GE Medical Systems
CT/i Options Installation

Copyright © 1997, 1998, 1999, 2000, 2002, by


General Electric Company, Inc.
All rights reserved.
Operating Documentation
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100

AVERTISSEMENT CE MANUEL DE MAINTENANCE N’EST DISPONIBLE QU’EN ANGLAIS.


SI LE TECHNICIEN DU CLIENT A BESOIN DE CE MANUEL DANS UNE AUTRE
LANGUE QUE L’ANGLAIS, C’EST AU CLIENT QU’IL INCOMBE DE LE FAIRE
TRADUIRE.
NE PAS TENTER D’INTERVENTION SUR LES ÉQUIPEMENTS TANT QUE LE
MANUEL SERVICE N’A PAS ÉTÉ CONSULTÉ ET COMPRIS.
LE NON-RESPECT DE CET AVERTISSEMENT PEUT ENTRAÎNER CHEZ LE
TECHNICIEN, L’OPÉRATEUR OU LE PATIENT DES BLESSURES DUES À DES
DANGERS ÉLECTRIQUES, MÉCANIQUES OU AUTRES.

WARNUNG DIESES KUNDENDIENST–HANDBUCH EXISTIERT NUR IN ENGLISCHER


SPRACHE.
FALLS EIN FREMDER KUNDENDIENST EINE ANDERE SPRACHE BENÖTIGT,
IST ES AUFGABE DES KUNDEN FÜR EINE ENTSPRECHENDE ÜBERSETZUNG
ZU SORGEN.
VERSUCHEN SIE NICHT, DAS GERÄT ZU REPARIEREN, BEVOR DIESES
KUNDENDIENST–HANDBUCH NICHT ZU RATE GEZOGEN UND VERSTANDEN
WURDE.
WIRD DIESE WARNUNG NICHT BEACHTET, SO KANN ES ZU VERLETZUNGEN
DES KUNDENDIENSTTECHNIKERS, DES BEDIENERS ODER DES PATIENTEN
DURCH ELEKTRISCHE SCHLÄGE, MECHANISCHE ODER SONSTIGE
GEFAHREN KOMMEN.

i
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100

AVISO ESTE MANUAL DE SERVICIO SÓLO EXISTE EN INGLÉS.


SI ALGÚN PROVEEDOR DE SERVICIOS AJENO A GEMS SOLICITA UN IDIOMA
QUE NO SEA EL INGLÉS, ES RESPONSABILIDAD DEL CLIENTE OFRECER UN
SERVICIO DE TRADUCCIÓN.
NO SE DEBERÁ DAR SERVICIO TÉCNICO AL EQUIPO, SIN HABER
CONSULTADO Y COMPRENDIDO ESTE MANUAL DE SERVICIO.
LA NO OBSERVANCIA DEL PRESENTE AVISO PUEDE DAR LUGAR A QUE EL
PROVEEDOR DE SERVICIOS, EL OPERADOR O EL PACIENTE SUFRAN
LESIONES PROVOCADAS POR CAUSAS ELÉCTRICAS, MECÁNICAS O DE
OTRA NATURALEZA.

ATENÇÃO ESTE MANUAL DE ASSISTÊNCIA TÉCNICA SÓ SE ENCONTRA DISPONÍVEL


EM INGLÊS.
SE QUALQUER OUTRO SERVIÇO DE ASSISTÊNCIA TÉCNICA, QUE NÃO A
GEMS, SOLICITAR ESTES MANUAIS NOUTRO IDIOMA, É DA
RESPONSABILIDADE DO CLIENTE FORNECER OS SERVIÇOS DE
TRADUÇÃO.
NÃO TENTE REPARAR O EQUIPAMENTO SEM TER CONSULTADO E
COMPREENDIDO ESTE MANUAL DE ASSISTÊNCIA TÉCNICA.
O NÃO CUMPRIMENTO DESTE AVISO PODE POR EM PERIGO A SEGURANÇA
DO TÉCNICO, OPERADOR OU PACIENTE DEVIDO A‘ CHOQUES ELÉTRICOS,
MECÂNICOS OU OUTROS.

AVVERTENZA IL PRESENTE MANUALE DI MANUTENZIONE È DISPONIBILE SOLTANTO IN


INGLESE.
SE UN ADDETTO ALLA MANUTENZIONE ESTERNO ALLA GEMS RICHIEDE IL
MANUALE IN UNA LINGUA DIVERSA, IL CLIENTE È TENUTO A PROVVEDERE
DIRETTAMENTE ALLA TRADUZIONE.
SI PROCEDA ALLA MANUTENZIONE DELL’APPARECCHIATURA SOLO DOPO
AVER CONSULTATO IL PRESENTE MANUALE ED AVERNE COMPRESO IL
CONTENUTO.
NON TENERE CONTO DELLA PRESENTE AVVERTENZA POTREBBE FAR
COMPIERE OPERAZIONI DA CUI DERIVINO LESIONI ALL’ADDETTO ALLA
MANUTENZIONE, ALL’UTILIZZATORE ED AL PAZIENTE PER FOLGORAZIONE
ELETTRICA, PER URTI MECCANICI OD ALTRI RISCHI.

WARNING THIS SERVICE MANUAL IS AVAILABLE IN ENGLISH ONLY.


IF A CUSTOMER’S SERVICE PROVIDER REQUIRES A LANGUAGE OTHER
THAN ENGLISH, IT IS THE CUSTOMER’S RESPONSIBILITY TO PROVIDE
TRANSLATION SERVICES.
DO NOT ATTEMPT TO SERVICE THE EQUIPMENT UNLESS THIS SERVICE
MANUAL HAS BEEN CONSULTED AND IS UNDERSTOOD.
FAILURE TO HEED THIS WARNING MAY RESULT IN INJURY TO THE SERVICE
PROVIDER, OPERATOR OR PATIENT FROM ELECTRIC SHOCK, MECHANICAL
OR OTHER HAZARDS.

ii
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100

CERTIFIED ELECTRICAL CONTRACTOR STATEMENT


All electrical Installations that are preliminary to positioning of the equipment at
the site prepared for the equipment shall be performed by licensed electrical
contractors. In addition, electrical feeds into the Power Distribution Unit shall be
performed by licensed electrical contractors. Other connections between pieces
of electrical equipment, calibrations and testing shall be performed by qualified
GE Medical personnel. The products involved (and the accompanying electrical
installations) are highly sophisticated, and special engineering competence is
required. In performing all electrical work on these products, GE will use its own
specially trained field engineers. All of GE’s electrical work on these products will
comply with the requirements of the applicable electrical codes.

The purchaser of GE equipment shall only utilize qualified personnel (i.e., GE’s
field engineers, personnel of third–party service companies with equivalent
training, or licensed electricians) to perform electrical servicing on the
equipment.

IMPORTANT...X–RAY PROTECTION
X–ray equipment if not properly used may cause injury. Accordingly, the
instructions herein contained should be thoroughly read and understood by
everyone who will use the equipment before you attempt to place this equipment
in operation. The General Electric Company, Medical Systems Group, will be glad
to assist and cooperate in placing this equipment in use.

Although this apparatus incorporates a high degree of protection against


x–radiation other than the useful beam, no practical design of equipment can
provide complete protection. Nor can any practical design compel the operator
to take adequate precautions to prevent the possibility of any persons carelessly
exposing themselves or others to radiation.

It is important that anyone having anything to do with x–radiation be properly


trained and fully acquainted with the recommendations of the National Council
on Radiation Protection and Measurements as published in NCRP Reports
available from NCRP Publications, 7910 Woodmont Avenue, Room 1016,
Bethesda, Maryland 20814, and of the International Commission on Radiation
Protection, and take adequate steps to protect against injury.

The equipment is sold with the understanding that the General Electric Company,
Medical Systems Group, its agents, and representatives have no responsibility
for injury or damage which may result from improper use of the equipment.

Various protective materials and devices are available. It is urged that such
materials or devices be used.

iii
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100

DAMAGE IN TRANSPORTATION
All packages should be closely examined at time of delivery. If damage is
apparent write ‘‘Damage In Shipment’’ on ALL copies of the freight or express bill
BEFORE delivery is accepted or ‘‘signed for’’ by a GE representative or hospital
receiving agent. Whether noted or concealed, damage MUST be reported to the
carrier immediately upon discovery, or in any event, within 14 days after receipt,
and the contents and containers held for inspection by the carrier. A
transportation company will not pay a claim for damage if an inspection is not
requested within this 14 day period.

Call Traffic and Transportation, Milwaukee, WI (414) 785 5052 or 8*323 5052
immediately after damage is found. At this time be ready to supply name of
carrier, delivery date, consignee name, freight or express bill number, item
damaged and extent of damage.

Complete instructions regarding claim procedure are found in Section S of the


Policy And Procedures Bulletins.
14 July 1993

iv
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 9 DIRECTION 2181737–100

REVISION HISTORY
REV DATE REASON FOR CHANGE
0 . . . . . . . . . . . . . . . . . . 28 February 1997 . . . . . Initial Release of document.
1 . . . . . . . . . . . . . . . . . . 13 May 1997 . . . . . . . . Added Bar Code Reader Option for ConnectPRO, 2nd Disk, Smart Beam
Software, Language Option & Dentascan Installation (Chaps.2 through
6).
2 . . . . . . . . . . . . . . . . . . 27–OCT–1997 . . . . . . . Added Smart Helical Option Installation (Chapter 7).
3 . . . . . . . . . . . . . . . . . . 03–DEC–1997 . . . . . . . Added BMD Option Installation (Chapter 8). Other miscellaneous
corrections & format modifications throughout.
4 . . . . . . . . . . . . . . . . . . 18–May–1998 . . . . . . . Chapters 9, 10, 11 & 12 added to support CT/i (Octane) Computer Op-
tions (memory, dual cpu, optional disk and mag tape)
5 . . . . . . . . . . . . . . . . . . 27–January–1999 . . . . . Chapter 8 (BMD) modified to support the new BMD Option Kit struc-
ture for 4.1 & 5.3 system software.
6 . . . . . . . . . . . . . . . . . . 22–April–1999 . . . . . . . CQA 992109.
7 . . . . . . . . . . . . . . . . . . 20–August–1999 . . . . . CQA’s 993783 & 995718.
8 . . . . . . . . . . . . . . . . . . 27–September–2000 . . Revised for software releases 4.4 & 6.1.
9 . . . . . . . . . . . . . . . . . . 11 June 2002 . . . . . . . . . CQA 1021005: Added 2nd disk option procedure.

LIST OF EFFECTIVE PAGES

PAGE REV PAGE REV PAGE REV PAGE REV PAGE REV
Cover/Blank . . . . 9 3–1 to 3–8 . . . . . . 4 7–2 to 7–3 . . . . . . 8 12–6 to 12–8 . . . . 8 ................
Warnings–i to iv 4 4–i to 4–ii . . . . . . . 4 7–4 to 7–6 . . . . . . 4 12–9 to 12–12 . . . 4 ................
History–v to vi . . 9 4–1 . . . . . . . . . . . . 4 8–i to 8–ii . . . . . . . 5 13-1 to 13-2 . . . . 9
TOC–vii to xv . . 5 4–2 to 4–3 . . . . . . 8 8–1 to 8–6 . . . . . . 5 Rear . . . . . . . . . . . 4
TOC xvi . . . . . . . 9 4–4 to 4–6 . . . . . . 4 9–i to 9–ii . . . . . . . 4 ................ ................
1–i to 1–ii . . . . . . 4 5–i to 5–ii . . . . . . . 4 9–1 to 9–16 . . . . . 4 ................ ................
1–1 to 1–8 . . . . . 4 5–1 to 5–4 . . . . . . 4 10–i to 10–ii . . . . . 4 ................ ................
2–i . . . . . . . . . . . 8 6–i to 6–ii . . . . . . . 4 10–1 to 10–16 . . . 4 ................ ................
2–ii . . . . . . . . . . . 4 6–1 to 6–2 . . . . . . 8 11–i to 11–ii . . . . . 4
2–1 to 2–5 . . . . . 4 6–3 to 6–4 . . . . . . 4 11–1 to 11–12 . . . 7 ................ ................
2–6 to 2–9 . . . . . 8 7–i to 7–ii . . . . . . . 4 12–i to 12–ii . . . . . 4 ................ ................
2–10 to 2–12 . . . 4 7–1 . . . . . . . . . . . . 4 12–1 to 12–5 . . . . 4
3–i to 3–ii . . . . . . 4

v
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 9 DIRECTION 2181737–100

NOTES

vi
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 5 DIRECTION 2181737–100

Table Of Contents

Chapter Description Tab

Chapter 1
CT/i CPU Memory Option . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.i
Section 1
CT/i CPU Memory Option Installation Preparation . . . . . . . . . . . . . . . . . 1.1
1.1 Shutdown the System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.1
1.2 Remove SGI Chassis . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.1

Section 2
Install Optional Memory SIMMs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.4
2.1 Install SIMMs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.4
2.2 Reinstall SGI Chassis . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.4

Section 3
Startup System and Verify Installed Memory . . . . . . . . . . . . . . . . . . . . . . . 1.6
3.1 Start System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.6
3.2 Verify Installed Memory . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.6

Section 4
Finish Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.7
4.1 Install Rating Plate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.7
4.2 Return Installation Card . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.7

Section 5
Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.8
5.1 B7510T Additional DPI Memory Option Kit Parts List . . . . . . . . . . . . 1.8

vii
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 5 DIRECTION 2181737–100

Chapter Description Tab

Chapter 2
CT/i ConnectPRO and Bar Code Reader Option . . . . . . . . . . . . . . . . . . . 2.i
Section 1
CT/i ConnectPRO Option Installation Preparation . . . . . . . . . . . . . . . . . . 2.1
1.1 Bar Code Reader Kit Hardware Installation . . . . . . . . . . . . . . . . . . . . . . 2.1
1.2 Console Hardware Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.2
1.3 Verify Bar Code Scanner Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.4

Section 2
ConnectPRO Software Option Installation & Checkout . . . . . . . . . . . . . . 2.6
2.1 Install and Setup ConnectPRO Option from MOD . . . . . . . . . . . . . . . . 2.6
2.2 ConnectPRO Feature Checkout . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.9
2.3 Save System State . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.9

Section 3
Finish Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.10
3.1 Install Rating Plate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.10
3.2 Return Installation Card . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.10

Section 4
Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.11
4.1 B7540RB Bar Code Reader Only Option Kit Parts List . . . . . . . . . . . . 2.11
4.2 B7500PM HIS/RIS Interface with Bar Code Reader Kit Parts List . . . . 2.11
4.3 Material List for Bar Code Reader Kit 2180690 . . . . . . . . . . . . . . . . . . 2.11

viii
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 5 DIRECTION 2181737–100

Chapter Description Tab

Chapter 3
CT/i Additional System Magnetic Disk Option Installation . . . . . . . . . . . 3.i
Section 1
Prepare for CT/i Additional Magnetic Disk Option Installation . . . . . . . 3.1
1.1 Prepare for Option Disk Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.1
1.2 Save System State Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.1
1.3 Shutdown the System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.2

Section 2
Install Additional Magnetic Disk Option Hardware . . . . . . . . . . . . . . . . . 3.3
2.1 Power Off Console and Install Optional Disk . . . . . . . . . . . . . . . . . . . . 3.3
2.2 Verify Hardware Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.3

Section 3
Load CT/i Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.6
3.1 Install Option Disk Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.6

Section 4
Finish Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.6
4.1 Install Rating Plate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.6
4.2 Return Installation Card . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.6

Section 5
Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.7
5.1 B7510S CT/i Additional System Magnetic Disk Option Kit Parts List . 3.7

ix
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 5 DIRECTION 2181737–100

Chapter Description Tab

Chapter 4
CT/i Smart Beam Option Software Installation . . . . . . . . . . . . . . . . . . . . 4.i

Section 1
CT/i Optional Smart Beam Software Installation . . . . . . . . . . . . . . . . . . . 4.1
1.1 Check Installed Collimator Type . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4.1
1.2 Add Smart Beam Software Option . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4.2

Section 2
Calibrate System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4.4

Section 3
Finish Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4.4
3.1 Install Rating Plate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4.4
3.2 Return Installation Card . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4.4

Section 4
Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4.5
4.1 B7530SB CT/i Smart Beam Software (w/o Collimator) Kit Parts List 4.5
4.2 B7530SC CT/i Smart Beam Collimator Option Kit Parts List . . . . . . . 4.5
4.3 B7530SS CT/i Smart Beam Software (with Collimator) Parts List . . . . 4.5

Chapter 5
CT/i Language Options Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.i

Section 1
Language Option Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.1
1.1 Shutdown the System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.1
1.2 Boot System and Stop For Maintenance . . . . . . . . . . . . . . . . . . . . . . . . 5.1
1.3 If you have a keyboard other than a US–English Keyboard. . . . . . . . . . 5.1

Section 2
Procedures for installing add–on packages . . . . . . . . . . . . . . . . . . . . . . . . . 5.2
2.1 Optional Package Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.2
2.2 Reconfig to Select Language . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.3
2.3 Reconfig SBC to Enable Language . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.3

