ST1800 to Aurora Replacement


Manual

Copyrighte 2000, 2001 by General Electric Company

Operating Documentation
GE MEDICAL SYSTEMS ST1800 TO AURORA REPLACEMENT MANUAL
REV 0 2280538

D THIS SERVICE MANUAL IS AVAILABLE IN ENGLISH ONLY.


WARNING D IF A CUSTOMER’S SERVICE PROVIDER REQUIRES A LANGUAGE OTHER
THAN ENGLISH, IT IS THE CUSTOMER’S RESPONSIBILITY TO PROVIDE
TRANSLATION SERVICES.
D DO NOT ATTEMPT TO SERVICE THE EQUIPMENT UNLESS THIS SERVICE
MANUAL HAS BEEN CONSULTED AND IS UNDERSTOOD.
D FAILURE TO HEED THIS WARNING MAY RESULT IN INJURY TO THE SERVICE
PROVIDER, OPERATOR OR PATIENT FROM ELECTRIC SHOCK,
MECHANICAL OR OTHER HAZARDS.

D CE MANUEL DE MAINTENANCE N’EST DISPONIBLE QU’EN ANGLAIS.


AVERTISSEMENT D SI LE TECHNICIEN DU CLIENT A BESOIN DE CE MANUEL DANS UNE AUTRE
LANGUE QUE L’ANGLAIS, C’EST AU CLIENT QU’IL INCOMBE DE LE FAIRE
TRADUIRE.
D NE PAS TENTER D’INTERVENTION SUR LES ÉQUIPEMENTS TANT QUE LE
MANUEL SERVICE N’A PAS ÉTÉ CONSULTÉ ET COMPRIS.
D LE NON-RESPECT DE CET AVERTISSEMENT PEUT ENTRAÎNER CHEZ LE
TECHNICIEN, L’OPÉRATEUR OU LE PATIENT DES BLESSURES DUES À DES
DANGERS ÉLECTRIQUES, MÉCANIQUES OU AUTRES.

D DIESES KUNDENDIENST–HANDBUCH EXISTIERT NUR IN


ENGLISCHER SPRACHE.
WARNUNG
D FALLS EIN FREMDER KUNDENDIENST EINE ANDERE SPRACHE BENÖTIGT,
IST ES AUFGABE DES KUNDEN FÜR EINE ENTSPRECHENDE ÜBERSETZUNG
ZU SORGEN.
D VERSUCHEN SIE NICHT, DAS GERÄT ZU REPARIEREN, BEVOR DIESES
KUNDENDIENST–HANDBUCH NICHT ZU RATE GEZOGEN UND VERSTANDEN
WURDE.
D WIRD DIESE WARNUNG NICHT BEACHTET, SO KANN ES ZU VERLETZUNGEN
DES KUNDENDIENSTTECHNIKERS, DES BEDIENERS ODER DES PATIENTEN
DURCH ELEKTRISCHE SCHLÄGE, MECHANISCHE ODER SONSTIGE
GEFAHREN KOMMEN.

D ESTE MANUAL DE SERVICIO SÓLO EXISTE EN INGLÉS.


AVISO D SI ALGÚN PROVEEDOR DE SERVICIOS AJENO A GEMS SOLICITA UN IDIOMA
QUE NO SEA EL INGLÉS, ES RESPONSABILIDAD DEL CLIENTE OFRECER UN
SERVICIO DE TRADUCCIÓN.
D NO SE DEBERÁ DAR SERVICIO TÉCNICO AL EQUIPO, SIN HABER
CONSULTADO Y COMPRENDIDO ESTE MANUAL DE SERVICIO.
D LA NO OBSERVANCIA DEL PRESENTE AVISO PUEDE DAR LUGAR A QUE EL
PROVEEDOR DE SERVICIOS, EL OPERADOR O EL PACIENTE SUFRAN
LESIONES PROVOCADAS POR CAUSAS ELÉCTRICAS, MECÁNICAS O DE
OTRA NATURALEZA.

