Microscope
ECLIPSE 50i
ECLIPSE 55i
Repair Manual
Volume-1
NIKON CORPORATION
PAGE REV.MARK REV. DATE PAGE REV.MARK REV. DATE PAGE REV.MARK REV. DATE
1 B 2005.3.2 46 B 2005.3.2 91 B 2005.3.2
2 1 47 B 2005.3.2 92 B 2005.3.2
3 B 2005.3.2 48 B 2005.3.2 93 B 2005.3.2
4 B 2005.3.2 49 B 2005.3.2 94 B 2005.3.2
5 B 2005.3.2 50 B 2005.3.2 95 B 2005.3.2
6 B 2005.3.2 51 B 2005.3.2 96 B 2005.3.2
7 B 2005.3.2 52 B 2005.3.2 97 B 2005.3.2
8 B 2005.3.2 53 B 2005.3.2 98 B 2005.3.2
9 B 2005.3.2 54 B 2005.3.2 99 B 2005.3.2
10 B 2005.3.2 55 B 2005.3.2 100 B 2005.3.2
11 B 2005.3.2 56 B 2005.3.2 101 B 2005.3.2
12 B 2005.3.2 57 B 2005.3.2 102 B 2005.3.2
13 B 2005.3.2 58 B 2005.3.2 103 B 2005.3.2
14 B 2005.3.2 59 B 2005.3.2 104 B 2005.3.2
15 B 2005.3.2 60 B 2005.3.2 105 B 2005.3.2
16 B 2005.3.2 61 B 2005.3.2 106 B 2005.3.2
17 B 2005.3.2 62 B 2005.3.2 107 B 2005.3.2
18 B 2005.3.2 63 B 2005.3.2 108 B 2005.3.2
19 B 2005.3.2 64 B 2005.3.2 109 B 2005.3.2
20 B 2005.3.2 65 B 2005.3.2 110 B 2005.3.2
21 B 2005.3.2 66 B 2005.3.2 111 B 2005.3.2
22 B 2005.3.2 67 B 2005.3.2 112 B 2005.3.2
23 B 2005.3.2 68 B 2005.3.2 113 B 2005.3.2
24 B 2005.3.2 69 B 2005.3.2 114 B 2005.3.2
25 B 2005.3.2 70 B 2005.3.2 115 B 2005.3.2
26 B 2005.3.2 71 B 2005.3.2 116 B 2005.3.2
27 B 2005.3.2 72 B 2005.3.2 117 B 2005.3.2
28 B 2005.3.2 73 B 2005.3.2 118 B 2005.3.2
29 B 2005.3.2 74 B 2005.3.2 119 B 2005.3.2
30 B 2005.3.2 75 B 2005.3.2 120 B 2005.3.2
31 B 2005.3.2 76 B 2005.3.2 121 B 2005.3.2
32 B 2005.3.2 77 B 2005.3.2 122 B 2005.3.2
33 B 2005.3.2 78 B 2005.3.2 123 B 2005.3.2
34 B 2005.3.2 79 B 2005.3.2 124 B 2005.3.2
35 B 2005.3.2 80 B 2005.3.2 125 B 2005.3.2
36 B 2005.3.2 81 B 2005.3.2 126 B 2005.3.2
37 B 2005.3.2 82 B 2005.3.2 127 A
38 B 2005.3.2 83 B 2005.3.2 128 A
39 B 2005.3.2 84 B 2005.3.2 129 A
40 B 2005.3.2 85 B 2005.3.2 130 A
41 B 2005.3.2 86 B 2005.3.2 131 A
42 B 2005.3.2 87 B 2005.3.2 132 A
43 B 2005.3.2 88 B 2005.3.2 133 A
44 B 2005.3.2 89 B 2005.3.2 134 A
45 B 2005.3.2 90 B 2005.3.2 135 A
2
R-4758.B
Table of Contents
2. Specifications................................................................................................................ P.9
2-1. 50i ......................................................................................................................... P.9
2-2. 55i ....................................................................................................................... P.13
3
R-4758.B
Table of Contents
4
Foreword R-4758.B
Foreword
1. This manual is for ECLIPSE 50i and ECLIPSE 55i.
Where an item is common for both ECLIPSE 50i and ECLIPSE 55i, “(For 50i/55i)” is included
in the title.
Where an item is exclusive to either ECLIPSE 50i or ECLIPSE 55i, “(For 50i)” or “(For 55i)” is
included in the title.
For the Stage and Ergonomic Binocular Tube, refer to the ECLIPSE 50i/ECLIPSE 55i
Repair Manual Volume-2.
2. This repair manual has been written to be used as the textbook for the repair training course.
This manual is organized so that trainees can understand the mechanical and optical
construction of the microscope by practicing repair procedures from the beginning to the end.
3. The table of contents of this manual includes every item that might require repair and can
efficiently be used as the guide for repair.
4. Unplug the power cable and wear work-gloves before starting work.
Take extra care to avoid injury from sharp edges or projections and electric shock by touching
live components.
Safety Precautions
WARNING (For ECLIPSE 50i)
The lamp and surrounding areas (including the lamp house) become very hot during and
immediately after a period of illumination.
- To avoid risk of burns, do not touch the lamp or surrounding areas during or
immediately after a period of illumination.
- Make sure the lamp and surrounding areas have cooled sufficiently before attempting
to the replace lamp.
- To avoid risk of fire, do not place fabric, paper or highly flammable volatile materials
such as gasoline, petroleum benzine, paint thinner or alcohol near the lamp house
while the lamp is lit or during a period of around thirty minutes after the lamp has been
turned off.
- To avoid risk of burns, wait at least thirty minutes after the lamp is turned off to give it
sufficient time to cool.
5
1. Tools R-4758.B
T92041 Tweezers AA
Handlap
E719 (Alcohol container)
6
1. Tools R-4758.B
Stage centering
J25298 Centering tool
adjustment
Stage centering
J25481 80i sub stage tool A
adjustment
Stage centering
J25299 Centering rod
adjustment
7
1. Tools R-4758.B
LED position
J25295 Standard straight tube
adjustment
LED position
J25172 Centering telescope
adjustment
LED position
J25260 Ph tool ring
adjustment
1-3. Preparations
Alcohol (or alcohol-ether mixture), Lens tissue, Blower, Gasoline (for cleaning),
Bamboo stick (for applying lubricant, adhesive), Brass (or tin) foil, Dial gauge
For Lubricant and Adhesive: Refer to Repair Parts Lists (50i: KR-7483.A, 55i: KR-7485.A,)
8
2. Specifications R-4758.B
2-1. 50i
I. Overviews
1. Model Number
MBA85010: ECLIPSE 50i, Using halogen lamp as light source L
MBA85020: ECLIPSE 50i, Using halogen lamp as light source R
3. Elevation Block
- Focusing system/Stage up-down system: Using roller races
- Stage up-down stroke: Upward 2 mm/Downward 28 mm
- Sub stage and up-down movement guide: DC integrated type
- Stage mounting system: Round dovetail (not compatible with E series)
- Condenser lens mounting system: Round dovetail (compatible with E series)
- Condenser lens up-down stroke: 27 mm
- Condenser lens stroke upper limit position: Adjustable (using tool)
- Condenser lens centering: 3 mm diameter
9
2. Specifications R-4758.B
2-1. 50i
4. Coarse/Fine Focus Block
- Coarse focus handle: 13.8 mm per rotation (reduction gear ratio: 3)
- Fine focus handle: 0.1 mm per rotation (reduction gear ratio: approx. 413)
- Minimum reading: 1 micron
- Coarse motion handle torque: Adjustable
- Refocusing mechanism: With one-touch refocusing mechanism (with
operation lever)
When the sub stage is lowered to the lower limit position by pushing the lever
down, the sub stage is locked to this position.
