CT/i Options Installation
i
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100
ii
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100
The purchaser of GE equipment shall only utilize qualified personnel (i.e., GE’s
field engineers, personnel of third–party service companies with equivalent
training, or licensed electricians) to perform electrical servicing on the
equipment.
IMPORTANT...X–RAY PROTECTION
X–ray equipment if not properly used may cause injury. Accordingly, the
instructions herein contained should be thoroughly read and understood by
everyone who will use the equipment before you attempt to place this equipment
in operation. The General Electric Company, Medical Systems Group, will be glad
to assist and cooperate in placing this equipment in use.
The equipment is sold with the understanding that the General Electric Company,
Medical Systems Group, its agents, and representatives have no responsibility
for injury or damage which may result from improper use of the equipment.
Various protective materials and devices are available. It is urged that such
materials or devices be used.
iii
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100
DAMAGE IN TRANSPORTATION
All packages should be closely examined at time of delivery. If damage is
apparent write ‘‘Damage In Shipment’’ on ALL copies of the freight or express bill
BEFORE delivery is accepted or ‘‘signed for’’ by a GE representative or hospital
receiving agent. Whether noted or concealed, damage MUST be reported to the
carrier immediately upon discovery, or in any event, within 14 days after receipt,
and the contents and containers held for inspection by the carrier. A
transportation company will not pay a claim for damage if an inspection is not
requested within this 14 day period.
Call Traffic and Transportation, Milwaukee, WI (414) 785 5052 or 8*323 5052
immediately after damage is found. At this time be ready to supply name of
carrier, delivery date, consignee name, freight or express bill number, item
damaged and extent of damage.
iv
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 9 DIRECTION 2181737–100
REVISION HISTORY
REV DATE REASON FOR CHANGE
0 . . . . . . . . . . . . . . . . . . 28 February 1997 . . . . . Initial Release of document.
1 . . . . . . . . . . . . . . . . . . 13 May 1997 . . . . . . . . Added Bar Code Reader Option for ConnectPRO, 2nd Disk, Smart Beam
Software, Language Option & Dentascan Installation (Chaps.2 through
6).
2 . . . . . . . . . . . . . . . . . . 27–OCT–1997 . . . . . . . Added Smart Helical Option Installation (Chapter 7).
3 . . . . . . . . . . . . . . . . . . 03–DEC–1997 . . . . . . . Added BMD Option Installation (Chapter 8). Other miscellaneous
corrections & format modifications throughout.
4 . . . . . . . . . . . . . . . . . . 18–May–1998 . . . . . . . Chapters 9, 10, 11 & 12 added to support CT/i (Octane) Computer Op-
tions (memory, dual cpu, optional disk and mag tape)
5 . . . . . . . . . . . . . . . . . . 27–January–1999 . . . . . Chapter 8 (BMD) modified to support the new BMD Option Kit struc-
ture for 4.1 & 5.3 system software.
6 . . . . . . . . . . . . . . . . . . 22–April–1999 . . . . . . . CQA 992109.
7 . . . . . . . . . . . . . . . . . . 20–August–1999 . . . . . CQA’s 993783 & 995718.
8 . . . . . . . . . . . . . . . . . . 27–September–2000 . . Revised for software releases 4.4 & 6.1.
9 . . . . . . . . . . . . . . . . . . 11 June 2002 . . . . . . . . . CQA 1021005: Added 2nd disk option procedure.
PAGE REV PAGE REV PAGE REV PAGE REV PAGE REV
Cover/Blank . . . . 9 3–1 to 3–8 . . . . . . 4 7–2 to 7–3 . . . . . . 8 12–6 to 12–8 . . . . 8 ................
Warnings–i to iv 4 4–i to 4–ii . . . . . . . 4 7–4 to 7–6 . . . . . . 4 12–9 to 12–12 . . . 4 ................
History–v to vi . . 9 4–1 . . . . . . . . . . . . 4 8–i to 8–ii . . . . . . . 5 13-1 to 13-2 . . . . 9
TOC–vii to xv . . 5 4–2 to 4–3 . . . . . . 8 8–1 to 8–6 . . . . . . 5 Rear . . . . . . . . . . . 4
TOC xvi . . . . . . . 9 4–4 to 4–6 . . . . . . 4 9–i to 9–ii . . . . . . . 4 ................ ................
1–i to 1–ii . . . . . . 4 5–i to 5–ii . . . . . . . 4 9–1 to 9–16 . . . . . 4 ................ ................
1–1 to 1–8 . . . . . 4 5–1 to 5–4 . . . . . . 4 10–i to 10–ii . . . . . 4 ................ ................
2–i . . . . . . . . . . . 8 6–i to 6–ii . . . . . . . 4 10–1 to 10–16 . . . 4 ................ ................
2–ii . . . . . . . . . . . 4 6–1 to 6–2 . . . . . . 8 11–i to 11–ii . . . . . 4
2–1 to 2–5 . . . . . 4 6–3 to 6–4 . . . . . . 4 11–1 to 11–12 . . . 7 ................ ................
2–6 to 2–9 . . . . . 8 7–i to 7–ii . . . . . . . 4 12–i to 12–ii . . . . . 4 ................ ................
2–10 to 2–12 . . . 4 7–1 . . . . . . . . . . . . 4 12–1 to 12–5 . . . . 4
3–i to 3–ii . . . . . . 4
v
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 9 DIRECTION 2181737–100
NOTES
vi
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 5 DIRECTION 2181737–100
Table Of Contents
Chapter 1
CT/i CPU Memory Option . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.i
Section 1
CT/i CPU Memory Option Installation Preparation . . . . . . . . . . . . . . . . . 1.1
1.1 Shutdown the System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.1
1.2 Remove SGI Chassis . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.1
Section 2
Install Optional Memory SIMMs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.4
2.1 Install SIMMs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.4
2.2 Reinstall SGI Chassis . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.4
Section 3
Startup System and Verify Installed Memory . . . . . . . . . . . . . . . . . . . . . . . 1.6
3.1 Start System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.6
3.2 Verify Installed Memory . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.6
Section 4
Finish Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.7
4.1 Install Rating Plate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.7
4.2 Return Installation Card . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.7
Section 5
Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.8
5.1 B7510T Additional DPI Memory Option Kit Parts List . . . . . . . . . . . . 1.8
vii
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 5 DIRECTION 2181737–100
Chapter 2
CT/i ConnectPRO and Bar Code Reader Option . . . . . . . . . . . . . . . . . . . 2.i
Section 1
CT/i ConnectPRO Option Installation Preparation . . . . . . . . . . . . . . . . . . 2.1
1.1 Bar Code Reader Kit Hardware Installation . . . . . . . . . . . . . . . . . . . . . . 2.1
1.2 Console Hardware Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.2
1.3 Verify Bar Code Scanner Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.4
Section 2
ConnectPRO Software Option Installation & Checkout . . . . . . . . . . . . . . 2.6
2.1 Install and Setup ConnectPRO Option from MOD . . . . . . . . . . . . . . . . 2.6
2.2 ConnectPRO Feature Checkout . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.9
2.3 Save System State . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.9
Section 3
Finish Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.10
3.1 Install Rating Plate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.10
3.2 Return Installation Card . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.10
Section 4
Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.11
4.1 B7540RB Bar Code Reader Only Option Kit Parts List . . . . . . . . . . . . 2.11
4.2 B7500PM HIS/RIS Interface with Bar Code Reader Kit Parts List . . . . 2.11
4.3 Material List for Bar Code Reader Kit 2180690 . . . . . . . . . . . . . . . . . . 2.11
viii
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 5 DIRECTION 2181737–100
Chapter 3
CT/i Additional System Magnetic Disk Option Installation . . . . . . . . . . . 3.i
Section 1
Prepare for CT/i Additional Magnetic Disk Option Installation . . . . . . . 3.1
1.1 Prepare for Option Disk Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.1
1.2 Save System State Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.1
1.3 Shutdown the System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.2
Section 2
Install Additional Magnetic Disk Option Hardware . . . . . . . . . . . . . . . . . 3.3
2.1 Power Off Console and Install Optional Disk . . . . . . . . . . . . . . . . . . . . 3.3
2.2 Verify Hardware Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.3
Section 3
Load CT/i Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.6
3.1 Install Option Disk Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.6
Section 4
Finish Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.6
4.1 Install Rating Plate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.6
4.2 Return Installation Card . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.6
Section 5
Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.7
5.1 B7510S CT/i Additional System Magnetic Disk Option Kit Parts List . 3.7
ix
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 5 DIRECTION 2181737–100
Chapter 4
CT/i Smart Beam Option Software Installation . . . . . . . . . . . . . . . . . . . . 4.i
Section 1
CT/i Optional Smart Beam Software Installation . . . . . . . . . . . . . . . . . . . 4.1
1.1 Check Installed Collimator Type . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4.1
1.2 Add Smart Beam Software Option . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4.2
Section 2
Calibrate System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4.4
Section 3
Finish Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4.4
3.1 Install Rating Plate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4.4
3.2 Return Installation Card . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4.4
Section 4
Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4.5
4.1 B7530SB CT/i Smart Beam Software (w/o Collimator) Kit Parts List 4.5
4.2 B7530SC CT/i Smart Beam Collimator Option Kit Parts List . . . . . . . 4.5
4.3 B7530SS CT/i Smart Beam Software (with Collimator) Parts List . . . . 4.5
Chapter 5
CT/i Language Options Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.i
Section 1
Language Option Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.1
1.1 Shutdown the System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.1
1.2 Boot System and Stop For Maintenance . . . . . . . . . . . . . . . . . . . . . . . . 5.1
1.3 If you have a keyboard other than a US–English Keyboard. . . . . . . . . . 5.1
Section 2
Procedures for installing add–on packages . . . . . . . . . . . . . . . . . . . . . . . . . 5.2
2.1 Optional Package Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.2
2.2 Reconfig to Select Language . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.3
2.3 Reconfig SBC to Enable Language . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.3
Section 3
Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.5
3.1 Material Lists for International Language Kits . . . . . . . . . . . . . . . . . . . 5.5
x
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 5 DIRECTION 2181737–100
Chapter 6
CT/i Dentascan Option Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.i
Section 1
Install Dentascan Option Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.1
1.1 Add Software Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.1
Section 2
Procedures for installing Dentascan filming option . . . . . . . . . . . . . . . . . . 6.3
Section 3
Finish Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.3
3.1 Install Rating Plate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.3
3.2 Return Installation Card . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.4
Section 4
Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.4
4.1 B7540N CT/i Dentascan Option Kit Parts List . . . . . . . . . . . . . . . . . . . 6.4
Chapter 7
CT/i Smart Helical Option Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . 7.i