a
GE MEDICAL SYSTEMS ST1800 TO AURORA REPLACEMENT MANUAL
REV 0 2280538

D ESTE MANUAL DE ASSISTÊNCIA TÉCNICA SÓ SE ENCONTRA


DISPONÍVEL EM INGLÊS.
ATENÇÃO
D SE QUALQUER OUTRO SERVIÇO DE ASSISTÊNCIA TÉCNICA, QUE NÃO A
GEMS, SOLICITAR ESTES MANUAIS NOUTRO IDIOMA, É DA
RESPONSABILIDADE DO CLIENTE FORNECER OS SERVIÇOS DE TRADUÇÃO.
D NÃO TENTE REPARAR O EQUIPAMENTO SEM TER CONSULTADO E
COMPREENDIDO ESTE MANUAL DE ASSISTÊNCIA TÉCNICA.
D O NÃO CUMPRIMENTO DESTE AVISO PODE POR EM PERIGO A SEGURANÇA
DO TÉCNICO, OPERADOR OU PACIENTE DEVIDO A‘ CHOQUES ELÉTRICOS,
MECÂNICOS OU OUTROS.

D IL PRESENTE MANUALE DI MANUTENZIONE È DISPONIBILE


SOLTANTO IN INGLESE.
AVVERTENZA
D SE UN ADDETTO ALLA MANUTENZIONE ESTERNO ALLA GEMS RICHIEDE IL
MANUALE IN UNA LINGUA DIVERSA, IL CLIENTE È TENUTO A PROVVEDERE
DIRETTAMENTE ALLA TRADUZIONE.
D SI PROCEDA ALLA MANUTENZIONE DELL’APPARECCHIATURA SOLO DOPO
AVER CONSULTATO IL PRESENTE MANUALE ED AVERNE COMPRESO IL
CONTENUTO.
D NON TENERE CONTO DELLA PRESENTE AVVERTENZA POTREBBE FAR
COMPIERE OPERAZIONI DA CUI DERIVINO LESIONI ALL’ADDETTO ALLA
MANUTENZIONE, ALL’UTILIZZATORE ED AL PAZIENTE PER
FOLGORAZIONE ELETTRICA, PER URTI MECCANICI OD ALTRI RISCHI.

b
GE MEDICAL SYSTEMS ST1800 TO AURORA REPLACEMENT MANUAL
REV 0 2280538

IMPORTANT! . . . X-RAY PROTECTION


X-ray equipment if not properly used may cause injury. Accordingly, the instructions herein contained should
be thoroughly read and understood by everyone who will use the equipment before you attempt to place this
equipment in operation. The General Electric Company, Medical Systems Group, will be glad to assist and
cooperate in placing this equipment in use.
Although this apparatus incorporates a high degree of protection against x-radiation other than the useful beam, no
practical design of equipment can provide complete protection. Nor can any practical design compel the operator to
take adequate precautions to prevent the possibility of any persons carelessly exposing themselves or others to
radiation.
It is important that everyone having anything to do with x-radiation be properly trained and fully acquainted with the
recommendations of the National Council on Radiation Protection and Measurements as published in NCRP Reports
available from NCRP Publications, 7910 Woodmont Avenue, Room 1016, Bethesda, Maryland 20814, and of the
International Commission on Radiation Protection, and take adequate steps to protect against injury.
The equipment is sold with the understanding that the General Electric Company, Medical Systems Group, its agents,
and representatives have no responsibility for injury or damage which may result from improper use of the equipment.
Various protective material and devices are available. It is urged that such materials or devices be used.

CERTIFIED ELECTRICAL CONTRACTOR STATEMENT


All electrical installations that are preliminary to positioning of the equipment at the site prepared for the equipment shall be
performed by licensed electrical contractors. In addition, electrical feeds into the Power Distribution Unit shall be performed
by licensed electrical contractors. Other connections between pieces of electrical equipment, calibrations, and testing shall
be performed by qualified GE Medical personnel. The products involved (and the accompanying electrical installations) are
highly sophisticated, and special engineering competence is required.

In performing all electrical work on these products, GE will use its own specially trained field engineers. All of GE’s electrical
work on these products will comply with the requirements of the applicable electrical codes.

The purchaser of GE equipment shall only utilize qualified personnel (i.e., GE’s field engineers, personnel of third-party
service companies with equivalent training, or licensed electricians) to perform electrical servicing on the equipment.

c
GE MEDICAL SYSTEMS ST1800 TO AURORA REPLACEMENT MANUAL
REV 0 2280538

DAMAGE IN TRANSPORTATION

All packages should be closely examined at time of delivery. If damage is apparent, have notation “damage in
shipment” written on all copies of the freight or express bill before delivery is accepted or “signed for” by a General
Electric representative or a hospital receiving agent. Whether noted or concealed, damage MUST be reported to the
carrier immediately upon discovery, or in any event, within 14 days after receipt, and the contents and containers held
for inspection by the carrier. A transportation company will not pay a claim for damage if an inspection is not requested
within this 14 day period.

