Intrinsically safe load cell interface

PR1626/60

Instruction manual

Instruction manual

9499 050 62613

Third edition

05.11.2004

Sartorius Hamburg GmbH , Meiendorfer Str. 205, D 22145 Hamburg, Germany Tel: +49.40.67960.303 Fax: +49.40.67960.383

Please note In correspondence concerning this instrument, please quote the type number and serial number as given on the type plate. Bitte beachten Bei Schriftwechsel ber dieses Gert wird gebeten, die Typennummer und die Gertenummer anzugeben. Diese befinden sich auf dem Typenschild an der Rckseite des Gertes. Noter s.v.p. Dans votre correspondance et dans vos rclamations se rapportant cet appareil, veuillez toujours indiquer le numro de type et le numro de srie qui sont marqus sur la plaquette de caractristiques.

Important As the instrument is an electrical apparatus, it may be operated only by trained personnel. The maintenance may be carried out only by qualified personnel. Wichtig Da das Gert ein elektrisches Betriebsmittel ist, darf die Bedienung nur durch eingewiesenes Personal erfolgen. Die Wartung darf nur von geschultem, fach- und sachkundigem Personal durchgefhrt werden. Important Comme l'instrument est un quipement lectrique, le service doit tre assur par du personnel qualifi. De mme, l'entretien est confier aux personnes suffisamment qualifies.

INSTRUCTION MANUAL

PR1626/60

List of Contents
1 2 2.1 2.2 2.3 2.4
2.4.1 2.4.2 2.4.3

DESCRIPTION .......................................................................................................................................................2 SAFETY INSTRUCTIONS .........................................................................................................................................4 Explosion Protection ........................................................................................................................................... 4 Initial Inspection .................................................................................................................................................. 4 Before Commissioning........................................................................................................................................ 4 Maintenance & Repair........................................................................................................................................ 4
Failure & excessive stress ......................................................................................................................................................5 Static sensitive components ................................................................................................................................................5 Replacing fuses.........................................................................................................................................................................5

3 3.1 3.2 3.3 3.4


3.4.1 3.4.2 3.4.3

TECHNICAL DATA.................................................................................................................................................6 Performance Characteristics & Specifications............................................................................................ 6 General Data.......................................................................................................................................................... 6 Supply...................................................................................................................................................................... 6 Interface To The Hazardous (Ex-) Area.......................................................................................................... 6
Supply voltage output ...........................................................................................................................................................6 Measuring voltage output (from the load cells)...........................................................................................................7 Sense voltage input (from the load cells)........................................................................................................................7

3.5
3.5.1 3.5.2

Interface To The Normal (Safre) Area............................................................................................................ 7


Measuring voltage output (to the indicator/transmitter) .........................................................................................7 Sense voltage output (to the indicator/transmitter)...................................................................................................7

3.6 3.7 3.8 3.9 4 4.1 4.2


4.2.1 4.2.2

Enviromentals ....................................................................................................................................................... 7 Electromagentic Compatibility ........................................................................................................................ 8 Dimensions & Weights ....................................................................................................................................... 8 Approvals................................................................................................................................................................ 8 MOUNTING & CONNECTIONS ..............................................................................................................................9 Mounting................................................................................................................................................................ 9 Connections .........................................................................................................................................................10
Potential equalization......................................................................................................................................................... 11 Terminal allocation............................................................................................................................................................... 12

5 5.1 5.2
5.2.1

SETTINGS.......................................................................................................................................................... 13 Opening the instrument ..................................................................................................................................13 Settings .................................................................................................................................................................14


Interface Unit (PCB 2) ......................................................................................................................................................... 14

6 6.1 6.2 7 8

MAINTENANCE ................................................................................................................................................. 15 General..................................................................................................................................................................15 Fuses.......................................................................................................................................................................15 FIGURES/INTERCONNECTION EXAMPLE .............................................................................................................. 16 EC-TYPE EXAMINATION CERTIFICATE ............................................................................................................... 18

