Operation Manual
ACOMA'S HI 300
Operation Manual
Document No. :
Issued Date: Dec.-2009
Revision: 0.0
Document No:
1/42
ACOMA'S HI 300
Operation Manual
Table of Content
1. General Information
1-1.
1-2.
Warranty
1-3..
1-4.
2-1.
Composition
10
2-2.
Specification
12
2-3.
Electrical Requirements
14
2-4.
Proper Environment
15
2-5.
Dimension
16
17
3-1.
Operation Panel
3-2.
3-3.
3-4.
3-5.
3-6.
Setting APR
3-7.
29
3-8.
Radiography
31
4. Troubleshooting
32
4-1.
18
19
23
24
25
27
33
5. Maintenance
36
5-1.
5-2.
5-3.
Periodic Inspection
Document No:
37
38
41
2/42
ACOMA'S HI 300
Operation Manual
Section 1
General Information
Document No:
3/42
ACOMA'S HI 300
Operation Manual
Document No:
4/42
ACOMA'S HI 300
Operation Manual
1-2. WARRANTY
Scope and Period
-
All ACOMA'S Products are guaranteed warrant of 1(one) year for any
disorder or discrepancy occurred from normal use and operation of product.
Note that defects not related to the main functions of the product are not eligible for this
warranty.
Stop using the product when a problem is found, and check the manual for a solution.
Turn off the product before service request, and check the model number, serial number,
the date of purchase before calling the service center.
Document No:
5/42
ACOMA'S HI 300
Operation Manual
Dangerous voltage:
Document No:
Protective earth:
Danger:
Warning:
Caution:
Note:
6/42
ACOMA'S HI 300
Operation Manual
The ACOMA'S HI 300 is designed and manufactured with due consideration given to the safety of
the operator and subject, and also to the reliability of the equipment. This equipment must be
used in accordance with all safety procedures described in this manual. The following
precaution must be observed for additional safety.
l The operator /owner needs to read a handling description carefully and understand the
contents enough, before using this equipment. Particular attention must be paid to all
warning, cautions and notes incorporated herein.
l The responsibility about use of medical apparatus and management of maintenance is in
a user side.
l Compliance with applicable statutory and regulatory requirements is the responsibility of
the owner/operator.
l The equipment must be operated only by or under a supervision or qualified person.
l The equipment is designed for the purpose of radiography X-ray equipment therefore the
unit shall not be used for fluoroscopy and associated application.
l The equipment is designed for the diagnosis therefore the equipment shall not be used
for therapy.
l Do not arbitrarily modify the equipment. If any modification is desired, ask ACOMA'S
or its authorized dealer for the service.
l In radiograph use, according to the exams, the lowest possible values in kV, in mA and in
mAs.
l The equipment has been factory -adjustment for optimum performance. Do not attempt to
adjust any present controls or switches except those specified in this manual for
operation.
Document No:
7/42
ACOMA'S HI 300
Operation Manual
.
l For safe operation of the product
(1) Do not exceed operation time and workload of the product.
l After operation
(1) Do not apply too much force to disconnect cord and cable.
(2) Clean the product after use.
l Power ON/OFF operation
As long as there is no problem on safety, please perform ON/OFF operation of a power
supply with the interval of 30 seconds or more.
l If there is any problem with the product, mark the area of problem to avoid careless
hands, and contact ACOMA'S or authorized dealer for the service.
l If you have not use the product for a long time, check if the product is normal and safe to
use.
Document No:
8/42
ACOMA'S HI 300
Operation Manual
Section 2
Composition and Specification
Document No:
9/42
ACOMA'S HI 300
Operation Manual
2-1. Composition
The electric composition block diagram of this High-Voltage Generating equipment is shown in
Fig 2-1.
Power
Supply
EM switch
Delay
EM switch
AC/DC
Convertor
HF
DC/AC Inverter
HV
TANK
X-ray
TUBE
IGBT
DRVIVER
Stater
OP
PANEL
HT CONT
Collimater
PWR
FILAMENT
I/F
Photo Sensor
or
IC AEC
Fig 2-1
AC input voltage is converted into DC voltage by a single -phase all wave rectifier circuit.
This DC voltage is inverted in high - frequency voltage of about 20 Khz by IGBT inverter.
