Académique Documents
Professionnel Documents
Culture Documents
%$ #
Revision 3
!"
#$$ %
!"$ %$$
GE MEDICAL SYSTEMS HELICAL UPGRADE INSTALLATION
REV 2 2151639
a
GE MEDICAL SYSTEMS HELICAL UPGRADE INSTALLATION
REV 2 2151639
b
GE MEDICAL SYSTEMS HELICAL UPGRADE INSTALLATION
REV 2 2151639
In performing all electrical work on these products, GE will use its own specially trained field engineers. All of GE’s electrical
work on these products will comply with the requirements of the applicable electrical codes.
The purchaser of GE equipment shall only utilize qualified personnel (i.e., GE’s field engineers, personnel of third-party
service companies with equivalent training, or licensed electricians) to perform electrical servicing on the equipment.
c
GE MEDICAL SYSTEMS HELICAL UPGRADE INSTALLATION
REV 2 2151639
DAMAGE IN TRANSPORTATION
All packages should be closely examined at time of delivery. If damage is apparent, have notation “damage in
shipment” written on all copies of the freight or express bill before delivery is accepted or “signed for” by a General
Electric representative or a hospital receiving agent. Whether noted or concealed, damage MUST be reported to the
carrier immediately upon discovery, or in any event, within 14 days after receipt, and the contents and containers held
for inspection by the carrier. A transportation company will not pay a claim for damage if an inspection is not requested
within this 14 day period.
Call Traffic and Transportation, Milwaukee, WI (414) 827–3449 / 8*285–3449 immediately after damage is found. At
this time be ready to supply name of carrier, delivery date, consignee name, freight or express bill number, item
damaged and extent of damage.
Complete instructions regarding claim procedure are found in Section “S” of the Policy & Procedure Bulletins.
GE personnel, please use the GEMS CQA Process to report all omissions, errors, and defects in this documentation.
Customers, please contact your GE Sales or Service represenatives.
d
GE MEDICAL SYSTEMS HELICAL UPGRADE INSTALLATION
REV 2 2151639
CAUTION
Do not use the following devices near this equipment. Use of these devices near this equipment could cause
this equipment to malfunction.
Devices which intrinsically transmit radio waves such as; cellular phone, radio transceiver, mobile radio transmitter,
radio–controlled toy, etc.
Keep power to these devices turned off when near this equipment.
Medical staff in charge of this equipment is required to instruct technicians, patients and other people who may be
around this equipment to fully comply with the above regulation.
e
GE MEDICAL SYSTEMS HELICAL UPGRADE INSTALLATION
blank 2151639
f
GE MEDICAL SYSTEMS HELICAL UPGRADE INSTALLATION
REV 3 2151639
REVISION HISTORY
REV DATE PRIMARY REASON FOR CHANGE
0 . . January 15, 1996 . . . initial release.
1 . . . . . Sep. 30, 1996 . . . SIMM P/N changed, others
2 . . October 15, 1997 . . . For Renaissance 3
3 . . . August 31, 1998 . . . XG CPU PATCH
PAGE REV PAGE REV PAGE REV PAGE REV PAGE REV
Title page . . . . . . . . . . 3 B . . . . . . . . . . . . . . blank 1–2 . . . . . . . . . . . . blank 3–3 . . . . . . . . . . . . . . . . 0 3–10 . . . . . . . . . . . . . . . 2
a to e . . . . . . . . . . . . . . 2 i...................3 2–1 to 2–2 . . . . . . . . . . 2 3–4 . . . . . . . . . . . . . . . 3 3–11 to 3–17 . . . . . . . 3
f . . . . . . . . . . . . . . blank ii . . . . . . . . . . . . . . blank 3–1 . . . . . . . . . . . . . . . . 1 3–5 . . . . . . . . . . . . . . . . 1 3–18 . . . . . . . . . . . blank
A ..................3 1–1 . . . . . . . . . . . . . . . . 0 3–2 . . . . . . . . . . . . . . . 3 3–6 to 3–9 . . . . . . . . . . 0 4–1 to 4–2 . . . . . . . . . . 2
A
GE MEDICAL SYSTEMS HELICAL UPGRADE INSTALLATION
blank 2151639
B
GE MEDICAL SYSTEMS HELICAL UPGRADE INSTALLATION
REV 3 2151639
TABLE OF CONTENTS
SECTION PAGE
i
GE MEDICAL SYSTEMS HELICAL UPGRADE INSTALLATION
blank 2151639
ii
GE MEDICAL SYSTEMS HELICAL UPGRADE INSTALLATION
REV 0 2151639
SECTION 1 – INTRODUCTION
This Option kit can extend the slice numbers in Helical scan as follows:
1–1
GE MEDICAL SYSTEMS HELICAL UPGRADE INSTALLATION
blank 2151639
1–2
GE MEDICAL SYSTEMS HELICAL UPGRADE INSTALLATION
REV 2 2151639
2–1
GE MEDICAL SYSTEMS HELICAL UPGRADE INSTALLATION
REV 2 2151639
2–2
GE MEDICAL SYSTEMS HELICAL UPGRADE INSTALLATION
REV 1 2151639
3–1 PREPARATION
1. Verify that the contents of this option kit are correct according to the Material List. See Section 2.
Note
Some kit contain the two types of SIMMs. Do not mix their SIMMs from the kit, and the SIMMs which
are currently installed on the IPU2 Board;
Keep these SIMMs separately with their part numbers as follows until they are installed into the cor-
rect sockets, since the provided SIMMs are not labelled, and should be installed into the specified
sockets according to the SIMM part number.
2150858 SIMMs
2150860 SIMMs
Currently installed SIMMs
The pouches or packages for the SIMMs carry their labels; follow these labels. If NOT labelled, differ-
entiate the SIMMs as follows (if not labelled ONLY):
2150858 SIMMs: ICs are mounted on one side of the SIMM board
2150860 SIMMs: ICs are mounted on both sides of the SIMM board
Since the above mount types may change in the future, follow the labels and IGNORE the mount
types.
3–1
GE MEDICAL SYSTEMS HELICAL UPGRADE INSTALLATION
REV 3 2151639
MEMORY CHECK
D PAGING TEST
NOTICE
3–2
GE MEDICAL SYSTEMS HELICAL UPGRADE INSTALLATION
REV 0 2151639
a. According to the table below, remove the SIMMs from the socket.
MD EXT SW
(S7A1)
J10 12 14
11 13 J15
J20 22 24
21 23 J25
SIMM SOCKETS
3–3
GE MEDICAL SYSTEMS HELICAL UPGRADE INSTALLATION
REV 3 2151639
c. Set up the SIMMs such that the metal latches hold each SIMM securely (See Illustrations 3–3).
ÊÊ
Ê ÉÉÉ
ÉÉÉÉ ÊÊÊÊ
Ê
ÊÊ ÉÉÉ
ÉÉÉÉ ÊÊÊÊ
Ê
ÊÊ ÉÉÉ
ÉÉÉÉ ÊÊÊÊ
Ê
ÊÊ ÉÉÉ
ÉÉÉÉ ÊÊÊÊ
Ê
ÊÊ
Ê
ÊÊ ÉÉ
ÉÉÉ
ÉÉ
É
ÉÉÉÉ
ÉÉ ÊÊ
ÊÊÊÊ
ÊÊ
Ê
ÊÊ
Ê
ÊÊ ÉÉ
ÉÉÉ
ÉÉÉÉ ÊÊÊÊ
ÉÉÉÉ
É
ÉÉÉ
ÊÊ
Ê
ÊÊ
J10
11
12
13
14
J15
ÉÉÉÉ
J10
11
12
13
14
J15
ÊÊÊÊ
ÊÊ
J10
11
12
13
14
J15
Ê
ÊÊ ÉÉÉ
ÉÉÉÉ ÊÊÊÊ
Ê
ÊÊ ÉÉÉ
ÉÉÉÉ ÊÊÊÊ
Ê
ÊÊ
Ê
ÊÊ ÉÉ
ÉÉÉ
ÉÉ
É
ÉÉÉÉ
ÉÉ ÊÊ
ÊÊÊÊ
ÊÊ
Ê
ÊÊ ÉÉÉ
ÉÉÉÉ ÊÊÊÊ
Ê
ÊÊ ÉÉÉ
ÉÉÉÉ ÊÊÊÊ
Ê
ÊÊ
Ê
ÊÊ ÉÉ
ÉÉÉ
ÉÉ
É
ÉÉÉÉ
ÉÉ ÊÊ
ÊÊÊÊ
ÊÊ
Ê
ÊÊ
J20
21
22
23
24
J25
ÉÉÉÉ
ÉÉÉ
J20
21
22
23
24
J25
ÊÊÊÊ
J20
21
22
23
24
J25
ÊÊ
ÊÊ SOCKETS FOR 2150858 SIMMs
ÉÉ
NO SIMM IS INSTALLED.
