Technical

Publications

5116386
Revision 1

HiSpeed Dual, ProSpeed FII Series


Advanced Installation

Copyright © 2004 by General Electric Company

Advanced Service Documentation


Property of GE
For GE Service Personnel Only
No Rights Licensed - Do Not Use or Copy
Disclosure to Third Parties Prohibited
HiSpeed Dual, ProSpeed FII Series
GE MEDICAL SYSTEMS ADVANCED INSTALLATION
REV 1 5116386

• THIS SERVICE MANUAL IS AVAILABLE IN ENGLISH ONLY.


• IF A CUSTOMER’S SERVICE PROVIDER REQUIRES A LANGUAGE OTHER
THAN ENGLISH, IT IS THE CUSTOMER’S RESPONSIBILITY TO PROVIDE
WARNING TRANSLATION SERVICES.
• DO NOT ATTEMPT TO SERVICE THE EQUIPMENT UNLESS THIS SERVICE
MANUAL HAS BEEN CONSULTED AND IS UNDERSTOOD.
• FAILURE TO HEED THIS WARNING MAY RESULT IN INJURY TO THE SERVICE
PROVIDER, OPERATOR OR PATIENT FROM ELECTRIC SHOCK,
MECHANICAL OR OTHER HAZARDS.

• CE MANUEL DE MAINTENANCE N’EST DISPONIBLE QU’EN ANGLAIS.


• SI LE TECHNICIEN DU CLIENT A BESOIN DE CE MANUEL DANS UNE AUTRE
LANGUE QUE L’ANGLAIS, C’EST AU CLIENT QU’IL INCOMBE DE LE FAIRE
AVERTISSEMENT TRADUIRE.
• NE PAS TENTER D’INTERVENTION SUR LES ÉQUIPEMENTS TANT QUE LE
MANUEL SERVICE N’A PAS ÉTÉ CONSULTÉ ET COMPRIS.
• LE NON-RESPECT DE CET AVERTISSEMENT PEUT ENTRAÎNER CHEZ LE
TECHNICIEN, L’OPÉRATEUR OU LE PATIENT DES BLESSURES DUES À DES
DANGERS ÉLECTRIQUES, MÉCANIQUES OU AUTRES.

• DIESES KUNDENDIENST–HANDBUCH EXISTIERT NUR IN ENGLISCHER


SPRACHE.
• FALLS EIN FREMDER KUNDENDIENST EINE ANDERE SPRACHE BENÖTIGT,
WARNUNG IST ES AUFGABE DES KUNDEN FÜR EINE ENTSPRECHENDE ÜBERSETZUNG
ZU SORGEN.
• VERSUCHEN SIE NICHT, DAS GERÄT ZU REPARIEREN, BEVOR DIESES
KUNDENDIENST–HANDBUCH ZU RATE GEZOGEN UND VERSTANDEN
WURDE.
• WIRD DIESE WARNUNG NICHT BEACHTET, SO KANN ES ZU VERLETZUNGEN
DES KUNDENDIENSTTECHNIKERS, DES BEDIENERS ODER DES PATIENTEN
DURCH ELEKTRISCHE SCHLÄGE, MECHANISCHE ODER SONSTIGE
GEFAHREN KOMMEN.

• ESTE MANUAL DE SERVICIO SÓLO EXISTE EN INGLÉS.


• SI ALGÚN PROVEEDOR DE SERVICIOS AJENO A GEMS SOLICITA UN IDIOMA
QUE NO SEA EL INGLÉS, ES RESPONSABILIDAD DEL CLIENTE OFRECER UN
AVISO SERVICIO DE TRADUCCIÓN.
• NO SE DEBERÁ DAR SERVICIO TÉCNICO AL EQUIPO, SIN HABER
CONSULTADO Y COMPRENDIDO ESTE MANUAL DE SERVICIO.
• LA NO OBSERVANCIA DEL PRESENTE AVISO PUEDE DAR LUGAR A QUE EL
PROVEEDOR DE SERVICIOS, EL OPERADOR O EL PACIENTE SUFRAN
LESIONES PROVOCADAS POR CAUSAS ELÉCTRICAS, MECÁNICAS O DE
OTRA NATURALEZA.

a
HiSpeed Dual, ProSpeed FII Series
GE MEDICAL SYSTEMS ADVANCED INSTALLATION
REV 1 5116386

• ESTE MANUAL DE ASSISTÊNCIA TÉCNICA SÓ SE ENCONTRA DISPONÍVEL


EM INGLÊS.
• SE QUALQUER OUTRO SERVIÇO DE ASSISTÊNCIA TÉCNICA, QUE NÃO A
ATENÇÃO GEMS, SOLICITAR ESTES MANUAIS NOUTRO IDIOMA, É DA
RESPONSABILIDADE DO CLIENTE FORNECER OS SERVIÇOS DE TRADUÇÃO.
• NÃO TENTE REPARAR O EQUIPAMENTO SEM TER CONSULTADO E
COMPREENDIDO ESTE MANUAL DE ASSISTÊNCIA TÉCNICA.
• O NÃO CUMPRIMENTO DESTE AVISO PODE POR EM PERIGO A SEGURANÇA
DO TÉCNICO, OPERADOR OU PACIENTE DEVIDO A‘ CHOQUES ELÉTRICOS,
MECÂNICOS OU OUTROS.

