R-4758.

Microscope
ECLIPSE 50i
ECLIPSE 55i

Repair Manual
Volume-1
NIKON CORPORATION

C 2004 NIKON CORPORATION


All rights reserved
R-4758.B

REV. RECORD R-4758 REV. NUMBER:1

PAGE REV.MARK REV. DATE PAGE REV.MARK REV. DATE PAGE REV.MARK REV. DATE
1 B 2005.3.2 46 B 2005.3.2 91 B 2005.3.2
2 1 47 B 2005.3.2 92 B 2005.3.2
3 B 2005.3.2 48 B 2005.3.2 93 B 2005.3.2
4 B 2005.3.2 49 B 2005.3.2 94 B 2005.3.2
5 B 2005.3.2 50 B 2005.3.2 95 B 2005.3.2
6 B 2005.3.2 51 B 2005.3.2 96 B 2005.3.2
7 B 2005.3.2 52 B 2005.3.2 97 B 2005.3.2
8 B 2005.3.2 53 B 2005.3.2 98 B 2005.3.2
9 B 2005.3.2 54 B 2005.3.2 99 B 2005.3.2
10 B 2005.3.2 55 B 2005.3.2 100 B 2005.3.2
11 B 2005.3.2 56 B 2005.3.2 101 B 2005.3.2
12 B 2005.3.2 57 B 2005.3.2 102 B 2005.3.2
13 B 2005.3.2 58 B 2005.3.2 103 B 2005.3.2
14 B 2005.3.2 59 B 2005.3.2 104 B 2005.3.2
15 B 2005.3.2 60 B 2005.3.2 105 B 2005.3.2
16 B 2005.3.2 61 B 2005.3.2 106 B 2005.3.2
17 B 2005.3.2 62 B 2005.3.2 107 B 2005.3.2
18 B 2005.3.2 63 B 2005.3.2 108 B 2005.3.2
19 B 2005.3.2 64 B 2005.3.2 109 B 2005.3.2
20 B 2005.3.2 65 B 2005.3.2 110 B 2005.3.2
21 B 2005.3.2 66 B 2005.3.2 111 B 2005.3.2
22 B 2005.3.2 67 B 2005.3.2 112 B 2005.3.2
23 B 2005.3.2 68 B 2005.3.2 113 B 2005.3.2
24 B 2005.3.2 69 B 2005.3.2 114 B 2005.3.2
25 B 2005.3.2 70 B 2005.3.2 115 B 2005.3.2
26 B 2005.3.2 71 B 2005.3.2 116 B 2005.3.2
27 B 2005.3.2 72 B 2005.3.2 117 B 2005.3.2
28 B 2005.3.2 73 B 2005.3.2 118 B 2005.3.2
29 B 2005.3.2 74 B 2005.3.2 119 B 2005.3.2
30 B 2005.3.2 75 B 2005.3.2 120 B 2005.3.2
31 B 2005.3.2 76 B 2005.3.2 121 B 2005.3.2
32 B 2005.3.2 77 B 2005.3.2 122 B 2005.3.2
33 B 2005.3.2 78 B 2005.3.2 123 B 2005.3.2
34 B 2005.3.2 79 B 2005.3.2 124 B 2005.3.2
35 B 2005.3.2 80 B 2005.3.2 125 B 2005.3.2
36 B 2005.3.2 81 B 2005.3.2 126 B 2005.3.2
37 B 2005.3.2 82 B 2005.3.2 127 A
38 B 2005.3.2 83 B 2005.3.2 128 A
39 B 2005.3.2 84 B 2005.3.2 129 A
40 B 2005.3.2 85 B 2005.3.2 130 A
41 B 2005.3.2 86 B 2005.3.2 131 A
42 B 2005.3.2 87 B 2005.3.2 132 A
43 B 2005.3.2 88 B 2005.3.2 133 A
44 B 2005.3.2 89 B 2005.3.2 134 A
45 B 2005.3.2 90 B 2005.3.2 135 A

2
R-4758.B

Table of Contents

Foreword ..................................................................................................................... P.5


Notes for Repair .......................................................................................................... P.5
Safety Precautions ...................................................................................................... P.5

1. Tools ........................................................................................................................... P.6


1-1. Tools for General Use and Preparations ............................................................... P.6
1-2. Tools for Special Purposes ................................................................................... P.7
1-3. Preparations ......................................................................................................... P.8

2. Specifications................................................................................................................ P.9
2-1. 50i ......................................................................................................................... P.9
2-2. 55i ....................................................................................................................... P.13

3. Optical Systems ........................................................................................................ P.20


3-1 50i ....................................................................................................................... P.20
3-2. 55i ....................................................................................................................... P.21

4. Connection Diagrams ............................................................................................... P.22


4-1. 50i ....................................................................................................................... P.22
4-2. 55i ....................................................................................................................... P.23

5. Disassembling/Assembling/Adjusting ....................................................................... P.24


5-1. Disassembling/Assembling/Adjusting Flow ......................................................... P.24
5-2. Removing Main Unit ........................................................................................... P.26
5-3. Disassembling/Assembling Elevation Block (For 50i/55i) ................................... P.27
5-3-1 Disassembling ................................................................................................ P.27
5-3-2. Assembling .................................................................................................... P.33
5-4. Disassembling/Assembling Condenser Holder Block (For 50i/55i) ..................... P.42
5-4-1. Disassembling ................................................................................................ P.42
5-4-2. Assembling .................................................................................................... P.46
5-5. Disassembling/Assembling Fine/Coarse Focus Block (For 50i/55i) .................... P.52
5-5-1. Disassembling ................................................................................................ P.52
5-5-2. Assembling .................................................................................................... P.59
5-6. Disassembling/Assembling Field Stop Block (For 50i/55i) .................................. P.68
5-6-1. Disassembling ................................................................................................ P.68
5-6-2. Assembling .................................................................................................... P.71
5-7. Disassembling/Assembling ND Filter Block (For 50i) .......................................... P.75
5-7-1. Disassembling ................................................................................................ P.75
5-7-2. Assembling .................................................................................................... P.77

3
R-4758.B

Table of Contents

5-8. Disassembling/Assembling Lighting Block (For 50i) ........................................... P.79


5-8-1. Disassembling .............................................................................................. P.79
5-8-2. Assembling ................................................................................................... P.82
5-9. Disassembling/Assembling Power Supply Assembly (For 50i) ........................... P.85
5-9-1. Disassembling .............................................................................................. P.85
5-9-2. Assembling ................................................................................................... P.87
5-10. Disassembling/Assembling Brightness Control Board (For 50i) .......................... P.89
5-10-1. Disassembling .............................................................................................. P.89
5-10-2. Assembling ................................................................................................... P.91
5-11. Disassembling/Assembling Seesaw Switch (For 50i) ......................................... P.93
5-11-1. Disassembling .............................................................................................. P.93
5-11-2. Assembling ................................................................................................... P.94
5-12. Disassembling/Assembling Color Compensation Filter Unit (For 55i) ................. P.95
5-12-1. Disassembling .............................................................................................. P.95
5-12-2. Assembling ................................................................................................... P.97
5-13. Disassembling/Assembling Diffuser Unit (For 55i) .............................................. P.99
5-13-1. Disassembling .............................................................................................. P.99
5-13-2. Assembling ................................................................................................. P.100
5-14. Disassembling/Assembling Light Source Unit (For 55i) .................................... P.102
5-14-1. Disassembling ............................................................................................ P.102
5-14-2. Assembling ................................................................................................. P.104
5-15. Disassembling/Assembling Switch Board (For 55i) .......................................... P.106
5-15-1. Disassembling ............................................................................................ P.106
5-15-2. Assembling ................................................................................................. P.108
5-16. Disassembling/Assembling Seesaw Switch (For 55i) ........................................P.110
5-16-1. Disassembling .............................................................................................P.110
5-16-2. Assembling .................................................................................................. P.111
5-17. Stage Tilt Adjustment (For 50i/55i) .....................................................................P.112
5-18. Stage Centering Adjustment (For 50i/55i) ..........................................................P.115
5-19. Condenser Holder Tilt Adjustment (For 50i/55i) .................................................P.117
5-20. Condenser Holder Upper Limit Stopper Position Adjustment (For 50i/55i) ....... P.121
5-21. LED Position Adjustment (For 55i) .................................................................... P.122

4
Foreword R-4758.B

Foreword
1. This manual is for ECLIPSE 50i and ECLIPSE 55i.
Where an item is common for both ECLIPSE 50i and ECLIPSE 55i, “(For 50i/55i)” is included
in the title.
Where an item is exclusive to either ECLIPSE 50i or ECLIPSE 55i, “(For 50i)” or “(For 55i)” is
included in the title.
For the Stage and Ergonomic Binocular Tube, refer to the ECLIPSE 50i/ECLIPSE 55i
Repair Manual Volume-2.

2. This repair manual has been written to be used as the textbook for the repair training course.
This manual is organized so that trainees can understand the mechanical and optical
construction of the microscope by practicing repair procedures from the beginning to the end.

3. The table of contents of this manual includes every item that might require repair and can
efficiently be used as the guide for repair.

4. Unplug the power cable and wear work-gloves before starting work.
Take extra care to avoid injury from sharp edges or projections and electric shock by touching
live components.

Notes for Repair:


- Use an appropriate screwdriver for the screw head slot. (Especially take care of the (+) slot
head screw).
- The plastic parts must be cleaned with a neutral solvent.
If an organic solvent (ether, thinner, alcohol, etc.) is used, discoloring or deterioration of the
part might occur.
- When cleaning the glass parts, take care to remove dust using a blower, etc. and wipe the
parts with a lens tissue moistened with alcohol or alcohol-ether mixture.

Safety Precautions
WARNING (For ECLIPSE 50i)
The lamp and surrounding areas (including the lamp house) become very hot during and
immediately after a period of illumination.

- To avoid risk of burns, do not touch the lamp or surrounding areas during or
immediately after a period of illumination.
- Make sure the lamp and surrounding areas have cooled sufficiently before attempting
to the replace lamp.
- To avoid risk of fire, do not place fabric, paper or highly flammable volatile materials
such as gasoline, petroleum benzine, paint thinner or alcohol near the lamp house
while the lamp is lit or during a period of around thirty minutes after the lamp has been
turned off.
- To avoid risk of burns, wait at least thirty minutes after the lamp is turned off to give it
sufficient time to cool.
5
1. Tools R-4758.B

1-1. Tools for General Use and Preparations

Tool No Name Illustration Remarks

J21084 Hex-key set Set (x6)

T91400 (+) Screwdriver set Set (x4)

T91000 (-) Screwdriver set Set (x6)

T91204 (+) Wood handle driver Diameter 6 mm

T91102 (-) Wood handle screwdriver Width 5.5 mm


φ2.5mm
(+) Screwdriver bit
T91305 φ115mm
(For JCIS screw M2, M2.6)
90mm
φ8mm
T91320 (+) Screwdriver bit holder

T92041 Tweezers AA

T92162 Compass spanner A

T92163 Compass spanner B

Handlap
E719 (Alcohol container)

6
1. Tools R-4758.B

1-2. Tools for Special Purposes

Tool No Name Illustration Remarks

Stage tilt adjustment


J28005 1” reading collimator Condenser holder tilt
adjustment

Stage tilt adjustment


J28003 Parallel mirror Condenser holder tilt
adjustment

Stage centering
J25298 Centering tool
adjustment

Stage centering
J25481 80i sub stage tool A
adjustment

Stage centering
J25299 Centering rod
adjustment

7
1. Tools R-4758.B

1-2. Tools for Special Purposes/Preparations

Tool No Name Illustration Remarks

LED position
J25295 Standard straight tube
adjustment

LED position
J25172 Centering telescope
adjustment

LED position
J25260 Ph tool ring
adjustment

Spring scale Torque adjustment

1-3. Preparations
Alcohol (or alcohol-ether mixture), Lens tissue, Blower, Gasoline (for cleaning),
Bamboo stick (for applying lubricant, adhesive), Brass (or tin) foil, Dial gauge
For Lubricant and Adhesive: Refer to Repair Parts Lists (50i: KR-7483.A, 55i: KR-7485.A,)

8
2. Specifications R-4758.B

2-1. 50i

I. Overviews
1. Model Number
MBA85010: ECLIPSE 50i, Using halogen lamp as light source L
MBA85020: ECLIPSE 50i, Using halogen lamp as light source R

II. Main Items


1. Mounting System
- For nosepiece: Square dovetail (compatible with E series)
Can be attached and removed using tool.
- For eyepiece tube: Round dovetail (compatible with E series)
Can be attached and removed using clamp screw.
- For intermediate attachments: Eyepiece tube mounting dovetail and main
body top surface (Using M4 screw (4 pcs.))

2. Main Body Top Arm and Stand Blocks


- Height from table top to eyepiece mounting surface: 330 mm
- Height from table top to stage top: 199 mm
- Max. width: 208 mm
- Foot: With vibration-proof damper (4 pcs.)

3. Elevation Block
- Focusing system/Stage up-down system: Using roller races
- Stage up-down stroke: Upward 2 mm/Downward 28 mm
- Sub stage and up-down movement guide: DC integrated type
- Stage mounting system: Round dovetail (not compatible with E series)
- Condenser lens mounting system: Round dovetail (compatible with E series)
- Condenser lens up-down stroke: 27 mm
- Condenser lens stroke upper limit position: Adjustable (using tool)
- Condenser lens centering: 3 mm diameter

9
2. Specifications R-4758.B

2-1. 50i
4. Coarse/Fine Focus Block
- Coarse focus handle: 13.8 mm per rotation (reduction gear ratio: 3)
- Fine focus handle: 0.1 mm per rotation (reduction gear ratio: approx. 413)
- Minimum reading: 1 micron
- Coarse motion handle torque: Adjustable
- Refocusing mechanism: With one-touch refocusing mechanism (with
operation lever)
When the sub stage is lowered to the lower limit position by pushing the lever
down, the sub stage is locked to this position.
When the sub stage is lowered 2 mm further from this position by pushing
the lever down, the lock mechanism is canceled. The sub stage is returned
to the former position.
The rotary damper controls the return of the sub stage at a constant speed.

5. Lighting Block
- Composed of collector lens, field lens (f = 89.7), field stop, and filter
- Filter: ND8 (Detachable)
- Lemon-skin filter (2 pcs.): Built-in (Not detachable)
- Diameter of field stop: 1 to 30 mm diameter (design value: 0.8 to 30.5 mm
diameter)
- Field stop control: Right-hand side of the microscope base
- Filter control: Right-hand side of the microscope base
- Additional filter (diameter: 45 mm, thickness: 3 mm): Up to two filters can be
attached tilting 5 degrees above the field lens.

6. Light Source Block


- Lamp ratings and type: 6V/30W halogen lamp, PHILIPS 5761
- Average lamp life: 100 hours (listed on product brochure)
- Lamp house detachable

10
2. Specifications R-4758.B

2-1. 50i

III. Configurations and Functions


1. System Diagram
(Described for eyepiece tube, intermediate attachments, nosepiece, stage,
Condenser lens, and polarizer)

C-TEP DSC Port


:Option
(MBB76600)

Binocular Tube Trinocular Tube F Trinocular Tube T C-TE Ergonomic Tube


(MBB72100/72105) (MBB73100/73105) (MBB73110) (MBB76500)

J-FL 50i55i Epi-Fluorescence


Teaching Unit F to F
Attachment (MBE84000)
(MBB75100) Double Port 0/100
Teaching Unit S by S type B
(MBB74105)
(MBB75110) Double Port
Main Teaching Unit (MBB74100)
(MBB75120) Double Port 0/100
Teaching Unit S by S type A (MBB74105)
(MBB75130) Magnification Module
Intermediate Tube w/Simple (MBB75500)
Analyzer (MBB75370) Drawing Tube
(MBB75400)

Intermediate Tube Eyelevel Riser


Objective Lens
w/Analyzer (MBB75360) (MBB75710)
(CFI60)
Condenser (dry)
(MBL12000)
C-N Sextuple Nosepiece
(MBP71300) Condenser (oil)
(MBL12001)
J-CY Cytodiagnostic Unit
(MBP89000) A/A Condenser
(MBL71400)

Swing-out Condenser
(MBL11300)

50i Halogen Lamp L


C-SR Left Handle Abbe Condenser
C-SR Right Handle
(MBA85010)
Stage (MBC71610) (MBL71100)
Stage (MBC71600)
50i Halogen Lamp R
(MBA85020) LWD Achromat
Condenser (MBL75000)
C-HC1 Specimen
Holder 1C (MBC75040) Achromat Swing-out
Simple Polarizer Condenser 1-100X
(MBB75370) (MBL71300)
C-HR2 Specimen C-HL2 Specimen
Holder 2R (MBC75010) Holder 2L (MBC75020) Polarizer with First-order PH Turret Condenser
Red Tint Plate (MBL73100)
(MBB75380)

11
2. Specifications R-4758.B

2-1. 50i
2. Restrictions (Following products are not installable.)
- Condenser lens
D-CUD-E Automatic U condensor (dry) (MBL78500)
D-CUD U condensor (dry) (MBL78600)
- Intermediate attachments
D-FL U-EPI Fluorescence (MBE74100)
D-DH-E DIH E (MBD76100)
D-DH DIH M (MBD76200)
- Stage
C-SRR Right handle R stage (MBC71500)
C-SLR Left handle R stage (MBC71600)
D-S-E Automatic XY stage (MBC76100)

12
2. Specifications R-4758.B

2-2. 55i
I. Overviews
1. Model Number
MBA85040: ECLIPSE 55i, Using LED as light source L
MBA85050: ECLIPSE 55i, Using LED as light source R

II. Main Items


1. Mounting system
- For nosepiece: Square dovetail (compatible with E series)
Can be attached and removed using tool.
- For eyepiece tube: Round dovetail (compatible with E series)
Can be attached and removed using clamp screw.
- For intermediate attachments: Eyepiece tube mounting dovetail and main
body top surface (Using M4 screw (4 pcs.))