Section 3
Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.5
3.1 Material Lists for International Language Kits . . . . . . . . . . . . . . . . . . . 5.5

x
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 5 DIRECTION 2181737–100

Chapter Description Tab

Chapter 6
CT/i Dentascan Option Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.i
Section 1
Install Dentascan Option Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.1
1.1 Add Software Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.1

Section 2
Procedures for installing Dentascan filming option . . . . . . . . . . . . . . . . . . 6.3

Section 3
Finish Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.3
3.1 Install Rating Plate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.3
3.2 Return Installation Card . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.4

Section 4
Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.4
4.1 B7540N CT/i Dentascan Option Kit Parts List . . . . . . . . . . . . . . . . . . . 6.4

Chapter 7
CT/i Smart Helical Option Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . 7.i
Section 1
Install CT/i 4.X System Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7.1
1.1 Install 4.X Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7.1

Section 2
PROCEDURES FOR INSTALLING SMART HELICAL SOFTWARE . 7.2

Section 3
Finish Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7.4
3.1 Install Rating Plate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7.4
3.2 Return Installation Card . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7.4

Section 4
Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7.5
4.1 B7530TL Smart Helical Option Kit Parts List (systems not at 4.X) . . . 7.5
4.2 B7530TK Smart Helical Option Kit Parts List (systems at 4.X) . . . . . . 7.5

xi
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 5 DIRECTION 2181737–100

Chapter Description Tab

Chapter 8
CT/i BMD Option Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8.i
Section 1
        
   . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8.1
1.1 " ! !# . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8.1

Section 2
PROCEDURES FOR INSTALLING BMD OPTION SOFTWARE . . . . 8.1
2.1 Install BMD Option Key and Load BMD Option Software . . . . . . . . . . 8.1

Section 3
Special Instructions for Transferring the BMD Database from an Advantage
Windows Workstation to the CT/i . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8.2
3.1 Transfer the BMD Database from AWW to Ct/i . . . . . . . . . . . . . . . . . . 8.2
3.2 Verify that the data has been successfully transferred . . . . . . . . . . . . . . 8.3

Section 4
Backup Database and Deliver Books & Phantoms to Customer . . . . . . . . 8.3
4.1 Backup BMD Database (Save To Mod) . . . . . . . . . . . . . . . . . . . . . . . . . 8.3
4.2 Deliver Books & BMD Phantom to Customer . . . . . . . . . . . . . . . . . . . . 8.3

Section 5
Finish Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8.4
5.1 Install Rating Plate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8.4
5.2 Return Installation Card . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8.4

Section 6
Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8.5
6.1



  ! ! !  ! # ! . 8.5
6.2
 
  ! ! ! !  ! #! ! 8.5

xii
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 5 DIRECTION 2181737–100

Chapter Description Tab

Chapter 9
CT/i (OCTANE) Memory Option . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9.i
Section 1
CT/i (OCTANE) CPU Memory Option Installation Preparation . . . . . . . 9.1
1.1 Shutdown the System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9.1
1.2 Preparing the Workstation to Install the optional memory . . . . . . . . . . . 9.1
1.3 Removing the System Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9.3

Section 2
Installing and Removing Memory . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9.6
2.1 About Memory . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9.6
2.2 Installing Memory . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9.8
2.3 Replacing the System Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9.8
2.4 Powering On the OCTANE Workstation . . . . . . . . . . . . . . . . . . . . . . . . 9.11

Section 3
Startup System and Verify Installed Memory . . . . . . . . . . . . . . . . . . . . . . . 9.12
3.1 Start System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9.12
3.2 Verify Installed Memory . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9.12

Section 4
Finish Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9.13
4.1 Install Rating Plate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9.13
4.2 Return Installation Card . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9.14

Section 5
Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9.15
5.1 Field Replaceable Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9.15
5.2 B7710M Memory Option Kit – Parts List . . . . . . . . . . . . . . . . . . . . . . . 9.15

xiii
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 5 DIRECTION 2181737–100

Chapter Description Tab

Chapter 10
CT/i (OCTANE) Dual Processor Option . . . . . . . . . . . . . . . . . . . . . . . . . . 10.i
Section 1
CT/i OCTANE Dual Processor Option Installation Preparation . . . . . . . 10.1
1.1 Shutdown the System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10.1
1.2 Removing the System Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10.2

Section 2
Installing The Optional Dual Processor . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10.6
2.1 Removing The Single Processor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10.6
2.2 Installing The Optional Dual Processor . . . . . . . . . . . . . . . . . . . . . . . . . 10.7
2.3 Replacing the System Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10.9
2.4 Powering On the OCTANE Workstation . . . . . . . . . . . . . . . . . . . . . . . . 10.12

Section 3
Startup System and Verify Installed Dual Processor . . . . . . . . . . . . . . . . . 10.13
3.1 Start System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10.13
3.2 Verify Installed Dual Processor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10.13

Section 4
Finish Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10.14
4.1 Install Rating Plate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10.14
4.2 Return Installation Card . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10.15

Section 5
Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10.16
5.1 Field Replaceable Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10.16
5.2 B7710JP Dual Processor Option Kit – Parts List . . . . . . . . . . . . . . . . . . 10.16

xiv
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 5 DIRECTION 2181737–100

Chapter Description Tab

Chapter 11
CT/i (OCTANE) Additional Magnetic Disk Option Installation . . . . . . . 11.i
Section 1
Prepare for CT/i (Octane) Additional Magnetic Disk Option Installation 11.1
1.1 Prepare for Option Disk Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . 11.1
1.2 Save System State Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11.1
1.3 Shutdown the System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11.2

Section 2
Install Additional Magnetic Disk Option Hardware . . . . . . . . . . . . . . . . . 11.3
2.1 About Internal Drives . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11.3
2.2 Power Off Console and Install Optional Disk . . . . . . . . . . . . . . . . . . . . 11.3
2.3 Removing the Bezel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11.4
2.4 Installing an Internal Option Drive . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11.7
2.5 Removing the Blank Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11.8
2.6 Replacing the Bezel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11.10
2.7 Verify Hardware Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11.11

Section 3
Load CT/i Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11.11
3.1 Install Option Disk Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11.11

Section 4
Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11.11
4.1 B7710JD CT/i (Octane) Additional Sys. Mag Disk Option Kit . . . . . . . 11.11

xv
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 9 DIRECTION 2181737-100

Chapter Description Tab

Chapter 12
CT/i (Octane) ConnectPRO and Bar Code Reader Option . . . . . . . . . . . 12.i
Section 1
CT/i (Octane) ConnectPRO Option Installation Preparation . . . . . . . . . . 12.1
1.1 Bar Code Reader Kit Hardware Installation . . . . . . . . . . . . . . . . . . . . . . 12.1
1.2 Console Hardware Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12.1
1.3 Verify Bar Code Scanner Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12.4

Section 2
ConnectPRO Software Option Installation & Checkout . . . . . . . . . . . . . . 12.6
2.1 Install and Setup ConnectPRO Option from MOD . . . . . . . . . . . . . . . . 12.6
2.2 ConnectPRO Feature Checkout . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12.9
2.3 Save System State . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12.9

Section 3
Finish Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12.10
3.1 Install Rating Plate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12.10
3.2 Return Installation Card . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12.10

Section 4
Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12.11
4.1 B7540RB Bar Code Reader Only Option Kit Parts List . . . . . . . . . . . . 12.11
4.2 B7500PM HIS/RIS Interface w/Bar Code Reader Option Kit Parts List 12.11
4.3 Material List for Bar Code Reader Kit 2180690 . . . . . . . . . . . . . . . . . . 12.11

Chapter 13
2nd Host Disk Option . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13.1

xvi
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 5 DIRECTION 2181737–100

List Of Illustrations

Illustration Page

Illustration 1–1
SGI Chassis Cover Removal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.2
Illustration 1–2
Disconnect SGI Midplane Power Cables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.3
Illustration 1–3
SGI Chassis Cover Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.4
Illustration 1–4
CT/i Host Computer Memory Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.5
Illustration 1–5
Rating Plate Location . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.7
Illustration 2–1
Bar Code Reader Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.2
Illustration 2–2
Bar Code Reader Installation (detail) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.3
Illustration 2–4
Options Window . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.6
Illustration 2–5
Sample Software Options Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.7
Illustration 2–6
ConnectPRO Setup screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.8
Illustration 2–7
Rating Plate Location . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.10
Illustration 3–1
Option Disk Installation Drawing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.4
Illustration 3–2
User Drive Assembly Exploded View . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.5
Illustration 3–3
Rating Plate Location . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.6
Illustration 4–1
Software Options Installation Window . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4.2
Illustration 4–2
Software Options Available / Installed Screen Example . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4.3
Illustration 4–3
Rating Plate Location . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4.4
Illustration 6–1
Software Options Installation Window . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.1
Illustration 6–2
Software Options Available / Installed Screen Example . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.2
Illustration 6–3
Rating Plate Location . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.3

xvii
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 5 DIRECTION 2181737–100

Illustration Page

Illustration 7–1
Software Options Installation Window . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7.2
Illustration 7–2
Software Options Available / Installed Screen Example . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7.3
Illustration 7–3
Rating Plate Location . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7.4
Illustration 8–1
Rating Plate Location . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8.4
Illustration 9–1
LOck Bar Removal (If Applicable) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9.2
Illustration 9–2
Removing the System Module Screws . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9.3
Illustration 9–3
Locating the Compression Connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9.3
Illustration 9–4
Releasing the System Module form the Front Plane . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9.4
Illustration 9–5
Identifying DIMM Sockets and DIMM Banks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9.7
Illustration 9–6
LOcating DIMMs on the System Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9.7
Illustration 9–7
Locating the Notches on a DIMM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9.8
Illustration 9–8
Removing the Compression Connector Cap . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9.8
Illustration 9–9
Supporting the System Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9.9
Illustration 9–10
Replacing the System Module in the OCTANE Workstation . . . . . . . . . . . . . . . . . . . . . . . . . . 9.10
Illustration 9–11
Locking the System Module to the Frontplane . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9.10
Illustration 9–12
Replacing the System Module Screws . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9.11
Illustration 9–13
Rating Plate Location . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9.13
Illustration 10–1
LOck Bar Removal (If Applicable) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10.2
Illustration 10–2
Removing the System Module Screws . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10.3
Illustration 10–3
Locating the Compression Connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10.3
Illustration 10–4
Placing the Compression Cap on the Compression Connector . . . . . . . . . . . . . . . . . . . . . . . . . 10.5
Illustration 10–5
Removing the Single Processor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10.6
Illustration 10–6
Viewing The Connectors on the Underside of the CPUs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10.7

xviii
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 5 DIRECTION 2181737–100

Illustration Page

Illustration 10–7
Installing the Dual Processor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10.8
Illustration 10–8
Removing the Compression Connector Cap . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10.9
Illustration 10–9
Supporting the System Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10.10
Illustration 10–10
Replacing the System Module in the OCTANE Workstation . . . . . . . . . . . . . . . . . . . . . . . . . . 10.11
Illustration 10–11
Locking the System Module to the Frontplane . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10.11
Illustration 10–12
Replacing the System Module Screws . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10.12
Illustration 10–13
Rating Plate Location . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10.14
Illustration 11–1
Locating the Lockbar on the Front of the OCTANE Workstation . . . . . . . . . . . . . . . . . . . . . . . 11.3
Illustration 11–2
Locating the Lockbar on the Octane Workstation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11.4
Illustration 11–3
Unlocking and Removing the Lockbar . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11.5
Illustration 11–4
Pressing the Bezel Release Buttons . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11.5
Illustration 11–5
Removing the Bezel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11.6
Illustration 11–6
Identifying the Option Drive Bays . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11.7
Illustration 11–7
Sliding In an Optional Drive . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11.7
Illustration 11–8
Locking the Drive . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11.8
Illustration 11–9
Opening the Door on the Bezel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11.8
Illustration 11–10
Pushing the Blank Panel Free . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11.9
Illustration 11–11
Replacing the Bezel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11.10
Illustration 11–12
Replacing The LockBar . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11.10
Illustration 12–1
Bar Code Reader Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12.2
Illustration 12–2
Bar Code Reader Instalation (Detail) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12.3
Illustration 12–3
Bar CoDE SAMPLEs TO SCAN . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12.5
Illustration 12–4
Options Window . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12.6

xix
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 5 DIRECTION 2181737–100

Illustration Page

Illustration 12–5
Sample Software Options Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12.7
Illustration 12–6
ConnectPRO Setup screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12.8
Illustration 12–7
Rating Plate Location . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12.10

xx
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100

Chapter 1
CT/i CPU Memory Option

Table Of Contents

Section Description Page

Section 1
CT/i CPU Memory Option Installation Preparation . . . . . . . . . . . . . . . . . 1.1
1.1 Shutdown the System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.1
1.2 Remove SGI Chassis . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.1

Section 2
Install Optional Memory SIMMs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.4
2.1 Install SIMMs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.4
2.2 Reinstall SGI Chassis . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.4

Section 3
Startup System and Verify Installed Memory . . . . . . . . . . . . . . . . . . . . . . . 1.6
3.1 Start System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.6
3.2 Verify Installed Memory . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.6

Section 4
Finish Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.7
4.1 Install Rating Plate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.7
4.2 Return Installation Card . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.7

Section 5
Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.8
5.1 B7510T Additional DPI Memory Option Kit Parts List . . . . . . . . . . . . 1.8

1.i
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100

List Of Illustrations

Illustration Page

Illustration 1–1
SGI Chassis Cover Removal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.2
Illustration 1–2
Disconnect SGI Midplane Power Cables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.3
Illustration 1–3
SGI Chassis Cover Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.4
Illustration 1–4
CT/i Host Computer Memory Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.5
Illustration 1–5
Rating Plate Location . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.7

1.ii
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100

SECTION 1
CT/I CPU MEMORY OPTION INSTALLATION PREPARATION

The CT/i CPU Memory Option, B7510T, provides an additional 64 Mb of CPU memory for the
Silicon Graphics Computer. This memory allows more efficient execution of system software
by reducing the amount of disk swap activity and allowing larger amounts of data to be available
in memory at any given moment.

1.1 Shutdown the System


1. Select the (square red) Shutdown Button (shown below) on the Image Display Monitor.
Select OK to shutdown Applications. This will stop both CT/i Applications and UNIX
software and bring you to the boot prompt.

2. After the software shuts down, the following message will appear.

i Okay to power off the system now.


Press any key to restart
Restart

3. Turn off the power switch on the console.

1.2 Remove SGI Chassis

CAUTION THE CIRCUIT BOARDS ARE VERY STATIC–SENSITIVE!!


Static–sensitive components may be damaged if not handled in a static–free
Static Sensitive Devices environment. Take appropriate precautions (e.g., wear properly grounded wrist strap)
when handling all boards.

1. Remove the lower front cover of the console.

1.1
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100

2. Remove the lower right cover over the Scan Chassis.


3. Loosen the single captive screw holding the cable bulkhead panel to the side of the scan
chassis and rotate the bulkhead cable assembly out of the console and onto the floor.
4. Loosen the six (6) captive screws holding the SGI CPU Chassis into the console.
Note 1 It may be necessary to disconnect some of the cables from the front of the SGI Chassis in
order to remove it from the console. Whether or not this is required depends upon cable
routing within the console.
5. Being careful of the cable connections, slide the SGI CPU Chassis out of the console
base and lay on it’s side. Refer to Illustration 1–1.
6. Follow the steps in Illustration 1–1 to remove the cover from the SGI CPU Chassis.
7. Follow the steps in Illustration 1–2 and disconnect the Midplane Power Connections in
order to allow access to the Motherboard SIMM Slots.
Illustration 1–1
SGI CHASSIS COVER REMOVAL

Cover Removal

(1) Remove 6 screws from cover.

(2) Lay chassis on right side.

(3) Rotate cover up and lift finger tabs


out of chassis.

(4) Set cover aside.

1.2
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100

CAUTION THE CIRCUIT BOARDS ARE VERY STATIC–SENSITIVE!!


Static–sensitive components may be damaged if not handled in a static–free
Static Sensitive Devices environment. Take appropriate precautions (e.g., wear properly grounded wrist strap)
when handling all boards.

Illustration 1–2
DISCONNECT SGI MIDPLANE POWER CABLES

ÏÏÏÏÏ Prepare for Memory

ÏÏÏÏÏ Installation/Removal

ÏÏÏÏÏ (1) Unplug Midplane Power cable from

ÏÏÏÏÏ backside of midplane at J10.

ÏÏÏÏÏ (2) Move the disconnected end of the

ÏÏÏÏÏ cable up over the power supply and

ÏÏÏÏÏ
out of the way.

(1) (3) Unplug Midplane Power Sense cable


from backside of midplane at J9.
J10
(4) Move the disconnected end of the
(3) cable up over the power supply and
Verify out of the way.
connections are J9 BACK
properly seated VIEW
when re–installed.

CAUTION THE CIRCUIT BOARDS ARE VERY STATIC–SENSITIVE!!