• IL PRESENTE MANUALE DI MANUTENZIONE È DISPONIBILE SOLTANTO IN


INGLESE.
• SE UN ADDETTO ALLA MANUTENZIONE ESTERNO ALLA GEMS RICHIEDE IL
AVVERTENZA MANUALE IN UNA LINGUA DIVERSA, IL CLIENTE È TENUTO A PROVVEDERE
DIRETTAMENTE ALLA TRADUZIONE.
• SI PROCEDA ALLA MANUTENZIONE DELL’APPARECCHIATURA SOLO DOPO
AVER CONSULTATO IL PRESENTE MANUALE ED AVERNE COMPRESO IL
CONTENUTO.
• NON TENERE CONTO DELLA PRESENTE AVVERTENZA POTREBBE FAR
COMPIERE OPERAZIONI DA CUI DERIVINO LESIONI ALL’ADDETTO ALLA
MANUTENZIONE, ALL’UTILIZZATORE ED AL PAZIENTE PER
FOLGORAZIONE ELETTRICA, PER URTI MECCANICI OD ALTRI RISCHI.

b
HiSpeed Dual, ProSpeed FII Series
GE MEDICAL SYSTEMS ADVANCED INSTALLATION
REV 1 5116386

IMPORTANT! . . . X-RAY PROTECTION


X–ray equipment if not properly used may cause injury. Accordingly, the instructions herein contained
should be thoroughly read and understood by everyone who will use the equipment before you attempt to
place this equipment in operation. The General Electric Company, Medical Systems Group, will be glad to
assist and cooperate in placing this equipment in use.
Although this apparatus incorporates a high degree of protection against x–radiation other than the useful beam,
no practical design of equipment can provide complete protection. Nor can any practical design compel the opera-
tor to take adequate precautions to prevent the possibility of any persons carelessly exposing themselves or others
to radiation.
It is important that everyone having anything to do with x–radiation be properly trained and fully acquainted with the
recommendations of the National Council on Radiation Protection and Measurements as published in NCRP
Reports available from NCRP Publications, 7910 Woodmont Avenue, Room 1016, Bethesda, Maryland 20814, and
of the International Commission on Radiation Protection, and take adequate steps to protect against injury.

The equipment is sold with the understanding that the General Electric Company, Medical Systems Group, its
agents, and representatives have no responsibility for injury or damage which may result from improper use of the
equipment.

Various protective material and devices are available. It is urged that such materials or devices be used.

CERTIFIED ELECTRICAL CONTRACTOR STATEMENT

All electrical installations that are preliminary to positioning of the equipment at the site prepared for the equipment
shall be performed by licensed electrical contractors. In addition, electrical feeds into the Power Distribution Unit
shall be performed by licensed electrical contractors. Other connections between pieces of electrical equipment,
calibrations, and testing shall be performed by qualified GE Medical personnel. The products involved (and the
accompanying electrical installations) are highly sophisticated, and special engineering competence is required.

In performing all electrical work on these products, GE will use its own specially trained field engineers. All of GE’s
electrical work on these products will comply with the requirements of the applicable electrical codes.

The purchaser of GE equipment shall only utilize qualified personnel (i.e., GE’s field engineers, personnel of third–
party service companies with equivalent training, or licensed electricians) to perform electrical servicing on the
equipment.

c
HiSpeed Dual, ProSpeed FII Series
GE MEDICAL SYSTEMS ADVANCED INSTALLATION
REV 1 5116386

DAMAGE IN TRANSPORTATION
All packages should be closely examined at time of delivery. If damage is apparent, have notation “damage in
shipment” written on all copies of the freight or express bill before delivery is accepted or “signed for” by a Gen-
eral Electric representative or a hospital receiving agent. Whether noted or concealed, damage MUST be reported
to the carrier immediately upon discovery, or in any event, within 14 days after receipt, and the contents and con-
tainers held for inspection by the carrier. A transportation company will not pay a claim for damage if an inspection
is not requested within this 14 day period.

Call Traffic and Transportation, Beijing, China, 800-810-8188 immediately after damage is found. At this time be
ready to supply name of carrier, delivery date, consignee name, freight or express bill number, item damaged and
extent of damage.

Complete instructions regarding claim procedure are found in Section “S” of the Policy & Procedure Bulletins.

OMISSIONS & ERRORS


GE personnel, please use the GEMS CQA Process to report all omissions, errors, and defects in this documenta-
tion. Customers, please contact your GE Sales or Service represenatives.

CAUTION
Do not use the following devices near this equipment. Use of these devices near this equipment could cause
this equipment to malfunction.

Devices not to be used near this equipment:

Devices which intrinsically transmit radio waves such as; cellular phone, radio transceiver, mobile radio transmit-
ter, radio-controlled toy, etc.

Keep power to these devices turned off when near this equipment.

Medical staff in charge of this equipment is required to instruct technicians, patients and other people who may be
around this equipment to fully comply with the above regulation.

d
HiSpeed Dual, ProSpeed FII Series
GE MEDICAL SYSTEMS ADVANCED INSTALLATION
REV 1 5116386

LIST OF EFFECTIVE PAGES

Tab - Section Page REV (Revision Information) A to B 1

Title page 1 (Table of Contents) i 1


-
Title page rear blank
1 (Installation Procedure) 1-1 to 1-26 1
(Warning and other
important information)
a to d 1 2 (Other Report) 2-1 to 2-5 1

- Blank/Rear cover -

A
HiSpeed Dual, ProSpeed FII Series
GE MEDICAL SYSTEMS ADVANCED INSTALLATION
REV 1 5116386

REVISION HISTORY
REV Date Primary Reason for Change
1 10/15/04 Initial Release.

B
HiSpeed Dual, ProSpeed FII Series
GE MEDICAL SYSTEMS ADVANCED INSTALLATION
REV 1 5116386

TABLE OF CONTENTS

SECTION PAGE

SECTION 1 - INSTALLATION PROCEDURE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1


1-1 POSITIONING SUBSYSTEMS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
1-2 CONNECTING POWER, GROUND, AND SIGNAL CABLES . . . . . . . . . . . . . . . . . . . 1-4
1-3 SYSTEM POWER-UP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-5
1-4 LEVELING OF GANTRY AND TABLE AND EMERGENCY OFF TEST . . . . . . . . . . . 1-6
1-5 OC INTEGRATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-7
1-6 GANTRY AND TABLE INTEGRATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-8
1-7 SYSTEM TEST . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-19
1-7-1 Image Performance Test Check Sheet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-23
1-7-2 Network Data Sheet. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-26

SECTION 2 - OTHER REPORT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1


2–1 TOOLS AND TEST EQUIPMENTS REQUIRED FOR CALIBRATION . . . . . . . . . . . . 2-1
2–2 OPTION INSTALLATION REPORT. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
2–3 INSTALLATION TIME TABLE (ACTUAL TIME) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3
2–4 TOTAL MAN-HOUR. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-4
2–5 COMPLETION CHECK SHEET. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-5

i
HiSpeed Dual, ProSpeed FII Series
GE MEDICAL SYSTEMS ADVANCED INSTALLATION
blank

ii
HiSpeed Dual, ProSpeed FII Series
GE MEDICAL SYSTEMS ADVANCED INSTALLATION
REV 1 5116386

SECTION 1 - INSTALLATION PROCEDURE

Note
The check that performs X-ray exposure is filled in gray.