When the sub stage is lowered 2 mm further from this position by pushing
the lever down, the lock mechanism is canceled. The sub stage is returned
to the former position.
The rotary damper controls the return of the sub stage at a constant speed.
5. Lighting Block
- Composed of collector lens, field lens (f = 89.7), field stop, and filter
- Filter: ND8 (Detachable)
- Lemon-skin filter (2 pcs.): Built-in (Not detachable)
- Diameter of field stop: 1 to 30 mm diameter (design value: 0.8 to 30.5 mm
diameter)
- Field stop control: Right-hand side of the microscope base
- Filter control: Right-hand side of the microscope base
- Additional filter (diameter: 45 mm, thickness: 3 mm): Up to two filters can be
attached tilting 5 degrees above the field lens.
10
2. Specifications R-4758.B
2-1. 50i
Swing-out Condenser
(MBL11300)
11
2. Specifications R-4758.B
2-1. 50i
2. Restrictions (Following products are not installable.)
- Condenser lens
D-CUD-E Automatic U condensor (dry) (MBL78500)
D-CUD U condensor (dry) (MBL78600)
- Intermediate attachments
D-FL U-EPI Fluorescence (MBE74100)
D-DH-E DIH E (MBD76100)
D-DH DIH M (MBD76200)
- Stage
C-SRR Right handle R stage (MBC71500)
C-SLR Left handle R stage (MBC71600)
D-S-E Automatic XY stage (MBC76100)
12
2. Specifications R-4758.B
2-2. 55i
I. Overviews
1. Model Number
MBA85040: ECLIPSE 55i, Using LED as light source L
MBA85050: ECLIPSE 55i, Using LED as light source R
3. Elevation Block
- Focusing system/Stage up-down system: Using roller races
- Stage up-down stroke: Upward 2 mm/Downward 28 mm
- Sub stage and up-down movement guide: DC integrated type
- Stage mounting system: Round dovetail (not compatible with E series)
- Condenser lens mounting system: Round dovetail (compatible with E series)
- Condenser lens up-down stroke: 27 mm
- Condenser lens stroke upper limit position: Adjustable (using tool)
- Condenser lens centering: 3 mm diameter
13
2. Specifications R-4758.B
2-2. 55i
4. Coarse/Fine Focusing Block
- Coarse focus handle: 13.8 mm per rotation (reduction gear ratio: 3)
- Fine focus handle: 0.1 mm per rotation (reduction gear ratio: approx. 413)
- Minimum reading: 1 micron
- Coarse motion handle torque: Adjustable
- Refocusing function: Refocusing with a single touch of the lever (with
operation lever)
When the sub stage is lowered to the lower limit position by pushing the lever
down, the sub stage is locked to this position.
When the sub stage is lowered 2 mm further from this position by pushing
the lever down, the lock mechanism is canceled. The sub stage is returned
to the former position.
The rotary damper controls the return of the sub stage at a constant speed
5. Lighting Block
- Composed of collector lens, field lens (f = 89.7), field stop, and filter
- Filter: LA60 (Detachable)
- Lemon-skin filter (2 pcs.): Built-in (Not detachable)
- Diameter of field stop: 1 to 30 mm diameter (design value: 0.8 to 30.5 mm
diameter)
- Field stop control: Right-hand side of the microscope base
- Filter control: Right-hand side of microscope base
- Additional filter (diameter: 45 mm, thickness: 3 mm): Up to two filters can be
attached tilting 5 degrees above the field lens.
7. Electric System
- Brightness control dial: Rotary system, located on the right-hand side of the
microscope base.
- Preset switch: Push-button, located on the right-hand side of the microscope
base.
- Lithium-ion battery mounting port: Located on the back of the LED array board.
- Input connector: Located on the back.
- Power connector for Ergo-View cytodiagnostic unit: Located on the back.
- Signal connector for Ergo-View cytodiagnostic unit: Located on the back.
- Hand-switch connector for Ergo-View cytodiagnostic unit: Located on the back.
14
2. Specifications R-4758.B
2-2. 55i
15
2. Specifications R-4758.B
2-2. 55i
7-6-1.Replacing Battery
Remove the lamp house cover and replace the battery in the battery
pocket with a new one.
16
2. Specifications R-4758.B
2-2. 55i
7-10. LED Current Control by Varying Magnification of Ergo-View Cytodiagnostic Unit
When the magnification of the Ergo-View cytodiagnostic unit is set to a
high value, the LED current increases to approx. 3.2 times, which makes
the LED brighter. However, the magnification is limited to the maximum
LED current 22 mA.
The LED current does not depend on the magnification of the Ergo-View
cytodiagnostic unit while the preset switch is ON.
7-13. Indicators
- Power ON/OFF indicator LED (Green): Located on the right side
operation panel of the microscope base
- Lamp house high temperature caution label
- Rated current indication seal
17
2. Specifications R-4758.B
2-2. 55i
J-FL 50i55i
Epi-Fluorescence
Double Port 0/100
Teaching Unit F to F
(MBB74105)
(MBB75100)
Double Port
Teaching Unit S by S type B
(MBB74100)
(MBB75110)
Double Port 0/100
Main Teaching Unit
(MBB74105)
(MBB75120) Magnification
Teaching Unit S by S type A
Module
(MBB75130) Drawing Tube
(MBB75400)
A/A Condenser
C-N Sextuple (MBL71400)
Nosepiece
Swing-out Condenser
J-CY Cytodiagnostic (MBL11300)
Unit (MBP89000)
A/A Condenser
(MBL71400)
LWD Achromat
Condenser
C-SR Left Handle C-SR Right Handle
Stage (MBC71610) Stage (MBC71600) 55i LED L Achromat Swing-out
(MBA85040) Condenser 1-100X
55iLED R (MBL71300)
C-HC1 Specimen (MBA85050)
Holder 1C (MBC75040)
18
2. Specifications R-4758.B
2-2. 55i
2. Restrictions (Following products are not installable.)
- Condenser lens
D-CUD-E Automatic U condenser (dry) (MBL78500)
D-CUD U condensor (dry) (MBL78600)
DF condenser (dry) (MBL12000)
DF condenser (oil) (MBL12001)
PH condenser (MBL73100)
- Intermediate attachments
D-FL U-EPI Fluorescence (MBE74100)
D-DH-E DIH E (MBD76100)
D-DH DIH M (MBD76200)
C-ISA simple analyzer (MBB75350)
C-IA Analyzer (MBB75360)
- Stage
C-SRR Right handle R stage (MBC71500)
C-SLR Left handle R stage (MBC71600)
D-S-E Automatic XY stage (MBC76100)
- Others
Simple polarizer (MBB75310)
Polarizer with first-order red tint plate (MDW10200)
19
3. Optical Systems R-4758.B
3-1. 50i
The following optical system is based on the ECLIPSE 50i.