Static–sensitive components may be damaged if not handled in a static–free
Static Sensitive Devices environment. Take appropriate precautions (e.g., wear properly grounded wrist strap)
when handling all boards.

1.3
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100

SECTION 2
INSTALL OPTIONAL MEMORY SIMMS

2.1 Install SIMMs


1. Follow the steps for SIMM installation in Illustration 1–4, and place the four (4) new
SIMM modules in Bank A, Slots 1, 2, 3, and 4 of the SGI mother board.
Note 2 SIMMs must be installed in groups of four (4) and the groups of four (4) must be of the same
memory capacity. i.e. four 16Mb SIMMs in slots 1, 2, 3, and 4.
2. Re–install Midplane Power Cables by reversing the steps in Illustration 1–2.

2.2 Reinstall SGI Chassis


1. Re–install the SGI CPU Chassis cover. Refer to Illustration 1–3.
2. Being careful of the cable connections, slide the SGI CPU Chassis into the console base.
Note 3 If any cables where disconnected while removing the SGI Chassis from the console,
re–connect the cables to the chassis noting connector labels on cables and chassis.
3. Re–install the cable bulkhead assembly into the side of the Scan Chassis and tighten the
captive screw.
4. Replace the lower right cover over the Scan Chassis.
5. Replace the lower front cover of the console.
Illustration 1–3
SGI CHASSIS COVER INSTALLATION

(4)
(2)

(3) SGI Chassis Cover Installation


(3)
(1) Insert finger tabs into slots in
(1) chassis as shown (at an angle).

(2) Rotate the cover closed.

(3) Install 2 screws to draw chassis


closed.

(4) Install four screws at.

(5) Tighten all 6 screws.

1.4
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100

Illustration 1–4
CT/I HOST COMPUTER MEMORY REPLACEMENT

Use a
wrist strap at TO REMOVE SIMMS:
a grounded
work station. 1. Always use proper static precautions!
2. Locate the memory SIMM you want to remove.
WITH THE I/O CONNECTORS TOWARDS YOU, YOU WILL NEED
TO ALSO REMOVE THE SIMMS IN FRONT OF THE ONE YOU
WANT TO REMOVE.

3. Slide a small, flat blade screwdriver in between the SIMM and


the latch. The SIMM will release and lean forward.

ÏÏÏÏÏÏÏ 4. Repeat step 3. on the other SIMM latch.

ÏÏÏÏÏÏÏ
ÏÏÏÏÏÏÏ
TO INSTALL SIMMS:

ÏÏÏÏÏÏÏ 1. Insert SIMM into connector at an angle,

ÏÏÏÏÏÏÏ 2. swing to right until SIMM ‘clicks’ into

ÏÏÏÏÏÏÏ 32Mb SIMMS


(Bank C, SIMM
place

Slots 9,10, 11,


12)

16Mb SIMMS Verify


(Bank B, SIMM positioning
Slots 5,6,7,8)
of notch.
Verify
16Mb SIMMS pins are aligned
Optional Memory prior to final installation
(Bank A, SIMM
Slots 1,2,3,4)

BASE CT/I MEMORY CONFIGURATION IS 192MB


32 Mb SIMMS (8 SIMMS – 4 X 32 MB IN BANK C + 4 X 16 MB IN
currently have devices on BANK B)
both sides, but this may
change as memory CT/I SUPPORTS AN ADDITIONAL 64 MB OF
density increases. PURCHASED OPTION MEMORY
(4 X 16 MB) IN BANK A, SIMM SLOTS 1, 2, 3, 4

1.5
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100

SECTION 3
STARTUP SYSTEM AND VERIFY INSTALLED MEMORY

3.1 Start System


1. Turn on the console power switch.
2. The following message will appear. Select the Stop for Maintenance button.

Starting up the system...


Stop for Maintenance

3.2 Verify Installed Memory


The system Maintenance menu appears.
1. Select “Enter Command Monitor”.
2. At the shell prompt type:
>> hinv
3. Verify that the Main memory size: is 256Mb.
System : IP22
Processor : 200Mhz R4400, with FPU
Primary I–cache : 16 kbytes
Primary D–cache : 16 Kbytes
Secondary cache size : 2048 Kbytes
Memory size : 256Mbytes
Graphics : High Impact
Graphics : MG1,0 Impact
SCSI Disk : scsi(0)disk(1)
SCSI Disk : scsi(1)disk(1)
SCSI Device : Controller 1ID3
SCSI Device : Controller 1ID5
SCSI CDROM : scsi(1)cdrom(6)
Audio : Iris Audio Processor: version A2 revision 1.1.0

If the reported Memory size is not 256Mbytes, open the SGI chassis and check/re–seat the
memory SIMMs.
4. Exit the shell:
>> exit
5. Select “Start System”

1.6
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100

The system software will start normally.


6. After the system is started, open a shell tool and re–check the hardware environment:
{ctuser@somesystem} [1] hinv
Iris Audio Processor: version A2 revision 1.1.0
1 200 MHZ IP22 Processor
FPU: MIPS R4000 Floating Point Coprocessor Revision: 0.0
CPU: MIPS R4400 Processor Chip Revision: 6.0
On–board serial ports: 2
On–board bi–directional parallel port
Data cache size: 16 Kbytes
Instruction cache size: 16 Kbytes
Secondary unified instruction/data cache size: 2 Mbytes on Processor 0
Main memory size: 256 Mbytes
EISA bus: adapter 0
Integral Ethernet: et0, IO0
Integral Ethernet: ec0, version 1
Integral SCSI controller 1: Version WD33C93B, revision D
CDROM: unit 6 on SCSI controller 1
Optical disk: unit 5 on SCSI controller 1
Optical disk: unit 3 on SCSI controller 1
Disk drive: unit 1 on SCSI controller 1
Integral SCSI controller 0: Version WD33C93B, revision D
Graphics board: Solid Impact
Graphics board: High Impact

If the reported Memory size is not 256Mbytes, open the SGI chassis and check/re–seat the
memory SIMMs.

SECTION 4
FINISH INSTALLATION

4.1 Install Rating Plate


1. Locate the Modification Rating plate, to the left of the Rating Plates already on the rear
of the cabinet. Refer to Illustration 1–5.

CT/I CONSOLE REAR VIEW

NEW RATING PLATE TO THE


LEFT OF EXISTING RATING
PLATES.

Illustration 1–5
RATING PLATE LOCATION

4.2 Return Installation Card


1. Complete/Verify the site specific information on the Red Installation Card and return to
the appropriate Location printed on the front of the card.

1.7
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100

SECTION 5
PARTS

5.1 B7510T Additional DPI Memory Option Kit Parts List


Item Part Number Name SPP Quantity Description (Remarks)
1 2115457–9 Memory Collector 1 1 Collector containing four 16Mb SIMMs

2 46–302200P3 Serial Rating Plate N 1 Option Kit Rating Plate

1.8
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 8 DIRECTION 2181737–100

Chapter 2
CT/i ConnectPRO and Bar Code Reader Option

Table Of Contents

Section Description Page

Section 1
CT/i ConnectPRO Option Installation Preparation . . . . . . . . . . . . . . . . . . 2.1
1.1 Bar Code Reader Kit Hardware Installation . . . . . . . . . . . . . . . . . . . . . . 2.1
1.2 Console Hardware Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.2
1.3 Verify Bar Code Scanner Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.4

Section 2
ConnectPRO Software Option Installation & Checkout . . . . . . . . . . . . . . 2.6
2.1 Install and Setup ConnectPRO Option from MOD . . . . . . . . . . . . . . . . 2.6
2.2 ConnectPro Feature Checkout . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.8
2.3 Save System State . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.9

Section 4
Finish Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.10
4.1 Install Rating Plate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.10
4.2 Return Installation Card . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.10

Section 4
Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.11
4.1 B7540RB Bar Code Reader Only Option Kit Parts List . . . . . . . . . . . . 2.11
4.2 B7500PM HIS/RIS Interface with Bar Code Reader
Option Kit Parts List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.11
4.3 Material List for Bar Code Reader Kit 2180690 . . . . . . . . . . . . . . . . . . 2.11

2.i
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100

List Of Illustrations

Illustration Page

Illustration 2–1
Bar Code Reader Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.2
Illustration 2–2
Bar Code Reader Installation (Detail) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.3
Illustration 2–3
Bar Code Samples to Scan . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.5
Illustration 2–4
Options Window . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.6
Illustration 2–5
Sample Software Options Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.7
Illustration 2–6
ConnectPRO Setup Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.8
Illustration 2–7
Rating Plate Location . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.10

2.ii
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100

SECTION 1
CT/i ConnectPRO OPTION INSTALLATION PREPARATION

Two versions of the CT/i ConnectPRO Option exist:


B7540RB Bar Code Reader Only
Use the hand–held bar code scanner to directly input information to the Exam Rx patient
information screen. The Bar Code Reader hardware option reads information from the paper
schedule or patient wrist strap bar codes.
B7500PM HIS/RIS Interface software WITH Bar Code Reader
Use the HIS/RIS software option to network patient information directly to the CT/i Exam Rx
desktop from the Hospital Information System (HIS) and Radiology Information Systems (RIS)
database. This option kit also contains the hand–held bar code reader.
Before you start the option installation, take an inventory of the supplied kit.
Refer to the “PARTS” Section on page 2.11 for the corresponding list(s).

1.1 Bar Code Reader Kit Hardware Installation


1. Shutdown the System:
a. Select the (square red) Shutdown Button (shown below) on the Image Display
Monitor.
b. Select OK to shutdown Applications. Shutdown stops the CT/i Applications and the
UNIX software, then displays the boot prompt.

2. After the software shuts down, the following message appears:

i Okay to power off the system now.


Press any key to restart
Restart

3. Turn off the power switch on the console.

2.1
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100

1.2 Console Hardware Installation


1. Remove the lower front cover of the CT/i console.
2. Route the bar code scanner cable through the opening in the rear of the console, and run
the cable connector to the front of the console. (Refer to Illustration 2–1 for details)
3. Attach the Y–cable assembly (2180690–2) to the bar code scanner cord.
(Refer to Illustration 2–2 for details)
4. Unplug the existing keyboard connector from J42 on the SGI chassis.
(Refer to Illustration 2–2 for details)
5. Attach keyboard connector J42 to the short length female connector on the Y–Cable
assembly. (Refer to Illustration 2–2 for details)
6. Connect the long length male connector from the Y–Cable assembly to J42 on the SGI
chassis. (Refer to Illustration 2–2 for details)
7. Optional: Mount the bar code scanner’s holder to the left side of the Exam Rx monitor
–or– to a location selected by the site Technologist. Use the double–sided tape
(supplied with kit) to attach the holder.
Illustration 2–1
BAR CODE READER INSTALLATION

Bar Code Scanner


Route Cable through the
opening in the rear of the console

CONSOLE
Items removed for clarity

See Illustration 2–2 for Detail Enlargement

2.2
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100

Illustration 2–2
BAR CODE READER INSTALLATION (DETAIL)

J42

Existing Cable
Disconnect from J42

Y–Cable assembly
(2180690–3)

From Bar Code Scanner

2.3
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100

1.3 Verify Bar Code Scanner Operation


Note 1 Do not replace the console front cover until you verify the bar code scanner operation.
1. With the console front cover still removed, restore power to the console and boot up the
system to the scan user level.
2. Select Patient Schedule and Add Patient on the Exam Rx desktop.
3. Use the hand–held bar code scanner to scan the PATIENT ID# bar code on one of the
Bar code examples in Illustration 2–3. Make sure the screen updates with the
corresponding number, either 555 666 777 or 777 888 999.
Note 2 The bar code scanner beeps when it successfully reads a bar code.
Recommended: New users, take some time to practice with the bar code scanner and bar
code samples. Delete the screen entry and reread the bar codes until you feel comfortable
with the bar code scanner operation.
A User’s Manual comes with the scanner kit. It contains detailed usage information, as well
as additional sample bar codes. If you encounter difficulties, refer to the Operation User
Maintenance and Troubleshooting section of this manual. If you still cannot successfully
scan the sample bar codes, select the Troubleshoot menu from the service desktop, and run
the Keyboard Confidence Test diagnostic to identify a bad cable connection or failed FRU.
4. Select the Requisition # data field.
5. Use the hand held bar code scanner to scan the REQUISITION# bar code
on one of the Bar code examples in Illustration 2–3. Make sure the screen updates
with the corresponding number, either 508 or 510.
6. Manually type/enter the Patient Name, and select Accept.
7. Select View Schedule
8. Make sure the Schedule screen displays the test patient entry.
9. Select Delete.
10. Select Quit.
11. This completes the bar code scanner and keyboard function test.
12. Reinstall the console front cover.
Note 3 For customer option #B7500PM ONLY, proceed to Section 2.

2.4
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100

Illustration 2–3
BAR CODE SAMPLES TO SCAN

REQUISITION #

508

PATIENT ID #

555 666 777

REQUISITION #

510

PATIENT ID #

777 888 999

2.5
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 8 DIRECTION 2181737–100

SECTION 2
ConnectPRO SOFTWARE OPTION INSTALLATION & CHECKOUT

Note 4 Before you install the ConnectPRO option software, obtain the following information
from the hospital’s network administrator:

HIS Server Ip Address ________________________________


HIS Server AE Title ________________________________
HIS Server AE Port# ________________________________

CT Server AE Title ________________________________

2.1 Install and Setup of ConnectPRO Option from MOD


1. Open the Service Desktop.
2. Select the Utilities Menu.

3. Select installOptions to display the options window shown in Illustration 2–4.


Illustation 2–4
OPTIONS WINDOW

2.6
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 8 DIRECTION 2181737–100

4. Insert the option MOD into the drive.


5. Select Install.

6. Select Permanent.

7. Select MOD.

8. Select OK to display a software options screen similar to the one shown in Illustration 2–5.
Illustation 2–5
SAMPLE SOFTWARE OPTIONS MANUAL

9. After the ConnectPRO label appears in the “Installed Options” list, a “ConnectPRO
Setup” screen appears similar to the one in Illustration 2–6.
10. Type/enter the information you recorded on page2.6 into the corresponding data fields
on the ConnectPRO Setup screen.
a. If you need additional information, select Help to display a HELP window
explaining each data field.
b. Select Dismiss to exit the ConnectPRO Setup screen without any changes.

2.7
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 8 DIRECTION 2181737–100

Illustation 2–6
ConnectPRO SETUP SCREEN

11. Select Accept to check the valid setup entries, execute a perl script that makes changes
to the configure files for HIS/RIS, and return to the Software Options screen, shown in
Illustration 2–5, on page 2.7.
12. Refer to Illustration 2–5.
a. Select Quit.

b. Select ok.

13. Remove the MOD, and write protect the side containing the ConnectPRO option.
Note 5 The initial install requires a write enabled MOD; but you may subsequently install the
option(s) with a write protected MOD.

2.2 ConnectPRO Feature Checkout


1. Start up the Exam Rx applications.
2. Query the HIS/RIS system to verify operation.
(Refer to the User Manual/CBT for details.)
Note 6 To troubleshoot the network hardware: Open a Shell Tool, and type:
more /usr/g/service/networkHelp.txt
to display the network diagnostics help menu, and follow the instructions displayed on the
screen.

2.8
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 8 DIRECTION 2181737–100

2.3 Save System State


The following procedure saves Characterization Data, Calibration Data, Protocols, and
reconfig information, to your System State MOD. If you don’t have a System State MOD,
skip this section.
Note 7 Do not display any images on Image Works when you save the System State. Exit to the
Browser level before you Save System State.
1. Insert the Save System State MOD into the MOD drive.
2. Open the Service desktop.
3. Select the PM option from the list.
4. Select System State.

5. Select the Start button.


6. Select All.

7. Select Save.

Review the output for errors or missing files. You may use the scroll bar on the right
after the tool completes its tasks.
8. When Save completes, select Dismiss.

2.9
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100

SECTION 3
FINISH INSTALLATION

3.1 Install Rating Plate


1. Refer to Illustration 2–7.
2. Locate the current Modification Rating plate(s) on the rear of the CT/i console.
3. Fasten the Option Disk rating plate to the left of the existing rating plates.
Illustration 2–7
RATING PLATE LOCATION

CT/I CONSOLE REAR VIEW

NEW RATING PLATE TO THE


LEFT OF EXISTING RATING
PLATES.

3.2 Return Installation Card


1. Complete and verify the site–specific information on the Red Installation Product
Locator Card.
2. Return the card to the Mailing Address that is printed on the front of the card.

