DIRECTION 5197053-100, REVISION 1 LOGIQ™ E PERIPHERALS INSTALLATION MANUAL
iv
GE HEALTHCARE
DIRECTION 5197053-100, REVISION 1 LOGIQ™ E PERIPHERALS INSTALLATION MANUAL
The purchaser of GE equipment shall only utilize qualified personnel (i.e., GE’s
field engineers, personnel of third-party service companies with equivalent training,
or licensed electricians) to perform electrical servicing on the equipment.
v
GE HEALTHCARE
DIRECTION 5197053-100, REVISION 1 LOGIQ™ E PERIPHERALS INSTALLATION MANUAL
GE Medical Systems employees should use the iTrak System to report all
documentation errors or omissions.
vi
GE HEALTHCARE
DIRECTION 5197053-100, REVISION 1 LOGIQ™ E PERIPHERALS INSTALLATION MANUAL
Revision History
7-1 to 7-8 1
- vii
GE HEALTHCARE
DIRECTION 5197053-100, REVISION 1 LOGIQ™ E PERIPHERALS INSTALLATION MANUAL
viii -
GE HEALTHCARE
DIRECTION 5197053-100, REVISION 1 LOGIQ™ E PERIPHERALS INSTALLATION MANUAL
Table of Contents
CHAPTER 1
Introduction
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 1
Purpose . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 1
Peripherals Covered by this Manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 2
Typical Users of the LOGIQ™ e Peripherals Installation Manual . . . 1 - 2
Important Conventions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 3
Conventions Used in Book . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 3
Icons . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 3
Safety Precaution Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 3
Standard Hazard Icons . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 4
Safety Considerations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 5
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 5
Human Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 5
Mechanical Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 5
Electrical Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 6
Safe Practices . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 6
Dangerous Procedure Warnings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 6
Lockout/Tagout Requirements (For USA Only) . . . . . . . . . . . . . . . . . . . . . . 1 - 7
Returning/Shipping Probes and Repair Parts . . . . . . . . . . . . . . . . . . . . . . . 1 - 7
Customer Assistance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 9
Contact Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 9
System Manufacture . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 10
1 Table of Contents
GE HEALTHCARE
DIRECTION 5197053-100, REVISION 1 LOGIQ™ E PERIPHERALS INSTALLATION MANUAL
CHAPTER 2
Preparation
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 1
Purpose of Chapter 2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 1
Furnished materials . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 3
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 3
B/W USB Printer (Sony UP-D897) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 3
Materials supplied with the B/W USB Printer . . . . . . . . . . . . . . . . . . 2 - 3
Materials supplied with scanner . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 3
DVD-RW (LITEON DVD-RW) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 4
Materials supplied with the DVD-RW . . . . . . . . . . . . . . . . . . . . . . . . 2 - 4
Materials supplied in the scanner . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 4
USB Wireless LAN (Linksys WUSB54G) . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 4
Materials supplied with the Wireless LAN . . . . . . . . . . . . . . . . . . . . . 2 - 4
Materials supplied in the scanner . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 4
Color USB Printer (UP-D23MD/HP460/HP K550) . . . . . . . . . . . . . . . . . . . . 2 - 5
Materials supplied with the Color USB Printer . . . . . . . . . . . . . . . . . 2 - 5
Materials supplied with scanner . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 5
3 Pedal Footswitch (MKF 2-MED GP26)/1 Pedal Footswitch (Whanam FSU2001) 2 -
5
Materials supplied with the Footswitch . . . . . . . . . . . . . . . . . . . . . . . 2 - 5
Materials supplied in the scanner . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 5
Peripherals/Accessories Connector Panel . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 6
Rear Panel Connector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 6
Left Panel Connector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 6
2 Table of Contents
GE HEALTHCARE
DIRECTION 5197053-100, REVISION 1 LOGIQ™ E PERIPHERALS INSTALLATION MANUAL
CHAPTER 3
B/W USB Printer Installation
Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 1
Purpose of Chapter 3 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 1
Table of Contents 3
GE HEALTHCARE
DIRECTION 5197053-100, REVISION 1 LOGIQ™ E PERIPHERALS INSTALLATION MANUAL