The cemented lens surfaces are omitted from the drawing because the lens direction only
is of concern here.
Field stop
6V 30W
Diffuser ND8
20
3. Optical Systems R-4758.B
3-2. 55i
Evaporated
surface
LED arrays Mirror
21
4. Connection Diagrams R-4758.B
4-1. 50i
22
4. Connection Diagrams R-4758.B
4-2. 55i
23
5. Disassembling/Assembling/Adjusting R-4758.B
85 67 68
Removing power supply assembly Fine torque adjustment Removing field stop block
29 89
Removing elevation block Disassembling/Assembling
brightness control board
NOTE: The number at upper right of the frame shows the page of this item.
24
5. Disassembling/Assembling/Adjusting R-4758.B
103 67 68
Removing control board Fine torque adjustment Removing field stop block
29 107
Disassembling/Assembling
Removing elevation block
switch board
Condenser holder block
29 42
Disassembling/Assembling Color compensation filter unit
Removing condenser holder condenser holder 95
Removing color compensation
filter unit
30 47
Disassembling/Assembling Knob/rack engaging adjustment
elevation block 96
Disassembling/Assembling
34 color filter unit
38
Light source Unit Diffuser unit
Installing elevation block 103 99
Removing control board Removing diffuser unit
113
104 100
Stage tilt adjustment
Disassembling/Assembling Disassembling lens/diffuser
LED board
116
105 101
Stage centering adjustment Cleaning adhesive coating surface
Installing control board
of mirror/diffuser
118 101
123
Condenser holder tilt adjustment LED position adjustment Installing diffuser unit
39 69 74
Installing fine/coarse focus block Disassembling/Assembling blade Cleaning adhesive coating surface
of mirror/lens
40 72
Elevation block movement check 74
Blade mount position adjustment
using fine/coarse focus handle Cementing mirror/lens
72
40
Installing field stop block
Re-focusing mechanism
operation check
73
Diaphragm open-shut dial
operation check
NOTE: The number at upper right of the frame shows the page of this item.
25
5. Disassembling/Assembling/Adjusting R-4758.B
Loosen the eyepiece clamp screw beneath Loosen the clamp screw on the side of the
the eyepiece tube, and remove the nosepiece, and remove the nosepiece.
eyepiece tube.
Nosepiece
Loosen the set screw on the front side of Loosen the set screw on the condenser
the stage using a hex-key, and remove the receptacle using a hex-key, and remove the
stage. condenser lens.
26
5. Disassembling/Assembling/Adjusting R-4758.B
- Disassembling the elevation block is not usually necessary, but if liquid (water etc.)
has dropped into the microscope, it should be disassembled.
- The two roller races (#A3) fixed to the attaching plate must not to be disassembled at
the same time. When reassembling one roller race, the other (assembled) roller race
is used as a reference for adjustment.
- Some screws are fixed with adhesive. Apply alcohol to melt the adhesive on these
screws.
5-3-1. Disassembling
(a)
(a) (b)
27
5. Disassembling/Assembling/Adjusting R-4758.B
7. In the case of 50i, disconnect the lamp Plastic cover (#202) Cap (#233)
(#227) to prevent damage.
8. Pull out the hexagonal driver (#248).
Hexagon
9. Remove the four hexagon socket head cap socket head
bolts (M3 x 6) (#254), and remove the cap bolts
plastic cover (#202). (M3 x 6)(#254)
Fine/coarse
focus block
14. Remove the two hexagon socket head cap bolts (M3 x 10) (#51), and remove the
plastic cover (#411).
15. Insert a (-) 2 mm screwdriver into the spaces on both sides of the plastic cover
shown in the photograph to release the lock of the claws.
[NOTE]
When releasing the lock, take care not to break the claws.
Hexagon socket head cap bolts (M3 x 10) (#51) (-) screwdriver Claws of plastic cover
28
5. Disassembling/Assembling/Adjusting R-4758.B
19. Remove the three hexagon socket head Hexagon socket head
Condenser holder cap bolts (M3 x 8) (#442)
cap bolts (M3 x 8) (#442), and remove the
condenser holder block.
29
5. Disassembling/Assembling/Adjusting R-4758.B
20. Remove the two hexagon socket head cap bolts (M3 x 6) (#441) on the front and
two hexagon socket head cap bolts (M3 x 8) (#442) on the rear. Remove the plastic
cover (#410).
Hexagon socket head cap bolts (M3 x 8) (#442)
Hexagon socket
head cap bolts
(M3 x 6) (#441)
30
5. Disassembling/Assembling/Adjusting R-4758.B
Rack (#406)
27. Remove the rack, coil spring (#405), coil
spring (#404) and spring guide (#403). Coil spring
(#404)
Coil spring
(#405)
Spring guide
(#403)
Sub body
Sub body
31
5. Disassembling/Assembling/Adjusting R-4758.B
29. Place the elevation block right-hand side Set screws (#447) Elevation block
down, and fully loosen the four set screws
(#447). Sub body
30. Remove the four hexagon socket head
cap bolts (M3 x 16) (#418), and remove
the roller race (#A3) on the left side of the
elevation block. Remove the sub body
from the elevation Block.
[NOTE]
When removing the roller race, the
Roller race (#A3) Hexagon socket head cap
cylindrical rollers (12 pcs each on up and bolts (M3 x 16) (#418)
lower sides) might fall. Be careful not to
lose them.
The guide rails (#417) and separators
(#408) are also removed at the same time.
32
5. Disassembling/Assembling/Adjusting R-4758.B
5-3-2. Assembling
Attaching plate
1. Paste the two guide rails to the two roller Guide rails (#417)
(#402)
races using lubricant L4010 as a substitute Sub body
for adhesive.
2. Paste the guide rails to the races on both
sides of the sub body using lubricant
L4010 in the same way as step 1.
4. Apply lubricant L4010 to the roller race on the sub body, and place the separator and
twelve cylindrical rollers on the race.
5. Assemble the sub body to the roller race on the attaching plate, and lightly tighten the
roller race with the four hexagon socket head cap bolts (M3 x 16) while pressing it to
the sub body.
Cylindrical rollers
Sub body
(12 pcs) (#409) Roller race (#A3)
Sub body
Separator
(#408)
33
5. Disassembling/Assembling/Adjusting R-4758.B
6. Move the sub body fully right and left, and adjust the position of the separator so that
the edges of the separator and the roller race coincide.
Also adjust the position of the separator on the opposite side in the same way.
7. Move the sub body so that the set screw Set screws (#447)
on the attaching plate is adjacent to the
cylindrical roller.
8. Tighten the set screw close to the
cylindrical roller with the following
tightening torque using torque driver.
Tightening torque: 1 kg x cm
9. Repeat steps 7 and 8, and tighten the
remaining three set screws in the same
way. Hexagon socket head
Roller race (#A3)
10. Tighten the four hexagon socket head cap cap bolts (M3 x 16)(#418)
bolts to fix the roller race.
