Vous êtes sur la page 1sur 90

GE Healthcare

Technical
Publication

Direction Number 5266280-100


Revision 3
GE Healthcare
LOGIQ C Series
Peripherals Installation Manual

Copyright©2008 - 2009 by GE Healthcare


GE HEALTHCARE LOGIQ C SERIES
DIRECTION 5266280-100, REVISION 3 PERIPHERALS INSTALLATION MANUAL

Important Precautions

• THIS SERVICE MANUAL IS AVAILABLE IN ENGLISH ONLY.


• IF A CUSTOMER’S SERVICE PROVIDER REQUIRES A LANGUAGE OTHER THAN
ENGLISH, IT IS THE CUSTOMER’S RESPONSIBILITY TO PROVIDE
TRANSLATION SERVICES.
WARNING • DO NOT ATTEMPT TO SERVICE THE EQUIPMENT UNLESS THIS SERVICE
MANUAL HAS BEEN CONSULTED AND IS UNDERSTOOD.
• FAILURE TO HEED THIS WARNING MAY RESULT IN INJURY TO THE SERVICE
PROVIDER, OPERATOR OR PATIENT FROM ELECTRIC SHOCK, MECHANICAL
OR OTHER HAZARDS.

• CE MANUEL DE MAINTENANCE N’EST DISPONIBLE QU’EN ANGLAIS.


• SI LE PRESTATAIRE DE SERVICES DU CLIENT A BESOIN DE CE MANUEL DANS
UNE AUTRE LANGUE QUE L’ANGLAIS, IL INCOMBE AU CLIENT DE LE FAIRE
TRADUIRE.
AVERTISSEMENT • NE PAS TENTER D’INTERVENTION SUR LES é QUIPEMENTS TANT QUE LE
MANUEL DE MAINTENANCE N’A PAS é Té CONSULTé ET COMPRIS.
• LE NON-RESPECT DE CET AVERTISSEMENT PEUT ENTRAîNER CHEZ LE
TECHNICIEN, L’OPé RATEUR OU LE PATIENT DES BLESSURES DUES à DES
DANGERS é LECTRIQUES, Mé CANIQUES OU AUTRES.

• DIESES KUNDENDIENST-HANDBUCH EXISTIERT NUR IN ENGLISCHER


SPRACHE.
• FALLS EIN FREMDER KUNDENDIENST EINE ANDERE SPRACHE BENö TIGT, IST
ES AUFGABE DES KUNDEN, Fü R EINE ENTSPRECHENDE ÜBERSETZUNG ZU
SORGEN.
• WARTEN SIE DIESES GERÄT NUR, WENN SIE DIE ENTSPRECHENDEN
WARNUNG ANWEISUNGEN IM KUNDENDIENST-HANDBUCH GELESEN HABEN UND
NACHVOLLZIEHEN KÖNNEN.
• WIRD DIESE WARNUNG NICHT BEACHTET, SO KANN ES ZU VERLETZUNGEN
DES KUNDENDIENSTTECHNIKERS, DES BEDIENERS ODER DES PATIENTEN
DURCH ELEKTRISCHE SCHLä GE, MECHANISCHE ODER SONSTIGE
GEFAHREN KOMMEN.

- iii
GE HEALTHCARE LOGIQ C SERIES
DIRECTION 5266280-100, REVISION 3 PERIPHERALS INSTALLATION MANUAL

• ESTE MANUAL DE SERVICIO Só LO ESTÁ DISPONIBLE EN INGLé S.


• SI ALGú N PROVEEDOR DE SERVICIOS AJENO A GEMS SOLICITA UN IDIOMA
QUE NO SEA EL INGLé S, LA TRADUCCIÓN ES RESPONSABILIDAD DEL
CLIENTE.
• NO SE DEBERá DAR SERVICIO Té CNICO AL EQUIPO SIN HABER
AVISO CONSULTADO Y COMPRENDIDO ESTE MANUAL DE SERVICIO.
• LA NO OBSERVANCIA DEL PRESENTE AVISO PUEDE DAR LUGAR A QUE EL
PROVEEDOR DE SERVICIOS, EL USUARIO O EL PACIENTE SUFRAN LESIONES
PROVOCADAS POR DESCARGAS ELé CTRICAS, PROBLEMAS MECÁNICOS O
PELIGROS DE OTRA NATURALEZA.

• ESTE MANUAL DE ASSISTê NCIA Té CNICA Só SE ENCONTRA DISPONíVEL EM


INGLê S.
• SE QUALQUER OUTRO SERVIç O DE ASSISTê NCIA Té CNICA, QUE Nã O A
GEMS, SOLICITAR ESTES MANUAIS NOUTRO IDIOMA, é DA
RESPONSABILIDADE DO CLIENTE FORNECER OS SERVIç OS DE TRADUç ã O.
ATENÇÃO • Nã O TENTE REPARAR O EQUIPAMENTO SEM TER CONSULTADO E
COMPREENDIDO ESTE MANUAL DE ASSISTê NCIA Té CNICA.
• O Nã O CUMPRIMENTO DESTE AVISO PODE PÔR EM PERIGO A SEGURANç A
DO Té CNICO, OPERADOR OU PACIENTE DEVIDO A CHOQUES ELé TRICOS,
MECâ NICOS OU OUTROS.

• IL PRESENTE MANUALE DI MANUTENZIONE è DISPONIBILE SOLTANTO IN


INGLESE.
• SE UN ADDETTO ALLA MANUTENZIONE ESTERNO ALLA GEMS RICHIEDE IL
MANUALE IN UNA LINGUA DIVERSA, IL CLIENTE è TENUTO A PROVVEDERE
DIRETTAMENTE ALLA TRADUZIONE.
• SI PROCEDA ALLA MANUTENZIONE DELL’APPARECCHIATURA SOLO DOPO
AVVERTENZA AVER CONSULTATO IL PRESENTE MANUALE ED AVERNE COMPRESO IL
CONTENUTO.
• NON TENERE CONTO DELLA PRESENTE AVVERTENZA POTREBBE FAR
COMPIERE OPERAZIONI DA CUI DERIVINO LESIONI ALL’ADDETTO ALLA
MANUTENZIONE, ALL’UTILIZZATORE ED AL PAZIENTE PER FOLGORAZIONE
ELETTRICA, PER URTI MECCANICI OD ALTRI RISCHI.

• KÄESOLEV TEENINDUSJUHEND ON SAADAVAL AINULT INGLISE KEELES.


• KUI KLIENDITEENINDUSE OSUTAJA Nõ UAB JUHENDIT INGLISE KEELEST
ERINEVAS KEELES, VASTUTAB KLIENT Tõ LKETEENUSE OSUTAMISE EEST.
• ä RGE ü RITAGE SEADMEID TEENINDADA ENNE EELNEVALT Kä ESOLEVA
HOIATUS TEENINDUSJUHENDIGA TUTVUMIST JA SELLEST ARU SAAMIST.
• Kä ESOLEVA HOIATUSE EIRAMINE Võ IB Põ HJUSTADA TEENUSEOSUTAJA,
OPERAATORI Võ I PATSIENDI VIGASTAMIST ELEKTRILö ö GI, MEHAANILISE
Võ I MUU OHU TAGAJä RJEL.

iv -
GE HEALTHCARE LOGIQ C SERIES
DIRECTION 5266280-100, REVISION 3 PERIPHERALS INSTALLATION MANUAL

• TÄMÄ HUOLTO-OHJE ON SAATAVILLA VAIN ENGLANNIKSI.


• JOS ASIAKKAAN PALVELUNTARJOAJA VAATII MUUTA KUIN
ENGLANNINKIELISTä MATERIAALIA, TARVITTAVAN Kä ä NNö KSEN
HANKKIMINEN ON ASIAKKAAN VASTUULLA.
• ä Lä YRITä KORJATA LAITTEISTOA ENNEN KUIN OLET VARMASTI LUKENUT
VAROITUS JA YMMä RTä NYT Tä Mä N HUOLTO-OHJEEN.
• MIKä LI Tä Tä VAROITUSTA EI NOUDATETA, SEURAUKSENA VOI OLLA
PALVELUNTARJOAJAN, LAITTEISTON Kä YTTä Jä N TAI POTILAAN
VAHINGOITTUMINEN Sä HKö ISKUN, MEKAANISEN VIAN TAI MUUN
VAARATILANTEEN VUOKSI.

• ΤΟ ΠΑΡΟΝ ΕΓΧΕΙΡΙΔΙΟ ΣΕΡΒΙΣ ΔΙΑΤΙΘΕΤΑΙ ΣΤΑ ΑΓΓΛΙΚΑ ΜΟΝΟ.


• ΕΑΝ ΤΟ ΑΤΟΜΟ ΠΑΡΟΧΗΣ ΣΕΡΒΙΣ ΕΝΟΣ ΠΕΛΑΤΗ ΑΠΑΙΤΕΙ ΤΟ ΠΑΡΟΝ
ΕΓΧΕΙΡΙΔΙΟ ΣΕ ΓΛΩΣΣΑ ΕΚΤΟΣ ΤΩΝ ΑΓΓΛΙΚΩΝ, ΑΠΟΤΕΛΕΙ ΕΥΘΥΝΗ ΤΟΥ
ΠΕΛΑΤΗ ΝΑ ΠΑΡΕΧΕΙ ΥΠΗΡΕΣΙΕΣ ΜΕΤΑΦΡΑΣΗΣ.
• ΜΗΝ ΕΠΙΧΕΙΡΗΣΕΤΕ ΤΗΝ ΕΚΤΕΛΕΣΗ ΕΡΓΑΣΙΩΝ ΣΕΡΒΙΣ ΣΤΟΝ ΕΞΟΠΛΙΣΜΟ
ΠΡΟΕΙΔΟΠΟΙΗΣΗ ΕΚΤΟΣ ΕΑΝ ΕΧΕΤΕ ΣΥΜΒΟΥΛΕΥΤΕΙ ΚΑΙ ΕΧΕΤΕ ΚΑΤΑΝΟΗΣΕΙ ΤΟ ΠΑΡΟΝ
ΕΓΧΕΙΡΙΔΙΟ ΣΕΡΒΙΣ.
• ΕΑΝ ΔΕ ΛΑΒΕΤΕ ΥΠΟΨΗ ΤΗΝ ΠΡΟΕΙΔΟΠΟΙΗΣΗ ΑΥΤΗ, ΕΝΔΕΧΕΤΑΙ ΝΑ
ΠΡΟΚΛΗΘΕΙ ΤΡΑΥΜΑΤΙΣΜΟΣ ΣΤΟ ΑΤΟΜΟ ΠΑΡΟΧΗΣ ΣΕΡΒΙΣ, ΣΤΟ ΧΕΙΡΙΣΤΗ Ή
ΣΤΟΝ ΑΣΘΕΝΗ ΑΠΟ ΗΛΕΚΤΡΟΠΛΗΞΙΑ, ΜΗΧΑΝΙΚΟΥΣ Ή ΑΛΛΟΥΣ ΚΙΝΔΥΝΟΥΣ.