1-1 POSITIONING SUMSYSTEMS

Item Specifications Result X-Ray ?


1. Site Construction Check:
Check the site to verify the completion of duct work, raceway, runs, contractor - OK NG No
installed wiring, and telephone extension in the control room.
2. Tools and Test Equipment Check:
Prepare the tools and test equipments required for installation. - OK NG No
Especially the System software CD-ROM must be prepared.
3. Network Data Check:
If the system is connected to the network, fill network data sheet, section 1–7– - OK NG No
2, with Site network information.
4. Environment Check Scan Room:
Scan Room:
20 ~ 28°C / 30 ~ 70%RH ________°C, ________%RH
No
Operation Room: Operation Room:
15 ~ 30°C / 20 ~ 80%RH
________°C, ________%RH
5. PDB Voltage Check: Line Voltage (VAC):
Monitor mean voltage, surges, sags, short term impulses, and frequency drift 200, 380, 400, 415,
Live Voltage:
at the Power Distribution Box (on terminals L1, L2, and L3) (hospital supply). 440, 460, or 480
________________ VAC
Daily Voltage Varia-
tion (%): Daily Voltage Variation:
No
+10~-5 (50Hz) ________________ % (_____Hz)
+6~-5 (60Hz) Frequency:
Frequency (Hz): ________________ Hz
50 or 60 ± 2

1-1
HiSpeed Dual, ProSpeed FII Series
GE MEDICAL SYSTEMS ADVANCED INSTALLATION
REV 1 5116386

1-1 POSITIONING SUBSYSTEMS (CONTINUED)

Item Specifications Result X-Ray ?


6. Accessory Check:
- OK NG No
Check if all accessories to be required have been shipped with the system.
7. Visually check if the all sub systems are damaged. - OK NG No
8. Sub System Positioning:
- OK NG No
Place each sub system to a proper position according to site planning.
9. Remove ALL Gantry covers. - OK NG No
10. Gantry Slip Ring Cleaning:
Clean the slip rings and ring grooves using Kim Wipe with ethyl alcohol at a - OK NG No
pure of 99.5%.
11. Gantry Brush Block Installation: (ETC slip ring ONLY. For HiSpeed Dual
48 kg-cm (M6 screw)
system, skip this step.) OK NG No
20 kg-cm (M4 screw)
Install the Brush Block to the Gantry.
12. Gantry Tilt Locks:
- OK NG No
Remove the four Gantry tilt locks from the Gantry.
13. Gantry Azimuth Locking Pin:
- OK NG No
Disengage the azimuth locking pin of the Gantry.
14. Table and Gantry Connection:
Remove the table dolly, then connect the table base to the Gantry base using
- OK NG No
the table mounting bolts securely.
Connect the joint valves of the Table and Gantry securely.
15. Remove the OC front and rear covers. - OK NG No
16. OC SCSI Bay Shipping Lock (for small OC):
- OK NG No
Remove the OC right side cover. Then, remove the SCSI bay shipping lock.

1-2
HiSpeed Dual, ProSpeed FII Series
GE MEDICAL SYSTEMS ADVANCED INSTALLATION
REV 1 5116386

1-1 POSITIONING SUBSYSTEMS (CONTINUED)

Item Specifications Result X-Ray ?


17. OC Installation:
Install the following to the OC.
• CRT - OK NG No
• Keyboard
• Mouse
18. PDU, Setting Coupling Terminal for Input Voltage:
Remove the rear cover of PDU2 (2298852 / 2298849).
Check the supplied voltage at Power Distribution Box (refer to step 5. of sec-
Setting Voltage (VAC): PDU setting:
tion 1-1, Positioning Subsystem).
200, 380, 400, 415, 440, No
Then set the proper PDU input rating voltage by changing the voltage selec-
460, or 480 ________________ VAC
tion cable at the transformer in the PDU.
Note: Regarding the voltage, there are two types of PDUs:
one for 480V~380V sytem, the other for 200V system.

1-3
HiSpeed Dual, ProSpeed FII Series
GE MEDICAL SYSTEMS ADVANCED INSTALLATION
REV 1 5116386

1-2 CONNECTING POWER, GROUND, AND SIGNAL CABLES

Item Specifications Result X-Ray ?


1. Power and Ground Cable Connection:
Should be connected
Connect the cables to the proper connectors labeled on the cable. Route the OK NG No
securely and properly.
PDU cables through proper cable hole.
2. Signal Cable Connection:
Should be connected
Connect the cables to the proper connectors labeled on the cable. Route the OK NG No
securely and properly.
PDU cables through proper cable hole.
3. Ground Continuity Check:
Switch OFF all the breakers of the PDB, measure the DC resistance between
the following components in the system by using an earth bonding test meter
or equivalent device.
• GND (PDB) - GND (PDU) Less than 0.1 Ω OK NG No
• GND (PDU) - GND (Gantry)
• GND (Gantry) - GND (Table)
• GND (PDU) - GND (OC)

1-4
HiSpeed Dual, ProSpeed FII Series
GE MEDICAL SYSTEMS ADVANCED INSTALLATION
REV 1 5116386

1-3 SYSTEM POWER-UP

Item Specifications Result X-Ray ?


1. Power-up Test:
Switch ON all the breakers of the PDB, and switch ON the OC to verify that - OK NG No
the system starts up with no error.
2. Grade Key Installation:
Install the system grade key DVD-RAM.

Select Service -> Utilities -> Shutdown -> Application Shutdown. Click - OK NG No
on Install Option, then insert the Grade key DVD-RAM. Select Options to be
installed using Available Options box. Select Install -> Quit -> OK -> Star-
tup.
3. Option(s) Installation:
Install the option DVD-RAM.

Select Service -> Utilities -> Shutdown -> Application Shutdown. Click - OK NG No
on Install Option, then insert the Option DVD-RAM. Select Options to be
installed using Available Options box. Select Install -> Quit -> OK -> Star-
tup.