11. Move the sub body, and check the
following: Sub body
- The cylindrical roller and separator are
still in position.
- The sub body moves smoothly and Spring scale
lightly without slack.
Push the sub body and check, using a
spring scale, that 50 to 100g pressure is
required to effect movement. If not to meet
the specification, repeat steps 7 to 10.
[NOTE]
When the four hexagon socket head cap bolts securing the roller race are firmly
tightened, the sub body sometimes loses its movement. In this case, perform steps
7 to 10 again.
34
5. Disassembling/Assembling/Adjusting R-4758.B
14. Put the spring guide into the hole of the sub body, and insert its tip into the center
hole on the spring retainer (#432).
[NOTE]
Be sure to insert the spring guide tip into the spring retainer hole.
15. Apply a light coating of lubricant L3047 to the whole surfaces of the two kinds of coil
springs.
16. Fit the lubricated coil springs into the hole of the sub body.
Tip of the Spring retainer (#432)
Coil spring (#404)
spring guide
Sub body
35
5. Disassembling/Assembling/Adjusting R-4758.B
Rack (#406)
Set screw (#452)
19. Install the shaft supporter to the sub body while inserting the spring guide tip into the
center hole of the shaft supporter, and tighten it using the two hexagon socket head
cap bolts (M3 x 6).
[NOTE]
After tightening the bolts, check again that the spring guide tip is inserted into the
spring retainer hole on the opposite side.
Tip of the
spring guide
(5013_1)
Sub body
36
5. Disassembling/Assembling/Adjusting R-4758.B
21. Tighten the hexagon socket head cap bolt (M3 x 12) of the stroke limitation screw.
22. Tighten the plate spring (with rubber plate) using the two hexagon socket head cap
bolts (M3 x 6).
When tightening, adjust the position of the plate spring so that the edge of the plate
spring is parallel to the edge of the boss on the attaching plate.
23. Tighten the cap and rotary damper using the two hexagon socket head cap bolts
(M3 x 8).
Hexagon socket
Hexagon socket head cap
head cap bolts Make parallel the edges of the plate spring
bolt (M3 x 12) (#443)
(M3 x 8) (#442) with the boss on the attaching plate.
(Stroke limitation screw)
24. Tighten the plastic cover using the two hexagon socket head cap bolts (M3 x 6) on
the front and two hexagon socket head cap bolts (M3 x 8) on the rear.
Hexagon socket
head cap bolts
(M3 x 6) (#441)
37
5. Disassembling/Assembling/Adjusting R-4758.B
26. While holding the elevation block so that it does not drop, tighten it using the six
hexagon socket head cap bolts (M4 x 16).
[NOTE]
Be sure to tighten the bolts while pushing the elevation block in the direction of the
arrows.
Hexagon socket head cap bolts (M4 x 16) (#53) Elevation block
27. While holding the elevation block, place the microscope upside down.
[NOTE]
Since the fine/coarse focus block is removed, the elevation block is not fixed.
Therefore, when placing the microscope to the original position, hold the elevation
block by hand.
38
5. Disassembling/Assembling/Adjusting R-4758.B
[Assembling positions]
With the main body turned upside down (photo):
・ Elevator: Lowest position
・ Coarse/fine focus drive: Where the limit
screw of the gear contacts the lower limit,
as shown in the photograph on the right.
(This applies to both right-handle and Limit screw: To be assembled
left-handle designs.) at this position
Bottom view
After these adjustments are completed,
place the plastic cover under the fine/coarse focus block, and hold these parts by hand firmly.
While engaging the gear of the fine/coarse focus block with the rack of the elevation
block, insert the two pins on the microscope into the two standard holes on the
fine/coarse focus block. Tighten the fine/coarse focus block using the two hexagon
socket head cap bolts (M3 x 12).
[NOTE]
The fine/coarse focus block can be installed to the microscope in both (right and left)
directions. Be sure to install it in the same orientation as disassembling.
Hexagon socket
Standard holes head cap bolts
(M3 x 12) (#52)
Pins
39
5. Disassembling/Assembling/Adjusting R-4758.B
elevator
1mm #402
30. Put the two claws on right and left sides of the plastic cover (#411) into the two
longitudinal holes of the plastic cover (#410).
31. Tighten the plastic cover using the two hexagon socket head cap bolts (M3 x 10).
Longitudinal holes
Claws
40
5. Disassembling/Assembling/Adjusting R-4758.B
Lamp (#227)
41
5. Disassembling/Assembling/Adjusting R-4758.B
- Disassembling the condenser holder block is not usually necessary, but if liquid (water
etc.) has dropped into the microscope, it should be disassembled.
- The two dovetails (#437 and #438) fixed to the condenser holder block must not be
disassembled at the same time. When re-assembling the dovetail, the other
(assembled) dovetail is used as a reference for assembling.
- Some screws are fixed with adhesive. Apply alcohol to melt the adhesive on these
screws.
5-4-1. Disassembling
Hexagon socket
Condenser head cap bolts
1. Remove the elevation block from the
holder block (M3 x 8)(#442)
microscope base according to the steps 1
to 18 of “Disassembling” in elevation block.
(Refer to Section 5-3.)
2. Remove the three hexagon socket head
cap bolts (M3 x 8) (#442), and remove the
condenser holder block from the elevation
block.
Set screws
(#446)
42
5. Disassembling/Assembling/Adjusting R-4758.B
Hexagon socket
Block (#422) head cap bolt (#445)
(Stroke limitation screw)
8. If there is a rusty section on the mounting surface of the dovetail and block, scrape it
using a scraper.
43
5. Disassembling/Assembling/Adjusting R-4758.B
13. Remove the two hexagon socket head cap bolts (M3 x 6) (#441), and remove the
spring retainer (#436).
14. Remove the condenser receptacle (#421) from the condenser holder.
[NOTE]
The plate spring (#435) is removed at the same time.
Spring retainer (#436)
Plate spring (#435)
Condenser
receptacle
(#421)
Condenser holder
Hexagon socket head
Condenser holder cap bolts (M3 x 6) (#441)
44
5. Disassembling/Assembling/Adjusting R-4758.B
45
5. Disassembling/Assembling/Adjusting R-4758.B
5-4-2. Assembling
When the male dovetail, rack, condenser receptacle, and knob were removed, assemble
these parts first, and then assemble the condenser holder according to the steps 1 to 13.
The assembling procedures for the male dovetail, rack, condenser receptacle and knob
are as follows:
- Assembling for male dovetail and rack: steps 14 to 17
- Assembling for condenser receptacle: steps 18 to 22
- Assembling for knob: steps 23 to 28
46
5. Disassembling/Assembling/Adjusting R-4758.B
Hexagon socket
Block (#422) head cap bolt (#445)
(Stroke limitation screw)
Block (#422)
ブロック(#422)
Knob block
7. Put the rod into the hole of the shaft Knob (#428) Rod
supporter through the hole of the block.
While holding the knob, move the rod in
the direction shown in the photograph. The
gear should engage firmly with the rack.
Check that the condenser holder moves
smoothly by turning the knob.