• EZEN KARBANTARTÁSI KÉZIKÖNYV KIZÁRÓLAG ANGOL NYELVEN ÉRHETŐ


EL.
• HA A VEVŐ SZOLGÁLTATÓJA ANGOLTÓL ELTÉRŐ NYELVRE TART IGÉNYT,
AKKOR A VEVŐ FELELŐSSÉGE A FORDÍTÁS ELKÉSZÍTTETÉSE.
FIGYELMEZTETÉS
• NE PRÓBÁLJA ELKEZDENI HASZNÁLNI A BERENDEZÉST, AMÍG A
KARBANTARTÁSI KÉZIKÖNYVBEN LEÍRTAKAT NEM ÉRTELMEZTÉK.
• EZEN FIGYELMEZTETÉS FIGYELMEN KÍVÜL HAGYÁSA A SZOLGÁLTATÓ,
MŰKÖDTETŐ VAGY A BETEG ÁRAMÜTÉS, MECHANIKAI VAGY EGYÉB
VESZÉLYHELYZET MIATTI SÉRÜLÉSÉT EREDMÉNYEZHETI.

• ÞESSI ÞJÓNUSTUHANDBÓK ER EINGÖNGU FÁANLEG Á ENSKU.


• EF ÞJÓNUSTUAÐILI VIÐSKIPTAMANNS ÞARFNAST ANNARS TUNGUMÁLS EN
ENSKU, ER ÞAÐ Á ÁBYRGÐ VIÐSKIPTAMANNS AÐ ÚTVEGA ÞÝÐINGU.
• REYNIÐ EKKI AÐ ÞJÓNUSTA TÆKIÐ NEMA EFTIR AÐ HAFA SKOÐAÐ OG SKILIÐ
VIÐVÖRUN ÞESSA ÞJÓNUSTUHANDBÓK.
• EF EKKI ER FARIÐ AÐ ÞESSARI VIÐVÖRUN GETUR ÞAÐ VALDIÐ MEIÐSLUM
ÞJÓNUSTUVEITANDA, STJÓRNANDA EÐA SJÚKLINGS VEGNA RAFLOSTS,
VÉLRÆNNAR EÐA ANNARRAR HÆTTU.

- v
GE HEALTHCARE LOGIQ C SERIES
DIRECTION 5266280-100, REVISION 3 PERIPHERALS INSTALLATION MANUAL

• TENTO SERVISNÍ NÁVOD EXISTUJE POUZE V ANGLICKé M JAZYCE.


• V Př íPADě , ŽE POSKYTOVATEL SLUŽEB ZÁKAZNÍKŮM POTř EBUJE Ná VOD V
JINé M JAZYCE, JE ZAJIšTě Ní Př EKLADU DO ODPOVíDAJíCíHO JAZYKA
ú KOLEM Zá KAZNíKA.
• NEPROVÁDĚJTE ú DRŽBU TOHOTO ZAř íZENí, ANIŽ BYSTE SI Př Eč ETLI
VÝSTRAHA
TENTO SERVISNÍ NÁVOD A POCHOPILI JEHO OBSAH.
• V Př íPADě NEDODRŽOVá Ní Té TO VýSTRAHY Mů ŽE DOJíT ÚRAZU
ELEKTRICKÁM PROUDEM PRACOVNíKA POSKYTOVATELE SLUŽEB,
OBSLUŽNé HO PERSONá LU NEBO PACIENTů VLIVEM ELEKTRICKé HOP
PROUDU, RESPEKTIVE VLIVEM K RIZIKU MECHANICKÉHO POŠKOZENÍ NEBO
JINÉMU RIZIKU.

• DENNE SERVICEMANUAL FINDES KUN PÅ ENGELSK.


• HVIS EN KUNDES TEKNIKER HAR BRUG FOR ET ANDET SPROG END
ENGELSK, ER DET KUNDENS ANSVAR AT SØRGE FOR OVERSÆTTELSE.
• FORSØG IKKE AT SERVICERE UDSTYRET MEDMINDRE
ADVARSEL DENNE SERVICEMANUAL ER BLEVET LÆST OG FORSTÅET.
• MANGLENDE OVERHOLDELSE AF DENNE ADVARSEL KAN MEDFØRE SKADE
PÅ GRUND AF ELEKTRISK, MEKANISK ELLER ANDEN FARE FOR TEKNIKEREN,
OPERATØREN ELLER PATIENTEN.

• DEZE ONDERHOUDSHANDLEIDING IS ENKEL IN HET ENGELS


VERKRIJGBAAR.
• ALS HET ONDERHOUDSPERSONEEL EEN ANDERE TAAL VEREIST, DAN IS DE
KLANT VERANTWOORDELIJK VOOR DE VERTALING ERVAN.
• PROBEER DE APPARATUUR NIET TE ONDERHOUDEN VOORDAT DEZE
WAARSCHUWING ONDERHOUDSHANDLEIDING WERD GERAADPLEEGD EN BEGREPEN IS.
• INDIEN DEZE WAARSCHUWING NIET WORDT OPGEVOLGD, ZOU HET
ONDERHOUDSPERSONEEL, DE OPERATOR OF EEN PATIËNT GEWOND
KUNNEN RAKEN ALS GEVOLG VAN EEN ELEKTRISCHE SCHOK,
MECHANISCHE OF ANDERE GEVAREN.

• šĪ APKALPES ROKASGRĀMATA IR PIEEJAMA TIKAI ANGĻU VALODĀ.


• JA KLIENTA APKALPES SNIEDZĒJAM NEPIECIEŠAMA INFORMĀCIJA CITĀ
VALODĀ, NEVIS ANGĻU, KLIENTA PIENĀKUMS IR NODROŠINĀT TULKOŠANU.
• NEVEICIET APRĪKOJUMA APKALPI BEZ APKALPES ROKASGRĀMATAS
BRĪDINĀJUMS IZLASĪŠANAS UN SAPRAŠANAS.
• ŠĪ BRĪDINĀJUMA NEIEVĒROŠANA VAR RADĪT ELEKTRISKĀS STRĀVAS
TRIECIENA, MEHĀNISKU VAI CITU RISKU IZRAISĪTU TRAUMU APKALPES
SNIEDZĒJAM, OPERATORAM VAI PACIENTAM.

vi -
GE HEALTHCARE LOGIQ C SERIES
DIRECTION 5266280-100, REVISION 3 PERIPHERALS INSTALLATION MANUAL

• ŠIS EKSPLOATAVIMO VADOVAS YRA IŠLEISTAS TIK ANGLŲ KALBA.


• JEI KLIENTO PASLAUGŲ TEIKĖJUI REIKIA VADOVO KITA KALBA – NE ANGLŲ,
VERTIMU PASIRŪPINTI TURI KLIENTAS.
• NEMĖGINKITE ATLIKTI ĮRANGOS TECHNINĖS PRIEŽIŪROS DARBŲ, NEBENT
ĮSPĖJIMAS VADOVAUTUMĖTĖS ŠIUO EKSPLOATAVIMO VADOVU IR JĮ SUPRASTUMĖTE
• NEPAISANT ŠIO PERSPĖJIMO, PASLAUGŲ TEIKĖJAS, OPERATORIUS AR
PACIENTAS GALI BŪTI SUŽEISTAS DĖL ELEKTROS SMŪGIO, MECHANINIŲ AR
KITŲ PAVOJŲ.

• DENNE SERVICEHÅNDBOKEN FINNES BARE PÅ ENGELSK.


• HVIS KUNDENS SERVICELEVERANDØR TRENGER ET ANNET SPRÅK, ER DET
KUNDENS ANSVAR Å SØRGE FOR OVERSETTELSE.
• IKKE FORSØK Å REPARERE UTSTYRET UTEN AT DENNE
ADVARSEL SERVICEHÅNDBOKEN ER LEST OG FORSTÅTT.
• MANGLENDE HENSYN TIL DENNE ADVARSELEN KAN FØRE TIL AT
SERVICELEVERANDØREN, OPERATØREN ELLER PASIENTEN SKADES PÅ
GRUNN AV ELEKTRISK STØT, MEKANISKE ELLER ANDRE FARER.

• NINIEJSZY PODRĘCZNIK SERWISOWY DOSTĘPNY JEST JEDYNIE W JĘZYKU


ANGIELSKIM.
• JEś LI FIRMA ś WIADCZĄCA KLIENTOWI USłUGI SERWISOWE WYMAGA
UDOSTę PNIENIA PODRę CZNIKA W Ję ZYKU INNYM NIŻ ANGIELSKI,
OBOWIĄZEK ZAPEWNIENIA STOSOWNEGO TłUMACZENIA SPOCZYWA NA
KLIENCIE.
OSTRZEŻENIE • NIE PRó BOWAć SERWISOWAć NINIEJSZEGO SPRZę TU BEZ UPRZEDNIEGO
ZAPOZNANIA SIę Z PODRę CZNIKIEM SERWISOWYM.
• NIEZASTOSOWANIE SIę DO TEGO OSTRZEŻENIA MOżE GROZIć
OBRAŻENIAMI CIAłA SERWISANTA, OPERATORA LUB PACJENTA W WYNIKU
PORAŻENIA PRĄDEM, URAZU MECHANICZNEGO LUB INNEGO RODZAJU
ZAGROŻEń .

• ACEST MANUAL DE SERVICE ESTE DISPONIBIL NUMAI ÎN LIMBA ENGLEZĂ.


• DACĂ UN FURNIZOR DE SERVICII PENTRU CLIENŢI NECESITĂ O ALTĂ LIMBĂ
DECÂT CEA ENGLEZĂ, ESTE DE DATORIA CLIENTULUI SĂ FURNIZEZE O
TRADUCERE.
ATENŢIE • NU ÎNCERCAŢI SĂ REPARAŢI ECHIPAMENTUL DECÂT ULTERIOR
CONSULTĂRII ŞI ÎNŢELEGERII ACESTUI MANUAL DE SERVICE.
• IGNORAREA ACESTUI AVERTISMENT AR PUTEA DUCE LA RĂNIREA
DEPANATORULUI, OPERATORULUI SAU PACIENTULUI ÎN URMA
PERICOLELOR DE ELECTROCUTARE, MECANICE SAU DE ALTĂ NATURĂ.

- vii
GE HEALTHCARE LOGIQ C SERIES
DIRECTION 5266280-100, REVISION 3 PERIPHERALS INSTALLATION MANUAL

• Да н н о е рук о в о дс т в о п о о бс луж и в а н и ю ПРЕДОСТАВЛЯЕТСЯ


т о ль к о н а а н г ли й с к о м Яз ы к е .
• Ес ли с е рв и с н о МУ ПЕРСОНАЛУ к ли е н т а н е о б х о ди мо рук о в о дс т в о
н е н а а н г ли й с к о м ЯЗЫКЕ, к ли е н т у с ле дуе т с а мо с т о Ят е ль н о
ОБЕСПЕЧИТЬ п е ре в о д.
ОСТОРОЖНО! • ПЕРЕД ОБСЛУЖИВАНИЕМ ОБОРУДОВАНИЯ ОБЯЗАТЕЛЬНО ОБРАТИТЕСЬ
К ДАННОМУ РУКОВОДСТВУ И ПОЙМИТЕ ИЗЛОЖЕННЫЕ В НЕМ СВЕДЕНИЯ.
• НЕСОБЛЮДЕНИЕ УКАЗАННЫХ ТРЕБОВАНИЙ МОЖЕТ ПРИВЕСТИ К ТОМУ,
ЧТО СПЕЦИАЛИСТ ПО ТЕХОБСЛУЖИВАНИЮ, ОПЕРАТОР ИЛИ ПАЦИЕНТ
ПОЛУЧАТ УДАР ЗЛЕКТРИЧЕСКИМ ТОКОМ, МЕХАНИЧЕСКУЮ ТРАВМУ ИЛИ
ДРУГОЕ ПОВРЕЖДЕНИЕ.

• TÁTO SERVISNÁ PRÍRUČKA JE K DISPOZíCII LEN V ANGLIČTINE.