2.10
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100

SECTION 4
PARTS

4.1 B7540RB Bar Code Reader Only Option Kit Parts List
Item Part Number Name SPP Quantity Description (Remarks)
1 2180690 Bar Code Reader Kit N 1 Bar Code Reader Kit including Y–Cable and rubber mount. 2

2 46–302200P3 Serial Rating Plate N 1 Option Kit Rating Plate

4.2 B7500PM HIS/RIS Interface with Bar Code Reader Option Kit Parts List
Item Part Number Name SPP Quantity Description (Remarks)
1 2180690 Bar Code Reader Kit N 1 Bar Code Reader Kit including Y–Cable and rubber mount

2 2182070 Option MOD N 1 HIS/RIS Option MOD

4 46–302200P3 Serial Rating Plate N 1 Option Kit Rating Plate

4.3 Material List for Bar Code Reader Kit 2180690


Item Part Number Name SPP Quantity Description (Remarks)
1 2180690–2 Bar Code Scanner 1 1 Hand Held Bar Code Scanner ONLY

2 2180690–3 Y–cable assembly. 1 1 Console/Keyboard/Scanner interface cable ONLY

3 2180690–4 Scanner Holder 2 1 Plastic holder ONLY (to store bar code scanner when not in use)

2.11
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100

NOTES

2.12
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100

Chapter 3
CT/i Additional System Magnetic Disk Option
Installation

Table Of Contents

Section Description Page

Section 1
Prepare for CT/i Additional Magnetic Disk Option Installation . . . . . . . 3.1
1.1 Prepare for Option Disk Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.1
1.2 Save System State Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.1
1.3 Shutdown the System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.2

Section 2
Install Additional Magnetic Disk Option Hardware . . . . . . . . . . . . . . . . . 3.3
2.1 Power Off Console and Install Optional Disk . . . . . . . . . . . . . . . . . . . . 3.3
2.2 Verify Hardware Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.3

Section 3
Load CT/i Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.6
3.1 Install Option Disk Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.6

Section 4
Finish Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.6
4.1 Install Rating Plate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.6
4.2 Return Installation Card . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.6

Section 5
Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.7
5.1 B7510S CT/i Additional System Magnetic Disk Option Kit Parts List . 3.7

3.i
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100

List Of Illustrations

Illustration Page

Illustration 3–1
Option Disk Installation Drawing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.4
Illustration 3–2
User Drive Assembly Exploded View . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.5
Illustration 3–3
Rating Plate Location . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.6

3.ii
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 6 DIRECTION 2181737–100

SECTION 1
PREPARE FOR CT/I ADDITIONAL MAGNETIC DISK OPTION INSTALLATION

This section describes the correct installation and configuration for B7510S,
CT/i Additional System Magnetic Disk Option. (Pioneer is not an Option)

1.1 Prepare for Option Disk Installation

WARNING THE INSTALLATION OF THE OPTIONAL MAGNETIC DISK REQUIRES A LOAD


FROM COLD OF THE SYSTEM SOFTWARE.

WARNING MAKE SURE THAT ALL IMAGES HAVE BEEN RECONSTRUCTED,


ARCHIVED, FILMED, AND NETWORKED BEFORE PERFORMING THIS
PROCEDURE. THIS PROCEDURE WILL RE–INITIALIZE ALL
SYSTEM DATA DISK, ERASING ALL IMAGES AND SCAN DATA.

1.2 Save System State Data


Before beginning the software install, you will want to, save the system configuration information, character-
ization, calibration, protocols, etc . . .
(1) Bring the system up if it is not already up.
(2) Select the Service desktop.
(3) Select PM.
(4) Highlight System State.
(5) Select the “Start ...” button.
(6) Select “All” (which will highlight Cals, Characterization, Reconfig Info, etc).
(7) Select Save. (The save will take a few minutes)
Note 1 Review the output for errors or missing files; the scroll bar on the right works only when the
tool isn’t busy performing some task –– it takes a little while.
If System State reported no problems, then you may skip to next section, else if you see any
missing files or failures, then refer to the note below.
Note 2 SAVE/RESTORE SYSTEM STATE ERRORS
Save/Restore System State reports status on each file specification listed in the
/usr/g/config/SysState.cfg file. The status may be:
› saved (or restored)
› does not exist, not saving (or restoring)
› FAILED
A status of “does not exist” or “FAILED”, indicates potential problems. Please review the
status for each file to see if any errors are reported.

3.1
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 6 DIRECTION 2181737–100

WARNING MAKE SURE THAT ALL IMAGES HAVE BEEN RECONSTRUCTED,


ARCHIVED, FILMED, AND NETWORKED BEFORE PERFORMING THIS
PROCEDURE. THIS PROCEDURE WILL RE–INITIALIZE ALL
SYSTEM DATA DISK, ERASING ALL IMAGES AND SCAN DATA.

(8) Select Dismiss.

1.3 Shutdown the System


To shut down system software:
(1) Select the (square red) Shutdown button (shown below).
(2) Press OK to shutdown Applications.

This will take down both CT/i Applications and UNIX and put you at the boot prompt.

3.2
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 6 DIRECTION 2181737–100

SECTION 2
INSTALL ADDITIONAL MAGNETIC DISK OPTION HARDWARE

2.1 Power Off Console and Install Optional Disk


(1) After the software is stopped, shut off the console power switch.
(2) Remove the CT/i Console Front Cover.
(3) Remove the EMC cover over the Scan Chassis and User Drive Assemblies.
(4) Remove two screws holding the User Drive Assembly to the chassis.

WARNING THE USER DRIVE ASSEMBLY WEIGHS APPROXIMATELY 10 KILOGRAMS.


SUPPORT THE WEIGHT OF THE ASSEMBLY WHEN REMOVING THE
ASSEMBLY FROM THE CHASSIS.

(5) Keeping the cables clear of the assemblies, pull the User Drive Assembly out of the
chassis and set on the floor in front of the console.
(6) Refer to Illustration 3–1. Verify the new drive jumper settings.
› Option Disk Drive SCSI ID 2.
(7) Refer to Illustration 3–1, Notes 1 thru 4. Install the Optional Disk using the kit
hardware as shown in Illustration 3–1.
(8) Reverse steps (5) through (2) above.

2.2 Verify Hardware Installation


(1) Turn on the CT/i console.
(2) Watch the Display Screen and select “Stop for Maintenance” when displayed.
(3) Select “Command Shell” from the menu.
(4) Enter:
hinv <ENTER>
(5) Refer to the following example output. Verify the Output contains the information
regarding Disk Drive Unit 2 on SCSI Controller 1:
Integral Ethernet: ec0, version 1
Integral SCSI controller 1: Version WD33C93B, revision D
CDROM: unit 6 on SCSI controller 1
Optical disk: unit 5 on SCSI controller 1
Optical disk: unit 3 on SCSI controller 1
* Disk drive: unit 2 on SCSI controller 1
Disk drive: unit 1 on SCSI controller 1
Integral SCSI controller 0: Version WD33C93B, revision D
Graphics board: Solid Impact
Graphics board: High Impact

3.3
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 6 DIRECTION 2181737–100

Illustration 3–1
OPTION DISK INSTALLATION DRAWING

3.4
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 6 DIRECTION 2181737–100

Illustration 3–2
USER DRIVE ASSEMBLY EXPLODED VIEW

User Drive Assembly


SBC HSD
SBC LSD

OC System Disk

OC Optional System
Disk

MaxOptics MOD

CDROM Drive
SCSI TERMINATOR

Pioneer MOD-NOT OPTIONAL

3.5
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 6 DIRECTION 2181737–100

SECTION 3
LOAD CT/I SOFTWARE

3.1 Install Option Disk Software


(1) Refer to Direction 2168070–100, HSA CT/i – Load From Cold Procedure and reload
software.
Note 3 The Optional Disk will be automatically partitioned and made available to the system
software during the software load.

SECTION 4
FINISH INSTALLATION

4.1 Install Rating Plate


1. Refer to Illustration 3–3.
2. Locate the current Modification Rating plate(s) on the rear of the CT/i console.
3. Fasten the Option Disk rating plate to the left of the existing rating plates.
Illustration 3–3
RATING PLATE LOCATION

CT/I CONSOLE REAR VIEW

NEW RATING PLATE TO THE


LEFT OF EXISTING RATING
PLATES.

4.2 Return Installation Card


1. Complete and verify the site–specific information on the Red Installation Product
Locator Card.
2. Return the card to the Mailing Address that is printed on the front of the card.

3.6
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 6 DIRECTION 2181737–100

SECTION 5
PARTS

5.1 B7510S CT/i Additional System Magnetic Disk Option Kit Parts List
Item Part Number Name SPP Quantity Description (Remarks)

1 2180997 4GB DISK DRIVE 1 1 OC System First or Second Optional software disk, Compatible
with Revision 3.6 or later software.
2 46–170015P10 Screw 2 4 6–32 X 1/4” LG BIND HD ST Screw

3 2150539IDW Drawing N 1 Option Disk Installation Drawing (same as Ill. 3–1)

4 46–302200P3 Serial Rating Plate N 1 Option Kit Rating Plate

3.7
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 6 DIRECTION 2181737–100

NOTES

3.8
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100

Chapter 4
CT/i Smart Beam Option Software Installation

Table Of Contents
Section Description Page

Section 1
CT/i Optional Smart Beam Software Installation . . . . . . . . . . . . . . . . . . . 4.1
1.1 Check Installed Collimator Type . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4.1
1.2 Add Smart Beam Software Option . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4.2

Section 2
Calibrate System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4.4

Section 3
Finish Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4.4
3.1 Install Rating Plate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4.4
3.2 Return Installation Card . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4.4

Section 4
Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4.5
4.1 B7530SB CT/i Smart Beam Software (w/o Collimator) Kit Parts List 4.5
4.2 B7530SC CT/i Smart Beam Collimator Option Kit Parts List . . . . . . . 4.5
4.3 B7530SS CT/i Smart Beam Software (w/Collimator) Kit Parts List . . . 4.5

4.i
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100

List Of Illustrations
Illustration Page

Illustration 4–1
Software Options Installation Window . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4.2
Illustration 4–2
Software Options Available / Installed Screen Example . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4.3
Illustration 4–3
Rating Plate Location . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4.4

4.ii
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100

SECTION 1
CT/I OPTIONAL SMART BEAM SOFTWARE INSTALLATION

Notice This Software Option Installation assumes that the system that this will be installed on
already has the correct collimator type required for this option to operate properly.

The correct collimator type is: 46–296300G4 or 46–296300G5. If this system does not
have either of these collimators installed, do not install this option software.

See the Material Lists in Section 4 of this chapter for Catalog Numbers for ordering the
Smart Beam collimator.

1.1 Check Installed Collimator Type


The Smart Beam Option requires that the correct collimator type be installed in the CT System’s
Gantry.

CAUTION Image Artifacts may result if the wrong collimator type is used with the Smart
Beam Option Software. Ensure that the correct collimator type is installed
before installing the Smart Beam Option Software.

(1) Remove the gantry right and left side covers.


(2) Disable the Gantry HVDC and Axial Drive using the switches on the Service switch
box in the gantry.
(3) Prop open the gantry top cover.
(4) Loosen the hand–wheel that retains the front cover support rod on the right side of the
gantry.
(5) Loosen/remove the hand–wheel that hold the front cover in place and swing the front
cover open.
(6) Rotate the gantry frame until the X–Ray tube is at the 3 o’clock position.
(7) Look for the Model/Serial Number tag on the front face of the collimator and verify the
model number is one of:
› 46–296300G4
or
› 46–296300G5
(8) If the collimator is either a G4 or a G5 collimator reverse steps (5) thru (1) and restore
the gantry to operation.
Note 4 If the collimator is not either 46–296300G4 or 46–296300G5 STOP this procedure and DO
NOT install the Smart Beam Option Software.
If the collimator is 46–296300G1, G2, or G3, you must first acquire, install, and correctly
align an 46–296300G4 or G5 collimator before installing this software option.
The optional Smart Beam Collimator is available via an Option Catalog Order (B7530SC).
See the Material Lists in Section 4 for further information.

4.1
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 8 DIRECTION 2181737–100

1.2 Add Smart Beam Software Option


Software Options must be loaded/re–loaded every time the OC software is reloaded.
1. Open the Service Desktop.
2. Select the Utilities Menu.

3. Select installOptions to display the options window shown in Illustration 4–1.


Illustation 4–1
SOFTWARE OPTIONS INSTALLATION WINDOW

4.2
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 8 DIRECTION 2181737–100

4. Insert the option MOD into the drive.


5. Select Install.

6. Select Permanent.

7. Select MOD.

8. Select OK to display a software options screen similar to the one shown in Illustration 4–2.
Illustation 4–2
SOFTWARE OPTIONS AVAILABLE/INSTALLED EXAMPLE SCREEN

9. After all the options are installed, select Quit, then OK.
10. Remove the MOD & write protect the side with options.
Note 5 The initial install requires the MOD to not be write protected; subsequent installs can be done
with the MOD write protected
11. If you have another option MOD to load, go back to step 1.

4.3
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100

SECTION 2
CALIBRATE SYSTEM

Because the Smart Beam Option makes use of different X–Ray Beam Shaping
filters to improve dose utilization, the system must have new calibration
vectors created for scanning.
(1) From the ExamRx Desktop, perform a complete set of Air and Phantom Calibrations
after the option software has been installed.
Note 6 Smart Beam Calibrations will take additional scans and perform additional processing in
order to accommodate the second X–Ray Beam profile.

SECTION 3
FINISH INSTALLATION

3.1 Install Rating Plate


1. Refer to .Illustration 4–3
2. Locate the current Modification Rating plate(s) on the rear of the CT/i console.
3. Fasten the Smart Beam Option rating plate to the left of the existing rating plates.
Illustration 4–3
RATING PLATE LOCATION

CT/I CONSOLE REAR VIEW

NEW RATING PLATE TO THE


LEFT OF EXISTING RATING
PLATES.

3.2 Return Installation Card


1. Complete and verify the site–specific information on the Red Installation Product
Locator Card.
2. Return the card to the Mailing Address that is printed on the front of the card.

4.4
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100

SECTION 4
PARTS

4.1 B7530SB CT/i Smart Beam Software (w/o Collimator) Option Kit Parts List

Item Part Number Name SPP Quantity Description (Remarks)


1 2148904 Option MOD 2 1 Smart Beam Option MOD.

2 46–302200P7 Serial Rating Plate N 1 Option Kit Rating Plate

4.2 B7530SC CT/i Smart Beam Collimator Option Kit Parts List
Item Part Number Name SPP Quantity Description (Remarks)
1 46–296300G4 Smart Beam Collimator N 1 Smart Beam ”G4” Collimator

2 2145015–100 Technique Chart N 1 Smart Beam Technique Charts

3 2143100–100 Install Procedure N 1 Smart Beam Collimator Kit Installation Procedure

4.3 B7530SS CT/i Smart Beam Software (with Collimator) Option Kit Parts List
Item Part Number Name SPP Quantity Description (Remarks)
1 B7530SB Smart Beam Software N 1 Smart Beam Software Option Kit

2 B7530SC Smart Beam Collimtor N 1 Smart Beam Collimator Option Kit

4.5
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100

NOTES

4.6
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100

Chapter 5
CT/i Language Options Installation

Table Of Contents

Section Description Page

Section 1
Language Option Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.1
1.1 Shutdown the System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.1
1.2 Boot System and Stop For Maintenance . . . . . . . . . . . . . . . . . . . . . . . . 5.1
1.3 If you have a keyboard other than a US–English Keyboard. . . . . . . . . . 5.1

Section 2
Procedures for installing add–on packages . . . . . . . . . . . . . . . . . . . . . . . . . 5.2
2.1 Optional Package Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.2
2.2 Reconfig to Select Language . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.3
2.3 Reconfig SBC to Enable Language . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.3

Section 3
Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.5
3.1 Material Lists for International Language Kits . . . . . . . . . . . . . . . . . . . 5.5

5.i
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100

NOTES

5.ii
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100

SECTION 1
LANGUAGE OPTION INSTALLATION

1.1 Shutdown the System


Click on the (square red) Shutdown button (shown below). Press OK to shutdown Applications.
This will take down both CT/i Applications and UNIX and put you at the boot prompt.

1.2 Boot System and Stop For Maintenance


(1) A message should be on the screen that says “ok to power off the system now”. The
next step requires two selections in quick succession (you may only have three seconds
between selections); please read carefully before proceeding:
You will first be selecting “Restart”, to bring up the system. A message may appear
about running power up diagnostics. Then a window will come up about starting up the
system. There will be a small button in the right hand corner: “Stop for Maintenance”.
You will use the mouse to quickly press this button.
You should now select Restart and Stop for Maintenance.
(2) A screen will pop up with several icons:
› Start System

› Install System Software

› Run Diagnostics

› Recover System

› Enter Command Monitor

› Select Keyboard Layout

1.3 If you have a keyboard other than a US–English Keyboard.


(1) Use the mouse and click on Select Keyboard Layout
(2) Select the appropriate keyboard type. Supported keyboards are:
› English – US (default)
› German – DE
› Swedish – SE
› French – FR
(3) Select Apply to reconfigure.
(4) Use the mouse to select Start System

(5) As the system comes up, a box will appear indicating the operator has 5 seconds to stop
application bring up. STOP applications start up when this box appears.