Section 1
Install CT/i 4.X System Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7.1
1.1 Install 4.X Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7.1
Section 2
PROCEDURES FOR INSTALLING SMART HELICAL SOFTWARE . 7.2
Section 3
Finish Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7.4
3.1 Install Rating Plate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7.4
3.2 Return Installation Card . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7.4
Section 4
Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7.5
4.1 B7530TL Smart Helical Option Kit Parts List (systems not at 4.X) . . . 7.5
4.2 B7530TK Smart Helical Option Kit Parts List (systems at 4.X) . . . . . . 7.5
xi
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 5 DIRECTION 2181737–100
Chapter 8
CT/i BMD Option Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8.i
Section 1
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8.1
1.1 " ! !# . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8.1
Section 2
PROCEDURES FOR INSTALLING BMD OPTION SOFTWARE . . . . 8.1
2.1 Install BMD Option Key and Load BMD Option Software . . . . . . . . . . 8.1
Section 3
Special Instructions for Transferring the BMD Database from an Advantage
Windows Workstation to the CT/i . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8.2
3.1 Transfer the BMD Database from AWW to Ct/i . . . . . . . . . . . . . . . . . . 8.2
3.2 Verify that the data has been successfully transferred . . . . . . . . . . . . . . 8.3
Section 4
Backup Database and Deliver Books & Phantoms to Customer . . . . . . . . 8.3
4.1 Backup BMD Database (Save To Mod) . . . . . . . . . . . . . . . . . . . . . . . . . 8.3
4.2 Deliver Books & BMD Phantom to Customer . . . . . . . . . . . . . . . . . . . . 8.3
Section 5
Finish Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8.4
5.1 Install Rating Plate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8.4
5.2 Return Installation Card . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8.4
Section 6
Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8.5
6.1
!
! ! ! # ! . 8.5
6.2
! !
! ! ! #! ! 8.5
xii
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 5 DIRECTION 2181737–100
Chapter 9
CT/i (OCTANE) Memory Option . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9.i
Section 1
CT/i (OCTANE) CPU Memory Option Installation Preparation . . . . . . . 9.1
1.1 Shutdown the System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9.1
1.2 Preparing the Workstation to Install the optional memory . . . . . . . . . . . 9.1
1.3 Removing the System Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9.3
Section 2
Installing and Removing Memory . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9.6
2.1 About Memory . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9.6
2.2 Installing Memory . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9.8
2.3 Replacing the System Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9.8
2.4 Powering On the OCTANE Workstation . . . . . . . . . . . . . . . . . . . . . . . . 9.11
Section 3
Startup System and Verify Installed Memory . . . . . . . . . . . . . . . . . . . . . . . 9.12
3.1 Start System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9.12
3.2 Verify Installed Memory . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9.12
Section 4
Finish Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9.13
4.1 Install Rating Plate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9.13
4.2 Return Installation Card . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9.14
Section 5
Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9.15
5.1 Field Replaceable Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9.15
5.2 B7710M Memory Option Kit – Parts List . . . . . . . . . . . . . . . . . . . . . . . 9.15
xiii
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 5 DIRECTION 2181737–100
Chapter 10
CT/i (OCTANE) Dual Processor Option . . . . . . . . . . . . . . . . . . . . . . . . . . 10.i
Section 1
CT/i OCTANE Dual Processor Option Installation Preparation . . . . . . . 10.1
1.1 Shutdown the System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10.1
1.2 Removing the System Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10.2
Section 2
Installing The Optional Dual Processor . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10.6
2.1 Removing The Single Processor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10.6
2.2 Installing The Optional Dual Processor . . . . . . . . . . . . . . . . . . . . . . . . . 10.7
2.3 Replacing the System Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10.9
2.4 Powering On the OCTANE Workstation . . . . . . . . . . . . . . . . . . . . . . . . 10.12
Section 3
Startup System and Verify Installed Dual Processor . . . . . . . . . . . . . . . . . 10.13
3.1 Start System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10.13
3.2 Verify Installed Dual Processor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10.13
Section 4
Finish Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10.14
4.1 Install Rating Plate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10.14
4.2 Return Installation Card . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10.15
Section 5
Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10.16
5.1 Field Replaceable Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10.16
5.2 B7710JP Dual Processor Option Kit – Parts List . . . . . . . . . . . . . . . . . . 10.16
xiv
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 5 DIRECTION 2181737–100
Chapter 11
CT/i (OCTANE) Additional Magnetic Disk Option Installation . . . . . . . 11.i
Section 1
Prepare for CT/i (Octane) Additional Magnetic Disk Option Installation 11.1
1.1 Prepare for Option Disk Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . 11.1
1.2 Save System State Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11.1
1.3 Shutdown the System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11.2
Section 2
Install Additional Magnetic Disk Option Hardware . . . . . . . . . . . . . . . . . 11.3
2.1 About Internal Drives . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11.3
2.2 Power Off Console and Install Optional Disk . . . . . . . . . . . . . . . . . . . . 11.3
2.3 Removing the Bezel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11.4
2.4 Installing an Internal Option Drive . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11.7
2.5 Removing the Blank Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11.8
2.6 Replacing the Bezel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11.10
2.7 Verify Hardware Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11.11
Section 3
Load CT/i Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11.11
3.1 Install Option Disk Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11.11
Section 4
Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11.11
4.1 B7710JD CT/i (Octane) Additional Sys. Mag Disk Option Kit . . . . . . . 11.11
xv
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 9 DIRECTION 2181737-100
Chapter 12
CT/i (Octane) ConnectPRO and Bar Code Reader Option . . . . . . . . . . . 12.i
Section 1
CT/i (Octane) ConnectPRO Option Installation Preparation . . . . . . . . . . 12.1
1.1 Bar Code Reader Kit Hardware Installation . . . . . . . . . . . . . . . . . . . . . . 12.1
1.2 Console Hardware Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12.1
1.3 Verify Bar Code Scanner Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12.4
Section 2
ConnectPRO Software Option Installation & Checkout . . . . . . . . . . . . . . 12.6
2.1 Install and Setup ConnectPRO Option from MOD . . . . . . . . . . . . . . . . 12.6
2.2 ConnectPRO Feature Checkout . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12.9
2.3 Save System State . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12.9
Section 3
Finish Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12.10
3.1 Install Rating Plate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12.10
3.2 Return Installation Card . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12.10
Section 4
Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12.11
4.1 B7540RB Bar Code Reader Only Option Kit Parts List . . . . . . . . . . . . 12.11
4.2 B7500PM HIS/RIS Interface w/Bar Code Reader Option Kit Parts List 12.11
4.3 Material List for Bar Code Reader Kit 2180690 . . . . . . . . . . . . . . . . . . 12.11
Chapter 13
2nd Host Disk Option . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13.1
xvi
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 5 DIRECTION 2181737–100
List Of Illustrations
Illustration Page
Illustration 1–1
SGI Chassis Cover Removal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.2
Illustration 1–2
Disconnect SGI Midplane Power Cables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.3
Illustration 1–3
SGI Chassis Cover Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.4
Illustration 1–4
CT/i Host Computer Memory Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.5
Illustration 1–5
Rating Plate Location . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.7
Illustration 2–1
Bar Code Reader Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.2
Illustration 2–2
Bar Code Reader Installation (detail) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.3
Illustration 2–4
Options Window . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.6
Illustration 2–5
Sample Software Options Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.7
Illustration 2–6
ConnectPRO Setup screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.8
Illustration 2–7
Rating Plate Location . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.10
Illustration 3–1
Option Disk Installation Drawing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.4
Illustration 3–2
User Drive Assembly Exploded View . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.5
Illustration 3–3
Rating Plate Location . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.6
Illustration 4–1
Software Options Installation Window . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4.2
Illustration 4–2
Software Options Available / Installed Screen Example . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4.3
Illustration 4–3
Rating Plate Location . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4.4
Illustration 6–1
Software Options Installation Window . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.1
Illustration 6–2
Software Options Available / Installed Screen Example . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.2
Illustration 6–3
Rating Plate Location . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.3
xvii
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 5 DIRECTION 2181737–100
Illustration Page
Illustration 7–1
Software Options Installation Window . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7.2
Illustration 7–2
Software Options Available / Installed Screen Example . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7.3
Illustration 7–3
Rating Plate Location . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7.4
Illustration 8–1
Rating Plate Location . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8.4
Illustration 9–1
LOck Bar Removal (If Applicable) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9.2
Illustration 9–2
Removing the System Module Screws . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9.3
Illustration 9–3
Locating the Compression Connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9.3
Illustration 9–4
Releasing the System Module form the Front Plane . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9.4
Illustration 9–5
Identifying DIMM Sockets and DIMM Banks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9.7
Illustration 9–6
LOcating DIMMs on the System Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9.7
Illustration 9–7