If the rotation of the knob is heavy, perform
the adjustment again using the rod.
47
5. Disassembling/Assembling/Adjusting R-4758.B
10. Install the condenser holder block, and Hexagon socket head
Condenser holder cap bolts (M3 x 8) (#442)
firmly move it in the direction of the arrow.
Tighten it with three hexagon socket head
cap bolts (M3 x 8) to the elevation block.
[NOTE]
Be sure to install the condenser block
while moving it in the direction of the arrow.
11. While holding the elevation block by hand so that it does not drop, tighten it using the
six hexagon socket head cap bolts (M4 x 16).
[NOTE]
Be sure to install the elevation block while pushing it in the direction of the arrows.
Hexagon socket head
cap bolts (M4 x 16) (#43) Elevation block
48
5. Disassembling/Assembling/Adjusting R-4758.B
ラック(#456)
Rack (#456)
Special hexagon
bolts (M3 x 4) (#457)
49
5. Disassembling/Assembling/Adjusting R-4758.B
16. Put the two parallel rods (2 mm diameter) into the grooves of the male dovetail and the
holes of the condenser holder. This sets the relative positions of the male dovetail and
the condenser holder.
17. Firmly tighten the two hexagon socket head cap bolts lightly tightened in step 15.
After tightening the bolts, pull out the two parallel rods.
Condenser holder
50
5. Disassembling/Assembling/Adjusting R-4758.B
28. Tighten the set screw to fix the knob with the gear.
After tightening the set screw, apply adhesive #757 to the screw threads.
51
5. Disassembling/Assembling/Adjusting R-4758.B
5-5-1. Disassembling
- Some screws are fixed with adhesive. Apply alcohol to melt the adhesive on these
screws.
- The following steps 1) to 8) describe the procedure with the fine/coarse focus block
installed to the microscope. But steps 1) to 8) can also be performed when the
fine/coarse focus block has been removed.
Fine motion
handle (#324)
52
5. Disassembling/Assembling/Adjusting R-4758.B
4. Remove the fine motion handle (#303) (with the fine motion handle rubber cap (#301))
on the opposite side from the microscope base.
The fine motion axle (#323) is removed at the same time.
If replacement of the fine motion handle or the fine motion axle is required, replace
these parts after removing the fine motion handle rubber cap (#301) and circular nut
(#302).
Fine motion
axle (#323)
Fine motion
Fine motion handle (#303) (with handle (#303) Fine motion handle
the fine motion handle rubber cap) rubber cap (#301)
53
5. Disassembling/Assembling/Adjusting R-4758.B
8. Pull out the coarse motion handle (#304) Remove the friction disk (#318)
from the microscope. and two wave washers (#316).
The coarse motion handle with the gear
Coarse motion
case (#307) and coarse motion axle (#310)
handle (#304)
are removed.
Remove the friction disk (#318) and two
wave washers (#316).
[NOTE]
When removing the coarse motion handle,
the ball bearing (#337) may be detached at
the same time. Remove the ball bearing
using tweezers etc.
Check the rotation of the removed ball
bearing using tweezers etc.
If replacement of the gear case or the
coarse motion axle is required, hold the
replacement parts firmly, and turn the other
parts counterclockwise to remove.
Perform the following steps when the fine/coarse focus block is removed from the
microscope base.
Special hexagon bolts
9. Remove the eyepiece tube. (Refer to Bottom cover (M3 x 6) (#239)
Section 5-2.)
10. Remove the objective lens. Special
11. Remove the condenser lens. hexagon bolts
(Refer to Section 5-2.) (M3 x 6) (#239)
12. Remove the stage. (Refer to Section 5-2.)
13. Place the microscope upside down. Fine/coarse
focus block
14. Remove the five special hexagon bolts
(M3 x 6) (#239), and remove the bottom
cover. Plastic cover (#411)
54
5. Disassembling/Assembling/Adjusting R-4758.B
15. Remove the two hexagon socket head cap bolts (M3 x 10) (#51), and remove the
plastic cover (#411).
16. Insert a (-) 2 mm screwdriver into the spaces on both sides of the plastic cover shown
in the photograph to release the lock of the claws.
[NOTE]
When releasing the lock, take care not to break the claws.
Hexagon socket head cap bolts (M3 x 10) (#51) (-) screwdriver Claws of plastic cover
55
5. Disassembling/Assembling/Adjusting R-4758.B
19. Remove the Teflon washer (#336), gear Gear (#326) Teflon washer
(#326) and Teflon washer (#336) from the (#336)
removed gear axle.
Remove the hexagon socket head cap bolt
Gear
Gear axle
(#354) tightened to the gear. (#327)
Teflon washer
(#336) Hexagon
Hexagon socket
socket
head
head
capcap
boltbolt
(#354)
(#354)
Pinion (#309)
56
5. Disassembling/Assembling/Adjusting R-4758.B
22. Remove the two kinds of ball bearings Ball bearing (#338) and
(#337 and #338) and special washer special washer (#339)
(#339) from the pinion using tweezers etc.
Check the rotation of the removed ball
bearings using tweezers etc.
[NOTE]
- Special washer sometimes remains in the
pinion housing. When removing, be very
careful.
Pinion (#309) Ball bearing
- When removing the coarse motion axle, (#337)
the ball bearing (#337) is sometimes
removed at the same time.
23. Remove the three hexagon socket head Three hexagon socket head
cap bolts (M3 x 8), and remove the coarse cap bolts (M3 x 8) (#351)
57
5. Disassembling/Assembling/Adjusting R-4758.B
Gear (#319)
58
5. Disassembling/Assembling/Adjusting R-4758.B
5-5-2. Assembling
Pinion (#309)
59
5. Disassembling/Assembling/Adjusting R-4758.B
Teflon washer
(#336)
Teflon washer
Gear (#326)
(#336)
7. Tighten the gear axle to the pinion housing Gear axle (#327)
using the two hexagon socket head cap
bolts (M3 x 10).
Although the following steps can be performed where the fine/coarse focus block is
removed from the microscope base, the following describes the assembling steps where
the fine/coarse focus block is attached to the microscope base.
60
5. Disassembling/Assembling/Adjusting R-4758.B
8. Place the plastic cover under the fine/coarse focus block, and hold these parts by
hand firmly. While engaging the gear of the fine/coarse focus block with the rack of the
elevation block, insert the two pins on the microscope into the two standard holes on
the fine/coarse focus block. Tighten the fine/coarse focus block using the two hexagon
socket head cap bolts (M3 x 12).
[NOTE]
The fine/coarse focus block can be installed to the microscope in both (right and left)
directions. Be sure to install it in the same orientation as disassembling.
Standard Gear Hexagon socket head
holes cap bolt (M3 x 12) (#52)
Pins
Rack
9. Put the two claws on right and left sides of the plastic cover into the two longitudinal
holes of the plastic cover.
10. Tighten the plastic cover using the two hexagon socket head cap bolts (M3 x 10).
Hexagon socket head cap bolts (M3 x 10) (#51)
61
5. Disassembling/Assembling/Adjusting R-4758.B
11. Tighten the bottom cover using the five Special hexagon bolts (M3 x 6) (#239)
Bottom cove
special hexagon bolts (M3 x 6).