2. Main Body Top Arm and Stand Blocks


- Height from table top to eyepiece mounting surface: 330 mm
- Height from table top to stage top: 199 mm
- Max. width: 208 mm
- Foot: With vibration-proof damper (4 pcs.)

3. Elevation Block
- Focusing system/Stage up-down system: Using roller races
- Stage up-down stroke: Upward 2 mm/Downward 28 mm
- Sub stage and up-down movement guide: DC integrated type
- Stage mounting system: Round dovetail (not compatible with E series)
- Condenser lens mounting system: Round dovetail (compatible with E series)
- Condenser lens up-down stroke: 27 mm
- Condenser lens stroke upper limit position: Adjustable (using tool)
- Condenser lens centering: 3 mm diameter

13
2. Specifications R-4758.B

2-2. 55i
4. Coarse/Fine Focusing Block
- Coarse focus handle: 13.8 mm per rotation (reduction gear ratio: 3)
- Fine focus handle: 0.1 mm per rotation (reduction gear ratio: approx. 413)
- Minimum reading: 1 micron
- Coarse motion handle torque: Adjustable
- Refocusing function: Refocusing with a single touch of the lever (with
operation lever)
When the sub stage is lowered to the lower limit position by pushing the lever
down, the sub stage is locked to this position.
When the sub stage is lowered 2 mm further from this position by pushing
the lever down, the lock mechanism is canceled. The sub stage is returned
to the former position.
The rotary damper controls the return of the sub stage at a constant speed

5. Lighting Block
- Composed of collector lens, field lens (f = 89.7), field stop, and filter
- Filter: LA60 (Detachable)
- Lemon-skin filter (2 pcs.): Built-in (Not detachable)
- Diameter of field stop: 1 to 30 mm diameter (design value: 0.8 to 30.5 mm
diameter)
- Field stop control: Right-hand side of the microscope base
- Filter control: Right-hand side of microscope base
- Additional filter (diameter: 45 mm, thickness: 3 mm): Up to two filters can be
attached tilting 5 degrees above the field lens.

6. Light Source Block


- White LED array (7 LEDs are used.)
- Lamp house detachable

7. Electric System
- Brightness control dial: Rotary system, located on the right-hand side of the
microscope base.
- Preset switch: Push-button, located on the right-hand side of the microscope
base.
- Lithium-ion battery mounting port: Located on the back of the LED array board.
- Input connector: Located on the back.
- Power connector for Ergo-View cytodiagnostic unit: Located on the back.
- Signal connector for Ergo-View cytodiagnostic unit: Located on the back.
- Hand-switch connector for Ergo-View cytodiagnostic unit: Located on the back.

14
2. Specifications R-4758.B

2-2. 55i

7-1 Power Input


The microscope uses power from an AC adapter or from a battery, as follows.
When the power is supplied from an AC adapter, this has priority over any battery.
1. AC adapter input: Option
Input voltage: 12 V/1.5 ADC
Specified AC adapter: Manufacturer: ILAN ELECTRONICS LTD
Model: F1650K
- Rated input voltage: 100 to 240 VAC/1.2 A max. (when inputting 115
VAC), 47 to 63 Hz
2. Specified Lithium-ion rechargeable battery: Option
Manufacturer: Nikon: Code: EN-EL1
Rating: 7.4 V/720 mAh
Specified battery charger: MH-53 or similar
Note: UR-421 (SANYO Electric Co., Ltd) can be used as the equivalent
battery charger.

7-2 Power Output (For Ergo-View cytodiagnostic unit)


Supply power to Ergo-View cytodiagnostic unit.
- DC 12 V/1 A max.: Current limitation (approx. 1A within 1 sec)
- DC 12V/0.5 A max.: Continuous supply

7-3 Battery Run Time


1. When Ergo-View cytodiagnostic unit is not used
4.5 hours minimum: When observed continuously at an LED light source at
maximum light intensity using a new rated-capacity battery
10 hours minimum: When observed continuously at 40% of the maximum current
The following are assumed for this case:
- Maximum illuminance using the Ergo-View cytodiagnostic unit with
PlanApo40x and a CB filter attached
- 4 lx (eyepiece illuminator reading)
2. When Ergo-View cytodiagnostic unit is used (Reference value)
Approx. 3.5 hours: Run time at maximum light intensity assuming that the
magnification of the Ergo-View cytodiagnostic unit is changed 5 times a
minute

7-4. Battery Charge Time (Reference value)


Approx. 100 minutes: When a specified battery charger is used

7-5. LOW-Battery Detection


“LOW-Battery” is displayed when the battery voltage lowers to 7.13 +/-0.2 V
during operation. The Power LED blinks at a cycle of approx. 5 Hz.

15
2. Specifications R-4758.B

2-2. 55i

7-6. Battery Shutdown Detection


When the battery voltage reaches 5.75 +/-0.16 V during operation, the
microscope shuts down to prevent the lithium battery from over discharging.
It takes 3 to 30 minutes (reference value) from “LOW-Battery” detection for
the microscope to shutdown.
The time depends mainly on the number of charge/discharge repetition times,
ambient temperature, and LED luminance setting.
To reboot the microscope, replace the battery with a charged new battery or
switch the power source to the AC adapter.

7-6-1.Replacing Battery
Remove the lamp house cover and replace the battery in the battery
pocket with a new one.

7-7. White LEDs


Seven white LEDs NSPW510BS (Nichia Corporation) are used
(luminous intensity rank = S or R, color tone rank = b1 or b2, maximum
drive current = 22 mA/LED).

7-8. LED Current Variation Range


Approx. 0.05 to 22 mA per LED
Current variation dynamic range: Approx. 440 times
Total current of seven LEDs is controlled at a constant value.
Current variation at maximum light intensity: 5% (reference value)

7-9. Brightness Control


The LED’s light intensity is adjusted with the master brightness control
knob on the operation panel.
The current variation characteristics (B characteristics) vary with the
master brightness control knob characteristics (A characteristics).
When the alternate preset switch is pressed, the light intensity adjustment
is switched to the preset brightness control which can be adjusted using a
(-) screwdriver through the adjustment hole of the operation panel.
The preset brightness control knob is visually set to the neutral position
(half of maximum light intensity) at the factory prior to shipping.

16
2. Specifications R-4758.B

2-2. 55i
7-10. LED Current Control by Varying Magnification of Ergo-View Cytodiagnostic Unit
When the magnification of the Ergo-View cytodiagnostic unit is set to a
high value, the LED current increases to approx. 3.2 times, which makes
the LED brighter. However, the magnification is limited to the maximum
LED current 22 mA.
The LED current does not depend on the magnification of the Ergo-View
cytodiagnostic unit while the preset switch is ON.

7-11. Life of White LED (Reference data)


According to the reference data provided by the manufacturer, the relation
between LED operating time and light intensity is as shown below. The life
of white LEDs is calculated assuming that the light intensity at the
beginning of test = 100% and operating time = 8 hours/day, 250 days/year

Ambient Operating Operating Light


Life of LED
temperature current time intensity
Ta = 25 °C IF = 20 mA 3500 hours 70% 1.75 years
Ta = 25 °C IF = 20 mA 8000 hours 55% 4 years
Ta = 25 °C IF = 10 mA 8000 hours 90% 4 years

7-12. Input/Output Connectors


- Input connector for cytodiagnostic unit switch: 3.5 diameter mini-jack
Connect the cable of the cytodiagnostic unit hand switch.
- Input/output connector for cytodiagnostic unit signal: 3.5 diameter mini-jack
Connect the signal cable of the cytodiagnostic unit.
Input/output the hand switch signal and magnification signal.
- Power output connector for cytodiagnostic unit: DC jack
Connect the cytodiagnostic unit power supply cable.
Note: The shape of this connector differs from the power input
connector.
- Power input connector: DC jack
Connect the power output plug of AC adapter.

7-13. Indicators
- Power ON/OFF indicator LED (Green): Located on the right side
operation panel of the microscope base
- Lamp house high temperature caution label
- Rated current indication seal

17
2. Specifications R-4758.B

2-2. 55i

III. Configurations and Functions


1. System Diagram
(Described for eyepiece tube, intermediate attachments, nosepiece, stage,
Condenser lens, and polarizer)

C-TEP DSC Port


:Option
(MBB76600)

Binocular Tube Trinocular Tube F Trinocular Tube T C-TE Ergonomic Tube


(MBB72100/72105) (MBB73100/73105) (MBB73110) (MBB76500)

J-FL 50i55i
Epi-Fluorescence
Double Port 0/100
Teaching Unit F to F
(MBB74105)
(MBB75100)
Double Port
Teaching Unit S by S type B
(MBB74100)
(MBB75110)
Double Port 0/100
Main Teaching Unit
(MBB74105)
(MBB75120) Magnification
Teaching Unit S by S type A
Module
(MBB75130) Drawing Tube
(MBB75400)

C-ER Eyelevel Riser


(MBB75710)
Objective Lens
(CFI60)

A/A Condenser
C-N Sextuple (MBL71400)
Nosepiece
Swing-out Condenser
J-CY Cytodiagnostic (MBL11300)
Unit (MBP89000)
A/A Condenser
(MBL71400)

LWD Achromat
Condenser
C-SR Left Handle C-SR Right Handle
Stage (MBC71610) Stage (MBC71600) 55i LED L Achromat Swing-out
(MBA85040) Condenser 1-100X
55iLED R (MBL71300)
C-HC1 Specimen (MBA85050)
Holder 1C (MBC75040)

C-HR2 Specimen C-HL2 Specimen


Holder 2R (MBC75010) Holder 2L (MBC75020)

18
2. Specifications R-4758.B

2-2. 55i
2. Restrictions (Following products are not installable.)
- Condenser lens
D-CUD-E Automatic U condenser (dry) (MBL78500)
D-CUD U condensor (dry) (MBL78600)
DF condenser (dry) (MBL12000)
DF condenser (oil) (MBL12001)
PH condenser (MBL73100)
- Intermediate attachments
D-FL U-EPI Fluorescence (MBE74100)
D-DH-E DIH E (MBD76100)
D-DH DIH M (MBD76200)
C-ISA simple analyzer (MBB75350)
C-IA Analyzer (MBB75360)
- Stage
C-SRR Right handle R stage (MBC71500)
C-SLR Left handle R stage (MBC71600)
D-S-E Automatic XY stage (MBC76100)
- Others
Simple polarizer (MBB75310)
Polarizer with first-order red tint plate (MDW10200)

19
3. Optical Systems R-4758.B

3-1. 50i
The following optical system is based on the ECLIPSE 50i.
The cemented lens surfaces are omitted from the drawing because the lens direction only
is of concern here.

Field stop

6V 30W

Diffuser ND8

20
3. Optical Systems R-4758.B

3-2. 55i

The following optical system is based on the ECLIPSE 55i.


The cemented lens surfaces are omitted from the drawing because the lens direction only
is of concern here.

Mat surface Evaporated surface Field stop

Evaporated
surface
LED arrays Mirror

Lemon skin Color compensation filter

21
4. Connection Diagrams R-4758.B

4-1. 50i

22
4. Connection Diagrams R-4758.B

4-2. 55i

23
5. Disassembling/Assembling/Adjusting R-4758.B

5-1. Disassembling/Assembling/Adjusting Flow

< ECLIPSE 50i >


26
Removing Stage/Eyepiece tube/Nosepiece/Condenser lens

Elevation Block Fine/coarse focus block Brightness control board


29 52 89
Removing fine/coarse focus block Disassembling/Assembling
fine/coarse focus block Removing brightness control dial

85 67 68
Removing power supply assembly Fine torque adjustment Removing field stop block

29 89
Removing elevation block Disassembling/Assembling
brightness control board

Condenser holder block


29 42 ND filter block
Disassembling/Assembling 75
Removing condenser holder condenser holder Disassembling/Assembling
ND filter block
30 47
Disassembling/Assembling Knob/rack engaging adjustment
elevation block Power supply assembly
85
Removing lamp
34
Roller race movement adjustment 86
Disassembling/Assembling
power supply assembly
38
Installing condenser holder
Seesaw switch
93
Disassembling/Assembling
38 seesaw switch
Installing elevation block
Lighting block
79 80
113 Removing lamp Disassembling/Assembling
diffuser
Stage tilt adjustment
80 81
Disassembling/Assembling
116 Removing lighting block collector lens
Stage centering adjustment
81
Disassembling/Assembling
118 cable with lamp socket
Condenser holder tilt adjustment
Field stop block
68 70
122 Removing field stop block Disassembling mirror/lens
Condenser holder upper
limit stopper position adjustment
69 74
Disassembling/Assembling blade Cleaning adhesive coating surface
39
of mirror/lens
Installing fine/coarse focus block
72
Blade mount position adjustment 74
40 Cementing mirror/lens
Elevation block movement check
using fine/coarse focus handle 72
Installing field stop block
40
Re-focusing mechanism 73
operation check Diaphragm open-shut dial
operation check

NOTE: The number at upper right of the frame shows the page of this item.

24
5. Disassembling/Assembling/Adjusting R-4758.B

5-1. Disassembling/Assembling/Adjusting Flow

< ECLIPSE 55i >


26
Removing Stage/Eyepiece tube/Nosepiece/Condenser lens

Elevation Block Fine/coarse focus block Switch board


29 52 107
Removing fine/coarse focus block Disassembling/Assembling
Removing brightness control dial
fine/coarse focus block

103 67 68
Removing control board Fine torque adjustment Removing field stop block

29 107
Disassembling/Assembling
Removing elevation block
switch board
Condenser holder block
29 42
Disassembling/Assembling Color compensation filter unit
Removing condenser holder condenser holder 95
Removing color compensation
filter unit
30 47
Disassembling/Assembling Knob/rack engaging adjustment
elevation block 96
Disassembling/Assembling
34 color filter unit

Roller race movement adjustment


Seesaw switch
111
38 Disassembling/Assembling
seesaw switch
Installing condenser holder

38
Light source Unit Diffuser unit
Installing elevation block 103 99
Removing control board Removing diffuser unit
113
104 100
Stage tilt adjustment
Disassembling/Assembling Disassembling lens/diffuser
LED board
116
105 101
Stage centering adjustment Cleaning adhesive coating surface
Installing control board
of mirror/diffuser
118 101
123
Condenser holder tilt adjustment LED position adjustment Installing diffuser unit

122 Field stop block


Condenser holder upper 68 70
limit stopper position adjustment Removing field stop block Disassembling mirror/lens

39 69 74
Installing fine/coarse focus block Disassembling/Assembling blade Cleaning adhesive coating surface
of mirror/lens
40 72
Elevation block movement check 74
Blade mount position adjustment
using fine/coarse focus handle Cementing mirror/lens
72
40
Installing field stop block
Re-focusing mechanism
operation check
73
Diaphragm open-shut dial
operation check

NOTE: The number at upper right of the frame shows the page of this item.