The high -frequency voltage is converted into the 40(+20kV/ -20kV) to 125kV
(+62.5kV/-62.5kV) DC high voltage by a high -voltage transformer and the multi -voltage
rectifier circuit. This DC high voltage is supplied to an X -ray tube through a high -voltage
cable, and PWM INV controls this X-ray tube voltage. Modulating the Filament voltage
controls the tube current.
.
The operation panel can operate selection in each mode of operation and control,
selection and setting of conditions. These operation conditions are transmitted by the
communication system from OP-CONT to HT-CONT.
HT-CONT controls the output of a high -voltage generator according to the instructions
from OP-CONT.
Document No:
10/42
ACOMA'S HI 300
Operation Manual
The flow of the control signal of X-ray exposure is shown in the following Fig 2-2.
Operation Panel
D/A CONV.
A/D CONV.
kV FB
PWM
Filament
mA FB
kV, mA FB Amp.
rms/DC CONV.
AEC
Photo timer
Driv circuit
FIL. INV.
HV Trans.
kV
FIL. Trans.
A
X-ray TUBE
mA
C
FIL.
Photo Detector
Fig 2-2
Document No:
11/42
ACOMA'S HI 300
Operation Manual
2-2. Specification
1. Rating Out-put
2. Tube Voltage kV
40 125 kV
3.
Tube Current
mA
10 ~ 400 mA
4. kV Accuracy
5%+1kV
5. mA Accuracy
10%+1mA
6. Timer range
7. mAs range
8. MAs Accuracy
9. Display method
14. AEC
Document No:
0.010,
0.063
0.400
3.20
Push switch
Error Code indication
Manual, APR, AEC
216 exam techniques
12/42
ACOMA'S HI 300
Operation Manual
digit
LED
mA
3 digit
LED
mAs
4 digit
LED
Sec
4 digit
LED
LED( 36 )
Alarm
LED( 2 )
16. Weight
High Voltage generator
OP Panel
Document No:
62 kg
2 kg
13/42
ACOMA'S HI 300
Operation Manual
Document No:
ACOMA'S HI 300
1 phase 215V/ 220V/ 240V
10%
50 Hz / 60 Hz
21 kVA
Less than 0.12O
Based on the local regulation
14/42
ACOMA'S HI 300
Operation Manual
For normal operation, you must keep away from the place with a strong vibration and
maintain the following range of temperature and humidity.
l
Operation Environment
Temperature range : 10C ~ 40C,
Relative Humidity Range : 30% ~ 75% RH
Atmospheric pressure : 700hPa ~
1060hPa
The most optimal recommended rage of temperature and humidity
Temperature range: 17C ~ 23C,
Relative Humidity Range : 40% ~ 60% RH
n For storage and transportation condition, you must maintain the following range of
temperature and humidity and atmosphere.
l Storage/Transport Environment
Temperature range : -25C ~ +60C,
Relative Humidity Range : 10% ~ 95% RH
Atmospheric pressure : 700hPa ~
1060hPa
Document No:
15/42
ACOMA'S HI 300
Operation Manual
2-5. Dimension
Document No:
16/42
ACOMA'S HI 300
Operation Manual
Section 3
Function and Operation
Document No:
17/42
ACOMA'S HI 300
Operation Manual
23
13
14
18
15
12
19
20
11
24
21
17
2
3
4
5
6
7
8
9
10
11
12
Power ON switch
Power OFF switch
kV Display / Setting switch
mA Display / Setting switch
mAs/sec Display / Setting switch
mAs/sec Selection switch/Display
Preparation switch for Exposure
X-ray Exposure switch
OVER Display (Red)
ERROR Display (Red)
Ready condition Display (Green)
X-ray Exposure Display (Yellow)
13
14
15
16
17
18
19
20
21
22
23
24
18/42
ACOMA'S HI 300
(1)
Operation MDnual
(2)
Pressing
decreases 1 kV.
If you hold the pressing buttons, the value increases or decreases continuously.
The set-up tube voltage is displayed on kV display.
(4) mA Display / Setting switch (Fig 3-1-? )
Pressing
If you hold the pressing buttons, the value increases or decreases continuously.