Special hexagon
bolts (M3 x 6) (#239)
Gear (#319)
14. Put the gear retaining plate to the gear Hexagon socket head cap bolts (M3 x 8) (#351)
case so that the slightly curved surface of
the gear retaining plate faces to the gear,
and tighten it using the two hexagon
socket head cap bolts (M3 x 8).
[NOTE]
When assembling, be sure to install each
shaft of the three gears into each hole on
the gear retaining plate.
Gear retaining plate (#308)
After assembling, check that all gears
rotate smoothly.
62
5. Disassembling/Assembling/Adjusting R-4758.B
63
5. Disassembling/Assembling/Adjusting R-4758.B
Four holes
Friction disk (#318)
Friction disk (#318)
19. Install the coarse motion axle (with coarse Pinion housing
motion handle) to the pinion housing.
64
5. Disassembling/Assembling/Adjusting R-4758.B
Grooved surface
Outward this side
Coarse
Coarse motion
motion axle
axle (#310)
(#310)
23. Apply lubricant L3047 to the surface with V-groove of the thrust retainer.
24. Screw the thrust retainer to the coarse motion axle until it stops. Insert a hex-key into
the hole, and tighten the two set screws.
25. Apply adhesive #757 to the threads of the two set screws.
Hex-key
Apply lubricant to the V-grooved surface
65
5. Disassembling/Assembling/Adjusting R-4758.B
26. Apply lubricant L3047 to the specified portions of the fine motion axle (with the fine
motion handle) as shown in the photograph, and insert it into the gear case.
Apply lubricant
微動軸(#323)
Fine motion axle (#323)
Fine motion
handle (#324)
66
5. Disassembling/Assembling/Adjusting R-4758.B
Specification: 11.5 to 21 g
(Fine motion torque: 35 to 65 gf x cm)
Wrap the string around the fine focus handle cover, and hook the spring
scale on the end of the string. Move the scale in the direction of the arrow. Fine focus handle
cover (#325)
31. After adjustment is completed, remove the fine focus handle cover once, and then
firmly tighten the circular nut.
After tightening, check that the fine motion handle is not slack even if the fine motion
handle on the opposite side is turned.
32. Apply adhesive #757 to the threads of the
circular nut.
33. Install the fine focus handle cover to the
fine motion handle.
67
5. Disassembling/Assembling/Adjusting R-4758.B
- Some screws are fixed with adhesive. Apply alcohol to melt the adhesive on these
screws.
5-6-1. Disassembling
- If the mirror or the lens attached on the lens holder of the field stop block is
disassembled, perform steps 1 to 9 and steps 13 to 17.
Lever (#210)
68
5. Disassembling/Assembling/Adjusting R-4758.B
9. Remove the three hexagon socket head Hexagon socket head cap Field stop block
cap bolts (M3 x 12) (#256) fixing the field bolts (M3 x 12) (#256)
stop block.
10. Remove the four hexagon socket head cap bolts (M3 x 6) (#254), and remove the
blade mount (#207).
[NOTE]
When the blade mount is removed, the eight blades (for 50i; #A1, for 55i: #260) are
removed at the same time.
Be careful not to lose them.
Be very careful not to bend or apply lubricant to the blade.
69
5. Disassembling/Assembling/Adjusting R-4758.B
13. Scrape off the adhesive joints between the Mirror (#2) Adhesive
mirror (#2) and the lens holder (#204).
15. Scrape off the adhesive from the perimeter Lens (#3)
of the lens (#3). Lens holder (#204)
Adhesive
17. Remove any remaining adhesive on the lens holder after the lens or the mirror is
removed completely. Then, clean the surface with alcohol.
70
5. Disassembling/Assembling/Adjusting R-4758.B
5-6-2. Assembling
- The assembling procedures for the lens and mirror are explained in steps 22 to 28.
Shaft
2. Insert the shaft of the blade guide way into Blade guide way (#206)
the groove of the lens holder, and fully turn
it in the direction of the arrow.
71
5. Disassembling/Assembling/Adjusting R-4758.B
4. Install the blade mount so that the slightly 2 mm dia. rod Projection of the blade
curved surface of it faces the blade. 2 mm dia.
Blade mount rod
[NOTE]
When assembling, be sure to insert the
projection of the blade into the hole of the
blade mount.
72
5. Disassembling/Assembling/Adjusting R-4758.B
Special
washer
(#209)
Lever (#210)
U-groove
73
5. Disassembling/Assembling/Adjusting R-4758.B
28. After the lens and mirror are cemented, blow away any dust from the lens and the
mirror using an airbrush or the like.
74
5. Disassembling/Assembling/Adjusting R-4758.B
5-7-1. Disassembling
Special
hexagon bolts
(M3 x 6) (#239)
75
5. Disassembling/Assembling/Adjusting R-4758.B
10. Release three claws of the ND filter holder ND filter holder (#215)
(#215), and remove the ND filter (#6). Claws
ND filter (#6)
11. Remove the projections on the ND filter holder from the holes on the lever (#216) and
the lever supporter (#217).
Lever (#216)
Projection
Projections
76
5. Disassembling/Assembling/Adjusting R-4758.B
5-7-2. Assembling
2. Install the ND filter holder to the two levers and lever supporter in the direction as
shown in the photograph.
3. Put the projections on the ND filter holder into each hole on the two levers and the
lever supporter.
ND filter (#6)
(An evaporated surface faces upward.)
77
5. Disassembling/Assembling/Adjusting R-4758.B
5. Put the harness into the hole on the ND Clamper ND filter block
filter block, and secure it using the clamper.
Harness connector
6. Tighten the ND filter block using a hexagon Hexagon socket head cap ND filter block
socket head cap bolt (M3 x 8). bolt (M3 x 8) (#255)
7. Secure the harness using a hexagon head Brightness
cap bolt (M3 x 8). controller
8. Connect the harness connector to the board (#501)
brightness controller board.
Special hexagon
9. Tighten the bottom cover using the five Bottom cover bolts (M3 x 6) (#239)
special hexagon bolts (M3 x 6).
Special
hexagon bolt
(M3 x 6) (#239)
78
5. Disassembling/Assembling/Adjusting R-4758.B
- Some screws and diffuser are fixed with adhesive. Apply alcohol to melt the
adhesive on these.
5-8-1. Disassembling
(a)
(a) (b)
Lamp (#227)
79
5. Disassembling/Assembling/Adjusting R-4758.B
13. Remove the four hexagon socket head cap Hexagon socket head
Lighting block cap bolts (M3 x 8) (#255)
bolts (M3 x 8) (#255), and remove
the lighting block.
14. Remove the two diffusers (#4 and #5) from Diffuser (#5) Diffuser (#4)
the lens holder (#221).
80
5. Disassembling/Assembling/Adjusting R-4758.B
16. Remove the four hexagon socket head cap bolts (M3 x 6), and remove the collector
lens (#1) and lamp socket holder (#226).
[NOTE]
When removing the lamp socket holder, the collector lens is removed at a same time.
Be careful not to drop or scratch the lens.