25
5. Disassembling/Assembling/Adjusting R-4758.B

5-2. Removing Main Unit

Before disassembling and assembling the inside mechanisms of the microscope,


remove the main unit. This will make it easier to disassemble and assemble the internal
parts, and re-adjust the optical system.

1. Eyepiece tube 2. Nosepiece

Loosen the eyepiece clamp screw beneath Loosen the clamp screw on the side of the
the eyepiece tube, and remove the nosepiece, and remove the nosepiece.
eyepiece tube.

Nosepiece

Eyepiece tube Clamp screw Clamp screw

3. Stage 4. Condenser lens

Loosen the set screw on the front side of Loosen the set screw on the condenser
the stage using a hex-key, and remove the receptacle using a hex-key, and remove the
stage. condenser lens.

Set screw Stage


Set screw
Condenser lens

26
5. Disassembling/Assembling/Adjusting R-4758.B

5-3-1.Disassembling Elevation Block (For 50i/55i)

- Disassembling the elevation block is not usually necessary, but if liquid (water etc.)
has dropped into the microscope, it should be disassembled.
- The two roller races (#A3) fixed to the attaching plate must not to be disassembled at
the same time. When reassembling one roller race, the other (assembled) roller race
is used as a reference for adjustment.
- Some screws are fixed with adhesive. Apply alcohol to melt the adhesive on these
screws.

WARNING (For ECLIPSE 50i)


- When disassembling the elevation block, remove the lamp cover from the lamp
house.
The lamp and surrounding areas (including the lamp house) become very hot
during and immediately after a period of illumination.
To avoid risk of burns, do not touch the lamp or surrounding areas during or
immediately after a period of illumination.
- To avoid risk of fire, do not place fabric, paper or highly flammable volatile
materials such as gasoline, petroleum benzine, paint thinner or alcohol near the
lamp house while the lamp is lit or during a period of around thirty minutes after
the lamp has been turned off.
- To avoid risk of burns, wait at least thirty minutes after the lamp is turned off to
give it sufficient time to cool.

5-3-1. Disassembling

1. Remove the eyepiece tube. (Refer to Section 5-2.)


2. Remove the objective lens.
3. Remove the condenser lens. (Refer to Section 5-2.)
4. Remove the nosepiece. (Refer to Section 5-2.)
5. Remove the stage. (Refer to Section 5-2.)
6. While pressing the two portions of the lamp
cover (#223) in the arrow directions (a)
shown in the photograph, move it in the
Lamp cover
arrow direction (b) to remove.
(#223)

(a)

(a) (b)

27
5. Disassembling/Assembling/Adjusting R-4758.B

5-3-1.Disassembling Elevation Block (For 50i/55i)

7. In the case of 50i, disconnect the lamp Plastic cover (#202) Cap (#233)
(#227) to prevent damage.
8. Pull out the hexagonal driver (#248).
Hexagon
9. Remove the four hexagon socket head cap socket head
bolts (M3 x 6) (#254), and remove the cap bolts
plastic cover (#202). (M3 x 6)(#254)

10. Remove the cap (#233).


11. In the case of 50i, remove the power Lamp (#227)

supply assembly. (Refer to Section 5-9.)


In the case of 55i, remove the control board
of the light source unit. (Refer to Section Hexagonal driver (#248)
5-14.) Special hexagon bolts
12. Place the microscope upside down. Bottom cover (M3 x 6) (#239)
13. Remove the five special hexagon bolts
(M3 x 6) (#239), and remove the Special
bottom cover. hexagon bolt
(M3 x 6) (#239)

Fine/coarse
focus block

Plastic cover (#411)

14. Remove the two hexagon socket head cap bolts (M3 x 10) (#51), and remove the
plastic cover (#411).
15. Insert a (-) 2 mm screwdriver into the spaces on both sides of the plastic cover
shown in the photograph to release the lock of the claws.
[NOTE]
When releasing the lock, take care not to break the claws.

Hexagon socket head cap bolts (M3 x 10) (#51) (-) screwdriver Claws of plastic cover

28
5. Disassembling/Assembling/Adjusting R-4758.B

5-3-1.Disassembling Elevation Block (For 50i/55i)

16. Hold the fine/coarse focus block firmly by


hand, and remove the two hexagon socket Hexagon socket
head cap bolts (M3 x 12) (#52). head cap bolts
Remove the fine/coarse focus block and (M3 x 12) (#52)
the plastic cover from the microscope
base.

Fine/coarse focus block

17. While holding the elevation block by hand,


Hexagon socket head cap bolts (M4 x 16) (#53)
place the microscope to its original position.
[NOTE]
Since the fine/coarse focus block is
removed, the movement of the elevation
block is not fixed.
Therefore, when the microscope is placed
to its original position, be sure to keep hold
of the elevation block. Hexagon socket head
Elevation block cap bolts (M4 x 16) (#53)

18. While holding the elevation block firmly by


hand so that it does not drop, remove the six
hexagon socket head cap bolts (M4 x 16)
(#53) to remove the elevation block.

19. Remove the three hexagon socket head Hexagon socket head
Condenser holder cap bolts (M3 x 8) (#442)
cap bolts (M3 x 8) (#442), and remove the
condenser holder block.

29
5. Disassembling/Assembling/Adjusting R-4758.B

5-3-1.Disassembling Elevation Block (For 50i/55i)

20. Remove the two hexagon socket head cap bolts (M3 x 6) (#441) on the front and
two hexagon socket head cap bolts (M3 x 8) (#442) on the rear. Remove the plastic
cover (#410).
Hexagon socket head cap bolts (M3 x 8) (#442)

Hexagon socket
head cap bolts
(M3 x 6) (#441)

Plastic cover (#410)


Plastic cover (#410)

Hexagon socket head Hexagon socket head


21. Remove a hexagon socket head cap bolt cap bolt (M3 x 12) (#443) cap bolts (M3 x 8)(#442)
(M3 x 12) (#443) of the stroke limitation (Stroke limitation screw)
screw.
22. Remove the two hexagon socket head cap
bolts (M3 x 6) (#441), and remove the
plate spring (#412) (with the rubber plate).
23. Remove the two hexagon socket head cap
bolts (M3 x 8) (#442), and remove the cap Cap (#415), Rotary
damper (#414)
(#415) and rotary damper (#414).
Plate spring (#412)
Hexagon socket head
cap bolts (M3 x 6) (#441)

24. Tighten a set screw (#452) on the side of


the sub body. Hexagon socket head cap bolts (M3 x 6) (#441)
25. Remove the two hexagon socket head cap
bolts (M3 x 6) (#441), and remove the
shaft supporter (#407).
Shaft supporter
(#407)

Set screw (#452)

30
5. Disassembling/Assembling/Adjusting R-4758.B

5-3-1. Disassembling Elevation Block (For 50i/55i)

26. While holding the top surface of the rack


(#406) by finger, loosen the set screw
tightened in step 24.
[NOTE]
When the set screw is loosened, the rack
will pop out by the power of the two coil
springs.
Be sure to hold the top surface of the rack
to prevent injury or damage to the
surroundings when the rack pops out.
Set screw (#452) Rack (#406)

Rack (#406)
27. Remove the rack, coil spring (#405), coil
spring (#404) and spring guide (#403). Coil spring
(#404)
Coil spring
(#405)

Spring guide
(#403)

Sub body

28. Remove the two plate washers from the


hole of the sub body.
[NOTE]
The two plate washers are only installed
on some microscope. If they are not
Plate washers (2 pcs)
installed, go to step 29.

Sub body

31
5. Disassembling/Assembling/Adjusting R-4758.B

5-3-1. Disassembling Elevation Block (For 50i/55i)

29. Place the elevation block right-hand side Set screws (#447) Elevation block
down, and fully loosen the four set screws
(#447). Sub body
30. Remove the four hexagon socket head
cap bolts (M3 x 16) (#418), and remove
the roller race (#A3) on the left side of the
elevation block. Remove the sub body
from the elevation Block.
[NOTE]
When removing the roller race, the
Roller race (#A3) Hexagon socket head cap
cylindrical rollers (12 pcs each on up and bolts (M3 x 16) (#418)
lower sides) might fall. Be careful not to
lose them.
The guide rails (#417) and separators
(#408) are also removed at the same time.

31. If there is a rusty or worn section on the


Roller race (#A3) Mounting surface
cylindrical rollers, separator and/or guide
rails (#417), replace it with a new one. Attaching plate
32. If there is a rusty section on the mounting (#402)
surface of the roller races and attaching
plate (#402), scrape it using a scraper.

Roller race (#A3)

The disassembly of the elevation block is completed.


The disassembled parts should be washed and cleaned.

32
5. Disassembling/Assembling/Adjusting R-4758.B

5-3-2. Assembling Elevation Block (For 50i/55i)

5-3-2. Assembling
Attaching plate
1. Paste the two guide rails to the two roller Guide rails (#417)
(#402)
races using lubricant L4010 as a substitute Sub body
for adhesive.
2. Paste the guide rails to the races on both
sides of the sub body using lubricant
L4010 in the same way as step 1.

Roller race (#A3) Roller race (#A3)

3. Apply lubricant L4010 to the roller race on Attaching plate (#402)


the attaching plate assembled in step 1, Cylindrical rollers
and place the separator and twelve (12 pcs) (#409)
cylindrical rollers on the roller race.
Separator (#408)

Roller race (#A3)

4. Apply lubricant L4010 to the roller race on the sub body, and place the separator and
twelve cylindrical rollers on the race.
5. Assemble the sub body to the roller race on the attaching plate, and lightly tighten the
roller race with the four hexagon socket head cap bolts (M3 x 16) while pressing it to
the sub body.

Cylindrical rollers
Sub body
(12 pcs) (#409) Roller race (#A3)
Sub body
Separator
(#408)

Attaching Hexagon socket head


plate (#402) cap bolts (M3 x 16) (#418)

33
5. Disassembling/Assembling/Adjusting R-4758.B

5-3-2. Assembling Elevation Block (For 50i/55i)

6. Move the sub body fully right and left, and adjust the position of the separator so that
the edges of the separator and the roller race coincide.
Also adjust the position of the separator on the opposite side in the same way.

Sub body Sub body

Roller race (#A3)

Roller race (#A3)


Edges of the separator and Edges of the separator and
the roller race are coincided. the roller race are coincided.

7. Move the sub body so that the set screw Set screws (#447)
on the attaching plate is adjacent to the
cylindrical roller.
8. Tighten the set screw close to the
cylindrical roller with the following
tightening torque using torque driver.
Tightening torque: 1 kg x cm
9. Repeat steps 7 and 8, and tighten the
remaining three set screws in the same
way. Hexagon socket head
Roller race (#A3)
10. Tighten the four hexagon socket head cap cap bolts (M3 x 16)(#418)
bolts to fix the roller race.
11. Move the sub body, and check the
following: Sub body
- The cylindrical roller and separator are
still in position.
- The sub body moves smoothly and Spring scale
lightly without slack.
Push the sub body and check, using a
spring scale, that 50 to 100g pressure is
required to effect movement. If not to meet
the specification, repeat steps 7 to 10.
[NOTE]
When the four hexagon socket head cap bolts securing the roller race are firmly
tightened, the sub body sometimes loses its movement. In this case, perform steps
7 to 10 again.

34
5. Disassembling/Assembling/Adjusting R-4758.B

5-3-2. Assembling Elevation Block (For 50i/55i)

12. Apply adhesive #757 to the threads of the Apply lubricant


four set screws tightened in steps 8 and 9.
13. After inserting the two plate washers into
the 13 mm diameter hole of the sub body, Plate washers
apply lubricant L4010 to the inner surface (2 pcs)
of the hole.
[NOTE]
The two plate washers are only installed
on some microscope. If the plate washers
were present during disassembling, be Sub body
sure to insert them in this step.

14. Put the spring guide into the hole of the sub body, and insert its tip into the center
hole on the spring retainer (#432).
[NOTE]
Be sure to insert the spring guide tip into the spring retainer hole.

15. Apply a light coating of lubricant L3047 to the whole surfaces of the two kinds of coil
springs.
16. Fit the lubricated coil springs into the hole of the sub body.
Tip of the Spring retainer (#432)
Coil spring (#404)
spring guide

Coil spring (#405)

Spring guide (#403)

Sub body Sub body

17. Install the rack to the sub body while


fitting the two coil springs into the hole of Rack (#406)
the rack.
[NOTE]
Be sure to install the rack in the direction
Two coil springs
as shown in the photograph.
and spring guide

Sub body
35
5. Disassembling/Assembling/Adjusting R-4758.B

5-3-2. Assembling Elevation Block (For 50i/55i)

18. After pushing the rack into the hole so that


the edges of the rack and the sub body
coincide, lightly tighten a set screw.

Rack (#406)
Set screw (#452)

19. Install the shaft supporter to the sub body while inserting the spring guide tip into the
center hole of the shaft supporter, and tighten it using the two hexagon socket head
cap bolts (M3 x 6).
[NOTE]
After tightening the bolts, check again that the spring guide tip is inserted into the
spring retainer hole on the opposite side.

Shaft supporter Hexagon socket head Spring retainer Tip of the


(#407) cap bolts (M3 x 6) (#441) (#432) spring guide

Tip of the
spring guide

(5013_1)
Sub body

20. Loosen the set screw lightly tightened in


step 18.
[NOTE]
When the set screw is firmly tightened, the
refocusing function cannot operate.

Loosen this set screw.

36
5. Disassembling/Assembling/Adjusting R-4758.B

5-3-2. Assembling Elevation Block (For 50i/55i)

21. Tighten the hexagon socket head cap bolt (M3 x 12) of the stroke limitation screw.
22. Tighten the plate spring (with rubber plate) using the two hexagon socket head cap
bolts (M3 x 6).
When tightening, adjust the position of the plate spring so that the edge of the plate
spring is parallel to the edge of the boss on the attaching plate.
23. Tighten the cap and rotary damper using the two hexagon socket head cap bolts
(M3 x 8).
Hexagon socket
Hexagon socket head cap
head cap bolts Make parallel the edges of the plate spring
bolt (M3 x 12) (#443)
(M3 x 8) (#442) with the boss on the attaching plate.
(Stroke limitation screw)

Plate spring (#412)

Cap (#415) and


Rotary damper
(#414)
Hexagon socket
Plate spring (#412) Hexagon socket
Attaching plate (#402) head cap bolts
head cap bolts
(M3 x 6) (#441)
(M3 x 6) (#441)

24. Tighten the plastic cover using the two hexagon socket head cap bolts (M3 x 6) on
the front and two hexagon socket head cap bolts (M3 x 8) on the rear.

Hexagon socket head cap bolts (M3 x 8) (#442)

Hexagon socket
head cap bolts
(M3 x 6) (#441)

Plastic cover (#410) Plastic cover (#410)

37
5. Disassembling/Assembling/Adjusting R-4758.B

5-3-2. Assembling Elevation Block (For 50i/55i)


Hexagon socket head
25. Install the condenser holder block, and Condenser holder block
cap bolts (M3 x 8) (#442)
firmly move it in the direction of the arrow.
Tighten it with three hexagon socket head
cap bolts (M3 x 8).
[NOTE]
Be sure to tighten the condenser block
while moving it in the direction of the arrow.

26. While holding the elevation block so that it does not drop, tighten it using the six
hexagon socket head cap bolts (M4 x 16).
[NOTE]
Be sure to tighten the bolts while pushing the elevation block in the direction of the
arrows.

Hexagon socket head cap bolts (M4 x 16) (#53) Elevation block

Hexagon socket head


Elevation block cap bolts (M4 x 16) (#53)

27. While holding the elevation block, place the microscope upside down.
[NOTE]
Since the fine/coarse focus block is removed, the elevation block is not fixed.
Therefore, when placing the microscope to the original position, hold the elevation
block by hand.

28. Perform the following adjustments:


- Stage tilt adjustment (Refer to Section 5-17.)
- Stage centering adjustment (Refer to Section 5-18.)
- Condenser holder tilt adjustment (Refer to Section 5-19.)
- Condenser holder upper limit stopper position adjustment (Refer to Section 5-20.)