Sartorius

PR1626/60

INSTRUCTION MANUAL

DESCRIPTION

The PR1626/60 is an intrinsically safe interface unit. It has been designed to interface load cells situated in a hazardous (Ex) area with a weighing indicator/transmitter situated in a safe area. It is classified to be an associated apparatus (with intrinsically safe circuits) and complies with class [EEx ib] IIB and IIC. The Certificate of Conformity has the PTB-Number Ex-91.C.2098 X, it is to be found under paragraph 8. The unit can be used for weighing applications as well for all other applications which require this kind of intrinsically safe interfacing. This is achieved by use of double provided current and voltage limiters in the supply circuit and barriers in the measuring and sense circuits of the PR1626/60.

The unit provides the intrinsically safe interfacing of: - 1 voltage output for the supply of e.g. load cells situated in a hazardous (Ex)-area, - 1 measuring voltage input from the hazardous (Ex)-area, - 1 sense voltage input from the hazardous (Ex)-area. This is achieved by use of double provided current and voltage limiters in the supply circuit and barriers in the measuring and sense circuits of the PR1626/60. In addition the PR1626/60 has an approval in conjunction with Sartorius indicators: PR1612/02, PR1613, PR1713, PR5210, PR5510, PR5610 & PR5710.

Sartorius

INSTRUCTION MANUAL
The load cell supply voltage can be set to:

PR1626/60

Either nominal 12V DC / 80 mA e.g. for the connection of 4 load cells of 650 each, connected in parallel (162.5 ) or nominal 7.5 V DC/ 92 mA e.g. for the connection of 8 load cells of 650 each, connected in parallel (81,3 ) or for the connection of 4 load cells of 350 each, connected in parallel (87.5 ). The housing of the PR1626/60 contains provisions for wall or rail mounting. The interface can be supplied by either a 24 V DC or a 24 V AC power supply unit. Connections are to be made via screw terminals up to 4mm2. Terminal PE must be connected to the potential equalization system, which has to be installed and earthed according to VDE 0100 and its part 540, or/and acc. to respective local regulations.

Sartorius

PR1626/60

INSTRUCTION MANUAL

2
2.1

SAFETY INSTRUCTIONS
Explosion Protection

The instrument fulfils the basic requirements of EN 50014 and EN 50020. It complies with the requirements for ignition protection class II(2)G [EEx ib]II B/C and has the type approval certificate number PTB 02 ATEX 2056. This approval number is printed on one side of the instrument. A copy of the approval can be looked up in chapter 8 of this manual. This instrument has been supplied in a safe condition. The following important information and warnings must be followed by the user in order to ensure a safe operation and to maintain the instrument in a safe condition.

2.2

Initial Inspection

Check the contents of the shipment for completeness and note whether any damage has occurred during transport. If the contents are incomplete, or there is damage, a claim should be filed with the carrier immediately, and the local organisation of Sartorius should be notified in order to facilitate the repair or replacement of the instrument.

2.3

Before Commissioning

Before putting the instrument into operation after storage or transport, visually check it for physical damage. Mounting, installation, and earthing: Mounting, installation, earthing and interconnection of the instrument must always comply with the requirements stated in the Certificate of Conformity. Over and above this, it has to comply with the local NATIONAL rules for the installation of electrical systems in hazardous areas (in Germany e.g. acc. to VDE 0165). For installation and connection refer to paragraph 4, to the application drawings of this manual, and observe the required environmental conditions as stated in the technical data. CAUTION Terminal PE must be connected to the potential equalization system. Before actual commissioning, the system has to be tested by a person who is authorized by the local authority to perform such a test.

2.4

Maintenance & Repair

Repairs can only be carried out at Sartorius Hamburg GmbH, Hamburg and are subject to checking. In case of defect or functional trouble, please, contact your local Sartorius organization for repair. When returning the instrument for repair, an exact and complete fault description must be supplied with the defective instrument.

Sartorius

INSTRUCTION MANUAL

PR1626/60

Maintenance work may be carried out only by a trained technician aware of the involved hazards, whereby the relevant precautions must be taken in account.

