
D Large flat screwdriver
JOB CARDS
SECTION 3
SAFETY PRECAUTIONS
None
SECTION 4
PREREQUISITES
None
SECTION 5
REMOVAL AND REPLACEMENT OF SIDE COVERS AND TRIM PANELS
This section is divided into two subsections, one for removal and replacement of generator side covers and
trim panels, and one for removal and replacement of gantry side covers and trim panels.
5.1 Remove/replace generator side covers and trim panels (see illustration 1).
5.1.1 Remove all power from the Senographe DMR by turning off the wall–mounted circuit breaker.
5.1.2 Remove the four plastic plugs from screws of generator top panel.
5.1.3 Use an Allen wrench to remove the four generator top panel retaining screws (5).
3–203
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
2. Disconnect the ground wire FROM THE GENERATOR END to avoid RISK OF CREATING
SHORT CIRCUITS.
5.1.6 Perform the procedure in reverse order for replacement. BE SURE TO RE–CONNECT THE
GROUND WIRE ON EACH OF THE FOUR SIDE PANELS.
5.2 Remove/replace gantry side covers and trim panels (see illustration 1).
5.2.3 Remove plastic plugs and screws (2) and (3) from each column half–cover.
5.2.4 Do not touch screws (4). Undo hex bolt (6) by 2 or 3 turns (accessible through a slot at rear of the
base).
(this method keeps alignment of covers with safety stop switches).
5.2.6 Perform the procedure in reverse order for replacement. BE CAREFUL NOT TO PINCH OR
DAMAGE CABLES.
3–204
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
ILLUSTRATION 1
SCREW LOCATIONS FOR REMOVING/REPLACING SIDE COVERS AND TRIM PANELS
JOB CARDS
2
5
1
4
6
3–205
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
Blank page
JOB CARDS
3–206
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
SECTION 1
SUPPLIES
None
SECTION 2
TOOLS
D Digital mAs meter (GSPONER type) able to display tube current
D Compression measurement tool: GAMMEX RMI model 817 or Mammo compression scale
JOB CARDS
# 46–194427P407 (digital display)
D Plexiglass in thickness increments of 0.5 cm
SECTION 3
SAFETY PRECAUTIONS
SECTION 4
PREREQUISITES
None
3–207
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
SECTION 5
GENERATOR TESTING
The objective is to check quickly the generator kV, mA and mAs calibrations, which have already been done
in the factory. At the same time, these procedures will warm up the X–ray tube slightly in preparation for the
AEC calibration.
5.1.3 Connect the digital mAs meter to kV mA board 701–PL1 (on HV tank unit 701) between terminals
MES_MA_POS and MES_MA_NEG. Disconnect the jumper that is normally between these same
two terminals.
5.1.5 Take 5 exposures at approximately 20 second intervals, incrementing the HV by 2 kV each time
(i.e. 22 – 24 – 26 – 28 – 30 kV).
5.1.6 Take 5 more exposures at the same interval, leaving the HV set at 30 kV.
5.1.7 The X–ray tube current value displayed on the console (Idisplay) must be 100 mA ± 5 mA AND in
agreement with the value read on the mAs meter (Imeasure) ± 5% (this value is available both for
GS412 and Maxiray 70TH–F tubes).
IF Idisplay ≠ Imeasure then perform Job Card IST 016 ”CALIBRATION OF X–RAY TUBE MA
MEASUREMENT” followed by Job Card IST 003 ”CALIBRATION OF X–RAY TUBE
HEATER CURRENT”.
5.1.8 Take 2 exposures at 40 kV and 50 mAs at an interval of 30 seconds. No arcing should be detected.
If arcing is detected, perform IST 020 ”AUTOMATIC X–RAY TUBE WARM–UP”.
3–208
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
5.1.9 Re–configure the Senographe DMR to 30 kV and 60 mAs. Take one exposure for each of the three
following focal spot/track combinations and verify the expected X–ray tube current reading on the
console display:
The values below apply for tube Maxiray 70TH–F
SMALL FOCAL SPOT/MO TRACK => I = 40 mA ± 2 mA
SMALL FOCAL SPOT/RH TRACK => I = 35 mA ± 2 mA
LARGE FOCAL SPOT/RH TRACK => I = 62 mA ± 4 mA
JOB CARDS
5.1.10 Disconnect the mAs meter from the Senographe DMR and re–connect the jumper on kV mA board
701–PL1.
5.1.11 Verify kVp using a non–invasive kVp meter, and if needed, recalibrate kVp with IST 017 (this step
may be necessary, due to local regulation)
SECTION 6
GANTRY TESTING
The objective is to check quickly the basic functions of the gantry.
Note that the following functions are tested by the software during each power–up sequence:
D Filter holder rotation
D Lateral, front and rear collimator blade movement
D X–ray tube housing movement
D Compression paddle movement to home position
D Column movement limit stops
D Coherency between focal spot size, magnification and X–ray tube housing angulation
D Grid movement
D Generator–gantry communications link
D Gantry CPU board
3–209
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
Note: X–Ray format readout displays the actual X–ray field depending on margins settings configured
in steering guide section 14.v.
6.2.1 With the compression paddle out of compression, check that all five column up/down control
buttons are operational (check each button for up and down movement of column).
6.2.2 Move column to its upper and lower limits. Optical sensors should stop movement. If optical sen-
sors have failed, mechanical safety stops halt movement. Total one–way movement time
JOB CARDS
(going either up or down) between upper and lower limits should be 16 ± 2 seconds. This time
includes initial acceleration which lasts 3 seconds.
6.3.1 With the compression paddle out of compression, check that all four arm rotation control buttons
are operational (check each button for release and re–locking of arm rotation).
6.3.2 Rotate arm through its full range of ± 180°. Check that electro–brake is sufficient by manual action
on the handles.
6.4.1 Press one of the light centering device activation buttons. The light centering device should light
for 30 seconds for large formats (eg: 18x24, 24x30). The light centering device should light for
120 seconds for medium and small formats.
6.6.2 With magnification device removed, the gantry readout should show a magnification of 1.0.
6.6.3 Install the magnification device into the lower, middle and upper holes successively. The gantry
readout should show a magnification of 1.5, 1.8 and 1.9, respectively.
6.6.4 Leaving the magnification device in the upper holes, change the SID to 610, 560 and 510 mm
successively. The gantry readout should show a magnification of 1.7, 1.6 and 1.5, respectively.
3–210
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
6.7.1 With Stereotix 2 NOT installed, activate the tube housing arm locking lever and rotate the tube
housing arm to its ± 15° and ± 27° detents. When tube housing arm is locked in each of these four
positions, the error message ”Incorrect X_Ray Incidence” should appear on the console display.
6.7.2 Rotate the tube housing arm so that it is ”floating” between detents. After five seconds, the error
message ”Tube Housing Arm Unlocked” should appear on the console display and ”Arm
Unlocked” should appear on the gantry readout.
6.8.1 Verify that both sets of compression/decompression pedals and the manual compression knobs
located on the paddle holder all function correctly (i.e. the paddle goes up and down in response
to each command).
JOB CARDS
6.8.2 Use the mammo compression scale placed between the cassette holder and the compression paddle
to check calibration of the compression force display on the gantry readout. Measure one point at
5 daN and one point at 15 daN. The value on the gantry readout must agree within ± 1 daN. If it does
not, perform IST 023 ”CALIBRATION OF COMPRESSION FORCE DETECTOR”.
Note: If the recommended compression measurement tool is not available, this verification can be done
by setting SID to 660 mm, lowering the paddle close to the top surface of the cassette holder, instal-
ling the magnification device in the upper pair of holes, and connecting a spring dynamometer
(calibrated from 0 to 15 kg) between the compression paddle and the magnification device.
6.9.1 Set SID to 660 mm, install the 18x24 bucky and place 10 mm of plexiglass in the center of the bucky.
6.9.2 Lower the compression paddle into contact with the plexiglass.
6.9.3 Slowly increase compression force on the plexiglass while watching the breast thickness display
on the gantry readout. The displayed value should remain in the range 10 mm ± 3 mm for any value
of compression force (as displayed on the gantry readout) between 5 and 15 kg.
6.9.4 Replace 10 mm of plexiglass with 50 mm of plexiglass and repeat the above check. Now the dis-
played value should remain in the range 50 mm ± 3 mm for any value of compression force
(as displayed on the gantry readout) between 5 and 15 kg.
6.9.5 If either or both of the above checks does not give a satisfactory result, perform Job Card IST 025
”CALIBRATION OF BREAST THICKNESS MEASUREMENT”.
3–211
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
6.11.1 With the bucky installed, press the 1st trigger button on the console. There should be grid
movement.
3–212
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
SECTION 1
SUPPLIES
D Supply of undeveloped film of type used by the customer
SECTION 2
TOOLS
D Sensitometer
JOB CARDS
D Densitometer
D Opaque sheet
SECTION 3
SAFETY PRECAUTIONS
None
SECTION 4
PREREQUISITES
None
SECTION 5
CHECKING FILM PROCESSING
The objective is to check the light–tightness of the darkroom, the influence of the safelighting and correct
operation of the automatic film processor.
5.1.1 Enter the darkroom and load a cassette with the film currently used for clinical purposes, taking care
to protect it from possible fogging from the darkroom safe light or leakage.
Note 1 : The emulsion side of the film must be positioned facing the screen side of the cassette (away from
the tube)
Note 2 : For Kodak Ortho–MA film, the emultion side is underneath when the notch is top left.
5.1.2 In 2–point mode, (HV: 28 kV, mAs=70, 4 cm of plexiglas) expose the cassette to the x–rays in such
a way as to obtain a film with an optical density of between 0.8 and 1.4 (change mAs as necessary
and repeat exposure with a fresh film).
3–213
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
5.1.3 Reload the cassette and make an exposure using the same factors; do not develop the film.
5.1.4 Unload the cassette in the darkroom and place the film on the work surface, emulsion side upwards
and completely covered by an opaque sheet.
Note: for KODAK ORTHO–MA films, the emulsion side is uppermost when the notch is top right.
5.1.5 Prepare the second opaque sheet on the work surface, on the left.
5.1.6 Move the opaque sheet covering the film to the right, uncovering half the film (see illustration 2)
and simultaneously activate the timer.
5.1.7 If this is the first maintenance inspection (reference test) proceed to the next step (5.1.8).
If this is a periodic maintenance check, go to step (5.1.11).
5.1.8 After 15 seconds, cover about a quarter of the bottom lefthand part of the film with the second
JOB CARDS
5.1.9 After 30 seconds, move the second opaque sheet upwards in such a way as to cover about half of
the lefthand part of the film (2nd area).
5.1.10 After one minute, move the second opaque sheet upwards to cover about three quarters of the left-
hand part of the film (3rd area).
5.1.11 After 2 minutes, cover the film completely with the first opaque sheet.
Note 1 : if the film is to be handled in the darkroom for long periods (more than one minute), it should be
tested in 5 positions and left for 2 extra minutes in the last position.
Note 2 : if this is a periodic maintenance check, one exposure of the film (of 2 to 4 minutes according to
requirements) will be sufficient.
5.1.12 Develop the film, protecting it with the opaque sheet until it is inserted in the automatic processor.
5.1.14 Look for the first area on the film in which the frame line is visible (see illustration 2) and note down
the time corresponding to the area concerned on the film.
Note: the ”laboratory safety time”, defined as half the period of time corresponding to the first visible
area, should also be at least one minute.
3–214
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
5.1.16 Decisions
1. If the time corresponding to the first visible area is greater than 2 minutes, go on to the next pro-
cedure.
3. If the time is less than 1 minute, priority must be given to solving the problem to enable the tests
to continue.
ILLUSTRATION 2
JOB CARDS
2 minutes 4th area
Frame line
5.1.17 Use the densitometer to measure the relative optical density on both parts of the film.
CRITERION 1: Any difference between the relative optical densities measured must be less
than 0.1.
CRITERION 2: Neither optical density measured must exceed 0.2 (except for certain fast films).
3–215
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
5.1.18 If either or both of the above two criteria are not met:
1. Check the light–tightness of the darkroom by repeating sections 5.1.1 to 5.1.17 with the safe-
light turned OFF.
5.2.1 Perform a sensitometry test on the type of film being used by the customer.
5.2.4 Use the densitometer to find the first step whose relative optical density is greater than or equal to
1. Let N be the step found and D(N) its relative optical density.
5.2.5 Measure the relative optical density for step (N+2). Let this measured value be D(N+2).
Note: A gamma value less that 3.0 cannot produce a good image quality in mammography.
5.2.7 If the above criterion is not met, check the automatic film processor (temperature, cleanliness,
change in chemicals, etc.).
3–216
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
Note: This Job Card apply for systems equipped with new centering devices. For old cen-
tering devices, consult Job Card IST 035B.
SECTION 1
SUPPLIES
D New centering device lamp (only genuine part number 2188568 must be used, to ensure correct light
intensity, easy filament centering and reliability).
SECTION 2
JOB CARDS
TOOLS
D Standard screwdrivers, open–end wrenches, etc.
D Ruler calibrated in mm
SECTION 3
SAFETY PRECAUTIONS
None
SECTION 4
PREREQUISITES
None
SECTION 5
REPLACEMENT OF CENTERING DEVICE LAMP
The objective is to replace the centering device lamp and set it to the correct vertical height.
5.1.3 Remove the anti–scatter box (2 parts); first remove the side part then loosen slightly the attachment
screw (at center underside) of the box and remove the anti–scatter box by pulling it gently forward.
5.1.4 Remove collimator assembly retaining pins by first removing the 2 cotter pins from one end of each
retaining pin. Pivot the collimator assembly downwards.
3–217
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
5.1.5 Loosen slightly the two attachment screws on the light box and remove the light box by pulling it
gently forward.
5.1.6 Remove the old lamp by pulling firmly but gently downwards.
Note: When handling the new lamp, be sure not to touch the glass surface with your fingers. Always use
a clean dry cloth or paper handkerchief.
5.2.1 Insert the leads of the new lamp into its socket SLOWLY, and PUSH THEM IN ALL THE WAY.
JOB CARDS
Note: The distance of 30 mm between filament center and pin end is guaranteed by the lamp manufac-
turer, see Illustration 1.
5.2.2 Re–install the light box without tighten its attachment screews.
CAUTION The light box is for protection in case any explosion of the new lamp being
installed occurs.
5.2.3 Turn on the Senographe DMR and verify that the new lamp lights when activated, then turn the
Senographe DMR off again.
5.2.5 Check that the CENTER of the filament is aligned with the CENTER of the small slot in the light
box when your eye is at the level of the small slot (see illustration 1). If not, start 5.2 again.
5.2.6 Pivot the collimator assembly back up and re–install its retaining pins and cotter pins.
5.2.7 Proceed directly to Job Card IST 036 ”CENTERING DEVICE LAMP ADJUSTMENT”, leaving
anti–scatter box and gantry head half–covers removed.
3–218
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
ILLUSTRATION 1
CORRECT LAMP FILAMENT HEIGHT
VISUAL INSPECTION:
30 ± 0.5 mm CHECK THAT CENTER OF
FILAMENT IS ALIGNED
JOB CARDS
WITH CENTER OF LIGHT
BOX SLOT.
3–219
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
Blank page
JOB CARDS
3–220
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
SECTION 1
SUPPLIES
D New centering device lamp
SECTION 2
TOOLS
D Standard screwdrivers, open–end wrenches, etc.
D Ruler calibrated in mm
JOB CARDS
SECTION 3
SAFETY PRECAUTIONS
None
SECTION 4
PREREQUISITES
None
SECTION 5
REPLACEMENT OF CENTERING DEVICE LAMP
The objective is to replace the centering device lamp and set it to the correct vertical height.
5.1.3 Loosen slightly the attachment screw on the protection box and remove the protection box by pull-
ing it gently forward.
5.1.4 Remove collimator assembly retaining pin by first removing the cotter pin from one of its ends.
Pivot the collimator assembly downwards.
3–221
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
5.1.5 Loosen slightly the two attachment screws on the light box and remove the light box by pulling it
gently forward.
5.1.6 Remove the old lamp by pulling firmly but gently downwards.
Note: When handling the new lamp, be sure not to touch the glass surface with your fingers. Always use
a clean dry cloth or paper handkerchief.
5.2.1 Study the new lamp from a profile perspective (i.e. filament axis perpendicular to the eye) and
decide if the filament is centered (from front to back when mounted in the Senographe DMR) in
the glass envelope. If it is NOT centered, be sure to orient the new lamp WITH THE FILAMENT
TOWARDS THE FRONT OF THE SENOGRAPHE DMR (i.e. towards the patient), when instal-
ling it.
JOB CARDS
5.2.2 After orienting the new lamp (if necessary) as described above, insert the leads of the new lamp into
its socket SLIGHTLY, just enough to hold the lamp in place. DO NOT PUSH THEM IN ALL THE
WAY.
5.2.3 Turn on the Senographe DMR and verify that the new lamp lights when activated, then turn the
Senographe DMR off again.
