LOGIQ P9/P7
Upgrade Kit Installation
Important Precautions
- i
GE PROPRIETARY TO GE LOGIQ P9/P7
DIRECTION 5736810, REVISION 6 UPGRADE KIT INSTALLATION
ii -
GE PROPRIETARY TO GE LOGIQ P9/P7
DIRECTION 5736810, REVISION 6 UPGRADE KIT INSTALLATION
- iii
GE PROPRIETARY TO GE LOGIQ P9/P7
DIRECTION 5736810, REVISION 6 UPGRADE KIT INSTALLATION
iv -
GE PROPRIETARY TO GE LOGIQ P9/P7
DIRECTION 5736810, REVISION 6 UPGRADE KIT INSTALLATION
- v
GE PROPRIETARY TO GE LOGIQ P9/P7
DIRECTION 5736810, REVISION 6 UPGRADE KIT INSTALLATION
vi -
GE PROPRIETARY TO GE LOGIQ P9/P7
DIRECTION 5736810, REVISION 6 UPGRADE KIT INSTALLATION
- vii
GE PROPRIETARY TO GE LOGIQ P9/P7
DIRECTION 5736810, REVISION 6 UPGRADE KIT INSTALLATION
(JA)
Traditional
Chinese
viii -
GE PROPRIETARY TO GE LOGIQ P9/P7
DIRECTION 5736810, REVISION 6 UPGRADE KIT INSTALLATION
(ZH-CN)
(KO)
- ix
GE PROPRIETARY TO GE LOGIQ P9/P7
DIRECTION 5736810, REVISION 6 UPGRADE KIT INSTALLATION
DAMAGE IN TRANSPORTATION
All packages should be closely examined at time of delivery. If damage is apparent write “Damage In
Shipment” on ALL copies of the freight or express bill BEFORE delivery is accepted or “signed for” by
a GE representative or hospital receiving agent. Whether noted or concealed, damage MUST be
reported to the carrier immediately upon discovery, or in any event, with
4 days after receipt, and the contents and containers held for inspection by the carrier. A transportation
company will not pay a claim for damage if an inspection is not requested within this 14 day period.
The purchaser of GE equipment shall only utilize qualified personnel (i.e., GE’s field engineers,
personnel of third-party service companies with equivalent training, or licensed electricians) to perform
electrical servicing on the equipment.
GE employees should contact the product's Lead Service Integrator (LSI) to report service
documentation issues. DO NOT use TrackWise for this purpose. The LSI will then use the internal
problem reporting tool to communicate these issues to the writer.
WARNING Use all Personal Protection Equipment (PPE) such as gloves, safety shoes, safety
glasses, and kneeling pad, to reduce the risk of injury.
For a complete review of all safety requirements, see the Chapter 1, Safety Considerations section of
the latest version of LOGIQ P9/P7 Proprietary Service Manual.
x -
GE PROPRIETARY TO GE LOGIQ P9/P7
DIRECTION 5736810, REVISION 6 UPGRADE KIT INSTALLATION
LEGAL NOTES
The contents of this publication may not be copied or duplicated in any form, in whole or in part, without
prior written permission of GE.
GE may revise this publication from time to time without written notice.
PROPRIETARY TO GE
Permission to use this Advanced Service Software and related documentation (herein called the
material) by persons other than GE employees is provided only under an Advanced Service Package
License relating specifically to this Proprietary Material. This is a different agreement from the one under
which operating and basic service software is licensed. A license to use operating or basic service
software does not extend to or cover this software or related documentation.
If you are a GE employee or a customer who has entered into such a license agreement with GE to use
this proprietary software, you are authorized to use this Material according to the conditions stated in
your license agreement.
However, you do not have the permission of GE to alter, decompose or reverse-assemble the software,
and unless you are a GE employee, you MAY NOT COPY the Material. The Material is protected by
Copyright and Trade Secret laws; the violation of which can result in civil damages and criminal
prosecution.