5.1
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100

SECTION 2
PROCEDURES FOR INSTALLING ADD–ON PACKAGES

2.1 Optional Package Installation


Note 7 Only the Italian Language Option requires additional Software to be loaded (using this
procedure). For English, French, and German, and languages, the language is selected
through Config/Reconfig. See Section 2.2.
(1) Open a UNIX shell, become root: su – <Enter>
(then enter the root passwd)
Note 8
(2) Type ’installpkg’
A software packages window appears:

(3) Insert the Software Packages Media into the CD–ROM drive.
(4) Click on the CD–ROM media selection.

5.2
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100

Example screen showing the packages available on sample Packages Media:

(5) Select the packages to be installed (e.g.: click & drag down to highlight all the available
options, then release).
(6) Click on the Install button.
A box appears while the software packages are loading.
(7) After all the packages are installed, select Quit, then OK.
(8) Leave the shell window open.

2.2 Reconfig to Select Language


(1) Before the options can be used, reconfig must be run to choose the correct language. In
the same shell window used for Option Package Installation use the following steps:
(2) Type cd /usr/g/scripts <enter>
(3) Type reconfig <enter> (The reconfig user interface window will open)
(4) Select config (with the mouse button)
(5) Select the Preferences tab with the mouse button.
(6) Select the preferred language in the lower left hand corner of the Preferences screen.
The selections available are English, French, German, and Italian (if the Italian option
has been installed).
(7) Select Accept at the top of the reconfig box.. This will take a few moments to
complete. There will be messages displayed on the screen.
(8) A prompt will appear on the screen for the operator to reboot; select Yes in the reboot
pop–up window.

2.3 Reconfig SBC to Enable Language


(1) After the system reboots, the SBC needs to be reconfigured.
(2) Bring up a UNIX Shell either from the service desktop or the toolbox.
(3) Type rsh sbc <enter>
(4) Log in as root; type su <enter>
(5) Start reconfig; type reconfig <enter>
(6) After the system finishes reconfig; type st <enter> to start applications.

5.3
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100

SECTION 3
PARTS

3.1 Material Lists for International Language Kits


Item Part Number Name SPP Quantity Description (Remarks)
French Language Kit B7500LF

1 2114561–27 Keyboard 1 1 French Keyboard


2 2139579 Interconnect 1 1 Keyboard Interconnect
3 2152628 Template 2 1 Keyboard Template
4 2142878–101 Manual N 1 French Technical Reference Manual
5 2168774–101 Manual N 1 French Technical Reference Mnaual

German Language Kit B7500LG

1 2114561–30 Keyboard 1 1 German Keyboard


2 2139579 Interconnect 1 1 Keyboard Interconnect
3 2152627 Template 2 1 Keyboard Template
4 2142878–108 Manual N 1 German Technical Reference Manual
5 2168774–108 Manual N 1 German Quick Reference Guide

Italian Language Kit B7500LH

1 2114561–2 Keyboard 1 1 English Keyboard


2 2139579 Interconnect 1 1 Keyboard Interconnect
3 2152629 Template 2 1 Keyboard Template
4 2188654 Language MOD 2 1 International Language software MOD
5 2142878–111 Manual N 1 Italian Technical Reference Manual
6 2168774–111 Manual N 1 Italian Quick Reference Guide

Scandinavian Language Kit B7500LS

1 2114561–33 Keyboard 1 1 Scandinavian Keyboard


2 2139579 Interconnect 1 1 Keyboard Interconnect
3 2144029 Template 2 1 Keyboard Template

5.4
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100

Chapter 6
CT/i Dentascan Option Installation

Table Of Contents
Section Description Page

Section 1
Install Dentascan Option Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.1
1.1 Add Software Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.1

Section 2
Procedures for installing Dentascan filming option . . . . . . . . . . . . . . . . . . 6.3

Section 3
Finish Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.3
3.1 Install Rating Plate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.3
3.2 Return Installation Card . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.4

Section 4
Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.4
4.1 B7540N CT/i Dentascan Option Kit Parts List . . . . . . . . . . . . . . . . . . . 6.4

6.i
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100

NOTES

List Of Illustrations
Illustration Page
Illustration 6–1
Software Options Installation Window . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.1
Illustration 6–2
Software Options Available / Installed Screen Example . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.2
Illustration 6–3
Rating Plate Location . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.3

6.ii
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 8 DIRECTION 2181737–100

SECTION 1
INSTALL DENTASCAN OPTION SOFTWARE

1.1 Add Software Options


Software Options must be loaded/re–loaded every time the OC software is reloaded.
1. Open the Service Desktop.
2. Select the Utilities Menu.

3. Select installOptions to display the options window shown in Illustration 6–1.


Illustation 6–1
SOFTWARE OPTIONS INSTALL WINDOW

6.1
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 8 DIRECTION 2181737–100

4. Insert the option MOD into the drive.


5. Select Install.

6. Select Permanent.

7. Select MOD.

8. Select OK to display a software options screen similar to the one shown in Illustration 6–2.
Illustation 6–2
SOFTWARE OPTIONS AVAILABLE/INSTALLED SAMPLE SCREEN

9. After all the options are installed, select Quit, then OK.
10. Remove the MOD & write protect the side with options. (Note: The initial install
requires the MOD to not be write protected; subsequent installs can be done with the
MOD write protected)
11. If you have another option MOD to load, go back to step 1.

Notice Once you have installed Dentascan, you need to follow the procedures in, Section 2, for
installing Dentascan filming option, before proceeeding to step 12..

12. Go to the Service Desktop –>Utilities –>Applications Shutdown –>Start


13. After Applications have shut down, type st <enter> to start up applications.
14. When the system applications return, the system is ready to use.

6.2
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100

SECTION 2
PROCEDURES FOR INSTALLING DENTASCAN FILMING OPTION

(1) Open UNIX shell from the toolchest.


(2) Type cd ~ctuser/install <Enter>
(3) Type installcamera.denta <Enter>
Message appears: Sending a film to the camera
...
Enter the distance Dist 1 in (mm):

Notice After the film is developed, use a metric scale to measure the distance between the vertical
lines for “Dist 1” and “Dist 2”. Enter the value(s) in MILLIMETERS (mm).

(4) Enter the measurement of Dist 1: XXX.X <Enter>


(5) If Zoom is supported by the type of Camera, the message will appear:
Enter the distance Dist 2 in (mm):
Enter the measurement of Dist 2: YYY.Y <Enter>
(6) Message appears: DentaScan filming option installed.
(7) Type ‘exit’ in the UNIX shell.
(8) Goto Section 1 step 9 for continuing installation.

SECTION 3
FINISH INSTALLATION

3.1 Install Rating Plate


1. Refer to Illustration 6–3
2. Locate the current Modification Rating plate(s) on the rear of the CT/i console.
3. Fasten the Dentascan Option rating plate to the left of the existing rating plates.
Illustration 6–3
RATING PLATE LOCATION

CT/I CONSOLE REAR VIEW

NEW RATING PLATE TO THE


LEFT OF EXISTING RATING
PLATES.

6.3
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100

3.2 Return Installation Card


1. Complete and verify the site–specific information on the Red Installation Product
Locator Card.
2. Return the card to the Mailing Address that is printed on the front of the card.

SECTION 4
PARTS

4.1 B7540N CT/i Dentascan Option Kit Parts List


Item Part Number Name SPP Quantity Description (Remarks)
1 2184748 CBT CD–ROM N 1 Navigator On Line Operators Manual
2 2184750 Options MOD N 1 Navigator Option Software

6.4
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100

Chapter 7
CT/i Smart Helical Option Installation

Table Of Contents
Section Description Page

Section 1
Install CT/i 4.X System Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7.1
1.1 Install 4.X Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7.1

Section 2
PROCEDURES FOR INSTALLING SMART HELICAL SOFTWARE . 7.2

Section 3
Finish Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7.4
3.1 Install Rating Plate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7.4
3.2 Return Installation Card . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7.4

Section 4
Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7.5
4.1 B7530TL Smart Helical Option Kit Parts List (systems not at 4.X) . . . 7.5
4.2 B7530TK Smart Helical Option Kit Parts List (systems at 4.X) . . . . . . 7.5

7.i
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100

List Of Illustrations
Illustration Page

Illustration 7–1
Software Options Installation Window . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7.2
Illustration 7–2
Software Options Available / Installed Screen Example . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7.3
Illustration 7–3
Rating Plate Location . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7.4

NOTES

7.ii
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100

SECTION 1
INSTALL CT/I 4.X SYSTEM SOFTWARE

1.1 Install 4.X Software

Notice The prerequisite to loading this Smart Helical Option software is that
you must first have Version 4.0 or higher System Software loaded on
your scanner. Do not attempt to load/enable this option without first
loading the latest version of 4.X Operating Software (included in this
Option Kit).

1. Refer to the Load From Cold Manual, 2168070–100, included with this Option Kit, for
instructions for loading the System Software included with this kit.
2. If 4.X software is already loaded on your system, go to Section 2 to install the Option
Software.

7.1
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 8 DIRECTION 2181737–100

SECTION 2
PROCEDURES FOR INSTALLING SMART HELICAL OPTION SOFTWARE

Software Options must be loaded/re–loaded every time the OC software is reloaded.


1. Open the Service Desktop.
2. Select the Utilities Menu.

3. Select installOptions to display the options window shown in Illustration 7–1.


Illustation 7–1
SOFTWARE OPTIONS INSTALLATION WINDOWS

4. Insert the option MOD into the drive.


5. Select Install.

6. Select Permanent.

7. Select MOD.

8. Select OK to display a software options screen similar to the one shown in Illustration 7–2.

7.2
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 8 DIRECTION 2181737–100

Illustation 7–2
SOFTWARE OPTIONS AVAILABLE/INSTALLED SAMPLE SCREEN

9. After all the options are installed, select Quit, then OK.
10. Remove the MOD & write protect the side with options.
Note 9 The initial install requires the MOD to not be write protected; subsequent installs can be done
with the MOD write protected
11. If you have another option MOD to load, go back to step 1.
12. Go to the Service Desktop –>Utilities –>Applications Shutdown –>Start
13. After Applications have shut down, type st <enter> to start up applications.
14. When the system applications return, the system is ready to use.

7.3
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100

SECTION 3
FINISH INSTALLATION

3.1 Install Rating Plate


1. Refer to Illustration 7–3
2. Locate the current Modification Rating plate(s) on the rear of the CT/i console.
3. Fasten the Smart Helical Option rating plate to the left of the existing rating plates.
Illustration 7–3
RATING PLATE LOCATION

CT/I CONSOLE REAR VIEW

NEW RATING PLATE TO THE


LEFT OF EXISTING RATING
PLATES.

3.2 Return Installation Card


1. Complete and verify the site–specific information on the Red Installation Product
Locator Card.
2. Return the card to the Mailing Address that is printed on the front of the card.

7.4
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100

SECTION 4
PARTS

4.1 B7530TL Smart Helical Option Kit Parts List (systems not at 4.X)
Item Part Number Name SPP Quantity Description (Remarks)
1 2180724 Options MOD N 1 Smart Helical Option
2 2181969 HiSpeed CT/I 1 1 IRIX Operating System Software
IRIX 6.2 CDROM
3 2200557 HiSpeed CT/I 4.0 1 1 4.0 Applications Software
Applications CDROM
4 2154721 Class A Service/ N 1 Operators & Service Manual Collector
Operators Manuals

4.2 B7530TK Smart Helical Option Kit Parts List (systems at 4.X)
Item Part Number Name SPP Quantity Description (Remarks)
1 2180724 Options MOD N 1 Smart Helical Option

7.5
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100

NOTES

7.6
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 5 DIRECTION 2181737–100

Chapter 8
CT/i BMD Option Installation
Table Of Contents
Section Description Page

Section 1
        
   . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8.1
1.1 " ! !# . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8.1

Section 2
PROCEDURES FOR INSTALLING BMD OPTION SOFTWARE . . . . 8.1
2.1 Install BMD Option Key and Load BMD Option Software . . . . . . . . . . 8.1

Section 3
Special Instructions for Transferring the BMD Database from an Advantage
Windows Workstation to the CT/i . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8.2
3.1 Transfer the BMD Database from AWW to Ct/i . . . . . . . . . . . . . . . . . . 8.2
3.2 Verify that the data has been successfully transferred . . . . . . . . . . . . . . 8.3

Section 4
Backup Database and Deliver Books & Phantoms to Customer . . . . . . . . 8.3
4.1 Backup BMD Database (Save To Mod) . . . . . . . . . . . . . . . . . . . . . . . . . 8.3
4.2 Deliver Books & BMD Phantom to Customer . . . . . . . . . . . . . . . . . . . . 8.3

Section 5
Finish Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8.4
5.1 Install Rating Plate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8.4
5.2 Return Installation Card . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8.4

Section 6
Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8.5
6.1



  ! ! !  ! # ! . 8.5
6.2
 
  ! ! ! !  ! #! ! 8.5

8.i
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 5 DIRECTION 2181737–100

List Of Illustrations
Illustration Page

Illustration 8–1
Rating Plate Location . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8.4

NOTES

8.ii
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 5 DIRECTION 2181737–100

SECTION 1
SYSTEM SOFTWARE PREREQUISITE

1.1 Prerequisite Software

Notice The prerequisite to loading this BMD Option software is that you must
first have Version 4.1, or Version 5.3 (or higher) System Software loaded
on your scanner. Do not attempt to load/enable this option without first
loading the latest version of 4.X or 5.X Operating Software.

1. If 4.1 or 5.3 (or higher) software is already loaded on your system, go to Section 2 to
install the Option Software.
2. If 4.1 or 5.3 (or higher) software is NOT loaded on your system, you must first perform
FMI 25245 (4.1 Software) or FMI 25249 (5.3 Software) before attempting to load this
option software

SECTION 2
PROCEDURES FOR INSTALLING BMD OPTION SOFTWARE

2.1 Install BMD Option Key and Load BMD Option Software
1. Please refer to Chapter 3 of the QCT Bone Densitometry User’s Guide for
instructions.
2. You will first load the BMD Option Key MOD, to enable loading of the BMD
software, and then you will load the BMD software from the QCT–5000 Bone
Densitometry CDROM.
3. Be sure to follow all of the instructions carefully, to ensure that the Option Software is
properly loaded and set up.

Notice The ”eject” softkey that is displayed on the screen may not work to eject the
CDROM when you have completed loading the BMD Option Software. If
you find this to be the case, perform the following steps to eject the CDROM:

a. In a UNIX shell window, type : cd / [Enter]


unmount CD [Enter]
b. Then press the eject pushbutton on the front of the CDROM Drive.
4. Reboot the system after the BMD Software is loaded, to allow the software to properly
complete the installation of BMD.

8.1
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 5 DIRECTION 2181737–100

SECTION 3
SPECIAL INSTRUCTIONS FOR TRANSFERRING THE BMD DATABASE FROM AN
ADVANTAGE WINDOWS WORKSTATION TO THE CT/i
Notice Use this procedure only if the customer has been using an Advantage Windows Workstation
to perform BMD Calculations on CT Images, and wants to transfer the BMD Database to
the CT/i System for all future BMD Calculations.
If the BMD Database is not transferred to the CT/i prior to performing BMD Calculations, it
will be impossible later to merge the AWW BMD Database with the CT/i BMD Database.

3.1 Transfer the BMD Database from AWW to CT/i


1. Determine the IP Address of the Advantage Windows Workstation, and verify that it is
networked to the CT/i Scanner.
c. At the AWW, open the Console Window, and type:
% more /etc/hosts <enter>
d. Look for a line similar to the following:
3.7.51.2 loghost <–record the number displayed for use in the next steps.
Note 10 You must also determine the username login and password for the AWW.
2. At the CT/i console, open a Shell Tool, and type the following:
{ctuser}[1] cd /export/home/sdc/bmd/database <enter>
3. Type:
{ctuser}[2] ftp (IP Address from step 1.) <enter>

Example: {ctuser}[2] ftp 3.7.51.2 <enter>


4. You will be prompted to enter a username and password (that you determined in step 1.)
in order to log in to the remote AWW .

Example: {ctuser}[2] ftp 3.7.51.2 <enter>


Connected to 3.7.51.2
220 awexpc0 FTP server (UNIX(r) System V Release 4.0) ready.
Name (3.7.51.2:ctuser): sdc <enter>
331 Password required for sdc.
Password: adw2.0 <enter> <– Note: The password will not display.
230 User sdc logged in.
ftp>
5. Type the following commands to set up and transfer the BMD Database files from the
AWW to the CT/i:
a. ftp> binary <enter>
b. ftp> cd /export/home/sdc/bmd/database <enter>
c. ftp> mget *.*
d. You will be prompted with each filename. Answer Y <enter>
There should be a total of seven (7) files transferred.
e. ftp> quit <enter>
6. In the CT/i Shell Tool window, type the following command to ensure that all of the
database files were transferred:
{ctuser}[3] ls <enter>
The following seven (7) filenames should be listed:
bdblok.dat exam.dat exam.idx pat.dat pat.idx
vert.dat vert.idx
7. Close the Shell Tool.

8.2
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 5 DIRECTION 2181737–100

3.2 Verify that the data has been successfully transferred


1. At the CT/i Console, switch to the Image Works screen.
In the upper right of the Browser, you should see a new button labelled BMD.
2. Select the BMD function. A BMD toolbar GUI should pop up.
3. On the BMD popup, select Browse BMD Database.