5.2.4 Take the ruler and measure the vertical distance between the BOTTOM EDGE of the filament and
the edge of the lamp mounting frame as shown in illustration 1. The objective is to attain a distance
of 16 ± 0.5 mm.
5.2.5 If this distance is not in the range 16 ± 0.5 mm, push the new lamp further into its socket CARE-
FULLY until the correct distance is obtained.
DO NOT push the new lamp too far into its socket because it is very difficult
CAUTION to pull it out again without damaging it.
5.2.6 Re–install the light box and tighten its two attachment screws.
5.2.7 Check that the BOTTOM EDGE of the filament is aligned with the bottom edge of the small slot
in the light box when your eye is at the level of the small slot (see illustration 1). If this is not the
case, remove the light box and repeat steps 5.2.4 to 5.2.5.
5.2.8 Pivot the collimator assembly back up and re–install its retaining pin and cotter pin.
5.2.9 Proceed directly to Job Card IST 036 ”CENTERING DEVICE LAMP ADJUSTMENT”, leaving
protection box and gantry head half–covers removed.
3–222
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
ILLUSTRATION 1
CORRECT LAMP FILAMENT HEIGHT
VISUAL INSPECTION:
CHECK THAT BOTTOM OF
FILAMENT IS ALIGNED
WITH BOTTOM OF LIGHT
JOB CARDS
16 ± 0.5 mm BOX SLOT.
3–223
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
Blank page
JOB CARDS
3–224
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
SECTION 1
SUPPLIES
D ”Ready Pack” or ”cardboard cassette” films (24 x 30 format)
SECTION 2
TOOLS
D Standard screwdrivers, open–end wrenches, etc.
D Ruler calibrated in mm
JOB CARDS
SECTION 3
SAFETY PRECAUTIONS
SECTION 4
PREREQUISITES
D Job Card IST 035 “REPLACEMENT OF CENTERING DEVICE LAMP” (protection box and gantry
head half–covers are still removed) or IST 035B “REPLACEMENT OF CENTERING DEVICE LAMP
(OLD VERSION)”
SECTION 5
ADJUSTMENT OF CENTERING DEVICE LAMP
The objective is to adjust the position of the centering device lamp (in all three dimensions, if necessary) so
that the lit field coincides, within acceptable tolerances, with the exposed X–ray field. This is done by first
exposing a film and then illuminating it on the image receptor support to check the coincidence of the lit field
with the exposed area on the X–ray film. The following properties of this coincidence are measured and
adjusted, if necessary, in the following order:
Lateral centering
Field width coincidence
Field depth coincidence
Field front light alignment
Note: Field width, depth and front can be fine–tuned using procedure detailed in IST 028.
3–225
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
Select SETUP/COVERAGE
JOB CARDS
Use + & – keys or rotate console kV button to return the displayed ”COVERAGE” value to zero.
5.1.3 Take a new cardboard cassette or Ready Pack and insert three thumb tacks through it, FROM THE
SIDE THAT WILL FACE THE X–RAY TUBE, in the following way:
– Two thumb tacks inserted along the (long) edge of the film that will be at the front edge
of the image receptor support, at approximately 2 cm in from this edge and approxi-
mately 4 cm inwards from the two shorter edges
– The third thumb tack along the (short) edge of the film that will be on the left side of
the image receptor support, approximately 2 cm in from this edge and approximately
12 cm from the rear edge of the film.
Refer to illustration 1 below.
3–226
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
ILLUSTRATION 1
PLACEMENT OF THREE THUMB TACKS TO BE INSERTED IN THE READY PACK
APPROX 12 CM
READY PACK
TUBE SIDE
JOB CARDS
APPROX 2 CM
APPROX 4 CM
APPROX 4 CM
APPROX 2 CM
APPROX 2 CM
Note: It is NOT important to insert the three thumb tacks in a perfectly ”squared” alignment with the
edges of the Ready Pack.
5.1.4 Place the Ready Pack on the image receptor support so that the points of the two ”front” thumb tacks
butt perfectly up against the front edge of the image receptor support and the third thumb tack butts
perfectly up against the left edge of the image receptor support. THE POSITIONING OF THE
THREE THUMB TACKS AGAINST THE IMAGE RECEPTOR SUPPORT IS VERY IMPOR-
TANT TO THE SUCCESS OF THESE CHECKS.
3–227
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
5.2.1 Using normal application mode, set up a manual (2–point) exposure as follows:
– 18 x 24 format (collimator blades fully open)
– Mo track
– Mo filter
– Large focal spot
– 25 kV
– 50 mAs
Note: It is advisable to turn off all room lighting before proceeding to the following steps.
5.3.1 Replace the three thumb tacks in the holes in the developed film.
5.3.2 Place the developed film back onto the image receptor support, with the same orientation as during
the exposure. Carefully align the thumb tacks against the edges of the image receptor support,
exactly as during the exposure.
5.3.3 Keeping the same format size as during the exposure, turn on the light centering device lamp.
5.3.4 Study the relationship between the lit zone from the light centering device and the darkened area
on the film.
5.3.5 Determine if the two areas are centered laterally (illustration 4 shows correct lateral centering).
Note: The lit zone may be wider or narrower than the darkened area on the film at this point.
Illustration 4 shows the lit zone wider than the darkened area on the film.
5.3.6 If the two areas are not centered laterally, turn adjustment screw 1 (see illustration 3) until the lit
zone is centered laterally with the darkened area on the film.
5.4.1 Keep the film aligned on the image receptor support as above.
5.4.2 If the lit zone is wider (narrower) than the darkened area on the film, reduce (increase) the width
of the lit zone by turning knurled nut 2 in the positive (negative) direction (see illustration 3).
Illustration 5 shows correct field width coincidence.
3–228
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
5.5.1 Keep the film aligned on the image receptor support as above.
5.5.2 If the rear edge of the lit zone is not aligned with the rear edge of the darkened area on the film, turn
adjustment screw 3 (see illustration 3) until alignment is achieved. Illustration 6 shows correct field
depth coincidence.
Note: The only mecanical adjustment for alignment of the front edge of the let zone is by adjusting the
mirror.
D Adjust the angle of the mirror with the angle adjustment screw (see illustration 2 ) in order to align the
front blade with the darkered area on the film.
Tilt the mirror slightly otherwise a major adjustement may distort the light
JOB CARDS
CAUTION field.
ILLUSTRATION 2
TUBE HOUSING ARM WITH MIRROR INSTALLATION DETAIL
MIRROR FRAME
TUBE HOUSING ARM
3–229
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
ÊÊÊ
LIGHT BOX
ÊÊÊ
ÊÊÊ
JOB CARDS
1
(LATERAL CENTERING ADJUSTMENT)
3
(REAR BLADE ADJUSTMENT)
3–230
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
ILLUSTRATION 4
CORRECT LATERAL ALIGNMENT OF ZONE LIT BY LIGHT CENTERING DEVICE WITH EXPOSED AREA OF X–RAY FILM
DEVELOPED READY PACK FILM IMAGE RECEPTOR SUPPORT ZONE LIT BY LIGHT
CENTERING DEVICE
JOB CARDS
LATERAL CENTERING IS
ACHIEVED WHEN THESE
TWO DISTANCES ARE EQUAL
3–231
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
ILLUSTRATION 5
WIDTH COINCIDENCE BETWEEN ZONE LIT BY LIGHT CENTERING DEVICE AND EXPOSED AREA OF X–RAY FILM
DEVELOPED READY PACK FILM IMAGE RECEPTOR SUPPORT ZONE LIT BY LIGHT
CENTERING DEVICE
JOB CARDS
3–232
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
ILLUSTRATION 6
DEPTH COINCIDENCE BETWEEN ZONE LIT BY LIGHT CENTERING DEVICE AND EXPOSED AREA OF X–RAY FILM
DEVELOPED READY PACK FILM IMAGE RECEPTOR SUPPORT ZONE LIT BY LIGHT
CENTERING DEVICE
JOB CARDS
DEPTH COINCIDENCE IS ACHIEVED
WHEN REAR EDGE OF LIT AREA IS
ALIGNED WITH REAR EDGE OF
DARKENED AREA ON FILM
3–233
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
Blank page
JOB CARDS
3–234
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
SECTION 1
SUPPLIES
None
SECTION 2
TOOLS
None
JOB CARDS
SECTION 3
SAFETY PRECAUTIONS
None
SECTION 4
PREREQUISITES
None
SECTION 5
JUMPERS AND SWITCHES
The objective is to verify the position of jumpers and switches. Jumpers and switches are divided into three
groups: generator, gantry and console.
5.1.1 Check that the jumper on generator CPU board 400–PL3 is installed in position X2.
See illustration 1.
This jumper must NEVER be removed from position X2 and must NEVER
CAUTION be placed in position TP1. If you find this jumper either removed or placed
in position TP1, the generator CPU RAMs are totally erased, necessitating
a TOTAL RE–CALIBRATION of the generator and AEC.
3–235
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
5.2.1 Check that the jumper on gantry CPU board 800–PL3 is installed in position X2. See illustration
1.
This jumper must NEVER be removed from position X2 and must NEVER
CAUTION be placed in position TP1. If you find this jumper either removed or placed
in position TP1, the gantry CPU RAMs are totally erased, necessitating a
TOTAL RE–CALIBRATION of the gantry.
ILLUSTRATION 1
POSITION OF JUMPER ON GENERATOR CPU BOARD 400–PL3 AND GANTRY CPU BOARD 800–PL3
GENE CPU BOARD 400–PL3 or
GANTRY CPU BOARD 800–PL3
!
JOB CARDS
B46
B41
X2
BT1
!
P1 P2
5.3.1 Check that console self–test activation switch 100 S1 is in its ”application position”. See
illustration 2.
Note: When this switch is in its ”unused” position, the console display stays blank and ”beeps” at
1–second intervals after power on is pressed.
ILLUSTRATION 2
CONSOLE SELF–TEST ACTIVATION SWITCH 100 S1
SENOGRAPHE DMR CONSOLE SIDE VIEW
CONSOLE AUTOTEST POSITION
100PL2 XJ3
APPLICATION POSITION UNUSED POSITION
100PL2 XJ5
100 S1
CONSOLE SELF–TEST ACTIVATION
SWITCH 100 S1 (LOCATED NEXT TO
CONSOLE CABLE CONNECTOR 100 XJ3)
3–236
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
SECTION 1
SUPPLIES
None
SECTION 2
TOOLS
None
JOB CARDS
SECTION 3
SAFETY PRECAUTIONS
SECTION 4
PREREQUISITES
None
SECTION 5
SAFETY LABELLING
The objective is to select and attach appropriate safety labels to the Senographe DMR, according to local safety
regulations concerning the use of X–ray producing equipment.
3–237
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
SAFETY LABELLING
Job Card IST 038 2 of 4
5.1 Locate the envelop (which is taped underneath the console) containing the 3 possible safety labels and select the
one that corresponds to local regulations (see illustration 1 below).
ILLUSTRATION 1
SAFETY LABELS
REDAC (CANADA)
JOB CARDS
CDRH (USA)
Note: The CDRH (USA) safety label is already affixed on the console.
3–238
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
SAFETY LABELLING
Job Card IST 038 3 of 4
5.2 Peel off the self–adhesive backing from the selected safety label, and stick the label over the CDRH (USA) label
in the lower right–hand corner of the console (see illustration 2 below).
ILLUSTRATION 2
PLACEMENT OF SAFETY LABEL ON CONSOLE
JOB CARDS
STICK LABEL HERE
3–239
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
SAFETY LABELLING
Job Card IST 038 4 of 4
Blank page
JOB CARDS
3–240
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
SECTION 1
SUPPLIES
D Normal cassette of type used in the screen pair being calibrated
D Film of type used in the screen pair being calibrated
D Plexiglass in thickness increments of 0.5 cm (minimum plexiglass dimensions 20 x 20 cm to insure com-
plete covering of the photo cell)
JOB CARDS
SECTION 2
TOOLS
None
SECTION 3
SAFETY PRECAUTIONS
SECTION 4
PREREQUISITES
Generator calibration must be successfully completed for the screen pair in question before this Job Card is
performed.
Note: The AOP strategy parameter is determined automatically at the end of IST013 or IST014 refer-
ence energy calibration. So the manual setting of the AOP strategy (IST039) is no longer neces-
sary.
However, if there are several senographe DMR installed on the same site, it may be necessary to
“equalize” all the strategies to the same value, to avoid unjustified mAs deviation between identi-
cal senographes. For this purpose only, it is still possible to read/set the strategy value.
This equalization must only be done if the DMRs are using the same film/screen couple, same film
processor, same SID.
3–241
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
SECTION 5
DETERMINE AOP STRATEGY
This Job Card provides the procedure for determining the AOP strategy parameter for a given screen pair. This
is done by taking a reference X–ray exposure in 1–point mode and choosing the AOP strategy parameter
according to the resulting mAs from the exposure.
GENERAL INSTRUCTIONS TO BE OBSERVED DURING ALL PROCEDURES GIVEN BELOW IN
THIS SECTION:
D The photo cell must be in its most forward position (towards the patient) throughout the procedures in
this Job Card.
D When placing plexiglass on the cassette holder, the plexiglass must always overlap the front edge
(towards the patient) by about 1 cm to insure that the photo cell is fully covered.
5.1.2 Load cassette with undeveloped film into cassette holder and take an exposure. Note down the
resulting mAs.
5.1.3 Determine AOP strategy parameter value according to the resulting mAs value:
mAs < 30 => strategy = SUPER FAST
30 < mAs < 50 => strategy = VERY FAST
50 < mAs < 70 => strategy = FAST
70 < mAs < 100 => strategy = MEDIUM
100 < mAs < 150 => strategy = SLOW
mAs > 150 => strategy = VERY SLOW
3–242
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
5.2 Enter the AOP strategy parameter value into the Senographe DMR.
5.2.1 Starting from application mode, change the position of the installation menu enable switch (switch
B1 on generator CPU board 400–PL3, see ”Accessing the Generator or Gantry Installation Mode
from the Console” in chapter 1).
Note: This must only be done if the Senographe DMR has been switched off since the last access to SET-
UP/GENE/INSTAL or SETUP/ARM/VALID/INSTAL
5.2.3 Select the AOP strategy as determined above (SUPER FAST, VERY FAST, FAST, MEDIUM,
SLOW, or VERY SLOW) on the console.
JOB CARDS
5.2.4 Perform a CKSUM and return to application mode.
5.2.5 If you are turning the machine back over to the user at this point, switch the Senographe DMR off,
then on again.
2) Or by reducing the dynamic range of the image by increasing the kV value. It is obtained if you modify the
strategy of film exposure; reducing its speed corresponds to an increase of the kV value:
– Access to menu /AOP/FSCx/ and set strategy to the immediat lower speed.
Changing the strategy will reduce image quality. It may be done only with the
CAUTION physician agreement.
3–243
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
3–244
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
Time: Personnel: 1
SECTION 1
SUPPLIES
None.
SECTION 2
TOOLS
D Screwdriver 4 mm.
D Allen wrench 2.5 mm.
JOB CARDS
SECTION 3
SAFETY PRECAUTIONS
D No specific safety precautions are applicable.
SECTION 4
PREREQUISITES
None.
SECTION 5
PROCEDURE
D Measure room overall height, floor to ceiling.
D Remove left cover from column housing (see IST032).
Note: The Senographe DMR is delivered with the optical sensor is installed in this position.
2. Ceiling height 2.4m (7.9ft.) or greater, but less than 2.5m (8.22ft.) :
3–245
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
3. Ceiling height 2.3m (7.5ft.) or greater, but less than 2.4m (7.9ft.) :
– Install the optical sensor in the lower mounting hole (3 in Illust. 1).
Note: The adjustment of the metal plate which blanks off the upper and lower sensors only affects the
elevator lower travel limit.
3. Remove the optical sensor and reinstall it in its new position, as determined in 5.1, above.
4. Reinstall and tighten the two 2.5 mm Allen screws, in the new position.
6. Switch on the Senographe DMR, carry out a column raising and lowering test and check that the tube hous-
ing cover does not collide with the ceiling (check minimum clearance for safety of at least 8 cm to 9 cm).
3–246
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
ILLUSTRATION 1
SENOGRAPHE DMR, SHOWING OPTICAL SENSOR POSITIONS
JOB CARDS
1
1
POSITIONS FOR 2
2 UPPER OPTICAL
SENSOR
3
3
3–247
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
Blank page
JOB CARDS
3–248
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
SECTION 1
SUPPLIES
None.
SECTION 2
TOOLS
None.
JOB CARDS
SECTION 3
SAFETY PRECAUTIONS
None.
SECTION 4
PREREQUISITES
None.
3–249
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
SECTION 5
GENERAL ERRORS DURING THE CALIBRATION
3–250
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
JOB CARDS
graph 6–4–1, copy succes-
sively from C (no screen)
to A (no screen)
and C (with screen)
to A (with screen).
3–251
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
NO ARM REPLY Gantry failure (arm) Check the gantry and its
connection with the genera-
tor (optical fibers)
ACCESS REFUSED The command has not been Software error ––> inform
taken into account the Central Support
CONNECT ERROR Unknown answer form the Software error ––> inform
JOB CARDS
3–252
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
JOB CARDS
check the photocell board
800–PL4. Change photocell
if necessary.
