FICHE DE CONTRÔLE
MONITORING
LOCATION : PARC DE PRET

ETABLISSEMENT : HEMAC MEDICAL

SERVICE : PRET

FABRICANT : NIHON KOHDEN N° DE SERIE : 01282

MODELE : PVM-2701 N° INTERNE :

MATERIELS DE TEST

X FLUKE PROSIM 8 NS 3681053 X FLUKE ESA 620 NS 1019011

CONTRÔLES GENERAUX

ETAT
NON
POINTS DE CONTRÔLE VALIDE NA DEGRAD
VALIDE
É
ETAT GENERAL X
ETAT DES CONNECTEURS D'ACCESSOIRES X
ETAT DES CONNECTEURS D'ALIMENTATION ELECTRIQUE X
ETAT DES CORDONS ET DES ACCESSOIRES X
ETAT CORDON SECTEUR - ALIMENTATION EXTERNE X

POINTS DE CONTRÔLE NON


VALIDE NA
VALIDE
DEROULEMENT DE L'AUTOTEST X
FONCTIONNEMENT SUR SECTEUR X
FONCTIONNEMENT SUR BATTERIE - PILE X
FONCTIONNEMENT ECRAN - AFFICHEUR - VOYANT X
FONCTIONNEMENT BOUTONS - CLAVIERS - DALLE TACTILE X
HORLOGE INTERNE X

POINTS DE CONTRÔLE DE SECURITE NON


VALIDE NA
VALIDE
FONCTIONNEMENT PASSAGE SUR BATTERIE X
ALARME DECHARGE - NIVEAU BATTERIE X
FONCTIONNEMENT DE L'ALARME SONORE ET VISUELLE X

CONTRÔLES DES FONCTIONS

NON NON
POINTS DE CONTRÔLE ECG VALIDE NA POINTS DE CONTRÔLE RESPIRATION VALIDE NA
VALIDE VALIDE

PRISE DE MESURE DE LA FREQUENCE CARDIAQUE X PRISE DE MESURE DE LA FREQUENCE RESPIRATOIRE


QUALITE D'AFFICHAGE DU TRACE X QUALITE D'AFFICHAGE DU TRACE
SELECTION DES DERIVATIONS ECG X SELECTION DES DERIVATIONS RESP
ALARME DE SEUILS ET D'ARYTHMIE X ALARME DE SEUILS
ALARME DEBRANCHEMENT ACCESSOIRE X ALARME D'APNEE
ALARME DEBRANCHEMENT ACCESSOIRE

NON NON
POINTS DE CONTRÔLE SPO2 VALIDE NA POINTS DE CONTRÔLE CO2 VALIDE NA
VALIDE VALIDE

PRISE DE MESURE DU TAUX OXYGENE X PRISE DE MESURE DU NIVEAU DE CO2


QUALITE D'AFFICHAGE DU TRACE X QUALITE D'AFFICHAGE DU TRACE
ALARME DE SEUILS X CALIBRATION DU CAPTEUR
ALARME DEBRANCHEMENT ACCESSOIRE X ALARME DE SEUILS
ALARME DEBRANCHEMENT ACCESSOIRE

NON NON
POINTS DE CONTRÔLE PNI VALIDE NA POINTS DE CONTRÔLE GAZ HALOGENÉS VALIDE NA
VALIDE VALIDE

PRISE DE MESURE DE LA TENSION ARTERIELLE X PRISE DE MESURE DE LA CONCENTRATION DES GAZ


LINEARITE DU CAPTEUR X QUALITE D'AFFICHAGE DU TRACE
SECURITE SURPRESSION X CALIBRATION DU BANC
ABSENCE DE FUITE X ALARME DE SEUILS
ALARME DE SEUILS X ALARME DEBRANCHEMENT ACCESSOIRE
ALARME DEBRANCHEMENT ACCESSOIRE X

NON NON
POINTS DE CONTRÔLE PI VALIDE NA POINTS DE CONTRÔLE CARDIO-TOCO VALIDE NA
VALIDE VALIDE

PRISE DE MESURE DE LA TENSION ARTERIELLE DETECTION SIGNAL US


QUALITE D'AFFICHAGE DU TRACE PRISE DE MESURE DE LA FREQUENCE CARDIAQUE US
CALIBRATION DU CAPTEUR - MISE A ZÉRO MISE A ZERO CAPTEUR TOCO
ALARME DE SEUILS PRISE DE MESURE TOCO
ALARME DEBRANCHEMENT ACCESSOIRE ALARME DE SEUILS

NON NON
POINTS DE CONTRÔLE TEMPERATURE VALIDE NA POINTS DE CONTRÔLE IMPRESSION VALIDE NA
VALIDE VALIDE

PRISE DE MESURE DE LA TEMPERATURE IMPRESSION


ALARME DE SEUILS ALARME FIN PAPIER - FIN DE RUBAN
ALARME DEBRANCHEMENT ACCESSOIRE DETECTION DES REPERES PAPIERS

TESTS DE SECURITE ELECTRIQUE

NF EN 62353 NON
TESTS REALISES SELON LA NORME VALIDE NA
VALIDE
RESULTAT DU CONTRÔLE X

OBSERVATIONS :
REPARAMETRAGE SELON CONFIGURATION INITIALE CLIENT

CONCLUSION

DISPOSITIF CONFORME : X OUI NON ETAT DEGRADE

ACCESSOIRES CONFORMES : X OUI NON NA ETAT DEGRADE

DATE : 2021/02/25 VISA OPERATEUR : DP