• AK ZÁKAZNÍKOV POSKYTOVATEĽ SLUŽIEB VYŽADUJE INÝ JAZYK AKO
ANGLIČTINU, POSKYTNUTIE PREKLADATEĽSKÝCH SLUŽIEB JE
ZODPOVEDNOSŤOU ZÁKAZNÍKA.
• NEPOKÚŠAJTE SA VYKONÁVAŤ SERVIS ZARIADENIA SKÔR, AKO SI
UPOZORNENIE NEPREČÍTATE SERVISNÚ PRÍRUČKU A NEPOROZUMIETE JEJ.
• ZANEDBANIE TOHTO UPOZORNENIA Mô ŽE VYÚSTIŤ DO ZRANENIA
POSKYTOVATEĽA SLUŽIEB, OBSLUHUJú CEJ OSOBY ALEBO PACIENTA
ELEKTRICKýM PRú DOM, PRÍPADNE DO MECHANICKé HO ALEBO INé HO
NEBEZPEč ENSTVA.

• DEN HÄR SERVICEHANDBOKEN FINNS BARA TILLGÄNGLIG PÅ ENGELSKA.


• OM EN KUNDS SERVICETEKNIKER HAR BEHOV AV ETT ANNAT SPRÅK ÄN
ENGELSKA ANSVARAR KUNDEN FÖR ATT TILLHANDAHÅLLA
ÖVERSÄTTNINGSTJÄNSTER.
VARNING • FÖRSÖK INTE UTFÖRA SERVICE PÅ UTRUSTNINGEN OM DU INTE HAR LÄST
OCH FÖRSTÅR DEN HÄR SERVICEHANDBOKEN.
• OM DU INTE TAR HÄNSYN TILL DEN HÄR VARNINGEN KAN DET RESULTERA I
SKADOR PÅ SERVICETEKNIKERN, OPERATÖREN ELLER PATIENTEN TILL
FÖLJD AV ELEKTRISKA STÖTAR, MEKANISKA FAROR ELLER ANDRA FAROR.

• BU SERVİS KILAVUZU YALNIZCA İNGİLİZCE OLARAK SAĞLANMIŞTIR.


• EĞER MÜŞTERİ TEKNİSYENİ KILAVUZUN İNGİLİZCE DIŞINDAKİ BİR DİLDE
OLMASINI İSTERSE, KILAVUZU TERCÜME ETTİRMEK MÜŞTERİNİN
SORUMLULUĞUNDADIR.
DİKKAT • SERVİS KILAVUZUNU OKUYUP ANLAMADAN EKİPMANLARA MÜDAHALE
ETMEYİNİZ.
• BU UYARININ GÖZ ARDI EDİLMESİ, ELEKTRİK ÇARPMASI YA DA MEKANİK
VEYA DİĞER TÜRDEN KAZALAR SONUCUNDA TEKNİSYENİN, OPERATÖRÜN
YA DA HASTANIN YARALANMASINA YOL AÇABİLİR.

viii -
GE HEALTHCARE LOGIQ C SERIES
DIRECTION 5266280-100, REVISION 3 PERIPHERALS INSTALLATION MANUAL

- ix
GE HEALTHCARE LOGIQ C SERIES
DIRECTION 5266280-100, REVISION 3 PERIPHERALS INSTALLATION MANUAL

DAMAGE IN TRANSPORTATION
All packages should be closely examined at time of delivery. If damage is apparent write “Damage In
Shipment” on ALL copies of the freight or express bill BEFORE delivery is accepted or “signed for” by
a GE representative or hospital receiving agent. Whether noted or concealed, damage MUST be
reported to the carrier immediately upon discovery, or in any event, within 14 days after receipt, and the
contents and containers held for inspection by the carrier. A transportation company will not pay a claim
for damage if an inspection is not requested within this 14 day period.

OMISSIONS & ERRORS


If there are any omissions, errors or suggestions for improving this documentation, please contact the
GE Healthcare Global Documentation Group with specific information listing the system type, manual
title, part number, revision number, page number and suggestion details. Mail the information to:
Service Documentation, 9900 Innovation Drive (RP-2123), Wauwatosa, WI 53226, USA.

GE Healthcare employees should use the iTrak System to report all documentation errors or omissions.

SERVICE SAFETY CONSIDERATIONS


DANGER DANGEROUS VOLTAGES, CAPABLE OF CAUSING DEATH, ARE PRESENT IN
THIS EQUIPMENT. USE EXTREME CAUTION WHEN HANDLING, TESTING AND
ADJUSTING.
WARNING Use all Personal Protection Equipment (PPE) such as gloves, safety shoes, safety
glasses, and kneeling pad, to reduce the risk of injury.
For a complete review of all safety requirements, see the Chapter 1, Safety Considerations section of
the LOGIQ C Series Peripheral Installation Manual (5266280-100).

x -
GE HEALTHCARE LOGIQ C SERIES
DIRECTION 5266280-100, REVISION 3 PERIPHERALS INSTALLATION MANUAL

LEGAL NOTES

The contents of this publication may not be copied or duplicated in any form, in whole or in part, without
prior written permission of GE Healthcare.

TRADEMARKS
All products and their name brands are trademarks of their respective holders.

COPYRIGHTS
All Material Copyright© 2008-2009 by General Electric Inc. All Rights Reserved.

- xi
GE HEALTHCARE LOGIQ C SERIES
DIRECTION 5266280-100, REVISION 3 PERIPHERALS INSTALLATION MANUAL

Revision History

Revision Date Reason for change


1 July 07, 2008 Initial Release
2 Jan 21, 2009 Add CKD Information
3 Aug 11, 2009 Add LOGIQ C3 Premium and LOGIQ C5 Premium Information

List of Effected Pages

Pages Revision Pages Revision Pages Revision


Chapter 7 - ECG Installation
Chapter 2 - Preparation
Title Page 3 3 pages 8-1 to 8-6 3
pages 2-1 to 2-7

Chapter 3 - B/W USB Printer Chapter 8 - Seagate HDD


Important Precautions
3 Installation 3 Installation pages 9-1 to 9-2 3
pages -iii to -xi
pages 3-1 to 3-4
Rev History Chapter 4 - Footswitch Installation Chapter 9 - Set up and check-
3 3 3
pages xii pages 4-1 to 4-4 out pages 10-1 to 10-22
Table of Contents Chapter 5 - Color USB Printer
3 3 Back Cover 3
pages 1 to 10 Installation pages 6-1 to 6-6
Chapter 1 - Introduction Chapter 6 - Sandisk USB Stick
3 3
pages 1-1 to 1-10 Installation pages 7-1 to 7-2

xii -
GE HEALTHCARE LOGIQ C SERIES
DIRECTION 5266280-100, REVISION 3 PERIPHERALS INSTALLATION MANUAL

Table of Contents
CHAPTER 1
Introduction
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 1
Purpose . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 1
Peripherals Covered by this Manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 2
Typical Users of the LOGIQ C Series . . . . . . . . . . . . . . . . . . . . . . 1 - 2

Important Conventions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3


Conventions Used in Book . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
Icons . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
Safety Precaution Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
Standard Hazard Icons . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-4

Safety Considerations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 5
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 5
Human Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 5
Mechanical Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 5
Electrical Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 6
Safe Practices . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 6
Dangerous Procedure Warnings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 6
Returning/Shipping Probes and Repair Parts . . . . . . . . . . . . . . . . . . . . . . . 1 - 6

EMC, EMI, and ESD . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-7


Electromagnetic Compatibility (EMC) and Interference (EMI) . . . . . . . . . . . 1-7
Electrostatic Discharge (ESD) Prevention . . . . . . . . . . . . . . . . . . . . . . . . . . 1-7
CE Compliance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-7

Customer Assistance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-8


Contact Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-8
System Manufacture . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-9
Factory Site . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-9

1 Table of Contents
GE HEALTHCARE LOGIQ C SERIES
DIRECTION 5266280-100, REVISION 3 PERIPHERALS INSTALLATION MANUAL

CHAPTER 2
Preparation
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 1
Purpose of Chapter 2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 1

Power Shut Down . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 2

Furnished materials . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 3
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 3
B/W USB Printer (Sony UP-D897) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 3
Materials supplied with the B/W USB Printer . . . . . . . . . . . . . . . . . . 2 - 3
Materials supplied with scanner . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 3
Color USB Printer (UP-D23MD/HP470) . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 4
Materials supplied with the Color USB Printer . . . . . . . . . . . . . . . . . 2 - 4
Materials supplied with scanner . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 4
1 Pedal Footswitch (Whanam FSU2001) . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 4
Materials supplied with the Footswitch . . . . . . . . . . . . . . . . . . . . . . . 2 - 4
Materials supplied in the scanner . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 4
Seagate Removable HDD . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 5
Materials supplied with the Seagate Removable HDD . . . . . . . . . . . 2 - 5
Materials supplied in the scanner . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 5
Peripherals/Accessories Connector Panel . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 5
Rear Panel Connector for LOGIQ C2/C2PRO,LOGIQ C3/C3PRO
and LOGIQ C5/C5PRO . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 5
Rear Panel Connector for LOGIQ C3 Premium and
LOGIQ C5 Premium . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 6
Left Panel Connector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 6
Located on the left side panel is one USB Port. . . . . . . . . . . . . . . . . 2 - 7

1-2 -
GE HEALTHCARE LOGIQ C SERIES
DIRECTION 5266280-100, REVISION 3 PERIPHERALS INSTALLATION MANUAL

CHAPTER 3
B/W USB Printer Installation
(Sony UP-D897)
Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 1
Purpose of Chapter 3 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 1

Installing the B/W USB Printer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2


Tools . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2
Manpower . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2
Preparations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2
Installation procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2

- 1-3
GE HEALTHCARE LOGIQ C SERIES
DIRECTION 5266280-100, REVISION 3 PERIPHERALS INSTALLATION MANUAL

CHAPTER 4
B/W USB Printer Installation
(Sony UP-897MD)
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 1
Purpose of Chapter 4 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 1

Installing the B/W Printer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 2


Tools . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 2
Manpower . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 2
Preparations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 2
Installation procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 2

1-4 -
GE HEALTHCARE LOGIQ C SERIES
DIRECTION 5266280-100, REVISION 3 PERIPHERALS INSTALLATION MANUAL

CHAPTER 5
Footswitch Installation
Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 1
Purpose of Chapter 5 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 1

Installing Footswitch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-2


Tools . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-2
Manpower . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-2
Preparations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-2
Installation Procedure for 1-Pedal Footswitch (Whanam FSU2001) . . . . . 5-2
Installation Procedure for 1-Pedal Footswitch (Whanam FSU1000) . . . . . 5-3
Installation Procedure for 3-Pedal Footswitch . . . . . . . . . . . . . . . . . . . . . . 5-3

Configuring Footswtich . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 4

- 1-5
GE HEALTHCARE LOGIQ C SERIES
DIRECTION 5266280-100, REVISION 3 PERIPHERALS INSTALLATION MANUAL

CHAPTER 6
Color USB Printer Installation
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 1
Purpose of Chapter 6 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 1

Installing the Color USB Printer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 2


Tools . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 2
Manpower . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 2
Preparations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 2
Installation procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 2

Setting for B/W and Color printer concurrent working . . . . . . . . . . . . . . . . . . . . . . . 6 - 3

1-6 -
GE HEALTHCARE LOGIQ C SERIES
DIRECTION 5266280-100, REVISION 3 PERIPHERALS INSTALLATION MANUAL

CHAPTER 7
SanDisk USB Stick Installation
Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 - 1
Purpose of Chapter 7 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 - 1