IST 005 NOT ENOUGH PLEXI PM current is too high Check the thickness of the
PHOTOCELL plexi.
CALIBRATION Increase the thickness of the
plexi according to the
Job Card and do the calibra-
tion again.
This problem may come
from the hardware ––>
check the photocell board
800–PL4. Change photocell
if necessary.
OUT OF ORDER Failure Go back to application mode
and note (fix) the error code.
ARM ABSENT Default generator configura- Select ”arm present” in the
tion CONFIG menu of the gener-
ator
3–253
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
3–254
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
JOB CARDS
CALCUL. ERROR Calculation error of the non– Redo the calibration with the
reciprocity coefficients root parameters
(A0, A1, A2) indicated in
the Job Card
CHECK PLEXI CM If the measured thickness is Check the thichness of the
too much different from the plexi.
plexi thickness required Ignore the message if the
(+/– 0.5 cm) measure is right
IST 014 OD_MS = TOO LOW The mAs used for the expo- Change the mAs value or
REFERENCE OD_MS = TOO HIGH sure are too low or too hight. enter the right value of the
ENERGY This may also be an error in optical density
CALIBRATION the measurement of the opti-
cal density.
ARCING STOP Too many arcings : No specific action : try to
calibration stopped make the exposure again
IST 015 REDUCING HT Information message : if the Press the exposure button to
MAGNIFICATION RISING HT exposure button has been continue the calibration
CORRECTION relased during the calibration
IST 016 OUT OF RANGE The entered mA value is out Do the calibration again or
mA CALIBRATION of acceptable range. check the Hardware
(40 mA mini – 60 mA maxi)
3–255
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
3–256
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
SECTION 1
SUPPLIES
None.
SECTION 2
TOOLS
None.
JOB CARDS
SECTION 3
SPECIAL SAFETY PRECAUTIONS
None.
SECTION 4
PREREQUISITES
None.
SECTION 5
PROCEDURE
Use this form to note down all the data collected during the DMR installation. This form has sufficient room
for the calibration of 2 screen / film couples. You can also add your personnal notes and remarks at the end
of this form, to help you during future maintenances.
3–257
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
3–258
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
JOB CARDS
3–259
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
3–260
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
JOB CARDS
Delta (X cm) = Radiological Thickness – Plexiglass Thickness
Expected Results:
Each Radiological thickness must be equal to the Plexiglass thickness +/– 2 mm
Delta (4cm) must be greater than Delta (2cm) and Delta (6cm)
absolute value (Delta (2cm) – Delta (6cm)) < 1.5 mm.
3–261
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
32 / 1,5 cm
28 / 5.5 cm
3–262
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
JOB CARDS
32 / 1,5 cm
28 / 5.5 cm
3–263
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
3–264
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
JOB CARDS
3–265
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
3–266
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
JOB CARDS
32 / 1,5 cm
28 / 5.5 cm
3–267
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
32 / 1,5 cm
JOB CARDS
28 / 5.5 cm
3–268
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
JOB CARDS
3–269
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
Blank page
JOB CARDS
3–270
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
SECTION 1
SUPPLIES
None.
SECTION 2
TOOLS
None.
JOB CARDS
SECTION 3
SPECIAL SAFETY PRECAUTIONS
None.
SECTION 4
PREREQUISITES
None.
SECTION 5
PADDLE DEFORMATION PARAMETERS
The objective is to check the values of the paddle deformation parameters and to modify them if necessary.
These parameters alter the X–Ray exposure behaviour in AOP mode.
5.1.2 Rotate the KV button to see either the offset or the gain values. Check agains table 1 below. change
the value(s) if necessary.
3–271
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
5.2.2 Rotate the KV button to see either the offset or the gain values. Check against table 2 below. Change
the value(s) if necessary.
5.3.2 Rotate the KV button to see either the offset or the gain values. Check against table 3 below. Change
the value(s) if necessary.
5.4.2 Rotate the kV button to see either the offset or the gain values. Check against table 4 below.
Changes the value(s) if necessary.
3–272
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
SECTION 1
SUPPLIES
None.
SECTION 2
TOOLS
None.
JOB CARDS
SECTION 3
SPECIAL SAFETY PRECAUTIONS
None.
SECTION 4
PREREQUISITES
None.
3–273
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
PREVENTIVE MAINTENANCE
FOLLOW–UP Job Card IST 045 2 of 4
SECTION 5
PREVENTIVE MAINTENANCE
Despite its inherent reliability, the Senographe DMR requires certain minimum maintenance. A Maintenance
Program, whose frequency may vary according to the type of use of the equipment, is outlined in Table 1.
Use the Maintenance Follow–up Table to track maintenance operations over time.
Frequency
Action Description
Procedure
Systematic
8 3 years 3 years - Replace batteries in the two microprocessor boards.
replacement
Systematic
- Replace the X1 relay (91682082) on the 200 PL1 Mains
9 replacement in 3 years 3 years
Distribution Board.
Generator Cabinet
3–274
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
PREVENTIVE MAINTENANCE
FOLLOW–UP Job Card IST 045 3 of 4
JOB CARDS
Compression ... ... ... ... ...
2
system
3–275
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
PREVENTIVE MAINTENANCE
FOLLOW–UP Job Card IST 045 4 of 4
1 Elevator
3–276
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
SECTION 1
SUPPLIES
D Normal cassette of type used in the screen pair being calibrated
D Film of type used in the screen pair being calibrated
SECTION 2
TOOLS
D Plexiglass in thickness increments of 10 mm (minimum plexiglass dimensions 200 x 200 mm to ensure
JOB CARDS
complete coverage of the photo-cell).
Tool reference (plexi plates) : 45 203 014 or 46–286893 (US pole).
Thickness accuracy of plexiglass plates is mandatory (10 0.2 mm).
For memo: chemical formula of plexiglass is C5H8O2, different from Lexan and Nylon!
SECTION 3
SAFETY PRECAUTIONS
SECTION 4
PREREQUISITES
D About checksum, see chapter 1, para. 3–7.
D kV accuracy check (if applicable) followed by kVp recalibration must be performed before this job
card (see IST 033, para. 5.1.11).
D Job Card IST 006 SCREEN PAIR SELECTION AND CONFIGURATION
D Job Card IST 007 must have been performed recently.
D In case of error messages during this calibration, see IST 041 for explanations.
3–277
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
CHECKS ON PRE-EXPOSURE
Job Card IST 046 2 of 6
SECTION 5
CHECKS ON PRE–EXPOSURE IN AOP MODE
START
Make 5 exposures
AEC 28kV
No
Result analysis good
Yes
Final adjustment
Turn P1
END
3–278
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
CHECKS ON PRE-EXPOSURE
Job Card IST 046 3 of 6
JOB CARDS
So we need to perform at least 5 pre–shot to get a sufficient accuracy; (du to the variability in pre–shot length)
3–279
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
CHECKS ON PRE-EXPOSURE
Job Card IST 046 4 of 6
2. Press 2nd trigger to take pre–exposure ONLY ; RELEASE 2nd trigger and Prep buttons BE-
FORE actual exposure starts. Clear the Abort sound.
4. Record this value in Table 1, value THp1 (which stands for Thickness, pre–exposure N°1 )
Note: Be sure to record all digits on the console (ex: 4.17 cm)
5. Redo the same exposure, but this time, let the system finish the main exposure.
7. Record this value in Table 1, value THe1 (which stands for Thickness, exposure N°1 ),
8. Repeat steps ”1” to ”7” FOUR TIMES and each time report values in Table 1 (THp2 to THp5
and THe2 to THe5). Observe carefully a waiting time between exposures.
10. Report Values AV_THp , AV_THe and AV_ ∆ Table 2. (AV stands for Average)
Table 1 : Pre–exposure test
(note Thickness values in cm)
3–280
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
CHECKS ON PRE-EXPOSURE
Job Card IST 046 5 of 6
JOB CARDS
Note: P1 is the frequency compensation adjustment of the DMR HV probe (HV connection +
wire…) if you turn P1 clockwise you go to overshoot, if you turn P1 to reverse clockwise you
go to undershoot.
Examples of adjustment for 701PL1 –P1:
First example :
In this case, turn P1 by 3 turns clockwise: this will increase THp by 0.03cm approximately.
Refine if necessary.
Second example :
In this case, when average delta is equals or smaller than –0.06, turn P1 with a maximum of 6 turns re-
verse clockwise by step: this will reduce THp by 0.06cm approximately; average delta is not linear.
Refine if necessary.
3–281
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
CHECKS ON PRE-EXPOSURE
Job Card IST 046 6 of 6
3–282
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
SECTION 1
SUPPLIES
None
SECTION 2
TOOLS
D Luxmeter (Range 1 to 10 000lux) suitable for photometric measurements.
JOB CARDS
D Luminancemeter (range 10 to 10 000 cd/m2 or nit) suitable for photometric measurements.
SECTION 3
SAFETY PRECAUTIONS
None
SECTION 4
PREREQUISITES
None.
3–283
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
CHECK OF VIEWBOX(ES)
Job Card IST 047 2 of 4
SECTION 5
PROCEDURE
• Position the luxmeter photocell at a distance of 30 cm from the viewbox, in direction opposite to the
latter (to measure the light received on viewbox surface),
• Record the Lux value (on datasheet) and proceed the same way for other viewboxes.
Pass criterion: Each measured Lux value of ambient light must be < 50 Lux.
• Five test points are identified in each diagnostic area (one test point at the center, and four test points at
the corner of an area of 35 X 43).
C1 C2 C3 C4
Viewbox
5 test points in
each area
5.2.2 Procedure
• Switch on the viewbox to be tested and wait for 5 minutes,
3–284
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
CHECK OF VIEWBOX(ES)
Job Card IST 047 3 of 4
• Position the luminancemeter towards the five test points of a diagnostic area and record the correspond-
ing luminance values (in cd/m2) values on datasheet,
JOB CARDS
3–285
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
CHECK OF VIEWBOX(ES)
Job Card IST 047 4 of 4
Blank page
JOB CARDS
3–286
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
SECTION 1
SUPPLIES
None
SECTION 2
TOOLS
D Densitometer, Sensitometer, Marker pen, Gloves, Film, Test Cassette.
JOB CARDS
SECTION 3
SAFETY PRECAUTIONS
SECTION 4
PREREQUISITES
3–287
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
SECTION 5
PROCEDURE
5.1.1 Purpose:
• The purpose is to check the compliance of system condition with respect to a rated and/or reference condition,
i.e. gamma as a function of chemical/Film/Film processor parameters.
• It could help to find out the optimum O.D. which allows a good completion of IST 013.
5.1.2 Procedure:
a) expose the film with the sensitometer set to green, with the film right way up (if single emulsion is used)
• For most of the films, the emulsion is facing down when the notch is on the upper left.
JOB CARDS
• For an X–RITE 333 sensitometer, the emulsion must be facing down with the film up against the stop.
b) Develop the film and write in all data necessary for its identification: site, date and time,
Note: Calibration of LNRT and Reference Energy are done using net Optical Density values (without base +
fog).
d) Measure and record O.D. levels for each step ( 1 through 21) in Table 3–1, then draw the sensitometric curve,
e) Note the first step whose relative density is greater or equal to 1 (Sensitivity Index). Let N be the step found and
D(N) its relative optical density, and record it in Table 3–2.
f) Note the relative optical density for step (N + 2). Let this measured value be D(N + 2).
g) Calibrate the densitometer to zero without film, then measure the raw O.D. for step 1 (Base + Fog density). Re-
cord it in Table 3–2.
3–288
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
JOB CARDS
Maximum Density = Density of step 21
Table 3–2: Sensitometric calculations
Note: The Gamma value calculated by the DMR must be close to the Gamma value calculated manually to
get good results about optical density.
3–289
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
3–290
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
SECTION 1
PURPOSE
This procedure applies to the DMR mammographic equipment line.
The standard adjustment made using Job cards IST007 and IST013 guarantees an acceptable Optical Density
variation for each Track/Filter combination.
The two purposes of this additional Job Card IST050 are the following:
1) to check the Optical Density variation for a plexiglass thickness varying from 2 to 6 cm,
2) to reduce the Optical Density variations when switching from one combination to another among the three
JOB CARDS
possible (Mo/Mo, Mo/Rh and Rh/Rh), and allows to be compliant with the specifications of MQSA 2002 stan-
dards (US Regulation). This is achieved by optimizing the values of the coefficients A0 without screen and
with grid coming from IST007 calibration, for Mo/Rh and Rh/Rh Track/Filter combinations.
SECTION 2
WHERE/WHEN
1) For already installed systems:
Where
When
Where
When
This procedure should be performed once for each Film/Screen combination during the first installation of the
system (refer to the “flowchart of AEC/AOP calibration”).
It must be repeated each time Job Card IST007, IST013 or IST014 is redone, for example when the user intro-
duces a new Film /Screen combination
3–291
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
SECTION 3
PROCEDURE
Copy the 2283482–200_IST050_DMR_A0_OPTIMIZATION.xls file from the CD 2283482–200 onto
your PC hard disk, then run the file from the hard disk. Follow instruction given in the .xls file
Note: If you are not provided with Excel*, all necessary formulas are included in the
2283482–200_IST050_DMR_A0_OPTIMIZATION.pdf file.
3–292
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
SECTION 1
PURPOSE
This procedure applies to the DMR mammographic equipment line.
The standard adjustment made using Job cards IST007, IST013 and IST046 guarantees an acceptable Optical
Density variation for each Track/Filter combination.
The goal of this additional Job card IST051 is to reduce the Optical Density variations when switching from
one combination to another among the three possible (Mo/Mo, Mo/Rh and Rh/Rh) inside the magnification
mode, and allows to be compliant with the countries regulations (e.g.: MQSA 2002).
JOB CARDS
This procedure does not reduce the shift between the contact mode and the magnification mode.
This is achieved by optimizing the values of the coefficients A0 without screen and without grid coming from
IST007 calibration, for Mo/Rh and Rh/Rh Track/Filter combination, the Mo/Mo combination being taken as
a reference.
SECTION 2
WHERE/WHEN SHOULD THIS JOB CARD BE PERFORMED?
Where
If you need to reduce the optical density variation between Mo/Mo, Mo/Rh and Rh//Rh inside the
magnification mode.
When
It must be repeated each time Job Card IST007, IST013 or IST014 is/are redone, for example when the user
introduces a new Film/Screen combination.
3–293
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
SECTION 3
PROCEDURE
Copy the 2283482–200_IST051_DMR_A0_OPTIMIZATION.xls file from the CD 2283482–200 onto your PC
hard disk, then run the file from the hard disk. Follow instruction given in the .xls file
Note: If you are not provided with Excel*, all necessary formulas are included in the
2283482–200_IST051_DMR_A0_OPTIMIZATION.pdf file.
3–294
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
SECTION 1
ERROR CODE STRUCTURE
The structure of the error messages, as displayed on the DMR console (e.g. E55 [text]
161/013), is shown below:
Error
E 5 5 text 1 6 1 0 1 3
a bb ccc ddd
I : Warning without exposure inhibition
S : Warning with exposure inhibition
a : type of error
E : Fault with exposure inhibition
K : Fault generating power cut–off (safety)
bb: message number
ccc: subassembly ref. (hardware or software)
ddd: number indicating origin of fault
ERROR CODES
inhibition.
161/013 indicates no reference crossing during movement.
E55 is the error code type listed in the Operator Manual. This code is easy to use and
communicate, and is considered sufficient for customer use.
161/013 is the ”technical code”, recommended for GE internal communication.
The error code listing currently contains almost 500 codes. Around 200 of these codes
concern software errors which, by definition, should never occur.
If a software error should occur, especially if it occurs several times, please inform Service
Engineering, giving as many details as possible.
The error code listing, below, only contains codes ccc/ddd, to facilitate technical
communication.
Note: *** : these 3 stars indicates an explanation at the end of this chapter (section 3) for a
possible known root cause.
4–1
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
Abbreviations used:
IST: Installation
MNT: Maintenance
PM: Photomultiplier
HTPM: PM high tension
IPM: PM current
VRTX: Real time operating system
Where an error message refers to a signal with a ”French signal name”, the corresponding
English language signal name is shown between parentheses.
