GE MEDICAL SYSTEMS ADVANCED INSTALLATION
REV 1 5116386
b
HiSpeed Dual, ProSpeed FII Series
GE MEDICAL SYSTEMS ADVANCED INSTALLATION
REV 1 5116386
The equipment is sold with the understanding that the General Electric Company, Medical Systems Group, its
agents, and representatives have no responsibility for injury or damage which may result from improper use of the
equipment.
Various protective material and devices are available. It is urged that such materials or devices be used.
All electrical installations that are preliminary to positioning of the equipment at the site prepared for the equipment
shall be performed by licensed electrical contractors. In addition, electrical feeds into the Power Distribution Unit
shall be performed by licensed electrical contractors. Other connections between pieces of electrical equipment,
calibrations, and testing shall be performed by qualified GE Medical personnel. The products involved (and the
accompanying electrical installations) are highly sophisticated, and special engineering competence is required.
In performing all electrical work on these products, GE will use its own specially trained field engineers. All of GE’s
electrical work on these products will comply with the requirements of the applicable electrical codes.
The purchaser of GE equipment shall only utilize qualified personnel (i.e., GE’s field engineers, personnel of third–
party service companies with equivalent training, or licensed electricians) to perform electrical servicing on the
equipment.
c
HiSpeed Dual, ProSpeed FII Series
GE MEDICAL SYSTEMS ADVANCED INSTALLATION
REV 1 5116386
DAMAGE IN TRANSPORTATION
All packages should be closely examined at time of delivery. If damage is apparent, have notation “damage in
shipment” written on all copies of the freight or express bill before delivery is accepted or “signed for” by a Gen-
eral Electric representative or a hospital receiving agent. Whether noted or concealed, damage MUST be reported
to the carrier immediately upon discovery, or in any event, within 14 days after receipt, and the contents and con-
tainers held for inspection by the carrier. A transportation company will not pay a claim for damage if an inspection
is not requested within this 14 day period.
Call Traffic and Transportation, Beijing, China, 800-810-8188 immediately after damage is found. At this time be
ready to supply name of carrier, delivery date, consignee name, freight or express bill number, item damaged and
extent of damage.
Complete instructions regarding claim procedure are found in Section “S” of the Policy & Procedure Bulletins.
CAUTION
Do not use the following devices near this equipment. Use of these devices near this equipment could cause
this equipment to malfunction.
Devices which intrinsically transmit radio waves such as; cellular phone, radio transceiver, mobile radio transmit-
ter, radio-controlled toy, etc.
Keep power to these devices turned off when near this equipment.
Medical staff in charge of this equipment is required to instruct technicians, patients and other people who may be
around this equipment to fully comply with the above regulation.
d
HiSpeed Dual, ProSpeed FII Series
GE MEDICAL SYSTEMS ADVANCED INSTALLATION
REV 1 5116386
- Blank/Rear cover -
A
HiSpeed Dual, ProSpeed FII Series
GE MEDICAL SYSTEMS ADVANCED INSTALLATION
REV 1 5116386
REVISION HISTORY
REV Date Primary Reason for Change
1 10/15/04 Initial Release.
B
HiSpeed Dual, ProSpeed FII Series
GE MEDICAL SYSTEMS ADVANCED INSTALLATION
REV 1 5116386
TABLE OF CONTENTS
SECTION PAGE
i
HiSpeed Dual, ProSpeed FII Series
GE MEDICAL SYSTEMS ADVANCED INSTALLATION
blank
ii
HiSpeed Dual, ProSpeed FII Series
GE MEDICAL SYSTEMS ADVANCED INSTALLATION
REV 1 5116386
Note
The check that performs X-ray exposure is filled in gray.
1-1
HiSpeed Dual, ProSpeed FII Series
GE MEDICAL SYSTEMS ADVANCED INSTALLATION
REV 1 5116386
1-2
HiSpeed Dual, ProSpeed FII Series
GE MEDICAL SYSTEMS ADVANCED INSTALLATION
REV 1 5116386
1-3
HiSpeed Dual, ProSpeed FII Series
GE MEDICAL SYSTEMS ADVANCED INSTALLATION
REV 1 5116386
1-4
HiSpeed Dual, ProSpeed FII Series
GE MEDICAL SYSTEMS ADVANCED INSTALLATION
REV 1 5116386
Select Service -> Utilities -> Shutdown -> Application Shutdown. Click - OK NG No
on Install Option, then insert the Grade key DVD-RAM. Select Options to be
installed using Available Options box. Select Install -> Quit -> OK -> Star-
tup.
3. Option(s) Installation:
Install the option DVD-RAM.
Select Service -> Utilities -> Shutdown -> Application Shutdown. Click - OK NG No
on Install Option, then insert the Option DVD-RAM. Select Options to be
installed using Available Options box. Select Install -> Quit -> OK -> Star-
tup.
1-5
HiSpeed Dual, ProSpeed FII Series
GE MEDICAL SYSTEMS ADVANCED INSTALLATION
REV 1 5116386
1-6
HiSpeed Dual, ProSpeed FII Series
GE MEDICAL SYSTEMS ADVANCED INSTALLATION
REV 1 5116386
1-5 OC INTEGRATION
1-7
HiSpeed Dual, ProSpeed FII Series
GE MEDICAL SYSTEMS ADVANCED INSTALLATION
REV 1 5116386
1-8
HiSpeed Dual, ProSpeed FII Series
GE MEDICAL SYSTEMS ADVANCED INSTALLATION
REV 1 5116386
1-9
HiSpeed Dual, ProSpeed FII Series
GE MEDICAL SYSTEMS ADVANCED INSTALLATION
REV 1 5116386
1-10
HiSpeed Dual, ProSpeed FII Series
GE MEDICAL SYSTEMS ADVANCED INSTALLATION
REV 1 5116386
1-11
HiSpeed Dual, ProSpeed FII Series
GE MEDICAL SYSTEMS ADVANCED INSTALLATION
REV 1 5116386
If CHANGE LED is lit, press FAST and RANGE keys together on the
Gantry Switch panel.