38
5. Disassembling/Assembling/Adjusting R-4758.B

5-3-2. Assembling Elevation Block (For 50i/55i)

29. Set the elevator and the coarse/fine focus


drive in the following assembly positions:

[Assembling positions]
With the main body turned upside down (photo):
・ Elevator: Lowest position
・ Coarse/fine focus drive: Where the limit
screw of the gear contacts the lower limit,
as shown in the photograph on the right.
(This applies to both right-handle and Limit screw: To be assembled
left-handle designs.) at this position
Bottom view
After these adjustments are completed,
place the plastic cover under the fine/coarse focus block, and hold these parts by hand firmly.
While engaging the gear of the fine/coarse focus block with the rack of the elevation
block, insert the two pins on the microscope into the two standard holes on the
fine/coarse focus block. Tighten the fine/coarse focus block using the two hexagon
socket head cap bolts (M3 x 12).
[NOTE]
The fine/coarse focus block can be installed to the microscope in both (right and left)
directions. Be sure to install it in the same orientation as disassembling.
Hexagon socket
Standard holes head cap bolts
(M3 x 12) (#52)

Pins

Gear Rack Fine/coarse focus block


Fine/coarse
focus block Plastic cover

39
5. Disassembling/Assembling/Adjusting R-4758.B

5-3-2. Assembling Elevation Block (For 50i/55i)

[Checking the assembly position of coarse/fine focus drive]


When the elevator is at the highest position,
its edge should be approximately 1 mm
higher than the edge of #402.

elevator
1mm #402

Lateral view Edge of #402 Edge of elevator

30. Put the two claws on right and left sides of the plastic cover (#411) into the two
longitudinal holes of the plastic cover (#410).
31. Tighten the plastic cover using the two hexagon socket head cap bolts (M3 x 10).

Hexagon socket head cap bolts (M3 x 10)(#51)

Longitudinal holes
Claws

Special hexagon bolts


32. Tighten the bottom cover using the five (M3 x 6) (#239)
Bottom cover
special hexagon bolts (M3 x 6).

Special hexagon bolt


(M3 x 6) (#239)

40
5. Disassembling/Assembling/Adjusting R-4758.B

5-3-2. Assembling Elevation Block (For 50i/55i)

33. Return the microscope to the horizontal position.


34. When the coarse focus knob and the fine focus knob are turned, check that the
elevation block moves smoothly.
35. When the elevation block is pushed down by hand, check that it moves smoothly.
Check that the elevation block is locked when the elevation block is fully pushed down,
and that the elevation block is moved upward when the elevation block is pushed
down again.
36. In the case of 50i, assemble the power supply assembly. (Refer to Section 5-9.)
In the case of 55i, assemble the control board of the light source unit. (Refer to
Section 5-14.)
37. In the case of 55i, perform the LED position adjustment. (Refer to Section 5-21.)
38. Tighten the plastic cover using the four hexagon socket head cap bolts (M3 x 6).
39. Install the cap.
40. In the case of 50i, connect the lamp to the lamp socket.
41. Move the lamp cover in the opposite direction of the arrow, and install it to the plastic
cover.
42. Install the hexagonal driver to the plastic cover.
Plastic cover (#202) Cap (#233)

Special Lamp cover (#223)


hexagon bolts
(M3 x 6) (#254)

Lamp (#227)

Hexagonal driver (#248)

43 Install the stage. (Refer to Section 5-2.)


44. Install the nosepiece. (Refer to Section 5-2.)
45. Install the condenser lens. (Refer to Section 5-2.)
46. Install the objective lens.
47. Install the eyepiece tube. (Refer to Section 5-2.)

41
5. Disassembling/Assembling/Adjusting R-4758.B

5-4-1. Disassembling Condenser Holder Block (For 50i/55i)

- Disassembling the condenser holder block is not usually necessary, but if liquid (water
etc.) has dropped into the microscope, it should be disassembled.
- The two dovetails (#437 and #438) fixed to the condenser holder block must not be
disassembled at the same time. When re-assembling the dovetail, the other
(assembled) dovetail is used as a reference for assembling.
- Some screws are fixed with adhesive. Apply alcohol to melt the adhesive on these
screws.

WARNING (For ECLIPSE 50i)


- When disassembling the condenser holder, remove the elevation block from the
microscope base.
Be sure to keep the items described in “WARNING” of the elevation block.
(Refer to Section 5-3.)

5-4-1. Disassembling
Hexagon socket
Condenser head cap bolts
1. Remove the elevation block from the
holder block (M3 x 8)(#442)
microscope base according to the steps 1
to 18 of “Disassembling” in elevation block.
(Refer to Section 5-3.)
2. Remove the three hexagon socket head
cap bolts (M3 x 8) (#442), and remove the
condenser holder block from the elevation
block.

3. Loosen the set screw (#446), and pull out


the lever (#439).
Lever (#439)
4. Loosen the set screw (#446), and pull out
the knob block of the condenser holder
block.

Set screws
(#446)

Knob Condenser holder

42
5. Disassembling/Assembling/Adjusting R-4758.B

5-4-1. Disassembling Condenser Holder Block (For 50i/55i)

5. Remove the hexagon socket head cap bolt


(#445) of the stroke limitation screw. Condenser holder

6. Slide the condenser holder block to


remove it from the block (#422).

Hexagon socket
Block (#422) head cap bolt (#445)
(Stroke limitation screw)

7. Remove the two hexagon socket head cap


bolts (M3 x 6) (#441), and remove the Dovetail (#438) Dovetail (#437)
dovetail (#437 or #438).
[NOTE]
When reassembling the dovetail, the other
(assembled) dovetail is used as the
reference for adjustment.

Hexagon socket head


Block (#422)
cap bolts (M3 x 6) (#441)

8. If there is a rusty section on the mounting surface of the dovetail and block, scrape it
using a scraper.

Disassembling male dovetail (#455) and rack (#456)

9. Remove the two hexagon socket head cap Condenser holder


bolts (M3 x 8) (#442), and remove the
male dovetail (#455) from the condenser
holder block.

Hexagon socket head cap


Male Dovetail (#455)
bolts (M3 x 8) (#442)

43
5. Disassembling/Assembling/Adjusting R-4758.B

5-4-1. Disassembling Condenser Holder Block (For 50i/55i)

10. Remove the two special hexagon bolts


(M3 x 4) (#457), and remove the rack Male Dovetail (#455)
(#456) from the removed male dovetail.

Rack (#456) Special hexagon bolts


(M3 x 4)(#457)

Disassembling condenser receptacle (#421)

11. Remove the set screw (#444).


Condenser holder
12. Fully loosen the two knobs (#A6).

Set screw (#444) Knobs (#A6)

13. Remove the two hexagon socket head cap bolts (M3 x 6) (#441), and remove the
spring retainer (#436).
14. Remove the condenser receptacle (#421) from the condenser holder.
[NOTE]
The plate spring (#435) is removed at the same time.
Spring retainer (#436)
Plate spring (#435)

Condenser
receptacle
(#421)

Condenser holder
Hexagon socket head
Condenser holder cap bolts (M3 x 6) (#441)

44
5. Disassembling/Assembling/Adjusting R-4758.B

5-4-1. Disassembling Condenser Holder Block (For 50i/55i)

Disassembling knob block


Remove two special washers (#429),
Belleville spring (#430) and special
15. Loosen the set screw (#446) fixing the washer (#440).
knob (#428). Knob (#428)
16. Turn the knob counterclockwise to
Gear (#426)
remove it from the gear (#426).
[NOTE]
When removing the knob, the two special
washers (#429), Belleville spring (#430)
and special washer (#440) might fall. Be
careful not to lose them.
Set screw (#446)

17. Remove the shaft supporter (#424) from


the gear.

Gear (#426) Shaft supporter (#424)

The disassembly of the condenser holder block is completed.


The disassembled parts should be washed and cleaned.

45
5. Disassembling/Assembling/Adjusting R-4758.B

5-4-2. Assembling Condenser Holder Block (For 50i/55i)

5-4-2. Assembling

When the male dovetail, rack, condenser receptacle, and knob were removed, assemble
these parts first, and then assemble the condenser holder according to the steps 1 to 13.
The assembling procedures for the male dovetail, rack, condenser receptacle and knob
are as follows:
- Assembling for male dovetail and rack: steps 14 to 17
- Assembling for condenser receptacle: steps 18 to 22
- Assembling for knob: steps 23 to 28

1. Lightly tighten the removed dovetail (#437


Dovetail (#438) Dovetail (#437)
or #438) to the block using the two hexagon
socket head cap bolts (M3 x 6).

Block (#422) Hexagon socket head


cap bolts (M3 x 6) (#441)

2. Insert the male dovetail on the condenser Condenser holder


コンデンサホルダ
holder to the block. While pushing the two
dovetails inside, tighten the two hexagon
socket head cap bolts. Male dovetail (#455)
[NOTE]
Move the condenser holder block in the
direction of the arrows, and tighten the
hexagon socket head cap bolts.
Dovetail (#437)

Block (#422) Dovetail (#438)


3. Move the condenser holder block in the
arrow directions, and check that the
condenser holder block moves smoothly
and lightly without slack.
If the condenser holder block does not
move smoothly and lightly without slack,
perform step 2 again.
4. Apply lubricant G7826 to the sliding
surfaces of the dovetail.

46
5. Disassembling/Assembling/Adjusting R-4758.B

5-4-2. Assembling Condenser Holder Block (For 50i/55i)

5. Tighten the hexagon socket head cap bolt Condenser holder


of the stroke limitation screw.

Hexagon socket
Block (#422) head cap bolt (#445)
(Stroke limitation screw)

6. Install the knob of the condenser holder to


the block, and engage the gear of the knob
and the rack of the condenser holder
block.

Block (#422)
ブロック(#422)

Knob block

7. Put the rod into the hole of the shaft Knob (#428) Rod
supporter through the hole of the block.
While holding the knob, move the rod in
the direction shown in the photograph. The
gear should engage firmly with the rack.
Check that the condenser holder moves
smoothly by turning the knob.
If the rotation of the knob is heavy, perform
the adjustment again using the rod.

Block (#422) Shaft supporter (#424)

47
5. Disassembling/Assembling/Adjusting R-4758.B

5-4-2. Assembling Condenser Holder Block (For 50i/55i)

8. Fix the knob using the set screw.


Lever (#439)
Apply adhesive #757 to the screw threads. Lever (#439)
9. Install the lever to the block, and tighten it
using the set screw.
Apply adhesive #757 to the screw threads.

Set screws (#446)


Set screws (#446)
Knob Condenser holder
Knob Condenser holder

10. Install the condenser holder block, and Hexagon socket head
Condenser holder cap bolts (M3 x 8) (#442)
firmly move it in the direction of the arrow.
Tighten it with three hexagon socket head
cap bolts (M3 x 8) to the elevation block.
[NOTE]
Be sure to install the condenser block
while moving it in the direction of the arrow.

11. While holding the elevation block by hand so that it does not drop, tighten it using the
six hexagon socket head cap bolts (M4 x 16).
[NOTE]
Be sure to install the elevation block while pushing it in the direction of the arrows.
Hexagon socket head
cap bolts (M4 x 16) (#43) Elevation block

Hexagon socket head


Elevation block cap bolts (M4 x 16) (#43)

48
5. Disassembling/Assembling/Adjusting R-4758.B

5-4-2. Assembling Condenser Holder Block (For 50i/55i)

12. After the installation is completed, perform the following adjustments.


- Stage tilt adjustment (Refer to Section 5-17.)
- Condenser holder tilt adjustment (Refer to Section 5-19.)
- Condenser holder upper limit stopper position adjustment (Refer to Section 5-20.)
13. After adjustments are completed, assemble the elevation block according to the steps
29 and later of “Assembling” in elevation block. (Refer to Section 5-3.)

Assembling male dovetail (#455) and rack (#456)

14. Tighten the rack to the male dovetail using


雄アリ(#455)
the two special hexagon bolts (M3 x 4). Male dovetail (#455)

ラック(#456)
Rack (#456)
Special hexagon
bolts (M3 x 4) (#457)

15. Lightly tighten the male dovetail to the


Condenser holder
condenser holder using the two hexagon
socket head cap bolts (M3 x 8).

Male dovetail (#455) Hexagon socket head


cap bolts (M3 x 8) (#442)

49
5. Disassembling/Assembling/Adjusting R-4758.B

5-4-2. Assembling Condenser Holder Block (For 50i/55i)

16. Put the two parallel rods (2 mm diameter) into the grooves of the male dovetail and the
holes of the condenser holder. This sets the relative positions of the male dovetail and
the condenser holder.
17. Firmly tighten the two hexagon socket head cap bolts lightly tightened in step 15.
After tightening the bolts, pull out the two parallel rods.

Male dovetail (#455) 2 mm dia. rod Male dovetail (#455)


2 mm dia. rod

Condenser holder Condenser holder

Assembling condenser receptacle (#421)

18. Assemble the condenser receptacle and Plate spring (#435)


plate spring to the condenser holder.
Condenser
Receptacle (#421)

Condenser holder

19. Apply lubricant G7826 to the specified Apply lubricant


portion of the spring retainer as shown in
the photograph.
20. Install the spring retainer to the condenser
holder as shown in the photograph, and
tighten it using the two hexagon socket
head cap bolts (M3 x 6).

Spring retainer Hexagon socket head


(#436) cap bolts (M3 x 6) (#441)

50
5. Disassembling/Assembling/Adjusting R-4758.B

5-4-2. Assembling Condenser Holder Block (For 50i/55i)

21. Apply lubricant G6053 to the tip of the set


Condenser holder
screw, and then thread it to the condenser
holder.
22. Screw the two knobs, and adjust the
position of the condenser receptacle so
that it is mostly located in the center of the
condenser holder.

Set screw (#444) Knob (#A6)


Assembling knob block

23. Apply lubricant L4010 to the specified


Apply lubricant
portion of the gear as shown in the
photograph.
24. Insert the gear into the shaft supporter.

Gear (#426) Shaft supporter (#424)

Insert special washer (#440), special washer (#429),


25. Apply lubricant L4010 to the end of the Belleville spring (#430) and special washer (#429) in order.
shaft supporter.
26. Apply lubricant L4010 on both surfaces of Knob (#428)
special washer (#440), special washer Gear (#426)
(#429), Belleville spring (#430) and special
washer (#429), and install them to the shaft
of the gear in order.
27. Screw the knob to the gear shaft until it
stops.
[NOTE] Set screw (#446)
Take care not to pinch the washers
assembled in step 26.

28. Tighten the set screw to fix the knob with the gear.
After tightening the set screw, apply adhesive #757 to the screw threads.

51
5. Disassembling/Assembling/Adjusting R-4758.B

5-5-1. Disassembling Fine/Coarse Focus Block (For 50i/55i)

5-5-1. Disassembling

- Some screws are fixed with adhesive. Apply alcohol to melt the adhesive on these
screws.
- The following steps 1) to 8) describe the procedure with the fine/coarse focus block
installed to the microscope. But steps 1) to 8) can also be performed when the
fine/coarse focus block has been removed.

1. Remove the fine focus handle cover


(#325).

Fine focus handle cover (#325)

2. Remove the circular nut (#302), and then


turn the fine motion handle (#324)
counterclockwise to remove.
When removing the fine motion handle,
grasp the fine motion handle on the
opposite side.

Fine motion
handle (#324)

Circular nut (#302)

3. Remove the Delrin washer (#314),


Remove the Delrin washer (#314), Belleville
Belleville spring (#313) and Delrin washer
spring (#313) and Delrin washer (#314).
(#314) from the fine motion axle (#323).

Fine motion axle (#323)

52
5. Disassembling/Assembling/Adjusting R-4758.B

5-5-1. Disassembling Fine/Coarse Focus Block (For 50i/55i)

4. Remove the fine motion handle (#303) (with the fine motion handle rubber cap (#301))
on the opposite side from the microscope base.
The fine motion axle (#323) is removed at the same time.
If replacement of the fine motion handle or the fine motion axle is required, replace
these parts after removing the fine motion handle rubber cap (#301) and circular nut
(#302).

Circular nut (#302)

Fine motion
axle (#323)

Fine motion
Fine motion handle (#303) (with handle (#303) Fine motion handle
the fine motion handle rubber cap) rubber cap (#301)

5. Insert a hex-key into the specified hole


Hex-key
shown in the photograph, and loosen the
two set screws (#353) fixing the thrust
retainer (#334).
6. Turn the thrust retainer (#334) counter-
clockwise to remove.