ERROR CODES
4–2
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
SECTION 2
ERROR CODE LIST
ERROR CODES
Gene interface: read/write error on timer PTM1 (see Chap. 1 Sect. 3) 004/002
Gene interface: read/write error on timer PTM1 (see Chap. 1 Sect. 3) 004/003
Gene interface: read/write error on timer PTM2 (see Chap. 1 Sect. 3) 004/004
Gene interface: read/write error on timer PTM2 (see Chap. 1 Sect. 3) 004/005
Gene interface: interrupt error on DUART1 (see Chap. 1 Sect. 3) 004/006
Gene interface: PTM1 timer 1 defective (see Chap. 1 Sect. 3) 004/007
Gene interface: PTM1 timer 2 defective (see Chap. 1 Sect. 3) 004/008
Gene interface: PTM1 timer 3 defective (see Chap. 1 Sect. 3) 004/009
Gene interface: PTM2 timer 1 defective (see Chap. 1 Sect. 3) 004/010
Gene interface: PTM2 timer 2 defective (see Chap. 1 Sect. 3) 004/011
Gene interface: PTM2 timer 3 defective (see Chap. 1 Sect. 3) 004/012
Gene interface: interrupt error on PTM1 timer 1 (see Chap. 1 Sect. 3) 004/013
Gene interface: interrupt error on PTM1 timer 2 (see Chap. 1 Sect. 3) 004/014
Gene interface: interrupt error on PTM1 timer 3 (see Chap. 1 Sect. 3) 004/015
Gene interface: interrupt error on PTM2 timer 1 (see Chap. 1 Sect. 3) 004/016
Gene interface: interrupt error on PTM2 timer 2 (see Chap. 1 Sect. 3) 004/017
Gene interface: interrupt error on PTM2 timer 3 (see Chap. 1 Sect. 3) 004/018
4–3
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
4–4
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
Heater software error: Heater scaling parameters (ICH) or heater values are not valid 021/003
Heater track 1: signal Pres_Chauf (Heater_Presence) ’high’ in standby status 022/001
Heater track 1: signal Pres_Chauf (Heater_Presence) stays ’low’ when heater started 022/002
Heater track 1: signal Pres_Chauf (Heater_Presence) goes ’low’ during pre–heater phase 022/003
Heater track 1: signal Pres_Chauf (Heater_Presence) goes ’low’ during 1st trigger 022/004
Heater track 1: signal Pres_Chauf (Heater_Presence) goes ’low’ during exposure 022/005
Heater track 1: mA value not significant 022/006
Heater track 1: signal Pres_Chauf (Heater_Presence) goes ’low’ after exposure 022/007
Heater track 1: signal Pres_Chauf (Heater_Presence) stays ’high’ when heater stopped 022/008
Heater track 2: signal Pres_Chauf (Heater_Presence) ’high’ in standby status 023/001
Heater track 2: signal Pres_Chauf (Heater_Presence) stays ’low’ when heater started 023/002
Heater track 2: signal Pres_Chauf (Heater_Presence) goes ’low’ during pre–heater phase 023/003
Heater track 2: signal Pres_Chauf (Heater_Presence) goes ’low’ during 1st trigger 023/004
Heater track 2: signal Pres_Chauf (Heater_Presence) goes ’low’ during exposure 023/005
Heater track 2: mA value not significant 023/006
Heater track 2: signal Pres_Chauf (Heater_Presence) ’low’ after exposure 023/007
Heater track 2: signal Pres_Chauf (Heater_Presence) stays ’high’ when heater stopped 023/008
ERROR CODES
Grid sync software error: ACQ_NTR or ACQ_OVR received or unknown reply when 026/001
starting
Grid sync software error: ACQ_NTR or ACQ_OVR received or unknown reply when 026/002
stopping
Grid sync software error: unknown command 026/003
Grid sync software error: unknown task status 026/004
Grid sync: signal Synchro_Bras (Synchro_Arm) ’high’ in standby status 027/001
Grid sync: grid start signal not received or grid fault detected 027/002
Grid sync: signal Synchro_Bras (Synchro_Arm) ’low’ after grid start 027/003
Grid sync: grid stop signal not received or grid fault detected 027/004
Arm fatal failure: acknowledgment of the Grid Stop Motion command sent by the 027/005
generator to the gantry not received by the generator
Arm fatal failure: Grid Stop Power command not received on time from the exposure 027/006
automaton (time out on the reception 15s)
Generator: delay task software error: unknown command 031/001
Generator: delay task software error: unknown task status 031/002
HTPM/IPM software error: ACQ_NTR or ACQ_OVR received or unknown reply 036/001
HTPM/IPM software error: ACQ_NTR or ACQ_OVR received or unknown reply 036/002
HTPM/IPM software error: ACQ_NTR or ACQ_OVR received or unknown reply 036/003
4–5
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
4–6
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
ERROR CODES
Gene software error (IST, standby): bias calibration task problem 073/005
Gene software error (IST, standby): heater scale calibration task problem 073/006
Gene software error (IST, standby): HV/DAC calibration task problem 073/007
Gene software error (IST, standby): FREQ/HV calibration task problem 073/008
Gene software error (IST, standby): PM calibration task problem 073/009
Gene software error (IST, standby): PM calibration task problem 073/010
Gene software error (IST, standby): PM calibration task problem 073/011
Gene software error (IST, standby): PM calibration task problem 073/012
Gene software error (IST, standby): gantry connection task problem 073/013
Gene software error (IST, standby): gantry connection fault 073/014
Gene software error (IST, standby): gantry connection fault 073/015
Gene software error (IST, standby): gantry connection fault 073/016
Gene software error (IST, standby): mA measurement calibration problem 073/017
Gene software error (IST, exposure): unknown exposure end 074/001
Gene software error (IST, exposure): unknown gantry reply 074/002
Gene software error (IST, exposure): unknown gantry reply 074/003
Gene software error (IST, exposure): unknown exposure status 074/004
4–7
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
4–8
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
ERROR CODES
Gene: ’Endurance’ error code 082/001
Gene: ’Endurance’ error code 082/010
Gene: ’Endurance’ error code 082/020
Gene: ’Endurance’ error code 082/030
Gene: ’Endurance’ error code 082/040
Communication error Digital system 083/000
Failure before exposure: rising Synchro_Grid signal not received from Gantry (after 084/008
Graphy pushed)
Failure before exposure: rising Synchro_Grid signal not received from Gantry (after 084/009
Graphy pushed and exposure start status received from the generator driver)
Generator: PROM checksum error 090/000
U–turn before end of pre–shot: The exposure time for the pre–shot (AOP mode) is greater 097/026
than the grid motion duration until the first turn–back point.
U–turn before end of exposure: The exposure time for the exposure (manual and AOP 097/027
mode) is greater than the grid motion duration until the first turn–back point.
Gantry software error: VRTX error during init 100/001
Gantry interface: Reset error (see Chap. 1 Sect. 3) 101/001
Gantry interface: read/write error on timer PTM1 (see Chap. 1 Sect. 3) 101/002
Gantry interface: read/write error on timer PTM1 (see Chap. 1 Sect. 3) 101/003
4–9
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
Gantry interface: read/write error on timer PTM2 (see Chap. 1 Sect. 3) 101/004
Gantry interface: read/write error on timer PTM2 (see Chap. 1 Sect. 3) 101/005
Gantry interface: interrupt error on DUART1 (see Chap. 1 Sect. 3) 101/006
Gantry interface: PTM1 timer 1 defective (see Chap. 1 Sect. 3) 101/007
Gantry interface: PTM1 timer 2 defective (see Chap. 1 Sect. 3) 101/008
Gantry interface: PTM1 timer 3 defective (see Chap. 1 Sect. 3) 101/009
Gantry interface: PTM2 timer 1 defective (see Chap. 1 Sect. 3) 101/010
Gantry interface: PTM2 timer 2 defective (see Chap. 1 Sect. 3) 101/011
Gantry interface: PTM2 timer 3 defective (see Chap. 1 Sect. 3) 101/012
Gantry interface: interrupt error on PTM1 timer 1 (see Chap. 1 Sect. 3) 101/013
Gantry interface: interrupt error on PTM1 timer 2 (see Chap. 1 Sect. 3) 101/014
Gantry interface: interrupt error on PTM1 timer 3 (see Chap. 1 Sect. 3) 101/015
Gantry interface: interrupt error on PTM2 timer 1 (see Chap. 1 Sect. 3) 101/016
Gantry interface: interrupt error on PTM2 timer 2 (see Chap. 1 Sect. 3) 101/017
Gantry interface: interrupt error on PTM2 timer 3 (see Chap. 1 Sect. 3) 101/018
Gantry interface: PROM checksum error 101/020
Gantry interface: unexpected interrupt received (level 2) 101/023
ERROR CODES
4–10
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
ERROR CODES
Gantry software error: unknown acknowledgement (power calculation) 105/066
Gantry software error: unknown event (power calculation) 105/067
Gantry software error: unknown tube type 105/068
Gantry software error: unknown focus change code 105/069
Gantry software error: unknown filter code requested by generator 105/070
Gantry software error: unknown code for ’Exposure’/’No Exposure’ status management 105/071
Gantry software error: unknown exposure end code 105/072
Gantry software error: unknown ’Endurance’ code 105/073
Gantry software error: unknown ’Endurance’ event 105/074
Gantry/tube housing type configuration incompatibility (angulation/angulation) 105/080
Gantry/tube housing type configuration incompatibility (angulation/focus) 105/081
Gantry software error: column: unknown task return 106/002
Gantry software error: column: unknown task status 106/003
Gantry software error: column: unknown task status 106/004
Gantry software error: column: unknown task status 106/005
Gantry software error: column: unknown task status 106/006
Gantry software error: column: unknown order transmitted to task 106/007
4–11
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
Gantry software error: compression: too long between task calls 107/001
Gantry software error: compression: unknown task return 107/002
Gantry software error: compression: unknown task status 107/003
Gantry software error: compression: unknown order transmitted to task 107/004
Gantry software error: lateral collimator: positioning timeout 108/001
Gantry software error: lateral collimator: unknown task return 108/002
Gantry software error: lateral collimator: unknown task status 108/003
Gantry software error: lateral collimator: unknown task status 108/004
Gantry software error: lateral collimator: unknown task status 108/005
Gantry software error: lateral collimator: unknown order transmitted to task 108/006
Gantry software error: lateral collimator: unknown order transmitted to task 108/007
Gantry software error: lateral collimator: position reference value is outside tolerance 108/008
Gantry software error: rear collimator: positioning timeout 109/001
Gantry software error: rear collimator: unknown task return 109/002
Gantry software error: rear collimator: unknown task status 109/003
Gantry software error: rear collimator: unknown task status 109/004
Gantry software error: rear collimator: unknown task status 109/005
ERROR CODES
Gantry software error: rear collimator: unknown order transmitted to task 109/006
Gantry software error: rear collimator: unknown order transmitted to task 109/007
Gantry software error: rear collimator: position reference value is outside tolerance 109/008
Gantry software error: tube angulation: unknown task status 111/002
Gantry software error: tube angulation: unknown task return 111/003
Gantry software error: tube angulation: unknown task status 111/004
Gantry software error: tube angulation: unknown task status 111/005
Gantry software error: tube angulation: unknown order transmitted to task 111/006
Gantry software error: tube angulation: unknown order transmitted to task 111/007
Gantry software error: tube angulation: position reference value is outside tolerance 111/008
Gantry software error: tube angulation: task triggered by timeout 111/017
Gantry software error: filter wheel: unknown task return 112/000
Gantry software error: filter wheel: unknown task status 112/001
Gantry software error: filter wheel: unknown task status 112/002
Gantry software error: filter wheel: unknown task status 112/003
Gantry software error: filter wheel: 112/004
Gantry software error: filter wheel: unknown order transmitted to task 112/005
4–12
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
Gantry software error: light centering: unknown order transmitted to task 114/000
Gantry software error: unknown set–up message 117/000
Gantry set–up access: generator end of exposure timeout 117/002
Gantry set–up access: communication sync problem at set–up access 117/003
Gantry software error: grid: task triggered by timeout (calibration time too long) 118/000
Gantry software error: grid: unknown task return 118/001
Gantry software error: grid: unknown task status 118/002
Gantry software error: grid: unknown task status 118/003
Gantry software error: grid: unknown task status 118/004
Gantry software error: grid: unknown order transmitted to task 118/005
Gantry software error: grid: unknown order transmitted to task 118/006
Grid movement: overcurrent detected 118/007
Grid movement: grid calibration time too long 118/008
Grid movement: overcurrent detected 118/009
Grid movement: invalid offset 118/010
Grid movement: invalid gain 118/011
Grid movement: limit of travel not found 118/012
ERROR CODES
Gantry software error: arm lock: unknown task return 120/000
Gantry software error: arm lock: unknown task status 120/001
Gantry software error: arm lock: unknown task status 120/002
Gantry software error: arm lock: unknown order transmitted to task 120/003
Gantry software error: rot. brake: unknown order transmitted to task 121/000
Gantry software error: rot. brake: unknown task status 121/001
Gantry software error: rot. brake: unknown task status 121/002
Gantry software error: rot. brake: unknown task return 121/003
Gantry software error: rot. brake: unknown order transmitted to task 121/004
Gantry software error: Reset triggered by Watch_dog_soft 122/007
Gantry software error: grid sensor Presence_Trame (Grid_Presence) not ’high’ or ’low’ 122/013
Gantry software error: tube housing thermal contact not ’high’ or ’low’ 122/014
Gantry: checksum error on battery saved memory 151/000
Gantry interface: unexpected interrupt received (level 0) 151/021
Gantry interface: unexpected interrupt received (level 1) 151/022
Gantry ’Endurance’ error code: synchronisation 153/075
Gantry ’Endurance’ error code: column 153/076
4–13
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
4–14
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
Column up/down: column is driven down but measured speed = 0 cm/s 156/025
Column up/down: emergency stop produced if movement continues more than 500 mS 156/027
after command release
Compression: motor slipping during reference search *** 157/005
Compression: abnormal immobility of paddle *** 157/006
Compression: loss of steps detected at reference crossing *** 157/007
Compression: abnormal step generation detected (during init) 157/008
Compression: signal Valid_Comp (Comp_Valid) cannot inhibit paddle movement (during 157/009
init)
Compression: counter used by force sensor no longer incremented *** 157/010
Compression: position sensor closed too long during compression 157/011
Compression: no reference crossing during decompression *** 157/012
Compression: signal Pres_Oper (Operator_Presence) cannot inhibit paddle movement 157/013
(during init)
Compression: signal Pres_Ajust (Comp_Adjust) too long in ’low’ status 157/014
Compression: signal Pres_Ajust (Comp_Adjust) too long in ’high’ status 157/015
Compression: signal Pres_Log_Comp (Comp_Soft_Pres) cannot inhibit paddle movement 157/016
(during init)
Compression: signal Pres_Log_Comp (Comp_Soft_Pres) too long in ’low’ status 157/017
ERROR CODES
Compression: signal Pres_Log_Comp (Comp_Soft_Pres) too long in ’high’ status 157/018
Compression: power supply problem (signal Alim_OK (Power_Supply_Active)) 157/020
Compression: phase1 & phase2 cannot be ’low’ at the same time (during init) 157/021
Compression: signal Pres_Log_Comp (Comp_Soft_Pres) ’high’ (during init) 157/022
Compression: slipping during decompression movement *** 157/023
Lateral collimator motor: loss of steps detected at reference crossing 158/009
Lateral collimator motor: loss of steps detected at reference crossing 158/010
Lateral collimator motor: loss of steps detected at reference crossing 158/011
Lateral collimator motor: loss of steps detected at reference crossing 158/012
Lateral collimator: no reference crossing during movement 158/013
Lateral collimator: no reference crossing during movement 158/014
Lateral collimator: no reference crossing during movement 158/015
Lateral collimator: no reference crossing during movement 158/016
Lateral collimator: optical sensor detection problem 158/017
Lateral collimator: optical sensor detection problem 158/018
Lateral collimator: slipping during collimator movement *** 158/019
Rear collimator motor: loss of steps detected at reference crossing 159/009
4–15
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
4–16
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
ERROR CODES
Power shutdown: uncontrolled movement risk (arm lock); R_On_Verrou 170/013
(On_Lock_Return) ’high’ without legal command
Arm rot. brake: signal Pres_Log_Frein_Rot (Rot_Brake_Soft_Pres) from compression arm 171/005
too long in ’low’ status
Arm rot. brake: signal Pres_Log_Frein_Rot (Rot_Brake_Soft_Pres) from compression arm 171/006
too long in ’high’ status
Arm rot. brake: signal R_On_Frein (On_Brake_Return) ’high’, but signal Eta_Rot_Off 171/007
(Rotat_Power_Inactive) ’low’
Arm rot. brake: Pres_Log_Frein_Rot*Pres_Oper*On_Frein_Rot 171/008
(Rot_Brake_Soft_Pres*Operator_Present*On_Rotation_Brake) = 1 & R_On_Frein_Rot
(On_Rotation_Brake_Return) = 0
Arm rot. brake: signal R_On_Frein_Rot (On_Rotation_Brake_Return) ’high’ without legal 171/009
command
Power shutdown: uncontrolled movement risk (arm brake); Eta_Rot_Off 171/010
(Rotat_Power_Inactive) & R_On_Frein_Rot (On_Rotation_Brake_Return) are ’low’
Power shutdown: uncontrolled movement risk (arm brake); Eta_Rot_Off 171/011
(Rotat_Power_Inactive) ’low’ & comp. arm defective
Power shutdown: uncontrolled movement risk (arm brake); Eta_Rot_Off 171/012
(Rotat_Power_Inactive) ’low’ during init
Arm rot. brake: signal R_On_Frein_Rot (On_Rotation_Brake_Return) ’high’ during init 171/013
Power shutdown: uncontrolled movement risk (arm brake); R_On_Frein_Rot 171/014
(On_Rotation_Brake_Return) ’high’ without legal command
4–17
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
4–18
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
ERROR CODES
4–19
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
SECTION 3
ERROR CODES: KNOWN ROOT CAUSES
071/081: Generator checksum error: check + 5.1V (see Job Card IST022) and Lithium
battery.