CHAPTER 4
DVD-RW Installation
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 1
Purpose of Chapter 4 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 1
4 Table of Contents
GE HEALTHCARE
DIRECTION 5197053-100, REVISION 1 LOGIQ™ E PERIPHERALS INSTALLATION MANUAL
CHAPTER 5
Wireless LAN Adapter Installation
Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 1
Purpose of Chapter 5 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 1
Table of Contents 5
GE HEALTHCARE
DIRECTION 5197053-100, REVISION 1 LOGIQ™ E PERIPHERALS INSTALLATION MANUAL
CHAPTER 6
Footswitch Installation
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 1
Purpose of Chapter 6 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 1
Installing Footswitch. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 2
Tools . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 2
Manpower . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 2
Preparations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 2
Installation Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 2
Configuring Footswtich . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 3
6 Table of Contents
GE HEALTHCARE
DIRECTION 5197053-100, REVISION 1 LOGIQ™ E PERIPHERALS INSTALLATION MANUAL
CHAPTER 7
Color USB Printer Installation
Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 - 1
Purpose of Chapter 7 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 - 1
Table of Contents 7
GE HEALTHCARE
DIRECTION 5197053-100, REVISION 1 LOGIQ™ E PERIPHERALS INSTALLATION MANUAL
CHAPTER 8
USB Memory Device Installation
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 - 1
Purpose of Chapter 8 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 - 1
8 Table of Contents
GE HEALTHCARE
DIRECTION 5197053-100, REVISION 1 LOGIQ™ E PERIPHERALS INSTALLATION MANUAL
CHAPTER 9
Measure Transfer Installation
Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 - 1
Purpose of Chapter 9 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 - 1
Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 - 2
Physical Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 - 2
Serial Communication Characteristics . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 - 2
Table of Contents 9
GE HEALTHCARE
DIRECTION 5197053-100, REVISION 1 LOGIQ™ E PERIPHERALS INSTALLATION MANUAL
CHAPTER 10
USB Hub Installation
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 1
Purpose of Chapter 10 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 1
10 Table of Contents
GE HEALTHCARE
DIRECTION 5197053-100, REVISION 1 LOGIQ™ E PERIPHERALS INSTALLATION MANUAL
CHAPTER 11
ECG Installation
Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11 - 1
Purpose of Chapter 11 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11 - 1
Table of Contents 11
GE HEALTHCARE
DIRECTION 5197053-100, REVISION 1 LOGIQ™ E PERIPHERALS INSTALLATION MANUAL
CHAPTER 12
Set up and Check-out
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12 - 1
Purpose of Chapter 12 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12 - 1
Footswitch Checks. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12 - 17
12 Table of Contents
GE HEALTHCARE
DIRECTION 5197053-100, REVISION 1 LOGIQ™ E PERIPHERALS INSTALLATION MANUAL
Chapter 1
Introduction
Section 1-1
Overview
1-1-1 Purpose
This Installation Manual describes how to install and configure the peripherals validated for the
LOGIQ™ e.
PAGE
SECTION DESCRIPTION NUMBER
1-1 Overview 1-1
WEIGHT
PART NUMBER DESCRIPTION (KG) POWER CONTROL MODEL
H41642LY(US) 100-120VAC/220-
H41642LZ(EU) 240VAC
Sony B/W USB Printer 3.4 USB port 1 or 2 UP-D897
H41652LA(CN) 50/60Hz
H41652LB(JP) 1.5A/0.8A
H41482LF(US)
H41482LG(EU) 100-120VAC/220-
Color USB Printer 6.5 240VAC USB port 1 or 2 UP-D23MD
H41482LE(CN)
1.8A/1.0A
H41482LH(JP)
H41862LC(US)
100-120VAC/220-
H41862LD(EU)
12.5 240VAC USB port 1 or 2 HP K550
H41862LE(CN)
1.8A/1.0A
H41552LL(JP)
Deskjet Color Printer
H41852LZ(US)
100-120VAC/220-
H41862LA(EU)
4.0 240VAC USB port 1 or 2 HP460+PIT
H41862LB(CN)
1.8A/1.0A
H41552LK(JP)
H41642LM (US)
H41642LP (EU) LITEON DVD-RW 100-240VAC
1.3 USB port 1 or 2 LITEON SHM-165P6SX
H41642LN (CN) (Power Cord) 1.5A
H41672LL (JP)
H41852LT Linksys Wireless LAN Adapter 0.05 5V USB port 1 or 2 Linksys WUSB54G
H41852LY 512M USB Memory Stick 0.05 5V USB port 1 or 2 Kingston Data Traveller II
H41462LL(US) 0.215
H41462LM(CN) 0.195
Power Cable for printer
H41462LN(EU) 0.205
H41462LP(JP) 0.19
1-2 -
GE HEALTHCARE
DIRECTION 5197053-100, REVISION 1 LOGIQ™ E PERIPHERALS INSTALLATION MANUAL
Section 1-2
Important Conventions
1-2-1 Conventions Used in Book
1-2-1-1 Icons
Pictures, or icons, are used wherever they will reinforce the printed message. The icons, labels and
conventions used on the product and in the service information are described in this chapter.