Call Traffic and Transportation, Milwaukee, WI (414) 827–3449 / 8*285–3449 immediately after damage is found. At
this time be ready to supply name of carrier, delivery date, consignee name, freight or express bill number, item
damaged and extent of damage.

Complete instructions regarding claim procedure are found in Section “S” of the Policy & Procedure Bulletins.

OMISSIONS & ERRORS

GE personnel, please use the GEMS CQA Process to report all omissions, errors, and defects in this documentation.
Customers, please contact your GE Sales or Service represenatives.

CAUTION

Do not use the following devices near this equipment. Use of these devices near this equipment could cause
this equipment to malfunction.

Devices not to be used near this equipment:

Devices which intrinsically transmit radio waves such as; cellular phone, radio transceiver, mobile radio transmitter,
radio–controlled toy, etc.

Keep power to these devices turned off when near this equipment.

Medical staff in charge of this equipment is required to instruct technicians, patients and other people who may be
around this equipment to fully comply with the above regulation.

d
GE MEDICAL SYSTEMS ST1800 TO AURORA REPLACEMENT MANUAL
REV 1 2280538

REVISION HISTORY

REV DATE PRIMARY REASON FOR CHANGE


0 . . . . . Sep. 30, 2000 . . . Initial release.
1 . . . . . Sep. 12, 2001 . . . General Update

LIST OF EFFECTIVE PAGES

PAGE REV PAGE REV PAGE REV PAGE REV PAGE REV
Title page . . . . . . . . . 1 i.................. 0 Tab 2 Tab 3
Title page rear . blank 2–1 . . . . . . . . . . . . . . . 1 3–1 to 3–17 . . . . . . . . 0 Blank/Rear cover . . . –
a to d . . . . . . . . . . . . . 0 Tab 1
A ................. 1 1–1 . . . . . . . . . . . . . . . 0

A
GE MEDICAL SYSTEMS ST1800 TO AURORA REPLACEMENT MANUAL
blank 2280538

B
GE MEDICAL SYSTEMS ST1800 TO AURORA REPLACEMENT MANUAL
REV 0 2280538

TABLE OF CONTENTS

SECTION PAGE

SECTION 1 – INTRODUCTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1–1

SECTION 2 – MATERIAL LIST . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2–1

SECTION 3 – INSTALLATION PROCEDURE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3–1


3-1 INSTALLING THE AURORA FOR HISPEED SERIES . . . . . . . . . . . . . . . . . . . . . . . . . . 3–1
3-1-1 Removing the ST1800 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3–1
3-1-2 Installing the Aurora . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3–3
3-2 INSTALLING THE AURORA FOR CT/E, PROSPEED AI/FI SERIES . . . . . . . . . . . . . 3–10
3-2-1 Removing the ST1800 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3–10
3-2-2 Installing the Aurora . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3–13
3-3 INSTALLING PATCH SOFTWARE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3–16
Patch Verification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3–16
Reconfigration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3–17

i
GE MEDICAL SYSTEMS ST1800 TO AURORA REPLACEMENT MANUAL
blank 2280538

ii
GE MEDICAL SYSTEMS ST1800 TO AURORA REPLACEMENT MANUAL
REV 0 2280538

SECTION 1 – INTRODUCTION

This manual describes serial port expander replacement from ST1800 to Aurora.

FLOWCHART
The flowchart below shows the main flow of the replacement procedures of this kit.

START

Removing the ST1800

Installing the Aurora

Installing the Patch Software

Reconfiguration

END

1–1
GE MEDICAL SYSTEMS ST1800 TO AURORA REPLACEMENT MANUAL
blank 2280538

1–2
GE MEDICAL SYSTEMS ST1800 TO AURORA REPLACEMENT MANUAL
REV 1 2280538

SECTION 2 – MATERIAL LIST

Serial Expander Replacement Kit (for HiSpeed Series)


2280219

ITEM PART NO. DESCRIPTION QTY.