1-5
HiSpeed Dual, ProSpeed FII Series
GE MEDICAL SYSTEMS ADVANCED INSTALLATION
REV 1 5116386

1-4 LEVELING OF GANTRY AND TABLE AND EMERGENCY OFF TEST

Item Specifications Result X-Ray ?


1. Tilt Roller Clearance Check:
Verify that the four tilt rollers of the Gantry should NOT be rotated by hand.
If some rollers are rotated, the clearance adjustment must be performed to
eliminate clearance beteen the tilt rollers and tilt arms:
• Remove the Gantry right lower bottom cover.
• Install the Zero degree adjustment tools.
No Clearance OK NG No
• Tilt the Gantry all the way BWD.
• Loosen the right/front tilt roller mounting nut.
• Swing the right/front tilt roller upward by hex wrench until it touches the
tilt arm.
• While holding the tilt roller using a hex wrench, tighten the tilt roller
mounting nut securely.
2. To mount the table to the Gantry base, hard-tighten the two table mounting
- OK NG No
bolts / washers with the cable support plate.
3. Gantry Leveling:
Place the bubble level at the left, right, and rear positions on the Gantry base - OK NG No
to level the Gantry using the four Gantry height adjusters.
4. Table Leveling:
Place the bubble level at the front, right, and rear positions on the table base - OK NG No
to level the Table using the four Table height adjusters.
5. Emergency OFF Test: Emergency Stop
Press the emergency stop switch of the OC keyboard and verify that table Switch Enable: The
and Gantry can NOT move. Then press the reset switch of the TGP board or table and Gantry can
OK NG No
the keyboard and verify that table and Gantry operates normally. NOT move.
Check all the emergency stop switches of the Gantry and PDU for proper Reset: The table and
operation above. Gantry move normally.

1-6
HiSpeed Dual, ProSpeed FII Series
GE MEDICAL SYSTEMS ADVANCED INSTALLATION
REV 1 5116386

1-5 OC INTEGRATION

Item Specifications Result X-Ray ?


1. Monitor Adjustment:
Display SMPTE pattern on the monitor and enlarge the image to full screen. - OK NG No
Then adjust the contrast value to the level an user desires.
2. Time and Date Check:
Select Service -> System Func. Check -> Generator Test -> Set JEDI time
- OK NG No
to verify that time and date info. is correct. If it is not, set them to the correct
one.
3. Laser Imager Installation (Option) - OK NG No
4. Advantage Workstation Installation (Option) - OK NG No
5. Connecting the CT System to the network:
• Connect the 10/100/1000Base - T cable to the Ethernet port of Host
computer.
• Network Setting:
Select Service -> Utilities -> Shutdown -> Application Shutdown.
Click on Reconfig to enter System Configuration screen.
Select Network Setting to enter Network Setting screen.
Select “Network on”, then enter all parameters for OC parameters. - OK NG No
If CT is not connected to site subnet, OC Net Mask and Gateway
parameters are set to the default. If connected, enter all parameters for
Gateway Parameters.
Select Accept -> Yes. Then the system restart automatically.
• Remote Host Setting:
Select ImageWorks -> Network -> Select remote host -> Add.
Enter the parameters, then click on Save.

1-7
HiSpeed Dual, ProSpeed FII Series
GE MEDICAL SYSTEMS ADVANCED INSTALLATION
REV 1 5116386

1-6 GANTRY AND TABLE INTEGRATION

Item Specifications Result X-Ray ?


1. Gantry Rotation Velocity Offset Adjustment:
1. Set the G6 Gantry dip switch (switch S2) to ON on the TGP Board.
2. Push the TEST1 switch to enter TGP off-line function #400 (Gantry
azimuth position display mode).
3. Set the TGP service switches to SVE/CONT/MNL.
4. Verify that the seven segment display of the Servo Amp shows 0 (Servo Time from the first
ON). change to next change
5. Push TEST2 switch on the TGP. on Gantry Position Dis- OK NG No
play : more than 4 sec-
6. Wait until Right most digit of the Position Display stops to blink.
onds
7. Measure the time from the first change to next change on Gantry Position
display. If it is less than 4 seconds, go to step 5. Otherwise proceed to
next step.
8. If CHANGE LED lits on, press switches FAST and RANGE together in
order to store new adjustment data.
9. Verify that CHANGE LED is OFF.

1-8
HiSpeed Dual, ProSpeed FII Series
GE MEDICAL SYSTEMS ADVANCED INSTALLATION
REV 1 5116386

1-6 GANTRY AND TABLE INTEGRATION (CONTINUED)

Item Specifications Result X-Ray ?


2. Gantry Rotation Speed (Velocity Gain) Adjustment:
1. Set the TGP service switch SYS/OFF/MNL to OFF.
2. Enter the TGP off-line function #401 (Gantry rotation speed display mode):
a. Set the G6 Gantry dip switch (switch S2) to ON on the TGP Board.
b. Push the TGP reset switch. The Gantry rotates to the home position, and
the Gantry Display panel displays a test number 400.
c. Press the test switch once on the TGP Board. The Gantry Display will
change from 400 to 401.
(401: Gantry rotation speed display mode)
3. Rotate the Gantry at a specified speed and check/adjust the Gantry rotation Gantry Position Display: OK NG No
speed displayed on the Gantry Display: 1000 ±5
a. Set the TGP service switches to FIX/SPD2/SLOW/CONT/MNL to rotate
the Gantry (1.0 sec/Rot).
b. After Gantry rotation is stabilized (a few rotations are required), push
TEST2 switch.
4. Check if the rotation time shown in the Position Display shows a value 1000
±5. If not, return to step 3-b. If it is in the specification, proceed to next step.
5. If CHANGE LED lits on, press switches FAST and RANGE together in order
to store new adjustment data.
6. Verify that CHANGE LED is OFF.

1-9
HiSpeed Dual, ProSpeed FII Series
GE MEDICAL SYSTEMS ADVANCED INSTALLATION
REV 1 5116386

1-6 GANTRY AND TABLE INTEGRATION (CONTINUED)

Item Specifications Result X-Ray ?