Locating the Notches on a DIMM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9.8
Illustration 9–8
Removing the Compression Connector Cap . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9.8
Illustration 9–9
Supporting the System Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9.9
Illustration 9–10
Replacing the System Module in the OCTANE Workstation . . . . . . . . . . . . . . . . . . . . . . . . . . 9.10
Illustration 9–11
Locking the System Module to the Frontplane . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9.10
Illustration 9–12
Replacing the System Module Screws . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9.11
Illustration 9–13
Rating Plate Location . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9.13
Illustration 10–1
LOck Bar Removal (If Applicable) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10.2
Illustration 10–2
Removing the System Module Screws . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10.3
Illustration 10–3
Locating the Compression Connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10.3
Illustration 10–4
Placing the Compression Cap on the Compression Connector . . . . . . . . . . . . . . . . . . . . . . . . . 10.5
Illustration 10–5
Removing the Single Processor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10.6
Illustration 10–6
Viewing The Connectors on the Underside of the CPUs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10.7
xviii
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 5 DIRECTION 2181737–100
Illustration Page
Illustration 10–7
Installing the Dual Processor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10.8
Illustration 10–8
Removing the Compression Connector Cap . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10.9
Illustration 10–9
Supporting the System Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10.10
Illustration 10–10
Replacing the System Module in the OCTANE Workstation . . . . . . . . . . . . . . . . . . . . . . . . . . 10.11
Illustration 10–11
Locking the System Module to the Frontplane . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10.11
Illustration 10–12
Replacing the System Module Screws . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10.12
Illustration 10–13
Rating Plate Location . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10.14
Illustration 11–1
Locating the Lockbar on the Front of the OCTANE Workstation . . . . . . . . . . . . . . . . . . . . . . . 11.3
Illustration 11–2
Locating the Lockbar on the Octane Workstation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11.4
Illustration 11–3
Unlocking and Removing the Lockbar . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11.5
Illustration 11–4
Pressing the Bezel Release Buttons . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11.5
Illustration 11–5
Removing the Bezel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11.6
Illustration 11–6
Identifying the Option Drive Bays . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11.7
Illustration 11–7
Sliding In an Optional Drive . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11.7
Illustration 11–8
Locking the Drive . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11.8
Illustration 11–9
Opening the Door on the Bezel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11.8
Illustration 11–10
Pushing the Blank Panel Free . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11.9
Illustration 11–11
Replacing the Bezel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11.10
Illustration 11–12
Replacing The LockBar . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11.10
Illustration 12–1
Bar Code Reader Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12.2
Illustration 12–2
Bar Code Reader Instalation (Detail) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12.3
Illustration 12–3
Bar CoDE SAMPLEs TO SCAN . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12.5
Illustration 12–4
Options Window . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12.6
xix
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 5 DIRECTION 2181737–100
Illustration Page
Illustration 12–5
Sample Software Options Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12.7
Illustration 12–6
ConnectPRO Setup screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12.8
Illustration 12–7
Rating Plate Location . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12.10
xx
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100
Chapter 1
CT/i CPU Memory Option
Table Of Contents
Section 1
CT/i CPU Memory Option Installation Preparation . . . . . . . . . . . . . . . . . 1.1
1.1 Shutdown the System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.1
1.2 Remove SGI Chassis . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.1
Section 2
Install Optional Memory SIMMs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.4
2.1 Install SIMMs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.4
2.2 Reinstall SGI Chassis . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.4
Section 3
Startup System and Verify Installed Memory . . . . . . . . . . . . . . . . . . . . . . . 1.6
3.1 Start System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.6
3.2 Verify Installed Memory . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.6
Section 4
Finish Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.7
4.1 Install Rating Plate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.7
4.2 Return Installation Card . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.7
Section 5
Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.8
5.1 B7510T Additional DPI Memory Option Kit Parts List . . . . . . . . . . . . 1.8
1.i
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100
List Of Illustrations
Illustration Page
Illustration 1–1
SGI Chassis Cover Removal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.2
Illustration 1–2
Disconnect SGI Midplane Power Cables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.3
Illustration 1–3
SGI Chassis Cover Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.4
Illustration 1–4
CT/i Host Computer Memory Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.5
Illustration 1–5
Rating Plate Location . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.7
1.ii
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100
SECTION 1
CT/I CPU MEMORY OPTION INSTALLATION PREPARATION
The CT/i CPU Memory Option, B7510T, provides an additional 64 Mb of CPU memory for the
Silicon Graphics Computer. This memory allows more efficient execution of system software
by reducing the amount of disk swap activity and allowing larger amounts of data to be available
in memory at any given moment.
2. After the software shuts down, the following message will appear.
1.1
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100
Cover Removal
1.2
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100
Illustration 1–2
DISCONNECT SGI MIDPLANE POWER CABLES
ÏÏÏÏÏ Installation/Removal
ÏÏÏÏÏ
out of the way.
1.3
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100
SECTION 2
INSTALL OPTIONAL MEMORY SIMMS
(4)
(2)
1.4
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100
Illustration 1–4
CT/I HOST COMPUTER MEMORY REPLACEMENT
Use a
wrist strap at TO REMOVE SIMMS:
a grounded
work station. 1. Always use proper static precautions!
2. Locate the memory SIMM you want to remove.
WITH THE I/O CONNECTORS TOWARDS YOU, YOU WILL NEED
TO ALSO REMOVE THE SIMMS IN FRONT OF THE ONE YOU
WANT TO REMOVE.
ÏÏÏÏÏÏÏ
ÏÏÏÏÏÏÏ
TO INSTALL SIMMS:
1.5
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100
SECTION 3
STARTUP SYSTEM AND VERIFY INSTALLED MEMORY
If the reported Memory size is not 256Mbytes, open the SGI chassis and check/re–seat the
memory SIMMs.
4. Exit the shell:
>> exit
5. Select “Start System”
1.6
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100
If the reported Memory size is not 256Mbytes, open the SGI chassis and check/re–seat the
memory SIMMs.
SECTION 4
FINISH INSTALLATION
Illustration 1–5
RATING PLATE LOCATION
1.7
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100
SECTION 5
PARTS
1.8
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 8 DIRECTION 2181737–100
Chapter 2
CT/i ConnectPRO and Bar Code Reader Option
Table Of Contents
Section 1
CT/i ConnectPRO Option Installation Preparation . . . . . . . . . . . . . . . . . . 2.1
1.1 Bar Code Reader Kit Hardware Installation . . . . . . . . . . . . . . . . . . . . . . 2.1
1.2 Console Hardware Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.2
1.3 Verify Bar Code Scanner Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.4
Section 2
ConnectPRO Software Option Installation & Checkout . . . . . . . . . . . . . . 2.6
2.1 Install and Setup ConnectPRO Option from MOD . . . . . . . . . . . . . . . . 2.6
2.2 ConnectPro Feature Checkout . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.8
2.3 Save System State . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.9
Section 4
Finish Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.10
4.1 Install Rating Plate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.10
4.2 Return Installation Card . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.10
Section 4
Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.11
4.1 B7540RB Bar Code Reader Only Option Kit Parts List . . . . . . . . . . . . 2.11
4.2 B7500PM HIS/RIS Interface with Bar Code Reader
Option Kit Parts List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.11
4.3 Material List for Bar Code Reader Kit 2180690 . . . . . . . . . . . . . . . . . . 2.11
2.i
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100
List Of Illustrations
Illustration Page
Illustration 2–1
Bar Code Reader Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.2
Illustration 2–2
Bar Code Reader Installation (Detail) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.3
Illustration 2–3
Bar Code Samples to Scan . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.5
Illustration 2–4
Options Window . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.6
Illustration 2–5
Sample Software Options Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.7
Illustration 2–6
ConnectPRO Setup Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.8
Illustration 2–7
Rating Plate Location . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.10
2.ii
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100
SECTION 1
CT/i ConnectPRO OPTION INSTALLATION PREPARATION
2.1
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100
CONSOLE
Items removed for clarity
2.2
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100
Illustration 2–2
BAR CODE READER INSTALLATION (DETAIL)
J42
Existing Cable
Disconnect from J42
Y–Cable assembly
(2180690–3)
2.3
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100
2.4
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100
Illustration 2–3
BAR CODE SAMPLES TO SCAN
REQUISITION #
508
PATIENT ID #
REQUISITION #
510
PATIENT ID #
2.5
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 8 DIRECTION 2181737–100
SECTION 2
ConnectPRO SOFTWARE OPTION INSTALLATION & CHECKOUT
Note 4 Before you install the ConnectPRO option software, obtain the following information
from the hospital’s network administrator:
2.6
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 8 DIRECTION 2181737–100
6. Select Permanent.
7. Select MOD.
8. Select OK to display a software options screen similar to the one shown in Illustration 2–5.
Illustation 2–5
SAMPLE SOFTWARE OPTIONS MANUAL
9. After the ConnectPRO label appears in the “Installed Options” list, a “ConnectPRO
Setup” screen appears similar to the one in Illustration 2–6.
10. Type/enter the information you recorded on page2.6 into the corresponding data fields
on the ConnectPRO Setup screen.
a. If you need additional information, select Help to display a HELP window
explaining each data field.
b. Select Dismiss to exit the ConnectPRO Setup screen without any changes.