CAUTION Caution is used to indicate the presence of a hazard that will or can cause minor personal injury
and property damage if instructions are ignored.
NOTE: Notes are used to provide important information about an item or a procedure.
Be sure to read the notes; the information contained in a note can often save you time or effort.
LASER
LIGHT
Other hazard icons make you aware of specific procedures that should be followed.
AVOID STATIC ELECTRICITY TAG AND LOCK OUT WEAR EYE PROTECTION
TAG
&
LOCKOUT EYE
PROTECTION
Signed Date
Section 1-3
Safety Considerations
1-3-1 Introduction
The following safety precautions must be observed during all phases of operation, service and repair of
this equipment. Failure to comply with these precautions or with specific warnings elsewhere in this
manual, violates safety standards of design, manufacture and intended use of the equipment.
Only personnel who have participated in a LOGIQ™ e Training Seminar are authorized to service the
equipment.
WARNING ALWAYS CLOSE THE LCD COVER BEFORE MOVING THE SCANNER AROUND.
• The system is equipped with a three-conductor AC power cable. This must be plugged into an
approved electrical outlet with protective ground.
• The power outlet used for this equipment should not be shared with other types of equipment.
• Both the system power cable and the power connector meet international electrical standards.
GEMS policy states that body fluids must be properly removed from any part or equipment prior to
shipment. GEMS employees, as well as customers, are responsible for ensuring that parts/equipment
have been properly decontaminated prior to shipment. Under no circumstance should a part or
equipment with visible body fluids be taken or shipped from a clinic or site (for example, body coils or
an ultrasound probe).
The purpose of the regulation is to protect employees in the transportation industry, as well as the
people who will receive or open this package.
NOTE: The US Department of Transportation (DOT) has ruled that “items that were saturated and/or
dripping with human blood that are now caked with dried blood; or which were used or intended
for use in patient care” are “regulated medical waste” for transportation purposes and must be
transported as a hazardous material.
Section 1-4
EMC, EMI, and ESD
1-4-1 Electromagnetic Compatibility (EMC) and Interference (EMI)
Electromagnetic compatibility describes a level of performance of a device within its electromagnetic
environment. This environment consists of the device itself and its surroundings including other
equipment, power sources and persons with which the device must interface. Inadequate compatibility
results when a susceptible device fails to perform as intended due interference from its environment or
when the device produces unacceptable levels of emission to its environment. This interference is often
referred to as radio–frequency or electromagnetic interference (RFI/EMI) and can be radiated through
space or conducted over interconnecting power of signal cables. In addition to electromagnetic energy,
EMC also includes possible effects from electrical fields, magnetic fields, electrostatic discharge and
disturbances in the electrical power supply.
WARNING DO NOT TOUCH ANY BOARDS WITH INTEGRATED CIRCUITS PRIOR TO TAKING THE
NECESSARY ESD PRECAUTIONS:
1.) FOLLOW GENERAL GUIDELINES FOR HANDLING OF ELECTROSTATIC SENSITIVE
EQUIPMENT.
1-4-3 CE Compliance
The LOGIQ™ e unit conforms to all applicable conducted and radiated emission limits and to immunity
from electrostatic discharge, radiated and conducted RF fields, magnetic fields and power line transient
requirements.