3. Table Height Down Limit Adjustment:
1. Move the Cradle to the out limit position.
2. Set the T3 switch of the TGP S1 dip-switch to ON position.
3. Lower the Table to the mechanical down limit position.
4. Set the T6 switch of the TGP S1 dip-switch to ON position.
5. Push TEST1 switch on the TGP board, until the height display on the
Gantry Display indicates ’009’. Gantry Display: 1 OK NG No
6. Check the value of position display indicates ’1’.
If not, adjust the position display to ’1’ by pressing the ’INLM (Internal
Landmark)’ or ’EXLM (External Landmark)’ keys.
7. Push TEST2 switch on the TGP board.
If CHANGE LED is lit, press FAST and RANGE keys together on the
Gantry Switch Panel.
Check that CHANGE LED is OFF.

1-10
HiSpeed Dual, ProSpeed FII Series
GE MEDICAL SYSTEMS ADVANCED INSTALLATION
REV 1 5116386

1-6 GANTRY AND TABLE INTEGRATION (CONTINUED)

Item Specifications Result X-Ray ?


4. Table Height UP Limit Adjustment:
1. Raise the Table to the mechanical up limit position.
2. Set the T6 switch of the TGP S1 dip-switch to ON position.
3. Push TEST1 switch on the TGP board, until the height display on the
Gantry Display indicates ’010’.
4. Gantry Display:
OK NG No
- For LT Table (2320315): 1177, for LC Table (2320315)
Check the value of position display indicates ’1177’.
5. Push TEST2 switch on the TGP board.
If CHANGE LED is lit, press FAST and RANGE keys together on the
Gantry Switch Panel.
Check that CHANGE LED is OFF.
6. Set the T3 and T6 switch of the TGP S1 dip-switch to OFF position.

1-11
HiSpeed Dual, ProSpeed FII Series
GE MEDICAL SYSTEMS ADVANCED INSTALLATION
REV 1 5116386

1-6 GANTRY AND TABLE INTEGRATION (CONTINUED)

Item Specifications Result X-Ray ?


5. Table Cradle Absolute Position (OUT Limit Adjustment):
1. Set the Gantry to the 0 degree Tilt position.
2. Set the table to: ‘100’mm (for LC Table 2320315) height position (shown
on Gantry display).
3. Move the cradle to the mechanical out limit position by hand.
4. Set the T6 switch of the TGP S1 dip-switch to ON. Gantry Display: 0 OK NG No
5. Push TEST1 switch on the TGP board, until the height display on the
Gantry Display indicates ’003’.
6. Verify that the position display indicates ’0’, then push TEST2 switch on
the TGP board.
If the CHANGE LED is lit, press the RANGE and FAST keys together on
the Gantry Switch panel.
Verify that the CHANGE LED is OFF.
6. Table Cradle Absolute Position (IN Limit Adjustment):
1. Verify that the cradle is still at mechanical out limit position.
2. Push TEST1 switch on the TGP board, until the height display on the
Gantry Display reads ’004’.
3. Verify that the position display indicates 0.
Then move the cradle IN by pushing IN button until the position display
indicates ’3040 ± 10’.
Gantry Display:
If cradle stopped due to interlock and the position display indicated OK NG No
smaller than ì3040î, move the cradle IN until the position display indicates 3040 ± 10
’3040 ±10’ by hand.
4. Push TEST2 switch on the TGP board.

If CHANGE LED is lit, press FAST and RANGE keys together on the
Gantry Switch panel.
Verify that the CHANGE LED is OFF.
5. Set the T6 switch of the TGP S1 dip-switch to OFF position.

1-12
HiSpeed Dual, ProSpeed FII Series
GE MEDICAL SYSTEMS ADVANCED INSTALLATION
REV 1 5116386

1-6 GANTRY AND TABLE INTEGRATION (CONTINUED)

Item Specifications Result X-Ray ?


7. Cradle Compensation:
1. Raise/lower the Table to 50mm height (shown on Gantry display).
2. Move the cradle IN approx. 400mm from the out limit.
3. Push the positioning light switch and mark the position on the cradle
where is hit by positioning light beam.
4. Set the T6 switch of the TGP S1 dip-switch to ON.
5. Push TEST1 switch on the TGP board until the height display on the
Gantry Display indicates ’012’ .
- OK NG No
6. Push TEST2 switch on the TGP board.
7. Lower the Table approx. 150mm (using a measure).
8. Push the positioning light switch again, and move the cradle until the
mark made in step 3 is hit by positioning light beam.
9. Push TEST2 switch on the TGP board.
If CHANGE LED is lit, press FAST and RANGE keys together on the
Gantry Switch panel.
Verify that CHANGE LED is OFF.
10. Set the T6 switch of the TGP S1 dip-switch to OFF position.

1-13
HiSpeed Dual, ProSpeed FII Series
GE MEDICAL SYSTEMS ADVANCED INSTALLATION
REV 1 5116386

1-6 GANTRY AND TABLE INTEGRATION (CONTINUED)

Item Specifications Result X-Ray ?


8. Height Compensation between Table and Gantry:
1. Raise the Table up to 50 mm ~ 0 mm down from the Up limit position
(shown on Gantry display).
2. Perform an axial scan
a. Advance the cradle into the Gantry so that the position of 20 cm ~ 40
cm from the cradle edge can be scanned.
b. Perform an axial scan of the cradle.
3. Measure distance (X) between scan center and cradle upper surface.
4. Set the T6 switch of the TGP S1 dip-switch to ON position. Gantry Display:
1000 + 2X
5. Push TEST1 switch on the TGP board until the height display on the
(X : distance between OK NG Yes
Gantry Display indicates ’013’.
scan center and cra-
6. Adjust the position display XXXX to {1000 + 2X} by pressing the ’INLM’ dle upper surface)
or ’EXLM’ keys.
e.g.: When the cradle is positioned at 10mm above the scan center,
adjust display to1020 (1000 + 2x10).
When the cradle is positioned at 10mm under the scan center, adjust
display to 980 (1000 + 2x (-10)).
7. Push TEST2 switch on the TGP board.
If CHANGE LED is lit, press FAST and RANGE keys together on the
Gantry Switch Panel.
Check that CHANGE LED is OFF.
8. Set the T6 switch of the TGP S1 dip-switch to OFF position.