157/005: check for play between compression motor axis and pulley (screw not tightened).
157/010: loss of compression feedback; a square wave should be always present between TP
14 and TP1 (0V) of 800PL2.
157/023: check for play between compression motor axis and pulley (screw not tightened).
158/019: check for play between collimator lateral motor axis and pulley (screw not
tightened). Possible problem with 800PL1 (Gantry power board).
ERROR CODES
159/019: check for play between collimator rear motor axis and pulley (screw not tightened).
160/019: check for mechanical play between collimator front motor axis and blade (screw
not tightened).
check for poor position of the flag in optocoupler fork.
check the calibration of front blade in IST030.
171/019 and 171/020:
reset DMR.
check for friction point in sensor; change if necessary.
If problem persists, note software version and call your OLC.
171/021 and 171/024:
reset DMR.
check sensor, connections, and board 800PL5, change if necessary.
172/006: if appearing at power on, the gantry interface board is probably defective.
4–20
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
SIGNAL LISTS
5–i
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
Page blank
SIGNAL LISTS
5–ii
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
GLOSSARY
ENGLISH/FRENCH
GENERATOR (GENERATEUR)
SIGNAL LISTS
5–1
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
Page blank
SIGNAL LISTS
5–2
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
SIGNAL LISTS
-15V_Vk 701PL1 - 5/A1 ...... - 15VK
-15V_Vm 701PL1 - 5/A3 ...... - 15VM
-VE (-8V) 200PL1 - 3/C10 - VE (-8V)
0VD 701PL1 - 5/F3 ...... 0VD
0VE 200PL1 - 3/B10 0VE
0VLC 200PL2 - 2/B4 0VLC
0VV 200PL1 - 3/E10 200PL2 - 1/D2 0VV
0V_27V 200PL2 - 2/F3 0V_27V
0V_250V 200PL2 - 2/B3 0V_250V
0V_600VF 300PL1 - 1/H1 200PL1 - 1/E11 0V_600VF
0V_CONSOLE 200PL2 - 2/F11 400PL1 - 4/C4 0V_PUPITRE
0VCPU 400PL3 - 8/A2 0VCPU
5–3
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
5–4
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
SIGNAL LISTS
#
#
#
#
#
#
#
#
#
#
5–5
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
$! !
$!
!
$!() ) )!")
$!()
)
)!")
$!())
)!")
$!' !'
$#
"! #
SIGNAL LISTS
$#"#))
)#"#)
$#"#)) )#"#) # $"
$!!#)"
!"#)$!#
)#
)#
)#
)#
)#)
)#
)#
)#
)#
)#
)#
)#
&$#)#
) #
"!)!""#!
!)!
5–6
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
SIGNAL LISTS
%&"!!
"!&
&
&
!&
&"
!&
&"
!&
&"
!&"
"&"
!&"!&
&
5–7
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
'$
" ' '
!'#'
" ' #!" #"' #'
"'# "'!# ' !''
"'!'" "' "
"'!'" "
"'!'"'
"' '
"'!'"'
"' '
"'!'"' '"'
" "
"' "'
"''
"''
"''
"''
"' "'
"' '
"' '
SIGNAL LISTS
5–8
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
*&)
*&)" )
%$)*&)
)%$)*&) %$
*&)" )
*&)
*&)'
*&)'
*&)')
*&)')
*&)')
)*&)') *&)')
*&)')
*&)')
*&)')
)*&)') *&)')
*&)#%"$
#%")*&
*&)""
#)*&
*&)"!%#$ %$)"%
*&)"!%#$) %$)"%)
*&)"!%#$) %$)"%)
##)$)*&)
*&))) *&)
##)$)*&)
*&)))
##)$)*&)
*&))) *&)
##)$)*&)
*&))) *&
"$
) ) )*&))
) ) )*&)
)*&)
)*&) )
)*&)
)*&) )
&)#% ())
%$))$
SIGNAL LISTS
+
+
#) "# "#)#$%"
$)"$" $"$
+# +#
+)
+) #%")+)
+))
+) #%")+)
+) ' & %$ "$%")+
+)'
+)'
+)') +)')
+)')
+)')
+)#%"$)
#%")+) +
+)#%"$)
#%")) +)
5–9
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
")
)"$!#
"$!)")
" "
&!
$""
&!)#!)$!!#
$"")$!#)
&!) "#
&!))! $!#
)"#
&!))&#
###)"#
&!)"$') !"))
&!)"$') !!#))
&!)"$') )
&!)"$')!" !")
&!)"$') !")
&!)"$') !")
SIGNAL LISTS
5–10
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
SIGNAL LISTS
$%#%#*#'* !
*% &#
$%# *+'* $%# *+'*
$%# *+'* !
! $%# * $%# *+'*
$%# *+'* $%# *+'*
$%# *+'* !
! $%# * $%# *+'*
$) !
$)
$)# *# !
$)# *#$
$)*# % #* !
#**$)
%!#*#*%*%(
!
%%*$&!*%(
%!#*$$*%*%(
!
%%**%(
%$%*
+'
! %$%*
+'
5–11
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
$"*
*$"
%* #$*
$"*
* $"* *
$"*
*
$"*
*
$"*
*
$"*
*
$( #"
$(
$(*"$"# #
$(*"$%" $(
$( #"
$(
$(*"$"# #
$(*"$%" $(
$(*" #"
$(*"#
$(*"#
#"& $" #
$(*"#
$*##*$*,
$**,
&
&
&
&
&*" )
&*"
&* "
&* "
&
& &
&
&
&
SIGNAL LISTS
&
&
'$*#$$
$$$*#$
'"
%$
*( #%" * #
*$*&"
*#%""*$
*&*#% )*
*%$**$
* *+&*$#$ *$#$*+&*&
* *"%$$"
**%
*&"*%""$*$$
*#%"$*"*
* '"**"!%"$ **#$
*$"*
**
5–12
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
GLOSSARY
ENGLISH/FRENCH
GANTRY (STATIF)
SIGNAL LISTS
5–13
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
Page blank
SIGNAL LISTS
5–14
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
NOM DU SIGNAL
SIGNAL NAME (GANTRY) FROM TO
(STATIF)
800PL4 - 1/E7 ...... 0VISOA
800PL1 -21/H1 ...... 0V_27V_CL
+15ISO (800PL4) 800PL4 - 1/G8 ...... +15ISO (800PL4)
+15V (800PL1) 800PL1 -20/D3 ...... +15V (800PL1)
+15V (800PL2) 800PL2 - 3/F2 ...... +15V (800PL2)
+5ISO (800PL4) 800PL4 - 1/E8 ...... +5ISO (800PL4)
-15ISO (800PL4) 800PL4 - 1/G7 ...... -15ISO (800PL4)
-5V (800:PL1.PL2) 800PL1 -20/B3 -5V (800:PL1.PL2)
-FILTER_STANDBY 800PL2 - 3/B5 800PL1 -15/H3 -STANDBY_FILTRE
-HOUSING_STANDBY 800PL2 - 3/B5 800PL1 -14/H3 -STANDBY_GAINE
-LAT_DIAPH_STANDBY 800PL2 - 3/B5 800PL1 -17/H3 -STANDBY_DIAPH_L
-REAR_DIAPH_STANDBY_A 800PL2 - 3/B5 800PL1 -16/H3 -STANDBY_DIAPH_A
0V_STEREOTIX 800PL1 -21/F4 STEREOTIX 0V_STER
12V (800PL1) NOT USED 800PL1 -21/C4 800PL2 - 3/F2 12V (800PL1) NON UTILISE
12V (800PL2) NOT USED ...... ...... 12V (800PL2) NON UTILISE
15V = +15V ON 800PL1 800PL1 -21/A4 800PL2 - 3/F1 15V = +15V DE 800PL1
24 x 30_SIZE 800PL10- 1/B2 800PL2 - 9/H8 FORMAT_24*30
24 X 30_SIZE_A 800PL2 - 9/D8 ...... FORMAT_24*30_A
27V 800PL1 -20/H4 ...... 27V
27V_BRAKE 800PL1 -22/A2 27V_FR
27V_CL (LOGIC BOARDS) 800PL1 -22/D4 ...... 27V_CL (CARTES LOGIQUES)
27V_COLUMN 800PL1 -20/D2 ...... 27V_ASC
27V_COMP 800PL1 -22/F4 ...... 27V_COMP
27V_D 800PL1 -21/E2 800PL2 - 7/G2 27V_D
27V_EXT (EXTERNAL) 800PL1 -22/A3 800PL5 -80/0L 27V_EXT
SIGNAL LISTS
27V_GRID 800PL1 -22/D2 ...... 27V_TRAME
27V_LIGHT_CENTERING 800PL1 -22/F3 ...... 27V_CENTR
27V_LOCK 800PL1 -22/F2 ...... 27V_VERROU
27V_P 800PL1 -20/G4 ...... 27V_P
27V_STEPPER 800PL1 -22/D3 ...... 27V_PAP
27V_STEREOTIX 800PL1 -20/A2 STEREOTIX 27V_STER
5VEXT (EXTERNAL) 800PL1 -21/C3 800PL2 - 3/G2 5VEXT
5V_CL (LOGIC BOARDS) 800PL1 -21/C3 800PL2 - 7/H3 5V_CL
5V_STEREOTIX STEREOTIX 800PL12- 1/A4 5V_STER
ARM_SYNCHRO 800PL2 - 7/E3 400PL2 - 6/B4 SYNCHRO_BRAS
ARM_TILT_±27 800PL15- 1/B3 800PL2 - 9/H3 INCLIN_GARAGE
ARM_TILT_±27_A 800PL2 - 9/C5 ...... INCLIN_GARAGE_A
5–15
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
NOM DU SIGNAL
SIGNAL NAME (GANTRY) FROM TO
(STATIF)
'!!'
'
'!!'' ''
'!!'
'
'!!'' ''
'!!' '
'!!'' ''
'" "'
''" ''!
"&&'
'"&&
!' '
'
!' '
'
'
!!'!! ' '
!!'!! ' ' !!
'' ''
'' ''
' !!% !!%
' !
'"
'"
'" '"
"''
' '
"''
' '
"'
"'''
' ''
"'''!
' ''
"'''
' ''
"'''!
' ''
SIGNAL LISTS
5–16
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
NOM DU SIGNAL
SIGNAL NAME (GANTRY) FROM TO
(STATIF)
!& *' "
'*$
!! *#!& *"!(+
"
"
!& *$$*"&$$
#*!#
!"* "
!"*
!"*
"
!"*
!"*&$%*"#$ "
"#$*&$%*!"
!"** "
*!"*
!"*&## %*#*'+
"
% $**!"
&*
!"*&## %*'&* "
% $* *!"
!"*&## %*'&*
"
% $*
*!"
!"*&## %*'&* "
% $**!"
!"*#%! "
$ $*!"
!"*#!& "
" $$*!"
!"*'%*' "
'*'%*!"
!"*! * " !"*! *
!"*! *
" !"*! *
!"*!"#%!#*"#$ "
"#$*!"*!"
!"*" "
"
"*!"
!"*"$
"
!%!# "
"$*!"*
!"*"$* " !%# " "$*!"*
!"*"$*
"
!%!# "
"$*!"*
!"*"$*
" !%!# " "$*!"*
!"*"#$ *" "
"#$*"*!"
!"*#*#!& "
" $$*! $ *!"
!"*#*"* "
"
! $ *!"*
SIGNAL LISTS
!"*#*"*
"
" ! $ *!"*
$&#%*!" !%
!"*$%)
&$
!"*$%)* $&#%*!"*
!"*$!%*"#$ "
"#$*!*!"
!"*$"*! "
!*""*!"
!"*$&"") '
' "
" *!" '
'
!"*' "
"&
'*%&*%%#) "
*$&'*
'
"& #$$ &$ "
"
"& #$$ &$ "
"
"& #$$ &$ "
"
5–17
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
NOM DU SIGNAL
SIGNAL NAME (GANTRY) FROM TO
(STATIF)
!% "## %# !
!
!%
"## %# !
!
!%(
$ %# !
!
(
5–18
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
NOM DU SIGNAL
SIGNAL NAME (GANTRY) FROM TO
(STATIF)
!%( $ %#
!
!
(
!((
! ##( !
!(!
!
! !( !
"'(#$ !("'
!
!
! "(! #($"
($"(&
$"(( %#(#%!!'
!
!
&
$"(" %("$%"
!
! "$ %"(##($"(
$"( $ "(
!
! $($"(
$"( $ "(("$%"
!
! $($"(
$"( $ "(("$%"
! $ " ! !#($"(
$"(!#(
! $ " ! !#($"(
$"(!#( ! $ " ! !#($"(
$"(!#(
! $ " ! !#($"(
$"(! #("
!
! "(! #($"
$"(! #("(
!
"(#$'($"
$"(#$(" %
!
! ##(#$($"
$"(#%!!'
!
!
!
#( "
"(" &
!
!
#( "
" $(!( $((")
SIGNAL LISTS
$%"
" $(!( $((")
!
! $(!(&(
" $(!( $ "(
!
! $(!(&(
" $
" $(!(!#(
! !#(!(&(
$ "
" $
" $(!(!#(
! !#(!(&(
$ "
" $
" $(!(!#(
! !#(!(&(
$ "
" $
" $(!(!#(
! !#(!(&(
$ "
5–19
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
NOM DU SIGNAL
SIGNAL NAME (GANTRY) FROM TO
(STATIF)
#! %*"*"!$*# "
" #*"!$*"*'
#! %*"*$% ) " "
$% )*"*'
#! %*"*$% +
"
#*$% )*"*'
)*#%&#
#! %*"*$&"") " "
*"*' ' '
&$*! "
"
&$*!
&$*! "
&$*!*
# "
%#
#*# "
#*%#
#*!"*#!& *#+
" "
#%!&#*$$*!"*%#
%&#
#*!%!# "
# !%!# !%*%#
#*!%!# "
# !%!# !%*%#
#*! *#!& " "
$$*! *%#
#*"!$%! "
"
"!$*%#
#*"!$%! *#!& "
"
$$*"!$*%#
#*"#$ "
" "#$ *%#
#*"#$ * "
"#$ *%#*
#*## "
"
! $ *%#
#*#*#!& " "
$$*! $ *%#
#*$%!" "
" &%*%#
#*$%!"* " &%*%#*
!&$ *"!$*# "
" #*"!$*
!&$ *!%!#*
" "
!%* *
!&$ *!%!#*
*#+
"
#*!%* *
%&#
!&$ *!%!#* " "
!%* *
SIGNAL LISTS
5–20
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
NOM DU SIGNAL
SIGNAL NAME (GANTRY) FROM TO
(STATIF)
! * #
'"*## " "
%"
#*"*"!$*"*
"
" *"!$*"!%
"#$**$!
$!*$&"")*%$% "
"
%$%**$!
%*"*"!$*# ' "
" #*"!$*"*
%*"*!%!#*
" "
!%*"**
SIGNAL LISTS
%* %# "
%#&#
%* %# "
%# & %#&#
%* %# "
%# & %#&#
%* %# * "
" %#&#*& &(
%* %# ** " %#&#*& &(*
%* %#+
" "
$$* %#&#
*#!&
%* %# *! "
"
%#&#*!
%* %# *!* "
%#&#*!*
#%!&#*$$* +
%*#!& *#%&# " "
%#&#
! "
! '##!&
5–21
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
NOM DU SIGNAL
SIGNAL NAME (GANTRY) FROM TO
(STATIF)
#"
%
%#"
%
#"
%%
%#"%
%"
"%#"
%"
%#"
%$%!#
!%#"%
% !%
%%#"
%#
#%#"
%
"%
%%
"%%
!
%
!
%
!%
%
!%%
%%
!%%
%%
!%%
%%
%" !
%"
%
%
%
%
!%%
!%%
!%%%!"
%!%%
!%%
!%%
!%%%!"
%!%%
%"
%
SIGNAL LISTS
5–22
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
NOM DU SIGNAL
SIGNAL NAME (GANTRY) FROM TO
(STATIF)
"#$&"$ #$%!&+
#
# #$%+"#+$
"#$&"$+#$%!&
# #$%+"#$
"#$&"$+#$%!&
#
#
#$%+"#+$
+&
"$+'%+
#
"'+
#+#"%+$
(
#
# $+#"%+#"&
#+#"%+#+
#
# #"%+#"&
#+#"%+#+
(
#
# #"%+#"&
#+#"%+#++
# #"%+#"&+
#+#"%+#+
#
# #"%+#"&
#+#"%+#+
(
#
# #"%+#"&
#+#"%+#++
# #"%+#"&+
#+#"%+$
#
# $+#"%+#"&
#+#"%+$+
# $+#"%+#"&+
#+" '!
