Académique Documents
Professionnel Documents
Culture Documents
2
Erreur de la sonde de température (Pt100) /°C
εPt100 /°C
0 -EMTTHR /°C
+EMTTHR /°C
-1
-2
-3
-20 0 20 40 60 80 100 120
consigne /°C Valeur conventionnelle /°C lTHR / °C -EMTTHR /°C +EMTTHR /°C
0 -0.042 -2.9 -2.5 2.5
20 20.0782 17.14 -2.5 2.5
40 40.181 37.3 -2.5 2.5
60 60.298 57.5 -2.5 2.5
80 80.4478 77.7 -2.5 2.5
100 100.575 97.96 -2.5 2.5
étalon
Uétalon 0.076
uétalon 0.038
après ajustage
εTHR /°C S(THR) /°C urepTHR /°C uTHR/°C UTHR/°C lPt100 /°C εPt100 /°C
-2.8578 0.00E+00 0.00E+00 3.80E-02 7.60E-02 0.1 -0.1422
-2.9382 5.48E-02 2.45E-02 4.52E-02 9.04E-02 20.1 -0.0218
-2.881 0.00E+00 0.00E+00 3.80E-02 7.60E-02 40.1 0.081
-2.798 0.00E+00 0.00E+00 3.80E-02 7.60E-02 60.1 0.198
-2.7478 0.00E+00 0.00E+00 3.80E-02 7.60E-02 80.1 0.3478
-2.615 5.48E-02 2.45E-02 4.52E-02 9.04E-02 100.18 0.395
ge
1
Erreur du thermocouple /°C
0
+EMTTHR /°C
-EMTTHR /°C
εTHR /°C
Linear (εTHR /°C)
-1
-2
-4
-20 0 20 40 60 80 100 120
2
Erreur corrigée du thermocouple /°C
εcorrigé/°C
-EMTTHR /°C
0
+EMTTHR /°C
-1
-2
-3
-4
-20 0 20 40 60 80 100 120