Collector lens
Lamp socket
(#1)
holder (#226)
Hexagon socket head cap bolts
(M3 x 6)(#254)on both sides
81
5. Disassembling/Assembling/Adjusting R-4758.B
5-8-2. Assembling
Collector lens
Lamp socket
(#1)
holder (#226)
Hexagon socket head cap bolts
M3 x 6)(#254)on both sides
4. Apply adhesive #757 (Clear) to the side Diffuser (#5) Diffuser (#4)
edges of the two diffusers which contact
the lens holder. Mat surface
5. Install the two diffusers to the lens holder in
the orientation shown in the photograph.
[NOTE]
Be sure to install the mat surface of the
diffusers in the orientation shown in the Mat surface
photograph.
Lens holder (#221)
82
5. Disassembling/Assembling/Adjusting R-4758.B
Special
hexagon bolts
(M3 x 6) (#239)
Lamp (#227)
83
5. Disassembling/Assembling/Adjusting R-4758.B
84
5. Disassembling/Assembling/Adjusting R-4758.B
5-9-1. Disassembling
(a)
(a) (b)
Lamp (#227)
6. Remove the four hexagon socket head Hexagon socket head cap
cap bolts (M3 x 6) (#254), and remove the bolts (M3 x 6) (#254)
85
5. Disassembling/Assembling/Adjusting R-4758.B
Connector (CN3)
Connector (CN2)
86
5. Disassembling/Assembling/Adjusting R-4758.B
5-9-2. Assembling
Connector (CN3)
Connector (CN2)
4. Tighten the attaching plate (#231) using Hexagon socket head cap
the four hexagon socket head cap bolts bolts (M3 x 6) (#254)
(M3 x 6).
Attaching plate
Attaching plate (#231)
(#231)
87
5. Disassembling/Assembling/Adjusting R-4758.B
Lamp (#227)
(a)
(a) (b)
88
5. Disassembling/Assembling/Adjusting R-4758.B
5-10-1. Disassembling
8. Remove the field stop block. (Refer to Brightness controller Connector (J1)
Section 5-6.) board (#501)
9. Disconnect the harness connector (J1) on
the brightness controller board (#501).
10. Remove the two hexagon socket head cap
bolts (M3 x 6) (#254), and remove the
brightness controller board.
89
5. Disassembling/Assembling/Adjusting R-4758.B
11. Remove the two special hexagon bolts Special hexagon bolts (M3 x 6) (#239)
(M3 x 6) (#239), and remove the two
bending plates from the brightness
controller board.
90
5. Disassembling/Assembling/Adjusting R-4758.B
5-10-2. Assembling
Brightness controller
2. Tighten the brightness controller board Connector (J1)
board (#501)
using the two hexagon socket head cap
bolts (M3 x 6).
3. Connect the harness connector to the
connector (J1) on the brightness controller
board.
Special
hexagon bolts
(M3 x 6) (#239)
91
5. Disassembling/Assembling/Adjusting R-4758.B
92
5. Disassembling/Assembling/Adjusting R-4758.B
5-11-1. Disassembling
side
Claw
93
5. Disassembling/Assembling/Adjusting R-4758.B
5-11-2. Assembling
1. Install the seesaw switch to the microscope, and secure it by the claws located on
both sides.
[NOTE]
Take care to install the correct orientation of the seesaw switch.
2. Connect the terminals to the seesaw switch.
[CAUTION]
Make sure you always connect to the correct wire to each pin.
1 pin: black wire of cable with AC socket (#503)
2 pin: black wire of cable with connector (#506)
3 pin: white wire of cable with AC socket (#503)
4 pin: white wire of cable with connector (#506)
Terminal (1 pin)
side
side side
side
Terminals location
Claw
Special
hexagon bolts
(M3 x 6) (#239)
94
5. Disassembling/Assembling/Adjusting R-4758.B
5-12-1. Disassembling
8. Remove a hexagon socket head cap bolt Color compensation filter unit
(M3 x 8) (#255), and remove the color
compensation filter unit.
95
5. Disassembling/Assembling/Adjusting R-4758.B
10. Remove the projections on the ND filter holder from each hole on the two levers
(#216) and the lever supporter (#217).
96
5. Disassembling/Assembling/Adjusting R-4758.B
5-12-2. Assembling
2. Install the ND filter holder to the lever supporter in the direction shown in the
photograph.
3. Put the projections on the ND filter holder into each hole on the two levers.
Projections
97
5. Disassembling/Assembling/Adjusting R-4758.B
5. Put the connector of the flat cable through Color compensation filter unit
the hole on the color compensation filter
unit.
6. Tighten the color compensation filter unit
using a hexagon socket head cap bolt
(M3 x 8).
Special hexagon
8. Tighten the bottom cover using the five Bottom cover bolts (M3 x 6) (#239)
special hexagon bolts (M3 x 6).
Special
hexagon bolt
(M3 x 6) (#239)
98
5. Disassembling/Assembling/Adjusting R-4758.B
5-13-1. Disassembling
99
5. Disassembling/Assembling/Adjusting R-4758.B
5-13-2. Assembling
1. Clean the adhesive coating surface of the Diffuser (#4) (Mat surface is lowered.)
diffuser on the lens holder using alcohol.
2. Apply adhesive #2061 to the adhesive
Adhesive
coating surfaces of the diffuser on the lens
holder.
[NOTE]
Do not mistake the installing position of the
diffuser and the lens to the lens holder.
Lens (#1)
Large curved surface
Adhesive
100
5. Disassembling/Assembling/Adjusting R-4758.B
Special
hexagon bolt
(M3 x 6) (#239)
101
5. Disassembling/Assembling/Adjusting R-4758.B
5-14-1. Disassembling
(a)
(a) (b)
Connector (J4)
102
5. Disassembling/Assembling/Adjusting R-4758.B
7. Remove the two special hexagon bolts Special hexagon bolts (M3 x 6)(#239)
(M3 x 6) (#239) fixing the LED board
(#503).
Connector
9. While pressing the claw on the battery case (#224), remove the LED board from the
battery case.
[NOTE]
When removing the LED board, be careful not to deform the terminals on the LED
board.
LED board (#503) LED board (#503)
Claw
Terminals
Battery case (#224)
103
5. Disassembling/Assembling/Adjusting R-4758.B
5-14-2. Assembling
2. Connect the connector (J2) on the LED board to the control board.
3. Tighten the LED board to the control board using the two special hexagon bolts
(M3 x 6).
Connector
Connector
Connector (J4)
(J4)
Control
Control board
board (#509)
(#509)
104
5. Disassembling/Assembling/Adjusting R-4758.B
Hexagon socket
head cap bolts
(M3 x 6) (#254)
9. After the assembling is completed, perform the LED position adjustment. (Refer to
Section 5-21.)
105
5. Disassembling/Assembling/Adjusting R-4758.B
5-15-1. Disassembling
Special hexagon
6. Place the microscope upside down. Bottom cover bolts (M3 x 6) (#239)
7. Remove the five special hexagon bolts
(M3 x 6) (#239), and remove the
Special
bottom cover.
hexagon bolts
(M3 x 6) (#239)
8. Remove the field stop block. (Refer to Switch board (#502) Connector (J1)
Section 5-6.)