SIMM LOCATIONS
ILLUSTRATION 3–1
3–4
GE MEDICAL SYSTEMS HELICAL UPGRADE INSTALLATION
REV 1 2151639
SIMM
METAL LATCH
3–5
GE MEDICAL SYSTEMS HELICAL UPGRADE INSTALLATION
REV 0 2151639
D MD EXT SW (S7A1) – Set the switches to the proper setting as shown below:
a. On the CPW Board (slot #1), set the Dip SW 8A1 as follows:
ON
ON : 1
OFF: 0
1 2 3 4 5 6 7 8
SW 8A1
b. Switch ON the OC power. It takes approx. 14 minutes for the P–ON test level 2 to finish. The start up screen
will appear when the P–ON test has been completed normally.
c. After verification of the P–ON test level 2, set the SW 8A1 to the default setting.
ON
ON : 1
OFF: 0
1 2 3 4 5 6 7 8
SW 8A1
Note:
There is no need to turn OFF the OC power when the SW8A1 setting is changed. The new setting is
available at next power ON sequence.
3–6
GE MEDICAL SYSTEMS HELICAL UPGRADE INSTALLATION
REV 0 2151639
1. Perform 60 axial scans using the water phantom. Select the minimum scan time (do not select the partial scan)
and the minimum MA. The “Auto View” function should be ON. Write down the Study and Series Nos.
Check that the scans are normally performed.
- Paging Test
2. Magnify the image as follows.
a. Press [Crsr] to display the crosshair. Use the trackball to move the cursor into an appropriate position.
e. Enter the image IDs of the former half of the acquired images (i.e., 30; the total is 60) so that AAA (# of Images
selected) equals to BBB (Total # available; BBB: 30) as in
PAGING
(AAA/BBB)
Check that the Paging function is normally performed and the images are normal.
Perform these checks also for the latter half of the acquired images.
3–7
GE MEDICAL SYSTEMS HELICAL UPGRADE INSTALLATION
REV 0 2151639
Select No. =
9. Enter 10 for “NO. OF IMAGE”. Use the default values for the other parameters.
14. Enter the image ID A.B.C;2. The displayed image is subtracted by the image entered.
Verify that a nil image is left with the settings of Window Width “2” and Window Level “0”.
18. Press [Image Delete] and delete the images A.B.C;1 ~ A.B.C;11 and images created by the test scans.
3–8
GE MEDICAL SYSTEMS HELICAL UPGRADE INSTALLATION
REV 0 2151639
BOOT
6. Remove the floppy, insert the CLV5 (Cooling Version 5) floppy, and press the <Enter>
key.
The software will be loaded to the system so that “x” replaces the number on the FD CONTENTS menu.
9. Verify that the system starts up normally, with no errors during the power–on self test.
3–9
GE MEDICAL SYSTEMS HELICAL UPGRADE INSTALLATION
REV 2 2151639
OGP BOARD
FRONT VIEW
3–10
GE MEDICAL SYSTEMS HELICAL UPGRADE INSTALLATION
REV 3 2151639
For systems with XG CPU V3.10 or lower, when scanning with ‘140kV/160mA/Zj–a:Large Focus (Qj: 3 ~ 10 mm
thickness) /72 or 743 sec or more scan time’ is performed, XG ERROR 371 often occurs, causing scanning to be
canceled.
This error does NOT occur when small focus (or 1~2mm thickness) technique, ‘71 sec or less’ scan time, or kV/mA
other than 140/160 is used. However, for site with 90 slice helical upgrade option, there is a possibility that this error
occurs.
1. Does XG ERROR 371 occur when scanning of ‘140kV/160mA/10mm thickness/90 sec/1sec scan time’ is per-
formed?
NOTICE
Normally, Firmware Version can be checked on the CRT monitor by using Generator test, but
in this time you MUST remove CPU board and check its part number to know CPU version.
This is why firmware version V3xx or V4xx might be displayed in complex indication (ex:
04V496 for V401...).
a. Turn the ‘Rotate’, ‘XG’, and ‘Slip Ring’ switches OFF at the rear base of the Gantry.
