Vous êtes sur la page 1sur 5

IEC 61010-2-032:2019/COR1:2020 –1–

 IEC 2020

INTERNATIONAL ELECTROTECHNICAL COMMISSION


COMMISSION ÉLECTROTECHNIQUE INTERNATIONALE
____________

IEC 61010-2-032 IEC 61010-2-032


Edition 4.0 2019-06 Édition 4.0 2019-06

SAFETY REQUIREMENTS FOR ELECTRICAL EXIGENCES DE SÉCURITÉ POUR APPAREILS


EQUIPMENT FOR MEASUREMENT, CONTROL ÉLECTRIQUES DE MESURAGE, DE RÉGULATION
AND LABORATORY USE – ET DE LABORATOIRE –

Part 2-032: Particular requirements for HAND ­ HELD Partie 2-032: Exigences particulières pour les
and hand-manipulated current sensors for capteurs de courant, PORTATIFS et manipulés
electrical test and measurement manuellement, pour essai électrique et mesurage

CORRIGENDUM 1

Table 101 – C LEARANCES and CREEPAGE Tableau 101 – D ISTANCES D ' ISOLEMENT et
DISTANCES for measuring circuit LIGNES DE FUITE des BORNES d'un circuit
TERMINALS with HAZARDOUS LIVE de mesure ayant des parties
conductive parts up to 1 000 V a.c. or conductrices sous TENSION DANGEREUSE
1 500 V d.c. qui peuvent atteindre 1 000 V en
courant alternatif ou 1 500 V en courant
continu

Replace the content "≥ 30 ≥ 300" of the Remplacer le contenu "≥ 30 ≥ 300" de la
cell located in the first column, second cellule située à la première colonne,
row, with the following content: deuxième ligne, par le contenu suivant:
"≥ 30 ≤ 300". "≥ 30 ≤ 300".

Table K.105 – Minimum values for Tableau K.105 – Valeurs minimales des
distance or thickness of solid distances ou de l'épaisseur de
insulation in measuring circuits RATED l'isolation solide dans les circuits de
FOR MEASUREMENT CATEGORIES III and IV mesure dont les CATÉGORIES DE MESURES
III et IV sont des CARACTÉRISTIQUES
ASSIGNÉES

Replace the title of Table K.105 with the Remplacer le titre du Tableau K.105 par le
following title: titre suivant:
IEC 61010-2-032:2019-06/COR1:2020-02(en-fr)

Table K.105 – Minimum values for Tableau K.105 – Valeurs minimales des
distance or thickness of solid distances ou de l'épaisseur de
insulation in measuring circuits RATED l'isolation solide dans les circuits de
FOR MEASUREMENT CATEGORIES mesure dont les CATÉGORIES DE MESURES
sont des CARACTÉRISTIQUES ASSIGNÉES

Figure DD.1 – Requirements for Figure DD.1 – Exigences relatives à la


CLEARANCE , CREEPAGE DISTANCE and solid DISTANCE D ' ISOLEMENT , à la LIGNE DE FUITE
insulation et à l'isolation solide
–2– IEC 61010-2-032:2019/COR1:2020
 IEC 2020
Replace in the last column on the right, 4th box, the text "Moulded and potted parts, PCBs,
thin films: Table K.14" with "Moulded and potted parts, PCBs, thin films: Table K.105" as
follows:
IEC 61010-2-032:2019/COR1:2020 –3–
 IEC 2020
Key

ACC R ATED altitude correction of CLEARANCE

ACV Site altitude correction of test voltage

& Both required

ACM Alternative conformity means


As applicable

Optional test path

Figure DD.1 – Requirements for CLEARANCE , CREEPAGE DISTANCE and solid insulation
–4– IEC 61010-2-032:2019/COR1:2020
 IEC 2020
Remplacer dans la dernière colonne à droite, 4 e symbole, le texte "Pièces moulées et
empotées, circuits imprimés (PCB), couches minces: Tableau K.14" par "Pièces moulées et
empotées, circuits imprimés (PCB), couches minces: Tableau K.105", comme suit:
IEC 61010-2-032:2019/COR1:2020 –5–
 IEC 2020
Légende

ACC Correction de la DISTANCE D ' ISOLEMENT suivant l'altitude ASSIGNEE (R ATED altitude correction of
CLEARANCE )

ACV Correction de la tension d'essai suivant l'altitude du site (Site altitude correction of test voltage)

& Les deux corrections sont exigées

ACM Autres méthodes de conformité (Alternative conformity means)

Selon le cas

Chemin d'essai facultatif

Figure DD.1 – Exigences relatives à la DISTANCE D ' ISOLEMENT , à la LIGNE DE FUITE et à


l'isolation solide