Installing the SanDisk USB Stick . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-2


Tools . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-2
Manpower . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-2
Preparations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-2
Installation procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-2

- 1-7
GE HEALTHCARE LOGIQ C SERIES
DIRECTION 5266280-100, REVISION 3 PERIPHERALS INSTALLATION MANUAL

CHAPTER 8
ECG Installation
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 - 1
Purpose of Chapter 8 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 - 1

Installing the ECG . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 - 2


Tools . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 - 2
Manpower . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 - 2
Preparations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 - 2
Installation procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 - 3
ECG Intended Use. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 - 4
Patient Preparation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 - 4
Technical Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 - 5

1-8 -
GE HEALTHCARE LOGIQ C SERIES
DIRECTION 5266280-100, REVISION 3 PERIPHERALS INSTALLATION MANUAL

CHAPTER 9
Seagate Removable HDD Installation
Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 - 1
Purpose of Chapter 9 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 - 1

Installing the Seagate Removable HDD . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-2


Tools . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-2
Manpower . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-2
Preparations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-2
Installation procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-2

- 1-9
GE HEALTHCARE LOGIQ C SERIES
DIRECTION 5266280-100, REVISION 3 PERIPHERALS INSTALLATION MANUAL

CHAPTER 10
Set up and Check-out
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 1
Purpose of Chapter 10 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 1

B/W USB Printer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 2


Control assignment for B/W USB Printer . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 2
Control assignment procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 2
Set Up Digital Peripheral . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 4

HP 470 office jet Printer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 8


Control assignment for HP 470 officejet Printer . . . . . . . . . . . . . . . . . . . . . . 10 - 8
Control assignment procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 8
Set Up Digital Peripheral . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 11

Sony UP-D23 MD Printer. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 14


Control assignment for Sony UP-D23 MD Printer . . . . . . . . . . . . . . . . . . . . 10 - 14
Control assignment procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 14
Set Up Digital Peripheral . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 17

B/W USB Printer Checks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 20

Color Printer Checks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 20

Footswitch Checks. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 21

USB Stick Checks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 21

1 - 10 -
GE HEALTHCARE LOGIQ C SERIES
DIRECTION 5266280-100, REVISION 3 PERIPHERALS INSTALLATION MANUAL

Chapter 1
Introduction

Section 1-1
Overview
1-1-1 Purpose
This Installation Manual describes how to install and configure the peripherals validated for the
LOGIQ C Series.

NOTE: LOGIQ C series contain LOGIQ C2/C2 PRO,LOGIQ C3/C3PRO,LOGIQ C5/C5 PRO,LOGIQ C3
Premium and LOGIQ C5 Premium.

• Chapter 1 - Introduction: Contains a content summary and warnings.


• Chapter 2 - Preparation: System power shut down and Materials furnished.
• Chapter 3 - B/W USB Printer Installation (Sony UP-D897): Installation procedure for B/W USB
Printer.
• Chapter 4 - B/W USB Printer Installation (Sony UP-897MD): Installation procedure for B/W USB
Printer.
• Chapter 5 - Footswitch Installation: Installation procedure for Footswitch.
• Chapter 6 - Color USB Printer Installation: Installation procedure for Color USB Printer
Installation.
• Chapter 7 - SanDisk USB Stick Installation: SanDisk USB Stick Installation
• Chapter 8 - ECG Installation: ECG Installation
• Chapter 9 - Seagate Removable HDD Installation: Seagate Removable HDD Installation
• Chapter 10 - Set up and Check-out: Final setup for peripherals and operational check-out.

Table 1-1 Contents in Chapter 1

PAGE
SECTION DESCRIPTION NUMBER
1-1 Overview 1-1
1-2 Important Conventions 1-3

1-3 Safety Considerations 1-5

1-4 EMC, EMI, and ESD 1-7


1-5 Customer Assistance 1-8

Chapter 1 - Introduction 1-1


GE HEALTHCARE LOGIQ C SERIES
DIRECTION 5266280-100, REVISION 3 PERIPHERALS INSTALLATION MANUAL

1-1-2 Peripherals Covered by this Manual

Table 1-2 LOGIQ C Series Peripherals

WEIGHT
PART NUMBER DESCRIPTION (KG) POWER CONTROL MODEL

100-120VAC/220-
240VAC
H48112LG Sony B/W USB Printer 3.4 USB ports UP-D897
50/60Hz
1.5A/0.8A

100-120VAC/220-
240VAC
Sony B/W Printer 3.4 UP-897MD
50/60Hz
1.5A/0.8A

H41482LE(CN) 100-120VAC/220-
H48272LR(EU) Color USB Printer 6.5 240VAC USB ports UP-D23MD
H48272LS(US) 1.8A/1.0A

H48242LD(CN) 100-120VAC/220-
H48272LT(EU) Deskjet Color Printer 4.0 240VAC USB ports HP470
H48272LW(US) 1.8A/1.0A

H41412LH 1-Pedal Footswitch 0.8 5V USB ports Whanam FSU2001

H41882LD 1-Pedal Footswitch 0.8 5V USB ports Whanam FSU1000

H41642LS 3-Pedal Footswitch 0.8 5V USB ports MKF 2-MED GP26

H48272LY SanDisk USB Stick 2G 0.05 USB ports CRUZER

Seagate 80G removable


H48172LW Seagate Removable HDD 0.3 USB ports
HDD

H41852LL
H41852LM ECG USB ports
H41852LK

1-1-2-1 Typical Users of the LOGIQ C Series


• Service Personnel (installation, maintenance, etc.).
• Hospital’s Service Personnel

1-2 -
GE HEALTHCARE LOGIQ C SERIES
DIRECTION 5266280-100, REVISION 3 PERIPHERALS INSTALLATION MANUAL

Section 1-2
Important Conventions
1-2-1 Conventions Used in Book

1-2-1-1 Icons
Pictures, or icons, are used wherever they will reinforce the printed message. The icons, labels and
conventions used on the product and in the service information are described in this chapter.

1-2-1-2 Safety Precaution Messages


Various levels of safety precaution messages may be found on the equipment and in the service
information. The different levels of concern are identified by a flag word that precedes the precautionary
message. Known or potential hazards are labeled in one of three ways:

DANGER DANGER IS USED TO INDICATE THE PRESENCE OF A HAZARD THAT WILL


CAUSE SEVERE PERSONAL INJURY OR DEATH IF THE INSTRUCTIONS ARE
IGNORED.
WARNING WARNING IS USED TO INDICATE THE PRESENCE OF A HAZARD THAT CAN CAUSE
SEVERE PERSONAL INJURY AND PROPERTY DAMAGE IF INSTRUCTIONS ARE
IGNORED.

CAUTION Caution is used to indicate the presence of a hazard that will or can cause minor personal injury
and property damage if instructions are ignored.

NOTICE Equipment damage possible


Notice is used when a hazard is present that can cause property damage but has absolutely no
personal injury risk.

Example: Disk drive will crash.

NOTE: Notes are used to provide important information about an item or a procedure.
Be sure to read the notes; the information contained in a note can often save you time or effort.

Chapter 1 - Introduction 1-3


GE HEALTHCARE LOGIQ C SERIES
DIRECTION 5266280-100, REVISION 3 PERIPHERALS INSTALLATION MANUAL

1-2-2 Standard Hazard Icons


Important information will always be preceded by the exclamation point contained within a triangle, as
seen throughout this chapter. In addition to text, several different graphical icons (symbols) may be used
to make you aware of specific types of hazards that could cause harm.

Table 1-3 Standard Hazard Icons

ELECTRICAL MECHANICAL RADIATION

LASER HEAT PINCH

LASER
LIGHT

Other hazard icons make you aware of specific procedures that should be followed.

Table 1-4 Standard Icons Indicating a Special Procedure Be Used

AVOID STATIC ELECTRICITY TAG AND LOCK OUT WEAR EYE PROTECTION

TAG
&
LOCKOUT EYE
PROTECTION
Signed Date

1-4 Section 1-2 - Important Conventions


GE HEALTHCARE LOGIQ C SERIES
DIRECTION 5266280-100, REVISION 3 PERIPHERALS INSTALLATION MANUAL

Section 1-3
Safety Considerations
1-3-1 Introduction
The following safety precautions must be observed during all phases of operation, service and repair of
this equipment. Failure to comply with these precautions or with specific warnings elsewhere in this
manual, violates safety standards of design, manufacture and intended use of the equipment.

1-3-2 Human Safety


Operating personnel must not remove the system covers.

Servicing should be performed by authorized personnel only.

Only personnel who have participated in a LOGIQ C Series Training Seminar are authorized to service
the equipment.

1-3-3 Mechanical Safety

Chapter 1 - Introduction 1-5


GE HEALTHCARE LOGIQ C SERIES
DIRECTION 5266280-100, REVISION 3 PERIPHERALS INSTALLATION MANUAL

1-3-4 Electrical Safety

1-3-4-1 Safe Practices


Follow these guidelines to minimize shock hazards whenever you are using the scanner;

• The system is equipped with a three-conductor AC power cable. This must be plugged into an
approved electrical outlet with protective ground. If an extension cord is used with the system, make
sure that the total current rating of the system does not exceed the extension cord rating.
• The power outlet used for this equipment should not be shared with other types of equipment.
• Both the system power cable and the power connector meet international electrical standards.

1-3-5 Dangerous Procedure Warnings


Warnings, such as the example below, precede potentially dangerous procedures throughout this
manual. Instructions contained in the warnings must be followed.

DANGER DANGEROUS VOLTAGES, CAPABLE OF CAUSING DEATH, ARE PRESENT IN THIS


EQUIPMENT. USE EXTREME CAUTION WHEN HANDLING, TESTING AND ADJUSTING.

WARNING EXPLOSION WARNING


DO NOT OPERATE THE EQUIPMENT IN AN EXPLOSIVE ATMOSPHERE. OPERATION
OF ANY ELECTRICAL EQUIPMENT IN SUCH AN ENVIRONMENT CONSTITUTES A
DEFINITE SAFETY HAZARD.

WARNING DO NOT SUBSTITUTE PARTS OR MODIFY EQUIPMENT


BECAUSE OF THE DANGER OF INTRODUCING ADDITIONAL HAZARDS, DO NOT
INSTALL SUBSTITUTE PARTS OR PERFORM ANY UNAUTHORIZED MODIFICATION
OF THE EQUIPMENT.

1-3-6 Returning/Shipping Probes and Repair Parts


Equipment being returned must be clean and free of blood and other infectious substances.

GEMS policy states that body fluids must be properly removed from any part or equipment prior to
shipment. GEMS employees, as well as customers, are responsible for ensuring that parts/equipment
have been properly decontaminated prior to shipment. Under no circumstance should a part or
equipment with visible body fluids be taken or shipped from a clinic or site (for example, body coils or
an ultrasound probe).

The purpose of the regulation is to protect employees in the transportation industry, as well as the
people who will receive or open this package.

1-6 Section 1-3 - Safety Considerations


GE HEALTHCARE LOGIQ C SERIES
DIRECTION 5266280-100, REVISION 3 PERIPHERALS INSTALLATION MANUAL

Section 1-4
EMC, EMI, and ESD
1-4-1 Electromagnetic Compatibility (EMC) and Interference (EMI)
Electromagnetic compatibility describes a level of performance of a device within its electromagnetic
environment. This environment consists of the device itself and its surroundings including other
equipment, power sources and persons with which the device must interface. Inadequate compatibility
results when a susceptible device fails to perform as intended due interference from its environment or
when the device produces unacceptable levels of emission to its environment. This interference is often
referred to as radio–frequency or electromagnetic interference (RFI/EMI) and can be radiated through
space or conducted over interconnecting power of signal cables. In addition to electromagnetic energy,
EMC also includes possible effects from electrical fields, magnetic fields, electrostatic discharge and
disturbances in the electrical power supply.