Thrust retainer (#334)

7. Remove the Delrin washer (#315) and


Remove the Delrin washer
spacer (#312) from the coarse motion axle
(#315) and spacer (#312).
(#310).

Coarse motion axle (#310)

53
5. Disassembling/Assembling/Adjusting R-4758.B

5-5-1. Disassembling Fine/Coarse Focus Block (For 50i/55i)

8. Pull out the coarse motion handle (#304) Remove the friction disk (#318)
from the microscope. and two wave washers (#316).
The coarse motion handle with the gear
Coarse motion
case (#307) and coarse motion axle (#310)
handle (#304)
are removed.
Remove the friction disk (#318) and two
wave washers (#316).
[NOTE]
When removing the coarse motion handle,
the ball bearing (#337) may be detached at
the same time. Remove the ball bearing
using tweezers etc.
Check the rotation of the removed ball
bearing using tweezers etc.
If replacement of the gear case or the
coarse motion axle is required, hold the
replacement parts firmly, and turn the other
parts counterclockwise to remove.

Gear case (#307)

Coarse motion axle (#310)

Perform the following steps when the fine/coarse focus block is removed from the
microscope base.
Special hexagon bolts
9. Remove the eyepiece tube. (Refer to Bottom cover (M3 x 6) (#239)
Section 5-2.)
10. Remove the objective lens. Special
11. Remove the condenser lens. hexagon bolts
(Refer to Section 5-2.) (M3 x 6) (#239)
12. Remove the stage. (Refer to Section 5-2.)
13. Place the microscope upside down. Fine/coarse
focus block
14. Remove the five special hexagon bolts
(M3 x 6) (#239), and remove the bottom
cover. Plastic cover (#411)

54
5. Disassembling/Assembling/Adjusting R-4758.B

5-5-1. Disassembling Fine/Coarse Focus Block (For 50i/55i)

15. Remove the two hexagon socket head cap bolts (M3 x 10) (#51), and remove the
plastic cover (#411).
16. Insert a (-) 2 mm screwdriver into the spaces on both sides of the plastic cover shown
in the photograph to release the lock of the claws.
[NOTE]
When releasing the lock, take care not to break the claws.
Hexagon socket head cap bolts (M3 x 10) (#51) (-) screwdriver Claws of plastic cover

17. Hold the fine/coarse focus block firmly by


hand so that it does not drop, and remove
the two hexagon socket head cap bolts
(M3 x 12) (#52). Remove the fine/coarse
focus block and the plastic cover.

Hexagon socket head cap


bolts (M3 x 12) (#52)

Fine/coarse focus block

18. Remove the two hexagon socket head cap


Gear axle (#327)
bolts (M3 x 10) (#352), and remove the
gear axle (#327) from the removed
fine/coarse focus block.

Hexagon socket head


cap bolts (M3 x 10) (#352)

Pinion housing (#305)

55
5. Disassembling/Assembling/Adjusting R-4758.B

5-5-1. Disassembling Fine/Coarse Focus Block (For 50i/55i)

19. Remove the Teflon washer (#336), gear Gear (#326) Teflon washer
(#326) and Teflon washer (#336) from the (#336)
removed gear axle.
Remove the hexagon socket head cap bolt
Gear
Gear axle
(#354) tightened to the gear. (#327)

Teflon washer
(#336) Hexagon
Hexagon socket
socket
head
head
capcap
boltbolt
(#354)
(#354)

20. Remove the pinion (#309) from the pinion


Pinion housing (#305)
housing (#305).

Pinion (#309)

21. Turn the torque adjustment ring (#306)


Pinion housing (#305)
counterclockwise to remove it from the Torque adjustment
pinion housing. ring (#306)

56
5. Disassembling/Assembling/Adjusting R-4758.B

5-5-1. Disassembling Fine/Coarse Focus Block (For 50i/55i)

22. Remove the two kinds of ball bearings Ball bearing (#338) and
(#337 and #338) and special washer special washer (#339)
(#339) from the pinion using tweezers etc.
Check the rotation of the removed ball
bearings using tweezers etc.
[NOTE]
- Special washer sometimes remains in the
pinion housing. When removing, be very
careful.
Pinion (#309) Ball bearing
- When removing the coarse motion axle, (#337)
the ball bearing (#337) is sometimes
removed at the same time.

Disassembling gears inside gear case

23. Remove the three hexagon socket head Three hexagon socket head
cap bolts (M3 x 8), and remove the coarse cap bolts (M3 x 8) (#351)

motion handle (#304).

Coarse motion handle (#304)

24. Remove the two hexagon socket head cap


Hexagon socket head cap bolts (M3 x 8) (#351)
bolts (M3 x 8) (#351), and remove the gear
retaining plate (#308).

Gear retaining plate (#308)

57
5. Disassembling/Assembling/Adjusting R-4758.B

5-5-1. Disassembling Fine/Coarse Focus Block (For 50i/55i)

25. Remove the three gears (#319, #320 and


Gear case (#307)
#321) from the gear case.

Gear (#319)

Gear (#320) Gear (#321)

The disassembly of the fine/coarse focus block is completed.


The disassembled parts should be washed and cleaned.

58
5. Disassembling/Assembling/Adjusting R-4758.B

5-5-2. Assembling Fine/Coarse Focus Block (For 50i/55i)

5-5-2. Assembling

1. Fit the ball bearings (#337 and #338) into


Ball bearing (#338)
the pinion.

Pinion (#309) Ball bearing


(#337)

2. Assemble the pinion into the pinion


Pinion housing (#305)
housing.

Pinion (#309)

Hexagon socket head


3. Tighten the hexagon socket head cap Gear axle (#327)
cap bolt (#354)
bolt (#354) to the gear.
4. Apply lubricant G9110 to both ends of the
gear.
5. Apply lubricant G9110 to the specified
portions of the gear axle as shown in the Apply lubricant
photograph, and insert the gear axle into
the hole of the gear.
Gear (#326)
Apply lubricant

59
5. Disassembling/Assembling/Adjusting R-4758.B

5-5-2. Assembling Fine/Coarse Focus Block (For 50i/55i)

6. Apply lubricant G9110 to the two Teflon


washers, and insert them to both ends of
the gear axle.

Teflon washer
(#336)

Teflon washer
Gear (#326)
(#336)

7. Tighten the gear axle to the pinion housing Gear axle (#327)
using the two hexagon socket head cap
bolts (M3 x 10).

Hexagon socket head cap


bolts (M3 x 10) (#352)

Pinion housing (#305)

Although the following steps can be performed where the fine/coarse focus block is
removed from the microscope base, the following describes the assembling steps where
the fine/coarse focus block is attached to the microscope base.

60
5. Disassembling/Assembling/Adjusting R-4758.B

5-5-2. Assembling Fine/Coarse Focus Block (For 50i/55i)

8. Place the plastic cover under the fine/coarse focus block, and hold these parts by
hand firmly. While engaging the gear of the fine/coarse focus block with the rack of the
elevation block, insert the two pins on the microscope into the two standard holes on
the fine/coarse focus block. Tighten the fine/coarse focus block using the two hexagon
socket head cap bolts (M3 x 12).
[NOTE]
The fine/coarse focus block can be installed to the microscope in both (right and left)
directions. Be sure to install it in the same orientation as disassembling.
Standard Gear Hexagon socket head
holes cap bolt (M3 x 12) (#52)

Pins

Rack

Fine/coarse Plastic cover (#411)


focus block

9. Put the two claws on right and left sides of the plastic cover into the two longitudinal
holes of the plastic cover.
10. Tighten the plastic cover using the two hexagon socket head cap bolts (M3 x 10).
Hexagon socket head cap bolts (M3 x 10) (#51)

Claws Longitudinal holes

61
5. Disassembling/Assembling/Adjusting R-4758.B

5-5-2. Assembling Fine/Coarse Focus Block (For 50i/55i)

11. Tighten the bottom cover using the five Special hexagon bolts (M3 x 6) (#239)
Bottom cove
special hexagon bolts (M3 x 6).

Special hexagon
bolts (M3 x 6) (#239)

Assembling gears to gear case

12. Return the microscope to its original Gear case (#307)


position.
13. Apply lubricant G9110 to both ends of
each gear (#320, #321 and #319), and
insert them to the gear case in order.

Gear (#319)

Gear (#320) Gear (#321)

14. Put the gear retaining plate to the gear Hexagon socket head cap bolts (M3 x 8) (#351)
case so that the slightly curved surface of
the gear retaining plate faces to the gear,
and tighten it using the two hexagon
socket head cap bolts (M3 x 8).
[NOTE]
When assembling, be sure to install each
shaft of the three gears into each hole on
the gear retaining plate.
Gear retaining plate (#308)
After assembling, check that all gears
rotate smoothly.

62
5. Disassembling/Assembling/Adjusting R-4758.B

5-5-2. Assembling Fine/Coarse Focus Block (For 50i/55i)

15. Tighten the coarse motion handle using


Hexagon socket head cap bolts (M3 x 8) (#351)
the three hexagon socket head cap bolts
(M3 x 8).

Coarse motion handle (#304)

16. Apply lubricant L3047 to the threads of


Apply lubricant
the pinion housing and torque adjustment
ring, and screw the torque adjustment
ring to the pinion housing until it stops.

Pinion housing (#305)

Torque adjustment ring (#306)

17. Apply lubricant L3047 to both surfaces of


the two wave washers, and install them
to the pinion housing.

Wave washers (#316)

63
5. Disassembling/Assembling/Adjusting R-4758.B

5-5-2. Assembling Fine/Coarse Focus Block (For 50i/55i)

18. Install the friction disk to the pinion housing.


When installing the friction disk, face its four Protrusion Key way on the
holes outward, and install its protrusion into pinion housing
the key way on the pinion housing.

Four holes
Friction disk (#318)
Friction disk (#318)

19. Install the coarse motion axle (with coarse Pinion housing
motion handle) to the pinion housing.

Coarse motion axle

20. Install the special washer to the coarse


motion axle.
Coarse motion axle (#310)

Special washer (#339)

64
5. Disassembling/Assembling/Adjusting R-4758.B

5-5-2. Assembling Fine/Coarse Focus Block (For 50i/55i)

21. Install the spacer to the coarse motion axle


Spacer (#312)
into the pinion housing. When installing the
spacer, face the grooved surface of the
spacer outward as shown in the Coarse motion
photograph. axle (#310)

Grooved surface
Outward this side

22. Apply lubricant L3047 to both surfaces of Delrin


Delrin washer
washer (#315)
(#315)
the Delrin washer, and fit it to the coarse
motion axle.

Coarse
Coarse motion
motion axle
axle (#310)
(#310)

23. Apply lubricant L3047 to the surface with V-groove of the thrust retainer.
24. Screw the thrust retainer to the coarse motion axle until it stops. Insert a hex-key into
the hole, and tighten the two set screws.
25. Apply adhesive #757 to the threads of the two set screws.

Hex-key
Apply lubricant to the V-grooved surface

Thrust retainer (#334) Thrust retainer (#334)

65
5. Disassembling/Assembling/Adjusting R-4758.B

5-5-2. Assembling Fine/Coarse Focus Block (For 50i/55i)

26. Apply lubricant L3047 to the specified portions of the fine motion axle (with the fine
motion handle) as shown in the photograph, and insert it into the gear case.

Apply lubricant

Fine motion axle (with the fine motion handle)


Fine motion handle

27. Apply lubricant L3047 to the two Delrin デルリンワッシャ(#314)、皿ば


Fit the Delrin washer (#314), Belleville spring
washers and Belleville spring, and fit them ね(#313)、デルリンワッシャ
(#313) and Delrin washer (#314) in order.
to the fine motion axle. (#314)の順に組み込む

微動軸(#323)
Fine motion axle (#323)

28. Screw the fine motion handle to the fine


motion axle, and tentatively fix them with
the circular nut.

Fine motion
handle (#324)

Circular nut (#302)

66
5. Disassembling/Assembling/Adjusting R-4758.B

5-5-2. Assembling Fine/Coarse Focus Block (For 50i/55i)

29. Tentatively install the fine focus handle cover.


30. Wrap the string around the fine focus handle cover as shown in the photograph, and
hook the spring scale on the end of the string. Pull the spring scale in the direction of
the arrow.
Check that the value of the spring scale meets the following specification when the
fine focus handle starts turning.
If not to meet the specification, change the installing position of the fine motion handle
screwed in step28, then check again.

Specification: 11.5 to 21 g
(Fine motion torque: 35 to 65 gf x cm)

Wrap the string around the fine focus handle cover, and hook the spring
scale on the end of the string. Move the scale in the direction of the arrow. Fine focus handle
cover (#325)

31. After adjustment is completed, remove the fine focus handle cover once, and then
firmly tighten the circular nut.
After tightening, check that the fine motion handle is not slack even if the fine motion
handle on the opposite side is turned.
32. Apply adhesive #757 to the threads of the
circular nut.
33. Install the fine focus handle cover to the
fine motion handle.

Fine focus handle cover (#325)

34. Install the stage. (Refer to Section 5-2.)


35. Install the objective lens.
36. Install the condenser lens. (Refer to Section 5-2.)
37. Install the eyepiece tube. (Refer to Section 5-2.)

67
5. Disassembling/Assembling/Adjusting R-4758.B

5-6-1. Disassembling Field Stop Block (For 50i/55i)

- Some screws are fixed with adhesive. Apply alcohol to melt the adhesive on these
screws.

5-6-1. Disassembling

- If the mirror or the lens attached on the lens holder of the field stop block is
disassembled, perform steps 1 to 9 and steps 13 to 17.

1. Remove the eyepiece tube. (Refer to Section 5-2.)


2. Remove the objective lens.
3. Remove the condenser lens. (Refer to Section 5-2.)
4. Remove the stage. (Refer to Section 5-2.) Special hexagon bolts
5. Place the microscope upside down. (M3 x 6) (#239)
Bottom cover
6. Remove the five special hexagon bolts (M3
x 6) (#239), and remove the Special
bottom cover. hexagon bolts
(M3 x 6) (#239)

7. Remove special screw (#208) fixing the Center shaft (#214)


中心軸(#214)
Special screw
lever (#210). (#208)
When removing the lever, the special Diaphragm
Special
washer (#209) might fall. open-shut dial
washer
(#205)
Be careful not to lose it. (#209)
8. Remove the center shaft (#214) to remove
the diaphragm open-shut dial (#205).

Lever (#210)

68
5. Disassembling/Assembling/Adjusting R-4758.B

5-6-1. Disassembling Field Stop Block (For 50i/55i)

9. Remove the three hexagon socket head Hexagon socket head cap Field stop block
cap bolts (M3 x 12) (#256) fixing the field bolts (M3 x 12) (#256)
stop block.

Hexagon socket head cap bolts (M3 x 12) (#256)

10. Remove the four hexagon socket head cap bolts (M3 x 6) (#254), and remove the
blade mount (#207).
[NOTE]
When the blade mount is removed, the eight blades (for 50i; #A1, for 55i: #260) are
removed at the same time.
Be careful not to lose them.
Be very careful not to bend or apply lubricant to the blade.

Hexagon socket head cap bolts (M3 x 6) (#254) Blade (50i:#A1、55i:#260)

Blade mount (#207)

11. Remove the eight blades from the lens


holder (#204). Lens holder (#204) Blade guide way (#206)
12. Remove the blade guide way (#206) from
the lens holder (#204).

69
5. Disassembling/Assembling/Adjusting R-4758.B

5-6-1. Disassembling Field Stop Block (For 50i/55i)

Disassembling mirror and lens

13. Scrape off the adhesive joints between the Mirror (#2) Adhesive
mirror (#2) and the lens holder (#204).

14. Remove the mirror from the lens holder,


with care to avoid damage to the mirror.
If the adhesive cannot be removed, warm
it using a heat gun or the like.

Adhesive Lens holder (#204)

15. Scrape off the adhesive from the perimeter Lens (#3)
of the lens (#3). Lens holder (#204)
Adhesive

16. Remove the lens from the lens holder.


If the adhesive cannot be removed, warm
it using a heat gun or the like.

17. Remove any remaining adhesive on the lens holder after the lens or the mirror is
removed completely. Then, clean the surface with alcohol.

The disassembly of the field stop block is completed.


The disassembled parts should be washed and cleaned.