2.4.1 Failure & excessive stress


If the instrument is suspected of being unsafe, take it out of operation permanently. This is the case when the instrument - shows physical damage - does not function anymore - is stressed beyond the tolerable limits (e.g. during storage and transportation)

2.4.2 Static sensitive components


The instrument contains electrical components which are sensible to damage from static discharge. Therefore it is strictly forbidden to open the instrument, as also unintentional touch of components can destroy them.

2.4.3 Replacing fuses


Fuses shall only be renewed by a qualified person who is aware of the hazard involved. The apparatus must be disconnected from all voltage sources when a fuse is to be renewed, or when the apparatus is to be adapted to different mains voltages.

Sartorius

PR1626/60

INSTRUCTION MANUAL

3
3.1

TECHNICAL DATA
Performance Characteristics & Specifications

Properties expressed in numerical values with stated tolerances are guaranteed by the manufacturer. Specified non-tolerance numerical values indicate those that could be nominally expected from the mean of a range of identical instruments. This specification is valid after the instrument has warmed-up for at least 0.5 hours (reference temperature 23C).

3.2

General Data
PR1626/60 9405 316 26601 II(2)G [EEx ib ] II B/C PTB 02 ATEX 2056 35000 hours 3000 d according to OIML class III in conjunction with Sartorius weighing instruments of types PR16../.., PR17../..., PR5210/.., PR5510/.., PR5610/.., PR5710/..

Type Order number Ex approval Number MTBF Accuracy

3.3

Supply
two wire, via 4mm2 screw terminals via 4mm2 screw terminal PE (see also paragraphs 2.3 and 4.2.1.) by either 24 V DC, or 24 V AC DC = 18..33 V, AC = 18..28 V 3.8 W, 6.6 VA at 24 V 50/60 Hz, 2 Hz class I, acc. to IEC 348, and/or VDE 100 part 410, and VDE 804

Connection Earth connection Nominal voltage Tolerance range Power consumption Frequency Electrical safety

3.4

Interface To The Hazardous (Ex-) Area

A blue coloured block with 15 screw terminals of up to 4mm2 is provided for all 2 connections to the Exarea.

3.4.1 Supply voltage output


Max. load Voltage Symmetry Tolerance range Max. current Tk Connection 81 settable to either 6 V DC 80 mA nominal, or 3.75 V DC 92 mA nominal. Both symmetrically to ground. 24 mV 5.5 V... 6.3 V DC; 3.25 V... 3.95 V DC 110 mA short circuit +8 mV/K at 6 V, +3 mV/K at 3.75 V. via two 4mm2 screw terminals.

Sartorius

INSTRUCTION MANUAL 3.4.2 Measuring voltage output (from the load cells)
Vmeas. (max) Connection 37.2 mV (differential) via two 4mm2 screw terminals.

PR1626/60

3.4.3 Sense voltage input (from the load cells)


Vsense (max) Connection either 6 V nominal (with ref. to ground), or 3.75 V nominal (with ref. to ground). via two 4mm2 screw terminals.

3.5

Interface To The Normal (Safe) Area

A black coloured block with 15 screw terminals of up to 4mm2 is provided for all 2 connections to the safe area.

3.5.1 Measuring voltage output (to the indicator/transmitter)


Vmeas. (max) Tk0 TKSpan Connection 37.2 mV (differential) 0.05 V/K (at 0..+40C and 1mV/V) 5 ppm/K (at +23C, 30000d and 1mV/V) via two 4mm2 screw terminals.

3.5.2 Sense voltage output (to the indicator/transmitter)


Vsense (max) Connection either 6 V nominal (with ref. to ground), or 3.75 V nominal (with ref. to ground). via two 4mm2 screw terminals.