1-4-2 Electrostatic Discharge (ESD) Prevention

WARNING DO NOT TOUCH ANY BOARDS WITH INTEGRATED CIRCUITS PRIOR TO TAKING THE
NECESSARY ESD PRECAUTIONS:
1.) ALWAYS CONNECT YOURSELF, VIA AN ARM-WRIST STRAP, TO THE ADVISED
ESD CONNECTION POINT LOCATED ON THE REAR OF THE SCANNER (TO THE RIGHT
OF THE POWER CONNECTOR).
2.) FOLLOW GENERAL GUIDELINES FOR HANDLING OF ELECTROSTATIC SENSITIVE
EQUIPMENT.

1-4-3 CE Compliance
The LOGIQ C Series unit conforms to all applicable conducted and radiated emission limits and to
immunity from electrostatic discharge, radiated and conducted RF fields, magnetic fields and power line
transient requirements.

Applicable standard is: IEC 601–1–2.

NOTE: For CE Compliance, it is critical that all covers, screws, shielding, gaskets, mesh, clamps, are
in good condition, installed tightly without skew or stress. Proper installation following all
comments noted in this service manual is required in order to achieve full EMC performance.

Chapter 1 - Introduction 1-7


GE HEALTHCARE LOGIQ C SERIES
DIRECTION 5266280-100, REVISION 3 PERIPHERALS INSTALLATION MANUAL

Section 1-5
Customer Assistance
1-5-1 Contact Information
If this equipment does not work as indicated in this service manual or in the User Manual, or if you
require additional assistance, please contact the local distributor or appropriate support resource, as
listed below.

Prepare the following information before you call:

- System ID serial number, --see label on rear side of unit.


- Software version.

Table 1-5 Phone Numbers for Customer Assistance

Location Phone Number

USA
Service: On-site 1-800–437–1171
GE Healthcare
Ultrasound Service Engineering
Service: Parts 1-800-558-2040
9900 Innovation Drive
Wauwatosa, WI 53226
Applications support 1-800-682-5327 or 1-262-524-5698

Canada 1-800-668-0732

Service 1-800-321-7937
Latin America
Applications support 1-262-524-5698

Europe
GE Ultraschall Deutschland GmbH& Co. KG Phone: +33 (0)130-831-300 (General Imaging and Cardiac)
BeethovenstraBe 239
Postfach 11 05 60, D-42665 Solingen Fax: +49 (0)212-2802-431
Germany

Asia (Singapore)
Tel: +65 291-8528
GE Ultrasound Asia
Service Department - Ultrasound
298 Tiong Bahru Road #15-01/06
Central Plaza
Fax: +65 6291-7006
Singapore 169730

Phone: 81-42-648-2944
Japan Support Center
Fax: 81-42-648-2905

1-8 Section 1-5 - Customer Assistance


GE HEALTHCARE LOGIQ C SERIES
DIRECTION 5266280-100, REVISION 3 PERIPHERALS INSTALLATION MANUAL

1-5-2 System Manufacture


Table 1-6 System Manufacture

Manufacturer Phone Number

GE Healthcare (China) Co., Ltd.


TEL: +86 510-85225888
No. 19 Changjiang Road, Wuxi National Hi-Tech Development Zone, Jiangsu, P.R. China
FAX: +86 510-85226688
214028

1-5-3 Factory Site


Table 1-7 Factory Site

Factory Site Phone Number

GE Healthcare (China) Co., Ltd.


TEL: +86 510-85225888
No. 19 Changjiang Road, Wuxi National Hi-Tech Development Zone, Jiangsu, P.R. China
FAX: +86 510-85226688
214028

Wirpo GE Healthcare Pvt. Ltd., TEL: +91-80-41801000


No. 4, Kadugodi Industrial Area, Bangalore. 560067 Karnataka. India. FAX: +91-80-28452924

Chapter 1 - Introduction 1-9


GE HEALTHCARE LOGIQ C SERIES
DIRECTION 5266280-100, REVISION 3 PERIPHERALS INSTALLATION MANUAL

This page was intentionally left blank.

1 - 10 Section 1-5 - Customer Assistance


GE HEALTHCARE LOGIQ C SERIES
DIRECTION 5266280-100, REVISION 3 PERIPHERALS INSTALLATION MANUAL

Chapter 2
Preparation

Section 2-1
Overview
2-1-1 Purpose of Chapter 2
This chapter includes procedures to conduct before installation of the Peripherals.

Table 2-1 Contents in Chapter 2

Section Description Page Number


2-1 Overview 2-1

2-2 Power Shut Down 2-2


2-3 Furnished materials 2-3

Chapter 2 - Preparation 2-1


GE HEALTHCARE LOGIQ C SERIES
DIRECTION 5266280-100, REVISION 3 PERIPHERALS INSTALLATION MANUAL

Section 2-2
Power Shut Down

Table 2-2 Power Shut Down

Step Task Expected Result(s)


1. The SYSTEM - EXIT Menu appears on the screen.

Press the ON/OFF key on the Control Panel once to display the LOGIQC2/C2PRO/C3/C3
System - Exit Menu on the monitor. PRO/C5/C5PRO

Press this key.

LOGIQ C3/C5 Premium

The SYSTEM - EXIT Menu gives you the following choices:


LOGOFF
Use this button to log off the current user. (The system remains on
and ready for a new user to log on.)
SHUTDOWN
Use this button to shut down the system.
CANCEL
Use this button if you don’t want to log off or shut down the system.
2. Backend Processor will first turn OFF the Scanner activity.
Then it starts to shut down itself.
The time to turn down the Scanner including the Backend Processor,
Click SHUTDOWN on the SYSTEM-EXIT Menu
may vary from 10 seconds up to approximately 1 minute.
The last thing that shut down, is the light in the LCD displays,
indicating that you can continue to next step.

2-2 Section 2-2 - Power Shut Down


GE HEALTHCARE LOGIQ C SERIES
DIRECTION 5266280-100, REVISION 3 PERIPHERALS INSTALLATION MANUAL

Section 2-3
Furnished materials
2-3-1 Overview
This section describes the materials furnished with the Peripherals and with the System.

Retain the original carton and packing materials in case transport is needed in the future.

2-3-2 B/W USB Printer (Sony UP-D897)

2-3-2-1 Materials supplied with the B/W USB Printer

Table 2-3 Materials furnished with B/W USB Printer

Item Description Quantity Note


1. B/W USB Printer 1

2. Paper Roll 1

3. Operating Instructions 1

4. USB cable 1

5. Printer Convert Power cable 1

6. Printer Bracket 1

7. Printer_Bracket_Cushion 4
8. D2 Screw SJ2836-87 M3X8 4

2-3-2-2 Materials supplied with scanner

Table 2-4 B/W USB Printer: Materials furnished with the Scanner

Item Description Quantity Note


1. The Usb port located in the Printer box
USB port 1
is only for B/W printer.

2. The Power Output port located in the


Power Output port 1
Printer box is only for B/W printer.

Chapter 2 - Preparation 2-3


GE HEALTHCARE LOGIQ C SERIES
DIRECTION 5266280-100, REVISION 3 PERIPHERALS INSTALLATION MANUAL

2-3-3 Color USB Printer (UP-D23MD/HP470)

2-3-3-1 Materials supplied with the Color USB Printer

Table 2-5 Materials furnished with Color USB Printer

Item Description Quantity Note


1. Color USB Printer 1

2. Paper Roll 1

3. Operating Instructions 1

4. AC Power Cord 1

5. USB cable 1

2-3-3-2 Materials supplied with scanner

Table 2-6 Color USB Printer: Materials furnished with the Scanner

Item Description Quantity Note


1. USB ports 3
2. Power Output port 1 Located on Power Box

2-3-4 1 Pedal Footswitch (Whanam FSU2001)

2-3-4-1 Materials supplied with the Footswitch

Table 2-7 Materials furnished with the Footswitch

Item Description Quantity Note


1. Footswitch 1

2-3-4-2 Materials supplied in the scanner

Table 2-8 Footswitch: Materials furnished with the System

Item Description Quantity Note


1. USB port 3

2-4 Section 2-3 - Furnished materials


GE HEALTHCARE LOGIQ C SERIES
DIRECTION 5266280-100, REVISION 3 PERIPHERALS INSTALLATION MANUAL

2-3-5 Seagate Removable HDD

2-3-5-1 Materials supplied with the Seagate Removable HDD

Table 2-9 Materials furnished with the USB 2.0 HDD

Item Description Quantity Note


1. Seagate Removable HDD 1

2. USB Cable 1

2-3-5-2 Materials supplied in the scanner

Table 2-10 Seagate Removable HDD: Materials furnished with the scanner

Item Description Quantity Note


1. USB port 3

2-3-6 Peripherals/Accessories Connector Panel


LOGIQ C Series peripherals and accessories can be properly connected using the rear connector
panel.

2-3-6-1 Rear Panel Connector for LOGIQ C2/C2PRO,LOGIQ C3/C3PRO and LOGIQ C5/C5PRO
Located on the rear panel are Network port, 2 USB port, S-Video Out, VGA, Audio Out, Composite Out.

Figure 2-1 Rear Connector Panel

1.) Network Port


2.) 2 USB ports (Printers, Footswitch, etc.)
3.) S-Video Out
4.) VGA
5.) Audio Out
6.) Composite Out

NOTE: Audio Out is not supported by LOGIQ C2/C2 PRO.

Chapter 2 - Preparation 2-5


GE HEALTHCARE LOGIQ C SERIES
DIRECTION 5266280-100, REVISION 3 PERIPHERALS INSTALLATION MANUAL

2-3-6-2 Rear Panel Connector for LOGIQ C3 Premium and LOGIQ C5 Premium

5 6

Figure 2-2 Rear Connector Panel

1.) 2 USB ports (Printers, Footswitch, etc.)


2.) Network Port
3.) VGA
4.) Audio
5.) S-Video Out
6.) VGA
7.) Composite Out

2-3-6-3 Left Panel Connector


One USB Port is located on the left side panel.

Figure 2-3 Left Connector Panel

1.) 1 USB Port (Printers, Foot Switch, etc).

2-6 Section 2-3 - Furnished materials


GE HEALTHCARE LOGIQ C SERIES
DIRECTION 5266280-100, REVISION 3 PERIPHERALS INSTALLATION MANUAL

2-3-6-4 Located on the left side panel is one USB Port.


One USB port is Located on the Printer Box Rear Panel connector.

Figure 2-4 Printer Box Rear Connector panel

1.) 1 USB Ports only for B/W Printer

NOTICE Power on LOGIQ C Series to work first, then connect USB devices and power on USB devices.

NOTE: Each outer (case) ground line of peripheral/accessory connectors is protectively grounded.
Signal ground lines are not isolated.

Chapter 2 - Preparation 2-7


GE HEALTHCARE LOGIQ C SERIES
DIRECTION 5266280-100, REVISION 3 PERIPHERALS INSTALLATION MANUAL

This page was intentionally left blank.