NOTE: For CE Compliance, it is critical that all covers, screws, shielding, gaskets, mesh, clamps, are
in good condition, installed tightly without skew or stress. Proper installation following all
comments noted in this service manual is required in order to achieve full EMC performance.
Section 1-5
Customer Assistance
1-5-1 Contact Information
If this equipment does not work as indicated in this service manual or in the User Manual, or if you
require additional assistance, please contact the local distributor or appropriate support resource, as
listed below.
USA
Service: On-site 1-800–437–1171
GE Medical Systems
Ultrasound Service Engineering
Service: Parts 1-800-558-2040
9900 Innovation Drive
Wauwatosa, WI 53226
Applications support 1-800-682-5327 or 1-262-524-5698
Canada 1-800-668-0732
Service 1-800-321-7937
Latin America
Applications support 1-262-524-5698
Europe
GE Ultraschall Deutschland GmbH& Co. KG Phone: +33 (0)130-831-300 (General Imaging and Cardiac)
BeethovenstraBe 239
Postfach 11 05 60, D-42665 Solingen Fax: +49 (0)212-2802-431
Germany
Asia (Singapore)
Tel: +65 291-8528
GE Ultrasound Asia
Service Department - Ultrasound
298 Tiong Bahru Road #15-01/06
Central Plaza
Fax: +65 6291-7006
Singapore 169730
Phone: 81-42-648-2944
Japan Support Center
Fax: 81-42-648-2905
Chapter 2
Preparation
Section 2-1
Overview
2-1-1 Purpose of Chapter 2
This chapter includes procedures to conduct before installation of the Peripherals.
Section 2-2
Power Shut Down
Press the ON/OFF key on the Control Panel once to display the
System - Exit Menu on the monitor.
Section 2-3
Furnished materials
2-3-1 Overview
This section describes the materials furnished with the Peripherals and with the System.
Retain the original carton and packing materials in case transport is needed in the future.
2. Paper Roll 1
3. Operating Instructions 1
4. USB cable 1
Table 2-4 B/W USB Printer: Materials furnished with the Scanner
2. Operating instruction 1
3. AC power cord 1
4. AC Power Adapter 1
5. USB cable 1
2. Operating Instructions 1
2. Paper Roll 1
3. Operating Instructions 1
5. USB cable 1
Table 2-10 Color USB Printer: Materials furnished with the Scanner
2-3-6 3 Pedal Footswitch (MKF 2-MED GP26)/1 Pedal Footswitch (Whanam FSU2001)
2
3
4
1
Figure 2-1 Rear Connector Panel
1 2
1.) Two interchangeable USB Port (Digital Printer, DVD-RW and/or Foot Switch, etc).
2.) Earphone Port.
NOTICE The USB devices should be connected to LOGIQ™ e first, power on USB devices before turning
LOGIQ™ e to work.
NOTE: Each outer (case) ground line of peripheral/accessory connectors is protectively grounded.
Signal ground lines are not isolated.
Chapter 3
B/W USB Printer Installation
Section 3-1
Overview
3-1-1 Purpose of Chapter 3
This chapter describes the installation procedure for the Sony UP-D897 B/W USB Printer.
Section 3-2
Installing the B/W USB Printer
3-2-1 Tools
No special tools needed.
3-2-2 Manpower
One person 5 min.
3-2-3 Preparations
1.) Unpack the B/W USB Printer.
USB cable
Power cord
Figure 3-2 B&W Printer USB cable and Power cord connection
USB port
Chapter 4
DVD-RW Installation
Section 4-1
Overview
4-1-1 Purpose of Chapter 4
This chapter describes the installation of the LITEON DVD-RW SHM-165P6SX.
DVD-RW
Section 4-2
LITEON DVD-RW drive recommended media list
Media issues are common throughout the DVD-RW drive industry. Because DVD media vendors often
change disc suppliers, quality levels may change due to manufacturing differences. This means that you
may encounter DVD/CD creation problems with media that may have worked successfully before.
Overall system configuration and other factors may also affect the success of creating a DVD/CD.