3.6

Enviromentals

The following environmental data are valid only if the instrument is checked in accordance with the official checking procedure. Details of these procedures and failure criteria are supplied on request from the local organization of Sartorius Hamburg GmbH. Ambient temperature, reference value Nominal operating temperature range Limit range for storage and transport Relative humidity: Reference range Nominal operational range Operational limits Limits for storage and transport Warming-up time Mounting position Electrical safety Protection class +23C -10C ... +55C 0C ... +40C (for 3000 d) -40C ... +70C 45 ... 75 % } 20 ... 80 % } no condensation 10 ... 90 % } 5 ... 95 % } approx. 30 min any. The instrument is designed for either wall- or railmounting. class I, acc. to IEC 348, and/or VDE 100 part 410, and VDE 804 acc. to IP 20 (DIN 40050)

Sartorius

PR1626/60
Housing Interference voltage

INSTRUCTION MANUAL
lower part macrolon 8020, upper (connection) part macrolon 2800. acc. DEUTSCHE-POST-VERORDNUNG VfG 1064; Amtsblatt Nr. 163/1984(VDE 0871, limit value class B)

3.7

Electromagentic Compatibility
acc. IEC 801-2, level 3 (8 kV with 5 mJ) 100 kHz...500 MHz at 10 V/m, 500 MHz...1 GHz at 1 V/m. 60 A/m at 50 Hz 100 % with 10 ms 50 % with 20 ms 20 % with 50 ms 2000 V 5 ns 100 ns 2000 V 100 ns 10 s 2000 V 5 ns 50 ns 1000 V 5 ns 50 ns } } } } } } } } } } } }

Static discharges Resistance to irradiation: Electrical fields Magnetic fields Mains voltage sags

Rapid low-energy transients: Amplitude Rise time Duration High-energy transients to power line: Amplitude Rise time Duration Burst to data lines: Amplitude Rise time Duration Burst to data lines: Amplitude Rise time Duration

positiv/negativ

positiv/negativ

acc. to IEC 801-4

acc. to IEC 801-4

3.8
Dimensions Weights

Dimensions & Weights


H = 75 mm, D = 110 mm, W = 100 mm net = approx. 500 g gross = approx. 1000 g

3.9
Ex approval

Approvals
II(2)G [EEx ib] II B/C PTB 02 ATEX 2056

Sartorius

INSTRUCTION MANUAL

PR1626/60

4
4.1

MOUNTING & CONNECTIONS


Mounting

The instrument is suitable for mounting in any position. Before mounting, the instructions in paragraph 2.3 have to be observed. It can either be wall mounted by means of two screws (acc. to DIN 46121 and 43660), or be fixed to a 35 mm rail acc. to DIN 46277 and 50022.

Sartorius

PR1626/60 4.2

INSTRUCTION MANUAL

Connections
All screw terminals of the blue coloured block are provided for the connections to the Exarea. All screw terminals of the black coloured block are provided for the connections to the safe area. The screw terminals for interfacing and terminal PE are provided for 4mm2 at maximum. The screens of the cables are to be connected on the shortest way to the respective screw terminals. In intrinsically safe lines with a cable length of 50 m, the screen of the cables may only be connected at one end to the potential equalization line. In non-intrinsically safe lines the screen of the cables may generally be connected only at one end to the potential equalization system (see sketch below).

To avoid any danger in the event of wires becoming loose and moving into the intrinsically safe connection area of the module causing short circuiting, all wires of the safe side have to be secured together in a cable harness.

10

Sartorius

INSTRUCTION MANUAL 4.2.1 Potential equalization

PR1626/60

According to VDE 0165/6.80, potential equalization is to be ensured within an intrinsically safe installation, especially in hazardous areas of zone 0 and 1. Otherwise impermissible sparking can occur between the particular conducting parts. Therefore terminal PE of the PR1626/60 must be connected to the potential equalization system, which has to be installed and earthed according to VDE 0100 and its part 540, or/and acc. to respective local regulations.