If you are not party to such a license agreement or a GE Employee, you must exit this Material now.
TRADEMARKS
All products and their name brands are trademarks of their respective holders.
COPYRIGHTS
All Material Copyright© 2017 by General Electric Company Inc. All Rights Reserved.
- xi
GE PROPRIETARY TO GE LOGIQ P9/P7
DIRECTION 5736810, REVISION 6 UPGRADE KIT INSTALLATION
Revision History
4 19 Jan 2017 Add note "Upgrade UDI label is not applicable to LOGIQ P7"
Rev History/LOEP 6
xii -
GE PROPRIETARY TO GE LOGIQ P9/P7
DIRECTION 5736810, REVISION 6 UPGRADE KIT INSTALLATION
Chapter 1
LOGIQ P9/P7 Upgrade Kit Installation
Section 1-1
LOGIQ P9/P7 UPGRADE KIT Installation
NOTE: Before shutting down the LOGIQ P9/P7 R2 system, please execute the backup procedure.
This upgrade kit installation manual is applicable to H42832LC and H42832LE.
2.) Unscrew 5 screws to remove the REAR CABLE COVER BRKT. Then disconnect all cables.
4.) Unscrew 7 screws to remove the NEST EMI BRKT and then remove the MCB ASSY.
Figure 1-4 Screw points on Nest Box and Separation Board Assy
Figure 1-7 Lifting up MPI and position of Stacking up connector on MPI assy
11.)Remove Stud on MCB assy and carefully lift up the MCB assy. Refer to following figure.
13.)Disconnect fan cable and carefully detach COM Express module. Refer to following figure.
NOTE: Be careful not to damage DDR3 4G SO-DIMM Memory from ESD. Carefully hold edge of DDR3
4G SO-DIMM Memory. Refer to following figure.
17.)Assemble the POWER CORD FIX BRKT, REAR COVER ASSY, REAR CABLE COVER
BRKT, CABLES, RS PROBE KNOBS, NEST EMI BACK BRKT, MPI Board Assy, MDC Board
Assy, MCB Board Assy, MCWD Board, COM Express and COM Express Bracket.
Reassembling procedure is reverse order of the removal procedure.
CAUTION When reassembling COM Express, Push the COM Express in both side of MCB Assy
like as below Figure 1-18
Figure 1-18 Push the COM Express in both side of MCB Assy
• Configuring BIOS
1.) Power ON the scanner and press “DEL” (If AN key is installed) or press left bottom key of trackball
(If there’s no AN Key) to enter BIOS.
If there’s no AN key, please use trackball to move menu in BIOS mode and click
NOTICEleft bottom key of trackball to select menu in BIOS and double-click to set or
change the value in BIOS. Right bottom key of trackball is ‘ESC’ in BIOS
Figure 1-19
NOTE: Check the installed memory amount on the windows system information.
7.) On "Maintenance Access" window, type in service password (user name not required).
NOTE: Service Password is rotating password that changes on 6 months period. Only the GE service
personnel or GE authorized dealers have access to maintenance password.
NOTE: Maintenance Password is function of date/time. Therefore, if BIOS settings differ from actual
date/time, then the maintenance password may not be recognized.
9.) Click the Start Menu and then Control Panel and System.
10.)Check that 8GB installed memory is shown as below Figure.
NOTE: The backup and restore procedures described in this section are divided into two parts. The first
part describes procedures to backup and restore patient image data using EZBackup/EZMove.
The second part describes procedures to backup and restore system and user-defined
configurations.
Depending on the system, you can use either a CD-R, DVD-R, USB Flash Drive, or USB Hard Disk for
system backup/restore. For the sake of simplicity, we have used the CD-R in the following examples.
NOTE: The system ONLY supports CD-R / DVD-R and DOES NOT support CD-RW / DVD+R.
1.) Prepare unformatted media or the USB HDD before starting EZBackup/EZMove.