2.7
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 8 DIRECTION 2181737–100
Illustation 2–6
ConnectPRO SETUP SCREEN
11. Select Accept to check the valid setup entries, execute a perl script that makes changes
to the configure files for HIS/RIS, and return to the Software Options screen, shown in
Illustration 2–5, on page 2.7.
12. Refer to Illustration 2–5.
a. Select Quit.
b. Select ok.
13. Remove the MOD, and write protect the side containing the ConnectPRO option.
Note 5 The initial install requires a write enabled MOD; but you may subsequently install the
option(s) with a write protected MOD.
2.8
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 8 DIRECTION 2181737–100
7. Select Save.
Review the output for errors or missing files. You may use the scroll bar on the right
after the tool completes its tasks.
8. When Save completes, select Dismiss.
2.9
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100
SECTION 3
FINISH INSTALLATION
2.10
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100
SECTION 4
PARTS
4.1 B7540RB Bar Code Reader Only Option Kit Parts List
Item Part Number Name SPP Quantity Description (Remarks)
1 2180690 Bar Code Reader Kit N 1 Bar Code Reader Kit including Y–Cable and rubber mount. 2
4.2 B7500PM HIS/RIS Interface with Bar Code Reader Option Kit Parts List
Item Part Number Name SPP Quantity Description (Remarks)
1 2180690 Bar Code Reader Kit N 1 Bar Code Reader Kit including Y–Cable and rubber mount
3 2180690–4 Scanner Holder 2 1 Plastic holder ONLY (to store bar code scanner when not in use)
2.11
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100
NOTES
2.12
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100
Chapter 3
CT/i Additional System Magnetic Disk Option
Installation
Table Of Contents
Section 1
Prepare for CT/i Additional Magnetic Disk Option Installation . . . . . . . 3.1
1.1 Prepare for Option Disk Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.1
1.2 Save System State Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.1
1.3 Shutdown the System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.2
Section 2
Install Additional Magnetic Disk Option Hardware . . . . . . . . . . . . . . . . . 3.3
2.1 Power Off Console and Install Optional Disk . . . . . . . . . . . . . . . . . . . . 3.3
2.2 Verify Hardware Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.3
Section 3
Load CT/i Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.6
3.1 Install Option Disk Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.6
Section 4
Finish Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.6
4.1 Install Rating Plate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.6
4.2 Return Installation Card . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.6
Section 5
Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.7
5.1 B7510S CT/i Additional System Magnetic Disk Option Kit Parts List . 3.7
3.i
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100
List Of Illustrations
Illustration Page
Illustration 3–1
Option Disk Installation Drawing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.4
Illustration 3–2
User Drive Assembly Exploded View . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.5
Illustration 3–3
Rating Plate Location . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.6
3.ii
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 6 DIRECTION 2181737–100
SECTION 1
PREPARE FOR CT/I ADDITIONAL MAGNETIC DISK OPTION INSTALLATION
This section describes the correct installation and configuration for B7510S,
CT/i Additional System Magnetic Disk Option. (Pioneer is not an Option)
3.1
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 6 DIRECTION 2181737–100
This will take down both CT/i Applications and UNIX and put you at the boot prompt.
3.2
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 6 DIRECTION 2181737–100
SECTION 2
INSTALL ADDITIONAL MAGNETIC DISK OPTION HARDWARE
(5) Keeping the cables clear of the assemblies, pull the User Drive Assembly out of the
chassis and set on the floor in front of the console.
(6) Refer to Illustration 3–1. Verify the new drive jumper settings.
› Option Disk Drive SCSI ID 2.
(7) Refer to Illustration 3–1, Notes 1 thru 4. Install the Optional Disk using the kit
hardware as shown in Illustration 3–1.
(8) Reverse steps (5) through (2) above.
3.3
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 6 DIRECTION 2181737–100
Illustration 3–1
OPTION DISK INSTALLATION DRAWING
3.4
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 6 DIRECTION 2181737–100
Illustration 3–2
USER DRIVE ASSEMBLY EXPLODED VIEW
OC System Disk
OC Optional System
Disk
MaxOptics MOD
CDROM Drive
SCSI TERMINATOR
3.5
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 6 DIRECTION 2181737–100
SECTION 3
LOAD CT/I SOFTWARE
SECTION 4
FINISH INSTALLATION
3.6
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 6 DIRECTION 2181737–100
SECTION 5
PARTS
5.1 B7510S CT/i Additional System Magnetic Disk Option Kit Parts List
Item Part Number Name SPP Quantity Description (Remarks)
1 2180997 4GB DISK DRIVE 1 1 OC System First or Second Optional software disk, Compatible
with Revision 3.6 or later software.
2 46–170015P10 Screw 2 4 6–32 X 1/4” LG BIND HD ST Screw
3.7
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 6 DIRECTION 2181737–100
NOTES
3.8
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100
Chapter 4
CT/i Smart Beam Option Software Installation
Table Of Contents
Section Description Page
Section 1
CT/i Optional Smart Beam Software Installation . . . . . . . . . . . . . . . . . . . 4.1
1.1 Check Installed Collimator Type . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4.1
1.2 Add Smart Beam Software Option . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4.2
Section 2
Calibrate System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4.4
Section 3
Finish Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4.4
3.1 Install Rating Plate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4.4
3.2 Return Installation Card . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4.4
Section 4
Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4.5
4.1 B7530SB CT/i Smart Beam Software (w/o Collimator) Kit Parts List 4.5
4.2 B7530SC CT/i Smart Beam Collimator Option Kit Parts List . . . . . . . 4.5
4.3 B7530SS CT/i Smart Beam Software (w/Collimator) Kit Parts List . . . 4.5
4.i
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100
List Of Illustrations
Illustration Page
Illustration 4–1
Software Options Installation Window . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4.2
Illustration 4–2
Software Options Available / Installed Screen Example . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4.3
Illustration 4–3
Rating Plate Location . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4.4
4.ii
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100
SECTION 1
CT/I OPTIONAL SMART BEAM SOFTWARE INSTALLATION
Notice This Software Option Installation assumes that the system that this will be installed on
already has the correct collimator type required for this option to operate properly.
The correct collimator type is: 46–296300G4 or 46–296300G5. If this system does not
have either of these collimators installed, do not install this option software.
See the Material Lists in Section 4 of this chapter for Catalog Numbers for ordering the
Smart Beam collimator.
CAUTION Image Artifacts may result if the wrong collimator type is used with the Smart
Beam Option Software. Ensure that the correct collimator type is installed
before installing the Smart Beam Option Software.
4.1
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 8 DIRECTION 2181737–100
4.2
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 8 DIRECTION 2181737–100
6. Select Permanent.
7. Select MOD.
8. Select OK to display a software options screen similar to the one shown in Illustration 4–2.
Illustation 4–2
SOFTWARE OPTIONS AVAILABLE/INSTALLED EXAMPLE SCREEN
9. After all the options are installed, select Quit, then OK.
10. Remove the MOD & write protect the side with options.
Note 5 The initial install requires the MOD to not be write protected; subsequent installs can be done
with the MOD write protected
11. If you have another option MOD to load, go back to step 1.
4.3
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100
SECTION 2
CALIBRATE SYSTEM
Because the Smart Beam Option makes use of different X–Ray Beam Shaping
filters to improve dose utilization, the system must have new calibration
vectors created for scanning.
(1) From the ExamRx Desktop, perform a complete set of Air and Phantom Calibrations
after the option software has been installed.
Note 6 Smart Beam Calibrations will take additional scans and perform additional processing in
order to accommodate the second X–Ray Beam profile.
SECTION 3
FINISH INSTALLATION
4.4
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100
SECTION 4
PARTS
4.1 B7530SB CT/i Smart Beam Software (w/o Collimator) Option Kit Parts List
4.2 B7530SC CT/i Smart Beam Collimator Option Kit Parts List
Item Part Number Name SPP Quantity Description (Remarks)
1 46–296300G4 Smart Beam Collimator N 1 Smart Beam ”G4” Collimator
4.3 B7530SS CT/i Smart Beam Software (with Collimator) Option Kit Parts List
Item Part Number Name SPP Quantity Description (Remarks)
1 B7530SB Smart Beam Software N 1 Smart Beam Software Option Kit
4.5
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100
NOTES
4.6
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100
Chapter 5
CT/i Language Options Installation
Table Of Contents
Section 1
Language Option Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.1
1.1 Shutdown the System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.1
1.2 Boot System and Stop For Maintenance . . . . . . . . . . . . . . . . . . . . . . . . 5.1
1.3 If you have a keyboard other than a US–English Keyboard. . . . . . . . . . 5.1
Section 2
Procedures for installing add–on packages . . . . . . . . . . . . . . . . . . . . . . . . . 5.2
2.1 Optional Package Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.2
2.2 Reconfig to Select Language . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.3
2.3 Reconfig SBC to Enable Language . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.3
Section 3
Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.5
3.1 Material Lists for International Language Kits . . . . . . . . . . . . . . . . . . . 5.5
5.i
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100
NOTES
5.ii
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100
SECTION 1
LANGUAGE OPTION INSTALLATION
› Run Diagnostics
› Recover System
(5) As the system comes up, a box will appear indicating the operator has 5 seconds to stop
application bring up. STOP applications start up when this box appears.
5.1
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100
SECTION 2
PROCEDURES FOR INSTALLING ADD–ON PACKAGES
(3) Insert the Software Packages Media into the CD–ROM drive.
(4) Click on the CD–ROM media selection.