The set-up tube current is displayed on mA display.
(5) mAs/sec Display / Setting switch (Fig 3-1-? )
If you hold the pressing buttons, the value increases or decreases continuously.
The set-up mAs or time is displayed.
Document No:
19/42
ACOMA'S HI 300
Operation Manual
is pressed, the rotor is rotated and the filament of X-ray tube is preheated,
then the preparation for X-ray exposure is performed. As same as time the Ready condition
indicator (Fig 3-1-? ) is switched on.
(8) X-ray exposure switch (Fig 3-1-? )
After Ready LED is switched on, the switch
If X-ray Exposure Indicator (Fig 3-1-? ) is switched off, you can release from the switch.
(9) OVER Indication (Fig 3-1-? , Red LED)
The LED is switched on when operation exceeds radiography conditions.
During AEC mode selection, when radiography time exceeds BACKUP time, it blinks.
BACKUP release switch (Fig 3-1-? ) cancels OVER display at the time of AEC mode.
Radiography cannot be performed during a lamp display.
(10) ERROR Indicator (Fig 3-1-? , Green LED)
The LED is switched on when an error occurs.
(11) Ready Condition Indicator (Fig 3-1-? , Green LED)
The LED is switched on when X-ray exposure changes into a possible state.
(12) X-ray Exposure Indicator (Fig 3-1-? , Yellow LED)
The light is switched on during X-ray is exposed.
Document No:
20/42
ACOMA'S HI 300
Operation Manual
The direction of radiography is set up. LED displays the selected direction.
(16) Setting switch of a Patient Size and Display and Radiography condition memory switch
(Fig 3-1-? )
If the region, the radiography direction, and patient size are chosen at the time of APR
mode, radiography conditions (kV, mA, mAs/sec) are set up automatically.
Each of photographic conditions can be changed and the condition is memorized by
Patient Size setting switch (condition memory switch).
Document No:
21/42
ACOMA'S HI 300
Operation Manual
Document No:
22/42
ACOMA'S HI 300
Operation Manual
NOTE: Please turn OFF a power supply except the time of use.
Document No:
23/42
ACOMA'S HI 300
Operation Manual
Document No:
24/42
ACOMA'S HI 300
Operation Manual
Document No:
25/42
ACOMA'S HI 300
Operation Manual
Document No:
26/42
ACOMA'S HI 300
Operation Manual
15 Radiography
The radiography direction is displayed on the right-hand side of the selected region display.
Push the direction setting switches 1 2 3 4 , and choose it from them.
4) Selection of Patient Size
A patient size is chosen with Size 16 setting switch.
The patient size that can be chosen is 6 kinds of LL L M S SS BABY.
5) Display of Radiography Conditions
Radiography conditions (Tube voltage, Tube current, mAs or sec, and BUCKY) are displayed
after selection of the region, the radiography direction, and patient size.
6) Selection of DENSITY
The set-up value of mAs/sec step is changed according as a step by pushing 19
DENSITY switch . The step of mAs/sec is based on 1-2.Specification.
DENSITY: range 1 to -3: Step down
DENSITY: 0
: No change
DENSITY: range+1 to +3: Step up
?
27/42
ACOMA'S HI 300
Operation Manual
Document No:
If the patient size switch not chosen is pushed, it will be set as other
radiography conditions and memory will not be carried out.
It does not memorize in the following condition.
(1) OVER display lights up.
(2) DENSITY setting 0
28/42
ACOMA'S HI 300
Operation Manual
Document No:
29/42
ACOMA'S HI 300
Operation Manual
If patient size is changed after FIELD setup, FIELD setup will return to
a default number.
7) Selection of DENSITY
Radiography Density can be fluctuated with DENSITY19 setting switch.
Radiography Density will become deep if DENSITY=+1 - +3.
Radiography Density will become light if DENSITY=-1 - -3.
8) Display in AEC mode
8-1) ACTUAL-mAs display
In AEC mode, ACTUAL-mAs value is displayed on a mAs/sec display 5 part after a
radiographic end.
However, it is not displayed when radiography is stopped on the way.