70
5. Disassembling/Assembling/Adjusting R-4758.B

5-6-2. Assembling Field Stop Block (For 50i/55i)

5-6-2. Assembling

- The assembling procedures for the lens and mirror are explained in steps 22 to 28.

1. Apply lubricant G7813 to the specified Blade guide way (#206)


portion of the blade guide way as shown in
the photograph.

Shaft

Apply lubricant outside surface

2. Insert the shaft of the blade guide way into Blade guide way (#206)
the groove of the lens holder, and fully turn
it in the direction of the arrow.

Lens holder (#204) Shaft of the blade guide way

3. Install the eight blades to the blade guide Blades (8pcs)


in the open state while putting the
projection of the blade into the groove of
the blade guide.

Lens holder (#204)

71
5. Disassembling/Assembling/Adjusting R-4758.B

5-6-2. Assembling Field Stop Block (For 50i/55i)

4. Install the blade mount so that the slightly 2 mm dia. rod Projection of the blade
curved surface of it faces the blade. 2 mm dia.
Blade mount rod
[NOTE]
When assembling, be sure to insert the
projection of the blade into the hole of the
blade mount.

5. Insert the two 2 mm diameter rods into the


two specified holes shown in the
photograph so that the relative position of Hexagon socket head cap bolts (M3 x 6) (#254)
the lens holder and the blade mount is
aligned.
6. Tighten the blade mount using the four hexagon socket head cap bolts (M3 x 6).
7. Remove the two rods inserted in step 5.
8. When the blade guide way is turned by hand, check that the blades open and shut
smoothly.
9. While opening or shutting the blade 10 to 20 times, blow away any dust from the
blades using an airbrush or the like.
10. Tighten the field stop block using the three Hexagon socket head Field stop block
hexagon socket head cap bolts (M3 x 12). cap bolt (M3 x 12) (#256)

Hexagon socket head cap bolt (M3 x 12) (#256)

11. Apply lubricant G7813 to the specified Center shaft (#214)


Apply lubricant
portion of the center shaft, and apply
adhesive #757 to the tip of its screw as Diaphragm
shown in the photograph. open-shut
12 Tighten the diaphragm open-shut dial to dial (#205)
the lens holder using the center shaft.
Apply lubricant

72
5. Disassembling/Assembling/Adjusting R-4758.B

5-6-2. Assembling Field Stop Block (For 50i/55i)

13. Apply lubricant G7813 to the shaft of the special screw.


14. While inserting the shaft of the blade guide way into the U-groove of the lever,
assemble the lever to the lens holder, and tighten it with special screw and
special washer.

Special screw (#208)

Special
washer
(#209)
Lever (#210)

U-groove

Shaft of the blade guide way Lever (#210)

Special hexagon bolts


15. When the lever is moved by hand, check Bottom cover (M3 x 6) (#239)
that the blades open and shut smoothly.
16. Tighten the bottom cover using the
Special
five special hexagon bolts (M3 x 6). hexagon bolts
(M3 x 6) (#239)

17. Return the microscope to its original position.


18. Install the stage. (Refer to Section 5-2.)
19. Install the condenser lens. (Refer to Section 5-2.)
20. Install the objective lens.
21. Install the eyepiece tube. (Refer to Section 5-2.)

73
5. Disassembling/Assembling/Adjusting R-4758.B

5-6-2. Assembling Field Stop Block (For 50i/55i)

Assembling mirror and lens

22. Clean the adhesive coating surfaces on Mirror (#2) Adhesive


the lens holder using alcohol.
23. Apply adhesive #2057 to the adhesive
coating surfaces of the mirror on the lens
holder.
24. Cement the mirror while pushing it against
the lens holder.

Adhesive Lens holder (#204)

25. Clean the lens with alcohol.


26. Place the lens onto the lens holder, with the less curved side facing the lens holder.
27. Bond the lens by applying adhesive (#2057) to a few positions around the lens and
shedding ultraviolet rays.

Lens (#3) Lens holder (#204) Small curved


surface of the lens
Adhesive

28. After the lens and mirror are cemented, blow away any dust from the lens and the
mirror using an airbrush or the like.

74
5. Disassembling/Assembling/Adjusting R-4758.B

5-7-1. Disassembling ND Filter Block (For 50i)

5-7-1. Disassembling

1. Remove the eyepiece tube. (Refer to Section 5-2.)


2. Remove the objective lens.
3. Remove the condenser lens. (Refer to Section 5-2.)
4. Remove the stage. (Refer to Section 5-2.)
5. Place the microscope upside down.
Special hexagon
6. Remove the five special hexagon bolts (M3 Bottom cover bolts (M3 x 6) (#239)
x 6) (#239), and remove the bottom cover.

Special
hexagon bolts
(M3 x 6) (#239)

7. Disconnect the connector connected to the brightness controller board (#501).


8. Remove a hexagon socket head cap bolt (M3 x 8) (#255) fixing the harness, and
release the harness clamper.
9. Remove a hexagon socket head cap bolt (M3 x 8) (#255), and remove the ND filter
block.
Remove the harness from the clamper on the ND filter.

Hexagon socket head


cap bolt (M3 x 8) (#255) ND filter block Clamper ND filter block
Harness
connector
Brightness
controller
board (#501)

Hexagon socket head cap bolt (M3 x 8) (#255)

75
5. Disassembling/Assembling/Adjusting R-4758.B

5-7-1. Disassembling ND Filter Block (For 50i)

10. Release three claws of the ND filter holder ND filter holder (#215)
(#215), and remove the ND filter (#6). Claws

ND filter (#6)

11. Remove the projections on the ND filter holder from the holes on the lever (#216) and
the lever supporter (#217).

Lever (#216)
Projection
Projections

ND filter holder (#215) Lever supporter (#217) ND filter holder (#215)

12. Pull out the two levers from the lever


supporter. Levers (#216)

Lever supporter (#217)

76
5. Disassembling/Assembling/Adjusting R-4758.B

5-7-2. Assembling ND Filter Block (For 50i)

5-7-2. Assembling

1. Apply lubricant G9124 to the specified


Apply lubricant Levers (#216)
portions of the two levers as shown in the
photograph, and insert them into the holes
on the lever supporter.
[NOTE]
Do not apply lubricant other than the
specified portions.

Lever supporter (#217) Apply lubricant

2. Install the ND filter holder to the two levers and lever supporter in the direction as
shown in the photograph.
3. Put the projections on the ND filter holder into each hole on the two levers and the
lever supporter.

Lever (#216) Projection


Projections

ND filter holder (#215) Lever supporter ND filter holder (#215)

4. Install the ND filter to the ND filter holder ND filter holder (#215)


so that an evaporated surface is in the Claws
orientation shown in the photograph.

ND filter (#6)
(An evaporated surface faces upward.)

77
5. Disassembling/Assembling/Adjusting R-4758.B

5-7-2. Assembling ND Filter Block (For 50i)

5. Put the harness into the hole on the ND Clamper ND filter block
filter block, and secure it using the clamper.
Harness connector

6. Tighten the ND filter block using a hexagon Hexagon socket head cap ND filter block
socket head cap bolt (M3 x 8). bolt (M3 x 8) (#255)
7. Secure the harness using a hexagon head Brightness
cap bolt (M3 x 8). controller
8. Connect the harness connector to the board (#501)
brightness controller board.

Hexagon socket head cap bolt (M3 x 8) (#255)

Special hexagon
9. Tighten the bottom cover using the five Bottom cover bolts (M3 x 6) (#239)
special hexagon bolts (M3 x 6).

Special
hexagon bolt
(M3 x 6) (#239)

10. Return the microscope to its original position.


11. Install the stage. (Refer to Section 5-2.)
12. Install the condenser lens. (Refer to Section 5-2.)
13. Install the objective lens.
14. Install the eyepiece tube. (Refer to Section 5-2.)

78
5. Disassembling/Assembling/Adjusting R-4758.B

5-8-1. Disassembling Lighting Block (For 50i)

- Some screws and diffuser are fixed with adhesive. Apply alcohol to melt the
adhesive on these.

5-8-1. Disassembling

1. Remove the eyepiece tube. (Refer to Section 5-2.)


2. Remove the objective lens.
3. Remove the condenser lens. (Refer to Section 5-2.)
4. Remove the stage. (Refer to Section 5-2.)
5. While pressing the two portions of the lamp
cover (#223) in the arrow direction (a)
shown in the figure, move it in the arrow
direction (b) to remove. Lamp cover
(#223)

(a)

(a) (b)

6. Disconnect the lamp (#227).


Plastic cover (#202) Cap (#233)
7. Pull out the hexagonal driver (#248).
8. Remove the four hexagon socket head cap
bolts (M3 x 6) (#254), and remove the Hexagon
plastic cover (#202). socket head
9. Remove the cap (#233). cap bolts
(M3 x 6) (#254)

Lamp (#227)

Hexagonal driver (#248)

79
5. Disassembling/Assembling/Adjusting R-4758.B

5-8-1. Disassembling Lighting Block (For 50i)

10. Remove the power supply assembly.


(Refer to Section 5-9.)
11. Place the microscope upside down.
12. Remove the five special hexagon bolts Special hexagon
Bottom cover bolts (M3 x 6) (#239)
(M3 x 6) (#239), and remove the bottom
cover.
Special
hexagon bolts
(M3 x 6) (#239)

13. Remove the four hexagon socket head cap Hexagon socket head
Lighting block cap bolts (M3 x 8) (#255)
bolts (M3 x 8) (#255), and remove
the lighting block.

14. Remove the two diffusers (#4 and #5) from Diffuser (#5) Diffuser (#4)
the lens holder (#221).

Lens holder (#221)

80
5. Disassembling/Assembling/Adjusting R-4758.B

5-8-1. Disassembling Lighting Block (For 50i)

15. Remove the two screws (#257), and


remove the cable with lamp socket (#505)
Screws (#257)
from the lamp socket holder (#226).

Cable with lamp


Lamp socket socket (#505)
holder (#226)

16. Remove the four hexagon socket head cap bolts (M3 x 6), and remove the collector
lens (#1) and lamp socket holder (#226).
[NOTE]
When removing the lamp socket holder, the collector lens is removed at a same time.
Be careful not to drop or scratch the lens.

Collector lens
Lamp socket
(#1)
holder (#226)
Hexagon socket head cap bolts
(M3 x 6)(#254)on both sides

81
5. Disassembling/Assembling/Adjusting R-4758.B

5-8-2. Assembling Lighting Block (For 50i)

5-8-2. Assembling

1. Clean the collector lens.


2. After installing the collector lens to the lens holder in the direction as shown in the
photograph, tighten the lamp socket holder using the four hexagon socket head cap
bolts (M3 x 6).

Collector lens
Lamp socket
(#1)
holder (#226)
Hexagon socket head cap bolts
M3 x 6)(#254)on both sides

3. Pass the cable of the cable with lamp


socket through the hole on the lamp socket Screws (#257)
holder, and tighten the socket with the two
screws.
Apply adhesive #2014 to the spaces
between the screws and the lamp socket.

Cable with lamp


Lamp socket socket (#505)
holder (#226)

4. Apply adhesive #757 (Clear) to the side Diffuser (#5) Diffuser (#4)
edges of the two diffusers which contact
the lens holder. Mat surface
5. Install the two diffusers to the lens holder in
the orientation shown in the photograph.
[NOTE]
Be sure to install the mat surface of the
diffusers in the orientation shown in the Mat surface

photograph.
Lens holder (#221)

82
5. Disassembling/Assembling/Adjusting R-4758.B

5-8-2. Assembling Lighting Block (For 50i)


Hexagon socket head
6. Place the microscope upside down. Lighting block cap bolts (M3 x 8) (#255)
7. Install the lighting block, and firmly move it
in the direction of the arrows. Tighten it
with four hexagon socket head cap bolts
(M3 x 8).

Special hexagon bolts


8. Tighten the bottom cover using the five
Bottom cover (M3 x 6) (#239)
special hexagon bolts (M3 x 6).

Special
hexagon bolts
(M3 x 6) (#239)

9. Return the microscope to its original position.


10. Install the power supply assembly. (Refer to Section 5-9.)
11. Tighten the plastic cover using the four hexagon socket head cap bolts (M3 x 6).
12. Install the cap.
13. Connect the lamp to the lamp socket.
14. Move the lamp cover in the opposite direction of the arrow, and install it to the plastic
cover.
15. Install the hexagonal driver (#248) to the plastic cover.
Plastic cover (#202) Cap (#233)

Hexagon Lamp cover


socket head (#223)
cap bolts
(M3 x 6) (#254)

Lamp (#227)

Hexagonal driver (#248)

83
5. Disassembling/Assembling/Adjusting R-4758.B

5-8-2. Assembling Lighting Block (For 50i)

16. Install the stage. (Refer to Section 5-2.)


17. Install the condenser lens. (Refer to Section 5-2.)
18. Install the objective lens.
19. Install the eyepiece tube. (Refer to Section 5-2.)

84
5. Disassembling/Assembling/Adjusting R-4758.B

5-9-1. Disassembling Power Supply Assembly (For 50i)

5-9-1. Disassembling

1. While pressing the two portions of the


lamp cover (#223) in the arrow direction (a)
shown in the figure, move it in the arrow Lamp cover
direction (b) to remove. (#223)

(a)

(a) (b)

2. Disconnect the lamp (#227). Plastic cover (#202) Cap (#233)


3. Pull out the hexagonal driver (#248).
4. Remove the four hexagon socket head cap
bolts (M3 x 6) (#254), and remove the Hexagon
plastic cover (#202). socket head
cap bolts
5. Remove the cap (#233). (M3 x 6) (#254)

Lamp (#227)

Hexagonal driver (#248)

6. Remove the four hexagon socket head Hexagon socket head cap
cap bolts (M3 x 6) (#254), and remove the bolts (M3 x 6) (#254)

attaching plate (#231).

Attaching plate (#231)

85
5. Disassembling/Assembling/Adjusting R-4758.B

5-9-1. Disassembling Power Supply Assembly (For 50i)

7. Disconnect the harness from the Harness connector


connector (CN1) on the power supply Power supply
assembly (#502). assembly (#502)
8. Remove the three hexagon socket
head cap bolts (M3 x 6) (#254) Hexagon socket
fixing the power supply assembly. head cap bolts
(M3 x 6) (#254)

9. Lift up the power supply assembly, and


disconnect the harness connectors (CN2
Power supply
and CN3). Then, remove the power supply
assembly (#502)
assembly.

Connector (CN3)

Connector (CN2)

86
5. Disassembling/Assembling/Adjusting R-4758.B

5-9-2. Assembling Power Supply Assembly (For 50i)

5-9-2. Assembling

1. Connect the harness connectors to CN2


Power supply
and CN3 on the power supply assembly.
assembly (#502)

Connector (CN3)

Connector (CN2)

2. Tighten the power supply assembly using Connector (CN1)


the three hexagon socket head cap bolts Power supply
(M3 x 6). assembly (#502)
3. Connect the harness connector
Hexagon socket
to CN1 located at the top of the
head cap bolts
power supply assembly. (M3 x 6) (#254)

4. Tighten the attaching plate (#231) using Hexagon socket head cap
the four hexagon socket head cap bolts bolts (M3 x 6) (#254)
(M3 x 6).

Attaching plate
Attaching plate (#231)
(#231)

87
5. Disassembling/Assembling/Adjusting R-4758.B

5-9-2. Assembling Power Supply Assembly (For 50i)

5. Install the cap (#233). Plastic cover (#202) Cap (#233)


6. Tighten the plastic cover using the four
hexagon socket head cap bolts (M3 x 6).
7. Install the hexagonal driver. Hexagon
socket head
8. Connect the lamp. cap bolts
(M3 x 6) (#254)

Lamp (#227)

Hexagonal driver (#248)


9. Move the lamp cover in the opposite
direction of the arrow, and install it to the
plastic cover.
Lamp cover
(#223)

(a)

(a) (b)

88
5. Disassembling/Assembling/Adjusting R-4758.B

5-10-1. Disassembling Brightness Control Board (For 50i)

5-10-1. Disassembling

1. Remove the eyepiece tube. (Refer to Section 5-2.)


2. Remove the objective lens.
3. Remove the condenser lens. (Refer to Section 5-2.)
4. Remove the stage. (Refer to Section 5-2.)
5. Loosen a set screw (#258), and remove
the brightness control dial (#241). Set screw (#258)

Brightness control dial (#241)


6. Place the microscope upside down.
7. Remove the five special hexagon bolts Special hexagon
Bottom cover bolts (M3 x 6) (#239)
(M3 x 6) (#239), and remove the bottom
cover.
Special
hexagon bolts
(M3 x 6) (#239)

8. Remove the field stop block. (Refer to Brightness controller Connector (J1)
Section 5-6.) board (#501)
9. Disconnect the harness connector (J1) on
the brightness controller board (#501).
10. Remove the two hexagon socket head cap
bolts (M3 x 6) (#254), and remove the
brightness controller board.