1-14
HiSpeed Dual, ProSpeed FII Series
GE MEDICAL SYSTEMS ADVANCED INSTALLATION
REV 1 5116386

1-6 GANTRY AND TABLE INTEGRATION (CONTINUED)

Item Specifications Result X-Ray ?


9. Table Height DOWN Limit Adjustment for Short Foot Print mode:
1. Move the Cradle to the out electrical limit position.
2. Lower the Table to the lowest position desired.
3. Set the T6 switch of the TGP S1 dip-switch to ON position.
4. Push TEST1 switch on the TGP board until the height display on the
Gantry Display indicates ’035’.
5. Set the T2 switch of the TGP S1 dip-switch to ON position.
6. Push TEST2 switch on the TGP board. - OK NG No
If CHANGE LED is lit, press FAST and RANGE keys together on the
Gantry Switch Panel.
Check that CHANGE LED is OFF.
7. Set the T6 switch of the TGP S1 dip-switch to OFF position. ( Do NOT
touch the T2 switch )
8. Press the Reset switch to reset the TGP board.
9. Move the table up and down to verify that the Table lowest position is set
to the desired one.

1-15
HiSpeed Dual, ProSpeed FII Series
GE MEDICAL SYSTEMS ADVANCED INSTALLATION
REV 1 5116386

1-6 GANTRY AND TABLE INTEGRATION (CONTINUED)

Item Specifications Result X-Ray ?


10. Cradle IN Limit Adjustment for Short Foot Print mode:
1. Move the Table to the highest position.
2. Move the Cradle to the IN-limit position desired.
3. Set the T6 switch of the TGP S1 dip-switch to ON position.
4. Push TEST1 switch on the TGP board until the height display on the
Gantry Display indicates ’036’.
5. Set the T4 switch of the TGP S1 dip-switch to ON position.
6. Push TEST2 switch on the TGP board. - OK NG No
If CHANGE LED is lit, press FAST and RANGE keys together on the
Gantry Switch Panel.
Check that CHANGE LED is OFF.
7. Set the T6 switch of the TGP S1 dip-switch to OFF position. ( Do NOT
touch the T4 switch )
8. Press the Reset switch to reset the TGP board.
9. Move the cradle OUT and IN to verify that Cradle IN-limit position is set
to the desired one.

1-16
HiSpeed Dual, ProSpeed FII Series
GE MEDICAL SYSTEMS ADVANCED INSTALLATION
REV 1 5116386

1-6 GANTRY AND TABLE INTEGRATION (CONTINUED)

Item Specifications Result X-Ray ?


11. Tilt Angle (FWD) Adjustment:
1. Set Dip Switch T6 on the TGP board to ON.
2. Push TEST1 switch on the TGP board to select test program 000 (height
display of the Gantry display panel shows 000).
3. Tilt the Gantry forward to +20 degree position (by tilt mark on the tilt arm).
4. Set the position display to 5 using INTERNAL LANDMARK and Gantry Display : 5 OK NG No
EXTERNAL LANDMARK keys on the Gantry Switch Panel.
5. Push TEST2 switch on the TGP board.
6. If CHANGE LED is lit, press FAST and RANGE keys (Gantry Switch
Panel) together. (If CHANGE LED is not lit, adjustment is not needed).
7. Verify that CHANGE LED is OFF.
12. Tilt Angle (BWD) Adjustment:
1. Push TEST1 switch on the TGP board to select test program 001 (height
display of the Gantry display panel shows 001).
2. Tilt the Gantry backward to -20 degree position (by tilt mark on the tilt
arm).
3. Set the position display to 405 using INTERNAL LANDMARK and
EXTERNAL LANDMARK keys on the Gantry Switch Panel. Gantry Display : 405 OK NG No

4. Push TEST2 switch on the TGP board.


5. If CHANGE LED is lit, press FAST and RANGE keys (Gantry Switch
Panel) together. (If CHANGE LED is not lit, adjustment is not needed).
6. Verify that CHANGE LED is OFF.
7. Set Dip Switch T6 on the TGP board to OFF.

1-17
HiSpeed Dual, ProSpeed FII Series
GE MEDICAL SYSTEMS ADVANCED INSTALLATION
REV 1 5116386

1-6 GANTRY AND TABLE INTEGRATION (CONTINUED)

Item Specifications Result X-Ray ?


13. Table to Gantry Alignment:
1. Perform an axial scan of the head holder:
a. Mount the axial head holder to the end of the cradle.
b. Switch ON the positioning lights.
c. Advance the cradle into the Gantry until the internal light cuts the
approximately middle section of the head holder.
d. Perform an axial scan of the head holder using the following Distance between cen-
technique; 35 cm FOV, 3 sec, 120 kV, 60 mA, 10 mm Thickness ters of the head holder
OK NG Yes
2. Check whether the Table center line and the scan center are aligned: and image :
Verify that the difference in the distance between centers of the head Within 1 mm
holder and image is within 1 mm. If not, proceed to next step.
3. Adjust the Table right - left position:
a. Loosen the two Table base mounting nuts.
b. Move the Table to center the head holder in the scan plane.
c. Tighten the Table base mounting nuts securely.
d. Return to step 1.
14. Anchoring of Gantry and Table: - OK NG No
(4 pcs for Gantry, 4 pcs for Table)
15. Anchoring of OC and PDU (Option): - OK NG No
(2 pcs for OC, 2 pcs for PDU)
16. Fix OC and PDU using their adjusters. - OK NG No

1-18
HiSpeed Dual, ProSpeed FII Series
GE MEDICAL SYSTEMS ADVANCED INSTALLATION
REV 1 5116386

1-7 SYSTEM TEST

Item Specifications Result X-Ray ?


1. Positioning Light Check - OK NG No
2. Intercommunication Test:
Press Talk (intercom) button on the OC keyboard to verify that your voice can - OK NG No
be heard from the speaker of the table.
3. Table Latch Check:
1. Move the cradle to the Out limit position by the ’OUT’ button.
2. While moving the cradle in the In direction by continuously pushing the
’IN’ button, perform the following:
a. Push the Unlatch switch (’Free’).
Verify that the cradle stops moving, and that the Gantry Display goes
blank; this indicating that the cradle is unlatched. - OK NG No
b. Push the Unlatch switch again (’Latch’).
c. Release the ëINí button, and then push it.
Verify that the cradle starts moving and the Gantry Display changes
to normal (displaying a position in millimeter).
3. Repeat the above same check for the Unlatch switch on the opposite
side.

