GE MEDICAL SYSTEMS ST1800 TO AURORA REPLACEMENT MANUAL
REV 0 2280538
a
GE MEDICAL SYSTEMS ST1800 TO AURORA REPLACEMENT MANUAL
REV 0 2280538
b
GE MEDICAL SYSTEMS ST1800 TO AURORA REPLACEMENT MANUAL
REV 0 2280538
In performing all electrical work on these products, GE will use its own specially trained field engineers. All of GE’s electrical
work on these products will comply with the requirements of the applicable electrical codes.
The purchaser of GE equipment shall only utilize qualified personnel (i.e., GE’s field engineers, personnel of third-party
service companies with equivalent training, or licensed electricians) to perform electrical servicing on the equipment.
c
GE MEDICAL SYSTEMS ST1800 TO AURORA REPLACEMENT MANUAL
REV 0 2280538
DAMAGE IN TRANSPORTATION
All packages should be closely examined at time of delivery. If damage is apparent, have notation “damage in
shipment” written on all copies of the freight or express bill before delivery is accepted or “signed for” by a General
Electric representative or a hospital receiving agent. Whether noted or concealed, damage MUST be reported to the
carrier immediately upon discovery, or in any event, within 14 days after receipt, and the contents and containers held
for inspection by the carrier. A transportation company will not pay a claim for damage if an inspection is not requested
within this 14 day period.
Call Traffic and Transportation, Milwaukee, WI (414) 827–3449 / 8*285–3449 immediately after damage is found. At
this time be ready to supply name of carrier, delivery date, consignee name, freight or express bill number, item
damaged and extent of damage.
Complete instructions regarding claim procedure are found in Section “S” of the Policy & Procedure Bulletins.
GE personnel, please use the GEMS CQA Process to report all omissions, errors, and defects in this documentation.
Customers, please contact your GE Sales or Service represenatives.
CAUTION
Do not use the following devices near this equipment. Use of these devices near this equipment could cause
this equipment to malfunction.
Devices which intrinsically transmit radio waves such as; cellular phone, radio transceiver, mobile radio transmitter,
radio–controlled toy, etc.
Keep power to these devices turned off when near this equipment.
Medical staff in charge of this equipment is required to instruct technicians, patients and other people who may be
around this equipment to fully comply with the above regulation.
d
GE MEDICAL SYSTEMS ST1800 TO AURORA REPLACEMENT MANUAL
REV 1 2280538
REVISION HISTORY
PAGE REV PAGE REV PAGE REV PAGE REV PAGE REV
Title page . . . . . . . . . 1 i.................. 0 Tab 2 Tab 3
Title page rear . blank 2–1 . . . . . . . . . . . . . . . 1 3–1 to 3–17 . . . . . . . . 0 Blank/Rear cover . . . –
a to d . . . . . . . . . . . . . 0 Tab 1
A ................. 1 1–1 . . . . . . . . . . . . . . . 0
A
GE MEDICAL SYSTEMS ST1800 TO AURORA REPLACEMENT MANUAL
blank 2280538
B
GE MEDICAL SYSTEMS ST1800 TO AURORA REPLACEMENT MANUAL
REV 0 2280538
TABLE OF CONTENTS
SECTION PAGE
i
GE MEDICAL SYSTEMS ST1800 TO AURORA REPLACEMENT MANUAL
blank 2280538
ii
GE MEDICAL SYSTEMS ST1800 TO AURORA REPLACEMENT MANUAL
REV 0 2280538
SECTION 1 – INTRODUCTION
This manual describes serial port expander replacement from ST1800 to Aurora.
FLOWCHART
The flowchart below shows the main flow of the replacement procedures of this kit.
START
Reconfiguration
END
1–1
GE MEDICAL SYSTEMS ST1800 TO AURORA REPLACEMENT MANUAL
blank 2280538
1–2
GE MEDICAL SYSTEMS ST1800 TO AURORA REPLACEMENT MANUAL
REV 1 2280538
2–1
GE MEDICAL SYSTEMS ST1800 TO AURORA REPLACEMENT MANUAL
blank 2280538
2–2
GE MEDICAL SYSTEMS ST1800 TO AURORA REPLACEMENT MANUAL
REV 0 2280538
Front Cover
EMC Cover F–A
4. Disconnect the following Seven cables from the ST1800 and discard them.
3–1
GE MEDICAL SYSTEMS ST1800 TO AURORA REPLACEMENT MANUAL
REV 0 2280538
6. Remove the ST1800 bracket (with Power Supply) from the OC by unscrewing its 2 screws, remove the ST1800
Assy
7. Cut the tie–wraps holding the power supply to the ST1800 bracket.
Power Supply
Screw (x2)
ST1800 Bracket
Tie–wrap Base
3–2
GE MEDICAL SYSTEMS ST1800 TO AURORA REPLACEMENT MANUAL
REV 0 2280538
NOTICE
Before touching any of the components, attach the wrist strap to your wrist and to the metal
part of the chassis.
1. Install the Aurora (Serial port expander; 2243265) into No.5 slot of the Nest Assy, then fix it using the screw
(N9408LK) shipped with this kit.