Hexagon socket head cap bolts (M3 x 6) (#254)

89
5. Disassembling/Assembling/Adjusting R-4758.B

1 5-10-1. Disassembling Brightness Control Board (For 50i)

11. Remove the two special hexagon bolts Special hexagon bolts (M3 x 6) (#239)
(M3 x 6) (#239), and remove the two
bending plates from the brightness
controller board.

Brightness controller Bending


Bending plates
plates (#218)
(#218)
board (#501)

90
5. Disassembling/Assembling/Adjusting R-4758.B

5-10-2. Assembling Brightness Control Board (For 50i)

5-10-2. Assembling

1. Tighten the two bending plates to the


Special hexagon bolts (M3 x 6) (#239)
brightness controller board using the two
special hexagon bolts (M3 x 6).

Brightness controller board (#501) Bending plates (#218)

Brightness controller
2. Tighten the brightness controller board Connector (J1)
board (#501)
using the two hexagon socket head cap
bolts (M3 x 6).
3. Connect the harness connector to the
connector (J1) on the brightness controller
board.

Hexagon socket head cap bolts (M3 x 6) (#254)

4. Install the field stop block. (Refer to


Section 5-6.)
Special hexagon
5. Tighten the bottom cover using the Bottom cover bolts (M3 x 6) (#239)
five special hexagon bolts (M3 x 6).

Special
hexagon bolts
(M3 x 6) (#239)

91
5. Disassembling/Assembling/Adjusting R-4758.B

5-10-2. Assembling Brightness Control Board (For 50i)

6. Return the microscope to its original position.


7. Turn the shaft of the variable resistor so that the D-cut of the shaft faces top.
Install the brightness control dial to the shaft so that a set screw aligns with the
D-cut of the shaft, and then tighten the set screw.
When the brightness control dial is turned fully clockwise and counterclockwise,
check that the indication of the brightness control dial is aligned with the
indication on the cover.

Set screw (#258)

Shaft of the variable resistor Brightness control dial (#241)

8. Install the stage. (Refer to Section 5-2.)


9. Install the condenser lens. (Refer to Section 5-2.)
10. Install the objective lens.
11. Install the eyepiece tube. (Refer to Section 5-2.)

92
5. Disassembling/Assembling/Adjusting R-4758.B

5-11-1. Disassembling Seesaw Switch (For 50i)

5-11-1. Disassembling

1. Remove the eyepiece tube. (Refer to Section 5-2.)


2. Remove the objective lens.
3. Remove the condenser lens. (Refer to Section 5-2.)
4. Remove the stage. (Refer to Section 5-2.)
5. Place the microscope upside down. Special hexagon
6. Remove the five special hexagon bolts Bottom cover bolts (M3 x 6) (#239)
(M3 x 6) (#239), and remove the bottom
cover. Special
hexagon bolts
(M3 x 6) (#239)

7. Disconnect the terminals connected to the


Seesaw switch (#511)
seesaw switch (#511).
8. Release the claws on both sides of the
seesaw switch, and remove it outward.
Terminal (1 pin)
side

side

Claw

93
5. Disassembling/Assembling/Adjusting R-4758.B

5-11-2. Assembling Seesaw Switch (For 50i)

5-11-2. Assembling

1. Install the seesaw switch to the microscope, and secure it by the claws located on
both sides.
[NOTE]
Take care to install the correct orientation of the seesaw switch.
2. Connect the terminals to the seesaw switch.
[CAUTION]
Make sure you always connect to the correct wire to each pin.
1 pin: black wire of cable with AC socket (#503)
2 pin: black wire of cable with connector (#506)
3 pin: white wire of cable with AC socket (#503)
4 pin: white wire of cable with connector (#506)

Seesaw switch (#511)

Bottom cover side

Terminal (1 pin)
side
side side

side

Terminals location

Claw

3. Tighten the bottom cover using the five Special hexagon


Bottom cover bolts (M3 x 6) (#239)
special hexagon bolts (M3 x 6).

Special
hexagon bolts
(M3 x 6) (#239)

4. Return the microscope to its original position.


5. Install the stage. (Refer to Section 5-2.)
6. Install the condenser lens. (Refer to Section 5-2.)
7. Install the objective lens.
8. Install the eyepiece tube. (Refer to Section 5-2.)

94
5. Disassembling/Assembling/Adjusting R-4758.B

5-12-1. Disassembling Color Compensation Filter Unit (For 55i)

5-12-1. Disassembling

1. Remove the eyepiece tube. (Refer to Section 5-2.)


2. Remove the objective lens.
3. Remove the condenser lens. (Refer to Section 5-2.)
4. Remove the stage. (Refer to Section 5-2.)
5. Place the microscope upside down. Special hexagon
6. Remove the five special hexagon bolts Bottom cover bolts (M3 x 6) (#239)
(M3 x 6) (#239), and remove the bottom
cover.
Special
hexagon bolts
(M3 x 6) (#239)

7. Disconnect the flat cable connector (#507) Switch board (#502)


from the switch board (#502).
Connector

8. Remove a hexagon socket head cap bolt Color compensation filter unit
(M3 x 8) (#255), and remove the color
compensation filter unit.

Hexagon socket head cap bolt (M3 x 8) (#255)

95
5. Disassembling/Assembling/Adjusting R-4758.B

5-12-1. Disassembling Color Compensation Filter Unit (For 55i)

9. Release three claws of the ND filter holder


(#215), and remove the color compensa- Claws
tion filter (#5).

Color compensation filter (#5)

ND filter holder (#215)

10. Remove the projections on the ND filter holder from each hole on the two levers
(#216) and the lever supporter (#217).

Lever (#216) Projection


Projections

ND filter holder (#215) Lever supporter (#217) ND filter holder (#215)

11. Pull out the two levers from the lever


Levers (#216)
supporter.

Lever supporter (#217)

96
5. Disassembling/Assembling/Adjusting R-4758.B

5-12-2. Assembling Color Compensation Filter Unit (For 55i)

5-12-2. Assembling

1. Apply lubricant G9124 to the specified


Apply lubricant Levers (#216)
portions of the two levers as shown in the
photograph, and insert them into the holes
on the lever supporter.
[NOTE]
Do not apply lubricant other than to the
specified portions.

Lever supporter (#217) Apply lubricant

2. Install the ND filter holder to the lever supporter in the direction shown in the
photograph.
3. Put the projections on the ND filter holder into each hole on the two levers.

Levers (#216) Projection

Projections

ND filter holder (#215) Lever supporter (#217) ND filter holder (#215)

4. Install the color compensation filter to the


ND filter holder so that an evaporated Claws
surface of the filter is in the orientation
shown in the photograph.

Color compensation filter (#5)


(An evaporated surface faces upward.)

ND filter holder (#215)

97
5. Disassembling/Assembling/Adjusting R-4758.B

5-12-2. Assembling Color Compensation Filter Unit (For 55i)

5. Put the connector of the flat cable through Color compensation filter unit
the hole on the color compensation filter
unit.
6. Tighten the color compensation filter unit
using a hexagon socket head cap bolt
(M3 x 8).

Hexagon socket head cap bolt (M3 x 8) (#255)

7. Connect the connector of the flat cable to Switch board (#502)


the switch board.
Connector

Special hexagon
8. Tighten the bottom cover using the five Bottom cover bolts (M3 x 6) (#239)
special hexagon bolts (M3 x 6).

Special
hexagon bolt
(M3 x 6) (#239)

9. Return the microscope to its original position.


10. Install the stage. (Refer to Section 5-2.)
11. Install the condenser lens. (Refer to Section 5-2.)
12. Install the objective lens.
13. Install the eyepiece tube. (Refer to Section 5-2.)

98
5. Disassembling/Assembling/Adjusting R-4758.B

5-13-1. Disassembling Diffuser Unit (For 55i)

5-13-1. Disassembling

1. Remove the eyepiece tube. (Refer to Section 5-2.)


2. Remove the objective lens.
3. Remove the condenser lens. (Refer to Section 5-2.)
4. Remove the stage. (Refer to Section 5-2.)
5. Place the microscope upside down. Special hexagon
6. Remove the five special hexagon bolts Bottom cover bolts (M3 x 6) (#239)
(M3 x 6) (#239), and remove the bottom
cover.
Special hexagon
bolt
(M3 x 6) (#239)

Hexagon socket head cap


7. Remove the two hexagon socket head cap
Diffuser unit bolts (M3 x 8) (#255)
bolts (M3 x 8) (#255), and remove the
diffuser unit.

99
5. Disassembling/Assembling/Adjusting R-4758.B

5-13-2. Assembling Diffuser Unit (For 55i)

5-13-2. Assembling

1. Clean the adhesive coating surface of the Diffuser (#4) (Mat surface is lowered.)
diffuser on the lens holder using alcohol.
2. Apply adhesive #2061 to the adhesive
Adhesive
coating surfaces of the diffuser on the lens
holder.
[NOTE]
Do not mistake the installing position of the
diffuser and the lens to the lens holder.

3. Cement the diffuser while pushing it Lens holder (#221-1)


against the lens holder.
4. Clean the adhesive coating surface of the lens on the lens holder using alcohol.
5. Apply adhesive #2061 to the adhesive coating surfaces of the lens on the lens holder.
6. Install the lens to the lens holder so that the small curved surface of the lens faces the
lens holder. Cement the lens to the lens holder while pushing it against the lens
holder.

Lens (#1)
Large curved surface
Adhesive

Lens holder (#221-1)

7. After the lens and/or diffuser is cemented, Hexagon socket head


blow away any dust from the lens and Diffuser unit cap bolts (M3 x 8)
diffuser using the airbrush or the like.
8. Install the diffuser unit, and firmly move it in
the direction of the arrows. Tighten it with
two hexagon socket head cap bolts (M3 x
8).

Push it in these directions.

100
5. Disassembling/Assembling/Adjusting R-4758.B

5-13-2. Assembling Diffuser Unit (For 55i)


Special hexagon
9. Tighten the bottom cover using the five Bottom cover bolts (M3 x 6) (#239)
special hexagon bolts (M3 x 6).

Special
hexagon bolt
(M3 x 6) (#239)

10. Return the microscope to its original position.


11. Install the stage. (Refer to Section 5-2.)
12. Install the condenser lens. (Refer to Section 5-2.)
13. Install the objective lens.
14. Install the eyepiece tube. (Refer to Section 5-2.)

101
5. Disassembling/Assembling/Adjusting R-4758.B

5-14-1. Disassembling Light Source Unit (For 55i)

5-14-1. Disassembling

1. While pressing the two portions of the


lamp cover (#223) in the arrow direction
(a) shown in the photograph, move it in the Lamp cover
arrow direction (b) to remove. (#223)

(a)

(a) (b)

2. Pull out the hexagonal driver (#248). Cap (#233)


3. Remove the four hexagon socket head cap
bolts (M3 x 6) (#254), and remove the
plastic cover (#202).
Hexagon socket
4. Remove the cap (#233). head cap bolts
(M3 x 6) (#254)

Plastic cover (#202)

Special hexagon bolts


5. Remove the four special hexagon bolts (M3 x 6) (#239)
(M3 x 6) (#239), and remove the control
board (#509).
6. Disconnect the flat cable from the
connector (J4) on the main board block of
the control board.

Connector (J4)

Control board (#509)

102
5. Disassembling/Assembling/Adjusting R-4758.B

5-14-1. Disassembling Light Source Unit (For 55i)

7. Remove the two special hexagon bolts Special hexagon bolts (M3 x 6)(#239)
(M3 x 6) (#239) fixing the LED board
(#503).

LED board (with battery case) (#503)

8. Disconnect the LED board from the Control board


connector (J2) on the control board.

Connector

Connector (J2) LED board

9. While pressing the claw on the battery case (#224), remove the LED board from the
battery case.
[NOTE]
When removing the LED board, be careful not to deform the terminals on the LED
board.
LED board (#503) LED board (#503)

Claw
Terminals
Battery case (#224)

103
5. Disassembling/Assembling/Adjusting R-4758.B

5-14-2. Assembling Light Source Unit (For 55i)

5-14-2. Assembling

1. While pushing the claw on the battery case,


LED board (#503)
move the LED board in the opposite
direction of the arrow to install to the
battery case.
[NOTE]
When installing the LED board, be careful
not to deform the terminals on the LED
board.
Claw

Battery case (#224)

2. Connect the connector (J2) on the LED board to the control board.
3. Tighten the LED board to the control board using the two special hexagon bolts
(M3 x 6).

Control board Special hexagon bolts (M3 x 6)(#239)

Connector

LED board (with battery case) (#503)


Connector (J2) LED board

Special hexagon bolts


4. Connect the flat cable to the connector (M3 x 6) (#229)
(J4) on the main board block of the control
board.
5. Tighten the control board using the four
special hexagon bolts (M3 x 6).

Connector
Connector (J4)
(J4)

Control
Control board
board (#509)
(#509)

104
5. Disassembling/Assembling/Adjusting R-4758.B

5-14-2. Assembling Light Source Unit (For 55i)

6. Tighten the plastic cover using the four


Cap (#233)
hexagon socket head cap bolts (M3 x 6).

Hexagon socket
head cap bolts
(M3 x 6) (#254)

Plastic cover (#202)

7. Move the lamp cover in the opposite


direction of the arrow, and install it to the
plastic cover.
Lamp cover
8. Install the hexagonal driver. (#223)

9. After the assembling is completed, perform the LED position adjustment. (Refer to
Section 5-21.)

105
5. Disassembling/Assembling/Adjusting R-4758.B

5-15-1. Disassembling Switch Board (For 55i)

5-15-1. Disassembling

1. Remove the eyepiece tube. (Refer to Section 5-2.)


2. Remove the objective lens.
3. Remove the condenser lens. (Refer to Section 5-2.)
4. Remove the stage. (Refer to Section 5-2.)
5. Loosen a set screw (#258) fixing the Set screw (#258)
brightness control dial (#241), and remove
the brightness control dial.

Brightness control dial (#241)

Special hexagon
6. Place the microscope upside down. Bottom cover bolts (M3 x 6) (#239)
7. Remove the five special hexagon bolts
(M3 x 6) (#239), and remove the
Special
bottom cover.
hexagon bolts
(M3 x 6) (#239)

8. Remove the field stop block. (Refer to Switch board (#502) Connector (J1)
Section 5-6.)
9. Disconnect the flat cable with connector
(#507) from the connector (J1) on the
switch board (#502).
10. Remove the two hexagon socket head cap
bolts (M3 x 6) (#254), and remove the Flat cable with
switch board. connector
(#507)

Hexagon socket head cap bolts (M3 x 6) (#254)

106
5. Disassembling/Assembling/Adjusting R-4758.B

5-15-1. Disassembling Switch Board (For 55i)

11. Remove the two special hexagon bolts Special hexagon bolts (M3 x 6) (#239)
(M3 x 6) (#239), and remove the two
bending plates from the switch board.

Bending plates (#218)


Switch board (#502)

107
5. Disassembling/Assembling/Adjusting R-4758.B

5-15-2. Assembling Switch Board (For 55i)

5-15-2. Assembling

1. Tighten the two bending plates to the Special hexagon bolts (M3 x 6) (#239)
switch board using the two special
hexagon bolts (M3 x 6).

Bending plates (#218)


Switch board (#502)

2. Tighten the switch board using the two Switch board (#502) Connector (J1)
hexagon socket head cap bolts (M3 x 6).
3. Connect the flat cable with connector to
the connector (J1) on the switch board.