9. Disconnect the flat cable with connector
(#507) from the connector (J1) on the
switch board (#502).
10. Remove the two hexagon socket head cap
bolts (M3 x 6) (#254), and remove the Flat cable with
switch board. connector
(#507)
106
5. Disassembling/Assembling/Adjusting R-4758.B
11. Remove the two special hexagon bolts Special hexagon bolts (M3 x 6) (#239)
(M3 x 6) (#239), and remove the two
bending plates from the switch board.
107
5. Disassembling/Assembling/Adjusting R-4758.B
5-15-2. Assembling
1. Tighten the two bending plates to the Special hexagon bolts (M3 x 6) (#239)
switch board using the two special
hexagon bolts (M3 x 6).
2. Tighten the switch board using the two Switch board (#502) Connector (J1)
hexagon socket head cap bolts (M3 x 6).
3. Connect the flat cable with connector to
the connector (J1) on the switch board.
108
5. Disassembling/Assembling/Adjusting R-4758.B
109
5. Disassembling/Assembling/Adjusting R-4758.B
5-16-1. Disassembling
side
Claw
110
5. Disassembling/Assembling/Adjusting R-4758.B
5-16-2. Assembling
1. Install the seesaw switch to the microscope, and secure it by the claws located on both
sides.
[NOTE]
Take care to install the correct orientation of the seesaw switch.
Claw
Special hexagon
3. Tighten the bottom cover using the five Bottom cover bolts (M3 x 6) (#239)
special hexagon bolts (M3 x 6).
Special
hexagon bolt
(M3 x 6) (#239)
111
5. Disassembling/Assembling/Adjusting R-4758.B
Sub Stage
Parallel mirror
X direction
Specifications:
X direction: within +/-1”
Y direction: within +/-1”
Center (O) of the collimator
If the specification is met, perform the stage centering adjustment. (Refer to Section
5-18.)
If not, perform the adjustment according to steps 6 and later.
Be sure to adjust Y direction first.
112
5. Disassembling/Assembling/Adjusting R-4758.B
Scale image is shifted to (a) direction Scale image is shifted to (b) direction
Paste the
brass foil.
9. Install the elevation block, stage, parallel mirror, and 1” reading collimator.
10. Check that the amount of deviation for Y direction meets the specification.
If the specification is not met, perform steps 6 to 10 again.
If the specification is met, check that the amount of deviation for X direction meets the
specification.
113
5. Disassembling/Assembling/Adjusting R-4758.B
114
5. Disassembling/Assembling/Adjusting R-4758.B
[NOTE]
Before performing this adjustment, the stage tilt adjustment should be completed. (Refer
to Section 5-17.)
115
5. Disassembling/Assembling/Adjusting R-4758.B
If the rod is shifted to (b) direction: Shift of (a) direction Shift of (b) direction
Adjust the centering according to steps 5
and later.
8. Install the elevation block, centering rod, and 80i sub stage tool A.
9. Check that the specification is met.
If the specification is not met, perform steps 4 to 8 again.
If the specification is met, perform the stage tilt adjustment again because the
elevation block is removed from the microscope base. (Refer to Section 5-17.)
10. After the stage tilt adjustment and stage centering adjustment are completed, be
sure to adjust the condenser holder tilt. (Refer to Section 5-19.)
116
5. Disassembling/Assembling/Adjusting R-4758.B
[NOTE]
Before performing this adjustment, the stage tilt adjustment (Section 5-17) and stage
centering adjustment (Section 5-18) should be completed.
Parallel mirror
Condenser receptacle
X direction
117
5. Disassembling/Assembling/Adjusting R-4758.B
If the specification is met, perform the condenser holder upper limit stopper position.
(Refer to Section 5-20)
If not, perform the adjustment according to steps 6 and later.
Be sure to adjust Y direction first.
118
5. Disassembling/Assembling/Adjusting R-4758.B
Scale image is shifted to (a) direction Scale image is shifted to (b) direction
Paste the brass foil.
10. Install the condenser holder block, parallel mirror, and 1” reading collimator.
11. Check that the amount of deviation for Y direction meets the specification.
If the specification is not met, perform steps 6 to 10 again.
If the specification is met, check that the amount of deviation for X direction meets the
specification.
12. Loosen the three hexagon socket head Hexagon socket head cap bolts (M3 x 8)(#442)
cap bolts fixing the condenser holder block
to the elevation block.
13. Tilt the condenser holder block in the
direction shown in the photograph to meet
the specification for the amount of
deviation for X direction.
14. Tighten the three hexagon socket head
cap bolts. Condenser
holder block
119
5. Disassembling/Assembling/Adjusting R-4758.B
120
5. Disassembling/Assembling/Adjusting R-4758.B
5-20. Condenser Holder Upper Limit Stopper Position Adjustment (For 50i/55i)
[NOTE]
Before performing this adjustment, the stage tilt adjustment (Section 5-17), stage
centering adjustment (Section 5-18), and condenser holder tilt adjustment (Section 5-19)
should be completed.
121
5. Disassembling/Assembling/Adjusting R-4758.B
(a)
(a) (b)
Special hexagon
9. Place the microscope upside down. Bottom cover bolts (M3 x 6) (#239)
10. Remove the five special hexagon bolts (M3
x 6) (#239), and remove the
Special
bottom cover.
hexagon bolts
(M3 x 6) (#239)
122
5. Disassembling/Assembling/Adjusting R-4758.B
11. Remove the two hexagon socket head cap Hexagon socket head cap
Diffuser unit bolts (M3 x 8) (#255)
bolts (M3 x 8) (#255), and remove the
diffuser unit.
123
5. Disassembling/Assembling/Adjusting R-4758.B
15. Install the standard straight tube (J25295) to the microscope, and secure it using the
clamp screw as shown in the photograph.
16. Set the Ph tool ring (J25260) to the centering telescope (J25172), and set them to the
standard straight tube.
[Note]
The binocular tube installation is acceptable instead of installing the standard straight
tube, Ph tool ring, and centering telescope to the microscope, as shown in the
photograph.
Centering telescope
Ph tool ring Binocular tube
Control board
Specifications:
Deviation between centers of the cross
Central LED
line and the central LED: Within 1.5 dev.
of scale
19. After the adjustment is completed, tighten the control board using the four special
hexagon bolts.
20. Re-check that the specification is met.
124
5. Disassembling/Assembling/Adjusting R-4758.B
After adjustment, assemble the parts and units in the following steps.
21. Turn OFF the power, and disconnect the AC adapter or the battery.
22. Remove the standard straight tube, centering telescope, and Ph tool ring.
23. Place the microscope upside down.
Hexagon socket head
24. Install the diffuser unit, and firmly move it in
Diffuser unit cap bolts (M3 x 8)
the direction of the arrows. Tighten it with
two hexagon socket head cap bolts (M3 x
8).
25. Tighten the bottom cover using the five Special hexagon
Bottom cover bolts (M3 x 6) (#239)
special hexagon bolts (M3 x 6).
Special
hexagon bolts
(M3 x 6) (#239)
Hexagon socket
head cap bolts
(M3 x 6) (#254)
125
5. Disassembling/Assembling/Adjusting R-4758.B
126