(1)

Conductor for additional potential equalization Leiter fr zustzlichen Potentialausgleich Conducteur pour une liaison quipotentielle additionnel Conductor for potential equalization Potentialausgleichsleiter Conducteur de liaison quipotentielle Conductor for potential equalization Potentialausgleichsleiter Conducteur de liaison quipotentielle Main conductor for potential equalization Hauptpotentialausgleichsleiter Conducteur principale de liaison quipotentielle Earthing conductor Erdungsleiter Conducteur de terre Sub potential equalization terminal Potentialausgleichsklemme (-schiene) Borne (barre) de liaison quipotentielle Main potential equalization terminal Hauptpotentialausgleichsschiene Barre principale de liaison quipotentielle To earth electrode Zum Haupterder (Elektrode) Vers la prise de terre

2,5mm Cu

(2)

4mm Cu

(3)

10mm Cu

(4)

10mm Cu

(5)

10mm Cu

(A) (B) (E)

Sartorius

11

PR1626/60 4.2.2 Terminal allocation

INSTRUCTION MANUAL

12

Sartorius

INSTRUCTION MANUAL

PR1626/60

SETTINGS

Opening and setting of the instrument may only be done after having switched-off the supply voltage, and by observing the instructions as given in paragraph 2.

5.1

Opening the instrument

Upon delivery from the factory the instrument is not completely closed. Sliding of the electronic part into the housing will be prevented by a small sponge, which only has to be removed after setting and connecting the instrument. When closed, and one needs to get access to the inside, the instrument must be opened carefully by spreading out slightly the side parts of the housing, without breaking them. This can be done by using e.g. a small screwdriver. As soon as the little plastic noses become released pull out the upper part.

After opening, the Power Supply Unit (PCB 1), and the Interface Unit (PCB 2) become visible. Both PCBs are fixed to each other by means of 3 bolts and interconnected via flat cable connection.

Sartorius

13

PR1626/60 5.2

INSTRUCTION MANUAL

Settings

5.2.1 Interface Unit (PCB 2)


The Interface Unit can be set by means of the following solder-links which are situated at the component side of the board: MX 7 : MX 8 : Reserved for factory-internal adjustments. Supply voltage to the load cells Closed: 7,5 V DC nominal, if 80 RLC < 150 Open: 12 V DC nominal, if RLC > 150 (delivery state) Type of the connected instrument Closed: for PR1612/00, PR1612/01, PR 614/01 and PR1615 Open: (delivery state) for PR1612/02, PR1613/.., PR1713/.., PR5210/.., PR5510/.., PR5610/.. and PR5710/..

MX 12:

14

Sartorius

INSTRUCTION MANUAL

PR1626/60

6
6.1

MAINTENANCE
General

It is recommended to check the module for proper functioning from time to time. In case of a fault check the connections for possible loose contacts check fuses and replace them if necessary check the supply voltage for correct value observe the maximum loading of inputs/outputs. In doing so, the safety instructions in paragraph 2 must be followed. Should the module be subject to failure, call the Customer Support of your local organization of Sartorius for assistance.

6.2

Fuses

All fuses belong to those parts which influence the intrinsic safety of the PR1626/60. Before exchanging when defective it is a must to evaluate the reason of the defect, and to observe the instructions in paragraph 2. make sure that only fuses with the given nominal current and of the given type are used. Using repaired fuses and/or short-circuiting the fuse holder is strictly forbidden. Fuses shall only be renewed by a qualified person who is aware of the hazard involved. The power supply unit PCB1 is protected by a medium time delayed 450mA fuse, fixed by a holder. The interface unit PCB2 is protected as follows: Each line of the measuring input by 1 medium time delayed 32mA fuse, each fixed by a holder, Each line of the sense voltage input by 1 medium time delayed 32mA fuse, each fixed by a holder.

Sartorius

15

PR1626/60

INSTRUCTION MANUAL

FIGURES/INTERCONNECTION EXAMPLE

16

Sartorius

INSTRUCTION MANUAL

PR1626/60

Sartorius

17

PR1626/60

INSTRUCTION MANUAL

EC-TYPE EXAMINATION CERTIFICATE

18

Sartorius

INSTRUCTION MANUAL

PR1626/60

Sartorius

21

Sartorius Hamburg GmbH Meiendorfer Strae 205 22145 Hamburg, Germany Tel.: +49.40.67960.303 Fax: +49.40.67960.383 www.sartorius.com

Sartorius Hamburg GmbH All rights are strictly reserved Printed in Germany