5.2
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100
(5) Select the packages to be installed (e.g.: click & drag down to highlight all the available
options, then release).
(6) Click on the Install button.
A box appears while the software packages are loading.
(7) After all the packages are installed, select Quit, then OK.
(8) Leave the shell window open.
5.3
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100
SECTION 3
PARTS
5.4
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100
Chapter 6
CT/i Dentascan Option Installation
Table Of Contents
Section Description Page
Section 1
Install Dentascan Option Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.1
1.1 Add Software Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.1
Section 2
Procedures for installing Dentascan filming option . . . . . . . . . . . . . . . . . . 6.3
Section 3
Finish Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.3
3.1 Install Rating Plate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.3
3.2 Return Installation Card . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.4
Section 4
Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.4
4.1 B7540N CT/i Dentascan Option Kit Parts List . . . . . . . . . . . . . . . . . . . 6.4
6.i
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100
NOTES
List Of Illustrations
Illustration Page
Illustration 6–1
Software Options Installation Window . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.1
Illustration 6–2
Software Options Available / Installed Screen Example . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.2
Illustration 6–3
Rating Plate Location . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.3
6.ii
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 8 DIRECTION 2181737–100
SECTION 1
INSTALL DENTASCAN OPTION SOFTWARE
6.1
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 8 DIRECTION 2181737–100
6. Select Permanent.
7. Select MOD.
8. Select OK to display a software options screen similar to the one shown in Illustration 6–2.
Illustation 6–2
SOFTWARE OPTIONS AVAILABLE/INSTALLED SAMPLE SCREEN
9. After all the options are installed, select Quit, then OK.
10. Remove the MOD & write protect the side with options. (Note: The initial install
requires the MOD to not be write protected; subsequent installs can be done with the
MOD write protected)
11. If you have another option MOD to load, go back to step 1.
Notice Once you have installed Dentascan, you need to follow the procedures in, Section 2, for
installing Dentascan filming option, before proceeeding to step 12..
6.2
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100
SECTION 2
PROCEDURES FOR INSTALLING DENTASCAN FILMING OPTION
Notice After the film is developed, use a metric scale to measure the distance between the vertical
lines for “Dist 1” and “Dist 2”. Enter the value(s) in MILLIMETERS (mm).
SECTION 3
FINISH INSTALLATION
6.3
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100
SECTION 4
PARTS
6.4
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100
Chapter 7
CT/i Smart Helical Option Installation
Table Of Contents
Section Description Page
Section 1
Install CT/i 4.X System Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7.1
1.1 Install 4.X Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7.1
Section 2
PROCEDURES FOR INSTALLING SMART HELICAL SOFTWARE . 7.2
Section 3
Finish Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7.4
3.1 Install Rating Plate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7.4
3.2 Return Installation Card . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7.4
Section 4
Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7.5
4.1 B7530TL Smart Helical Option Kit Parts List (systems not at 4.X) . . . 7.5
4.2 B7530TK Smart Helical Option Kit Parts List (systems at 4.X) . . . . . . 7.5
7.i
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100
List Of Illustrations
Illustration Page
Illustration 7–1
Software Options Installation Window . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7.2
Illustration 7–2
Software Options Available / Installed Screen Example . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7.3
Illustration 7–3
Rating Plate Location . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7.4
NOTES
7.ii
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100
SECTION 1
INSTALL CT/I 4.X SYSTEM SOFTWARE
Notice The prerequisite to loading this Smart Helical Option software is that
you must first have Version 4.0 or higher System Software loaded on
your scanner. Do not attempt to load/enable this option without first
loading the latest version of 4.X Operating Software (included in this
Option Kit).
1. Refer to the Load From Cold Manual, 2168070–100, included with this Option Kit, for
instructions for loading the System Software included with this kit.
2. If 4.X software is already loaded on your system, go to Section 2 to install the Option
Software.
7.1
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 8 DIRECTION 2181737–100
SECTION 2
PROCEDURES FOR INSTALLING SMART HELICAL OPTION SOFTWARE
6. Select Permanent.
7. Select MOD.
8. Select OK to display a software options screen similar to the one shown in Illustration 7–2.
7.2
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 8 DIRECTION 2181737–100
Illustation 7–2
SOFTWARE OPTIONS AVAILABLE/INSTALLED SAMPLE SCREEN
9. After all the options are installed, select Quit, then OK.
10. Remove the MOD & write protect the side with options.
Note 9 The initial install requires the MOD to not be write protected; subsequent installs can be done
with the MOD write protected
11. If you have another option MOD to load, go back to step 1.
12. Go to the Service Desktop –>Utilities –>Applications Shutdown –>Start
13. After Applications have shut down, type st <enter> to start up applications.
14. When the system applications return, the system is ready to use.
7.3
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100
SECTION 3
FINISH INSTALLATION
7.4
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100
SECTION 4
PARTS
4.1 B7530TL Smart Helical Option Kit Parts List (systems not at 4.X)
Item Part Number Name SPP Quantity Description (Remarks)
1 2180724 Options MOD N 1 Smart Helical Option
2 2181969 HiSpeed CT/I 1 1 IRIX Operating System Software
IRIX 6.2 CDROM
3 2200557 HiSpeed CT/I 4.0 1 1 4.0 Applications Software
Applications CDROM
4 2154721 Class A Service/ N 1 Operators & Service Manual Collector
Operators Manuals
4.2 B7530TK Smart Helical Option Kit Parts List (systems at 4.X)
Item Part Number Name SPP Quantity Description (Remarks)
1 2180724 Options MOD N 1 Smart Helical Option
7.5
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100
NOTES
7.6
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 5 DIRECTION 2181737–100
Chapter 8
CT/i BMD Option Installation
Table Of Contents
Section Description Page
Section 1
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8.1
1.1 " ! !# . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8.1
Section 2
PROCEDURES FOR INSTALLING BMD OPTION SOFTWARE . . . . 8.1
2.1 Install BMD Option Key and Load BMD Option Software . . . . . . . . . . 8.1
Section 3
Special Instructions for Transferring the BMD Database from an Advantage
Windows Workstation to the CT/i . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8.2
3.1 Transfer the BMD Database from AWW to Ct/i . . . . . . . . . . . . . . . . . . 8.2
3.2 Verify that the data has been successfully transferred . . . . . . . . . . . . . . 8.3
Section 4
Backup Database and Deliver Books & Phantoms to Customer . . . . . . . . 8.3
4.1 Backup BMD Database (Save To Mod) . . . . . . . . . . . . . . . . . . . . . . . . . 8.3
4.2 Deliver Books & BMD Phantom to Customer . . . . . . . . . . . . . . . . . . . . 8.3
Section 5
Finish Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8.4
5.1 Install Rating Plate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8.4
5.2 Return Installation Card . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8.4
Section 6
Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8.5
6.1
!
! ! ! # ! . 8.5
6.2
! !
! ! ! #! ! 8.5
8.i
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 5 DIRECTION 2181737–100
List Of Illustrations
Illustration Page
Illustration 8–1
Rating Plate Location . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8.4
NOTES
8.ii
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 5 DIRECTION 2181737–100
SECTION 1
SYSTEM SOFTWARE PREREQUISITE
Notice The prerequisite to loading this BMD Option software is that you must
first have Version 4.1, or Version 5.3 (or higher) System Software loaded
on your scanner. Do not attempt to load/enable this option without first
loading the latest version of 4.X or 5.X Operating Software.
1. If 4.1 or 5.3 (or higher) software is already loaded on your system, go to Section 2 to
install the Option Software.
2. If 4.1 or 5.3 (or higher) software is NOT loaded on your system, you must first perform
FMI 25245 (4.1 Software) or FMI 25249 (5.3 Software) before attempting to load this
option software
SECTION 2
PROCEDURES FOR INSTALLING BMD OPTION SOFTWARE
2.1 Install BMD Option Key and Load BMD Option Software
1. Please refer to Chapter 3 of the QCT Bone Densitometry User’s Guide for
instructions.
2. You will first load the BMD Option Key MOD, to enable loading of the BMD
software, and then you will load the BMD software from the QCT–5000 Bone
Densitometry CDROM.
3. Be sure to follow all of the instructions carefully, to ensure that the Option Software is
properly loaded and set up.
Notice The ”eject” softkey that is displayed on the screen may not work to eject the
CDROM when you have completed loading the BMD Option Software. If
you find this to be the case, perform the following steps to eject the CDROM:
8.1
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 5 DIRECTION 2181737–100
SECTION 3
SPECIAL INSTRUCTIONS FOR TRANSFERRING THE BMD DATABASE FROM AN
ADVANTAGE WINDOWS WORKSTATION TO THE CT/i
Notice Use this procedure only if the customer has been using an Advantage Windows Workstation
to perform BMD Calculations on CT Images, and wants to transfer the BMD Database to
the CT/i System for all future BMD Calculations.
If the BMD Database is not transferred to the CT/i prior to performing BMD Calculations, it
will be impossible later to merge the AWW BMD Database with the CT/i BMD Database.
8.2
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 5 DIRECTION 2181737–100
Ask the Operators to review the displayed BMD Database to ensure that all of the BMD
Studies appear. Or else, go back to the AWW, and from the Browser, perform these
same steps, and compare the two BMD Database displays to ensure that all of the BMD
Studies were transferred.