Flat cable with


connector
(#507)
Hexagon socket head cap bolts (M3 x 6) (#254)

4. Install the field stop block. (Refer to Section Special hexagon


Bottom cover bolts (M3 x 6) (#239)
5-6.)
5. Tighten the bottom cover using the
five special hexagon bolts (M3 x 6). Special
hexagon bolt
(M3 x 6) (#239)

108
5. Disassembling/Assembling/Adjusting R-4758.B

5-15-2. Assembling Switch Board (For 55i)

6. Return the microscope to its original position.


7. Turn the shaft of the variable resistor so that the D-cut of the shaft faces top. Install the
brightness control dial to the shaft so that a set screw aligns with the D-cut of the shaft,
and then tighten the set screw.
When the brightness control dial is turned fully clockwise and counterclockwise, check
that the indication of the brightness control dial is aligned with the indication on the
cover.
Set screw (#258)

Brightness control dial (#241)


Shaft of the variable resistor

8. Install the stage. (Refer to Section 5-2.)


9. Install the condenser lens. (Refer to Section 5-2.)
10. Install the objective lens.
11. Install the eyepiece tube. (Refer to Section 5-2.)

109
5. Disassembling/Assembling/Adjusting R-4758.B

5-16-1. Disassembling Seesaw Switch (For 55i)

5-16-1. Disassembling

1. Remove the eyepiece tube. (Refer to Section 5-2.)


2. Remove the objective lens.
3. Remove the condenser lens. (Refer to Section 5-2.)
4. Remove the stage. (Refer to Section 5-2.)
5. Place the microscope upside down. Special hexagon
6. Remove the five special hexagon bolts Bottom cover bolts (M3 x 6) (#239)
(M3 x 6) (#239), and remove the
bottom cover.
Special
hexagon bolt
(M3 x 6) (#239)

7. Disconnect the terminals connected to the Seesaw switch (#506)


seesaw switch (#506).
8. Release the claws on both sides of the Terminal (1 pin)
seesaw switch, and remove it outward.
side

side

Claw

110
5. Disassembling/Assembling/Adjusting R-4758.B

5-16-2. Assembling Seesaw Switch (For 55i)

5-16-2. Assembling

1. Install the seesaw switch to the microscope, and secure it by the claws located on both
sides.
[NOTE]
Take care to install the correct orientation of the seesaw switch.

2. Connect the terminals of the cable to 3-pin Seesaw switch (#506)


and 4-pin of the seesaw switch.
[CAUTION]
Terminal (1 pin)
Be sure to connect the terminals to the
side
pins as far as they will go.
side

Claw

Special hexagon
3. Tighten the bottom cover using the five Bottom cover bolts (M3 x 6) (#239)
special hexagon bolts (M3 x 6).

Special
hexagon bolt
(M3 x 6) (#239)

4. Return the microscope to its original position.


5. Install the stage. (Refer to Section 5-2.)
6. Install the condenser lens. (Refer to Section 5-2.)
7. Install the objective lens.
8. Install the eyepiece tube. (Refer to Section 5-2.)

111
5. Disassembling/Assembling/Adjusting R-4758.B

5-17. Stage Tilt Adjustment (For 50i/55i)

1. Set the 1” reading collimator (J28005) to


1” reading collimator
the dovetail for eyepiece tube mounting.
2. Set the parallel mirror (J28003) on the
stage.

Sub Stage

Parallel mirror

3. Set the collimator so that the X direction


scale is parallel with the traveling direction Traveling direction of the X-axis of the stage

of the X-axis of the stage.

X direction

4. Measure the amount of deviation of the


center (O’) of the scale image from the Y direction Center (O’) of
the scale image
center (O) of the collimator. Check the
amount of deviation meets the following
specification.
X direction

Specifications:
X direction: within +/-1”
Y direction: within +/-1”
Center (O) of the collimator
If the specification is met, perform the stage centering adjustment. (Refer to Section
5-18.)
If not, perform the adjustment according to steps 6 and later.
Be sure to adjust Y direction first.
112
5. Disassembling/Assembling/Adjusting R-4758.B

5-17. Stage Tilt Adjustment (For 50i/55i)

If Y direction does not meet the specification:

5. Check that the scale image is shifted in


either direction.

Shift of (a) direction Shift of (b) direction

6. Remove the 1” reading collimator, parallel mirror, and stage.


7. Remove the elevation block from the microscope base.
(Refer to Section 5-3.)
8. If the scale image is shifted to (a) direction:
Paste the brass foil to the specified portion of the elevation block as shown in the
photograph using adhesive #350.
If the scale image is shifted to (b) direction:
Paste the brass foil to the specified portion of the elevation block as shown in the
photograph using adhesive #350.

Scale image is shifted to (a) direction Scale image is shifted to (b) direction

Paste the
brass foil.

Paste the brass foil.

9. Install the elevation block, stage, parallel mirror, and 1” reading collimator.
10. Check that the amount of deviation for Y direction meets the specification.
If the specification is not met, perform steps 6 to 10 again.
If the specification is met, check that the amount of deviation for X direction meets the
specification.

113
5. Disassembling/Assembling/Adjusting R-4758.B

5-17. Stage Tilt Adjustment (For 50i/55i)

If X direction does not meet the specification:

11. Loosen the six hexagon socket head cap


bolts fixing the elevation block to the
microscope base.
12. Tilt the elevation block in the direction
shown in the photograph to meet the
amount of deviation for X direction meets
the specification.

13. Tighten the six hexagon socket head cap bolts.


14. Check that the amount of deviation for X direction meets the specification.
If the specification is not met, perform steps 12 to 14 again.
If the specification is met, re-check the amount of deviations for both Y and X
directions meet the specifications.
15. After the stage tilt adjustment is completed, be sure to adjust the centering of the
stage. (Refer to Section 5-18.)

114
5. Disassembling/Assembling/Adjusting R-4758.B

5-18. Stage Centering Adjustment (For 50i/55i)

[NOTE]
Before performing this adjustment, the stage tilt adjustment should be completed. (Refer
to Section 5-17.)

1. Set the 80i sub stage tool A (J25481) to the


dovetail on the elevation block.
Centering tool
2. Set the centering tool (J25298) to the
round dovetail for the eyepiece tube
mounting, and secure it using the clamp
screw.
80i sub stage tool A

3. Insert the centering rod (J25299) to the


centering tool, and check that the tip of the Centering rod
rod is inserted into the hole on the 80i sub
Centering tool
stage tool A.
If the rod is inserted: perform the conden-
ser holder tilt adjustment. (Refer to Section
5-19.)
If not, perform the confirmation and
adjustment according to steps 4 and later.

80i sub stage tool A

115
5. Disassembling/Assembling/Adjusting R-4758.B

5-18. Stage Centering Adjustment (For 50i/55i)

4. Check that the centering rod is shifted


against to the 80i sub stage tool in either
direction.
If the rod is shifted to (a) direction:
Centering rod
This shift is generated by the accuracy of
machine when manufactured, and cannot
be adjusted to zero. 80i sub
Therefore, continue with the condenser stage tool A
holder tilt adjustment. (Refer to Section
5-19.)

If the rod is shifted to (b) direction: Shift of (a) direction Shift of (b) direction
Adjust the centering according to steps 5
and later.

Center of the 80i sub stage tool A

5. Remove the centering rod and 80i sub


stage tool A.
6. Remove the elevation block from the
microscope base.
7. Paste the brass foil to the specified portion
of the elevation block as shown in the
photograph using adhesive #350.

Paste the brass foil in round marks position.

8. Install the elevation block, centering rod, and 80i sub stage tool A.
9. Check that the specification is met.
If the specification is not met, perform steps 4 to 8 again.
If the specification is met, perform the stage tilt adjustment again because the
elevation block is removed from the microscope base. (Refer to Section 5-17.)
10. After the stage tilt adjustment and stage centering adjustment are completed, be
sure to adjust the condenser holder tilt. (Refer to Section 5-19.)

116
5. Disassembling/Assembling/Adjusting R-4758.B

5-19. Condenser Holder Tilt Adjustment (For 50i/55i)

[NOTE]
Before performing this adjustment, the stage tilt adjustment (Section 5-17) and stage
centering adjustment (Section 5-18) should be completed.

1. Remove the stage.


2. Set the 1” reading collimator (J28005) to 1” reading collimator
the dovetail for eyepiece tube mounting.
3. Set the parallel mirror (J28003) on the
condenser receptacle of the elevation
block.

Parallel mirror
Condenser receptacle

4. Set the collimator so that the X direction


Traveling direction of the X-axis of the stage
scale is parallel with the traveling direction
of the X-axis of the stage.

X direction

117
5. Disassembling/Assembling/Adjusting R-4758.B

5-19. Condenser Holder Tilt Adjustment (For 50i/55i)

5. Measure the amount of deviation of the


center (O’) of the scale image from the Y direction
Center (O’) of
center (O) of the collimator. Check the the scale image
amount of deviation meets the following
specification.
Specifications:
X direction: within +/-7”
Y direction: within +/-7”

Center (O) of the collimator

If the specification is met, perform the condenser holder upper limit stopper position.
(Refer to Section 5-20)
If not, perform the adjustment according to steps 6 and later.
Be sure to adjust Y direction first.

If Y direction does not meet the specification:

6. Check that the scale image shifts in


either direction.

Shift of (a) direction Shift of (b) direction

7. Remove the 1” reading collimator and


parallel mirror.
8. Remove the three hexagon socket head Hexagon socket head
cap bolts, and remove the condenser cap bolts (M3 x 8) (#442)
holder block from the elevation block. Elevation block

Condenser holder block

118
5. Disassembling/Assembling/Adjusting R-4758.B

5-19. Condenser Holder Tilt Adjustment (For 50i/55i)

9. If the scale image is shifted to (a) direction:


Paste the brass foil to the specified portion of the condenser holder block as shown in
the photograph using adhesive #350.

If the scale image is shifted to (b) direction:


Paste the brass foil to the specified portion of the condenser holder block as shown in
the photograph using adhesive #350.

Scale image is shifted to (a) direction Scale image is shifted to (b) direction
Paste the brass foil.

Paste the brass foil.

10. Install the condenser holder block, parallel mirror, and 1” reading collimator.
11. Check that the amount of deviation for Y direction meets the specification.
If the specification is not met, perform steps 6 to 10 again.
If the specification is met, check that the amount of deviation for X direction meets the
specification.

If X direction does not meet the specification:

12. Loosen the three hexagon socket head Hexagon socket head cap bolts (M3 x 8)(#442)
cap bolts fixing the condenser holder block
to the elevation block.
13. Tilt the condenser holder block in the
direction shown in the photograph to meet
the specification for the amount of
deviation for X direction.
14. Tighten the three hexagon socket head
cap bolts. Condenser
holder block

119
5. Disassembling/Assembling/Adjusting R-4758.B

5-19. Condenser Holder Tilt Adjustment (For 50i/55i)

15. Re-check the amount of deviation for X direction.


If the specification is not met, perform steps 12 to 14 again.
If the specification is met, check that the amounts of deviation for both Y and X
directions meet the specifications.
16. After the condenser holder tilt adjustment is completed, be sure to adjust the upper
limit stopper position of the condenser holder. (Refer to Section 5-20.)

120
5. Disassembling/Assembling/Adjusting R-4758.B

5-20. Condenser Holder Upper Limit Stopper Position Adjustment (For 50i/55i)

[NOTE]
Before performing this adjustment, the stage tilt adjustment (Section 5-17), stage
centering adjustment (Section 5-18), and condenser holder tilt adjustment (Section 5-19)
should be completed.

1. Loosen a set screw (#446) of the


condenser holder block. (The straight pin Straight pin of the stopper
of the stopper will pop up by the force of
the coil spring.)

Set screw (#446)


Condenser holder block
2. Install the stage.
3. Install the condenser lens.
4. Set the specimen slide to the specimen
holder on the stage. Paper (approx. 0.1
Specimen slide
5. Cut a sheet of paper (approx. 0.1 mm mm thick)
thick) to about the same size as the tip of
the condenser lens, and place it on that tip.
6. While observing the specimen slide, turn Straight pin
Stage
the knob of the condenser holder upward
until the paper on the condenser lens just Set screw Condenser lens
comes in contact with the specimen slide
on the stage.
7. While maintaining the position adjusted in
step 6, tighten the set screw loosened in
step 1.
8. Remove the specimen slide.
9. Remove the paper from the tip of the condenser lens.
10. Turn the knob of the condenser holder once to lower.
11. Turn the knob upward until it stops.
12. Place the specimen slide on the stage again, and check that the specimen slide does
not contact the tip of the condenser lens.

121
5. Disassembling/Assembling/Adjusting R-4758.B

5-21. LED Position Adjustment (For 55i)

1. Remove the eyepiece tube. (Refer to Section 5-2.)


2. Remove the objective lens.
3. Remove the condenser lens. (Refer to Section 5-2.)
4. Remove the stage. (Refer to Section 5-2.)
5. Pull out the hexagonal driver (#248).
6. While pressing the two portions of the
lamp cover (#223) in the arrow direction
(a) shown in the figure, move it in the
Lamp cover
arrow direction (b) to remove.
(#223)

(a)

(a) (b)

7. Remove the four hexagon socket head Cap (#233)


cap bolts (M3 x 6) (#254), and remove the
plastic cover (#202).
8. Remove the cap (#233). Hexagon socket
head cap bolts
(M3 x 6) (#254)

Plastic cover (#202)

Special hexagon
9. Place the microscope upside down. Bottom cover bolts (M3 x 6) (#239)
10. Remove the five special hexagon bolts (M3
x 6) (#239), and remove the
Special
bottom cover.
hexagon bolts
(M3 x 6) (#239)

122
5. Disassembling/Assembling/Adjusting R-4758.B

5-21. LED Position Adjustment (For 55i)

11. Remove the two hexagon socket head cap Hexagon socket head cap
Diffuser unit bolts (M3 x 8) (#255)
bolts (M3 x 8) (#255), and remove the
diffuser unit.

12. Return the microscope to its original


Special hexagon bolts (M3 x 6) (#229)
position.
13. Loosen the four special hexagon bolts
(M3 x 6) fixing the control board.
14. Connect the AC adapter or the battery to
the microscope.

Control board (#509)

123
5. Disassembling/Assembling/Adjusting R-4758.B

5-21. LED Position Adjustment (For 55i)

15. Install the standard straight tube (J25295) to the microscope, and secure it using the
clamp screw as shown in the photograph.
16. Set the Ph tool ring (J25260) to the centering telescope (J25172), and set them to the
standard straight tube.
[Note]
The binocular tube installation is acceptable instead of installing the standard straight
tube, Ph tool ring, and centering telescope to the microscope, as shown in the
photograph.

Centering telescope
Ph tool ring Binocular tube

Standard straight tube

Control board

17. Turn ON the power switch. The LEDs light


up.
18. Adjust the position of the control board so
that the center of the cross line on the
centering telescope and the center of the
central LED coincide.

Specifications:
Deviation between centers of the cross
Central LED
line and the central LED: Within 1.5 dev.
of scale
19. After the adjustment is completed, tighten the control board using the four special
hexagon bolts.
20. Re-check that the specification is met.
124
5. Disassembling/Assembling/Adjusting R-4758.B

5-21. LED Position Adjustment (For 55i)

After adjustment, assemble the parts and units in the following steps.
21. Turn OFF the power, and disconnect the AC adapter or the battery.
22. Remove the standard straight tube, centering telescope, and Ph tool ring.
23. Place the microscope upside down.
Hexagon socket head
24. Install the diffuser unit, and firmly move it in
Diffuser unit cap bolts (M3 x 8)
the direction of the arrows. Tighten it with
two hexagon socket head cap bolts (M3 x
8).

Push it in these directions.

25. Tighten the bottom cover using the five Special hexagon
Bottom cover bolts (M3 x 6) (#239)
special hexagon bolts (M3 x 6).

Special
hexagon bolts
(M3 x 6) (#239)

26. Return the microscope to its original


position.
27. Tighten the plastic cover using the four
Cap (#233)
hexagon socket head cap bolts (M3 x 6).

Hexagon socket
head cap bolts
(M3 x 6) (#254)

Plastic cover (#202)

125
5. Disassembling/Assembling/Adjusting R-4758.B

5-21. LED Position Adjustment (For 55i)

28. Move the lamp cover in the opposite


direction of the arrow, and install it to the
plastic cover.
29. Install the hexagonal driver. Lamp cover
(#223)

30. Install the stage. (Refer to Section 5-2.)


31. Install the condenser lens. (Refer to Section 5-2.)
32. Install the objective lens.
33. Install the eyepiece tube. (Refer to Section 5-2.)

126