2.) Specify the EZBackup/EZMove setup on the Utility-->System --> Backup/Restore page.
3.) Go to the Patient menu and select EZBackup/EZMove. The EZBackup/EZMove Wizard starts.
NOTE: If you use the USB HDD, some wizards and the pop-up messages DO NOT appear.
4.) Verify the information on the first page of the EZBackup/EZMove Wizard, Select Full backup
and then press Next button.
5.) Please check that backup span and perform the backup procedures.
NOTE: Patient Archive, Report Archive, Service does not compatible with LOGIQ P9/P7 R2 system.
The user should be executing backup and restore only User defined configuration data.
1.) Backup at Utility -> System -> Backup/Restore. Enable the only checkboxes User defined
configuration.
2.) Press the “Backup” button. -> Press “OK” to execute backup function.
Always keep a paper record of the settings for the LOGIQ P9/P7. Verify that it is current before starting
NOTICE software load!
a
1-1-3-2-1 Overview
See the following pages to record data.
These subsections include descriptions for recording/setting up data for the following screens.
With the system Powered On, follow the procedures below to document basic system information and
back up Presets, and if required, Patient Archive and Patient Images prior to software load.
Hardcopy forms are provided here for convenience, but content will be included if selected during
Backup process. Backing up this information is only required if an upgrade specifies the information be
backed up or as an extra precaution against loss due to a corrupt back-up disk.
Parameter Value
Computer Name
IP Settings
IP-Address
Subnet Mask
Default Gateway
MAC Address
Network Speed
Parameter Value
Network Authentication
Data Encryption
Network key
Key Index
EAP TYPE
Parameter Value
AE Title
Port Number
Table 1-4 Record settings for each device listed in Device table
Name IP Address
2.) Record all settings in Table 1-5 for each Destination Device and Service.
2.) Record all settings in Table 1-6 for each Destination Device and Service.
ON / OFF
ON / OFF
ON / OFF
ON / OFF
ON / OFF
ON / OFF
ON / OFF
ON / OFF
ON / OFF
ON / OFF
ON / OFF
ON / OFF
ON / OFF
ON / OFF
ON / OFF
ON / OFF
ON / OFF
Default Dataflow
Compression
Parameter Format Image Frames Compression Printflow View
Quality
Print1
Print2
Print3
Print4
Print5
Print6
PrintScreen
Still Images
Clip/Volume
Clips Add
ON/OFF
Multiframe Data
Quality
Volume File
1 - Standard DICOM / 2 - Standard DICOM with Raw
Format
Services (Types)
Destination Device Name Properties Name Properties
for each Destination Device
Patient/Exam Menu Optioans ON/OFF or Value Print and Store Options ON/OFF or Value
Use birthdate P[1-6] Key Sound
Export to Network Storage: Create Allow press and hold print key to
DICOMDIR replace and image
WARNING Make
WARNING sure that the Software Option Keys (alphanumeric passwords) have been
recorded correctly. If the key is incorrect, you will not be able to log on after the
software installation has been completed. The password is case sensitive. Hyphens
must also be recorded. There may be more than one password.
R2 Option key is not compatible with R1 Option key.
Group
User ID Prefix Last Name First Name Op Rights
Membership
EZMove
EZBackup
Parameter Value
Standard Printer
Default Printer
NOTE: DVR configuration values are not compatible with LOGIQ P9/P7 R2 system.
Figure 1-38
Parameter Value
Parameter Value
Parameter Value
Parameter Value
Parameter Value
B\B-Flow
CF\TVI\PDI
PW\CW\TVD
Software Version
Software Revision
Parameter Value
Hospital
Department
Preset Region
Language
Units
PoC
MyTraniner
Parameter Value
Single Screen
Figure 1-43
Parameter Value
OB Type
EFW GP
Parameter Value
OB Range Type
Parameter Value
Parameter Value
Parameter Value
Active Room
NOTE: Firmware upgrade ONLY supports in software 2.0.2 and later versions.