Verify that the CHANGE LED is OFF.
5. Set the T6 switch of the TGP S1 dip-switch to OFF position.
1-12
HiSpeed Dual, ProSpeed FII Series
GE MEDICAL SYSTEMS ADVANCED INSTALLATION
REV 1 5116386
1-13
HiSpeed Dual, ProSpeed FII Series
GE MEDICAL SYSTEMS ADVANCED INSTALLATION
REV 1 5116386
1-14
HiSpeed Dual, ProSpeed FII Series
GE MEDICAL SYSTEMS ADVANCED INSTALLATION
REV 1 5116386
1-15
HiSpeed Dual, ProSpeed FII Series
GE MEDICAL SYSTEMS ADVANCED INSTALLATION
REV 1 5116386
1-16
HiSpeed Dual, ProSpeed FII Series
GE MEDICAL SYSTEMS ADVANCED INSTALLATION
REV 1 5116386
1-17
HiSpeed Dual, ProSpeed FII Series
GE MEDICAL SYSTEMS ADVANCED INSTALLATION
REV 1 5116386
1-18
HiSpeed Dual, ProSpeed FII Series
GE MEDICAL SYSTEMS ADVANCED INSTALLATION
REV 1 5116386
1-19
HiSpeed Dual, ProSpeed FII Series
GE MEDICAL SYSTEMS ADVANCED INSTALLATION
REV 1 5116386
1-20
HiSpeed Dual, ProSpeed FII Series
GE MEDICAL SYSTEMS ADVANCED INSTALLATION
REV 1 5116386
1-21
HiSpeed Dual, ProSpeed FII Series
GE MEDICAL SYSTEMS ADVANCED INSTALLATION
REV 1 5116386
1-22
HiSpeed Dual, ProSpeed FII Series
GE MEDICAL SYSTEMS ADVANCED INSTALLATION
REV 1 5116386
< 4.0
25 cm WTR Largel 120 1.5 250 7 2i 0±2 < 4.0 0±2 5 1 ±0.1
25 cm WTR Large 140 1.5 160 10 2i 0±2 0±2 5
< 5.0
25 cm WTR Small 120 1 200 10 2i 0±2 < 5.0 0±2 5 1 ±0.1
25 cm WTR Small 120 1 200 10 1i 0±2 < 5.0
1-23
HiSpeed Dual, ProSpeed FII Series
GE MEDICAL SYSTEMS ADVANCED INSTALLATION
REV 1 5116386
1-24
HiSpeed Dual, ProSpeed FII Series
GE MEDICAL SYSTEMS ADVANCED INSTALLATION
REV 1 5116386
ROI Circle
(Area of 674.11 mm2 or
+Y 1994.30 mm2)
Image
-X +X
Mc (0.0 cm) Mo (10.0 cm)
SD (0.0 cm)
-Y
1-25
HiSpeed Dual, ProSpeed FII Series
GE MEDICAL SYSTEMS ADVANCED INSTALLATION
REV 1 5116386
1-26
HiSpeed Dual, ProSpeed FII Series
GE MEDICAL SYSTEMS ADVANCED INSTALLATION
REV 1 5116386
2-1
HiSpeed Dual, ProSpeed FII Series
GE MEDICAL SYSTEMS ADVANCED INSTALLATION
REV 1 5116386
Installation Date Option Name Kit Number Hour (h) Members Man-
hour
2002/03/28 Slave Monitor Option 2249058 0.5 2 1
2-2
HiSpeed Dual, ProSpeed FII Series
GE MEDICAL SYSTEMS ADVANCED INSTALLATION
REV 1 5116386
/ / H H H H H
/ / H H H H H
/ / H H H H H
/ / H H H H H
/ / H H H H H
/ / H H H H H
/ / H H H H H
H H
2-3
HiSpeed Dual, ProSpeed FII Series
GE MEDICAL SYSTEMS ADVANCED INSTALLATION
REV 1 5116386
Site Information
System Number: Site Name:
Address:
Phone:
Hours
Hours
Other Man-Hour
For Option For Missing com- For DOA/Trouble For Siting For Safety Test For Operation
ponents explanation
Total: C
System Man-Hour
Calculation:
B - C = System Man-Hour
Hours
2-4
HiSpeed Dual, ProSpeed FII Series
GE MEDICAL SYSTEMS ADVANCED INSTALLATION
REV 1 5116386
Questions Answer
Could you check [OK] for all items of Section 1 - Installation Procedure? Yes No
Is the proper reason described if it does not?
Is the measured value within allowable specifications for items of Section 1 - Installation Proce- Yes No
dure?
Is the proper reason described if it is not?
Could you fill all the items of Section 2 - Other Report? Yes No
When all answers of the three questions above are “Yes”, sign your name on the Installation Completion sheet below.
Date:
2-5
HiSpeed Dual, ProSpeed FII Series
GE MEDICAL SYSTEMS ADVANCED INSTALLATION
2-6
GE Medical Systems: Telex 3797371
P.O. Box 414, Milwaukee, Wisconsin 53201 U.S.A.