The following media types have been tested. LITEON highly recommends that you use the media types
on this list when creating your DVDs/CDs.
Not all brands of media have been tested and, therefore, you may encounter success with other brands
not listed. This list will be updated as other media is tested and approved.
SONY DVD-R
SONY CD-R
SONY DVD-RW
Mitsubishi CD-R
Mitsubishi CD-RW
Mitsubishi DVD-R
Mitsubishi DVD-RW
JVC DVD-R
MBI CD-R
MBI DVD-R
4-2 -
GE HEALTHCARE
DIRECTION 5197053-100, REVISION 1 LOGIQ™ E PERIPHERALS INSTALLATION MANUAL
Section 4-3
Installing the DVD-RW
4-3-1 Tools
No special tools need.
4-3-2 Manpower
One person 5 min.
4-3-3 Preparations
1.) Unpack the DVD-RW.
2.) Ensure no physical damage.
3.) Connect the power cord with outlet on the wall, then turn on the printer.
USB port
WARNING
Chapter 5
Wireless LAN Adapter Installation
Section 5-1
Overview
5-1-1 Purpose of Chapter 5
This chapter describes how to install Linksys Wireless LAN Adapter.
Section 5-2
Installing Wireless LAN Adapter
5-2-1 Tools
No special tools needed.
5-2-2 Manpower
One person 5 min.
5-2-3 Preparations
1.) Unpack the Wireless LAN Adapter.
2.) Ensure no physical damage.
USB Port
2.) Press Start menu, then click on Network Connections to active Wireless application window.
3.) Config the wireless Lan setting with local IT person, using network connections control panel.
Chapter 6
Footswitch Installation
Section 6-1
Overview
6-1-1 Purpose of Chapter 6
This chapter describes how to install Footswitch. LOGIQ™ e supports two types of footswitch: 3-pedal
footswitch and 1-pedal footswitch.
Section 6-2
Installing Footswitch
6-2-1 Tools
No special tools needed.
6-2-2 Manpower
One person 5 min.
6-2-3 Preparations
1.) Unpack the Footswitch.
2.) Ensure no physical damage.
3-pedal
Footswitch
Section 6-3
Configuring Footswtich
Footswtich supports six configurations: Non function, acting as Freeze button, P1 button, P2 button, P3
button or B pause button.
For 1-pedal footswitch, the default setting is acting as Freeze button; for 3-pedal footswitch, the default
setting is acting as Print 1 for Left pedal, B Pause for Middle pedal and Freeze button for Right pedal.
To change them, press Utility button on the key board then select Application-->Settings. (Figure 6-4 on
page 6-3 )
Chapter 7
Color USB Printer Installation
Section 7-1
Overview
7-1-1 Purpose of Chapter 7
This chapter describes the installation procedure for Color USB Printers. LOGIQ™ e supports three
types of printers for printing: Sony UP-D23MD Digital Color USB Printer, HP460 Deskjet Color Printer,
HP K550 Deskjet Color Printer. For all three printers, the connection to LOGIQ™ e is same. Here we
take Sony UP-D23MD as an example. Please make sure the printers are Power On.
Section 7-2
Installing the Color USB Printer
7-2-1 Tools
No special tools needed.
7-2-2 Manpower
One person 5 min.
7-2-3 Preparations
1.) Unpack the color USB Printer.
USB cable
Figure 7-6 Color Printer USB cable and Power cord connection
USB port
Section 7-3
Setting for B/W and Color printer concurrent working
Example below points B/W printer (UP-D897) to [P2] button on the keyboard and Color printer (UP-
D23MD) to [P3].
Section 7-3 Setting for B/W and Color printer concurrent working (cont’d)
4.) Choose [connectivity] -> [Service], select [Standard Print] in list, see Figure 7-9 on page 7-4 .
5.) Set standard printer as “Sony UP-D897”, change the name as “UP-D897” in properties area. Save
the setting.
6.) Choose [Connectivity] -> [Button], Select [Print2] in Button list, and then press “<<“ to remove all
the original setting.
Section 7-3 Setting for B/W and Color printer concurrent working (cont’d)
7.) Select UP-D897, press “>>” to add to the Printflow View.
8.) [P2] setting for UP-D897 shows as below.