NOTE: Service Password is rotating password that changes on 6 months period. Only the GE service
personnel or GE authorized dealers have access to maintenance password.
NOTE: Maintenance Password is function of date/time. Therefore, if BIOS settings differ from actual
date/time then the maintenance password may not be recognized.
Figure 1-46
Figure 1-47
Figure 1-48
Figure 1-49
14.)Verify that chip configuration list appears after disappearing the chip configuration list in the first
open window. And then show the green light like the figures below. (if not, please reboot the
system).
Figure 1-50
Figure 1-51
Figure 1-52
Byte Value
0 2
1 2
2 72
3 27
4 44
• Back up the entire Patient Archive and Reports (you can import the database from previous version
into this version).
• Back up all presets, including Service.
• Record Connectivity settings (TCP/IP).
• Load Base System Software and install Application Software:
- Reload the entire drive (all partitions)
• The Hard Disk Drive (HDD) on LOGIQ P9/P7 R2 is divided into four visible partitions.
C: SYSTEM
D: USER
E: ARCHIVE
V: DVR
Z: REPOSIT
NOTE: Before loading the Base System Software, you must perform the following steps:
• Contact the customer to allow enough lead time to back up the system.
• Be sure no images are in the clipboard or opened exams.
• Review troubleshooting section to collect information for future analysis if you are troubleshooting
loss of image or patient data.
• Disconnect the system from the network and remove all transducers from the system.
• Save all User Presets.
NOTE: After a full or partial system software load, be sure to check the system date, time and time zone
are set correctly.
1.) Follow the procedures in section 8-3-7 Software Installation from DVD in service manual and
select [1] “Load the complete disk” during Base Image Software installation
2.) After loading the base image, follow the Application Software Installation procedures in service
manual.
Figure 1-53
Figure 1-54
NOTE: If yo do not have permission to access OKOS, or generate a software specific "Permanent key".
Please contact your On Line Center.
1.) Start the Internet explorer (VPN access to the GE Intranet is required) and open the OKOS -
Option Key Order System page: http://3.187.187.9/OKOS and login with your Username and
Password.
Figure 1-55
Figure 1-56
c.) Type the systems serial number (e.g., 208405SU1) into the "Search for" field.
d.) Click the SEARCH button. (The appearance of the screen changes again).
Figure 1-57
3.) From the "History" drop down menu select the line without demo.
Item Value
FDO no.
Figure 1-58
5.) Check the Existing options same as option names in Table 1-12.
6.) Check the Upgrade to LOGIQ P9/P7 R2 check box. And Enter recorded FDO no.
Figure 1-59
9.) Print out or write down the "Permanent Key" and then click OK.
Item Value
Permanent Key
NOTE: NOTE: R1 OSD control values are not compatible with R2 OSD control values. See the section
6-3 LCD Monitor Adjustment in service manual to refer R2 monitor control in detail.
NOTE: Please refer to the table below to compare R1 and R2 Color Temp. values.
NOTE: The calculation for the number of backup CDs is only an estimate. Allow for one additional CD
when performing an EZBackup/EZMove.
You can review backed up media via the Patient Menu, Import, and the DICOM Read dataflow.
1.) Select the patient on the Patient Menu (on the same system where the EZMove was performed).
2.) Insert the media volume indicated on the Patient->Data Transfer Menu and Select the backup CD
in “Form” data source menu.
3.) View the exam from the media and Press “Transfer” button to execute Import image data.
NOTE: Use Import, from as many media as you have for that patient. However, take care not to import
studies over existing studies; duplicate or missing images may result. Delete the existing exam
first.
NOTE: Results of the tests have to be always submitted with this section; Information of Parts and Software
you used is necessary for the test results.