2-8 Section 2-3 - Furnished materials


GE HEALTHCARE LOGIQ C SERIES
DIRECTION 5266280-100, REVISION 3 PERIPHERALS INSTALLATION MANUAL

Chapter 3
B/W USB Printer Installation
(Sony UP-D897)

Section 3-1
Overview
3-1-1 Purpose of Chapter 3
This chapter describes the installation procedure for the Sony UP-D897 B/W USB Printer.

Table 3-1 Contents in Chapter 3

Section Description Page Number


3-1 Overview 3-1
3-2 Installing the B/W USB Printer 3-2

B/W USB Printer

Figure 3-1 The B/W USB Printer on the LOGIQ C Series

Chapter 3 - B/W USB Printer Installation (Sony UP-D897) 3-1


GE HEALTHCARE LOGIQ C SERIES
DIRECTION 5266280-100, REVISION 3 PERIPHERALS INSTALLATION MANUAL

Section 3-2
Installing the B/W USB Printer
3-2-1 Tools
No special tools needed.

3-2-2 Manpower
One person 5 min.

3-2-3 Preparations
1.) Unpack the B/W USB Printer.

3-2-4 Installation procedure


1.) Place in a suitable place.
2.) Stick 4 Printer Bracket Cushions on the back of Printer Bracket, refer to Figure 3-2 on page 3-2 .

NOTE: The side with 2 juts is the back of Printer Bracket, the Printer Bracket Cushions should be stuck
on the back of Printer Bracket; the Printer should be fixed on the front of the Printer Bracket.

3
1

4 2

Figure 3-2 The B/W USB Printer installation page 1

3-2 Section 3-2 - Installing the B/W USB Printer


GE HEALTHCARE LOGIQ C SERIES
DIRECTION 5266280-100, REVISION 3 PERIPHERALS INSTALLATION MANUAL

3-2-4 Installation procedure (cont’d)


3.) Put the B/W Printer on the Printer Bracket and screw 4 screws (SJ2836-87 M3X8) to fix the B/W
Printer on the Printer Bracket, refer to Figure 3-3 on page 3-3 .

2 1

3 4

Figure 3-3 The B/W USB Printer installation page 2

4.) Connect the Power cord and USB Cable to the Printer Box and B/W printer, refer to Figure 3-
4 on page 3-3

USB Cable

Power Cable

Figure 3-4 The B/W USB Printer installation page 3

Chapter 3 - B/W USB Printer Installation (Sony UP-D897) 3-3


GE HEALTHCARE LOGIQ C SERIES
DIRECTION 5266280-100, REVISION 3 PERIPHERALS INSTALLATION MANUAL

3-2-4 Installation procedure (cont’d)

5.) Put the B/W Printer into the Printer Box, refer to Figure 3-5 on page 3-4

NOTE: When putting the B/W Printer into the Printer Box, the jut of the Printer Bracket should be put in
corresponding hole in the Printer Box, which will let B/W Printer fit well.

Figure 3-5 The B/W USB Printer installation page 4

3-4 Section 3-2 - Installing the B/W USB Printer


GE HEALTHCARE LOGIQ C SERIES
DIRECTION 5266280-100, REVISION 3 PERIPHERALS INSTALLATION MANUAL

Chapter 4
B/W USB Printer Installation
(Sony UP-897MD)

Section 4-1
Overview
4-1-1 Purpose of Chapter 4
This chapter describes the installation procedure for the Sony UP-897MD B/W Printer.

Table 4-1 Contents of Chapter 4

Section Description Page Number


4-1 Overview 4-3
4-2 Installing the B/W Printer 4-2

B/W Printer

Figure 4-6 B/W Printer Connection for LOGIQ C3 Premium and LOGIQ C5 Premium

NOTE: Sony UP-897MD B/W Printer is only available for LOGIQ C3 Premium and LOGIQ C5 Premium.

Chapter 4 - B/W USB Printer Installation (Sony UP-897MD) 4-1


GE HEALTHCARE LOGIQ C SERIES
DIRECTION 5266280-100, REVISION 3 PERIPHERALS INSTALLATION MANUAL

Section 4-2
Installing the B/W Printer
4-2-1 Tools
No special tools needed.

4-2-2 Manpower
One person 5 min.

4-2-3 Preparations
1.) Unpack the B/W Printer.

4-2-4 Installation procedure


1.) Connect the one side of the cable to Composite Port in Rear Panel Connector of the system and
the other side to the Video In Port of the the Printer.

Figure 4-7 Printer Connection

2.) Connect the Power Cable to AC IN Port of the Printer and the Electric Outlet, refer to Figure 3-2 on
page 3-2

Figure 4-8 Power Cable Connection

4-2 Section 4-2 - Installing the B/W Printer


GE HEALTHCARE LOGIQ C SERIES
DIRECTION 5266280-100, REVISION 3 PERIPHERALS INSTALLATION MANUAL

Chapter 5
Footswitch Installation

Section 5-1
Overview
5-1-1 Purpose of Chapter 5
This chapter describes how to install Footswitch. LOGIQ C Series supports 1-pedal footswitch and 3-
Pedal footswitch.

Table 5-2 Contents in Chapter 5


Section Description Page Number

5-1 Overview 5-1

5-2 Installing Footswitch 5-2

5-3 Configuring Footswtich 5-4

Chapter 5 - Footswitch Installation 5-1


GE HEALTHCARE LOGIQ C SERIES
DIRECTION 5266280-100, REVISION 3 PERIPHERALS INSTALLATION MANUAL

Section 5-2
Installing Footswitch
5-2-1 Tools
No special tools needed.

5-2-2 Manpower
One person 5 min.

5-2-3 Preparations
1.) Unpack the Footswitch.
2.) Ensure no physical damage.

5-2-4 Installation Procedure for 1-Pedal Footswitch (Whanam FSU2001)


1.) Connect to the USB ports located on the LOGIQ C Series left side or rear panel. (see the figure
below)
.

1-pedal
Footswitch

Figure 5-9 Footswitch connection

5-2 Section 5-2 - Installing Footswitch


GE HEALTHCARE LOGIQ C SERIES
DIRECTION 5266280-100, REVISION 3 PERIPHERALS INSTALLATION MANUAL

5-2-5 Installation Procedure for 1-Pedal Footswitch (Whanam FSU1000)


1.) Connect to the USB ports located on the LOGIQ C Series left side or rear panel. (see the figure
below)

Figure 5-10 Footswitch connection

5-2-6 Installation Procedure for 3-Pedal Footswitch


1.) Connect to the USB ports located on the LOGIQ C Series left side or rear panel. (see the figure
below

3-pedal
Footswitch

Figure 5-11 Footswitch connection

Chapter 5 - Footswitch Installation 5-3


GE HEALTHCARE LOGIQ C SERIES
DIRECTION 5266280-100, REVISION 3 PERIPHERALS INSTALLATION MANUAL

Section 5-3
Configuring Footswtich
Footswtich supports five configurations: Non function, acting as Freeze button, Print button, Store
button or B pause button.

NOTE: Acting as B pause button configuration of footswitch is not supported by LOGIQ C2/C2 PRO.

The default setting is acting as Freeze button; To change them, press Utility button on the key board
then select Application-->Settings. (see the figure below)

Figure 5-12 Default Setting

NOTE: The footswitch function is in Middle, please ignore the Left and right.

5-4 Section 5-2 - Installing Footswitch


GE HEALTHCARE LOGIQ C SERIES
DIRECTION 5266280-100, REVISION 3 PERIPHERALS INSTALLATION MANUAL

Chapter 6
Color USB Printer Installation

Section 6-1
Overview
6-1-1 Purpose of Chapter 6
This chapter describes the installation procedure for Color USB Printers. LOGIQ C Series supports two
types of printers for printing: Sony UP-D23MD Digital Color USB Printer, HP470 Deskjet Color Printer.

NOTE: Color Printer Sony UP-D23MD is only supported by LOGIQ C5/C5 PRO and LOGIQ C5
Premium; HP470b printer is supported by LOGIQ C2/C2 PRO, LOGIQ C3/C3 PRO ,LOGIQ C5
/C5 PRO,LOGIQ C3 Premium and LOGIQ C5 Premium.

For this two type printers, the connection to LOGIQ C Series is same. Here we take HP470b Deskjet
Color Printer as an example. Please make sure the printers are Power On.

Table 6-3 Contents in Chapter 6

Section Description Page Number


6-1 Overview 6-1

6-2 Installing the Color USB Printer 6-2

Color USB Printer

Figure 6-13 The Color USB Printer on the LOGIQ C Series

Chapter 6 - Color USB Printer Installation 6-1


GE HEALTHCARE LOGIQ C SERIES
DIRECTION 5266280-100, REVISION 3 PERIPHERALS INSTALLATION MANUAL

Section 6-2
Installing the Color USB Printer
6-2-1 Tools
No special tools needed.

6-2-2 Manpower
One person 5 min.

6-2-3 Preparations
1.) Unpack the color USB Printer.

6-2-4 Installation procedure


1.) Place in a suitable place.
2.) Connect power cord and USB cable on the Printer as shown in Figure 6-14 on page 6-2 .

USB cable
Power cord

Figure 6-14 Color Printer connection page 1

3.) Connect power cord and USB cable to the system as shown in Figure 6-15 on page 6-2

NOTE: The Color Printer Power cable should be connected to the Output Port on the Power Box.

USB cable

Power cord

Figure 6-15 Color Printer connection page 2

6-2 Section 6-2 - Installing the Color USB Printer


GE HEALTHCARE LOGIQ C SERIES
DIRECTION 5266280-100, REVISION 3 PERIPHERALS INSTALLATION MANUAL

6-2-4 Installation procedure (cont’d)


4.) The Color printer has been Installed successfully, see Figure 6-16 on page 6-3 .

Figure 6-16 Color Printer connection

Section 6-3
Setting for B/W and Color printer concurrent working
Example below points B/W printer (UP-D897) to print button on the keyboard and Color printer (HP470)
to Store button.

1.) Power on LOGIQ C Series system and wait until the system booting up.
1.) Connect B/W printer with USB ports of LOGIQ C Series and power on printer.
2.) Choose [Utility] -> [System] -> [Peripherals], press [Printer] button, then check the highlight printer
-- Sony UP-D897.

Figure 6-17 Peripherals

Chapter 6 - Color USB Printer Installation 6-3


GE HEALTHCARE LOGIQ C SERIES
DIRECTION 5266280-100, REVISION 3 PERIPHERALS INSTALLATION MANUAL

Section 6-3 Setting for B/W and Color printer concurrent working (cont’d)
3.) Choose [connectivity] -> [Service], select [Standard Print] in list, see Figure 6-18 on page 6-4 .
4.) Set standard printer as “Sony UP-D897”, change the name as “UP-D897” in properties area. Save
the setting.

Figure 6-18 Service Page

5.) Choose [Connectivity] -> [Button], Select [Print] in Button list, and then press “<<“ to remove all the
original setting.

Figure 6-19 Button page

6-4 Section 6-2 - Installing the Color USB Printer


GE HEALTHCARE LOGIQ C SERIES
DIRECTION 5266280-100, REVISION 3 PERIPHERALS INSTALLATION MANUAL

Section 6-3 Setting for B/W and Color printer concurrent working (cont’d)
6.) Select Sony UPD897, press “>>” to add to the Printflow View.
7.) [Print] setting for UP-D897 shows as below.

Figure 6-20 Print Button setting

8.) Connect Color printer (HP 470b) to the USB ports of LOGIQ C Series.
9.) Power on the Color printer.
10.)Choose [connectivity] -> [Service], select [Standard Print] in list, see Figure 6-21 on page 6-5 .
11.)Set standard printer as “HP Officejet H470 Series”, change the name as “HP Officejet H470 Series”
in properties area. Save the setting.