8-2) BACKUP-mAs display
In AEC mode, if mAs value reaches BACKUP-mAs, X-ray exposure will be carried out,
and ACTUAL-mAs value is displayed on a mAs/sec display 5 part
Document No:
30/42
ACOMA'S HI 300
Operation Manual
3-8. Radiography
1) Photography preparation
After setting up radiography conditions, the cassette equipped with the film is set to a
radiography position.
If pushing the X -ray irradiation field lamp switch attached to the variable X -rays beam limiting
device, the lamp lights up.
X -ray irradiation field is adjusted with a radiography position, and radiography preparation is
ended.
Document No:
31/42
ACOMA'S HI 300
Operation Manual
Section 4
Troubleshooting
Document No:
32/42
ACOMA'S HI 300
Operation Manual
Error Code
The error code is displayed when problems, such as abnormalities, occur to equipment.
While the error code is displayed, radiography operation is forbidden for safety.
A power supply is re-switched off and on, then an error code is cleared. If the problem of
equipment is not solved at the time of a power-supply re-switched on, the display is lighted on
again.
Since it is failure of equipment when an unusual state is not recovered, please inform the
service section of our company or our agent.
Check code
Check codes are warning displays when an error is on operation of equipment unlike an error
code.
During a check code display, X-ray radiography operation is forbidden for safety.
The method of lifting pushes BACKUP 18 release switch from Ch_02 to Ch_06.
Release of the state where it generated clears Ch_01, Ch_08, and Ch_09.
Ch_07 are cleared by power-supply re-switch.
Document No:
33/42
ACOMA'S HI 300
Operation Manual
Description
Err 12
Err 13
Err 14
Err 15
Err 16
WRONG mA FB AT STANDBY
Err 17
NA
Err 18
Err 19
Err 20
Err 21
Err 22
Err 23
Err 24
HV INTERLOCK
Err 25
NO ZERO-CROSS OR NO INPUT
Err 26
Err 27
Err 28
Err 29
Err 30
Err 31
Err 32
Err 33
Err 34
Action
Check line voltage
Document No:
34/42
ACOMA'S HI 300
Operation Manual
DICRIPTION
TUBE THERMOSTAT OPERATION
When another switch is pushed in exposure
When PREP. switch on OP PANEL is detached in exposure
When X-RAY switch on OP PANEL is detached in exposure
When a hand switch (PREP) is detached in exposure
When a hand switch (X-RAY) is detached in exposure
Shortage of power-supply capacity for exposure
Ch_08
Ch_09
CODE
Document No:
Measure
Reset
Release by Backup
?18 switch
Power-supply
re-switch
Reset
35/42
ACOMA'S HI 300
Operation Manual
Section 5
Maintenance
Document No:
36/42
ACOMA'S HI 300
Operation Manual
Document No:
37/42
ACOMA'S HI 300
Operation Manual
Turn off the switch of a switchboard and carrying out after progress
of 1 minute or more, when removing the cover of equipment.
Keep an operation room and a radiography room in mind to always keep it clean.
The ingredient that makes silver etc. sulfur rate may be contained in
the exhaust gas of an automatic development machine.
Please install the independent air conditioner.
Document No:
38/42
ACOMA'S HI 300
Operation Manual
Operation Panel
It checks whether each switch of an operation panel lights up.
When cleaning the surface of an operation panel, it is wiped with the soft and fine cloth.
When using a cleaner, it is made to contain in cloth and is used.
Document No:
39/42
ACOMA'S HI 300
Operation Manual
Interval
Daily
Air Conditioner
Temperature and Humidity
?
?
Operation Panel
Radiographic Condition
?
?
Weekly
Result
Good/No Good
Memo:
Document No:
40/42
ACOMA'S HI 300
Operation Manual
It is a check report by the expertise person. The periodic inspection is performed according to this
item.
Date:
Performed by:
Client name :
Address :
Installation date: /
/
Last-time checked date :
/
/
Check classification
Document No:
Approval by a user
Result
Result
41/42
ACOMA'S HI 300
Operation Manual
Normal.
Adjusting
Repair
Supplement
T
E
C
O
Result
Result of
maintenance
Tighten
Replacement
Cleaning
Oiling
Sign
M
S
Result of
maintenance
Measurement
Data Store
Memo:
Document No:
Maker/Model/S.N.
Term of validity
42/42