SECTION 4
BACKUP DATABASE AND DELIVER BOOKS & PHANTOMS TO CUSTOMER
8.3
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 5 DIRECTION 2181737–100
SECTION 5
FINISH INSTALLATION
8.4
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 5 DIRECTION 2181737–100
SECTION 6
PARTS
6.1 B7710MB CT/i BMD Basic Option Kit Parts List (w/o Phantom)
Item Part Number Name SPP Quantity Description (Remarks)
1 2206382 Options MOD N 1 BMD Option Key MOD
2 2222885 BMD Software Package 2 1 BMD Software Package (w/o Phantoms)
6.2 B7710MC CT/i BMD + Phantom Option Kit Parts List (with Phantom)
Item Part Number Name SPP Quantity Description (Remarks)
1 2206382 Options MOD N 1 BMD Option Key MOD
2 2222885–2 BMD Option Package N 1 BMD Software Package (with Phantom)
8.5
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 5 DIRECTION 2181737–100
NOTES
8.6
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100
Chapter 9
CT/i (OCTANE) Memory Option
Table Of Contents
Section Description Page
Section 1
CT/i (OCTANE) CPU Memory Option Installation Preparation . . . . . . . 9.1
1.1 Shutdown the System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9.1
1.2 Preparing the Workstation to Install the optional memory . . . . . . . . . . . 9.1
1.3 Removing the System Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9.3
Section 2
Installing and Removing Memory . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9.6
2.1 About Memory . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9.6
2.2 Installing Memory . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9.8
2.3 Replacing the System Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9.8
2.4 Powering On the OCTANE Workstation . . . . . . . . . . . . . . . . . . . . . . . . 9.11
Section 3
Startup System and Verify Installed Memory . . . . . . . . . . . . . . . . . . . . . . . 9.12
3.1 Start System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9.12
3.2 Verify Installed Memory . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9.12
Section 4
Finish Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9.13
4.1 Install Rating Plate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9.13
4.2 Return Installation Card . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9.14
Section 5
Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9.15
5.1 Field Replaceable Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9.15
5.2 B7710M Memory Option Kit – Parts List . . . . . . . . . . . . . . . . . . . . . . . 9.15
9.i
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100
List Of Illustrations
Illustration Page
Illustration 9–1
LOck Bar Removal (If Applicable) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9.2
Illustration 9–2
Removing the System Module Screws . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9.3
Illustration 9–3
Locating the Compression Connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9.3
Illustration 9–4
Releasing the System Module form the Front Plane . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9.4
Illustration 9–5
Identifying DIMM Sockets and DIMM Banks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9.7
Illustration 9–6
LOcating DIMMs on the System Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9.7
Illustration 9–7
Locating the Notches on a DIMM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9.8
Illustration 9–8
Removing the Compression Connector Cap . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9.8
Illustration 9–9
Supporting the System Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9.9
Illustration 9–10
Replacing the System Module in the OCTANE Workstation . . . . . . . . . . . . . . . . . . . . . . . . . . 9.10
Illustration 9–11
Locking the System Module to the Frontplane . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9.10
Illustration 9–12
Replacing the System Module Screws . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9.11
Illustration 9–13
Rating Plate Location . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9.13
9.ii
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100
SECTION 1
CT/I (OCTANE) CPU MEMORY OPTION INSTALLATION PREPARATION
The CT/i CPU Memory Option, B7510T, provides an additional 64 Mb of CPU memory for the
Silicon Graphics Computer. This memory allows more efficient execution of system software
by reducing the amount of disk swap activity and allowing larger amounts of data to be available
in memory at any given moment.
2. After the software shuts down, the following message will appear.
T prepare the workstation for installation of optional memory, follow these steps:
1. Open the cover and push the power button to power off the OCTANE workstation.
2. Unplug the power cord from the electrical outlet and from the OCTANE workstation.
3. Press the power switch on the monitor.
4. Wait 5 minutes before removing the system module.
9.1
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100
CAUTION The heat sinks on the CPU and system module become very hot. Wait
5minutes after powering off the OCTANE workstation before you remove the
system module. Test before touching the CPU or heat sinks.
Lock
Lock Bar
Illustration 9–1
LOCK BAR REMOVAL (IF APPLICABLE)
5. If applicable, remove the lockbar which locks the system module and bezel to the
chassis.
› Unlock the lock at the back of the OCTANE workstation.
› Pull the lockbar out of the OCTANE workstation.
6. Face the rear of the workstation.
7. Remove the cables attached to the system module.
9.2
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100
Illustration 9–2
REMOVING THE SYSTEM MODULE SCREWS
CAUTION The heat sinks on the CPU and system module become very hot. Wait
5minutes after powering off the OCTANE workstation before you remove the
system module. Test before touching the CPU or heat sinks.
1. To remove the system module, you need to power off the system, wait 5 minutes after
powering off the workstation to allow the heat sinks to cool, and attach the wrist strap.
2. Loosen the captive screws that hold the sliding handles to the OCTANE workstation.
Compression Connectors
Bristled pad
Illustration 9–3
LOCATING THE COMPRESSION CONNECTORS
9.3
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100
Look for the two compression connectors when you take the system module out of the
chassis, and do not touch them. They are on the back of the system module, and
connect to the chassis when the system module is installed in the workstation.
Notice The gold bristled pad on the compression connectors on the back of the system module is
very delicate and easily damaged. Bristled pad
3. Pull both sliding handles simultaneously until they are completely extended. This action
releases the system module from the workstation.
Illustration 9–4
RELEASING THE SYSTEM MODULE FORM THE FRONT PLANE
4. Pull the system module from the chassis by grasping the module handle with your left
hand and bracing your right hand against the top of the workstation.
Support the bottom of the module as you pull it out.
5. Place it on a dry, antistatic surface with the CPU and DIMMs facing up. Your desktop
works well.
Notice Do not touch the compression connectors as you remove the system module. The side of the
compression connector that connects to the workstation is extremely delicate.
9.4
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100
6. Place a cap on each compression connector on the back of the system module.
(Compression connector caps come with the workstation.)
CAUTION Warning: The heat sinks on the CPU and system module become very hot.
Wait 5 minutes after powering off the OCTANE workstation before you remove
the system module. Test before touching the CPU or heat sinks.
9.5
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100
SECTION 2
INSTALLING AND REMOVING MEMORY
To install or remove memory, you must power off the workstation, wait 5 minutes after powering off the workstation to allow the heat
sinks to cool, attach the wrist strap, and remove the system module.
9.6
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100
Processor
Banks
S1 1
System module S2
S3 2
DIMM S4
sockets
S5 3
S6
S7 4
S8
Illustration 9–5
IDENTIFYING DIMM SOCKETS AND DIMM BANKS
DIMMs
Bank 1
Bank 2
Bank 3
Bank 4
Illustration 9–6
LOCATING DIMMS ON THE SYSTEM MODULE
9.7
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100
Notches
Illustration 9–7
LOCATING THE NOTCHES ON A DIMM
DIMMs are notched on the bottom so that they cannot be inserted incorrectly.
1. Insert the DIMM into the socket, gently but firmly. You hear a click as it is seated, and
the latch on the end of the socket moves up.
2. Check to be sure both sockets in the bank are full. DIMMs must be installed in pairs.
You have finished installing memory and are ready to replace the system module.
Illustration 9–8
REMOVING THE COMPRESSION CONNECTOR CAP
1. Remove the caps on the compression connectors on the back of the system module. Do
not touch the system module compression connectors as you lift the module into the
workstation.
9.8
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100
Notice Do not insert the system module into the workstation unless the power cord is unplugged
from the electrical socket. Inserting the system module into an OCTANE workstation with
the power cord connected to an electrical socket causes damage to the OCTANE
workstation.
Illustration 9–9
SUPPORTING THE SYSTEM MODULE
2. Grasp the system module by its immovable handle and support it with one hand as you
slide it into the chassis.
3. Push the module completely into the chassis using the immovable handle. (Both sliding
handles protrude.) The system module stops when it is slightly out of the chassis.
9.9
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100
Illustration 9–10
REPLACING THE SYSTEM MODULE IN THE OCTANE WORKSTATION
4. Push the sliding handles simultaneously to connect the system module to the
workstation.
Illustration 9–11
LOCKING THE SYSTEM MODULE TO THE FRONTPLANE
9.10
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100
5. Push firmly on the handles to completely lock the system module to the workstation.
The handles are completely recessed and the system module is flush with the chassis
when it is fully seated and locked to the workstation.
Illustration 9–12
REPLACING THE SYSTEM MODULE SCREWS
6. Tighten the captive screws in the sliding handles until the system module is attached to
the chassis.