#
# " '!+#
(
#+#$%!
#
# #$%+#
#++" #+'$,
#
+#%+
$!&+ %
#++" #+'$,
# +#%+
$!&+ %
%+&# +# $%
# (+# + %+
#
#
%'$
# (+# + %+
#
#
#
!
#")$+$+"
#
#")+&$ +"+
SIGNAL LISTS
#")$+%'##*+&(
#
#
+" ( (
#")$+%'##*+&(+
#
+"+
#+" #+
# #+" #+
#+" #+
# #+" #+
#+" #+
# #+" #+
#+" #+
# #+" #+
$$ #+%'##*+
#
#
# $+#"%+#+
$$+ + &+# %'#, +#++ (
#
#
#* (
$$+#+$"'!+$,
#
#
$&"'$+ %%+#+
&'$!
5–23
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
NOM DU SIGNAL
SIGNAL NAME (GANTRY) FROM TO
(STATIF)
""*!* $ "*
! !
$*!**
$%"
""*!* $**",
!
"* $*!**
$%"
""*!*!#*
!
$ " !#*!**
""*!*!#*
!
$ " !#*!**
""*!*!#* !
$ " !#*!**
""*!*!#* !
$ " !#*!**
""*!*! #*" !
! "*! #*!*
""*!*! #*"** ! "*! #*!**
""*!*#$)*",
!
"*#$)*!*
$%"
""*!*#$*" % ! !
##*#$*!*
" $$ *" !
" "*" $
" $$ *" !
" "*" $
" $$ *"*& !
&*"*" $
" $$ *! '"*,
!
$*&"" %* *
$&*
" $$*! '"*$& !
!
$*" $*
" $$*! '"*$&* !
$*" $* *
" $*"* ! "*" $
" $*"** ! *"*" $*
" $*"**" *
!
! *"*" $*"
" $*"*" % ! !
##*"*" $
SIGNAL LISTS
5–24
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
NOM DU SIGNAL
SIGNAL NAME (GANTRY) FROM TO
(STATIF)
SID_2_ (560mm) 800PL5 - 2/F4 800PL2 - 9/H6 FOCALE_2 (560mm)
SID_2_A 800PL2 - 9/D8 ...... FOCALE_2_A
SID_3* (0V) 800PL5 - 2/F4 800PL2 - 9/H5 FOCALE_3*
SID_3_ (510mm) 800PL5 - 2/F4 800PL2 - 9/H5 FOCALE_3 (510mm)
SID_3_A 800PL2 - 9/D8 ...... FOCALE_3_A
CDE_FORMAT_A (COLLIMAĆ
SISE_CMD_A 800PL2 - 8/B6 ......
TOR)
CDE_FORMAT (COLLIMAĆ
SIZE_CMD 800PL6 - 1/D4 800PL2 - 8/G3
TOR)
SOFT_CMD_START_COLĆ
800PL2 - 4/E8 ...... START_LOG_ASC
UMN_
SOFT_CMD_START_COMP 800PL2 - 4/E4 ...... START_LOG_COMP
SOFT_CMD_START_LOCK 800PL2 - 4/E4 ...... START_LOG_VERROU
SOFT_CMD_START_ROT_BRA
800PL2 - 4/E4 ...... START_LOG_FREIN_ROT
KE
STEREOTIX_AUTHORIZAĆ
STEREOTIX 800PL2 -10/H2 AUTORISATION_STEREOTIX
TION
STEREOTIX_AUTHORIZAĆ AUTORISATION_STEREOĆ
800PL2 -10/C2 ......
TION_A TIX_A
STEREOTIX_SUPPLY_PRESĆ
800PL2 - 6/C5 ...... PRES_ALIM_STER
ENCE
TEMPER_HOUSING_SAFETY 800PL6 - 1/D4 800PL2 - 8/G5 SECU_GAINE
TEMPER_HOUSING_SAFEĆ
800PL2 - 8/C6 ...... SECU_GAINE_A
TY_A
TEST_COLUMN_STOPS 800PL2 - 3/B7 ...... TEST_BUT_ASC
TUBE_ARM_DISTR_GND_REĆ
800PL1 -19/G3 800PL2 - 8/G5 RETOUR_MASSE_RG
TURN
TX-PID (PATI.IDENT.DEVIC) 800PL3 - 3/B3 800PL1 -21/G4 TX+PID (T1B2)
TX-PID (PATI.IDENT.DEVIC) 800PL3 - 3/B3 800PL1 -21/G4 TX-PID
SIGNAL LISTS
TX_GENE_(OPT.SERIAL LINK) 800PL3 - 3/B3 800PL2 - 7/E2 TX_GENE (T1A2)
UCOM_GRID 800PL1 -10/A2 ...... UCOM_TRAME
VCC (+5V) 800PL1 800PL1 -20/C4 ......
VCC (+5V) 800PL1
VIA 800PL1 - 2/E2 ...... VIA
W18_GROUND_RETURN 800PL5 - 1/E4 800PL2 - 8/G2 RETOUR_MASSE_W18
W19_GROUND_RETURN 800PL5 - 1/C4 800PL2 - 8/G3 RETOUR_MASSE_W19
_12V_DROP 800PL2 - 7/D6 ...... _CHUTE_12V
_15V_DROP 800PL2 - 7/D7 ...... _CHUTE_15V
_27V_DROP 800PL2 - 7/A1 ...... _CHUTE_27V
_COLUMN_LOWER_STOP_A LOWER LIMIT 800PL2 -11/B4 _BUTE_ASC_BAS_A
5–25
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
NOM DU SIGNAL
SIGNAL NAME (GANTRY) FROM TO
(STATIF)
"" "#
""
""""
"
"""
""
!
"
!"
""
""
"" "
" ""
"" "
""
"" "
""
""#
""""
"
"
""#
""""
"
" "!"
""
"" "
""
SIGNAL LISTS
5–26
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
GLOSSAIRE
FRANÇAIS/ANGLAIS
GENERATEUR (GENERATOR)
SIGNAL LISTS
5–27
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
Page blank
SIGNAL LISTS
5–28
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
SIGNAL LISTS
-15V (CPU NOT USED) - 15Vcpu
-15VK 701PL1 - 5/A1 ...... - 15V_Vk
-15VM 701PL1 - 5/A3 ...... - 15V_Vm
-VE (-8V) 200PL1 - 3/C10 - VE (-8V)
0V-VENTILATEUR 200PL2 - 2/H10 0V_FAN
0V232_1 400PL3 - 3/C4 CPU:0V232_1
0VD 701PL1 - 5/F3 ...... 0VD
0VE 200PL1 - 3/B10 0VE
0VK 701PL1 - 5/A2 ...... 0V_kV
0VLC 200PL2 - 2/B4 0VLC
0VM 701PL1 - 5/A4 0V_mA
0VV 200PL1 - 3/E10 200PL2 - 1/D2 0VV
5–29
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
'# "
"
'# '#%#
*%!&#$ " " *#"
+'* +'* ********* "
"
+'*
+'* +'* ** "
"
+'**
'
" "
' '#%#
**!
" " * '*&## %
**!
" " * '*&## %
*"&$$*! " " '*$&"")*!
*"&$$*!* "
" '*$&"")*!*
*"&$$*!* "
" '*$&"")*!*
*"&$*! " " '*$&"")*!
5–30
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
SIGNAL LISTS
$#)!
!)$##
$#)!
!)$##
)#
)#
)#
)#
)#
)#
)#
)#
)#
)#
)#
)#
)##$! ##!)
!#(*
)!!
#)
5–31
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
5–32
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
SIGNAL LISTS
% %
#
%
%
#%
% % %%
%
%%
%%&
% %
% %
%%
%
%
%
%!%
%
%
%
%
% $
%
% $
%
% $
5–33
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
5–34
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
SIGNAL LISTS
!(( (!# "(
!(( (!# "(
$(# "(!# +
!((
"
!" % (
(#
" (
!(
" !(
!(
" !(
!(
" !(
!"(
" (
5–35
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
5–36
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
SIGNAL LISTS
T1A2 400PL3 - 3/B3 CPU:T1A2 (SERIAL LINK)
T1B1 400PL3 - 3/B4 CPU:T1B1 (SERIAL LINK)
T2A1 400PL3 - 3/B4 CPU:T2A1
T5000 400PL2 - 4/B4 400PL1 -25/H4 5000_RPM
TEMPS_POSE_MAX 400PL1 - 9/A2 400PL2 - 5/H4 EXPOSURE_TIME_MAX
TEST_62KV5 701PL1 - 1/G3 TEST_62KV5
TEST_kV_VIDE 400PL1 - 10/F3 ...... NO LOAD_kV_TEST
TEMĆ
TETA_INF_TMAX 200PL2 - 4/D11 ......
PER_LESS_THAN_TMAX
TEMPER_HIGHĆ
TETA_SUP_TMAX 200PL2 - 4/F9 ......
ER_THAN_TMAX
THERMAL
THERMOSTAT+ 200PL2 - 2/F10 THERMOSTAT+
SWITCH
5–37
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
#!(
(
#!(
(
#!(
(
#!(
(
#&
#& "!
#&(!#!*
#&(!#$! #&
" "
#&
#& "!
#&(!#!*
#&(!#$! #&
" "
#&(!"
#&(! "!
#&(!"
"!% #!
#&(!"
"
#(()
#(""(#()
#(! #!
#("(#
#(!( #("(# (
#(!( #("(# (
#("$ ()
#(!(#()
%
%
SIGNAL LISTS
%
%
%(!
%(! '
%( !
%( !
%
% %
%
%
%
%
%
%'#(
(#($##
%'#(
(#($##
%'#( "(
& "$!(#(
%'#( "(
& "$!(#(
($#((#
(%("$ '(
5–38
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
SIGNAL LISTS
5–39
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
Page blank
SIGNAL LISTS
5–40
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
SIGNAL LISTS
5–41
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
Page blank
SIGNAL LISTS
5–42
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
DESTINAĆ
NOM DU SIGNAL (STATIF) ORIGINE SIGNAL NAME (GANTRY)
TION
$ ! !
$ !
' ! !
' !
' !
!
' !
$ ! !
$ !
$ ! !
$ !
' ! !
!
' ! !
'$ !
')
') !
')$%# !
$%# %( ')$%# %(
' ! &%$ !
!
' ! % &$
' !
&%$
' !
% &$
'
' ! !
!
'
' !
' !
'
')$ !
') &
')%# !
')%)%#
') #$
') ! !
') !
') !
!
')
')(% !
!
')(% (%#
')# !
')#
')! !
')!
')!! !
')$%!!#
')$%# !
$%# %(
')$%# %(
')%# !
')#
')
SIGNAL LISTS
'(% !
!
'(% (%#
') !
!
') #$
')$%# $%# %( !
')$%# %(
!
!
!& #$$ &$
!
!
!& #$$ &$
!
!
!&
#$$ &$
!
!
!& #$$ &$
!
!
!& #$$ &$
!
!
!&
#$$ &$
!
!
!& #$$ &$
5–43
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
DESTINAĆ
NOM DU SIGNAL (STATIF) ORIGINE SIGNAL NAME (GANTRY)
TION
A7 800PL3 - 5/B4 800PL2 - 1/G6 CPU:A7 (ADDRESS BUS G64)
A8 800PL3 - 5/B4 800PL2 - 1/G6 CPU:A8 (ADDRESS BUS G64)
A9 800PL3 - 5/B4 800PL2 - 1/G6 CPU:A9 (ADDRESS BUS G64)
A10 800PL3 - 5/B4 800PL2 - 1/G6 CPU:A10 (ADDRESS BUS G64)
A11 800PL3 - 5/B4 800PL2 - 1/G6 CPU:A11 (ADDRESS BUS G64)
ALIM_COMP (VCC: +5V) 800PL2 - 3/G2 800PL1 - 7/G4 COMP_SUPPLY (VCC:+5V)
ALIM_DIAPH_A (VCC: +5V) 800PL2 - 3/G2 800PL1 -16/H4 REAR DIAPH_SUPPLY_
ALIM_DIAPH_A_L (VCC: +5V) 800PL2 - 3/G2 800PL1 -19/B3 REAR_ &_ LAT_DIAPH SUPPLY
ALIM_DIAPH_L (VCC: +5V) 800PL2 - 3/G2 800PL1 -17/H4 LAT_DIAPH_SUPPLY
ALIM_DIAPH_V (VCC: +5V) 800PL2 - 3/B7 800PL1 -26/C2 FRONT_DIAPH_SUPPLY
ALIM_FILTRE (VCC: +5V) 800PL2 - 3/G2 800PL1 -15/H4 FILTER_SUPPLY
ALIM_FILTRE_GAINE(VCC
800PL2 - 3/G2 800PL1 -14/H4 FILTER_&_HOUSING_SUPPLY
+5V
ALIM_GAINE (VCC: +5V) 800PL2 - 3/G2 800PL1 -14/H4 HOUSING_POWER
ALIM_OK (VCC: +5V) 800PL1 - 1/A3 800PL2 -10/G5 POWER_SUPPLY_ACTIVE
ALIM_OK_A 800PL2 -10/D5 ...... POWER_SUPPLY_ACTIVE_A
ALIM_SAUV_+5V 800PL3 - 9/E2 ...... CPU:+5V_LITHIUM_BATTERY
ASC 800PL2 - 4/E3 ...... COLUMN_CMD
ASC_INACTIF 800PL1 - 1/A2 800PL2 -10/D4 COLUMN_POWER_INACTIVE
ASC_INACTIF_A 800PL2 -10/D3 ...... COLUMN_POWER_INACTIVE_A
AUTORISATION_STEREOTIX STEREOTIX 800PL2 -10/H2 STEREOTIX_AUTHORIZATION
AUTORISATION_STEREOĆ STEREOTIX_AUTHORIZAĆ
800PL2 -10/C2 ......
TIX_A TION_A
BATTERIE_OK 800PL3 - 9/A1 ...... CPU:BATTERY_OK
BOUCLAGE_OK 800PL1 - 1/A3 800PL2 -10/G4 LOOPS_OK
BOUCLAGE_OK_A 800PL2 -10/D4 ...... LOOPS_OK_A
SIGNAL LISTS
5–44
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
DESTINAĆ
NOM DU SIGNAL (STATIF) ORIGINE SIGNAL NAME (GANTRY)
TION
*!# ,
")*
#!
*!#* ,
""**
#!
*!*!#*
!#*!**
*!*!#*!
!#*!**!*
*!"* $!(* *
*!"*
$!(*
*
*!"*
$!(**
*#"#*
$#"#* *
*#"#*
$#"#*
*
*#"#*
$#"#**
*#"#* $#" $#"#** # $"
*%!!$
*
*%!!$*
**
**
**
**
**
#!$!
#*#!
#!
#!$!
#*#!
$
#!
#!$!
#*#!
$
#!$!*$$'
#*#!*
#!$!*$$'*
#*#!**
#!$!*
#*#!*
#!$!**
#*#!**
*"#! "#!#'
*"#!#'
SIGNAL LISTS
+ $ +
+ $ +
+ $+
*$
*$
*$
*$
* *
*" *
!*#!
!*!
* !
% *!!
#
*!$* &,
$*""* $""
!*!
$* %
*$
5–45
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
DESTINAĆ
NOM DU SIGNAL (STATIF) ORIGINE SIGNAL NAME (GANTRY)
TION
COMP_1 800PL2 -12/A3 ...... COMP_1
COMP_2 800PL2 -12/A3 ...... COMP_2
COMP_ON_1 800PL1 - 7/D3 ...... COMP_ON_1
COMP_ON_2 800PL1 - 7/D3 ...... COMP_ON_2
CONSIGNE_COMP_1 800PL2 - 5/A6 800PL1 - 5/H3 COMP_REF_PH_1
CONSIGNE_COMP_2 800PL2 - 5/A8 800PL1 - 6/H3 COMP_REF_PH_2
CONSIGNE_HTPM(COMMANĆ
800PL2 - 5/F1 800PL4 - 1/E8 PMHV_REFERENCE
DE DAC
CONSIGNE_TRAME 800PL2 - 5/A4 800PL1 -10/H2 GRID_REFERENCE
CONS_FORCE 800PL8 - 1/B3 800PL2 -12/H3 FORCE_REF
CONS_FORCE* 800PL8 - 1/B3 800PL2 -12/H3 FORCE_REF* (0V)
EN_STER (NOT USED) EN_STER (NOT USED)
ETA_ROT_OFF 800PL1 - 1/A3 800PL2 -10/H5 ROTAT_POWER_INACTIVE
ETA_ROT_OFF_A 800PL2 -10/D5 ...... ROTAT_POWER_INACTIVE_A
ETA_VERROU_OFF 800PL1 - 1/A3 800PL2 -10/H5 LOCK_POWER_INACTIVE
ETA_VERROU_OFF_A 800PL2 -10/D5 ...... ROTATION_POWER_INACTIVE_A
FOCALE_0 (660mm) 800PL5 - 2/G4 800PL2 - 9/H6 SID_0_ (660mm)
FOCALE_0* 800PL5 - 2/G4 800PL2 - 9/H6 SID_0* (0V)
FOCALE_0_A 800PL2 - 9/D8 ...... SID_0_A
FOCALE_1 (610mm) 800PL5 - 2/F4 800PL2 - 9/H6 SID_1_ (610mm)
FOCALE_1* 800PL5 - 2/F4 800PL2 - 9/H6 SID_1* (0V)
FOCALE_1_A 800PL2 - 9/D8 ...... SID_1_A
FOCALE_2 (560mm) 800PL5 - 2/F4 800PL2 - 9/H6 SID_2_ (560mm)
FOCALE_2* 800PL5 - 2/F4 800PL2 - 9/H5 SID_2* (0V)
FOCALE_2_A 800PL2 - 9/D8 ...... SID_2_A
FOCALE_3 (510mm) 800PL5 - 2/F4 800PL2 - 9/H5 SID_3_ (510mm)
SIGNAL LISTS
5–46
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
DESTINAĆ
NOM DU SIGNAL (STATIF) ORIGINE SIGNAL NAME (GANTRY)
TION
&)'## & !