Section 7-3 Setting for B/W and Color printer concurrent working (cont’d)
10.)Power on the Color printer.
11.)Choose [connectivity] -> [Service], select [Standard Print] in list, see Figure 7-13 on page 7-6 .
12.)Set standard printer as “UP-D23MD”, change the name as “UP-D23MD” in properties area. Save
the setting.
13.)Go to [Connectivity] -> [Button], choose [Print3], and then remove all the [Print3] original setting.
Section 7-3 Setting for B/W and Color printer concurrent working (cont’d)
14.)Select UP-D23MD, press “>>” to add to the Printflow View.
15.)[P3] setting for UP-D23MD shows as below.
Chapter 8
USB Memory Device Installation
Section 8-1
Overview
8-1-1 Purpose of Chapter 8
This chapter describes the installation procedure for USB Memory Device.
Section 8-2
Installing the USB Memory Device
8-2-1 Tools
No special tools needed.
8-2-2 Manpower
One person 1 min.
8-2-3 Preparations
1.) Unpack the USB Memory Device.
Chapter 9
Measure Transfer Installation
Section 9-1
Overview
9-1-1 Purpose of Chapter 9
This chapter describes the installation procedure for measure transfer function. TO use this function,
one USB-to-RS232C Converter is needed.
Section 9-2
Description
9-2-1 Physical Interface
The connection between LOGIQ™ e and Peripheral Device uses two cables: USB-to-RS232C
Connector and RS232C Cross Cable (Female 9P-Female 9P) as shown in Figure 9-18 on page 2 .
Data Bits 8
Stop Bit 1
Parity None
Section 9-3
Install Measure Transfer Function
9-3-1 System Preparations
The system shall be connected to the serial peripheral device by serial cable. Port setting is static.
CAUTION On LOGIQ™ e, it is important that the converter shall be connected before booting the system.
If not, LOGIQ™ e does not recognize the converter and no error messages are displayed.
CAUTION Only one USB port on LOGIQ™ e can be used for this. USB port 2 as shown in Figure 9-19 on
page 3 .
Figure 9-19 Connect USB end of USB-to-RS232 converter to LOGIQ™ e USB port 2
3.) Connect RS232C Cross Cable to the other end of USB-to-RS232 Converter.
Figure 9-20 Connect RS232 end of USB-to-RS232 converter to a RS232C Cross Cable
4.) Connect the other end of RS232C Cross Cable to Serial Peripheral Device (e.g. Sonultra’s Ultra32
reporting station)
Section 9-4
Patient Data Output
9-4-1 User Interface
1.) Turn on the serial peripheral device, then keep on the state of receiving the patient data. Press the
“Meas. Transf” key on Worksheet in the Ultrasound system to transfer the patient data from the
Ultrasound system to the serial peripheral device.
2.) When the transfer is complete correctly, the Ultrasound system does not display any message. But
the peripheral device indicates a successful transfer by beep.
At the end of serial peripheral device:
You can use different receiving program at the serial peripheral device end to receive data. However,
you have to give Acknowledgement signal back to LOGIQ™ e for success transfer no matter which
receiving program you choose. For example, if you use HyperTerminal (from Microsoft), you have to
type letter "a" immediately after you press "Meas. Transf" button to tell LOGIQ™ e to send data. After
the transmissin, you have to type letter "a" again to confirm the transmission is successful.
Figure 9-22
If the system receives any undesired response, the following message shall be displayed.
9-1-4-0-3 No acknowledgment
Whenever there is no acknowledgment from the serial peripheral device within the desired period
after finishing to send the data from system, the following message shall be displayed.
Message: "Transfer failed. No acknowledgment from peripheral device after tyransmitting
the data."
Chapter 10
USB Hub Installation
Section 10-1
Overview
10-1-1 Purpose of Chapter 10
This chapter describes the installation procedure for the USB Hub.
Section 10-2
Installing the USB Hub
10-2-1 Tools
No special tools needed.
10-2-2 Manpower
One person 1 min.
10-2-3 Preparations
1.) Unpack the USB Hub.
2.) Ensure no physical damage.
NOTE: Any Four kinds of peripheral listed above can be used on USB HUB.