Ask the Operators to review the displayed BMD Database to ensure that all of the BMD
Studies appear. Or else, go back to the AWW, and from the Browser, perform these
same steps, and compare the two BMD Database displays to ensure that all of the BMD
Studies were transferred.

SECTION 4
BACKUP DATABASE AND DELIVER BOOKS & PHANTOMS TO CUSTOMER

4.1 Backup BMD Database (Save To Mod)


1. Follow the instructions contained in the QCT Bone Densitometry User’s Guide,
Chapter 8, to save the BMD Database to MOD.
Note 11 You must use a blank MOD Cartridge (or one that has previously been initialized for saving
System State information) to save the BMD Database. You cannot use a cartridge that has
been initialized for Image Storage.
2. Give this BMD Database MOD to the Operators, and instruct them to perform this
Database Backup procedure at least once a week.
3. Remember anytime you perform a Load From Cold on this system, to separately
Backup the BMD Database before loading software, because the BMD Database IS
NOT saved on the System State MOD.

4.2 Deliver Books & BMD Phantom to Customer


1. Deliver the BMD Phantom (if your kit contains one – see section 6 – Kit Parts Lists),
and the BMD Users Guide to the Operators.

8.3
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 5 DIRECTION 2181737–100

SECTION 5
FINISH INSTALLATION

5.1 Install Rating Plate


1. Refer to Illustration 8–1
2. Locate the current Modification Rating plate(s) on the rear of the CT/i console.
3. Fasten the BMD Option rating plate to the left of the existing rating plates.
Illustration 8–1
RATING PLATE LOCATION

CT/I CONSOLE REAR VIEW

NEW RATING PLATE TO THE


LEFT OF EXISTING RATING
PLATES.

5.2 Return Installation Card


1. Complete and verify the site–specific information on the Red Installation Product
Locator Card.
2. Return the card to the Mailing Address that is printed on the front of the card.

8.4
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 5 DIRECTION 2181737–100

SECTION 6
PARTS

6.1 B7710MB CT/i BMD Basic Option Kit Parts List (w/o Phantom)
Item Part Number Name SPP Quantity Description (Remarks)
1 2206382 Options MOD N 1 BMD Option Key MOD
2 2222885 BMD Software Package 2 1 BMD Software Package (w/o Phantoms)

6.2 B7710MC CT/i BMD + Phantom Option Kit Parts List (with Phantom)
Item Part Number Name SPP Quantity Description (Remarks)
1 2206382 Options MOD N 1 BMD Option Key MOD
2 2222885–2 BMD Option Package N 1 BMD Software Package (with Phantom)

8.5
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 5 DIRECTION 2181737–100

NOTES

8.6
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100

Chapter 9
CT/i (OCTANE) Memory Option
Table Of Contents
Section Description Page

Section 1
CT/i (OCTANE) CPU Memory Option Installation Preparation . . . . . . . 9.1
1.1 Shutdown the System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9.1
1.2 Preparing the Workstation to Install the optional memory . . . . . . . . . . . 9.1
1.3 Removing the System Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9.3

Section 2
Installing and Removing Memory . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9.6
2.1 About Memory . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9.6
2.2 Installing Memory . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9.8
2.3 Replacing the System Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9.8
2.4 Powering On the OCTANE Workstation . . . . . . . . . . . . . . . . . . . . . . . . 9.11

Section 3
Startup System and Verify Installed Memory . . . . . . . . . . . . . . . . . . . . . . . 9.12
3.1 Start System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9.12
3.2 Verify Installed Memory . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9.12

Section 4
Finish Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9.13
4.1 Install Rating Plate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9.13
4.2 Return Installation Card . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9.14

Section 5
Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9.15
5.1 Field Replaceable Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9.15
5.2 B7710M Memory Option Kit – Parts List . . . . . . . . . . . . . . . . . . . . . . . 9.15

9.i
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100

List Of Illustrations
Illustration Page

Illustration 9–1
LOck Bar Removal (If Applicable) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9.2
Illustration 9–2
Removing the System Module Screws . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9.3
Illustration 9–3
Locating the Compression Connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9.3
Illustration 9–4
Releasing the System Module form the Front Plane . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9.4
Illustration 9–5
Identifying DIMM Sockets and DIMM Banks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9.7
Illustration 9–6
LOcating DIMMs on the System Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9.7
Illustration 9–7
Locating the Notches on a DIMM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9.8
Illustration 9–8
Removing the Compression Connector Cap . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9.8
Illustration 9–9
Supporting the System Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9.9
Illustration 9–10
Replacing the System Module in the OCTANE Workstation . . . . . . . . . . . . . . . . . . . . . . . . . . 9.10
Illustration 9–11
Locking the System Module to the Frontplane . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9.10
Illustration 9–12
Replacing the System Module Screws . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9.11
Illustration 9–13
Rating Plate Location . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9.13

9.ii
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100

SECTION 1
CT/I (OCTANE) CPU MEMORY OPTION INSTALLATION PREPARATION

The CT/i CPU Memory Option, B7510T, provides an additional 64 Mb of CPU memory for the
Silicon Graphics Computer. This memory allows more efficient execution of system software
by reducing the amount of disk swap activity and allowing larger amounts of data to be available
in memory at any given moment.

1.1 Shutdown the System


1. Select the (square red) Shutdown Button (shown below) on the Image Display Monitor.
Select OK to shutdown Applications. This will stop both CT/i Applications and UNIX
software and bring you to the boot prompt.

2. After the software shuts down, the following message will appear.

i Okay to power off the system now.


Press any key to restart
Restart

3. Turn off the power switch on the console.

1.2 Preparing the Workstation to Install the optional memory

Notice THE CIRCUIT BOARDS ARE VERY STATIC–SENSITIVE!!


Static–sensitive components may be damaged if not handled in a static–free environment.
Static Sensitive Devices Take appropriate precautions (e.g., wear properly grounded wrist strap) when handling all
boards.

T prepare the workstation for installation of optional memory, follow these steps:
1. Open the cover and push the power button to power off the OCTANE workstation.
2. Unplug the power cord from the electrical outlet and from the OCTANE workstation.
3. Press the power switch on the monitor.
4. Wait 5 minutes before removing the system module.

9.1
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100

CAUTION The heat sinks on the CPU and system module become very hot. Wait
5minutes after powering off the OCTANE workstation before you remove the
system module. Test before touching the CPU or heat sinks.

Lock

Lock Bar

Illustration 9–1
LOCK BAR REMOVAL (IF APPLICABLE)

5. If applicable, remove the lockbar which locks the system module and bezel to the
chassis.
› Unlock the lock at the back of the OCTANE workstation.
› Pull the lockbar out of the OCTANE workstation.
6. Face the rear of the workstation.
7. Remove the cables attached to the system module.

9.2
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100

1.3 Removing the System Module

Notice THE CIRCUIT BOARDS ARE VERY STATIC–SENSITIVE!!


Static–sensitive components may be damaged if not handled in a static–free environment.
Take appropriate precautions (e.g., wear properly grounded wrist strap) when
Statichandling allDevices
Sensitive
boards.

Captive Screws (2)

Illustration 9–2
REMOVING THE SYSTEM MODULE SCREWS

CAUTION The heat sinks on the CPU and system module become very hot. Wait
5minutes after powering off the OCTANE workstation before you remove the
system module. Test before touching the CPU or heat sinks.

1. To remove the system module, you need to power off the system, wait 5 minutes after
powering off the workstation to allow the heat sinks to cool, and attach the wrist strap.
2. Loosen the captive screws that hold the sliding handles to the OCTANE workstation.

Compression Connectors

Bristled pad

Illustration 9–3
LOCATING THE COMPRESSION CONNECTORS

9.3
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100

Look for the two compression connectors when you take the system module out of the
chassis, and do not touch them. They are on the back of the system module, and
connect to the chassis when the system module is installed in the workstation.

Notice The gold bristled pad on the compression connectors on the back of the system module is
very delicate and easily damaged. Bristled pad

3. Pull both sliding handles simultaneously until they are completely extended. This action
releases the system module from the workstation.

Illustration 9–4
RELEASING THE SYSTEM MODULE FORM THE FRONT PLANE

4. Pull the system module from the chassis by grasping the module handle with your left
hand and bracing your right hand against the top of the workstation.
Support the bottom of the module as you pull it out.
5. Place it on a dry, antistatic surface with the CPU and DIMMs facing up. Your desktop
works well.

Notice Do not touch the compression connectors as you remove the system module. The side of the
compression connector that connects to the workstation is extremely delicate.

9.4
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100

6. Place a cap on each compression connector on the back of the system module.
(Compression connector caps come with the workstation.)

CAUTION Warning: The heat sinks on the CPU and system module become very hot.
Wait 5 minutes after powering off the OCTANE workstation before you remove
the system module. Test before touching the CPU or heat sinks.

9.5
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100

SECTION 2
INSTALLING AND REMOVING MEMORY

To install or remove memory, you must power off the workstation, wait 5 minutes after powering off the workstation to allow the heat
sinks to cool, attach the wrist strap, and remove the system module.

Notice DIMMS ARE VERY STATIC–SENSITIVE!!


Static–sensitive components may be damaged if not handled in a static–free environment.
Static Sensitive Devices Take appropriate precautions (e.g., wear properly grounded wrist strap) when handling all
boards. Dual in line memory modules (DIMMs) are extremely sensitive to static electricity.
Handle the DIMMs carefully and wear the wrist strap to avoid the flow of static electricity.

2.1 About Memory


Before you install DIMMs in your OCTANE workstation, read the following information.
» Bank 1 (sockets 1 and 2) must always be filled.
› Minimum memory configuration is 64 MB in a bank. (2 x 32 MB DIMM)
› Maximum configuration is 512 MB in a bank. (2 x 256 MB DIMM)
» Banks are filled sequentially; when Bank 1 is full, fill Bank 2. Do not skip banks.
» Each bank is either empty, or contains two DIMMs, one in each of the two sockets.
» Each banks of DIMMS must contain two DIMMs of the same capacity or density and of
the same type.
Capacity refers to the number of megabytes in a DIMM: 32, 64, 128, and so on.
Type refers to the construction of the DIMM. Because DIMM construction cannot be
determined visually, it is important to know that two DIMMs are the same type before
you begin installation. If you insert two DIMMs of different types, the system does not
work. Banked DIMMs and stacked DIMMs are two examples of many different types of
memory.
» Minimum memory capacity in the OCTANE workstation is 64 MB. (2 x 32 MB
DIMMs)
» Maximum memory capacity in the OCTANE workstation is 2 GB. (8 x 256 MB
DIMMs) Installing Memory

9.6
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100

Processor

Banks

S1 1

System module S2

S3 2
DIMM S4
sockets
S5 3

S6

S7 4

S8

Illustration 9–5
IDENTIFYING DIMM SOCKETS AND DIMM BANKS

DIMMs

Bank 1
Bank 2
Bank 3
Bank 4
Illustration 9–6
LOCATING DIMMS ON THE SYSTEM MODULE

9.7
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100

2.2 Installing Memory

Notches

Illustration 9–7
LOCATING THE NOTCHES ON A DIMM

DIMMs are notched on the bottom so that they cannot be inserted incorrectly.
1. Insert the DIMM into the socket, gently but firmly. You hear a click as it is seated, and
the latch on the end of the socket moves up.
2. Check to be sure both sockets in the bank are full. DIMMs must be installed in pairs.
You have finished installing memory and are ready to replace the system module.

2.3 Replacing the System Module

Caution – do not touch

Illustration 9–8
REMOVING THE COMPRESSION CONNECTOR CAP
1. Remove the caps on the compression connectors on the back of the system module. Do
not touch the system module compression connectors as you lift the module into the
workstation.

9.8
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100

Notice Do not insert the system module into the workstation unless the power cord is unplugged
from the electrical socket. Inserting the system module into an OCTANE workstation with
the power cord connected to an electrical socket causes damage to the OCTANE
workstation.

Illustration 9–9
SUPPORTING THE SYSTEM MODULE

2. Grasp the system module by its immovable handle and support it with one hand as you
slide it into the chassis.
3. Push the module completely into the chassis using the immovable handle. (Both sliding
handles protrude.) The system module stops when it is slightly out of the chassis.

9.9
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100

Illustration 9–10
REPLACING THE SYSTEM MODULE IN THE OCTANE WORKSTATION

4. Push the sliding handles simultaneously to connect the system module to the
workstation.

Illustration 9–11
LOCKING THE SYSTEM MODULE TO THE FRONTPLANE

9.10
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100

5. Push firmly on the handles to completely lock the system module to the workstation.
The handles are completely recessed and the system module is flush with the chassis
when it is fully seated and locked to the workstation.

Illustration 9–12
REPLACING THE SYSTEM MODULE SCREWS

6. Tighten the captive screws in the sliding handles until the system module is attached to
the chassis.
7. Remove the wrist strap.

2.4 Powering On the OCTANE Workstation


1. Plug in the power cord.
2. Push the power button on the workstation.
3. Replace the lockbar.
› Open the cover on the front of the OCTANE workstation.
› Slide the lockbar into the lockbar slot.
› Insert the lock through the lockbar at the back of the workstation.

9.11
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100

SECTION 3
STARTUP SYSTEM AND VERIFY INSTALLED MEMORY

3.1 Start System


1. Turn on the console power switch.
2. The following message will appear. Select the Stop for Maintenance button.

Starting up the system...


Stop for Maintenance

3.2 Verify Installed Memory


The system Maintenance menu appears.
1. Select “Enter Command Monitor”.
2. At the shell prompt type:
>> hinv
3. Verify that the Main memory size: is 256Mb.
Note 12 If the amount of memory is incorrect, follow the directions at the beginning of this chapter to
power off your system and check the installation procedure.
» Check the angle of the DIMMs. They should be upright and completely inserted.
» Check that each bank is populated with two DIMMs, and that they are of the same type.
You must have an even number of DIMMs installed.
4. Exit the shell:
>> exit
5. Select “Start System”

9.12
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100

The system software will start normally.


6. After the system is started, open a shell tool and re–check the hardware environment:
{ctuser@somesystem} [1] hinv
1 195 MHZ IP30 Processor
CPU: MIPS R10000 Processor Chip Revision: 2.7
FPU: MIPS R10010 Floating Point Chip Revision: 0.0
Main Memory size: 256 Mbytes
Instruction cache size: 32 Kbytes
Data cache size: 32 Kbytes
Secondary unified instruction/data cache size: 1 Mbytes
Intergal SCSI controller 0: Version QL1040B (rev. 2)
Disk drive: unit 1 on SCSI controller 0
Integral SCSI controller 1: Version QL1040B (rev. 2)
Optical disk: unit 4 on SCSI controller 1
Comm device: unit 4 on SCSI controller 1
Comm device: unit 4, lin 1 on SCSI controller 1
CDROM: unit 6 on SCSI controller 1
IOC3 serial port: tty1
IOC3 serial port: tty2
IOC3 parallel port: plp1
Graphics board: SI with texture option
Graphics board: SI
Integral Fast Ethernet: ef0, version 1
Integral Ethernet: et0, IO0
Iris Audio Processor: version RAD revision 12.0, number 1
PCI card, bus 0, slot 221, Vendor 0x0, Device 0x0

SECTION 4
FINISH INSTALLATION

4.1 Install Rating Plate


1. Locate the Modification Rating plate, to the left of the Rating Plates already on the rear
of the cabinet. Refer to Illustration 1–5.

CT/I CONSOLE REAR VIEW

NEW RATING PLATE TO THE


LEFT OF EXISTING RATING
PLATES.

Illustration 9–13
RATING PLATE LOCATION

9.13
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100

4.2 Return Installation Card


1. Complete/Verify the site specific information on the Red Installation Card and return to
the appropriate Location printed on the front of the card.