1 2251601 SP Assy 1
2 2243265 Serial Port Expander 1
3 2247648 Serial Cable Assy 4
4 2247454 SCSI Cable Assy 1
5 2219693–2 Service Key Cable Assy 1
6 2297143 CD–ROM: Patch Software V/R 4.16 1
7 2280538 Service Manual (This manual) 1
8 N9408LK Screw (for Serial Port Expander and Support Bracket) 3
9 U0508AC Edge Guard 2
10 U0014AC Clamp 1
11 U0202AC Tie–wrap 10
12 2281422 Support Bracket (for Mobile System Only) 1

Serial Expander Replacement Kit (for CT/e, ProSpeed AI/FI Series)


2280220

ITEM PART NO. DESCRIPTION QTY.


1 2265429 SP Assy 1
2 2243265 Serial Port Expander 1
3 2249698 Serial Cable Assy 4
4 2156306–2 SCSI Cable Assy 1
5 2219693–2 Service Key Cable Assy 1
6 2297143 CD–ROM: Patch Software V/R 4.16 1
7 2280538 Service Manual (This manual) 1
8 N9408LK Screw (for Serial Port Expander) 1
9 U0202AC Tie–wrap 10

2–1
GE MEDICAL SYSTEMS ST1800 TO AURORA REPLACEMENT MANUAL
blank 2280538

2–2
GE MEDICAL SYSTEMS ST1800 TO AURORA REPLACEMENT MANUAL
REV 0 2280538

SECTION 3 – INSTALLATION PROCEDURE

3-1 INSTALLING THE AURORA FOR HISPEED SERIES


Understand and follow the Safety Guidelines Manual.

3-1-1 Removing the ST1800


1. Turn OFF the OC main power.

2. Remove the following covers from the OC:

D OC Front Cover ..................... Four(4) Screws


D EMC Cover F–A ..................... Twelve(12) Screws
D EMC Cover F–B ..................... Ten(10) Screws

EMC Cover F–B

Front Cover
EMC Cover F–A

3. Remove seven screws, then slide out the SCSI bay.

4. Disconnect the following Seven cables from the ST1800 and discard them.

D Four serial cables between Rear CN2 (J3–J6) and ST1800

D One serial cable between Service key connector and ST1800

D One SCSI cable between CPU and ST1800

D One SCSI cable between NPSC Assy (J14) and ST1800

3–1
GE MEDICAL SYSTEMS ST1800 TO AURORA REPLACEMENT MANUAL
REV 0 2280538

3-1-1 Removing the ST1800 (Continued)


5. Disconnect the power cable from the AC1 on the AC PS Assy.

6. Remove the ST1800 bracket (with Power Supply) from the OC by unscrewing its 2 screws, remove the ST1800
Assy

7. Cut the tie–wraps holding the power supply to the ST1800 bracket.

Power Supply
Screw (x2)
ST1800 Bracket

Tie–wrap Base

Serial Port Expander


(ST1800)

3–2
GE MEDICAL SYSTEMS ST1800 TO AURORA REPLACEMENT MANUAL
REV 0 2280538

3-1-2 Installing the Aurora

NOTICE
Before touching any of the components, attach the wrist strap to your wrist and to the metal
part of the chassis.

1. Install the Aurora (Serial port expander; 2243265) into No.5 slot of the Nest Assy, then fix it using the screw
(N9408LK) shipped with this kit.

Screw (N9408LK)

Aurora (2243265)

For Mobile System Only:


Install the support bracket to hold the aurora using the three screws (N9408LK).

Screws (N9408LK)

Support Bracket (2281422)

3–3
GE MEDICAL SYSTEMS ST1800 TO AURORA REPLACEMENT MANUAL
REV 0 2280538

3-1-2 Installing the Aurora (Continued)


2. Loosen the two fan mounting screws, then install the SP Assy bracket using their screws.

3. Install the SP Assy using the two screws (N9408ZX) shipped with this kit.

Fan mounting screws

SP Assy bracket

SP Assy
Screws (N9408ZX)

3–4
GE MEDICAL SYSTEMS ST1800 TO AURORA REPLACEMENT MANUAL
REV 0 2280538

3-1-2 Installing the Aurora (Continued)


4. Clean any dirt off the attaching surface of the cable clamp (U0014AC), shipped with this kit, using alcohol and
attach the clamp onto the CPU.