Screw (N9408LK)
Aurora (2243265)
Screws (N9408LK)
3–3
GE MEDICAL SYSTEMS ST1800 TO AURORA REPLACEMENT MANUAL
REV 0 2280538
3. Install the SP Assy using the two screws (N9408ZX) shipped with this kit.
SP Assy bracket
SP Assy
Screws (N9408ZX)
3–4
GE MEDICAL SYSTEMS ST1800 TO AURORA REPLACEMENT MANUAL
REV 0 2280538
Clamp
3–5
GE MEDICAL SYSTEMS ST1800 TO AURORA REPLACEMENT MANUAL
REV 0 2280538
D Four serial cables between SP Assy (No.1 – No.4) and Rear CN2 panel (J3 – J6), 2247648
D One serial cable between SP Assy (No.5) and Service Key connector, 2219693–2
3–6
GE MEDICAL SYSTEMS ST1800 TO AURORA REPLACEMENT MANUAL
REV 0 2280538
NPSC Assy
J14
3–7
GE MEDICAL SYSTEMS ST1800 TO AURORA REPLACEMENT MANUAL
REV 0 2280538
3–8
GE MEDICAL SYSTEMS ST1800 TO AURORA REPLACEMENT MANUAL
REV 0 2280538
10. Apply a fast setting adhesive to the grooves of the edge guards, then install them to the OC center main frame
so that they are touched to the bottom frame of the OC.
11. Then set the serial cables between SP Assy and Rear CN2 to the back side of the center main frame.
12. Install the SCSI bay and OC covers in the reverse order of removal.
3–9
GE MEDICAL SYSTEMS ST1800 TO AURORA REPLACEMENT MANUAL
REV 0 2280538
2. Remove the front cover, rear cover, and left side cover.
3. Disconnect the following Seven cables from the ST1800 and discard them.
3–10
GE MEDICAL SYSTEMS ST1800 TO AURORA REPLACEMENT MANUAL
REV 0 2280538
Nuts
3–11
GE MEDICAL SYSTEMS ST1800 TO AURORA REPLACEMENT MANUAL
REV 0 2280538
5. Cut the tie–wraps that hold the cables connected to the ST1800 Serial Port Expander.
6. Remove the ST1800 bracket from the OC by unscrewing its 4 screws, remove the ST1800 Assy.
Screws (x4)
3–12
GE MEDICAL SYSTEMS ST1800 TO AURORA REPLACEMENT MANUAL
REV 0 2280538
NOTICE
Before touching any of the components, attach the wrist strap to your wrist and to the metal
part of the chassis.
1. Install the Aurora into No.5 slot of the Nest Assy, then fix it using the screw (N9408LK) shipped with this kit.
2. Install the SP Assy using the two screws (N9408HR) shipped with this kit.
Screws (N9408HR)
SP Assy
3–13
GE MEDICAL SYSTEMS ST1800 TO AURORA REPLACEMENT MANUAL
REV 0 2280538
D Four serial cables (2249698) between SP Assy (No.1 – No.4) and Rear CN2 panel (J3 – J6):
Rear
Front
OC Top View
Serial Cables of the Bind the serial
Rear CN2 cables using
this tie–wrap Bind the four se-
base. rial cables using
three tie–wraps.
Rear CN2
SP ASSY
Rear CN2
3–14
GE MEDICAL SYSTEMS ST1800 TO AURORA REPLACEMENT MANUAL
REV 0 2280538
Serial Cable
(2219693–2)
D One SCSI cable (2156306–2) between the CPU SCSI terminal and NPSC Assy J14
Tie–wrap
5. Install the SCSI bay and OC covers in the reverse order of removal.
3–15
GE MEDICAL SYSTEMS ST1800 TO AURORA REPLACEMENT MANUAL
REV 0 2280538
2. Approx. 30 seconds later the desktop menu and “Auto–Start” red window appears. Click on Cancel in the Auto–
Start window to enter the desktop menu.
Note
If you miss clicking the button, press Shift + 7 in the next black screen.
3. Insert the CD–ROM (Patch Software 4.14) into the CD–ROM drive of the OC.
4. On the desktop menu, click on Install Software located on the upper right of the screen.
5. “Do you install software from CD–ROM?” message appears. Click on Yes.
8. After completion of the patch installation, Click on Shell located on the upper right of the screen.
Patch Verification
1. Click on List S/W Package.
2. Click on Patch.
3. Verify that information for the patch installed are displayed. (Example)
******************************* Displayed *******************************
Patch NPP41001: HiSpeed, ProSpeed, CT/e Series Patch Software 4.11
Patch NPP41002: HiSpeed, ProSpeed, CT/e Series Patch Software 4.12
Patch NPP41003: HiSpeed, ProSpeed, CT/e Series Patch Software 4.13
Patch NPP41004: HiSpeed, ProSpeed, CT/e Series Patch Software 4.14
************************************************************************
4. Click on Close.
3–16
GE MEDICAL SYSTEMS ST1800 TO AURORA REPLACEMENT MANUAL
REV 0 2280538
Reconfigration
Before performing reconfiguration, the system must be rebooted again.
2. Verify that the message “Okay to power off the system now. Press any key to restart.” appears.
3. Click on Restart.
4. Approx. 30 seconds later the desktop menu and “Auto–Start” red window appears. Click on Cancel in the Auto–
Start window to enter the desktop menu.
Note
If you miss clicking the button, press Shift + 7 in the next black screen.
5. On the desktop menu, click on Reconfig located on the upper right of the screen.
This System configuration screen appears.
6. Click on Hardware.
This Hardware setting screen appears.
8. Click on Accept.
The system is rebooted. It takes approx. 7 minutes.
3–17
GE MEDICAL SYSTEMS ST1800 TO AURORA REPLACEMENT MANUAL
blank 2280538