9.14
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100

SECTION 5
PARTS

5.1 Field Replaceable Parts


Item Part Number Name SPP Quantity Description (Remarks)
1 2199806 64 mb DIMM 1 1 64 mb SDRAM DIMMs (2x32 mb)

2 2199806–2 128 mb DIMM 1 1 128 mb SDRAM DIMMs (2x64 mb)

5.2 B7710M Memory Option Kit – Parts List


Item Part Number Name SPP Quantity Description (Remarks)
1 B7710M Memory Collector N 1

9.15
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100

9.16
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100

Chapter 10
CT/i (OCTANE) Dual Processor Option
Table Of Contents
Section Description Page

Section 1
CT/i OCTANE Dual Processor Option Installation Preparation . . . . . . . 10.1
1.1 Shutdown the System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10.1
1.2 Removing the System Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10.2

Section 2
Installing The Optional Dual Processor . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10.6
2.1 Removing The Single Processor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10.6
2.2 Installing The Optional Dual Processor . . . . . . . . . . . . . . . . . . . . . . . . . 10.7
2.3 Replacing the System Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10.9
2.4 Powering On the OCTANE Workstation . . . . . . . . . . . . . . . . . . . . . . . . 10.12

Section 3
Startup System and Verify Installed Dual Processor . . . . . . . . . . . . . . . . . 10.13
3.1 Start System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10.13
3.2 Verify Installed Dual Processor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10.13

Section 4
Finish Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10.14
4.1 Install Rating Plate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10.14
4.2 Return Installation Card . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10.15

Section 5
Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10.16
5.1 Field Replaceable Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10.16
5.2 B7710JP Dual Processor Option Kit – Parts List . . . . . . . . . . . . . . . . . . 10.16

10.i
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100

List Of Illustrations
Illustration Page

Illustration 10–1
LOck Bar Removal (If Applicable) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10.2
Illustration 10–2
Removing the System Module Screws . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10.3
Illustration 10–3
Locating the Compression Connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10.3
Illustration 10–4
Placing the Compression Cap on the Compression Connector . . . . . . . . . . . . . . . . . . . . . . . . . 10.5
Illustration 10–5
Removing the Single Processor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10.6
Illustration 10–6
Viewing The Connectors on the Underside of the CPUs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10.7
Illustration 10–7
Installing the Dual Processor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10.8
Illustration 10–8
Removing the Compression Connector Cap . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10.9
Illustration 10–9
Supporting the System Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10.10
Illustration 10–10
Replacing the System Module in the OCTANE Workstation . . . . . . . . . . . . . . . . . . . . . . . . . . 10.11
Illustration 10–11
Locking the System Module to the Frontplane . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10.11
Illustration 10–12
Replacing the System Module Screws . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10.12
Illustration 10–13
Rating Plate Location . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10.14

10.ii
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100

SECTION 1
CT/I OCTANE DUAL PROCESSOR OPTION INSTALLATION PREPARATION

The CT/i Octane Dual Processor Option, B7710JP, replaces the single 195 Mhz CPU with a dual
195 Mhz CPU or the Octane Silicon Graphics Computer.

1.1 Shutdown the System


1. Select the (square red) Shutdown Button (shown below) on the Image Display Monitor.
Select OK to shutdown Applications. This will stop both CT/i Applications and UNIX
software and bring you to the boot prompt.

2. After the software shuts down, the following message will appear.

i Okay to power off the system now.


Press any key to restart
Restart

3. Turn off the power switch on the console.


4. Open the cover and push the power button to power off the OCTANE workstation.
5. Unplug the power cord from the electrical outlet and from the OCTANE workstation.
6. Press the power switch on the monitor.
7. Wait 5 minutes before removing the system module.

CAUTION The heat sinks on the CPU and system module become very hot. Wait
5minutes after powering off the OCTANE workstation before you remove the
system module. Test before touching the CPU or heat sinks.

10.1
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100

Lock

Lock Bar

Illustration 10–1
LOCK BAR REMOVAL (IF APPLICABLE)

8. If applicable, remove the lockbar which locks the system module and bezel to the
chassis.
› Unlock the lock at the back of the OCTANE workstation.
› Pull the lockbar out of the OCTANE workstation.
9. Face the rear of the workstation.
10. Remove the cables attached to the system module.

1.2 Removing the System Module

Notice THE CIRCUIT BOARDS ARE VERY STATIC–SENSITIVE!!


Static–sensitive components may be damaged if not handled in a static–free environment.
Static Sensitive Devices Take appropriate precautions (e.g., wear properly grounded wrist strap) when handling all
boards.

10.2
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100

Captive Screws (2)

Illustration 10–2
REMOVING THE SYSTEM MODULE SCREWS

CAUTION The heat sinks on the CPU and system module become very hot. Wait
5minutes after powering off the OCTANE workstation before you remove the
system module. Test before touching the CPU or heat sinks.

1. To remove the system module, you need to power off the system, wait 5 minutes after
powering off the workstation to allow the heat sinks to cool, and attach the wrist strap.
2. Loosen the captive screws that hold the sliding handles to the OCTANE workstation.

Compression Connectors

Bristled pad

Illustration 10–3
LOCATING THE COMPRESSION CONNECTORS
Look for the two compression connectors when you take the system module out of the
chassis, and do not touch them. They are on the back of the system module, and
connect to the chassis when the system module is installed in the workstation.

Notice The gold bristled pad on the compression connectors on the back of the system module is
very delicate and easily damaged. Bristled pad

3. Pull both sliding handles simultaneously until they are completely extended. This action
releases the system module from the workstation.

10.3
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100

4. Pull the system module from the chassis by grasping the module handle with your left
hand and bracing your right hand against the top of the workstation.
Support the bottom of the module as you pull it out.
5. Place it on a dry, antistatic surface with the CPU and DIMMs facing up. Your desktop
works well.

Notice Do not touch the compression connectors as you remove the system module. The side of the
compression connector that connects to the workstation is extremely delicate.

10.4
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100

Caution – do not touch

Illustration 10–4
PLACING THE COMPRESSION CAP ON THE COMPRESSION CONNECTOR

6. Place a cap on each compression connector on the back of the system module.
(Compression connector caps come with the workstation.)

CAUTION Warning: The heat sinks on the CPU and system module become very hot.
Wait 5 minutes after powering off the OCTANE workstation before you remove
the system module. Test before touching the CPU or heat sinks.

10.5
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100

SECTION 2
INSTALLING THE OPTIONAL DUAL PROCESSOR

2.1 Removing The Single Processor

Single processor

Illustration 10–5
REMOVING THE SINGLE PROCESSOR

1. Loosen the four captive Phillips screws holding the CPU in place. The dual processor
has six screws.

Notice Do not remove the four inner screws (not Phillips) holding the heat sink to the CPU.

2. Slide your fingers under the edge of the single processor closest to the back of the
module, and push up to release it. You may need to use two hands to lift it out.

10.6
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100

2.2 Installing The Optional Dual Processor


Connectors

Single Processor
Dual processor

Illustration 10–6
VIEWING THE CONNECTORS ON THE UNDERSIDE OF THE CPUS

1. Turn over the CPU to determine where the connectors are located. Align the connectors
on the base of the CPU with the connectors on the system board.
2. Place the dual processor on the side of the system module closest to the panel of
connectors.

10.7
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100

Dual Processor

Illustration 10–7
INSTALLING THE DUAL PROCESSOR

3. Lower the CPU onto the standoffs and connectors as shown


4. Tighten the six captive screws to the standoffs.

10.8
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100

2.3 Replacing the System Module

Caution – do not touch

Illustration 10–8
REMOVING THE COMPRESSION CONNECTOR CAP

1. Remove the caps on the compression connectors on the back of the system module. Do
not touch the system module compression connectors as you lift the module into the
workstation.

Notice Do not insert the system module into the workstation unless the power cord is unplugged
from the electrical socket. Inserting the system module into an OCTANE workstation with
the power cord connected to an electrical socket causes damage to the OCTANE
workstation.

10.9
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100

Illustration 10–9
SUPPORTING THE SYSTEM MODULE

2. Grasp the system module by its immovable handle and support it with one hand as you
slide it into the chassis.
3. Push the module completely into the chassis using the immovable handle. (Both sliding
handles protrude.) The system module stops when it is slightly out of the chassis.

10.10
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100

Illustration 10–10
REPLACING THE SYSTEM MODULE IN THE OCTANE WORKSTATION

4. Push the sliding handles simultaneously to connect the system module to the
workstation.

Illustration 10–11
LOCKING THE SYSTEM MODULE TO THE FRONTPLANE

10.11
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100

5. Push firmly on the handles to completely lock the system module to the workstation.
The handles are completely recessed and the system module is flush with the chassis
when it is fully seated and locked to the workstation.

Illustration 10–12
REPLACING THE SYSTEM MODULE SCREWS

6. Tighten the captive screws in the sliding handles until the system module is attached to
the chassis.
7. Remove the wrist strap.

2.4 Powering On the OCTANE Workstation


1. Plug in the power cord.
2. Push the power button on the workstation.
3. Replace the lockbar.
› Open the cover on the front of the OCTANE workstation.
› Slide the lockbar into the lockbar slot.
› Insert the lock through the lockbar at the back of the workstation.

10.12
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100

SECTION 3
STARTUP SYSTEM AND VERIFY INSTALLED DUAL PROCESSOR

3.1 Start System


1. Turn on the console power switch.
2. The following message will appear. Select the Stop for Maintenance button.

Starting up the system...


Stop for Maintenance

3.2 Verify Installed Dual Processor


The system Maintenance menu appears.
1. Select “Enter Command Monitor”.
2. At the shell prompt type:
>> hinv
3. Verify that the processor is: ”2 195 mhz IP30 Processors”
4. Exit the shell:
>> exit
5. Select “Start System”

10.13
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100

The system software will start normally.


6. After the system is started, open a shell tool and re–check the hardware environment:
{ctuser@somesystem} [1] hinv
2 195 MHZ IP30 Processor
CPU: MIPS R10000 Processor Chip Revision: 2.7
FPU: MIPS R10010 Floating Point Chip Revision: 0.0
Main Memory size: 256 Mbytes
Instruction cache size: 32 Kbytes
Data cache size: 32 Kbytes
Secondary unified instruction/data cache size: 1 Mbytes
Integral SCSI controller 0: Version QL1040B (rev. 2)
Disk drive: unit 1 on SCSI controller 0
Integral SCSI controller 1: Version QL1040B (rev. 2)
Optical disk: unit 4 on SCSI controller 1
Comm device: unit 4 on SCSI controller 1
Comm deveice: unit 4, lin 1 on SCSI controller 1
CDROM: unit 6 on SCSI controller 1
IOC3 serial port: tty1
IOC3 serial port: tty2
IOC3 parallel port: plp1
Graphics board: SI with texture option
Graphics board: SI
Integral Fast Ethernet: ef0, version 1
Integral Ethernet: et0, IO0
Iris Audio Processor: version RAD revision 12.0, number 1
PCI card, bus 0, slot 221, Vendor 0x0, Device 0x0

SECTION 4
FINISH INSTALLATION

4.1 Install Rating Plate


Locate the Modification Rating plate, left of the Rating Plates already on the rear of the cabinet.

CT/I CONSOLE REAR VIEW

NEW RATING PLATE TO THE


LEFT OF EXISTING RATING
PLATES.

Illustration 10–13
RATING PLATE LOCATION

10.14
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100

4.2 Return Installation Card


Complete/Verify the site specific information on the Red Installation Card and return to the
appropriate Location printed on the front of the card.

10.15
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100

SECTION 5
PARTS

5.1 Field Replaceable Parts


Item Part Number Name SPP Quantity Description (Remarks)
1 2169940–18 195 Mhz Processor 1 1 Option 195 Mhz Processor Module

2 2169940–20 Dual 195 Mhz Processor 1 1 Option Dual 195 Mhz Processor Module

5.2 B7710JP Dual Processor Option Kit – Parts List


Item Part Number Name SPP Quantity Description (Remarks)
1 B7710JP Dual Processor Option N

10.16
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100

Chapter 11
CT/i (OCTANE) Additional Magnetic Disk Option
Installation
Table Of Contents
Section Description Page

Section 1
Prepare for CT/i (Octane) Additional Magnetic Disk Option Installation 11.1
1.1 Prepare for Option Disk Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . 11.1
1.2 Save System State Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11.1
1.3 Shutdown the System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11.2

Section 2
Install Additional Magnetic Disk Option Hardware . . . . . . . . . . . . . . . . . 11.3
2.1 About Internal Drives . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11.3
2.2 Power Off Console and Install Optional Disk . . . . . . . . . . . . . . . . . . . . 11.3
2.3 Removing the Bezel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11.4
2.4 Installing an Internal Option Drive . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11.7
2.5 Removing the Blank Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11.8
2.6 Replacing the Bezel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11.10
2.7 Verify Hardware Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11.11

Section 3
Load CT/i Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11.11
3.1 Install Option Disk Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11.11

Section 4
Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11.11
4.1 B7710JD CT/i (Octane) Additional Sys Mag Disk Option Kit . . . . . . . 11.11

11.i
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100

List Of Illustrations
Illustration Page

Illustration 11–1
Locating the Lockbar on the Front of the OCTANE Workstation . . . . . . . . . . . . . . . . . . . . . . . 11.3
Illustration 11–2
Locating the Lockbar on the Octane Workstation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11.4
Illustration 11–3
Unlocking and Removing the Lockbar . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11.5
Illustration 11–4
Pressing the Bezel Release Buttons . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11.5
Illustration 11–5
Removing the Bezel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11.6
Illustration 11–6
Identifying the Option Drive Bays . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11.7
Illustration 11–7
Sliding In an Optional Drive . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11.7
Illustration 11–8
Locking the Drive . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11.8
Illustration 11–9
Opening the Door on the Bezel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11.8
Illustration 11–10
Pushing the Blank Panel Free . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11.9
Illustration 11–11
Replacing the Bezel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11.10
Illustration 11–12
Replacing The LockBar . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11.10

11.ii
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 7 DIRECTION 2181737–100

SECTION 1
PREPARE FOR CT/I (OCTANE) ADDITIONAL MAGNETIC DISK OPTION
INSTALLATION

This section describes the correct installation and configuration for B7710JD,
CT/i (Octane)Additional System Magnetic Disk Option.

1.1 Prepare for Option Disk Installation

WARNING THE INSTALLATION OF THE OPTIONAL MAGNETIC DISK REQUIRES A LOAD


FROM COLD OF THE SYSTEM SOFTWARE.

WARNING MAKE SURE THAT ALL IMAGES HAVE BEEN RECONSTRUCTED,


ARCHIVED, FILMED, AND NETWORKED BEFORE PERFORMING THIS
PROCEDURE. THIS PROCEDURE WILL RE–INITIALIZE ALL
SYSTEM DATA DISK, ERASING ALL IMAGES AND SCAN DATA.

1.2 Save System State Data


Before beginning the software install, you will want to, save the system configuration information, character-
ization, calibration, protocols, etc . . .
(1) Bring the system up if it is not already up.
(2) Select the Service desktop.
(3) Select PM.
(4) Highlight System State.
(5) Select the “Start ...” button.
(6) Select “All” (which will highlight Cals, Characterization, Reconfig Info, etc).
(7) Select Save. (The save will take a few minutes)
Note 1 Review the output for errors or missing files; the scroll bar on the right works only when the
tool isn’t busy performing some task –– it takes a little while.
If System State reported no problems, then you may skip to next section, else if you see any
missing files or failures, then refer to the note below.
Note 2 SAVE/RESTORE SYSTEM STATE ERRORS
Save/Restore System State reports status on each file specification listed in the
/usr/g/config/SysState.cfg file. The status may be:
› saved (or restored)
› does not exist, not saving (or restoring)

11.1
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 7 DIRECTION 2181737–100

› FAILED
A status of “does not exist” or “FAILED”, indicates potential problems. Please review the
status for each file to see if any errors are reported.

WARNING MAKE SURE THAT ALL IMAGES HAVE BEEN RECONSTRUCTED,


ARCHIVED, FILMED, AND NETWORKED BEFORE PERFORMING THIS
PROCEDURE. THIS PROCEDURE WILL RE–INITIALIZE ALL
SYSTEM DATA DISK, ERASING ALL IMAGES AND SCAN DATA.

(8) Select Dismiss.

1.3 Shutdown the System


To shut down system software:
(1) Select the (square red) Shutdown button (shown below).
(2) Press OK to shutdown Applications.

This will take down both CT/i Applications and UNIX and put you at the boot prompt.

11.2
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 7 DIRECTION 2181737–100

SECTION 2
INSTALL ADDITIONAL MAGNETIC DISK OPTION HARDWARE

2.1 About Internal Drives


» Unsupported devices may not work with the OCTANE workstation, and may cause
problems with supported products.
» All internal devices have a 3.5–inch form factor.
» Internal SCSI devices (drives) are on a separate bus and are automatically given an
address when they inserted. Each bus has a unique address: Bus 0 (internal devices) has
addresses such as 0–1, 0–2, 0–3; the system drive is SCSI ID 0–1, the option drive
above it is SCSI ID 0–2, and the unused option drive slot at the top of the Octane
workstation is SCSI ID 0–3. Bus 1 (external devices) has addresses such as 1–1, 1–2, 1-3.

Notice Drives can be easily damaged. Handle drive carefully; do not drop it or handle it roughly.

2.2 Power Off Console and Install Optional Disk


(1) After the software is stopped, shut off the console’s power switch
(2) Remove the CT/i (Octane) console front cover.
(3) Open the Octane computer front cover
Illustration 11–1
LOCATING THE LOCKBAR ON THE FRONT OF THE OCTANE WORKSTATION

B A Power button

Power cord

C Power switch

11.3
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 7 DIRECTION 2181737–100

(4) Press and release the power button to shut down the software and power off the system.
The system powers off automatically. See illustration 11–1 (A).
(5) Press the monitor power switch to turn off your monitor. (C)
(6) Disconnect the power cable from the back of the workstation. (B)

2.3 Removing the Bezel


Illustration 11–2
LOCATING THE LOCKBAR ON THE OCTANE WORKSTATION

Lockbar slot

(1) Locate the lockbar. If no lockbar is in place, skip to step 4.