3–11
GE MEDICAL SYSTEMS HELICAL UPGRADE INSTALLATION
REV 3 2151639
f. Remove the CPU board from the Electronic rack and Check Part No. of the CPU Board.
TO TABLE
ELECTRONIC RACK
3–12
GE MEDICAL SYSTEMS HELICAL UPGRADE INSTALLATION
REV 3 2151639
a. Turn the ‘Rotate’ switch OFF at the rear base of the Gantry.
Note
The ‘Slip Ring 115V’ switch must remain ON and the TGP Board Service Switch must remain set to
‘SYS’.
***GENERATOR TEST***
c. Enter the following in response to ‘TO:’ for entering Parameter Change Mode.
TO:dc750 ‘CR’
TO:SS0 ‘CR’
Note
For ‘dc’, the lower case must be entered.
NOTICE
If ‘0710’ or ‘0730’ is NOT displayed as DARAM229, cancel this solution procedures (Enter ‘N’
+ ‘CR’ to exit from this mode) .
3–13
GE MEDICAL SYSTEMS HELICAL UPGRADE INSTALLATION
REV 3 2151639
Note
For ‘m’ and ‘d’, the lower case must be entered.
TO:SSW ‘CR’
TO:N ‘CR’ FROM:WRITE EEP OK
TO:dc550 ‘CR’
TO:SS0 ‘CR’
Note
For ‘dc’, the lower case must be entered.
Note
For ‘m’ and ‘d’, the lower case must be entered.
TO:SSW ‘CR’
TO:N ‘CR’ FROM:WRITE EEP OK
3–14
GE MEDICAL SYSTEMS HELICAL UPGRADE INSTALLATION
REV 3 2151639
a. Press <End> key to terminate the Generator Test mode and pull the reset switch under the Operator Console.
b. Enter GENERATOR TEST mode again and verify that d282 or d229 address has new data:
TO:dc750 ‘CR’
TO:SS0 ‘CR’
TO:dc550 ‘CR’
TO:SS0 ‘CR’
d. Perform continuous scans on condition of 140kV/160mA/10mm thickness/90 sec/1sec scan time. Verify that
all scans can be performed without XG Err 371.
3–15
GE MEDICAL SYSTEMS HELICAL UPGRADE INSTALLATION
REV 3 2151639
[Preset Prmtr] ”
SCAN PARAMETER ”
I–EXTR (Example; if the customer uses all of menu No. in this anatomic location, select other location.) ”
MULTI/DYNAMIC Menu ”
No.10 (Example; if the customer uses this menu No., select another menu No.)
2. Modify only the items in bold and underlined to the following values.
.
.
.
4. Verify that ‘NOW COOLING CHECK’ message appears on lower rigtht side of the screen, then return to the
preset scan parameters screen.
Note
This means that cooling check can have been completed successfully. If cooling check error occured,
‘**COOLING OVER LIMIT’ message will appear.
3–16
GE MEDICAL SYSTEMS HELICAL UPGRADE INSTALLATION
REV 3 2151639
3. Press the following keys on the LCD screen in the order indicated.
4. The [Multi/Dynamic Scan] technique page appears. Touch the [Scan Screen] key to change the scan parameters
as described below. For the other parameters, use the defaults.
Scan Parameter to be 2150486 Upgrade Kit 2150488 Upgrade Kit 2150491 Upgrade Kit
changed
Slice Count 60 90 90
Time 1s 1s 1s
kV 80 80 80
mA 60 60 60
Thickness 10 mm 10 mm 10 mm
Interval 0 mm 0 mm 0 mm
6. Press the following scan buttons under the scan panel in the order that these buttons blink:
7. Verify that no error occurs during the scans and that images are displayed one after another.
Note
The ring artifacts due to low dose can be ignored.
3–17
GE MEDICAL SYSTEMS HELICAL UPGRADE INSTALLATION
blank 2151639
3–18
GE MEDICAL SYSTEMS HELICAL UPGRADE INSTALLATION
REV 2 2151639
Note
If other optional software have been installed in the system, other rating plates should have been
attached. In this case, attach this rating plate above the others.
OPTION SOFTWARE
RATING PLATE
b. Instruct the Operators as to the appropriate operation procedures of this software described in the Operator’s
manual.
4–1
GE MEDICAL SYSTEMS HELICAL UPGRADE INSTALLATION
REV 2 2151639
2. Send the Product locator card to the appropriate addresses provided below.
4–2