1-19
HiSpeed Dual, ProSpeed FII Series
GE MEDICAL SYSTEMS ADVANCED INSTALLATION
REV 1 5116386

1-7 SYSTEM TEST (CONTINUED)

Item Specifications Result X-Ray ?


4. System Reconfiguration:
Select Service -> System Tools -> Application Shutdown -> OK -> Recon-
fig. After reconfiguration. Select Accept.
At least the followings must be configured:
• System setting:
Hospital Name, Time Zone, and Machine #
• Preference setting:
Units for patient weight, Date Format, Auto mA
• Language setting: - OK NG No
Language and Default Autovoice Language
• Hardware setting:
Filter type, Start angle for Head scan, Keyboard, Start angle for Head
scan, and Diagnosis setting
• Camera Setup (Option)
• Network setting (if required)
• InSite setting (if required)

1-20
HiSpeed Dual, ProSpeed FII Series
GE MEDICAL SYSTEMS ADVANCED INSTALLATION
REV 1 5116386

1-7 SYSTEM TEST (CONTINUED)

Item Specifications Result X-Ray ?


5. Tube Alignment:
• Plane of Rotation
• Detector Beam on Window - OK NG Yes
• ISO Center Alignment
• Filter Center Alignment
6. Calibration:
Perform the followings in this order. Use Service menu -> Service Adjust-
ment -> Service Calibration.
• Q-cal
• XT-cal
- OK NG Yes
• AV-cal
• DG-cal
• Air calibration
• Phantom calibration
7. CT Number Adjustment: 0 ± 3 (water)
Use Service menu -> Service Adjustment -> CT Number Adjustment. OK NG Yes
-1000 ± 5 (air)
8. Image Performance Check:
1. Select Abdomen, then center the phantom. See list in
OK NG Yes
section 1-7-1.
2. Perform scans listed in 1-7-1 Image Performance Check Sheet, and
positron a circular ROI and write down the data on this sheet.
9. Scan Test: Scans complete with
OK NG Yes
Perform Warm up scan, Scoutview scan, and Axial scan. no error.
10. Recording Slice Number:
- _________________Slice No
Slice counter location: upper counter next to sub board

1-21
HiSpeed Dual, ProSpeed FII Series
GE MEDICAL SYSTEMS ADVANCED INSTALLATION
REV 1 5116386

1-7 SYSTEM TEST (CONTINUED)

Item Specifications Result X-Ray ?


11. Recording Gantry Rotation Number
Rotation counter location: lower counter next to sub board - No
__________________Rotation
12. Gantry, Table, and OC cover Installation - OK NG No
13. Gantry Tilt Speed Adjustment:
1. Remove the table front bottom cover.
2. Rotate both FWD and BWD adjusters of the hydraulic valve adjuster all 37 +/- 3 sec.
the way clockwise (S limit position). (BWD20 -> FWD20)
OK NG No
3. Mesure the time of movement between BWD 20 deg. and FWD 20 deg. 37 +/- 3 sec.
4. Rotate adjusters so that the tilt speed is adjusted within specifications. To (FWD20 -> BWD20)
faster speed, rotate them to the F direction. (The FWD adjuster must be
rotated by one scale interval. The BWD must be rotated by 1/5 scale
interval.
14. Accessory Installation - OK NG No
15. Attaching the Labels and Emblems:
• Data Plate to PDB
- OK NG No
• OC Warning Label to Keyboard
• Gantry Emblem to GANTRY
16. Tube Usage File:
Select Service Menu -> Utility -> Update System Log to enter each param- - OK NG No
eter, then click on Create to create the tube usage file.
17. System Data Back-up:
1. Insert a MOD into the DVD drive of the SCSI Tower.
- OK NG No
2. Select Service -> Utilities -> savestate.
3. Click on Continue several times.
18. System and Room Cleaning - OK NG No

1-22
HiSpeed Dual, ProSpeed FII Series
GE MEDICAL SYSTEMS ADVANCED INSTALLATION
REV 1 5116386

1-7-1 Image Performance Test Check Sheet

For CT/e Dual, ProSpeed AII/FII Series


PTM FOV*1 KV TIME MA THK Mode SPECIFICATION DATA Description

(cm) (sec) (mm) 1i mode 2i mode


A B |Delta SD
(A–B)| ratio
MC SD MC SD MC SD (A/B) MC SD

42 cm PP Large 120 1.5 170 10 2i No Rings/


42 cm PP Large 120 1.5 170 10 1i Bands checked
42 cm PP Large 120 1.5 120 1 2i visually
< 3.5
25 cm WTR Large 120 1 200 10 2i 0±2 < 3.5 0±2 5 1 ±0.1
25 cm WTR Large 120 1 200 10 1i 0±2 < 3.15 -

< 4.0
25 cm WTR Largel 120 1.5 250 7 2i 0±2 < 4.0 0±2 5 1 ±0.1
25 cm WTR Large 140 1.5 160 10 2i 0±2 0±2 5

25 cm WTR Largel 140 1 160 7 2i 0±2 0±2 5

< 5.0
25 cm WTR Small 120 1 200 10 2i 0±2 < 5.0 0±2 5 1 ±0.1
25 cm WTR Small 120 1 200 10 1i 0±2 < 5.0

25 cm WTR Large 120 1 60 10 2i 0±2 0±2 5

25 cm WTR Large 120 1 80 7 2i 0±2 0±2 5

25 cm WTR Large 120 1 100 5 2i 0±2 0±2 5

25 cm WTR Large 120 1 120 3 2i 0±2 0±2 5

*1: Large: 43 cm Small: 25 cm

1-23
HiSpeed Dual, ProSpeed FII Series
GE MEDICAL SYSTEMS ADVANCED INSTALLATION
REV 1 5116386

1-7-1 Image Performance Test Check Sheet (Continued)