)&$
#) !
! % )
#) ) ! % ) )
#)
!
! %
#)
) ! % )
)
#) !
! %
#)) ! % ))
)
!
!
#)%%)
)
) ! #)%%)
)
)
!
!
#)%%)
)
) ! #)%%)
)
) !
! #)%%)
) ) ! #)%%) )
)# !
! #)%%)*
)#) ! #)%%)*)
)! $)! %
!
! #)!)! $)!)
)! $)! %
!
! #)!)! $)!)
'
)! $)! %
) ! #)!)! $)!)
)
)! $)! % !
!
#)!)! $)!)
)! $)! % !
!
#)!)! $)!) '
)! $)! %) ! #)!)! $)!))
%$) ) ! !
!)&##%)#)'&)
%$)
) ! !
!)&##%)'&)
SIGNAL LISTS
)!$)
!
!)
) !)&##%)$
)!$) !
!)) !)&##%)$
#)&((# !
&((#)
$$)$ ! !
&)# &
$$)%#&# ! !
%)%#)# &
$$) ! ! !
!)# &
$$) $) ! ! !
!)#)# &
$$) $)%# ! !
#)#)# &
$$)#)! % ! !
#)!)# &
$$)#)# % ! !
# %)#)# &
$$)"& #% # !
)# &
$$) )%# ! !
#) )# &
5–47
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
DESTINAĆ
NOM DU SIGNAL (STATIF) ORIGINE SIGNAL NAME (GANTRY)
TION
""( "(#! !( "#(!$
""("#( (
!!( ("#(!$
""("#( (
#( ("#(!$
""("#(#!
#!("#(!$
""("#(
$"("#(!$
""(%!!$
(!$
"(# ( !" )
%( ("(
"$!
(
(
(
(
#("
$(#!
#("
#! $(#!
#("
#! $(#!
#( (
#( (
#( (
#( (
#(#!(
#!(#!(
#(#!(
#!(#!(
#((
$"(#!(
#((
$"(#!(
! )
#(#! !(#!
SIGNAL LISTS
#!
! )
#(#! !(#!
#!
((
( &!("$ '(
("(#
($( &!(#%
(#!$!(
(#( &!(#%
(#(!# (!##( &!(#%
(#(%!!$ (( &!(#%
($"(
($"(
( &(#!(
(!( &!(
("
($
(#!$!
(# #!
5–48
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
SIGNAL LISTS
$
##"# ###
$
##"# ###
# #
##
# #
##
# # ##
# # ##
##
!##
##
!##
##
!##
##
!##
5–49
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
DESTINAĆ
NOM DU SIGNAL (STATIF) ORIGINE SIGNAL NAME (GANTRY)
TION
! #*#* * ! !
%*! #*!* *
! #*#*
* ! !
%*! #*!*
*
! #*! $ !
!
!*! #*!*
! #*! $ !
!
!*! #*!* &
! #*! $ * !
!*! #*!* *
! #*! $
!
!
!*! #*!*
! #*! $
!
!
!*! #*!*
&
! #*! $
* !
!*! #*!*
*
! #*$" ! ! "*! #$
! '*$"* ! ! '"*"*
! '*$"* * !
! '"*"* *
*!"#*##$$ ! !
##$$*$$ "
!"#*$" !
!
"*!"#
!"#*$"* !
"*!"#*
!"#* !
!
##$*!"#*
!"#** !
!
##$*!"#*
!"#*%#$* ! !
!*%#$*!"#
!"#**# !
!
# *#%!!)*!"#
!"#**#$" !
#$" $(*#%!!)*!"#
!"#* *# !
%*# $*!"#
!"#* * ! !
!*# $*!"#
!"#* *"*" $ !
" $*"*# $*!"#
!"#* *&"" % !
*# $*!"#
!"#* !" !
!"$ "*!"#$
!"#* !* ! !
!* !"$ "*!"#
SIGNAL LISTS
5–50
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
DESTINAĆ
NOM DU SIGNAL (STATIF) ORIGINE SIGNAL NAME (GANTRY)
TION
#
#)+ "&#
#
#)" "&#
# #)" "&#
#+"!$+"+ "
" ##+"+"!$+#
#+"!$+"+ "
" ##+"+"!$+#+ '
#+"!$+"++ " ##+"+"!$+#++
#+"!$+"+ "
" %+"+"!$+#
#+"!$+"+ "
" %+"+"!$+# '
#+"!$+"++ " %+"+"!$+#+
#+"!$+"+' "
" #! %+"+"!$+#
#+"!$+%# "
" %#+"!$+#
#+"!$+%# "
" %#+"!$+#
#+"!$+%#+ " %#+"!$+#+
#+"!$+ "
" !&$ +"!$+#
#+"!$+ "
" !&$ +"!$+#
#+"!$+ + " !&$ +"!$+#+
#+"!$+"!% "
"
"+"!$+#
#+"!$+"!% "
"
"+"!$+# '
#+"!$+"!%+ "
"+"!$+#+
#$+!&"+"! "
"
# *+$%!"+#*
#$+$&# "
"
"'+"+$+
#%!&#+$$+ %#&# " "
%+#!& +#%&#
#%!&#+$$+!"+%# " "
#+!"+#!& +#%&#
##+"+#!& +#,
#%!&#+$$+"+ " "
%&#
#%!&#+$$+%#+ " "
%#+#!& +#%&#
SIGNAL LISTS
%&+#+$%#+ +#,
#%!&#+$$+# "
"
%&#
#%!&#+$$+(
"
" (
+#!& +#%&#
#%!&#+$$+(
"
" (
+#!& +#%&#
#%!&#+'##!&+# + " "
#!%+#+!+!&
#)" #
" "
#)" "% %'
#)" # " "
#)" "% %'
#)+ #
" " #)+ !"%$#
5–51
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
DESTINAĆ
NOM DU SIGNAL (STATIF) ORIGINE SIGNAL NAME (GANTRY)
TION
%!%% !%%%!"
%!%% !%% %!"
%!%%%
%%!%%!"
%!%%%
%%!% %!"
%!%%%
!%%!%%!"
%!%%% !%%!% %!"
%!%%#%
!%%!%%!"
%!%%#%
!%%!% %!"
%!%!%
!%!%%!"
%!%!% !%!% %!"
%!%%
" %!%%!"
%!%% " %!% %!"
%%
%"%!"
%!%!%&
%%!"
!"
%%
%%!"
%%%!
%!!%%!"
%%! %%!"
%%#"
%%!"
%% !$%&
% !$%%
!"
% !$%%
!%% !$%!"
!%% !$%&
% !$%%#
!"
% !$%!
!% !$%!"
% !$%
" % !$%!"
"!% ! " % !$
SIGNAL LISTS
5–52
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
DESTINAĆ
NOM DU SIGNAL (STATIF) ORIGINE SIGNAL NAME (GANTRY)
TION
"#!#))"
"#))"#!#)$)
"#!#))
"#))"#!#)
"#!#))!)!#
"#))"#!#)!#)!
"#!#))%!!$
"#))"#!#)
"(!)!"
!)"(!
#
#') $#
#
#' $#
#"#))"
")"$ ()#"#
#"#)$#)"
#"#)$)"# "
#!
!
#' #
#' ##%
#'
#' ##%
#') #
#')) #"!
$)#!
$)!
%)"
$)%
%)
)%
%)!)!#
!##)!)%
%)%#)
)%#)%
%)%!!$
)%
%
%
%
%
%!!$
)
%!!$
%!!$
%
%
)$#)")") &! #
)$)&!)"# )
SIGNAL LISTS
)$#)")$#) $ ! #
)$)$ !)"# )
)$#)
%
)
%)!
)$#)
%
)
%)!
)$#)
%
)
%)!
)
$) # $"
)
$) # $"
)
$) # $"
)
$) # $"
)
$)
# $"
)
$)
# $"
5–53
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
DESTINAĆ
NOM DU SIGNAL (STATIF) ORIGINE SIGNAL NAME (GANTRY)
TION
#
!#
!
#
!# !
#
!# !
#
!#
!
#
!# !
#
!# !
#
!# !
#### # # ##
####
# # #
# "#
## "
SIGNAL LISTS
5–54
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
CONTENTS
ABBREVIATIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6–1
6–1 LOCATION OF SUBASSEMBLIES . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6–2
6–2 LOCATION OF BOARDS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6–4
6–3 SENOGRAPHE DMR 36005705 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6–7
6–4 GENERATOR CABINET 36005958 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6–9
6–5 GENERATOR CABINET SEMI–ASSEMBLED 36002118 . . . . . . . . . . . . . . . . . . . . . . . . . 6–11
6–6 701–HV UNIT ASSEMBLY 45440275 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6–13
6–7 200–GENERATOR LINE SUPPLY UNIT 45440274 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6–15
6–8 400–GENERATOR CONTROL PLATE 36005963 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6–19
6–9 300–GENERATOR POWER PLATE 45440269 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6–21
6–10 GENERATOR CABLE ASSEMBLY SET 45441072 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6–23
6–11 100–CONSOLE CDRH 45561630 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6–25
6–12 SET OF COVERS 45560678 – COVERS MOUNTING HARDWARE 45560702 . . . . . . . 6–27
6–13 800–GANTRY 36005747 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6–29
6–14 800–COLUMN UNIT 45559482 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6–31
6–15 SET OF CABLES, COLUMN 45560239 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6–33
6–16 800–ASSEMBLED GANTRY BOARDS 2102287 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6–35
6–17 800–READOUT UNIT 36005259 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6–37
6–18 800–SET OF CABLES, GANTRY 45560263 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6–39
6–19 800–MAIN HARNESS 45560261 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6–41
6–20 800–EXAMINATION ARM 45560086 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6–43
6–21 800–EXAMINATION ARM 45560086 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6–45
6–22 800–EXAMINATION ARM 45560086 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6–49
6–23 800–EXAMINATION ARM 45560086 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6–51
RENEWAL PARTS
6–i
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
6–ii
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
ABBREVIATIONS
ITEM
NO.
– Not illustrated.
1 Critical.
2 Not critical.
N Not available.
REP
Y Repairable.
QTY
Viewing direction.
RENEWAL PARTS
6–1
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
ILLUSTRATION 6–1
LOCATION OF SUBASSEMBLIES
SENO DMR
ILL. 4–3
701 – HV UNIT
ASSEMBLY
ILL. 6–6
200 – GENERATOR
LINE SUPPLY UNIT
ILL. 6–7
400 – GENERATOR
CONTROL PLATE
ILL. 6–8
300 – GENERATOR
POWER PLATE
ILL. 6–9
GENERATOR ACCESSORIES
CABLE ASSEMBLY 100–CONSOLE ILL. 6–32
SET CDRH
ILL. 6–10 ILL. 6–11
SET OF COVERS
ILL. 6–12
COVERS MOUNTING
HARDWARE
ILL. 6–12
RENEWAL PARTS
GENERATOR CABINET
ILL. 6–4
GENERATOR CABINET
SEMI–ASSEMBLED
ILL. 6–5
6–2
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
800–GANTRY
ILL. 6–13
SET OF HANDLES
ILL. 6–31
800–READOUT UNIT
ILL. 6–17
PADDLE FASTENER
ASSEMBLY
800–COLUMN UNIT ILL. 6–28
ILL. 6–14
SET OF CABLES,
COLUMN IMAGE RECEPTOR
ILL. 6–15 ILL. 6–26
ELECTROMAGNET PHOTOCELL
MOUNT ASSEMBLY ASSEMBLY
ILL. 6–29 ILL. 6–27
800–ASSEMBLED 800–EXAMINATION
GANTRY BOARDS ARM
ILL. 6–16 ILL. 6–20 thru 6–23
SET OF CABLES,
COMPRESSION ARM
800–MAIN HARNESS ILL. 6–30
ILL. 6–19
RENEWAL PARTS
800–SET OF CABLES,
GANTRY
ILL. 6–18
6–3
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
ILLUSTRATION 6–2
LOCATION OF BOARDS
6–4
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
800–PL16 INTER–ARM–LOCK
COMMAND BOARD
800–PL12 GANTRY
READOUT
BOARD
800–PL9 ENCODER
800–PL5 COMPRESSION BOARD
ARM
DISTRIBUTION 800–PL8 PADDLE
BOARD BOARD
800–PL14 MAGNIFICATION
SENSOR 800–PL15 INTER
BOARD ARM TILT
SENSOR
800–PL7 SID BOARD
SENSOR
BOARD
800–PL10 IMAGE
800–PL3 GANTRY RECEPTOR
CPU BOARD
BOARD
800–PL4 PHOTOCELL
BOARD
6–5
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
ILLUSTRATION 6–3
SENOGRAPHE DMR 36005705
4 2
3
6
RENEWAL PARTS
6–6
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
ILLUSTRATION 6–3
SENOGRAPHE DMR 36005705
A
ITEM
PART NO. FRU REP DESCRIPTION QTY P
NO.
P
1 2 3 4 5
– N SENOGRAPHE DMR 1
6–7
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
ILLUSTRATION
GENERATOR CABINET 36005958
6 6
1 7
8
6
6
4
5
3 2
RENEWAL PARTS
6–8
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
ILLUSTRATION 6–4
GENERATOR CABINET 36005958
A
ITEM
PART NO. FRU REP DESCRIPTION QTY P
NO.
P
1 2 3 4 5
– N GENERATOR CABINET PL
RENEWAL PARTS
6–9
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
ILLUSTRATION 6–5
GENERATOR CABINET, SEMI–ASSEMBLED 36002118
1 2
3
RENEWAL PARTS
6–10
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
ILLUSTRATION 6–5
GENERATOR CABINET SEMI–ASSEMBLED 36002118
A
ITEM
PART NO. FRU REP DESCRIPTION QTY P
NO.
P
1 2 3 4 5
RENEWAL PARTS
6–11
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
ILLUSTRATION 6–6
701–HV UNIT ASSEMBLY 45440275
RENEWAL PARTS
1 2 3
701–PL1
6–12
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
ILLUSTRATION 6–6
701–HV UNIT ASSEMBLY 45440275
A
ITEM
PART NO. FRU REP DESCRIPTION QTY P
NO.
P
1 2 3 4 5
RENEWAL PARTS
6–13
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
ILLUSTRATION 6–7
200–GENERATOR LINE SUPPLY UNIT 45440274
4
5
7 3
17
18 6
19
2
200–PL1
26
T1
27
28
14
D25
14
D24
200 PL2 XJ2
15
25 D26
W12
15
D27
24
W10 16
200–PL2
20
F1 13
RENEWAL PARTS
21
F2
12
22
F3
10
9 11
6–14
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
ILLUSTRATION 6–7
200–GENERATOR LINE SUPPLY UNIT 45440274
A
ITEM
PART NO. FRU REP DESCRIPTION QTY P
NO.
P
1 2 3 4 5
6–15
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
LLUSTRATION 6–7
200–GENERATOR LINE SUPPLY UNIT 45440274 (Continued)
4
5
7 3
17
18 6
19
2
200–PL1
26
T1
27
28
14
D25
14
D24
200 PL2 XJ2
15
25 D26
W12
15
D27
24
W10 16
200–PL2
20
F1 13
RENEWAL PARTS
21
F2
12
22
F3
10
9 11
6–16
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
ILLUSTRATION 6–7
200–GENERATOR LINE SUPPLY UNIT 45440274 (Continued)
A
ITEM
PART NO. FRU REP DESCRIPTION QTY P
NO.
P
1 2 3 4 5
RENEWAL PARTS
6–17
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
ILLUSTRATION 6–8
400–GENERATOR CONTROL PLATE 36005963
A B
400–PL3
400–PL1 400–PL2
14
13 12
A B
1 4
2
3 8
or
15
400–PL3
14 7
RENEWAL PARTS
400–PL1 400–PL2
6 5
6–18
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
ILLUSTRATION 6–8
400–GENERATOR CONTROL PLATE 36005963
A
ITEM
PART NO. FRU REP DESCRIPTION QTY P
NO.