Figure 1-60 Attach the LOGIQ P9 R1 to R2 UPG UDI label or LOGIQ P9 R1 to R2 UPG UDI label 2 on the left
bottom side of the system
Chapter 2
Functional Check after Software Installation
Section 2-1
Software Load Functional Tests
2-1-1 Power ON/Boot up
NOTE: If the system has been already turned on after upgrading software and have been up and
running normally, this test result can be “Pass”.
Results
Steps Expected Outcomes Comments
(Pass/Fail/NA)
• Connect the probe to one of probe ports.
• The system boots up
• Press once on the ON/OFF button on the
successfully. Pass / Fail / NA
operator panel to boot the system.
• Probe marker can be seen.
• After booting up and see the main screen.
Results
Steps Expected Outcomes Comments
(Pass/Fail/NA)
• From the touch panel, select Utility ->
System-> About. • Software Version is the
same as the number written Pass / Fail / NA
• See “Software Version” in Software group,
in 1-1-5-6-1
and note the version _______
• See “Image Part Number” in System Image, • Part Number is the same as
and note the version _______ the part number and Pass / Fail / NA
revision written in 1-1-5-6-1
Results
Steps Expected Outcomes Comments
(Pass/Fail/NA)
• Make sure that the probe is connected
properly. • Starts B-mode scan. Pass / Fail / NA
• Select the probe on the touch panel.
• Press [M] key on the operator panel. • Starts M-mode scan. Pass / Fail / NA
• Press [PW] on the operator panel. • Starts B/CFM/PWD scan. Pass / Fail / NA
Results
Steps Expected Outcomes Comments
(Pass/Fail/NA)
• Make sure that the probe, which supports
B-Flow, is connected properly.
• From the touch panel, select Utility ->
Imaging -> General.
• Change the BF/BFC button to “BF”.
• Starts B-Flow scan. Pass / Fail / NA
2-1-5 CW (Optional)
Results
Steps Expected Outcomes Comments
(Pass/Fail/NA)
• Make sure that the probe, which supports
CWD, is connected properly.
• Starts CW scan. Pass / Fail / NA
• Select the probe on the touch panel and
press [CW] key on the operator panel.
Results
Steps Expected Outcomes Comments
(Pass/Fail/NA)
• Connect a service dongle to USB port and
press a power button on console panel.
• Select Exit and enter the user name and
password. Choose Maintenance > Exit to
Windows. The Windows desktop will
appear. • NETGEAR A6210 WiFi
Pass / Fail / NA
USB3.0 Adapter is listed.
• Click the Device Manager icon on the status
bar. A Device Manager window will open.
NOTE: This test is only applied if the DVR option has been installed.
Results
Steps Expected Outcomes Comments
(Pass/Fail/NA)
• Make sure “Show Video Tab” is selected in
Utility > Application > Settings > Video. • Video tab is displayed in
touch panel. Pass / Fail / NA
• See the touch panel.
2-1-8 4D (Optional)
NOTE: This test is only applied if the 4D options (both hardware and software) have been installed.
Results
Steps Expected Outcomes Comments
(Pass/Fail/NA)
• Prepare a 4D probe that the version of
software has supported.
• Make sure that the probe is connected
properly.
• Select the probe on the touch panel. • Static 3D image is acquired. Pass / Fail / NA
• Press 3D/4D key
• Press Static 3D
• Press Start Set key
NOTE: This test is only applied if the Battery option has been installed.
Results
Steps Expected Outcomes Comments
(Pass/Fail/NA)
• Battery icon is displayed in
• Make sure that the Battery option is
status bar and the color is Pass / Fail / NA
installed.
green or yellow, not orange.
Figure 2-1
Figure 2-2
Figure 2-3
Figure 2-4
Results
Steps Expected Outcomes Comments
(Pass/Fail/NA)
• Press Power ON/OFF button on the • “SYSTEM EXIT” dialog
Pass / Fail / NA
operator panel to display Exit dialog. appears.
Section 2-2
Software Load Functional Checks - Results Sheet
Console Serial Number: _________________________________
Test Result
www.gehealthcare.com