For CT/e Dual, ProSpeed AII/FII Series


PTM FOV*1 KV TIME MA THK Mode SPECIFICATION DATA Description

(cm) (sec) (mm) 1i mode 2i mode


A B |Delta SD
(A–B)| ratio
MC SD MC SD MC SD (A/B) MC SD

25 cm WTR Large 120 1 160 1 2i 0±3 0±3 5

25 cm WTR Small 120 1 60 10 2i 0±2 0±2 5

25 cm WTR Small 120 1 80 7 2i 0±2 0±2 5

25 cm WTR Small 120 1 100 5 2i 0±2 0±2 5

25 cm WTR Small 120 1 120 3 2i 0±2 0±2 5

25 cm WTR Smalll 120 1 140 2 2i 0±3 0±3 5

25 cm WTR Small 120 1 160 1 2i 0±3 0±3 5

25 cm WTR Largel 140 1 60 10 1i 0±2


25 cm WTR Largel 140 1 100 6 1i 0±2
25 cm WTR Large 140 1 120 4 1i 0±2
25 cm WTR Large 140 1 160 2 1i 0±3

25 cm WTR Small 140 1 60 10 1i 0±2


25 cm WTR Small 140 1 100 6 1i 0±2
25 cm WTR Small 140 1 120 4 1i 0±2
25 cm WTR Small 140 1 160 2 1i 0±3

*1: Large: 43 cm Small: 25 cm

1-24
HiSpeed Dual, ProSpeed FII Series
GE MEDICAL SYSTEMS ADVANCED INSTALLATION
REV 1 5116386

1-7-1 Image Performance Test Check Sheet (Continued)

ROI Circle
(Area of 674.11 mm2 or
+Y 1994.30 mm2)
Image

-X +X
Mc (0.0 cm) Mo (10.0 cm)
SD (0.0 cm)

-Y

1-25
HiSpeed Dual, ProSpeed FII Series
GE MEDICAL SYSTEMS ADVANCED INSTALLATION
REV 1 5116386

1-7-2 Network Data Sheet

Site Name: Address: Network Administrator: FE name:


Site Contact:
Phone: City: Phone: Phone:
State/ZIP: Site Notes: Note:

Product: Product: Product:


Host Name: Host Name: Host Name:
IP Address: IP Address: IP Address:
Netmask: Netmask: Netmask:
AE title: AE title: AE title:

Product: CT HiSpeed Product: Product: Product:


Host Name: Host Name: Host Name: Host Name:
IP Address: IP Address: IP Address: IP Address:
Netmask: Netmask: Netmask: Netmask:
AE title: AE title: AE title: AE title:

DASM type: Laser Camera


Manufacturer:
VDB
or LCAM Model Name:

1-26
HiSpeed Dual, ProSpeed FII Series
GE MEDICAL SYSTEMS ADVANCED INSTALLATION
REV 1 5116386

SECTION 2 - OTHER REPORT

2-1 TOOLS AND TEST EQUIPMENTS REQUIRED FOR CALIBRATION


Write down in the table below the tools and test equipments required for calibration among ones used in this Installation.

Item Manufacturer / Model Name Serial No. Valid until:


(Year/Month)
1. Digital Multimeter
2. Oscilloscope
3. Dial Gauge
4. HV bleeder
5.
6.
7.
8.
9.
10.
11.
12.
13.
14.
15.
16.

2-1
HiSpeed Dual, ProSpeed FII Series
GE MEDICAL SYSTEMS ADVANCED INSTALLATION
REV 1 5116386

2-2 OPTION INSTALLATION REPORT


Write down options installed at this installation in the table below. (Slave Monitor Option is one example of description.)

Installation Date Option Name Kit Number Hour (h) Members Man-
hour
2002/03/28 Slave Monitor Option 2249058 0.5 2 1

Option Installation Total Man-hour →

2-2
HiSpeed Dual, ProSpeed FII Series
GE MEDICAL SYSTEMS ADVANCED INSTALLATION
REV 1 5116386

2-3 INSTALLATION TIME TABLE (ACTUAL TIME)

Date Job Remarks Man-Hour


(Y/M/D)
Worker 1 Worker 2 Backup OJT OJT

/ / H H H H H

/ / H H H H H

/ / H H H H H

/ / H H H H H

/ / H H H H H

/ / H H H H H

/ / H H H H H

Total of each person →


H H H H H

Installation Total Man-hour OJT Total Man-hour

H H

2-3
HiSpeed Dual, ProSpeed FII Series
GE MEDICAL SYSTEMS ADVANCED INSTALLATION
REV 1 5116386

2-4 TOTAL MAN-HOUR (SAMPLE)

Site Information
System Number: Site Name:
Address:
Phone:

Installation Total Man-Hour


Input from
Input from Installation Time Table (Section 2–3): B Note:

CAPS: A These two data (A and B) must be identical.

Hours
Hours

Other Man-Hour
For Option For Missing com- For DOA/Trouble For Siting For Safety Test For Operation
ponents explanation
Total: C

Hours Hours Hours Hours Hours Hours Hours

System Man-Hour
Calculation:

B - C = System Man-Hour
Hours

2-4
HiSpeed Dual, ProSpeed FII Series
GE MEDICAL SYSTEMS ADVANCED INSTALLATION
REV 1 5116386

2-5 COMPLETION CHECK SHEET

Questions Answer
Could you check [OK] for all items of Section 1 - Installation Procedure? Yes No
Is the proper reason described if it does not?
Is the measured value within allowable specifications for items of Section 1 - Installation Proce- Yes No
dure?
Is the proper reason described if it is not?
Could you fill all the items of Section 2 - Other Report? Yes No

When all answers of the three questions above are “Yes”, sign your name on the Installation Completion sheet below.
Date:

Chief FE’s Name:

Chief FE’s ID No.:

2-5
HiSpeed Dual, ProSpeed FII Series
GE MEDICAL SYSTEMS ADVANCED INSTALLATION

2-6
GE Medical Systems: Telex 3797371
P.O. Box 414, Milwaukee, Wisconsin 53201 U.S.A.
(Asia, Pacific, Latin America, North America)

GE Medical Systems - Europe: Telex 698626


283, rue de la Miniére, B.P. 34, 78533 Buc Cedex, France