2.) Connect the peripheral to the USB HUB port. (see Figure 10-3 on page 3 )
NOTICE Don’t connect more than two removable disks to USB ports at the same time.
Chapter 11
ECG Installation
Section 11-1
Overview
11-1-1 Purpose of Chapter 11
This chapter describes the installation procedure for the ECG.
Section 11-2
Installing the ECG
11-2-1 Tools
No special tools needed.
11-2-2 Manpower
One person 1 min.
11-2-3 Preparations
1.) Unpack the ECG.
2.) Ensure no physical damage (refer to Figure 11-2 on page 2 ).
2.) Wait 5 seconds for ECG function start automatically. (Refer to Figure 11-4 on page 3 ).
AHA IEC
RA R
I Right Arm
(White) (Red)
LA L
II Left Arm
(Black) (Yellow)
RL N
III Right Leg
(Green) (Black)
Feature Specification
Current 80 mA ± 10%
CMMR > 80 dB
Chapter 12
Set up and Check-out
Section 12-1
Overview
12-1-1 Purpose of Chapter 12
This chapter describes the Final Setup for the Peripherals and the Operational Check-out.
Section 12-2
B/W USB Printer
12-2-1 Control assignment for B/W USB Printer
The B/W USB Printers are controlled from the P2 on the System’s Control Panel. The factory setting is
P2 only for the B/W USB Printer.
Figure 12-1 The Utility window (with the General/System sheet displayed)
Figure 12-2 The Connectivity window (with the Connectivity/Service sheet displayed)
5.) Select CONNECTIVITY and then choose BUTTON in the upper menu (see Figure 12-3 on page
12-3 ).
6.) In the Button Field, select the Control Panel Key to configure: P2.
7.) Select Standard Print, press “>>” to add to the Printflow View. (see Figure 12-3 on page 12-3 )
Figure 12-3 The Connectivity window (with the Connectivity/Button sheet displayed)
Section 12-3
Color USB Printer
12-3-1 Control assignment for Color USB Printer
The B/W USB Printers are also controlled from the P2 on the System’s Control Panel. But the factory
setting is P2 which is only for the B/W USB Printer, to assign P2 for color printer, please refer to Section
7-3.
Figure 12-4 The Utility window (with the General/System sheet displayed)
4.) Choose [Connectivity] -> [Service], select [Standard Print] in list, see Figure 12-6 on page 12-9 .
5.) Set standard printer as “Sony UP-D23MD”. Save the setting.
8.) In the Button Field, select the Control Panel Key to configure: P2 (see Figure 12-8 on page 12-10 ).
9.) Select Standard Print, press “>>” to add to the Printflow View.
10.)If another device is selected, use “<<“ to clear window.
12-10 Section 12-3 - Color USB Printer
GE HEALTHCARE
DIRECTION 5197053-100, REVISION 1 LOGIQ™ E PERIPHERALS INSTALLATION MANUAL
Section 12-4
DVD-RW Settings and Check-out
12-4-1 Check procedure
1.) Connect and power on USB DVD-RW drive with one DVD/CD.
2.) Press Utility -> Connectivity -> Removable Media, see Figure 12-9 on page 12-14 .
3.) Click VERIFY to check the media’s status, see Figure 12-10 on page 12-14 .
Section 12-5
B/W USB Printer Checks
Make sure Printer Function is assigned to P2 key. (The factory default is P2 for the B/W USB Printer.)
When scanning in B mode, press “Freeze” to Image scanning stops with the last picture on the
1
stop image acquisition screen.
Section 12-6
Color Printer Checks
Make sure Printer Function is assigned to P2 key (The factory default is P2 for the B/W USB Printer),
to assign P2 for color printer, please refer to Section 7-3.
When scanning in B mode, press “Freeze” to Image scanning stops with the last picture on the
1
stop image acquisition screen.
Section 12-7
Wireless LAN Adapter Checks
Table 12-3 Wireless LAN Adapter Checks
Select “Network Connections” to open network Wireless Network Connection displays in the
2
connection control panel. window, See Figure 12-12.
Section 12-8
Footswitch Checks
Section 12-9
USB Memory Device Checks