7. Remove the wrist strap.
9.11
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100
SECTION 3
STARTUP SYSTEM AND VERIFY INSTALLED MEMORY
9.12
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100
SECTION 4
FINISH INSTALLATION
Illustration 9–13
RATING PLATE LOCATION
9.13
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100
9.14
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100
SECTION 5
PARTS
9.15
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100
9.16
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100
Chapter 10
CT/i (OCTANE) Dual Processor Option
Table Of Contents
Section Description Page
Section 1
CT/i OCTANE Dual Processor Option Installation Preparation . . . . . . . 10.1
1.1 Shutdown the System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10.1
1.2 Removing the System Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10.2
Section 2
Installing The Optional Dual Processor . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10.6
2.1 Removing The Single Processor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10.6
2.2 Installing The Optional Dual Processor . . . . . . . . . . . . . . . . . . . . . . . . . 10.7
2.3 Replacing the System Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10.9
2.4 Powering On the OCTANE Workstation . . . . . . . . . . . . . . . . . . . . . . . . 10.12
Section 3
Startup System and Verify Installed Dual Processor . . . . . . . . . . . . . . . . . 10.13
3.1 Start System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10.13
3.2 Verify Installed Dual Processor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10.13
Section 4
Finish Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10.14
4.1 Install Rating Plate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10.14
4.2 Return Installation Card . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10.15
Section 5
Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10.16
5.1 Field Replaceable Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10.16
5.2 B7710JP Dual Processor Option Kit – Parts List . . . . . . . . . . . . . . . . . . 10.16
10.i
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100
List Of Illustrations
Illustration Page
Illustration 10–1
LOck Bar Removal (If Applicable) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10.2
Illustration 10–2
Removing the System Module Screws . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10.3
Illustration 10–3
Locating the Compression Connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10.3
Illustration 10–4
Placing the Compression Cap on the Compression Connector . . . . . . . . . . . . . . . . . . . . . . . . . 10.5
Illustration 10–5
Removing the Single Processor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10.6
Illustration 10–6
Viewing The Connectors on the Underside of the CPUs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10.7
Illustration 10–7
Installing the Dual Processor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10.8
Illustration 10–8
Removing the Compression Connector Cap . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10.9
Illustration 10–9
Supporting the System Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10.10
Illustration 10–10
Replacing the System Module in the OCTANE Workstation . . . . . . . . . . . . . . . . . . . . . . . . . . 10.11
Illustration 10–11
Locking the System Module to the Frontplane . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10.11
Illustration 10–12
Replacing the System Module Screws . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10.12
Illustration 10–13
Rating Plate Location . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10.14
10.ii
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100
SECTION 1
CT/I OCTANE DUAL PROCESSOR OPTION INSTALLATION PREPARATION
The CT/i Octane Dual Processor Option, B7710JP, replaces the single 195 Mhz CPU with a dual
195 Mhz CPU or the Octane Silicon Graphics Computer.
2. After the software shuts down, the following message will appear.
CAUTION The heat sinks on the CPU and system module become very hot. Wait
5minutes after powering off the OCTANE workstation before you remove the
system module. Test before touching the CPU or heat sinks.
10.1
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100
Lock
Lock Bar
Illustration 10–1
LOCK BAR REMOVAL (IF APPLICABLE)
8. If applicable, remove the lockbar which locks the system module and bezel to the
chassis.
› Unlock the lock at the back of the OCTANE workstation.
› Pull the lockbar out of the OCTANE workstation.
9. Face the rear of the workstation.
10. Remove the cables attached to the system module.
10.2
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100
Illustration 10–2
REMOVING THE SYSTEM MODULE SCREWS
CAUTION The heat sinks on the CPU and system module become very hot. Wait
5minutes after powering off the OCTANE workstation before you remove the
system module. Test before touching the CPU or heat sinks.
1. To remove the system module, you need to power off the system, wait 5 minutes after
powering off the workstation to allow the heat sinks to cool, and attach the wrist strap.
2. Loosen the captive screws that hold the sliding handles to the OCTANE workstation.
Compression Connectors
Bristled pad
Illustration 10–3
LOCATING THE COMPRESSION CONNECTORS
Look for the two compression connectors when you take the system module out of the
chassis, and do not touch them. They are on the back of the system module, and
connect to the chassis when the system module is installed in the workstation.
Notice The gold bristled pad on the compression connectors on the back of the system module is
very delicate and easily damaged. Bristled pad
3. Pull both sliding handles simultaneously until they are completely extended. This action
releases the system module from the workstation.
10.3
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100
4. Pull the system module from the chassis by grasping the module handle with your left
hand and bracing your right hand against the top of the workstation.
Support the bottom of the module as you pull it out.
5. Place it on a dry, antistatic surface with the CPU and DIMMs facing up. Your desktop
works well.
Notice Do not touch the compression connectors as you remove the system module. The side of the
compression connector that connects to the workstation is extremely delicate.
10.4
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100
Illustration 10–4
PLACING THE COMPRESSION CAP ON THE COMPRESSION CONNECTOR
6. Place a cap on each compression connector on the back of the system module.
(Compression connector caps come with the workstation.)
CAUTION Warning: The heat sinks on the CPU and system module become very hot.
Wait 5 minutes after powering off the OCTANE workstation before you remove
the system module. Test before touching the CPU or heat sinks.
10.5
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100
SECTION 2
INSTALLING THE OPTIONAL DUAL PROCESSOR
Single processor
Illustration 10–5
REMOVING THE SINGLE PROCESSOR
1. Loosen the four captive Phillips screws holding the CPU in place. The dual processor
has six screws.
Notice Do not remove the four inner screws (not Phillips) holding the heat sink to the CPU.
2. Slide your fingers under the edge of the single processor closest to the back of the
module, and push up to release it. You may need to use two hands to lift it out.
10.6
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100
Single Processor
Dual processor
Illustration 10–6
VIEWING THE CONNECTORS ON THE UNDERSIDE OF THE CPUS
1. Turn over the CPU to determine where the connectors are located. Align the connectors
on the base of the CPU with the connectors on the system board.
2. Place the dual processor on the side of the system module closest to the panel of
connectors.
10.7
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100
Dual Processor
Illustration 10–7
INSTALLING THE DUAL PROCESSOR
10.8
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100
Illustration 10–8
REMOVING THE COMPRESSION CONNECTOR CAP
1. Remove the caps on the compression connectors on the back of the system module. Do
not touch the system module compression connectors as you lift the module into the
workstation.
Notice Do not insert the system module into the workstation unless the power cord is unplugged
from the electrical socket. Inserting the system module into an OCTANE workstation with
the power cord connected to an electrical socket causes damage to the OCTANE
workstation.
10.9
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100
Illustration 10–9
SUPPORTING THE SYSTEM MODULE
2. Grasp the system module by its immovable handle and support it with one hand as you
slide it into the chassis.
3. Push the module completely into the chassis using the immovable handle. (Both sliding
handles protrude.) The system module stops when it is slightly out of the chassis.
10.10
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100
Illustration 10–10
REPLACING THE SYSTEM MODULE IN THE OCTANE WORKSTATION
4. Push the sliding handles simultaneously to connect the system module to the
workstation.
Illustration 10–11
LOCKING THE SYSTEM MODULE TO THE FRONTPLANE
10.11
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100
5. Push firmly on the handles to completely lock the system module to the workstation.
The handles are completely recessed and the system module is flush with the chassis
when it is fully seated and locked to the workstation.
Illustration 10–12
REPLACING THE SYSTEM MODULE SCREWS
6. Tighten the captive screws in the sliding handles until the system module is attached to
the chassis.
7. Remove the wrist strap.
10.12
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100
SECTION 3
STARTUP SYSTEM AND VERIFY INSTALLED DUAL PROCESSOR
10.13
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100
SECTION 4
FINISH INSTALLATION
Illustration 10–13
RATING PLATE LOCATION
10.14
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100
10.15
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100
SECTION 5
PARTS
2 2169940–20 Dual 195 Mhz Processor 1 1 Option Dual 195 Mhz Processor Module
10.16
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100
Chapter 11
CT/i (OCTANE) Additional Magnetic Disk Option
Installation
Table Of Contents
Section Description Page
Section 1
Prepare for CT/i (Octane) Additional Magnetic Disk Option Installation 11.1
1.1 Prepare for Option Disk Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . 11.1
1.2 Save System State Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11.1
1.3 Shutdown the System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11.2
Section 2
Install Additional Magnetic Disk Option Hardware . . . . . . . . . . . . . . . . . 11.3
2.1 About Internal Drives . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11.3
2.2 Power Off Console and Install Optional Disk . . . . . . . . . . . . . . . . . . . . 11.3
2.3 Removing the Bezel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11.4
2.4 Installing an Internal Option Drive . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11.7
2.5 Removing the Blank Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11.8
2.6 Replacing the Bezel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11.10
2.7 Verify Hardware Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11.11
Section 3
Load CT/i Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11.11
3.1 Install Option Disk Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11.11
Section 4
Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11.11
4.1 B7710JD CT/i (Octane) Additional Sys Mag Disk Option Kit . . . . . . . 11.11
11.i
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100
List Of Illustrations
Illustration Page
Illustration 11–1
Locating the Lockbar on the Front of the OCTANE Workstation . . . . . . . . . . . . . . . . . . . . . . . 11.3
Illustration 11–2
Locating the Lockbar on the Octane Workstation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11.4
Illustration 11–3
Unlocking and Removing the Lockbar . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11.5
Illustration 11–4
Pressing the Bezel Release Buttons . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11.5
Illustration 11–5
Removing the Bezel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11.6
Illustration 11–6
Identifying the Option Drive Bays . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11.7
Illustration 11–7
Sliding In an Optional Drive . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11.7
Illustration 11–8
Locking the Drive . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11.8
Illustration 11–9
Opening the Door on the Bezel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11.8
Illustration 11–10
Pushing the Blank Panel Free . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11.9
Illustration 11–11
Replacing the Bezel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11.10
Illustration 11–12
Replacing The LockBar . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11.10
11.ii
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 7 DIRECTION 2181737–100
SECTION 1
PREPARE FOR CT/I (OCTANE) ADDITIONAL MAGNETIC DISK OPTION
INSTALLATION
This section describes the correct installation and configuration for B7710JD,
CT/i (Octane)Additional System Magnetic Disk Option.
11.1
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 7 DIRECTION 2181737–100
› FAILED
A status of “does not exist” or “FAILED”, indicates potential problems. Please review the
status for each file to see if any errors are reported.
This will take down both CT/i Applications and UNIX and put you at the boot prompt.
11.2
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 7 DIRECTION 2181737–100
SECTION 2
INSTALL ADDITIONAL MAGNETIC DISK OPTION HARDWARE
Notice Drives can be easily damaged. Handle drive carefully; do not drop it or handle it roughly.