Clamp

5. Connect the SP Assy serial cable to the Aurora.

SP Assy Serial Cable

3–5
GE MEDICAL SYSTEMS ST1800 TO AURORA REPLACEMENT MANUAL
REV 0 2280538

3-1-2 Installing the Aurora (Continued)


6. Connect the following new cables:

D Four serial cables between SP Assy (No.1 – No.4) and Rear CN2 panel (J3 – J6), 2247648

D One serial cable between SP Assy (No.5) and Service Key connector, 2219693–2

D One SCSI cable to the CPU SCSI terminal, 2247454

SP Assy (Serial Cables, 2247648)


Rear CN2 (Serial
Cables, 2247648)

Service key Connecter (Serial Cable,


2219693–2)

CPU (SCSI Cable, 2247454)

The four serial cables, connected to


the Rear CN2 Assy, must be held by
the clamp installed in step 4.

3–6
GE MEDICAL SYSTEMS ST1800 TO AURORA REPLACEMENT MANUAL
REV 0 2280538

3-1-2 Installing the Aurora (Continued)


7. Connect the SCSI Cable (2247454) from the CPU to the NPSC J14.

NPSC Assy
J14

3–7
GE MEDICAL SYSTEMS ST1800 TO AURORA REPLACEMENT MANUAL
REV 0 2280538

3-1-2 Installing the Aurora (Continued)


8. Clamp the cables properly using the tie–wraps, shipped with this kit.

D (1) Three tie–wraps


used to bind four serial cables between SP Assy No.1~No.4 connectors and Rear CN2.

D (2) One tie–wrap


used to bind one serial cable between SP Assy No.5 connector and service key connector.

D (3) Two tie–wraps


used to bind SCSI cable 2247454.

(1) Three tie–wraps (2) One tie–wrap (3) Two tie–wraps

3–8
GE MEDICAL SYSTEMS ST1800 TO AURORA REPLACEMENT MANUAL
REV 0 2280538

3-1-2 Installing the Aurora (Continued)


9. Cut the edge guard to two 315mm (12.4 in) pieces of the guard.

10. Apply a fast setting adhesive to the grooves of the edge guards, then install them to the OC center main frame
so that they are touched to the bottom frame of the OC.

11. Then set the serial cables between SP Assy and Rear CN2 to the back side of the center main frame.

Center Main Frame

Edge Guard Edge Guard


(315mm) (315mm)

The end face must be touched to the


bottom frame.

12. Install the SCSI bay and OC covers in the reverse order of removal.

3–9
GE MEDICAL SYSTEMS ST1800 TO AURORA REPLACEMENT MANUAL
REV 0 2280538

3-2 INSTALLING THE AURORA FOR CT/e, PROSPEED AI/FI SERIES


Understand and follow the Safety Guidelines Manual.

3-2-1 Removing the ST1800


1. Turn OFF the OC main power.

2. Remove the front cover, rear cover, and left side cover.

3. Disconnect the following Seven cables from the ST1800 and discard them.

D Four serial cables between Rear CN2 (J3–J6) and ST1800:


Remove the two Rear CN2 mounting bolts and pull out the board to access the connectors.

Serial Cables of the Rear CN2

3–10
GE MEDICAL SYSTEMS ST1800 TO AURORA REPLACEMENT MANUAL
REV 0 2280538

3-2-1 Removing the ST1800 (Continued)


D One serial cable between Service key connector and ST1800:
Remove the four service connector mouting bolts and remove the two nuts to remove the serial cable from
the cover.

Serial Cable Cover

Nuts

D One SCSI cable between CPU and ST1800

D One SCSI cable between NPSC Assy (J14) and ST1800

3–11
GE MEDICAL SYSTEMS ST1800 TO AURORA REPLACEMENT MANUAL
REV 0 2280538

3-2-1 Removing the ST1800 (Continued)


4. Disconnect the cables and power cable connectors from the ST1800 Serial Port Expander.

5. Cut the tie–wraps that hold the cables connected to the ST1800 Serial Port Expander.

6. Remove the ST1800 bracket from the OC by unscrewing its 4 screws, remove the ST1800 Assy.

Serial Port Expander


(ST1800)

Screws (x4)

3–12
GE MEDICAL SYSTEMS ST1800 TO AURORA REPLACEMENT MANUAL
REV 0 2280538

3-2-2 Installing the Aurora

NOTICE
Before touching any of the components, attach the wrist strap to your wrist and to the metal
part of the chassis.