P
1 2 3 4 5
RENEWAL PARTS
6–19
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
ILLUSTRATION 6–9
300–GENERATOR POWER PLATE 45440269
8 20
19 10 21
20 18 11 22
300–PL1
17
15
16
300–PL2
D1 D2 D3
A
300–PL3
6 6 5 7 12 13 14 9 21 25 5 2 4
10 300–PL1 26 6 300–PL2 300–PL3
11 27 7
B
1
RENEWAL PARTS
23 23 22 24 B A
6–20
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
ILLUSTRATION 6–9
300–GENERATOR POWER PLATE 45440269
A
ITEM
PART NO. FRU REP DESCRIPTION QTY P
NO.
P
1 2 3 4 5
RENEWAL PARTS
6–21
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
ILLUSTRATION 6–10
GENERATOR CABLE ASSEMBLY SET 45441072
2
4
5
6
1
RENEWAL PARTS
6–22
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
ILLUSTRATION 6–10
GENERATOR CABLE ASSEMBLY SET 45441072
A
ITEM
PART NO. FRU REP DESCRIPTION QTY P
NO.
P
1 2 3 4 5
RENEWAL PARTS
6–23
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
ILLUSTRATION 6–11
100–CONSOLE CDRH 45561630
9 A 11
12
10
8
3
100–PL1
8 7 2
100–PL2
A B
RENEWAL PARTS
6–24
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
ILLUSTRATION 6–11
100–CONSOLE CDRH 45561630
A
ITEM
PART NO. FRU REP DESCRIPTION QTY P
NO.
P
1 2 3 4 5
– 1 N S CONSOLE
2 45561530 1 Y S S 100–PL2 CONSOLE CONTROL BOARD 1
RENEWAL PARTS
6–25
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
ILLUSTRATION 6–12
SET OF COVERS 45560678 – COVERS MOUNTING HARDWARE 45560702
2 1
11
5 6
3 4
RENEWAL PARTS
12
13 7 8
10
6–26
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
ILLUSTRATION 6–12
SET OF COVERS 45560678 – COVERS MOUNTING HARDWARE 45560702
A
ITEM
PART NO. FRU REP DESCRIPTION QTY P
NO.
P
1 2 3 4 5
– N SET OF COVERS PL
RENEWAL PARTS
6–27
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
ILLUSTRATION 6–13
800–GANTRY 36005747
7
8
2 6
RENEWAL PARTS
6–28
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
ILLUSTRATION 6–13
800–GANTRY 36005747
A
ITEM
PART NO. FRU REP DESCRIPTION QTY P
NO.
P
1 2 3 4 5
– N 800–GANTRY PL
RENEWAL PARTS
6–29
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
ILLUSTRATION 6–14
800–COLUMN UNIT 45559482
A B
24
A
4 24
5
6
2
3
B 19
800–PL13
7 17 20
8 25
26
21
22
23
RENEWAL PARTS
12
13
14 10
15 1 16 11 18
6–30
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
ILLUSTRATION 6–14
800–COLUMN UNIT 45559482
A
ITEM
PART NO. FRU REP DESCRIPTION QTY P
NO.
P
1 2 3 4 5
– N 800–COLUMN UNIT PL
RENEWAL PARTS
6–31
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
ILLUSTRATION 6–15
SET OF CABLES, COLUMN 45560239
2
RENEWAL PARTS
6–32
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
ILLUSTRATION 6–15
SET OF CABLES, COLUMN 45560239
A
ITEM
PART NO. FRU REP DESCRIPTION QTY P
NO.
P
1 2 3 4 5
RENEWAL PARTS
6–33
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
ILLUSTRATION 6–16
800–ASSEMBLED GANTRY BOARDS 2102287
2 1
800 PL–3 800 PL–4
5
800 PL–2
6 7
8
800 PL–1
RENEWAL PARTS
6–34
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
ILLUSTRATION 6–16
800–ASSEMBLED GANTRY BOARDS 2102287
A
ITEM
PART NO. FRU REP DESCRIPTION QTY P
NO.
P
1 2 3 4 5
RENEWAL PARTS
6–35
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
ILLUSTRATION 6–17
800–READOUT UNIT 36005259
2
RENEWAL PARTS
6–36
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
ILLUSTRATION 6–17
800–READOUT UNIT 36005259
A
ITEM
PART NO. FRU REP DESCRIPTION QTY P
NO.
P
1 2 3 4 5
RENEWAL PARTS
6–37
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
ILLUSTRATION 6–18
800–SET OF CABLES, GANTRY 45560263
V1
V2
4
RENEWAL PARTS
6–38
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
ILLUSTRATION 6–18
800–SET OF CABLES, GANTRY 45560263
A
ITEM
PART NO. FRU REP DESCRIPTION QTY P
NO.
P
1 2 3 4 5
RENEWAL PARTS
6–39
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
ILLUSTRATION 6–19
800–MAIN HARNESS 45560261
10 B
7 5 4
W10 W8 W7
8
W59
9 6
W60 W9
2
1
RENEWAL PARTS
6–40
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
ILLUSTRATION 6–19
800–MAIN HARNESS 45560261
A
ITEM
PART NO. FRU REP DESCRIPTION QTY P
NO.
P
1 2 3 4 5
– N 800–MAIN HARNESS PL
RENEWAL PARTS
6–41
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
ILLUSTRATION 6–20
800–EXAMINATION ARM 45560086
2
3
4
5
11
12
13
14
6
7
8
9
15
16
17
18
19
RENEWAL PARTS
6–42
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
ILLUSTRATION 6–20
800–EXAMINATION ARM 45560086
A
ITEM
PART NO. FRU REP DESCRIPTION QTY P
NO.
P
1 2 3 4 5
– N 800–EXAMINATION ARM PL
RENEWAL PARTS
6–43
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
ILLUSTRATION 6–21
800–EXAMINATION ARM 45560086
A
20
21
22 C
C D
B
19
23 25
24
D
7 8 10 9 B
26
1 11
12
13
5 3 6 4 14
RENEWAL PARTS
16
800–PL7
15
17
18
6–44
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
ILLUSTRATION 6–21
800–EXAMINATION ARM 45560086
A
ITEM
PART NO. FRU REP DESCRIPTION QTY P
NO.
P
1 2 3 4 5
– N EXAMINATION ARM PL
6–45
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
ILLUSTRATION 6–21
800–EXAMINATION ARM 45560086 (Continued)
A
20
21
22 C
C D
B
19
23 25
24
D
7 8 10 9 B
26
1 11
12
13
5 3 6 4 14
RENEWAL PARTS
16
800–PL7
15
17
18
6–46
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
ILLUSTRATION 6–21
800–EXAMINATION ARM 45560086 (Continued)
A
ITEM
PART NO. FRU REP DESCRIPTION QTY P
NO.
P
1 2 3 4 5
– N S COMPRESSION BELT 1
– N S BENT PLATE 1
– N S FLAT WASHER Z4N PASS SST 2
RENEWAL PARTS
6–47
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
ILLUSTRATION 6–22
800–EXAMINATION ARM 45560086
9
800–PL5
10
11
A
C
15 16
D
14
B
8 3 1
800–PL11 4 2
5 13 12
RENEWAL PARTS
6 13
7
12
6–48
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
ILLUSTRATION 6–22
800–EXAMINATION ARM 45560086
A
ITEM
PART NO. FRU REP DESCRIPTION QTY P
NO.
P
1 2 3 4 5
– N 800–EXAMINATION ARM PL
6–49
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
ILLUSTRATION 6–23
800–EXAMINATION ARM 45560086
5 2 3 1
C D
9
800–PL15
10
11
7
800–PL14
RENEWAL PARTS
6–50
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
ILLUSTRATION 6–23
800–EXAMINATION ARM 45560086
A
ITEM
PART NO. FRU REP DESCRIPTION QTY P
NO.
P
1 2 3 4 5
– N 800–EXAMINATION ARM PL
1 N S PIN 1
2 N S OFF–CENTERED SHAFT 1
3 N S HEX NUT M8 PASS SST 2
4 N S EXT CIRCLIP D=6 TYPE 81 2
5 99146577 1 S BALL BEARING 6X19X6 2
6 N S SPACER 2
7 45553870 1 S 800–PL14 MAGNIFICATION SENSOR BOARD 1
8 N S HEX SKT HD CAP SCREW M3X8/8 PASS SST 2
9 45554408 1 S 800–PL15 INTER ARM TILT SENSOR BOARD 1
10 N S SPACER 2
11 N S HEX SKT HD CAP SCREW M3X8/8 PASS SST 2
RENEWAL PARTS
6–51
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
ILLUSTRATION 6–24
TUBE HOUSING SPACER 2101067
A B
35
37
10
38
35
36
28
33
800–PL6
29
A 4 30
31
6 32
4
7
3 8
9
11
12
13
B
C 1
21 22
20 24
25
5
RENEWAL PARTS
16 26
17
27 14 15 23
18
19
6–52
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
ILLUSTRATION 6–24
TUBE HOUSING SPACER 2101067
A
ITEM
PART NO. FRU REP DESCRIPTION QTY P
NO.
P
1 2 3 4 5
6–53
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
ILLUSTRATION 6–25
SET OF CABLES, TUBE HOUSING SPACER 2109334
3
W49
1
W36
2
W48
4
W53
5
W56
RENEWAL PARTS
6
W75
W76
6–54
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
ILLUSTRATION 6–25
SET OF CABLES, TUBE HOUSING SPACER 2109334
A
ITEM
PART NO. FRU REP DESCRIPTION QTY P
NO.
P
1 2 3 4 5
RENEWAL PARTS
6–55
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
ILLUSTRATION 6–26
IMAGE RECEPTOR 45554608
A B 25
26
15 29 27
28
24
2
3
C
ÉÉ
B 16
É
11
800–PL10
21
13
14
C
35
1 4 5
6
9 32
30 10 7 13 33
31 12 18 17 8 14 34
22
23
RENEWAL PARTS
A
30
31
20
6–56
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
ILLUSTRATION 6–26
IMAGE RECEPTOR 45554608
A
ITEM
PART NO. FRU REP DESCRIPTION QTY P
NO.
P
1 2 3 4 5
– N IMAGE RECEPTOR PL
6–57
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
ILLUSTRATION 6–27
PHOTOCELL ASSEMBLY 2102432
1
RENEWAL PARTS
6–58
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
ILLUSTRATION 6–27
PHOTOCELL ASSEMBLY 2102432
A
ITEM
PART NO. FRU REP DESCRIPTION QTY P
NO.
P
1 2 3 4 5
1 45067100 1 S PHOTOMULTIPLIER 1
2 N S PLUG 1
3 N S STOPPER 1
RENEWAL PARTS
6–59
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
ILLUSTRATION 6–28
PADDLE FASTENER ASSEMBLY 45560085
A
19
2 3 4 1 5 6
10
7 9 20
8
18
11
15 16
16 17
RENEWAL PARTS
800–PL9 800–PL8
13 12
14 14
6–60
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
ILLUSTRATION 6–28
PADDLE FASTENER ASSEMBLY 45560085
A
ITEM
PART NO. FRU REP DESCRIPTION QTY P
NO.
P
1 2 3 4 5
1 N S COMMAND SHAFT 1
2 N S SELF–LUBRICATING BUSHING D=6X10X6 BRONZE 4
3 N S WASHER ”BELLEVILLE” 12.5X6.2X0.5 STL 10
4 N S EXT CIRCLIP D=6 TYPE 81 2
5 36005384 1 S COMMAND BUTTON 2
6 N S HEX SKT HDLS SCREW CUP END M3X8/8 PASS SST 2
7 45554615 1 S ENCODER SUPPORT GLUED 1
8 N S HDLS SCREW M3X4 BLACK STL CUP END 1
9 N S PCB SUPPORT 1
10 N S HEX SKT HD CAP SCREW M4X8/8 PASS SST 2
11 91671876 2 S MAGNET D=4X20 1
12 45561125 2 S PADDLE FASTENER RIGHT COVER 1
13 45561126 2 S PADDLE FASTENER LEFT COVER 1
14 N S FLSTR HD SCREW M4X10/10 PASS SST 2
15 45554409 1 S 800–PL9 ENCODER BOARD 1
16 N S HEX SCREW M3X8 PASS SST 5
17 45554410 1 S 800–PL8 PADDLE BOARD 1
18 45553536 1 S CABLE (PADDLE FINE–MOVEMENT) 1
19 N S FLAT WASHER Z6 U GALV STL 4
20 N S FLAT WASHER M4 N PASS SST 2
RENEWAL PARTS
6–61
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
ILLUSTRATION 6–29
ELECTROMAGNET MOUNT ASSEMBLY 45560090
7 4 8 9
6 5 3 1 10
RENEWAL PARTS
6–62
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
ILLUSTRATION 6–29
ELECTROMAGNET MOUNT ASSEMBLY 45560090
A
ITEM
PART NO. FRU REP DESCRIPTION QTY P
NO.
P
1 2 3 4 5
RENEWAL PARTS
6–63
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
ILLUSTRATION 6–30
SET OF CABLES, COMPRESSION ARM 45560231
3
W39
1
W37
2
W38
6
W43
7
W44
8
W45
RENEWAL PARTS
9
W47
6–64
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
ILLUSTRATION 6–30
SET OF CABLES, COMPRESSION ARM 45560231
A
ITEM
PART NO. FRU REP DESCRIPTION QTY P
NO.
P
1 2 3 4 5
RENEWAL PARTS
6–65
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
ILLUSTRATION 6–31
SET OF HANDLES 45475381
1 1
2 2
4 4
RENEWAL PARTS
3 3
6–66
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
ILLUSTRATION 6–31
SET OF HANDLES 45475381
A
ITEM
PART NO. FRU REP DESCRIPTION QTY P
NO.
P
1 2 3 4 5
– 1 SET OF HANDLES PL
RENEWAL PARTS
6–67
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
ILLUSTRATION 6–32
ACCESSORIES 45560675
3
4
5 24
6 25
29 31
30 32
9
13
8 16 10 12 26
14
21
17
28
RENEWAL PARTS
20
6–68
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
ILLUSTRATION 6–32
ACCESSORIES 45560675
A
ITEM
PART NO. FRU REP DESCRIPTION QTY P
NO.
P
1 2 3 4 5
– – ACCESSORIES PL
6–69
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
ILLUSTRATION 6–32
ACCESSORIES 45560675 (Continued)
3
4
5 24
6 25
29 31
30 32
9
13
8 16 10 12 26
14
21
17
28
RENEWAL PARTS
20
6–70
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
ILLUSTRATION 6–32
ACCESSORIES 45560675 (Continued)
A
ITEM
PART NO. FRU REP DESCRIPTION QTY P
NO.
P
1 2 3 4 5
– N S S JACK SUPPORT 4
– N S S COVER 7
– 45559940 1 S S PLASTIC COVER 30
– N S S HEX SKT HD CAP SCREW M10X80/80 BLACK STL 3
– N S S FLAT WASHER M10 N PASS SST 3
– N S S GRID ADJUSTMENT TOOL 1
– N S S WASHER 8
– N S S STABILIZER 4
– N S S EXPANSION BOLT M10 3
Note: Recommended glue for bucky plastic corners is: Loctite 406
RENEWAL PARTS
6–71
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
ILLUSTRATION 6–33
EMERGENCY MAINTENANCE KIT 45560671
RENEWAL PARTS
6–72
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
ILLUSTRATION 6–33
EMERGENCY MAINTENANCE KIT 45560671
A
ITEM
PART NO. FRU REP DESCRIPTION QTY P
NO.
P
1 2 3 4 5
– 45560509 N S COVER 6
– 45066644 N S HALOGEN LAMP 12V 100W 1
– 58079627 N S FAST FUSE 6X32 2.0A 250V C 100A ULL 2
– 58079166 N S FAST FUSE 6X32 10A 380V C 10KA ULL 2
– 58079685 N S FAST FUSE 6X32 0.5A 250V C 35A ULL 5
6–73
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
ILLUSTRATION 6–33
EMERGENCY MAINTENANCE KIT 45560671 (Continued)
RENEWAL PARTS
6–74
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
ILLUSTRATION 6–33
EMERGENCY MAINTENANCE KIT 45560671 (Continued)
A
ITEM
PART NO. FRU REP DESCRIPTION QTY P
NO.
P
1 2 3 4 5
RENEWAL PARTS
6–75
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
ILLUSTRATION 6–34
OPTIONAL ACCESSORIES
13
14
16
4 2
19
18
3
RENEWAL PARTS
6–76
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
ILLUSTRATION 6–34
OPTIONAL ACCESSORIES
A
ITEM
PART NO. FRU REP DESCRIPTION QTY P
NO.
P
1 2 3 4 5
– OPTIONAL ACCESSORIES
– 7 N S ACCESSORY TROLLEY 1
8 45554823 1 S S ACCESSORY TROLLEY WHEEL 4
9 N S STEREOTIX CDRH 1
6–77
GE Medical Systems SENOGRAPHE DMR (V1_P)
REV 0 sm 2308529–100
Blank page
RENEWAL PARTS
6–78
Technical
Publications
2308529–100
Revision 0
do not duplicate