Figure 6-21 HP Officejet H470 Series

Chapter 6 - Color USB Printer Installation 6-5


GE HEALTHCARE LOGIQ C SERIES
DIRECTION 5266280-100, REVISION 3 PERIPHERALS INSTALLATION MANUAL

Section 6-3 Setting for B/W and Color printer concurrent working (cont’d)
12.)Go to [Connectivity] -> [Button], choose [Store], and then remove all the [Store] original setting.

Figure 6-22 Button page

13.)Select Stand Print, press “>>” to add to the Printflow View.


14.)[Store] setting for HP470 shows as below.

Figure 6-23 Store Button setting

6-6 Section 6-2 - Installing the Color USB Printer


GE HEALTHCARE LOGIQ C SERIES
DIRECTION 5266280-100, REVISION 3 PERIPHERALS INSTALLATION MANUAL

Chapter 7
SanDisk USB Stick Installation

Section 7-1
Overview
7-1-1 Purpose of Chapter 7
This chapter describes the installation procedure for SanDisk USB Stick.

Table 7-4 Contents in Chapter 7

Section Description Page Number


7-1 Overview 7-1

7-2 Installing the SanDisk USB Stick 7-2

SanDisk USB Stick

Figure 7-24 SanDisk USB Stick on the LOGIQ C Series

Chapter 7 - SanDisk USB Stick Installation 7-1


GE HEALTHCARE LOGIQ C SERIES
DIRECTION 5266280-100, REVISION 3 PERIPHERALS INSTALLATION MANUAL

Section 7-2
Installing the SanDisk USB Stick
7-2-1 Tools
No special tools needed.

7-2-2 Manpower
One person 1 min.

7-2-3 Preparations
1.) Unpack the SanDisk USB Stick

7-2-4 Installation procedure


Connect the SanDisk USB Stick with USB ports on the LOGIQ C Series as shown in the figure below.

Figure 7-25 SanDisk USB Stick connection

7-2 Section 7-2 - Installing the SanDisk USB Stick


GE HEALTHCARE LOGIQ C SERIES
DIRECTION 5266280-100, REVISION 3 PERIPHERALS INSTALLATION MANUAL

Chapter 8
ECG Installation

Section 8-1
Overview
8-1-1 Purpose of Chapter 8
This chapter describes the installation procedure for the ECG.

Table 8-5 Contents in Chapter 8

Section Description Page Number


8-1 Overview 8-1

8-2 Installing the ECG 8-2


8-3 ECG Intended Use 8-4

8-4 Patient Preparation 8-4

8-5 Technical Specifications 8-5

Figure 8-26 The ECG on the LOGIQ C5/C5 PRO

NOTE: Only LOGIQ C5/C5 PRO and LOGIQ C5 Premium supports ECG.

Chapter 8 - ECG Installation 8-1


GE HEALTHCARE LOGIQ C SERIES
DIRECTION 5266280-100, REVISION 3 PERIPHERALS INSTALLATION MANUAL

Section 8-2
Installing the ECG
8-2-1 Tools
No special tools needed.

8-2-2 Manpower
One person 1 min.

8-2-3 Preparations
1.) Unpack the ECG.
2.) Ensure no physical damage (refer to Figure 8-27 on page 8-2 ).

Figure 8-27 ECG

8-2 Section 8-2 - Installing the ECG


GE HEALTHCARE LOGIQ C SERIES
DIRECTION 5266280-100, REVISION 3 PERIPHERALS INSTALLATION MANUAL

8-2-4 Installation procedure


2.) Connect to the USB ports located on the LOGIQ C Series rear panel or left side.(see Figure 8-
28 on page 8-3 ).

Figure 8-28 ECG Installation

3.) Wait 5 seconds for ECG function start automatically. (Refer to Figure 8-29 on page 8-3 ).

Figure 8-29 ECG Mode

Chapter 8 - ECG Installation 8-3


GE HEALTHCARE LOGIQ C SERIES
DIRECTION 5266280-100, REVISION 3 PERIPHERALS INSTALLATION MANUAL

Section 8-3ECG Intended Use


ECG is intended to disclose either normal condition or patterns of arrhythmia, myocardial ischemia, rate
abnormalities, or features of prognostic value in the following cases:

• Patients with suspected cardiac abnormalities


• Populations of patients at an age or period in which a routine baseline evaluation of ECG
characteristics is desired.

Section 8-4Patient Preparation


The ECG traces quality depends very much on the stability and conductivity of the electrodes during the
test, especially during high stages when the patient movements can cause artifacts. Here are some
basic rules to ensure good electrical contact:

• Shave hair at the electrode contact points.


• Use a special shirt that attached the electrodes and lead wires to the body.
• Use high quality liquid gel electrodes.
• Make sure that the lead wires do not swing.
Detachable 3 leads each 100cm with clip connectors on the patient end designed for use with
disposable ECG electrodes. Furthermore, equip 3 repeatable limbs electrode clamps which can
connect with clip connector. Limbs electrodes are white, red and green (AHA\IEC).

Table 8-6 Patient cable marking

Lead Patient cable MArking Position on Patient

AHA IEC

RA R
I Right Arm
(White) (Red)

LA L
II Left Arm
(Black) (Yellow)

RL N
III Right Leg
(Green) (Black)

8-4 Section 8-2 - Installing the ECG


GE HEALTHCARE LOGIQ C SERIES
DIRECTION 5266280-100, REVISION 3 PERIPHERALS INSTALLATION MANUAL

Section 8-5Technical Specifications

Feature Specification

Size [mm] 85x45x22 mm (+/- 0.5mm)

Weight of ECG Module (gram) 50 (+5gr)

Weight of Patient cable (gram) 90 (+5gr)

Power consumption USB 5 V supply;

Current 80 mA ± 10%

Dynamic range +5mV to -5mV (refer to input)

ECG samples/sec 1000

A/D bits 12 (2.44 µV/LSB)

Defibrillation protection Built in

Patient Leads Detachable 3 leads (IEC or AHA)

CMMR > 80 dB

Input resistance > 10 Mohm

Signal snamic range 10mV (-5 to +5 mV)

Patient isolation: 4000V

Patient leakage current: < 10 µA

Frequency range (-3db) 0.05 - 150 Hz

Safety standard IEC 60601-1, EN 60601-1-2, IEC 60601-2-27, and EC11

Safety class IIa CF type

Operating temperature 0 to 70oC

Storing temperature (-65) to 150oC

Humidity < 80%

Table 8-7 Technical Specifications

Chapter 8 - ECG Installation 8-5


GE HEALTHCARE LOGIQ C SERIES
DIRECTION 5266280-100, REVISION 3 PERIPHERALS INSTALLATION MANUAL

This page was intentionally left blank.

8-6 Section 8-2 - Installing the ECG


GE HEALTHCARE LOGIQ C SERIES
DIRECTION 5266280-100, REVISION 3 PERIPHERALS INSTALLATION MANUAL

Chapter 9
Seagate Removable HDD Installation

Section 9-1
Overview
9-1-1 Purpose of Chapter 9
This chapter describes the installation procedure for Seagate Removable HDD.

Table 9-1 Contents in Chapter 9

Section Description Page Number


9-1 Overview 9-1

9-2 Installing the Seagate Removable HDD 9-2

Seagate Removable HDD

Figure 9-1 Seagate Removable HDD on the LOGIQ C Series

Chapter 9 - Seagate Removable HDD Installation 9-1


GE HEALTHCARE LOGIQ C SERIES
DIRECTION 5266280-100, REVISION 3 PERIPHERALS INSTALLATION MANUAL

Section 9-2
Installing the Seagate Removable HDD
9-2-1 Tools
No special tools needed.

9-2-2 Manpower
One person 1 min.

9-2-3 Preparations
1.) Unpack the Seagate Removable HDD.

9-2-4 Installation procedure


Connect the Seagate Removable HDD on the LOGIQ C Series as shown in Figure 9-2 on page 9-2 .

Figure 9-2 Seagate Removable HDD connection

9-2 Section 9-2 - Installing the Seagate Removable HDD


GE HEALTHCARE LOGIQ C SERIES
DIRECTION 5266280-100, REVISION 3 PERIPHERALS INSTALLATION MANUAL

Chapter 10
Set up and Check-out

Section 10-1
Overview
10-1-1 Purpose of Chapter 10
This chapter describes the Final Setup for the Peripherals and the Operational Check-out.

Table 10-1 Contents in Chapter 10

Section Description Page Number


10-1 Overview 10-1

10-2 B/W USB Printer 10-2


10-3 HP 470 office jet Printer 10-8

10-5 B/W USB Printer Checks 10-20

10-6 Color Printer Checks 10-20

10-7 Footswitch Checks 10-21


10-8 USB Stick Checks 10-21

Chapter 10 - Set up and Check-out 10 - 1


GE HEALTHCARE LOGIQ C SERIES
DIRECTION 5266280-100, REVISION 3 PERIPHERALS INSTALLATION MANUAL

Section 10-2
B/W USB Printer
10-2-1 Control assignment for B/W USB Printer
The B/W USB Printers are controlled from the Print on the System’s Control Panel. The factory setting
is Print for all USB Printer.

10-2-1-1 Control assignment procedure


The Control Assignment for the B/W USB Printers is done in the Utility Menu.

1.) Turn Peripherals on and power up the System.


2.) Press UTILITY on the Keyboard. The Utility window is displayed (see Figure 10-1 on page 10-2 ).

Figure 10-1 The Utility window (with the General/System sheet displayed)

10 - 2 Section 10-2 - B/W USB Printer


GE HEALTHCARE LOGIQ C SERIES
DIRECTION 5266280-100, REVISION 3 PERIPHERALS INSTALLATION MANUAL

10-2-1-1 Control assignment procedure (cont’d)


3.) Choose [Connectivity] -> [Service], select [Standard Print] in list, see Figure 10-2 on page 10-3 .
4.) Set standard printer as “Sony UP-D897”. Save the setting

Figure 10-2 The Connectivity window (with the Connectivity/Service sheet displayed)

5.) Select CONNECTIVITY and then choose BUTTON in the upper menu (see Figure 10-
3 on page 10-3 ).
6.) In the Button Field, select the Control Panel Key to configure: Print.
7.) Select Standard Print, press “>>” to add to the Printflow View. (see Figure 10-3 on page 10-3 )

Figure 10-3 The Connectivity window (with the Connectivity/Button sheet displayed)

8.) If another device is selected, use “<<“ to clear window.

Chapter 10 - Set up and Check-out 10 - 3


GE HEALTHCARE LOGIQ C SERIES
DIRECTION 5266280-100, REVISION 3 PERIPHERALS INSTALLATION MANUAL

10-2-2 Set Up Digital Peripheral


You set up digital peripherals from the Utility -> System -> Peripherals menu.
1. Click the Printers Button.
2. Left-click on the desired printer icon.
The following digital printer is supported: UP-D897 Sony printer.

Sony UP-D897 Printer

Sony UP-D897 Printer

Table-11: Sony UP-D897 Printer

Utility Screen Instructions

• Make sure that the settings match


the settings on this example.
• Click the Sharing tab.

10 - 4 Section 10-2 - B/W USB Printer


GE HEALTHCARE LOGIQ C SERIES
DIRECTION 5266280-100, REVISION 3 PERIPHERALS INSTALLATION MANUAL

Table-11: Sony UP-D897 Printer

Utility Screen Instructions

• The default printer settings is Not


shared; however, you can
configure the printer to be shared
if needed.
• Click the Ports tab.