11.4
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 7 DIRECTION 2181737–100

Illustration 11–3
UNLOCKING AND REMOVING THE LOCKBAR

(2) Unlock and remove the lockbar, if it is in place. (A)


(3) Close the door on the front of the OCTANE workstation. (B)
Illustration 11–4
PRESSING THE BEZEL RELEASE BUTTONS

(4) Press both bezel release buttons. Removing the Bezel as shown above.

11.5
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 7 DIRECTION 2181737–100

Illustration 11–5
REMOVING THE BEZEL

(5) Remove the bezel by tilting it forward, then lifting it from its release hooks at the
bottom.

11.6
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 7 DIRECTION 2181737–100

2.4 Installing an Internal Option Drive


Illustration 11–6
IDENTIFYING THE OPTION DRIVE BAYS

Option drive bays

System drive

Light module

The option drive bays are the two top drive bays. Install the option drive only in the middle
bay. The OCTANE workstation automatically assigns an address to the internal drives. The
system drive is SCSI ID 0–1, the drive directly above it is SCSI ID 0–2.

Illustration 11–7
SLIDING IN AN OPTIONAL DRIVE

Handle

Option drive bays

11.7
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 7 DIRECTION 2181737-100

(1) Lift the handle on the drive to a horizontal position, and slide the drive into the middle drive
bay. The drive handle is the same color as the drive.
(2) Slide the drive in gently until you feel solid resistance. The drive is connected to the
frontplane when the front of the drive is flush with the chassis.

Notice Pushing the drive into the frontplane with force can damage the drive.

Notice The option drive MUST be installed in the middle drive bay, or the application software
will not recognize it. The following illustration shows the drive in the upper slot to
clarify drive handle locking.
Illustration 11-8
LOCKING THE DRIVE

Drive handle

(3) With the drive completely inserted, rotate the handle downward to lock the drive in the
workstation.
You have completed installing an option drive and are ready to remove the blank panel on the
bezel.

2.5 Removing the Blank Panel


Illustration 11-9
OPENING THE DOOR ON THE BEZEL

Vent

Blanking panels

After installing a new option drive, you remove a blank panel from the bezel before replacing
the bezel.

11.8
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 7 DIRECTION 2181737–100

(1) Place the bezel on a flat surface so that the door is facing up.
(2) Open the door.
(3) Locate the blank panel for the drive bay in which you installed the option drive.
Illustration 11–10
PUSHING THE BLANK PANEL FREE

Vent

Blanking panels

(4) Press down on the blank panel. It pops out the back of the bezel.
(5) Keep the blank panel. You must replace it if you ever remove the drive.

Notice To prevent damage to your OCTANE workstation, you must place a blank panel in the bezel
if you have removed a drive. The blank panel ensures proper airflow through the
workstation.

11.9
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 7 DIRECTION 2181737–100

2.6 Replacing the Bezel


Illustration 11–11
REPLACING THE BEZEL

(1) Replace the Bezel


» Lower the bezel into the lower edge of the chassis
» Hook the bezel into place
» Rotate the bezel back and into the closed position
You can hear the latches snap into place when the bezel is fully installed.
Illustration 11–12
REPLACING THE LOCKBAR

A Push in

(2) OPen the door on the front of the Octane workstation

11.10
GE MEDICAL SYSTEMS CT/I OPTIONS INST ALLATION
REV 7 DIRECTION 2181737-100

(3) Insert the lockbar into the front of the workstation with the angled end up.(A)
(4) Insert and close the lock at the back of the Octane workstation.(B)
You are now ready to power on the Octane workstation.

2.7 Verify Hardwar e Installation


(1) Turn on the CT/i console.
(2) Watch the Display Screen and select "Stop for Maintenance" when displayed.
(3) Select "Command Shell" from the menu.
(4) Enter:
hinv
(5) Verify the hinv output contains the information regarding Disk Drive Unit 2 on SCSI
controller 0.

SECTION 3
LOAD CT/I SOFTWARE

3.1 Install Option Disk Software


(1) Refer to Load From Cold Procedure to reload software.
The Optional Disk will be automatically partitioned and made available to the system software
during the software load.

Notice Only by performing a complete LFC can you ensure that the new drive is properly
set-up and that the Image Database integrity is assured.

SECTION 4
PARTS

4.1 B7710JD CT/i (Octane) Additional Sys. Mag Disk Option Kit
Item Part Number Name SPP Quantity Description (Remarks)
1 2169940-10 4GB DISK DRIVE 1 1 Second Optional 4GB Disk Drive

11.11
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 7 DIRECTION 2181737–100

NOTES

11.12
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100

Chapter 12
CT/i (Octane) ConnectPRO and Bar Code Reader
Option

Table Of Contents

Section Description Page

Section 1
CT/i (Octane) ConnectPRO Option Installation Preparation . . . . . . . . . . 12.1
1.1 Bar Code Reader Kit Hardware Installation . . . . . . . . . . . . . . . . . . . . . . 12.1
1.2 Console Hardware Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12.1
1.3 Verify Bar Code Scanner Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12.4

Section 2
ConnectPRO Software Option Installation & Checkout . . . . . . . . . . . . . . 12.6
2.1 Install and Setup ConnectPRO Option from MOD . . . . . . . . . . . . . . . . 12.6
2.2 ConnectPRO Feature Checkout . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12.9
2.3 Save System State . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12.9

Section 3
Finish Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12.10
3.1 Install Rating Plate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12.10
3.2 Return Installation Card . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12.10

Section 4
Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12.11
4.1 B7540RB Bar Code Reader Only Option Kit Parts List . . . . . . . . . . . . 12.11
4.2 B7500PM HIS/RIS Interface w/Bar Code Reader Option Kit Parts List 12.11
4.3 Material List for Bar Code Reader Kit 2180690 . . . . . . . . . . . . . . . . . . 12.11

12.i
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100

List Of Illustrations

Illustration Page

Illustration 12–1
Bar Code Reader Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12.2
Illustration 12–2
Bar Code Reader Instalation (Detail) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12.3
Illustration 12–3
Bar CoDE SAMPLEs TO SCAN . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12.5
Illustration 12–4
Options Window . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12.6
Illustration 12–5
Sample Software Options Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12.7
Illustration 12–6
ConnectPRO Setup screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12.8
Illustration 12–7
Rating Plate Location . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12.10

12.ii
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100

SECTION 1
CT/i (Octane) ConnectPRO OPTION INSTALLATION PREPARATION

Two versions of the CT/i (Octane) ConnectPRO Option exist:


B7540RB Bar Code Reader Only
Use the hand–held bar code scanner to directly input information to the Exam Rx patient
information screen. The Bar Code Reader hardware option reads information from the paper
schedule or patient wrist strap bar codes.
B7500PM HIS/RIS Interface software WITH Bar Code Reader
Use the HIS/RIS software option to network patient information directly to the CT/i Exam Rx
desktop from the Hospital Information System (HIS) and Radiology Information Systems (RIS)
database. This option kit also contains the hand–held bar code reader.
Before you start the option installation, take an inventory of the supplied kit.
Refer to the “PARTS” Section 4 for the corresponding list(s).

1.1 Bar Code Reader Kit Hardware Installation


1. Shutdown the System:
a. Select the (square red) Shutdown Button (shown below) on the Image Display
Monitor.
b. Select OK to shutdown Applications. Shutdown stops the CT/i Applications and the
UNIX software, then displays the boot prompt.

2. After the software shuts down, the following message appears:

i Okay to power off the system now.


Press any key to restart
Restart

3. Turn off the power switch on the console.

1.2 Console Hardware Installation


1. Remove the lower front cover of the CT/i (Octane) console.
2. Route the bar code scanner cable through the opening in the rear of the console. (Refer
to Illustration 12–1)
3. Attach the Y–cable assembly (2180690–2) to the bar code scanner cord.
(Refer to Illustration 12–2 for details)

12.1
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100

4. Unplug the existing keyboard connector from rear of Octane Computer.


(Refer to Illustration 12–2 for details)
5. Attach keyboard connector to the short length female connector on the Y–Cable
assembly. (Refer to Illustration 12–2 for details)
6. Connect the long length male connector from the Y–Cable assembly to keyboard
connection in the back of the Octane Computer. (Refer to Illustration 12–2 for details)
7. Optional: Mount the bar code scanner’s holder to the left side of the Exam Rx monitor
–or– to a location selected by the site Technologist. Use the double–sided tape
(supplied with kit) to attach the holder.

Console Rear View


Bar Code Scanner

Monitor Monitor

Route Cable through the


opening in the rear of the console

Illustration 12–1
BAR CODE READER INSTALLATION

12.2
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100

Front View of Console Back View of Octane

J21
3rd from top
(in front of console)

2 1

Keyboard Connection
in back of Octane (Console)

Existing Cable
Disconnected
From J21

Y–Cable assembly
(2180690–3)

From Bar Code Scanner


Illustration 12–2
BAR CODE READER INSTALATION (DETAIL)

12.3
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100

1.3 Verify Bar Code Scanner Operation


Note 1 Do not replace the console front cover until you verify the bar code scanner operation.
1. Restore power to the console and boot up the system to the scan user level.
2. Select Patient Schedule and Add Patient on the Exam Rx desktop.
3. Use the hand–held bar code scanner to scan the PATIENT ID# bar code on one of the
Bar code examples in Illustration 12–3. Make sure the screen updates with the
corresponding number, either 555 666 777 or 777 888 999.
Note 2 The bar code scanner beeps when it successfully reads a bar code.
Recommended: New users, take some time to practice with the bar code scanner and bar
code samples. Delete the screen entry and reread the bar codes until you feel comfortable
with the bar code scanner operation.
A User’s Manual comes with the scanner kit. It contains detailed usage information, as well
as additional sample bar codes. If you encounter difficulties, refer to the Operation User
Maintenance and Troubleshooting section of this manual. If you still cannot successfully
scan the sample bar codes, select the Troubleshoot menu from the service desktop, and run
the Keyboard Confidence Test diagnostic to identify a bad cable connection or failed FRU.
4. Select the Requisition # data field.
5. Use the hand held bar code scanner to scan the REQUISITION# bar code
on one of the Bar code examples in Illustration 12–3. Make sure the screen updates
with the corresponding number, either 508 or 510.
6. Manually type/enter the Patient Name, and select Accept.
7. Select View Schedule
8. Make sure the Schedule screen displays the test patient entry.
9. Select Delete.
10. Select Quit.
11. This completes the bar code scanner and keyboard function test.
12. Reinstall the console front cover.
Note 3 For customer option #B7500PM ONLY, proceed to Section 2.

12.4
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100

Illustration 12–3
BAR CODE SAMPLES TO SCAN

REQUISITION #

508

PATIENT ID #

555 666 777

REQUISITION #

510

PATIENT ID #

777 888 999

12.5
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 8 DIRECTION 2181737–100

SECTION 2
ConnectPRO SOFTWARE OPTION INSTALLATION & CHECKOUT

Note 4 Before you install the ConnectPRO option software, obtain the following information
from the hospital’s network administrator:

HIS Server Ip Address ________________________________


HIS Server AE Title ________________________________
HIS Server AE Port# ________________________________

CT Server AE Title ________________________________

2.1 Install and Setup ConnectPRO Option from MOD


1. Open the Service Desktop.
2. Select the Utilities Menu.

3. Select installOptions to display the options window shown in Illustration 12–4.


Illustation 12–4
OPTIONS WINDOW

12.6
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 8 DIRECTION 2181737–100

4. Insert the option MOD into the drive.


5. Select Install.

6. Select Permanent.

7. Select MOD.

8. Select OK to display a software options screen similar to the one shown in Illustration 12–5.
Illustation 12–5
SAMPLE SOFTWARE OPTIONS MANUAL

9. After the ConnectPRO label appears in the “Installed Options” list, a “ConnectPRO
Setup” screen, similar to the one in Illustration 12–6, appears.
10. Type/enter the information you recorded on page 12.6 into the corresponding data
fields on the ConnectPRO Setup screen.
a. If you need additional information, select Help to display a HELP window
explaining each data field.
b. Select Dismiss to exit the ConnectPRO Setup screen without any changes.

12.7
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 8 DIRECTION 2181737–100

Illustation 12–6
ConnectPRO SETUP SCREEN

11. Select Accept to check the valid setup entries, execute a perl script that makes changes
to the configure files for HIS/RIS, and return to the Software Options screen, shown in
Illustration 12–5, on page 12.7.
12. Refer to 12–5.
a. Select Quit.

b. Select ok.

13. Remove the MOD, and write protect the side containing the ConnectPRO option.
Note 5 The initial install requires a write enabled MOD; but you may subsequently install the
option(s) with a write protected MOD.

12.8
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 8 DIRECTION 2181737–100

2.2 ConnectPRO Feature Checkout


1. Start up the Exam Rx applications.
2. Query the HIS/RIS system to verify operation.
(Refer to the User Manual/CBT for details.)
Note 6 To troubleshoot the network hardware: Open a Shell Tool, and type:
more /usr/g/service/networkHelp.txt
to display the network diagnostics help menu, and follow the instructions displayed on the
screen.

2.3 Save System State


The following procedure saves Characterization Data, Calibration Data, Protocols, and
reconfig information, to your System State MOD. If you don’t have a System State MOD,
skip this section.
Note 7 Do not display any images on Image Works when you save the System State. Exit to the
Browser level before you Save System State.
1. Insert the Save System State MOD into the MOD drive.
2. Open the Service desktop.
3. Select the PM option from the list.
4. Select System State.

5. Select the Start button.


6. Select All.

7. Select Save.

Review the output for errors or missing files. You may use the scroll bar on the right
after the tool completes its tasks.
8. When Save completes, select Dismiss.

12.9
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100

SECTION 3
FINISH INSTALLATION

3.1 Install Rating Plate


1. Refer to Illustration 12–7.
2. Locate the current Modification Rating plate(s) on the rear of the CT/i console.
3. Fasten the Option Disk rating plate to the left of the existing rating plates.
Illustration 12–7
RATING PLATE LOCATION

CT/I CONSOLE REAR VIEW

NEW RATING PLATE TO THE


LEFT OF EXISTING RATING
PLATES.

3.2 Return Installation Card


1. Complete and verify the site–specific information on the Red Installation Product
Locator Card.
2. Return the card to the Mailing Address that is printed on the front of the card.

12.10
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100

SECTION 4
PARTS

4.1 B7540RB Bar Code Reader Only Option Kit Parts List
Item Part Number Name SPP Quantity Description (Remarks)
1 2180690 Bar Code Reader Kit N 1 Bar Code Reader Kit including Y–Cable and rubber mount. 2

2 46–302200P3 Serial Rating Plate N 1 Option Kit Rating Plate

4.2 B7500PM HIS/RIS Interface w/Bar Code Reader Option Kit Parts List
Item Part Number Name SPP Quantity Description (Remarks)
1 2180690 Bar Code Reader Kit N 1 Bar Code Reader Kit including Y–Cable and rubber mount

2 2182070 Option MOD N 1 HIS/RIS Option MOD

4 46–302200P3 Serial Rating Plate N 1 Option Kit Rating Plate

4.3 Material List for Bar Code Reader Kit 2180690


Item Part Number Name SPP Quantity Description (Remarks)
1 2180690–2 Bar Code Scanner 1 1 Hand Held Bar Code Scanner ONLY

2 2180690–3 Y–cable assembly. 1 1 Console/Keyboard/Scanner interface cable ONLY

3 2180690–4 Scanner Holder 2 1 Plastic holder ONLY (to store bar code scanner when not in use)

12.11
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 4 DIRECTION 2181737–100

NOTES

12.12
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 9 DIRECTION 2181737-100

Chapter 13
2nd Host Disk Option

With the system shutdown and the power off, install it behind CDROM and MOD.
If it's an addition to the Octane, simply slide it into the middle or top slot with the power off.
Re-power the system, reload software or open a shell and run this script:
/usr/g/scripts/install2Disk
Use:SERVICE, LOG VIEWING, OC HARDWARE INFO to verify host recognizes the new disk. On
the original console, it will look like this:
EXAMPLE: Integral Ethernet: ec0, version 1
OC HW INFO Integral SCSI controller 1: Version WD33C93B, revision D
CDROM: unit 6 on SCSI controller 1
Optical disk: unit 5 on SCSI controller 1
Optical disk: unit 3 on SCSI controller 1
* Disk drive: unit 2 on SCSI controller 1
Disk drive: unit 1 on SCSI controller 1
Integral SCSI controller 0: Version WD33C93B, revision D
Graphics board: Solid Impact
Graphics board: High Impact

13.1
GE MEDICAL SYSTEMS CT/I OPTIONS INSTALLATION
REV 9 DIRECTION 2181737-100

NOTES

13.2
GE Medical Systems: Telex 3797371
P.O. Box 414, Milwaukee, Wisconsin 53201 U.S.A.
(Asia, Pacific, Latin America, North America)

GE Medical Systems – Europe: Telex 698626


283, rue de la Miniére, B.P.34, 78533 Buc Cedex, France