1. Install the Aurora into No.5 slot of the Nest Assy, then fix it using the screw (N9408LK) shipped with this kit.

2. Install the SP Assy using the two screws (N9408HR) shipped with this kit.

3. Connect the SP Assy serial cable to the Aurora.

Screws (N9408HR)
SP Assy

The SP Assy con-


tains the screw
hole here to mount
the SP Assy, but
ignore this when
installing the SP
Assy.

Screw (N9408LK) Aurora

3–13
GE MEDICAL SYSTEMS ST1800 TO AURORA REPLACEMENT MANUAL
REV 0 2280538

3-2-2 Installing the Aurora (Continued)


4. Connect the following new cables and hold them using tie–wraps:

D Four serial cables (2249698) between SP Assy (No.1 – No.4) and Rear CN2 panel (J3 – J6):

Rear

Front
OC Top View
Serial Cables of the Bind the serial
Rear CN2 cables using
this tie–wrap Bind the four se-
base. rial cables using
three tie–wraps.
Rear CN2
SP ASSY

OC Left Side View

Rear CN2

3–14
GE MEDICAL SYSTEMS ST1800 TO AURORA REPLACEMENT MANUAL
REV 0 2280538

3-2-2 Installing the Aurora (Continued)


D One serial cable (2219693–2) between SP Assy (No.5) and Service Key connector

Hold the serial cable onto the tie–wrap base located on


the back side of the upper frame using the tie–waps.

Serial Cable
(2219693–2)

D One SCSI cable (2156306–2) between the CPU SCSI terminal and NPSC Assy J14

NPSC Assy J14

Tie–wrap

CPU SCSI Terminal

To CPU SCSI Terminal

5. Install the SCSI bay and OC covers in the reverse order of removal.

3–15
GE MEDICAL SYSTEMS ST1800 TO AURORA REPLACEMENT MANUAL
REV 0 2280538

3-3 INSTALLING PATCH SOFTWARE


1. Power ON the main switch of the OC.

2. Approx. 30 seconds later the desktop menu and “Auto–Start” red window appears. Click on Cancel in the Auto–
Start window to enter the desktop menu.

Note
If you miss clicking the button, press Shift + 7 in the next black screen.

3. Insert the CD–ROM (Patch Software 4.14) into the CD–ROM drive of the OC.

4. On the desktop menu, click on Install Software located on the upper right of the screen.

5. “Do you install software from CD–ROM?” message appears. Click on Yes.

6. “Do you install selected Patch?” message appears. Click on Yes.


The patch installation procedures are automatically started.

7. Click on OK in response to “Installation Successful. xxxxxxx”.

8. After completion of the patch installation, Click on Shell located on the upper right of the screen.

9. Type the following to eject the CD–ROM.

> eject <Enter>

> exit <Enter>

Patch Verification
1. Click on List S/W Package.
2. Click on Patch.
3. Verify that information for the patch installed are displayed. (Example)
******************************* Displayed *******************************
Patch NPP41001: HiSpeed, ProSpeed, CT/e Series Patch Software 4.11
Patch NPP41002: HiSpeed, ProSpeed, CT/e Series Patch Software 4.12
Patch NPP41003: HiSpeed, ProSpeed, CT/e Series Patch Software 4.13
Patch NPP41004: HiSpeed, ProSpeed, CT/e Series Patch Software 4.14
************************************************************************

4. Click on Close.

3–16
GE MEDICAL SYSTEMS ST1800 TO AURORA REPLACEMENT MANUAL
REV 0 2280538

3-4 Installing Patch Software (Continued)

Reconfigration
Before performing reconfiguration, the system must be rebooted again.

1. On the desktop menu, click on Shutdown.


Some test programs run. The NO button to be displayed is clicked.

2. Verify that the message “Okay to power off the system now. Press any key to restart.” appears.

3. Click on Restart.

4. Approx. 30 seconds later the desktop menu and “Auto–Start” red window appears. Click on Cancel in the Auto–
Start window to enter the desktop menu.

Note
If you miss clicking the button, press Shift + 7 in the next black screen.

5. On the desktop menu, click on Reconfig located on the upper right of the screen.
This System configuration screen appears.

6. Click on Hardware.
This Hardware setting screen appears.

7. Select Aurora for Serial Expander.

8. Click on Accept.
The system is rebooted. It takes approx. 7 minutes.

3–17
GE MEDICAL SYSTEMS ST1800 TO AURORA REPLACEMENT MANUAL
blank 2280538

3–18