• Make sure that the settings match


the settings on this example.
• Click the Advanced tab.

Chapter 10 - Set up and Check-out 10 - 5


GE HEALTHCARE LOGIQ C SERIES
DIRECTION 5266280-100, REVISION 3 PERIPHERALS INSTALLATION MANUAL

Table-11: Sony UP-D897 Printer

Utility Screen Instructions

• Make sure that the settings match


the settings on this example.
• Click the Security tab.

• Make sure that the settings match


the settings on this example.
• Click the General tab, then select
Print Preferences.

10 - 6 Section 10-2 - B/W USB Printer


GE HEALTHCARE LOGIQ C SERIES
DIRECTION 5266280-100, REVISION 3 PERIPHERALS INSTALLATION MANUAL

Table-11: Sony UP-D897 Printer

Utility Screen Instructions

• Select Portrait orientation.


• All Portrait and Landscape
orientation and paper size show
good image quality. The reason
portrait is recommended is
because this is the orientation
used for the analog printer.
• Select 1920x1280 from the Paper
Size drop-down menu.
• Click on Density Adjust.

• Change the Gamma setting to


Soft (Tone3), Advanced to -10
Dark to 0, Light to 0, and
Sharpness to 0.
• Click on the Save Button and save
your settings *.897 file (default
save location recommended).

Chapter 10 - Set up and Check-out 10 - 7


GE HEALTHCARE LOGIQ C SERIES
DIRECTION 5266280-100, REVISION 3 PERIPHERALS INSTALLATION MANUAL

Section 10-3
HP 470 office jet Printer
10-3-1 Control assignment for HP 470 officejet Printer
The HP 470 officejet Printer are controlled from the Print on the System’s Control Panel. The factory
setting is Print for all USB Printer.

10-3-1-1 Control assignment procedure


The Control Assignment for the color USB Printers is done in the Utility Menu.

1.) Turn Peripherals on and power up the System.


2.) Press UTILITY on the Keyboard.The Utility window is displayed (see Figure 10-4 on page 10-8 ).

Figure 10-4 The Utility window (with the General/System sheet displayed)

10 - 8 Section 10-3 - HP 470 office jet Printer


GE HEALTHCARE LOGIQ C SERIES
DIRECTION 5266280-100, REVISION 3 PERIPHERALS INSTALLATION MANUAL

10-3-1-1 Control assignment procedure (cont’d)


3.) Select PERIPHERALS in the upper menu (see Figure 10-5 on page 10-9 ) and press Printer.The
Peripherals window is displayed.

Figure 10-5 Peripherals Window

4.) Choose [Connectivity] -> [Service], select [Standard Print] in list, see Figure 10-6 on page 10-9 .
5.) Set standard printer as “HP Officejet H470 Series”. Save the setting.

Figure 10-6 Color Printer Set

6.) Select SAVE button.

Chapter 10 - Set up and Check-out 10 - 9


GE HEALTHCARE LOGIQ C SERIES
DIRECTION 5266280-100, REVISION 3 PERIPHERALS INSTALLATION MANUAL

10-3-1-1 Control assignment procedure (cont’d)


7.) Select CONNECTIVITY and then choose BUTTON in the upper menu (see Figure 10-
7 on page 10-10 ). The Button window is displayed.

Figure 10-7 Connectivity Window

8.) In the Button Field, select the Control Panel Key to configure: Print (see Figure 10-8 on page 10-10
).

Figure 10-8 The Button configuration screen

9.) Select Standard Print, press “>>” to add to the Printflow View.
10.)If another device is selected, use “<<“ to clear window.

10 - 10 Section 10-3 - HP 470 office jet Printer


GE HEALTHCARE LOGIQ C SERIES
DIRECTION 5266280-100, REVISION 3 PERIPHERALS INSTALLATION MANUAL

10-3-2 Set Up Digital Peripheral


You set up digital peripherals from the Utility -> System -> Peripherals menu.
1. Click the Printers Button.
2. Left-click on the desired printer icon.

The following digital printer is supported: HP 470 officejet printer.

HP 470 officejet Printer

HP 470 officejet Printer

Table-12: HP 470 officejet Printer

Utility Screen Instructions

• Make sure that the settings match


the settings on this example.
• Click the Printing Shortcuts tab.

Chapter 10 - Set up and Check-out 10 - 11


GE HEALTHCARE LOGIQ C SERIES
DIRECTION 5266280-100, REVISION 3 PERIPHERALS INSTALLATION MANUAL

Table-12: HP 470 officejet Printer

Utility Screen Instructions

• Make sure that the settings match


the setting on this example.
• Click the Features tab.

• Make sure that the settings match


the setting on this example.
• Click the Color tab.

10 - 12 Section 10-3 - HP 470 office jet Printer


GE HEALTHCARE LOGIQ C SERIES
DIRECTION 5266280-100, REVISION 3 PERIPHERALS INSTALLATION MANUAL

Table-12: HP 470 officejet Printer

Utility Screen Instructions

• Make sure that the settings match


the settings on this sample.

Chapter 10 - Set up and Check-out 10 - 13


GE HEALTHCARE LOGIQ C SERIES
DIRECTION 5266280-100, REVISION 3 PERIPHERALS INSTALLATION MANUAL

Section 10-4
Sony UP-D23 MD Printer
10-4-1 Control assignment for Sony UP-D23 MD Printer
The Sony UP-D23 MD Printer are controlled from the Print on the System’s Control Panel. The factory
setting is Print for all USB Printer.

10-4-1-1 Control assignment procedure


The Control Assignment for the color USB Printers is done in the Utility Menu.

1.) Turn Peripherals on and power up the System.


2.) Press UTILITY on the Keyboard.The Utility window is displayed (see Figure 10-4 on page 10-8 ).

Figure 10-9 The Utility window (with the General/System sheet displayed)

10 - 14 Section 10-4 - Sony UP-D23 MD Printer


GE HEALTHCARE LOGIQ C SERIES
DIRECTION 5266280-100, REVISION 3 PERIPHERALS INSTALLATION MANUAL

10-3-1-1 Control assignment procedure (cont’d)


3.) Select PERIPHERALS in the upper menu (see Figure 10-5 on page 10-9 ) and press.The
Peripherals window is displayed.

Figure 10-10 Peripherals Window

4.) Choose [Connectivity] -> [Service], select [Standard Print] in list, see Figure 10-6 on page 10-9 .
5.) Set standard printer as “Sony UP-D23MD”. Save the setting.

Figure 10-11 Color Printer Set

6.) Select SAVE button.

Chapter 10 - Set up and Check-out 10 - 15


GE HEALTHCARE LOGIQ C SERIES
DIRECTION 5266280-100, REVISION 3 PERIPHERALS INSTALLATION MANUAL

10-3-1-1 Control assignment procedure (cont’d)


7.) Select CONNECTIVITY and then choose BUTTON in the upper menu (see Figure 10-
7 on page 10-10 ). The Button window is displayed.

Figure 10-12 Connectivity Window

8.) In the Button Field, select the Control Panel Key to configure: Print (see Figure 10-8 on page 10-10
).

Figure 10-13 The Button configuration screen

9.) Select Standard Print, press “>>” to add to the Printflow View.
10.)If another device is selected, use “<<“ to clear window.

10 - 16 Section 10-4 - Sony UP-D23 MD Printer


GE HEALTHCARE LOGIQ C SERIES
DIRECTION 5266280-100, REVISION 3 PERIPHERALS INSTALLATION MANUAL

10-4-2 Set Up Digital Peripheral


You set up digital peripherals from the Utility -> System -> Peripherals menu.
1. Click the Printers Button.
2. Left-click on the desired printer icon.

The following digital printer is supported by LOGIQ C5/C5 PRO: UP-D23MD Sony printer.

Sony UP-D23 MD Printer

Sony UP-D23MD Printer

Table-13: Sony UP-D23MD Printer

Utility Screen Instructions

• Select Landscape orientation.


• Click the Paper tab.

Chapter 10 - Set up and Check-out 10 - 17


GE HEALTHCARE LOGIQ C SERIES
DIRECTION 5266280-100, REVISION 3 PERIPHERALS INSTALLATION MANUAL

Table-13: Sony UP-D23MD Printer

Utility Screen Instructions

• Make sure that UPC-21L is


highlighted.
• Click the Gray Balance Tab.

• Make sure that the settings match


the setting on this example.
• Click the Graphics tab.

10 - 18 Section 10-4 - Sony UP-D23 MD Printer


GE HEALTHCARE LOGIQ C SERIES
DIRECTION 5266280-100, REVISION 3 PERIPHERALS INSTALLATION MANUAL

Table-13: Sony UP-D23MD Printer

Utility Screen Instructions

• Make sure that the settings match


the settings on this sample.

Chapter 10 - Set up and Check-out 10 - 19


GE HEALTHCARE LOGIQ C SERIES
DIRECTION 5266280-100, REVISION 3 PERIPHERALS INSTALLATION MANUAL

Section 10-5
B/W USB Printer Checks
Make sure Printer Function is assigned to Print key. (The factory default is Print for all USB Printer.)

Table 10-1 B/W USB Printer checks

Step Task Expected Result(s)

When scanning in B mode, press “Freeze” to Image scanning stops with the last picture on the
1
stop image acquisition screen.

The image displayed on the screen is printed on B/


2 Press “Print” on the Keyboard. W USB Printer, depending on the key assignment
configuration.

3 Check if the print quality is of expected quality.

Section 10-6
Color Printer Checks
Make sure Printer Function is assigned to Print key. (The factory default is Print for all USB Printer.)

Table 10-2 Color USB Printer checks

Step Task Expected Result(s)

When scanning in B mode, press “Freeze” to Image scanning stops with the last picture on the
1
stop image acquisition screen.

The image displayed on the screen is printed on B/


2 Press “Print” on the Keyboard. W USB Printer, depending on the key assignment
configuration.

3 Check if the print quality is of expected quality.

10 - 20 Section 10-4 - Sony UP-D23 MD Printer


GE HEALTHCARE LOGIQ C SERIES
DIRECTION 5266280-100, REVISION 3 PERIPHERALS INSTALLATION MANUAL

Section 10-7
Footswitch Checks
Table 10-1 Footswitch Checks

Step Task Expected result(s)

Connect Footswitch with one USB port. When


Image scanning stops with the latest picture on
1 scanning in B mode, click the button of
the screen.
Footswitch once.

2 Click the button of Footswitch once again. Image scanning start.

Section 10-8
USB Stick Checks
Table 10-1 USB Stick Checks

Step Task Expected result(s)

"New hardware device detector" icon display


1 Connect USB Stick with the USB Port on the status bar

Check whether the device driver installed


2 Click the "New hardware device detector" icon
successfully

Chapter 10 - Set up and Check-out 10 - 21


GE HEALTHCARE LOGIQ C SERIES
DIRECTION 5266280-100, REVISION 3 PERIPHERALS INSTALLATION MANUAL

This page was intentionally left blank.

10 - 22 Section 10-4 - Sony UP-D23 MD Printer


GE HEALTHCARE

GE Healthcare: Telex 3797371


P.O. Box 414; Milwaukee, Wisconsin 53201, U.S.A.
(Asia, Pacific, Latin America, North America)

GE Ultraschall:Tel: +33 (0) 130 831 300


Deutschland GmbH & Co KG
Beethovenstrabe 239, Postfach 11 05 60
D-42655 Solingen, Germany

Vous aimerez peut-être aussi