B A Power button
Power cord
C Power switch
11.3
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 7 DIRECTION 2181737–100
(4) Press and release the power button to shut down the software and power off the system.
The system powers off automatically. See illustration 11–1 (A).
(5) Press the monitor power switch to turn off your monitor. (C)
(6) Disconnect the power cable from the back of the workstation. (B)
Lockbar slot
11.4
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 7 DIRECTION 2181737–100
Illustration 11–3
UNLOCKING AND REMOVING THE LOCKBAR
(4) Press both bezel release buttons. Removing the Bezel as shown above.
11.5
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 7 DIRECTION 2181737–100
Illustration 11–5
REMOVING THE BEZEL
(5) Remove the bezel by tilting it forward, then lifting it from its release hooks at the
bottom.
11.6
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 7 DIRECTION 2181737–100
System drive
Light module
The option drive bays are the two top drive bays. Install the option drive only in the middle
bay. The OCTANE workstation automatically assigns an address to the internal drives. The
system drive is SCSI ID 0–1, the drive directly above it is SCSI ID 0–2.
Illustration 11–7
SLIDING IN AN OPTIONAL DRIVE
Handle
11.7
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 7 DIRECTION 2181737-100
(1) Lift the handle on the drive to a horizontal position, and slide the drive into the middle drive
bay. The drive handle is the same color as the drive.
(2) Slide the drive in gently until you feel solid resistance. The drive is connected to the
frontplane when the front of the drive is flush with the chassis.
Notice Pushing the drive into the frontplane with force can damage the drive.
Notice The option drive MUST be installed in the middle drive bay, or the application software
will not recognize it. The following illustration shows the drive in the upper slot to
clarify drive handle locking.
Illustration 11-8
LOCKING THE DRIVE
Drive handle
(3) With the drive completely inserted, rotate the handle downward to lock the drive in the
workstation.
You have completed installing an option drive and are ready to remove the blank panel on the
bezel.
Vent
Blanking panels
After installing a new option drive, you remove a blank panel from the bezel before replacing
the bezel.
11.8
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 7 DIRECTION 2181737–100
(1) Place the bezel on a flat surface so that the door is facing up.
(2) Open the door.
(3) Locate the blank panel for the drive bay in which you installed the option drive.
Illustration 11–10
PUSHING THE BLANK PANEL FREE
Vent
Blanking panels
(4) Press down on the blank panel. It pops out the back of the bezel.
(5) Keep the blank panel. You must replace it if you ever remove the drive.
Notice To prevent damage to your OCTANE workstation, you must place a blank panel in the bezel
if you have removed a drive. The blank panel ensures proper airflow through the
workstation.
11.9
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 7 DIRECTION 2181737–100
A Push in
11.10
GE MEDICAL SYSTEMS CT/I OPTIONS INST ALLATION
REV 7 DIRECTION 2181737-100
(3) Insert the lockbar into the front of the workstation with the angled end up.(A)
(4) Insert and close the lock at the back of the Octane workstation.(B)
You are now ready to power on the Octane workstation.
SECTION 3
LOAD CT/I SOFTWARE
Notice Only by performing a complete LFC can you ensure that the new drive is properly
set-up and that the Image Database integrity is assured.
SECTION 4
PARTS
4.1 B7710JD CT/i (Octane) Additional Sys. Mag Disk Option Kit
Item Part Number Name SPP Quantity Description (Remarks)
1 2169940-10 4GB DISK DRIVE 1 1 Second Optional 4GB Disk Drive
11.11
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 7 DIRECTION 2181737–100
NOTES
11.12
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100
Chapter 12
CT/i (Octane) ConnectPRO and Bar Code Reader
Option
Table Of Contents
Section 1
CT/i (Octane) ConnectPRO Option Installation Preparation . . . . . . . . . . 12.1
1.1 Bar Code Reader Kit Hardware Installation . . . . . . . . . . . . . . . . . . . . . . 12.1
1.2 Console Hardware Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12.1
1.3 Verify Bar Code Scanner Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12.4
Section 2
ConnectPRO Software Option Installation & Checkout . . . . . . . . . . . . . . 12.6
2.1 Install and Setup ConnectPRO Option from MOD . . . . . . . . . . . . . . . . 12.6
2.2 ConnectPRO Feature Checkout . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12.9
2.3 Save System State . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12.9
Section 3
Finish Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12.10
3.1 Install Rating Plate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12.10
3.2 Return Installation Card . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12.10
Section 4
Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12.11
4.1 B7540RB Bar Code Reader Only Option Kit Parts List . . . . . . . . . . . . 12.11
4.2 B7500PM HIS/RIS Interface w/Bar Code Reader Option Kit Parts List 12.11
4.3 Material List for Bar Code Reader Kit 2180690 . . . . . . . . . . . . . . . . . . 12.11
12.i
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100
List Of Illustrations
Illustration Page
Illustration 12–1
Bar Code Reader Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12.2
Illustration 12–2
Bar Code Reader Instalation (Detail) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12.3
Illustration 12–3
Bar CoDE SAMPLEs TO SCAN . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12.5
Illustration 12–4
Options Window . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12.6
Illustration 12–5
Sample Software Options Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12.7
Illustration 12–6
ConnectPRO Setup screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12.8
Illustration 12–7
Rating Plate Location . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12.10
12.ii
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100
SECTION 1
CT/i (Octane) ConnectPRO OPTION INSTALLATION PREPARATION
12.1
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100
Monitor Monitor
Illustration 12–1
BAR CODE READER INSTALLATION
12.2
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100
J21
3rd from top
(in front of console)
2 1
Keyboard Connection
in back of Octane (Console)
Existing Cable
Disconnected
From J21
Y–Cable assembly
(2180690–3)
12.3
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100
12.4
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100
Illustration 12–3
BAR CODE SAMPLES TO SCAN
REQUISITION #
508
PATIENT ID #
REQUISITION #
510
PATIENT ID #
12.5
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 8 DIRECTION 2181737–100
SECTION 2
ConnectPRO SOFTWARE OPTION INSTALLATION & CHECKOUT
Note 4 Before you install the ConnectPRO option software, obtain the following information
from the hospital’s network administrator:
12.6
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 8 DIRECTION 2181737–100
6. Select Permanent.
7. Select MOD.
8. Select OK to display a software options screen similar to the one shown in Illustration 12–5.
Illustation 12–5
SAMPLE SOFTWARE OPTIONS MANUAL
9. After the ConnectPRO label appears in the “Installed Options” list, a “ConnectPRO
Setup” screen, similar to the one in Illustration 12–6, appears.
10. Type/enter the information you recorded on page 12.6 into the corresponding data
fields on the ConnectPRO Setup screen.
a. If you need additional information, select Help to display a HELP window
explaining each data field.
b. Select Dismiss to exit the ConnectPRO Setup screen without any changes.
12.7
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 8 DIRECTION 2181737–100
Illustation 12–6
ConnectPRO SETUP SCREEN
11. Select Accept to check the valid setup entries, execute a perl script that makes changes
to the configure files for HIS/RIS, and return to the Software Options screen, shown in
Illustration 12–5, on page 12.7.
12. Refer to 12–5.
a. Select Quit.
b. Select ok.
13. Remove the MOD, and write protect the side containing the ConnectPRO option.
Note 5 The initial install requires a write enabled MOD; but you may subsequently install the
option(s) with a write protected MOD.
12.8
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 8 DIRECTION 2181737–100
7. Select Save.
Review the output for errors or missing files. You may use the scroll bar on the right
after the tool completes its tasks.
8. When Save completes, select Dismiss.
12.9
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100
SECTION 3
FINISH INSTALLATION
12.10
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100
SECTION 4
PARTS
4.1 B7540RB Bar Code Reader Only Option Kit Parts List
Item Part Number Name SPP Quantity Description (Remarks)
1 2180690 Bar Code Reader Kit N 1 Bar Code Reader Kit including Y–Cable and rubber mount. 2
4.2 B7500PM HIS/RIS Interface w/Bar Code Reader Option Kit Parts List
Item Part Number Name SPP Quantity Description (Remarks)
1 2180690 Bar Code Reader Kit N 1 Bar Code Reader Kit including Y–Cable and rubber mount
3 2180690–4 Scanner Holder 2 1 Plastic holder ONLY (to store bar code scanner when not in use)
12.11
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100
NOTES
12.12
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 9 DIRECTION 2181737-100
Chapter 13
2nd Host Disk Option
With the system shutdown and the power off, install it behind CDROM and MOD.
If it's an addition to the Octane, simply slide it into the middle or top slot with the power off.
Re-power the system, reload software or open a shell and run this script:
/usr/g/scripts/install2Disk
Use:SERVICE, LOG VIEWING, OC HARDWARE INFO to verify host recognizes the new disk. On
the original console, it will look like this:
EXAMPLE: Integral Ethernet: ec0, version 1
OC HW INFO Integral SCSI controller 1: Version WD33C93B, revision D
CDROM: unit 6 on SCSI controller 1
Optical disk: unit 5 on SCSI controller 1
Optical disk: unit 3 on SCSI controller 1
* Disk drive: unit 2 on SCSI controller 1
Disk drive: unit 1 on SCSI controller 1
Integral SCSI controller 0: Version WD33C93B, revision D
Graphics board: Solid Impact
Graphics board: High Impact
13.1
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 9 DIRECTION 2181737-100
NOTES
13.2
GE Medical Systems: Telex 3797371